3T MRI Systems, Issue 2. Comparative Report On Five MRI Systems.
3T MRI Systems, Issue 2. Comparative Report On Five MRI Systems.
net/publication/350054257
CITATIONS READS
0 340
4 authors, including:
All content following this page was uploaded by Ioannis Delakis on 14 March 2021.
August 2005
αβχ
MHRA
MHRAlogo
logo
αβχ
Report 05048
3T MRI systems
Issue 2
Department of Health
logo
best choice
DH log o
best practice
Visit www.mhra.gov.uk to download publications
Department of Health
logo
About evaluation reports
The Device Evaluation Service (DES) provides independent and objective
evaluations of medical devices available on the UK market. Specialist centres,
mainly in NHS Trusts, do the evaluations under contract to the Medicines and
Healthcare products Regulatory Agency (MHRA). Results are available on the
MHRA website (www.mhra.gov.uk). Our evaluations are usually of products
supplied by the manufacturer. We expect these products to be representative
of those on the market but cannot guarantee this. Prospective purchasers
should satisfy themselves about any modifications that might have been made
after our evaluation.
All evaluation reports published after 2002 are available in full colour to
download from the website:
David Price,
Caroline Renaud, Ioannis Delakis, James Williams
MagNET
Department of Bioengineering
Imperial College London
Exhibition Road
London
SW7 2AZ
Information on reproduction outside these terms can be found on the HMSO website
(www.hmso.gov.uk) or email: [email protected].
Summary ....................................................................... 1
Purpose of this report ............................................... 1
Comparative specifications....................................... 1
Technical evaluation................................................. 1
Introduction................................................................... 2
Structure................................................................... 2
Evaluated systems ................................................... 3
Comparative specifications ......................................... 4
Magnet system ......................................................... 4
RF system ................................................................ 7
Gradient system ....................................................... 8
Sequence information............................................... 9
Computer system ................................................... 18
Radio-frequency (RF) coils..................................... 20
Technical evaluation................................................... 29
Introduction to evaluation ....................................... 29
Quadrature head coil evaluation............................. 30
In-built body coil evaluation .................................... 44
Multi-channel head coil evaluation.......................... 48
Multi-channel body coil evaluation.......................... 52
2D and 3D imaging speed ...................................... 55
Acoustic noise ........................................................ 59
List of safety information......................................... 61
Acknowledgements .................................................... 63
References .................................................................. 64
Appendix ..................................................................... 65
Manufacturers’ comments ...................................... 65
2D imaging speed: sequence parameters .............. 66
3D imaging speed: sequence parameters .............. 68
Report 05048: 3.0 T MRI systems - issue 2
Comparative specifications
The comparative specification is presented as a side-by-side summary
comparison of the specifications of each scanner and related equipment. It is
grouped into a series of sub-sections relating to different aspects of the
scanner, such as magnet, gradients, coils etc. Manufacturers supplied the
specification data in response to a template issued by MagNET. The data have
not been verified by MagNET.
Technical evaluation
The technical evaluation section presents a detailed analysis of the head and
body coils available from each manufacturer. Parameters such as signal-to-
noise ratio, uniformity, resolution are included in the evaluation. Furthermore
dedicated sections on parallel imaging and acoustic noise have also been
incorporated. The data are published after consultation with manufacturers.
Their comments are included in the appendix.
Report 05048: 3.0 T MRI systems - issue 2
1
Introduction
MagNET comment
This report does not attempt to explain MRI. Readers who are unfamiliar with
this modality may have difficulty understanding the results presented in this
report. In this case, advice should be sought from a suitably qualified MRI
specialist.
Structure
The main body of the report is divided into two sections, the first presents
system specifications, system information and safety evaluation, and the
second gives technical evaluation results. The results contained in this report
are published after consultation with the manufacturer. Their comments are
included in the Appendix.
2
Evaluated systems
This report presents system specification and technical evaluation data of the
MR systems listed below and shown in Figure 1.
