NX 3.0.9000 - Chapter 09 - Troubleshooting
NX 3.0.9000 - Chapter 09 - Troubleshooting
NX 3.0.9000
Type 4406/303
► Document History
► Referenced Documents
Document ID Title
57082877 NX 3.0.9000 Service Manual
Edition 1, Revision 0
12-2016 printed in Germany Document Node ID: 57096423
troubleshooting_chapter_e_template_v01_modified
Agfa Company Confidential Copyright © 2016 Agfa HealthCare N.V.
Troubleshooting
► Manufacturer
Agfa HealthCare N.V.
Published by
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright 2016 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium, or its
affiliates.
All other trademarks mentioned in this document are held by Agfa HealthCare N.V. or
the respective owners and are used in an editorial fashion with no intention of
infringement.
Nothing contained in this legal notice nor in any text in this document shall be
construed as granting by implication, estoppel or otherwise, any license or right to use
any of the trademarks, service marks, trade names or logos appearing in this document
without the express prior written consent of their respective owner.
WARNING:
Improper operation or service activities may cause damage or injuries.
(1) Read the "Generic Safety Directions" prior to attempting any operation, repair or
maintenance task on the equipment.
Refer to Document ID 11849633.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
IMPORTANT:
The installation and service of the product(s) described herein is to be performed
by qualified personnel who are employed by Agfa HealthCare N.V or one of its affiliates
or who are otherwise authorized by Agfa HealthCare N.V. or one of its affiliates to
provide such services.
LIST OF CONTENTS
1.1.2 Sometimes it takes a long Time to write an Image Plate (Cassette) ........................................ 8
1.2.1 Print Sheet Text Boxes empty and inaccessible in NX Configuration Tool .............................. 9
1.2.2 Configuration Tool start-up Error "Missing or wrong initial configuration of the database" ..... 9
1.2.3 Activating a Configuration gives Error about wrong Settings for partial Exposure ................. 10
1.2.4 When using the Digitizer Emergency Button, Images do not arrive on NX Workstation ........ 10
1.2.8 Changes made in the Agfa Dicom Bridge (ADB) Tool running on Windows 7 are active after
a Reset only ............................................................................................................................ 11
1.2.9 Error Messages when starting NX Viewer or Configuration Tool after Installation or Upgrade12
1.2.10 Configuration Tool not startable after Power Cut during Configuration Restore .................... 12
1.2.11 Test Connection does not work after AE Title or Port Number Change of the MUSICA MCE
Engine ..................................................................................................................................... 13
1.2.12 Images from Digitizer arrive but cannot be modified: All Buttons grayed out ......................... 13
1.3 Upgrades................................................................................................................................. 15
1.3.2 NX doesn't start when logging in as User with Operator Rights ............................................. 15
1.3.5 Configuration Data Import Message: Exception caught during import [Padding is invalid and
cannot be removed.] ............................................................................................................... 17
1.3.6 Recovering Data after failed automatic Import during Software Upgrade .............................. 17
1.3.7 Error when loading the License File before Activation ........................................................... 19
1.3.8 Failure to start Licensing Service and to import Configuration, Image and Dose Data during
Software Upgrade starting from NX 2.0.8000 / 3.0.8000 ........................................................ 19
1.3.9 NX upgrade or NX installation fails with SQLServer 2014 SP2 error message ..................... 21
1.4.1 RIS Query may become slow after Upgrade from NX 1.0 ...................................................... 22
1.4.4 Accession Number cannot be edited after Image has been received .................................... 24
1.5.4 Wrong Characters on PACS (e.g. ' º ', ' cm² ') ........................................................................ 25
1.7.1 NX Application does not start up anymore due to a corrupted Database .............................. 29
1.7.5 Fast Preview Image is not shown: Error Messages about AE Title are displayed ................. 31
1.7.7 Missing W/L Markers in the Histogram of MUSICA 2 processed Images ............................... 31
1.7.9 After initiating a Shutdown of the NX Workstation the System keeps in Shut-Down State .... 32
1.9.3 Error about “License service is not running” when going back to previous Windows Restore
Point ........................................................................................................................................ 37
1.9.5 NX Workstation in grace period after the Network Adapter name has been renamed........... 39
1.11.2 Additional Internet Explorer Windows are displayed when retrieving Priors in Web1000 ..... 41
1.11.3 "Fatal exception" Error when refreshing the Internet Explorer ............................................... 41
1.12.2 Barco MFCD1219 TS does not work due to wrong Driver ..................................................... 44
1.14.1 Mammograms processed with "Mammo" or "Mammo T2" sporadically show Black Burn
Areas in the Skin Line ............................................................................................................. 45
1.14.2 NX User Manual does not give Instructions how to proceed if "Autocollimation" fails with
MUSICA 2................................................................................................................................ 46
3.6 Checking Version Information with “About NX” (only available as of NX 2.0.6805) .............. 58
3.10.1 Lock the Screen by Password Protection of the current User (default Configuration after
Installation) .............................................................................................................................. 61
3.10.2 Configuring the Screensaver to log out the current User ....................................................... 62
3.13 Saving the Adjustments of Musica 1 Settings (via NX GUI) to the Backup File ..................... 64
3.16.3 Import of the Configuration Data fails due to Verification Errors in the Data .......................... 70
4 FAULT FINDING.....................................................................................................................72
4.1 NX System Snapshot .............................................................................................................. 73
1.1 Digitizer/ID-Tablet
Symptom For example, a scrambled image is received from the digitizer. When trying to
resend it an error occurs.
Cause The error occurs because the software is implemented to ignore images being
sent twice.
Solution (1) Remove the original image from the study via image transfer and then resend
the image from the digitizer. NX gives following message:
Figure 1
1.2.1 Print Sheet Text Boxes empty and inaccessible in NX Configuration Tool
Symptom In the NX Configuration Tool, the print sheet text box configuration is lost and
configuration is not accessible anymore.
Cause Software bug
Solution Restore the factory default text box configuration:
(1) Go to: Device Configuration Printers
Figure 2
Figure 3
Symptom When starting the Configuration Tool, this error appears: "Missing or wrong initial
configuration of the database".
Cause Possible cause is a interrupted procedure to restore configuration
(e.g. power failure).
Solution (1) Go to: Start Agfa NX Service Install tools
Reset NX Configuration Factory Defaults
(2) Start the Configuration Tool.
(3) Select Restore again.
1.2.3 Activating a Configuration gives Error about wrong Settings for partial Exposure
1.2.4 When using the Digitizer Emergency Button, Images do not arrive on NX
Workstation
Symptom When using the digitizer emergency button, the images are not arriving on the NX
workstation.
Cause The digitizer CPF file contains the destination on the NX where emergency images
should be sent to. If the hostname of the NX Workstation is changed and the CPF
file is updated using the NX Configuration Tool, the emergency destination is not
updated and the digitizer cannot send to this emergency destination anymore.
Solution After changing the hostname of an NX workstation that is configured as emergency
destination, update the emergency destination in the CPF file manually, to reflect the
new hostname.
Symptom When trying to start the Agfa DICOM Bridge (ADB) Tool, it cannot be started and an
error message is displayed.
Cause The logged on user has not sufficient rights to access the ADB tool folder.
Solution Log in using the “crservice” account to be able to use the ADB tool.
1.2.8 Changes made in the Agfa Dicom Bridge (ADB) Tool running on Windows 7 are
active after a Reset only
Symptom Changes made in the Agfa Dicom Bridge (ADB) Tool running on Windows 7 are not
taken into account.
