Chemistry Analyzer

Technical Specifications

System function Reaction System


Fully automated, discrete, random access, Reaction rotor: Rotating tray, 124 cuvettes with automatic
STAT, urine and homogeneous immunoassays;STAT sample priority washing
Throughput: 600 photometric tests/hour, up to 770 tests/hour for Cuvette: Optical length 5mm
ISE Reaction volume: 100~300µl
Measuring principles: Absorbance Photometry, Turbidimetry Operating temperature: 37°C
Methodology: End-point, Fixed-time, Kinetic, optional ISE Mixing system: 2 independent mixers
Single/Dual/Triple/Quadruple reagent chemistries,
Monochromatic/Bichromatic Optical System
Programming: User defined profiles and calculation Light Source: Halogen-tungsten lamp
Photometer: Reversed optics, grating photometry
Sample Handling Wavelength: 340nm, 380nm, 412nm, 450nm, 505nm, 546nm,
Sample tray: 90 positions for primary or secondary tubes and 570nm, 605nm, 660nm, 700nm, 740nm, 800nm
sample cups Absorbance range: 0~3.3Abs (10mm conversion)
Sample volume: 1.5~45 µl, step by 0.1µl Resolution: 0.0001Abs
Sample probe: Liquid level detection, clot detection and collision
protection Control and Calibration
Probe cleaning: Interior and exterior automatic probe washing Calibration mode: Linear (one-point, two-point and
carry-over < 0.05% multi-point),Logit
Automatic sample dilution, Pre-dilution and post-dilution -Log 4P, Logit-Log 5P, Spline, exponential, Polynomial,Parabola
Control rules: Westgard multi-rule, Levy-Jennings, Cumulative
Internal bar code reader (optional) sum check, Twin plot
Sample/Reagent barcode reading – applicable to various bar code
systems including Codabar, ITF (Interleaved Two of Five), code128, Operation Unit
code39, Operation system: Window XP Professional or Windows 7
UPC/EAN, Code93; Bi-directional LIS Interface transmission Professional or Windows 8,
Interface: RS-232, Network Port, USB/ parallel port
ISE Module (optional)
Optional selection of K+, Na+, Cl Throughput: Up to255 tests per
hour Working Condition BS-600
Reagent Handling
Power Supply: 220V,50/60Hz;≤1700VA
Temperature: 15~30°C Chemistry Analyzer
Reagent tray: 80 positions in refrigerated compartment (2~10°C) Humidity: 35~85%
Reagent volume: 10~200µl,step by 0.5µL Water consumption: ≤28L/H, De-ionized water
Reagent probe: Liquid level detection, collision protection and Dimension: 1190mm(W)*1155mm(H)*720mm(D)
inventory check, reagent bubble detection Weight: 300 Kg
Probe cleaning: Interior and exterior automatic probe washing

Mindray Building, Keji 12th Road South,


2015 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved. Specifications subject to changes without prior notice.
High-tech Industrial Park, Nanshan, Shenzhen 518057, P.R. China
P/N: ENG-BS-600-210285x8-20150514
Tel: +86 755 8188 8998 Fax: +86 755 26582680
E-mail: [email protected] www.mindray.com
Mindray is listed on the NYSE under the symbol”MR”
BS-600
Chemistry Analyzer

Easy maintenance
All containers & maintenance kits are located in the
front of the analyzer
Easily accessible for part replacement, routine
maintenance or troubleshooting

Reagent bubble detection


Accurate
Able to detect bubbles in reagent bottles, as well as
Complete metrological
traceability detect real liquid level before reagent aspiration
Highly precision sampling
Reagent bubble detection

Whole blood HbA1C testing*


No need pretreatment of the sample
Proven precision and specificity

Economical No interference from hemoglobin variants


Traceable to IFCC/NGSP reference methods
Light-spot flatting optical
system- lower the reaction
volume to 100µL
Minimum sample
volume:1.5µL
Cost-effective
Sampling precision up to 1.5μL
Perfect match between instrumentation and
dedicated reagents
100μL minimum reaction volume ensures
more effective cost per test
Intelligent
Whole blood HbA1C testing*
Advanced software Lower carry over
Easy maintenance
High pressure washing for interior and water fall
washing for exterior
Carry over < 0.05%

* in development
BS-600 Advanced software Mindray solution for clinical chemistry
Chemistry Analyzer After more than 10 years of research and development on reagents, Mindray can now provide 61
parameters of dedicated reagents, covering hepatic, renal, cardiac, lipids, diabetes, pancreatitis,
inorganic ions and immunoassays, etc., together with original calibrators
User-friendly interface with metrological traceability as well as controls for BS-600 chemistry analyzer.
Touch screen
Share the same platform with BS-2000 series and BS-800 series
Real-time status monitoring between analytical unit and carousels

Auto Chemistry
Analyzer
Original Quality
Control
Convenient R/S continuous loading and offloading
Remaining time counter makes it easy for operator to follow
the instructions Mindray
solution for
clinical chemistry

Intelligent reagent management


Real-time indicating the test number, inventory and expiry date
Improve work efficiency

