Atlan A300
Atlan A300
Atlan A300
Atlan
A300, A300 XL, A350, A350 XL
2 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Contents
Contents
2 Safety-related information....................................................................... 9
2.1 Intended use ................................................................................... 9
2.2 Indications....................................................................................... 9
2.3 Contraindications ............................................................................ 9
2.4 Environments of use ....................................................................... 10
2.5 Essential performance features ...................................................... 10
2.6 User group requirements ................................................................ 11
2.7 Information on safety instructions and precautionary statements... 11
2.8 Safety instructions .......................................................................... 12
2.9 Additional information ..................................................................... 14
3 Overview ................................................................................................... 15
3.1 Hardware ........................................................................................ 15
3.2 Functional scope............................................................................. 31
3.3 Gas flow diagram............................................................................ 42
5 Operating concept.................................................................................... 79
5.1 Screen ............................................................................................ 79
5.2 Color concept.................................................................................. 83
5.3 Selecting and setting ...................................................................... 84
6 Getting started.......................................................................................... 86
6.1 Safety instructions .......................................................................... 86
6.2 Turning on the device ..................................................................... 86
6.3 Checking the device configuration.................................................. 86
6.4 Checking the operational readiness ............................................... 87
6.5 Emergency start-up ........................................................................ 88
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 3
Contents
7 Operation .................................................................................................. 89
7.1 Safety instructions .......................................................................... 89
7.2 Starting the therapy ........................................................................ 93
7.3 Adjusting the therapy ...................................................................... 96
7.4 Special forms of therapy ................................................................. 100
7.5 Using fields with special functions .................................................. 103
7.6 Customizing the screen display ...................................................... 107
7.7 Displaying additional data............................................................... 113
7.8 Setting the volume .......................................................................... 115
7.9 Adjusting the alarms ....................................................................... 116
7.10 Changing the patient data............................................................... 119
7.11 Exporting data................................................................................. 119
7.12 Other settings ................................................................................. 120
7.13 Ending the therapy.......................................................................... 122
7.14 Change of patient ........................................................................... 124
11 Configuration............................................................................................ 165
11.1 Device settings ............................................................................... 165
11.2 Setting the date and time................................................................ 165
11.3 Specifying the start settings............................................................ 166
11.4 Transferring device configurations.................................................. 181
11.5 Activating software options ............................................................. 182
11.6 Overview of configurable screen contents...................................... 184
4 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Contents
13 Reprocessing............................................................................................ 222
13.1 Safety instructions .......................................................................... 222
13.2 Information on reprocessing ........................................................... 223
13.3 Classifications for reprocessing ...................................................... 224
13.4 Reprocessing list ............................................................................ 225
13.5 Reprocessing procedures............................................................... 225
13.6 Preparation ..................................................................................... 227
13.7 Surface disinfection with cleaning................................................... 229
13.8 Disassembly ................................................................................... 230
13.9 Disassembling the breathing system .............................................. 238
13.10 Machine reprocessing..................................................................... 242
13.11 Fitting and assembly....................................................................... 249
13.12 Preparation before the next use ..................................................... 258
13.13 Brief instructions for reprocessing .................................................. 260
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 5
Contents
6 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Information about this document
1.3 Illustrations
Illustrations of products and screen content in this document may differ from the
actual products depending on configuration and design.
1.4 Trademarks
1.4.1 Trademarks owned by Dräger
Trademark
Atlan®
AutoFlow®
Infinity®
D-Vapor®
Drägersorb®
MEDIBUS®
ServiceConnect®
WaterLock®
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 7
Information about this document
Trademark
E-Vent®
The following web page provides a list of the countries in which the trademarks are
registered: www.draeger.com/trademarks
Korsolex®
Neodisher
Dr. Weigert
Mediclean®
acryl-des®
Mikrozid® Schülke & Mayr
Perform®
Actichlor®
Ecolab USA
OxyCide®
BruTab 6S® Brulin
Dispatch® Clorox
Klorsept® Medentech
Descogen®
Antiseptica
Oxygenon®
Virkon® DuPont
SteriMax® Aseptix
8 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Safety-related information
2 Safety-related information
2.1 Intended use
This device is intended for use in anesthetizing adults, pediatric patients, and
neonates. The device can be used for mechanical ventilation, manual ventilation,
pressure-supported spontaneous breathing, and spontaneous breathing.
The device is equipped with the following basic functions:
– Ventilation monitoring
– Inspiratory O2 measurement
– Device monitoring
– Anesthetic gas receiving system
The following options are additionally available:
– Patient-gas measurement module for O2, CO2, N2O, and anesthetic gases
– O2 insufflation
Anesthesia is achieved through a mixture of pure oxygen and Air (medical
compressed air) or pure oxygen and nitrous oxide, with the addition of volatile
anesthetic agents.
Ventilation is accomplished on the patient through a laryngeal mask, a breathing
mask, or an endotracheal tube.
The integrated breathing system can be used with partial rebreathing (low-flow or
minimum-flow).
A non-rebreathing system, such as the Bain, Mapleson, Kuhn, or Waters system,
may be used at the external fresh-gas outlet.
2.2 Indications
The device is specified for inhalational anesthesia and/or patient ventilation in
accordance with the intended use during surgical or diagnostic interventions.
2.3 Contraindications
The device has no product-specific contraindications.
It is the responsibility of the user to select the appropriate treatment for the patient’s
underlying disease. The patient's condition must be continually monitored for any
potential changes.
The safety information must be followed for patients suspected of malignant
hyperthermia, patients with ketoacidosis, and patients who are under the influence
of alcohol. Observe the following information: "Therapy and applications", page 89.
The device administers medical gases such as O2, N2O, Air (medical
compressed air), and volatile anesthetic agents (halothane, enflurane, isoflurane,
sevoflurane, desflurane). For contraindications to the applied medical gases, strictly
follow the instructions for use of the medical gases.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 9
Safety-related information
General
– Supply of the anesthesia workstation with O2:
If the O2 supply (central gas supply or gas cylinder) fails, an alarm is issued.
– Supply of the patient with adequately oxygenated breathing gas:
If the breathing gas contains insufficient levels of O2, an alarm is issued.
– Supply of an adequate anesthetic gas concentration to the patient:
When the anesthetic gas is measured by means of an integrated patient-gas
measurement module, an alarm will be generated if the anesthetic gas
concentrations are too high.
– Monitoring of the airway pressure:
Alarms are issued depending on the set alarm limits.
Gas measurement
– Breathing gas monitoring:
– Set values for FiO2
– Inspiratory and expiratory measured values for O2, CO2, N2O, and
anesthetic gas; automatic anesthetic agent identification (patient-gas
measurement module)
The gas composition is measured with ISO accuracy.
– Monitoring of breathing gas concentrations:
Alarms will be issued depending on the set alarm limits or if the gas
measurement fails.
10 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Safety-related information
Reprocessing personnel
This user group performs reprocessing activities.
Service personnel
This user group installs the product and performs service activities.
If product-specific skills or tools are required, then the service activities must be
performed by specialized service personnel. The specialized service personnel
have been trained by Dräger to perform these specific service activities on this
specific product.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 11
Safety-related information
The following warning signs and signal words indicate precautionary statements
and differentiate the possible consequences of non-compliance.
12 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Safety-related information
2.8.6 Device
Penetrating liquid
Penetrating liquid may cause the following:
– Damage to the device
– Electric shock
– Device malfunctions
Personal injury and property damage may occur as a consequence.
► Ensure that no liquid penetrates the device.
► Do not place any containers containing liquids above or on the device.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 13
Safety-related information
Housing
Under the housing, there are live electrical components, which may cause an
electric shock.
► The housing may only be opened by those user groups that are assigned to that
particular measure.
2.8.7 Service
If service activities are not performed regularly or properly, malfunctions may occur
that can result in personal injury and property damage.
► Perform the service in accordance with the chapter "Service".
2.8.8 Reprocessing
Reusable products must be reprocessed, otherwise there is an increased risk of
infection.
► Perform the reprocessing in accordance with the chapter "Reprocessing".
2.9.2 Training
Training for users is available via the Dräger organization responsible (see
www.draeger.com).
14 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
3 Overview
3.1 Hardware
3.1.1 Front
3.1.1.1 Large version
36437
No. Designation Description
1 Screen Enables user inputs by touchscreen and
rotary knob.
2 Plug-in connectors for vaporiz- Enable the connection of up to 3 vaporizers.
ers
3 Backup manual switch (behind Used to switch to backup manual mode in
the flap) the event of malfunctions of the device.
4 Gas mixing unit Produces a gas mixture from the selected
gases (e.g., O2 and Air).
5 Work surface Used for storing, e.g., documents.
6 Pull-out writing tray (option) Provides additional work surface.
7 Viewing window for piston ven- Allows visual checking of the movement of
tilator the piston ventilator.
8 Lockable drawer Provides additional storage space.
9 Trolley with footrest and cen- Used for moving the device. The central
tral brake brake locks the two front castors.
10 CO2 absorber Absorbs CO2 from the patient’s breathing
gas.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 15
Overview
36438
16 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
3.1.2 Screen
36441
No. Designation Description
1 Touchscreen Calls up functions or dialogs when touched.
2 Alarm silence key Suppresses the alarm tones of all active
alarms for 2 minutes.
3 Rotary knob Used for selecting, adjusting, and confirm-
ing settings. Lights up in color in certain sit-
uations.
4 Working light Illuminates the work surface.
5 key Turns the working light on or off. Dims the
illuminance in 3 steps (dark, medium, and
bright).
6 key Turns the device on or off.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 17
Overview
36446
No. Designation Description
1 Pressure gauge (option) Indicates the airway pressure in the internal
breathing system.
2 Inspiratory port Used to connect the inspiratory hose to the
device.
3 Bag elbow with circuit plug Used to connect the breathing bag hose.
The circuit plug is used to seal the Y-piece
during an automatic test.
4 APL valve Limits the maximum airway pressure in the
Manual / Spontaneous mode.
5 Breathing system cover Protects the breathing system underneath
and provides climate control for the breath-
ing system.
6 Expiratory port Used to connect the expiratory hose to the
device.
7 Holder Used for parking the breathing bag hose.
8 Handles Used for aligning the device and for stowing
used breathing hoses.
9 CO2 absorber Absorbs CO2 from the patient’s breathing
gas.
10 Anesthetic gas receiving sys- Used for suctioning and scavenging excess
tem anesthetic gas and breathing gas, to reduce
the anesthetic gas concentration released
into the environment by the anesthesia
machine, and to scavenge the sample
gases from an anesthetic gas monitor.
18 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
36447
No. Designation Description
1 O2 sensor Measures the inspiratory O2 concentration
in the breathing gas.
2 Sealing cap Seals the O2 sensor port during calibration
of the O2 sensor.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 19
Overview
36462
No. Designation Description
1 Castor brake Prevents movement of the rear castor.
2 Gas cylinder holder (option) Secures the gas cylinders.
3 Rail Used for fastening additional components.
4 Hanger yoke system (option) Enables the connection of gas cylinders
with a pin-index connector.
20 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
36468
No. Designation Description
1 Standard rail with handle Allows the device to be maneuvered during
intrahospital transport and also the attach-
ment of accessories.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 21
Overview
36469
No. Designation Description
1 Column cover Depending on the version, allows a patient
monitor or other workplace components to
be mounted.
2 Holder for hoses and cables Used for storing and winding up central sup-
ply hoses and cables.
3 Cable holder with cable For passing hoses and cables through.
channels
22 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
3.1.7 Rear
3.1.7.1 Version with screw connections for standing gas cylinders
36470
No. Designation Description
1 Storage compartment Can be used as storage space.
2 Holder for hoses and cables Used for storing and winding up central sup-
ply hoses and cables.
3 Gas supply block Provides connectors for gases from the
central gas supply system and for gas cylin-
ders.
4 Gas cylinder holder (option) Secures the gas cylinders.
5 Strain relief for compressed Protects the compressed gas hoses from
gas hoses loosening inadvertently.
6 Auxiliary power sockets Allow other devices to be connected.
(option)
7 Vent Discharges warmed air from the device into
the environment.
8 Ventilation slot Feeds ambient air to the device for ventila-
tion.
9 Connectors Provides connectors for power cable,
potential equalization, and interfaces. Used
for data exchange between external
devices, additional components, and net-
works.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 23
Overview
3.1.7.2 Version with pin-index connector for suspended gas cylinders (option)
36471
No. Designation Description
1 Storage compartment Can be used as storage space.
2 Holder for hoses and cables Used for storing and winding up central sup-
ply hoses and cables.
3 Gas supply block Provides connectors for gases from the
central gas supply system and for gas cylin-
ders.
4 Hanger yoke system (option) Enables the connection of gas cylinders
with a pin-index connector.
5 Strain relief for compressed Protects the compressed gas hoses from
gas hoses loosening inadvertently.
6 Vent Discharges warmed air from the device into
the environment.
7 Gas cylinder holder (option) Secures the gas cylinders.
8 Auxiliary power sockets Allow other devices to be connected.
(option)
9 Ventilation slot Feeds ambient air to the device for ventila-
tion.
10 Connectors Provides connectors for power cable,
potential equalization, and interfaces. Used
for data exchange between external
devices, additional components, and net-
works.
24 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
3.1.8 Connectors
36474
No. Designation Description
1 Serial port (COM 1 and This serial port (RS232) supports the MEDI-
COM 2) BUS.X protocol for data exchange between
the anesthesia machine and external
devices.
2 Main switch Turns the device off and minimizes the cur-
rent consumption when the power plug is
pulled out. To be used when the device is to
be disconnected from the power supply for
longer than 2 weeks and during service
activities.
3 Connector for workplace light Used for connecting an external workplace
(option) light.
4 Power inlet (connector for Used to connect the device to the mains
power cable) power supply.
5 Potential equalization pin Used for connecting a potential equalization
cable. This will minimize differences in elec-
trical potential.
6 Network port Enables data transfer within an IT network.
7 USB port Used to transfer data to a USB mass stor-
age device.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 25
Overview
36475
No. Designation Description
1 Connectors for pressure mea- Used for connecting pressure measuring
suring lines for gas cylinders lines of the pressure reducers on the gas
(option) cylinders.
2 Connectors for central gas Enable the device to be supplied with gases
supply system from the central supply.
3 Connectors for gas cylinders Enable the device to be supplied with gases
(option) from the gas cylinders.
4 Label Advanced Cylinder Sup- On devices that are equipped with
port Advanced Cylinder Support, the gas cylin-
der valves can also remain open during
operation with the central supply.
26 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
36476
No. Designation Description
1 Protection bar Protects the connectors for the gas cylin-
ders from damage.
2 Hanger yoke system with three Enables the connection of gas cylinders
pin-index connections (option) with a pin-index connector.
3 Wrench For opening and closing the gas cylinder
valves.
4 Connecting cable Enables electronic gas pressure measure-
ment.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 27
Overview
36477
No. Designation
1 Fuses, 2 each per power socket
2 Auxiliary power sockets, 4 pcs.
3 Main fuse
28 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
36479
No. Designation
1 Status display
2 Symbols for mains power supply and power supply from internal battery
3 Symbols for gas supply (O2, Air, N2O or O2, Air) from central supply and gas
cylinders
4 O2 flowmeter (for O2 insufflation Aux. O2 and emergency O2 delivery
Add. O2)
5 O2 switch (for switching between O2 insufflation Aux. O2 and emergency
O2 delivery Add. O2)
6 Outlet for O2 insufflation, e.g., for nasal cannula
7 Display of pressure in the internal breathing system, see page 18
8 O2+ key (O2 flush)
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 29
Overview
36480
No. Designation
1 Status display
2 Symbols for mains power supply and power supply from internal battery
3 Symbols for gas supply (O2, Air, N2O or O2, Air) from central supply and gas
cylinders
4 O2 flowmeter (for O2 insufflation Aux. O2) (option)
5 Outlet for O2 insufflation, e.g., for nasal cannula (option)
6 Flow control valves (O2, Air, N2O or O2, Air)
7 Total flow tube
8 Display of the set fresh-gas flows
9 Display of pressure in the internal breathing system, see page 18
10 O2+ key (O2 flush)
30 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
Hardware options
Hardware options are listed in the following table:
Name Description
Compact version (model – Version with small trolley for environments of use
designation A300, A350) with constricted space
– 1 large drawer
– Central brake
– Available with plug-in connectors for 1 or 2 vaporiz-
ers
– The work surface can be enlarged with a folding
table extension on the side (option).
– When heavy workplace components, monitors, or
syringe pumps are attached, a counterweight is
required for increased tipping stability (option).
Large version (model des- – Version with large trolley for normal OR environ-
ignation A300 XL, ments with adequate space
A350 XL) – 1 large and 2 small drawers
– Available with plug-in connectors for 2 or 3 vaporiz-
ers
– Can be fitted with a pull-out writing tray and a fold-
ing table extension on the side (option).
Mechanically controlled – The fresh-gas delivery is adjusted by means of
gas mixer (model desig- manually operated needle valves.
nation A300) – The individual fresh-gas flows are measured elec-
tronically and displayed on both the status display
and the screen.
– The total fresh-gas flow is indicated on the inte-
grated total flow tube.
– Available as a 2-gas version (O2/Air) or a 3-gas
version (O2/Air/N2O)
– Available with O2 flowmeter for O2 insufflation
(option)
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 31
Overview
Name Description
Electronically controlled – The fresh-gas delivery is adjusted using the screen
gas mixer (model desig- with the aid of settable parameters for O2 concen-
nation A350) tration in % and for fresh-gas flow in L/min.
– Depending on the version, Air or N2O can be
selected as the carrier gas.
– With O2 flowmeter for O2 insufflation
– With integrated, mechanical emergency O2 deliv-
ery Add. O2
External fresh-gas outlet – Allows use with external non-rebreathing systems,
e.g.:
– Mapleson
– Kuhn
– Bain
– Magill
– Waters
Advanced Cylinder Sup- On devices that are equipped with Advanced Cylinder
port Support, the gas cylinder valves can also remain open
during operation with the central supply.
Integrated O2 monitoring – Measurement, monitoring, and display of the inspi-
ratory O2 concentration
– The measurement is performed by the O2 sensor
integrated in the breathing system.
Integrated patient-gas – Measurement, monitoring, and display of the inspi-
measurement module ratory and expiratory gas concentrations of O2,
anesthetic gases, CO2, and N2O
– Detection and indication of anesthetic gas mixtures
– Display of the xMAC
O2 detection – During the self-test, a check is made using the inte-
grated patient-gas measurement module to ensure
that the connected O2 supply is actually delivering
O2.
Support of Infinity ID Allows the use of Dräger Infinity ID accessories with
accessories the following functions:
– Generates a message when the maximum period
of use is exceeded for the breathing circuit, the
water traps, the CO2 absorber, and the flow sensor
– Generates a notice when the breathing circuit is
incorrectly connected
– Generates a message if the CO2 absorber is not
present or is not locked
Active anesthetic gas Anesthetic gas scavenging with flow indicator for use
scavenging with an active disposal system with a wall terminal unit
Passive anesthetic gas Anesthetic gas scavenging for gas disposal without an
scavenging active disposal system
32 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
Software options
Software options are listed in the following table:
Name Description
Spontaneous breathing Allows assisted ventilation with pressure support and,
support in controlled ventilation modes, synchronization with
the patient’s inspiratory effort.
Includes the following ventilation modes:
– CPAP / PSV
– PC - SIMV
– PC - SIMV / PS
– VC - SIMV
– VC - SIMV / PS
– VC - CMV / AutoFlow (requires AutoFlow option)
– VC - SIMV / AutoFlow (requires AutoFlow option)
– VC - SIMV / PS / AutoFlow (requires AutoFlow
option)
AutoFlow With AutoFlow, the set tidal volume VT is applied with
all mandatory volume-controlled breaths at the lowest
required pressure.
