Vitrea Installation Guide For Workstation Deployment
Vitrea Installation Guide For Workstation Deployment
Workstation Deployment
VPMC-16207C
Copyright © 1997 -2021 Vital Images, Inc. All rights reserved. Date of Publication: 2-2021
2
VPMC-16207C
Australian Sponsor: Brazil Importer and Distributor:
Canon Medical Systems ANZ Pty Limited Canon Medical System do Brasil
Level 2, Building C Avenida Ceci, 328, Tamboré,
12-24 Talavera Road Barueri/SP
North Ryde NSW 2113 Brasil
Vitrea: ARTG 241833 ANVISA registration number:
MDSS GmbH
Vitrea View: ARTG 262479 10295030090
Schiffgraben 41
30175 Hannover, Germany
Canon Medical Systems ANZ Pty Limited is the authorized sponsor in Australia and acts on behalf of Vital Images, Inc. in the
communication of safety-related incidents and regulatory matters with Therapeutic Goods Administration in Australia.
Distributors are still the first line of communication with their customers regarding service and complaints.
Manufactured by: Vital Images, Inc.; 5850 Opus Parkway, Suite 300; Minnetonka, MN, US; 55343; Phone
866.433.4624
C A U TION
Federal law restricts this device to sale by or on the order of a physician, as directed by
21 CFR 801.109(b)(1).
N OTE
While every effort has been made to ensure the accuracy of the content in this document,
you may notice slight differences between screen captures and the actual software
interface.
Contact Us
l For general, non-technical support questions, contact us through our website: www.vitalimages.com.
l For customer technical support, contact us:
In the U.S., call the Customer Support line at 1.800.208.3005.
l For a printed version of the Release Notes, Education and Reference Guide, or Installation Guides,
contact Customer Support at 1.800.208.3005.
3
VPMC-16207C
UDI
Locate the Vitrea unique device identifier (UDI) on the Help Central page. This identifier contains the
software version information and manufacture date.
1. Click at the top of the window to display the Global Options menu.
Release Notes
Vitrea Release Notes contain late-breaking information not available at the time the Education and
Reference Guide was released. This document is available from your system administrator or from Vital
Images.
4
VPMC-16207C
Contents
5
VPMC-16207C
Appendix D: Windows 7 Setup 97
Install the Graphics Driver 97
Set Microsoft Windows Basic 97
Disable Microsoft User Account Control (UAC) 97
Enable Microsoft Internet Information Services (IIS) and Microsoft Message Queue (MSMQ)
Server 100
Set Up Best Performance and Page File 103
Install Microsoft Windows Updates 107
Set Power Options 107
Appendix E: Windows 10 Setup 108
Before You Start 108
Install the Graphics Driver 108
Disable Microsoft User Account Control (UAC) 108
Enable Microsoft Internet Information Services (IIS) and Microsoft Message Queue (MSMQ)
Server 109
Set Up Best Performance and Page File 113
Install Microsoft Windows Updates 117
Set Power Options 117
Install Language Packs for Chinese, Japanese, and Korean Workstation Installations 118
Appendix F: Configure Vitrea Local Database Users 119
Add Users 119
Disable BUILTIN Users Group in Vitrea application 124
Configure Local Database Groups 126
Modify Local Users 130
Appendix G: Vitrea Installer Command Line Options 132
Command Line Options 132
Appendix H: PowerScribe Configuration 133
Set Up SSL on the PowerScribe Server 133
Verify Name Resolution 137
Configure PowerScribe Server Integration 138
Configure PowerScribe Nodule Export Custom Fields 139
Appendix I: Set up Snapshot Converter, Angio Viewer, CT Dental, and Filming
applications 141
Set up the Snapshot Converter application 141
Set up the Angio Viewer application 142
Set up the CT Dental application 145
Set up the Filming application 147
Appendix J: Supported Monitor Resolutions for VSTP Applications 150
Appendix K: Configure the CSMC Floating License Server 152
Verify the license server is installed 152
Upload the license file 153
Activate the Floating License Server 154
Verify the license is active 155
Configure the TCP port 155
6
VPMC-16207C
Direct Workstations to the Floating License Server 158
Floating License Server Post Setup 159
Troubleshooting the Floating License Server 160
7
VPMC-16207C
Overview
C A U TION
l Federal law restricts this device to sale by or on the order of a physician, as directed
by 21 CFR 801.109(b)(1).
l Vitrea can be configured to lock out users after multiple incorrect passwords are
entered. If this is enabled, it is possible for all administrative users to be locked out.
l A malicious administrator could configure Vitrea software to use a custom log-in
script that compromises user or system security.
l When configuring storage locations during installation, overly long path names may
prevent some applications from saving files correctly.
l All users with permission to log into Vitrea software are able to access data for all
patients stored on the system.
l Application failure may cause temporary files containing patient information to be
stored on disk. Vitrea software temporary files should be manually deleted after an
application failure.
N OTE
It is the customer’s responsibility to ensure the system configuration and logging
capabilities are set up in such a manner to facilitate compliance with any applicable data
privacy laws.
Purpose
This document comprises the installation for the Vitrea workstation deployment, the current release of the
Vitrea® medical diagnostic tool that allows physicians to access 2D and 3D advanced visualization. The
goal of the software is to enable users to measure, rotate, analyze and segment images.
Scope
The scope of the document is to provide a detailed description of the Vitrea software installation and
configuration processes.
Intended Audience
This document is intended for people who wish to install the Vitrea software. Vital assumes you know how to
complete a standard installation process, have a basic understanding of the Windows Operating Systems,
and know how to navigate in the administrative tools of the Operating System.
Required Background
Vital has made every attempt to make this a step-by-step guide. However, some familiarity with a Windows
operating system is assumed. Vital assumes that you have a basic familiarity with using the Vitrea software.
Also, Vital assumes that you are familiar with hardware and software that will be used to complete the
8
VPMC-16207C
installation and/or configuration of the Vitrea software.
9
VPMC-16207C
Install the Vitrea software
C A U TION
Vital recommends you do not disable the Windows Firewall while you install the Vitrea
software.
In This Topic
N OTE
l Unless otherwise noted, the procedures listed below are mandatory and must be
completed to install a fully functional Vitrea system.
l Refer to the Administration Guide for Extend and Workstation Deployments for
optional administrative and troubleshooting procedures.
10
VPMC-16207C
Prepare the Workstation
N OTE
SSO is enabled by default, but may be disabled (refer to the “To Disable the
singleSignOnEnabled setting” section in the Administration Guide for Extend and
Workstation Deployments).
The following procedures assume you are using IWA on either a workgroup or domain setting. Usage of the
Vitrea database users should be reserved for cases in which IWA is not possible. Refer to Appendix D:
Configure Vitrea Local Database Users in this guide for details on configuring Vitrea local database users.
1. Select the Start button and select Settings | System and Security | System.
11
VPMC-16207C
4. In the Computer name field, enter the desired Host Name. Do not include an underscore "_". Click OK.
5. Click OK.
12
VPMC-16207C
• For Windows 7, select the Start button and select Settings | Control Panel | Network and
Sharing Center.
• For Windows 10, right-click the Start button and select Network Connections.
4. On the Network Connections screen, right-click the desired connection and select Properties.
13
VPMC-16207C
8. In the IP address field, enter the static IP address.
9. Enter the desired values in the Subnet mask field and in the Default gateway field.
10. Select the Use the following DNS server addresses radio button.
11. Enter the desired values in the Preferred DNS server field and in the Alternate DNS server field.
Accounts to create
N OTE
Create an account type of each account (shown below) on each machine.
Vitrea-Admin Administrators
N OTE
For all user accounts, set the password to not be changed on login and to never expire.
• For Windows 7, select the Start button and select Settings | Control Panel | Administrative
Tools.
• For Windows 10, right-click the Start button and select Control Panel | System and Security |
Administrative Tools.