Original System
Pack Reagent Original
Reflex function Calibrators
Pre-defined reflexive assays will be performed automatically when with Traceability
preset criteria is met
Each assay may involve multiple reflexive criteria
Each criteria may initiate up to a maximum of 20 relevant assays
Standard reference system
Adopt JCTLM reference system
IFCC primary method for enzyme, ID/MS method for substrate
NIST, IRMM reference materials
Test Summary
Test summary report calibration, QC, sample, validation test and
JCTLM, Joint Committee On Traceability In Laboratory Medicine
rerun tests can be generated
NIST, National Institute of Standards and Technology,USA
Facilitate computation of total test costs
Error Log Export function -facilitate error report to engineers IRMM, Institute for Reference Materials and Measurements,EU
Results Archive can be transferred to engineers for evaluation IFCC, International Federation of Clinical Chemistry and Laboratory Medicine
Complete traceability process External quality assurance for Reagent menu
Complete calibration hierarchy and traceability chain based reference measurement
on ISO standard (EN/ISO17511) from reference system to Mindray participates RELA ( External quality control for reference laboratory ) Hepatic Panel Cardiac panel
routine measurement system and CAP ( College of American Pathologists external quality control ) Alanine Aminotransferase (ALT) Creatine Kinase (CK)
Aspartate Aminotransferase (AST) Creatine Kinase-MB (CK-MB)
Alkaline Phosphatase (ALP) Lactate Dehydrogenase (LDH)
Traceability chain of Mindray measurement system (Glu) EQA for Mindray Reference laboratory—— Rela γ-GlutamylTransferase (γ-GT) α-Hydroxybutyrate Dehydrogenase(α-HBDH)
Traceability Material Calibration / Procedure Implementation Uncertainty Mindray reference laboratory has passed Rela for 6 continuous years Direct Bilirubin (D-Bil) DSA Method High sensitive C-reaction protein( HS-CRP)
Value Assignment Uc(y)
Direct Bilirubin (D-Bil)VOX Method Homocysteine (HCY)

SI unit,mmol/L
Total Bilirubin (T-Bil) DSA Method Myoglobin(MYO)
Total Bilirubin (T-Bil)VOX Method D-Dimer(D-Dimer)
ALT GGT
B B Total Protein (TP)
Reference material Primary Reference 3.4

SRM 917C measurement procedure 2.8 Albumin (ALB) Inorganic & Anemia
3.2

Total Bile Acids (TBA) Iron (Fe)


3 2.6

Mindray Master CDC Hexokinase Prealbumin (PA) Ferritin (FER)


Calibrator reference method 2.8
2.4
Cholinesterase (CHE) Transferrin (TRF)
2.6

2.2
α-L-fucosidase (AFU) Calcium (Ca)
2.4
Mindray Product Mindray standard
Calibrator Measurement system 1.5 1.6 1.7 1.8 1.9 2 1.68 1.76 1.64 1.92 2 2.08 5'-nucleotidase (5'-NT) Magnesium (Mg)
A [umol/l] A [mmol/l]
limits of equivalence = +5,25% limits of equivalence = +5,25%
Phosphate Inorganic (P)

End user’ Routine


Renal Panel Unsaturated iron binding capacity (UIBC)
Routine Sample More Rela results please refer to: www.dgkl-rfb.de/81
Mindray Urea (UREA) Glucose-6-phosphate dehydrogenase (G6PD)
Rela
All the items Mindray participate Rela

ALT AMY ALP CK GGT GLU LDH TB TP UA UREA


Result Creatinine (CREA)Sarcosine Oxidase Method Lipid Panel
Uric Acid (UA) Total Cholesterol (TC)
Carbon dioxide (CO2) Triglycerides (TG)
Microalbumin HDL-Cholesterol (HDL-C)
β2-Microglobulin (β2-MG) LDL-Cholesterol (LDL-C)
EQA for Mindray Testing System—— CAP
Cystatin C (CysC) Apolipoprotein A1 (ApoA1)
Mindray testing system has passed CAP for 6 continuous years
Retinol binding protein( RBP) Apolipoprotein B (ApoB)
Lipoprotein(a) [Lp(a)]
CAP Number: 7198395-01 Kit# 1 Kit ID: 25733824 Immune Panel
Institution: Shenzhen Mindray Biomed Elec Co Ltd Kit Mailed: 6/3/2013
Attention: Lixing Liu MD Original Evaluation: 7/8/2013 Immunoglobulin A (IgA) Rheumatism Panel
City / State: Hongkong HK CH 518055
EVALUATION Immunoglobulin G (IgG) C-reactive protein (CRP)
ORIGINAL
C-B 2013 Chemistry
Immunoglobulin M (IgM) Rheumatoid Factor (RF)
CAP

CAP #: 7198395 Subspecialty : Routine Chemistry Immunoglobulin E (IgE) Antibodies Against Streptolysin O (ASO)
Proficiency Event Proficiency Event Proficiency Event
2012 3 2013 1 2013 2 Current Event Cumulative CLIA Complement C3 (C3)
Performance '88 Performance
Regulated Analyte Test Event Score % Test Event Score % Test Event Score % Interpretation Interpretation
ALT C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
Complement C4 (C4) Lung Panel
Albumin C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
Alkaline Phosphatase C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
Adenosine Deaminase (ADA)
Amylase C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
AST C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful Diabetes Panel Angiotensin Converting Enzyme(ACE)
Bilirubin, Total C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
Calcium, Total C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful Glucose (Glu) GOD-POD Method
Chloride C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
Cholesterol, Total C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful Glucose (Glu) HK Method Pancreatitis Panel
Cholesterol, HDL C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
Creatine Kinase C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful Hemoglobin A1c (HbA1c) α-Amylase (α-AMY)
Creatinine C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
Glucose C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful Fructosamine (FUN) Lipase (LIP)
Iron, Total C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
LD C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful β-Hydroxybutyrate(β-HB)
Magnesium C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
Potassium C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
Sodium C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
Protein, Total C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
Triglycerides C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
Urea Nitrogen C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful
Uric Acid C-C 5 /5 100 C-A 5 /5 100 C-B 5 /5 100 Satisfactory Successful