ReferenceGuide E
ReferenceGuide E
Product Overview
(CR-IR 391CL)
Confirming by
Viewing Indicators
Setting
Study Utilities
Configuring
Client Utilities
Customizing
Main Window
Setting
User Utilities
Troubleshooting
Daily Inspection
and Maintenance
Appendix
Maintenance and
Inspection
Index
This Reference Guide describes details on how to operate the FCR PRIMA
Console (CR-IR 391CL) and cautions to be observed when operating it.
Please read the Reference Guide thoroughly before actually operating the
FCR PRIMA Console (CR-IR 391CL) system. After reading this guide, store
it nearby the FCR PRIMA Console (CR-IR 391CL) so that you can see it
whenever necessary.
897N101440
ii FCR PRIMA Console Reference Guide 897N101440
Introduction
This Reference Guide applies to the following software.
• FCR PRIMA Console Software V1.1(B)
FCR PRIMA Console (CR-IR 391CL) is a Fuji Computed Radiography image processing unit.
This Reference Guide offers the reader a summary insight into the FCR PRIMA Console (CR-IR
391CL) and proceeds to present information prerequisite to using the product, including operating
instructions and precautions, and tips on daily inspections and care.
FCR PRIMA Console is a CR workstation intended to provide diagnostic quality images and
measuring function to aid the doctor in diagnosis, and also is a viewer for US referral images.
CAUTION
1 No part or all of this guide may be reproduced in any form without prior permission.
2 The information contained in this guide may be subject to change without prior notice.
3 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
installation, relocation, remodeling, maintenance, and repair performed by other than
dealers specified by FUJIFILM Corporation.
4 FUJIFILM Corporation shall not be liable for malfunctions and damages of FUJIFILM Corporation
products due to products of other manufacturers not supplied by FUJIFILM Corporation.
5 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
remodeling, maintenance, and repair using repair parts other than those specified by
FUJIFILM Corporation.
6 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
negligence of precautions and operating methods contained in this guide.
7 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from use
under environment conditions outside the range of using conditions for this product such
as power supply, installation environment, etc. contained in this guide.
8 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
natural disasters such as fires, earthquakes, floods, lightning, etc.
Caution: Rx Only in the United States (Federal law restricts this device to sale by or on the
order of a physician)
This system is classified as a medical device under EC Directive 93/42/EEC.
This equipment has been designed on the assumption that the patient would not come into direct
contact with it or for operation by appropriately trained operator.
Process waste correctly, as stipulated by local law or any regulations that apply.
When discarding the CR-IR 391CL that incorporates the lithium battery, be sure to contact a
licensed waste disposal contractor because it cannot be disposed of as a general waste.
Trademark
FCR is a trademark or a registered trademark of FUJIFILM Corporation.
Product Overview
Chapter 2 Introduces the features of the FCR PRIMA Console, the names and functions of its components and
basic operational sequences, such as starting up and shutting down the product.
Starting a Study
Chapter 3 Explains a sequence of operational steps for starting a study, from entering patient
information to exiting the X-ray study image reading process, and how to use a main window.
Troubleshooting
Chapter 10 Describes what actions should be taken to respond to faults as they occur in the FCR PRIMA
Console and defines the error messages that may be displayed on occurrence of faults.
Chapter 10 Troubleshooting
10.1 Troubleshooting����������������������������������������������������������������������������������������������������� 10-1
10.2 PC Does Not Start������������������������������������������������������������������������������������������������� 10-2
10.3 PC is Inoperative��������������������������������������������������������������������������������������������������� 10-3
Safety Information
Before using this equipment, please read “Precautions Before Operating This Equipment”
carefully so that you can operate it correctly.
Whenever you operate this equipment, be sure to observe those precautions. Failure to do so
may cause you to subject to injuries or property damage to occur.
The institution where the equipment is installed is responsible for its use and
maintenance.
In addition, this equipment should not be used by persons other than doctors
or suitably trained staff.
Safety Information
9 Other
Abide by instructions in the Operation Manual for proper equipment handling.
1 Preventing Explosion
WARNING
Safety Information
Because this equipment is not explosion-proof, do not use combustible and explosive gases
near the equipment.
WARNING
To prevent electrical shock, observe these precautions by all means.
● Never open the equipment covers. Hands touching a high-voltage zone inside can cause
electrical shock.
● Install the equipment where it will not be splashed with water.
● Check that the equipment is fully grounded.
● Check that all cords are connected correctly and safely.
● While this equipment is in operation, pay due attention not to have the patient come into
direct contact with it.
Connection Instruction
CAUTION
Use external devices of specified types to connect to the equipment.
This epuipment complies with the following electromagnetic compatibility (EMC) standard by comforting to
the requirements of IEC 60601-1-2:2001+A1:2004/EN 60601-1-2:2001+A1:2006.
EMC Standard:
CISPR 22:2005/EN 55022:2006
CISPR 24:1997+A1:2001+A2:2002/EN 55024:1998+A1:2001+A2:2003
1
IEC 61000-3-2:2005/EN 61000-3-2:2006
IEC 61000-3-3:1994+A1:2001+A2:2005
Safety Information
However, there is no guarantee that harmful electromagnetic interference will not occur in any installation
environment.
This equipment generates, uses and can radiate radio frequency energy. In a particular environment,
e.g., the equipment is not installed and used in accordance with the instructions or any device that is not
compliant with the EMC Standard is used in a surrounding area, harmful interference may result, causing
the equipment or surrounding device to malfunction. If this equipment causes harmful interference to other
equipment or receives harmful interference from other equipment, you are encouraged to try to correct the
interference by one or more of the following measures:
If the problem persists, discontinue the use of the equipment and consult the manufacturer of the affected
equipment or our official dealer.
WARNING
Keep the equipment away from sources of strong electromagnetic radiation.
CISPR22
Electric field noise strength Class B
Safety Information
CISPR22
Harmonic emissions Class D
EN61000-3-2
IEC61000-3-2
Voltage fluctuations/ Complies
flicker emissions
EN61000-3-3
IEC61000-3-3
<5% UT <5% UT
(>95% dip in UT) (>95% dip in UT)
for 250 cycles for 250 cycles
Power frequency (50/60Hz) 1 A/m 1 A/m
magnetic field
EN61000-4-8
IEC61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Safety Information
EN61000-4-3 80 MHz to 1 GHz
IEC61000-4-3
film-based diagnostics.
• Verify the information that accompanies images and use it to aid in their reading. Pay particular attention
to patient information.
• Print and retain original images on a film (called an “original film”). The hard disk or disk mounted on the
FCR PRIMA Console (CR-IR 391CL) unit is not intended for permanent storage of diagnostic images.
• Images displayed in the “Image Viewer window” are only for viewing. Do not use these images for the
diagnosis.
• When exposing an area where it is difficult to distinguish the exposed direction and the symmetric body
part, mark it using markers for example to avoid the confusion of the exposed direction and body part.
• When reading an image, be sure to verify the image flip information (front or back) and the actual
exposure condition of patient.
• Be careful not to confuse the patient when printing several images on a film for several patients and
reading of images on the film should be carried out at the responsibility of a user who is conscious of the
intended differences from the image quality by displayed image size.
• When importing image or study data into FCR PRIMA Console (CR-IR 391CL) (Optional) or saving from
FCR PRIMA Console (CR-IR 391CL) to media such as CD, be very careful to check the image or study data
and patient information.
• Configure the image processing parameters for the FCR PRIMA Console (CR-IR 391CL) to suit display
images.
Improperly configured image processing parameters could make correct image reading difficult.
• Luminance and contrast can be changed depending on the monitor types but do not change.
If it changed, an image may not be confirmed properly.
• Use the FCR PRIMA Console (CR-IR 391CL) with proper screen contrast (high/low) settings.
Improper screen contrast (high/low) settings could make image reading difficult.
• Before reading images on FCR PRIMA Console (CR-IR 391CL), make sure that the monitor is set at the
right angle for viewing. If the monitor is not set at the right angle, adjust it. An improperly angled monitor
could make it difficult to read images properly.
• Adjust the FCR PRIMA Console (CR-IR 391CL) to display images in a proper size.
The FCR PRIMA Console (CR-IR 391CL) may subject original images to interpolation processing (display
processing) to enable you to view the images in an optional display size. As a result, images displayed in
an improper size may not impart their intrinsic information.
• Do not change the display setting (for portrait or landscape) of the LCD monitor that was set at time of
installation. If you need to change the display orientation setting, contact our authorized dealer.
• Use a screensaver when displaying the same image for extended periods of time. If your screensaver fails
to launch, consult our official dealer.
If the same image is displayed for a long time, the burn-in or residual image of the previous image may
result when switching the display on the monitor.
• Evaluation of the results of measurement performed on the FCR PRIMA Console (CR-IR 391CL) is your
responsibility.
Images used for measurement on the FCR PRIMA Console (CR-IR 391CL) may contain errors in relation to
the actual length of the subject, depending on how the images have been captured and other conditions.
• When specifying coordinates on the FCR PRIMA Console (CR-IR 391CL) monitor, specify them correctly.
Incorrectly specified coordinates will deliver incorrect measurement results.
• Do not turn off the PC while FCR PRIMA Console (CR-IR 391CL) is in use. Images saved on the PC or DVD
drive might be damaged.
• When connecting FCR PRIMA Console and other study equipment, use the UTP (Unshield Twist Pair)
cable.
1
about 10 minutes to resume. If you shut down the FCR PRIMA Console (CR-IR 391CL) or Image Reader,
the images being read might be lost.
• The FCR PRIMA Console (CR-IR 391CL) might be accidentally powered off due, for example, to a power
failure while image reading is in progress. When this happens, restart the PC and log in again to reload
Safety Information
the images. Make sure that the correct images have been read in response to your exposure menu choice.
After starting up and loggin in the FCR PRIMA Console (CR-IR 391CL), when double-clicking and starting
to expose the study which was reading images at the time of shuting down from the patient list on the
“Main window”, images can be started to read again.
If the images loaded do not match your exposure menu choice, exchange or otherwise manipulate the
images to synchronize the exposure menu choice and the images with each other.
• Operate the disk drive using FCR PRIMA Console (CR-IR 391CL) applications.
• Do not use the disks that are dedicated to image storage on the FCR PRIMA Console (CR-IR 391CL) for
any other use. Images stored on those disks would be made inaccessible for browsing.
• Use disks that meet the requirements listed in “1.5 Disk Handling”.
• Images that have been saved with “Processed” being selected can be reloaded into the FCR PRIMA
Console (CR-IR 391CL), but cannot be printed on a film (option).
• Be careful not to confuse the DVD disk with the CD-R or DVD-R disk for PDI because the CD-R or DVD-R
disk for PDI to be given to the patient is created using the DVD drive to create DVD disks.
When creating a CD-R or DVD-R disk for PDI, write information (patient name, patient ID, birth date, disk
creation date, study date, facility name) on the front of the CD-R or DVD-R disk in order to distinguish it
from DVD+RW/-RW disk. When giving the created disk for PDI to the patient, advise the patient that the
contents of the disk is for reference.
• Confirm that time and date which display on the “Indicator window” is correct before using the FCR
PRIMA Console (CR-IR 391CL).
The date and time appearing in the “Indicator window” are recorded as the date and time of image
recording on a disk.
• Do not modify the Windows setting of the FCR PRIMA Console (CR-IR 391CL). Image input/output might
be disabled.
• While using the FCR PRIMA Console (CR-IR 391CL), do not double-click any file associated with any other
application software.
If you attempt to open a file associated with any other application software by double-clicking it, the file
would take a considerable time to open. If it is absolutely necessary to open that file, launch the
associated application software first and then control the application software to open the target file.
• Do not handle the FCR PRIMA Console (CR-IR 391CL) in a Windows Explorer operation, unless otherwise
specified in this Reference Guide. If you eject a disk in a Windows Explorer operation, the data stored on
the disk could be corrupted as a consequence of the disk having been ejected while images were being
written to or read from it, or your application might not be able to eject it.
Further, when you browse through information stored on the disk, the information might take time before
it shows up.
• Before deleting images on FCR PRIMA Console (CR-IR 391CL), make sure that there is no problem if the
images are deleted. Although FCR PRIMA Console (CR-IR 391CL) allows you to delete images, if you
delete required images by mistake, you would have to make a re-exposure of the subject and recapture
the images.
• Before you use an uninterruptible power supply with the FCR PRIMA Console (CR-IR 391CL), verify its
performance characteristics, such as the battery life. Replace the battery when needed.
If the FCR PRIMA Console (CR-IR 391CL) is forced to shut down while it is in operation due to electric
power failure or other causes, data could be lost.
You need to have a security, backup or any other relevant procedure programmed separately to save
images safely. Use of an uninterruptible power supply is recommended to protect against emergencies.
• Do not connect any additional hardware to the FCR PRIMA Console (CR-IR 391CL) in addition to the
hardware already connected to it. It will result in malfunction.
