Sop-Exl 200
Sop-Exl 200
Sop-Exl 200
STANDARD OPERATING
PROCEDURE
(BIOCHEMISTRY)
PREPARED AS PER ISO 15189:2012
ISSUE NO: 04
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 1 of 1
AMENDMENT SHEET
TABLE OF CONTENTS
LAST
CHAPTER SL. REVISION
TITLE REVISION PAGE No.
No. No. NO
MADE
GLUCOSE
1 00 00 02 – 05
UREA
2 00 00 06 – 08
CREATININE
3 00 00 09 - 11
TOTAL BILIRUBIN
4 00 00 12 - 14
DIRECT BILIRUBIN
5 00 00 15 - 17
TOTAL PROTEIN
6 00 00 18 - 20
ALBUMIN
7 00 00 21 - 23
ALKALINE
8 PHOSPHATASE 00 00 24 - 27
ALT (SGPT)
I 9 00 00 28 - 31
AST (SGOT)
10 00 00 32 - 34
CHOLESTEROL
11 00 00 35 - 38
TRIGLYCERIDE
12 00 00 39 - 42
HDL CHOLESTEROL
13 00 00 43 - 46
URIC ACID
14 00 00 47 - 50
PHOSPHOROUS
15 00 00 51 - 54
CALCIUM
16 00 00 55 - 58
GAMMA-GT
17 00 00 59 - 62
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 1 of 2
LAST
CHAPTER SL. REVISION
TITLE REVISION PAGE No.
No. No. NO
MADE
18 LDL CHOLESTEROL 00 00 63 - 66
SODIUM, POTASSIUM & CHLORIDE
19 00 00 67 – 69
MACHINE OPERATION
II 00 00 1–2
PROCEDURE
SAFETY PRECAUTION
V 00 00 1–1
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 2 of 2
Method: Hexokinase.
c) Performance characteristics
The test linearity is 0 – 500 mg/dL.
Analytical Sensitivity: 1 mg/dL.
The analytical sensitivity represents the lowest concentration of GLUC that
can be distinguished from zero.
e) Patient preparation:
QMSP 15
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 2 of 69
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
Hemoglobin (hemolysate) at 1000 mg/dL [0.62 mmol/L] decreases a GLUC result at
50 mg/dL [2.8 mmol/L] by 11%.
Bilirubin (unconjugated) at 60 mg/dL [1026 μmol/L] increases a GLUC result at 50
mg/dL [2.8 mmol/L] by 13%.
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 3 of 69
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 4 of 69
u) Document Reference:
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
a. ADA Guideline
b. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 5 of 69
a) Purpose of examination
The BUN method used on the Dimension® clinical chemistry system is an in
vitro diagnostic test intended for the quantitative determination of urea
nitrogen in human serum, plasma and urine.
e) Patient preparation:
QMSP 15
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 6 of 69
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
The conversion factor of 2.14 must be applied to the BUN calibrator bottle
values (mg/dL) prior to calibrating, to express the results of urea nitrogen as
urea.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 7 of 69
u) Document Reference:
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
a. Kit Literature.
b. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 8 of 69
a) Purpose of examination
The CREA method used on the Dimension® clinical chemistry system is an in
vitro diagnostic test intended for the quantitative determination of creatinine
in human serum, plasma and urine.
c) Performance characteristics
The method has a linearity range of 0 – 20.0 mg/dL.
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter specific
calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
Interfering Substances
Bilirubin (unconjugated) at 20 mg/dL [342 µmol/L] and 40 mg/dL [684
µmol/L] decreases a CREA result of 1.7 mg/dL [150.3 µmol/L] by 12% at
standard sample volume.
Bilirubin (unconjugated) at 60 mg/dL [1026 µmol/L] decreases a CREA result
of 1.7 mg/dL [150.3 µmol/L] by 57% at standard sample volume.