Cause Different behavior of Windows 7 compared to Windows XP.
Solution After changes in the ADB Tool on Windows 7 either restart the ADB Service or
reboot the NX Workstation.
1.2.9 Error Messages when starting NX Viewer or Configuration Tool after Installation
or Upgrade
Symptom After the installation or upgrade starting the NX viewer or the Configuration Tool
fails. One of these messages is displayed:
• Windows Event properties dialog:
".NET Runtime 2.0 Error for the System.data.sqlclient.sql"
• MESS279 "NX has encountered an internal error"
• MESS310 "The system has encountered an unexpected error"
• "Provider services not yet ready. Please try restarting the application."
• "An exception has occurred, please check the following error messages"
Cause After installation or upgrade there may be a delay in the startup of the database.
This causes this error.
Solution Start the NX viewer or Configuration Tool after some minutes waiting time.
1.2.10 Configuration Tool not startable after Power Cut during Configuration Restore
Symptom If a power cut occurs during restoring a configuration in the NX Configuration Tool,
the Configuration tool cannot be started anymore after the reboot. An Application
Error is displayed.
Cause Cutting of the power during the restore process is in progress leaves the database in
an inconsistent state.
Solution Reset the configuration data to factory settings and repeat the restore procedure:
(1) Go to: Start Agfa NX Service Backup Restore Restore
Configuration Data
(2) Follow the instructions in the DOS prompt.
(3) Repeat the restore procedure in the NX Configuration Tool.
1.2.11 Test Connection does not work after AE Title or Port Number Change of the
MUSICA MCE Engine
Symptom After adapting the AETitle or Port Number in the configuration of the MUSICA MCE
Engine as an Archive (Refer to Mammo Solution Manual chapter 3.04 - Appendices -
Configuration of the CR Mammography Musica MCE Image Processing Software) the
Test Connection does not work.
Note: Normally it is not required to change the default AE title or port number.
Changing the port number might be required if e.g. a port number conflict occurs (port
is also used for another application).
Cause The MUSICA MCE Engine must be restarted to start using the adapted AETitle or
Port Number
Solution For NX 2.0.8300 / 3.0.8300: Do not change the port number. This is not
supported.
For NX ≥ 2.0.8400 / 3.0.8400: Activate the configuration after change of AE-title
and port number. After this the test connection of the
archive will work.
1.2.12 Images from Digitizer arrive but cannot be modified: All Buttons grayed out
Symptom The images from the digitizer (DICOM digitizer DX-G, DX-M, CR 30-X 2nd) arrive
but cannot be modified: All Buttons are greyed out.
Cause In the “Workstation settings” in the configuration tool the “Performed Station AE
Title” is set equal to the “AE Title” for Fast Preview. In this case only the fast preview
images arrive on the NX.
Solution For DICOM digitizer DX-G, DX-M and CR 30-X 2nd: Give different names for
“Performed Station AE Title” and “AE Title” for fast Preview.
As of NX 2.0.8600 / 3.0.8600 following text is displayed in configuration tool menu
General Settings > Workstation settings:
(see red marked text in Figure 4)
Figure 4
Reference HQ_1011090001
Symptom Following error is displayed if starting the CCM (Configuration and Customization
Manager) tool on a 64 bit Windows operating system:
The version of this file is not compatible with the version of
Windows you're running. Check your computer's system
information to see whether you need an x86(32-bit) or x64 (64-
bit) version of the program, and then contact the software
publisher.
Cause CCM tool not compatible with Windows 7, 64 bit
Solution Install the CCM tool on 32 bit operating systems only.
1.3 Upgrades
Symptom Upon an upgrade from NX 1.0 (SU1) to a higher NX version, it is possible that the
system is initially a bit slower because it has a lot of clean-up work to do.
Cause On some sites, the cleanup in NX 1.0 (SU1) fails to work because of duplicate
date/time combinations.
Solution After one day, performance is back to normal.
The clean-up can be monitored in the 'closed exam' screen where the number of
sessions decreases.
Symptom After upgrade of NX 1.0.3203 to , when logging in to the NX workstation as user with
operator rights. The NX application doesn't start up and the error message 278 is
shown.
Cause This error happens sporadically and is probably related to corrupted
Regional Settings data.
Solution Workaround 1:
(1) Log in to the defective user account.
(2) Open Regional Settings Panel.
(3) Select in the language drop-down any language, other than the intended
language.
(4) Select in the language drop-down the intended language for the user.
(5) Click <Apply>.
(6) Click <OK>.
It is required to select another language to force windows to correct the underlying
data for the correct language.
Workaround 2:
Remove and recreate the user with operator rights on Windows level.
Reference HQ_0706080013
Symptom The shortcut to "Activate XRDI logging" is not available in the Start menu.
Cause The XRDI (X-Ray Device Integration) installer adds the following item to the start
menu "Activate XRDI logging". When activating the NX workstation, this item is
removed from the Start menu within the list of possible debug loggings.
Solution This logging can still be activated via the folder C:\Agfa\Healthcare\NX\Bin\Service.
Symptom When the ADB tool is in use, incorrect accession numbers are generated after the
upgrade to ≥ .
Cause If no accession number is present in the image sent from NX, the default values in
the ADB configuration file (database.txt) can cause incorrect accession numbers to
be generated.
Solution (1) Log in as “crservice”.
(2) Go to: C:Agfa\Healthcare\ADB
(3) Open the ADB configuration file database.txt in a text editor.
(4) Remove the lines that mention the attributes:
ACCESSION-NUMBER-PREFIX and ACCESSION-NUMBER-MAPPING
(5) Save and close the file: database.txt
NOTE:
This issue can be seen on all versions of NX, but from
on, the default settings in the ADB configuration file do not contain these
attributes. However when upgrading an NX workstation to , the old settings
are copied.
1.3.5 Configuration Data Import Message: Exception caught during import [Padding is
invalid and cannot be removed.]
1.3.6 Recovering Data after failed automatic Import during Software Upgrade
Symptom The upgrade installer wizard or the windows registry ImportAfterUpgradeInfo informs
about a failed automatic import of data.
Cause The automatic import of data was not completed successfully.
Solution Recover the system and import the data manually according to the failed import type.
IMPORTANT:
Perform the following actions only if the respective data import has failed:
Symptom When you load the license file before upgrading to NX 2.0.8900 / 3.0.8900 following
error message appears: “License Mgr Exception: There is an error in XML document”.
Cause License Manger of software < NX 2.0.8800 / 3.0.8800 not compatible with license file
for ≥ NX 2.0.8800 / 3.0.8800.
Solution Do not load the license file before starting the upgrade, but load it during activation. For
details refer to NX Service Manual chapter 7, Upgrade Installation.
1.3.8 Failure to start Licensing Service and to import Configuration, Image and Dose
Data during Software Upgrade starting from NX 2.0.8000 / 3.0.8000
Symptom When performing a software upgrade starting from NX 2.0.8000 / 3.0.8000 following
software issue may appear:
NX Activation Utility Helper brings up messages:
- “Unable to connect to the Agfa Licensing service !”
- “ActivationHelper failed to install additional software” and only the close button is
enabled
After clicking the close button, the Activation Utility continuous.
At the end of the installation, failure to import configuration, image and dose data is
reported (Reason: Cannot start service Agfa.Licensing on computer).