1
Images stored on the FCR PRIMA Console (CR-IR 391CL) hard drive or disks could be lost as a result of
equipment or media faults. Further, some characteristics of a disk could make images stored on the disk
inaccessible for reading or writing if the disk has deteriorated over an extended period of usage, even
though it has been stored properly. An improperly stored disk could make images stored on it
Safety Information
1
Precaution in Handling Disks
Before using disks, read instructions on the disk package.
Safety Information
Precautions in Handling Files Stored on Disks
• Do not save new files to a disk on which other files are already stored.
• Do not delete, move or modify files stored on the disks.
Failure to abide by these precautions could result in data unknown to the FCR PRIMA Console (CR-IR 391CL) or
result in a failure to save data or loss of saved data.
Unformatted disks cannot be used. To use an unformatted disk with the FCR PRIMA Console (CR-IR 391CL),
format the disk in the “Indicator dialog”.
If you find any unusable disk even after it is formatted, contact our official dealer.
For instructions about how to format new disks, see “2.5.2 Starting Disk Handling”.
Ejecting Disks
Do not force the ejection of a disk by pressing the disk drive EJECT button.
If it is necessary to eject a disk, follow the instructions in “2.5.3 Ejecting Disk”.
If you use an application other than the FCR PRIMA Console to eject the disk, the PC can freeze or the disk can
become unusable.
Disk Management
Before using a new disk, you are advised to fill out the label seal with the save start date.
For disks that are not used for further writing, enter the label name (media identification number) in the label seal.
You should also enter the save end date in the label seal.
Windows Explorer
Do not attempt to access the disk drive from Windows Explorer. The drive might take a considerable time to
access.
Disk Identification
When a disk is inserted in the disk drive, it will take about 30 seconds before it is identified to the FCR PRIMA
Console.
Product Overview
The FCR PRIMA Console is a diagnostic imaging workstation that not only simplifies the workflow of
diagnosing X-ray images, but also offers a fully integrated solution to managing modality, as well as X-ray,
study images.
X-ray images captured by the image reader are transmitted to the FCR PRIMA Console (CR-IR 391CL) .
Transmitted or captured images are displayed on the FCR PRIMA Console (CR-IR 391CL) monitor screen,
which provides information to aid in reading the images.
1 Digital image enhancements make X-ray images easier to view and better suited for diagnostic
purposes.
2 The FCR PRIMA Console (CR-IR 391CL) lets you enter patient information and view images. For
X-ray studies, you can enter exposure information (such as exposed body parts and exposure
menus) , as well as patient information.
3 Image data can be backed up to DVD.
4 The FCR PRIMA Console (CR-IR 391CL) keeps images obtained from various studies under fully
integrated management.
5 Study images can be saved on CD-R or DVD-R in the PDI format and made available to the patients
for collaborative diagnostic purposes.
Image Reader
2
Product Overview
PC Imager
FCR PRIMA Console
(CR-IR 391CL)
Hub
NAS
HINT
NAS is the hard disk for the image data storing. When NAS is installed, the image data can be saved in NAS but
not in DVD.
Images Explanations
Unprocessed FCR study X-ray images created by FUJIFILM CR equipment have the FUJIFILM-unique
images image processing parameters added on the image data. Unprocessed FCR
images are images on which only the parameters were adjusted to enable image
editing such as changing gradation (contrast or density) and reversing the display
for obtaining a better view on screen or film, while keeping the original image
data as is. Therefore, the original images before editing can be restored.
These images cannot be displayed in the DICOM viewer.
Processed FCR study images X-ray images created by FUJIFILM CR equipment have the FUJIFILM-unique
image processing parameters added on the image data. The Processed FCR
images are images whose original data was processed along with the parameter
change. Therefore the original images cannot be restored.
Images can be viewed in the DICOM viewer.
Other than FCR study images X-ray images loaded from a DICOM study disk created in another facility.
Patient
2
Operation for information
main window edit button
Product Overview
Toolbar
Input and refer to
patient information Image pane
Study pane
Study operations
FCR PRIMA Console (CR-IR 391CL) Thumbnail images Study status ([Suspend] [Verify])
menu bar for operation
2
Product Overview
Product Overview
Before starting up FCR PRIMA Console, make HINT
sure that all peripheral devices connected to If the client application has not launched
the system are turned on. automatically, the user menu is not displayed.
To launch the client application, double-click the
1 Turn on the monitor. FCR Web icon on the desktop.
When the client application launches, it displays a * For information about starting up the peripheral
“User Selection window” to signify that FCR PRIMA devices connected to the FCR PRIMA Console
Console is ready to work. system, refer to the Operation Manual pertaining
to the peripheral devices.
Guidance
2
* Logging off from and restarting Windows
When the FCR PRIMA Console server If no operation window (“User Selection
application launches, it displays server window”) is displayed while the server
application icon in the task tray in the lower application icon appears in the PC’s task
right-hand corner of the screen. tray, start the client application. When the
Client Application starts, the “User Selection
Double-click the server application icon placed window” opens. To start the client application,
in the task tray to display the “Select Server
Application Behavior dialog”.
follow these steps:
2
Select desired operation, and click [OK]. The User Utility can be configured for the client
application to start automatically.
Product Overview
For details, see “9.8 Setting Automatic Start-up of
Client Application”.
Product Overview
The “Main window” closes and the “Select Server
The “Select Server Application Behavior dialog” Application Behavior dialog” appears.
appears.
Product Overview
(1) DVD disk These disks are used to temporally store
images from the FCR PRIMA Console system
With FCR PRIMA Console, DVD disks are used
for such purposes, such as the provision of
to store images.
images to other facilities and collection of
The setup necessary to use DVD disk with FCR
images depicting various diseases.
PRIMA Console is called “The Start of Disk
Handling”. During this, the disk usage must be * FCR PRIMA Console does not remember which
specified according to the purpose for saving images are stored on which work disks. FCR
the image. Disk usage is classified as follows PRIMA Console treats the images imported
from the work disk as unrelated images with
according to the purpose for saving:
those in the FCR PRIMA Console system.
1) For image storage FCR PRIMA Console also treats disks storing
images provided by other facilities and DICOM-
Disks intended for image storage are called
compatible format disks storing images
storage disks. captured by the FCR system as work disks.
These disks are used to manage images. FCR
PRIMA Console remembers which images are Non-DICOM-compatible format
stored on which storage disks. Even though Disks in a non-DICOM-compatible format are
images are deleted from the FCR PRIMA called other disks.
These disks are used to store images in the
Console database, FCR PRIMA Console can
JPEG, bitmap or DICOM format for purposes,
still import the copies of the same images
such as presentations, etc.
from the storage disks.
FCR PRIMA Console cannot display the names
With images imported from disks, you cannot: of such disks and the amount of available
• Rotate or reverse the images, space on them.
• Save FCR image processing parameters,
For the procedure for starting disk handling, see
• Edit markers,
“2.5.2 Starting Disk Handling”.
• Update study information, and
• Verify or suspend studies.
* As the size of free space on the PC falls below
a certain limit, stored images are automatically
deleted from the FCR PRIMA Console database
in sequence in ascending time order of the
browsed studies.
• Being displayed in the viewer, Operation window” of the client utility “Start
• Printing or being saved, or Disk Handling” is performed on the control
• Locked. drive.
Studies deleted from the FCR PRIMA Console ● Disk for Normal drive
database can be reinstated in the FCR PRIMA
Normal drive can write data to other disks and
Console database by performing the operation
described in “Disk Restoration”. PDI disk, but cannot write data to storage disk
or work disk. Normal drive is not displayed in
For details, see “7.7 Handling Disks”. the “Disk window” of the indicator nor the
Images and study information associated “Disk Operation window” of the client utility.
with the images are stored on a storage disk,
Images in the disk can be loaded from both the
thereby allowing the studies to be reinstated in
control drive and normal drive.
the FCR PRIMA Console database.
1) For PDI
Called “PDI disk”.
CD-R or DVD-R disk for providing study
result images to the patient for diagnostic
cooperation with other facilities.
For instructions on how to save images to PDI disks,
see “3.5.3 Saving to PDI Disk”.
Product Overview
Confirm the DVD drive where the DVD disk
is inserted in “Drive” area of the appeared
“UDF Format Tool” dialog box.
2 disabled.
If no disk is inserted in the disk drive, insert
a disk into the disk drive by performing the 3 Click [Replace].
procedure below to start disk handling. Do
Product Overview
Device Status:
Shows the name of a disk and the amount of free
space (%) on it.
When the formatting process completes, the “Set Check that information about the disk whose
Disk Usage dialog” appears. Continue with Step 6 . handling has started is displayed.
HINT 2
6 Select the way you want to handle the disk
When the disk replacement operation completes,
inserted in the disk drive and click [Next]. remember to click [Update] to check that the disk
Product Overview
has been replaced. Simply replacing a disk will
not refresh the disk information.
3 Click [Eject].
Goes off
(for an image storage disk)
2
First, view the “Event window” to confirm that
Product Overview
the disk is full. Then, eject the full disk and
start the handling of a new disk. Whenever
a disk becomes full, take relevant action as
instructed on the right.
4 Remove the full disk from the disk drive
To start the handling of a new disk, follow these tray.
steps:
5 Start the handling of a new disk.
1 Select the event indicator to display the
Event tab. Handle disks pursuant to the instructions in
Event indicator “2.5.2 Starting Disk Handling”.
HINT
Simply placing a disk in the disk drive and closing
the tray won’t start the handling of the disk. See
2 Select the latest warning on display Disk instructions in “2.5.2 Starting Disk Handling”.
storing failed and click [Details].
Product Overview
then click [Eject]. illuminates green, click [OK].
To cancel the disk copying process, click
[Cancel].
2 Place a source disk in the disk drive tray 5 When the message appears, click [OK].
and close the tray. The target drive tray opens and a message appears.
HINT
When the disk access lamp is ON, never press
the Eject button on the disk drive.
Starting a Study
on the “Input ID window” of “Main window”, ID.
and select the column except the patient ID
in the “Input ID window”.
Enter the patient ID with single-byte alphanumeric
characters. The allowable character length is
determined by a system setting.
The following characters cannot be used:
“, #, %, &, ‘, +, ?, \, /, *, @, <
A patient ID consisting solely of spaces is not
allowed.
Patient Name:
Enter a name of patient.
Description for columns changes to “Register new
The following characters cannot be used:
patient”.
*, @, “, #, %, &, ‘, +, ?, \, /, <, =, 2-byte character.
★ If “Input ID tab” has not been selected when
Sex:
“Main window” is started, select “Input ID tab” and
Select a sex from the pull-down menu.
perform the operation 1 .
Date of Birth:
In the utilities customize, whether to display the Enter a date of birth in the same format as the
window at the time of starting the “Main window” sample entry appearing on the screen.
can be set.
Body length:
For details, see “8.2 Setting the General Enter a length of body.
Display Contents”.
Weight:
HINT Enter a weight.
When the patient ID which already registered in Patient comment:
the FCR PRIMA Console database is entered Enter a comment.
in the patient ID column, an applicable patient
information displays.
In such cases, confirm the patient information and
start the exposure.
3
Starting a Study
Starting a Study
To search using either of method, it determines
When the study performing with FCR
depends on the settings at the time of
PRIMA Console or the patient information
installation.
of study displayed in the study list is
Explains the procedures with using the
changed with MWM server, it will not
example of give priority to the MWM server and
reflect to the study list. In this case,
search.
also change the patient information of
1 Enter the patient ID into the Patient ID applicable study.
column in the “Main window” and select
the other column of patient information. For method of how to edit patient information,
see “7.2 Registering, Editing or Deleting
Enter the patient ID with single-byte alphanumeric Patient Information”.
characters. The allowable character length is
determined by a system setting.
The following characters cannot be used: 2 Confirm that the displayed patient
“, #, %, &, ‘, +, ?, \, /, *, @, <. A patient ID consisting information is correct, and select .
solely of spaces is not allowed. Further, if Automate
The patient information will be registered to FCR
Saving in General-Purpose File Format (optional) is
PRIMA Console.
selected, the characters :, > and | cannot be used
either.
MWM server is searched by entered patient ID.
For editing the exposure menu in user utilities, see selected, select the “Take X-rays tab”, and then
“9.2 Setting up the Exposure Menu and Study perform 1 operation.
Menu”.
★ When the menu you wish to expose is not existed
In the following procedures, select the exposure
in the displayed exposure menu, select the
menu and start the studies. [Others], and then select a menu.
Starting a Study
the left-hand side of the “Patient List window”, you
can view only those studies that are in the
Scheduled study status in the Patient list.
3
Starting a Study
Starting a Study
is blinking.
Cassette guide
Barcode window
Insert straight and carefully
3 Indicators
[ (Restore)]
Starting a Study
Indicators
Rotate and/or reverse the image in the rotation Displays the “Main Menu Modification window”.
direction of the selected button. An example of From the “Main Menu Modification window”,
the rotation of an image following the selection you can perform the following operations on
of [ (Rotate Left 90º)] is shown below. the current study:
• Add Menu
The rotation or reversal of an image is reflected
• Change Menu
in the image when it is printed on a film
• Delete menu, inactivate image, set re-
(optional), when it is saved on the disk, or when
exposure
it is sent to the DICOM server (optional).