Lipemia (Intralipid®) interference testing at levels 600 mg/dL [6.78 mmol/L]
and greater tripped a test report message; therefore the magnitude of the
interference could not be determined.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 10 of 69
a) Purpose of examination
The TBI method for the Dimension® clinical chemistry system is an in vitro
diagnostic test intended to quantitatively measure total bilirubin in human
serum and plasma. Measurements of total bilirubin are used in the diagnosis
and treatment of liver, hemolytic, hematological, and metabolic disorders,
including hepatitis and gallbladder disease.
c) Performance characteristics
The Linearity is 0.1 – 25.0 mg/dL
e) Patient preparation:
QMSP 15
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter specific
calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
The TBI method was evaluated for interference according to CLSI/NCCLS EP7-A.10 Bias is
the difference in the results between the control sample (without the interferent) and the test
sample (contains the interferent) expressed in mg/dL [µmol/L]. Bias exceeding 10% is
considered interference.
Concentration
Interferent SI Units Total Bilirubin Bias Bias (%)
Levodopa 300 µg/mL 1.1 mg/dL +2.6 mg/dL +236
[1.52 mmol/L] [19 µmol/L] [44 µmol/L]
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 13 of 69
o) Reportable Interval
0-25 mg/dl
p) Instruction for determining quantitative results when a result is not
within the measurement interval
Sample having Total Bilirubin value more than 25 mg/dL should be auto
diluted and re-run.
q) Alert/Critical values, where applicable
Newborn : > 15 mg/dl
r) Laboratory clinical interpretation: Not done
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 15 of 69
a) Purpose of examination
The DBIL method used on the Dimension® clinical chemistry system is an in
vitro diagnostic test intended for the quantitative determination of direct
(conjugated) bilirubin in human serum and plasma.
Method: DPD
c) Performance characteristics
The Linearity is 0.00 – 20.00 mg/dL.
e) Patient preparation:
QMSP 15
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 16 of 69
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
Lipemia (Intralipid®) of 600 mg/dL [6.78 mmol/L] and greater at a DBIL
value of 0.17 mg/dL [2.19 µmol/L] tripped a test report message; therefore
the magnitude of the interference could not be determined.
Hemolysis may depress DBIL results. Follow your laboratory’s procedures for
reporting results when the sample is hemolyzed.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 17 of 69
u) Document Reference:
1. Principle and procedure:
c. Kit Instruction.
d. John Bernard Henry, M.D.
2. Biological reference Interval :
c. Kit Literature.
d. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 18 of 69
a) Purpose of examination
The TP method used on the Dimension ® clinical chemistry system is an in
vitro diagnostic test intended for the quantitative determination of total
protein in human serum and heparinized plasma. Measurements of total
protein are used in the diagnosis and treatment of a variety of diseases
involving the liver, kidney or bone marrow as well as metabolic or nutritional
disorders.
Cupric ion (Cu++) reacts with the peptide linkages of protein in a basic solution.
The blue copper (II) protein complex thus formed is proportional to the total protein concentration in
the sample and is measured using a bichromatic (540, 700 nm) endpoint technique.
c) Performance characteristics
The Linearity is 2.0 – 12.0 g/dL.
e) Patient preparation:
QMSP 15
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 19 of 69
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
Dextran 40 of 1500 mg/dL [375 μmol/L] increases a TP result of 7.0 g/dL [70 g/L]
by 17%.
Immunoglobulin G of 2.5 g/dL [25 g/L] increases a TP result of 7.0 g/dL [70 g/L] by
25%.
Using standard sample size (15 μL):
Hemoglobin (hemolysate) of 500 mg/dL [0.31 mmol/L] (monomer) increases a TP
result of 3.9 g/dL [39 g/L] by 11%.
Bilirubin (unconjugated) of 20 mg/dL [342 μmol/L] decreases a TP result of 3.8 g/dL
[38 g/L] by -11%.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 20 of 69
Lipemia (Intralipid®) of 600 mg/dL [6.78 mmol/L] and above tripped an error flag on
this method, so the magnitude of the interference is not available.
m) Principle of Procedure for calculating result including, relevant, the
measurement uncertainty of measured quality values
Instrument automatically calculates the result. CV% is calculated from
internal QC data. MU% is calculated using the formula CV% X 1.96.
n) Biological Reference intervals
Adults : 6.6-8.3 g/dL
Children (1-18 y) : 5.7-8.0 g/dl
New- born (1-30 d) : 4.1-6.3 g/dl
o) Reportable interval of examination results:
2-12 gm/dl
p) TAT
Regular: 8 hrs
Emergency: 2 hrs
q) Instruction for determining quantitative results when a result is not
within the measurement interval
Sample having Total Protein value more than 12.0 g/dl should be auto
diluted and re-run.
r) Alert/Critical values, where applicable
Not Applicable.
s) Laboratory clinical interpretation: Not done
t) Potential source of variability
Demographic variations and diseased conditions.
