Operating Manual: Nuclear Gauge Calibration and Verification System
Operating Manual: Nuclear Gauge Calibration and Verification System
OPERATING MANUAL
www.InstroTek.com
1. Table of Contents
****************************************
1. INTRODUCTION 2
3. SOFTWARE INSTALLATION 15
4. SYSTEM OPERATION 17
5. THEORY OF OPERATION 36
6. INDEX 38
7. WARRANTY 40
1. Introduction
We are happy and proud to share this Nuclear Moisture/Density Gauge
verification and calibration technology with you. ValiDator is the first and the
only traceable portable verification and calibration system in the industry.
InstroTek, Inc. would like to thank you for selecting the ValiDator as your field
verification and calibration device. The ValiDator Technology is covered by US
Patent numbers 5,923,726 and 6,050,725 including International patents.
We value your input and would like to hear from you with comments and
suggestions on how we can better serve you with this or other products.
This guide is designed as a reference for the operation of the Verification and
Calibration System. A thorough understanding of the operation and application
of Nuclear Moisture/Density gauges is necessary.
Gauge owners have never had the option before of performing the verification
and re-calibration in the field. The ValiDator from InstroTek, Inc. is designed to
optimize the management of your gauge resources and help reduce
inconsistencies when multiple gauges are on-site. Multiple gauges calibrated
on the same device will reduce conflicts between results.
It is important when you verify or calibrate your gauge that the correct chart
and values are selected. If there are gauges in your fleet that are not supported
on this list, you may establish guidelines for Verification of these gauges.
(Reference Verification Procedures for Gauges Not Supported page 20)
CPN
MC3, MC1 DR: Gauges manufactured before approximately 1996 with one GM
detector. InstroTek currently does not support these gauges. Standard counts
are generally between 10000 & 20000. Please contact InstroTek for more
information on these gauge models or reference Verification Procedures for
Gauges Not Supported page 20.
MC1: InstroTek currently does not support verification of this gauge. Please
contact InstroTek for more information on this gauge model or reference
Verification Procedures for Gauges Not Supported page 20.
MC2: InstroTek currently does not support verification of this gauge. Please
contact InstroTek for more information on this gauge model or refer to manual
section Verification Procedures for Gauges Not Supported page 20.
Humboldt
5001 B, 5001C, 5001P, 5001 EZ and 5001 SD are supported.
InstroTek
Model 3500 Xplorer and 3500 Xplorer 2 are supported.
Troxler
3401 & 3411: These gauges are only different in electronic functionality. Both
these gauges are supported by the ValiDator.
3430 & 3440: These gauges are only different in electronic functionality. Both
these gauges are supported by the ValiDator.
3411 New: (Very few gauges were produced, serial numbers greater than 17000)
3450: Please contact InstroTek for instructions on how to calibrate these gauges.
These gauges, like 4640, operate with two independent measurement systems.
A special optional software application is required for these gauges.
4640: Not supported by the ValiDator I. The ValiDator II available from InstroTek
is specifically designed for verification and calibration of 4640B gauges and 3450
in the Thinlayer mode. Please contact InstroTek for information on how these
gauges can be verified and calibrated.
3430 and 3440 Plus Gauges: Please contact Instrotek for more information on
these gauge models, or reference Verification Procedures on page 18.
Seaman Nuclear
These gauges are not supported.
Verification: The process of using the Validation System to verify the calibration
accuracy of a nuclear density gauging device.
Safety
Always operate in accordance with the nuclear
gauge manufacturer recommendations.
Stand opposite the source rod when operating the
ValiDator Device.
Move three feet (one meter) back when the
source rod is extended into the measurement
position.
Use proper lifting techniques when loading and
unloading the ValiDator.
Be aware of heavy machinery operating in the
area of testing.
The quick start instructions will benefit the new and the experienced operators
using the ValiDation System. The System Operation portion of this manual
contains detailed information about all aspects of the ValiDator system. The
Validation system is designed for use with various nuclear gauges manufactured
by CPN, Humboldt, InstroTek, and Troxler.