With switchable synchronization and settable pressure
support (requires the spontaneous breathing support
option)
Advanced trends Includes the following functions:
– Graphical trends of measured values
– Mini-trends next to the waveforms
– Export of trend data to USB mass storage device
Advanced ventilation Includes the following functions:
monitoring – Display of patient compliance with trend
– Display of loops (Pressure-Volume and Flow-Vol-
ume)
– Volumeter (bar graphic for monitoring the inspira-
tory and expiratory tidal volumes)
– Display of the patient-triggered, mechanically sup-
ported minute volume compared with the manda-
tory minute volume
Advanced gas monitoring Includes the following functions:
– Indicator and trend for efficiency of fresh-gas set-
ting and anesthetic agent consumption, (econom-
eter and low-flow wizard (no trend))
– Display of gas consumption
– Display of anesthetic agent consumption and
uptake
– Display of MV×CO2 with trend
– Display of O2 uptake with trend
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 33
Overview
Name Description
Advanced neonatal sup- Includes advanced ventilation functions and monitor-
port ing functions for the ventilation of neonates:
– Minimum settable tidal volume of 5 mL
– Higher sweep speed
– Increased flow measurement sensitivity for more
precise ventilation monitoring
Expert view Provides the following advanced views:
– 4 waveforms
– 3 waveforms and one row with parameter fields
34 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
– VC - SIMV / AutoFlow
– VC - SIMV / PS / AutoFlow
For a detailed description of the ventilation modes and the additional settings, see
page 298.
3.2.6 Monitoring
The device can monitor the following:
– Airway pressure
– Minute volume
– Inspiratory O2 concentration
– Inspiratory and expiratory anesthetic gas concentrations (only available with the
integrated patient-gas measurement module)
– Inspiratory and expiratory CO2 concentrations (only available with the integrated
patient-gas measurement module)
– Inspiratory and expiratory N2O concentrations (only available with the integrated
patient-gas measurement module)
– Apnea (pressure, flow, and CO2)
– Occurrence of anesthetic gas mixtures (only available with the integrated
patient-gas measurement module)
– Lack of fresh gas in the breathing system and the breathing circuit
3.2.8 Logbook
The device can capture and store the following data, among other things:
– Measured values
– Set values and related changes
– Patient data
– Ventilation modes
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 35
Overview
– Events (e.g., alarms, confirmed alarms, switch-on time and switch-off time)
– Test results
– Gas consumption
– Anesthetic agent consumption
Central
Gas Gas cylinders
supply
O2 Yes Permanently mounted Dräger pressure reducer
(option)
or
Third-party manufacturer pressure reducer
Air Yes Permanently mounted Dräger pressure reducer
(option)
N 2O Yes (option) or
Third-party manufacturer pressure reducer
36 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
If the connected network offers an NTP service, the time on the device can be
synchronized with the time on the NTP server.
For further information see: "Connections to IT networks", page 294.
CAUTION
Risk of increased anesthetic agent delivery
When the emergency O2 delivery (Add. O2) is in use, anesthetic agent continues to
be delivered into the breathing system in accordance with the vaporizer setting.
When the emergency O2 delivery is used during low-flow anesthesia or minimal-
flow anesthesia, an increased quantity of anesthetic agent may enter the breathing
system. This may lead to an increased anesthetic gas concentration.
► Carefully monitor the gas mixture.
36034
2
3. Open the flow control valve (2) on the O2 flowmeter and set the desired flow.
This O2 flow flows through the vaporizer.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 37
Overview
40676
2. Activate the backup manual switch (2). Follow the instructions on the product
label (3) or the screen (4).
39762
38 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
40677
2. Activate the backup manual switch (2). Follow the instructions on the product
label (3) or the screen (4).
39763
3. Ventilate the patient manually.
a. During mechanical ventilation or in the Man/Spon ventilation mode:
Ventilate manually with the breathing bag.
b. When using the external fresh-gas outlet:
Ventilate manually with the breathing bag on the non-rebreathing system.
4. Set the fresh-gas flow.
5. Check the vaporizer setting.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 39
Overview
3.2.12.3 Overview
The following table gives an overview of the integrated safety functions which come
into effect if problems occur during operation:
40 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 41
Overview
36482
No. Designation
1 Fresh gas (O2, Air, N2O) and anesthetic gas
2 CO2 absorber
3 Changeover between mechanical ventilation and Manual / Spontaneous
4 Anesthetic gas receiving system
5 Anesthetic gas scavenging valve
6 APL valve
7 Breathing bag
8 PEEP/Pmax valve
9 Patient-gas measurement module
10 Expiratory valve
11 Expiratory pressure measurement
12 Pressure gauge (option)
13 Patient
14 Expiratory flow sensor
15 Inspiratory flow sensor
16 Inspiratory valve
17 Inspiratory pressure measurement
18 Inspiratory O2 sensor
19 Ventilation drive
20 Fresh-gas decoupling valve
42 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
36484
No. Designation
1 Gas supply (central supply or gas cylinders)
2 Gas mixer
3 O2 flowmeter
4 O2 switch
5 Vaporizer
6 Ejector
7 O2 flush
8 Breathing system
9 Switch-over valve
10 External fresh-gas outlet
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 43
Overview
36485
No. Designation
1 Gas supply (central supply or gas cylinders)
2 Flow control valves
3 Minimum O2 delivery
4 Total flow tube
5 Vaporizer
6 Ejector
7 O2 flush
8 Breathing system
9 Switch-over valve
10 External fresh-gas outlet
11 O2 flowmeter (option)
44 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 45
Assembly and preparation
CO2 absorber
Moisture losses occur if fresh gas is continuously passed through the soda lime. If
the moisture falls below the minimum level, the following adverse reactions occur
regardless of the type of soda lime and inhalational anesthetic agent used:
– Reduced CO2 absorption and consequently an increase in inspiratory CO2
values
– Increased generation of heat in the CO2 absorber and consequently increased
breathing gas temperature
– Formation of carbon monoxide
– Absorption and/or degradation of the inhalational anesthetic agent
► Check the soda lime for color changes regularly and replace if necessary,
especially if the inspiratory CO2 value increases unexpectedly.
► Do not use unnecessarily high fresh-gas flows.
► Only use the O2 flush when it is required.
► With electronically controlled gas mixer:
Only use the emergency O2 delivery when it is required.
► With mechanically controlled gas mixer:
Do not leave the flow control valves open for an unnecessarily long period of
time.
► Use a suitable soda lime such as Drägersorb Free. Do not use any soda lime
based on potassium hydroxide.
Soda lime is caustic and is a strong irritant for eyes, skin, and airway.
► Handle the soda lime carefully and do not spill it.
46 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
Battery supply
If the battery is not sufficiently charged, it may not be possible to maintain operation
for long enough if the mains power supply fails. If there is a mains power supply
failure, devices connected to the auxiliary power sockets are not supplied from the
internal battery. As a result, the patient could be put at risk.
► Before first operation or after storage, charge the battery for at least 8 hours.
► Check the functional integrity of the battery by performing regular inspections.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 47
Assembly and preparation
Interfaces
Connecting devices to the data interfaces (serial ports and network ports) can lead
to an increased leakage current. If the protective ground conductor of one of these
devices fails, the patient leakage current may rise above the permissible values.
This may result in an electric shock and the user and patient may be at risk.
► Only use USB devices that do not have their own power supply.
► Only connect devices or networks to a serial port, or to the network port, that
have a maximum nominal voltage of 24 V DC and meet one of the following
standards:
– IEC 60950-1: Ungrounded SELV circuits
– IEC 60601-1 (as of 2nd edition): Touchable secondary circuits
► Have the leakage current checked by service personnel.
► If the permissible value is exceeded, disconnect the devices from the serial
ports.
► Do not touch the interface ports and the patient simultaneously.
Flow sensors
Residues which are not removed during reprocessing can damage the measuring
wires in the flow sensor or cause fire. As a result, user and patient could be put at
risk.
► Check the flow sensor before insertion and at regular intervals afterwards for
visible damage, soiling, and particles.
► Replace flow sensors when damaged, soiled, or not particle-free.
Drugs or other substances based on readily flammable substances may be ignited
by the flow sensor. As a result, user and patient could be put at risk.
► Do not nebulize drugs or other substances based on readily flammable
substances or spray them into the device.
► Do not use any substances containing alcohol.
► Do not allow any combustible or explosive substances to get into the breathing
system or the breathing circuit.
► Do not use cyclopropane or ether.
48 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
Pressure reducers
Pressure reducers have an internal pressure release valve. If a fault occurs, gas
may escape into the ambient air. Personal injury or property damage may result as
a consequence.
► Do not block or cover the pressure release valve.
Strangulation
Negligent placement of hoses, cables, and similar device components can put the
patient at risk.
► Use particular caution when establishing connections to the patient.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 49
Assembly and preparation
Transport
If the device collides with an obstacle during transport, the pressure reducers may
be damaged. Take the following measures before transporting:
► Align the pressure reducers so that they are protected from collisions.
► Close the valves on the gas cylinders.
During transport, the device may tip over due to incorrect handling or carelessness.
Personal injury or property damage may result as a consequence.
► To push or pull the device, hold it by the standard rail with the handle (right-hand
side of device).
► The device may only be moved by persons who have the physical ability to do
so.
► Always have the device moved by 2 persons for better maneuverability and
when it is being transported over sloping surfaces.
► When the device is moved over inclines, around corners, or over thresholds
(e.g., through doors or in elevators), make sure that it does not bump against
anything.
► Do not pull the device over hoses, cables, or other obstacles lying on the floor.
► Do not operate either the central brake or any of the castor brakes while the
device is being moved.
► Do not lean against the device.
50 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
Devices connected to the gas supply may be damaged by inadequate gas quality.
The use of non-medical gas can result in gas compositions that impair the
functional integrity of the device.
► Use only medical gases.
► Follow the national and international standards regarding the use of medical
gases.
Impermissible supply pressures or using oxygen supplies with less than 100 % O2
may cause incorrect gas composition. As a result, the patient could be put at risk.
► Check the supply pressures of the central supply and of the gas cylinders before
operation.
► When using O2, use only 100 % O2.
The following effects may occur when using O2 concentrators and may put the
patient at risk:
– Deviations between the set value and the actual value for fresh-gas flow and
O2 concentration in the fresh gas
– Inaccurate measured values for volume, anesthetic agent consumption,
econometer, and low-flow wizard
– Accumulation of argon in low-flow operation and minimal-flow operation
► Do not use any O2 concentrators.
A failure of the gas supply can result in a risk of patient injury. In the following
cases, the availability of the gas cylinders cannot be monitored and the backup
functionality may be put at risk:
– Instead of a Dräger pressure reducer, a pressure reducer without the required
pressure sensor is used.
– A central supply hose is connected to the connector for the compressed gas
cylinders.
► If monitoring of the gas cylinders is not available, suitable pressure monitoring
conforming to ISO 80601-2-13 must be used. This will allow the user to read the
cylinder pressures from the user's operating location.
► Do not connect central supply hoses to the connectors for gas cylinders.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 51
Assembly and preparation
Column cover
The column cover may be loaded with a maximum of 15 kg (33 lbs).
Rear
The rear may be loaded with a maximum of 40 kg (88 lbs). In addition to the
accessories, this must also take account of the weight of gas cylinders including the
pressure reducers and the hanger yoke system.
Other components
Component Maximum weight
Writing tray 20 kg (44 lbs)
Standard rail 10 kg (22 lbs)
Large drawer 3 kg (6.6 lbs)
Small drawer 2 kg (4.4 lbs)
52 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
Column cover
The column cover may be loaded with a maximum of 15 kg (33 lbs).
Rear
The rear may be loaded with a maximum of 40 kg (88 lbs). In addition to the
accessories, this must also take account of the weight of gas cylinders including the
pressure reducers and the hanger yoke system.
Other components
Component Maximum weight
Writing tray 20 kg (44 lbs)
Standard rail 10 kg (22 lbs)
Large drawer 3 kg (6.6 lbs)
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 53
Assembly and preparation
Column cover
The column cover may be loaded with a maximum of 15 kg (33 lbs).
Rear
The rear may be loaded with a maximum of 40 kg (88 lbs). In addition to the
accessories, this must also take account of the weight of gas cylinders including the
pressure reducers and the hanger yoke system.
Other components
Component Maximum weight
Writing tray 20 kg (44 lbs)
Standard rail 10 kg (22 lbs)
Large drawer 3 kg (6.6 lbs)
54 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
The breathing system must be reprocessed before the device is operated for the
first time.
39637
1
After the O2 sensor cell has been inserted, the initialization phase of the O2
sensor takes place. The initialization lasts 30 minutes. The 30 minutes start when
the mains power supply is established.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 55
Assembly and preparation
2. Turn the knurled nut (2) about 90° clockwise. Turn it until the palpable resistance
is overcome and a click is heard. The sensor cap is now fitted.
37179
3. Insert the O2 sensor (3) into the sensor port.
38271
4. Turn the O2 sensor (4) clockwise.
38272
56 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
5. Put the breathing system cover (5) in place and click it into position.
36505
4.3.3 Establishing the mains power supply
The mains voltage must correspond to the voltage range indicated on the rating
plate on the rear of the device.
To protect from inadvertent disconnection of the power cable, the power inlet of the
device is secured with a guard plate.
1. WARNING
Risk due to incorrect mains voltage or missing protective ground
conductor
If the device is connected to a power socket with incorrect mains voltage or a
power socket without a protective ground conductor, an electric shock may
occur.
► Connect the device only to power sockets with correct mains voltage and a
protective ground conductor.
WARNING
Risk of electric shock
If the device is connected to an additional power socket strip, this may lead to
increased leakage current. The leakage current may exceed the permissible
values.
► Do not connect the device to additional power socket strips.
► Do not connect additional power socket strips to the integrated auxiliary
power sockets on the device, but rather to separate power sockets on the
wall.
Plug the power plug into the power socket.
The power plug must be readily accessible so that the power supply to the
device can be interrupted quickly in the event of a device malfunction.
2. Using the symbols on the status display, check that the power supply is
established.
3. Turn on the device.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 57
Assembly and preparation
WARNING
Risk of fire
Components such as power supply units that heat up are unable to cool down in
enclosed storage locations and may cause a fire.
► Do not keep components that heat up in the drawers or in the storage
compartment at the rear of the device.
► WARNING
Risk of overloading the network
The following hazardous situations may occur if the network does not possess
the required characteristics:
► Exported patient-related data (age, weight, height) and therapy-related data
may be intercepted, falsified, or damaged.
► An overload of the device due to high network load (e.g., caused by denial-
of-service attacks) may lead to a shut-down of the device’s network port. The
network port will not be available again until the device is restarted.
Connect the device to a network or a computer.
An illustration of the ports can be found on page 25.
58 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
To use the anesthetic gas compensation - see page 98 - connect an anesthetic gas
monitor to the COM 2 port.
Only use the cables from the list of accessories.
For further information on configuring the particular interface, see page 179.
2. WARNING
Danger to the patient and user
If the strain relief for the compressed gas hoses is not used, the device may be
damaged.
► Use the strain relief for the compressed gas hoses.
Insert the compressed gas hoses into the strain relief. Tighten the strain relief,
see page 23.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 59
Assembly and preparation
1. WARNING
Danger to the patient and user
If the strain relief for the compressed gas hoses is not used, the device may be
damaged.
► Use the strain relief for the compressed gas hoses.
WARNING
Risk of fire
Ignition sources together with oxygen can cause fires.
► Do not position oxygen sources in the vicinity of ignition sources, e.g.,
electrical connectors.
► Always fit the oxygen cylinder in the right-hand fixing position on the rear.
Insert the compressed gas hoses into the strain relief (1). Tighten the strain
relief.
36490
2. Check that the pressure measuring lines above the gas inlets are correctly
connected (2).
3. Place the gas cylinders (3) in the gas cylinder holders. Secure with hook-and-
loop straps.
60 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
4. WARNING
Risk of personal injury and damage to the device
Pressure reducers have an internal release valve. If a fault occurs, gas may
escape into the ambient air.
► Do not block or cover the release valve.
NOTICE
Risk of damage to the device
► When connecting the pressure reducers, ensure that they do not protrude
beyond the device.
Tightly screw the pressure reducers (4) to the gas cylinder valves. The
connectors must fit each other directly. Do not use an adapter.
36492
2. Connect the pressure measurement lines to the connectors (2).
3. The gas cylinder holder can be fastened at 2 different heights (not shown in this
illustration). Adjust the position of the gas cylinder holder to the size of the gas
cylinder in use. Contact service personnel to do this.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 61
Assembly and preparation
36491
2. Make sure that both pin-index pins (2) are present below the gas inlet (3).
3. Align the gas cylinder (4) so that the pin-index holes on the head of the
cylinder (1) are pointing towards the pin-index pins (2).
4. Insert the head of the cylinder (1) into the cylinder holder (5) from below.
5. Allow the pin-index pins (2) to engage in the pin-index holes.
6. Turn the handle (6) clockwise until the threaded stud is slightly screwed into the
visible recess on the head of the cylinder.
Align the gas cylinder (4) so that it is hanging vertically.
7. Tighten the handle (6).
8. Secure the gas cylinders (4) with hook-and-loop straps (not shown here).
40163
62 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
► WARNING
Risk of explosion
When pressurized, O2 is self-igniting in combination with oil or grease.
► Do not oil or grease the gas cylinder valve or the pressure reducer on the O2
cylinder. Do not touch with oily or greasy fingers.
The gas cylinder valves must only be opened and closed slowly. Do not use any
tools with the screw connection variants.
Have service personnel replace any leaky or stiff gas cylinder valves.
WARNING
Risk due to improperly mounted vaporizers
Incorrectly mounted vaporizers can cause leakage. This can cause the fresh-gas
delivery to be too low or contaminate the ambient air. Patient and user can be
endangered.
► Make sure that the connected vaporizers are hanging vertically.
► When using D-Vapor vaporizers, make sure that the power cable is not pinched.
► After mounting the vaporizers, perform a leakage test.
1. Set all the vaporizers upright and securely on the plug-in adapter.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 63
Assembly and preparation
2. Turn the locking levers (1) clockwise. The vaporizers are locked when the levers
are pointing to the left.
The illustration shows vaporizers of the type Dräger-Vapor 3000.
36493
3. WARNING
Risk due to incorrect anesthetic agent delivery
If the vaporizer is filled with the wrong anesthetic agent or if it is not filled
sufficiently, incorrect anesthetic gas concentrations or concentrations that are
too low can occur as a result.
► Compare the color code and labeling on the vaporizer used with the
anesthetic agent bottle.
► Follow the instructions for use for the vaporizer.
Check the filling levels in the sight glasses (2). Fill the vaporizers if required.
4. Turn the control dial on each vaporizer to the 0 position. The key (3) engages.
64 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
1. CAUTION
Risk of ambient air contamination
If the anesthetic gas receiving system is not connected to the disposal system,
contamination of the ambient air with anesthetic gas may result.
► Connect the anesthetic gas receiving system correctly to the disposal
system.
Connect the scavenging hose (1) to the nozzle on the AGS.
36494
2. Secure the scavenging hose with the clip (2).
3. Connect the gas probe of the scavenging hose to the terminal unit (3) of the
disposal system.
36495
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 65
Assembly and preparation
40960
4
5
If the flow indicator (4) is floating in the normal range (5), the anesthetic gas
scavenging system is functional.
If the flow indicator is floating in the restricted range (6), certain fresh-gas flows
should not be exceeded, see "Anesthetic gas receiving system" in chapter
"Technical data".