Increased levels are observed in dehydration, multiple myeloma, chronic liver
diseases, chronic infections. Decreased values are observed in renal diseases,
malnutrition, albuminuria and terminal liver failure.
u) Document Reference:
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
a. Kit Literature.
b. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 21 of 69
a) Purpose of examination
The ALB method used on the Dimension® clinical chemistry system is an
in vitro diagnostic test intended for the quantitative determination of
albumin in human serum and plasma.
e) Patient preparation:
QMSP 15
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 22 of 69
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
CMPF (3-carboxy-4-methyl-5-propyl-2-furanpropanoic acid) present in sera of
patients with renal failure has been reported to give falsely low albumin values.
Lipemia (Intralipid®) at 1000 mg/dL [11.3 mmol/L] and above tripped a test report
message; therefore the magnitude of the interference could not be determined.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 23 of 69
a) Purpose of examination
The ALP method used on the Dimension® clinical chemistry system is an in
vitro diagnostic test intended for the quantitative determination of total
protein in human serum and plasma.
ALP
p-NPP + AMP p-NP + AMP +PO4
pH 10.35 Mg/Zn
c) Performance characteristics .
The linearity range of the test is from 10-1000 U/L
e) Patient preparation:
QMSP 15
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 25 of 69
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
Results of studies conducted to evaluate the susceptibility of the method
to interference were as follows:
Icterus Interference less than 10% up to 28 mg/dl or 479 µmol/L
bilirubin
Haemolysis Interference less than 10% up to 4.5 g/L haemoglobin
lipemia Interference less than 3% up to 1000 mg/dl Intralipid
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 25 of 69
u) Document Reference:
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
a. Kit Literature.
b. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 27 of 69
a) Purpose of examination
The alanine aminotransferase (ALTI) method is an in vitro diagnostic test for
the quantitative measurement of alanine aminotransferase activity in human
serum or plasma on the Dimension® clinical chemistry system.
Measurements of alanine aminotransferase are used in the diagnosis and
treatment of certain liver diseases and heart diseases.
c) Performance characteristics
The linearity range of the test is 6 – 1000 U/L.
Analytical Sensitivity: 6 U/L
The analytical sensitivity represents the lowest concentration of albumin that
can be distinguished from zero.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 28 of 69
e) Patient preparation:
QMSP 15
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
Bilirubin (unconjugated) at 60 mg/dL [1026 μmol/L] decreases ALTI results
at an activity of 68 U/L[1.14 μkat/L] by -11%.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 29 of 69
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 31 of 69
u) Document Reference:
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
a. Kit Literature.
b. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 32 of 69
a) Purpose of examination
The ASTI method is an in vitro diagnostic test for the quantitative
measurement of AST activity in human serum or plasma on the Dimension ®
clinical chemistry system.
MDH
Oxaloacetate + NADH Malate +NAD
c) Performance characteristics
Test has a linearity range of 6 – 1000 U/L.
Analytical Sensitivity: 6 U/L
The analytical sensitivity represents the lowest concentration of albumin that
can be distinguished from zero.
d) Primary Sample System
Serum.
e) Patient preparation:
QMSP 15
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 33 of 69
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
Results of studies conducted to evaluate the susceptibility of the method to
interference were as follows:
Icterus: Interference less than 10% up to 40 mg/dL or 684 μmol/L bilirubin
Lipemia: Interference less than 5% up to 300 mg/dL Intralipid®
Pyruvate: Interference less than 10% up to 1 mmol/L pyruvate
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 34 of 69
u) Document Reference:
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
a. Kit Literature.
b. Ref: Jacques Wallach, Md
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 35 of 69
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 36 of 69
a) Purpose of examination
The CHOL method used on the Dimension® clinical chemistry system is an in
vitro diagnostic test intended for the quantitative determination of total
cholesterol in human serum and plasma.