Standard Counts for Daily or Rapid Verification: If you are evaluating your
gauges frequently for Quality Assurance, or will not be calibrating on-site, a daily
4 minute standard count may be used for the Standard counts on the
Evaluation Form for Verification. Take one 4-minute count at all other depths.
Standard count should be taken on a solid surface such as soil, asphalt or
concrete base. Do not take standard counts with the standard block on the
ValiDator, second floor, or the tailgate of a truck.
I. Verification Process
A. Align gauge on the ValiDator. For lowering the Source rod, slide the
gauge back until the rod is over the access hole. (See Gauge
Placement page 8)
B. Set count time for four (4) minutes. Note: All moisture verification is
done at the Backscatter position.
C. Backscatter measurements should be taken with the gauge aligned
with the surface stops; two four-min. counts should be averaged for
better precision.
D. Direct Transmission measurements require that the gauge
placement be adjusted to minimize the void between the source
rod and the inside of the ValiDator. After lowering the source rod
move it towards the control panel end of the gauge and to the left
ValiDator Device:
Equipment
Gauge Placement
All Backscatter (Depth 0) and Moisture readings should be taken with the gauge
flush with the stop at the top and on the left as shown below.
Notes/Re-counts:
10
11
2) Model
3) Serial Number
5) Calibration Date
12
Before using the ValiDator, it is recommended that the user read this manual
and understand the operation of this system.
User Guide
ValiDator (Calibration Box)
ValiDator Density Data (Laminated Sheets)
Quick Start Instructions
NIST Certificate
Note: Small scratches should be expected on the ValiDator working surface due
to extensive quality control process utilizing measurements with multiple gauges.
Application Overview
The steps below describe a procedure within the Verification and Calibration
process. Density and Moisture gauge data is required by the VeriFier Software
and should reflect the most recent calibration. If not available, this data may
be obtained from the organization that completed this calibration.
13
Calibration constants may be entered into a gauge (CPN MC3, MC3 Elite, MC1
Elite, Humboldt EZ and SD, InstroTek 3500 and Troxler 3430, 3440, 3430 Plus and
3440 Plus) using the operating keypad. Units that employ EPROM’s (Troxler 3411
and Humboldt 5001 series) may operate using density tables while an EPROM is
programmed and mailed for installation. InstroTek can program EPROM’s for
your gauges. Simply fax or mail the Calibration Constants Report and InstroTek
will program and ship your EPROM by Priority Mail for a nominal charge. You
can also purchase a Prom Programmer Package to burn proms at your facility.
VeriFier allows creation of hexfiles for Troxler 3411B and Humboldt 5001C gauges.
There is an option available on 3401 and 3411 gauges that you can use to allow
direct entry of constants through the keypad. SmartPanel from InstroTek is a
“drop in” replacement for the existing 3401 and 3411 scalars. SmartPanel
upgrades the 3401 and 3411 gauges with many features, including direct entry
of calibration constants through the keypad. For more information on the
InstroTek SmartPanel, please visit our website at www.InstroTek.com.
14
3. Software Installation
Object: Allow the operator to access the software and re-calibrate a nuclear
gauge.
Locate your flash drive included with your ValiDator system. Please
read license agreement before opening software package. This
software should only be loaded on one computer. Access codes will
be needed to complete installation.
NOTE: ValiDator and its’ associated software is licensed to a single user. Use
of the ValiDator and its’ associated software is not allowed for
calibration of gauges from other organizations.
Windows
You may load the new software version directly over the existing version and not
lose your existing data.
15
Follow prompts.
Moving Software
16
4. System Operation
Verification Process
Verification
Calibration
Note: The VeriFier Software used for calibrating your gauge assumes that the
current gauge calibration information is obtained and entered
accurately.
17
Moisture Calibration
*Density Gauge: A gauge qualified for evaluation should meet the requirements
below:
Gauge base should be free of asphalt or debris that would cause the
unit to rock,
Pass the manufacturers recommended statistical stability and drift
tests, and
Pass a standard count at the evaluation site.