1. WARNING
Risk of negative pressure
If a passive AGS is connected to an active anesthetic gas scavenging system, a
negative pressure may arise in the patient’s lungs.
► Connecting a passive AGS to an active anesthetic gas scavenging system is
not allowed.
WARNING
Risk of overpressure
If the overpressure valve in the passive AGS or the scavenging hose is blocked,
overpressure in the breathing circuit and in the patient’s lungs will occur.
► Only connect the passive AGS to kink-proof and pressure-tight scavenging
hoses.
WARNING
Risk of ambient air contamination
The ambient air will be contaminated if the passive AGS feeds excess
anesthetic gas to a ventilation system with circulating air.
► Use the passive AGS only with ventilation systems that work without
circulating air.
66 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
36494
2. Secure the scavenging hose with the clip (2).
3. Lay the scavenging hose so that the gas is disposed of, e.g., by a ventilation
system.
4. Fasten the end of the hose.
The device can be used with Infinity ID breathing hoses or conventional breathing
hoses. If no leakage test has yet been performed after switching on the device,
hose compliance and hose resistance will automatically be adopted when Infinity ID
breathing hoses are connected.
Do not use any inspiratory or expiratory bacteria filters if the ID functionality of
the Infinity ID breathing circuit is to be used. In this case, fit a filter to the Y-piece.
In cases which preclude use of a bacteria filter at the Y-piece, the Infinity ID function
of the Infinity ID breathing circuit cannot be used.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 67
Assembly and preparation
2. WARNING
Risk due to particles and dust
In order to protect the patient from particles and dust, a filter must be used
between the inspiratory limb of the breathing system and the patient.
► Use a filter at the Y-piece or at the inspiratory port.
WARNING
Risk of infection
The breathing system may be contaminated with infectious agents. The
following causes may be present:
– No bacteria filters have been used at the Y-piece or at the expiratory port.
– The breathing system is being used for the first time.
Perform the following measures:
► Reprocess the breathing system before the first use.
► Reprocess the breathing system if necessary.
► To prevent future contamination, use bacteria filters close to the patient.
CAUTION
Risk due to too low tidal volumes
The hose configuration can influence the values for leakage, compliance,
resistance, and the therapy. Consequently, the tidal volume, for example, may
be too high or too low.
► Perform a leakage test after replacing breathing hoses, particularly
extendable hoses.
► Do not use extendable hoses to ventilate neonates.
68 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
Assemble the breathing circuit and connect it to the Y-piece and nozzles on the
breathing system. Observe the following information: "Permissible hose
configurations", page 69.
35636
When attaching or removing the breathing hoses, always hold them by the
connection sleeve and not by the hose itself.
Breathing hoses, sample line, and filters, etc., must be arranged carefully and
adapted to the patient, particularly for neonates and pediatric patients. For further
information see: "Improving the CO2 measurement by means of an HME filter",
page 308.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 69
Assembly and preparation
The permissible hose configurations are marked with an "X" in the following table:
70 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
37137
2. WARNING
Risk due to pinched breathing bag
If the breathing bag is pinched, excessive airway pressures or a lack of fresh
gas may result.
► When attaching, ensure the following:
– The breathing bag is not pinched.
– The breathing bag can inflate freely.
Lay the breathing bag hose in the holder (1) in a loop so that the bag hangs on
the left of the holder.
Make sure that the breathing bag is not impeded by breathing hoses or cables
when inflating.
4.8.2.2 Attaching the breathing bag to the breathing bag arm (option)
► Attach the breathing bag to the elbow.
37138
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 71
Assembly and preparation
RExpiration =
RBreathing system_exp + RExp hose + RExp filter (port) + RExp filter (Y-piece)
If necessary, take into consideration other parts such as water traps or additional
hoses. The specifications for the resistance of the breathing system can be found
on page 290. The specifications for all other accessories can be found in the
respective instructions for use.
In these instructions for use, the specifications for the resistance in the inspiratory
limb are regarded as negative values. The resistance values given in the
instructions for use for the accessories must therefore be subtracted from the
inspiratory resistance of the breathing system.
72 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
Since RInspiration is greater than –6 hPa (cmH2O) and RExpiration is less than 6 hPa
(cmH2O), this configuration may be used.
Depending on the breathing circuit in use, the connected accessories, and the
resistance of the patient's airways, air trapping (incomplete expiration) may occur
with some ventilation settings. Air trapping can be recognized on the flow waveform
by the fact that the inspiration begins before the expiration has ended.
The effects are, for example, a reduced minute volume in pressure-controlled
ventilation or higher mean airway pressures and peak pressures in volume-
controlled ventilation.
Air trapping can be prevented on the anesthesia machine by the following
measures:
– Adjusting the respiratory rate and inspiratory time
– Changing the configuration of breathing hoses and accessories that carry gases
during the expiratory time
It is the responsibility of the user of the medical device to select the most suitable
remedial measure.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 73
Assembly and preparation
1. WARNING
Risk of excessively high airway pressure
Without a pressure release valve or breathing bag, airway pressure may
become too high.
► Only non-rebreathing systems with breathing bags or pressure release
valves which comply with ISO 8835-2 may be connected.
Select a suitable non-rebreathing system.
2. WARNING
Risk of faulty gas delivery
Faulty gas delivery and insufficient gas supply when a non-rebreathing system is
used may put the patient at risk. Thus O2, CO2, and any anesthetic gases must
also be monitored for non-rebreathing systems.
► The sample line must be connected to the mask and the water trap on the
anesthesia machine.
► Provide for suitable gas monitoring conforming to ISO 80601-2-55.
► Provide suitable O2 monitoring for devices with inspiratory O2 measurement.
Screw the sample line securely to the mask (1) on the non-rebreathing system
and to the water trap (2).
37868
For masks without a connector for the sample line, optionally proceed as
follows:
► Place the T-piece with a T-piece filter directly on the elbow. Screw the sample
line firmly onto the T-piece filter. The part numbers for the T-piece and the T-
piece filter are listed in the list of accessories.
► If necessary, connect the sample line to a filter on the Y-piece.
Ensure the correct course of the sample line. Do not use adapters.
3. Connect the fresh-gas hose of the non-rebreathing system to the external fresh-
gas outlet (3).
4. Remove the sealing plug from the inlet nozzle (4) on the AGS.
5. Use the transfer hose to connect the non-rebreathing system to the inlet nozzle
on the AGS (4).
74 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
2. CAUTION
Risk due to leakage from an open AGS inlet nozzle
► To prevent contamination of the ambient air with anesthetic gases, press the
sealing plug back into the inlet nozzle after using a non-rebreathing system.
Seal the inlet nozzle (4) on the AGS again with a sealing plug.
37107
3. CAUTION
Risk of chemical burns
Soda lime is caustic and is a strong irritant for eyes, skin, and airway.
► Handle the soda lime carefully and do not spill it.
Empty out the used soda lime and dispose of it according to its instructions for
use.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 75
Assembly and preparation
4. Remove the absorber insert (3) from the absorber container. The sealing ring (4)
remains on the absorber insert.
39487
3
41120
5
2. CAUTION
Risk of chemical burns
Soda lime is caustic and is a strong irritant for eyes, skin, and airway.
► Handle the soda lime carefully and do not spill it.
Fill the CO2 absorber with fresh soda lime to the upper mark.
Recommendation: Use Drägersorb 800 Plus or Drägersorb Free.
3. WARNING
Risk of hypoventilation and incorrect gas measurement
Reuse of the disposable dust filter can increase the filter resistance and impair
the ventilation function of the device.
► If soda lime from third-party manufacturers is used, e.g., granular soda lime,
use a disposable dust filter and replace it with every change of the soda lime.
If soda lime from third-party manufacturers is used, insert a new disposable dust
filter (7). Only use dust filters from the list of accessories.
76 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
41121
4. Attach the CO2 absorber (8) to the breathing system from below. Rotate it in the
direction of the arrow until it reaches the stop.
Follow the instructions for use for the particular soda lime.
► WARNING
Risk due to occluded components in the breathing circuit
In the following cases, the sample gas flow can immediately cause negative
pressure in the lungs:
– Filters, hoses, or endotracheal tubes are blocked.
– The sample gas is being extracted between the patient and an occluded
component.
Ensure the following when ventilating pediatric patients and neonates:
► When fine pored filters are used, do not connect the sample line between the
tube and the filter.
► If the sample line is connected close to the patient, set the alarm limits for
MV low and Paw high to suitable values. Observe the following information:
"Improving the CO2 measurement by means of an HME filter", page 308.
WARNING
Risk due to incorrectly connected sample line
If the sample line is connected to the wrong connectors, e.g., connectors on
infusion pumps, fluids may be drawn in instead of sample gas. Consequently,
the gas measurement may not display correct values.
► When connecting the sample line, take care that it is correctly connected.
WARNING
Risk of incorrect measured gas values
Blocked water traps or blocked sample lines prevent correct gas measurement.
As a result, incorrect measured gas values could be displayed.
► Only use Dräger sample lines.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 77
Assembly and preparation
Screw the sample line on to the water trap (1) and to the Y-piece, HME filter or
hose adapter (2).
Ensure the following:
– Ensure the correct course of the sample line.
– Do not use adapters.
– Particularly when ventilating pediatric patients and neonates, a sample line
connection close to the patient can improve the quality of the measured CO2
values. Observe the following information: "Improving the CO2 measurement
by means of an HME filter", page 308.
– When tube adapters are used for the sample line connection, make sure the
volume (dead space) is small.
37741
78 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operating concept
5 Operating concept
5.1 Screen
5.1.1 Main screen
The main screen displays the most important information regarding anesthesia and
ventilation.
34746
1
1:35
No. Designation
1 Header bar The header bar shows the following information:
– Patient category
– Patient data
– System information (date, time, device name)
– Alarms, messages, and notifications
– Information regarding temporarily deactivated alarms
2 Main menu bar The main menu bar contains buttons to open dialogs
and activate functions.
These buttons are assigned to various groups.
For further information, see the following chapter:
"Overview of the menu structure", page 323
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 79
Operating concept
No. Designation
3 Monitoring area The following information is displayed in the monitoring
area:
– Gas measurement
– Waveforms
– Parameter fields
– Loops (Pressure-Volume and Flow-Volume)
– Mini-trends
– Virtual flow tubes
For further information see: "Customizing the current
view", page 109.
4 Therapy bar The ventilation settings can be adjusted in the therapy
bar.
Electronically controlled gas mixer:
– Ventilation modes
– Ventilation parameters
– Fresh-gas delivery
Mechanically controlled gas mixer:
– Ventilation modes
– Ventilation parameters
5.1.2 Dialogs
Dialogs consist of one or more pages which are displayed by touching the
corresponding horizontal or vertical tab.
34766
1 2
No. Designation
1 Title of the dialog
2 Button for closing the dialog
3 Vertical tab to open subordinate structures
4 Horizontal tab to open a page
80 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operating concept
34752
1 2 3 4
5
4
7 6
No. Designation
1 Name of active ventilation mode
2 Tabs
3 Notification field
4 Buttons to expand and collapse the therapy bar
5 Field with additional information:
– More values
– Spontaneous breathing activity of the patient
6 Therapy controls for ventilation parameters and buttons for synchronizing
the breaths
7 With electronically controlled gas mixer:
Therapy controls and buttons for fresh-gas delivery
Start values
Arrows on the scales of the therapy controls mark the start values resulting from
the patient data and start settings. The start values can be configured, see
page 166.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 81
Operating concept
Setting ranges
Some settable parameters may be limited or be mutually restricted so that certain
combinations of therapy settings are not possible, e.g., Ti 6.9 s at RR 100 /min.
If a state is reached in which a parameter cannot be changed any more, the device
displays a corresponding message in the notification field (3).
34767
Quick setup
This dialog can be opened by touching the corresponding parameter field in the
monitoring area. This dialog can be configured so that it opens automatically in the
event of an alarm, see page 170.
82 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operating concept
Color Meaning
Blue Therapy in progress.
Yellow A function or setting must be confirmed.
Flashing yellow A function or setting, which is still not confirmed, will
be reset within the next 5 seconds.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 83
Operating concept
34777
1
2. Set
Turn the rotary knob. For some therapy controls, faster turning raises the
increment value.
3. Confirm
Press the rotary knob. The color of the control element changes to green.
In the subsequent chapters of this document, these steps will be written in
simplified form as follows:
– "Set the value."
– "Touch the button."
84 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operating concept
► Select another parameter. This selection resets the change made to the
previous parameter.
► Do not press the rotary knob. After 15 seconds, the change is reset and signal
tones sound during the last 5 seconds (timeout).
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 85
Getting started
6 Getting started
6.1 Safety instructions
Checking the safety systems
Some safety systems are only checked during start-up. If this check is not
performed regularly, a device malfunction may occur, putting the patient at risk.
► Restart the device at least once per month so that the safety systems will be
checked regularly.
35026
1
1. Connect the device to the mains power supply.
2. Set the main switch to position I.
3. CAUTION
Risk of device malfunction
Condensed water may form when the device is brought from a cold storage
location into a warm environment.
► To prevent condensation and resulting failures of electrical components, do
not turn on the device after abrupt temperature changes for 1 to 2 hours.
Press the key (1).
The device starts. The Standby page is displayed.
If there is sufficient battery charge, the device will also start without the power plug
being plugged in.
86 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Getting started
35034
1
2
5
Color Meaning
Green System is fully operational.
Yellow System is operational with limitations.
There are functional restrictions.
Take further measures to ensure patient safety (e.g., external
monitoring).
Red System is not operational.
Contact service personnel if necessary.
If the device is not fully operational, the most important irregularities (5) are
displayed along with a recommendation to perform a specific test (3).
Additionally, the current system leakage is displayed in area (5).
To view details regarding the status of the device, touch the Details... button (2) or
the Tests... button (4), see page 126.
Dräger recommends performing the system test every 24 hours. Otherwise, it will
not be possible to ensure that the device is functional.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 87
Getting started
35041
4
1
3
1. Adjust the APL valve.
2. Set the O2 switch (3) to the Add. O2 position.
3. Open the flow control valve (2) and set the desired O2 flow. If necessary, press
the O2+ key (4) to quickly fill the breathing bag.
4. Monitor the set flow on the O2 flowmeter (1).
5. Set the anesthetic gas concentration on the vaporizer.
6. Ventilate the patient manually.
7. Turn on the device.
8. As soon as the Standby page is displayed, start the therapy, see page 93.
9. Set the O2 switch (3) to Aux. O2.
10.Close the flow control valve (2).
35054
3
1
1. Adjust the APL valve.
2. Open the flow control valve (1) and set the desired O2 flow. If necessary, press
the O2+ key (3) to quickly fill the breathing bag.
3. Monitor the flow on the total flow tube (2).
4. Set the anesthetic gas concentration on the vaporizer.
5. Ventilate the patient manually.
6. Turn on the device.
7. As soon as the Standby page is displayed, start the therapy, see page 93.
88 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
7 Operation
7.1 Safety instructions
7.1.1 Alarms
Alarm volume
If the alarm volume is too low, alarm signals may not be heard. The patient may be
put at risk.
► Set the alarm volume loud enough so that the alarm signals can be heard in the
environment where the device is located.
► The user must remain within earshot of the alarm signals.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 89
Operation
90 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
Measured gas values and waveforms such as the CO2 waveform are determined
on the basis of the composition of the sample gas. The composition of the sample
gas is affected by many factors and their interactions, especially in patients with low
body weight. This may result in biased measured values or waveforms and thus to
misinterpretations. As a result, the patient could be put at risk.
The following factors affect the sample gas measurement:
– Dead space
– Airway resistance of the patient
– Compliance of the patient
– Type of surgical procedure
– Gas sampling site
– Breathing circuit, filter, sample line, tube
– Ventilation settings and the resulting ventilation
– Leakage
– Spontaneous breathing
– Cardiogenic oscillations
– I:E ratio and the respiratory rate
► Adhere to the following:
– Do not make therapeutic decisions based solely on individual measured
values or parameters.
– If possible, minimize the effects of the factors described above, e.g., take the
sample gas from a gas sampling site close to the patient, minimize leakage,
adjust the ventilation settings.
– The measured values are not meaningful during the warm-up time of the
patient-gas measurement module.
Data (e.g., measured values, alarms) which the anesthesia machine transfers to
other systems such as patient monitors or EMR systems may be displayed there
incompletely or incorrectly and may thus put the patient at risk. Consequently, the
data are intended to be used only for information purposes.
► Do not use data displayed on other devices for patient monitoring or device
monitoring.
► Do not use data displayed on other devices for diagnostic or therapeutic
decisions.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 91
Operation
92 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
35280
3
1
Quick start
Electronically controlled gas mixer:
The fresh-gas settings (3) are displayed.
► Touch the Quick start Man/Spon button (1).
Mechanically controlled gas mixer:
► Open the flow control valves. The device switches automatically to the
Manual / Spontaneous mode. Take care that the quick start does not happen
unintentionally.
Or:
► Touch the Quick start Man/Spon button (1) and open the flow control valves.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 93
Operation
35297
1 2
Start
Select patient and check data
Select mode
94 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
35298
1
► WARNING
Risk due to incorrect settings
Different standard alarm limits or therapy settings might be configured for
medical devices within the same area. The user must observe the following:
► Make sure that the set values and alarm limits are selected to suit the patient.
► Make sure that the alarm system is neither rendered useless by setting
extreme values for the alarm limits nor deactivated by switching off the
alarms.
► Check the therapy settings, the start settings for alarms, and the alarm
settings during a change of ventilation mode.
► Only turn off alarms if the safety of the patient will not be compromised as a
result.
CAUTION
Risk due to incorrect setting for patient age
Incorrectly setting the patient age can lead to incorrect xMAC values and thus to
an incorrect anesthetic agent delivery.
► Always set the patient age correctly.
Adjust the patient data (1).
The device will suggest appropriate therapy settings for these data, e.g., for tidal
volume, respiratory rate, and alarm limits. For more information, see page 311.
35300
Select mode
5 4 3 1 2
1. Select the ventilation mode (1).
The following ventilation modes are available:
– Man / Spon
– VC
– PC
– VC - AF
– PSV
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 95
Operation
2. WARNING
Risk of patient injury
The use of minimal-flow settings or low-flow settings can lead to the following
problems:
– Accumulation of metabolic by-products in the breathing system
– Condensed water in hoses
– Condensed water in the piston diaphragm
► Follow the recommendations of professional societies (e.g., regular flushing
of the breathing system).
► Check for condensed water in the hoses and remove if necessary.
► Use water traps in the breathing hoses.
Set the fresh-gas delivery.
The device is equipped with a minimum O2 delivery function which ensures that
a minimum quantity of oxygen is delivered. For further information see:
"Minimum O2 delivery", page 310.
Electronically controlled gas mixer:
► Select the carrier gas (3).
Set the O2 concentration (5) and fresh-gas flow (4).
Mechanically controlled gas mixer:
► Open and set the flow control valves for the required gases.
Also use the total flow tube to check the total flow set, see page 29.
3. Adjust the ventilation settings (2).
4. Press the rotary knob. The therapy starts and a signal tone sounds.
WARNING
Risk of excessively high airway pressures
If the ventilator fails, the device switches into the Man / Spon ventilation mode.
► Also set the APL valve to a value suitable for the patient when using mechanical
ventilation modes. If the ventilator fails, ventilate the patient manually.
96 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
35256
10 20 30
35264
In the Manual / Spontaneous mode, lifting the valve relieves pressure from the
breathing system.
35265
The dots are aligned vertically over one another. The valve lifts.
The pressure limitation is canceled and the valve is open for free spontaneous
breathing.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 97
Operation
The gas concentration can change abruptly when the O2 flush is used.
