Urea 501
Urea 501
UREAL
Urea/BUN
Order information
Analyzer(s) on which cobas c pack(s) can be used
04460715 190 Urea/BUN 500 tests System‑ID 07 6303 9 Roche/Hitachi cobas c 311, cobas c 501/502
10759350 190 Calibrator f.a.s. (12 x 3 mL) Code 401
12149435 122 Precinorm U plus (10 x 3 mL) Code 300
12149443 122 Precipath U plus (10 x 3 mL) Code 301
10171743 122 Precinorm U (20 x 5 mL) Code 300
10171735 122 Precinorm U (4 x 5 mL) Code 300
10171778 122 Precipath U (20 x 5 mL) Code 301
10171760 122 Precipath U (4 x 5 mL) Code 301
05117003 190 PreciControl ClinChem Multi 1 (20 x 5 mL) Code 391
05947626 190 PreciControl ClinChem Multi 1 (4 x 5 mL) Code 391
05117216 190 PreciControl ClinChem Multi 2 (20 x 5 mL) Code 392
05947774 190 PreciControl ClinChem Multi 2 (4 x 5 mL) Code 392
04489357 190 Diluent NaCl 9 % (50 mL) System‑ID 07 6869 3
English elevations may also be seen during periods of high protein intake.
Unpredictable levels occur with liver diseases.
System information
For cobas c 311 analyzer: Test principle
UREAL: ACN 418 (serum/plasma) Kinetic test with urease and glutamate dehydrogenase.2,3,4,5
U‑BUN: ACN 421 (serum/plasma) Urea is hydrolyzed by urease to form ammonium and carbonate.
URELU: ACN 417 (urine) Urease
UBUNU: ACN 428 (urine) Urea + 2 H2O 2 NH4+ + CO32-
SUREA: ACN 419 (STAT, reaction time: 4, serum/plasma)
In the second reaction 2‑oxoglutarate reacts with ammonium in the
SUBUN: ACN 427 (STAT, reaction time: 4, serum/plasma) presence of glutamate dehydrogenase (GLDH) and the coenzyme NADH to
SUREU: ACN 420 (STAT, reaction time: 4, urine) produce L‑glutamate. In this reaction two moles of NADH are oxidized to
SBUNU: ACN 429 (STAT, reaction time: 4, urine) NAD+ for each mole of urea hydrolyzed.
For cobas c 501 analyzer: GLDH
UREAL: ACN 418 (serum/plasma/urine) NH4+ + 2‑oxoglutarate + NADH L‑glutamate + NAD+ + H2O
U‑BUN: ACN 421 (serum/plasma/urine)
The rate of decrease in the NADH concentration is directly proportional to
SUREA: ACN 419 (STAT, reaction time: 4, serum/plasma/urine) the urea concentration in the specimen and is measured photometrically.
SUBUN: ACN 427 (STAT, reaction time: 4, serum/plasma/urine)
Reagents - working solutions
For cobas c 502 analyzer:
UREAL: ACN 8418 (serum/plasma) R1 NaCl 9 %
U‑BUN: ACN 8421 (serum/plasma) R2 TRIS buffer: 220 mmol/L, pH 8.6; 2‑oxoglutarate: 73 mmol/L;
URELU: ACN 8417 (urine) NADH: 2.5 mmol/L; ADP: 6.5 mmol/L; urease (jack bean):
UBUNU: ACN 8428 (urine) ≥ 300 µkat/L; GLDH (bovine liver): ≥ 80 µkat/L; preservative;
SUREA: ACN 8419 (STAT, reaction time: 4, serum/plasma) nonreactive stabilizers
SUBUN: ACN 8427 (STAT, reaction time: 4, serum/plasma) R1 is in position C and R2 is in position B.
SUREU: ACN 8420 (STAT, reaction time: 4, urine) Precautions and warnings
SBUNU: ACN 8429 (STAT, reaction time: 4, urine) For in vitro diagnostic use.
Exercise the normal precautions required for handling all laboratory
Intended use reagents.
In vitro test for the quantitative determination of urea/urea nitrogen in Disposal of all waste material should be in accordance with local guidelines.
human serum, plasma and urine on Roche/Hitachi cobas c systems. Safety data sheet available for professional user on request.
Summary1 Reagent handling
Urea is the major end product of protein nitrogen metabolism. It is Ready for use
synthesized by the urea cycle in the liver from ammonia which is produced
by amino acid deamination. Urea is excreted mostly by the kidneys but Storage and stability
minimal amounts are also excreted in sweat and degraded in the intestines
by bacterial action. UREAL
Determination of blood urea nitrogen is the most widely used screening test Shelf life at 2‑8 °C: See expiration date
for renal function. When used in conjunction with serum creatinine on cobas c pack
determinations it can aid in the differential diagnosis of the three types of label.
azotemia: prerenal, renal and postrenal.
On‑board in use and refrigerated on the analyzer: 8 weeks
Elevations in blood urea nitrogen concentration are seen in inadequate
renal perfusion, shock, diminished blood volume (prerenal causes), chronic Diluent NaCl 9 %
nephritis, nephrosclerosis, tubular necrosis, glomerular nephritis (renal
causes) and urinary tract obstruction (postrenal causes). Transient
UREAL
Urea/BUN
UREAL
Urea/BUN
Reagent pipetting Diluent (H2O) In very rare cases, gammopathy, in particular type IgM (Waldenström’s
macroglobulinemia), may cause unreliable results.10
R1 10 µL 90 µL
Urine
R2 38 µL 110 µL
Drugs: No interference was found at therapeutic concentrations using
Sample volumes Sample Sample dilution common drug panels.9
Sample Diluent (NaCl) For diagnostic purposes, the results should always be assessed in
conjunction with the patient’s medical history, clinical examination and other
Normal 2 µL 3 µL 147 µL findings.
