15EC34T
15EC34T
Measurements play a very important role in every branch of scientific research and engineering. The
whole area of automation is based on measurements. The very concept of control is based on the
comparison of the actual condition and the desired performance. The measurement confirms the validity
of a theory and also adds to its understanding. This eventually leads to new discoveries. Through
measurement a product can be designed or a process be operated with maximum efficiency, minimum
cost and with desired degree of reliability and maintainability.
In the direct comparison method of measurement, we compare the quantity directly with the
primary or secondary standard. For example, if we have to measure the height of a person, we do it with
the help of the measuring tape or scale that acts as the secondary standard. Here we are comparing the
quantity to be measured (height) directly with the standard.
Instrument: It is defined as a device for determining the value or magnitude of a quantity or variable.
Accuracy: It is defined as the closeness with which an instrument reading approaches the true value of
the quantity being measured.
Precision: It is defined as how exactly the result is determined. i.e given a fixed value of the quantity,
precision is a measure of the degree of agreement within a group of measurements.
Sensitivity: It is defined as the ratio of the magnitude of output signal to the input signal or response of
measuring system to the quantity being measured.
Resolution: It is defined as the smallest change in measured quantity that causes a visible change in its
output.
Dynamic characteristics of a measuring instrument refer to the case where the measured quantity
changes rapidly with time.
1. Speed of response
2. Measuring Lag
3. Fidelity
4. Dynamic error
Speed of Response (desirable): It is defined as the speed with which an instrument or measurement
system responds to changes in measured quantity.
Response Time (desirable): It is the time required by instrument or system to settle to its final steady
position after the application of the input.
Fidelity: Fidelity of a system is defined as the ability of the system to reproduce the output in the same
form as the input. It is the degree to which a measurement system indicates changes in the measured
quantity without any dynamic error.
Dynamic error: It is difference between the true value of the quantity changing with time and the value
indicated by the measurement system if no static error is assumed.
**Note: Static error is the difference between the true value and the measured value of a quantity.
The measurement of a given quantity is the result of comparison between the quantity (whose
magnitude is unknown) & a predefined Standard. Since two quantities are compared, the result is
expressed in numerical values.
The unknown quantity under measurement makes its first contact with the primary sensing
element of a measurement system. The sensing elements sense the condition, state or value by taking out
a small part of energy from the measured (the unknown quantity which is to be measured), and then
produce an output.
The output from the primary sensing element may require to be converted to a more suitable form
while saving its information contents. This conversion is performed by the variable conversion element
called transducer.
The function of this element is to manipulate the signal presented to it preserving the original
nature of the signal. Some non-linear processes like modulation, detection, sampling, filtering, etc., are
performed on the signal to bring it to the desired form to be accepted by the next stage of measurement
system.
This element transmits the signal from one location to another without changing the physical
nature of the variable.
This element presents a display record or indication of the output from the manipulation elements to
the person handling the instrument.
1.2 Errors
Error is defined as the difference between the actual value of a quantity and the value obtained by a
measurement.
A study of errors is the first step in finding ways to reduce them. Errors may arise from different
sources and they are mainly classified as shown below,
This type of error occurs due to human mistakes, while reading, recording and calculating
measurement results. For example the observer due to an oversight, may read the temperature as 30.50C
while the actual reading may be 30.20 C, there is 0.30C error in the reading.
Gross errors may be of any amount and therefore their mathematical analysis is impossible. But, the
following precautions can be taken to avoid such errors. They are:
2. More than one reading should be taken for the quantity under measurement preferably by
different observers.
Systematic errors occur usually from the measuring instruments. They may occur because there is
something wrong with the instrument or its data handling system, or because the instrument is wrongly
used by the experimenter.
These errors can be found by conducting repeated measurements under different conditions or with
different equipment and if possible by entirely different method.
These errors are further classified as follows,
Instrumental errors:
These errors arise due to following reasons:
1. Due to inbuilt shortcomings in the instruments.
2. Due to misuse of the instruments.
3. Due to loading affects the instruments.
These errors can be minimised by using the following methods,
1. Measurement procedure must be carefully planned.
2. Correction factors should be adopted after finding the instrumental errors.
3. Instrument must be re-calibrated carefully.
Environmental errors:
These errors arise due to conditions external to the measuring device (e.g. effects of temperature,
pressure, humidity, dust etc.)
These errors can be minimised by using the following methods,
1. Temperature controlled enclosure can be used to avoid temperature variations.
2. The effect of humidity, dust etc. Can be entirely eliminated by sealing the equipment in an
airtight container.
3. By providing shields the instrument can be protected against external magnetic and electrostatic
fields.
Observational errors:
These errors arise due wrong observations. The Observational errors arise due to following
reasons:
4. Mechanical vibrations
3. Poor design.
Arithmetic mean:
It is the ratio of sum of readings taken to the total no. of readings.
Arithmetic Mean = (Sum of readings)/ (Number of readings)
Deviation:
Deviation is the departure of the given reading from the arithmetic mean of the group of readings.
Let the deviation of the first reading 1 be d1 and that of second reading 2 be d2 and so on.
Then the deviation from the mean is expressed as
Average deviation:
It is the ratio of sum of the absolute values of deviations to the no. of readings.
Limiting errors:
Limiting error is used to indicate the accuracy of an instrument. The limiting error (or guarantee error) is
given by the manufacturer to define the maximum limit of the error that may occur in the instrument. For example,
if the resistance of a resistor is given as 50Ω ± 5%, it means that the resistance value falls between the limits 45Ω
and 55Ω.In other words the manufacturer of the resistor guarantees its value lie between 45 Ω to 55 Ω.
Probable errors:
It defines the half-range of an interval about a central point for the distribution, such that half of
the values from the distribution will lie within the interval and half outside. Thus it is equivalent to half
the interquartile range, or the median absolute deviation. The probable error can also be expressed as a
multiple of standard deviation σ,
Variance:
The square of the standard deviation is called variance.
i.e. V= (standard deviation)2
V = σ2
Problem 1:
A circuit was tuned for resonance by eight different students and the values of resonance
frequency in KHz were recorded as 532,548,543,535,546,531,543 and 536. Calculate the arithmetic
mean, average deviation, standard deviation and variance.
=4314
8
= 539.25
2. Average deviation ( ̅ ) = d1+d2+d3+d4+d5+d6+d7+d8
8
=
=5.75KHZ
4. Variance (V) = σ2
= (6.54)2
=42.77KHZ
1.4 Standards
A standard is a physical representation of unit of measurement. Standards have been developed for all the
fundamental units as well as some of the derived mechanical and electrical units.
1. Primary standards:
Primary standards are standards of such high accuracy which can be used as ultimate reference
standards. These standards are maintained by national standard laboratories in different parts of the world.
2. Secondary standards:
Secondary standards are basic reference standards used by measurement and calibration
laboratories. It is obtained by comparing with primary standard. For measurement of a quantity using
secondary standard instrument, pre-calibration is required. Calibration of a secondary standard is made by
comparing the results with a primary standard instrument or with an instrument having high accuracy or
with a known input source.
3. Working standards:
These standards are used to check and calibrate general laboratory instrument for their accuracy
and performance. Working standards are checked against the secondary standards.
4. International standards:
One of the most important standards is the IEEE 4888 digits interface for programmable
instrumentation for test and other equipment standardising.
AC BRIDGES DC BRIDGES
The AC bridges are used to The DC bridges are used to
measure the impedances measure resistances.
consisting of capacitance and
inductances.
The AC bridges use the The DC bridges use the DC voltage
alternating voltage as the as exciting voltage.
exciting voltage.
The four arms of bridge The four arms of bridge consists of
consists of resistors, inductors, pure resistors.
capacitors or their
combinations.
The balancing equation for AC The balancing equation for DC
bridges are bridges is
1. Z1Z4=Z2Z3, for magnitude R1R4=R2R3
balance.
2. ɸ1+ɸ3= ɸ2+ɸ4, for phase
angle balance
Examples of AC bridges are, Examples of AC bridges are,
Maxwell’s bridge, Wein Wheatstone bridge, Kelvin bridge,
bridge, etc,. etc,.
The bridge has four arms together with a source of EMF (V0) and a null detector i.e galvanometer
(G). The unknown resistor is Rx, the resistor Rk is known value, and the resistors R1 and R2 have a
known ratio R2/R1.A galvanometer (G) measures voltage difference VAB between points A and B.
When VAB=0 the bridge is said to be “balanced” and no current flows through the galvanometer (G).
Since VAB=0, the voltage drop from C to A must be equal to the voltage drop from C to B.VCA =
VCB. Likewise, we must have VAD = VBD. So we can write,
(1)
(2) .
(3) .
Thus, the unknown resistance Rx can be computed from the known resistance Rk and the known
ratio R2/R1. The resistors R1 and Rk are called Ratio arms, while the resistor R2 is called standard arm of
the bridge.
08 Hours
Necessity of electrical transducers, selection of a transducer, active, passive, analog and digital
transducers. Strain gauge-principle, gauge factor, features of bonded, unbonded, wire and foil type strain
gauges, load cell. Principle of working & features of capacitive transducer, Hall effect type, LVDT,
thermistor, thermocouple, piezoelectric, proximity sensors, digital optical encoders & PIR sensors.
2. Sensitivity: The transducer should give a sufficient output signal per unit of measured input in order
to yield meaningful data.
5. Stability and reliability: The transducer should exhibit a high degree of stability during its operation
and storage life. Reliability should be assured so that the instrument continues uninterrupted.
6. Insensitivity to unwanted signals: The transducer should be minimally sensitive to unwanted signal
and highly sensitive to wanted signal.
