Lampara Hen Reichert-Xcel-255-User-Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 24

XCEL™ 255

Slit Lamp

User’s Guide
©2013 AMETEK, Inc.

Reichert and Reichert Technologies are registered trademarks of Reichert, Inc.

Xcel is a trademark of Reichert, Inc.

AMETEK is a registered trademark of AMETEK, Inc.

All other trademarks are property of their respective owners.

The information contained in this document was accurate at time of publication. Specifications subject to change without notice. Reichert,
Inc. reserves the right to make changes in the product described in this manual without notice and without incorporating those changes in any
products already sold.

ISO 9001/13485 Certified – Reichert products are designed and manufactured under quality processes meeting
ISO 9001/13485 requirements.

No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means, electronic, mechanical,
recording, or otherwise, without the prior written permission of Reichert, Inc.

Caution: Federal law restricts this device to sale by or on the order of a licensed physician. Rx only.

2 15120-101 Rev. D
Table of Contents
Warnings and Cautions.................................................................................... 4
Symbol Information........................................................................................... 6
Introduction....................................................................................................... 7
Indications for Use...................................................................................... 7
Contraindications........................................................................................ 7
Setup ............................................................................................................. 8
Parts Identification...................................................................................... 8
Xcel 255 Package Contents....................................................................... 8
Accessories................................................................................................ 8
Unpacking & Installation............................................................................. 9
Application of Input Power........................................................................ 11
Disconnection of Input Power................................................................... 11
Instructions for Use......................................................................................... 12
Operation.................................................................................................. 12
Slit Length................................................................................................. 13
Filter Dial................................................................................................... 13
Slit Rotation.............................................................................................. 13
Cleaning & Maintenance................................................................................ 14
External Cleaning..................................................................................... 14
Forehead / Chinrest Preparation.............................................................. 14
Cleaning the Glide Plate........................................................................... 14
Changing the Halogen Bulb...................................................................... 14
Fuse Replacement.................................................................................... 15
Troubleshooting.............................................................................................. 16
Chart of Common Errors........................................................................... 16
Specifications................................................................................................. 17
Disposal.................................................................................................... 18
Software Revision..................................................................................... 18
Guidance Tables............................................................................................. 19
Warranty......................................................................................................... 23

15120-101 Rev. D 3
Warnings & Cautions
Reichert Technologies (Reichert) is not responsible for the safety and reliability of this instrument when:
• Assembly, disassembly, repair, or modification is made by unauthorized dealers or persons.
• Instrument is not used in accordance with this User’s Guide.

WARNING: AN INSTRUCTION THAT DRAWS ATTENTION TO RISK OF INJURY OR DEATH.

WARNING: United States federal law and European regulations require that this
device be purchased only by a physician or a person acting on behalf of a physician.

WARNING: This instrument should be used in strict accordance with the instructions
outlined in this user’s guide. The safety of the operator and the performance of
the instrument cannot be guaranteed if used in a manner not specified by Reichert
technologies.

WARNING: do not REPAIR OR SERVICE This instrument without authorization from


the manufacturer. ANY REPAIR OR SERVICE TO This instrument MUST BE PERFORMED BY
EXPERIENCED PERSONNEL OR DEALERS WHO ARE TRAINED BY REICHERT or serious INJURY TO
THE OPERATOR OR PATIENT MAY OCCUR.

WARNING: Modifications to this instrument are not allowed. Any modification to this
unit must be authorized by Reichert or serious INJURY TO THE OPERATOR OR PATIENT
MAY OCCUR.

Warning: If this instrument is modified, appropriate inspection and testing must be


conducted to ensure continued safe use of this instrument.

WARNING: to avoid risk of electric shock, this equipment must only be connected to
a supply mains with protective earth OR DAMAGE TO THis instrument AND/OR INJURY TO
THE OPERATOR OR PATIENT MAY OCCUR.

WARNING: ensure that the voltage applied to the unit is the same as the voltage that
is indicated on the data plate or damage to the unit may occur.

WARNING: THis INSTRUMENT MUST BE PLUGGED INTO AN OUTLET WITH AN EARTH GROUND. DO
NOT REMOVE OR DEFEAT THE EARTH GROUND CONNECTION ON POWER INPUT CONNECTOR
OR THE UNIT’S POWER CORD of THis instrument OR DAMAGE TO it AND/OR INJURY TO THE
OPERATOR OR PATIENT MAY OCCUR.

Warning: THE EQUIPMENT OR SYSTEM SHOULD NOT BE USED ADJACENT TO OR STACKED WITH
OTHER EQUIPMENT AND THAT IF ADJACENT OR STACKED USE IS NECESSARY, THE EQUIPMENT OR
SYSTEM SHOULD BE OBSERVED TO VERIFY NORMAL OPERATION IN THE CONFIGURATION IN WHICH
IT WILL BE USED.