Decreased 2 µL 2 µL 178 µL ACTION REQUIRED
Special Wash Programming: The use of special wash steps is mandatory
Increased 2 µL – – when certain test combinations are run together on Roche/Hitachi
Calibration cobas c systems. The latest version of the carry‑over evasion list can be
found with the NaOHD/SMS/Multiclean/SCCS or the
Calibrators S1: H2O NaOHD/SMS/SmpCln1+2/SCCS Method Sheets. For further instructions
refer to the operator’s manual. cobas c 502 analyzer: All special wash
S2: C.f.a.s. programming necessary for avoiding carry‑over is available via the
Calibration mode Linear cobas link, manual input is not required.
Where required, special wash/carry‑over evasion programming must
Calibration frequency 2‑point calibration be implemented prior to reporting results with this test.
• after 4 weeks on board
• after reagent lot change Limits and ranges
• as required following quality control Measuring range
procedures Serum/plasma
Traceability: This method has been standardized against ID/MS. 0.5‑40 mmol/L (3.0‑240 mg/dL urea, 1.4‑112 mg/dL urea nitrogen)
Determine samples having higher concentrations via the rerun function.
Quality control Dilution of samples via the rerun function is a 1:3 dilution. Results from
Serum/plasma samples diluted using the rerun function are automatically multiplied by a
For quality control, use control materials as listed in the "Order information" factor of 3.
section. Urine
In addition, other suitable control material can be used. 1‑2000 mmol/L (6‑12000 mg/dL urea, 2.8‑5600 mg/dL urea nitrogen)
Urine Determine samples having higher concentrations via the rerun function.
Quantitative urine controls are recommended for routine quality control. Dilution of samples via the rerun function is a 1:1.8 dilution. Results from
samples diluted using the rerun function are automatically multiplied by a
The control intervals and limits should be adapted to each laboratory’s factor of 1.8.
individual requirements. Values obtained should fall within the defined
limits. Each laboratory should establish corrective measures to be taken if Determine samples having concentrations lower than the technical limit of
values fall outside the defined limits. 40 mmol/L (240 mg/dL urea and 112 mg/dL urea nitrogen) via the rerun
function. Samples are measured undiluted.
Follow the applicable government regulations and local guidelines for
quality control. Lower limits of measurement
Calculation Lower detection limit of the test
Roche/Hitachi cobas c systems automatically calculate the analyte Serum/plasma
concentration of each sample. 0.5 mmol/L (3.0 mg/dL urea, 1.4 mg/dL urea nitrogen)
The lower detection limit represents the lowest measurable analyte level
Conversion factors: mmol/L urea x 6.006 = mg/dL urea that can be distinguished from zero. It is calculated as the value lying 3
mmol/L urea x 0.06006 = g/L urea standard deviations above that of the lowest standard (standard 1 + 3 SD,
repeatability, n = 21).
mmol/L urea nitrogen x 2.801 = mg/dL urea nitrogen
Urine
mmol/L urea nitrogen x 0.02801 = g/L urea nitrogen 1 mmol/L (6 mg/dL urea, 2.8 mg/dL urea nitrogen)
mg/dL urea x 0.467 = mg/dL urea nitrogen The lower detection limit represents the lowest measurable analyte level
When 24‑hour urine is used as the specimen, multiply the result by the that can be distinguished from zero. It is calculated as the value lying 3
24‑hour volume to obtain values in g or mmol/24 hours. standard deviations above that of the lowest standard (standard 1 + 3 SD,
repeatability, n = 21).
Limitations - interference
Expected values
Criterion: Recovery within ± 10 % of initial value at a urea concentration of
8.3 mmol/L (49.8 mg/dL urea, 23.2 mg/dL urea nitrogen). Urea:
Serum/plasma Serum/plasma11
Icterus:7 No significant interference up to an I index of 60 for conjugated
and unconjugated bilirubin (approximate conjugated and unconjugated Adults 2.76‑8.07 mmol/L (16.6‑48.5 mg/dL)
bilirubin concentration: 1026 µmol/L (60 mg/dL)). Urine
Hemolysis:7 No significant interference up to an H index of 1000 24‑hour urine12 428‑714 mmol/24 h (25.7‑42.9 g/24 h),
(approximate hemoglobin concentration: 621 µmol/L (1000 mg/dL)).
corresponding to
Lipemia (Intralipid):7 No significant interference up to an L index of 1000. 286‑595 mmol/L (1.71‑3.57 g/dL)a
There is poor correlation between the L index (corresponds to turbidity) and
a) Based on average urine output of 1.2‑1.5 L/24 h
triglycerides concentration.
Ammonium ions may cause erroneously elevated results. Urea nitrogen (BUN):
Drugs: No interference was found at therapeutic concentrations using
common drug panels.8,9 Serum/plasma12
Adults (18‑60 years) 2.14‑7.14 mmol/L 6‑20 mg/dL
UREAL
Urea/BUN
UREAL
Urea/BUN
Symbols
Roche Diagnostics uses the following symbols and signs in addition to
those listed in the ISO 15223‑1 standard.
Contents of kit
Volume after reconstitution or mixing
GTIN Global Trade Item Number