Manual Instrument - Manager
Manual Instrument - Manager
User Manual
version 1.3
INSTRUMENT
MANAGER
KEY TO SYMBOLS
Please see below all the symbols that will be used in the manual to draw the reader’s attention:
Further information.
TABLE OF CONTENTS
INTRODUCTION .................................................................................................................... 1
INSTALLATION AND START-UP ............................................................................................ 1
SOFTWARE CONFIGURATION .............................................................................................. 2
CONFIGURATIONS ............................................................................................................... 3
CONFIGURATION DATA ......................................................................................................... 3
IDENTITY PARAMETERS ................................................................................................................ 3
OPERATING PARAMETERS ........................................................................................................... 3
SAVING PARAMETERS................................................................................................................... 4
CREATING A NEW CONFIGURATION ................................................................................... 5
OPENING AN EXISTING CONFIGURATION ........................................................................... 6
EDITING A CONFIGURATION ................................................................................................. 7
SAVING A NEW CONFIGURATION ...................................................................................... 10
SAVING AN EXISTING CONFIGURATION............................................................................ 10
EXPORTING A CONFIGURATION ........................................................................................ 10
IMPORTING A CONFIGURATION ......................................................................................... 11
COMPARING CONFIGURATIONS ........................................................................................ 11
PRINTOUTS ........................................................................................................................... 12
PROFILES............................................................................................................................ 13
PROFILE DATA ..................................................................................................................... 13
CREATING A NEW PROFILE ................................................................................................ 13
EDITING AN EXISTING PROFILE ......................................................................................... 14
CONNECTION...................................................................................................................... 15
SERIAL SETTINGS ................................................................................................................ 15
CONNECT .............................................................................................................................. 16
REAL-TIME OPERATIONS .................................................................................................. 17
DASHBOARD......................................................................................................................... 17
CHANNELS ............................................................................................................................ 19
CHANNEL SELECTION ..................................................................................................................20
EQUALISATION ..............................................................................................................................21
REAL CALIBRATION............................................................................................................. 22
COMMANDS .......................................................................................................................... 23
FIRMWARE UPDATE .......................................................................................................... 24
ERRORS DURING UPDATE .................................................................................................. 26
INSTRUMENT RESTORE ...................................................................................................... 27
QUALIFIED ACCESS........................................................................................................... 28
LOG EXPORT ...................................................................................................................... 29
SOFTWARE BACKUP ......................................................................................................... 29
INTRODUCTION
The Instrument Manager software allows you to set, update and monitor parameters, to be managed
from a PC for an instrument connected via a PC serial port.
Software features:
- Creation of instrument configurations.
- Configuration uploading and downloading to/ from instruments.
- Profiles and personal configuration database.
- Configuration comparison and printout.
- Real-time instrument management via serial port.
- Firmware update.
- Instrument real calibration.
- Equalisation and channel selection (multichannel instruments only).
- Restoring non-operating instruments.
- Qualified access for legally relevant operations.
You may need to perform the operation as an administrator. In this case, right-click on the
installation file and select “Run as administrator”.
Once installed, start Instrument Manager from the Windows Start menu or double-click the icon on
your desktop.
-1-
From the welcome window, you may perform various quick operations by clicking on the displayed
icons and fields:
- New: click to create a new configuration (see section CREATING A NEW CONFIGURATION).
- Browse: click to open an existing configuration (see section OPENING AN EXISTING
CONFIGURATION).
- Connect: click to open the connection window in real time (see section CONNECT).
- Help: click to open this user manual.
- Recent configurations: double-click on an item on the list or select and click Open to open the
configuration.
SOFTWARE CONFIGURATION
Select the menu Advanced Settings to access the Instrument Manager software settings window.
-2-
CONFIGURATIONS
Configurations are sets of parameters that define the identity and behaviour of an instrument.
Configuration data consist of various types of parameters:
- identity parameters;
- operating parameters;
- saving parameters.
You can create new configurations or open previously created and saved configurations.
CONFIGURATION DATA
IDENTITY PARAMETERS
OPERATING PARAMETERS
The operating parameters are the set of parameters that the instrument uses while performing its
tasks. They can be set directly from the instrument menu.
For example: sensitivity, theoretical full scale, setpoint and weight filter, etc.
-3-
SAVING PARAMETERS
The saving parameters are the software data configurations being stored and identified within the
software:
-4-
CREATING A NEW CONFIGURATION
You can create a new configuration by manually defining all the identity parameters or by using the
identity parameters of a connected instrument.
To create a new configuration by manually defining all the identity parameters, proceed as follows:
The configuration created may only be loaded onto an instrument with the same identity
parameters.
Create a new configuration using the identity parameters of a connected instrument as follows:
- connect the instrument to the PC;
- select the menu Connection Connect
- the software will show the connection window with the list of instruments detected;
o if the window fails to show the connected instruments, check the serial settings (see
section SERIAL SETTINGS) and click on Refresh;
- select the instrument from which you want to load the identity parameters;
- click on New to confirm your selection;
- the software will generate a new configuration with the default operating parameters.
