Automatic Biochemistry Analyzer BK-200 User Manual
Automatic Biochemistry Analyzer BK-200 User Manual
Automatic Biochemistry Analyzer BK-200 User Manual
BK-200 (SAPPHIRE)
Content
Preface................................................................................................................................................. 1
Guarantee.................................................................................................................................... 2
Product manual.......................................................................................................................... 4
Transportation and storage......................................................................................................... 5
Safety Information...................................................................................................................... 7
Safety Symbols........................................................................................................................... 7
Safety Precautions...................................................................................................................... 8
Operation Precautions...............................................................................................................11
Instrument operation.................................................................................................................12
Chapter 1 Main Introduction.............................................................................................................17
1.1 System Composition...........................................................................................................17
1.2 Instrument parameters........................................................................................................ 17
1.3 Introduction of analysis parts............................................................................................. 20
1.3.1 Instrument Appearance........................................................................................... 20
1.3.2 Instrument roof structure.........................................................................................23
1.3.3 Instrument Structure................................................................................................24
1.4 Structure and function........................................................................................................ 24
1.4.1 Operation Part......................................................................................................... 25
1.4.2 Analysis Part............................................................................................................25
1.4.2.1 Reagent & Sample tray...............................................................................26
1.4.2.2 Dispensing System......................................................................................27
1.4.2.3 Reaction Tray System................................................................................. 30
1.4.2.4 Stirring arm................................................................................................. 32
1.4.2.5 Washing Arm...............................................................................................34
1.4.2.6 Cooling system........................................................................................... 34
1.4.2.7 Optical inspection system........................................................................... 36
1.4.2.8 water system.............................................................................................. 36
1.4.2.9 Alarm system............................................................................................. 38
Chapter 2 Installation........................................................................................................................ 39
2.1 Unpacking...........................................................................................................................39
2.1.1 Unpacking Steps......................................................................................................39
2.1.2 Instrument Transporting..........................................................................................40
2.2 Installation Requirement.................................................................................................... 40
Automatic biochemistry analyzer BK-200 user manual
Preface
Thank you for purchasing BK-200 automatic biochemistry analyzer
Intended use
The instrument for quantitative analysis of human serum, plasma, urine, cerebration
fluid and other samples of the clinical chemical composition. Do not use for other
purposes.
Using object
This manual is intended for the clinical laboratory technologists operating this
instrument.
Before using the product, please carefully read the contents of this manual, the correct
use of the product. Keep this manual in a safe place for easy reference. If you do not
follow the precautions described in this manual, we will not guarantee the
maintenance.
Software used in the product
Auto Biochemical Analyzer software for Jinan BIOBASE Biotech Co., Ltd. to obtain
software license software license.
1
Automatic biochemistry analyzer BK-200 user manual
Guarantee
Warranty period
One year from the date of completion of the installation.
Ensure that content
In the warranty period, if any failure caused by the design, manufacturing defects,
will be repaired free of charge, our company will take the appropriate content of the
corresponding countermeasures.
Statement
Jinan BIOBASE Biotech Co., Ltd. (hereinafter referred to as "Jinan BIOBASE ") has
the final interpretation of this manual.
The Company shall be responsible for the safety, reliability and performance of the
product only if all of the following requirements happened:
1. Assembly operations, expansion, re-adjustment, improvement and repair by the
Company recognized professionals;
2. All repairs involving replacement parts and supporting the use of accessories,
supplies are original of the Company (original) or approved by the Company;
3. the related electrical equipment is according to national standards and the use of the
manual requirements;
4. Product operation is carried out according to the instruction manual.
Disclaimer
The Company shall not be liable for any equipment failure or damage, or for any
direct or indirect damage that may occur during the use of the equipment.
1. Malfunction or damage due to violation of the instructions, precautions, and
intended use of this manual.
2. Malfunction or damage caused by repair or alteration of the other company.
3. Malfunction or damage caused by use instruments of other company at the same
time .
4. Malfunction or damage caused by operating environment not corresponding to the
specified operating environment (power conditions, installation environment, etc).
5. Malfunction or damage caused by natural disasters such as earthquakes and floods.
6. Malfunction or damage caused by the company unaware of the movement or
transfer (transport) after installation.
2
Automatic biochemistry analyzer BK-200 user manual
3
Automatic biochemistry analyzer BK-200 user manual
Product manual
Dimensions and weight of the instrument
BK-200: Dimensions: 950mm (length) × 603mm (width) × 505mm (height);
Weight: 65Kg
BK-200(with bottom cabinet): Size: 950mm (length) × 603mm (width) × 1075mm
(height); Weight: 110Kg
Description of service life
The service life of this instrument is 5 years from the date of completion of
installation and debugging of the instrument. After 5 years, the instrument will be
discarded. Please follow the relevant regulations.
Product Categories
The classification criteria are described below:
1. Overvoltage Category: Overvoltage Category (Class II)
2. Pollution Degree: Pollution Degree (Class Ⅱ)
3. Installation of environmental conditions:
a) Indoor use;
b) Altitude not exceeding 2000m;
c) The temperature range of 15 ℃ ~ 30 ℃;
d) The maximum relative humidity is 85% when the temperature is below 30 ℃;
e) The supply voltage fluctuates not more than ± 10% of the nominal voltage;
f) Typical transient overvoltage present on the mains supply;
g) Note: The transient over-voltage of the nominal rating of GB 16895.12 specified
pulse withstand voltage (over-voltage) Category Ⅱ.
h) The applicable rated pollution level.
4. Equipment categories: laboratory equipment
5. Connection to the network power supply: Removable power cord
6. Operating conditions: continuous
4
Automatic biochemistry analyzer BK-200 user manual
Transport
In the packaging states, the instruments conduct transportation according to the
requirements of the transport contract, in the transport process to prevent rain and sun
exposure, to prevent severe impact, weight and dumping.
Note: If the instrument has been unpacked, before moving the instrument, please re-
packaging equipment before transport.
Store
Packaged instruments should be stored at 0℃ to 40℃ and relative humidity not more
than 85%, non-corrosiveness gases and in a well-ventilated environment.
5
Automatic biochemistry analyzer BK-200 user manual
6
Automatic biochemistry analyzer BK-200 user manual
Safety Information
This chapter introduces the safety symbols and their meanings, summarizes the safety
hazards and precautions for using the instrument, and the specific meaning of the
label affixed to the instrument. It also lists the content of toxic and hazardous
substances contained in each part of the instrument is in accordance with the relevant
standards or not.
Safety Symbols
Various safety symbols are used in this instruction manual to remind you of the
precautions to be taken during operation. As shown in the following table:
Symbols Slogan Introduction
Biological infection Indicates a risk of biological infection, if not
in accordance with the instructions, there may
be the risk of biological infection.
Electric shock Indicates a risk of electric shock. If exposure
can result in personal injury.
7
Automatic biochemistry analyzer BK-200 user manual
Safety Precautions
For the safe use of this instrument, please carefully read the following safety
precautions. Any violation of the following safety precautions may result in personal
injury or damage to the instrument.
Warning:
Do not open the back cover or the side cover when the
main power is turned on.
Reagent and sample are spilled in the instrument, which
may cause malfunction and electric shock. Do not place
samples or reagents on the instrument. If splash occurs,
please turn off the power immediately and contact
Service.
8
Automatic biochemistry analyzer BK-200 user manual
9
Automatic biochemistry analyzer BK-200 user manual
Waste treatment
To prevent waste liquid from causing environmental pollution and personal injury,
observe the following precautions when treating waste liquid.
Biological infection:
10
Automatic biochemistry analyzer BK-200 user manual
Operation Precautions
To use the instrument correctly and effectively, please read the following precautions
carefully.
