Electrical Measuring Instruments
Electrical Measuring Instruments
10
Measuring Instruments
10.1 INTRODUCTION
The instruments which are used to measure the electrical quantities like curent.
voltage, power, energy, etc. are known as electrical measuring instruments.
Instruments can be divided broadly as absolute and secondary instruments
Absolute instruments are those, which give the value of the quantity to be
measured in terms of the deflection and instrument constant. Tangent
galvanometer and Rayleigh's current balance are few examples of absolute
instruments. These instruments are used in standard laboratories and institutions
as standardizing instruments.
Secondary instruments are those, in which the value of quantity to be meas
ured is determined from the deflection of the pointer, as the scale is calibrated
in terms of the unit of quantity. These instruments are commonly used inla
ratories, institutions, industries and power stations, etc. They are further ckass-
fied into three groups, that is
indicating instruments, recording instrumens
integrating instruments.
Indicating instruments are those which indicate the magnitude ofne meas
etc
uring quantity at that time. Instruments like ammeter, voltmeter, wattmetc
those
are some examples of indicating instruments. Recording instruments arc
The
deflecting device of the measuring instrument converts the quantitu
measured (electric current or potential) into a mechanical force to be
generally kna
as deflecting force which is
responsible for the deflection of the pointer. known
10.5 CONTROLLING DEVICE
The controlling device of an indicating instrument serves the following
es. pupos
avoid
the effect of temperature B
To
ed.
riation he
upon the length of the spring.
1a
springs A
and
and B are wound in opposite
directions.
wind. up
sprin.
twist in the springs pro
The
combined
duce the the desired controlling torque,
of
proportional to the angle
which is deflection of the
angular
twist. The
proportional to the
nointer is directly Fig. 10.1 Spring control
As the angle of twist of
deflecting torque.
he snrings depends upon the angular deflection of the pointer, the controlling
provided by the springs can be made equal and opposite to the deflect-
torque
ing torque, so that the pointer detlects to a final steady position for a particular
magnitude of current.
10.5.3
0.5.3 Comparison Between Spring and Gravity Control
OWing points clearly give the comparison between spring and gravity con-
trol.
The controlling torque increases very slowly in gravity control as it is
proportional to the sine of angular deflection, whereas the increase is
much more in spring eontrol, as the controlling torque in this case is pro-
deflection. Thus the scale is more at the
portional
to the angular
ower end in case of instruments using gravity control. Hence spring con
cramped
trol is better than the contro. gravity
312 Basic Electrical Engineering
c
on
based
Under-damped
Critically-damped
Over-damped
Time
Fig. 10.3 Damping system
the principles of
electromagnetic
induction.
Figure 10.4 illustrates one form of eddy cur-
P
rent damping, where a thin disc D N
of alu-
minium is mounted on a D
spindle P, which
can move between the
poles of a permanent
magnet M. When the spindle moves clock-
M
wise, the disc D cuts through the field of the
permanent magnet M, thereby an emf 1s
induced in the disc. The emf induced in the
disc causes eddy currents as shown by dot-
ted lines in
Fig. 10.4. Eddy currents circulat-
ng in the disc produce the necessary
ing torque. With constant field strength, the
damp M
damping torque produced by eddy currents D
Is
proportional to the velocity of the conduc-
tor. This is the most
efficient type of damp- Fig. 10.4 Eddy current damping9
ing and is used for hot wire,
moving coil and induction type instruments.
10.6.2 Air Damping
Figure 10.5 shows air damping,
where a thin metal vane V is attached
to a spindle S. The vane V can move
in the
sector-shaped box C. The ten-
dency of moving system to oscillate
about its final
steady position is
damped by the action of air on the
Vane. This method of Fig. 10.5 Air damping
ve
y
damping is
Simple and quite inexpensive. However precautions must be taken against
314 Basic Electrical Engineering
the bending of the vane. Normally hot wire and moving iron inct.
provided with this type of damping. aments ate
10.7 TYPES OF INDICATING INSTRUMENTS
Table 10.l gives a brief and clear picture ot various typesof
of electrical :indica
ing instruments along with their methods of control and damping
moving iron ammeters and voltmeters are most suited for ac measurements.