• Relocate Display Order
• Exchange Images
• Output/Storage
• Edit Patient Information
For details on the “Main Menu Modification
window”, see “3.3.4 Editing Study Information”.
Starting a Study
The previous or next page appears on the
The “Image Viewer window” lets you magnify
current page to manipulate hidden registered
views of images or adjust their image quality.
exposure menus into view.
For details on the “Image Viewer window”, see
[Simple] “Chapter 4 Using the Image Viewer”.
[Suspend] [Complete]
3
Starting a Study
Starting a Study
To add a menu, follow these steps:
The selected menu is added to the selected list of
1 Click [Add Menu] in the “Main Menu menus on the right-hand side of the window.
Modification window”.
4 Click [Finish] in the lower right-hand corner
of the window.
Starting a Study
The “Deleting Menus/Inactivating Images window”
appears. [Undo Delete]
Click to disable read images from being saved Click to inactivate a read image and select a re-
to disk or printing on a film with the film print exposure menu.
option.
This is called image inactivation. Enabling 1 Select the exposure menu for a read image
inactivated images, for example, to be saved to and click [Re-exposure]
disk is called image activation.
3
A re-exposure mark is attached to the exposure
menu. This image is inactive and the same exposure
Starting a Study
[Back]
Starting a Study
3 Click [Finish] in the lower right-hand corner
The “Exposure Menu Relocate window” appears. of the window.
be exchanged.
Starting a Study
★ If you choose not to save images to the storage
disk, leave the DVD check box cleared.
HINT
Once images have their pixel density set to Save
by Default Pixel Count on FCR PRIMA Console,
the pixel density is kept as is even if Save by
Max Pixel Count is specified after reloading the
images into FCR PRIMA Console and saving
them in the disk.
The “Print Destination Detail window” appears. Click [Finish] in the lower right-hand corner of the
window to save the changes.
Starting a Study
(output destination) in DICOM server to which
images are transmitted to DICOM server and
the pixel density (output density) with which
the images are transmitted.
[Upon Completion]:
Prints images on a film when their study status
changes to Completed.
[Upon Verification]:
Prints images on a film when their study status
changes to Verified.
HINT
DICOM server that has already been selected as
a transmission destination (output destination)
does not appear on the list.
HINT
[Back]
Starting a Study
The following message is displayed:
Modification of patient information for this study The following message is displayed:
has been completed. Contents of modification will Are you sure you want to apply the changes to the
be reflected in the Patient Information Database. database?
Do you wish to proceed?
★ To apply the edits made to the detailed patient
★ To apply the modified patient ID not only to this information not only to this study but also to the
study but also to the patient information stored in patient information stored in the database, click
the database, click [OK]. [OK].
★ To apply the modified patient ID to this study only ★ To apply the edits made to the detailed patient
and not to the patient information stored in the information to this study only and not to the
database, click [Cancel]. patient information stored in the database, click
[Cancel].
The “Read Image window” appears.
Check that the patient ID has been modified. The “Read Image window” appears.
Check that the detailed patient information has been
modified.
HINT
If the patient information matching the patient ID
entered in the “Search for ID Number window”
is not registered in the FCR PRIMA Console
database, no detailed patient information is
displayed in the “Search for and Edit Patient
Information window”.
To register new patient information in the FCR
PRIMA Console database, proceed to enter
detailed patient information.
Starting a Study
search requirement
2) Search panel
3) Patient list
Starting a Study
finishing date are specified, the specified period will
be the search requirement.
Patient ID or Patient name:
Enter the patient ID or patient name.
In the default, the study that the registration
date is set to today is displayed in the list. CAUTION
When specifying the registration date
1 Select [Search Requirement display and study date, the calendar displays.
column]. When click another area on the
window without selecting the date in
the calendar, display of calendar may
Search panel displays.
remain. To hide the calendar, select the
date in the calendar.
3 Select .
The study which matches with the selected search
requirement displays as a patient list.
Starting a Study
+Do you wish to continue?
Check the image reading status to make sure that
the study for which you are trying to start an
exposure operation is displayed in a “Read Image
window” by another user. Then, take either of the
following actions:
★ To continue with the exposure operation for The search requirements are saved.
this study, click [OK] to display an “Read Image
window”.
★ To cancel the exposure operation for this study,
click [Cancel].
- If the following message is displayed –
A study is already in progress.
*Carry it out again after a while.
Click [OK] to close the message.
An exposure operation is being performed by a
different user or on the other PC connected.
Let the other user or PC finish its exposure operation
before starting your exposure operation.
Set the search panel to close automatically at Switch the study information of list to display/
the time of searching. hide.
1 Select the “Close this panel automatically” 1 Select with the condition which
checkbox. the patient list is displayed.
3
Starting a Study
When the search button is selected, the search The patient information of patient list will be grayed
panel is automatically closed. out.
When the “Privacy” is selected again, the patient
Initialize the search requirement information is displayed.
When the “Privacy” is selected with the condition
Reset the requirement selected in the column which the patient is selected, the patient information
for each search requirement, and initialize. The except the selected patient will be hidden.
default requirements shows the conditions that
the registration date is set to “Today” and other
requirements are not selected.
Set the items based on the display order of Update the patient list manually.
patient list, and the list can be arranged.
1 Select .
1 Select the heading of item based on the
display order.
The ascending order (▲) and descending order (▼)
are changed by selecting each heading.
Starting a Study
The patient list is updated with the search
requirement displayed in the column which displays
the search requirements.
HINT
Items or orders to display in the patient list can be
set in the customize.
CAUTION
When launching is not displayed
at the lower of the “Refer window”,
operation items must be added in the
customize.
For details on searching method, see “3.4 Using The selected study displays in the “Image Viewer
a Patient List”. window”.
- If no disk is mounted -
3
The following message is displayed:
Preparing to retrieve images
Starting a Study
When the disk mounting process completes, the
“Images retrieve dialog” appears, enabling you to
confirm the status of image loading.
• button
Switch the auto image reference function to ON/
OFF.
CAUTION
3 When is not displayed, operation
items must be added in the customize.
Starting a Study
Select an operational menu from the following For instructions on how to edit studies, see
“3.3.4 Editing Study Information”.
choices and configure it to conduct an additional or
re-exposure operation. To close the edit main menu, click [Close].
Starting a Study
The selected study returns to the Started study For re-study/edit in the study utilities, see
“6.1.4 (2) Reconduct a study and editing
status, enabling you to start an exposure operation.
information”.
For instructions on how to start an exposure
operation, see “3.2 Starting an exposure
operation”.
HINT
If [Output after Re-verified] in the user utility
printer settings has been unchecked (setting for
no film printing at the time of study re-verification),
perform print output manually in image units at
the time of study menu addition.
CAUTION
If an already verified study (which is
in the Verified study status) has been
reconducted, verify the study again when
the study is completed.
• Change Menu
Changes the menu for a read image and
reprocesses the image according to the new menu.
If the image has been rotated or reversed, its
status is updated according to the new menu.
• Relocate Display Order
Changes the order of appearance of a menu in the
“Read Image window”.
• Exchange Images
Exchanges images to correct the order in which
they are read.
• Output/Storage
Modifies the settings for saving images on a
storage disk and printing them on a film (optional).
• Edit Patient Information
Edits patient information. Edits made to the patient
3 WARNING
Before saving study image in PDI disk,
write the following information on
the surface of the PDI disk to prevent
Starting a Study
CAUTION
When is not displayed, operation
items must be added in the customize.
The “PDI Creation Wizard window” appears. - When the following message is displayed - 3
You can view the write preparation status.
No disk is inserted.
★ To cancel the PDI file creation, click [Cancel].
Starting a Study
Unable to write to this disk.
★ To output images with annotations embedded, ★ If a PDI file is created in the storage disk drive,
select [Yes] in the “Output images with set the storage disk in the drive.
annotations embedded”.
HINT
PDI can be created in the study utilities.
CAUTION
Starting a Study
All imgs included in a study:
Print all the images in the study on a film. Images
already printed on a film are reprinted.
Study imgs yet-to-be output:
Print only those images in a film that have not been
printed on a film.
7 Select a priority.
8 Select [Output].
HINT
Printing on a film can be performed in the study
utilities.
CAUTION
4 Select a priority.
3
LOW : Gives higher priority to other save
operations.
Starting a Study
MEDIUM : Save with the ordinary priority.
HIGH : Starts save immediately.
5 Select [Output].
HINT
When the disk access lamp is ON, never press
the Eject button on the disk drive.
HINT
Saving to a storage disk can be performed in the
study utilities.
CAUTION
3 When is not displayed, operation
items must be added in the customize.
Starting a Study
HINT
Transmitting to a DICOM server can be
- If the following message is displayed -
performed in the study utilities.
3
Some of the studies cannot be handled.
● Do you wish to continue excluding the studies For re-study/edit in the study utilities, see
which are not able to process? “6.1.4 (12) Transmitting studies to DICOM
server (optional)”.
★ To transmit only those studies that contain
Starting a Study
images to DICOM server among all selected
studies, click [OK].
★ To cancel the transmission of the selected study
to DICOM server, click [Cancel].
4 Select a destination.
6 Select a priority.
Images read from the “Read Image window” are ★ To delete the exposure menus from which images
displayed in the Image Viewer. The Image Viewer are to be read and use the Image Viewer, click
can be launched after all the images in a study [OK].
are read. When the Image Viewer launches, the ★ To choose not to use the Image Viewer, click
study is updated to have the completed status. [Cancel].
CAUTION
When is not displayed at the lower
part of “Refer window”, operation items
must be added in the customize.
4 HINT
The same operation can be performed in the
following.
Using the Image Viewer
4
The “Requesting for disk insert dialog” appears.
Using the Image Viewer
4
3) Toolbar
(Study window bar)
Using the Image Viewer
4) Study window
5) Image pane
6) Thumbnail images
Thumbnail images of all images in the study.
Double-click a thumbnail image to display it in the
image pane.
Maximize button Storage/print button An image having a mis-exposure mark shown in
its lower right-hand corner is an invalidated image.
• Maximize button
Click the Mis-exposure mark to view the image in
Enlarges an image to fill the window.
the “Mis-exposed Image Display window”.
For details, see “4.2.3 Magnifying Images”.
• Storage/print button
Selects an image to be saved or printed.
For details, see “4.5 Printing or Saving
Mis-exposure mark
Images”.
For details, see “4.7.1 Thumbnail Images and
❍ The text appearing in the top of each image
Study Status ([Suspend][Verify])”.
pane provides information about the image,
such as the menu name and magnification/
reduction ratio.
7) Study status ([Suspend] [Verify])
For details, see “4.7.2 Viewing Image Update a study that has the completed study status
Information”. to either the verified or suspended study status.
Click [Suspend] to leave the study having the
completed study status. Click [Verify] to update the
study to the verified study status.
For details, see “4.7.6 Suspending and Verifying a
Study”.
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Prev. Patient] or
[Next Patient].
The previous, next, or top image is displayed.
The buttons are not available when there is no
previous or next image.
You can do the same by right-clicking anywhere Move the mouse pointer over the image and it will
in the image pane and clicking [Prev. Study] or turn to .
[Next Study].
HINT
In stack display, the images can be displayed
fit to the window using the “AutoFit to Window”
function.
HINT
This image switching operation is also possible
when multiple image windows are displayed side
by side. In this case, drag the mouse on one of
the displayed images. During operation, only the
image with the mouse pointer on it is switched.
When mouse dragging is ended, that image and
the images linked with it are displayed.
4
Using the Image Viewer
Maximize button
HINT
When [Stack Display] has been selected, the
maximize button is not displayed and this function
cannot be used.
Restore button
study study
• Images 4 x 3
→ Displays 12 image panes arranged horizontally
and vertically.
Images are displayed starting from upper left to
upper right, and from top to bottom in sequence.
image image image image
• Images 4 x 4
→ Displays 16 image panes arranged horizontally
and vertically.
4
Images are displayed starting from upper left to
upper right, and from top to bottom in sequence.
• Images 4 x 6
→ Displays 24 image panes arranged horizontally
and vertically.
Images are displayed starting from upper left to
upper right, and from top to bottom in sequence.
image image image image
image image image image
image image image image
image image image image
image image image image
image image image image
• Images 6 x 4
→ Displays 24 image panes arranged horizontally
and vertically.
Images are displayed starting from upper left to
upper right, and from top to bottom in sequence.
image image image image image image
HINT
To view multiple images in an image pane, right-
click on the upper right-hand corner of an image
4 and choose from among the following:
[Image Pane Pattern] - [1 x 2]
[Image Pane Pattern] - [2 x 1]
Using the Image Viewer
2 Search through FCR PRIMA Console or a ★ To cancel the comparative image display, click
disk for past studies. [Cancel].
left-hand side of the “Image Viewer window”, with field. When the “Specifying the Date dialog”
past images appearing in the image pane on the appears, specify a date and click [Setup].
right-hand side. Check Study Date.
To narrow down the search or to change the Matching studies are narrowed down according to
display order in the “Acquire Comparative the conditions set.