14
VPMC-16207C
4. Right-click Users and select New User.
8. Clear the User must change password at next logon check box.
9. Check the User cannot change password and the Password never expires check boxes.
15
VPMC-16207C
10. Click Create.
15. Clear the User must change password at next logon check box.
16. Check the User cannot change password and the Password never expires check boxes.
1. Select Groups.
16
VPMC-16207C
2. Double-click the Administrators group.
4. In the Enter the object names to select field, add the Vitrea-Admin user account.
17
VPMC-16207C
5. Click Apply.
6. Click OK.
1. Select Groups.
2. Right-click the Remote Desktop Users group and select Add to Group.
4. In the Enter the object names to select field, add the Vitrea-User user account.
18
VPMC-16207C
5. Click Apply.
6. Click OK.
Accounts to create
The domain administrator creates these accounts in Active Directory.
Vitrea-User Users
Groups to create
The following groups must be created in Active Directory by an Active Directory administrator. You will use
these groups later to configure corresponding VSP user roles.
N OTE
The domain administrator should add any IT Administration Security Groups that would
be required at the customer site to the Administrators Group on the Workstation.
Vitrea Users Vitrea-User and all other application users (Radiologists, Cardiologist)
19
VPMC-16207C
Add the Vitrea-Admin user to the local Administrators group and to the
Vitrea Administrators group
N OTE
Add any relevant domain administrator users to the security group as needed.
5. Select Groups.
8. In the Enter the object names to select field, add the Vitrea-Admin user account.
9. Click OK.
12. In the Enter the object names to select field, add the Vitrea-Admin user account.
16. In the Enter the object names to select field, add the Vitrea-Admin user account.
5. Select Groups.
20
VPMC-16207C
7. Click the Add button.
8. In the Enter the object names to select field, add the Vitrea-User account.
9. Click OK.
2. Copy the contents of the product download into the Vital_Resources folder.
For Windows 7
1. Navigate to Vital_Resources\Utilities\OS Configuration Scripts\Windows 7.
The script will execute and configure the operating system, and then reboot the Vitrea software.
For Windows 10
To run the System Setup Script using a command line on Windows 10 only
The OS Setup Script has the following command line options:
• -rebootOption
• Possible Values are "prompt", "autoreboot", "noreboot"
5. On the Powershell screen, type A to select Yes to All and press Enter.
7. Type cd "
8. In the PowerShell window, right-click after the " symbol and select Paste.
9. Enter a closed " symbol after the pasted text and press Enter.
21
VPMC-16207C
10. Enter the Powershell command, type “.\Vitrea_Win10_Script.ps1”
The script executes and prepares the system for software installation. It reboots the computer when the
script is finished.
13. At the command prompt, type “Set-ExecutionPolicy Restricted” and press Enter.
3. Expand Security devices and ensure Trusted Platform Module 2.0 is installed and listed.
22
VPMC-16207C
Install BitLocker
1. From the Start menu, select Control Panel.
23
VPMC-16207C
3. Encrypt both the operating system drive and the fixed data drive.
4. For both the operating system drive and the fixed data drive, select the Use a password to unlock the drive
check box and define your password. Re-type the password to confirm and then click Next.
5. Select the Automatically unlock this drive on this computer check box.
6. Select the Save to a recovery key to a file option. Save the file to a safe network location (such as a
Shared Folder) to ensure you can retrieve the key.
24
VPMC-16207C
C A U TION
If you do not save the key somewhere remote and your system goes down, the system
will be unrecoverable.
8. You can choose to either encrypt used space or whole drive contents.
10. Click the Start encrypting button and wait for encryption to finish. This may take several minutes depending
on the size of the drive.
11. If you did not select the Automatically unlock this drive on this computer check box (in step 5 above) or if
you did not get this option, right-click on the drive that was encrypted and select Manage Bitlocker. Then
select the Turn on autounlock option.
N OTE
If you do not perform step 11, the SQL Server Instance is unable to access the ViDicom
database and the Vitrea application will not function after a reboot.
To Uninstall BitLocker
1. Verify BitLocker encryption was successful.
2. If the encryption was successful, BitLocker on displays next to the drive letter.
25
VPMC-16207C
N OTE
Make sure you decrypt before you reimage the workstation.
5. Once the decryption completes, you are ready to reimage the workstation.
1. Right-click the Start button and enter Edit group policy in the Search programs and files field.
26
VPMC-16207C
2. Expand Computer Configuration | Administrative Templates | Windows Components | Internet Explorer
| Internet Control Panel | Security Page.
27
VPMC-16207C
5. In the Options area, click the Show button.
6. In the Value Name field, enter http://<hostname> and enter 1 in the Value column to add
http://<hostname> to the local intranet zone.
7. Click OK.
28
VPMC-16207C
8. Click Apply.
C A U TION
Before you complete the steps below, first uninstall any existing NVIDIA graphics driver
(s). Then restart the machine.
N OTE
l Complete this procedure only on machines that use hardware-based visualization.
l Complete this procedure on machines that use any video card except the NVIDIA
Quadro 2000.
Determine the NVIDIA Driver Type installed on the system (Windows 10 Dell Powerstations and
Workstations only)
N OTE
You only need to complete this procedure if you are using a Dell Powerstation or a Dell
Workstation with Windows 10 installed upon it.
To determine which NVIDIA driver is installed on the system, complete the steps below:
a. Right-click the Desktop and select NVIDIA Control Panel.
b. Select the System Information link in the lower left corner of the screen.
29
VPMC-16207C
c. The Driver Type value lists whether you have a DCH or a Standard driver type installed.
1. For Powerstations or Dell Workstation systems that use Windows 10, complete the "Determine the
NVIDIA Driver Type installed on the system" procedure listed above. For Powerstations or Workstation
systems with Windows 7 installed or for a non-Dell Powerstation or non-Dell Workstation with Window 10
installed, skip to step 2.
30
VPMC-16207C
For Windows 10, the following screen displays.
5. Click OK.
31
VPMC-16207C
6. Click AGREE AND CONTINUE.
32
VPMC-16207C
8. In the Select driver components section, do not change the selected check boxes.
9. Select the Perform a clean installation check box and click the Next button.
33
VPMC-16207C
10. When the installation completes, click Restart Now to restart the machine.
34
VPMC-16207C
11. Click Agree And Continue.
12. Select the Custom (Advanced) radio button and click Next.
35
VPMC-16207C
13. In the Select driver components section, do not change the selected check boxes.
14. Select the Perform a clean installation check box and click the Next button.
36
VPMC-16207C
15. When the installation completes, click Restart Now to restart the machine.
N OTE
Complete this procedure only on machines that use hardware-based visualization.
N OTE
Complete this procedure on machines that use the NVIDIA Quadro 2000 video card.
37
VPMC-16207C
For Windows 10, the following screen displays.
3. Click OK.
38
VPMC-16207C
4. Click Agree And Continue.
39
VPMC-16207C
6. Select the Perform a clean installation check box and click the Next button.
7. When the installation completes, click Restart Now to restart the machine.
40
VPMC-16207C
Install Microsoft Windows Updates
N OTE
• Vital recommends you install Microsoft Windows monthly updates. Follow your site’s
policy regarding software updates.
• Vital recommends you run the .NET Framework patches when you install Microsoft
Windows monthly updates. If you do not have Internet access, complete the
following steps to access the patches:
a. Access the Vitrea product media.
b. Navigate to the Bundle\Prerequisites\NET480 subfolder.
c. Double-click the ndp48.exe to install the updates.
1. Click the Start button, select the Settings option, and select the Update & security option.
41
VPMC-16207C
All important updates and any optional updates will be installed.
For more information, refer to the “Installing the Access database engine” procedure in the CMSC
documents.
42
VPMC-16207C
Configure the firewall to provide access for Ultrasound applications
1. Click the Start button. Type Firewall and select the Windows Defender Firewall with Advanced Security
option.
4. Verify the Program radio button is selected and click the Next button.
5. Select the This Program path: radio button, type C: \CMSC US\CMSC\CMSCWSClinicalApp_
US\bin\Mecha4D.exe in the field and click the Next button.
6. Verify the Allow the connection radio button is selected and click the Next button.
8. In the Name field, enter a descriptive name such as "Ultrasound Software". Click the Finish button.
1. Right-click the InstallVitrea.exe file from the product media or from the Vital_Resources folder and select
Run as Administrator.
