C Runtime Error When Microlog Analyzer Transfer Window Is Opened

The document provides a solution to resolve a Microsoft C++ Runtime Error that appears when transferring data from a SKF Microlog Analyzer connected to a PC. The solution is to initialize both the Memory Card and Internal Memory on the Microlog Analyzer which will delete any saved data, but allow the user to then access the Transfer window without the error.

Uploaded by

Mohammed Abdalla
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
26 views1 page

C Runtime Error When Microlog Analyzer Transfer Window Is Opened

The document provides a solution to resolve a Microsoft C++ Runtime Error that appears when transferring data from a SKF Microlog Analyzer connected to a PC. The solution is to initialize both the Memory Card and Internal Memory on the Microlog Analyzer which will delete any saved data, but allow the user to then access the Transfer window without the error.

Uploaded by

Mohammed Abdalla
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

`

Knowledge Base Article


Product Group: Software
Product: CMSW7300 - @ptitude Analyst for Microlog Analyzer
Version: 8.0

Abstract
When a SKF Microlog Analyzer is connected to a PC and the Transfer >
Microlog Analyzer menu in SKF @ptitude Analyst is opened, a
Microsoft C++ Runtime Error is immediately displayed. This article
provides a solution to the error.

Overview
To resolve the C++ Runtime Error that appears as soon as the Transfer
window is opened, initialize both the Memory Card and Internal
Memory on the Microlog Analyzer.

NOTE: This procedure will delete any data saved on the


Microlog.

When that is complete, the user should be able to access the Transfer
window without the error appearing.

Clear and Config the Microlog Analyzer from within the Transfer
window.

----

For further assistance, please contact the Technical Support Group by


phone at 1-800-523-7514 option 8, or by e-mail at [email protected].

SKF Reliability Systems 3833 Rev A


5271 Viewridge Court * San Diego, California, 92123 USA Page 1 of 1