No other nuclear gauges should be within thirty (30) feet.
Make sure proper recommended maintenance steps are taken prior to
calibrating gauges.
**Stable Surface:
Be at least three (3) feet (1M) from any vertical structure.
Dry and level,
Have a density of greater than100pcf (1602 Kg/m3).
Parking lots and construction sites are ideal surfaces for testing.
Verification Procedures
Gauge Evaluation
The gauge base should be free of dirt and debris, no other nuclear gauges
near-by, 3 ft. (1 m) from any vertical structure, and on a stable surface.
Pass the manufacturers statistical stability test or ASTM and AASHTO
recommendations in D6938, D7759 and/or AASHTO T310.
Complete and pass a four (4) minute standard count in the area where
verification tests are being conducted.
Refer to the gauge operator’s manual for gauge cleaning, routine maintenance
and evaluation procedures.
18
Select a flat, level and dry surface with a density greater than 100 lb/ft3
(1602 Kg/m3) three (3) feet from any vertical structure. Note: Soil, asphalt or
concrete are good examples of the surface to place the ValiDator for
testing. Parking lots or open fields are recommended locations for testing.
Remove ValiDator from shipping case, place unit on selected area, and
position handle at bottom stop.
Take a Daily or Calibration Standard Count and record on ValiDator
Evaluation Form. Do not take the Standard Count on the ValiDator.
Align the density gauge on the top of the ValiDator using the specific guides
on top of the ValiDator device. (See Gauge Placement page 8) Backscatter
readings should be measured in contact with alignment stop. After lowering
the source rod for direct transmission, position the gauge to the rear of the
ValiDator device against side stops to reduce the air void between the
source and material being read.
Select the specific ValiDator Density data table for the model nuclear gauge
under test.
19
For gauges not supported you can establish a history and determine when each
individual gauge falls outside a ±1% limit. Observe the following procedure to
establish a verification process for gauges not supported.
Example:
Depth A B C D E F
Date 12/7/2017 6/7/2018 12/7/2017 6/7/2018
S/N 7022 S/N 7022 Difference Dif.
PCF E-D
B-A
BS 150.2 150.6 0.4 5.6 5.7 0.1
2 118.5 117.6 -0.9
4 136.2 135.2 -1.0
6 132.1 132.0 -0.1
8 125.0 125.5 0.5
Note: This procedure assumes that the Calibration received is accurate. For re-
calibrating these gauges contact InstroTek.
20
(A) Baseline (B) Test (C) % Diff. (D) (E) Test (F)
Date Depth Density Density B-A Baseline Moisture Diff.
Moisture E-D
21
Complete a single four (4) minute reading at the desired depth(s) under
evaluation, and record the Wet Density (WD) Density Count (DC) Moisture
(M) Moisture Count (MC) on the ValiDator Evaluation Form (Introduction
page 10). Note: For better precision at Depth 0 (BS) and 12, it is
recommended that you take (2)-four minute counts and record the average
on the form.
Use the ValiDator Density Table to evaluate the Wet Density and Moisture of
the depth being tested for the pass/fail range. If the Wet Density and
Moisture fall within this range the gauge passes the verification process.
Note: All gauge moisture measurements are performed from the
backscatter position. The ValiDator moisture value is determined using the BS
measurement stop.
22
To utilize the features of the VeriFier Software, you must enter the current gauge
calibration and ValiDator density data into the software. Follow Step I: Entering
Gauge Calibration Data on page 25 and Step II entering ValiDator Densities on
page 26.
1. Enter the current gauge calibration data into VeriFier software. (Follow
procedure Step I page 25)
2. Enter the ValiDator densities from laminated sheet. (Step II, page 26)
3. Enter counts from ValiDator Evaluation Form taken by the gauge on the
ValiDator.