Such discussions on ammeters and voltmeters will be limited to these two typE
connected in series with the circuit. The voltage drop across tneed by e
the ammeter should be as low as possible, so that the
umed
power coI l o w
meter is small. Hence the resistancé of the ammeters should be veryacrossany
Voltmeters are used in ment of voltage.
circuit for the measuremer
a thecircuit
-
Thus, the deflection of the moving coil instrument is directhy proportional to
ne current to be measured. Hence the measuring scale of these instruments is
unitormly calibrated. Because of uniform scale, the measurements are more
accurate in these types of instruments.
10.9.2 Construction
ure 10.6 shows the constructional details ofa permanent magnet moving coil
m e n t . As the name suggests, these instruments consist of a permanent
e t M and a rectangular coil C having a large number of turns of enamelled
h COvered copper wire. The coil is wound on aluminium former supported
i n t e bearings and controlled by spiral hairsprings AA. The euTent is led
nto and out
dout of the coil by spiral hairsprings. The danmping torque is produced
by thee motion of aluminium former in the magnetic field of the permanent
magnet.
the air The coil is free to move
in the field of the permanent magnet, that is in
by P between the pole pieces PP and the soft iron cylinder B supported
rdss plate (not shown in Fig. 10.6). Core B increases the field strength
and gives a radial
rad magnetic flux of uniform flux density. The permanent
agnetSEd
used in
in modern
1 instruments is relatively short in size, because of the
Basic Electrical Engineering
316
M
P
P
C
(a)
(6)
the variation of
temperature. The resistance of swamping resistom Strumet
nearly three limes that of the coil, so that the
will be reduced by about 75 the possible
errOr
per cent.
The moving coil instrument o dit
can also be converted into a
Ler by using shunts of
different values as shown in Fig. 10.7. of with ts
ferent values have been >i series
shunt
Thus IR, = V- V
or
R,- Fig. 10.9 Extension of range
Let the resistance of the voltmeter be R,
Then V = IR
or
1 RV
Thus R, (V-VRn
If n1, the
voltage multiplying factor
(10.2)
Then R, =(m-) R, can be cal-
Hence the proper value of the resistance to be connected in series
e
culated using Eq. (10.2) for a particular case.
Examples on Moving Coil Instruments
with
Example 10.1 A moving
coil instrument gives full scale
deflecuo"
20 mA. The resistance of the coil is 4 2. It is
u m e n t
into an anmeter to read upto 2 A. Find the resistance of the shunt to be connect
anmmeter
narallel with the instrument. Also determine the value of series resistance
edi
far the above instrument to read upto a voltage of 30 V.
Solution:
) Current through the coil = 20x 10 A
Resistance of the coil = 4 Q
Thus voltage across the coil = (20 x 10 ) x 4
= 0.08 V
Current through the shunt = total current to be measured - current through
the coil
Current through shunt = 2 - 0.02
= 1.98 A
(i) Let R be the resistance connected in series with the instrument. Then
resistance between the terminal points = R+4.
Voltage to be measured = 30 V
Current through the coil = 20 mA =0.02 A
30
Thus
0.02 R+4
30 = 1500 Q
or R+4
0.02
Resistance to be connected in series with the moving coil
= 1500-4 = 1496 2
2996 Q
Additional resistance needed = 2996 - 1496
= 1500 2
lotal resistance in series needed to read upto 90 V
90
-4 4500 4
0.02
= 4496 Q
Additional resistance = 4496 - 2996 = 1500 2
320 Basic Electrical Engineering
0.08
Resistance of the shunt for 6 A range = 0=0.01338 S2
Example 10.4 The coil of a moving coil instrument is wound with 50 turns,
the area of the coil being 5 cm, flux density in the gap is .12 Wb/mf. Find the
deflecting torque, when the coil is carrying a current of 15 mA. Also determine
the deflection of pointer, if the constant of the control is 18x 10 N
spring m
pe
degree.