8. Loading Effects: To avoid loading effect, it is necessary that a transducer has high input impedance
and low output impedance.
9. Physical Environment: The transducer selected should be able to withstand any change in
environmental conditions and maintain its output-input relationship.
A transducer, which develops its output in the form of electrical current or voltage without any
auxiliary source, is called active transducer. The energy required for this is absorbed from the physical
quantity which is being measured. Therefore, active transducers are also called as self generating type
transducers.
A transducer, which derives the power required for energy conversion from an external power
source is called as a passive transducer. Therefore, passive transducers are also called as externally
powered transducers.
Examples are linear variable differential transformer (LVDT), thermo-couple, strain gauge,
thermistor etc.
A transducer which converts the input physical quantity into an digital output which is in the form
of pulses is known as digital transducers.
The basic principle of operation of a strain gauge is simple: when strain is applied to a thin metallic
wire, its dimension changes, thus changing the resistance of the wire. The value of resistivity of conductor
also changes.When it is strained its property is called piezo-resistance. Therefore, resistance strain gauges
are also known as piezo-resistive gauges.
R=
Problem 1:
A strain gauge has an unstrained length of 10cm and resistance of 100KΩ. When its length
reduces to 9.9cm, the resistance decreases to 98KΩ. Estimate its gauge factor.
Solution: R= Initial Resistance = 100KΩ.
L= Initial length=10Cm
ΔR= change in initial resistance
=100x103-98x103
=2KΩ
ΔL = The change in length
= 10-9.9 = 0.1 cm
= = 20
1. The wire type strain gauge should have a high value of gauge factor.
3. The wire type strain gauge should have a low resistance temperature coefficient.
4. The wire type strain gauge should not have any hysteresis effect in its response.
6. The wire type strain gauge should have a good frequency response.
The features of foil type strain gauge are similar to those of wire wound strain gauges except
resistance value of foil gauge are available in between 50 and 1000ohms, and maximum gauge current is
about 30mamps.
Figure shows a strain gauge load cell. It consists of a steel cylinder, on which four identical strain
gauges. The gauges R1 and Rg are along the direction of applied load and the gauges R2 and R3 are
attached circumferentially to gauges R1 and Rg. All the four gauges are connected electrically to the four
limbs of a wheat stone bridge circuit.
When there is no load on the cell, all the four gauges have the same resistance
(R1=R2=R3=Rg).under these conditions the A and B terminals are at the same potential, the bridge is
balanced and the output voltage is zero.
Applications:
The capacitive transducer comprises of two parallel metal plates that are separated by a dielectric
material.
The principle of operation of capacitive transducers is based upon the equation for capacitance of
a parallel plate capacitor.
When a magnetic field is applied at right angles to the direction of electric current, an electric field
is setup which is perpendicular to both the direction of electric current and the applied magnetic field.
This phenomenon is called Hall effect.
Fig 2.8.1.1 Current carrying semiconductor bar subject to transverse magnetic field
VH
VH =
Where, RH = Hall co-efficient.
The Hall Effect element is located in the gap, adjacent to the permanent magnet and the field
strength produced in the gap, due to the permanent magnet, is changed by changing the position of the
ferromagnetic plate. The voltage output of the Hall effect element is proportional to the field strength of
the gap which is a function of the position of ferromagnetic plate with respect to the structure.Thus
displacement can be measured by the Hall effect transducer. Very small displacements (as small as 0.025
mm) can be measured by this method.
2.9.1 Construction
Figure 2.9.1.1 shows the construction of LVDT. LVDT consists of a cylindrical former where it is
surrounded by one primary winding in the centre of the former and the two secondary windings at the
sides. The number of turns in both the secondary windings are equal, but they are opposite to each other,
i.e., if the left secondary windings is in the clockwise direction, the right secondary windings will be in
the anti-clockwise direction, hence the net output voltages will be the difference in voltages between the
two secondary coil. The two secondary coils are represented as S1 and S2. An iron core is placed in the
centre of the cylindrical former which can move in to and fro motion as shown in the figure. The AC
excitation voltage is 5 to 12V and the operating frequency is given by 50 to 400 HZ.
As shown in the above figure, an ac voltage with a frequency between (50-400Hz) is supplied to the
primary winding. Thus, two voltages VS1 and VS2 are obtained at the two secondary windings S1 and S2
respectively. The output voltage will be the difference between the two voltages (VS1-VS2) as they are
combined in series. Let us consider three different positions of the soft iron core inside the former.
Null Position - This is also called the central position as the soft iron core will remain in the exact
center of the former. Thus the linking magnetic flux produced in the two secondary windings will be
equal. The voltage induced because of them will also be equal. Thus the resulting voltage VS1-VS2 = 0.
Right of Null Position - In this position, the linking flux at the winding S2 has a value more than
the linking flux at the winding S1. Thus, the resulting voltage VS1-VS2 will be in phase with VS2.
Left of Null Position - In this position, the linking flux at the winding S2 has a value less than the
linking flux at the winding S1. Thus, the resulting voltage VS1-VS2 will be in phase with VS1.
From the working it is clear that the difference in voltage, VS1-VS2 will depend on the right or left
shift of the core from the null position. Also, the resulting voltage is in phase with the primary winding
voltage for the change of the arm in one direction, and is 180 degrees out of phase for the change of the
arm position in the other direction.
The magnitude and displacement can be easily calculated or plotted by calculating the magnitude
and phase of the resulting voltage.
LVDT is used to measure displacement ranging from fraction millimetre to centimetre. Acting as a
secondary transducer, LVDT can be used as a device to measure force, weight and pressure, etc,.
Multiple LVDT‟s are used for measurement of pressure or weight applied by liquid in a tank.
2.10 Thermistors
Thermistors are transducers which are thermally sensitive variable resistance made of
semiconducting materials. A thermistor is a type of resistor whose resistance is dependent on temperature.
It exhibits high negative temperature coefficient of resistance.
The thermocouple works on the principle of Seebeck effect. Seebeck effect is the phenomenon in
which a voltage difference is produced between two dissimilar electrical conductors or semiconductors
due to temperature difference between the two substances. When heat is applied to one of the two
conductors or semiconductors, heated electrons flow towards the cooler one. Direct current will flow
through an electric circuit, if the pair is connected through it. Seebeck effect usually produces small
voltages that are a few microvolts per kelvin of temperature difference at junction.
2.11.1 Advantages
1. Low cost.
2. Small size.
3. Robust.
4. Wide range of operation.
5. Provide fast response.
6. Accurate for large temperature changes.
Piezoelectric effect:
Piezoelectric effect states that when mechanical stress or forces are applied on quartz, crystal,
produce electrical charges on quartz crystal surface. The rate of charge produced will be proportional to
rate of change of mechanical stress applied on it. Higher will be stress higher will be voltage. Certain
crystals namely Quartz, Rochelle salt and tourmaline, which exhibits piezoelectric effect are called
piezoelectric crystals.
The charge at the electrode gives rise to voltage (E) is given by,
E=
= gtP
P = pressure =
2.12.1 Advantages
This uses the effect of eddy (circular) currents to sense the proximity of non-magnetic but
conductive materials. A typical eddy current transducer contains two coils: an active coil (main coil) and
a reference coil. When a coil is supplied with an alternating current, an alternating magnetic field is
produced. If there is metal object in close proximity to this alternating magnetic field, then eddy currents
are induced in it.Therefore impedance of the coil changes thereby changing the amplitude of the
alternating current.
1. Non-contacting measurement.
2. High resolution.
3. High frequency response.
Disadvantages of Eddy current proximity sensor:
Figure 2.13.2 below shows capacitance proximity sensor. Capacitive Proximity Sensors detect
changes in the capacitance between the sensing object and the sensor. The amount of capacitance varies
depending on the size and distance of the sensing object. An ordinary capacitive proximity sensor is
similar to a capacitor with two parallel plates, where the capacity of the two plates is detected. One of the
plates is the object being measured (with an imaginary ground), and the other is the Sensor's sensing
surface. The changes in the capacity generated between these two poles are detected. The objects that can
be detected depend on their dielectric constant.
In the below figure a coil is wound around a core. When the end of a inductive coil is close to a
metal object its inductance changes. This change can be monitored by its effect on a oscillator circuit, and
this change is used to trigger a switch.
Pneumatic sensors:
Pneumatic sensors are used to measure displacement, as well as sense the objects close to it. The
displacement and proximity are transformed into change in air pressure. Figure below shows a schematic
of such a sensor. It comprises of three ports. Low pressure air is allowed to escape through port A. In the
absence of any obstacle / object, this low pressure air escapes and in doing so, reduces the pressure in the
port B. However, when an object obstructs the low pressure air (Port A), there is rise in pressure in output
port B. This rise in pressure is calibrated to measure the displacement.
Encoders have both linear and rotary configurations, but the most common type is rotary.
Absolute encoder where a unique digital word corresponds to each rotational position of the shaft,
The incremental encoder, which produces digital pulses as the shaft rotates, allowing
measurement of relative position of shaft. Most rotary encoders are composed of a glass or plastic code
disk with a photographically deposited radial pattern organized in tracks. As radial lines in each track
interrupt the beam between a photo emitter-detector pair, digital pulses are produced.
Figure above shows the construction of an optical encoder. It comprises of a disc with three
concentric tracks of equally spaced holes. Three light sensors are employed to detect the light passing
through the holes. These sensors produce electric pulses which give the angular displacement of the
mechanical element e.g. shaft on which the Optical encoder is mounted. The inner track has just one hole
which is used locate the „home' position of the disc. The holes on the middle track offset from the holes of
the outer track by one-half of the width of the hole. This arrangement provides the direction of rotation to
be determined. When the disc rotates in clockwise direction, the pulses in the outer track lead those in the
inner; in counter clockwise direction they lag behind.