CE
Cholesterol esters Cholesterol + Fatty Acids
CO
Cholesterol + O2 Cholest-4-ene-3-one + H2O2
HPO
2 H2O2 +
4 H2O + Oxidized DEA•HCl/AAP
DEA•HCl/AAP
Method: CHOD-POD
c) Performance characteristics
The linearity range of the test is 50 – 600 mg/dL.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 37 of 69
e) Patient preparation:
QMSP 15
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 38 of 69
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
Potassium Oxalate/Sodium Fluoride can decrease cholesterol results an average of
12%.
Li Heparin can depress cholesterol results by an average of 4 mg/dL [0.1 mmol/L] at
a level of 200 mg/dL [5.2 mmol/L].
Bilirubin (conjugated) of 8.1 mg/dL [139 µmol/L] and bilirubin (unconjugated) of 9.4
mg/dL [161 µmol/L] decrease the CHOL result by 15 mg/dL [0.4 mmol/L] at CHOL
concentration of 150 mg/dL [3.9 mmol/L].
Bilirubin (conjugated) of 12.8 mg/dL [219 µmol/L] and bilirubin (unconjugated) of
14.7 mg/dL [251 µmol/L] decrease the CHOL result by 25 mg/dL [0.7 mmol/L] at
CHOL concentration of 250 mg/dL [6.5 mmol/L].
Bilirubin (unconjugated) of 20 mg/dL [342 µmol/L] decreases a CHOL result of 178
mg/dL [4.6 mmol/L] by 15%.
Hemoglobin (hemolysate) of 1000 mg/dL [0.62 mmol/L] (monomer) decreases a
CHOL result of 177 mg/dL [4.6 mmol/L] by 15%.
Lipemia (Intralipid®) at 1000 mg/dL [11.3 mmol/L] tripped a test report message;
therefore the magnitude of the interference could not be determined.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 39 of 69
u) Document Reference:
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
c. Kit Literature.
d. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 40 of 69
a) Purpose of examination
The TGL method used on the Dimension ® clinical chemistry system is an in
vitro diagnostic test intended for the quantitative determination of
triglycerides in human serum and plasma. Measurements obtained are used
in the diagnosis and treatment of patients with diabetes mellitus, nephrosis,
liver obstruction, other diseases involving lipid metabolism, or various
endocrine disorders.
Method: GPO-POD
c) Performance characteristics
The Linearity is 15 - 1000 mg/dl.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 41 of 69
e) Patient preparation:
QMSP 15
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with CHEM II Calibrator is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 42 of 69
l) Interferences
Small amounts of free glycerol may be found in blood samples from healthy
individuals due to natural lipolysis. The concentration of free glycerol may be
increased by stress, disease states or administration of intravenous infusates. Free
glycerol or other polyols may cause a positive interference.
Glycerol-based quality control products should not be used with this method.
Hemoglobin (hemolysate) of 500 mg/dL [0.31 mmol/L] (monomer) will increase a
triglycerides result of 155 mg/dL [1.75 mmol/L] by 12%.
Bilirubin (unconjugated) of 20 mg/dL [342 μmol/L] will increase a triglycerides
result of 156 mg/dL [1.76 mmol/L] by 11%.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 43 of 69
u) Document Reference:
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
a. Kit Literature.
b. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 44 of 69
a) Purpose of examination
The AHDL method for the Dimension® clinical chemistry system is an in vitro
diagnostic test intended to quantitatively measure high density lipoprotein
cholesterol (HDL-C) in human serum and plasma. HDL-C measurements are
used as an aid in the diagnosis of lipid disorders.
Accelerator + CO
HDL, LDL,
Non-reactive LDL, VLDL, Chylomicrons
VLDL,
Chylomicrons DSBmT + Peroxidase
HDL Specific
Detergent
HDL HDL disrupted
Cholesterol esterase
HDL Cholesterol ∆4 Cholestenone + H2O2
Cholesterol oxidase
Peroxidase
H2O2 + DSBmT + 4-AAP Color development
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 45 of 69
e) Patient preparation:
QMSP 15
f) Performance characteristics
The Linearity is 10.0 - 150 mg/dl.