Intended use
Warning:
Operator
Warning:
Environment
Attention:
11
Automatic biochemistry analyzer BK-200 user manual
Instrument operation
Attention:
(2). Before using the instrument for the first time, please carry
out calibration, quality control, to confirm the instrument is
working properly.
12
Automatic biochemistry analyzer BK-200 user manual
Computer viruses can destroy software and data. Do not use the
computer for other purposes or connect to the Internet.
Instrument maintenance
Attention:
Before cleaning, turn off all power to the instrument and unplug
the power cord. Take necessary measures to prevent water
droplets from entering the instrument during cleaning. Doing so
13
Automatic biochemistry analyzer BK-200 user manual
Parameter settings
Attention:
Sample
Attention:
(1) Use the full separation serum samples and urine samples
that do not contain suspended solids. If the serum sample
contains fibrin, or urine samples containing suspended solids
and other insoluble impurities, are likely to block the needle,
affecting the analysis results.
14
Automatic biochemistry analyzer BK-200 user manual
(7) Before testing, make sure the sample is placed in the correct
sample position, otherwise the correct result may not be
obtained.
Reagents, calibrator, QC
Attention:
15
Automatic biochemistry analyzer BK-200 user manual
Data backup
Attention:
16
Automatic biochemistry analyzer BK-200 user manual
17
Automatic biochemistry analyzer BK-200 user manual
Wavelength
±2nm
accuracy
Reaction
37℃±0.1℃
Temperature
Test Item Up to 56 test items
dispensing probe
Inner and outer walls cleaning
cleaning
Reagent
Reagent dish with frozen;Liquid medium conduction
dish,reagent
cooling refrigeration; total of 56 reagent positions
position
Reagent
20~350ul(1ul stepping)
volume
Reagent Reagent bottle
18mL 、50mL
system Specifications
Reagent margin Reagent solution requires greater than 1mL
Liquid level detection,remaining detection,vertical
Reagent probe
collision detection
Reagent probe
Inner and outer walls cleaning
cleaning
18
Automatic biochemistry analyzer BK-200 user manual
Reagent storage
2℃~8℃ storage 24 hours working
temperature
Reaction system Discrete
Cuvette optical
6mm
path
The number of
6 groups, each group of 20, a total of 120
cuvettes
Reaction time About 10 minutes
The reaction
120~450ul
liquid volume
Light source 20W/12V Long life quartz halogen
Photoelectricity
8 imported photoelectricity receiver
receiver
Reaction
system Absorbance range 0~3.0Abs
19
Automatic biochemistry analyzer BK-200 user manual
system 950mm(L)×603mm(W)×505mm(H)(BK-
bottom cabinet)
Power(VA) 300VA(BK-200)
Water
5L/h
Consumption
Power supply AC110/220V±10%,60/50Hz
20
Automatic biochemistry analyzer BK-200 user manual
Figure 1-3 Right side view Figure 1-4 Left side view
21
Automatic biochemistry analyzer BK-200 user manual
No Name Remark
1 header Protect adding system;detection unit;sample and reagent
tray
2 window Watch adding system and detection units from here
3 Front panel Strike when maintenance the machine
4 alarm indicator Liquid level alarm and temperature alarm
lamp
5 Right side Strike when adding cooling oil and maintenance water
panel system
6 Power switch Operation switch and cooling switch
7 Left side panel power switch is located on the left side panel
8 L-type support Connect grounding line,water alarm float switch and water
tube connection
9 Power panel Open or close power supply and power line connection
10 Earth bolt Used to connect earth line
11 Aviation joint Connect water alarm float switch to alarm purifier and
waste water
12 Outfall Connect armored tube, empty waste water
13 Inlet Connect silicone tube, provide purifier water
14 Communicatio Used to connect analyzer and PC
n interface
15 Fuse Protect the safe operation of the circuit
16 Main switch Used to control the main power of the analyzer
17 Power line Connect power supply
22
Automatic biochemistry analyzer BK-200 user manual
23
Automatic biochemistry analyzer BK-200 user manual
24
Automatic biochemistry analyzer BK-200 user manual
25
Automatic biochemistry analyzer BK-200 user manual
As shown in figure 1-9, sample tray is divided into inner ring and outer ring and
sample positions, a total of 49 samples. We can put the standard cup that contain
sample, calibration liquid, quality control liquid in the designated position, through
the sample disc rotating to send samples to dispensing position of the dispensing arm.
1-Gear positioning plate; 2-Step motor; 3-Shaft; 4-Outline frame; 5-Support base
Figure1-10 Reagent&sample Tray Structure
26
Automatic biochemistry analyzer BK-200 user manual
27
Automatic biochemistry analyzer BK-200 user manual
28
Automatic biochemistry analyzer BK-200 user manual
Dispensing arm is used to take quantitative sample from sample cup and fill into the
reaction cuvettes. dispensing probe have liquid level detection function. If dispensing
probe collide, it will execute anti-collision protection procedure. The probe will return
to the highest place and stop. Operator need to power off and power on again. After
power on, need to check dispensing probe’s mechanical position. If the position is
correct , please run the test again.
Dispensing probe volume range is 2ul-70ul, 0.1ul stepping. The minimum liquid level
detection sample volume is 50ul, the minimum requirement of sample volume is
above 150ul.
Dispensing probe withdraws fix quantify reagent from the reagent bottle and then add
to reaction cuvettes. The probe also have level senor which can detect the remaining
volume of the reagent. As figure 1-14 shown:
29
Automatic biochemistry analyzer BK-200 user manual
After testing, if we need to add washing liquid to wash reaction tray cuvettes, select
“Clean fluid” in the “Washing and Blank ground” window. It can set the washing
liquid position in reagent tray and the amount and the reaction cuvettes that needed.
Generally, setting number is 28 and the volume is 300ul.
During testing, click “instrument check” button,as figure 1-15 shown, reagent
sample plate and rotating to a certain number,Click instrument check to test the
position of arms and trays, including the washing position, probe horizontal position
,probe vertical position and reaction cuvette position.
Figure 1-15 Sample arm adjustment interface Figure 1-16 Dispensing probe
component
As Figure 1-16 shown, the dispensing probe is consist of anti-collision probe, up fixed
screw, spring, adapter sleeve and nut.
1.4.2.3 Reaction Tray System
As figure1-17 shown, reaction tray system consists of tray, cover, cuvettes and
thermostat.
Reaction tray cover: rotated by stepper motor.
Cuvettes: cuvettes, total 120 units, can be replaced independently by micro organic
material, its standard optical path is 6 mm.
30
Automatic biochemistry analyzer BK-200 user manual
Reflection cup group: a total of 6, each part is installed 20 cuvettes, fixed on the
reaction tray cover by two pins, as shown in figure 1-18.
Thermostat bath: air heating incubation, make samples and reagents reaction at 37 ℃.
At the same time monitor the temperature through the temperature control system.
31
Automatic biochemistry analyzer BK-200 user manual
base;
drive motor, support pillar and optical fiber fixed plate etc.
After turning on the power, reaction tray rotates clockwise, move to the number 1
position.
When check movement, click “Instrument Check” ,choose
32
Automatic biochemistry analyzer BK-200 user manual
1-Spline shaft;2-Left and right belt wheel;3-Left and right driving motor;4-Up
33
Automatic biochemistry analyzer BK-200 user manual
1.4.2.5Washing Arm
;13-Block
Cooling system adapts oil cooling system,the system consists of reagent & sample
tray, coolant circulation container, magnetic pump, heat exchanger, fan, float switch,
tubes and so on. The pipeline diagram as shown in figure 1-22.