Figure 5
1.3.9 NX upgrade or NX installation fails with SQLServer 2014 SP2 error message
Symptom The installation of SQLServer 2014 SP2 fails sporadically, causing the NX installation
to stop with message that not all dependencies could be installed.
Cause Software issue.
Solution Restart the NX installation.
For upgrades from below NX 3.0.8950 reboot the PC before applying the update.
1.4 RIS
1.4.1 RIS Query may become slow after Upgrade from NX 1.0
Symptom This issue only occurs on networks that do not use a DNS server and consequently
have the IP addresses configured in the Device Configuration of the Configuration
Tool:
When performing a RIS worklist query on a , it may take longer than on NX 1.0
systems. Independent from the number of worklist items, the query time is longer by
20-30 seconds.
Cause Normally, DNS queries should only be done for hostnames, not for IP addresses.
performs a DNS query although the IP address is configured for the RIS.
Result: The DNS query fails as there is no DNS server (this action takes about 20 -
30 seconds). After the DNS query will use the configured IP address and the
connection succeeds. NX1.0 did not show this behavior.
The issue is linked to the installed version of the .NET framework that is delivered
with the NX Software.
Solution (1) The local hosts file of the must be updated with IP addresses and hostnames
of the RIS/PACS.
(2) The hostnames of the RIS/PACS must be used in the Device Configuration of
the Configuration Tool instead of the IP-addresses.
For the detailed procedure refer to Service Manual chapter 04 “Installation and
Configuration”, section “Configuring Devices”.
Reference HQ_0707230004
Symptom RIS Query results in MESS27. This mainly happens if automatic query is configured
in the NX and several RIS devices are queried.
Cause A new RIS query request was received while a previous one was still busy (due to a
slow RIS or slow network),
Solution Workaround: Close the error message and wait for the RIS query which is still
running to finalize and return its results.
Symptom The RIS option type “User program” is configured. When clicking the button
<RIS query> in the NX user interface to create the worklist, the user program does
not start up.
Cause The “User program” (.exe file) cannot be started from the NX application
(<RIS query> button).
Solution (1) Create a batch file somewhere on the NX that calls the executable file (exe) of
the RIS user program.
(2) Open the Configuration Tool and go to:
User Interface Configuration Configure External Applications
(3) Check the checkbox: Worklist Environment
(4) Enter a caption for the button that is related to the function of the
user program.
(5) Set the path to the batch file by clicking the <Browse> button.
(6) In the Configuration Tool go to: Device Configuration RIS
(7) Change the option in the drop-down menu from User Program to Data File.
(8) Check the checkbox: Remove DataFile
(9) If the path to the Data file is not entered yet, click the button <Browse> and
navigate to the data file that is generated by the User program.
(10) Activate the Configuration.
Test following procedure and inform the customer about the new procedure to query
the RIS:
(1) In the NX application Worklist window click the new button.
The user program is started.
(2) Perform the tasks requested from the user program (e.g. choose patient, enter
accession number, etc.). The user program generates the data file.
(3) Click the button <RIS Query>. The NX will fetch the data file and enter the
data into the Worklist.
Reference HQ_0610050006, HQ_0701310002
1.4.4 Accession Number cannot be edited after Image has been received
Symptom After the image has been received, it is not possible to modify the accession
number.
Cause This is normal behavior if MPPS is used.
The issue also appears if MPPS is not used on site but configured.
Solution Use the Configuration Tool to remove the configuration of the MPPS if it is not used
at the site.
Symptom When receiving the RIS list, the values in TAG 0010,2201 (Species Description) and
the TAG 0010,2292 (Breed Description) are correct. But when the values are shown
at the worklist or closed exams lists the values have changed place
Cause Software issue.
Solution In NX 2.0.8800 / 3.0.8800 SU1 the issue is solved. Note, that it is required to undo
the RIS mapping, if this was done to solve the issue in NX 2.0.8800 / 3.0.8800 (i.e. if
BreedDescription was explicitly mapped onto SpeciesDescription and vice versa).
1.5 Printing/Archiving
Symptom When configuring NX to send 12 Bit OD (optical density) image data to the printer,
the printed image quality is not optimal.
Cause Some printers don't support the default NX pixel representation of '1' for this output
type, e.g. Drystar 3000.
Solution Configure NX to send 8 Bit image data.
Symptom True size prints on Drystar 3000 are not true size.
Cause Drystar 3000 requires installation of PMS-file for true size printing from NX.
Solution Refer to the Drystar 3000 Service Documentation for installing the PMS-file.
Symptom After finishing a print job cycle, the log sometimes shows:
1.5.4 Wrong Characters on PACS (e.g. ' º ', ' cm² ')
Symptom Special characters may not be visible on the PACS, or may be displayed wrongly, or
may even cause the rest of a text to be cut off.
Cause For annotations the NX uses international symbols such as ° (degree sign), cm², ....
In some character sets this character is not available and thus cannot be translated
to DICOM.
Solution Configure the NX system to use extended character sets and combine the character
set to be used with ISO 2022 IR 6 (DICOM default repertoire, which contains these
characters).
An alternative is to use UTF-8 encoding (Unicode) ISO_IR 197, which also contains
these characters. Check whether the connected PACS system also supports this.
If one of the solutions described above are used, no information will be lost in the
GSPS when sending it to the PACS. It is still possible that the PACS viewer is using
a font that cannot display a certain character!
Symptom The Windows Level (W/L) settings of images on the PACS are not in a ratio of
“50% - 50%”. This can result in:
• A very bright and less in contrast image appearance on the PACS compared to
the appearance before the upgrade.
• A very bright and less in contrast image appearance of normal Mammography
images compared to Micro Calcification Enhancement (MCE) images on the
PACS.
Cause Enabling the license NX MUSICA2 VOILUT WINDOWLEVEL in combination with
PACS archives that do not support or follow the DICOM standard can lead to the
described symptoms regarding the image appearance on the PACS.
Solution Workaround:
(1) Check the Connectivity Release documents of the PACS and of the
NX 3.0.9000 for mismatching DICOM settings.
(2) If the PACS cannot be setup to support and follow the DICOM standard,
disable the NX MUSICA2 VOILUT WINDOWLEVEL license in the NX
Configuration Tool.
Reference HQ_1003010001, HQ_1003260002
Symptom Secure Dicom communication stops working (failing print/archive jobs). There are
different causes for this fault:
Cause 1 The public certificate of the NX has expired on the external Dicom device with which
the NX needs to communicate over SSL or has been removed accidentally. Refer to
section Fault Finding sub-section 4.14 “Check installed Certificates”.
Solution 1 Get a new certificate from a CA (Certification Authority) for the NX device and
upload it to the external DICOM device.
Cause 2 The public certificate of the external device with which the NX needs to
communicate over SSL has expired on the NX. Refer to section Fault Finding sub-
section 4.14 "Check installed Certificates”.
Solution 2 Get a new certificate from a CA (Certification Authority) for the external device and
install it on the NX (see chapter 5 Appendices, section ‘Secure Communication').
Symptom Corrupted NX database after power failure: The NX application does not start up anymore.
In the Windows Event logging (part of the NX System Snapshot or accessible via Control
Panel > Administrative Tools > Event Viewer) an error message is displayed indicating a
severe problem with the NX database "ADC_QSdat.mdf".