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with AHDL Calibrator is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 46 of 69
l) Interferences
The AHDL method was evaluated for interference from hemolysis, icterus and
lipemia according to CLSI/NCCLS EP7-P. Bias is the difference in the results
between the control sample (without the interferent) and the test sample
(contains the interferent) expressed in percent. Bias exceeding 10% is
considered interference.
Substance Test Concentration AHDL Concentration Biasi
Tested SI Units mg/dL [mmol/L] %
Hemoglobin 1000 mg/dL 29 [0.75] < 10
(hemolysate) [0.62 mmol/L] (monomer)
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 47 of 69
u) Document Reference:
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
a. AHA Guideline
b. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 48 of 69
a) Purpose of examination
The URCA method used on the Dimension ® clinical chemistry system is an in
vitro diagnostic test intended for the quantitative determination of uric acid in
human serum, plasma and urine.
Method: URICASE
c) Performance characteristics
Test has a linearity range of 0 – 20 mg/dl.
e) Patient preparation:
QMSP 15
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 49 of 69
Sample storage:
The sample is stored at 2 – 8ºC for 1 day.
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with CHEM I Calibrator is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
The URCA method (using the standard sample size of 17 μL) was evaluated for
interference from hemolysis, icterus and lipemia according to CLSI/NCCLS EP7-P.
Bias, defined as the difference between the control sample (does not contain
interferent) and the test sample (contains the interferent), is shown in the table
below. Bias exceeding 10% is considered “interference”.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 50 of 69
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 51 of 69
u) Document Reference:
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
a. Kit Literature.
b. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 52 of 69
a) Purpose of examination
The PHOS method is an in vitro diagnostic test for the quantitative measurement of
inorganic phosphorus in serum, plasma and urine on the Dimension® clinical chemistry
system. Measurements of phosphorus (inorganic) are used in the diagnosis and
treatment of bone, parathyroid and renal disease.
Method
Photometric UV test with endpoint determination.
c) Performance characteristics
The Linearity is 0.5 – 9.0 mg/dl.
e) Patient preparation:
QMSP 15
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 53 of 69
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with CHEM II Calibrator is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
The PHOS method was evaluated for interference according to CLSI EP07-
A2.8 Bias is the difference in results between the control sample (without
interferent) and the test sample (contains interferent) expressed in percent.
Bias exceeding 10% is considered interference.
Hemoglobin at 1000 mg/dL [0.62 mmol/L] increases PHOS results by 17% at
a phosphorus concentration of 2.5 mg/dL [0.81 mmol/L].
Lipemia (Intralipid®) at 3000 mg/dL [33.9 mmol/L] increases PHOS results
by 71% and 30% at a phosphorus concentration of 2.5 mg/dL [0.81 mmol/L]
and 6.5 mg/dL [2.1 mmol/L] respectively.
Albumin at 6.0 g/dL [60 g/L] increases PHOS results by 19% at a
concentration of 2.5 mg/dL [0.81 mmol/L].
Hemolyzed samples may give spuriously elevated phosphorus results. Bias
from hemolysis may result from inorganic phosphates produced by the action
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 54 of 69
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 55 of 69
u) Document Reference:
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
a. Kit Literature.
b. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 56 of 69
a) Purpose of examination
The CA method used on the Dimension ® clinical chemistry system is an in
vitro diagnostic test intended for the quantitative determination of calcium in
human serum, plasma and urine.
Method: OCPC
c) Performance characteristics
The method has a linearity range of 5.0 – 15.0 mg/dl.
e) Patient preparation:
QMSP 15
Sample storage:
The sample is stored at 2 – 8ºC for 1 day.
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with CHEM I Calibrator is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
Interference due to magnesium is negligible at magnesium levels normally
encountered in human serum. A maximum positive interference of 0.7 mg/dL
[0.17 mmol/L] occurs at a magnesium level of 7 mg/dL [2.9 mmol/L].
Calcium values may be falsely decreased in the presence of gadolinium-
containing contrast agents such as Omniscan™. Therefore the manufacturer
of this product recommends to avoid drawing samples for serum calcium
determination 24 hours after administration of Omniscan™.