34
Automatic biochemistry analyzer BK-200 user manual
35
Automatic biochemistry analyzer BK-200 user manual
components, used to provide enough intensity monochromatic light, after the light
source produce stable and continuous spectrum, polychromatic light divide into
monochromatic light through the Fiber optic components, then after filtering the
specific wavelengths of monochromatic light into the signal detection system.
Signal detection system includes photoelectric conversion section and AD collection
and processing. Its main function is to convert the light of the transmission to
electrical signals, then amplify, analysis and processing, acquiring the light path
signal.
Water system includes water pump liquid tube and liquid condensate drain tube,The
36
Automatic biochemistry analyzer BK-200 user manual
Pump and valve installation as shown in figure 1-25, seven diaphragm pump, five
Solenoid valve is fixed on the plate from left to right in turn.
37
Automatic biochemistry analyzer BK-200 user manual
38
Automatic biochemistry analyzer BK-200 user manual
Chapter 2 Installation
In order to ensure the normal operation of the instrument after installation, only
person authorized by BIOBASE is allowed to doing the initial installation and settings.
2.1 Unpacking
2.1.1 Unpacking Steps
After the goods arrival, please carefully check the instrument packaging, to see if
there is physical damage. If any damage, please contact with BIOBASE or local
distributor. After confirming no external damage, unpacking follow steps below:
●Up Place package box as the Up arrow
●Open the accessory box and check whether object is complete according to the
packing list or not. If there is any missing, please contact with BIOBASE or local
distributor.
●Carefully check the instrument appearance, if there is any damage, please contact
with BIOBASE or local distributor.
39
Automatic biochemistry analyzer BK-200 user manual
40
Automatic biochemistry analyzer BK-200 user manual
41
Automatic biochemistry analyzer BK-200 user manual
b). The distance between the rear panel against the wall should be greater than 50 cm;
c). The distance between front side of the analyzer and other instruments should be
greater than 100 cm;
d.) The space of the waste liquid discharging device and the pure water supplying
device should be ensured while installation.
42
Automatic biochemistry analyzer BK-200 user manual
Warning
The protective grounding must be well protected to against electrical shock and
instrument malfunction.
2.2.3 Environmental requirement
Environment temperature of 15 ℃ ~ 30 ℃;
Relative humidity is 40% ~ 85%;
Atmospheric pressure is 76 kpa to 106 kpa;
43
Automatic biochemistry analyzer BK-200 user manual
BK series full automatic biochemical analyzers are only in indoor use. The
environment should be clean, without mechanical vibration and noise source and the
power interference;
The ground should be flat and of sufficient strength.
Don't get analyzer close to brush type generator, flashing fluorescent, and often
switching electric contact device;
Avoid direct sunlight or in front of heat source and pouring;
Keep the instrument well ventilated.
Note:
The instrument will not guarantee the accuracy of the normal operation and test data
in the rugged environment mentioned above. If the temperature and humidity cannot
meet the above requirements, please use the air conditioning equipment.
In the working process, the instrument can produce heat and it is discharged through
the instrument back. Working environment should be kept well ventilation and
ventilation equipment can be used when necessary. Direct flow of air should be
avoided to the instrument, or it may affect the accuracy of the instrument test.
2.3 Water supply and drainage
The following requirements of water supply and drainage must be met before the
instrument is delivered:
Instrument requires deionized water, the conductivity below 1μs / cm, water
consumption is 5L per hour. We recommend to use 15L water machine.
Ensure that the instrument water supply hole and pipe installation should be
unimpeded. In addition, the inlet hole of L rack should be higher than the pure water
barrel, and difference of feed water level should be less than 50cm.
Ensure that the instrument drain hole and pipe installation should be unimpeded, and
the equipment L rack outlet hole should be higher than the waste barrel (or waste
discharge), waste pipe length should not exceed 2m.
2.4 Installation steps
2.4.1 Hardware installation
Company professionals handle the instrument hardware system installation. It is
recommended that the user only carry out normal maintenance, and do not allowed to
other disassembly and installation of the instrument
44
Automatic biochemistry analyzer BK-200 user manual
45
Automatic biochemistry analyzer BK-200 user manual
46
Automatic biochemistry analyzer BK-200 user manual
47
Automatic biochemistry analyzer BK-200 user manual
Step 6: Confirm the installation, click “Next” to install, the installation process shown
in Figure 2-14
Figure 2-13 Software Installation Procedure Figure 2-14 Software Installation Procedure
Step 7: Installation is complete, as shown in Figure 2-15, click "Close" button, close
the form.
48
Automatic biochemistry analyzer BK-200 user manual
49
Automatic biochemistry analyzer BK-200 user manual
50
Automatic biochemistry analyzer BK-200 user manual
51
Automatic biochemistry analyzer BK-200 user manual
52
Automatic biochemistry analyzer BK-200 user manual
Step 4: Select "Select from computer’s driver list" and click "Next" to open the sub-
interface as follows:
Step 6: Click the "Browse" button and then select the CD-ROM drive to find
“ ”, open the folder and select driver file slabvcp.inf, as shown in
Figure 2-26:
53
Automatic biochemistry analyzer BK-200 user manual
54
Automatic biochemistry analyzer BK-200 user manual
55
Automatic biochemistry analyzer BK-200 user manual
56
Automatic biochemistry analyzer BK-200 user manual
57
Automatic biochemistry analyzer BK-200 user manual
58
Automatic biochemistry analyzer BK-200 user manual
Run the desktop icon , When you run the software for the first time, it will tip
On line setup, as shown in Figure 3-1. Click "OK", the dialog box shown in Figure 3-
2 will pop up and then click the software configuration. If the following figure 3-3
dialog box appears, the software configuration is successful.
Run the desktop icon again, open the software, enter the login interface, as
shown in Figure 3-4, input the initial user name: 1000, the initial password: 1000,
click the button and the software opens.
59
Automatic biochemistry analyzer BK-200 user manual
60
Automatic biochemistry analyzer BK-200 user manual
change to "en", the software interface is English. Click "File / Save" is ok, as shown
in Figure 3-7.
"[Language]" of document.
61
Automatic biochemistry analyzer BK-200 user manual
the menu bar, toolbar, workspace, the status bar ,as shown in 3-10
62
Automatic biochemistry analyzer BK-200 user manual
2) Status Bar
Located in the top right of the main interface,real-time display the current user,
display computer system time, as well as current working state or testing process.
: Display the current operator's name, the user can be added and
deleted in “User setting”.
3)Tool bar
Located in the main interface of the upper left, several commonly used functions of
the software are placed in the toolbar. It is convenient to user for corresponding
operation, as shown in figure 3-12, selected by mouse click.
63
Automatic biochemistry analyzer BK-200 user manual
5)Version information
Software version shows in the status bar,as figure 3-14 shown.the way
64
Automatic biochemistry analyzer BK-200 user manual
65
Automatic biochemistry analyzer BK-200 user manual
66
Automatic biochemistry analyzer BK-200 user manual
-16:
Non-modal forms:After this form is opened, other forms can also be operated. As
67
Automatic biochemistry analyzer BK-200 user manual
68
Automatic biochemistry analyzer BK-200 user manual
Radio box:Choosing only one function among several functions, as shown in Figure
3-19:
69
Automatic biochemistry analyzer BK-200 user manual
This instrument uses the whole journey mode,when the reaction disk is rotated, the
absorbance of the pure water and the reaction solution in the cuvettes is determined
continuous.
70
Automatic biochemistry analyzer BK-200 user manual
The reaction disk rotates in 1 laps for about 18 seconds, and the absorbance value is
measured by the 120 cuvettes when the optical axis of a spectrophotometer is passed.
36 times (36 metering points) were determined for each cuvettes.