GE HD Philips Achieva 3T
Magnet system
Table 2. Magnet specification
Siemens
GE Philips Siemens Siemens
Trio,A Tim
HD Achieva Allegra Trio
system
RF frequency
127.7 127.7 123.2 123.2 123.2
MHz
Shielding Active Active Active Active Active
Homogeneity
(V-RMS)
0.35 0.6 N/A 0.1 0.1
40 cm DSV
ppm
Number of
measurement 12 12 18 24 24
planes
Number of
points per 12 12 24 20 20
plane
Field stability
<0.1 <0.1 <0.1 <0.1 <0.1
ppm/hr
Cooling
Liquid helium Liquid helium Liquid helium Liquid helium Liquid helium
system
Boil-off rate
<0.03 0.15 <0.1 <0.12 <0.12
l/hr
Report 05048: 3.0 T MRI systems - issue 2
4
Comparative specifications
5
Comparative specifications
6
Comparative specifications
RF system
Table 7. RF system specification
GE Philips Siemens Siemens Siemens
HD Achieva Allegra Trio Trio, A Tim system
Name/type/ HD Freewave iPAT iPAT iPAT plus Tim Tim Tim
version of the plus [102x8] [102x18] [102x32]
system
Number of 4/ 8 (std)/ 4 8 8 8 18 32
independent RF 16 16 (opt)
receiver
channels
(standard/
optional)
Bandwidth of 0.5 3 1 1 1 1 1 1
each
independent RF
receiver channel
(MHz)
Number of 1 1 1 1 1 1 1 1
Analog-to-Digital
Converters for
each
independent RF
channel
Sampling 1 80 10 10 10 10 10 10
frequency of Direct digital
each Analog-to- sampling with
Digital Converter no
(MHz) intermediate
frequency
modulation
Report 05048: 3.0 T MRI systems - issue 2
7
Comparative specifications
Gradient system
Table 8. Gradient system specification (x, y, z as defined in Figure 2)
GE Philips Siemens Siemens Siemens
HD Achieva Allegra Trio Trio, A Tim
system
Zoom Whole
mode body
mode
Shielding Active Active Active Active Active Active
Single axis maximum
amplitude mT/m
x (horizontal) 50 23 80 40 40 40
y (vertical) 50 23 80 40 40 40
z (along the bore axis) 50 23 80 40 40 45
Single axis slew rate
mT/m/ms
x (horizontal) 150 80 200 400 200 200
y (vertical) 150 80 200 400 200 200
z (along the bore axis) 150 80 200 400 200 200
Duty cycle 100 100 100 100 100 100
at max amplitude %
Z
X
8
Comparative specifications
Sequence information
MagNET comment
system
Minimum FOV mm 10 5 5 5 5
Maximum imaging 1024 x 1024 1024 x 1024 1024x1024 1024x1024 1024x1024
matrix
Minimum 2D slice 0.5 0.5 0.1 0.1 0.1
thickness mm
Minimum 3D slice 0.1 0.05 0.05 0.05 0.05
thickness mm
9
Comparative specifications
10
Comparative specifications
Paediatric suite (standard): Head and spine protocols divided according to age
groups. Cardiac morphology protocols according to age groups, optimised for a
small FoV and faster heart rates in congenital heart diseases (CHD). Imaging
protocols for ventricular function as well as valvular and septal defects. CE-MRA
as an adjuvant in the assessment of CHD and vasculature
11
Comparative specifications
12
Comparative specifications
13
Comparative specifications
14
Comparative specifications
15
Comparative specifications
16
Comparative specifications
Others: Real-time interactive imaging (provides the imaging techniques and user
interface elements for fluoroscopic MR imaging.)
Siemens Others: 3D Constructive Interference in the Steady State (CISS), 3D Dual Echo
Allegra Steady State (DESS) with/without water excitation, Online visualisation of
contrast enhancement as centre of k-space is filled as quickly as possible for
exact timing of CE MRA, Fast switch from 2D to 3D, Centric, elliptical k-space
filling
Siemens Others: 3D Constructive Interference in the Steady State (CISS), 3D Dual Echo
Trio Steady State (DESS) with/without water excitation, Online visualisation of
contrast enhancement as centre of k-space is filled as quickly as possible for
exact timing of CE MRA, Fast switch from 2D to 3D, Centric, elliptical k-space
filling
Siemens CISS & DESS (Double Echo Steady State): T2/T1-weighted. Fluid-cartilage
Trio, A differentiation in orthopedic imaging 3D CISS (Constructive Interference in
system Steady State).
Flow Quantification: quantitative flow determination studies: measuring
blood/CSF flow non-invasively Requires Physiological Measurement Unit (PMU)
option.