3.1. Daily Density and Moisture Standard Counts
3.2. Verification Date
3.3. Failure Cutoff, default value is 2pcf (32 Kg/m3)
4. Enter Gauge Counts from Evaluation Form
4.1. Enter Moisture Count at Backscatter
4.2. Enter Density Count at all depths
5. Click on Calculate to display the Calculated Density and Error. This density
should be similar to the WD from the ValiDator Evaluation form.
6. To change measurement units, click on Options and select desired units.
7. To print a Verification report, click on Report.
8. To customize your report click on Options, then Customize.
23
Calibration Procedures
The calibration process utilizes the optional VeriFier Software. Contact InstroTek
to obtain your specific password for operating this program. Note: Only trained,
qualified and experienced users should perform the Calibration and you should
follow the requirements of your nuclear materials license for operating,
maintaining and using your gauge.
Note: If the gauge is not maintained nor has service problems, Calibration is not
going to resolve any service or maintenance problems. Contact your service
center to address any service issues with the gauge.
24
(An example of the Gauge Calibration Sheet may be found on page 11)
1. From the Gauge pull down menu, select New if this is the first time this
gauge data is being entered.
2. Enter Password. (This was obtained from InstroTek on installation. Please
contact InstroTek if lost.)
3. Enter Gauge General Information; Make, Model (Select the model
defined on page 3), Serial Number, Factory Calibration Standard Count
(This is the Calibration Standard Count on the Gauge Calibration Sheet of
the Gauge under test.)
4. Enter Gauge Calibration Data.
25
(Use the laminated ValiDator Density Data supplied with your ValiDator. These
sheets are manufacturer, model, and ValiDator specific. For large organizations,
a multiple ValiDator function is available. Insure that before a calibration is
attempted that the correct ValiDator is selected.)
1. Select ValiDator from the pull down menu, and Select New. Enter the
ValiDator Serial number, and From the Model pull down, select the
appropriate model from the laminated ValiDator Calibration Table.
2. Enter the Moisture content from the laminated sheet and all the depths and
densities found on your sheet. You may enter Pcf or Kg3/m, but not both.
The options menu allows you to change units.
26
(This procedure will calculate new Calibration constants from the counts taken
on the ValiDator. It is extremely important to accurately input this data.)
1. From the Calibration pull down menu, select CaliBrate and enter the
Password at the prompt. Contact InstroTek for a misplaced password.
2. Enter the Calibration Standard Count that you established while collecting
Verification data.
3. Enter the Calibration date. This is the date you collected the data on the
ValiDator for calibration.
4. Enter the Gauge count for the appropriate depth in the first column. This
is the Density Count (DC) from the ValiDator Evaluation form.
5. Enter the Moisture count taken at Backscatter. This is the (MC) from the
evaluation form.
6. After the Gauge count data is entered, click on the 3 or 5 block calibrate
button depending on the number of blocks used in the previous
calibration. Generally, CPN, InstroTek and Troxler gauges use a three (3)
block calibration, while Humboldt gauges are five (5) block calibrated.
27
Calibration Date
This is the date that the calibration was done.
Gauge Count
Density Counts from each depth taken on
the ValiDator. Depths 0 and 12 are the
average of (2) four-minute readings.
28
ASTM DD7759 requires the reporting of verification and calibration uncertainties. The
software calculates and reports uncertainty budgets for each gauge verified and
calibrated.
29
30
31
MC3: Press the Step and Calib. key on the front panel. Select Wet Coefficients,
using the select key to change the depth. The A, B, C and A, B coefficients may
be entered by pressing the Enter key.
3430: Press the special key and use the arrow keys to select Calib. Constants.
Enter the code 4678 using the arrow keys. Enter the E and F constants (moisture
constants), select depth and enter the A, B and C constants. Use the arrow keys
to choose the depth.
3440: Press Shift-Special 1, then 9 and enter the code 4688. Some gauges
require a secondary code, which is 5936. Select Calibration Constants from
menu and follow the screen prompts and enter the E & F moisture constants and
select the depths for entering the A, B and C constants.
3430 and 3440 Plus: Contact Troxler to purchase codes in order to enter
constants into these gauges. Troxler charges their customers an additional fee
for performing this step.