Images window”, follow these steps:
● Narrowing down your search
Enter a search keyword in a keyword entry field.
HINT
To exit the gradation change mode, click
[Gradation] once again.
4
Using the Image Viewer
4
Select the direction you want the image
rotated. The frame of the selected image turns
4
Using the Image Viewer
● Magnifying an image
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Magnify]. 4
HINT
To exit the reduce mode, click [Reduce] once
again.
HINT
In stack display, the images can be displayed
4 fit to the window using the [AutoFit to Window]
function.
1 Click [Panning].
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Panning].
4
turn to .
HINT
To exit the image move mode, click [Panning]
DRC GP/MFP
❍: Exclusive control
Characteristics of each image processing
parameter are defined below, along with an
acceptable range of input values.
GT DRT
Characteristics Shape of the gray level curve. Characteristics This is a DR compression filter type.
Yields the same result as changing Sets the density region for
the γ-curve of X-ray films. processing.
Value range A to Z (26 types) Value range A to T (20 types)
GC DRE
Characteristics Center density of gray level curve Characteristics Enhancement of DR compression
slope.
Changes GA according to the density
filter processing.
Changes the enhancement level of
4
specified here. processing in accordance with the
specified value.
Value range 0.30 to 2.64 (2 decimal places)
MRT
Characteristics Shape of the enhanced curve
corresponding to the density of the
image.
By changing this parameter, the
enhancement level can be adjusted
within a certain density region.
Value range F, P to Z (12 types)
MRE
Characteristics Enhancement of frequency
processing.
Changes the enhancement level of [Cancel]
processing in accordance with the
specified value. Cancels the changes made to the image
Value range 0.0 to 9.9 (1 decimal place) processing parameters and closes the “Modify
10 to 16 (no decimal places) FCR Image Processing Parameters window”.
HINT
Once you click [Save], the image processing
parameters cannot be restored to those in effect
at the time of exposure. If it is necessary to
restore the image processing parameters to those
in effect at the time of exposure, click [Apply].
4
Using the Image Viewer
2 Click [Reverse].
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Processing] -
[Reverse].
5 Click [OK].
The preset settings have been registered.
HINT
When preset settings are assigned to keypad
keys, image processing can be done just by
pressing a keypad key.
settings
Preset names can be changed or deleted,
display or no display in the preset list can be
selected, and the display sequence can be
changed.
1 Click [Organize].
4 Click [Setting].
The preset settings have been changed.
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Annotation] - The added arrows disappear by clicking [Reset].
[Arrow]. To save the arrows, click [Save Annotation]. Once
the arrow is saved, it would not be undone if [Reset]
is clicked.
Move the mouse pointer over the image and it will 3 Select [Arrow] again to end arrow drawing
turn to . mode.
HINT
HINT You can do the same by right-clicking anywhere
To exit the arrow drawing mode, click [Arrow] in the image pane and clicking [Annotation] -
once again. [Arrow] once again.
2 Drag the mouse pointer from the start point The shape of the mouse pointer will return to .
to the end point of an arrow.
HINT
Drag To hide the arrows, click [Annotation] in selected
state.
You can do the same by right-clicking anywhere
in the image pane and clicking [Display/Hide] -
[Annotation] in selected state.
HINT
To exit the ellipse drawing mode, click [Ellipse]
once again.
Drag
HINT
To exit the rectangle drawing mode, click
[Rectangle] once again.
Drag
turn to .
HINT
HINT To hide the polygons, click the checked
[Annotation].
To exit the polygon drawing mode, click [Polygon] You can do the same by right-clicking anywhere
once again. in the image pane and clicking the checked
[Display/Hide]-[Annotation].
2 Click the mouse to specify an apex of a
polygon and double-click the mouse at the
last apex position.
(4) Double-Click
(1) Click
(2) Click
(3) Click
menu displayed by clicking on the right of the 5 To exit the text input mode, click [Text]
window. once again.
HINT HINT
You can do the same by right-clicking anywhere
You can do the same by right-clicking anywhere
in the image pane and clicking [Annotation] -
in the image pane and clicking [Annotation] -
[Text] once again.
[Text].
4
The shape of the mouse pointer will return to .
HINT
To hide the text, click the checked [Annotation].
HINT
To exit the text input mode, click [Text] once
again.
3 Input text.
4 1 Click [Properties].
HINT
A desired button may not display on the window. Text
Using the Image Viewer
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Properties].
Font typeface:
The “Annotation Properties dialog” appears. This is the font type.
Font style:
2 Select the tab of the object to be changed Selection is possible from regular, italic, bold, and
and set line style and line color or font bold italic.
typeface, style, and size. Font size:
Selection is possible from 5 to 96 points.
Arrow, Ellipse, Rectangle, and Polygon
Font color and background:
Selection is possible from the following.
3 Click [OK].
Line style:
Set the line style for each object. In case of an ★ To cancel changing the properties, click [Cancel]
arrow, the arrow tip becomes larger or smaller to close the “Annotation Properties dialog”.
proportional to the line style setting. By this setting, the properties will be changed.
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Annotation] -
[Marker].
4
A marker is displayed on the clicking point, and the
shape of the mouse pointer will return to .
dialog”. HINT
To hide the markers, click the checked
[Annotation].
You can do the same by right-clicking anywhere
in the image pane and clicking the checked
[Display/Hide]-[Annotation].
HINT
Markers can be system markers registered in
advance and user markers registered by a user.
Select [System] to use system markers or [User]
to use user markers and switch the displayed
markers.
The markers to be used can be registered and
deleted. System markers cannot be registered or
deleted.
HINT
Measurement results yielded by this processing
are presented only for reference purposes without
any warranties of their accuracy. It is the user’s
responsibility to evaluate the measurement Click [Reset] to return to the status in effect prior to
results. the length measurement.
The measured results are displayed to two digit of
decimal point by round to the thousandth digit. To save the measurement result, click [Save
Annotation]. Once the measurement result is saved,
1 Click [Length]. it would not be undone if [Reset] is clicked.
HINT
To hide the measurement results, click
[Annotation] in selected state.
You can do the same by right-clicking anywhere
Move the mouse pointer over the image and it will in the image pane and clicking [Display/Hide]-
turn to . [Annotation] in selected state.
HINT
To exit the linear measurement mode, click
[Length] once again.
Drag
HINT
Measurement results yielded by this processing
are presented only for reference purposes without
any warranties of their accuracy. It is the user’s
responsibility to evaluate the measurement
results. (1) Click (2) Click
The measured results are displayed to
thousandth digit of decimal point by round to the
ten-thousandth digit.
The maximum inside diameter of the thoracic cage
1 Select the image in which you want to is indicated (Displayed to thousandth digit of decimal
point by round to the ten-thousandth digit).
measure CTR.
HINT
To exit the CTR mode, click [CTR] once again.
First, measure the maximum inside diameter of the
thoracic cage.
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Angle between
the two line segments].
HINT
To exit the Angle Measurement mode, click
[Angle between the two line segments] once
again.
Drag HINT
To hide the measurement results, click the
checked [Annotation].
You can do the same by right-clicking anywhere
3 Drag the mouse pointer from the start point in the image pane and clicking the checked
to the end point of a straight line to draw [Display/Hide]-[Annotation].
the second line segment.
4
4
Using the Image Viewer
HINT
You can do the same by:
• Clicking [Edit Annotation] in the toolbar.
• Right-clicking anywhere in the image pane and
clicking [Save Annotation].
HINT
If a new annotation is added surrounding the
original annotation, the original annotation
cannot be selected. By moving or editing the
added annotation, the original annotation can be
selected again.
Move the mouse pointer over the image and it will The shape of the mouse pointer will return to .
turn to .
HINT
To exit the annotation edit mode, click [Edit
Annotation] once again.
Drag
HINT
You can do the same by right-clicking anywhere Click [Reset] to return to the status in which the
in the image pane and clicking [Edit Annotation]. annotations were last saved.
To save the annotation, click [Save Annotation].
Once the annotation is saved, it would not be
undone if [Reset] is clicked.
Move the mouse pointer over the image and it will For details, see “4.4.4 Saving Annotations”.
turn to .
HINT
To exit the annotation edit mode, click [Edit
Annotation] once again.
HINT
With selecting [Properties] after right-clicking an
arrow or a shape, the property of an arrow or a
shape can be changed.
2 Editing Annotations
● Editing an arrow
(1) Select the arrow you want to edit.
Select
Select Select
(2) Select a point indicating the range of the (2) Select an apex of a rectangle and drag it to a
ellipse and drag it to a desired position. desired position.
4
Drag
Drag
Select
4 Drag
Using the Image Viewer
menu displayed by clicking on the right of the Click [Reset] to return to the status in which the
window. annotations were last saved.
To save the annotation, click [Save Annotation].
HINT Once the annotation is saved, it would not be
You can do the same by right-clicking anywhere undone if [Reset] is clicked.
in the image pane and clicking [Edit Annotation].
For details, see “4.4.4 Saving Annotations”.
HINT
With selecting [Properties] after right-clicking the
text, the property of the text can be changed.
Double-Click
The insertion point of the character strings will
display.
HINT
You can do the same by right-clicking with
selecting the text and clicking [Edit Text].
2 Select the marker you want to edit. Font color and background:
Selection is possible from the following.
HINT
To exit the annotation edit mode, click [Edit
Annotation] once again.
Select
Select
Drag
Select
4
Drag
HINT
Using the Image Viewer
HINT
To exit the annotation edit mode, click [Edit
Annotation] once again.
HINT
When [Output after Re-verified] in the user utility
printer settings has been unchecked (setting
for no film printing at the time of inspection re-
verification), print output has to be performed
manually in image units at the time of adding
study menu.
7 Click [Output].
HINT
When the disk access lamp is ON, never press
the eject button on the disk drive.
CAUTION
When is not displayed in the
4 “Refer window”, operation items must be
added in the customize.
Using the Image Viewer
4
The indicator at the left bottom of the window can be
HINT
You can do the same by right-clicking on the
study window bar of the study to be printed and
selecting [Storage/Print] – [Freelayout].
1 Click [Close].
2 Click [OK].
The “Free layout print screen” is closed and the
display returns to the “Main window” or the “Image
Viewer window”.
4
Using the Image Viewer
1 Click .
The “Format information setting dialog” appears.
2 Click .
The “Input format name dialog” appears.
7 Click .
The selected items are added as auxiliary
information. The auxiliary information of the selected format
appears.
8 Click [OK].
★ To cancel auxiliary information format setting, The “Format information setting dialog” is closed.
select [Cancel] and close the “Format information
setting dialog”.
The auxiliary information is set.
HINT
The display contents for the auxiliary information
The “Format information setting dialog” is closed. of each image can be set even after image layout.
1 Click .
The “Format information setting dialog” appears.
HINT
When the cursor is moved onto the border of the
cell, it changes to or and the cell width or
height can be changed.
1 Click .
The “Footer information setting dialog” appears.
The “Footer information setting dialog” is closed.
HINT
1 Click .
The annotations appear.
4
Using the Image Viewer
4
● Reducing an image
1 Click .
Move the mouse pointer over the image and it will
turn to .
4
Using the Image Viewer
CAUTION
The deleted image cannot be undeleted.
Select an image again on the left side
of the screen and add it to the image
window.
1 Click .
The Confirmation dialog appears.
2 Click [OK].
Film printing is done with the layout of the image
window.
4
Using the Image Viewer
1 Click [Thumbnails/Status].
Thumbnail images Study status ([Suspend][Verify])
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Display/ Repeat the procedure above to view multiple images
Hide] - [Thumbnails/Status]. (The check-mark each in a desired image pane.
disappears.)
● Showing an invalidated image
Thumbnail images and study status are hidden.
1 Select the inactivated mark in the lower
4 Because this setting is saved, the Image Viewer
comes up with the thumbnail images and study right-hand corner of a thumbnail image.
analysis being hidden next time it launches.
Using the Image Viewer
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Display/Hide] -
[Image Information].
(The check-mark disappears.)
4
Image information is hidden. Because this setting is
saved, the Image Viewer comes up with the image
information being hidden next time it launches.
Image information Image information
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Display/Hide] -
[Image Information].
Tree View
Patient information
Study information HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Display/Hide] -
[Tree View].
(The check-mark disappears.)
4
The tree view is hidden.
Because this setting is saved, the Image Viewer
comes up with the tree view hidden next time it
launches.
Using the Image Viewer
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Display/Hide] - An image from the double-clicked study is displayed.
[Tree View].
4
mark
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Display/Hide] -
[Annotation].
(The check-mark disappears.)
Showing annotations
1 Click [Annotation].
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Display/Hide] -
[Annotation].
Toolbar
Display Settings
Displays the Display Settings tab. Configure 4
toolbars to display.
HINT
You can do the same by right-clicking anywhere
in the image pane and clicking [Toolbar Settings]
- [Toolbar Settings].
4 Buttons:
segments on the image)
❍ [Text]
❍ [Gradation]
(→ Displays character strings on the image.)