RetroGated Flow: Dynamic representation of temporally changing flow
Interactive Realtime: Real-time cardiac examinations. Real-time interactive
slice positioning and slice angulation. 3D Magellan SpaceMouse included
TGSE (Turbo Gradient Spin Echo): Hybrid Turbo Spin Echo/Gradient echo
used primarily for T2- weighted imaging
Report 05048: 3.0 T MRI systems - issue 2
17
Comparative specifications
Computer system
Table 17. Main computer system - architecture
GE Philips Siemens Siemens Siemens
HD Achieva Allegra Trio Trio, A Tim
system
Type Intel Xeon dual Dual Pentium 2 x Pentium Pentium Pentium IV
processors with IV Xeon IV/Intel Xeon IV/Intel Xeon Xeon
Hyperthreading processors
Operating Linux Windows XP Windows XP Windows XP Windows XP
system Professional, Professional,
sygno sygno
speaking speaking
CPU speed Dual 2.66GHz > 2x2.8 2x3 2x3.06 2x3.6
GHz >900
(SPECfp2000)
>900
(SPECint2000)
Word length 32 32 32 32 32
bit
Memory size 2 2 2 2 2
GB
Hard disk:
Software GB 1 x 32 36 36 36 (+ 36 36 (+ 36
database) database)
Images GB 2 x 32 36 73 73 73
†
Image > 490 000 250 000 110 000 110 000 110 000†
capacity
2562
images*
Archive
drive:
Drive MOD, DVD RW DVD RW or CDR, MOD CDR, MOD CDR
MOD (read only) (read only)
Report 05048: 3.0 T MRI systems - issue 2
18
Comparative specifications
19
Comparative specifications
Type* R PA R PA R PA L R PA L R PA R PA R PA
Q/L Q/L Q/L
Dimensions 24x24 23x22 Not supplied Not supplied 30 x 30 x 33 (l x w x h)
cm (diaxl) (dia x l)
# output 8 8 4 8 4 CP 8 CP 12
channels
# elements 0 0 0 0 12 12 12
for QD/CP (=4 CP) (=8 CP) (=12)
# elements 8 8 4 8 See See See
for LP above above above
*R=receive, T/R=transmit/receive, PA=phased array, Q=quadrature, L=linear, w=width, h=height, d=depth,
l=length, dia=diameter, circ=circumference
20
Comparative specifications
21
Comparative specifications
22
Comparative specifications
23
Comparative specifications
24
Comparative specifications
# output 8 6 - 8 12 CP 24 CP
channels
# elements 0 0 - 0 36 36
for QD/CP (=12CP) (=24CP)
# elements 8 6 - 8 36 36
for LP
*R=receive, T/R=transmit/receive, PA=phased array, Q=quadrature, L=linear, w=width, h=height, d=depth,
l=length, dia=diameter, circ=circumference
# elements - 4 0 4 4
for LP
*R=receive, T/R=transmit/receive, PA=phased array, Q=quadrature, L=linear, w=width, h=height, d=depth,
l=length, dia=diameter, circ=circumference
25
Comparative specifications
Type* R PA R PA - R PA L R PA L
Dimensions Not supplied 2 sizes - Not supplied Not supplied
cm
# output 3 4 - 4 4
channels
# elements 0 0 - 0 0
for QD/CP
# elements 3 4 - 4 4
for LP
*R=receive, T/R=transmit/receive, PA=phased array, Q=quadrature, L=linear, w=width, h=height, d=depth,
l=length, dia=diameter, circ=circumference
26
Comparative specifications
27
Comparative specifications
Dimensions - 14 (dia) - - -
cm
# output - 1 - - -
channels
# elements - 0 - - -
for QD/CP
# elements - 1 - - -
for LP
*R=receive, T/R=transmit/receive, PA=phased array, Q=quadrature, L=linear, w=width, h=height, d=depth,
l=length, dia=diameter, circ=circumference
28
Technical evaluation
Introduction to evaluation
The technical performance of the 3.0 T systems included in this evaluation is
provided in this section of the report. Table 28 provides information on these
systems and Table 29 shows MagNET's standard quality assurance protocol.
29
Technical evaluation
Analysis
The SNR was calculated using a subtraction method; two identical scans were
acquired and a difference image obtained. The mean signal was measured from
five regions of interest within the test object area in one of the acquired images
and the noise was measured from the standard deviation from these regions in
the difference image (Lerski 1998). The values obtained for SNR were
normalised for voxel size (including measured slice width), scan time, and
sampling bandwidth (no normalisation for coil loading).
Interpretation of results
The image SNR value obtained on a system is influenced by many factors. For
example, system factors such as the main magnetic field strength Bo and the
design of the radiofrequency receive and transmit systems can affect the SNR.
Other factors are the choice of sequence and imaging parameters.
The SNR could not be normalised for loading since some manufacturers were
unable to provide Q measurements. Readers should take this into account
when reading the report.
Table 30. Experimental conditions for quadrature head coil SNR test
Report 05048: 3.0 T MRI systems - issue 2
30
Technical evaluation
PH-ACH
SI-TRI
System
SI-ALL
GE-EXC
31
Technical evaluation
Analysis
The average of several intensity profiles is calculated in all three gradient
directions. The fractional uniformity is calculated for each gradient direction from
the fraction of the profile that lies within 10% of the mean value of a central ROI
(Lerski 1998). The optimum value is unity, indicating 100% of the signal is
considered uniform over the measured distance.
