Abbott i1000SR - PM Checklist

Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

ISA 117-022 i1000SR Preventive Maintenance and Total Call Procedure

i1000SR Preventative Maintenance Procedure

Customer Name Serial Number


Customer Number Ticket Number
City FSR Name
Telephone Date

Pre-site Investigative Procedures Completed


Interview customer for any on-going issues and order parts as necessary.
Note: If the instrument is experiencing ongoing assay imprecision or accuracy issues, perform
the ISA 117-019 (latest revision) Assay Troubleshooting i1000SR prior to completing the PM.
Ensure PM parts are available per the parts list at the end of this procedure.
Verify Abbott Link connection status & review message history log for issues.
On-site Investigative Procedures Completed
Verify TSB status of instrument.
Verify customer has performed a recent software back-up and current system back-up CD is
available. Perform system backup, if needed (M&D 6004)
Review Maintenance & Message History Logs and advise customer on performance of customer
maintenance procedures and inventory of customer maintenance parts.

The removal and replacement instructions for L/N parts are available on the System Control
Center (SCC) in the online Operations Manual.

Note: Follow local Service Organization processes if it is necessary to charge customer for LN
parts and/or labor.
SCC / Module Shutdown Completed
Perform unload of reagent kits from reagent carousel. Store in refrigerated area.
Check reagent cover LEDs on reagent distribution board to verify reagent cover battery
condition.
 Blinking ALIVE (upper) & STATUS (lower) LEDs indicate battery needs to be replaced
 NOTE: Replacement batteries (type BR2032) are not stocked by Abbott & should be
sourced locally
Inspect doors, covers, & hinges for misalignment, squeaks, etc.
Power off SCC & Processing Module P-248 Startup and Shutdown / P-248 Startup and
Shutdown

Lower Fluidics Area Completed


Inspect area for leaks & dry buffer build up. Clean as needed.
Verify all tubing connections are tight & secure.
Replace vacuum filter, as needed(R&R K2.06 Vacuum Pump Filter) /
(R&R K2.06 Vacuum Pump Filter Assembly)

1
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times. **
ISA 117-022 i1000SR Preventive Maintenance and Total Call Procedure

Replace reagent cooler condensate inline filter, as needed.


Replace waste pump head, as needed. (R&R K2.08 Waste Pump Head) / (R&R K2.08 Waste
Pump Head)
Replace in-line waste pump to pressure switch tubing check valve, as needed.
Inspect & secure cable connections on fluidics distribution board & sensors, as needed.
Perform MD 3181 Vacuum Diagnostics to check system health, if check fails:
 Check fittings on vacuum accumulator cap for cracks & kinked tubing.
 Flush out vacuum accumulator bottle if build-up is present. Ensure cap is tight

Bulk Solution / Waste Area Completed


Inspect area for leaks & dried buffer buildup; clean as needed.
Clean area of RVs, dust, etc.
Verify tubing connections are tight.
Verify waste platform movement is not obstructed.
Inspect Pre-Trig / Trig level sensor fittings for cracks and/or kinked tubing.
Clean out Pre-Trigger / Trigger Tray

2
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times. **
ISA 117-022 i1000SR Preventive Maintenance and Total Call Procedure

Lower Fluidics Area Completed


Inspect external waste pump (if present) tubing & connections. Power off external waste pump
and clean as needed.

Process Path Completed


Inspect & clean the Wash zone & PT / T manifold assemblies.
 Perform MD2057 WZ Pressure Test and replace any WZ valves that fail.
 Utilize ISA 117-007(latest revision) for cleaning tips & techniques
 Verify date codes for Trigger / Pretrigger manifold valves and replace if > 2years
 Write installation date on replacement Trigger / Pre-Trigger Valves for future 2 yr
date verification
Inspect & clean the Process Path cover.
 Clean all buffer build-up & ensure lane rotators do not bind
Remove, inspect, & clean (2) Vortexers
 Verify Vortexer cups move freely & cable connections are secure
Inspect for Trigger leakage around CMIA Reader & Shutter solenoid, clean as needed.
 If optics must be removed for cleaning, ensure the maintenance cap is applied to
the light pipe
Verify shutter solenoid movement is not obstructed
Clean & inspect process path disk, track, and home sensor
 Inspect disc for grooves, etching, & wear
 Common process path problem areas: wash zones, vortexers, diverters, & optics
Verify RV Present switch arm moves freely.
Clean & inspect process path LLS antenna; verify cable connections are secure.
(R&R E2.02 LLS Antenna Process Path) / (R&R E2.02 LLS Antenna Process Path)

