Anti-Hbs: Please Note: Both The Vial Labels, and The Additional Labels (If Available)
Anti-Hbs: Please Note: Both The Vial Labels, and The Additional Labels (If Available)
Anti-HBs
Antibody to hepatitis B surface antigen (anti-HBs)
Elecsys 2010
MODULAR ANALYTICS E170
11820524 122 100 cobas e 411
cobas e 601
cobas e 602
Anti-HBs
Antibody to hepatitis B surface antigen (anti-HBs)
Anti-HBs
Antibody to hepatitis B surface antigen (anti-HBs)
Limitations - interference After manual dilution, multiply the result by the dilution factor.
The assay is unaffected by icterus (bilirubin < 513 µmol/L or < 30 mg/dL), After dilution by the analyzers, the MODULAR ANALYTICS E170,
hemolysis (Hb < 0.93 mmol/L or < 1.5 g/dL), lipemia (Intralipid Elecsys 2010 and cobas e software automatically takes the dilution into
< 1500 mg/dL) and biotin (< 123 nmol/L or < 30 ng/mL). account when calculating the sample concentration.
Criterion: Recovery within ± 10 % of initial value. Manual dilution can also be made with negative human serum.
Samples should not be taken from patients receiving therapy with high Note: Antibodies to HBsAg are heterogenous. In some isolated cases, this
biotin doses (i.e. > 5 mg/day) until at least 8 hours following the last biotin may lead to non‑linear dilution behavior.
administration.
Expected values
No interference was observed from rheumatoid factors up to a
concentration of 2100 IU/mL. Interpretation of the results
A high-dose hook effectc) may be observed for samples with anti‑HBs Samples with concentrations < 10 IU/L are considered non‑reactive in the
concentrations > 150000 IU/L. In rare cases a high‑dose hook effect Elecsys Anti‑HBs assay.
< 150000 IU/L cannot be excluded. In case of an unexpected low result, i.e. Samples with concentrations ≥ 10 IU/L are considered reactive in the
after re‑vaccination, the sample should be diluted 1:100 (refer to chapter Elecsys Anti‑HBs assay.
“Dilution”) and tested again. Note: Due to the diversity of the antibodies, the measured anti‑HBs value
In vitro tests were performed on 17 commonly used pharmaceuticals. No can vary depending on the testing procedure used. Results obtained from a
interference with the assay was found. single sample using tests from different manufacturers can therefore differ
In rare cases, interference due to extremely high titers of antibodies to by up to a factor of 4 (or even a factor of 10 in rare cases).8 If there is a
streptavidin and ruthenium can occur. The test contains additives which change in the assay procedure used during the monitoring of vaccination
minimize these effects. protection, then the anti‑HBs values obtained upon changing over to the
new procedure must be confirmed by parallel measurements with both
MODULAR ANALYTICS E170, cobas e 601 and cobas e 602 analyzers: methods.
In case the Elecsys HBsAg II/Anti‑HBs and HBeAg/Anti‑HBe assay
combinations are processed, make sure that these assays are entered in Vaccination strategies in certain risk groups are based on the measured
the “Special Wash” section of the system software and “Step1” (wash anti‑HBs concentration.4 Respective recommendations are given by
execute) is checked. Please refer to the operator's manual. national or regional guidelines.
For diagnostic purposes, the results should always be assessed in Specific performance data
conjunction with the patient’s medical history, clinical examination and other Representative performance data on the analyzers are given below.
findings. Results obtained in individual laboratories may differ.
c) High-dose hook effect: A sample with a true concentration clearly above the measuring range, Precision
but found within the measuring range.
Precision was determined using Elecsys reagents, human sera and
Limits and ranges controls.
Measuring range
2.00-1000 IU/L (defined by the detection limit and the maximum of the Elecsys 2010 and cobas e 411 analyzers
master curve). Values below the detection limit are reported as < 2.00 IU/L. Repeatabilityd) Intermediate precisione)
Values above the measuring range are reported as > 1000 IU/L (or up to
100000 IU/L for 100‑fold diluted samples). Sample Mean SD CV Mean SD CV
Detection limit IU/L IU/L % IU/L IU/L %
≤ 2.0 IU/L HSf), negative 8.7 0.57 6.6 7.73 0.76 9.9
The detection limit represents the lowest measurable analyte level that can HS, weakly positive 15.4 0.52 3.4 12.9 1.42 11.0
be distinguished from zero. It is calculated as the value lying two standard
deviations above that of the lowest standard (master calibrator, standard HS, positive 603 9.30 1.5 605 20.6 3.4
1 + 2 SD, repeatability study, n = 21). A‑HBS Cal1 10.7 0.67 6.3 11.0 1.79 16.2
Sensitivity in diluted samples A‑HBS Cal2 498 14.1 2.8 514 18.4 3.6
Two human sera reactive for anti‑HBs were stepwise diluted with serum
negative for anti‑HBs and assayed in duplicate with the Elecsys Anti‑HBs d) Repeatability = within‑run precision (n = 21)
assay and a commercial comparison test: e) Intermediate precision = between‑run (n = 10)
f) HS = human serum, negative
HS 1 Elecsys Comparison HS 2 Elecsys Comparison
Dilution Anti-HBs test Dilution Anti-HBs test MODULAR ANALYTICS E170, cobas e 601 and cobas e 602 analyzers
IU/L IU/L IU/L IU/L Repeatabilityg) Intermediate precisionh)
1 > 1000 > 150 1 > 1000 > 150 Sample Mean SD CV Mean SD CV
1:100 > 1000 > 150 1:100 > 1000 > 150 IU/L IU/L % IU/L IU/L %
1:500 310 > 150 1:500 261 > 150 HS, negative < 2.00 - - < 2.00 - -
1:1000 161 109 1:1000 136 99.0 HS, borderline 9.42 0.41 4.3 8.52 1.18 13.9
1:2500 69.5 34.3 1:2500 59.0 36.5 HS, positive 916 11.2 1.2 910 37.3 4.1
1:5000 36.5 17.9 1:5000 31.8 17.1 PC A‑HBS1i) < 2.00 - - < 2.00 - -
1:10000 19.8 8.4 1:10000 16.5 9.0 PC A‑HBS2 89.6 0.99 1.1 86 4.87 5.7
1:20000 10.3 4.3 1:20000 8.5 4.3 g) Repeatability = within‑run precision (n = 21)
h) Intermediate precision = within‑laboratory (modified protocol (EP5‑A) of the CLSI (Clinical and
Dilution Laboratory Standards Institute): 6 times daily for 10 days (n = 60))
Samples with anti‑HBs concentrations above the measuring range can be i) PC = PreciControl
diluted with Diluent Universal. The recommended dilution is 1:100 (either Analytical specificity
automatically by the MODULAR ANALYTICS E170, Elecsys 2010 and No cross‑reactions with HAV, HCV, HEV, CMV, EBV, HIV, Rubella,
cobas e analyzers or manually). The concentration of the diluted sample Toxoplasma gondii, and Treponema pallidum were observed.
must be > 10 IU/L.
Anti-HBs
Antibody to hepatitis B surface antigen (anti-HBs)