Interpretation of results
A key factor affecting uniformity is the design of the radiofrequency coil, in this
case, the head coil. Most head coils are optimised to give a high value for the
transverse orientation. Poor uniformity often occurs in the bore direction
(z-gradient direction). This can be seen in the vertical values for the sagittal and
coronal plane. These values often lower the average for the three planes.
The fractional uniformity measurements in the two directions for each plane are
presented in Table 32. The comparison made in Graph 2 shows the mean
fractional uniformity for 3T systems tested by MagNET.
Report 05048: 3.0 T MRI systems - issue 2
32
Technical evaluation
SI-TRI
SI-ALL
System
PHI-ACH
GE-EXC
33
Technical evaluation
Analysis
The resolution is evaluated by calculating the 50% point on the modulation
transfer function (MTF) plot (Lerski 1998). This frequency is converted into pixel
resolution.
Interpretation of results
The resolution measurements are presented in Table 33. The measured
resolution should be equal to the nominal pixel dimension of 0.98mm.
Graph 3 shows a comparison of the average pixel dimension for both the PE
and FE directions (256 x 256 matrix) for 3 T systems tested by MagNET.
MagNET comment
34
Technical evaluation
Table 33. Pixel dimension measurements (mm) for 256 x 256 matrix
(nominal 0.98 mm)
Mean Mean
Mean ± SD
(in PE direction) (in FE direction)
GE-EXC 1.03 1.03 1.03±0.05
PH-ACH 1.05 1.06 1.06±0.01
SI-ALL 1.06 1.07 1.06±0.04
SI-TRI 1.03 1.08 1.06±0.03
Graph 3: Mean pixel dimension - 256 x 256 matrix (mean of three planes)* †
GE-EXC
PH-ACH
System
SI-ALL
SI-TRI
35
Technical evaluation
Analysis
The geometric linearity results are obtained from a set of horizontal and vertical
distance measurements in the acquired images (Lerski 1998).
Interpretation of results
These measurements are converted from pixels to millimetres and compared to
the actual separation distance of 120mm.
Table 34 shows a comparison of the mean geometric linearity for the three
planes.
MagNET comment
36
Technical evaluation
GE-EXC
PH-ACH
System
SI-ALL
SI-TRI
*The optimum value is 0 ± 1 mm. Systems are ranked in order of the absolute deviation from optimum.
Report 05048: 3.0 T MRI systems - issue 2
37
Technical evaluation
Analysis
The Coefficient of Variation (CV) indicates the degree of variation of the
distance measurements from one another. The coefficient of variation is defined
as (Lerski 1998):
standard deviation
CV = x 100%
mean
Interpretation of results
The lower the coefficient of variation, the lower the in-plane distortion and the
better the performance.
Graph 5 shows a comparison of the mean geometric distortion for the three
gradient directions.
Report 05048: 3.0 T MRI systems - issue 2
38
Technical evaluation
PH-ACH
GE-EXC
System
SI-ALL
SI-TRI
39
Technical evaluation
Analysis
The slice width is measured from the full width at half maximum (FWHM) of the
slice intensity profile (Lerski 1998).
Interpretation of results
The slice profiles should have minimum side lobes, no ringing, and no central
drop-out. The measured slice widths for the three imaging planes presented in
Table 36 should lie within 10% of the nominal slice width.
40
Technical evaluation
GE-EXC
PH-ACH
System
SI-TRI
SI-ALL
* The optimum value is 0 ± 0.3 mm. Systems are ranked in order of the absolute deviation from optimum.
GE-EXC
PH-ACH
System
SI-TRI
SI-ALL
Report 05048: 3.0 T MRI systems - issue 2
-1.2 -1 -0.8 -0.6 -0.4 -0.2 0 0.2 0.4 0.6 0.8 1 1.2
Error in measured slice width (mm)
* The optimum value is 0 ± 0.5 mm. Systems are ranked in order of the absolute deviation from optimum.
41
Technical evaluation
Analysis
Ghosting is calculated as the ratio of the maximum image ghost, minus the
background noise, to the image signal (Lerski 1998).
Interpretation of results
The lower the ghosting, the better the performance. Table 37 presents the
ghosting results for each echo.
Graph 8 and Graph 9 present results for these 3T systems for 1 and 2 NSA
respectively.
NSA = 1
NSA = 2
42
Technical evaluation
SI-TRI
2nd echo
3rd echo
4th echo
GE-EXC
PH-ACH
SI-TRI
2nd echo
3rd echo
4th echo
PH-ACH
GE-EXC
Report 05048: 3.0 T MRI systems - issue 2
43
Technical evaluation
Analysis
The SNR was calculated using a subtraction method; two identical scans were
acquired and a difference image obtained. The mean signal was measured from
five regions of interest within the test object area in one of the acquired images
and the noise was measured from the standard deviation from these regions in
the difference image (Lerski 1998). The values obtained for SNR were
normalised for voxel size (including measured slice width), scan time and
sampling bandwidth (no normalisation for coil loading).