43
VPMC-16207C
N OTE
l The default Vital data install location is the drive with the most free space. To change
this location, refer to the “Change the location where data is stored” procedure in
Appendix C: Vitrea Install Options.
l To review the install locations of the Vital applications and products, click the
Advanced button and select Install Options. For a detailed explanation of the
Advanced menu options, refer to Appendix C: Vitrea Install Options.
N OTE
The CSMC Cardiac Suite Floating License Server is not installed automatically. To set
up a Floating License Server to share a license between one or more systems, complete
the steps below.
b. Scroll down to Cedars Concurrent License Server and select the button.
c. After you install the Vitrea software, you will configure the CSMC Floating License Server.
44
VPMC-16207C
This screen lists installation progress. No action is required.
N OTE
It takes about 15-20 minutes to install the prerequisites and components.
45
VPMC-16207C
5. Click OK. The system restarts.
6. If you installed Ultrasound applications, ensure that you have disabled the Windows Firewall.
2. If the system restarts without executing the SetupBuildStartWS.js file, complete the steps below to manually
execute the SetupBuildStartWS.js file.
Create a License
1. To run the Host Information Tool, select the Start button and select All Programs | Vital Images | Vitrea
Generate License HostInfo.
46
VPMC-16207C
2. The license file (using a .lic file format) is automatically saved to the Desktop.
To acquire a license file, you must contact Vital Customer Support. Send an email to
[email protected] and include the following information:
• The System ID that is listed in the screen that displays on your system
• Site Name
• Vital ID
• The host name of the system that houses the license server
• The version of Vitrea software installed
Add a License
N OTE
Prior to installing a license, a “Your system is not licensed; Please contact your
administrator.” message displays on the Applications Home screen.
47
VPMC-16207C
2. Click the Make Changes button.
5. Click the Browse button and navigate to the location where the license has been saved to your system.
6. Click the Apply button. The license file provided by Vital is uploaded.
Now that the Vitrea software is licensed, the Applications Home screen displays with new options available.
48
VPMC-16207C
12. Restart the Vitrea software after the license has been applied.
2. Right-click the Start Vitrea desktop shortcut icon and select Run as Administrator.
49
VPMC-16207C
7. Click the Make Changes button in the left pane. Enter a reason and click OK.
Attach an Archive
This section explains how to add an archive using the System Configuration tab.
N OTE
Complete this procedure to create a DICOM Export/Archive destination.
1. Right-click the Start Vitrea desktop shortcut and select Run as administrator.
50
VPMC-16207C
5. Click the Add button.
6. In the Name field, enter a name for the DICOM device from which the Vitrea software will query/retrieve.
7. In the Key field, enter the name of the Query Node you want to create.
8. In the Host field, enter the Host or IP Address of the DICOM device.
10. In the AE Title field, enter the AE Title of the DICOM device.
N OTE
Use a unique AE Title value.
• If you select the Storage check box and the Query / Receive check box, the system will
automatically create the appropriate DICOM interfaces.
• If you want to set up an export destination for an XA device (DSA), complete the “Configure
Vitrea software for 3D Angio XA volumes” procedure in the Administration Guide for Extend
and Workstation Deployments.
12. Click the Test button to perform a C-ECHO to the archive location.
51
VPMC-16207C
16. Select the Applications Home tab and select Study List.
17. Verify the archive you created displays in the Search section in the upper left corner of the screen.
1. From the Start menu, select Control Panel | Windows Firewall | Advanced Settings.
3. Locate “Open Port 3003” in the Inbound Rules list and double-click the entry.
52
VPMC-16207C
4. Select the Scope tab.
53
VPMC-16207C
5. In the Remote IP Address section, click the Add button.
54
VPMC-16207C
6. Complete this screen based on the IP address location of the querying device.
• If the IP address of the querying device is inside the local subnet of the Vitrea system, select the
Predefined set of computers radio button, click the dropdown menu, and select Local subnet.
• If the IP address of the querying device is outside the local subnet, complete one of the methods below:
• For a single querying device, select the This IP address or subnet radio button and enter the IP
55
VPMC-16207C
address of the device from which to query.
• For multiple querying devices, select the This IP address range radio button and enter the
VLAN range from the query devices.
• If you will not have any devices querying the Vitrea system you can choose to block the port. Complete
one of the two methods below.
Method One
a. Launch the “Open Port 3003” rule from the Inbound Rules list.
b. On the General tab, select the Block the connection radio button.
c. Click the OK button.
Method Two
a. Access the Inbound Rules list.
b. Right-click the “Open Port 3003” rule and select Disable rule.
56
VPMC-16207C
2. Scroll to the <FileCache enabled="true" MininumRequestSizeInBytes="262144" /> line.
Configure DICOM
• Select the Start button and select All Programs | Vital Images | Vitrea Administration Tool.
57
VPMC-16207C
Configure DICOM Receive Storage settings
1. From within the Vitrea Administration Tool, click the Configuration Manager navigation button.
2. In the Navigation Pane, expand Profiles and select the machine profile.
3. In the main pane, select the Inherit from Enterprise link. This ensures all settings are configured in one
profile.
4. On the Action Request prompt that displays, click the Skip button.
58
VPMC-16207C
6. Select the DICOM Receive tab.
59
VPMC-16207C
7. Expand the Connection and Implementation section and complete the following steps:
a. Verify the value entered in the AE Title field matches the AE Title value that you wish to use
for the AETitle. If it does not, change the value and then click Save. Vital suggests that you
enter an AE Title using all capital letters.
N OTE
Use a unique AE Title value.
b. Verify the value entered in the Port field is 3002. If it is some other value, change it to 3002
and then click Save.
9. In the Root Directory field, verify the Root Directory for Temporary Storage is E:\DICOMReceive.
10. Select the Low Watermark spin box and change the value to 15GB.
11. Select the High Watermark spin box and change the value to 20GB.
60
VPMC-16207C
Configure Query Entity settings
1. Expand the Query Entity section.
2. The default AE Title field value is QRSCP. You can modify this value as you desire.
N OTE
To allow someone to query retrieve from the Vitrea system, you must add the host into
the Valid Host List.
2. In the field to the right of the New button, enter the desired IP address of the sending PACS DICOM Archive.
4. Click Save.
3. Click the Delete button. The IP address is removed from the table.
4. Click Save.
N OTE
If you modify any fields and then select another tab or Navigation Button, a Save
Configuration Changes screen displays.
61
VPMC-16207C
5. In the Message field, enter a reason for the change(s) you completed.
6. Click OK.
Configure AutoDelete
The AutoDelete service allows you to set up the rules by which studies from the Vitrea workstation
deployment are automatically deleted. In addition, use this service to define the autodelete behavior when
the disk is full and no studies are expired.
N OTE
The AutoDelete service deletes the oldest patients in the system first.
• For SQL Server Express Edition - there are two states in which the AutoDelete service will execute.
• When the database exceeds 80% of the 10GB database size in SQL Server 2014 SP3 Express
Edition.
• When the configured disk-based mark is reached.
N OTE
Disabling the AutoDelete rules puts the Vitrea workstation deployment in manual delete
mode.
To Configure AutoDelete
You must update the configuration setting for one of the following sections:
• Configure Rule-Based Marking Options
• Configure Disk-Based Marking Options
• Select the AutoDelete tab.
62
VPMC-16207C
Configure AutoDelete Markers
1. Expand the Markers section.
N OTE
For each study in the Vitrea workstation deployment cache, the expiration date is set
63
VPMC-16207C
based as follows:
• The expiration date is set when a study is received by the Vitrea workstation deployment.
• The expiration date is updated when a study is accessed by a user.
• The expiration date is set and updated in accordance with the rules for the modality of the study.
• If a study contains multiple modalities, it changes to the expiration date will always follow the
modality with the longest retention period.
2. Select the Enable rule-based marking check box to allow AutoDelete to mark studies for deletion based on
modality and/or time. The default value is cleared.
• The Modality table field lists the available modalities that were defined with Accepted Service
Object Pair Classes value on the DICOM Receive tab.
• The Retention Days After Receipt of Study table field lists the number of days a study should
be retained after receipt.
• The Retention Days After Last Access table field lists the number of days a study should be
retained after the last study access.