Light emitted from the light source is focused by a lens, first pass the cuvettes, then to
the photoelectric receiver transformed into electrical signals. After the two-stage
amplification and Log conversion obtained after the absorbance or absorbance rate of
71
Automatic biochemistry analyzer BK-200 user manual
72
Automatic biochemistry analyzer BK-200 user manual
Among them:
73
Automatic biochemistry analyzer BK-200 user manual
AL AL 1 ... AM 1 AM
AX
(L M )
(c) Concentration calculation
C X {K ( AX A1 )} IFA IFB
74
Automatic biochemistry analyzer BK-200 user manual
absorbance between these two metering points are used to calculate sample
concentration, called 2 end-point, reaction curve shows below:
【L】-【M】-【0】-【0】(1<L<M≤49)
Among them:
a
S Rj
j 1
b
S Ri
k= i 1
75
Automatic biochemistry analyzer BK-200 user manual
absorbance. IFA, IFB is a constant of the instrument, are represented by the slope and
intercept.
(a)Metering points:
【L】-【M】-【0】-【0】(1<L<M≤49)
(b)Calculation of absorbance
The metering points M and M-1 absorbance metering point average and the average
of L and L-1 absorbance subtraction, the difference divided by the time as absorbance,
is calculated as follows:
76
Automatic biochemistry analyzer BK-200 user manual
AM AM 1 AL AL 1
AX 2 2
t
Among them:
B is the calibration solution 1 (reagent blank) absorbance, IFA and IFB is a constant
of the instrument, are represented by the slope and intercept.
77
Automatic biochemistry analyzer BK-200 user manual
【L】-【M】-【0】-【0】(1<L<M≤49、 L+2<M)
AX=△A(M-L)
minute. IFA、IFB is a constant of the instrument, are represented by the slope and
intercept.
5.2.2 Calibration Method
(1) One point linear method (K factor method)
Though measuring the calibration solution 1’s absorbance and input K factor to get
working curve, shows as below:
78
Automatic biochemistry analyzer BK-200 user manual
79
Automatic biochemistry analyzer BK-200 user manual
80
Automatic biochemistry analyzer BK-200 user manual
Cri
n
Cr Ai A
X: i 1
Cri Cr
n
2
Y: i 1
n
n
Cr : Cri / n
A : Ai / n i 1
i 1
A1, A2 is the two measured values of calibration solution (1), n is the number of
calibration solution N × 2, Cri is the concentration calibration solution (i) .
(c) Calculation of concentration
81
Automatic biochemistry analyzer BK-200 user manual
C X K AX B C1 IFA IFB
(4)Logit-log4P(nonlinear method)
Suitable for the working curve that absorbance showed convergence with the
concentration increase, Logit-log4P(Nonlinear method) calibration curve shown in
Figure 5-10: applies to the concentration increases with the absorbance showed a
smooth work curve, Logit-log4P(nonlinear method) Calibration curve shows as 5-8:
Calibration Model:【Logit-log4P】
minute.
82
Automatic biochemistry analyzer BK-200 user manual
(c)calculation of concentration
C X (C C1 ) IFA IFB
K
AX B
1 aC b
1 K ( AX B )
C b
a AX B
(d)calculation of SD
N 2
A Ai ,
2
ij
i 1 j 1
SD
2N 4
(N=4~6、j=1or2)
(Aij-Ai’) is d-value of absorbance between Ai’ from fitting equation and Aij from
testing or the d-value between Aij and A12. Every calibration fluid test twice, and the
maximum value of Aij is 12.
(e)Applicable methods
(6)Logit-log5P(Nonlinear method)
Same characteristic with Logit-log4P, and Logit-log5p has one more calculate
parameter, so the result is more accuracy. Calibration is shown curve as Pic 5-9.
83
Automatic biochemistry analyzer BK-200 user manual
minute.
(c)Calculation of concentration
AX B
a b lnC c C ln 0
K AX B
Get C from Newton Approximation
C X (C C1) IFA IFB
K
AX B
1 exp a b l n C c C
84
Automatic biochemistry analyzer BK-200 user manual
(d)Calculation of SD
N 2
A Ai ,
2
ij
i 1 j 1
SD
2N 5
(N=5~6、j=1or2)
(Aij-Ai’) is d-value of absorbance between Ai’ from fitting equation and Aij from
testing or the d-value between Aij and A12. Every calibration fluid test twice, and the
maximum value of AU is 12.
(e)Applicable methods
Test from fluid1 to fluid 5 or fluid6, and get the curve line, and link those points with
straight line. As pic 5-10 shown.
85
Automatic biochemistry analyzer BK-200 user manual
S1ABS is average value of twice test for calibration fluid1 for absorbance or change
value of absorbance.
C2 C 1
K
A2 B
B: Absorbance or change rate of Calibration Fluid1.
A2: Absorbance or change rate of Calibration Fluid2.
C1: Concentration of Calibration Fluid1.
C2: Concentration of Calibration Fluid2.
Calculate K2, K3, K4,K5 with same method
(c)Calculation of concentration.
C X K N ( AX AN ) C N IFA IFB
(d)Applicable methods
In this line, every value of calibration linked to a complete curve, and the error is also
fitting in the curve, so the curve fitting is better than poly line. The calibration curve
is shown as pic 5-11.
86
Automatic biochemistry analyzer BK-200 user manual
(c)Calculation of concentration
3
AX a ( I ) b ( I ) C X C ( I ) c (I) C X C ( I )2 d (I) C X-C(I)
f (C X C ( I )) a ( I ) b ( I ) (C X C ( I )) d ( I ) (C X C ( I )) 2 d ( I ) (C X C ( I )) 3 AX
Get C from Newton Approximation
C X (C C1) IFA IFB
CX is concentration of testing sample, C1 is blank concentration. C2~CN is
concentration of calibration fluid. AX is absorbance of sample or the change value per
minute. A2~AN is absorbance of calibration fluid or the change value per minute.
IFA&IFB is coefficient of analyzer, which displays the slope and intercept.
(d)Calculation of SD
N 2
A Ai ,
2
ij
i 1 j 1
SD
2N 4
(N=5~6, j=1or2)
(Aij-Ai’) is d-value of absorbance between Ai’ from fitting equation and Aij from
testing or the d-value between Aij and A12. Every calibration fluid test twice, and the
maximum value of Aij is 12.
(e)Applicable methods
87
Automatic biochemistry analyzer BK-200 user manual
88
Automatic biochemistry analyzer BK-200 user manual
89
Automatic biochemistry analyzer BK-200 user manual
System
9. testing process
monitor
(1) system monitoring
Add pause / Test status of real time monitoring
(2) suspension and
continue instruments
continuation of adding
Emergency Edit the sample in the test process and
sample
stop click start testing.
(3) emergency stop
Sample
(4) additional sample
registration
90
Automatic biochemistry analyzer BK-200 user manual
Warning:
The BK series of cleaning solution should be regarded as liquid corrosive liquid.
Once you hurt the skin or eyes, rinse with plenty of water.
6.2.2 Power on and Login
(a) Power on the automatic biochemical analyzer. The switch on the left side of the
91
Automatic biochemistry analyzer BK-200 user manual
analysis apparatus is power switch, when a reagent on the reagent plate the switch
should be in the open state, to ensure the normal operation of the cooling system.
Switch on the right side of the equipment is power supply for analysis apparatus.
(b) Login biochemical analyzer application software, if you want to test, you should
first execute online instruction.
6.2.3 Light checking
(a) A/D Reading
In the light conditions, click the button in the menu button , after
entering the maintenance form, choose the "A/D reading" option in the maintenance
list, click "Y" button, the instrument will perform the inspection of light, the amount
of light with previous test results values are displayed in the result bar. As shown
below:
Click on the menu bar button , open the "absorbance test" window,
covers the cuvettes with an opaque material, fix the reaction tank , click the
button, The instrument starts reading, reading completed, shows the cup number and
the corresponding absorbance values, you can use the scroll bar up and down on the
right data interface to choose to read all the cup data.