(→ Changes gradation (GA and GS))
❍ [Marker]
❍ [Magnify]
Using the Image Viewer
Buttons: Buttons:
❍ [Prev. Patient] ❍ [Image 1 x 1]
(→ Displays a study for previous patient (→ Displays one image pane)
information)
❍ [Next Patient] Image
(→ Displays a study for next patient information)
❍ [Prev. Study] ❍ [Image 1 x 2]
(→ Displays the previous study for the patient (→ Displays two image panes arranged vertically)
information on display)
Image
❍ [Next Study]
(→ Displays the next study for the patient Image
information on display)
❍ [Prev. Image] ❍ [Images 2 x 1]
(→ Displays the previous image from the study on (→ Displays two image panes arranged
display) horizontally)
❍ [Next Image]
(→ Displays the next image from the study on Image Image 4
display)
❍ [Top Image] ❍ [Images 2 x 2]
Buttons:
❍ [Images 2 x 6]
❍ [Study 1 x 1]
(→ Displays 12 image panes arranged horizontally
(→ Displays one study window)
and vertically from upper left to upper right, and
from top to bottom in sequence.)
Study Image Image
Image Image
Image Image
Image Image
❍ [Study 2 x 1]
Image Image
(→ Displays two study windows) Image Image
Study Study
❍ [Image 4 x 1]
(→ Displays four image panes arranged Modify Study Info
horizontally)
Buttons:
❍ [Menu Change]
Image Image Image Image
(→ Changes menus in the “Menu Change window”)
❍ [Mis-exposure/Re-exposure]
❍ [Images 4 x 2] (→ Invalidates images or sets a re-exposure
4
(→ Displays eight image panes arranged operation in the “Delete/Inactivate Menu
horizontally and vertically: four horizontally and window”)
two vertically from upper left to upper right, and ❍ [Additional Exposures]
from top to bottom in sequence.)
(→ Registers additional exposure menus in the
Using the Image Viewer
4
1 Select a button style.
Adding buttons to Toolbar Favorites
1 Select a button to be deleted from Toolbar 1 Select the button you want to reorder in
Favorites and click [<-]. Toolbar Favorites and click [ (Move UP)]
or [ (Move DOWN)].
HINT
When a button is moved up, it moves to the left
on the toolbar. When a button is moved down, it
moves to the right on the toolbar.
4 HINT
You can do the same by double-clicking the
button to be deleted from Toolbar Favorites.
Using the Image Viewer
[Setup]
Saves changes made to the toolbar display/
hide, button display style and favorites settings
and closes the “Toolbar Settings dialog”.
[Cancel]
Closes the “Toolbar Settings dialog” without
saving the changes made to the toolbar
display/hide, button display style and favorites
settings.
HINT
If neither [Suspend] nor [Verify] is displayed under
an image pane, display them in either of the
following ways:
• Click [Thumbnails/Status] in the toolbar.
• Right-clicking anywhere in the image pane and
click [Display/Hide] - Click [Thumbnails/Status].
HINT
The exposure menu is displayed as image
information.
HINT
Re-verify the study.
★ To cancel the menu changes and return to
the “Changing Exposure Menu window”, click
[Cancel].
4
A mark appears indicating that the image is
Using the Image Viewer
inactivated.
2 Click [Mis-exposure/Re-exposure].
HINT
You can do the same by right-clicking in the
study window bar of the study for which you
want to set a mis-exposure or re-exposure and
clicking [Modify Study Info.] - [Mis-exposure/Re-
exposure].
[Activate]
The “Deleting Menus/Inactivating Images window” Select an exposure menu that has been
appears. inactivated by [Inactivate] or [Re-exposure] and
click [Activate] to activate the image.
Check the “Image Viewer window” to make sure
that the exposure menu for the image has been
modified.
HINT
A re-exposure mark is attached to the exposure Re-verify the study when it is finished.
menu. This image is inactivated and the same
exposure menu as the image is added to the study.
★ To cancel the programming of the re-exposure and
return to the “Delete/Inactivate Menu window”, click
4
[Cancel].
[Finish]
Click to save changes to the mis-exposure
(image invalidation) or re-exposure setting and
close the “Delete/Inactivate Menu window”
appears.
Check the “Image Viewer window” to make
sure that an image inactivation mark is
attached to the thumbnail image.
If a re-exposure is programmed, message
Being studied, is displayed in the study window
bar. The study returns to the Started study
status. Exit the Image Viewer and start a study.
For details, see “3.4.3 Manipulating Patient List”.
HINT
You can do the same by right-clicking in the study
window bar of the study to which you are about
to add a menu and click [Modify Study Info.] - The selected exposure menu is added to the
[Additional Exposures].
selected list of menus on the right-hand side of the
window.
The “Adding Exposure Menus window” appears. ★ To cancel the addition of the exposure menu,
click [Close] and close the “Adding exposure
menu window”.
HINT
Re-verify the study when it is finished. 4
★ To cancel the programming of the additional
exposure menu and return to the “Adding
4
Using the Image Viewer
The “Replacing Images window” appears. The “Replacing Images window” closes.
Check the “Image Viewer window” to make sure that
the images have been exchanged.
- If the following message is displayed -
Verified studies are about to be changed.
❍Do you wish to continue?
Notice: To continue, re-verify operations
necessary to verify studies.
If images in a verified study are exchanged, the
st u dy wo u l d r et u r n f r o m t h e Ve r i f i e d to t h e
Completed study status.
4
handling.
Select the DVD check box to save images to the
storage disk.
To change the output density of images saved to the
4
Using the Image Viewer
HINT
Images once saved by default pixel count on
FCR PRIMA Console, cannot be saved by max.
pixel count when reloaded. Save by Default Pixel
Count would be the only option.
1 Click [Details] at the rightmost end of the You can modify the transmission destination
film print indicating line. (output destination) in DICOM server to which
images are transmitted to DICOM server as well
as the pixel density (output density) with which
The “Setting Image Print/Storage window” for film the images are transmitted.
print setting appears.
1 Click [Details] at the rightmost end of the
DICOM output indicating line.
HINT
DICOM server already selected as a transmission
destination (output destination) does not appear
on the list.
system.
HINT
Images once saved by default pixel count on
FCR PRIMA Console, cannot be saved by max.
pixel count when reloaded. Save by Default Pixel
Count would be the only option.
★ To apply the edits made to the detailed patient The “Edit Patient Information window” appears.
information not only to this study but also to the
patient information stored in the database, click ❍ If the patient information located is not
[OK]. displayed in the “Edit Patient Information
★ To apply the edits made to the detailed patient window”
information to this study only and not to the It means that the patient information matching
patient information stored in the database, click
the patient ID entered in the “Input Patient ID
[Cancel].
window” is not registered in the FCR PRIMA
The “Input Patient Information window” closes. Console database. Retry the search with
Check the “Image Viewer window” to make sure that another patient ID or register it as a new patient
the detailed patient information has been modified. information.
- If the following message is displayed - ★ To retry with another patient ID, click [Back] to
return to the “Input Patient ID window”.
Verified studies are about to be changed.
★ To register new patient information, return to the
❍Do you wish to continue?
“Edit Patient Information window” in these steps:
Notice: To continue, re-verify operations
(1) Click [Back] in the “Modifying patient
necessary to verify studies.
information window”.
If you modify detailed patient information for a (2) Click [Back] in the “Search for ID Number
verified study, the study would return from the window”.
verified to the completed study status. (3) Return from the “Image Viewer window” to the
★ To modify the detailed patient information, click “Main window”.
[OK]. (4) Register new patient information.
HINT
Re-verify the study.
★ To cancel the edits made to the patient ID and
return to the “Modifying patient information
window”, click [Cancel].
HINT
When the Free Drawing function is started, the
pen color used last the previous time is selected.
1 Click [Exit].
HINT
(3) Capturing the drawn image When exiting the Free Drawing function, the
selected pen size and pen color are stored,
and they are set again when the Free Drawing
1 Click [Snapshot]. function is started again.
4
HINT
When images are displayed on multiple monitors,
the image data of both monitors are copied to the
clipboard as one file. In case of different screen
sizes, the images copied to the clipboard with
their upper edges aligned.
Monitor 2 Monitor 2
Monitor 1 Monitor 1
1 Click [Delete].
Indicators are separated from the “Main window” (1) Image Reader
or “Image Viewer window”, so it can be placed
in any spots. Indicator
Notice
state
Indicators Off No Image Reader is connected to the
FCR PRIMA Console.
Images cannot be read. Connect an
Image Reader.
On An Image Reader is connected to the
FCR PRIMA Console. 5
Images can be read.
Blinking An error has occurred with the Image
Reader connected to the FCR PRIMA
[Store-Disk List]
Confirming by Viewing Indicators
Status:
Shows the name of a disk and the amount of free
Disk Name:
space (%) on it.
Storage disk name.
Usage:
Indicates whether the disk is intended as a storage Write Start:
disk or work disk. The date and time at which writing to a storage disk
Nothing appears for any other type of disk. started.
For disk details, see “2.5.1 Disks and Drives Write End:
Used by FCR PRIMA Console”.
The date and time at which writing to a storage disk
Disk Write: ended.
Indicates whether a storage disk inserted is write-
enabled or read-only. Status:
Indicates whether the storage disk is currently write-
enabled or it is a read-only disk.
● DICOM-Compatible Format:
Use the disk as a work disk.
● Other:
Use the disk for any other purpose.
For disk usage details, see “2.5.1 Disks and
Drives Used by FCR PRIMA Console”.
HINT
To replace the disk, see “[Replace]” in this Check that the information of the disk being handled
subsection. is displayed.
To eject the disk, see “2.5.3 Ejecting Disk”.
Indicates whether the disk is intended as a storage Restores a storage or work disk after mismatches
disk or work disk. occurring in the image management file on the
Nothing appears for any other type of disk. disk resulting from a power failure while writing
to the disk or other adverse conditions.
Disk Write:
Indicates whether a storage disk inserted is write-
enabled or read-only. The following message is displayed if
mismatches occur in the image management
file:
A data mismatch is found.
❍Restore the disk.
[Replacing Disks]
Replaces the disk shown in the current “Disk
Operation dialog”. Do the same as [Replace] in
the Disk tab.
5
Confirming by Viewing Indicators
[Update]
Refreshes the view of the device status, usage
and the setting for writing data.
[OK]
Closes the Disk tab.
CAUTION
• When the HDD does not have enough
space, install another NAS. In such
cases, contact our official dealer.
• When an error of NAS is displayed in the
indicator for Event, contact our official
dealer to fix it.
HINT
Select tabs in the top of the dialog to confirm
the Image Reader, film printer, output status or
events.
[Update]
Refreshes the view of the device status.
[OK]
Closes the Printer tab.
HINT
Select tabs in the top of the window to confirm the
disk, film printer, output or event status.
[Force Output]
Outputs again an image that has once failed to
be output because of an error encountered.
HINT
[OK] Select tabs in the top of the window to confirm the
Image Reader, disk, film printer or events.
Closes the Output Status tab.
[Details]
Confirms details about an event occurring.
The Events tab appears, allowing you to confirm the
1 Select an event of interest and click
events (errors, warnings or information) that have
occurred on the FCR PRIMA Console system. The [Details]. 5
events are shown in the descending order of Date/
Time.
[One page]
Check all the events in the current page as being
confirmed.
Confirmed events are hidden when Unconfirmed
Only is selected in Status.
[Confirm all]
Check all the events as being confirmed.
Confirmed events are hidden when Unconfirmed
Only is selected in Status.
HINT
Select tabs in the top of the window to confirm the
Image Reader, disk, film printer or output status.
6
5) Study operations
2) Search queries
7) List operations
study.
Patient ID Indicates the ID number of • Undelivered
patient studied. Images are ready to be stored
and delivered.
Patient Name Indicates the name of patient
studied. • Delivering
Images are being delivered
Sex Indicates the sex of patient
from storage/printing.
studied.
• Delivered
Date of Birth Indicates the date of birth of
Storage/printing has ended.
patient studied.
Lock Indicates whether the study is
Study Instance UID Indicates numbers uniquely
locked or not.
assigned to each individual study.
• -
Date of Registration Indicates the date on which the
The image is not locked and
study was registered (scheduled).
the study can be deleted.
Time of Registration Indicates the time at which the
• Locked
study was registered (scheduled).
The image is locked and the
Study Date Indicates the date on which the study cannot be deleted.
exposure was started.
# of Img Indicates the number of images
Study Time Indicates the time at which the that have been read or that are
exposure was started. scheduled to be read.
Study Description Indicates the name of the study
menu that has been created with
the User Utility.
[Select All]:
Selects all the studies in the current page.
[Unselect All]:
Deselects all the studies in the current page.
[Update]:
Updates the display of the study list to the latest
status.
HINT
Click [Save as is] to overwrite the existing search
■ Date of Birth, Study Date, Date of Registration: queries with the new display order.
To create new search queries, display the “List The Search Items tab appears.
Search Conditions window” and then select Set the items to be used as search conditions.
desired tabs to enter required settings as
instructed below.
[Cancel]
Click to edit search queries. Click to delete search queries created, except
for the search queries All, Today, Scheduled,
1 Select a search query and click [Edit]. Started, Unverified, Undelivered and Temporal.
HINT
To delete the search query set to initial condition,
set another search query to initial condition; then,
the search query can be deleted.