3. In the Disk Full Actions section, select the desired radio button.
4. To create a new rule or update an existing rule, complete the following steps:
Field Action
Maximum Disk Limit (MB) The default value is 20000. Enter the identical value as the
spinbox value entered in the Minimum Disk Limit field.
64
VPMC-16207C
2. The Profiles Name dropdown menu shows three DICOM Modify profiles:
• FULL_ANONYMIZATION - edit existing DICOM Tags defined in the profile only - used during
export.
• EVIDENCE_EDIT - edit existing DICOM Tags defined in the profile only - used during export
• IMPORT_EDIT - edit existing DICOM Tags defined in the profile only - used during import.
2. In the table, select the DICOM Tag to edit and click the Edit button.
5. Click OK.
65
VPMC-16207C
6. Click Save to save your changes.
1. From the DICOM Modify Profiles table, click the Profiles Name dropdown menu and select PATIENT_EDIT.
3. Click the DICOM Tag dropdown menu and select the desired option.
6. Click OK.
The new DICOM Tag displays in the table on the DICOM Modify screen.
1. From the DICOM Modify Profiles table, click the Profiles Name dropdown menu and select PATIENT_EDIT.
2. Select the DICOM Tag to remove and click the Remove button.
The removed DICOM Tag no longer displays in the table on the DICOM Modify screen.
66
VPMC-16207C
system. When you save your changes on the Post Processing tab, the Vitrea
Administration Tool restarts the VIMS Post Processing Service and IIS which can cause a
user to lose an active session.
1. From within the Vitrea Administration Tool, click the Configuration Manager navigation button.
PowerScribe is a comprehensive radiology reporting solution that supports a robust integration platform
which not only allows third party integrators to push discrete data such as measurements directly into the
PowerScribe database, but also supports direct intake and parsing of DICOM SR documents.
Complete the steps below to configure PowerScribe integration.
2. In the Server BaseUrl field, enter the base URL to the PowerScribe AP.
EXAMPLE http://<PowerScribeServerName>/iRADPortal/Services.
3. In the Username field, enter the username used to call the PowerScribe API.
4. In the Password field, enter the password used to call the PowerScribe API.
5. Click the Save button in the upper left corner of the screen.
67
VPMC-16207C
Create a DICOM Printer
1. Select the Configuration Manager button.
2. In the left pane, right-click Remote DICOM Entries and select Create a DICOM Printer.
3. Select the Use the default printer settings check box (if desired) and click Next.
68
VPMC-16207C
N OTE
Use a unique AE Title value.
N OTE
Do not use an underscore (_) when using Host Names.
5. Vital recommends you click the Test DICOM C-ECHO button to determine if the device is available.
N OTE
If the DICOM Test fails, the following message displays.
6. Click Next.
7. Click Finish.
The Configuration Manager screen displays with the new entity included in the Remote DICOM Entities
section of the Navigation Pane.
Listed below is an example of a DICOM Printer once it has been added.
69
VPMC-16207C
Listed below is an example of the default DICOM Printer options. Edit the fields as desired.
1. Right-click the Start button and enter gpedit.msc in the Search programs and files field.
70
VPMC-16207C
3. Select the Enabled radio button and click the OK button to save the change.
The Windows Firewall: Define inbound port exceptions Properties screen displays.
71
VPMC-16207C
6. Click the Show button.
N OTE
There is a 256 character limit to the path when selecting data. When you access data
from a remote source via a UNC path, you should map a drive.
72
VPMC-16207C
3. In the left pane, browse to the directory where the patient data is located.
5. If necessary, update the VIMS Host field to the IP Address of the Vitrea workstation.
6. Modify the AE Title field value to match the AE Title of where the data is being pushed. This value is case
sensitive.
8. Click Push.
9. When the data for the specific patient is successfully pushed, it will be removed from the right window.
1. Launch the Media Manager. Select the Start button and select All Programs | Vital Images | Media
Manager.
2. Click the Browse button and browse to the source location where the data is present.
73
VPMC-16207C
Next Installation Steps
N OTE
l You have completed the setup and implementation of the Vitrea workstation
deployment.
l Refer to the Administration Guide for Extend and Workstation Deployments for
optional administrative and troubleshooting procedures.
74
VPMC-16207C
Appendix A: Database Backup
Procedures
Perform a Database Backup
Example:
Location: %allusersprofile%\Vital Images\Logs\{MachineName}.VI.VIMS.Services.Agent.
MonitorAgentManager.log config
75
VPMC-16207C
N OTE
The Vitrea installer will populate the variable %INSTALLER_PUT% automatically. If you
want the Database Backup to point to a different location, update the Parameters path
using the format listed above.
Job Configuration
When the VIMS Monitor Agent service starts, it reads job information from an XML section of a local
configuration file. On startup, the service calculates an offset value for each job defined in the file. This offset
represents the difference between the current time and the next scheduled execution of the job (time until
the next job execution).
The combination of ScheduleTime, ScheduleIteration and ScheduleAmount determines the job schedule.
Jobs are executed in intervals equal to ScheduleIteration X ScheduleAmount (ex:
ScheduleIteration=“HOUR” and ScheduleAmount =“4” executes the job every 4 hours).
The Enabled=“” attribute must be present for a job to execute. A job that includes the Immediate=“” attribute
is executed upon startup of the VIMS Monitor Agent service, as well as on its schedule.
Each scheduled job calls a stored procedure in the VIDICOM database, providing parameters as required.
Jobs that do not include Extended=“” attribute are executed using the VIDICOM SQL Server login, with
minimal database permissions. Jobs requiring elevated permissions for actions such as file system cleanup
include the Extended=“” attribute and are executed using the credentials of the VIMS Monitor Agent service.
A VIMS Monitor Agent service configuration file is present on the Vitrea software. Any change to the file
requires a restart of the VIMS Monitor Agent service. On the Vitrea installation, this file will be overwritten.
The value for the VIMSDatabaseBackup job parameter will be set to the VitreaData folder location
provided during a Vitrea installation.
The VIMS Monitor Agent service log file is located in the C:\ProgramData\Vitrea\logs\VIMS directory and is
named <hostname>.VI.VIMS.Services.Agent.MonitorAgentManager.
1. From the Start menu, select All Programs | Microsoft SQL Server (version) | SQL Server Management
Studio.
4. Expand VIDICOM.
5. Expand Programmability.
9. In the @BackupPath row, click the Value field and enter the path to which the database will reside
(E:\VitreaData\DBBackup).
10. In the @DaystoRetain row, click the Value field and enter the number of days you wish to retain the backup.
76
VPMC-16207C
Enter 5.
N OTE
If you use a value less than 5, you will lose your pre-existing backup files.
N OTE
l Login to the Vitrea software using the domain account under which the SQL Server
service resides. This account will be a member of the local Administrators security
group on the Vitrea software.
l The database schema is coupled to the build version. When attempting to restore a
database, you must be running the same version of the Vitrea software as was
installed when the database backup was created. There is currently no restriction in
place to prevent restoration of a database backup that does not match the installed
build version.
77
VPMC-16207C
2. In the Server name field, select the hostname\SQL Server instance from the dropdown menu. If the local
server is not present in the server name dropdown menu, click Browse for more from the list.
3. In the Authentication field, select Windows Authentication from the dropdown menu.
4. Click Connect.
6. In the Object Explorer pane, expand the server node and then expand the Databases node.
Database Backup
1. Right-click on the VIDicom database and select Tasks | Back Up.
2. Set the fields as shown in the screen above. Make sure the Backup type field is set to Full.
3. In the Destination section, remove any existing entries, then use Add to specify the location and filename of
the database backup destination. This may be a local drive or a UNC path. Include the .bak extension in the
backup filename.
4. Click OK.
The backup progress displays in the Progress dialog in the lower-right corner.
78
VPMC-16207C
Restore the Database (OPTIONAL)
N OTE
If the database upgrade fails, complete the following procedure to revert to the previous
version of the installation for the Database. The corresponding Vitrea software
installation should also be reverted.
