34450A Manual
34450A Manual
Multimeter
User’s Guide
Notices U.S. Government Rights Warranty
The Software is “commercial computer THE MATERIAL CONTAINED IN THIS
Copyright Notice software,” as defined by Federal DOCUMENT IS PROVIDED “AS IS,”
Acquisition Regulation (“FAR”) 2.101. AND IS SUBJECT TO BEING
© Keysight Technologies 2012-2020 Pursuant to FAR 12.212 and 27.405-3 CHANGED, WITHOUT NOTICE, IN
No part of this manual may be and Department of Defense FAR FUTURE EDITIONS. FURTHER, TO THE
reproduced in any form or by any Supplement (“DFARS”) 227.7202, the MAXIMUM EXTENT PERMITTED BY
means (including electronic storage U.S. government acquires commercial APPLICABLE LAW, KEYSIGHT
and retrieval or translation into a computer software under the same DISCLAIMS ALL WARRANTIES, EITHER
foreign language) without prior terms by which the software is EXPRESS OR IMPLIED, WITH REGARD
agreement and written consent from customarily provided to the public. TO THIS MANUAL AND ANY
Keysight Technologies as governed by INFORMATION CONTAINED HEREIN,
Accordingly, Keysight provides the
United States and international INCLUDING BUT NOT LIMITED TO THE
Software to U.S. government
copyright laws. IMPLIED WARRANTIES OF
customers under its standard MERCHANTABILITY AND FITNESS FOR
commercial license, which is embodied
Trademarks in its End User License Agreement
A PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS
Microsoft ® is a U.S. registered (EULA), a copy of which can be found OR FOR INCIDENTAL OR
trademark of Microsoft Corporation. at http://www.keysight.com/find/ CONSEQUENTIAL DAMAGES IN
sweula. The license set forth in the CONNECTION WITH THE
Manual Part Number EULA represents the exclusive authority FURNISHING, USE, OR
by which the U.S. government may use, PERFORMANCE OF THIS DOCUMENT
34450-90000
modify, distribute, or disclose the OR OF ANY INFORMATION
Edition Software. The EULA and the license set CONTAINED HEREIN. SHOULD
forth therein, does not require or KEYSIGHT AND THE USER HAVE A
Edition 7, March 2020 permit, among other things, that SEPARATE WRITTEN AGREEMENT
Keysight: (1) Furnish technical WITH WARRANTY TERMS COVERING
Printed in: information related to commercial THE MATERIAL IN THIS DOCUMENT
Printed in Malaysia computer software or commercial THAT CONFLICT WITH THESE TERMS,
computer software documentation that THE WARRANTY TERMS IN THE
Published by: is not customarily provided to the SEPARATE AGREEMENT SHALL
public; or (2) Relinquish to, or CONTROL.
Keysight Technologies
Bayan Lepas Free Industrial Zone, otherwise provide, the government Safety Information
11900 Penang, Malaysia rights in excess of these rights
customarily provided to the public to
Technology Licenses use, modify, reproduce, release, CAUTION
perform, display, or disclose
The hardware and/or software A CAUTION notice denotes a hazard. It
commercial computer software or
described in this document are calls attention to an operating
commercial computer software
furnished under a license and may be procedure, practice, or the like that, if
documentation. No additional
used or copied only in accordance with not correctly performed or adhered to,
government requirements beyond
the terms of such license.
those set forth in the EULA shall apply, could result in damage to the product
Declaration of Conformity except to the extent that those terms, or loss of important data. Do not
rights, or licenses are explicitly required proceed beyond a CAUTION notice
Declarations of Conformity for this from all providers of commercial until the indicated conditions are fully
product and for other Keysight computer software pursuant to the FAR understood and met.
products may be downloaded from the and the DFARS and are set forth
Web. Go to http://www.keysight.com/ specifically in writing elsewhere in the
go/conformity. You can then search by
product number to find the latest
EULA. Keysight shall be under no
obligation to update, revise or
WARNING
Declaration of Conformity. otherwise modify the Software. With A WARNING notice denotes a hazard. It
respect to any technical data as calls attention to an operating
defined by FAR 2.101, pursuant to FAR procedure, practice, or the like that, if
12.211 and 27.404.2 and DFARS not correctly performed or adhered to,
227.7102, the U.S. government
could result in personal injury or death.
acquires no greater than Limited Rights
as defined in FAR 27.401 or DFAR Do not proceed beyond a WARNING
227.7103-5 (c), as applicable in any notice until the indicated conditions are
technical data. fully understood and met.