PIRs are basically made of a pyroelectric sensor (which is shown in figure below as the round metal
can, with a rectangular crystal in the centre), which can detect levels of infrared radiation. The PIR sensor
itself has two slots in it, each slot is made of a special material that is sensitive to IR. The lens used here
consists of two slots that can 'see' out past some distance. When the sensor is idle, both slots detect the
same amount of IR. When a warm body like a human or animal passes by, it first intercepts one half of
the PIR sensor, which causes a positive differential change between the two halves. When the warm body
2.15.1 Applications
2.15.2 Features
Electronic voltmeters: Pros and cons, working of FET input, chopper type DC amplifier voltmeter,
solid-state voltmeter using op-amp, AC voltmeter using full-wave rectifier, Peak responding and true
RMS voltmeters. Ohmmeters series and shunt type. Concept of Calibration of meters.
3. ANALOG METERS:
An instrument which measures and indicates values by means of a continuous scale and a movable
pointer are called analog meters.
“Ammeters”, are connected in series in the circuit whose current is to be measured. “Voltmeters”
are connected in parallel with the circuit whose voltage is to be measured. “Ohmmeters” are used for
measurement of resistance.
3.1 PMMC meters
When current carrying conductor is placed in a magnetic field, a mechanical force acts on the
conductor, if it is attached to a moving system, with the coil movement, the pointer moves over the scale.
Thereby, the basic PMMC movement is called as D‟Arsonal movement. It can be used for D.C
measurements.
Construction:
It consists of a permanent horse shoe magnet with soft iron pole pieces attached to it.
A cylinder shaped soft iron core, is placed in between two pole pieces around which a coil of fine
wire moves wound on a light metal frame. A light pointer attached to the moving coil moves up-scale as
Working:
This meter movement works on “MOTOR PRINCIPLE” (when a current carrying conductor is
placed in a magnetic field, it is acted upon by a force which tends to move it to one side and out of the
field).When the instrument is connected in the circuit to measure current or voltage, the operating current
flows through the coil. Since the current carrying coil is placed in the magnetic field of the permanent
magnet, a mechanical torque acts on it. As a result of this torque, the pointer attached to the moving
system moves in clockwise direction over the graduated scale to indicate the value of current or voltage
being measured.
The deflecting torque is given by,
Td = NBldI
where N is number of turns,
B is magnetic flux density in air gap,
l is the length of moving coil, d is the width of the moving coil,
I is the electric current.
Thus, Td α I
The instrument is spring controlled so that,
Tc α θ
The pointer will comes to rest at a position, where
Td =Tc
Therefore, θ α I
Thus, the deflection is directly proportional to the operating current. Hence, such instruments have
uniform scale.
Advantages:
1. Uniform scale.ie, evenly divided scale.
2. Very effective eddy current damping.
3. High efficiency.
4. Require little power for their operation.
5. No hysteresis loss (as the magnetic field is constant).
6. Very accurate and reliable.
Disadvantages:
1. Cannot be used for a.c measurements.
The basic movement of DC ammeter is the PMMC D‟Arsonal movement. Since the coil winding
in PMMC meter is small and light, they can carry only small currents (μA-1mA). Measurement of large
current requires a shunt external resistor to connect with the meter movement, so only a fraction of the
total current will passes through the meter.
Rsh=shunt resistance
Thus, I= Im+Ish
Since voltage drop across the shunt and the meter is same,
Vsh=Vm
Ish*Rh=Im*Rm
Note: This is only Basic Information for students. Please refer
Rsh=
“Reference Books” prescribed as per syllabus
DEPARTMENT OF TECHNICAL EDUCATION E-CONTENT
Therefore,
A voltmeter is always connect in parallel with the element being measured, and measures the
voltage between the points across which it is connected. Most d.c voltmeter use PMMC meter with series
resistor as shown in figure below. The series resistance should be much larger than the impedance of the
circuit being measured, and they are usually much larger than Rm.
Rs =
Therefore,
Rs=
The current range of the DC ammeter can be extended by a number of shunts selected by a
switch(S).Such meter is called multirange ammeters.
- 1=
Therefore, Rsh =
Let m1, m2, m3, m4 be the shunt multiplying powers for current I1, I2, I3, I4.
R1=
R2=
R3=
R4=
The selector switch S, selects the appropriate shunt required to change the range of the meter.
When the position of the switch is '1' then the resistance R1 is in parallel with the series combination
of R2 , R3 and Rm. Hence current through the shunt is more than the current through the
meter, thus protecting the basic meter.
When the switch is in the position '2', then the series resistance of R1 and R2, is in parallel with the
series combination of R3 and Rm. The current through the meter is more than through the shunt in this
position.
When the switch is in the position '3', the resistances R1 , R2 and R3 are in series and acts as the
shunt. In this position, the maximum current flows through the meter. This increases the sensitivity of the
meter.
The voltage drop across the two parallel branches is always equal.
Thus, Ish Rsh = Im Rm
But in position 1, R1 is in parallel with R2 + R3 + Rm
In the above figure, the multipliers are connected in series with the meter. The selector switch is
used to select the required voltage range.
When the switch S is at position V1, R1 + R2 + R3 + R4 acts as a multiplier resistance. While when the
switch S is at position V4 then the resistance R4 only acts as multiplier resistance. The V4 is
the lowest voltage range while V1 is the maximum voltage range.
The multiplier resistances can be calculated as :
In position V4, the multiplier is R4 only. The total resistance of the circuit is say RT.
Using equations (1), (2), (3) and (4) multipliers can be designed.
Loading effect:
When selecting a meter for certain voltage measurement it is important to consider the sensitivity
of a DC voltmeter. A low sensitivity meter may give a correct reading if the circuit resistance is low but it
Voltmeter sensitivity:
In a, the ratio of the total resistance RT to the voltage range remains same. This ratio is nothing but
the reciprocal of the full scale deflection current of the meter i.e. 1/Im. This value is called sensitivity of
voltmeter.
Thus the sensitivity of voltmeter is defined as,
S= Ω/V
The sensitivity range is specified on the meter dial and it indicates the resistance of the meter for a
one volt range.
Where, S=sensitivity of the voltmeter (Ω/V)
V=Voltage rang, as set by the range switch
Rm=internal resistance of the meter
Rs=Resistance of the multiplier
RT=Total circuit resistance
RT = ------------------------------------------------------------(1)
S= ---------------------------------------------------------------(2)
3.4 Electrodynamometer
Construction:
The necessary field required for the operation of the instrument is produced by the fixed coils. A
uniform field is obtained near the centre of coil due to division of coil in two sections. These coils are air
cored. Fixed coils are wound with fine wire for using as voltmeter, while for ammeters and wattmeter‟s it
is wound with heavy wire. Ceramic is usually used for mounting supports. If metal parts would have been
used then it would weaken the field of the fixed coil. The moving coil is wound either as a self-sustaining
coil or else on a non-metallic former. If metallic former is used, then it would induce eddy currents in it.
The construction of moving coil is made light as well as rigid. It is air cored. The moving coil is mounted
on a aluminium spindle. The moving system carries a pointer.
The electrodynamometer is used as Ac voltmeter, ammeter and with a slight modification it can
also be used as a wattmeter, power factor meter, frequency meter.
3.4.1 Electrodynamometer type ammeter
In the above circuit arrangement, when the instrument is used as an ammeter, fixed coils and
moving coil are connected in series and therefore carry the same current
i.e.I1=I2=I
Hence, angular deflection, ɵ
To measure heavy currents shunt is used to limit current through the moving coil. For small currents
shunt is not needed.
In the above circuit arrangement, when the instrument is used as an voltmeter, fixed coils and
moving coil are connected in series along with a high resistance.
i.e I1=I2=I
I= in DC circuits
And I = in AC circuits
ɵ V2
The fixed coils are connected in series with the circuit, while the moving coil is connected
in parallel. Also, on analog wattmeters, the moving coil carries a needle that moves over a scale to
indicate the measurement. A current flowing through the fixed coil generates an electromagnetic
field around the coil. The strength of this field is proportional to the line current and in phase with it.
For a dc circuit the deflection of the needle is proportional to both current(I) and the voltage(V).
For an ac circuit the deflection is proportional to the average instantaneous product of voltage and
current,
Thus power, P=VI cos φ. Here, cosφ represents the power factor.
Figure above shows the circuit diagram of a Chopper type DC amplifier voltmeter.
Two photo diodes are used in input stage which acts as half-wave modulators because of its
alternate switching action by the neon lamps at the frequency of oscillator.
In dc voltmeter circuit two neon lamps are used, these are supplied by an oscillator for alternate half
cycles. Output of chopper modulator is a square wave voltage (proportional to the input signal) which is
supplied to the ac amplifier through a capacitor and the amplified output is again passed through a
capacitor and then fed to chopper demodulator. The capacitor is used to smooth the output from the
amplifier.
The Chopper demodulator gives a dc output voltage (proportional to the input voltage) which is
passed through the low pass filter to remove any residual ac component and this dc output voltage is
supplied to the PMMC meter for measurement of input voltage.
Advantages:
1. It has very high input impedance of the order of 10MΩ.
2. It allows input signal in the range of 0.01mV.
Figure above shows an electronic voltmeter using op-amp 741. This is directly coupled very high
gain amplifier. The gain can be adjusted to any suitable value by providing appropriate resistance
between its output terminal, Pin No.6, and inverting input pin No.2, to provide negative feedback. The
ratio R2/R1 determines the gain. The 0.1μF capacitor across the 100KΩ resistance R2 is used for stability
under stray pickups.