39628
4 2 3
2. CAUTION
Risk of inaccurate measured flow values
If the anesthetic gas compensation is incorrectly set, the resulting incorrect
measured flow values can lead to incorrect measured values for the tidal
volume.
► Make the settings for the anesthetic gas compensation carefully.
Choose one of the following selection options (3):
– Off: The measured flow values will not be corrected.
– Des: The measured flow values will be corrected for desflurane using an
average value. For more information, see table page 175.
– Sev: The measured flow values will be corrected for sevoflurane using an
average value. For more information, see table page 175.
98 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
35268
3 2 1 3 2
2. Adjust the therapy controls (3) or the buttons (2).
3. Activate the ventilation mode with the rotary knob. A signal tone is emitted when
the mode is changed.
35279
2 3
2. If required, show the extended therapy bar with one of the More buttons (3) or
(4). Then adjust the additional parameters (Trigger, ∆Psupp, etc.).
WARNING
Risk of insufficient ventilation
In ventilation modes in which breaths are to be triggered only by the patient (e.g.
PSV), adverse settings or sensor failure can lead to insufficient ventilation.
► To maintain a minimal ventilation of the patient, set the respiratory rate to a
suitable value.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 99
Operation
WARNING
Risk of insufficient ventilation
The spontaneous minute volume MVspon indicates the volume which results from
spontaneous breathing and from pressure-supported spontaneous breathing. If the
patient frequently triggers pressure-supported breaths as a result of small tidal
volumes, a large part of MVspon will be achieved by mechanical ventilation and not
by spontaneous breathing of the patient. In this case, MVspon shows a high value
although the actual spontaneous minute volume is very low.
► Do not base therapy decisions solely on the value displayed for MVspon.
100 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
1. CAUTION
Risk of faulty gas delivery
With non-rebreathing systems, O2, CO2, and any anesthetic gases must be
monitored.
► Connect the sample line to the non-rebreathing system and the integrated
patient-gas measurement module or the anesthetic gas monitor.
CAUTION
Risk of gas contamination
When the device’s integrated patient-gas measurement module is used, the
extracted sample gas is also returned to the internal breathing system during
operation with an external fresh-gas outlet.
► Using a breathing hose, establish a closed connection between the
inspiratory port and the expiratory port. Flush the breathing system each time
patients or anesthetic gas are changed!
Start the Ext. FGO operation mode.
2. Adjust the fresh-gas delivery. Set the vaporizer if required.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 101
Operation
Activating
1. Open the Alarms dialog.
2. Touch the Settings tab (1).
35307
Alarms 1
2 3
Deactivating
Deactivate the CBM mode optionally as follows:
► For Cardiac bypass mode (CBM), touch the Off button (2):
► In the main menu bar, touch the Exit CBM button.
102 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
in et
35310
O2
N2O
Iso
The age-adjusted MAC values are calculated according to the equation of W.W.
Mapleson (British Journal of Anaesthesia 1996, pp. 179-185).
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 103
Operation
For gas mixtures, the respective multiples for N2O and anesthetic agents are
summed up according to the following equation:
7.5.1.1 Example
Exp. Isoflurane = 0.65 Vol%
Exp. N2O = 69 %
Age = 32 years
MACage-adjusted for Iso: MAC2) = 1.21 Vol%
MACage-adjusted for N2O: MAC2) = 110 Vol%
xMAC = 0.54 + 0.63 = 1.2
The influence of other drugs (opioids or intravenous hypnotics) is not considered in
the xMAC calculation.
7.5.2 Econometer
Prerequisite: The device has the "Advanced gas monitoring" option.
During operation, the device monitors the breathing bag for sufficient filling.
35322
Surplus
Efficient
Deficit
Econometer
The bar graph indicates whether the device is supplied with sufficient fresh gas.
1) 40 years
2) 32 years
104 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
An insufficiently filled breathing bag can trigger the Fresh gas low or leakage or
Emergency air inlet activated alarms, for example.
7.5.3 Stopwatch
35548
Stopwatch
1
7.5.4 Timer
35552
Timer
2 1
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 105
Operation
7.5.5 Volumeter
The volumeter is used for observing and assessing the ventilation.
35553
1
Volumeter
2
VT
6 3
Volume
4
The bar graph indicates the inspiratory and expiratory tidal volume.
At the end of the inspiration, the delivered tidal volume is displayed as a bar (1).
At the end of the expiration, the difference between inspiratory and expiratory tidal
volumes (6) is displayed.
The expiratory tidal volume is displayed next to the bar graph (3).
Required FG flow
Total flow
Efficient
106 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
An insufficient fresh-gas flow may trigger the Fresh gas low or leakage or
Emergency air inlet activated alarms, for example.
Standard view
35683
1
Up to three waveforms (1), (2), and (3) are displayed along with their associated
parameter fields.
Expert view
35685
4 5 6 7
In addition to the standard view, the 4 additional parameter fields (4), (5), (6), and
(7) are displayed.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 107
Operation
– View button
35688
1
2
3
108 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
35686
Quick setup
2. WARNING
Risk due to inadequate monitoring
National and medical regulations may require certain parameters to be
displayed.
► Always consider the relevant regulations when configuring the screen layout.
For Content (1), select the desired content.
For a list of the possible screen content, see page 184.
3. For Scale (2), select the desired setting.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 109
Operation
Displaying loops
1. Open the Views dialog.
2. Touch the Loops button.
35866
1
35868
3
21
110 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
35870
2 1
1. Open the Views dialog.
2. Touch the Mini-trends button.
For larger and more detailed graphical and numerical trends, see page 114.
35874
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 111
Operation
35875
System setup
1
3 3
Setting the sweep speed:
► Touch the button (2) and set the sweep speed.
Adjusting the waveform scale:
► To change the scale, touch one of the buttons (3) and select the value.
35883
2
System setup
1
1
112 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
35886
1
Trends/Data
7.7.2 Logbook
The logbook can save up to a maximum of 20000 entries. The entries in the
logbook cannot be deleted and are retained even after the device has been turned
off and on again or following a power supply failure. When the storage limit is
reached, the oldest entries are overwritten. Logbook data are displayed in table
form.
1. Open the Trends/Data dialog.
2. Touch the Logbook tab (1).
35887
1 2
Trends/Data
Use the rotary knob or the arrow buttons (3) to scroll the cursor (2) up or down in
the logbook. To scroll quickly, touch the gray area (4).
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 113
Operation
7.7.3 Trends
Prerequisite: The device has the "Advanced trends" option.
Trends are displayed in the form of a graphic or a table.
1. Open the Trends/Data dialog.
2. Touch the Graphical trends tab (1) or the Tabular trends tab (2).
The following illustration shows the graphical trend:
35917
1 2
Trends/Data
7 6
Zooming
In both trend displays, the displayed time period can be enlarged or diminished.
Changing the time period:
► Touch the Zoom + or the Zoom - button (4):
Displaying the standard time period and the current point in time:
► Touch the Reset zoom button (5).
114 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
At Standby > System setup > System status > Consumption the gas
consumption levels can be reset to zero, see page 180.
2. CAUTION
Risk of an operating error
During operation in louder environments, the acoustic alarm signals might not be
heard.
► Always set the alarm tone to be sufficiently loud.
Touch the System > Sound volume tab (1).
35918
1
System setup
1
2
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 115
Operation
35922
Quick setup
Manual setting
In the Alarms dialog, the alarm limits can be set either manually or automatically.
1. Open the Alarms dialog.
2. Open the Limits tab (1).
35935
Alarms
1
5 4
3. Set the upper alarm limit (2).
4. Set the lower alarm limit (3).
116 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
Automatic setting
Alarm limits can be automatically adapted to current measured values or set values.
1. Open the Alarms > Limits dialog.
2. To adjust the alarm limits for one individual parameter, touch and confirm one of
the Autoset buttons (5).
To adjust the alarm limits for all parameters, touch and confirm the Autoset all
button (4).
The function for adjusting all alarm limits can be called up directly with the Autoset
limits button in the main menu bar.
Only use the automatic adjustment when measured values and set values are
stable.
The lower alarm limit for the xMAC level is also adjusted during automatic setting,
see page 118.
Configuration and algorithm, see page 169.
35936
1
Alarms
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 117
Operation
41088
[0$& LQVS[0$&
H[S[0$&
ORZHUDODUPOLPLW
If the xMAC value rises, the xMAC monitoring (1) is activated and the lower alarm
limit for the xMAC level is automatically adapted to the anesthetic gas
concentration. The lower alarm limit (2) can reach a maximum value of 1.0.
The lower alarm limit can be recalculated by touching the Autoset button (3). This
allows the alarm limit for the xMAC low alarm to be adjusted in special anesthesia
situations and to exceed the value of 1.0 if necessary.
If the expiratory xMAC value falls below the alarm limit (4 or 7), the device issues
the xMAC low alarm with low priority. If the alarm is not acknowledged with the
ALARM RESET button, the priority is raised to medium priority after 60 seconds.
118 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
35978
Patient
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 119
Operation
35980
1
Trends/Data
2
3. Touch the respective button (2).
The data will be saved as a ".txt" file.
120 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
35982
2
System setup
1
CAUTION
Risk due to faulty or switched-off breathing system warmer
Without the breathing system warmer, increased condensation may occur in the
breathing system.
► Increase the fresh-gas flow if necessary. Remove the condensed water from
hoses, water traps, and the breathing system regularly. Have service personnel
repair the faulty breathing system warmer.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 121
Operation
35981
1
System setup
7.13.2 O2 insufflation
7.13.2.1 Overview
On the electronically controlled gas mixer, O2 insufflation is performed using the O2
flowmeter. On the mechanically controlled gas mixer, O2 insufflation is performed
using an integrated O2 flowmeter or an external O2 flowmeter.
122 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
35294
2
1 3
35295
2
37140
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 123
Operation
1. WARNING
Risk of fire
In combination with oxygen or nitrous oxide, ignition sources such as
electrosurgical devices and laser surgical devices can cause fires.
► Prevent leakage, e.g., at endotracheal tubes, laryngeal mask airways, face
masks, Y-piece, breathing system including hoses, filters, and breathing bag,
at the external fresh-gas outlet, and at the outlet for O2 insufflation.
► Use only intact and leak-free hoses at the outlet for O2 insufflation.
► Before beginning laser surgery or electrosurgery, flush with sufficient air
(<25 % O2), and flush beneath the surgical drapes as well.
► If ignition sources are present, close the flow control valve on the
O2 flowmeter to the end position.
► When O2 outlets are in use (e.g., for insufflation), do not use any ignition
sources in the immediate vicinity.
► Do not position oxygen sources in the vicinity of ignition sources, e.g.,
electrical connectors.
Connect the appropriate accessories to the outlet for O2 insufflation (1).
2. WARNING
Risk due to overpressure
When the patient is connected to the outlet for O2 insufflation without a release
valve, increased pressure may be applied to the patient.
► Only connect the patient in a way that allows excess gas to escape (e.g.,
through a release valve).
Connect the patient using a mask or nasal cannula.
3. Open the flow control valve (2) to begin O2 insufflation.
Ending O2 insufflation:
► Close the flow control valve (2).
124 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
1. WARNING
Risk of infection
The water trap may contain infectious fluid.
► Proceed carefully when emptying and take protective measures if necessary.
► Follow the infection prevention policies and reprocessing regulations of the
healthcare facility.
Check the water trap of the patient-gas measurement module (PGM) for
leakage. If necessary, empty or replace the water trap.
2. WARNING
Risk of infection
Used sample lines and water traps may be infectious due to the breathing gases
that passed through them.
► Replace the sample line and the water trap regularly in the following
situations:
– If the sample line is connected to the filter on the Y-piece, replace it daily.
– If there is no filter fitted to the Y-piece and the sample line is connected
directly to the Y-piece, replace the sample line after every patient.
► Remove the sample line from the water trap.
► Initially leave the water trap fitted to prevent infectious fluid from spurting out.
Remove the water trap only after surface disinfection.
► Replace the water trap after each patient.
If no filter was used, replace the sample line and dispose of the used sample
line.
CO2 absorber
► Check the soda lime for discoloration and replace if necessary, see page 75.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 125
Tests
8 Tests
8.1 Status of the device functions
► In the standby mode, touch the Details... button or the Tests... button.
35069
1
The Test results list (1) shows the results of the test last performed. The following
information is displayed using different colors:
– Influence of the individual device functions on the functional integrity of the
device
– Leakage values
Color Meaning
Green Successfully tested, fully available
Yellow A non-critical fault has been detected. The device can be operated
with restricted function.
Red A serious fault has been detected. Operation is not possible or is for-
bidden.
Gray Not tested
126 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Tests
Leakage test
Type and duration: Automatic, approximately 2 min
Perform the test: – After filling the CO2 absorber
– After changing the hose configuration (e.g.,
changed hoses, changed lengths of extendable
hoses etc.)
– After replacing the breathing system
– After inserting the piston diaphragm
– After replacing the flow sensors
– If leakage at vaporizer is suspected (see "Check-
ing a vaporizer for leakage", page 130)
Description: – Determining leakage, system compliance, and
system resistance
– Calibration of valves and flow sensors, if required.
In this case, the test is extended by
approximately 3 minutes.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 127
Tests
1. WARNING
Risk of patient injury
During the system test, the device is pressurized.
► To prevent patient injury, do not perform the system test and leakage test if a
patient is connected.
Touch the button for the system test.
2. Complete the tabular checklist (1).
It is possible to switch from the tabular checklist to the walk-through mode, if
required. To do this, touch the button (2). For further information see: "Checklist
in the walk-through mode", page 136.
35072
1
System test
4 5
4. CAUTION
Risk of device malfunction and/or patient injury
Existing malfunctions cannot be detected if the system test is canceled. This
may result in device malfunctions and the patient may be put at risk.
► More attention is required when operating without a system test.
► Perform the system test every day. If the system test is canceled during
execution, perform it again as soon as possible.
128 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Tests
If all components are operational, touch the button (4). The automatic test
will start.
If a component is not operational, touch the button (5).
The walk-through mode will start.
35083
6
System test
7
The components (6) are polled one after the other.
The buttons (7) are used to document whether the check passed.
Button Meaning
Check passed
Check failed
The automatic test starts after all the checks in the walk-through mode are
complete.
9 35085
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 129
Tests
The test progress is displayed in area (8) while the automatic test is running. All test
results are displayed in area (9).
After the test, the final test result is displayed on the standby screen, see page 87.
35088
Test details
1
2
130 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Tests
– Gas mixer:
– With electronically controlled gas mixer:
The flow control valve of the O2 flowmeter is closed.
– With mechanically controlled gas mixer:
All flow control valves are closed.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 131
Tests
6. Press the O2+ key and keep it pressed for at least 5 seconds. The breathing
system will be flushed and the residual anesthetic agent will be disposed of in
the AGS.
Using the determined values, the user can decide whether the vaporizer can be
used.
39554
2 4
1 3
5
132 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Tests
39555
7
2 4
8
1
3
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 133
Tests
For leaking hoses, the complete leakage test can be used to determine how large
the leakage value is and whether the hose can continue to be used.
39554
2 4
1 3
5
134 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Tests
39555
7
2 4
8
1
3
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 135
Tests
Prerequisites
1. Connect the hoses (1).
2. Seal the Y-piece (2).
35102
2
35100
20 30 40
136 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Tests
35104
8.4.1 Vaporizers
For each vaporizer, check:
1. The locking lever points left, indicating the vaporizer is locked.
35105
2. The filling inlet is closed.
35108
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 137
Tests
3. The control dial is set to position 0 and the key is locked in placed.
35107
4. Check the filling level in the sight glass. Refill anesthetic agent if required.
35106
138 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Tests
35109
When using pressure reducers without electronic pressure measurement, read
the pressure from the pressure gauge.
2. Close the gas cylinder valves.
On devices that are equipped with Advanced Cylinder Support, the gas cylinder
valves can remain open during operation. These devices are identified by an
appropriate label near the gas inlets (see "Gas inlets", page 26).
35110
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 139
Tests
2. Open the flow control valve (1). Set the desired O2 flow.
35112
The O2 flowmeter indicates a flow, the breathing bag fills, and the inflow of gas is
audible.
3. Close the flow control valve (3).
35111
3
140 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Tests
35115
2
1 3
2. Extendable hoses (3) are extended to the length intended for use.
Do not change the length of the hoses after the test is done.
35117
4. Prerequisite:
– Patient-gas measurement module present
– Setting: Sample line is connected during test: Off
– Setting: Test breathing circuit: On
– Electronically controlled gas mixer or mechanically controlled gas mixer with
O2 detection
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 141
Tests
Press the O2+ key. The inflow of gas is audible. Keep the key pressed until the
pressure exceeds 15 hPa (cmH2O).
35118
5. Pull off the Y-piece and plug it on again. The pressure drops.
37961
8.4.3.1 Special test settings
1. Test breathing circuit
Prerequisite:
– Patient-gas measurement module present
– Sample line is connected during test: On
– Electronically controlled gas mixer or mechanically controlled gas mixer with
O2 detection
A check is made to ensure that the inspiratory hose, expiratory hose, and
breathing bag hose are connected correctly.
This can be deactivated in the System setup > System > System test dialog.
2. Verify O2 delivery
Prerequisite:
– Patient-gas measurement module present
– Sample line is connected during test: On
– Electronically controlled gas mixer or mechanically controlled gas mixer with
O2 detection
A check is made to ensure that the O2 supply is actually delivering O2. This is
checked at the Y-piece, using the connected sample line.
142 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Tests
8.4.4 Valves
8.4.4.1 Checking the O2 flush
1. Press the O2+ key and keep it pressed until the pressure stabilizes.
35118
The pressure does not exceed 45 hPa (cmH2O).
2. Release the O2+ key again and wait until the pressure stabilizes.
35118
The pressure does not fall below 15 hPa (cmH2O).
3. Lift the APL valve.
35120
20 30 40
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 143
Tests
37962
The non-return valves have been fitted.
8.4.5 Components
8.4.5.1 Loudspeakers
► Touch the button and wait for 2 different acoustic signals.
35123
If the acoustic signals are not emitted, contact service personnel.
144 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Tests
35121
2. Check the functional integrity of the manual resuscitator.
35125
1
2
If the flow indicator (1) is floating in range (3) ("restricted range"), certain fresh-
gas flows should not be exceeded, see "Anesthetic gas receiving system" in
chapter "Technical data". Contamination of the ambient air can be prevented by
limiting the fresh-gas flow.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 145
Tests
1. WARNING
Risk of overpressure
If the overpressure valve in the passive AGS or the scavenging hose is blocked,
overpressure in the breathing circuit and in the patient’s lungs will occur.
► Take care that the scavenging hose does not become blocked.
► Perform a visual inspection of the overpressure valve for damage and soiling.
Check that the scavenging hose is run correctly. The hose must not be blocked.
2. Perform the visual inspection of the overpressure valve (4).
42190
8.4.6 Accessories
8.4.6.1 Soda lime
► Make sure that the soda lime does not need to be exchanged. Change the soda
lime if it is discolored or when its maximum period of use has been reached, see
page 75.
35128
146 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Tests
35129
8.4.6.3 Water trap
1. Check the water level in the water trap.
35130
2. Check the period of use of the water trap. Replace the water trap when
necessary.
With Infinity ID function:
– Water traps of type Infinity ID WaterLock 2 will automatically be detected and
the replacement date will automatically be set.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 147
Tests
CAUTION
Risk of inaccurate measured values
During calibration, the sensor must be exposed to ambient air, i.e., 21 % oxygen
concentration. Fluctuating oxygen concentrations must be avoided. The following
notes must be adhered to:
► Do not fill vaporizers during the calibration.