Example error messages:
• SQL Server detected a logical consistency-based I/O error: incorrect checksum
(expected: 0xdfe952; actual: 0x1846eef2). It occurred during a read of page
(1:38305) in database ID 5 at offset 0x00000012b42000 in file
'C:\Agfa\Healthcare\NX\Configuration\SQLServerExpress\ADC_QSdat.mdf'.
Additional messages in the SQL Server error log or system event log may provide
more detail. This is a severe error condition that threatens database integrity and
must be corrected immediately. Complete a full database consistency check (DBCC
CHECKDB). This error can be caused by many factors; for more information, see
SQL Server Books Online.
• File ADC_QSdat.mdf appears to have been truncated by the operating system.
Cause The connected UPS is not configured properly or the user has shut down the system
without proper shutdown procedure.
Symptom When adding markers or annotations, a double cursor is seen (a copy of the correct
one but a little grayer). This is seen in the Editing environment and in the
Examination environment.
Cause Only applicable to Dell 670 with Matrox cards: The video card settings are wrong.
Solution (1) Go to Display Properties Settings and click <Advanced>.
(2) Set the Hardware Acceleration to the 3rd notch from the left
(Disable all DirectDraw and Direct3D Accelerations, etc.).
Symptom Scale factor is not shown in the status box when NX is configured.
Cause This happens if:
• "True Scale" is configured in the Device Configuration Printer Sheet Text
Box and
• "Show true scale factor" is checked in the User Interface Editing Settings
Configure Normal Mode & Print Mode.
This can happen on a workstation with Dutch language settings because the default
label text is too long.
Solution Change the label from "Schaal aanpass..." to "Schaal:".
1.7.5 Fast Preview Image is not shown: Error Messages about AE Title are displayed
Symptom After digitizing an exposure, the Fast Preview image is not shown. Following error
messages are displayed:
Figure 6
Cause A Fast Preview AE Title length of 16 characters generates this error.
Solution Workaround: Use Fast Preview AE Titles shorter than 16 characters in the
CCM Tool and the Configuration Tool.
Reference HQ_0710220006
Symptom Draw two (or more) collimation areas (any shape), invert all of them and apply
collimation.
Cause The collimation areas are drawn as if they were not inverted, but with an extra
border added.
Solution Only a single collimation area can be inverted.
Symptom The NX CMS is locking up when switching between thumbnails. The lock-up may
take half a minute or longer.
Cause Network adapter setting "Link speed & duplex" not set to "Auto Negotiation" but e.g.
to 100 Mbps Full Duplex at CMS and NX.
Solution Set parameter "Link speed & duplex" at all NX and CMS network adapters to "Auto
Negotiation".
Reference HQ_1004080003
1.7.9 After initiating a Shutdown of the NX Workstation the System keeps in Shut-
Down State
Symptom • After initiating a shutdown of the NX workstation, the system keeps in shut-
down state: The background is black and one can move the mouse. It is
possible only to switch-off the NX by holding the on/off switch on the PC front
panel long enough
or
• The NX workstation does not perform the shutdown or log-off command but
remains operational.
Cause Interaction with wrong LanSafe software.
Solution (1) De-install the LanSafe software that came with the UPS.
(2) Install the Lansafe software which is part of DVD2 of the corresponding NX
release: This is the NX compatible and tested Lansafe software.
Reference HQ_1009070003
1.8 Exporting/Importing
Symptom When exporting Dose Monitoring statistics, the export file has no default
filename extension.
Cause Software issue
Solution The export file contains data in XML format, so the user should explicitly provide a
filename with the extension ".xml".
Symptom DICOM Export fails with error message "Unknown error occurred during export."
Cause This happens when McAfee Anti-Virus v8.0 is installed because it defines a security
rule: Prevent remote creation of autorun.inf files. This means copying a file called
autorun.inf to a mapped drive is forbidden.
This is part of the DICOM export process.
Solution (1) Go to McAfee VirusScanConsole Access Protection File Share
Folder Protection.
(2) Disable the security rule Prevent remote creation of autorun.inf files.
Symptom Intermittently the DVD burning process started via the NX user interface stops and
the error message “Unknown error RefMess 163” is displayed.
Cause DVDs with insufficient quality have been used.
Solution Only use high-quality DVD media. For a recommendation of high-quality DVDs refer
to: http://www.digitalfaq.com/reviews/dvd-media.htm
Reference HQ_0910270001
Symptom Exporting DICOM images fails: No images on the DVD, despite the NX shows
“Done” at the end of the export process.
Cause Issue with DVD-RW media after a DVD-RW has been formatted.
Solution To prevent the fault observe following recommendations:
- Use of recordable media is preferred over rewriteable media
- DVD+R/RW is preferred over DVD-R/RW
- To reuse a DVD-RW, always choose "Format Disk" instead of "Erase Disk"
1.9 Licenses
Symptom When running the License Manager Tool, this error is displayed:
Figure 7
As a consequence of this error, the indication of remaining days in the Grace Period
may be wrong.
Cause Software issue
Solution Reboot the NX Workstation.
Symptom The user does not constantly use the application and removes the licensing dongle
for the time the application is not used.
The licenses for the application falls into status: Grace
Figure 8
Cause The Licensing service is automatically started as soon as the PC is started. Since
there is no dongle detected, the licenses fall into status Grace.
Solution A checkbox has been created in the License Manager tool to toggle between
enabled and disabled grace functionality.
Communicate following procedure to users who want to temporarily remove the
licensing dongle from the NX workstation:
Figure 9
Figure 10
NOTE:
When reconnecting the dongle, follow the procedure in reversed order:
(1) Reconnect the dongle.
(2) Check the checkbox: Enable Grace Functionality.
Reference HQ_0811200002
1.9.3 Error about “License service is not running” when going back to previous
Windows Restore Point
Symptom When going back to a previous Windows restore point, in some cases the licensing
service cannot be started.
NX does not start-up, but gives error message: License service is not running.
Please start the service.
When trying to start the licensing service manually, following error message is
displayed: “Windows could not start the Agfa License Service on local computer.
Error 1053: The service did not respond to the start or control request in a timely
fashion.”
Figure 11
Figure 12
Figure 13
Symptom NX does not start-up anymore after reboot of the PC. The error message displays
“internal error”.
Cause System clock set back to default (e.g. 1.1.1970) because battery on motherboard is
empty.
Solution Replace battery on motherboard. Consider to replace the PC. Refer to the existing
NX hardware upgrade programs.
1.9.5 NX Workstation in grace period after the Network Adapter name has been
renamed
Symptom After renaming the network adapter from default “Local Area Connection” for
example to “Connection to Hospital” the NX Workstation is in grace period.
The issue appears in NX 3.0.8950 and may also appear in NX 3.0.9000
Cause As of NX 3.0.8950 the software looks for the network adapter name "Local Area
Connection XYZ" to determine the MAC Address of the network adapter on the PC's
motherboard. In case of a different name the licensing service fails to retrieve the
MAC address of that adapter.
Solution When renaming the network adapter hosted by the motherboard make sure its name
starts with “Local Area Connection”, for example “Local Area Connection to hospital”.
Reference PRB0054680; NX 3.0.9000 SB01, Document ID 57096072.
NOTE:
In case several network adapters are installed on NX it can be useful to rename the
adapters to meaningful names. This makes it easy to identify the purpose of the network
adapter. The network adapter can be renamed in menu Control Panel > Network and
Sharing Center > Change Adapter Settings.
The adapter belonging to the motherboard is usually the first in the list. This can be double-
checked by executing "ipconfig /all" in a command window and checking the IP address
assignment for that adapter.