A 10KΩ potentiometer is connected between the offset null terminal 1 and 5 with its centre tap
connected to -5V supply for adjusting zero output for zero input conditions. The two diodes used are for
IC protection. If an excessive voltage say more than 100mV appears across them then depending on the
polarity of the voltage one of the diode conducts and protects the IC. A μA scale 50-1000μA full scale
deflection can be used as an indicator. R4 is adjusted to get maximum full scale deflection.
A full wave rectifier type AC voltmeter consists of four diodes and a PMMC meter as shown in
above fig. The current through the moving coil instrument flows in the same direction for both cycles of
the input voltage. The indication of instrument depends upon the mean value of the current flowing
through it.
Here, the meter reading would be 90% RMS i.e., 90% of the DC value. When the input is positive
D1 and D3 conducts, and the current flows through the meter from top to bottom. When the input is
negative D2 and D4 conducts through the meter from top to bottom. In both the cycles of the input
voltage current flows in same direction.
Advantages:
1. The frequency range extends from about 20Hz to high audio frequencies.
2. They have much lower operating current.
3. They have practically uniform scale for most ranges.
Figure above shows the block diagram of a RMS Reading Voltmeter. It consists of two
thermocouples called main thermocouple (MT) and balancing thermocouple (BT). BT is used in the
feedback loop to cancel out the non-linear effects of the MT.
The unknown A.C voltage is amplified and fed to the heating element of the main thermocouple.
The heat produced by the wire is sensed by the measuring thermocouple which produces a proportional
DC voltage. This DC voltage upsets the bridge balance. The unbalance voltage is amplified by the DC
amplifier and fed back to the heating element of the balancing thermocouple.
Bridge balance is reestablished when the two thermocouples produce the same output voltages. At
this point the DC current in the heating element of the feedback thermocouple is proportional to the AC
current in the input thermocouple i.e., the DC is proportional to the rms value of the input AC signal. This
DC value is indicated by the meter movement in the output circuit.
Advantages:
Disadvantages:
3. In case of highly non-linear waveforms, the crest factor puts the limitation on the meter reading.
3.6 Ohmmeter
An ohmmeter is an electrical instrument that measures electrical resistance.
A circuit of series type ohmmeter is shown in figure below. It consists of basic D‟Arsonval
movement connected in parallel with a shunting resistor R2. This parallel circuit is in series with
resistance R1 and a battery of emf (E). The series circuit is connected to the terminals A and B of
unknown resistor Rx.
A switch is used to turn on and off the circuit. The switch is opened when it is not in use.
For the figure,
Rse = zero adjusting resistor,
E = emf of internal battery,
Rm = internal resistance of D‟Arsonval movement,
R=unknown resistance.
3.7 Calibration
In measurement technology calibration is the comparison of measurement values delivered by
an instrument under test with those of a standard instrument of known accuracy.
Figure above shows calibration of DC voltmeter. The test voltmeter reading „V‟ is compared to the
voltage reading obtained by the standard instrument „M‟. Several points on the voltmeter scale can be
calibrated. A voltmeter tested with this method can be calibrated with an accuracy of + 0.01
The value of the current through the meter to be calibrated is determined by measuring the potential
difference across a standard resistor by the voltmeter and then calculating the current by Ohm‟s law.
The result of this calculation is compared to the actual reading of the ammeter under the calibration
inserted in the circuit. A rheostat is placed in the circuit to control the current to any desired value, so that
different points on the meter scale can be calibrated.
Problem - 1:
A basic D‟Arsonal movement with internal resistance, Rm=50Ω, and full scale deflection current
Ifsd=2mA, is to be converted into a multirange dc voltmeter with voltage range of 0-1V, 10V, 100V, and
250V. Draw the necessary circuit arrangement and find values of suitable multipliers.
Solution:
Multirange DC voltmeter
For 0-1V range (V4 position of range switch) the total circuit resistance.
RT = = 500Ω
R3= RT-(R4+Rm)
= 5000 -(450+50)
=5000-500=4500
R3=4.5K Ω
ii) For 100V range (V2 is the position of switch)
RT= =50 KΩ
R2= RT-(R3+R4+Rm)
= 50K -(4500+450+50)
=50K-5K=45KΩ
R2=45K Ω
ii) For 250V range (V1 is the position of switch)
R1 = =125K Ω
R1= RT-(R2+R3+R4+Rm)
= 125K -(45K+4.5K+450+50)
=125K-50K
R1=75K Ω
Problem - 1:
Design an Ayrton shunt to provide an ammeter with current ranges of 1A, 5A and 10A. A
D‟Arsonal movement with an internal resistance Rm=50 Ω and full scale deflection current of 1mA is
used in the configuration.
Therefore, -1=
= 1+
Multiplying factor= m = = 1+
Rsh =
Thus, R1 = = = 0.05 Ω
R2 = = = 0.01 Ω
R3 = = = 0.005 Ω
10 Hours
Cathode Ray Oscilloscope - block diagram, working of CRT, concept of dual tracing.CRO probes: direct,
high impedance, active and current probes. Applications of CRO-simple problems on voltage and frequency
measurements. DSO: block diagram, advantages and applications. Sampling oscilloscope: advantages and
applications. Function generator: block diagram, features and applications. Features of standard RF signal
generator and sweep frequency generator. Features of distortion analyzer, wave analyzers, and spectrum
analyzers.
The cathode ray oscilloscope (CRO) is an instrument which we use in laboratory to display, measure
and analyze various wave forms of various electrical and electronic circuits.
CRO is a very fast X-Y plotter that can display an input signal versus time or other signal.
Using CRO the wave shapes of alternating currents and voltages can be studied. It can also be used for
measuring voltage, current, power, frequency and phase shift.
Cathode ray oscilloscope uses luminous spot which is produced by striking the beam of electrons and
this luminous spot moves in response variation in the input quantity. CRT is the heart of oscilloscope.
2. Vertical Amplifier: In Vertical Amplifier, the input signals are amplified by the vertical amplifier which
is a wide band amplifier which passes the entire band of frequencies.
3. Delay Line: This circuit is used to delay the signal for a period of time in the vertical section of CRT. The
delay Line retards the arrival of the input waveform at the vertical deflection plates until the trigger and time
base circuits start the sweep of the beam.
5. Time Base Generator: It uses a uni junction transistor, which is used to produce the sweep. The saw tooth
voltage produced by the time base generator is required to deflect the beam in the horizontal section. The
spot is deflected by the saw tooth voltage at a constant time dependent rate.
6. Horizontal amplifier: The saw tooth voltage produce by the time base circuit is amplified by the
horizontal amplifier before it is applied to horizontal deflection plates
7. Power supply: A high voltage section is used to operate CRT and a low voltage section is used to supply
electronic circuitry of the oscilloscope such as horizontal amplifier, vertical amplifier, sweep generator etc,.
The CRT generates the electron beam which are accelerated to a high velocity and brought to focus on
a fluorescent screen which produces a visible spot where the electron beam strikes. By deflecting the beam
over the screen in response to the electrical signal, the electrons can be made to act as an electrical pencil of
light which produces a spot of light wherever it strikes.
Low voltage supply is required for the heater of the electron gun to generate the electron beam and high
voltage is required for the cathode ray tube to accelerate the beam. Normal voltage supply is required for
other control units of the oscilloscope. All these various powers are provided by power supply.
Horizontal and vertical deflection plates are fitted between the electron gun and the screen so that these
can deflect the beam according to the input signal.
To deflect the electron beam on the screen in horizontal direction i.e. X-axis with constant time
dependent rate, a time base generator is provided in the oscilloscope. The signal to be viewed is supplied to
the vertical deflection plate through the vertical amplifier, so that it can amplify the signal to a level that will
provide usable deflection of the electron beam. As the electron beam is deflected in X-axis as well as Y-axis,
a triggering circuit is provided for synchronizing these two types of deflections so that horizontal deflection
starts at the same point of the input vertical signal each time it sweeps. When the deflected electron beam
strikes the screen, the image is obtained as a spot of light.
The diagram shows that when delay line is not used, the initial part of the signal is lost and only part of
the signal is displayed. To counteract this disadvantage the signal is not passed directly to the vertical plates
but is passed through delay line circuit.
This gives time for sweep to start at the horizontal plates before the signal has reached the vertical
plates.
The trigger pulse is picked off at a t0 after the signal has passed through the main amplifier. The sweep
generator delivers the sweep to the horizontal amplifier and sweep starts at the HDP(horizontal deflection
plates) at time t0+80 ns. Hence sweep starts well in time, since the signal arrives at the VDP(vertical
deflection plates) at time t0+200 ns.
The main part of cathode ray oscilloscope is cathode ray tube which is also known as the heart of
cathode ray oscilloscope. The cathode ray tube or CRT is a vacuum tube of special geometrical shape which
converts an electrical signal into a visual one.
1. Electron gun
3. Fluorescent screen
4. Glass envelope
1. Electron Gun
The electron gun assembly consists of an indirectly heated cathode, a control grid, a focusing anode
and an accelerating anode and it is used to produce a focused beam of electrons.
A layer of barium oxide is deposited on cathode end which is indirectly heated at moderate temperature
for high emission of electrons, these electrons are then passed to control gird made up of nickel controls the
intensity of electrons emitted from cathode. The electrons are then accelerated by pre-accelerating and
accelerating anodes which are connected to common positive potential of 1500V. Focusing anode focuses the
electron beam produced. There are two methods of focusing. a) Electrostatic focusing b) Electromagnetic
focusing.
It consists of two sets of deflecting plates and is used for the deflection of the beam. a) Vertical
deflection plates b) Horizontal deflection plates.