Interpretation of results
The image SNR value obtained on a system is influenced by many factors.
Example system factors are the main magnetic field strength B0 and the design
of the radiofrequency receive and transmit systems. Other factors are the
choice of sequence and imaging parameters.
The SNR could not be normalised for loading since some manufacturers were
unable to provide Q measurements. Readers should take this into account
when reading the report.
The image normalised SNR results for the in-built body coil are presented in
Table 39. The comparison made in Graph 10 shows the image SNR normalised
for voxel size, scan time and sampling bandwidth for 3T systems.
Table 38. Experimental conditions for in-built body coil SNR test
Report 05048: 3.0 T MRI systems - issue 2
44
Technical evaluation
Table 39. Signal to noise ratio for the in-built body coil
Transverse Sagittal Coronal Mean
GE-EXC 437 449 499 462
PH-ACH 933 910 900 914
SI-TRI 417 427 414 419
Graph 10. Comparison of normalised SNR for in-built body coils (mean of
three planes)*
PH-ACH
System
GE-EXC
SI-TRI
45
Technical evaluation
Analysis
The average of several intensity profiles is calculated in all three gradient
directions. The fractional uniformity is calculated for each gradient direction from
the fraction of the profile that lies within 10% of the mean value of a central ROI
(Lerski 1998). The optimum value is unity, indicating 100% of the signal is
considered uniform over the measured distance.
Interpretation of results
A key factor affecting uniformity is the design of the radiofrequency coil. Most
body coils are optimised to give a high value for the transverse orientation. Poor
uniformity often occurs in the bore direction (z-direction for this system). This
can be seen in the vertical values for the sagittal and coronal planes. These
values often lower the average for the three planes.
The fractional uniformity measurements in the two directions for each plane are
presented in Table 40. The comparison made in Graph 11 shows the mean
fractional uniformity for 3T systems tested by MagNET.
Report 05048: 3.0 T MRI systems - issue 2
46
Technical evaluation
Table 40. Fractional uniformity for the in-built body coil (unfiltered)
x-direction y-direction z-direction Mean±SD
GE-EXC 0.97 0.88 0.75 0.87±0.10
PH-ACH 1.00 1.00 0.85 0.95±0.08
SI-TRI 0.95 1.00 1.00 0.96±0.04
SI-TRI
System
PHI-ACH
GE-EXC
47
Technical evaluation
Analysis
The SNR was calculated using a subtraction method; two identical scans were
acquired and a difference image obtained. The mean signal was measured from
five regions of interest within the test object area in one of the acquired images
and the noise was measured from the standard deviation from these regions in
the difference image (Dietrich 2005). The values obtained for SNR were
normalised for voxel size (including measured slice width), scan time and
sampling bandwidth (no normalisation for coil loading).
Interpretation of results
The image SNR value obtained on a system is influenced by many factors. For
example, system factors such as the main magnetic field strength Bo and the
design of the radiofrequency receive and transmit systems can affect the SNR.
Other factors are the choice of sequence and imaging parameters.
Table 42 shows the measured SNR values. The comparison made in Graph 12
shows the image SNR normalised for voxel size, scan time, and sampling
bandwidth for high-field systems.
Table 41. Experimental conditions for multi-channel head coil SNR test
GE-EXC PH-ACH SI-TRI
Temperature (°C) 24 23 21.5
Report 05048: 3.0 T MRI systems - issue 2
48
Technical evaluation
Graph 12. Multi-channel head coil normalised SNR (NSNR) (mean of three
planes, parallel imaging factors = off, 2)*
SI-TRI
factorOff
System
factor2
GE-EXC
PH-ACH
Report 05048: 3.0 T MRI systems - issue 2
49
Technical evaluation
Analysis
The average of several intensity profiles is calculated in all three gradient
directions. The fractional uniformity is calculated for each gradient direction from
the fraction of the profile that lies within 10% of the mean value of a central ROI
(Lerski 1998). The optimum value is unity, indicating 100% of the signal is
considered uniform over the measured distance.
The results for each gradient direction are presented in Table 43. The mean of
the gradient directions for each system is indicated in Graph 13.
Report 05048: 3.0 T MRI systems - issue 2
50
Technical evaluation
Graph 13. Multi-channel head coil uniformity (parallel imaging factor off)*
PHI-ACH
System
SI-TRI
GE-EXC
51
Technical evaluation
Analysis
The SNR was calculated using a subtraction method; two identical scans were
acquired and a difference image obtained. The mean signal was measured from
five regions of interest within the test object area in one of the acquired images
and the noise was measured from the standard deviation from these regions in
the difference image (Dietrich 2005). The values obtained for SNR were
normalised for voxel size (including measured slice width), scan time and
sampling bandwidth (no normalisation for coil loading).