3
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times. **
ISA 117-022 i1000SR Preventive Maintenance and Total Call Procedure

RV Loader Assembly Completed


Remove RV Loader
Inspect RV pickup screw for tightness.
Verify RV pickup movement is not obstructed or contacts the RV Picked sensor arm when
rotated. It is OK if the pick-up arm slightly touches the upper chute.
Clean all RV Loader sensors – Hopper (proximity sensor), RV Picked, & RV Loader Home
sensors.
Check pressure monitor connections for leakage and tighten as needed.
 Use absorbent material placed against fittings to check for wetness.
Reagent Carousel Area Completed
Replace v-wheels & shoulder screws, as needed.
(R&R D1.09 V-Wheel Lever and V-Wheels) / (R&R D1.09 V-Wheel Lever and V-Wheels)
Use a small amount of Green Grease / Aqua Lube on the shoulder screw threads
Clean & lightly grease hole in V-Wheel Adjust Lever with Super Lube Grease
Utilize Green Grease / Aqua Lube on new shoulder screw for lever
Use small amount of Super Lube Grease at interface between shoulder screw & lever
Clean dust & dried liquid from reagent cooling chamber.
Note: If the lab environment is dusty, refer to PM Addendum for additional cleaning instructions
Clean reagent carousel & inspect each position for proper slide / spring function.
Clean & inspect reagent antenna board
Inspect reagent carousel motor gear for wear
Clean all sensors in reagent cooling chamber
Inspect cooling chamber gaskets for tears & wear; replace as needed.
Inspect & clean cooler condensation collection ring.
Re-install reagent carousel. Before securing the motor latch verify the carousel rotates freely on
the v-wheels.

4
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times. **
ISA 117-022 i1000SR Preventive Maintenance and Total Call Procedure

Wash Cup / Upper Waste Manifold Completed


Wash Cup assembly
 Remove & clean wash cup baffle & check for leakage.
 Verify cable connections are secure.
Upper Waste Manifold
 Inspect thermistor tubing & connections.

5
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times. **
ISA 117-022 i1000SR Preventive Maintenance and Total Call Procedure

Pumps & Syringe Completed


Pumps
 Upper Pumps / Syringe
o Inspect Trigger / Pre-trigger pumps & Sample Syringe for leakage & check all
connections.
 Lower Pumps (located behind Trigger / Pre-trigger sliding tray).
o Inspect Buffer & Transfer pumps for leakage & check all connections.

Verify Instrument Performance Completed


Review PM checklist with customer.
Ensure appropriate Quality Control meets specification and calibrate as necessary.

6
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times. **
ISA 117-022 i1000SR Preventive Maintenance and Total Call Procedure

Information for Level 2 Follow Up

Comments: ___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________

Field Engineer Signature: _____________________________________

Customer Signature: _______________________________________________

Field Service PM Parts List


Part # Description Qty Criteria
Evaluate if kit is needed or if there
7-205087-04 PM Kit, i1000sr (RoHS) 1 will be cost savings by using
individual parts.
7-100302-01 Reagent Cooler In-line Filter 1 Contained in PM kit
7-200055-01 Reagent Bottle Rotator Wheel 1 Contained in PM kit
7-200370-02 Vacuum Filter Assembly 1 Contained in PM kit
7-200378-01 Pump Head, Liquid Waste 1 Contained in PM kit
7-202660-01 Tubing, Waste Pump to Pressure Switch 1 Contained in PM kit
Pinion Gear kit i1000sr (used with reagent
7-202411-04 carousel motor with replaceable gear 7-200635- 1 Replace if worn
00)
EOD

7
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times. **