The vertical deflection plates are mounted horizontally in the tube. On application of proper potential to
these plates, the electron beam can be made to move up and down vertically on the screen.
The horizontal deflection plates are mounted vertically in the tube. On application of proper potential to
these plates, the electron beam can be made to move right and left horizontally on the screen.
3. Fluorescent screen
The screen is coated with some fluorescent materials such as zinc orthosilicate, zinc oxide etc, and is the
inside face of the tube. When high velocity electron beam strikes the screen, a spot of light appears at the
point of impact. The colour of the spot depends upon the nature of fluorescent material.
It is a conical highly evacuated glass housing which maintains vacuum inside it and supports various
electrodes. The inner wall of CRT between the neck and screen is coated with a conducting material known
as aquadag, which is electrically connected to the accelerating anode so that the electrons which accidentally
strike the walls are returned to the anode. This prevents the walls from charging to a high negative potential.
http://www.electrical4u.com/cathode-ray-oscilloscope-cro/
https://www.elprocus.com/cathode-ray-tube/
https://youtu.be/U1amW7S1fcI
Dual-trace operation allows to view two independent signal sources as a dual display on a single CRT.
This operation allows an accurate means of making amplitude, phase, time displacement, or frequency
comparisons and measurements between two signals.
There are two separate vertical input channels A and B. These use two separate attenuators and pre-
amplifiers stages. Therefore the amplitude of each input as viewed on the oscilloscope can be individually
controlled. After pre-amplification the two channels meet at an electronic switch. This has the ability to pass
one channel at time into the vertical amplifier, via the delay line.
Switch S2 allows the circuit to be triggered on either the A or B channel waveforms, or on line
frequency, or on an external signal. The horizontal amplifier can be fed from the sweep generator or the B
channel via switch S1. This is the X - Y mode and the oscilloscope operates from channel A as the vertical
signal and channel B as the horizontal signal, giving very accurate X - Y measurements.
The dual trace oscilloscope is operated in two modes a) alternate mode b) chop mode. The operating
modes are selected form the instrument front panel.
Alternate mode: In case of alternate mode, electronic switch alternately connects the main vertical amplifier
to channel A and B and adds a different D.C. component to each signal. Due to this D.C. component the
beam alternately goes to the upper and lower half of the screen. Switching is carried out at the start of each
new sweep of the time base generator.
To obtain the cathode ray tube spot it traces the channel A signal on one sweep and the channel B
signal on the succeeding sweep. The switching rate of the electronic switch is synchronized to the sweep rate.
The alternate mode cannot be used for displaying very low frequency signals.
The function of Probe is to connect the test circuit to the oscilloscope without altering, loading or
disturbing the test circuit.
1. Direct probe
When direct probes are used the stray capacitance of probe adds with capacitance of oscilloscope.
Hence at high frequencies, the total input reactance becomes low. This decreases the input impedance and
causes a sensitive circuit to break into oscillations. This effect is dominant at high frequencies, hence such
probes are used only at low frequencies. These are also known as 1X probe.
These probes are used to increase the input resistance and reduce the effective input capacitance of an
oscilloscope. The capacitance is adjusted so that the elements of the bridge are balanced.
Therefore, X and Y are equipotential and the effect of the probe is equivalent to placing a potential
divider consisting of R1 and Rin across the input circuit. The attenuation of signal is 10:1 i.e, (R1+Rin)/Rin =
10:1 over a wide frequency range. Hence the name compensated 10:1 probe. Capacitor C 1 is adjusted by
connecting the probe tip to a square wave of 1KHz and observing the CRT display. When the CRT display
has optimum response the value of C1 is deemed to be appropriate. Then, Vout = 0.1Vin = ( Vin x Rin)/
(R1+Rin).
3. Active Probe
These probes are designed to provide an efficient method of coupling high frequency, fast rise time
signals to CRO input. They have very high input impedance with less attenuation than passive probes. These
The FET is used as the active element to amplify the input signal. The voltage gain of FET follower
circuit is unity but it provides a power gain so that input impedance can be increased.
4. Current Probe
This probe can be clamped around a wire carrying current without any physical contact to the probe
and the magnitude of current with frequency from 0(DC) to 50Hz can be measured in oscilloscope.
The current sensor consists of two parts: A conventional transformer for transforming alternating
current to voltage, and Hall effect device for converting direct current to a voltage.
http://www.physics.wisc.edu/undergrads/courses/fall2015/623/lab_equipment/MSO2014_scope_tutorials/Pro
bes/Probes-Lab.pdf
1. To trace and measure a signal throughout the RF, IF and AF channels of radio and television receivers.
2. It provides the only effective way of adjusting FM receivers, broadband high frequency RF amplifiers
3. To test AF circuits for different types of distortions and other spurious oscillations
4. To give visual display of wave shapes such as sine waves, square waves and their many different
combinations.
7. Amplifier Gain.
8. Modulation percentage.
http://hubpages.com/technology/Major-Applications-Uses-of-Cathode-Ray-Oscilloscope-CRO-explained-in-
detail
Peak and Peak to Peak Value (Vpp): - The maximum value, +ve or –ve of the alternating quantity is
known as its peak value. It is also called maximum value or amplitude of the alternating quantity. The total
voltage measured from –ve peak to +ve peak is called the Peak to Peak voltage. Measure the number of
divisions on the voltage axis and multiply it by the value indicated by the Volts/Div knob on the CRO. This
gives the Vpp value.
RMS Value (Root Mean Square Value) (Vrms): It is given by the steady dc current which when flowing
through a given circuit for a given time produces the same heat as produced by the alternating current which
when flowing through the same circuit for the same time. RMS value of alternating voltage is related to its
peak value by the relation Vrms= Vp/√2 where Vp = Vpp/2.
Measurement of frequency:
Measure the number of divisions for a single cycle on the time axis and multiply it by the value
indicated by the Times/Div knob on the CRO. This gives the Time Period of the alternating voltage.The
number of cycles completed in one second is called the frequency of the alternating voltage. Its unit is Hertz.
Frequency is given by the reciprocal of Time period T.
Example - 1:
Here Peak to peak voltage in CRO is 4 blocks and the Volts /Div knob is in 2V position.
Here Time period in CRO is 4 blocks and the Times/ Div Knob is in 2μs.
Example - 2:
Here Peak to peak voltage in CRO is 4.4 blocks and Volts/Div knob is at 5V position.
https://youtu.be/tCpChoMWhjE
A digital storage oscilloscope (DSO) is an oscilloscope which stores and analyses the signal digitally
rather than using analog techniques. It is now the most common type of oscilloscope in use because of its
advanced trigger, storage, display and measurement features.
A digital oscilloscope is an instrument which stores a digital copy of the waveform in the digital
memory which it analyses further using digital signal processing techniques rather than using analogue
techniques.
In DSO, the waveform to be stored is digitized, stored in a digital memory, and retrieved for display
on the storage oscilloscope.
1. Data acquisition
2. Storage
3. Data display
Data acquisition portion of the system consist of a Sample-and-Hold (S/H) circuit and an analog to
digital converter (ADC) which continuously samples and digitizes the input signal at a rate determined by the
sample clock and transmit the digitized data to memory for storage. The control circuit determines whether
the successive data points are stored in successive memory location or not, which is done by continuously
updating the memories.
When the memory is full, the next data point from the ADC is stored in the first memory location
writing over the old data. The data acquisition and the storage process is continues till the control circuit
receive a trigger signal from either the input waveform or an external trigger source. When the triggering
occurs, the system stops and enters into the display mode of operation in which all or some part of the
memory data is repetitively displayed on the cathode ray tube.
In display operation, two DACs are used which gives horizontal and vertical deflection voltage for the
CRT Data from the memory gives the vertical deflection of the electron beam, while the time base counter
gives the horizontal deflection in the form of staircase sweep signal. The screen display consist of discrete
dots representing the various data points but the number of dot is very large as 1000 or more that they tend to
blend together and appear to be a smooth continuous waveform. The display operation ends when the
operator presses a front-panel button and commands the digital storage oscilloscope to begin a new data
acquisition cycle.
1. Roll Mode - It is the most basic mode of operation which is similar to that of a general purpose CRO.
When an input is applied, its trace is displayed on the screen. A user can use this mode to keep an eye on
2. Refresh Mode - It is used when the sample rate of a waveform becomes too high, and when the waveform
of interest is repetitive or nearly so. The DSO produces a stale, triggered display with a higher sweep time.
3. Single Shot Mode - The DSO is in an inactive state in this mode, displaying the last trace captured, until a
4. Equivalent Time Mode - It basically arranges the ADC to work on limited sweeps, so as to allow
1. It can store the given signal indefinitely as long as the small amount of power is supplied to the
memory.
3. It employs normal CRT, hence the cost of the tube is much cheaper than the storage tube used in
9. The digital storage oscilloscope is also able to operate in a babysitting mode. When the scope is
4. Acting as a simple signal tracer, a DSO enables technicians to probe electronic device’s individual
5. In measuring the functions of the individual component of the device, the DSO locates where an
6. The DSO can also measure components’ minor variations in operations and alert the technician of
7. To prevent erroneous replacement of parts, the DSO also helps technicians identify the parts that
https://youtu.be/G8dNpWqk9-U
The sampling oscilloscope is a special type of digital sampling oscilloscope which is used to examine a
very fast signal. It works on the principle of stroboscopic light in which sample is taken at different portions
of the waveform, over successive cycles, and then the total picture is stretched, amplified by relatively low
bandwidth amplifiers, and display as the continuous wave on the screen.