► Close all the flow control valves.
► Do not blow into the sensor.
► Do not use any disinfectants.
39679
1
148 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Tests
39674
5. Follow the instructions on the screen.
The device will perform the calibration.
6. After the calibration is complete, remove the sealing cap and plug it back in the
holder.
Insert the O2 sensor (4) back into the sensor port and turn it clockwise until it
reaches the end position.
39680
4
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 149
Tests
39676
If the O2 calibration is possible despite the expired O2 sensor cell, the O2
measurement will continue to be available. With an expired O2 sensor cell, a yellow
test result (operational with limitations) can be achieved at best.
Replace the sensor cell when the O2 sensor can no longer be calibrated, see
page 55.
150 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Ending operation
9 Ending operation
9.1 At the end of the OR day
Dräger recommends shutting down the device during longer periods of non-use
such as overnight or on weekends. This can lower power consumption and prolong
the life span of the medical device without negatively influencing device availability.
1. Make sure that all flow control valves are closed.
2. Press the key.
3. Follow the instructions on the screen.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 151
Alarms
10 Alarms
10.1 Safety instructions
Alarm volume
If the alarm volume is too low, alarm signals may not be heard. The patient may be
put at risk.
► Set the alarm volume loud enough so that the alarm signals can be heard in the
environment where the device is located.
► The user must remain within earshot of the alarm signals.
152 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Alarms
35988
1 2
3
3
3
3
No. Designation
1 All alarms button
2 Alarm message field
3 Alarm-triggering parameters
In the event of an alarm, the device displays the relevant alarm message in the
alarm message field. For certain alarms, the parameter field of the parameter
triggering the alarm will flash.
In the alarm message field (2), up to 8 alarms can be displayed at a time. If more
alarms occur, the All alarms button (1) is displayed. Touching this button opens the
Alarms > Current alarms dialog with information about all active alarms, see
page 154.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 153
Alarms
36003
1
Alarms
4 3
2. In the list (2), touch the corresponding alarm or select it with the rotary knob.
3. Refer to the information under Cause (4) and Remedy (3) to remedy the error.
A list of all possible alarms can be found in chapter "Alarm – Cause – Remedy",
see page 201.
154 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Alarms
36004
1
► Press the alarm silence key (1) below the screen, see page 17.
The symbol and the remaining time for the silenced alarm tone are displayed in
the header bar.
If the cause of the alarm persists, the alarm tone starts again immediately after the
alarm silence ends.
During the alarm silence, only new alarms with a higher alarm priority or a higher
internal priority number compared with the silenced alarm are acoustically signaled.
For further information see: "Alarm – Cause – Remedy", page 201.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 155
Alarms
36008
1
Or
Option 2:
36011
2
Alarms
Option 1 Option 2
Touch the ALARM RESET button (1) in In the Alarms > Current alarms
the header bar and confirm. dialog (2), touch the Reset all button (3)
and confirm.
All the alarms displayed in the alarm All alarms will be downgraded or reset.
message field will be downgraded or
reset.
156 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Alarms
36010
1
Alarms
Use the rotary knob or the arrow buttons (2) to scroll the cursor up or down.
The alarm logbook is cleared when the device is shut down or a new case is
started.
36015
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 157
Alarms
36016
Quick setup
158 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Alarms
FiO2 low2)
Is adopted.
inAgent high2)
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 159
Alarms
160 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Alarms
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 161
Alarms
Ventilation alarms
162 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Alarms
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 163
Alarms
164 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
11 Configuration
11.1 Device settings
Some device functions are available as an option and consequently are only
available on appropriately equipped devices.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 165
Configuration
Color scheme Day light; Day dark; Night Sets the color scheme. Observe the
information in the following section:
"Daytime colors and nighttime
colors", page 84
Screen bright- 10 to 100 Sets the screen brightness.
ness 80
166 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
Sweep speed 6.25; 12.5; 6.25; 12.5; 6.25; 12.5; Specifies the sweep speed.
[mm/s] 25 25 25
VT scale Auto; Auto; Auto; Specifies the scale for the volumeter.
[mL] 0 to 10; 0 to 10; 0 to 10;
0 to 50; 0 to 50; 0 to 50;
0 to 150; 0 to 150; 0 to 150;
0 to 500; 0 to 500; 0 to 500;
0 to 1000; 0 to 1000; 0 to 1000;
0 to 2000 0 to 2000 0 to 2000
Flow scale Auto Auto Auto Specifies the scale for the flow wave-
[L/min] -5 to 5 -5 to 5 -5 to 5 form.
-10 to 10; -10 to 10; -10 to 10;
-30 to 30; -30 to 30; -30 to 30;
-60 to 60; -60 to 60; -60 to 60;
-120 to 120 -120 to 120 -120 to 120
O2 scale Auto; Specifies the scale for the O2 wave-
[%] 0 to 100; 15 to 35; 25 to 45; 35 to 55; form.
45 to 65; 55 to 75; 65 to 85; 75 to 95;
85 to 105
CO2 scale [%]; [kPa]: Auto; 0 to 6; 0 to 12 Specifies the scale for the CO2 wave-
[mmHg]: Auto; 0 to 50; 0 to 100 form.
Paw scale Auto; -5 to 20; -7.5 to 30; -10 to 40; Specifies the scale for the Paw wave-
[mbar]; [hPa]; -20 to 80 form.
[cmH2O]
Flow-volume loop ISO standard; Dräger Specifies coordinate axes for the Flow-
Volume loop.
CO2; Default color; color palette with Specifies the parameter colors.
Paw; 7 additional colors
Flow, volume
O 2; Default color; ISO color
Agent
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 167
Configuration
168 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 169
Configuration
170 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
MV high Yes; No
inAgent low Yes; No
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 171
Configuration
Default ventila- Buttons with available ventilation modes Specifies the default ventilation mode at
tion mode Man/Spon the start of therapy.
VT and RR start settings Specifies the tidal volume and the
Based on Patient category; Ideal body weight respiratory rate.
The start settings for VT and RR influence the start values of the alarm limits for MV
high, MV low, and VTi high:
The "offset" value corresponds to the respective offset setting for automatic alarm
adjustment. The "offset" value can be set in vertical tab System setup > Alarms >
Autoset.
172 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
Start settings for ventilation Specifies the start settings for the venti-
Pmax 12 to 80 12 to 80 12 to 80 lation.
[mbar]; [hPa]; 40 30 25
[cmH2O]
Pinsp 7 to 80 7 to 80 7 to 80
[mbar]; [hPa]; 15 15 15
[cmH2O]
∆Psupp Off; Off; Off;
[mbar]; [hPa]; 3 to 80 3 to 80 3 to 80
[cmH2O] 10 10 10
Insp term 5 to 80 5 to 80 5 to 80
[%] 25 25 25
PEEP Off; Off; Off;
[mbar]; [hPa]; 2 to 35 2 to 35 2 to 35
[cmH2O] 3 3 3
Slope 0 to 2 0 to 1.5 0 to 1.5
[s] 0.2 0.2 0.2
RRmin Off; Off; Off;
[/min] 3 to 25 3 to 25 3 to 25
6 10 15
% Tplat 0 to 60 0 to 60 0 to 60
[%] 20 20 20
Trigger 0.3 to 15 0.3 to 15 0.3 to 15
[L/min] 4.0 2.0 1.0
Sync. SIMV; CMV SIMV; CMV SIMV; CMV
11.3.4.3 Vertical tab "Fresh gas" (only with electronically controlled gas mixer)
Start settings for fresh gas Selects the start settings for the fresh-
gas delivery.
FG O2 21 to 100 21 to 100 21 to 100 Sets the O2 flow.
[%] 100 100 100
FG flow 0.20 to 0.20 to 0.20 to Sets the fresh-gas flow.
[L/min] 15.00 15.00 15.00
2.00 2.00 2.00
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 173
Configuration
Minimal O2 flow Off; 50 to Off; 50 to Off; 50 to Sets the minimal O2 flow that is deliv-
(carrier gas: Air) 300 300 300 ered when Air is used as carrier gas.
[mL/min] Off Off Off
Do not set the value too low, but suit-
able for the patient category. The rec-
ommendation is, e.g., 200 for adults,
100 for pediatric patients, and 50 for
neonates.
Minimal O2 flow 50 to 300 50 to 300 50 to 300 Sets the minimum O2 flow that is deliv-
(carrier gas: N2O) 200 200 200 ered when N2O is used as carrier gas.
[mL/min]
Do not set this value too small; recom-
mended is, e.g., 200 for adults, 100 for
pediatric patients, and 50 for neonates.
Carrier gas Air; N2O Air; N2O Air; N2O Sets the carrier gas.
174 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 175
Configuration
Gas supply
Disable N2O On; Off A device with connectors for nitrous
oxide can be configured so that nitrous
oxide is no longer displayed, nor can it
be selected as a carrier gas.
Prerequisite:
– The current carrier gas is Air.
– Nitrous oxide is not connected or
available.
Restart the device if necessary to
implement the settings.
On: Nitrous oxide cannot be delivered.
Off: Nitrous oxide can be delivered.
176 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
General
Always use walk- On; Off Sets the walk-through mode function.
through mode
When On is configured, system tests
and leakage tests will always be exe-
cuted in walk-through mode.
Sample line is On; Off Specifies whether the sample line is
connected during connected to the Y-piece or to the filter
test on the Y-piece during the automatic
tests.
When On is configured, the device can
automatically check whether, e.g.,
breathing hoses or central supply
hoses are incorrectly connected.
When Off is configured, additional
checks are required during the system
test.
Test gas supply
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 177
Configuration
Central O2 supply On; Off Specifies which gas supplies are tested
Central Air supply On; Off during the automatic system test.
A logbook entry with measured values is created Enables the creation of additional log-
book entries with measured values.
Every 1 min; 2 min; 5 min; 10 min; 15 min Generates periodic entries.
For all high-prior- On; Off Generates entries in the event of an
ity alarms alarm.
For all medium- On; Off
priority alarms
178 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 179
Configuration
180 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
36021
System setup
No. Designation
1 General information
– Installed software version
– Next maintenance date
2 QR code for further product information
This information can be retrieved:
– Device description
– Device options
– Available accessories
– Service options
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 181
Configuration
Prerequisite: USB mass storage device has been connected to the USB port.
► Open the System setup > Im/Export config. > Im/Export config. page (1).
36030
1
System setup
1
2
4
The configurations saved on the USB mass storage device are displayed in a
list (4). If not all of the configurations are displayed, delete all configurations from
the USB mass storage device that are not needed or move them to a subdirectory
on the USB mass storage device.
The following settings are neither imported nor exported:
– Device name
– Date and time
– IP address
182 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
36022
System setup
1
1
2
5
4
3
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 183
Configuration
36158
CO2 in/et
CO2
36156
CO2 in/et, RR
CO2
36157
Paw
Paw
36162
Paw (3)
Volume-controlled modes:
Parameters PIP, Pplat, PEEP
All other modes:
Parameters PIP, Pmean, PEEP
Paw
36163
Paw (4)
Parameters PIP, Pplat, Pmean, PEEP:
Paw
36164
Volume MV, VT
Volume
36214
184 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
36166
Volume VT, MV
Volume
36216
O2
O2
36160
O2, ∆
O2
36161
Primary agent
Iso
36155
"Empty"
36159
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 185
Configuration
RR
Volume-controlled modes:
Parameters PIP, Pplat,
PEEP
All other modes:
Parameters PIP, Pmean,
PEEP
MVmand, spon MV, VT, RR MV, VT
MV Volume Volume
Mand Spon MV VT MV VT
RR
Volume Volume O2
VT MV VT MV in
RR
O2 Iso RR
in ∆ in et
186 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
Volume
Paw
Set Start
Compliance
Compliance
Cdyn mean
PV loop
42195
Volume
Paw
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 187
Configuration
36178
Surplus
Efficient
Deficit
O2 N2O Econometer
42191
O2 Air N2O
188 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
When the PV loop parameter field is displayed, the flow tubes will be displayed at
reduced size:
42192
O2 Air N2O
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 189
Troubleshooting
12 Troubleshooting
12.1 Leakage
Leakage may result in the system not being operational or being ready for operation
with limitations only.
190 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
Component Measure
Sample line Remove the sample line and seal the Luer-Lock connector
on the Y-piece.
Breathing hoses Disconnect the breathing hoses. Connect the inspiratory
port and expiratory port with a hose that is known to be with-
out leakage. Connect the breathing bag directly to the
breathing system.
Vaporizers Remove the vaporizers.
O2 sensor Remove the O2 sensor. Seal the sensor port with the sealing
cap.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 191
Troubleshooting
WARNING
Risk of patient injury
If all power sources fail, the screen goes dark and mechanical ventilation ends.
► Ventilate the patient manually.
Further procedures:
1. Check the vaporizer setting.
2. Electronically controlled gas mixer:
Use emergency O2 delivery, see page 37
Mechanically controlled gas mixer:
Close the flow control valves for Air and N2O. Use only O2 as fresh gas.
3. Electronically controlled gas mixer:
Monitor the O2 flow with the O2 flowmeter.
Mechanically controlled gas mixer:
Monitor the O2 flow on the total flow tube.
4. Ventilate the patient manually.
5. Ensure corresponding substitute monitoring.
192 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
WARNING
Risk of patient recovering consciousness
If the gas supply fails completely, further operation takes place through gas supply
of the anesthesia machine with ambient air. The ventilation continues with the
ambient air, however no more anesthetic agent or additional oxygen is delivered.
The inspiratory anesthetic gas concentration and the inspiratory oxygen
concentration in the breathing gas fall.
► Monitor the gas mixture carefully and use intravenous anesthetic agents if
necessary.
If the central O2 supply and the central Air supply fail at the same time and no gas
cylinders are connected, the patient can nevertheless be mechanically ventilated.
This is possible because the piston ventilator does not require a drive gas.
1. CAUTION
Risk of increased anesthetic gas concentrations in the ambient air
If the breathing bag is not connected, expiratory anesthetic gases may escape
from the breathing system.
► Ensure adequate circulation of the ambient air.
Remove the breathing bag.
2. Continue the mechanical ventilation.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 193
Troubleshooting
When the breathing bag is removed, the missing fresh-gas volume will
automatically be filled by ambient air. Nevertheless, the Fresh gas low or leakage
alarm is likely to be triggered.
36243
Internal FG flow failure
In the event of a fault, illustrations and instructions showing how to start the
emergency O2 delivery are displayed in areas (1) and (2).
The emergency O2 delivery is started as follows:
WARNING
Risk of patient injury
If the gas mixer fails, no fresh gas is delivered.
► Check the vaporizer setting.
► Supply the patient with O2.
► Use the emergency O2 delivery.
1. Set the O2 switch upwards to the Add. O2 position. (Follow the illustration on the
screen.)
The Internal FG flow failure alarm will then be automatically downgraded.
2. Open the flow control valve on the O2 flowmeter. Set the desired flow. This O2
flow flows through the vaporizer.
194 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
1. WARNING
Risk due to faulty O2 sensor
If the O2 measurement fails, it will no longer be possible to monitor the patient
adequately.
► Make sure that substitute O2 monitoring conforming to the general safety
requirements is available.
Perform a system test and calibrate the O2 sensor.
2. If the calibration fails and the O2 sensor failure alarm is displayed, refer to the
following table:
Cause Remedy
During the calibration, the sensor was During calibration:
exposed to a gas mixture with a fluctu- – Remove the sensor from the breath-
ating O2 concentration. ing system and place it on the work
The O2 sensor was not correctly surface.
placed.
– Do not fill vaporizers during the cali-
The O2 sensor was placed close to
bration.
mobile radio equipment or similar
sources. – Close the additional O2 flow delivery
for the fresh gas mixer.
– Do not blow into the sensor.
– Do not use any disinfectants.
– Close off other gas sources in the
vicinity of the sensor.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 195
Troubleshooting
Cause Remedy
The O2 sensor cell is not present or is Check the O2 sensor cell:
not plugged in correctly. – Insert the sensor cell if it is not pres-
ent.
– Check that the sensor cell is cor-
rectly seated.
– Calibrate the sensor cell again.
An initialization phase of up to 30
minutes is required after a new O2
sensor cell is inserted. Only then is
O2 measurement and calibration
possible once more.
► WARNING
Risk due to gas measurement failure
If the gas measurement fails, the patient can no longer be adequately
monitored.
► Check the sample line and the water trap for damage or blockages. Pay
attention to the replacement intervals.
► Replace the patient-gas measurement module.
► Ensure corresponding substitute monitoring.
Arrange for appropriate substitute monitoring conforming to ISO 80601-2-55.
1. WARNING
Incorrect gas measurement
Faulty or non-functional PGMs can cause incorrect gas measurements. As a
result, the patient could be put at risk.
► Have the PGM replaced by service personnel.
196 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
At the left-hand side of the device, take hold of the cover (1) at the top and
bottom and pull it off directly to the front.
40886
2. On the underside, remove the water trap (2) and the RFID antenna, if present.
To do this, gently press the side surfaces (3) of the RFID antenna together. Pull
the RFID antenna off downwards and leave it on the device.
Do not pull on the antenna cable. Do not kink the antenna cable.
40837
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 197
Troubleshooting
WARNING
Risk of device malfunction
If the breathing bag does not fill with fresh gas, the patient cannot be sufficiently
ventilated.
► Check the oxygen supply. If necessary, open the gas cylinder valves.
► If fresh gas still is not delivered or manual ventilation is not possible, close the
flow control valve of the O2 flowmeter.
► Disconnect the patient from the device. Use a substitute device.
1. Activate the backup manual switch (see "Backup manual mode", page 38).
2. Follow the instructions on the product label.
3. If no ventilation of the patient is possible, use a manual resuscitator. Use a
substitute device.
198 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
12.10.1 Replacing the particle filter of the active anesthetic gas receiving
system (AGS)
1. Fold down the cover (1).
38127
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 199
Troubleshooting
4. CAUTION
Risk of contaminating the ambient air
A filter cover that is not properly closed can cause contamination of the ambient
air with anesthetic gases.
► Take care that the cover is properly closed.
Fold up the cover.
Disassembly
1. Disconnect the transfer hose (1).
38134
2. Disconnect the scavenging hose (2).
3. Unfasten the 2 screws (3).
4. Withdraw the AGS to the side.
5. Dispose of the AGS.
The AGS must not be reprocessed.
Assembly
1. Slide in the new AGS.
2. Fasten the AGS with 2 screws.
3. Connect the scavenging hose.
4. Connect the transfer hose.
200 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
In order to classify alarms having the same alarm priority, internal priority numbers
are given in the table below. The most critical alarm within an alarm priority is given
the number 255.
The following table lists the alarm messages in alphabetical order. If an alarm
occurs, the table helps to quickly identify causes and remedies. The possible
causes and remedial measures should be looked through in the order they are
listed until the alarm is resolved.
Some alarms are listed several times because their priority may change under
certain conditions, see page 160.
Some alarms only occur if the corresponding option is installed.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 201
Troubleshooting
202 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 203
Troubleshooting
204 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 205
Troubleshooting
206 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 207
Troubleshooting
208 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 209
Troubleshooting
210 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 211
Troubleshooting
212 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 213
Troubleshooting
214 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 215
Troubleshooting
216 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 217
Troubleshooting
218 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 219
Troubleshooting
220 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 221
Reprocessing
13 Reprocessing
13.1 Safety instructions
Reusable products
Reusable products must be reprocessed, otherwise there is an increased risk of
infection.
► Follow the infection prevention policies and reprocessing regulations of the
health-care facility.
► Follow the national infection prevention policies and reprocessing regulations.
► Use validated procedures for reprocessing.
► Reprocess reusable products after every use.