WARNING: THIS INSTRUMENT IS NOT SUITABLE FOR USE IN THE PRESENCE OF FLAMMABLE
ANESTHETIC MIXTURES, SUCH AS OXYGEN OR NITROUS OXIDE.

WARNING: Because prolonged intense light exposure can damage the retina, the
use of the device for ocular examination should not be unnecessarily prolonged,
and the brightness setting should not exceed what is needed to provide clear
visualization of the target structures. This device should be used with filters that
eliminate UV radiation <400 nm) and, whenever possible, filters that eliminate short-
wavelength blue light <420 nm).

Warning: The use of accessories or cables other than those specified, with the
exception of those sold by the manufacturer as replacement parts for the internal
components, may result in increased emissions or decreased immunity of the
equipment or system.

4 15120-101 Rev. D
Warnings & Cautions (continued)
CAUTION: AN INSTRUCTION THAT DRAWS ATTENTION TO THE RISK OF DAMAGE TO THE PRODUCT.

Caution: THE INTERNAL CIRCUITRY OF THE INSTRUMENT CONTAINS ELECTROSTATIC DISCHARGE


SENSITIVE DEVICES (ESDS) THAT MAY BE SENSITIVE TO STATIC CHARGES PRODUCED BY THE
HUMAN BODY. DO NOT REMOVE THE COVERS WITHOUT TAKING PROPER ESDS PRECAUTIONS.

CAUTION: DO NOT USE SOLVENTS OR STRONG CLEANING SOLUTIONS ON ANY PART OF THIS
INSTRUMENT AS DAMAGE TO THE UNIT MAY OCCUR. SEE MAINTENANCE SECTION FOR DETAILED
CLEANING INSTRUCTION.

CAUTION: MEDICAL ELECTRONIC EQUIPMENT NEEDS SPECIAL PRECAUTIONS REGARDING EMC


AND NEEDS TO BE INSTALLED AND PUT INTO SERVICE ACCORDING TO THE EMC INFORMATION
PROVIDED IN THE ACCOMPANYING DOCUMENTS.

CAUTION: PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT CAN AFFECT MEDICAL


ELECTRICAL EQUIPMENT.

CAUTION: THIS INSTRUMENT IS NOT TO BE USED NEAR HIGH-FREQUENCY EMITTING SURGICAL


EQUIPMENT.

CAUTION: THIS INSTRUMENT IS NOT intended to be connected to equipment outside the
control of reichert Technologies or must be tested to an applicable iec or iso
standards.

15120-101 Rev. D 5
Symbol Information
Symbol Information
The following symbols appear on the instrument:

Caution symbol indicating important operating and maintenance


instructions that are included in this User’s Guide

Type B Applied Part

Alternating Current Power

Protective Earth Connection

ON / OFF

Date of Manufacture
2013

REF Catalog Number

S/N Serial Number

Waste of Electrical and Electronic Equipment

Compliance to Medical Device Directive 93/42/EEC

Accompanying Documents must be consulted

Authorized Representative in European Community

Fragile Contents in Shipping Container - handle with care

Keep Dry - Package shall be kept away from rain

This Way Up - Indicates correct upright position of package

6 15120-101 Rev. D
Introduction
Congratulations on your purchase of the Xcel™ 255 Slit Lamp.

This User’s guide is designed as a training and reference manual for the operation and maintenance
of the instrument. We recommend that you read it carefully prior to use and follow the instructions to
ensure optimum performance of your new instrument. Properly trained eyecare professionals such as
ophthalmologists, optometrists, opticians and eye care technicians should operate this instrument.

Please retain this manual for future reference and to share with other users. Additional copies can be
obtained from your authorized Reichert dealer or from the Reichert Customer Service Department at:

Tel: 716-686-4500
Fax: 716-686-4555
Email: [email protected]

Indications for use


The Xcel 255 Slit Lamp is an AC-powered slit lamp biomicroscope that is intended for use in examining the
anterior segment, from the corneal epithelium to the posterior capsule. It is used to aid in the diagnosis of
diseases or trauma, which affect the structural properties of the anterior segment of the eye.

Contraindications
None.