92
Automatic biochemistry analyzer BK-200 user manual
Click the software Toolbar " " menu, open the temperature display, observe the
In the menu bar button, click the , enter the interface of the items
parameter, to determine the items parameter, as shown in figure 6-3.
93
Automatic biochemistry analyzer BK-200 user manual
add to the “optional items list", click" "to cancel relative item operation, then
click" save "to complete the operation.
94
Automatic biochemistry analyzer BK-200 user manual
95
Automatic biochemistry analyzer BK-200 user manual
Click the menu bar " "show "reagent information" operating interface, as
shown in Figure 6-6, scan the code with a sweep code reader barcode according to the
reagent panel .
96
Automatic biochemistry analyzer BK-200 user manual
Click the menu bar " software show " reagent information" interface, as
shown in Figure 6-7, Query reagent project according to the order of reagent disc ,
and select the type of reagent packaging specifications and click Modify, then you can
input.
97
Automatic biochemistry analyzer BK-200 user manual
Warn: No. 28 of reagent plate only can put BIOBASE series machine special
cleaning fluid.
6.2.6.3 Residual amount of reagent
In the "reagent information" form, click the " " key, the instrument
automatic detection reagent remaining amount, the remaining test number, the results
of the test information is displayed in the "reagent information" list box.
98
Automatic biochemistry analyzer BK-200 user manual
Factor — CK-MB
99
Automatic biochemistry analyzer BK-200 user manual
Click the menu bar " " software "calibration" interface, click on the lower
right side of the window, select " " , show “control parameter
"interface, as shown in figure 6-10.
The steps of adding new project quality control: first in the "quality control
parameters" click " " button, according to the quality manual batch fill in
"batch", as shown in Figure 6-11, choose the items need quality control and then click
100
Automatic biochemistry analyzer BK-200 user manual
After the preservation, the quality control parameters interface will be shown as figure
6-12.
Click the " " button in the toolbar, enter the "entry" form, as shown in
the figure below:
Single sample entry: click , and choose test items in the left column,
101
Automatic biochemistry analyzer BK-200 user manual
click " " to move test item to “selected test items "column, the sample number is
generally from the beginning of 1, followed by accumulation, a number can not be
repeated, and then fill in the information of patients" name "," gender "," age ", Edit
“inspection department "and" sample type "," submission of doctors' information,
the fill in the quantity and starting position number that to be tested,
sample position number imputing is from1 to 60, from 60 on, sample position number
plus 1, increasing gradually start from 1, save the selected items to be tested, complete
the operation.
Project portfolio login: if in the process of sample inputting, the project selected is the
combination that has been set up, you can directly select in the combination bar that
in the left side of interface, click directly add to the “selected project
determination "column.
You can edit the information you have input. In the landing list column, you can
choose one sample number to choose, which display the patient information and
testing items information, press" " or " " button, to edit, delete
102
Automatic biochemistry analyzer BK-200 user manual
set finished, click the start button , will pop up the dialogue as shown in the
Figure 6-14, choose “cleaning solution”, automatically adding cleaning solution to
reaction cuvette and cleaning the used reaction cuvette. Click “Start”, enter the
process of sample testing.
During the sample testing, click on the menu bar to perform real-time
monitoring of the instrument sample tray, reagent tray, reaction tray status, as shown
in Figure 6-15 b).
103
Automatic biochemistry analyzer BK-200 user manual
b) Status Monitor
Figure 6-15 Sample Test
Sample tray monitoring: click “sample trays status” under the “monitor” form, as
shown in Figure 6-16, wherein the sample tray with different colors to distinguish the
status of samples.
Reagent tray monitoring: click the “reagent tray status” in the "monitor" form,
respectively, the reagent plate test status will display in the left of the form, as shown
in figure 6-17:
Figure 6-16 sample plate of monitoring Figure 6-17 reagent disk monitoring
Reagent disk by serial number indicates that the reagent placed number, when the
instrument starts testing, reagent probe after taking the reagent, the software will
automatically calculate the residual of the reagent, and the reagent information
104
Automatic biochemistry analyzer BK-200 user manual
displayed on the inside. In the reagent tray: white indicates empty yellow indicates
less reagent, green indicates that the reagent sufficient, Red means no reagent.
The reaction disk detection: The reaction disk in real time display the status of the
reaction at the time of sample testing. As shown in Figure 6-18, where in the different
colors represent the cuvette in different working conditions.
Sample test in progress, click push button in the editor toolbar “ ”, “ ”or
Click " " to temporarily stop the test process, click ' ' to continue testing, click
105
Automatic biochemistry analyzer BK-200 user manual
Figure 6- 19 Print
As shown in the figure, directly select the date, in the "Print Report" ,you can query
historical test data. Query process, choose the corresponding date in the “date”,then
Data statistics include test statistics, workload statistics and cost statistics, as shown
below.
106
Automatic biochemistry analyzer BK-200 user manual
A) Test Statistics
B ) Workload Statistics
107
Automatic biochemistry analyzer BK-200 user manual
C ) Expensive statistics
“retest”.
In this interface, choose the retest items , firstly, click 'Edit' to select the re-testing
items, then, Click "Save" to complete the retest item input. If the sample are testing,
the retest items will be tested automatically. If the machine is not run, find the
“sample test” , restart the test task.
NOTE : Only the test data of the today , which can be re-tested
108
Automatic biochemistry analyzer BK-200 user manual
In software interface, click the , then, click the “ CalibrationPar ”, you can
check the calibration parameter, and change the calibration parameter.
7.1 Calibration
7.1.1 Calibration Parameters Settings
Click “ Item Setting”, then click the “CalibrationPar” , input the calibrator parameters.
109
Automatic biochemistry analyzer BK-200 user manual
After we finished the input reagents parameters, in this interface, select test items
which need be calibration. Then click on the right “ ”. After input the
correct parameters, in "Calibration type", "Standard counts", "Repeat counts",
"Sample type", "Concentration", "Cup No.", finally click "Save". "Concentration" and
"Cup No" , need according to the calibrator and the calibrators location number in
reagents tray , then click 'Edit'.
7.1.2 Calibration test
In the “ Program Input” interface, click the “ Calibration & QC Input ”, and select the
need calibrator test items, click the , then, click the “Save” , will get the a warn
and tell you “Save successful”, as the Figure 7-2 shown . After click the “ Sure” ,
110
Automatic biochemistry analyzer BK-200 user manual
111
Automatic biochemistry analyzer BK-200 user manual
In Curves interface, at “Test Date”, select the calibration date and the need queried
items. Then, we can get the calibration curve.
( b) Calibration result
Under “Item setting” interface , click " CalibrationPar " interface , Directly select the
test items result that you need to view, as shown in Figure 7-4.
As shown in Figure 7-4, in " CalibrationPar " interface , firstly, in the "item list"
select the items name that you want to view, then, in this interface, select the Date,
you query the calibration result.
112
Automatic biochemistry analyzer BK-200 user manual
In the original window, click the “ Data Processing” , then, click the “QC
results query", firstly select the QC items and “ QC Lot No” , According the “Test
Date”, we can get the QC Control result.
And, if you want get the trend of every time test result, Can select the “ Sandard QC
Chart” “Relative QC Chart” “ Display QC Value”
113
Automatic biochemistry analyzer BK-200 user manual
114
Automatic biochemistry analyzer BK-200 user manual
115
Automatic biochemistry analyzer BK-200 user manual
the advanced user login software, you can select a user left button, click "delete" to
delete the current user.