► Follow the manufacturer's instructions for cleaning agents, disinfectants, and
reprocessing devices.
Signs of wear, e.g., cracks, deformation, discoloration, or peeling, may occur with
reprocessed products.
► Check the products for signs of wear and replace them if necessary.
Disposable products
Disposable products such as, e.g., the sample line and the water trap, have been
designed, tested, and manufactured exclusively for single use. Reuse,
reprocessing, or sterilization can result in failure of the accessory, incorrect
measurements, and injury to the patient.
► Do not reuse disposable products.
► Do not reprocess disposable products.
► Do not use any disinfectants.
► Do not reprocess with elevated pressure from syringes or compressed air.
Risk of infection
The breathing system may be contaminated with infectious agents. The following
causes may be present:
– No bacteria filters have been used at the Y-piece or at the expiratory port.
– The breathing system is being used for the first time.
Perform the following measures:
► Use a reprocessed breathing system.
► Reprocess the breathing system if necessary.
► To prevent future contamination, use patient-side bacteria filters.
222 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
Flow sensors
High temperatures arise in the flow sensors, for example, in operation or while
calibrating during a system test. Due to the high temperatures, residual vapors of
flammable disinfectants (e.g., alcohols) and residues that were not removed during
reprocessing may ignite. As a result, user and patient could be put at risk.
► Ensure particle-free cleaning and disinfection.
► After disinfection, allow the flow sensor to air for at least 30 minutes.
► Before inserting the flow sensor, check for visible damage and soiling, such as
residual mucus, medication aerosols, and particles.
► Replace flow sensors when damaged, soiled, or not particle-free.
Incorrect reprocessing or soiling, e.g., residues or particles, can damage the flow
sensor. As a result, the patient could be put at risk.
► Do not use machine disinfection or cleaning
► Do not use plasma sterilization or radiation sterilization
► Do not use water jets, compressed air, brushes, or similar
► Do not use ultrasonic baths
► Do not use hot steam sterilization on the Spirolog and Infinity ID flow sensors
► Clean and disinfect the flow sensor in accordance with the associated
instructions for use.
► Use only clean disinfectant solutions to disinfect the flow sensor.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 223
Reprocessing
224 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 225
Reprocessing
13.5.2 Disinfectants
Use disinfectants that are nationally approved and are suitable for the particular
reprocessing procedure.
226 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
13.6 Preparation
13.6.1 Disassembling the patient-specific accessories
1. WARNING
Risk of infection
Used sample lines and water traps may be infectious due to the breathing gases
that passed through them.
► Replace the sample line and the water trap regularly in the following
situations:
– If the sample line is connected to the filter on the Y-piece, replace it daily.
– If there is no filter fitted to the Y-piece and the sample line is connected
directly to the Y-piece, replace the sample line after every patient.
► Remove the sample line from the water trap.
► Initially leave the water trap fitted to prevent infectious fluid from spurting out.
Remove the water trap only after surface disinfection.
► Replace the water trap after each patient.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 227
Reprocessing
37165
2. Remove the breathing bag (2) together with the breathing bag hose.
3. Remove the breathing circuit (3) and the filters at positions (4) or (5).
1. CAUTION
Risk of injury to patients due to failure of accessories
Disposable products have been designed, tested, and manufactured exclusively
for single use. Reuse, reprocessing or sterilization can result in failure of the
accessory and injury to the patient.
► Do not reuse, reprocess, or sterilize disposable products.
Dispose of the disposable accessories:
– Sample line
– Breathing circuit
– Breathing bag
– Filter
2. Reprocess the reusable accessories in accordance with their instructions for
use.
– Breathing circuit
– Breathing bag
228 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
2. WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
– Damage to the device
– Electric shock
– Device malfunctions
► Ensure that no liquid penetrates the device.
CAUTION
Risk of contaminating the device
If the water trap is removed, fluid may get into the main device while it is being
disinfected.
► Leave the water trap fitted during surface disinfection.
Perform the surface disinfection of the entire device by wiping backwards and
forwards at least 3 times:
– In particular, disinfect device surfaces and hand contact points that are
touched frequently.
– Follow the list of validated disinfectants (see "Validated reprocessing
procedures", page 225).
The following illustration shows the frequently touched surfaces and hand
contact points:
39472
No. Designation
1 Screen
2 Control dials of the vaporizers
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 229
Reprocessing
No. Designation
3 Rotary knob
4 Control elements on the gas mixing unit
5 Work surface, side table (option)
6 Handles
7 Handles on drawers
8 Absorber
9 Circuit plug on bag elbow
10 Water trap on the patient-gas measurement module (PGM)
11 APL valve
12 Breathing system cover
13 O2+ key
14 Flow control valve on external O2 flowmeter (option)
15 Pressure reducers (on the rear of the device, not shown)
13.8 Disassembly
13.8.1 Preparation
1. Switch off the device and all devices connected to it.
2. Disconnect all power plugs.
► WARNING
Risk of infection
The water trap may contain infectious fluid.
► Proceed carefully when emptying and take protective measures if necessary.
► Follow the infection prevention policies and reprocessing regulations of the
health-care facility.
230 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
39629
13.8.3 Disassembling the CO2 absorber
13.8.3.1 Reusable CO2 absorber
1. Unscrew the CO2 absorber (1) from the breathing system.
37107
2. Remove and dispose of the optional disposable dust filter (2).
3. CAUTION
Risk of chemical burns
Soda lime is caustic and is a strong irritant for eyes, skin, and airway.
► Handle the soda lime carefully and do not spill it.
Empty out the used soda lime and dispose of it according to its instructions for
use.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 231
Reprocessing
4. Remove the absorber insert (3) from the absorber container. The sealing ring (4)
remains on the absorber insert.
39487
3
38098
2. Withdraw the CLIC absorber (2) upwards from the holder (3).
3. Fold the holder back again until it engages audibly.
4. Dispose of the CLIC absorber in accordance with its instructions for use.
1. Unfasten the knurled screws on the attachment piece (1) for the breathing bag
arm.
2. Remove the attachment piece (1) with the breathing bag arm from the socket (2)
on the breathing system.
232 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
37062
2. Turn the O2 sensor (2) counterclockwise.
37079
3. Take out the O2 sensor (3).
37080
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 233
Reprocessing
37172
5. Remove the sensor cap (5).
37173
Do not remove the sensor cell. Otherwise, the calibration data will be lost.
This may result in a waiting time when putting the device back into operation.
6. WARNING
Risk of incorrect gas measurement
Reprocessing and sterilization can damage the O2 sensor cell. This may result
in a malfunctioning measurement and the patient may be put at risk.
► Do not disinfect or sterilize the O2 sensor cell, or immerse it in other fluids.
Place the O2 sensor as shown (6).
40978
234 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
41134
13.8.6 Removing the breathing system
To prevent accidental penetration of soda lime into the breathing system, make
sure that the reusable CO2 absorber has been removed.
37063
2. CAUTION
Risk of injury due to breathing system warmer
When the breathing system warmer is switched on, the bottom side of the
breathing system and the heating plate beneath it can become very hot.
► Allow the breathing system to cool off before removing.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 235
Reprocessing
Remove the breathing system (2) vertically upwards from the breathing system
mount.
37069
13.8.7 Removing the piston diaphragm
► Remove the piston diaphragm (1) from the ventilator. If necessary, remove any
residual moisture present.
37057
Prerequisite:
– The breathing system cover has been removed.
– The breathing system has been removed.
236 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
– Damage to the device
– Electric shock
– Device malfunctions
► Ensure that no liquid penetrates the device.
39534
3. Repeat steps 1 and 2 (4 times each).
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 237
Reprocessing
39538
No. Designation Remarks
1 Upper part of the breathing system
2 Non-return valve (yellow)
3 Non-return valve (blue)
4 Flow sensors Reprocessing is performed in accordance with
their own instructions for use.
5 Lower part of the breathing system
6 Non-return valve (black)
7 CLIC adapter (option)
8 Incident flow mesh, expiration Remains in the expiratory port.
9 Expiratory port
10 Inspiratory port
11 Bag elbow Remains at the lower part of the breathing system.
12 Breathing bag arm (option) When the breathing bag arm is used, there is no
bag elbow fitted to the lower part of the breathing
system.
13 Incident flow mesh, inspiration Remains in the lower part of the breathing system.
14 APL valve
238 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
37744
2. Press the release button (2). The holder (3) flips open.
37745
3. Push the cleaning plate (4) fully into the adapter.
4. Close the holder (3) so that it engages.
CAUTION
Risk of failure of flow measurement
Improper reprocessing and soiling, such as residues or particles, can damage the
flow sensor.
► No machine cleaning or disinfection
► No plasma sterilization or radiation sterilization
► No water jets, compressed air, brushes or the like
► No ultrasonic bath
► No hot-steam sterilization with Spirolog and Infinity ID flow sensors
► Clean and disinfect the flow sensor in accordance with the corresponding
instructions for use.
► For disinfecting the flow sensor use only clean disinfectant solutions.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 239
Reprocessing
37071
2. Remove the expiratory flow sensor (2).
3. Unfasten and remove the inspiratory port (3).
4. Remove the inspiratory flow sensor (4).
If the breathing system is equipped with a bag elbow, the bag elbow remains on the
breathing system.
37174
240 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
► WARNING
Risk of damage to breathing system
If the APL valve is not removed before the breathing system is reprocessed, it
may be damaged by the reprocessing measures. This can lead to leaks in the
breathing system.
► Always remove the APL valve prior to reprocessing.
Using the lower knurled nut (1), unscrew the APL valve from the upper part of
the breathing system.
39498
13.9.2.5 Non-return valves and fresh-gas decoupling valve
On the top side of the lower part of the breathing system:
► Remove the yellow inspiratory valve (1) and the blue expiratory valve (2) from
the lower part of the breathing system.
37058
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 241
Reprocessing
37060
13.9.2.6 List of disassembled components
After complete disassembly of the breathing system, the following machine-
reprocessable components are present:
– Upper part of the breathing system
– Lower part of the breathing system
– Sensor cap of the O2 sensor
– APL valve
– Inspiratory port
– Expiratory port
– Inspiratory valve (yellow)
– Expiratory valve (blue)
– Fresh-gas decoupling valve (black)
– Piston diaphragm
– Breathing bag arm (option)
Optionally present are:
– Absorber container and absorber insert
– CLIC adapter (option)
242 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 243
Reprocessing
38566
Lower part of the breathing system
► Position the lower part of the breathing system with a slight tilt.
39500
244 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
42213
2. Insert the sensor cap (2) into the flushing adapter.
APL valve
► Lay the APL valve in the basket.
38569
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 245
Reprocessing
39501
Breathing bag arm (option)
1. Connect the breathing bag arm (1) to a hose (2).
39502
2. Lay the breathing bag arm together with the hose in the load carrier.
3. Connect the hose (2) to a flushing nozzle (3).
Absorber insert
1. Connect the absorber insert (1) to a hose (2).
39503
2. Lay the absorber insert together with the hose in the load carrier.
3. Connect the hose (2) to a flushing nozzle (3).
246 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
Absorber container
► Place the absorber container with the opening facing down on an injector
nozzle.
38595
CLIC adapter (option)
► Lay the CLIC adapter in the basket with the cleaning plate facing upwards.
38594
13.10.3 Machine cleaning with thermal disinfection
Use a washer-disinfector that meets the requirements of the standard ISO 15883.
Dräger recommends the use of a load carrier for anesthesia accessories and
ventilation accessories.
Follow the list of validated disinfectants (see "Validated reprocessing procedures",
page 225).
Prerequisite: The components have been positioned appropriately in the load
carrier.
1. Use a suitable cleaning agent.
2. Select a suitable cycle.
Use a cycle with the following phases:
– Pre-wash
– Cleaning using a suitable cleaning agent in accordance with the
manufacturer’s instructions for use
– Final rinsing with demineralized water
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 247
Reprocessing
– Thermal disinfection
– Drying
Proceed as follows after the cycle has ended:
1. Check all components for visible soiling.
2. Repeat the cycle if necessary.
248 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
WARNING
Risk of fire
High temperatures arise in the flow sensors, for example, in operation or while
calibrating during a system test. Due to the high temperatures, residual vapors of
flammable disinfectants (e.g., alcohols) and residues that were not removed during
reprocessing may ignite.
► Ensure particle-free cleaning and disinfection.
► After disinfection, allow the flow sensor to air for at least 30 minutes.
► Before inserting the flow sensor, check for visible damage and soiling, such as
residual mucus, medication aerosols, and particles.
► Replace flow sensors when damaged, soiled, or not particle-free.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 249
Reprocessing
► WARNING
Risk of insufficient anesthetic gas concentration
If the component connections of the breathing system are not sufficiently leak-
tight, ambient air may get into the breathing gas.
► Make sure that all components of the breathing system are connected tightly.
Assemble the breathing system in accordance with the instructions in this
chapter.
36497
250 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
37097
13.11.2.3 Fitting the APL valve
1. WARNING
Risk of an incorrectly set pressure limitation
If the lower dot marked on the APL valve is not aligned correctly, the APL valve
may be set or read off incorrectly.
► When fitting, take care that the lower dot is facing the user during operation.
Align the APL valve correctly. The lower dot (1) must face the user during
operation.
39506
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 251
Reprocessing
2. Place the APL valve (2) vertically on the upper part of the breathing system (3).
39507
3. Tighten the knurled nut (4).
39505
2. Place the upper part of the breathing system (2) on the lower part of the
breathing system (3).
3. Turn the levers (1) by approximately 120° clockwise.
252 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
39508
2. Push in the expiratory port (2). Tighten the knurled nut.
3. Insert the inspiratory flow sensor (3).
4. Push in the inspiratory port (4). Tighten the knurled nut.
39509
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 253
Reprocessing
5. Check the correct orientation of the adapter (4). The two symbols (5) must
align vertically.
39511
13.11.3 Inserting the piston diaphragm
1. Insert the piston diaphragm (1) in the ventilator.
Make sure that the Dräger inscription on the base of the piston diaphragm can
be read from above.
36496
254 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
36503
2. Turn the levers (2) by approximately 120° clockwise. The breathing system is
now locked. On the product variant with integrated patient-gas measurement
module, seal the O2 sensor port with the sealing cap (3).
36504
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 255
Reprocessing
2. Turn the knurled nut (2) about 90° clockwise. Turn it until the palpable resistance
is overcome and a click is heard. The sensor cap is now fitted.
37179
3. Insert the O2 sensor (3) into the sensor port.
38271
4. Turn the O2 sensor (4) clockwise.
38272
256 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
5. Put the breathing system cover (5) in place and click it into position.
36505
13.11.6 Fitting the breathing bag arm (option)
1. Plug the breathing bag arm (1) onto the socket (2) on the breathing system.
Tighten the two knurled screws. Check that the breathing bag arm is held
securely.
36506
2. Align the breathing bag arm so that collisions with other components are
prevented.
39516
2. CAUTION
Risk of chemical burns
Soda lime is caustic and is a strong irritant for eyes, skin, and airway.
► Handle the soda lime carefully and do not spill it.
Fill the CO2 absorber with fresh soda lime to the upper mark.
Recommendation: Use Drägersorb 800 Plus or Drägersorb Free.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 257
Reprocessing
3. WARNING
Risk of hypoventilation and incorrect gas measurement
Reuse of the disposable dust filter can increase the filter resistance and impair
the ventilation function of the device.
► If soda lime from third-party manufacturers is used, e.g., granular soda lime,
use a disposable dust filter and replace it with every change of the soda lime.
If soda lime from third-party manufacturers is used, insert a new disposable dust
filter (3). Only use dust filters from the list of accessories.
Only use undamaged filters.
39515
4. Attach the CO2 absorber (4) to the breathing system from below. Rotate it in the
direction of the arrow until it reaches the stop.
Follow the instructions for use for the particular soda lime.
► WARNING
Risk due to particles and dust
In order to protect the patient from particles and dust, a filter must be used
between the inspiratory limb of the breathing system and the patient.
► Use a filter at the Y-piece or at the inspiratory port.
Complete the device with the following accessories:
– Breathing hoses
– Breathing bag
– Filter
– Water trap
– Sample line
Observe the following information: "Selecting and connecting patient-specific
accessories", page 67.
258 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 259
Reprocessing
Disassembling the patient-specific accessories d. Remove the absorber insert from the absorber con-
tainer.
1. Unscrew the sample line.
11. CLIC absorber (disposable absorber)
2. Remove the breathing bag together with the breathing
bag hose. a. Remove and dispose of the CLIC absorber.
3. Disassemble the breathing circuit and the filters. Disassembling the breathing bag arm (option)
Disposing of or reprocessing patient-specific accesso- 12. Unscrew the breathing bag arm (option).
ries
Disassembling the O2 sensor
4. Dispose of the following disposable accessories:
13. Disassemble the O2 sensor.
– Sample line
14. Remove the sensor cap.
– Breathing circuit
Removing the breathing system
– Breathing bag
15. Remove the breathing system.
– Filter
5. Reprocess the following reusable accessories: Removing the piston diaphragm
– Breathing circuit 16. Remove the piston diaphragm.
– Breathing bag Disinfecting the breathing system mount
Surface disinfection of the device surfaces 17. Disinfect the breathing system mount.
6. Disinfect the following surfaces on the device: Disassembling the breathing system
– Screen 18. CLIC adapter (option)
– Control dials of the vaporizers a. Unscrew the adapter.
– Rotary knob b. Insert the cleaning plate and close the holder.
– Control elements on the gas mixing unit 19. Disassembling the flow sensors and ports
– Work surface a. Remove the expiratory port.
– Handles b. Remove the expiratory flow sensor.
– Handles on drawers c. Remove the inspiratory port.
– Absorber d. Remove the inspiratory flow sensor.
– CLIC adapter (option) 20. Lift off the upper part of the breathing system.
– Circuit plug on bag elbow 21. Screw off the APL valve.
– Water trap on the integrated patient-gas measure- 22. Remove the non-return valves.
ment module (PGM)
– Inspiratory valve (yellow)
– APL valve
– Expiratory valve (blue)
– Breathing system cover
23. Remove the fresh-gas decoupling valve (black).
– O2+ key
– Flow control valve on external O2 flowmeter (option) Machine reprocessing
– Pressure reducers (on the rear of the device) 24. Position the following parts appropriately and perform
machine reprocessing:
Preparation for disassembly – Upper part of the breathing system
7. Switch off the device and all devices connected to it. – Lower part of the breathing system
8. Disconnect all power plugs. – Sensor cap of the O2 sensor
Disassembling the water trap – APL valve
9. Pull out and dispose of the water trap. – Inspiratory port
260 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 261
Service
14 Service
14.1 Safety instructions
Intervals and implementation
Wear and material fatigue of the components may lead to device failure and
malfunctions. If services activities are not performed regularly or properly,
malfunctions may occur, which can result in personal injury and property damage.
► Perform service at the specified intervals.
► Service activities must be performed by those user groups that are assigned to
the particular measure.
► Only perform service activities when there is no patient connected to the device.
► Before performing maintenance, disconnect all electrical connections from the
power supply and all gas connections from the gas supply.
► Perform a system test after service activity.
Housing
Under the housing, there are live electrical components, which may cause an
electric shock.
► The housing may only be opened by those user groups that are assigned to that
particular measure.
Risk of infection
If the product has not been reprocessed properly, it may be contaminated with
pathogens. As a result, persons could be put at risk.
Perform the following measure before servicing, before returning the device, and
before disposal:
► Reprocess the product in accordance with the chapter "Reprocessing".