Symptom After creating a new configuration file from scratch on the Offline Config Tool, this
error is listed when trying to save the file:
Figure 14
Cause Licenses are read from the database representation on the Offline Config Tool
system. Only when the file is deployed on an NX Workstation, the licenses are read
from the ALF.
Solution Workaround: Manually disable one of the licenses in the Offline Config Tool before
saving the configuration file.
Symptom Error message when viewing priors using Web1000: "Unable to log in: An unknown
error occurred, check the local log".
Cause Limitation on number of logged in users (max. 50), when using EPR demo license
on Web1000.
Solution Retry when less users are logged in.
1.11.2 Additional Internet Explorer Windows are displayed when retrieving Priors
in Web1000
Symptom When retrieving prior images from IMPAX using Web1000, additional
Internet Explorer windows are temporarily appearing.
Cause Normal behaviour
Solution None, the windows can be closed without disturbing the operation.
Symptom After retrieving the studies from IMPAX and transferring them to WEB1000, the prior
list is not automatically updated. Refreshing the Internet Explorer screen will end in a
"fatal exception" error.
Cause Normal behaviour
Solution (1) Close the Internet Explorer window.
(2) Press the <Priors> button again.
1.12 Monitors
Symptom After installing the drivers for a Barco diagnostic monitor, following error message is
displayed:
Figure 15
Figure 16
WARNING:
Electrical device. Electric shock possible.
Always switch off the equipment and unplug the power cord, before
disconnecting or connecting cables on printed circuit boards.
Reference HQ_0902230002
Symptom After installation the Barco MFCD1219 Touchscreen does not work.
Cause The wrong driver is installed. During installation of the driver the Windows dialog
New Hardware found is displayed. If multiple drivers are found, the wrong driver may
be highlighted by default.
Solution (1) Go to: Computer Management System Tools Device Manager Mice
and other pointing devices
(2) Uninstall the installed touchscreen driver.
(3) Install the correct driver according to the procedure in chapter 5 Monitor
Options.
1.13 Networking
Symptom The NX CMS is locking up when switching between thumbnails. The lock-up may
take half a minute or longer.
Cause Network adapter setting "Link speed & duplex" not set to "Auto Negotiation" but e.g.
to 100 Mbps Full Duplex at CMS and NX.
Solution Set parameter "Link speed & duplex" at all NX and CMS network adapters to "Auto
Negotiation".
Reference HQ_1004080003
Symptom A CMS is active in the LAN network. The DX-D100 starts up when LAN and WLAN
are connected. The CMS reports the following error: "The recieved in room station
IP address differs from retrieved address by in room hostname".
There is no connection possible CMS to NX.
Cause The NX gives the IP address of the WLAN istead of the LAN to the CMS when CMS
sends out his broadcast.
Solution Workaround to solve the problem: Disable the wireless connection at the NX in the
DX-D 100 and reboot the NX: Now the CMS can connect to the NX.
Symptom Case 1: No network communication with a DICOM digitizer, if two network interfaces
are used in NX (2nd network interface for connection to DR detector in a mixed
CR/DR system)
OR
Case 2: No network connection to a DICOM digitizer if used in a point-to-point
network connection (direct network cable connection between digitizer and NX).
Cause Case 1: Wrong assignment of IP addresses.
Case 2: In a point-to-point network connection with the digitizer, the NX PC has no
activate network connection at boot time if the digitizer is not switched on. Windows
reports the LocalHost 127.0.0.1 address as the system’s IP Address.
As a result, in both cases, the NXWorkStation service is listening on the wrong IP
address for SYNC messages coming from the digitizer.
Solution Following solution is available in NX < NX 3.0.8950 only (as of NX 3.0.8950 the
configuration of a fixed IP address is not required anymore): Set a fixed IP address
for the NX workstation in the Config Tool > General Settings >
Workstation Settings > IP address > Advanced ... window.
In case of a point-to-point network connection with the digitizer: Configure this fixed
IP address also in the Windows network properties of the NX.
Symptom Mammograms mostly of big breasts sporadically show black burn areas in the skin
line of the breast. Images as of NX 2008 (NX 3.0.7000) are affected, where
examinations are processed with package "Mammo" or "Mammo T2" and the
license NX_MAMMO_DARKENING is enabled.
This issue is visible on the NX workstation monitor, and in exceptional cases also
on hardcopy prints and on the softcopy monitor.
Cause Already saturated (overexposed) pixels e.g. in the skin line of big breasts, will be
highlighted when using “background darkening” function when examinations are
processed with the "Mammo" or "Mammo T2" package.
This issue will be even more highlighted if the Window/Level function is used. For
more information about the cause of this, read the following background
information:
Background The image processing for background darkening of a mammography image is
Information based on two techniques:
• If no Window/Level is applied or if the Window/Level is adjusted slightly, the
pixel padding range technique defines a range of pixel values containing the
pixel values of the background regions in the image.
• If Window/Level is applied extensively, the background region is defined by a
bitmask. I.e. a mask that indicates for every pixel in the image whether the pixel
is part of the background region or not.
If the image contains clipped pixels in the diagnostic area (e.g. the breast edge),
these pixels may be darkened together with the background when Window/Level is
adjusted slightly. This becomes visible if the image is inverted.
This behavior has no impact on diagnosis.
Solution Instruct the customer not to combine Window/Level with background darkening.
The NX user manual is adapted accordingly.
Reference HQ_0911050001
1.14.2 NX User Manual does not give Instructions how to proceed if "Autocollimation"
fails with MUSICA 2
Symptom Musica 2 image processing ("Autocollimation") may fail rendering very dark images.
With Musica 1 users were told to apply "Manual Collimation" to recover good image
quality, but this does not work with Musica 2.
Cause User Manual does not give instruction how to proceed if "Autocollimation" fails.
Solution The workaround to apply manual W/L to recover good image contrast is described in
the user manual for NX 2.0.8400 / 3.0.8400 (and higher versions).
Reference HQ_0907170001
NOTE:
• Musica pre-processing failures should rarely occur. However, it was of utmost
importance to foresee how the system will react on this, because otherwise the
system would come into a state where it constantly retries to acquire the image.
• In NX, a message will pop up on the screen asking the operator whether or not to
retry the image-processing, prior to actually refusing the image. If the issue is
caused by a temporal problem, the retry can work and the image will not be
refused. In case the issue is data-related, every subsequent retry will fail
(resulting in the refusal of the image).
Question 3 How to trace the CR Cassette/Detector an Image has been exposed on?
Answer If there are doubts about the origin of artifacts on an image, it is useful to determine
the respective CR cassette/detector, the image has been exposed on:
(1) Select the image in the Image Overview of the Main Menu in the NX Viewer.
(2) Click the link View All Image Attributes.
(3) Find the label Detector ID and note it down.
(4) For CR Cassettes:
Compare the Detector ID
with the Cassette Number
of the CR cassettes and
find the matching one.
The Cassette Number of
the CR cassettes can be
read out using the RF Tag
Initialization Tool:
Figure 17
NOTE:
For detailed information about the RF Tag Initialization Tool refer to the
Service Manual chapter 5 Using the RF Tag Initialization Tool.
NOTE:
For detailed information about the pane View All Image Attributes refer
to the NX User manual on NX Startekit DVD 1.
Result The CR cassette/detector that contains the matching Cassette Number/IP serial
number is the CR cassette/detector the image has been exposed on.
NOTE:
Do not try to clone a system using NX backup and restore. Instead, it is advised to use
the Configuration Tool:
NOTE:
The ‘clone’ system will only look ‘identical’ to the master system in case that on both
systems the same licenses are available. Also the authorization of the logged in user
can cause the system to look differently (certain options are disabled).
Example:‘Print’ view was configured on the master system ‘Normal’ view is shown on
the cloned system, because there is no Print license available on the machine.
In case of port conflicts, typically the NXWorkstation service will not start up, and when
opening the log file NX.Application.NonVisual.NXWorkstation.exe.log the error:
‘java.BindException’ can be found.
It is possible that after configuring NX one or more of its services are not able to open a
DICOM port for listening. This means that another program has taken control over one
of these ports, for example a RIS client.
TCP/IP port conflicts on the local NX PC can be caused by:
• Standard Windows applications such as Remote desktop.
• 3rd party applications (e.g. RIS) installed on the NX PC.
NOTE:
Many applications assign ports dynamically.
This does normally not cause a problem because the NX backend starts up when the
system starts up. But when started up from the Configuration Tool (or when another
application is using the same fixed port), a conflict can occur.
NX has a listening port for Fast–ID, storage SCP and storage/commit SCP, so it is
advised to check whether no other applications are claiming any of these ports.
Following simple DOS commands can help to find the conflicts:
(1) Go to Start Programs Accessories Command Prompt to open a DOS
command window.
(2) Type ‘netstat /a’ in a command window to see the list of ports in use.
If running this prior to running NX (make sure that the NX workstation service is
not running), the free ports on the system become visible:
Figure 19
(3) Type NetStat /a /o into a command window to display the process ID (PID)
next to each port that is used:
(4) Find out which process is using the ID in the Windows task manager:
The Image storage capacity depends on preferences, such as the set number of days
an image will be kept on the system.
NOTE:
• Sessions are deleted from the system if they are older than the configured
maximum number of days or if room on the disk is low, meaning less space
than 2 GB.
• Only closed sessions for which printing or archiving is done, not just requested,
can be deleted automatically.
Sessions cannot be deleted automatically if:
• They are locked.
• They are younger than 24 hours.
Manual cleanup may be necessary for sessions that do not properly end printing or
archiving (e.g. pending storage commitment) or that are younger than 24 hours in case
disk size is very critical.
IMPORTANT:
On manual cleanup:
• Pending jobs will be removed from the queues.
• Active jobs will be continued.
The maximum number of days to keep sessions can be configured in the
NX Configuration Tool Workflow Management Identification-Examination
Storage pane.
Windows Internet Explorer is started by running the Windows Explorer (file browser)
and typing an http URL (http://...) in the address bar.
NOTE:
When scheduling the automatic reboot, it should be set to a time when the NX will not
be used, e.g. during the night.
IMPORTANT:
Be aware, that it might be impossible to schedule this automatic reboot in emergency
departments where the system needs to be running at any moment.
Figure 22
Figure 23
Figure 24
(7) Give the task an appropriate name (e.g. “ShutdownNX”) and select the
frequency of the shutdown (Weekly).
Figure 25
(9) Select the start time (= time when the system is preferably not used or
rarely used).
Figure 26
(10) Finally, enter the user name and password under which this task may run.
Click <Next>:
Figure 27
NOTE:
• The task will run under this profile independent of the fact that
somebody is logged in or not. Preferably, use an administrator
account to run this task (e.g. crservice). Note that this does not
need to be a local account. In a domain setup, also a domain
account can be used. After the reboot, a user with log-in rights
must be present to log in.
• In case of password expiration, the task will cease to run.
(11) Check Open advanced properties for this task when I click Finish and
click <Finish>.
Figure 28
Figure 29
The “About NX” utility creates a “snapshot” of the currently installed system and lists all
version related information:
• The installed NX version
• Dose Type
• NX Usage Information (Disk space)
• The installed Hotfix / Documentation CD version
• The installed languages
(which were installed using the language CDs)
• NX Component versions
• Installed licenses
• XRDi version information
This tool can be accessed using: Start Agfa NX Service About NX …
(1) Search the NXWorkstation log files for text FirmwareInfo. For a detailed
description refer to section “Fault Finding” sub section “View Logging”.
(2) Find out the version number by the text similar to the following example:
1007 ID-TABLET 1.12.07 06/17/05 00005054 1.11.13
Explanation:
• 1.12.07 represents the version of the firmware.
• 5054 is the serial number of the reader.
• 1.11.13 represents the version of the driver on the PC.
Figure 30
(5) Locate the Mac adress in the license file. Do this by searching for the string
HardwareKey (in the XML section ‘FingerprintFile’). The Mac address is the
Value attribute of this XML tag.
Figure 31
(6) Locate the dongle serial number in the license file. Do this by searching for the
string ‘Dongle’ (in the XML section ‘FingerprintFile’). The dongle serial number is
the Serial number attribute of this XML tag.
(7) Locate the installed software version in the license file. Do this by searching for
the string InstalledSwComponent. The version number consists of three
numbers: X, Y and Z-version
(8) Now the information has to match:
• The Mac address of the machine to the Mac address' in the license file.
• The dongle serial number to the dongle serial number in the license file.
• The installed software version to the installed software version in the
license file.
NOTE:
If a license file is loaded and the system goes into grace as a result of the loading,
the user is warned and can cancel the operation via the following warning dialog:
Figure 32
In some situations sessions cannot be deleted using the Delete Examination function in
the Main Menu (cleanup date/time issues, exam cannot get into the 'closed'
status, etc).
Any Examination (open or closed) can be deleted by a Service user via the
Configuration Tool:
NOTE:
Make sure there are no jobs in the output queues for the session to be deleted.
These jobs will become faulty after the sessions is deleted.
Figure 33
(7) Select an examination and click <Delete>.
(8) If the NX software is running, a confirmation is requested to stop the
NX software.
(9) The examination is deleted.
(10) Restart the system.
NOTE:
Do not select Logoff Screen Saver!
(4) Set Wait time: Determines the time of non-activity before the screen saver
is activated.
(5) Set On resume, password protect: The user needs to enter his password to
unlock the screen saver.
(6) Click <OK>.
The service mode allows the service engineer to preserve the data as it was acquired
from the digitizer (without any processing). This can be enabled and disabled per
digitizer.
This is a feature that can be used when there is doubt that the digitizer is sending the
correct information to the NX workstation (can be added as an attachment in the
problem tracking system for example).
When enabled, a DICOM file is generated in the folder:
• D:\Agfa\Healthcare\NX\DataFiles\DicomFiles in case it is a DICOM digitizer.
• An xml file with the demographic data and the raw pixel data are stored in the folder
D:\Agfa\Healthcare\NX\DataFiles\TwainProcessedFiles for Twain digitizers (e.g.
DX-S). In a DICOM file, both are available in one file. The names of the files match
the SOP instance UID (unique identifier) of the image in NX.
To enable or disable the service mode, use the following procedure:
(1) Go to:
Start Agfa NX Service Install tools Set digitizer service mode
(2) All configured digitizers are listed.
(3) Select the digitizer for which the service mode needs to be changed.
Press 0 for cancelling this task.
Result The changes will be picked up when the NX backend service is restarted.
NOTE:
When the service mode is enabled for a digitizer, all pixel data will be stored twice on
the data harddisk. The service engineer is responsible to clean up the data when ready
with it. This cleanup happens automatically when disabling the service mode.
The service mode allows the service engineer to preserve the data as it was sent from
the NX Workstation to the Dicom archive. This can be enabled and disabled per
archive.
This is a feature that can be used when there is doubt that the NX is sending the
correct information to the Dicom archive (can be added as an attachment in the
problem tracking system for example). These Dicom objects can also be viewed using
any DICOM Viewer for further analysis or to compare them with what is visible on the
PACS.
When enabled, one or two (in case that the GSPS option was enabled for the archive)
DICOM file are generated in the ArchiveEnhancedServiceModeStorage folder per
archive job. The names of the files match the SOP instance UID (unique identifier) of
the image in NX. To enable or disable the service mode, use the following procedure:
(1) Go to:
Start Agfa NX Service Install tools Set archiving service mode.
(2) All configured archives are listed.
(3) Select the archive for which the service mode has to be changed. Press 0 for
cancelling this task.
Result The changes will be picked up when the NX backend service is restarted.
NOTE:
When the service mode is enabled for an archive, all pixel data will be stored twice on
the data harddisk. The service engineer is responsible to clean up the data when
he/she is ready with it. This cleanup happens automatically when disabling the
service mode.
3.13 Saving the Adjustments of Musica 1 Settings (via NX GUI) to the Backup
File
In the NX application (GUI), the Musica 1 settings can be adjusted and saved as
default. If these settings should be included into the backup configuration file, so that
they will not have to be repeated after a restore, follow these instructions:
On NX systems with RAID functionality, the system remains operational if one of both
disks fails. The user receives a warning that the RAID volume is degraded and should
inform Service to replace the failing disk. After replacing the disk, the RAID must be
activated again.
(1) Change the BOOT ORDER after replacing the disk:
• On a DELL PC:
During booting press <F12> to enter the BIOS setup.
Select the RAID Volume (e.g. Intel Volume_000) as primary boot media.
• Or on an HP PC:
Press <F10> during booting to enter the BIOS setup.
Go to Storage and select Boot Order.
Set the RAID Volume (e.g. Intel Volume_000) as primary boot media.
Press <F10> to accept changes.
Select Save Changes and Exit and press <F10> to accept.
Figure 34
(3) Wait until Windows is booted and log in using the “crservice” account.
For all PCs with RAID, except HP 8300 and HP rp5800:
Intel Matrix storage manager is informing about the fact that the RAID is
degraded an may be rebuild.
Figure 35
Figure 36
(4) Go to Start All Programs > Intel Matrix Storage Manager Intel Matrix
Storage Console to launch the Intel Storage Console.
Figure 37
Figure 38
Figure 39
IMPORTANT:
Do not perform a reboot of the system until the migration process has
finished completely.
NOTE:
The computer can be used normally during the migration process.
If there are symptoms for a corrupted database (See also section 1.7.1) proceed as
follows to recover from this error:
IMPORTANT:
This procedure can only be used at software NX 2.0.8700 / 3.0.8700 (and higher
versions if available).
For software versions < NX 2.0.8700 / 3.0.8700:
• Omit steps (2), (4) and (6): These functions are only available for
software ≥ NX 2.0.8700 / 3.0.8700.
• Step (5) is relevant only, if a Windows scheduled tasks exists to backup all data.
(1) A power failure may have caused corruption of the hard disk. Use the Windows
tooling (Windows7: http://www.w7forums.com/use-chkdsk-check-disk-t448.html ,
Windows XP: http://support.microsoft.com/kb/156571) to repair the disk.
(2) If the symptom for database corruption is the occurrence of error messages in
the NXWorkStation log file reporting database login failures of the user adcqs:
Repair the database by using the shortcut “ReSynchronize NX Configuration”.
Note, that in software versions < NX 2.0.8700 / 3.0.8700 using the function
"Restore NX configuration" always restores the configuration backup which was
made previously compared to the actual selection.
(4) Use the “Agfa > NX > Service > Backup Restore > Database Repair Utility”
shortcut. This utility provides some basic repair functions at SQL level:
• Check the database consistency
• Repair the database consistency without data loss
• Repair the database consistency allowing data loss
(5) Use the “Agfa > NX > Service > Backup Restore > Restore All Data” shortcut.
Note, that all changes since the latest backup will be lost.
As for NX 2.0.8700 / 3.0.8700 a backup in hourly rate is automatically created,
the maximum of data loss will be the data added the last hour.
(6) Use the “Agfa > NX > Service > Backup Restore > Recreate Database and
Restore All Data” shortcut to recreate the database and restore its content using
the latest backup.
Note, that all changes since the latest backup will be lost.
As for NX 2.0.8700 / 3.0.8700 a backup in hourly rate is automatically created,
the maximum of data loss will be the data added the last hour.
(7) Use the “Agfa > NX > Service > Install Tools > Reset NX Completely to Factory
Settings” shortcut to re-initialize the database.
Using this script results in a loss of all configuration, image and dose data.
• Configuration Data
• Image Data
• Dose Statistic Data
Be aware of the following best practices using the automated data import/export:
IMPORTANT:
Do not interrupt the upgrade installation.
If the amount of image data that needs to be preserved is very limited compared to the
total set on the system, it is not advised to use the automated export/import. Use the
manual export/import procedures instead.
The procedures are described in the NX 3.0.9000 Service Manual, chapter 05 Clean
Installation.
• Make sure all sessions are closed prior to starting the upgrade.
• Make sure there are no pending jobs prior to starting the upgrade.
If one of the exports fails a dialogue is shown reporting the outcome of each of the
exports and offering the possibility to cancel the upgrade process. In this case:
3.16.3 Import of the Configuration Data fails due to Verification Errors in the Data
If one the import of the configuration data fails due to verification errors in the data:
If after the upgrade one or more imports of image data have been interrupted, e.g.
because the system was switched off during import:
(1) Manually move (not copy and paste) the existing preprocessed pixel files
from the folder:
D:\Agfa\Healthcare\NX\DataFiles\PreprocessedPixelFiles
To the folder:
D:\Agfa\Healthcare\NX\DataFiles\UpgradeData\PreprocessedPixelFiles
(2) Run the NX Activation Utility to reset the system to factory defaults:
Start Agfa NX Service Install Tools NX NX Activation Utility
(3) Browse to the folder: C:\Agfa\Healthcare\NX\Bin
(4) Double click the file:
NX.Application.Visual.CommonTools.ImportAfterUpgrade.exe
An application opens.
(5) Use this application to restart the import of the data.
If after the upgrade, the images appear as red crosses in the NX Viewer (and reporting
about missing pixel files), it can be that the associated pixel files have not been moved
back to their original folder.
To repair this situation, use the Windows Explorer to manually move the pixel files from
the folder:
D:\Agfa\Healthcare\NX\DataFiles\UpgradeData\PreprocessedPixelFiles
To the folder:
D:\Agfa\Healthcare\NX\DataFiles\PreprocessedPixelFiles
4 Fault Finding
(1) Select START > All programs > Agfa > NX > Service > NX System
Snapshot
(2) Recommendation: Leave everything on default. For details to the different
settings refer to table next page.
Click <Make Snapshot>
(3) Wait until the snapshot is created. This takes a few minutes.
(4) Copy the snapshot from C:\Agfa\HealthCare\NX\SystemSnapshot to a
USB flash drive and forward it to the next support level.
NOTE:
If the NX System Snapshot is too large for sending via email, create a temporary FTP
account for file transfer. Refer to following document on the Agfa HealthCare Library:
How to get a temporary AGFA FTP account for uploading files,
Document ID 34721327.
NOTE:
The NX System Snapshot GUI shown in Figure 42 was introduced together with
software version NX 2.0.8900 / 3.0.8900. For more information to the NX System
Snapshot see also the Application Note - Functional Changes in NX 2.0.8900 -
3.0.8900, Document ID 47340162.
Figure 40
1 Deselect the options in the “snapshot actions” pane only, if instructed by the
support center.
The system snapshot tool for the user (CTRL-ALT-I) facilitate the user to export native
images and a limited system snapshot at the time of an incident.
This can be picked up later by the FSE.
Figure 41
Figure 42
When passing the NX logfiles to the RSN or GSN, inform about which attention key you
pressed with which event. Following attention keys are possible:
• Return key
• Figures 0 to 9. Do not use the numerical keypad to enter digits. Entering digits are
logged with a “D” in front of the digit. Example: CTRL-ALT-SHIFT-1 results in entry
“Warning #ATTENTION: User has pressed attention key
combination CTRL-ALT-SHIFT-D1”
• Characters A to Z
• F1 to F12 key
CAUTION:
Violation of patient privacy possible!
Since the patient privacy information is human readable, it must be treated as
confidential information and protected from unauthorized access.
IMPORTANT:
Do not alter the security settings of the logfiles, since the application needs to be able
to write to them.
For using the AGFA Healthcare Log Viewer read the CX.Logging.Viewer manual.
Access the AGFA Healthcare log viewer’s documentation by selecting Help User
manual in the Log viewer’s menu.
Logging is available on a per-process basis. For NX, logging is available for the
following processes:
• NX Workstation Windows service (NX backend, where most of the business logic is
running): NX.Application.NonVisual.NXWorkstation.exe.log
• Configuration Tool: NX.Application.Visual.Configuration.ConfigTool.exe.log
• QC/ID Viewer (NX viewer): NX.Application.Visual.QCID.exe.log
(1) Start up the logging viewer by clicking on Logging viewer item in the start
menu: Start Agfa NX Service Logging
(2) Select File new log4net reader. Browse to the location where the log
files are located (C:\agfa\healthcare\log\NX\xml).
(3) Select a file (see list of files above).
The logging components always log timestamps in GMT (Greenwich Mean Time)
format. Upon displaying log entries, the logging viewer will use the time zone of the PC.
This is not a problem in case the log file was created on a PC which is configured in the
same time zone.
When viewing a log file of another time zone the local time of the log entries might be
interesting. E.g. a log file entry created in the UK at 9 AM would be displayed on a
system located in Paris (GMT+1 time zone) as 10 AM.
To override this behavior follow this procedure:
In order to understand the logging information, the meaning of the namespaces need to
be known. The table below lists some of the most important namespaces.
Namespace Description
General
Agfa.Healthcare.NX.Business.CompositeInsta Fast Preview communication (high level,
nce.FastPreview.Facade abstracting from the protocol used =>
Common to Compact –family digitizers and
Stratus-family digitizers (e.g. DX-S).
com.agfa.med.DICOM.facility.DTFLogger When set to DEBUG, a very detailed
DICOM log is being produced, both for
MPPS, Dicom RIS communication and
archiving.
Agfa.Healthcare.NX.Data All database related actions and actions
pertaining structured data transfer.
Agfa.Healthcare.NX.Business.ImageAcquisitio Core of the system, including state
n.SessionManagement.QCIDSessionFacade management.
Specific to printing
Agfa.Healthcare.NX.Business.Printing Printing
Specific to Compact – family digitizers
Agfa.Healthcare.NX.Business.CompositeInsta Image acquisition for Compact-family
nce.Dicom.DicomStorageSCPGateway digitizers
Agfa.Healthcare.NX.Business.CompositeInsta Fast ID communication with DICOM
nce.DirectIDGateway digitizers (Compact family)
Agfa.Healthcare.NX.Business.CompositeInsta Fast preview communication for DICOM
nce.Dicom.DicomFastPreviewSCPGateway digitizers (Compact family) – low level
Specific to Stratus – family digitizers
Agfa.Healthcare.NX.Business.CompositeInsta Image acquisition using a DX-S, including
nce.Twain.Gateway low-level fast preview communication.
Specific to ID Tablet
Agfa.Healthcare.NX.Business.CompositeInsta ID Tablet communication
nce.IDTabletGateway
Specific to archiving
Agfa.Healthcare.NX.Business.CompositeInsta DICOM communication for archiving
nce.Dicom.DicomStorageSCUGateway
Agfa.Healthcare.NX.Business.Archiving High-level archiving communication
NOTE:
Apart from ‘Activate Standard logging’, all logging batch files that are available will
activate DEBUG-level logging for certain components.
IMPORTANT:
When debug logging is enabled, this can have a serious impact on the performance of
the NX application.
It is the intention that it is only used for a limited amount of time during the fault finding
process. When finished with troubleshooting switch back to the standard log
configuration.
Images can be exported and imported on another (local) NX workstation for further
problem analysis.
NOTE:
Not all information related to the image is exported. Some data on session-level will be
missing and rejected images will be visible as normal images after export and import.
In order to validate and test DICOM interfaces the DICOM Validation Tool (DVT) will be
used. This tool provides extensive logging information. DVT is programmed using a
simple scripting language and can be instructed to play either SCU or SCP.
Besides it can be used as a Storage SCP emulator, Print SCP emulator or to validate
Media Stored Files.
The service engineer can consult the operational statistics or infocounters. A total and
a relative counter have to be kept which is incremented each time the specified event
occurs.
Figure 43
The relative counter value can be reset to zero by the service engineer, normally after
a service intervention. The absolute counter value cannot be reset.
(1) To reset a counter, select the counter and right-click to select the Reset counter
option from the context menu. Enter the user name and the reason for the
intervention.
Figure 44
Figure 45
The ELMS also allows downloading the most recently generated license key file or
fingerprint file of each system that can be viewed using the License Manager Tool.
The database icon in the task bar indicates the state of the SQL Server database
service. Green indicates that it is properly working.
Note that when logged in as a normal user, the database icon in the taskbar may be
white. This does not mean the database is not running.
The NX health monitor icon in the task bar indicates the health of the NX backend
application.
Green means that the NX backend application is up and running and properly
initialized.
When it stays red, starting the NX Viewer will end up with an error. In order to find out
the details of the error state look at the log files.
The following procedure allows checking the presence and content of the installed
certificates. Note that this is only relevant in case SSL has been configured.
(1) Log in with a Windows account with administrator privileges (If possible, with the
same account that used to install the certificates, e.g. CRService).
(2) Start the certificate management tool:
Start Agfa NX Service Install Tools Manage Cerificates
(3) Open up the ‘Personal’ and ‘Trusted Root Authorities’ folders in the Certificate
management tool and check the presence of the required certificates.
(4) Double-click on the certificate to check the content of a certificate
(e.g. expiration date).
Figure 46
(5) Double-click on the certificate in the ‘Personal’ folder, select the ‘Details’ tab and
select the ‘Subject’ item to check the common name of the NX certificate (which
needs to be configured in the NX Configuration tool as well).
The text next to ‘CN = ‘ is the common name of the certificate and needs to match
the common name in the NX Configuration tool / Security section.
Figure 47