Solution: Deflecting torque in of
case moving coil instruments is given Dy
T =N xB xA xI
Number of turns in the coil, N= 50
Flux density in the gap, B =
0.12 Wb/m
Area of the coil, A 5 cm = 5 x
=
10 m
Current through the coil 15 mA = = 15 x 10 AA
Thus deflecting torque, T 50 x 0.12 x 5 x 10
=
x 15 x 10
= 45x 10° N - m
Controlling torque for spring controlled meters = C8
Constant for the spring, C 18 x 10"/ N
=
m per degree -
ample 10.5 The coil of a moving coil instrument is wound with 80 turns.
Exam and 3 wide. The flux
is
Thecoil ol
cm long cm
density in the air gap is 0.5 Wb/m.
spring exerts a torque of 2.0 x 1 0 Nm, when the deflection is 100
Asions on full scale. Estimate the resistance which must be connected in series
with the coil to give 2 V per division. The resistance of the instrument coil may
the series resistance to be connected.
he neglected compared with
Solution:
Deflecting torque in moving coil instruments is given by.
Ta =NxBxAxI
Number of turns in the coil, N 80 =
2
Thus 200 600 xR
200 x 600
= 60,000 2
Series resistance, R =
0.1 V
10) x 10 =
=
(10 x
be extended = 1A
lotal current for which the range is to
the shunt
hus c u e n t to be diverted through
= 1.0 0.01 = 0.99 A
S2
Let the resistance of the shunt be r
= 14.9 V
Ta = T
mov result,
sult, scale
these instruments will be operating current. ASa lower sca e n da. n d
noid.
ii) Repulsion type-in which two parallel strips of soft iron magnetized
inside a solenoid repel each other.
for the types (attraction and repulsion) is directly
The deflecting torque both
nroportional to the square of current for a given position of the moving system.
waveform and corresponding detlecting torque in a
Figure 10.10 shows current
moving iron instrument, which clearly indicates, that if the frequency of current
hundred times
wave is 50 Hz, torque varies between zero and maximum value,
a second.
Hence the moving system takes up a position corresponding to the
mean torque, because
of its inertia. In both types of moving iron instruments,
to the voltage to be meas-
the current to be measured (or a current proportional
ured) is passed through a coil. The number of turns in the coil will depend upon
the current passing through the coil. For proper operation of the instrument,a
certain number of ampere turns is needed, which can be made up by having a
Deflecting
torque
m
Mean torque
Time
Current
instrument
Fig. 10.10 Deflecting torque in moving iron
Airdamping
chamber
Pointer
Coil
winding
Balance
weight Movingiron
Control
0cor
weight
Fig. 10.11 Attraction type moving iron instrument
ber.
nents
) Non-uniform scale.
withmeasurement
the load and
of power. The fixed coil or the field coil is connected in series
fixed coil forms thuhus carries the current flowing in the load circuit. Hence the
the current coil of the wattmeter. The moving coil of the
wattmeter
the l
is con
connected across the load and thus carries a current proportional to
voltage.
In order
to limit the current through the moving coil, a high non-
ctive resistance Is connected in series with it. As the moving coil carries a
current prop
Sure coportional to the voltage across the load, it is normally called the pres-
coil of the
wattmeter.
The electrodynan
nd Odynamometer type wattmeter consist of two current cols CCi.
give CC, as
shown Fig. 10.13, which can be connected in series or parallel to
two
ived
current
Cnt ratings.
Is achieved.and if the
ratings. If the coils are connected in series, lower current range
curre
rent coils are connected in parallel higher current range
Basic Electrical Engineering
326
Pointer
Moving coil
Fixed coils
PC
CC& CC
CC CC2
O
Supply
A
scale. nected
Based on the above discussion, the two coils of the wattmeter are co
e load
connected in series witn u
properly for measuring power. The fixed coil is ortion
and moving coil across the supply. The current in the moving coil isprop the instant
Spindle
Magnetic core
Load
CC CC
is
voltage is to be measured, ange o fthe
(volti
h e volmmet