HINT
You may not be able to edit some tabs or items in
the predefined set of search queries depending
on the selected search queries.
[Save]
[Cancel]
6
[Set to Initial Condition]
status
You can narrow down your search through a
list of studies that have been selected to meet
a specified set of search queries or change the
order of appearance of the studies. Click [Save
as is] to overwrite the current set of search
queries with edited versions of the search
queries, including display width and sort order
changes.
HINT
You can do the same by:
• Selecting a study in the study list and clicking
[Image Viewer] on the right-hand side of the
window.
• Entering the number of the study (item at the
leftmost end of the list) you want to view in the
“Image Viewer window”.
★ To cancel the image loading process, click 1 Select a study from the list and [Edit/
[Cancel]. Re-Study] on the right-hand side of the
window.
- If no disk is mounted -
The following message is displayed:
Preparing to retrieve images
• Add Menu
You can register additional exposure menus.
• Delete/Inactivate Menu
You can delete menus that are no longer needed or
inactivate read images to configure a re-exposure
operation.
For the operating instructions, see “3.3.4
Editing Study Information”.
• Change Menu
Changes the menu for a read image and
reprocesses the image according to the new menu. The selected studies are locked and are marked
If the image has been rotated or reversed, its
status is updated according to the new menu.
“Locked” in the Lock column of the list, so they
cannot be deleted.
6
• Relocate Display Order
Changes the order of appearance of a menu in the
HINT
If menus for an already verified study (which
is in the Verified study status) are changed or
relocated, images in the study is replaced or its
patient information is edited, the study returns
to the Completed study status. Verify the study
again.
4
Using a Study Utilities
WARNING
Before saving study image in PDI disk,
write the following information on
★ To output images with annotations embedded,
the surface of the PDI disk to prevent select [Yes] in the “Output images with
confusing with study image for other annotations embedded”.
patient ID.
CAUTION
• Patient name
• Birth date Progress is displayed when write to
• Disk creation date CD-R or DVD-R disk starts. Do not press
• Study date (write all if more than one) the drive eject button or use any other
• Facility name program while writing. Otherwise, write
may not complete properly.
Note that once write starts, it cannot be
6
HINT
canceled.
The processing takes time when a large quantity
1 Select a study in the Verified study status - When the following message is displayed -
from the list and click [Create PDI] on the No disk is inserted.
right-hand side of the window.
Multiple studies can be selected. Unable to write to this disk.
1 Select a study in the Completed or Verified LOW : Gives higher priority to other print
study status from the list and click [Film operations.
Print] on the right-hand side of the window. MEDIUM : Print with the ordinary priority.
Multiple studies can be selected. HIGH : Starts print immediately.
6 Select [Output].
HINT 6
Although images that have been saved with
“Processed” being selected can be reloaded into
the FCR PRIMA Console, the images thus loaded
Uncompressed :
Saves images uncompressed.
JPEG Lossless :
Saves images compressed. The image file size
shrinks. Compressed images can fully be restored.
1
6 Select a study in the Verified study status
from the list and click [Storage Disk] on the
right-hand side of the window.
Multiple studies can be selected.
Using a Study Utilities
5 Select [Output].
The “Setting for DICOM Transmission dialog” ★ To cancel the transmission to the DICOM server, 6
appears. click [Cancel] to close the “Setting for DICOM
Transmission dialog”.
2 Select a destination.
4) Disk operations
6
Using a Study Utilities
6) U
pdate,
Image Viewer
3) Page navigation
6
exposure was started. you can edit, delete and move registered search
Study Time Indicates the time at which the queries. You can also set the search queries under
exposure was started.
which a search launches right after the startup of
Study Indicates the name of the study menu
FCR PRIMA Console.
Description that has been created with the User
: First page
: Previous page
: Next page
: Last page
[Image Viewer] :
Displays images in the “Image Viewer window”.
For details, see “6.2.3 Launching the Image
Viewer”.
[Registering studies stored on the disk] :
The “Load Disk Images dialog” appears, enabling
you to load images from a disk into the FCR PRIMA
Console system.
For details, see “6.2.4 Loading Images from a Disk
into the FCR PRIMA Console System”.
5) List operations
You can select and deselect all studies in the
current page, and update the display of studies in
6 the study list.
Selecting this portion hides the operational menu.
Using a Study Utilities
[Select All] :
Selects all the studies in the current page.
[Unselect All] :
Deselects all the studies in the current page.
[Update] :
Updates the display of the study list to the latest
status.
1 Select a study from the study list and click • Rotate or reverse the images,
[Image Viewer] in the lower right-hand • Save FCR image processing parameters,
corner of the window. • Edit markers,
Multiple studies can be selected. • Update study information, and
• Verify or suspend studies.
HINT
- If any of the following messages is displayed:
You can do the same by:
Study importing has completed.
• Selecting a study in the study list and clicking ● When importing from other than a disk for
1 Select a study from the list and click When the image loading process completes, the
[Registering studies stored on the disk] on message Study loading completed is displayed.
the right-hand side of the window. Click [OK].
Multiple studies can be selected. To manipulate the studies thus loaded, search 6
through the study utilities for target studies stored on
the PC.
6
Using a Study Utilities
• IP Use Frequency
Confirms the IP use frequency for image reading.
The “Main Menu dialog” appears. • User Settings (for General Users)
Edits information about the currently logged-in
user.
• Disk Operation
Replaces disks, configures disks for writing, and
restores disks or the database.
HINT
Guidance is displayed in the “Client Utilities
dialog” to lead the user through operating
procedures.
Guidance
7
Configuring by Client Utilities
HINT
Items other than ID and Sex are optional.
7
The “Patient Info Maintenance dialog” appears.
3 Enter ID.
An ID is a key used to identify patient information to
Configuring by Client Utilities
6 Enter DOB.
Enter a date of birth according to the sample entry.
7
The patient information is registered in the FCR
PRIMA Console database and the entry fields are
cleared. Continue to register additional patient
7
★ To specify a date, click the lower radio button and
Configuring by Client Utilities
6 Edit ID.
An ID is a key used to identify patient information to
FCR PRIMA Console. IDs of patient information
already registered in the FCR PRIMA Console
database cannot be used.
5 7
• ID
The patient information that completely matches
the ID as entered is located.
• Name
The patient information whose prefix matches the
Name as entered is located.
• Sex
Select from a pulldown menu. The patient
information that matches the selected Sex is
located.
• DOB
Enter a date of birth according to the sample
“Patient Info Maintenance dialog” appears.
entry to the right of the entry field. The patient
information that completely matches the DOB as
entered is located.
HINT
The data is displayed in ascending order of ID.
To change the display order, select the header of
the field to be used as a reference.
Each time you select the header, the display order
toggles between ascending (▲) and descending
(▼).
HINT
The data is displayed in ascending order of [IP
Use Frequency].
To change the display order, select the header of
the field to be used as a reference.
7
Each time you select the header, the display order
toggles between ascending (▲) and descending
8 Enter a comment.
HINT
Enter a comment within 50 characters. 7
Account names are information recorded in the
log. Do not register account names that pertain to
(1) Enter a new password in the New Password. Registration dialog” closes.
Enter a password within 20 characters. The
password is indicated as **** in the dialog.
HINT
Auto-Verify users are configured using the User
Utility.
7
2 Click [User Info Maintenance].
4 Click [OK] in the confirmation dialog.
Configuring by Client Utilities
The “Registered User List dialog” appears. The selected user information is deleted.
HINT
When registering a user, select a user group
suited to the user.
Image
4 Click [View].
7
★ To exit the User Settings (for Administrator), click
[Close].
Configuring by Client Utilities
3 Set an image.
(1) Click [Image Select].
7
Configuring by Client Utilities
• Comment
• Image
• Password
Perform Steps 2 t o 4 to make necessary edits
to the user information.
HINT The selected image is displayed in the “Modify
• Edits made to the User Selection box will currently logged in user information dialog”.
appear as shown below.
Image
Comment
• The entry of Comment and Password is
optional. When the entry of Password is
omitted, this user can log in without being
required to enter the password.
HINT
The data is displayed in descending order of Date
of Operation.
To change the display order, select the header of
the field to be used as a reference.
Each time you select the header, the display order
toggles between ascending (▲) and descending
(▼).
7
★ To exit the login history check, click [Close]. Then,
click [Close] in the “History Info dialog”.
5 Click [Search].
]
The “Study Access History dialog” appears.
HINT
The data is displayed in descending order of Date
of Access.
To change the display order, select the header of
the field to be used as a reference.
Each time you select the header, the display order
toggles between ascending (▲) and descending
(▼).
HINT
The data is displayed in descending order of Date
of Image Transfer.
To change the display order, select the header of
the field to be used as a reference.
Each time you select the header, the display order
toggles between ascending (▲) and descending
(▼).
Search for studies for which you want to view a
history of image transfer and list them.
4 Click [Search].
HINT
Configuring by Client Utilities
[Replacing Disks]
Replaces the disk displayed in the “Disk
Operation dialog”.
The “Disk Operation dialog” appears. 1 Place a disk in the disk drive tray and close
the tray.
★ To cancel the database restoration process, Closes the “Disk Operation dialog”.
click [Cancel] to close the “Database Restoration
dialog”.
Customize window is opened from the “Main 1 Select - [Customize] from the “Main
window”. window”.
The initial display for the “Main window” can be The shortcuts for the exposure menu can be
set. set.
Exposure menus not set to shortcuts can be
selected from [Others].
HINT
This item is set for the entire system.
ID input
Exposure · Operations
Patient List
Customizing Main Window
3 Click [OK].
8
Customizing Main Window
1 In the “Patient List window”, check the check 4 Set the maximum number of the list to be
box of items to be displayed in the patient displayed in the patient list.
list. Uncheck the check box of items not to
be displayed in the patient list.
The items are displayed left aligned in the patient list
according to the order of items on the customize
window.
8
Select either of the up or down button to change the
value.
Customizing Main Window
7 Click [OK].
1 Click [Select] in the “Exposure · Shortcut 4 Click [OK] at the right bottom of the window.
menus window”.
The selected menu is added.
5 Click [OK].
3 Click [OK].
8
Customizing Main Window
4 Click [OK].
1 Select an operation item to be added to a 1 Select the operation item to be removed from
block from the operation item list on the each block in the “Study list · Operations
left side of the window in the “Study list · window” and select [Remove].
Operations window”.
2 Click [OK].
3 Click [OK].
2 Click [OK].
8
Customizing Main Window
For details, see “9.4 Setting of Printing, For details, see “9.13 Setting Automatic
Saving and Transmitting Parameters”. Logoff”.
• Auto-Determine settings • History-Info Consistency Check
Set up automatic verification of the completed Check that the history of processes such as FCR
study. PRIMA Console login/logoff, access to studies,
image storing or printing, and backup or restoring,
For details, see “9.5 Setting the Time to
has not been altered.
Automatically Verify a Study”.
• Image Deletion Settings For details, see “9.14 History Information
Change the time to automatically delete images Consistency Check”.
read from a disk. • Marker Setting
Register and delete user markers.
For details, see “9.6 Changing the Time to
Automatically Delete Images Loaded from a For details, see “9.15 Setting Markers”.
Disk”.
• Setting Preset Settings
• Backup Notification Settings Assign preset settings to the keypad keys.
Set up to display a message prompting you to
make the backup. For details, see “9.16 Setting Preset Settings”.
9
Display Settings Exposure/Study Menu
column column
Setting User Utilities
Exposure Parameter 1
If the selected code has already been assigned to Displays the Exposure Parameter 1 tab.
another exposure menu, this exposure menu will be The setting items are described below.
displayed in the Related Menu field. In this case,
select another unused code so that the Related
1 Select whether or not to flip image at the
Menu field becomes empty.
time of exposure. Leave [Frontal Image]
selected by default for normal exposure.
< What is the MPM code? >
Select [Frontal Image] again (deselection)
The MPM code is a four-digit code number to
to flip image (to back image) for exposure.
manage the exposure menus. This code determines
Select an reversed display of the image.
EDR (function for automatic density and contrast
correction) and the image processing parameters.
HINT
You can do the same by entering up to 10
characters directly using the keyboard.
❍: Exclusive control
HINT
In the “Color window”, you can create a unique
color by adjusting a color selected from the Basic
Colors. You can add this newly created color
to the Custom Colors pallet by clicking [Add to 1 Select Image Position.
Custom Colors] to make it selectable as a display
Select an image to which the image processing
color for the exposure menu name.
parameter setting is to be applied.
(2) Select the desired color either in the Basic
Colors or Custom colors pallet in the “Color
window” and then click [OK].
The “Color window” and “Display Color Setting
dialog” close. An exposure menu name in the
Exposure Menu Name field on the “Exposure 2 Select whether to enable GP/RP or MFP
Menu Setting dialog” will be displayed by the (optional).
selected color.
Select GP (Gradation Processing)/RP (Spatial
★ To cancel the operation, click [Cancel] and close Frequency Processing) or MFP (Multi-objective
the “Color window”. Frequency Processing).