-5-
OPENING AN EXISTING CONFIGURATION
It is possible to open a configuration by dragging the .lac file into the software window.
To help the research of configurations, use the fields on top of the window to filter the items listed in
the table.
Filters Date and Last Edit allow you to search for configurations created and/or edited within a given
time frame.
Click on Clear Filters to reset the active filters and come back to display the active configurations.
-6-
EDITING A CONFIGURATION
After creating a new configuration or loading an existing configuration, the software will show a series
of sheets with a breakdown of the operating parameters. You may change the operating parameters
via the related fields.
When the open configuration has unsaved changes, the indication “(edited)” appears next to the
device name in the summary section at the top.
-7-
After each edit, the software will check the new value entered. If incorrect, the parameter will be
highlighted in red and the name of the tab it belongs to will be marked by a warning icon.
-8-
To display the type of error, place the mouse pointer onto the name of the parameter concerned.
Example: The maximum permitted value for sensitivity is 7. If set to 8 - a higher value - the parameter
will be highlighted in red. If the mouse pointer is positioned over the message Sensitivity, the display
will show the range of permitted values for this parameter.
If one or more operating parameters are in error, you cannot save the configuration nor
load it onto the instrument.
If the configuration has been set in a Legal for Trade state - e.g. OIML, OIML Multiple Range, OIML
Multi Interval, NTEP, NTEP Multiple Range or NTEP Multi Interval - you can only edit certain
parameters via a qualified access (see section QUALIFIED ACCESS). These parameters are
identified by the icon next to the name.
To close an open configuration and return to the main screen, press on the top right or select the
menu: Configuration Close.
-9-
SAVING A NEW CONFIGURATION
After creating and editing a new configuration, you can save it by defining its backup parameters.
To save a new configuration, proceed as follows:
EXPORTING A CONFIGURATION
- 10 -
IMPORTING A CONFIGURATION
You may import a previously exported configuration. Imported configurations may be used and edited
even if they are not in the database.
COMPARING CONFIGURATIONS
You may compare two configurations to highlight any differences between parameters.
- 11 -
The parameters and their values are shown in the table below.
Parameters with different values are highlighted in red.
To show only parameters with different values, select Only Differences in the Show drop-down menu.
To print out the comparison, click on Print.
You may compare instrument configurations with different identity parameters. In this case, the table
will only show the operating parameters of both configurations.
PRINTOUTS
After creating or editing a configuration, you may print out all its parameters.
- 12 -
PROFILES
Profiles are attributes that may be used to distinguish configurations of instruments belonging to
different customers or to subdivide configurations as desired. You may create and edit profiles to
which configurations may be assigned.
PROFILE DATA
Logo: image associated with the profile (optional). You may upload
an image by clicking Open or delete it by clicking Delete.
- 13 -
EDITING AN EXISTING PROFILE
- 14 -
CONNECTION
SERIAL SETTINGS
If the Baud Rate parameter is set to Auto, the software automatically searches for the value that
allows establishing a connection and uses it for subsequent operations on the instrument.
- 15 -
CONNECT
o New: will generate a new configuration where identity parameters will be identical to
those of the connected instrument, and operating parameters will be set by default (see
section CREATING A NEW CONFIGURATION).
o Upload Configuration: will upload the currently open configuration to the software on the
connected instrument. Each operating parameter on the instrument will be overwritten.
The configuration upload procedure ends with the restart of the instrument.
o Download Configuration: will generate and open a new configuration with all parameters
identical to those of the connected instrument. The parameters on the instrument are
not synchronised with those of the open configuration. To update them you need to
upload the configuration again.
o Connect: will activate a real time connection with the instrument (see section REAL-
TIME OPERATIONS).
o Search Instrument: will search again for connected instruments.
- 16 -
REAL-TIME OPERATIONS
All the operations described in this section are possible only if a real time connection with the
instrument is in progress (see the section CONNECT).
The instrument keypad will be locked during real-time connection. If the real-time connection is
unexpectedly interrupted, the instrument will need to be restarted to resume use.
DASHBOARD
The Dashboard is a window that shows the real-time trend of the data read by the connected instrument.
To access the Dashboard window, proceed as follows:
- establish a real time connection with the instrument (see section CONNECT)
- select the menu Operations Dashboard
or
click on the toolbar icon .
8 2 3 4
1. Weight graph: shows the weight trend displayed on the connected instrument display
2. Weight value read by the connected instrument.
3. Information on the status of the read weight: Stable, Net/Gross, Negative, Error.
4. Buttons for interacting with the connected instrument:
- PAUSE/RESUME: will stop the weight reading by the instrument. The graph and all read
values will remain unchanged at the last value read;
- Net: will perform a net operation. This operation may be repeated several times;
- Gross: will cancel all net operations performed;
- Zero: will send the semi-automatic zero command to the instrument.
- 17 -
5. Information on the status of the connected instrument inputs and outputs:
- Indicator off: Input or Output open;
- Indicator on: Input or Output closed;
6. Load graph: shows the percentage of current load, expressed in relation to the capacity of the load
cells connected to the instrument.