An ordinary oscilloscope has a bandwidth of 10 MHz the HF performance can be improved by means
of sampling the input waveform and reconstructing its shape from the sample, i.e. the signal to be observed is
sampled and after a few cycles sampling point is advanced and another sample is taken.
The shape of the wave form is reconstructed by joining the sample levels together. The sampling
frequency may be as low as 1/10th of the input signal frequency (if the input signal frequency is 100 MHz,
the bandwidth of the CRO vertical amplifier can be as low as 10 MHz). As many as 1000 samples are used to
reconstruct the original waveform.
The input is applied to the sampling gate. The input waveform is sampled whenever a sampling pulse
opens the sampling gate. The sampling must be synchronized with the input signal frequency. The signal is
delayed in the vertical amplifier, allowing the horizontal sweep to be initiated by the input signal. The
waveforms are shown in fig 4.3.1.2. At the beginning of each sampling cycle, the trigger pulse activates an
oscillator and a linear ramp voltage is generated. This ramp voltage is applied to a voltage comparator which
compares the ramp voltage to a staircase generate-When the two voltages are equal in amplitude, the
staircase advances one step and a sampling pulse is generated, which opens the sampling gate for a sample of
input voltage. The resolution of the final image depends upon the size of the steps of the staircase generator.
The smaller the size of the steps the larger the number of samples and higher the resolution of the image.
1. The digital sampling oscilloscope is intended for very high frequency operation.
2. They are used for looking at repetitive signals which have a higher frequency than the sample rate of
the scope.
3. They achieve their performance by collecting samples from several successive waveforms, and then
assembling them together to build up a picture of the waveform. This technique allows these digital
4. The dynamic range of this instrument is low (max voltage peat to peak is 3V ) but they are able to
5. Controlling the size of steps of the staircase generator, the number of samples and hence resolution
can be controlled.
7. The technique used in sampling oscilloscope allows the design of oscilloscope with wide band width,
http://www.indiastudychannel.com/resources/141738-Working-principle-sampling-oscilloscope.aspx
https://youtu.be/3Q7zPJcWFmk
https://youtu.be/IkBiI69_Mnw
A function generator is a electronic test equipment or software used to generate different types of
electrical wave forms over a wide range of frequencies. The most common wave forms produced by the
function generator are the sine, square, triangular and saw tooth shapes.
A function generator is a signal source that has the capability of producing different types of
waveforms as its output signal.
In this instrument the frequency is controlled by varying the magnitude of current that drives the
integrator. This instrument provides different types of waveforms (such as sinusoidal, triangular and square
waves) as its output signal with a frequency range of 0.01 Hz to 100 kHz.
The frequency controlled voltage regulates two current supply sources. Current supply source 1 supplies
constant current to the integrator whose output voltage rises linearly with time. Current supply source 2
supplies constant reverse current to the integrator whose output voltage falls linearly with time.
1
C
According to output equation of integrator Vout idt an increase or decrease in the current
increases or reduces the slope of the output voltage and thus controls the frequency.
The voltage comparator multivibrator changes state at a predetermined maximum level, of the integrator
output voltage. This change cuts-off the current supply from supply source 1 and switches to the supply
source 2. The current supply source 2 supplies a reverse current to the integrator so that its output drops
linearly with time. When the output attains a predetermined level, the voltage comparator again changes state
and switches on to the current supply source 1. The comparator output provides a square wave of the same
frequency as output.
The output of the integrator is a triangular wave whose frequency depends on the current supplied by
the constant current supply sources.
The resistance diode network changes the slope of the triangular wave as its amplitude changes and
produces a sinusoidal wave with less than 1% distortion.
The out put amplifiers provide impedance matching and control the output signal level. The out put
amplifier is a adjustable gain wide band amplifier.
4.4.2 Features
2. It can generate a wide range of frequencies. For example, the Tektronix FG 502 (ca 1974)
3. Frequency stability of 0.1 percent per hour for analog generators[4] or 500 ppm for a digital
generator.
4. Maximum sine wave distortion of about 1% (accuracy of diode shaping network) for analog
generators. Arbitrary waveform generators may have distortion less than -55 dB below 50
7. Amplitude can be modified, usually by a calibrated attenuator with decade steps and
8. Many function generators are also capable of generating two different waveforms
simultaneously (from different output terminals, of course). This can be a useful feature when
9. A function generator has several high-quality waveforms with accurate, stable frequency and is
and waveform.
4.4.3 Applications
need a signal that increases (or reduces) at a specific linear rate. They are used for driving sweep
4. Their common uses are in the field of education, repairing electrical and electronic equipment
5. Used in laboratory to measure frequency response of filters, measuring the ESR (Equivalent
RF signal Generator is used to generate RF or radio frequency signals. Radio frequency signal
generators (RF signal generators) are a particularly useful item of test equipment widely used in RF
microwave design and test applications.
Features:
1. Frequency range: RF signal generators typically range from a few kHz to 6 GHz, while microwave signal
generators cover a much wider frequency range, from less than 1 MHz to at least 20 GHz
2. Output level: The output range for an RF and microwave signal generator is controlled to a relatively
high degree of accuracy. The output within the generator itself is maintained at a constant level and then
3. Built in attenuator: common feature is a built-in attenuator which makes it possible to vary the signal’s
output power, Depending upon manufacturer model output powers can range from -135 to +30 dBm.
4. Modulation: Some RF or microwave signal generators have inbuilt oscillators that can apply modulation
to the output signal. Others also have the ability to apply modulation from an external source. Signal
generators that support complex modulation are often referred to as vector signal generators.
5. Sweep: On some RF signal generators it is necessary to sweep the signal over a range. Some generators
6. Control: They may have front panel control or remote control option for controlling RF and microwave
signal generators.
http://www.radio-electronics.com/info/t_and_m/generators/radio-frequency-rf-signal-generator.php
http://www.cosmanmedical.com/products/rf-generators/
https://youtu.be/YxCTicf3QVA
A sweep frequency generator is a signal generator which can automatically vary its frequency smoothly
and continuously over an entire frequency range and maintains a constant amplitude. Sweep frequency
generators are commonly used to test the frequency response of electronic filter circuits.
Features:
1. A sweep frequency generator is a type of signal generator that is used to generate a sinusoidal output.
Such an output will have its frequency automatically varied or swept between two selected frequencies.
One complete cycle of the frequency variation is called a sweep. depending on the design of a particular
instrument, either linear or logarithmic variations can be introduced to the frequency rate. However, over
the entire frequency range of the sweep, the amplitude of the signal output is designed to remain constant.
2. Sweep-frequency generators are primarily used for measuring the responses of amplifiers, filters, and
3. The frequency range of a sweep-frequency generator usually extends over three bands, 0.001 Hz – 100
kHz (low frequency to audio), 100 kHz – 1,500 MHz (RF range), and 1-200 GHz (microwave range).
5. The automatic level controller keeps power constant avoiding the source mismatch and loading effect.
Applications:
1. They are used to display the response curve of the various stages of frequency of television or radio
receivers.
2. They are used to determine the characteristics of a device over a wide range of frequencies
3. Sweep generators may also be employed for checking and repairing of amplifiers used in TV and radar
receivers.
The different instruments which are used for signal analysis are wave analyzers, distortion analyzers,
spectrum analyzers, audio analyzers and modulation analyzers. All signal analysis instruments measure the
basic frequency properties of a signal, but they use different techniques .
A spectrum analyzer sweeps the signal frequency band and displays a plot of amplitude versus frequency.
It has an operating range of about 0.02 Hz-250 GHz.
A wave analyzer is a voltmeter which can be accurately tuned to measure the amplitude of a single
frequency, within a band of about 10 Hz - 40 MHz.
An audio analyzer is similar to a distortion analyzer but can measure additional functions, such as noise.
Modulation analyzers tune to the required signal and recover the whole amplitude modulated (a.m.),
frequency modulated (FM) or phase modulation (PM) envelope for display or analysis.
Distortion analyzer measures the total harmonic power present in the test wave rather than the
distortion caused by each component. The simplest method is to suppress the fundamental frequency by
means of a high pass filter whose cut off frequency is a little above the fundamental frequency. This high
pass allows only the harmonics to pass and the total harmonic distortion can then be measured. A
Fundamental suppression distortion analyzer is used to measure distortion factor (THD).
1. The harmonic distortion generated within the instrument which affect THD measurement is very small
2. The selectivity requirement is not severe because only fundamental component is to be suppressed.
3. There are different types of distortion analyzer Fundamental suppression distortion analyzer, Heterodyne
Applications:
2. To determine THD ( total harmonic distortion) of an amplifier by using very low distortion sine wave
3. It is used in measurement of effectiveness of electronic filter with extremely narrow pass band such as
https://youtu.be/IzwEnEAu8Lo
There are two types of wave analyzers, depending upon the frequency ranges used,
1. Frequency Selective wave analyzer: This is used in measurements of audio frequency range (20Hz to
20kHz). This analyzer has a filter section with very narrow pass band which can be tuned to frequency of
Applications:
1. The wave analyzers have very important application in the fields of a) Electrical measurements b)
2. The wave analyzers are applied industrially in the field of reduction of sound and vibrations generated
http://www.myclassbook.org/heterodyne-wave-analyser/
https://prashanthrejintala.wordpress.com/2014/09/11/heterodyne-wave-analyzer/
http://www.slideshare.net/sumeetpatel21/lab-equipments-40835876
https://youtu.be/zc3HtEAjODM
A spectrum analyzer measures the magnitude of an input signal versus frequency within the full
frequency range of the instrument. The primary use is to measure the power of the spectrum of known and
unknown signals. The input signal that a spectrum analyzer measures is electrical.
Spectrum analysis of a signal gives the following information by displaying signal in frequency domain.