(B) Password Modify
Click the “Change User Password”, can change the being used account’s password,
8.2 About
"About" is a software version of the introduction.
116
Automatic biochemistry analyzer BK-200 user manual
Chapter 9 Maintenance
9.1 Prepare before System Maintenance
To ensure the precision and accuracy of the fully automatic biochemistry analyzer, the
operator must obey “BK-200 Fully Automatic Biochemistry Analyzer User Manual”
for the operation and required maintenance. Thus the instrument will provide reliable
results and ensure longer service life of the instrument.
9.1.1 Required Tools
Before the system maintenance, please prepare below tools:
1. Tools in Accessory box
cleaning probe)
In the daily operation and maintenance of the system, use deionized water, electrical
conductivity is less than 1 us/cm. When using pure water equipment, please do not
forget the regular maintenance and inspection of water purification equipment. Please
refer to the manual of the pure water device, or contact the manufacturer or distributor
of the water purifier.
9.1.3 Washing Solution
Washing solution is used for cleaning the equipment, should be provided by Biobase.
If replaced by other cleaning liquid, may cause a reaction cup, stir needle, needle,
needle cleaning and other cleaning is not clean, thus affecting the accuracy of test
117
Automatic biochemistry analyzer BK-200 user manual
results and precision. Biobase is not responsible for inaccurate measurement of the
equipment not use Biobase washing solution.
The fully automatic biochemistry analyzer is mainly used alkaline cleaning liquid, the
designated position is the reagent position 28, for cleaning the reaction cup.
9.2 Software Maintenance
9.2.1 Initialization
Click the software Toolbar "initialization", the instrument automatically back to zero
position. In the reset process, do not allow emergency stop. Other operations can be
carried out after the initialization is completed.
Attn: When the instrument is powered into the standby state, each position is not in
the zero position, and it is necessary to enter the software click initialization
command to perform an initialization operation to return to zero!
9.2.2 Instrument Adjustment
9.2.2 .1 Position Adjustment
Sheet 9-1 Adjustment Position
Component Cleaning Reaction Sample Reagent Reagent
name position disk disk inner ring outer ring
plate
dispensing Cleaning 1 1 1 29
arm position
Stirring arm Cleaning 105 # # #
position
Cleaning Cleaning 75 # # #
arm position
Attn:
1, Vertical cleaning position of the dispensing probe: thinner part of the needle should
be completely in the cleaning cup. Dispensing probe vertical depth of reaction cup:
point of probe is 30 steps distance to the bottom of the reaction cup. Reagent depth is
20 steps distance to the bottom of the reagent bottle; Sample depth is 10 steps distance
to the bottom of sample cup.
118
Automatic biochemistry analyzer BK-200 user manual
2, Cleaning position of the mixer should be enough deep but not touch the bottom of
the cleaning cup; Reaction position of mixer should be 20 steps distance to the bottom
of the reaction cup.
3, Cleaning arm position adjustment should make sure all the probe except the one
with dry unit, just reach the bottom of the cuvettes; the last probe should 2mm lower
than other probes.
9.2.3 Optical System Maintenance
9.2.3 .1 Light Spot Inspection
To take away the cuvettes which block aperture in the reaction disk, use darker paper
block the light source, the light spot formed on paper. The paper moves back and
forth to see the brightest spot on the track cup campaign. Spot shape around a long
5mm, rectangular spot width of about 1.5mm, If there is a problem spot brightness,
check halogen installation location is correct (This position is generally not adjusted).
9.2.3 .2 AD Stability Adjustment
After adjust the AD board voltage. Click Software Toolbar "instrument check" button,
enter "instrument check" interface. If the optical system △ OD has larger changes,
Please adjust as follows:
(A) Check the lamp is uniform ,spot size is suitable.
(B) 12V input voltage is stable, if separate power supply or not.
(C) Grounding line is good; requirements to external and internal ground only on a
single-point ground power supply; external power supply on the ground must be
connected.
9.2.3.3 AD Absorbance Test
With an opaque material to cover a joint cuvette, place the opaque cuvette into
reaction disk.
Click on the software menu bar “maintenance”, Display "Absorbance Test" interface.
As shown in Figure 9-1:
119
Automatic biochemistry analyzer BK-200 user manual
120
Automatic biochemistry analyzer BK-200 user manual
L frame has four waterways adapter, two small linker is water inlet. The installation
of water pipes should be as short as possible, and install water fall head, Ensure the
machine waterway is normal.
Attention: When installation of water pipes make sure the smooth flow of
waste water outlet.
Each cleaning probe connected with a corresponding waterway through a
polyethylene tube.
When added head or pumping head is abnormal, sequentially check the pipe joints are
connected, solenoid valves and micro diaphragm pump.
9.3 Check before test
9.3.1 Power on Checking
When finished the installation of all parts and water system, please power on the
machine and check the conditions:
As below figure 9-3 shows, power supply port includes USB power supply and three
pin power supply.
Attn: before connect USB wire, please install USB device drive.
121
Automatic biochemistry analyzer BK-200 user manual
and down again, and the air in the plunger pump and the pipeline is excluded.
During the air out procedure, do not suddenly stop. Execute other movements
when it finished.
Attn: After replacing piston pump or pipeline, execute air out procedure.
9.4 Software Problem
9.4.1 Printing Problem
9.4.1.1 change print paper type
Open the software installation folder, find “ ”and open it. Find [print
page], set the print paper type, A4, A5 and B6. For example the print paper type is
A5, .
122
Automatic biochemistry analyzer BK-200 user manual
123
Automatic biochemistry analyzer BK-200 user manual
124
Automatic biochemistry analyzer BK-200 user manual
125
Automatic biochemistry analyzer BK-200 user manual
Click “login” button and at the same time press the “Ctrl” button. Enter software
operation interface. Click “Ctrl + Alt + R”, then click “ok” as figure 9-16 shows and
click “exit” button to exist the software.
Click login again. Appears figure 9-17interface. Click software configuration, appears
Figure 9-18 interface, that means the software configuration is successful.
Attn: One computer can only control one biochemistry analyzer, if move
126
Automatic biochemistry analyzer BK-200 user manual
Figure 9-20,click the “Repair” button, waiting interface as Figure 9-21, finish
127
Automatic biochemistry analyzer BK-200 user manual
Regular clean, check and replace part, please see sheet 9-2(Supposed using the
6 Halogen lamp 1 2 ●
7 Cleaning probe ○
8 Mixer ○
9 Dispensing pump ●
Water supply
10 ○
filter net
128
Automatic biochemistry analyzer BK-200 user manual
Reagent tray
11 ○
cooling system
12 Cooling fan ○
Purified water
13 ○ ●
system
Waste water
14 ○
exhaust
Note:
(b)Halogen lamp life is 2000 hours. To ensure the accuracy and precision, suggest
(c)Instrument is suitable for wire printers, inkjet printers, and laser printers, the
(d)Execute save the blank weekly. Otherwise will indicate cell blank abnormal
alarm.
purifier consumables.
9.5.2 Users regular replacement accessories sheet
Please prepare below accessories, in case any faults need replacement, please check
sheet 9-3:
129
Automatic biochemistry analyzer BK-200 user manual
! Attention:
△
●Do not drop water, reagent or cleaning fluid to any mechanical or electrical parts,
avoid damage to the instrument.
●When operating, do not touch dispensing probe, mixer and cleaning probe. Avoid
the risk of infection or injury.
●During operation, should take preventive measures, with protective gloves, wear
work clothes. Otherwise, it is possible to contact contaminated areas, contaminated
fluid and infected, or contact with corrosive liquids and damage the skin. If there is
contamination or corrosive liquid accidentally touching the body, please immediately
washing by water and disinfection.