Interpretation of results
The image SNR value obtained on a system is influenced by many factors.
Example system factors are the main magnetic field strength B0 and the design
of the radiofrequency receive and transmit systems. Other factors are the
choice of sequence and imaging parameters.
The normalised SNR results for multi-channel body coils are presented in Table
45. The comparison made in Graph 14 shows the image SNR normalised for
voxel size, scan time, and sampling bandwidth for systems with equivalent field
strength.
Table 44. Experimental conditions for multi-channel body coil SNR test
GE-EXC PH-ACH SI-TRI
Report 05048: 3.0 T MRI systems - issue 2
Temperature (°C) 24 23 20
Bandwidth (kHz) 11.9 46.54 16.64
SE for factor off SE for factor off SE for factor off
Sequence TSE for other factors TSE for other factors TSE for other factors
Turbo factor = 2 Turbo factor = 2 Turbo factor = 2
Test object MAGFF-OIL MAGFF-OIL MAGFF-OIL
Loading None None None
52
Technical evaluation
Table 45. Signal to noise ratio for the multi-channel body coil
Parallel
imaging Transverse Sagittal Coronal Mean
factor
GE-EXC Off 2000 2323 2191 2171
2 2269 2178 2000 2149
PH-ACH Off 1179 1193 962 1111
1 1274 1247 1041 1187
2 823 872 757 818
3 462 472 417 450
4 244 303 233 260
6 51 45 39 45
SI-TRI Off 1181 1300 1033 1171
2 743 291 230 421
Data for further parallel imaging factors are not available due to time constraints at the time of system
technical assessment.
GE-EXC
factorOff
System
factor2
PH-ACH
SI-TRI
Report 05048: 3.0 T MRI systems - issue 2
53
Technical evaluation
Analysis
The average of several intensity profiles is calculated in all three gradient
directions. The fractional uniformity is calculated for each gradient direction from
the fraction of the profile that lies within 10% of the mean value of a central ROI
(Lerski 1998). The optimum value is unity, indicating 100% of the signal is
considered uniform over the measured distance.
The results for each gradient direction are presented in Table 46. The mean of
the gradient directions for each system is indicated in Graph 15.
PHI-ACH
Report 05048: 3.0 T MRI systems - issue 2
System
SI-TRI
GE-EXC
Interpretation of results
Both the 2D and 3D imaging speed tests require a fixed volume to be acquired
using 2D and 3D fast imaging sequences. The aim of these tests is to measure
data acquisition speed in voxels/second. The voxel size is defined by the image
matrix and the number of slices in the fixed range. Both the 2D and 3D imaging
speed tests allow for the use of parallel imaging techniques.
MagNET comment
Imaging speed has been calculated as the ratio of imaged voxels to total scan
time. The total scan time is the time displayed by the scanner, rather than the
number of phase-encoding steps × repetition time (TR).
The numerical results of the 2D and 3D imaging speed tests are presented in
the Appendix. The following graphs show the results for no parallel imaging
speeding (parallel imaging off) and for parallel imaging factor =2. It is important
to note that 2D FGE, 2D and 3D FSE imaging speeds have not been measured
on Siemens systems.
Report 05048: 3.0 T MRI systems - issue 2
55
Technical evaluation
PH-ACH
System
factor2
factorOff
GE-EXC
GE-EXC
System
factor2
factorOff
PH-ACH
voxels/sec
56
Technical evaluation
GE-EXC
System
factor2
factorOff
PH-ACH
voxels/sec
factor2
SI-TRI
factorOff
voxels/sec
Report 05048: 3.0 T MRI systems - issue 2
57
Technical evaluation
SI-TRI
SI-ALL
System
factor2
factorOff
GE-EXC
PH-ACH
PH-ACH
System
factor2
factorOff
GE-EXC
58
Technical evaluation
Acoustic noise
The acoustic noise levels measured in this assessment are provided for
indication only. Variations in factors such as room acoustics may mean that
similar noise levels may not be reproduced at a different site - even with an
identical pulse sequence. Acoustic noise levels are given in terms of the
continuous equivalent level, LAeq, which is A-weighted root-mean-square sound
pressure level (SPL) averaged over the measurement period of 1 minute. The
peak noise Lpeak measured on a linear weighting scale is also recorded. The
relevant safety levels follow.
MHRA
Hearing protection is recommended for all patients even when exposure is less
than 99 dB(A). Where sites can demonstrate noise levels significantly below 85
dB(A) then this requirement may be relaxed. Please refer to Guidelines for
Magnetic Resonance Diagnostic Equipment in Clinical Use: Medical Devices
Agency 2002.