Bilirubin (unconjugated) of 80 mg/dL [1368 μmol/L] decreases calcium at 6.4
mg/dL [1.60 mmol/L] by 11%.
Lipemia (Intralipid®) of 600 mg/dL [6.78 mmol/L] and above tripped a test
report message; therefore the magnitude of the interference could not be
determined.
EDTA when present at 200 mg/dL [2 g/L]e and potassium oxalate when
present at 500 mg/dL [5 g/L]e depresses the CA result to less than the assay
range of the method.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 58 of 69
p) TAT
Regular: 8 hrs
Emergency: 2 hrs
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 59 of 69
u) Document Reference:
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
a. Kit Literature.
b. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 60 of 69
a) Purpose of examination
The GGT method used on the Dimension ® clinical chemistry system is an in
vitro diagnostic test intended for the quantitative determination of γ-
glutamyl transferase activity in human serum and plasma.
e) Patient preparation:
QMSP 15
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 61 of 69
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 62 of 69
l) Interferences
Hemolysis (hemolysate) of 500 mg/dL [0.31 mmol/L]f(monomer) and above
tripped a test report message; therefore the magnitude of the interference
could not be determined.
Bilirubin (unconjugated) of 60 mg/dL [1024 μmol/L] and above tripped a test
report message; therefore the magnitude of the interference could not be
determined.
Triglycerides as (Intralipid®) of 200 mg/dL [2.299 mmol/L] decreases a GGT
at 128 U/L by 16%.
m) Principle of Procedure for calculating result including, relevant, the
measurement uncertainty of measured quality values
Instrument automatically calculates the result & CV%. MU% is calculated as
CV% X 1.96.
n) Biological Reference intervals
Male : < 55 U/L
Female: < 38 U/L
Children: Male Female
1 – 182 days 12 – 122 U/L 15 – 132 U/L
183 – 365 days 1 – 39 U/L 1 – 39 U/L
1 – 12 years 3 – 22 U/L 4 – 22 U/L
13 – 18 years 2 – 42 U/L 4 – 24 U/L
o) Reportable interval of examination results:
0-800 U/L
p) TAT
Regular: 8 hrs
Emergency: 2 hrs
q) Instruction for determining quantitative results when a result is not
within the measurement interval
Sample having GGT value more than 800 U/L should be auto diluted and re-
run.
r) Alert/Critical values, where applicable
Not Applicable
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 64 of 69
a) Purpose of examination
The ALDL method for the Dimension® clinical chemistry system is an in vitro
diagnostic test intended for the quantitative determination of low-density
lipoprotein cholesterol (LDL-C) in human serum and plasma.
LDL-C measurements are used in the diagnosis and treatment of lipid
disorders such as diabetes mellitus, atherosclerosis, and various liver and
renal diseases.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 65 of 69
Method
Direct Homogeneous Enzymatic Colorimetric (Without Precipitation)
c) Performance characteristics
Test has a linearity range of 5 - 300 mg/dl
The sensitivity of the ALDL method is 5 mg/dL and represents the lowest
concentration of LDL-C that can be distinguished from zero.
e) Patient preparation:
QMSP 15
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 66 of 69
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
Bilirubin (unconjugated) of 80 mg/dL [1368 μmol/L] will decrease an ALDL
result of 124 mg/dL
[3.2 mmol/L] by 10%.
Lipemia (Intralipid®) of 3000 mg/dL[33.9 mmol/L] will decrease an ALDL
result of 122 mg/dL [3.2 mmol/L] by 19%.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 67 of 69
u) Reference:
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
a. AHA Guideline
b. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 68 of 69
a. Purpose of examination:
Estimation of Serum Sodium, Potassium and Chloride
c. Performance specifications:
Test is measurable from detection limit:
Sodium: 20 – 200 mEq/L.
Potasium: 1 – 10 mEq/L.
Chloride: 25 – 200 mEq/L.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 69 of 69
SAMPLE STORAGE:
The sample is stored at 2 – 8ºC for 1 day.
i. Calibration procedure:
The dimension are IMT system will routinely perform a one point calibration with each sample
measurement.In addition,the system performs as two point automatic calibration in duplicate
every two hours,if no analysis is in progress.Auto calibration also occurs shortly after turn
on,with the changing of standards A,B or a sensor and when reset.Calibration can be initiated at
any time a sample is not being run.