3. In the Other options area, set the Restrict Access state to RESTRICTED_USER.
4. Click OK.
79
VPMC-16207C
7. In the Restore Database dialog, a list of recent backups displays in the Select the backup sets to restore
area. If your backup is in the list, select the Restore check box next to the backup. If your backup is not
listed, select the From device radio button and click the ellipsis button to use the Specify Backup dialog to
add the backup file to the list. This may be a local disk or UNC path and must include the backup filename
and extension. Once the backup is added to the list, select the Restore check box next to the backup.
80
VPMC-16207C
9. Select the Overwrite the existing database (WITH REPLACE) check box.
10. Select the Restrict access to the restored database (WITH RESTRICTED_USER) check box.
11. Verify the Restore As path of the database and log files in the Restore the database files as area. These
reflect the state of the database when the backup was created. If these are incorrect, modify the destinations
of the files in this list.
The restore progress displays in the Progress dialog in the lower-right corner.
14. In the Other options area, set the Restrict Access state to MULTI_USER.
81
VPMC-16207C
Appendix B: Hardware and Software
Requirements
N OTE
• If you are interested in purchasing hardware through Vital, please contact your Vital
Sales Representative.
• For the recommended hardware requirements, refer to the information below.
82
VPMC-16207C
HP Z620 Workstation (SKU #C3R24UC)
N OTE
Modern CPU’s are required to run several of the applications included as part of Vitrea
Advanced Visualization. Certain applications may depend on Advanced Vector
Extensions 2 (AVX2) which were first introduced with Intel Haswell processors in 2013
and AMD Excavator processors in 2015.
83
VPMC-16207C
Extensions 2 (AVX2) which were first introduced with Intel Haswell processors in 2013
and AMD Excavator processors in 2015.
84
VPMC-16207C
• NVIDIA Quadro K2000 2GB 1st GFX
• Windows 7 Operating System installed by default
Networking
Vital Images supports the use of a single network connection per device, except in the use case of multiple
network connections for redundancy so that only a Single IP address and MAC is used. If a network fails, the
system will still be available on the same IP address.
Recommended network connection speed is a 1 Gbps network connection.
C A U TION
Vitrea utilizes a network connection in order to process studies. As such, any Vitrea
system must have a network connection available. In the case of a system without a
network connection, you must install the Microsoft Loopback Adapter. For more
information, refer to Microsoft knowledge base article https://technet.microsoft.com/en-
us/library/cc708322(v=ws.10).aspx.
N OTE
l The recommended graphics card is the NVIDIA Quadro K2200/P2000/P2200/P5000.
l For the Global Illumination Rendering application, the minimum recommended
graphics card is the NVIDIA Quadro K2200.
NVIDIA Drivers
The Vital Rendering Team has evaluated and recommends the following drivers on all workstations:
All video cards except the NVIDIA Quadro 2000
• For Windows 7 using any video card except the NVIDIA Quadro 2000 - use 431.98
• For Windows 10 using any video card except the NVIDIA Quadro 2000 - use 452.39
85
VPMC-16207C
N OTE
• Supported Windows 7 drivers can be found on
Vital_Resources\Utilities\Video Card Drivers\431.98.
• Supported Windows 10 drivers can be found on
Vital_Resources\Utilities\Video Card Drivers\452.39.
N OTE
l The Quadro 2000 is an older card and may not be available in newer models of
hardware. Before installing the specified version of the driver, be sure to verify the
system has the Quadro 2000 rather than the P2000.
l Supported Windows 7 or Windows 10 drivers can be found on
Vital_Resources\Utilities\Video Card Drivers\377.83.
N OTE
For successful remote use of Vitrea, you must install the correct NVIDIA driver listed
above. Complete the steps listed in the “Configure Vitrea for a Single User Remote
Access Deployment” in the Post Installation Procedures topic. Failure to complete
these steps will result in lower performance and/or workflow anomalies.
Monitor Resolution
Vitrea software supports the following monitor setups (32 bit color and 60 hertz):
• a single 20” standard format monitor with a minimum resolution of 1280x1024 with Windows desktop
scaling set to 100% and the recommended resolution of 1600x1200 with Windows desktop scaling set
to 100%.
• a single 24” monitor with a recommended resolution of 1920x1200 with Windows desktop scaling set to
100%.
• a single 30” monitor with a minimum resolution of 1600x1200 with Windows desktop scaling set to
100% and the recommended resolution of 2560x1600 with Windows desktop scaling set to 100%.
• a single 4k monitor with a resolution of 3840x2160 and Windows desktop scaling set to 100%.
N OTE
CMSC Ultrasound applications cannot run on 4k monitors.
• For the Global Illumination Rendering application, a single 24” monitor with a recommended
resolution of 1024x1024 or 2048x2048 with Windows desktop scaling set to 100%.
Supported Monitor Resolutions for VSTP applications
• Refer to Appendix K for a complete list of the supported monitor resolutions for VSTP applications.
86
VPMC-16207C
Recommended Operating Systems
C A U TION
Only English Operating Systems are supported. You can change the Language Pack to
support other languages.
N OTE
The Vitrea 7.14.2 software release is the last version that will support Windows 7
operating systems.
Software Requirements
DNS Entry
Create a DNS entry for the Vitrea software.
Other software
• Microsoft .NET Framework 4.8
• Microsoft Word 2003-2010 (optional component for generating Word reports)
Process Paths
• C:\Program Files (x86)\Vital Images\Vitrea Remote Kiosk\VitreaKioskShell_Release_Win32.exe
• C:\Program Files\Vital Images\Application\bin\VspAppTomcat.exe
• C:\Program Files\Vital Images\Application\jre\bin\java.exe
• C:\Program Files\Vital Images\AppShell\AppShell_Release_x64.exe
• C:\Program Files\Vital Images\DatabaseDeploymentTool\jre\bin\java.exe
87
VPMC-16207C
• C:\Program Files\Vital Images\DicomExportService\VI.ServiceExtension.DicomExport.Service.exe
• C:\Program Files\Vital Images\DicomPrintService\VI.ServiceExtension.DicomPrint.Service.exe
• C:\Program Files\Vital Images\Help\Node.exe
• C:\Program Files\Vital Images\Home\Node.exe
• C:\Program Files\Vital Images\Management\bin\VspMgmtTomcat.exe
• C:\Program Files\Vital Images\Management\jre\bin\java.exe
• C:\Program Files\Vital Images\MigrationService\VI.VIMS.Migration.Service.exe
• C:\Program Files\Vital Images\Part19ToPart10Converter\jre\bin\java.exe
• C:\Program Files\Vital Images\SessionHelper\Node.exe
• C:\Program Files\Vital Images\StackDefinitionBuilder\StackDefinitionBuilderTMDICOM_Release_
x64.exe
• C:\Program Files\Vital Images\StudyList\Bonsai_Shell_Release.exe
• C:\Program Files\Vital Images\VIMS\ArchiveIntegrationAPI\VI.VIMS.ArchiveIntegration.App.exe
• C:\Program Files\Vital Images\VIMS\ArchiveIntegrationAPI\VI.VIMS.ArchiveIntegration.Service.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.DicomReceive.CdarIntegration_Release_x64.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.Enterprise.Servers.Acceptor_Release_x64.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.Enterprise.Servers.AssocServant_Release_x64.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.Enterprise.Servers.DICOMFileCopier_Release_
x64.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.Enterprise.Servers.FileCleanup_Release_x64.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.Enterprise.Servers.Rcv3DProcessor_Release_
x64.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.Enterprise.Servers.VolumeBuilder_Release_x64.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.VIMS.ArchiveIntegration.DarProcessor_Release_
x64.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.VIMS.Services.Agent.MonitorAgentManager.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.VIMS.Services.AutoDelete.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.VIMS.Services.DicomPrintSCU.ServiceManager.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.VIMS.Services.QRSCP.Service.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.VIMS.Services.StoreSCU.ServiceManager.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.VIMS.Services.StoreSCU.Worker.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.VIMS.Services.VitalSigns.Console.exe
• C:\Program Files\Vital Images\VIMS\Vims\bin\VI.VIMS.Services.VSPNotifications.exe
• C:\Program Files\Vital Images\VitreaCourierService\Vitrea.Courier.Service.exe
• C:\Program Files\Vital Images\VitreaLicenseStewardService\Vitrea.LicenseSteward_Release_x64.exe
• C:\Program Files\Vital Images\VitreaRemoteDesktopServices\RemoteDesktopInfo.Service.exe
• C:\Program Files\Vital Images\VitreaSharedImagePoolService\bin\Vitrea.SharedImagePool_Release_
x64.exe
• C:\Program Files\Vital Images\VitreaSharedImagePoolService\bin\Vitrea.SharedImagePoolWorker_
Release_x64.exe
• C:\Program Files\Vital Images\VitreaSPAInfoServices\Vitrea.SPA\SPATool_Release_x64.exe
• C:\Program Files\Vital Images\VitreaSPAInfoServices\Vitrea.SPA\Vitrea.SystemProvisionAllocator_
Release_x64.exe
• C:\Program Files\Vital Images\VitreaSPAInfoServices\Vitrea.SystemInfo\SystemInfo.Service.exe
• C:\Program Files\Vital Images\VitreaSystemHealthService\SystemHealth.Service.exe
• C:\Windows\System32\inetsrv\inetinfo.exe
• C:\Windows\System32\inetsrv\w3wp.exe
• C:\Windows\SysWOW64\inetsrv\w3wp.exe
88
VPMC-16207C
Files/Locations
• C:\pagefile.sys
• C:\ProgramData\Vitrea\temp\SIP\SIPCache
• C:\ProgramData\Vitrea\temp\SPA
• D:\DICOMReceive\DICOMInstanceData
• E:\VitreaData\Temp\SIPCache
• %allusersprofile%\Vital Images\
• **\*.AVOL (path)
• **\*.GPB.GZ (path)
• *.bak
• *.bin
• *.cdar
• *.dar
• *.dcm
• *.gpb
• *.gz
• *.json
• *.ldf
• *.mdf
• *.ndf
• *.png
• *.processed
• *.rxmsg
• *.series
• *.tmp
• *.tn1
• *.tn2
• *.xml
89
VPMC-16207C
Appendix C: Vitrea Install Options
Install Options
This topic describes the options available to you from the Advanced button on the Vitrea software install
screen.