3500, 3500 Xplorer 2 and SmartPanel: Press menu, use up and down keys, select
calibration constants, Enter code (4659) and press enter. Then enter constants
for moisture (E and F) and A, B and C constants for all depths.
Humboldt EZ:
32
Note: JPI is a small black plastic jumper located at the upper right-hand side
between keyboard and CPU board as shown below.
JPI
33
(Static charges may damage silicone chips. Ground yourself while installing EPROM.)
1. Loosen the front panel for the gauge requiring EPROM installation.
2. Locate the EPROM on the circuit board of the front panel. (IC that has a
sticker with the gauge serial number and last calibration date.)
3. Make a note to the position of the notch in the top of the chip.
4. Remove the old chip with a small flat blade screwdriver or chip removal tool.
5. Reverse this process for installation.
At any point that the Calibration is in error you may revert back to the old calibration by
pressing Ctrl+E.
Additional Features:
Select Gauge from the pull down menu to access the functions described
below.
Pick is used to choose another gauge from your database.
Modify is used to change or update the information for a specific gauge in the
database.
New is for entering new gauge data
Density Tables
Allow for the manual calculation of wet density using the gauge count ratio.
34
Calculate Density
Gauge Management
Management: This option will allow you to manage your gauges more
effectively. The software will report the last verification date, the last calibration
date and the next leak test date. The data is sorted by the gauge’s Next Leak
Test date, and highlighted if the date is over six months for leak test and one
year for calibration.
Leak Test: The software will allow you to enter the last leak test date for each
gauge in the database. The software will keep up with this information and will
alert you of the next leak test date (six months) under the Management report
option. An overdue leak test will be highlighted in red.
Expected Standard Count: This calculation is based on the decay of the CS-137
gamma source. The limit on these standard counts is based on the gauge
manufacturer’s limit of ± 1% of the actual standard count. The software will
automatically access the most recent calibration date and standard counts for
the calculations. Note: Gauge counts may change due to many different
factors. Radioactive decay is the only variation that could be estimated
accurately and does not necessarily correct for all deviations in the gauge.
Notes: The notes function will allow you to document repairs and items used to
identify service trends or any gauge related information.
35
5. Theory of Operation
Verification Theory
Nuclear density gauges utilize two systems that are used to calculate an in-
place density and moisture content of construction materials. The density
system generally uses Cesium 137 and detector tubes which output analog
signals that are converted to counts by a micro-processor inside the gauge.
Moisture systems use a neutron source of Americium-241: beryllium and a
detector tube that also generate signals that are converted to counts. The
calibration process establishes a relationship between these raw counts and
known densities and moisture content references. In the construction density
range, the higher the density, the lower the counts. For moisture, the counts
increase with increasing moisture content.
Density Calibration
The final stage of the gauge manufacturing process is the calibration. The
density calibration method used by most manufacturers utilizes an exponential
equation that models the relationship between the known densities and the
counts. CPN, Humboldt and Troxler use an equation such as:
CR A exp BD C
36
Where A, B and C are gauge parameters, CR is the count ratio and D is the
material density. The A, B and C values are commonly known as calibration
constants. In the field when a count is collected on the test material, the
resulting wet density displayed on the screen is calculated by:
1 A
D ln
B CR C
Moisture Calibration
To calculate the parameters E (intercept) and F (slope), MCR is the count ratio,
which is the ratio of the measured counts and the daily reference standard
count. The daily reference standard count is performed on a high-density
polyethylene block provided with each gauge. M is the test material moisture
content. The E and F parameters are placed in the gauge memory. They are
then used in the field with the count ratio obtained on the test material to
calculate the moisture content from the equation.
MCR E
M
F
37
6. Index
Accessories ....................................................... 13
C
Calibration......................................................... 17
Humboldt EZ ...................................................... 32
I
38
Precautions ..........................................................7
Readings ...............................................................8
S
39
7. Warranty
InstroTek will replace, free of charge, any part found to be defective within the
warranty period.
40
Contact Information
41