15120-101 Rev. D 7
Setup
Parts Identification
X
cel 255 Package Contents
1. On/Off Switch
2. Illumination Level Control • Xcel 255 Slit Lamp (15120)
3. Glide Plate • User's Guide (15120-101)
4. Joystick for horizontal and vertical movement
5. Bulb Access Door Accessories
6. Filter Dial • Focusing Rod (P/N 15120-226)
7. Slit Length Dial • Hex Wrench - 3mm (P/N X54264)
8. Breath Shield Mount • Hex Wrench - 4mm (P/N X54248)
9. Eyepieces • Hex Wrench - 5mm (P/N X54398)
10. Focusing Rings • Dust Cover (15120-225)
11. Fixation Light • Halogen Lamp, Main (15121)
12. Magnification Dial • Guide Rail Covers (15120-031)
13. Microscope Lock Knob • Replacement Fuses (RFAG20063)
14. Slit Rotation Scale • Chin Rest Paper - 1 pack (15120-051)
15. Slit Width/Rotation Knob
16. Illumination Arm Lock Knob
17. Microscope Arm Lock Knob
18. Instrument Base Lock Knob
19. Guide Rail Covers 11
20. Geared Rollers 10
21. Guide Rails 22
22. Chin Rest Assembly 9
23. Power Supply Assembly 12
24. Table Top 8
25. Patient Handles 13
14
7 15
6

5
16
17
25

4
18
20
21 19
3
23 24
2
1

8 15120-101 Rev. D
Setup (continued)
Unpacking and Installation

1. Open the outside shipping box and remove the


three (3) inner boxes.
2. Remove the User's Guide and read it.
3. Open the box with the Table Top and Electronics
in it. Refer to Figure SU-1.
4. Remove the Table Top from the box and install the
Table Top onto the instrument stand and secure
it into place as indicated in the user guide for the
stand. Figure SU-1 Table Top and Electronics
5. Open the box with the Chin Rest Assembly and
remove it. Refer to Figure SU-2.
6. Using the 3mm Hex Wrench, connect the Ground
Wire from the Power Supply to the Chin Rest
Assembly using the Ground Screw provided
(torque to 7.8 N•m). Refer to Figure SU-3.
7. Using the 5mm Hex Wrench, remove the two
Screws from the bottom of the Table Top and
attach the Chin Rest Assembly to the Table Top
using these Screws. Refer to Figure SU-4.
8. Using the 4mm Hex Wrench, adjust the Patient
Handles by loosening the Allen Cap screws that
are securing them to the Chinrest Posts. Slide the Figure SU-2 Chin Rest Assembly
Patient Handles up or down to the desired height,
and secure them in place by tightening the Allen
Cap Screws. Refer to Figure SU-5.

Ground
Screw
-continued-

Ground
Wire
Figure SU-3 Chin Rest Ground

Move
Handles Up
or Down

Allen Cap
Screws
Screws
Figure SU-4 Chin Rest Attaching Screws Figure SU-5 Adjusting Patient Handle Height

15120-101 Rev. D 9
Setup (continued)
Unpacking and Installation (continued) Chin Rest
Base
Lamp
Wire
Wire
9. Attach the Fixation Light Wire from the Chin Rest
Assembly into the connector on the back of the
Power Supply Assembly. Refer to Figure SU-6.
10. Open the box with the Microscope Assembly.
Remove the Microscope Assembly, Base Assembly,
and Accessories. Refer to Figure SU-7.
11. Using the 5mm Hex Screw remove the Allen
Screw on the bottom of the Illumination Assembly
and Arm. Figure SU-6 Connections

Note: The Illumination Assembly and Arm are


connected as one piece.

12. Mount the Illumination Assembly and Arm onto the


Base Assemnly and secure it with the Allen Screw Base
using the 5mm Hex Wrench. Refer to Figure SU-08.
Accessories
Note: There is a Notch in the Base Assembly, and Microscope
a Slot in the Illumination Assembly. Align the
Illumination Assembly and Arm so the Notch
goes into the Slot. If the Notch is not aligned
properly, the Slit Lamp will not sit flush, and
won’t be able to focus properly.
Figure SU-7 Microscope, Base, Accessories
Refer to Figure SU-09.

13. Install the Base Assembly onto the tracks of the


Table Top and slide the Guide Rail Covers around
the tracks. Refer to Figure SU-10.
14. Attach the Base Lamp Wire to the back of the Arm Illumination
Power Supply Assembly. Refer to Figure SU-6. Assembly

Allen
Screw
-continued-
Base
Assembly

Illumination
Assembly Figure SU-8 Allen Screw
and
Arm
Slot

Notch

Base
Slide

Figure SU-9 Line Up Notch Figure SU-10 Base Install

10 15120-101 Rev. D
Setup (continued)
Unpacking and Installation (continued) Slide
15. Install the Microscope Assembly onto the top of Microscope
the Arm by sliding it into position, making sure Assembly
it is up against the stop. Then, tighten the Lock
Knob located on the right side of the Microscope
Assembly. Refer to Figure SU-11.
Turn Counterclockwise
Note: Do not adjust the microscope stop knob
behind the base of the microscope or the Lock Knob
vertex distance will cause misalignment of
focus and require re-calibration of the slit lamp Figure SU-11 Microscope Install
assembly.