Staff
NOISE AT WORK REGULATIONS 1989 (UK)
Employers have legal duty to protect the hearing of their employees. Hearing
protection must be available for workers exposed to 85 dB(A) and must be worn
if levels exceed 90 dB(A). Employers are responsible for performing risk
assessments for employees exposed to noise. This would include staff present
in the MR scan room during imaging.
EU DIRECTIVE 2003/10/EC
This directive becomes effective in the UK in 2006. Compared to the regulations
Report 05048: 3.0 T MRI systems - issue 2
above action levels will be reduced by 5 dB(A). Substitution and control of noisy
equipment must be prioritised above the use of hearing protection. There will be
an exposure limit of 87 dB taking into account the noise reduction afforded by
protective equipment.
59
Technical evaluation
SI-TRI
Report 05048: 3.0 T MRI systems - issue 2
SI-ALL
System
PH-ACH
GE-EXC
0 20 40 60 80 100 120
Continuous equivalent level, LAeq, db(A)
60
Technical evaluation
General
Reference Manual for Magnetic Resonance Safety: F Shellock. Amirsys Inc,
Salt Lake City, 2003.
Safety notice MDA SN 9517. Risk of burns to patients, with attached monitoring
Report 05048: 3.0 T MRI systems - issue 2
61
Technical evaluation
Other bodies
IEC 60601-2-33:2002 (Medical Electrical Equipment. Part 2. Particular
Requirements for Safety. Section 2.33 Specifications for Magnetic Resonance
Equipment for Medical Diagnosis). www.iec.ch
NHS advice
Health Building Note 6, Supplement 1: Accommodation for Magnetic
Resonance Imaging. Produced by NHS Estates, published by HMSO PO Box
276, London SW8 5DT. ISBN 0113217307.
Health Building Note 6, Volume 3: Extremity and open MRI, magnetic shielding
and construction for radiation protection. Produced by NHS Estates, published
by HMSO PO Box 276, London SW8 5DT. ISBN 0113224869.
Report 05048: 3.0 T MRI systems - issue 2
62
Acknowledgements
MagNET would like to acknowledge the support of GE, Philips and Siemens
Medical Solutions during these evaluations. In particular we would like to thank
Mr Ricardo Becerra, Mr John Johnson from GE, Mr Frans Donders, Dr
Elizabeth Moore from Philips, Dr Eckart Stetter, Dr Thorsten Speckner, Mr
Patrick Revell from Siemens, for their expert assistance during the technical
tests.
Report 05048: 3.0 T MRI systems - issue 2
63
References
64
Appendix
Manufacturers’ comments
GE
The manufacturer had no additional comments
Philips
The manufacturer had no additional comments
Siemens
Comparative specifications
Key technical data (e.g. gradient strength and slew rate, image reconstruction speed) are based
on manufacturer information and are not cross-checked by MagNET applying independent
experimental procedures.
The reader of this document should be aware that some manufacturers define these
specifications in a different way making direct comparison invalid. Please note that Siemens
specifies gradient strength and gradient slew rate for each individual axis (and not as a vector
summation).
SNR-evaluation
Comparison of different SNR values acquired with not comparable loads are highly questionable.
The reader is advised to ask each manufacturer for SNR values obtained under clinical loading
conditions.
Spatial resolution
The MTF of a properly calibrated MRI system with 2D-spin single echo imaging is (in absence of
raw data filters) rectangular with a step from one to zero at pi/(nominal pixel size). To see this,
the analysis has to be performed on the complex image data prior to the application of the
magnitude operator [Steckner et al., Med. Phys. 21 (3), 1994 and examples provided therein].
The presented data are obtained from magnitude images. The differences of the calculated
values to the nominal pixel resolution may result from the applied magnitude operator on the
evaluated image and therefore probably will not represent a true resolution difference.
Parallel imaging SNR
Evaluation or comparison on images acquired using parallel imaging techniques requires to keep
in mind that the noise background is not uniform so the results are dependent on the ROI choice.