– Do not defeat the power cord safety ground feature. Plug in to a grounded
WARNING (earthed) outlet.
– Do not use the instrument in any manner that is not specified by the
manufacturer.
– To avoid electric shock or injury, do not operate the multimeter without
panels or case in place.
– Do not substitute parts or modify the instrument to avoid the danger of
introducing additional hazards. Return the instrument to Keysight
Technologies Sales and Service Office for service and repair to ensure the
safety features are maintained.
– Main Power and Test Input Disconnect: Unplug the instrument from the
wall outlet, remove the power cord, and remove all probes from all
terminals before servicing. Only qualified, service-trained personnel
should remove the cover from the instrument.
– Line and Current Protection Fuses: For continued protection against fire,
replace the line fuse and the current-protection fuse only with fuses of
the specified type and rating.
– IEC Measurement Category II. The HI and LO input terminals may be
connected to mains in IEC Category II installations for line voltages up to
300 VAC. To avoid the danger of electric shock, do not connect the inputs
to mains for line voltages above 300 VAC. See “IEC Measurement
Category II Overvoltage Protection” on the following page for further
information.
The Keysight 34450A 5½ Digit Multimeter complies with the following EMC and
NOTE safety requirements:
– IEC 61010-1:2001 / EN 61010-1:2001 (2nd Edition)
– IEC 61326-2-1:2005 / EN61326-2-1:2006
– CISPR 11:2003 / EN 55011:2007 Group 1 Class A
– Canada: ICES/NMB-001:Issue 4, June 2006
– Australia/New Zealand: AS/NZS CISPR 11:2004
Product category
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Operation
The Fine Tip, Mini Grabber, and SMT Grabber attachments plug onto the probe
end of the Test Leads.
Maintenance
If any portion of the Test Lead Set is worn or damaged, do not use. Replace with a
new Keysight 34138A Test Lead Set.
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Regulatory Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
General Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Waste Electrical and Electronic Equipment (WEEE) Directive
2002/96/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Product category . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Additional Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
3 Measurement Tutorial
DC Measurement Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . .94
Noise Rejection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .95
Measurement Speed Consideration . . . . . . . . . . . . . . . . . . . . . . . . . . . .98
Dual Measurement Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . .99
DC voltage dynamic range in dual measurement . . . . . . . . . . . . . . .99
Voltage and current in dual measurement . . . . . . . . . . . . . . . . . . .100
Resistance Measurement Considerations . . . . . . . . . . . . . . . . . . . . . .102
True RMS AC Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .105
Other Primary Measurement Functions . . . . . . . . . . . . . . . . . . . . . . . .109
Frequency measurement errors . . . . . . . . . . . . . . . . . . . . . . . . . . .109
DC current measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .110
Capacitance measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .111
Temperature measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113
Other Sources of Measurement Error . . . . . . . . . . . . . . . . . . . . . . . . .114
This chapter provides a tutorial on how to get started using the Keysight 34450A
5½ Digit Multimeter and using the front panel in order to make measurements.
19
1 Getting Started Tutorial
10 7 9 11
3 6 5 8
The system annunciators are described in Table 1-1. (See Table 2-2 on page 58
for Math Annunciators).
The keypad has been locked. Press the keys simultaneously for
more than 3 seconds to unlock
Error in queue
Triggering has been enabled and the meter is in the “waiting-for-trigger” state
Direct current
Alternating current
Press to access utility menu. See “Using the Utility Menu” on page 63
>
Press simultaneous for 3 seconds to lock and unlock the keypad
Measurement functions
> Press to access data log view menu. See “Viewing the log info” on page 86
Feature Upgrades
There are two licenses, listed in Table 1-3, which are available for purchase:
For the license upgrade procedure, refer to the instructions in the license
redemption e-mail.
3 4 6 10
1 2 5 7 7 8 9
Making Measurements
The following pages show how to make measurement connections and how to
select measurement functions from the front panel for each of the measurement
functions.
For remote operation, refer to the MEASure Subsystem in the Keysight 34450A
Online Programmer’s Reference Helpfile.
>
Digit masking
The navigation keypad provides a shortcut to mask (change the number of digits
displayed) the reading on the main display, easing readability. Masking digits only
affects what is displayed. It does not affect measurement speed or accuracy. It
applies to all functions except continuity, diode test, temperature, and
capacitance measurement. To enable masking, follow the instructions below:
O 1
> or
3 4
DC Voltage:
– Measurement Range: 100.000 mV, 1.00000 V, 10.0000 V, 100.000 V,
1000.00 V
– Speed: Slow, Medium, Fast
– Default Setting: Autoranging, Slow measurement speed
– Measurement Method: Sigma Delta A-to-D converter
– Input Impedance: >10 GΩ selected range (0.1 V and 1 V only) or ~10 MΩ all
ranges (typical)
– Input Protection: 1000 V on all ranges (HI terminal)
Figure 1-6 ACV rms and DCV terminal connection and display
Do not apply any voltage to the instrument inputs until all terminals are
WARNING properly connected. Plugging or unplugging the test lead while high voltage
is applied can cause instrument damage, and may increase the risk of
electric shock.
Measuring resistance
– Measurement Range: 100.000 Ω, 1.00000 kΩ, 10.0000 kΩ, 100.000 kΩ,
1.00000 MΩ, 10.0000 MΩ, 100.000 MΩ.
– Speed: Slow, Medium, Fast
– Default Setting: Autoranging, Slow measurement speed
– Measurement Method: 2-wire ohms or 4-wire ohms
– Input protection: 1000 V on all ranges (HI terminal)
1 2 3
I
1 2 3
I
Resistance
Press Twice
Figure 1-9 ACI rms or DCI (mA) terminal connection and display
Figure 1-10 ACI rms or DCI (A) terminal connection and display
1 2 4
Typical Frequency Measurement Display
Frequency
Source
1 2 4
Frequency 3
Source
Figure 1-12 Frequency terminal connection and display for ACI (mA)
1 2 4
Frequency 3
Source
Figure 1-13 Frequency terminal connection and display for ACI (A)
Testing continuity
– Measurement Method: 0.5 mA ± 0.2% constant current source
– Response Time: 165 samples/second with audible tone
– Continuity Threshold: 10 Ω fixed
– Input Protection: 1000 V (HI terminal)
Open or
Closed Circuit
Typical Closed Circuit Measurement Display
Checking diodes
– Measurement Method: Uses 0.5 mA ± 0.2% constant current source
– Response Time: 190 samples/ second with audible tone
– Input Protection: 1000 V (HI terminal)
1 2 3
I
Typical Forward-Biased Diode
Measurement Display
Forward
Bias
Press Twice
1 2 3
I
Typical Reverse-Biased Diode
Measurement Display
Reverse
Bias
Press Twice
Measuring temperature
– Measurement Range: –80.0 °C to 150.0 °C, –110.0 °F to 300.0 °F
– Measurement Method: 2-wire Ohms measurement of 5 kΩ thermistor sensor
(E2308A) with computed conversion
– Input Protection: 1000 V (HI terminal)
– Optional Accessory: E2308A Thermistor temperature probe
1 2 3
I
Measuring capacitance
– Measurement Range: 1.000 nF, 10.00 nF, 100.0 nF, 1.000 µF, 10.00 µF,
100.0 µF, 1.000 mF, 10.00 mF
– Default Setting: Autoranging
– Measurement Method: Computed from constant current source charge time.
Typical 0.12 V to 1.0 V AC signal level
– Input Protection: 1000 V (HI terminal)
1 2 3
Capacitance
Press Twice
Selecting a Range
You can let the multimeter automatically select the range using autoranging, or
you can select a fixed range using manual ranging. Autoranging is convenient
because the multimeter automatically selects the appropriate range for sensing
and displaying each measurement. However, manual ranging results in better
performance, since the multimeter does not have to determine which range to use
for each measurement.
1 2
5 6
>
Remote Operation
USB interface
1 2
3 4
5 6
To easily configure and verify an interface connection between the 34450A and
NOTE your PC, use the Automation-Ready CD, which is shipped with your 34450A. This
CD includes the Keysight IO Libraries Suite and the Keysight Connection Expert
application. For more information about Keysight's I/O connectivity software,
visit www.keysight.com/find/iolib.
Serial interface
The 5-pin male connector on the meter’s rear panel is a serial port or terminal to
form a minimal 3-wire RS-232 connection (TX, RX, GND).
In order to operate the multimeter via a host computer or terminal, the
parameters of the serial interface within the multimeter has to match the
parameters with the serial interface provided by the host or terminal.
The default settings of the meter are 9600 baud rate, non-parity, 8 data bits, and
1 stop bit (9600, n, 8, 1).
The connecting diagram and setup procedure are shown in Serial interface
connector diagram and “RS232 utility sub-menu” on page 70 below.
Instrument PC
Red 1 DCD
TX 1
2 RX
RS232 Protocol RX 2 3 TX
Yellow
GND 4 DTR
3
Grey 5 GND
TRIGGER IN 4 6 DSR
TRIGGER OUT 5 7 RTS
8 CTS
9 RI
Serial Serial DB9 DB9
Interface Interface Female Male
Male Female
SCPI commands
The Keysight 34450A complies with the syntax rules and conventions of SCPI
(Standard Commands for Programmable Instruments).
For a complete discussion of the 34450A SCPI syntax, refer to the Keysight
NOTE 34450A Programmer’s Reference Helpfile, This is provided in the 34450A
Product Reference CD-ROM
This chapter contains information on the functions and features of the Keysight
34450A 5½ Digit Multimeter and how to use the front panel to operate these
settings.
47
2 Functions and Features
Math Operations
Table 2-1 below describes the math operations that can be used with each
measurement function.
Null measurement
When making null measurements, also called relative measurement, each reading
is the difference between a stored null value and the input signal.
For example, this feature can be used to make more accurate resistance
measurements by nulling the test lead resistance.
Before performing null measurement, eliminate offset errors associated with test
lead resistance by following the steps below:
1 2
1Ω
3 4
5 6
1Ω
After you enable the null operation, the multimeter stores the next reading into
the offset register and immediately displays the null measurement:
Null measurement d isplay = Read ing – Offset
You can view and edit the offset value in the secondary display as described in
“Editing math functions reference values” on page 58.
Hold measurement
The hold feature allows you to capture and hold a stable reading on the front
panel display.
When a stable reading is detected, the multimeter emits a beep (if the buzzer is
enabled in the utility menu) and holds the reading on the primary display.
>
>
When enabled, the hold operation turns on the hold annunciator and begins
evaluating readings using the rules described below:
Primary Display = Read ingN IF Max() – Min() ≤ 0.1% × Read ingN
The decision to update a new reading value in the primary display is based upon
the box-car moving statistics of the present reading and the three previous
readings:
Max (Read ingN Read ingN–1 Read ingN–2 Read ingN–3)
Min (Read ingN Read ingN–1 Read ingN–2 Read ingN–3)
Limit measurement
The limit operation allows you to perform pass/fail testing against specified upper
and lower limits.
>
Press to edit Choose the limit mode that you want to change
5 6-a
1 2
1 2
Press to start
Select STATS (SINGLE) option
3
Statistical values
Present measurement
4 5 6
1 2
Press to start
Select STATS (ALL)
Note: Each time a new min or max value is stored, the multimeter will beep once (if the buzzer is
turned on in the utility menu) and the New annunciator is briefly turned on for 1 second in the
respective Min or Max box
Press to exit
Editing dB measurement
When enabled, the dB operation computes the dBm value for the next reading,
stores the dBm result into the dB Ref register, and immediately produces the
following calculation. The first displayed reading is always precisely 000.00 dB.
dB = 10 × Log10 [(Read ing2/RREF)/0.001 W]–dB Ref
1 2
Press to edit
Select dB option
Ongoing measurement
dB measurement
Reference value
0dB to ±120.0000 dBm
Default: R REF is 0 dBm
Press to exit
You can view and edit the dB reference value as described in “Editing math
functions reference values” on page 58.
1 2
Press to start
Ongoing measurement
dBm measurement
Reference value
Press to exit
You can view and select the reference value as described in “Editing math
functions reference values” on page 58.
Math annunciators
Table 2-2 below shows the possible math annunciators which can appear on the
display and the editable values.
Editing values
For math functions with editable values, the label “PRESS MATH TO EDIT” will be
shown at the bottom left of the display.
To edit math values, follow the steps below:
1 2
3 4-a
4-b
Press to edit
Press to change the value of selected digit
5-a 5-b
1 2
3 4
7-a 8
Select desired function
for the secondary display
>
7-b
Select desired range Press to turn off secondary display
1 2 3
6 7
Note : Press to turn off secondary display.
>
– Make sure you select the correct input terminal according to the input
WARNING range used.
– Do not apply current exceeding specified range to input terminals of mA
or A.
1 2 3
4 5
6 7
Note : Press to turn off secondary display.
>
Below are the steps you should follow if you want to edit any of the values on the
utility menu:
1 2
>
Press to edit
5 6 7
1 2
Press to edit
Select I/O option
3 4 5
6 7 8
1 2
Press to edit
Select I/O option
3 4 5
6 7 8
1 2
>
Typical display with error message Press to display the utility menu
3 4
5 6 7
The beeper
Normally, the multimeter beeps whenever certain conditions are met (for
example, the multimeter beeps when a stable reading is captured in reading hold
mode).
The beeper is factory set to ON, but may be disabled or enabled manually.
– Turning OFF the beeper does NOT disable the front panel keys beep tones.
– A beep tone is always emitted (even with the beep state turned OFF) in the
following cases:
– A continuity measurement is less than or equal to the continuity threshold.
– A SYSTem:BEEPer command is sent.
– An error is generated.
– In addition to the beep operations just described, when the beeper is ON, a
single beep occurs for the following cases (turning the beeper OFF disables
the beep for the following cases):
– When a new MIN or MAX value is stored
– When a new stable reading is updated on display for Math Hold operation
– When a measurement exceeds the HI or LO Limit value
– When a forward-biased diode is measured in the Diode function
1 2
6 7 8
Press to save the state into Display shows state is stored Move cursor to the location
selected memory location label at the recall box
9 10 11 12
For remote operation, refer to the MEMory Subsystem, the *SAV, and *RCL
commands in the Keysight 34450A Programmer’s Reference Helpfile.
Reset/Power-On State
The table below summarizes the 34450A settings as received from the factory,
following power cycling, and following the *RST command received over the USB
remote interface. Non-volatile, user customizable behavioral differences are
shown in BOLD.
Immediate Triggering
The immediate triggering mode is available from the remote interface only.
In the immediate trigger mode, the trigger signal is always present. When you
place the multimeter in the wait-for-trigger state, the trigger is issued
immediately. This is the default trigger source for remote interface operation.
– Remote Interface Operation: The following command selects the immediate
trigger source:
TRIGger:SOURce IMMediate
The CONFigure and MEASure? commands automatically set the trigger source
to IMMediate.
Refer to the Keysight 34450A Programmer’s Reference Helpfile for complete
description and syntax for these commands.
External Trigger
External triggering takes one reading (or the specified number of readings in data
logger) each time the multimeter receives a pulse on the rear-panel external
trigger connector.
The multimeter uses the rising edge (POS) of the external trigger signal to trigger
a reading. Diagram below shows the external trigger connector:
TRIG TRIG
Tx Rx IN OUT
– +
Figure 2-5 Trigger in connector
The trigger annunciator turns on when the multimeter is waiting for an external
trigger.
The rear panel trigger out connector provides a pulse after the completion of each
measurement. Trigger out and external trigger implement a standard hardware
handshake sequence between measurement and switching devices.
Trigger Out
The Trigger out signal cannot be configured and is implemented in four ways:
– In local mode, a trigger out signal is sent whenever a measurement is updated
on the front panel.
– In remote mode, a trigger out signal is sent whenever a measurement is taken
by the user through using command.
– In data log/external trigger mode, a trigger out signal is sent whenever a
measurement is log/triggered and updated on the front panel.
– In code compatibility mode, a trigger out signal is sent whenever a
measurement is updated on the front panel or a measurement is taken by the
user using command.The diagram below shows the trigger out connector:
TRIG TRIG
Tx Rx IN OUT
– +
Figure 2-6 Trigger out connector
Manual Trigger
Manual triggering takes one reading (or the specified number of readings in data
logger) each time the multimeter receives a pulse on the front-panel trigger
button. The diagram below shows the manual trigger operation:
The trigger annunciator turns on when the multimeter is waiting for a manual
trigger. The rear panel trigger out connector provides a pulse after the completion
of each measurement. Trigger out and manual trigger implement a standard
hardware handshake sequence between measurement and switching devices.
Data Logging
The data logger feature provides a front-panel interface that allows you to set up
data logging into the instrument’s non-volatile memory with programming, and
without a connection to a computer.
Once you have finished collecting data, you can view it from the front panel, or
you can also connect your computer and import the data using the DATA:DATA?
NVMEM command.
The instrument is connected through the enable log configuration and will begin
to log measurement data once an external pulse is received through the terminal
or the trigger button is pressed. Once data logging is enabled, all IO connectivity
will be disabled. The IO connectivity will be restored when the data logging has
been completed or aborted.
The 34450A has memory for up to 50,000 readings which is the maximum limit for
the data logger feature.
1 2
Press to enable data log menu Typical data log menu display
3 4 5
Move cursor to enable logging Press to execute Note: Enabling new log data will
automatically delete previous log
data.
6 7
or
Press the trigger button (manual triggering) or wait for an external pulse (external
triggering) to begin data logging
To edit the trigger delay, sample interval, sample count, and trigger count in the
data log menu, follow the steps below:
1 2
Press to enable
data log menu
Typical data log menu display
3 4-a
1 2
>
Press to display data log menu Typical log view menu display
3 4
5-a 6
1 2
>
Press to display data log menu Typical log view menu display
3 4 5
6 7 8
Press to switch pages Typical log list first page display when both Press to find specific
primary and secondary measurements are log data
performed during data log
9 10-a
10-b
Finding specific log data page display Press to change the value of selected digit
11 12
1 2
>
3 4
7
Typical histogram display for primary measurement with
secondary measurement active
Press to exit
1 2
Press to display data log menu Typical log view menu display
3 4
Press to select
Move cursor to the log statistics menu
5 6
Press to exit
1 2
>
4 5
3 Measurement Tutorial
DC Measurement Considerations 94
Noise Rejection 95
Measurement Speed Consideration 98
Dual Measurement Considerations 99
Resistance Measurement Considerations 102
True RMS AC Measurements 105
Other Primary Measurement Functions 109
Other Sources of Measurement Error 114
93
3 Measurement Tutorial
DC Measurement Considerations
Thermal EMF Errors
Thermoelectric voltages are the most common source of error in low–level DC
voltage measurements. Thermoelectric voltages are generated when you make
circuit connections using dissimilar metals at different temperatures. Each
metal-to-metal junction forms a thermocouple, which generates a voltage
proportional to the junction temperature. You should take the necessary
precautions to minimize thermocouple voltages and temperature variations in
low-level voltage measurements. The best connections are formed using
copper-to-copper crimped connections, as the multimeter’s input terminals are a
copper alloy. The table below shows common thermoelectric voltages for
connections between dissimilar metals.
Table 3-1 Common thermoelectric voltages for connections between dissimilar metals
Copper -to- Approx. mV / °C
Cadmium-Tin Solder 0.2
Copper <0.3
Gold 0.5
Silver 0.5
Brass 3
Beryllium Copper 5
Aluminum 5
Tin-Lead Solder 5
Kovar or Alloy 42 40
Silicon 500
Copper-Oxide 1000
Noise Rejection
Rejecting Power-Line Noise Voltages
A desirable characteristic of integrating analog-to-digital (A/D) converters is their
ability to reject power-line related noise present with DC input signals. This is
called normal mode noise rejection, or NMR. The multimeter achieves NMR by
measuring the average DC input by “integrating” it over a fixed period.
V f = F loat Voltage
R s = DUT Sour ce Resi stance
HI
I mbalance
R i = Multi meter I solati on Resi stance
V test Ideal (L O-E ar th)
Meter C i = Multi meter I nput Capaci tance:
Rs
Vf × R s
LO E r ror (v) =
Rs + R i
Vf Ci Ri
The best way to eliminate ground loops is to isolate the multimeter from earth by
not grounding the input terminals. If the multimeter must be earth-referenced,
connect it and the device-under-test to the same common ground point. Also
connect the multimeter and the device-under-test to the same electrical outlet
whenever possible.
1.2 V
0V
-1.2 V
100 kΩ 10 µA 10 µW
1 MΩ 1 µA 1 µW
10 MΩ 100 nA 100 nW
100 MΩ 100 nA / 10 MΩ 1 µW
Offset Compensation
A resistance measurement involves measuring a voltage (E) induced across the
resistance by a known current source.
R
Ideal
Meter I E
Thermal EMF caused by dissimilar metals can create a parasitic voltage in the
measurement circuit (VEMF). The thermal EMF can be caused by the input lead
connections or internally in resistor R. In general, this voltage will not change with
the current applied to the resistor.
VEMF VEMF
Ideal Ideal
Meter E=IR+VEMF Meter E=(0*R)+VEMF = V EMF
I=0 R I=0 R
The voltage measured, and so the resistance calculated, is in error by VEMF. Using
offset compensation can reduce the errors caused by VEMF. To make an offset
compensated measurement, the meter makes two voltage measurements, one
with the current source on and one with the current source off, and subtracts the
two measurements. The actual voltage drop across the resistor, and the
calculated resistance are obtained by:
First Reading -Second Reading = (I*R+VEMF) -VEMF = I*R
Offset compensation can be used in 2-Wire or 4-Wire ohms measurements (only
available on 100 Ω, 1 kΩ and 10 kΩ).
An AC-coupled true rms measurement is desirable when you are measuring small
AC signals in the presence of large DC offsets. For example, this situation is
common when measuring AC ripple present on DC power supplies. There are
situations, however, where you might want to know the AC+DC true RMS value.
You can determine this value by combining results from DC and AC
measurements, as shown below:
For the best AC noise rejection, you should perform the DC measurement at
s-mode.
Traditionally, digital multimeters include a crest factor derating table that applies
at all frequencies. The measurement algorithm used in the 34450A multimeter is
not inherently sensitive to crest factor, so no such derating is necessary. With this
multimeter, as discussed in the previous section, the focal issue is high-frequency
signal content which exceeds the multimeter’s bandwidth.
For periodic signals, the combination of crest factor and repetition rate can
suggest the amount of high-frequency content and associated measurement
error. The first zero crossing of a simple pulse occurs at
1
f1 =
tp
This gives an immediate impression of the high-frequency content by identifying
where this crossing occurs as a function of crest factor: f = CF 2 ⋅ prf
1
Table 3-3 below shows the typical errors for various pulse waveforms as a function
of input pulse frequency:
Table 3-3 Typical errors for various pulse waveforms as a function of input pulse frequency
Typical error for square wave, triangular wave, and pulse trains of CF=3, 5, or 10
prf
Square wave Triangle wave CF=3 CF=5 CF=10
200 –0.02% 0.00% –0.04% –0.09% –0.34%
1000 –0.07% 0.00% –0.18% –0.44% –1.71%
2000 –0.14% 0.00% –0.34% –0.88% –3.52%
The table above gives an additional error for each waveform, to be added to the
value from the accuracy table provided in Chapter 4, “Characteristics and
Specifications”.
Example: A pulse train with level 1 Vrms, is measured on the 1 V range. It has
pulse heights of 3 V (that is, a Crest Factor of 3) and duration 111 µs. The prf can
be calculated to be 1000 Hz, as follows:
Thus, from the table above, this AC waveform can be measured with 0.18 percent
additional error.
AC Filter
The multimeter’s AC voltage and AC current functions implement three
low-frequency comb filters. These filters allow you to trade-off minimum
measured frequency for faster reading speed. The “SLOW” mode filter notch at
2 Hz and useful for frequencies above 20 Hz. The “MEDIUM” filter notch at 20 Hz
and useful for frequencies above 200 Hz. The “FAST” filter notch at 200 Hz and
useful for frequencies above 1 kHz.
DC current measurements
When you connect the multimeter in series with a test circuit to measure current,
a measurement error is introduced. The error is caused by the multimeter’s series
burden voltage. A voltage is developed across the wiring resistance and current
shunt resistance of the multimeter, as shown below.
If you apply more than 5 A during current measurement, self-heating occurs in the
multimeter’s 10 A shunt resistor and internal signal-conditioning component. Do
allow few minutes stabling time for better accuracy for current measurement.
After applying more than 5 A current measurement, do allow a few minutes for
heat dissipation to ensure better accuracy for other measurement.
Capacitance measurements
The multimeter implements capacitance measurements by applying a known
current to the capacitor as shown below:
C offset
C RP C offset
V charged C
R'
1 µF 1 µA 0.25/second
10 µF 10 µA 0.25/second
100 µF 100 µA 0.25/second
1 mF 500 µA 0.25/second
10 mF 1 mA 0.15/second
The values of capacitance and loss resistance measured with the multimeter may
differ from the values measured using an LCR meter. This is to be expected, since
this is essentially a DC measurement method, while LCR measurement uses
applied frequencies anywhere from 100 Hz to 100 kHz. In most cases, neither
method measures the capacitor at its exact frequency of application.
For the best accuracy, take a zero null measurement with open probes, to null out
the test lead capacitance, before connecting the probes across the capacitor to be
measured.
Temperature measurements
The multimeter measures temperature by measuring the temperature sensitive
resistance of 5 kΩ thermistors.
Thermistors consist of semiconductor materials and provide roughly 10 times the
sensitivity of an RTD. Because they are semiconductors, their temperature range
is more limited, commonly to –80 °C to 150 °C. Thermistors have highly
non-linear, temperature-resistance relationships; therefore their conversion
algorithms are more complex. Keysight multimeters use the standard
Hart-Steinhart Approximation to provide accurate conversions.
R s = Source Resistance
F = Input Frequency
C in = Input Capacitance (100 pF ) Plus Cable
Capacitance
1
- f ( x ) dx
T
T
where f(x) is the function representing the signal waveform over a period of T.
Errors may occur when the pulse signal is measured at low voltage range due to
saturation of the multimeter’s analog-to-digital (ADC) rail voltage.
4 Characteristics and
Specifications
For the characteristics and specifications of the 34450A 5½ Digit Multimeter, refer
to the datasheet at
http://literature.cdn.keysight.com/litweb/pdf/5991-1133EN.pdf.
117
4 Characteristics and Specifications
*34450-90000*
34450-90000
www.keysight.com