Scanner dimensions:
125 cm125
(49")
cm (49")
64 cm (25")
64 cm (25")
287 cm287
(113")
cm (113") 131 cm131
(52")
cm (52")
3.2 m exam room with the included • Monitor 24" SVGA (minimum resolution 1920 x 1080 32-bit color)
workstation. No operator shielding • Archive 500 GB USB hard drive
or special site preparation beyond • Adobe® Reader® DC
a dedicated 100-127/200-240 VAC • Internet Explorer® 11
3.35 m (11') duplex outlet is usually required.2 • Two 10/100 Mbit Ethernet ports
The outlet should be placed near • Windows®-compatible printer
the desired location of the
operator’s console. Environmental specifications:
Power . . . . . . . . . . . . . . . . . . 100-127 VAC 50/60Hz 20A dedicated circuit
Scanner table specifications: 200-240 VAC 50/60Hz 10A dedicated circuit
Scanner size . . . . . . . . . . . . . . 2.87 m x 1.31 m x 1.25 m (113" x 52" x 49") Consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . idling 40VA, scanning 525VA
Scanner weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 353 kg (778 lbs) Distortion . . . . . . . . . . . . . . . . . . . sinusodal waveform, less than 5% THD
Patient table top height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 cm (25") Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20%-80% non-condensing
Maximum patient weight supported . . . . . . . . . . . . . . . . 204 kg (450 Ibs) Room temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . 18°C-27°C (65°F-81°F)
Drive system . . . . . . . . . . . . . . stepper motor with reinforced drive belts Scanner heat output . . . . . . . . idling 150 BTU/hr, scanning 1800 BTU/hr
Active scan area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198 cm x 66 cm Console heat output . . . . . . . . . . . approx. 200 BTU/hr with 24" monitor
Start position indicator . . . . . . . . cross laser light (class II, <1 mW power) Ventilation . . . . . . . . . . . . . . . . . all cooling vents must remain unblocked
Pad . . . . . . . . . . . . . washable patient mat, includes paper roll dispenser Dust, fumes, debris . . . . . . . . . . . install system in clean, ventilated area
Attenuation of patient support table . . . . . . . . . . . . . . . . . . . . <1.2 mm AL
Communication cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Ethernet
Scanner leakage current . . . . . . . . . meets IEC 60601-1 safety standard
Detector specifications:
Detector . . . . . . . . . . . . . . . . . . . . . . high-definition, direct-digital detector
3. Additional hardware may be required for fax capabilities. 4. German speaking countries only. 5. Bone-Lean-Tissue Color Coding,
Metabolic Results (ICW, ECW, TBW), Resting Metabolic Rate, Composer Style Sheets – Body Sports Medicine Segmental, Body Patient
Weight Loss. 6. Consult for market availability.