90
VPMC-16207C
Option Description
91
VPMC-16207C
Option Description
Install Options
92
VPMC-16207C
Option Description
next to the component name and select . After you click INSTALL, a
screen will display when the product you wish to modify is about to be
installed. Enter the desired install path for that specific product.
N OTE
Vitrea detects the largest partition on which to install.
Version Information
,
cont.
93
VPMC-16207C
Option Description
Installation Information
,
cont.
94
VPMC-16207C
Option Description
Click this button to access a drop down menu that lists the available
deployment options:
This option displays if you have not run the script file to configure the
Windows Operating System. Refer to the Install the Vitrea software topic.
95
VPMC-16207C
Change the location where data is stored (OPTIONAL)
If you need to change the location where data is stored to another location other than what is selected by
default, complete the steps below.
1. Double-click the InstallVitrea.exe file from the product media or from the Vital_Resources folder.
2. Click the Advanced button and select Install Options from the drop down menu.
A screen displays when the products you wish to modify are about to be installed.
96
VPMC-16207C
Appendix D: Windows 7 Setup
This topic describes the procedures that are completed by the System Setup Script to configure Windows 7
on the Vitrea software. To use a system setup script to automatically configure the majority of these
procedures, refer to the Install the Vitrea software topic in this guide.
3. In the Basic and High Contrast Themes section, select Windows 7 Basic.
97
VPMC-16207C
2. On the Control Panel screen, select User Accounts.
98
VPMC-16207C
4. Select Change User Account Control Settings.
N OTE
You can also access the User Account Control Settings screen by entering “uac” in the
Search Programs and Files field on the desktop.
99
VPMC-16207C
5. To turn off User Account Control, move the slider to Never notify position and click OK.
2. Select Programs.
100
VPMC-16207C
3. Under the Programs and Features heading, select Turn Windows features on or off.
N OTE
You can also access the Windows Features screen by entering “windows features” in the
Search Programs and Files field on the desktop.
101
VPMC-16207C
4. On the Windows Features screen, select the following features listed under the Internet Information
Services heading:
• Select the Web Management Tools feature check box and all its subcomponents.
• Select the World Wide Web Services feature check box.
• Select the .NET Extensibility, ASP, and ASP.NET check boxes.
• Select the Microsoft .NET Framework 3.5.1 feature check box and all its components.
• Expand the Microsoft Message Queue (MSMQ) Server feature list and select the MSMQ
DCOM Proxy subcomponent.
N OTE
Only select the features and components listed above.
5. Click OK.
102
VPMC-16207C
Set Up Best Performance and Page File
1. From the Start menu, select Control Panel.
3. Select System.
103
VPMC-16207C
5. Select the Advanced tab.
6. Click Settings.
104
VPMC-16207C
7. Select Adjust for best performance radio button.
8. Click Apply.
105
VPMC-16207C
N OTE
Microsoft Page Files are 1.5 times the total amount of RAM on the system. Therefore, if
your system has a lot of RAM and a small OS partition, you may see performance issues.
For systems with a C and E drive, the Page File should be disabled for the C drive and
enabled for the E drive.
11. The Automatically manage paging file size for all drives check box is selected by default. To change this
setting, you must clear that check box, choose your drive partition, and then select the Custom size radio
button, and enter the values in the Initial size and Maximum size fields based on the value listed in the
Recommended entry in the Total paging file size for all drives section.
N OTE
Only one page file should be used for the entire system no matter how many partitions
exist on Vitrea.
12. Select the correct drive and select System managed size radio button.
106
VPMC-16207C
14. Click OK. Then click OK twice.
15. Restart Vitrea.
• Complete the “Install Microsoft Windows Updates for Windows 7” procedure in the Install the Vitrea
software topic.
4. For the Turn off the display field, select 15 minutes from the dropdown menu.
5. For the Put the computer to sleep field, select Never from the dropdown menu.
6. Restart Vitrea.
107
VPMC-16207C
Appendix E: Windows 10 Setup
This topic describes the procedures that are completed by the System Setup Script to configure Windows 10
on the Vitrea software. To use a system setup script to automatically configure the majority of these
procedures, refer to the Install the Vitrea software topic in this guide.
108
VPMC-16207C
4. To turn off User Account Control, move the slider to Never notify position and click OK.
6. Restart Vitrea.
109
VPMC-16207C
4. Select the .NET Framework 3.5 (includes .NET 2.0 and 3.0) feature check box and all its subcomponents.
110
VPMC-16207C
9. Expand the Web Management Tools feature heading.
• Select the IIS Management Console, IIS Management Scripts and Tools, and the IIS
Management Service check boxes.
111
VPMC-16207C
• Select the .NET Extensibility 3.5, .NET Extensibility 4.7, ASP, ASP.NET 4.7, ISAPI
Extensions, and ISAPI Filters check boxes.
• Expand the Common HTTP Features subheading.
• Select all subcomponents except WebDAV Publishing.
• Expand the Health and Diagnostics subheading.
• Select the HTTP Logging, Logging Tools, Request Monitor, and the Tracing check
boxes.
• Expand the Performance Features subheading.
• Select the Static Content Compression check box.
• Expand the Security subheading.
• Select the Request Filtering check box.
112
VPMC-16207C
12. Expand the Microsoft Message Queue (MSMQ) Server heading.
13. Expand the Microsoft Message Queue (MSMQ) Server Core subheading.
• Select the Internet Printing Client and the Windows Fax and Scan check boxes.
16. Select the Remote Differential Compression API Support check box.
17. Select the SMB 1.0/CIFS File Sharing Support check box.
N OTE
Only select the features and components listed above.
3. Select System.
113
VPMC-16207C
5. Select the Advanced tab.
114
VPMC-16207C
8. Select Adjust for best performance of Programs radio button.
9. Click Apply.
115
VPMC-16207C
N OTE
Microsoft Page Files are 1.5 times the total amount of RAM on the system. Therefore, if
your system has a lot of RAM and a small OS partition, you may see performance issues.
For systems with a C and E drive, the Page File should be disabled for the C drive and
enabled for the E drive.