GS
HINT Characteristics Parallel shift of the gray level curve.
Changes the display density of the
DRC has priority over TAS in image processing.
entire image.
This means that setting DRC to ON automatically
sets TAS to OFF. Value range -1.44 to 1.44 (2 decimal places)
RT
6 Change the parameter values.
Characteristics Shape of the enhanced curve
corresponding to the density of the
★ To restore standard parameters set as default (by
image.
FUJIFILM Corporation), select Return to Default.
By changing this parameter, the
Please note that this button is disabled when enhancement level can be adjusted
the standard exposure menu set as default (by within a certain density region.
FUJIFILM Corporation) are not displayed in the
Related Menu field. Value range F, P to Z (12 types)
Characteristics Enhancement level of one-dimensional The parameters for GP (GA, GT, GC, GS) are
unsharpness mask processing.
the MFP gradation processing parameters. For
Changes the enhancement level of
processing in accordance with the characteristic, refer to the description for GP.
specified value. ● FNC (Frequency Noise Control Processing)
Value range 0.0 to 9.9 (1 decimal place) parameter (optional)
10 to 16 (no decimal places) FFC
Characteristics Parameter used to determine the
9 ● MFP (Multi-objective Frequency Processing) scale of the noise control processing
parameter (optional) according to the X-ray exposure dose.
A is for general exposure and is for linac
MRB
graphy.
Setting User Utilities
Standard Density: 9
Print or save the image with low pixel density.
This reduces the size of the stored image file.
Setting User Utilities
High Density:
Print or save the image without changing the pixel
density.
FR : Reading 5 lines/mm
SR : Reading 10 lines/mm
GR : Reading 10 lines/mm
2 Select whether to enable/disable GPR The color of the selected exposure menu reverses.
Processing (optional).
Select ON (enable) or OFF (disable) for GPR (Grid 3 Select [Edit].
Removal Processing ).
5
Setting User Utilities
HINT
Please note that you cannot restore the standard
exposure menu set as default (by FUJIFILM
Corporation) once you delete it. (User-created
exposure menus cannot be restored either and
need to be re-created.)
9
The color of the selected exposure menu reverses.
Setting User Utilities
3 Select [Delete].
9 .
2
Setting User Utilities
Select [New].
HINT
Dragging it across [ v (Previous) ] or [ w (Next) ]
will display the previous or next film images.
★ To cancel the operation, click [Cancel]. The dialog Setting User Utilities
returns to “Study Menu Setting dialog”.
The selected image is exchanged with the image
currently displayed at the desired position.
(Clear) HINT
Dragging it across [ v (Previous) ] or [ w (Next) ]
Clears the selected image from the layout of will display the film images of the previous or next
images. film.
The procedure to clear an image from the
layout is explained below.
[Cancel]
Cancels the operation and closes the “Film
9
Print Position Setting dialog”.
[OK]
Setting User Utilities
HINT
The color selected in the “Display Color Setting
dialog” is applied to the display of study menu
name in the Study Menu Name field when clicking
[OK] in the “Study Menu Setting dialog”.
The selected display color is not applied to the
study menu, if [Cancel] is selected in the “Study
Menu Setting dialog”.
9
● Setting a color not available in the “Display
Color Setting dialog”
The procedure to set a color not available in Setting User Utilities
the “Display Color Setting dialog” is described
below.
(1) In the “Display Color Setting dialog”, select
[More Colors...].
The “Color window” appears.
HINT
Please note that you cannot restore the standard
study menu set as default (by FUJIFILM
Currently registered study menus are displayed. Corporation) once you delete it. (User-created
study menus cannot be restored either and need
2 Select a study menu to be edited. to be re-created.)
3 Select [Edit].
Currently registered study menus are displayed.
4 Edit settings.
Edit the settings of an existing study menu using the
same procedure as creating a new study.
9
The color of the selected study menu reverses.
5
Setting User Utilities
HINT
Up to 16 display groups can be set.
When 16 display groups have already been set,
no further new display groups can be created. In
this case, you can delete display groups no longer
needed to create new groups.
The color of the selected tab reverses. 2 Select [Edit Display Group].
9 [Cancel].
The selected display group is deleted.
4 Click [OK].
1 Select the tab of the Display Group whose 1 Select the tab of the Display Group which
position you want to change. you want to exclude from display and click
[Change Display].
9
Setting User Utilities
A B B A
Move A to B Positions of A and B
position. are exchanged.
HINT
Dragging over the Change Page button allows
you to change the position across pages.
A B B A
Move A to B Positions of A and B
position. are exchanged.
HINT
Dragging over the Change Page button allows
9 you to change the position across pages.
HINT
Destination for data transfer is selected by
default.
The Timing of output is set as the Upon
On the CR Study tab, the setting tabs for Disk, verification of a study.
Printer, and DICOM Transmission can be displayed It cannot be changed.
by clicking on the respective tab.
9
Setting User Utilities
HINT
The Timing of output is set as the Upon
verification of a study.
It cannot be changed.
[OK]
9
Saves the settings for printing, saving,
and transmitting and closes the “Setting
Destination for Automatic Image Output Setting User Utilities
dialog”.
- When the following message appears -
Set up again so that CR study images are
stored on the image storage disk, printing film or
DICOM server.
9 HINT
Destination for data transfer is selected by
default.
Setting User Utilities
[Cancel]
The “Automatic verification of a study dialog” ★ To verify a study after a certain time elapsed
appears. after completing an exposure, select the Study is
verified after the set time followed by exposure
check box.
2 Select whether to enable/disable automatic ★ Set a time between 5 and 720 minutes. Enter only
verification. integral numbers.
The study will be automatically verified within a
★ To verify a study automatically after completion,
period equivalent to the entered time up to double
select the Automatically verifying a study check
that time at maximum.
9 box.
The completed study is automatically verified. Set
For example, if 30 is specified for the Time
required for automatic verification of a study field,
the time to automatically verify a study as
the study will be automatically verified in 30
described in 3 .
Setting User Utilities
HINT
To automatically verify the study, select the
auto verify user using the client utility. If you
do not select it, the study will not be verified
automatically.
HINT
While the study is shown on the viewer, the study is not
verified automatically even if automatic verification of
study is specified.
The study is not verified automatically while the
information about the study whose status is Finish
Exposure is being modified. When display on the
viewer is closed or the study information is being
modified, the study will be verified automatically
after lapse of the predetermined time.
9
Setting User Utilities
3 Click [OK].
1 Select [Backup Notification Settings] in the ★ To disable the backup prompt, deselect the
“Main Menu dialog”. Issuing a backup notification check box.
In this case the backup prompt will not be
displayed and you will have to make backups
regularly on your own.
Proceed to Step 4 .
4 Click [OK].
1 Select [Prefix Display Settings] in the “Main 3 Enter a prefix in the A title of respect of a
Menu dialog”. patient to be displayed field.
Enter up to 4 characters.
4 Click [OK].
HINT
1 Select [Image-incidental info Display
Information associated with the image can be set
Settings] in the “Main Menu dialog”.
individually by CR study or other studies.
7 Enter Prefix/Suffix.
Enter up to 15 characters in each field. For example,
if you input the following, the Patient ID 1234567890
will be displayed as Patient ID [1234567890].
Proceed to Step 8 .
9
5 Select Classification.
9
Setting User Utilities
• M
ake sure to enter the Patient ID. If the Patient ID
is not entered, the patient information in that row
The “Save As window” appears.
will not be registered.
(2) S
pecify the storage folder of the file in the
• <
Tab> and <Enter> refer respectively to the Tab
Save in.
and Enter key on the keyboard.
(3) F
or Save as type, select All.
• P
ress the Tab key on the keyboard each time (4) E
nter the extension .csv after the File name.
between Patient ID, Patient Information Name,
Patient Name, Sex and Date of Birth. A space
(Example: Patient.csv)
(5) S
elect [Save].
9
entered with the Space key on the keyboard will
cause an error when importing.
Setting User Utilities
• A
fter Date of Birth, press the Enter key to insert
a line feed. Enter the patient information for 1
individual in 1 line.
• T
o leave some items blank, press the Tab key on
the keyboard repeatedly.
Example: Leaving Patient Name
blank1234567890<Tab>Brown
James<Tab><Tab>M<Tab>20020705
1 2 3 4 5
1 Patient Patient Patient Sex Date of
ID information name birth
name
2 Patient Patient Patient Sex Date of
ID information name birth
name
3 Patient Patient Patient Sex Date of
ID information name birth
name
4 ••••
- If the following message is displayed -
• M ake sure to enter the Patient ID. If the Patient ID The selected file type does not support
is not entered, the patient information in that row workbooks that contain multiple sheets.
will not be registered.
• Enter the patient information for 1 individual in 1 Click [OK].
line. - If the following message is displayed -
• To leave some items blank, leave the …txt may contain features that are not
corresponding cell blank. compatible with Text (Tab delimited). Do you
Example: Leaving Patient Name blank want to keep the workbook in this format?
9
become unusable. Are you sure you want to
change it?
Setting User Utilities
9
Setting User Utilities
Insert the disk for consistency check into the The consistency check starts as soon as the storage
disk drive beforehand. disk is detected.
The “Consistency check - Consistency check in
1 Select [Media Consistency Check] in the progress dialog” is displayed for checking the drive
“Main Menu dialog”. name, disk name and the progress of the
consistency check.
9
Message Integrity of image information not
verified.
Setting User Utilities
4 Click [OK].
9
2 After verifying the history, click [OK].
Setting User Utilities
The “History-Info Consistency Check dialog” closes.
The “Add or Remove Marker dialog” appears. 3 Select the marker to register.
4 Click [Open].
The registered marker displays on the “Add or
Remove Marker dialog”.
9
5 Click [OK].
Setting User Utilities
4 Click [OK].
9
3 Select [Remove Marker].
9
Setting User Utilities
3 Click [OK].
10.1 Troubleshooting
When the equipment does not operate normally, see the relevant page to take remedial actions
according to the symptom shown below.
Should you encounter any trouble you cannot handle, contact our official dealer.
Troubleshooting
PC is Inoperative 10.3
Is the power cord Check that the power cord is firmly connected to the electrical outlet.
disconnected? Otherwise, your PC may not start. In such case, turn off the main power
switch, firmly insert the power cord in the outlet, and turn on the power
switch once more.
HINT
When AC mains power is supplied, the POWER lamp illuminates
green.
Is the monitor switch turned If the monitor switch is not turned ON, nothing will be displayed on the
ON? screen even if the PC power switch is turned ON. Turn ON the monitor
switch.
HINT
When monitor power is supplied, the monitor lamp illuminates blue.
10
Troubleshooting
Is the PC frozen? If nothing happens when you press a key or move the mouse, your PC is
frozen. Take the following actions:
(1) Press the Delete key while holding down Ctrl and Alt keys to display the
“Windows Task Manager window”.
Select the Application tab to display a list of currently running
applications. If the Status column of the displayed application list shows
No response, select that application and click [Terminate task]. Restart
your PC because Windows may be unstable.
HINT
If the mouse is disabled, use the Tab key and Enter key.
HINT
All current data are lost when you force shutdown your PC.
HINT
All current data are lost when you force shutdown your PC.
10
Troubleshooting
Is there a cassette in the Image FCR PRIMA Console cannot quit if there is a cassette remaining in the
Reader connected to the FCR Image Reader.
PRIMA Console? Remove the cassette from the Image Reader and quit FCR PRIMA
Console.
Is the PC frozen and inoperative If the PC cannot be shutdown, keep pressing the power button for at least 4
when you try to shutdown? seconds. The power will turn off.
If the power still does not turn off, pull out the power cord to turn OFF the
power.
HINT
If the PC could not shutdown by normal shutdown procedure,
SCANDISK will start automatically next time you turn on the PC.
This will take some time, but normal screen will eventually appear if
there is no problem.
10
Troubleshooting
Is the event indicator blinking? If the event indicator is blinking, there is an error.
Check the cause of error as follows and take the appropriate action:
Event indicator
10
(3) Check the detailed Information and take appropriate action.
Troubleshooting
An example of an error message displayed on the “Warning dialog” and a description of each
display area are given below:
(1)
(2)
(3)
(1)
(2)
(3)
Name Description
(1) Title Display Field The title of an error is displayed.
(2) Message Display Field Displays an error code and a message prompting the user to perform an operation.
(3) Selection-menu Display Selection menu is displayed. Select the button of the operation to be performed in
10 response to the message.
Troubleshooting
Streak
■ Remedial action
If a streak appear on the image, clean the scanner and IP.
• Scanner cleaning
Clean the scanner using the Utility.
For details on scanner cleaning, see “5.2 Confirming Image Reader Status”.
• IP cleaning
Remove dirt from the surface of the IP and then wipe with a dry cloth.
If the dirt cannot be removed by dry wiping, wipe the surface using a small amount of dehydrated
ethanol.
The IP should be dried completely before use.
10
HINT
• Be very careful not to damage IPs.
• Never use plastic erasers and other solvents.