MagNET comment
MagNET recognises that the image noise can be spatially dependant in images acquired using
parallel imaging techniques. An average is taken from 5 ROIs placed on the image. The
technique used by MagNET has been shown to be robust for parallel imaging acquisitions, as
shown recently in ISMRM (Influence of Parallel Imaging and Other Reconstruction Techniques
Report 05048: 3.0 T MRI systems - issue 2
Table 50. Scan parameters for 2D imaging speed evaluation on the GE MRI
system (Twinspeed HD gradients)
Sequence parameter Sequence parameter values
Imaging coil 8-channel high resolution brain coil
2D sequence 2D FGRE GRE-EPI FSE-XL
TE (ms) 1.6 113 57.4 3.3
TR (ms) 3.5 9300 4800 21000
Flip angle (degrees) 10 90 Not supplied
Report 05048: 3.0 T MRI systems - issue 2
NSA 1 1 1
Bandwidth (kHz) 200 Not measured 83.3
Echo Train Length Not applicable Not applicable 128
FOV (mm) 250 250 250
Matrix (PE x FE) 256 x 256 256 x 256 256 x 256
Slice width (mm) 5 5 4.2
Parallel imaging factors off 1.5 2.0 off 2.0 off 2.0
Range (mm) 200 200 201.6
Contiguous slices 40 40 48
Scan time (min:sec) 0:38 0:26 0:20 0:09 0:05 0:42 0:21
66
Appendix
Table 51. Scan parameters for 2D imaging speed evaluation on the Philips
Achieva MRI system (Quasar Dual gradients)
Sequence parameter Sequence parameter values
Imaging coil 8-channel SENSE head coil
2D sequence 2D-FFE FE-EPI 2D-TSE
TE (ms) 1.50 9.00 34.90
TR (ms) 119 1000 8100
Flip angle (degrees) 10 10 30
NSA 1 1 1
Bandwidth (kHz) Not measured Not measured Not measured
Echo Train Length Not applicable Not applicable Not measured
FOV (mm) 250 250 250
Matrix (PE x FE) 256 x 256 256 x 256 256 x 256
Slice width (mm) 5 5 2.09
Parallel imaging factors 1 2 4 8 1 2 4 8 1 2 8
Range (mm) 200 200 200
Contiguous slices 40 40 96
Scan time (min:sec) 0:30.8 0:15.3 0:07.6 0:04.0 0:21 0:12 0:08 0:06 2:09.6 1:04.8 0:16.2
67
Appendix
Table 54. Scan parameters for 3D imaging speed evaluation on the GE MRI
system (Twinspeed HD gradients)
Sequence parameter Sequence parameter values
Imaging coil 8-channel high resolution brain coil
3D sequence FSPGR FRFSE-XL
TE (ms) 0.9 17
TR (ms) 1.7 217
Flip angle (degrees) 10 Not measured
NSA 1 1
Report 05048: 3.0 T MRI systems - issue 2
68
Appendix
Table 55. Scan parameters for 3D imaging speed evaluation on the Philips
Achieva MRI system (Quasar Dual gradients)
Sequence parameter Sequence parameter values
Imaging coil 8-channel SENSE head Coil
3D sequence 3D FFE 3D TSE
TE (ms) 0.83 51
TR (ms) 1.73 268
Flip angle (degrees) 10 30
NSA 1 1
Bandwidth (kHz) 329.97 329.97
Echo Train Length Not applicable 34
FOV (mm) 250 x 250 x 199.68 250 x 250 x 199.68
Matrix (PE x FE) 128 x 128 128 x 128
Slice width (mm) 1.56 1.56
Parallel imaging factors 1 2 4 8 off 2 4 8
Range (mm) 199.68 199.68
Contiguous slices 128 128
Scan time (min:sec) 0:28.4 0:14.9 0:07.4 0:03.7 1:44 0:36 0:18.8 0:10.2
Report 05048: 3.0 T MRI systems - issue 2
69
Appendix
70
Appendix
Not Not
PH-ACH 37 131 48 545 97 090 388 361
measured measured
Not Not Not Not Not Not
SI-ALL
measured applicable measured measured measured applicable
Not Not Not Not Not Not
SI-TRI
measured applicable measured measured measured applicable
71
Appendix
72
Appendix
Background information
MagNET
The assessment reported on the preceding pages was carried out by the
Magnetic resonance National Evaluation Team (MagNET) based at Imperial
College London. The project is supported by the Department of Health to
assess the imaging performance of commercially available clinical MR systems.
Resources from the National Health Service fund the project.
Manufacturer involvement
The choice of MR system to be evaluated is influenced by both new system
releases and customer demand. Both the user and clinical evaluations are
performed with the active participation of the manufacturer. The clinical sites are
chosen by MagNET from a short-list provided by the manufacturer. The
technical evaluation is carried out at a site agreed with the manufacturer, where
the performance of the system is within specification and is representative of
that model. Acknowledgement is made to the company for arranging and
supporting the examination of MR equipment. A copy of each report is sent to
the manufacturer of the equipment for comment before publication. These
comments are included in the published report.
Report 05048: 3.0 T MRI systems - issue 2
73
Appendix
Contact details
MagNET
In addition to answering queries on the evaluation reports, MagNET can provide
advice on detailed specification, general MR technology, acceptance testing
and quality assurance.
MagNET
Department of Bioengineering
Bagrit Centre
Imperial College London
London SW7 2AZ
E-mail: [email protected]
Web page: http://www.magnet-mri.org/
Report 05048: 3.0 T MRI systems - issue 2
74