12. The Automatically manage paging file size for all drives check box is selected by default. To change this
setting, you must clear that check box, choose your drive partition, and then select the Custom size radio
button, and enter the values in the Initial size and Maximum size fields based on the value listed in the
Recommended entry in the Total paging file size for all drives section.
N OTE
Only one page file should be used for the entire system no matter how many partitions
exist on Vitrea.
13. Select the correct drive and select System managed size radio button.
116
VPMC-16207C
15. Click OK. Then click OK twice.
• Complete the “Install Microsoft Windows Updates for Windows 7” procedure in the Install the Vitrea
software topic.
4. For the Turn off the display field, select 15 minutes from the dropdown menu.
5. For the Put the computer to sleep field, select Never from the dropdown menu.
6. Restart Vitrea.
117
VPMC-16207C
Install Language Packs for Chinese, Japanese, and Korean
Workstation Installations
If you are pushing datasets to the Vitrea workstation deployment that will contain Chinese, Japanese, or
Korean characters in patient names, you must complete the following steps to install language packs on the
Vitrea workstation deployment to ensure the patient names display correctly in the Study Directory.
N OTE
This configuration will also impact the machine’s user interface to items such as Date
and Time. Refer to your Microsoft Windows documentation for more information.
5. Select the language you want to add from the list of input languages.
6. On the Time and Language screen, the newly-added language displays under the Languages heading.
118
VPMC-16207C
Appendix F: Configure Vitrea Local
Database Users
You may install the Vitrea workstation deployment to authenticate with Integrated Windows Authentication
(IWA), or via a Vitrea local user database. Use of IWA is strongly recommended, as it provides single sign-
on (SSO) support, and uses the standard Windows operating credential management for both workgroup
and domain (Active Directory) deployments.
N OTE
SSO is enabled by default, but may be disabled (refer to the “To Disable the
singleSignOnEnabled setting” section in the Administrative Procedures topic).
This appendix explains how to configure Vitrea local database users which would be performed for an
installation in which IWA is not possible.
Add Users
N OTE
Vitrea user configuration is only necessary if Integrated Windows Authentication is not
being used.
Users can either be a Windows user (local to the machine and part of the domain) or a local database user.
A local database user is a member of a local database group.
In this section, you will use the Local Database User Management subsystem within the System
Configuration tab to create users on the system.
N OTE
For more details, refer to the “Groups, Users, Roles, and Permissions Overview “section
in this topic.
119
VPMC-16207C
3. Click the Add button.
120
VPMC-16207C
Field Action
Display name field Enter the user's name, e.g., John Smith or Jane Smith.
Username field Enter the name used to sign into the application.
Enter the user's password. If the password does not meet the
Password field
requirements, the Password box outlines in red (shown below).
Confirm password
Re-enter the user's password.
field
Groups check boxes Select the group(s) to which the user belongs.
Select this box to enable the user account. Clear this box to disable
Enabled check box
the user account.
Password expires Select this box to force the password to expire after a given time
check box period.
Force user to
Select this box to force the user to change the password at the next
change password
sign in.
next login check box
5. Click Apply.
Disable Password for Guest User and Change Password Admin User
N OTE
Vital recommends you complete this procedure.
1. On the System Configuration screen, select Local Database User Management in the Subsystem menu.
121
VPMC-16207C
2. Double-click the guest Username.
122
VPMC-16207C
7. Select the Advanced dropdown button.
8. Enter the desired new password in the New Password and Confirm password fields.
123
VPMC-16207C
Disable BUILTIN Users Group in Vitrea application
N OTE
If you want to limit the access that specific users have to the Vitrea software, complete
this procedure to Disable the BUILTIN User group and ensure access is locked down.
124
VPMC-16207C
4. In the Domain field, click the dropdown menu and select the desired domain.
5. In the Search Terms field, enter Vitrea Administrators and select Search.
The External Group Search screen displays the results of the search.
6. For the Vitrea Administrators group, select the check box in the Admin column.
7. Click Add.
A revised version of the Group /Role Mapping screen displays with the newly add External Group name(s).
8. Clear the User check box for the BUILTIN Users group.
125
VPMC-16207C
Configure Local Database Groups
This section explains how to add or remove local database groups.
126
VPMC-16207C
Administrator account. When a computer joins a domain, the Domain Admins group is added to the
Administrators group. When a server becomes a domain controller, the Enterprise Admins group also is
added to the Administrators group.”
• Users - this group will start with a mapping to the “User” role. Microsoft describes this built-in group as
“After the initial installation of the operating system, the only member is the Authenticated Users group.
When a computer joins a domain, the Domain Users group is added to the Users group on the
computer.”
N OTE
Both the Administrators and the Users groups can be edited by a system administrator.
4. In the Group Name field, enter the desired fully qualified group name.
6. Click Add.
A revised version of the Group /Role Mapping screen displays with the newly created External Group name.
7. Select the Service, User, and/or Admin check box(es) to assign the desired permission(s).
127
VPMC-16207C
2. In the Subsystem menu, select Group / Role Mapping.
128
VPMC-16207C
3. Select the check box that corresponds to the role you want to add.
129
VPMC-16207C
Modify Local Users
The System Configuration tab is used to perform several tasks. In this section, you will use it to edit or
delete an existing user.
1. On the System Configuration screen, select Local Database User Management in the Subsystem menu.
2. Either double-click the desired user or right-click the desired user and select Edit User.
3. Modify the desired field and check box values and click the Save button.
130
VPMC-16207C
To delete a local user
1. On the System Configuration screen, select Local Database User Management in the Subsystem menu.
131
VPMC-16207C
Appendix G: Vitrea Installer Command
Line Options
Command Line Options
To access all available command line options, complete the following steps:
132
VPMC-16207C
Appendix H: PowerScribe Configuration
C A U TION
Only the customer administrator or the PowerScribe administrator should be configuring
SSL on the PowerScribe Server.
This appendix explains how to configure data export to PowerScribe via HTTPS.
133
VPMC-16207C
4. In the main workspace, in the IIS area, double-click the Server Certificates icon.
7. Click Next.
134
VPMC-16207C
8. Accept the default value for the Cryptographic service provider dropdown menu.
11. In the Specify a file name for the certificate request field, enter c:\certreq.txt.
135
VPMC-16207C
Enable HTTPS Binding
1. On the IIS Manager screen, select the server name in the left pane.
2. Expand the Sites folder and select the Default Web Site option.
By selecting All Unassigned, the site will respond to requests for all IP addresses on the port and optional
host name that you specify for this site, unless another site on the server has a binding on the same port but
with a specific IP address.
7. In the Port field, enter 443 as the outbound https port for PowerScribe integration.
8. Click the SSL certificate dropdown menu and select the certificate you want the site to use for SSL.
9. Click OK.
C A U TION
This should only be performed during planned downtime since this will disrupt
application usage.
136
VPMC-16207C
Verify Name Resolution
Complete this procedure to confirm the DNS entry for PowerScribe.
2. On the system you are integrating with PowerScribe, open a Command prompt.
11. Update the web.config file with the appropriate certificate issuer.
137
VPMC-16207C
12. Save the web.config field.
138
VPMC-16207C
5. In the Server BaseUrl field, enter the hostname to the PowerScribe API.
https://<PowerScribeServerName>/iRADPortal/Services.
6. In the Username field, enter the username used to call the PowerScribe API.
7. In the Password field, enter the password used to call the PowerScribe API.
Doubling Time
Vital_Nodule_DblTime_[n] ---
(days)
Comments Vital_Nodule_Comments_[n]
139
VPMC-16207C
where [ n ] = { 1 , 2 , 3 , .......... n } is the number of nodules.
If the site is doing comparison workflows then the following custom fields must also be added for however
many simultaneous comparisons the site envisions.
where [ v ] = { 1 , 2 , .......... v } is the number of volumes loaded into Lung Analysis Protocol.
N OTE
Custom fields should be added to the PowerScribe database prior to exporting nodules
from the Vitrea application.
140
VPMC-16207C
Appendix I: Set up Snapshot Converter,
Angio Viewer, CT Dental, and Filming
applications
Set up the Snapshot Converter application
To use the following applications, you must complete this procedure.