14.3 Inspection
Measure Interval User group
Inspection and safety check Every 12 months Service personnel
262 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Service
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 263
Service
264 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Service
14.4 Maintenance
Component Interval Measure User group
CO2 absorber / soda When colored violet or Replace, see page 75 Users
lime with dust filter according to the config-
ured Infinity ID replace-
ment interval
Water trap (option) As required, when Replace, see page 77 Users
soiled, or according to
the configured Infinity ID
replacement interval
Flow sensors As required, if calibra- Replace, see page 198 Users
tion is no longer possi-
ble, or according to the
configured Infinity ID
replacement interval
Piston diaphragm Annually Replace, see page 236 Users
Inspiratory O2 sensor Every 2 years Replace, see page 55 Users
(option)
AGS As required, when Replace Specialized service per-
soiled sonnel
AGS filter If required, if filter is Replace Users
soiled, or flow is no lon-
ger achieved
O-rings on the water Every 2 years Replace Service personnel
trap mount (option)
Pressure reducer Every 6 years Replace Specialized service per-
(option) sonnel
Filter cloth Every 2 years Replace Service personnel
– Above the gas mixer
– Power supply unit
CLIC adapter (option) Every 4 years Replace Users
Lead-gel battery Every 2 years Replace Service personnel
(2 pieces) Or after determination of Replace Specialized service per-
remaining capacity sonnel
14.5 Repair
Repairs may only be performed by specialized service personnel.
It is recommended that only original parts from Dräger be used for repairs and that
the repairs be performed by Dräger.
The replacement of the integrated patient-gas measurement module is described in
the chapter "Replacing the patient-gas measurement module (PGM)" (see
"Replacing the patient-gas measurement module (PGM)", page 196).
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 265
Disposal
15 Disposal
15.1 Safety instructions
Risk of infection
If the product has not been reprocessed properly, it may be contaminated with
pathogens. As a result, persons could be put at risk.
Perform the following measure before servicing, before returning the device, and
before disposal:
► Reprocess the product in accordance with the chapter "Reprocessing".
266 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
16 Technical data
16.1 Safety instructions
16.1.1 Electromagnetic compatibility (EMC)
Medical electrical equipment is subject to special precautionary measures
concerning electromagnetic compatibility. During installation and before initial
operation, follow the information in section: "EMC declaration" (page 291).
Electrostatic discharge
Measures to protect from electrostatic discharge must be adhered to when handling
components carrying the ESD warning symbol. Otherwise, malfunctions may occur
which could put the patient at risk.
To prevent malfunctions, observe the following measures and train the relevant
personnel:
► Observe the ESD protective measures, e.g.:
– Wear antistatic clothes and shoes.
– Use gloves that are electrically insulating and antistatic.
– When establishing connections, touch a potential equalization pin.
► Observe the requirements for the electromagnetic environment. Observe the
following section: "Electromagnetic environment" (page 292).
Electromagnetic disturbances
Wireless communication devices (e.g., cellular phones) and medical electrical
equipment (e.g., defibrillators, electrosurgical devices) emit electromagnetic
radiation. When such devices are operated too close to this device or its cables, the
functional integrity of this device may be compromised by electromagnetic
disturbances. As a result, the patient could be put at risk.
► Maintain a distance of at least 0.3 m (1.0 ft) between this device and wireless
communication devices, to ensure that the essential performance of this device
is fulfilled.
► Maintain an adequate distance between this device and other medical electrical
equipment.
Magnetic fields can adversely affect the functional integrity of the medical device
and thus put the patient or user at risk.
► Do not use the medical device in rooms where devices for magnetic field
applications are used (e.g., magnetic resonance imaging).
The medical device meets the applicable limit values for electromagnetic fields. The
functioning of pacemakers can nevertheless be impaired by emissions.
► All wearers of pacemakers should maintain a distance of at least 25 cm (10 in)
between pacemaker and medical device.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 267
Technical data
268 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
O2 concentration FG O2
Setting range 21 to 100 Vol% (carrier gas: Air)
25 to 100 Vol% (carrier gas: N2O)
Accuracy ±5 % or ±2 Vol% (the larger value applies)
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 269
Technical data
16.7 Ventilator
Design Electronically driven ventilator, fresh-gas
decoupled
Time-based settings
Respiratory rate RR 3 to 100 /min
±10 % of the set value or ±1 /min (the larger
value applies)
270 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 271
Technical data
272 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
APL valve
Setting range Open, 5 to 70 hPa (cmH2O)
Accuracy (at a flow of 20 ±1 L/min): ±20 % of the set value or ±3 hPa (the larger
value applies), but not more than +10 hPa
(cmH2O)
Pressure drop at 30 L/min (ATPD), fully opened Dry: 2.1 hPa (cmH2O)
Wet: 2.2 hPa (cmH2O)
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 273
Technical data
Suction flow
Normal range 30 to 50 L/min
At lower end of restricted range ≥10 L/min
274 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Scavenging hose
Maximum length 8 m (26 ft)
Minimum diameter 19 mm (0.75 in)
Airway pressure
Airway pressure Paw
Plateau pressure Pplat
Positive end-expiratory pressure PEEP
Peak inspiratory pressure PIP
Mean airway pressure Pmean
Range –20 to +99 hPa (cmH2O)
Accuracy ±4 % of the measured value or ±2 hPa
(cmH2O) (the larger value applies)
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 275
Technical data
Tidal volume
Expiratory VT
Range 0 to 2500 mL
Accuracy for gas measurement with PGM ±10 % of the measured value or ±15 mL (the
larger value applies)
Accuracy without PGM ±20 % of the measured value or ±40 mL (the
larger value applies)
Resolution of displayed value 1 mL
Minute volume
Total MV
Mandatory MVmand
Spontaneous MVspon
Range 0 to 40 L/min
Accuracy for gas measurement with PGM ±10 % of the measured value or ±0.1 L/min
(the larger value applies)
Accuracy without PGM ±20 % of the measured value or ±0.2 L/min
(the larger value applies)
Resolution of displayed value 0.01 L/min (MV <1 L/min) or 0.1 L/min
(MV ≥1 L/min)
276 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Low-flow wizard
Standardization The following data are standardized to STPD
conditions.
Range 0 to 8 L/min
Accuracy ±25 % of measured value or ±100 mL/min
(the larger value applies)
Respiratory rate
Total RR
Spontaneous RRspon
Mandatory RRmand
Range 0 to 100 /min
Accuracy ±10 % or ±1 /min (the larger value applies)
Resolution of displayed value 1 /min
T0...90 <45 s (RR ≥6 /min)
<105 s (RR <6 /min)
Dynamic compliance
Dynamic compliance Cdyn
Mean dynamic compliance Cdyn mean
Resistance R
Range 0 to 100 hPa/L/s (cmH2O/L/s)
Accuracy for gas measurement with PGM ±30 % of the measured value or ±3 hPa/L/s
(cmH2O/L/s) (the larger value applies)
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 277
Technical data
Elastance E
Range 0.005 to 10 hPa/mL (cmH2O/mL)
Resolution of displayed value 0.001 hPa/mL (cmH2O/mL)
Supply pressures
Central supply
Range 0 to 9.8 kPa x 100
0 to 140 psi
0 to 0.98 MPa
Accuracy (up to 7 kPa x 100) ±4 % or ±0.2 kPa x 100 (the larger value
applies)
±4 % or ±3 psi (the larger value applies)
±4 % or ±0.02 MPa (the larger value applies)
Resolution of displayed value 0.1 kPa x 100
1 psi
0.01 MPa
Gas cylinders
Range 0 to 250 kPa x 100
0 to 3600 psi
0 to 25 MPa
Accuracy ±4 % or ±6 kPa x 100 (the larger value applies)
±4 % or ±87 psi (the larger value applies)
±4 % or ±0.6 MPa (the larger value applies)
Resolution of displayed value 1 kPa x 100
1 psi
0.1 MPa
278 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 279
Technical data
O2
Range 0 to 100 Vol%
Accuracy ±(2.5 Vol% + 2.5 % relative)
Resolution of displayed value 1 Vol%
T10...90 <500 ms
CO2
Range 0 to 13.6 Vol%
0 to 13.6 kPa
0 to 102 mmHg
Accuracy ±(0.43 Vol% + 8 % relative)
±(0.43 kPa + 8 % relative)
±(3.3 mmHg + 8 % relative)
Resolution of displayed value 0.1 Vol%
0.1 kPa
1 mmHg
T10...90 <350 ms
N 2O
Range 0 to 100 Vol%
Accuracy ±(2 Vol% + 8 % relative)
Resolution of displayed value 1 Vol%
T10...90 <500 ms
Anesthetic gases
Range
Halothane 0 to 8.5 Vol% (kPa)
Isoflurane 0 to 8.5 Vol% (kPa)
Enflurane 0 to 10 Vol% (kPa)
Sevoflurane 0 to 10 Vol% (kPa)
Desflurane 0 to 20 Vol% (kPa)
280 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 281
Technical data
O2
Range 0 to 100 Vol%
Accuracy ±(2.5 Vol% + 2.5 % relative)
Resolution of displayed value 1 Vol%
Typical delays <15 s
Fresh-gas consumption
Standardization The following data are standardized to STPD
conditions.
Range 0 to 99999 L
282 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Scale
Airway pressure –20 to 80 hPa (cmH2O)
Flow –120 to 120 L/min
Volume 0 to 2000 mL
O2 0 to 100 Vol%
CO2 0 to 100 mmHg (0 to 12 Vol%, 0 to 12 kPa)
Anesthetic agent
Halothane 0 to 5 Vol% (kPa)
Enflurane 0 to 6 Vol% (kPa)
Isoflurane 0 to 5 Vol% (kPa)
Sevoflurane 0 to 10 Vol% (kPa)
Desflurane 0 to 20 Vol% (kPa)
Loops Pressure-Volume
Flow-Volume
Power cable
Maximum length 5 m (16.4 ft)
Protective ground resistance Maximum 0.1 Ω
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 283
Technical data
Power consumption
Standby (without charging the internal battery) <40 W
During mechanical ventilation (PC - CMV, <95 W
Pinsp = 15 hPa, RR = 8 /min, I:E = 1:2, PEEP =
0 hPa, FG flow 4 L/min O2) without charging the
internal battery
Maximum 400 W
Peak inrush current Approx. 8 to 14 A
Approx. 6 to 10 A quasi-RMS
Internal battery
Type Lead-gel battery
Sealed, maintenance-free
Capacity 7.2 Ah
Voltage 24 V
Fuse F15A 80V UL248-14, breaking capacity
1000 A, size 19.7 mm x 19 mm x 5 mm
Current Maximum 15 A
Backup time with new and fully charged battery
During mechanical ventilation (PC - CMV, At least 45 min
Pinsp = 15 hPa, RR = 8 /min, I:E = 1:2, Typically 120 min
PEEP = 0 hPa, FG flow 4 L/min O2)
During ventilation in Man / Spon At least 90 min
284 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Gas supply
Gas quality
Oil content <0.1 mg/m3
Dew point 5 °C (41 °F) at ambient temperature
Particle size Dust-free air (filtered with pore size <1 μm)
Supply pressure for O2, Air, N2O 2.7 to 6.9 kPa x 100
39 to 100 psi
0.27 to 0.69 MPa
Maximum short-term peak inlet flow at 6.9 kPa
x 100 (100 psi or 0.69 MPa) supply pressure
O2 135 L/min (applies only when there is no distri-
bution piece for the central O2 supply)
Air
Without bronchial suction system 50 L/min
Including a directly connected bronchial 130 L/min
suction system
N 2O 40 L/min
Drive gas Not needed
Gas supply connection Depending on configuration: NIST, DISS
(CGAV-5/B or CGAV-5/N), French standard
(NFS90-116)
Gas cylinders (dimensions)
Diameter 100 to 140 mm (3.94 to 5.51 in) for versions
with upright gas cylinders
100 to 102 mm (3.94 to 4.01 in) for versions
with hanger yoke system for gas cylinders with
pin-index connections
Maximum height 880 mm (34.64 in) for versions with upright gas
cylinders
757 mm (29.80 in) for versions with hanger
yoke system for gas cylinders with pin-index
connection
Pressure reducers
Version Conforms to DIN EN ISO 10524-1
Permissible inlet pressure (PV)
Air, O2 Up to 200 kPa x 100 (2900 psi, 20 MPa)
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 285
Technical data
286 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
RFID system
Operating frequency 13.56 MHz ± 50 ppm (wideband)
Transmitter power ≤42 dBμA/m (200 mW ± 1 dB)
Modulation ASK (amplitude shift keying)
Electromagnetic compatibility Tested in compliance with IEC 60601-1-2
Protection classes
Device I, in compliance with IEC 60601-1
Applied parts (connections for breathing hoses) TYPE BF
Degree of protection IP20 according to IEC 60529,
meets ISO 80601-2-13
Classification in compliance with Directive II b
93/42/EEC, Annex IX
UMDNS code 10-134
Use of latex The device is made without natural rubber
latex.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 287
Technical data
288 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 289
Technical data
16.21 Diagrams
16.21.1 Pressure characteristics and flow characteristics of the
breathing system
Breathing system with and without breathing hoses and filters (conforming
to ISO 8835-2 and ISO 80601-2-13):
41381
3UHVVXUH>K3DFP+2@
ಣ
ಣ
ಣ
ಣ
ಣ
ಣ
ಣ
ಣ
ಣ
ಣ
ಣ
ಣ
ಣ
ಣ
ಣ
)ORZ>/PLQ@
Breathing system, dry, with filled Peak flow in Resistance [hPa (cmH2O)]
reusable CO2 absorber and soda use [L/min] Man / Spon
lime dust filter MX50115 Inspiratory Expiratory
60 –3.8 4.5
Without breathing circuit and inspi- 30 –1.2 2.9
ratory filter 15 0.0 2.0
2.5 0.7 1.6
290 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Breathing system, dry, with filled Peak flow in Resistance [hPa (cmH2O)]
reusable CO2 absorber and soda use [L/min] Man / Spon
lime dust filter MX50115 Inspiratory Expiratory
With breathing circuit for adults 60 –5.6 5.5
MP00349, inspiratory filter
MP01730 30 –2.0 3.4
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 291
Technical data
Emissions Compliance
Radiated emissions Class A, group 1 (30 MHz to 1 GHz)
Conducted emissions Class A, group 1 (150 kHz to 30 MHz)
NOTICE
The emissions characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for
which CISPR 11 class B is normally required), this equipment might not offer
adequate protection to radio-frequency communication services.
► The user might need to take mitigation measures, such as relocating or re-
orienting the equipment.
292 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 293
Technical data
294 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 295
Technical data
296 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Principles of operation
17 Principles of operation
17.1 Safety instructions
Infinity ID components
It is only possible to use the additional functions of Infinity ID accessories if the
device has the additional functions for replacement interval monitoring and anti-
interchange security.
The use of these additional functions does not ensure the maximum period of use
for the accessories and any correctly connected hoses. The user is thus not
excused from regularly checking the accessories. If these accessories are not
checked, the patient may be put at risk.
► Use the appropriate Infinity ID accessories to utilize the additional functions.
► Check the current condition and the period of use of the accessories regularly.
► Check regularly that the hoses are correctly connected.
There are product-specific data saved on the Infinity ID accessories, which are
further processed by the device. If an unused Infinity ID accessory is in the
immediate vicinity of the device, values such as those for resistance and
compliance may be transferred inadvertently from this accessory. When
accessories are added to an Infinity ID breathing circuit, the values for compliance
and leakage may deviate from those saved on the breathing circuit.
► Do not keep unused Infinity ID accessories in the vicinity of the device.
► To determine the actual values for compliance and resistance, always perform
the leakage test before starting the therapy. If the test cannot be performed
because the patient is already connected, particular attention is required during
ventilation.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 297
Principles of operation
298 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Principles of operation
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 299
Principles of operation
300 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Principles of operation
Additional parame-
Basic parameter
ters
Group Tabs Ventilation mode (normal therapy
(expanded therapy
bar)
bar)
Volume-con- VC - AF VC - CMV / Auto- Pmax Slope
trolled ventilation Flow VT
RR
PEEP
Ti
CMV
VC - SIMV / Auto- Pmax Trigger
Flow VT ∆Psupp1) = Off
RR Slope
PEEP
Ti
SIMV
VC - SIMV / PS / Pmax Trigger
AutoFlow1) VT ∆Psupp1) >0
RR
Insp term1)
PEEP
Slope
Ti
SIMV
VC VC - CMV Pmax % Tplat
VT
RR
PEEP
Ti
CMV
VC - SIMV Pmax Trigger
VT ∆Psupp1) = Off
RR % Tplat
PEEP
Ti
SIMV
VC - SIMV / PS1) Pmax Trigger
VT ∆Psupp1) >0
RR
Insp term1)
PEEP
Slope
Ti
% Tplat
SIMV
1) Requires the PSV option
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 301
Principles of operation
36250
Paw ∆Psupp
PEEP
Slope 7LPH
1/RRmin
Flow
Trigger
Insp term
7LPH
3DWLHQWWULJJHUHGEUHDWK
0DFKLQHWULJJHUHGEUHDWK
302 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Principles of operation
36259
Paw Pinsp 7ULJJHUZLQGRZ
LQVS
PEEP
∆Psupp
Slope 7LPH
Slope
Ti
1/RR
Flow
Trigger
Trigger
Insp term
7LPH
0DFKLQHWULJJHUHGEUHDWK
3DWLHQWWULJJHUHGEUHDWK
17.2.5.2 PC - CMV
– Pressure-controlled
– Time-controlled
– Machine-triggered
The mandatory breaths are machine-triggered and are not triggered by the patient.
17.2.5.3 PC - SIMV
– Pressure-controlled
– Time-controlled
– Machine-triggered
– Synchronized inspiration
In PC - SIMV, the patient can breathe spontaneously at any time, while the number
of mandatory breaths is predefined. When synchronization is switched on, the
breaths are adapted to the spontaneous breathing efforts of the patient. If
spontaneous breathing effort by the patient is detected during the inspiratory trigger
window, a patient-triggered breath will be initiated.
17.2.5.4 PC - SIMV / PS
This mode is similar to PC - SIMV, except that the patient's spontaneous breathing
at the PEEP level during the expiratory phase is pressure-supported with ∆Psupp
when outside the trigger window.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 303
Principles of operation
17.2.6.2 VC
36287
7ULJJHUZLQGRZ
LQVS
Paw
Pmax
∆Psupp
PEEP
% Tplat 7LPH
Slope
Ti
1/RR
Flow
Trigger Trigger
Insp term
7LPH
0DFKLQHWULJJHUHGEUHDWK
3DWLHQWWULJJHUHGEUHDWK
17.2.6.3 VC - CMV
– Volume-controlled
– Pressure-limited
– Time-controlled
– Machine-triggered
– Constant inspiratory flow
In this volume-controlled ventilation mode, the patient receives the set tidal volume
VT with each mandatory breath.
304 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Principles of operation
17.2.6.4 VC - SIMV
– Volume-controlled
– Pressure-limited
– Time-controlled
– Machine-triggered or patient-triggered
– Constant inspiratory flow
– Synchronized inspiration
In VC - SIMV, the patient can breathe spontaneously at any time, while the number
of mandatory breaths is predefined. When synchronization is switched on, the
breaths are adapted to the spontaneous breathing of the patient. If inspiratory effort
by the patient is detected during the inspiratory trigger window, a patient-triggered
breath will be initiated.
17.2.6.5 VC - SIMV / PS
This mode is similar to VC - SIMV, except that the patient's spontaneous breathing
at the PEEP level during the expiratory phase is pressure-supported with ∆Psupp
when outside the trigger window.