OPERATION OF INSTRUMENT
Sample is collected, processed as per requirement and then the examination is carried out
following the operating procedure as described in chapter –II.
l. Interferences:
Clots and fibrin interfere.Prolonged tourniquet use and hand exercise when drawing blood falsely
increase potassium level. Na2EDTA as anticoagulant may interfere
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 70 of 69
o. TAT
Regular timing for reporting is 8 hrs.
Emergency – 2 hrs.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 71 of 69
s. Laboratory interpretation:
Sodium-as hypo and hypernatremia,
Potassium-as hypo and hyperkalemia
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 72 of 69
Review the instrument maintenance logs and the System counter screen for any
maintenance that may be overdue. Check the system count for the sample probe
tip, especially if the problem is on a method with a low sample volume.
Check that all temperatures are within range on the Daily maintenance screen.
Check the temperatures with a calibrated thermometer according to the
“Calibration Cuvettes system Temperature” “Calibration range system
Temperature “and “Calibration HM module Temperature” procedures in your
operator’s guide.
C. Shut Down :
Step 1.
After Pump prime step is completed, wait until the system is in STANBY status.
From the operating menu display, press the EXIT key two times.
The system will display the following message:
“Do you really want to exit? (Hit “EXIT” to confirm, “RETURN” to continue)”
Step 2
Press the EXIT key. The system will the following message.
‘If you really want to shut down the system down, type ‘y’.”
Step 3
Press y. The system will make a backup copy of all important data to the disks.
Step 4
Wait until both disk drive lights are stay off, and the message “ Rhug;” appears on
the computer display screen.
Step 5
Open the left front instrument door and turn off the main instrument power switch.
Dimension
C. Shut
Down Go back to main Menu
↓
Go back to maintenance & Start water save procedure
↓
Go to Exit & Shut Down the Windows
↓
Switch Off
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 1 of 4
Receive the sample and confirm Lot No and Expiry Date with reference values specific to
the machine.
Make a note of the storage temperature.
Controls are stored at 2-8oC in specified Biochemistry Refrigerator
Vials are kept at room temperature for a maximum period of 1 hour while running tests.
Record sample validity period after opening or reconstitution.
Prepare aliquots (if applicable) and store in recommended temperature.
Internal Controls are run on a daily basis immediately after the specific machines are switched
on and in perfect running mode. Only authorized personnel in each department run controls.
Values observed are recorded and plotted in L-J charts.
Observe for any violation of Multi QC Deviation Rule and proceed accordingly.
QC Results are reviewed & if any of the parameters run out of 2SD or do not follow any of the MULTI
QC RULES described below, then a QC Level is run for the second time. If the results still do not fall
within the expected range, then the test parameter is calibrated, QC checked & samples run. All the
QC values are checked and filed and reviewed by the Consultant Pathologist.
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 2 of 4
13S: One control observation exceeding the mean ± 3s; primarily sensitive to random error.
22S: Two consecutive control observation exceeding the same mean plus 2 s or mean minus 2 s
limit; sensitive to systematic error.
10X: 10 consecutive control observations falling on one side of the mean (above or below, with no
other requirement on size of the deviations); sensitive to systematic error.
41s: Four consecutive values on either side of mean are within 1SD to 2SD (41S).
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 3 of 4
D. Correlations of results:
Different Tests when results indicate the same pathology are used to confirm test results when
necessary. Both inter and intradepartmental correlation is done. Pathologist reviews the record
Issue No. 04 Issue Date: 20.01.2020 Prepared By: Copy No. Page 4 of 4
CALIBRATION PROCEDURE
EXL 200
Machine is calibrated by the manufacturer as and when required
Individual parameter calibration is done in-house by calibrators having
international traceability.
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 1 of 1
Issue No. 03 Issue Date: 11.09.2014 Prepared By: QM Copy No. Page 1 of 1
Ref: Wallsch. J interpretation of diagnostic test 7th edition page 29-31. Lippincort
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 1 of 1