• US Cardiac Fusion
• NM Cardiac Fusion
• SUREVolume Synthesis
2. Login as “admin”.
8. On the Scripts screen, click the Add button in the upper left corner.
141
VPMC-16207C
The Add Script screen displays.
11. In the Language field, click the dropdown menu and select groovy.
14. On the Scripts screen, verify SnapshotConverterLauncher is added and is listed as Active.
To confirm the C-arm type setting and the height of iso-center, complete the steps below.
a. Click the Utility tab of system console of the INFX-8000 system.
b. Click Service Tools-Service.
c. On the Service screen input the password. You need to confirm the password before this
operation.
d. Open config-system in the Service screen then confirm the information.
3. Login as “admin”.
142
VPMC-16207C
6. Select System Settings in the Subsystem menu.
8. Click the Make Changes button in the left pane. Enter a reason and click OK.
10. Enter one of the following values for single plane system or FPD of frontal arm in biplane system.
• “FPD8x8inch”
• “FPD12x12inch”
• “FPD16x12inch”
12. Enter one of the following values for FPD of lateral arm in biplane system. If the system is single plane, you
do not need to complete this step.
• “FPD8x8inch”
• “FPD12x12inch”
143
VPMC-16207C
• “FPD16x12inch”
If the correspondence relation between C-arm and height of iso-center corresponds to the table values
shown below, enter “default”.
CAS-810A 1050 mm
CAS-820B 1110 mm
CAS-880A 1110 mm
If the C-arm type is CAS-810A but the height of iso-center is NOT 1050 mm, enter the height of the iso-center
of the INFX-8000 system.
If there is another case, contact Vital Customer Support
144
VPMC-16207C
3. Select the DICOM radio button.
4. In the Name field, enter the desired name for the DICOM device.
7. In the Port field, enter the Port number of the DICOM device.
10. Click the Test button to perform a C-ECHO to the archive location.
2. Login as “admin”.
145
VPMC-16207C
5. Select System Settings in the Subsystem menu.
7. Click the Make Changes button in the left pane. Enter a reason and click OK.
8. Double-click the Value column in the following parameters and then enter the new value.
146
VPMC-16207C
Sample 2
frameLayoutForFilming: 6 x 3
147
VPMC-16207C
Vitrea Administration Tool procedures
Refer to the “Create a DICOM Printer” procedure within this guide.
The following table lists the notes on the print options supported by the Filming application.
Setting
Description
contents
8INX10IN
8_5INX11IN
10INX12IN
10INX14IN
11INX14IN
Print Size
11INX17IN
14INX14IN
14INX17IN
24CMX24CM
24CMX30CM
A4
A3
2. Login as “admin”.
148
VPMC-16207C
5. Select System Settings in the Subsystem menu.
7. Click the Make Changes button in the left pane. Enter a reason and click OK.
149
VPMC-16207C
Appendix J: Supported Monitor
Resolutions for VSTP Applications
Listed below are the supported monitor resolutions for all VSTP software applications.
3D Viewer O O X X
Angio Viewer O O X X
Cerebral
Aneurysm O O X X
Analysis
Coronary CTA
O O O O
Analysis
CT-FFR O O O O
CT Dental O O O O
CT Protocol
O O X O
Management
DE Composition
O O X X
Analysis
DE Stone
O O X X
Analysis
DTS Data
O O X X
Management
Dynamic
Myocardial O O X O
Perfusion
Embolization
O O X X
Plan
Filming O O O O
MR Coronary
O O X X
Tracking
MR Wall Motion
O O X X
Tracking
150
VPMC-16207C
VSTP 1280x1024 3840x2160
1920x1200 1600x1200
Application (*L) (*O)
Patient Dose
O O X O
Management
Respiratory
O O X O
Analysis
SUREVolume
O O X O
Synthesis
Utilization
O O X O
Analysis Tool
Ultrasound
Clinical
O X X X
Application
Package
US Cardiac
O O X X
Fusion
151
VPMC-16207C
Appendix K: Configure the
CSMC Floating License Server
This section explains how to configure the Cedars-Sinai Medical Center (CSMC) Cardiac Suite Floating
License Server. The CSMC Cardiac Suite applications are installed automatically on the Workstation.
For a Vitrea Workstation deployment, you can configure Cedars licensing using one of two scenarios:
Local Node-Locked - this option limits license usage to a single machine. If you choose Local Node-
Locked, you will install the license file on each Vitrea Workstation in your deployment. You will also need to
run the Host Information Tool on each Vitrea Workstation in your deployment.
To set up a local node-locked license on a Vitrea Workstation, complete the procedures listed below on
each Vitrea Workstation in your deployment.
Install Cedars Floating License Server on a Workstation - the necessary license server software is not
automatically installed with the Vitrea Workstation installation. You only need to complete the procedures
listed below on the system acting as the license server and not on each workstation in your deployment.
You should not set up a floating license server on a Vitrea Workstation.
N OTE
Be aware that the floating and node-locked license types are not interchangeable. Once
you receive the License File from Customer Support, you cannot change it.
C A U TION
l The CSMC Cardiac Suite Floating License Server requires a Call Home (i.e.,
constant internet access) to http://vm.csaim.com/. You may need to set up rules in
your proxy server to reach the Call Home destination. If the software cannot reach its
Call Home destination with 14 days, your floating license will shut down.
l If you use a proxy server, you may need to add a rule to the proxy server to allow
connections to http://vm.csaim.com/ or white list a connection to that Internet site.
l If you are using the CSMC applications, a new version of the CSMC Cardiac Suite
Floating License Server 2017 software and a new license is required.
N OTE
• The QGS/QPS and the QBS applications require both a Vital license and a Cedars
license. To configure the CSMC Floating License server and enable use of the
CSMC applications, you must perform this procedure on the Workstation and then
point any additional Workstations to the CSMC Floating License Server.
• To use the QGS/QPS and the QBS applications on the Workstation, you must first
configure the floating license server.
152
VPMC-16207C
complete the following procedure. Be aware that you must contact Vital Customer
Support to obtain a valid license file during this procedure. Plan your time accordingly
because Vital Customer Support typically requires 24-48 hours after being contacted to
send you the license file.
1. Complete the "Acquire the License File" section in the Prepare for Licensing topic.
4. To obtain a valid license file, you need to contact Vital Customer Support. Send an email to
[email protected] and include the following information:
• The System ID that is listed in the screen that displays on your system
• Site Name
• Vital ID
• The host name of the system that houses the license server (typically the Management Server)
• The version of Vitrea software installed
153
VPMC-16207C
4. Click the Choose File button and navigate to the location of the license file.
154
VPMC-16207C
3. Complete the Registered To, Company, and Email #1 fields.
3. Select the various tabs to ensure the license file is active. Or, navigate to C:\Program Files\CSMC Cardiac
Suite\bin\csconfig.exe and view the license status tab.
155
VPMC-16207C
2. In the left pane, select Inbound Rules.
5. Select the TCP radio button and add 6433 into the Specific local ports field.
6. Click Next.
156
VPMC-16207C
7. Select the Allow the connection radio button and click Next.
157
VPMC-16207C
9. In the Name field, enter Cedars Floating License Server and click Finish.
10. On the Windows Firewall with Advanced Security screen, double-click the new inbound rule you just
created.
12. In the Remote IP Address section, select the These IP Addresses radio button and add all other
Workstations that need to access the floating license server.
158
VPMC-16207C
2. In the Host field, enter the hostname or IP address of the Floating License Server.
4. Click OK.
159
VPMC-16207C
3. Click the Get Reset Code button. A code will be emailed to you.
4. Once you receive the reset code, complete steps 1 and 2 to access the screen shown above.
5. In the Reset Code field, enter the code you received in the email and click the Reset Server button.
1. Enable Firewall logging for dropped packets by using of the following methods.
The log file shows that a machine with IP address 172.30.17.50 tried to access 172.29.11.76:6433 (the
host:port of the Cedars Floating License Server) and the firewall dropped the packets.
Error displays when you activate the Floating License Server
If the error shown below displays when you activate the floating license server, you do not have the
necessary Internet access.
160
VPMC-16207C
161