9.5.3.1 Dispensing probe maintenance
If internal and external of the probe is polluted, it is easy to attach serum, reagent and
water drops, also easy to cause the blockage of the syringe tube inside, thus affecting
the instrument test results. Therefore, it is necessary to regularly check and timely
cleaning.
1. Clean up probe outside
(a) Close the total power switch of the instrument.
(b)Scrub the outer wall with cotton swab dipped in alcohol( During the process of
cleaning, do not bend the probe). As shown in figure 9-23
130
Automatic biochemistry analyzer BK-200 user manual
131
Automatic biochemistry analyzer BK-200 user manual
132
Automatic biochemistry analyzer BK-200 user manual
cannot pull liquid level signal lines, to prevent the terminal loss, as shown in figure 9-
28.
133
Automatic biochemistry analyzer BK-200 user manual
134
Automatic biochemistry analyzer BK-200 user manual
(c)A check of sample probe horizontal position: Click on the sample position “Y”,
sample probe turn left, when the sample probe stop at the upward side of sample
position of reagent sample plate, check whether the pinpoint of sample probe is in the
center of the sample cup, as shown in figure 9-35, then click the zero position "Y",
sample probe is placed back to the initial position.
135
Automatic biochemistry analyzer BK-200 user manual
Figure 9-35
Attention: If the sample with the tip of the needle is not in the center of the sample
cup, please contact maintenance personnel.
(d) Reagent inner ring level examination: Click the reagent inner ring "Y”, sample
probe turn left, when the sample probe stop at the upward side of reagent inner ring of
reagent sample plate, check whether the pinpoint of sample probe is in the center of
the reagent position of reagent inner ring, then click the zero position "Y", sample
probe is placed back to the initial position.
Attention: If the sample with the tip of the needle is not in the center of the reagent
bottle of reagent inner ring, please contact maintenance personnel.
(e) Reagent outer level examination: Click the reagent outer ring "Y”, sample probe
turn left, when the reagent probe stop at the upward side of reagent outer ring of
reagent sample plate, check whether the pinpoint of sample probe is in the center of
the reagent position of reagent outer ring, then click the zero position "Y", sample
probe is placed back to the initial position.
Attention: If the sample with the tip of the needle is not in the center of the reagent
bottle of reagent outer ring, please contact maintenance personnel.
(f) Reaction cup level examination: Click the reaction cup position "Y", sample probe
turn right, when the sample probe stop at the upward side of reaction cup of the
reaction plate, check whether the sample with the tip of the needle is in the center of
the reaction cup, as shown in figure 9-43, then click the zero position "Y", sample
probe is placed back to the initial position.
136
Automatic biochemistry analyzer BK-200 user manual
Figure 9-36
Attention: If the sample with the tip of the needle is not in the center of the reaction
cup, please contact maintenance personnel.
(g)Sample vertical position examination: Click the sample position "Y", sample probe
turn left and stop at the top of the sample position, click the vertical position "Y", the
sample probe will continue to drop until the needle tip touches the bottom of the
sample cup, then click the vertical position "Y", the needle is lifted above the sample
position, then click the zero position "Y", sample probe turn right and placed back to
the initial position.
(h) Reagent inner ring vertical position examination: Click the reagent inner ring "Y",
sample probe turn left and stop at the top of the reagent inner ring, click the vertical
position "Y", the sample probe will continue to drop until the needle tip touches the
bottom of the reagent cup, then click the vertical position "Y", the needle is lifted
above the reagent inner ring, then click the zero position "Y", sample probe turn right
and placed back to the initial position.
(i) Reagent outer ring vertical position examination: Click the reagent outer ring "Y",
sample probe turn left and stop at the top of the reagent outer ring, click the vertical
position "Y", the sample probe will continue to drop until the needle tip touches the
bottom of the reagent cup, then click the vertical position "Y", the needle is lifted
above the reagent outer ring, then click the zero position "Y", sample probe turn right
and placed back to the initial position.
(j) Reaction cup vertical position examination: Click the reaction cup position "Y",
sample probe turn right and stop at the top of the reaction cup, click the vertical
position "Y", the sample probe will go down, put the part of the probe which coated
with Teflon into the reaction cup, then click the vertical position "Y", the needle is
lifted above the reaction cup, then click the zero position "Y", sample probe turn left
and placed back to the initial position.
137
Automatic biochemistry analyzer BK-200 user manual
138
Automatic biochemistry analyzer BK-200 user manual
139
Automatic biochemistry analyzer BK-200 user manual
During the sample test, click on the menu bar , can realize real-time monitoring
of the reaction state of the instrument, if the dirty cup is detected during the test, the
instrument will automatically skip the dirty cup and use the next cup. In the
monitoring, the cup will show red, as shown in figure 9-41,
140
Automatic biochemistry analyzer BK-200 user manual
When there are too many dirty cup , test speed will be affected. If in the software
shows the dirty cup more than 1/3 of reaction tray, please replace new cuvette.
Operations are as follows:
(a)Turn off the instrument main power switch, remove the reaction plate tray;
(b)Wear protective gloves and unscrew the cuvette pins, as shown in Figure 4-42.
141
Automatic biochemistry analyzer BK-200 user manual
Attention1: If the used cuvette is exposed to air for long periods of time,
contaminants may build up on the cup wall. So should be promptly covered with
reaction tray cover. In addition, if the emergency stop during the test, should
clean cuvettes which not be cleaned or rinse them with pure water to avoid the
reaction fluid remain in the cuvette for a long time.
142
Automatic biochemistry analyzer BK-200 user manual
Attention: Do not touch the surface of the halogen lamp, otherwise the
amount of light will be affected. If you find fingerprints and other stains on the
surface, you can wipe with alcohol gauze.
(b) Turn off the instrument's main power switch, and after about 30 minutes (wait the
lamp compartment to cool completely), perform the next operation to avoid burns;
(c) Remove the three plugging plugs on the rear plate and unscrew the three screws
with a Phillips screwdriver. Operation as shown in Figure9-46;
143
Automatic biochemistry analyzer BK-200 user manual
144
Automatic biochemistry analyzer BK-200 user manual
145
Automatic biochemistry analyzer BK-200 user manual
146
Automatic biochemistry analyzer BK-200 user manual
147
Automatic biochemistry analyzer BK-200 user manual
148
Automatic biochemistry analyzer BK-200 user manual
Figure 9-55
NOTE: The pin must be securely fastened to prevent instrument malfunction.
(F) Raise the mixer arm to the top, and rotate the agitator arm in the direction of the
cuvettes by hand (same as reagent arm operation), as shown in Figure 9-56, so that the
height of the stirrer is about 1.3cm.
Figure 9-57
149
Automatic biochemistry analyzer BK-200 user manual
See the location of the pin in relation to the cuvette in a top-down view, as shown in
Figure 9-58:
Figure 9-58
Note: If not in the center of the reaction cup, please contact the service staff.
(i)Mixer probe’s vertical position check: In the "Instrument Check" window, click the
cuvette position "Y", the probe will move to the top of the cuvette, then click the
vertical position "Y", the mixer probe down. Click the vertical position "Y" again, the
mixer probe lift. Finally click the zero position "Y", the mixing probe will return to
the top of the cleaning cup.
9.5.3.6 Reagent sample tray maintenance
The instrument reagent sample tray has reagents refrigerated storage reagent function.
Reagent sample tray is divided into sample position, reagent inner ring and reagent
outer ring. 49 sample position, 56 reagent position, as shown in Figure 9-59.
150
Automatic biochemistry analyzer BK-200 user manual
Attention: Reagent can not be too full to ensure that below the top of the
reagent location as above figure, otherwise it will affect the level detection.
1. Replace the reagent bottle
When reagent less than the remaining time, please replace the reagent bottle. 50mL
reagent bottle can only be placed in the reagent inner circle, 20mL reagent bottle can
be set not only in the reagent outer ring, but also in the reagent inner ring. Place the
reagent bottle as follows:
(a) 50mL reagent bottle placed: bottle near the reagent outer ring;
(b) 20mL reagent bottle placed in the reagent outer ring: the mouth of the bottle near
the reagent inner ring;
(c) 20mL reagent bottle placed in the reagent inner ring: the bottle near the reagent
outer ring, and needed for fixed block (box attached) as shown in Figure 9-60. After
placing the reagent bottle, fix the reagent bottle with the fixing piece and then insert
the fixing block vertically into the bottom.
Attention: Must put the reagent bottle first, and then put a fixed block, to
prevent friction damage to the reagent bottle bar code.
2. Reagent sample tray clean
The sample tray will be contaminated with reagents, samples and dust for a long time
use, and condensation water will be generated during the cooling process, so it should
be cleaned once a day.
(a) Remove the reagent sample tray cover;
(b) Unscrew the reagent tray handle counterclockwise to remove the reagent sample
tray.
151
Automatic biochemistry analyzer BK-200 user manual
Figure 9- 63 Alarm
The procedure for adding the refrigerant is as follows:
(a) Turn off the main power switch of the instrument;
(b) Remove the three plugging plugs on the right side plate and unscrew the three
screws with a Phillips screwdriver, as shown in Figure 9-64;
152
Automatic biochemistry analyzer BK-200 user manual
153
Automatic biochemistry analyzer BK-200 user manual
154
Automatic biochemistry analyzer BK-200 user manual
155
Automatic biochemistry analyzer BK-200 user manual
156
Automatic biochemistry analyzer BK-200 user manual
157
Automatic biochemistry analyzer BK-200 user manual
158
Automatic biochemistry analyzer BK-200 user manual
Sheet 9-5 The analysis and solution of the instrument failure of alarm table
Alarm Alarm describe Treatment recommendations
number source
512 Sample The sample plate is Press the shortcut key [Ctrl]+[F3] to
plate not moved to the reset the device and try again. If the
specified position problem persists, contact your service
person
1290 Sample Not down Press the shortcut key [Ctrl]+[F3] to
arm reset the device
1802 R2 arm Not down Press the shortcut key [Ctrl]+[F3] to
reset the device
1803 Sample Sample needle’s make complete the sample quantity
needle pipetting not enough, and re test
position[{0}].
1804 R1 arm Not detect the Recommend the implementation of a
reagent, reagent residue detection to ensure that
position[{0}]. there are enough reagents in the
reagent bottle
1805 R2 arm Not detect the Recommend the implementation of a
reagent, reagent residue detection to ensure that
position[{0}]. there are enough reagents in the
reagent bottle
1806 Sample Sample needle block Recommend the implementation of a
needle reagent residue detection to ensure that
there are enough reagents in the
reagent bottle
2059 AD Unable to receive Please contact service person
module absorbance data of
module
2060 AD Data acquisition Please contact service person
module module interrupt
abnormal
P45
159
Automatic biochemistry analyzer BK-200 user manual
160
Automatic biochemistry analyzer BK-200 user manual
code={1}:Unknown
new reagent
200016 Reagent Reagent Please confirm the reagent bar
plate position={0},bar code and then scan it again
code={1}:Reagent bar
code is invalid
161
Automatic biochemistry analyzer BK-200 user manual
Item code={2},Repeat
number={3}。
300008 data processing An error occurred during Please check the previous
the automatic calibration project fitting parameters
of reagent {0}: {1}. and status, and to ensure
that the standard quantity is
sufficient for fitting the
current
300009 data processing Reagent bottle empty, Please check the location
162
Automatic biochemistry analyzer BK-200 user manual
163
Automatic biochemistry analyzer BK-200 user manual
164
Automatic biochemistry analyzer BK-200 user manual
165
Automatic biochemistry analyzer BK-200 user manual
166
Automatic biochemistry analyzer BK-200 user manual
disk
100025 Stirring Stirring arm initialization Please restart, or contact
arm error technical support person
100026 others Unable to get the basic Please check the database
information from the service is correct, or contact
database <= Alarm ID={0} technical support person
167
Automatic biochemistry analyzer BK-200 user manual
168
Automatic biochemistry analyzer BK-200 user manual
Note: In the figure, the dotted line box is reserved for defoaming function.
169
Appendix C Products supporting reagents
Specific table below (BIOBASE commonly used biochemical reagents parameter table)
ALP Alkaline Phosphatase 6 240 60 405 Rate method serum 0 U/L 0 3 22 30 0 0 135 45
ALT Alanine Amino Transferase 22 240 60 340 Rate method serum 1 U/L 0 3 23 33 0 0 41 0
ApoA1 Apolipoprotein A1 5 225 75 340 Two Point End serum 2 g/L 0 3 35 35 12 13 1.9 1.2
ApoB Apolipoprotein B 5 225 75 340 Two Point End serum 2 g/L 0 3 35 35 12 13 1.5 0.6
ASO Antistreptococcus O 5 240 60 578 Two Point End serum 0 IU/mL 0 3 28 29 19 20 166 0
AST Aspartate Amino Transferase 22 240 60 340 Rate method serum 1 U/L -1 3 22 33 0 0 40 0
BMG β2-Micro Globulin 5 225 75 578 Fixed time method serum 1 mg/L 0 3 21 28 0 0 1.8 0.8
CHE Cholinesterase 5 250 50 405 Rate method serum 0 U/L 0 3 22 30 0 0 12600 3930
CHO Cholesterol 4 300 0 510 One Point End serum 2 mmol/L 0 3 15 15 0 0 5.2 2.34
170
CK-MB Creatine Kinase Isozyme 15 240 60 340 Rate method serum 1 U/L 0 3 22 30 0 0 25 0
CRP C-reaction protein(normal) 20 225 75 340 Two Point End serum 2 mg/dL 0 3 35 35 12 13 0.8 0
GLU Glucose(Oxidase Method) 6 300 0 510 One Point End serum 2 mmol/L 0 3 34 35 0 0 6.4 3.89
HDL-C HDL-Cholesterol 5 225 75 546 Two Point End serum 2 mmol/L 0 3 35 35 12 13 2.25 0.77
LDH Lactic Dehydrogenase 6 240 60 340 Rate method serum 0 U/L 0 3 22 30 0 0 225 135
LDL-C LDL-Cholesterol 5 225 75 546 Two Point End serum 2 mmol/L 0 3 35 35 12 13 3.35 0
TBA Total Bile Acid 5 225 75 405 Fixed time method serum 1 umol/L 0 3 22 28 0 0 20 0
TG Triglyceride 5 300 0 510 One Point End serum 2 mmol/L 0 3 35 35 0 0 1.7 0.7
UA Uric Acid 5 240 60 546 Two Point End serum 0 umol/L 0 3 35 35 12 13 480 140
UREA Urea 5 225 75 340 Fixed time method serum 1 mmol/L 0 3 22 30 0 0 8.3 1.7
171
α-HBDH α-Hydroxybutyric Acid Dehydrogenase 5 240 60 340 Rate method serum 0 U/L 0 3 22 30 0 0 182 72
GGT/γ-
γ-GlutamoylTransferase 6 225 75 405 Rate method serum 0 U/L 0 3 22 30 0 0 47 0
GT
172
Appendix D Cross contamination reference table
The above cross-contamination only is taken for example when BIOBASE reagents
is tested on the BIOBASE series of automatic biochemical analyzer.
The reagent formula’s changing in cross-contamination, so the above test is only for
reference, if not, please refer to the actual test situation.
173