Medrad Stellant D PM Checklist
Medrad Stellant D PM Checklist
Medrad Stellant D PM Checklist
Original Parameters:
Phase 1: ____________________________________________________________________________________________
Phase 2: ____________________________________________________________________________________________
Phase 3: ____________________________________________________________________________________________
Phase 4: ____________________________________________________________________________________________
Phase 5: ____________________________________________________________________________________________
Phase 6: ____________________________________________________________________________________________
Display: Part # 3007032 Serial # : __________ Base: Part # 3007039 Serial # : __________
Injector Head: Part # (circle one) 3007033 (New) or 3007038 (Old) Serial # : __________
Size Sensor Assy A and B: Part # (circle one) 3013368 (New) or 3009234 (Old)
Power Regulator PCB (below Size Sensor Assy A): Part # 3007297 Serial # : __________
Power Drive Cards A and B: Part # 3007293 A Side Serial # : __________ B Side Serial # : __________
Interface PCB (underside of injector head): Part # (circle one) 3007294 (New) or 3011778 (Old) Serial # : ________
Servo PCB (under injector head display): Part # (circle one) 3007431 (New) or 3011675 (Old) Serial # : ________
____________________________________________________________________________________
Pass Fail
Operational Checkout Procedure
1.1. Remote Panel Screen
1.2. Coiled Start Hand Switch (if applicable)
1.3. Remote Cable
1.4. Base Unit Cables
1. Visually Inspect Remote Panel 1.4.1. Power Cord
1.4.2. Remote Panel Cord
1.4.3. Injector Cord
1.5. Remote Panel Securely Mounted
2.1. Integrity of Buttons
2.2. Pivot Knuckle
2. Visually Inspect Injector Head 2.3. Integrity of Syringe Wells
2.4. Injector Head Cable and Extension Cable
3.1. Pedestal
3.1.1. Head Swings Freely on Post
3.1.2. Weights are Secure
3.1.3. Moves Freely About the Room
3.1.4. Cable Neat and Orderly
3. Visually Inspect Head Mounting 3.2. Counter-Poise Arm
Device 3.2.1. Properly Installed; Secured to Unistrut
3.2.2. Screws are Tight
3.2.3. Injector Head Secured to Post
3.2.4. Arm Swings Smoothly; No Drift
3.2.5. Integrity of Spring Section
3.2.6. Cable Neat and Orderly
4.1. Remote Panel Powers Up Properly
4.2. Annotate Customer’s Last Settings
4.3. Proper Volumes Displayed
4.4. Proper General Injector Operation
4. Operational Checks
4.5. Proper Auto-Load Injector Operation
4.6. Proper Auto Advance Injector Operation
4.7. Smooth Plunger Motor Movement
4.8. Heat Maintainers Get Warm (if applicable)
Record Main System #, unit serial #s, local identification
5. Identification #s, and board serial numbers on Inventory Check Sheet Complete ______
Syringe A
10 secs +/- .5 secs A Time: ______ secs Pass _____ Fail _____
Syringe A
20 s ± 5 s Fast Forward A FF __________ secs Pass _____ Fail _____
20secs ± 5secs Fast Reverse B FR __________ secs Pass _____ Fail _____
< 100 PSI, +/- 20 __________ PSI Pass _____ Fail _____
13. A-Side Pressure Limit < 200 PSI, +/- 30 __________ PSI Pass _____ Fail _____
< 325 PSI, +/- 40 __________ PSI Pass _____ Fail _____
< 100 PSI, +/- 20 __________ PSI Pass _____ Fail _____
14. B-Side Pressure Limit < 200 PSI, +/- 30 __________ PSI Pass _____ Fail _____
< 325 PSI, +/- 40 __________ PSI Pass _____ Fail _____
1. Parts Installed:
2. Work Performed:
3. Next Year: