AB Scan
AB Scan
Feature:
Retiwave-1000 is digital ultrasonic system for ophthalmology. With advanced intelligent digital software the parameters
of freezed and stored images could be adjusted voluntarily.Scan A measures data for every part of the eyeball as anterior
chamber depth, lens thickness,axial length and so on which are needed in ophthalmic surgeries, and calculates IOL by
axial length. Scan B displays profile images of the eyeball clearly and directly. Scanning anatomical formsand nidi inside
the eyeball, doctors can diagnose accurately for examination of cataract, vitreousbody disease, ocular trauma,
detachment of retina or choroid, macula disease, and intraoculartumor, etc..
Feature:
Technical Data
A Probe frequency 10MHz
A Probe Measuring model Contact
A Probe Measuring range Over 5mm~40mm
A Probe Measuring error ≤0.05mm
A Probe Measuring type Length of optic axis, automatic calculation, result analysis
A Probe Image freezing Image freezing auto, auto+ preservable, manual
Intraocular lens formula SRK-T, SRK II, HOLLADAY,BINKHORST-IIHOFFER-Q, HAIGISI
Gain 20~110dB Technical Data
Time gain adjustment 0~30dB A Scan B Scan
B Probe Frequency 10Mhz ACD, Lens, Vitreous, Axial length measurement High resolution and contrast B scan
B Probing Depth ≥50mm IOL Formulas: SRK-II, SRK-T, BINKHORST, 8 kinds pseudo-color
Axial Resolution ≤ 0.2mm HOLLADAY,HOFFER-Q,HAIGIS
Lateral Resolution ≤0.3mm Direct contact/immersion measurement Multiple electronic measurements: Distance, Area,
Gain Adjustment 20-110db Perimeter, Volume
Time Gain Adjustment 0-30db Support Normal, Cataract, Aphakic and Special eye 8 kinds TGC adjustable
Operation Mode Eyecup, gel, aqueous capsule model
Post Processing Measuring Tools Area, Length, Mark and Annotation
Reviews 100 images available
Feature:
SAB-500 A/B Scan with normal, vitreous body enhancement, retina observation mode, mainly used for diagnosis of
intraocular diseases, display the location, shape range of the focus of infection and the relationship with the surrounding
tissue. Can be diagnosed vitreous opacity, retinal detachment, eye base tumors etc. eye diseases. A scan is used to
measure anterior chamber depth, lens thickness, axial length, calculate diopter of implant IOL as well.
Technical Data
A Scan B Scan
Frequency: 10MHz, with LED Frequency 10MHz/20MHz(optional) , magnetic
driven, noiseless
Depth 40mm Scanning Mode Sector Scanning
Precision ±0.05 mm Magnify Multicontinuous magnification,
real-time magnification
Measurement Anterior chamber depth, lens Resolution Lateral ≤0.3mm; Vertical ≤0.2mm
thickness, vitreous body length,
total length and average
Eye mode Phakic / Aphakic / Dense / Geometry position precision Lateral ≤10%; Vertical ≤5%
Various IOL
IOL Formula SRK-II, SRK-T, HOFFER-Q, Depth 60mm
HOLLADAY, BINKHORST-II,
HAIGIS
Stat. Average and standard deviation Enhance the part of vitreous body and retina
Calculation
Store 10 Scanning results for each Gain of probe 30dB-105dB
eye
Display Mode B, B+B, B+A, A Scanning Angle 53°
Hint preset keyword Gray Scale 256
Case Search Multi-keywords False Color Multi colors. OCT
Screen 15 inch LCD Measurement type Multigroup distances, perimeters and
areas
Built-in battery can use 4 hours Image post processing Multiple curves processing,
pseudo-color processing curve
User-defined report template poor credit secured Movies 100 images movie review, AVI JPG
loans format image output
Feature: Feature:
Technical Data
B Scanner
Probe Frequency 10MHz
Scanning Mode Mechanical sector scanning
Gray Scale 256 Levels
Zoom √
Total Gain 20- 105db
Measuring Depth ≥60mm
Scanning Angle 53°
TGC TGC 0-30db
Memory 20 Images Technical Data
Probe Frequency 10MHz Specification Clinical Accuracy: ±0.1mm, Electrical Accuracy: 0.0375mm, IOL Calculation in 0.5D Increments
Measuring Range 0-42mm
IOL Calculation Refractive: SRK-II, SRK-T, Binkhorst- II, Holladay, Hoffer-Q, Haigis, Post-refractive: Latkany
Biometry Accuracy 0.05mm
Formulas Myopic, Latkany Hyperopic, Double-K/SRK-T
EYE Mode 6 Different measuring mode, 3 Custom measuring mode
A-Scan Probe Hand-held, Immersion or Slit Lamp Mounted Applicable
LOL Formula 6 Calculation formulas with contrast calculation
B-Scan
Statistical Calculation Standard deviation, variance
Scan Modes B Mode, B+A Mode, B+B Mode
P Scanner
Probe Frequency 20MHz 1. Adjustable Zoom, Gain. 2. Variable Gain Control. 3. 256 Levels Grey Scale,
Features
Biometry Accuracy ≤ 0.01mm 4. Capture of Frames and Cline Loops Available. 5. Clinical Resolution: 0.2mm
Measuring Depth 0.3-1.2mm Probe Transducer Frequency: 12.5MHz, 53° Sector Scanning Method
Output Mode Vedio 20MHz FD 13mm for Anterior segment
Computer Parameter Special imbedded system, 12.1" display screen, touching type operation Optional 20MHz FD 25mm for Posterior segment
Printer Available Sony copy machine, normal Accuracy: 0.1mm
Feature:
Feature:
10.4’’ adjustable LED screen
Capable of capturing both still images and video clips, provides excellent image resolution with a full set of
Press one key could transport current image to PC workstation
features
Equipped with keyboard floodlight, used for dark room
Easy to use measurement calipers for distance and angle measurements
100 images permanent storage capacity.
Frame-by-frame review of video clips provided for selection
Standard workstation software, easily management and print data
of optimal image
Patient database and image library is provided with printing and download capabilities
Technical Data
A Scan B Scan
Probe Frequency 10MHZ B-scan mode Mechanic Sector scans
Method contact, immersion (optional) Scan angle 53°
Precision 0.05mm Detect Depth 50mm
Measurement ACD depth, lens, axial length, Probe frequency 10 MHZ Error±0.5 MHZ
parameters and its average
IOL Calculation SRK/II, SRK/T, Holladay, Resolution Vertical<0.5mm, Lateral<1.0mm
SCDK, Hoffer-Q, Haggis
Operation Automatic, Manual Gray Scale 256
Technical Data
Information storage Build-in Case data Display Mode B; B+B; B+A
A Scan B Scan
Printer Inner Thermal Printer Images reports External high definition ink-jet printer
ACD, Lens, Vitreous, Axial length measurement High resolution and contrast B scan
Display Monitor High Speed LCD Image Post Brightness, Contrast, Smooth, Sharp,
IOL Formulas: SRK-II, SRK-T, BINKHORST, 8 kinds pseudo-color
Processing stretch, Equalize, etc.
HOLLADAY,HOFFER-Q,HAIGIS
Report Data Patient’s name, ID, Right/left, Ultrasonic videoing Ultrasonic image dynamic playback
Direct contact/immersion measurement Multiple electronic measurements: Distance, Area,
Doctor's name, Date, IOL grey scale reaches
Perimeter, Volume
table. 256
Support Normal, Cataract, Aphakic and Special eye 8 kinds TGC adjustable
model
Feature:
Probe frequency 10MHz Probe 10MHz with fixation red light Scanning angle 53°
Resolution Axial ≤ 0.2mm, lateral≤ 0.4mm Biometry accuracy ≤±0.06mm Gain 0~98 dB adjustable
Zoom 34mm~60mm, 6 steps selectable Measuring range 15~39mm Gray scale 256 levels
Various area measurement Save up to 50 groups of results and 4 groups of IOL constants
Cineloop for 5.6sec Editable clinical report with A-scan/IOL results, B-scan images,
Accurate axial measurement under B-mode (SMI-5) and textual comments with configurable entries.
Reports in PDF format for sharing and print-out.
Manual correction of axial measurement under A-mode (SMI-5)
Compatibility with graph/text printer.
Multi-groups of calipers for distance measurement
Diverse connections via HDMI and USB 2.0 ports.
Measuring mode: contact and immersion
HDMI interface for double-screen showcase.
8 groups average with standard deviation (S.D.)
DICOM 3.0 protocol available via USB 2.0 port.
Save up to 50 groups of IOL calculation results
Massive storage capacity for over 20,000 exams.
Technical Data B-Scan Enhanced Visibility and User-friendliness
B-Scan: A-Scan: Image/video buffer slots for immediate click- and- -check,review and comparison.
Probe Images/videos captured in real time and saved in lossless format ( .bmp & avi ).
10MHz Probe 10MHz with fixation red light
frequency Zooming and scan depth adjustment for better ocular disease observation.
Scanning Vitreous- Enhanced Mode improves visibility at vitreous body.
53° Resolution 0.01mm
angle A-Scan & IOL Calculation Enhancements and Tight Integration
Biometry Biometry buffer slots for immediate click-and-check,review and comparison.
Resolution axial≤0.2mm, lateral≤0.4mm ≤±0.06mm
accuracy Improved accuracy and reliability with Average and Standard Deviation for
Total gain ≥98dB adjustable Gain ≥98dB with an adjustable range of 0~59dB
up to10 scans per exam.
34mm ~ 60mm, 6 steps Measuring
Zoom 15~39mm 6 popular IOL formulae and 5 Post-refractive IoL formulae available.
selectable range Full Reporting Capabilities
-30dB~0dB dynamic, 6 points Measuring anterior chamber depth, lens thickness, vitreous length
TGC Full report integrating A-scan/IOL results.
manually adjustable parameters and axial length B-scan images and doctor s comments.
5 different With customized dictionary of symptom entries.
Variable delay automatic (for normal, dense cataract, aphakic and
0~15mm measuring
depth pseudophakic); manual Comprehensive Data Archiving and Management
modes Simple UI allows users to load, search, print, etc..
4 groups of curves (linear, Unlimited storage capacity for over 20,000 exams(8 lossless images per exam).
Post IOL power SRK/T, SRK-II, BINKHOST-II, HOLLADAY, HOFFER-Q
logarithmic, exponential,
processing calculation and HAIGIS
s-curve)
Gray scale 256 levels 8 groups average with standard deviation (S.D)
Dynamic replay for 5.6 sec, 56 images Save up to 50 groups of IOL calculation results and 4 groups of IOL
displayed continually or separately constants
Multi-groups of electronic calipers for distance measurement
Various area measurement