Troubleshooting
The following message appears Click [OK] to display the “Login window” and log in once more.
when you try to operate the FCR
PRIMA Console after it is logged For instructions about how to log in, see “2.4.1 Starting Up FCR
off automatically.
PRIMA Console”.
Not logged on
Log in and then re-execute.
* Contact our official dealer if the
same symptom recurs even
after operation has been retried.
10
Troubleshooting
Did you click [X] on the title bar If you accidentally closed the client application, double-click the client
and close the window? application (FCR Web) icon to start it once more.
For details, see “Launching the Client Application with Only the
Server Application Running” in “2.4.1 Starting Up FCR PRIMA
Console”.
Is the window hidden behind Window hidden behind another window is displayed minimized on the
another window? taskbar (Example: ). Click (Example: ) on the
taskbar to redisplay the window.
10
Troubleshooting
Did you try to remove the disk On the PC’s control drive, be sure to eject the disk from FCR PRIMA
by pressing the disk drive Console because the eject button is locked while FCR PRIMA Console is
EJECT button? in operation.
If you use an application other than the FCR PRIMA Console to eject the
disk, the PC can freeze or the disk can become unusable.
For instructions about how to remove the disk, see “2.5.3 Ejecting
Disk”. If the disk cannot be removed, terminate FCR PRIMA Console. If
you cannot terminate it, see “10.3 PC is Inoperative”.
For details on disk handling, also see “2.5.3 Ejecting Disk”.
10
Troubleshooting
Did you display the “Toolbar Display the “Display Settings dialog” as follows and set the toolbar to
Settings dialog” from the display.
“Image Viewer window” and
uncheck each toolbar displayed (1) Right-click inside the image pane of the “Image Viewer window” to
on the “Display Settings dialog” display the shortcut menu.
of the Display Settings tab?
(2) Click [Toolbar settings] - [Toolbar settings] to display the “Toolbar
Settings dialog”.
(3) Select the Display Settings tab to display the “Display Settings dialog”.
(4) Check the toolbar to display and click [Setup].
HINT
Click [Toolbar settings] - [Reset Toolbar] to display the toolbar in the
initial state.
10
Troubleshooting
10
Troubleshooting
Is the search requirement for Confirm the search requirement display column for patient list.
patient list selected properly?
For details on searching method, see “3.4 Using a Patient List” and
“6.1.2 Searching for studies”.
10
Troubleshooting
Is the checkbox selected at the Select the checkbox at the bottom right-hand of the image displayed on the
bottom right-hand of the image “Image Viewer window” and then save, print or transmit the image.
you are trying to print, save, or
transmit from the “Image Viewer For details, see “4.5 Printing or Saving Images”.
window”?
10
Troubleshooting
■ Action
• When “Web page unavailable while offline dialog” is displayed
Check that the LAN cable is firmly connect to the PC and then select [Connect].
• When [Stay Offline] is selected on “Web page unavailable while offline dialog”
Close all FCR PRIMA Console windows and select [Start] - [Programs] - [Internet Explorer] to restart Internet
Explorer. Select and uncheck [File] - [Work Offline] from Internet Explorer and then restart FCR PRIMA
Console.
For information on how to start FCR PRIMA Console, see “2.4.1 Starting Up FCR PRIMA Console”.
10
Troubleshooting
As for the daily inspection, make certain that the system starts up normally and communicates
properly with devices connected to the system. Be sure to clean the monitor daily.
Also, back up the patient information and other data at least once a week.
Use the Maintenance Record when performing daily inspection and maintenance.
Enter the following items in the Maintenance Record:
HINT
When performing backup, be sure to check that no storage disk is inserted in the drive.
11
Daily Inspection and Maintenance
You can continue using the FCR PRIMA (1) Double-click the server application icon placed
Console after backup completes. in the task tray.
The “Select Server Application Behavior dialog”
For details, see “2.4.1 Starting Up FCR PRIMA
appears.
Console”.
Start the Backup Utility by one of the following (2) Select [Backup Utility] and click [OK].
two methods:
★ If you do not want to start the Backup Utility, click
[Cancel] to return to FCR PRIMA Console.
• Start from backup notification message
The backup notification message appears when A message is displayed and then the “User
backup notification is set in the User Utility. authentication dialog” appears.
For details, see “9.7 Setting the Backup
Notification”.
• Start from server application icon in the task
tray
HINT
You can also start FCR PRIMA Console by the
following method:
(1) Select [Start] - [Programs].
★ Click [Backup] to backup the patient information, (2) Select [Fujifilm] - [FCR Application].
setting information, and log information.
For details, see “11.2.2 Backup”.
★ Click [Restore] to restore the database.
For details, see “11.2.3 Restore”.
★ Click [Collecting logs for maintenance purpose]
in order to save the log information used for
maintenance requested by our official dealer.
* Perform only when requested by our official
dealer.
For details, see “11.2.4 Collecting
Maintenance Log”.
11
Daily Inspection and Maintenance
11
3 To delete the log information after backup
completes, check Delete log information
Daily Inspection and Maintenance
8 Click [OK].
6 Select a folder and then click [OK].
Returns to “Main Menu dialog”.
★ To cancel folder selection, click [Cancel] and
HINT
close the “Browse For Folder window”.
When backup completes, the BackupInfo.ini file
containing the backup status is saved.
BackupInfo.ini file contains the FCR PRIMA
Console version, backup date and time,
application name, backed up file name, and
backed up file version. 11
9 Remove the disk (or USB flash drive)
Daily Inspection and Maintenance
11 Click [Cancel].
HINT
When backup completes, replace with storage
disk.
11
Daily Inspection and Maintenance
1 Click [Restore] in the “Main Menu dialog”. The “Restore dialog” appears.
2 Click [Browse].
6 Click [Start].
11
Daily Inspection and Maintenance
HINT
• Perform this operation only when requested by
our official dealer.
• When you collect the log for maintenance, a
partial log file (Study.txt) containing only the
log data related to personal information is
created on the desktop, allowing you to check
whether personal information has been deleted 3 Click [Start].
by verifying its content. After completion of the
★ To cancel maintenance log collection, click
check, be sure to delete the Study.txt file.
[Cancel] to close the “Collecting Logs for
• If the log is to be collected by our official dealer,
Maintenance Purpose dialog”.
you do not need to perform log collection. In
this case, log will be collected after obtaining
your agreement.
The “Select Folder dialog” appears.
1 Click [Collecting logs for maintenance
purpose] in the “Main Menu dialog”.
11
Daily Inspection and Maintenance
● Check
Check Description Frequency User Check Result
/
/
Does the system start up
Daily /
normally?
/
/
/
Is the normal communication /
possible between the Daily /
connected equipment? /
/
● Cleaning
Equipment to be Cleaned Frequency Date of cleaning
/
/
Monitor Daily /
/
/
● Backup
Backup Item Frequency Backup Date
/
/
Patient information Weekly /
/
/
/
/
Setting information Weekly /
/
/
/
/
Log information Weekly / 11
/
/
Daily Inspection and Maintenance
■ PC
PC Power Supply Conditions
(DELL Optiplex 780DT) Input voltage : 115/230VAC
Input current : 5.0A / 2.5A
Frequency : 50Hz / 60Hz
Environmental Conditions
(1) Operating
Temperature : 10°C to 35°C
Humidity : 20%RH to 80%RH (No dew condensation)
(2) Not operating
Temperature : -40°C to 65°C
Humidity : 5%RH to 95%RH (No dew condensation)
Spec
Processor : Intel(R) Core(TM)2 Duo E7500/2.93GHz
Chipset : Intel Q45 Express Chipset w/ICH10DO
Memory : 2GB DDR3 Non-ECC SDRAM
System Board
Connectors : PCI Slots x2
Operating
System : Windows Vista(R) Business SP1 (32 bit)
Video Controller : Integrated Intel GMA 4500
External Ports : USB 2.0 x8 (2 Front, 6 Rear)
Network Adapter : 1000 BASE-T/100 BASE-TX/10 BASE-T
HDD : 250GB SATA 7200 RPM HDD
Environmental Conditions
(1) Operating
Temperature : 5°C to 35°C
Humidity : 10%RH to 80%RH (No dew condensation)
(2) Not operating
Temperature : -20°C to 60°C
Humidity : 5%RH to 90%RH (No dew condensation)
Environmental Conditions
(1) Operating
Temperature : 0°C to 35°C
Humidity : 30%RH to 80%RH (No dew condensation)
(2) Not operating
Temperature : -20°C to 60°C
Humidity : 30%RH to 80%RH (No dew condensation)
Environmental Conditions
(1) Operating
Temperature : 0°C to 35°C
Humidity : 30%RH to 80%RH (No dew condensation)
(2) Not operating
Temperature : -20°C to 60°C
Humidity : 30%RH to 80%RH (No dew condensation)
Environmental Conditions
(1) Operating
Temperature : 0°C to 35°C
Humidity : 30%RH to 80%RH (No dew condensation)
(2) Not operating
Temperature : -20°C to 60°C
Humidity : 30%RH to 80%RH (No dew condensation)
Environmental Conditions
(1) Operating
Temperature : 0°C to 35°C
Humidity : 30%RH to 80%RH (No dew condensation)
(2) Not operating
Temperature : -20°C to 60°C
Humidity : 30%RH to 80%RH (No dew condensation)
Standard Monitor
(1) PC
Appendix A Main Specifications
(Unit : mm)
399 353
114
* The illustration of the FCR PRIMA Console (CR-IR 391CL) may differ from the actual external view.
897N101440
Insert the PDI disc into the drive and start the Viewer. 3. Double-click "INDEX.HTM" and then click "View Image".
The viewer application operates in the following environment.
● OS : Windows XP, Windows VISTA, Windows 2000
● CPU : Intel Pentium 4 or better or equivalent CPU
● RAM: 512 MB or more
(1 GB or more in case of Windows VISTA)
PDI Disc Image 1. Insert the PDI disc into the drive.
Viewer starts automatically.
Display Method * When the Viewer does not start automatically,
proceed to Step 2.
2. Display the PDI disc contents from My Computer.
Viewer starts.
* If a security dialog is displayed before Viewer starts,
click "Run" or "Execute".
1 2 3
■ Select the image to be displayed. ■ Buttons on the study list window ■ Image display window
Select the image to be displayed study images can be confirmed.
on the study list window of Viewer.
A
2 1
●Initialize image button ●Previous page button ●Change window pattern button
When this button is clicked, the selected When this button is clicked, the previous When this button is clicked, the change
image returns to the state before window page is displayed. The previous page shows window pattern window is displayed When
level change. In case of zoomed-in, the image before the presently displayed the window has been displayed, drag the
zoomed-out, or panning display, the display image. This button cannot be clicked when mouse to the position of the pattern to be
returns to the previous state. there is no previous page. changed and click there.
●Previous image button ●Next page button * For change to a 3x3
When this button is clicked, the previous When this button is clicked, the next page is display, click at the
image is displayed. If there is no previous displayed. The next page shows the image position shown right.
image, the button cannot be clicked. following the presently displayed image. This
button cannot be clicked when there is no
next page.
●Next image button
●Switch attribute information display button
When this button is clicked, the next image is
When this button is clicked, the attribute information
displayed. If there is no next image, the
displayed on the image is hidden. When it is clicked
button cannot be clicked.
again, the attribute information is displayed.
●Exit image display button
When this button is clicked, the image display
window is closed and return is made to the
study list window.
10 11 12
897N101440
Appendix B PDI Disk Image Display
It is recommended to use Bucky’s device to remove scatter rays and obtain a high-quality image.
When carrying out exposure using a stationary grid, it is recommended to use a 60 lines/cm grid with
fewer artifacts.
Use a grid whose line direction is parallel to the cassette’s vertical direction (between top and bottom).
Green mark
Top
Bottom
Cassette
portions must be covered by the radiation IP, a split portion(s) may be left unexposed.
field. Guidance as to which to use: PRIEF4S or
PRIEF1S Table Z-1 shows the relationships
among anatomical regions to be radiographed,
radiographic exposure methods, PRIEF4S
and PRIEF1S. Explanations on PRIEF4S
and PRIEF1S are also provided along with
precautions to take with them.
IP center: 25 mm × 25 mm area (black
rectangles in examples above) located in the
center of the IP
Notes on PRIEF
[PRIEF 4S] Image area
[PRIEF 1]
Used to tomography menus for chest,
abdomen, pelvis, upper and lower extremities.
Femur side
1/12
1/4 1/4
b c
a
x
d e
E S
Editing and Saving FCR Saving to a storage disk.................. 3-38, 6-24
Image Processing Parameters............... 4-26 Saving to a Work Disk <Image Viewer>.... 4-57
Editing Information about Saving to a work disk <Study Utilitites>..... 6-23
the Logged-in User................................ 7-24 Searching for Patients <Main window>..... 3-25
Edit Patient Information............................. 3-20 Searching for Studies <Search in Disk>.... 6-29
Ejecting Disk.............................................. 2-16
T
Thumbnail Images and Study Status
([Suspend] [Verify])................................ 4-73
Transmitting studies to
DICOM server.............................. 3-40, 6-25
U
Using Preset Settings................................ 4-31