If the connected instrument is of the multichannel type, the section shows the status of each
channel (active, inactive or in error) and the percentage distribution of the load.
- 18 -
CHANNELS
3 5
- 19 -
CHANNEL SELECTION
The channel selection commands may be used to display and confirm the channels automatically
detected by the instrument. It is possible to establish active channels automatically or manually. For
further information on how to select channels, please refer to the instrument manual.
- 20 -
EQUALISATION
If an instrument has been set in a Legal for Trade state - e.g. OIML, OIML Multiple Range, OIML
Multi Interval, NTEP, NTEP Multiple Range or NTEP Multi Interval - to access this feature,
qualified access is required (see section QUALIFIED ACCESS).
The equalisation procedure allows the weight differences detected on a platform to be corrected,
weighed by more than one load cell.
For more information on how equalising works, please refer to the instrument manual.
E.g.:
OK OK NO
Click on Next to confirm the Click Next to equalise channel 3. Channel 1 has already been
weight with the scales unloaded. equalised and by proceeding,
you would get an error.
Move the sample weight to
channel 2 or 3 before
continuing.
To delete the equalisation and return to a non-equalised system, click on Delete active equalization.
- 21 -
REAL CALIBRATION
If an instrument has been set in a Legal for Trade state - e.g. OIML, OIML Multiple Range, OIML
Multi Interval, NTEP, NTEP Multiple Range or NTEP Multi Interval - to access this feature,
qualified access is required (see section QUALIFIED ACCESS).
It is possible to perform instrument real calibration using sample weights of a known value and, if
necessary, to correct indicated value deviations from the correct value.
The graph in the Real Calibration window shows the calibration curve currently active on the
instrument. The blue dot on the graph indicates the weight currently being read by the instrument.
- 22 -
To delete a calibration point, click on the corresponding icon in the Points pane.
To delete all calibration points, click Delete All in the Points pane.
Performing a reset will cancel all calibration points: so, this is the first operation you need
to perform in the real calibration procedure.
COMMANDS
You can send commands to the connected instrument: to access the command list, proceed as follows:
- establish a real time connection with the instrument (see section CONNECT);
- select the menu Operations Commands.
Command options:
- Restart instrument: the instrument is restarted;
- Factory Reset: all instrument's operating parameters will be set to the instrument's default
values (see instrument manual to acquaint the default values).
- 23 -
FIRMWARE UPDATE
- 24 -
After launching the update, follow the instructions on the monitor and proceed as follows:
1. The software will show the Confirm Boot Mode window.
o If the window requires the RS232 port to be specified, connect the instrument RS232
serial port to the PC and select the corresponding COM port from the drop-down menu
(for further information on connecting the serial port, consult the instrument manual);
o If the window does not require you to specify the RS232 port, the COM port will be
selected automatically;
o If you are prompted to perform preliminary operations before proceeding, follow the
directions of the wizard shown;
2. Activate the Boot mode on the instrument:
o Restart the instrument by holding down the ENTER key on the instrument keyboard
until “” or “” appears;
3. Click on Proceed to confirm the update;
4. Upon completion, the software will display a message showing the outcome of the update
procedure. In case of failure, see section ERRORS DURING UPDATE.
- 25 -
ERRORS DURING UPDATE
If the firmware update is unsuccessful the software automatically saves to file the identification data
of the instrument and the information needed for recovery.
- 26 -
INSTRUMENT RESTORE
The software will not check that the connected instrument corresponds with the selected one.
Pay special attention when selecting the instrument
- 27 -
QUALIFIED ACCESS
If an instrument has been set in a Legal for Trade state - e.g. OIML, OIML Multiple Range, OIML Multi
Interval, NTEP, NTEP Multiple Range or NTEP Multi Interval - you can edit certain parameters and
perform certain features only via a qualified access.
A customer password table is required to execute a Qualified Access Procedure, which is supplied by the
manufacturer to authorised service centres only. If you have the password table, proceed as follows:
- select the menu Advanced Qualified Access;
- the software will show the Qualified Access window;
- enter your identification code (user password) in the User field, shown in the table;
- refer to the table and record the password (4-digit number) corresponding to the value shown
in the Seed field;
- type the password in the Password field and click on Authenticate.
- 28 -
LOG EXPORT
In the event of a malfunction or a failed firmware update, you can export log files, containing diagnostic
information, to facilitate technical assistance in identifying the problem.
SOFTWARE BACKUP
To export a backup select the menu: ? Backup Save backup, the software will show the save
dialog box.
To import a backup select the menu: ? Backup Import from backup, select the path of the .zip
file you want to import.
WARNING: importing a backup overwrites all data contained in the current installation.
- 29 -
On our website www.laumas.com you can find videos on the correct installation of weighing
systems and tutorials on the configuration of our weight transmitters and indicators.
Comprehensive user manuals for all LAUMAS products can be found online. They can be
downloaded in PDF format from either the Products section or the Download Area of the website
www.laumas.com. Registration is required.