3. Band width.
4. Frequency stability.
6. Spectral purity.
There are two types of spectrum analyzer based on instrumentation frequency limitations
Audio Frequency Spectrum analyzer: It covers audio frequency range of 20Hz to 20KHz
There are two types of Spectrum analyzer based on operating technique used.
Applications:
Note: This is only Basic Information for students. Please refer
“Reference Books” prescribed as per syllabus
DEPARTMENT OF TECHNICAL EDUCATION E-CONTENT
1. In the Fields of biomedical electronics, geological surveying and oceanography, it is used to analyze
4. It is used to examine the vibration signals from automobiles, airplanes, space vehicles, bridges etc
5. It finds applications in the field of electronic testing related to troubleshooting and quality control.
Digital instruments - pros and cons, working of ramp and successive approximation type digital
voltmeters. Automatization in digital meters-mechanism of automatic zeroing, polarity indication and
auto ranging. Electronic counters-decade counters as an electronic counter, totalizing, frequency mode,
ratio mode, period mode and time interval mode. Digital meters: digital frequency meter, time interval
measurement, digital LCR meter, digital multimeter, microprocessor-based instruments, IEEE 488 GPIB
instruments.
Pros (Advantages):
1. Sometimes they tend to indicate erratic values due to faulty electronic circuit or damaged
Principle: The time taken for a linear ramp voltage to rise from 0 V to input voltage level, or to decrease
from the input voltage level to 0 V is measured with an electronic time-interval counter, and the count is
displayed as a number of digits on electronic indicating tubes (display).
At the start of measurement, a ramp voltage is initiated and this can be negative-going or positive-
going. As shown in figure 5.2.1.2, the negative-going ramp is continuously compared with the unknown
input voltage. At the instant that the ramp voltage equals the unknown voltage, a coincidence circuit, C1 (
input comparator) generates a pulse which opens a gate. The ramp voltage continues to decrease with
time and when it reaches 0 V (or ground potential), a second comparator (zero crossing detector), C2
generates an output pulse which closes the gate.
A clock oscillator generates clock pulses which are allowed to pass through the gate to a number
of decade counting units (DCUs) which totalise the number of pulses passed through the gate. The
decimal number, displayed by the display associated with the DCUs, is a measure of the magnitude of the
input voltage.
The sample-rate multivibrator determines the rate at which the measurement cycles are initiated
and this is connected to ramp generator which provides an initiating pulse for the ramp generator to start
next ramp voltage for the next measurement. It is also used to reset the counter before generating the next
ramp voltage.
Merits:
1. Low cost.
2. Simple and easy to design.
3. Long distance transmission of output pulse is possible.
Demerits:
DAC is used to provide the estimates. The "equal to or greater than" or "less than" decision is
made by a comparator. DAC provides the estimate and is compared to the analog input signal. A special
shift register called a successive-approximation register (SAR) is used to control DAC and the estimates.
At the beginning of the conversion all the outputs from the SAR are at logic zero. If the estimate
generated by DAC is greater than the input, the comparator output is high and the first SAR output
reverses state and the second output changes to logic 1. If the estimate generated by DAC is lower than
the input, the comparator output is low, the first output remains in the logic one state and the second
output assumes the logic state one. This continues to all the states until the conversion is complete.
For an N-bit conversion after N clocks, the actual value of input is known. The LSB is the state of
the comparator.
Merits:
1. Less expensive.
2. More effective type.
3. Good ratio of speed to power.
Before taking real measurements, switch S3 is closed for grounding the input, switch S4 is also
closed to provide a short RC time to integrator and switch S5 is closed to connect output of comparator to
capacitor. Then the capacitor is charged to total offset voltage which is equivalent to zero error. Now
switches S3, S4 and S5 are opened to take the real measurements by considering the offset voltage stored
in capacitor.
The polarity indication is obtained from the information in ADC. In integrating ADCs, the
polarity is measured at the very end of integration period. Counting the number of clock pulses, the
length of the integration period is determined. Polarity flip flop is set by using the output of integrator
which is stored in memory till next measurement is made.
It is used to obtain reading with optimum resolution in all conditions and to prevent overload and
reverse polarity connection. For example 150mV should be displayed as 150.0V and not as 0.150V. If
the value to be measured is within a certain limit, more sensitive range is selected and if higher value is to
be measured, the next less sensitive range is selected.
Consider an example of 3 ½ digit display. 149mV is displayed as 149.0. Any higher value than
150mV is displayed with less resolution as 151mV is read as 0.151. Small fluctuations around 150mV are
displayed successively as 149.9, 0.150 and 0.151 which is confusing. This can be avoided by setting
lower limit less than 150mV (say 120). All values are displayed in the same range as 0.149, 0.150 and
0.151. Values around lower limit gives stable display.
Principle: Giving the input frequency to the counter for a predetermined time.
To one input of AND gate, a pulse with known period is applied. Till the Gate signal is at logic
“1” the output of AND gate is same as unknown input. When the gate signal becomes logic “0”, the
output of AND gate is zero. Counters are cleared using reset signal. For every input pulse counter
increments. Strobe signal is applied at the end of counting and counted value is displayed on a display
device.
By applying a rectangular pulse of known duration to the AND gate, accurate control of time
interval is achieved. If the time interval in which the pulses are being totalized is accurately controlled,
the counter operates in frequency mode.
F= Hertz
Numerical value of the ratio of the frequencies of two signals is displayed in this mode. To
provide gate pulse, a low frequency signal is used instead of clock. During gate pulse, decade counter
counts the number of cycles of high frequency signal and is directly read as ratio of the frequencies.
By giving input signal as a gating pulse and counting the clock pulses, period can be easily
measured. Period is reciprocal of frequency. From the number of pulses of known frequency, period of
input signal is determined.
T=
Elapsed time between two events is measured in this mode. The start and stop inputs controls the
gate independently. Start input enables gate and stop disables gate and pulses are counted between the
time interval of start and stop inputs.
Frequency counters measure the number of oscillations or pulses during a given interval of time in a
repetitive electronic signal. As the preset period gets completed, the value in the counter display on the
screen and the counter reset to zero. For measuring low frequencies, the deflection type meters are widely
used.
A frequency counter is an electronic instrument, which is used for measuring pulse repetition
frequency of unknown signal.
Principle:
The input signal is converted into trigger pulses and applied continuously to one input of AND
gate. A pulse of required duration (say 1 sec) is applied to the other terminal and number of pulses
counted during this period indicates the frequency.
When the input signal of unknown frequency enters the frequency counter, it is fed into input signal
conditioning circuit where the signal is amplified and then it is converted into square wave by Schmitt
trigger circuit. Due to the presence of Schmitt trigger circuit, noise does not cause spurious edges that
would give rise to additional pulses that would be counted. The output from Schmitt trigger is a train of
pulses, one pulse for each cycle of the signal. The square wave is differentiated and clipped to produce a
train of positive pulses.
A crystal oscillator is used to provide an accurately timed gate signal that will allow through pulses
from the incoming signal. This is generated from the clock by dividing the clock signal by decade
dividers. The timebase selector output obtained from oscillator is passed through another Schmitt trigger
circuit to produce square output and then it is fed to frequency dividers connected in cascade. Thus a train
of pulses are obtained after each frequency divider section.
The frequency divider output is then differentiated and clipped to produce a train of positive pulses.
These pulses are applied to clock input of T flip-flop. This flip-flop changes its output state on the
application of each pulse. The first pulse sets the gate control flip-flop and provides an enable signal to
AND gate. When gate is enabled, the input pulses pass through it and fed directly to electronic counter
and counting starts. The second pulse resets the control flip-flop and removes the enable signal from the
AND gate. When the gate is disabled, the counter stops counting. The number of pulses counted between
the time interval of enabling and disabling of AND gate indicates the unknown frequency of the input
Advantages:
Uses:
Figure 5.5.2.1 Digital LCR meter Figure 5.5.2.2 Hand held digital LCR meter
Courtesy: aliexpress.com Courtesy: embeddedlab.com
Principle:
Measuring the voltage across and the current through the Component Under Test (CUT) when the
Operation:
The operation of LCR meter is explained with the help of the block diagram of LCR meter as
shown in figure 5.5.2.3. The processed voltage and current signals are fed to digital integrator which
enables a display unit to provide the value of test component.
An oscillator applies the measuring test signal to test component through a source resistor, R S with
a typical test frequency of 1 KHz. The signal current then flows through current to voltage converter
which is an op-amp with range resistor, RR connected in the feedback path. The op-amp drives the
junction of component and RR to virtual ground and hence RR does not change the current through the
component which develops a voltage E2 proportional to current through the component.
The voltages E1 and E2 are the vector quantities which defines the characteristics of component at
a particular test frequency and signal level.
Mathematically,
Therefore,
The above ratios are adopted in measurement modes. For inductance measurements, a series
equivalent of an inductor is assumed and for capacitance measurements, a parallel equivalent circuit of
capacitor is assumed.
The value of RS and RR are selected based on the impedance of the unknown component. For
inductance measurements, the component impedance is low and hence RS is chosen much higher. This
provides a constant current drive to the test component. The RS decides the value of current. For
capacitance measurements, the component impedance is high and hence R S is chosen much lower. The RS
decides the value of current. This provides a constant voltage drive to the test component.
The voltage E1 is fed as one of the input to differential amplifier and then it is fed to a control
switch along with E2. Greater of the two voltages is fed to Average Voltage Detector (AVD) and lesser is
fed to the the Phase Sensitive Detector (PSD) (phase sensitive selector).