17.2.6.6 VC - AF
36290
Paw 7ULJJHUZLQGRZ
LQVS
PEEP H[S
Pmax
∆Psupp
Flow
Trigger Trigger
Insp term
7LPH
0DFKLQHWULJJHUHGEUHDWK
3DWLHQWWULJJHUHGEUHDWK
With AutoFlow, the set tidal volume VT is applied for all mandatory volume-
controlled breaths with the lowest required pressure. The pressure patterns and
flow patterns of the mechanical inspiratory breaths correspond to those of pressure-
controlled ventilation.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 305
Principles of operation
Due to the patient's inspiratory effort or compliance changes in the lungs, the tidal
volume in an individual breath may deviate from the set tidal volume VT. However,
averaged over time a tidal volume corresponding to the set volume VT is applied.
If no mechanical ventilation has previously taken place, a volume-controlled test
breath with constant inspiratory flow is performed first when starting a ventilation
mode with AutoFlow in order to estimate the lung parameters. The inspiratory
pressure required at the start is determined from this test breath. Each additional
breath-related readjustment of the inspiratory pressure is limited to ±3 hPa
(cmH2O). The pressure difference (inspiratory pressure - PEEP) is at least 5 hPa
(cmH2O) and the upper inspiratory pressure limit is set by Pmax. If the set value for
VT is reduced, the inspiratory pressure will be reduced by a greater amount if
necessary.
306 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Principles of operation
40162
No. Designation
1 Piston diaphragm
2 Spindle
3 Electric motor
4 Piston
The piston drive is connected directly to the breathing system. The electric
motor (3) moves the piston (4) by means of a spindle (2). Thus, the fresh gas is
moved out of the piston diaphragm and through the breathing system to the patient.
The user can check through a viewing window whether the piston is moving. During
inspiration, the ventilator applies the tidal volume with the required pressure and at
a defined respiratory rate.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 307
Principles of operation
308 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Principles of operation
41098
*DVVDPSOLQJ
LQVS
H[S
3DWLHQW
As a consequence, the CO2 waveform may fall at the end of the expiration before
the inspiration is executed, in contrast to the normal course of the waveform.
37888
100 CO2
The unusual waveform shape results from the small dead space. The measured
values derived from the CO2 waveform are nevertheless correct, as the etCO2
value corresponds to the maximum expiratory value and is not subject to the
waveform fluctuations.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 309
Principles of operation
The following graphic illustrates the Paw high alarm limit as a suitable indicator in
the event of a soiled filter:
41394
&OHDU+0( 6RLOHG+0(
Paw high
310 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Principles of operation
VT and RR are only dependent on the ideal body weight when the Ideal body
weight function has been selected in System setup > Therapy, see page 172.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 311
Principles of operation
Changing the body weight during mechanical ventilation has no effect on the
current ventilation settings.
312 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Principles of operation
The described acoustic alarm signals are handled by a backup loudspeaker if the
primary acoustic alarm system fails. It plays the acoustic alarm signals of the
"Warning" and "Caution" alarm priorities at a constant tone frequency and
unchanged volume.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 313
Annex
18 Annex
18.1 Abbreviations
Abbreviation Explanation
%, Vol % Percentage gas ratio, related to total volume
A Ampere
Add. O2 Emergency O2 delivery
AGS Anesthetic gas receiving system
AGSS Anesthetic gas scavenging system
Air Medical compressed air
APL Adjustable Pressure Limitation
ASA American Society of Anesthesiologists
ATPS Ambient Temperature and Pressure, Saturated
Aux. O2 O2 insufflation
BIPAP Biphasic Positive Airway Pressure, spontaneous breathing
under continuous positive airway pressure with 2 different pres-
sure levels
BMI Body mass index
BTPS Body Temperature and Pressure, Saturated
37 °C (98,6 °F), ambient pressure, 100 % relative humidity
CAL Display when a measurement value is calibrated.
CBM mode Cardiac bypass mode
Cdyn Dynamic compliance (patient)
CISPR Comité International Spécial des Perturbations
Radioélectriques
International special committee on radio interference
cmH2O Centimeters of water
CMV Controlled Mandatory Ventilation
CO Carbon monoxide
CO2 Carbon dioxide
COM Serial port
CPAP Continuous Positive Airway Pressure
CSA Canadian Standards Agency
dB(A) Sound pressure level, A-weighted
Des Desflurane
∆O2 Difference between inspiratory and expiratory O2 concentration
∆Psupp Pressure support above PEEP
EMC Electromagnetic compatibility
Enf Enflurane
ERR Display when a measured value cannot be determined.
ESD Electrostatic Discharge
314 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Annex
Abbreviation Explanation
FG Fresh gas
FiO2 Inspiratory oxygen fraction
FTP File Transfer Protocol
GPL General Public Licence
Hal Halothane
HF High-frequency
HME Heat and moisture exchanger
HMEF HME filter
hPa Hectopascal
Hz Hertz
I:E Ratio of inspiratory time to expiratory time
ID Identification
Insp term Inspiration termination criterion in % based on peak inspiratory
flow
Iso Isoflurane
kg Kilogram
kPa Kilopascal
L Liter
LAN Local area network
lbs Pound; unit of mass
LED Light-emitting diode
LGPL Lesser General Public Licence
MAC Minimum Alveolar Concentration
Man/Spon
Manual / Manual ventilation / Spontaneous breathing
Spontaneous
mbar Millibar
MEDIBUS.X Communication protocol for medical devices with uniform data
definition for all devices
min Minute
mL Milliliter
mmHg Millimeter of mercury
MPa Megapascal
MV Minute volume
N 2O Nitrous oxide
NTP Network Time Protocol, standard for synchronizing clocks
O2 Oxygen
O2+ O2 flush
Pa Pascal; unit of pressure
Paw Airway pressure
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 315
Annex
Abbreviation Explanation
PEEP Positive end-expiratory pressure
PGM Patient-gas measurement module
Pinsp Inspiratory pressure
PIP Peak inspiratory pressure
Pmax Maximum pressure
Pmean Mean pressure
png Graphics format
Pplat Plateau pressure
ppm Parts per million
QR code Quick Response Code
R Resistance
RFID Radio Frequency Identification
RR Respiratory rate
RRmin Minimum respiratory rate
Sev Sevoflurane
Slope Pressure rise time
SNMP Simple Network Management Protocol
STAPD Standard Temperature, Ambient Pressure, Dry
20 °C (68 °F), dry gas
STPD Standard Temperature and Pressure, Dry
20 °C (68 °F), 1013 hPa
TC Time constant
Ti Inspiratory time
UMDNS Universal Medical Device Nomenclature System
Nomenclature for medical devices
USB Universal Serial Bus
V Volt
VT Tidal volume
xMAC Accumulated multiple of the MAC values of anesthetic agents
and N2O
316 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Annex
18.2 Symbols
Additional information about the symbols is available on the following web page:
www.draeger.com/md-symbols
Symbol Explanation
Manufacturer
WEEE marking
Risk of tipping over! Do not take hold of the device above this
mark in order to push it or pull it.
Group Views...
Group Trends/Data...
Group Alarms...
Group Procedures...
Group Start.../Standby...
Device on/Standby
Key: Start/Standby
Main switch on
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 317
Annex
Symbol Explanation
Alarm silence
Alarm inactive
Mains power
Key for switching on and off and dimming the workplace illumi-
nation
No alarm limit
318 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Annex
Symbol Explanation
Spontaneous breathing activity by the patient
Risk of crushing
Locked
Unlocked
Inspiration
Labeling on breathing system and breathing system cover
Expiration
Labeling on breathing system and breathing system cover
Breathing bag
Hot surface
Storage temperature
Relative humidity
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 319
Annex
Symbol Explanation
Ambient pressure
Do not reuse
Spare part
LAN connection
USB port
Enter key
Indicates a changed view which has not yet been saved
Identifies the tabs that lead to the page with the language set-
tings.
Read the flow at the center of the float.
Identifies O2 cylinders.
The color code conforms to the locally applicable standard.
Gas inlet
320 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Annex
Symbol Explanation
Display on the total flow tube indicating the cumulative value of
the individual flows
MR unsafe
Do not use this device in the vicinity of MRI scanners.
Weight: Load
Ventilate manually.
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 321
Annex
322 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Annex
Group
Group
View1)
1 2 3
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 323
Annex
Group
Vent. 2
Anesthesia
Tabular trends Overview Display trends of measured values in
Vent. 1 table form.
Vent. 2
Anesthesia
Values Ventilation 1) Display overview of current measured val-
ues.
Gases 1)
Device
Logbook Display the logbook.
Export 2) Export data to a USB mass storage
device.
1) Only during operation, not in Standby mode
2) Only in Standby mode
Group
Group
324 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Password
19 Password
19.1 Configuration password for Atlan A300, A300 XL, A350,
A350 XL Software 1.0n
Cut out from the instructions for use Atlan A300, A300 XL, A350,
A350 XL Software 1.0n
To prevent unauthorized alteration, the start settings for Atlan A300, A300 XL,
A350, A350 XL are protected by the following configuration password:
0000
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 325
This page has been left blank intentionally.
326 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Index
Index
A B
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . 314 Backup manual mode . . . . . . . . . . . . . . . . . . 38
Accessories Backup manual switch . . . . . . . . . . . . . . . . . 38
Connecting . . . . . . . . . . . . . . . . . . . . . . . . 67 Backup mode . . . . . . . . . . . . . . . . . . . . . . . . 38
Accessories and consumptions . . . . . . . . . . 115 Battery
Activation code . . . . . . . . . . . . . . . . . . . . . . 182 Charging. . . . . . . . . . . . . . . . . . . . . . . . . . 57
Advanced Cylinder Support . . . . . . . . . . . . . 139 Breathing bag
Alarm limits Connecting . . . . . . . . . . . . . . . . . . . . . . . . 71
Adjusting . . . . . . . . . . . . . . . . . . . . . . . . . 157 Breathing hoses
Automatic setting . . . . . . . . . . . . . . 117, 169 Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Displaying . . . . . . . . . . . . . . . . . . . . . . . . 111 Breathing system
Setting . . . . . . . . . . . . . . . . . . . . . . . . . . 116 Assembling. . . . . . . . . . . . . . . . . . . . . . . 250
Specifying . . . . . . . . . . . . . . . . . . . . . . . . 168 Checking . . . . . . . . . . . . . . . . . . . . . . . . 141
Alarm volume Inserting . . . . . . . . . . . . . . . . . . . . . . . . . 255
Safety instruction . . . . . . . . . . . . . . . 89, 152 Pressure/Flow characteristics . . . . . . . . 290
Setting . . . . . . . . . . . . . . . . . . 115, 169, 179 Warmer. . . . . . . . . . . . . . . . . . . . . . . . . . 120
Alarms Breaths
Acknowledging alarm messages . . . . . . 155 Synchronizing. . . . . . . . . . . . . . . . . . . . . . 99
Alarm settings when changing ventilation
modes. . . . . . . . . . . . . . . . . . . . . . . . . . . 159 C
Apnea alarms . . . . . . . . . . . . . . . . . . . . . 160
Breath detection . . . . . . . . . . . . . . . . . . . 163 Cardiac bypass mode . . . . . . . . . . . . . . . . . 102
Causes . . . . . . . . . . . . . . . . . . . . . . . . . . 201 CBM mode . . . . . . . . . . . . . . . . . . . . . . . . . 102
CBM mode . . . . . . . . . . . . . . . . . . . . . . . 171 Central gas supply system
Colors . . . . . . . . . . . . . . . . . . . . . . . 154, 201 Connecting . . . . . . . . . . . . . . . . . . . . . . . . 59
Details. . . . . . . . . . . . . . . . . . . . . . . 154, 201 Connectors. . . . . . . . . . . . . . . . . . . . . . . . 26
Displaying . . . . . . . . . . . . . . . . . . . . . . . . 152 Change of patient . . . . . . . . . . . . . . . . . . . . 124
Downgrading alarm messages . . . . . . . . 155 Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
Gas cylinders . . . . . . . . . . . . . . . . . . . . . 171 Circuit plug . . . . . . . . . . . . . . . . . . . . . . . . . . 18
MV alarms . . . . . . . . . . . . . . . . . . . . . . . 160 CLIC adapter
Priorities . . . . . . . . . . . . . . . . . . . . . 154, 201 Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . 253
Quick setup. . . . . . . . . . . . . . . . . . . 158, 170 Clinical users . . . . . . . . . . . . . . . . . . . . . . . . . 11
Remedy . . . . . . . . . . . . . . . . . . . . . . . . . 201 CO2 absorber . . . . . . . . . . . . . . . . . . . . . . . . 75
Setting the alarm behavior . . . . . . . . . . . 171 CO2 alarms
Silencing . . . . . . . . . . . . . . . . . . . . . . . . . 155 Adjusting . . . . . . . . . . . . . . . . . . . . . . . . 117
Viewing the alarm logbook . . . . . . . . . . . 157 CO2 measurement . . . . . . . . . . . . . 69, 77, 308
Anesthetic gas receiving system (AGS) 18, 145 Color scheme . . . . . . . . . . . . . . . . . . . . . . . 166
Connecting . . . . . . . . . . . . . . . . . . . . . . . . 65 Setting . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Replacing . . . . . . . . . . . . . . . . . . . . . . . . 200 Colors for parameters
Anti-interchange security . . . . . . . . . . . . . . . 312 Specifying. . . . . . . . . . . . . . . . . . . . . . . . 167
APL valve. . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Colors of the control elements . . . . . . . . . . . 83
Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . 251 COM 1, COM 2 . . . . . . . . . . . . . . . . . . . 25, 179
Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 Compliance compensation . . . . . . . . . 304, 308
Assisted ventilation with pressure support. . 302 Confirm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
AutoFlow . . . . . . . . . . . . . . . . . . . . . . . . . . . 305 Consumables
Automatic leakage compensation . . . . . . . . 191 Connecting . . . . . . . . . . . . . . . . . . . . . . . . 75
Auxiliary power socket Consumption
Connecting other devices . . . . . . . . . . . . . 58 Resetting . . . . . . . . . . . . . . . . . . . . . . . . 180
Consumptions
Displaying. . . . . . . . . . . . . . . . . . . . . . . . 115
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 327
Index
E H
Econometer . . . . . . . . . . . . . . . . . . . . . . . . . 104 Hose configurations . . . . . . . . . . . . . . . . . . . 69
Electromagnetic compatibility . . . . . . . . . . . 267 Hose set and filters
Emergency O2 delivery . . . . . . . . . . . . . . . . . 29 Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Checking. . . . . . . . . . . . . . . . . . . . . . . . . 139
Using . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 I
Emergency start-up . . . . . . . . . . . . . . . . . . . . 88
Environments of use . . . . . . . . . . . . . . . . . . . 10 Import of configurations. . . . . . . . . . . . . . . . 182
Essential performance features . . . . . . . . . . . 10 Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Expiratory port . . . . . . . . . . . . . . . . . . . . . . . . 18 Infinity ID . . . . . . . . . . . . . . . . . . . . 37, 180, 312
Export of configurations . . . . . . . . . . . . . . . . 182 Activating the function . . . . . . . . . . . . . . 180
External fresh-gas outlet . . . . . . . . . . . . . . . 100 Replacement interval monitoring . . . . . . 312
Replacement intervals . . . . . . . . . . . . . . 180
Support. . . . . . . . . . . . . . . . . . . . . . . . . . . 37
F Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . 262
Factory defaults . . . . . . . . . . . . . . . . . . 165, 180 Inspiratory port . . . . . . . . . . . . . . . . . . . . . . . 18
Resetting to factory defaults . . . . . . . . . . 176 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Failure IT networks . . . . . . . . . . . . . . . . . . . . . . . . . 294
Complete failure . . . . . . . . . . . . . . . . . . . 198
Flow measurement . . . . . . . . . . . . . . . . . 198 L
Fresh-gas delivery . . . . . . . . . . . . . . . . . 194
Gas supply . . . . . . . . . . . . . . . . . . . . . . . 193 Language
O2 sensor . . . . . . . . . . . . . . . . . . . . . . . . 195 Selecting . . . . . . . . . . . . . . . . . . . . . . . . 176
Piston ventilator . . . . . . . . . . . . . . . . . . . 195 Leakage
Screen . . . . . . . . . . . . . . . . . . . . . . . . . . 198 Causes . . . . . . . . . . . . . . . . . . . . . . . . . . 190
Filters Current values . . . . . . . . . . . . . . . . . . . . 126
Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67 Localizing . . . . . . . . . . . . . . . . . . . . . . . . 190
Flow measurement Leakage test . . . . . . . . . . . . . . . . . . . . . . . . 127
Failure. . . . . . . . . . . . . . . . . . . . . . . . . . . 198 Licenses . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
328 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Index
N Q
N2O-free device . . . . . . . . . . . . . . . . . . . . . . 175 Quick setup . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Name
Changing . . . . . . . . . . . . . . . . . . . . . . . . 176 R
Network
Configuring . . . . . . . . . . . . . . . . . . . . . . . 179 Remote service . . . . . . . . . . . . . . . . . . . . . . 263
Network port . . . . . . . . . . . . . . . . . . . . . . . . . 25 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 265
New case Reprocessing
Defining . . . . . . . . . . . . . . . . . . . . . . . . . . 94 Classifications . . . . . . . . . . . . . . . . . . . . 224
Non-rebreathing system Information . . . . . . . . . . . . . . . . . . . . . . . 223
Connecting . . . . . . . . . . . . . . . . . . . . . . . . 73 Reprocessing list . . . . . . . . . . . . . . . . . . . . . 225
Using . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 Reprocessing personnel . . . . . . . . . . . . . . . . 11
Reprocessing procedures . . . . . . . . . . . . . . 225
Validated procedures . . . . . . . . . . . . . . . 225
O Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . 72
O2 flowmeter . . . . . . . . . . . . . . . . . . . . . . 29, 30 Respiratory support . . . . . . . . . . . . . . . . . . . 299
O2 insufflation Rotary knob . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Using . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
O2 sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 S
Failure. . . . . . . . . . . . . . . . . . . . . . . . . . . 195
O2 switch . . . . . . . . . . . . . . . . . . . . . . . . . 29, 43 Safety checks . . . . . . . . . . . . . . . . . . . . . . . 263
O2 waveform scale Safety instruction
Specifying . . . . . . . . . . . . . . . . . . . . . . . . 167 Gas supply . . . . . . . . . . . . . . . . . . . . . . . . 50
O2+ . . . . . . . . . . . . . . . . . . . . . . . 29, 30, 43, 44 Operation . . . . . . . . . . . . . . . . . . . . . . . . . 49
Checking. . . . . . . . . . . . . . . . . . . . . . . . . 143 Risk of fire . . . . . . . . . . . . . . . . . . . . . . . . 48
Using . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 329
Index
330 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Index
W
Water trap . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Checking. . . . . . . . . . . . . . . . . . . . . . . . . 147
Emptying or exchanging . . . . . . . . . . . . . . 77
Waveforms
Configuring . . . . . . . . . . . . . . . . . . . . . . . 167
Overview. . . . . . . . . . . . . . . . . . . . . . . . . 184
Saving view . . . . . . . . . . . . . . . . . . . . . . 167
Specifying the scale . . . . . . . . . . . . . . . . 167
X
xMAC . . . . . . . . . . . . . . . . . . . . . . . . . . 103, 118
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 331
These instructions for use only apply to
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
with the Serial No.:
If no serial number has been filled in by Dräger,
these instructions for use are provided for
general information only and are not intended
for use with any specific medical device.
These instructions for use are provided for
customer information only and will only be
updated or exchanged upon customer request.
Manufacturer
Á9056001|È
9056001 – GA 6500.350 en
© Drägerwerk AG & Co. KGaA
Edition: 2 – 2019-04
(Edition: 1 – 2019-02)
Dräger reserves the right to make modifications
to the device without prior notice.