The signal given to AVD is also given to phase locked loop (PLL) and Voltage Controlled
Oscillator (VCO). PLL and VCO produce clock signals which are locked in phase with the reference
signal. Then the clock signal is divided into phase shifted 90 and 270 signals. These signals are given to
PSD which detects the phase angle.
The DC voltage outputs from AVD and PSD are given to digital integrator and finally to digital
display unit where it displays the component value.
Features:
1. High accuracy.
2. Multipoint measurement techniques.
3. Hold function which is used for selecting the components within a specified range.
4. High frequency range.
5. Auto-computing facility and auto ranging facility.
6. Series or parallel modes of operation.
Applications:
https://www.youtube.com/watch?v=-rNttPtL1ME
https://www.youtube.com/watch?v=BYjNkWE2v2E
Digital multimeter consists of the following blocks with their specific functions as follows:
Constant current source is used to supply constant current through the resistance which needs to be
measured and then measuring the voltage across resistance which is directly proportional to resistance
and displayed in ohms.
2. Buffer amplifier
A buffer amplifier is used to unaffect the constant current source with whatever voltages that the
load may produce.
3. Calibrated attenuator
Calibrated attenuator is used for measuring both DC and AC voltages. After measuring AC voltage,
it is passed through rectifier circuit to convert it in to DC voltage.
Note: This is only Basic Information for students. Please refer
“Reference Books” prescribed as per syllabus
DEPARTMENT OF TECHNICAL EDUCATION E-CONTENT
4. Current to voltage converter
Current to voltage converter is used for current measurements (both AC and DC). After measuring
AC current, it is passed through rectifier circuit to convert it in to DC current.
5. Rectifier circuit
For AC measurements, the rectifier circuit converts input AC voltage or current to DC voltage or
current respectively.
DMM is commonly used for the measurement of analog quantities. Hence the conversion from
analog to digital is done with the help of analog to digital converter.
7. Display unit
The output from analog to digital converter is displayed as numerals in the display unit. LCD or
LED displays can be used.
Features:
2. Auto-ranging.
3. dB readings.
4. Data acquisitions.
Advantages:
1. High resolution.
2. Automatic operation.
4. High accuracy.
5. Increased speed.
Applications:
6. It is a hand-held device used for basic fault finding and field service work
7. It can be used to troubleshoot electrical problems in a wide array of industrial and household
devices such as electronic equipment, motor controls, domestic appliances, power supplies, and
wiring systems.
http://www.slideshare.net/vivekpathak5686/digital-multimeter
http://www.slideshare.net/bapikumar144/digital-multimeters
https://www.youtube.com/watch?v=zb7WHaL_dz8
https://www.youtube.com/watch?v=AEq1wNIcWJM
The front end circuit applies the test signal to the unknown impedance and the standard impedance
provides an output signal proportional to the voltage across each, to the phase sensitive detector (PSD).
PSD which is controlled by microprocessor converts the AC inputs of impedance in vector form to
DC output.
The digital data given by analog to digital converter (A/D) is used by microprocessor to obtain the
value of unknown impedance and then it is displayed on a CRT or sent to IEEE 488 bus.
Figure 5.7.1 IEEE 488 GPIB card, GPIB cable and GPIB connector
GPIB interface is a parallel 24 conductor bus. It includes eight data lines for control messages that are
ASCCI encoded, and various management and handshake lines (Handshaking is used to transfer
messages between the PC and the instrument being controlled). This bus is called by three names:
The basic structure of IEEE 488 GPIB system has four different devices, namely computer, frequency
counter, signal generator and digital multimeter.
It has three major buses namely GIM, DIO and DBT and their functions are as follows:
GIM bus co-ordinates the whole system and ensures an orderly flow of data over DIO bus.
DIO bus which is bi-directional 8-bit data bus that carries data interface messages and device-
dependent messages between controllers, talkers and listeners. Asynchronous sending is performed in
byte serial format.
The equipments connected to IEEE 488 GPIB system falls into three categories, although items can
fulfil more than one function:
Listener - A listener accepts commands from other instruments and acts on the commands but it does not
send any data back to controller.
Features:
Parameter Details
3. Handshake lines 3
5. Maximum individual
2 meters average, 4 meters
distance between
maximum
instruments
7. Maximum number of
devices that can be 15
connected
Advantages:
1. Improved performance.
2. High reliability.
3. Increased productivity.
4. GPIB cards are relatively cheap.
5. Simple & standard hardware interface.
6. Interface present on many bench instruments.
Grounding or earthing means connecting exposed metal parts of equipments to ground (earth)
through a low resistance wire to prevent user contact with dangerous voltage when electrical
insulation fails OR
It is a collection of measures and precautions taken to prevent stray currents entering sensitive parts
of equipment by connecting appropriate part of the circuit to ground.
It prevents entering dangerous potentials on the equipment even during electrical faults and also
ensures proper operation of equipment during fault conditions. Grounding minimizes the susceptibility of
equipment to interference and also eliminates electrostatic buildup that can damage system components.
Grounding has the ability to absorb or dissipate an unlimited amount of electrical charge.
If grounding is not performed in equipments, then it may result in Stray voltages or earth potential
rise effects, which may create noise in signals or if large enough will produce an electric shock hazard.
Shielding means placing sensitive electronic parts and components in a metal casing to prevent
electric and magnetic fields entering that casing. OR
Shielding provides reduction or elimination of incident fields that can affect sensitive circuits as
well as to prevent the emission of components of the system from radiating outside the boundaries limited
by the shield.
Figure 6.2.2 Non - inverting amplifier placed in shielding cabinet (a) with parasitic feedback
and (b) without parasitic feedback
Consider a non-inverting amplifier placed in shielding cabinet as shown in figure (a). C1, C2 and C3
are the three parasitic capacitances between the terminals a, b and common terminal respectively and
shield. C1 and C2 are in the feedback path which results in undesired oscillators. This feedback path can
be eliminated by connecting the shield to common terminal and short circuiting C3 as shown in figure (b).
To isolate electrical devices from electromagnetic interference from external source, shielding of
cabinets (enclosures) which contain instruments is done.
Cabinet shielding reduce the coupling of radio waves, electromagnetic fields and electrostatic fields
An enclosure used to block electrostatic fields which is formed by conductive material or by a mesh
of such materials called Faraday cage is used.
1. Instruments get operating power from AC mains supply, the line voltage is the most common source of
interference. The AC mains supply includes voltage as parasitic capacitors are coupled with the system
or AC main source itself.
2. The temperature is one of the important physical parameter. It plays a vital role in electronic circuits
build with semiconductor devices. The temperature gradient can easily affect the biasing of active
elements used in electronic circuits.
3. The dry or poorly soldered electronic components on the PCB (printed circuit board) are considerably
affected by the mechanical shocks.
4. The ignition systems used in two wheelers and four wheelers can produce low or high frequency spikes
which greatly influence other electronic equipments in operation in the premises.
5. In digital equipments, short pulses called spikes can cause serious interference.
6. The components like thyristors, controlled rectifiers acting as switches in power distribution systems
are one of the sources of interference. Large pulses produced by such switches in system where number
of switches are to be controlled and activated simultaneously cause major part of interference.
7. In mechanical switches,a collector of ac series electrometer acting as fast switch are the sources of
interference due to mechanical influence.
8. A periodic discharge of gas discharge lamp is another source of interference.
9. The high magnitude voltages produce corona effect which finally results in interference.
1. Power quality problems with electronic equipment such as voltage drops, equipment failure, resets, etc.
2. Inaccurate readings and data from signal noise can negatively affect whole process or research, and
might not notice the errors if weren’t aware there was a signal problem in the first place. This symptom
is especially a threat to industrial processes given the sensitivity of newer electrical equipment.
6.6.3 Prevention of interference
2. Shielding: It is defined as method of placing electronic circuits or systems in a protective metal casing
such that the electrostatic and/or electromagnetic fields cannot interfere the sensitive components of
electronic or electrical circuits or systems.
For more information on shielding and grounding refer 6.1 and 6.2.
A functional block diagram in systems engineering and software engineering is a block diagram that
describes the functions and interrelationships of a system.
The block diagram can use additional schematic symbols to show particular properties.
Figure 6.6.4.1 FBD of altitude control and maneuvering electronics system of gemini space craft
Trouble shooting procedure must always start with preliminary analysis of the trouble symptoms,
from which various possibilities of malfunctions are deduced.
To troubleshoot a system, process, or equipment, start by collecting technical records from relevant
sources. These include the OEMs, suppliers, contractors, operators, and maintenance departments.
They are analyzed in order of probability and various checks are usually made to estimate or variety
mutual deduction. Figure 6.7.1shows the summary of generalized trouble shooting procedure.
Example 1: An electrician is troubleshooting a faulty light circuit, where the power source and light bulb
are far removed from one another:
Figure 6.7.2 Trouble shooting the fault between source and light
The electrician uses a voltmeter to check for the presence of voltage between pairs of terminals in
the circuit. Based on the voltage indications electrician can determine the location of the circuit fault.
Example 2:
In this Voltage divider circuit when no load connected, the voltage across the lower resistor would
be precisely 6 volts. With the load connected, the parallel combination of load and 1 kΩ resistor results in
an effective resistance of less than 1 kΩ on the lower half of the divider, resulting in a voltage of less than
half the total supply voltage.
Suppose that something goes wrong in this voltage divider circuit, and the load voltage suddenly
falls to zero. A technician following the “divide-and-conquer” troubleshooting strategy begins by
measuring voltage across the lower resistor (finding 0 volts), then measuring voltage across both resistors
(finding 12 volts).
http://www.interstates.com/newsletters/newsletters-detail/106/7_Steps_for_Electrical_Troubleshooting
https://www.myodesie.com/wiki/index/returnEntry/id/2956