16. Remove the accessories and store them in an


appropriate place so that when they are needed Focusing
they will be available. Refer to Figure SU-12. Rod
Power Dust
Application of Input Power Cord Cover Chin Paper
WARNING: Care must be taken to arrange the
cables for the accessories such that they do
not present a tripping hazard to the examiner Wrenches
or a danger to the patient. Guide Rails Replacement
WARNING: POSITION THIS INSTRUMENT SO THAT IT Bulb
IS NOT DIFFICULT TO OPERATE THE DISCONNECTION Figure SU-12 Accessories
DEVICE (PLUG).

1. After the unit is in its secure location, apply the


correct input voltage to the instrument using the
Power Cord from the Accessory Tray.
ON/OFF Switch
Note: The power inlet is located on the backside of
the Power Supply Assembly.

2. Press down on the “|” located on the ON/OFF


Switch. Refer to Figure SU-13.

Note: The ON/OFF Switch will illuminate green when Figure SU-13 Power Supply Assembly
there is power to the unit. When the ON/OFF
Switch is set to off, the green light will turn off.

Disconnection of Input Power


1. At any time, the power switch can be set to OFF.
The unit does not have a power down sequence.
To terminate operation of this instrument, press
the ON / OFF switch to the OFF position (O).
2. If this instrument is intended to be OFF for an
extended period of time, it can be disconnected
from power by detaching the power cord from the
its receptacle.

15120-101 Rev. D 11
Instructions for Use
Operation
1. Turn on the power using the On/Off switch located
on the front of the power supply. Brightness can Projected
Slit
be adjusted by rotating the illumination level knob.

Note: The maximum position is for intermittent use


only. Continuous use will shorten lamp life.

2. Insert the Focusing Rod in the pivot post of the


instrument body to make rough PD and focus
adjustments. Figure IN-1 Focus on Slit
3. Position the light onto the flat surface of the
Focusing Rod and adjust the pupillary distance Canthus
and focus of the eyepieces to suit the needs of the Mark
operator. Refer to Figure IN-1.
4. Using the Slit Width Knobs, adjust the projected
slit so that the thinnest slit is shown on the
Focusing Rod. Refer to Figures IN-1 and IN-4.

Note: The thinnest line will allow for greater accuracy. Chinrest
Elevation
5. Remove the Focusing Rod. Handle
6. To position a patient, adjust the chinrest height
by turning the Chinrest Elevation Handle on the
post of the Chin Rest Assembly until the patient’s
canthus is in line with the Canthus Mark on the
chin rest post. Refer to Figure IN-2. Figure IN-2 Adjust Patient Height
7. Microscope elevation is adjusted by rotating the
Joystick and observing the slit image through the
Microscope Assembly until the slit is centered on
the patient's cornea. Refer to Figure IN-3.
8. Move the slit lamp with the Joystick held firmly
and slightly angled toward the patient, until the slit
Adjust Lamp Lock
appears sharply on the cornea. Height Base
Note: The accuracy of this rough adjustment should
be checked by the naked eye. The fine
adjustment is performed while observing the
Figure IN-3 Adjust Height
slit through the microscope.

9. Tilt the Joystick, which is now held lightly at its


upper end, until the slit appears sharply at the
depth of the eye which is to be observed.
10. The horizontal motion of the base can be locked Slit
by tightening the Base Locking Screw. Refer to Width/Rotation
Figure IN-3. Knob
Note: Lock the base whenever the lamp is not in use.

11. The slit width can be adjusted by rotating the


Slit Width/Rotation Knob on either side of the
instrument. Refer to Figure IN-4. Figure IN-4 Adjust Slit Width

12 15120-101 Rev. D
Instructions for Use (continued)
Operation (continued)
12. The angle between the illumination system and the Magnification
microscope can be varied between 0° and 90° to Dial
either the left or to the right. Refer to Figure IN-5.
13. The illumination angle is indicated on the Scale of
the slit lamp arm. Refer to Figure IN-6.
14. Magnification is altered by rotating the
Magnification Dial on the Microscope Assembly.
Refer to Figure IN-5.

Slit Length
The slit length is adjusted by rotating the Slit Length Figure IN-5 Illumination Angle
Dial. The dial has five stops for adjustments. They
are 0.6, 5.8, 9, 13.5 mm diameter and continuous
length. They index into place. Refer to Figure IN-7.

Filter Dial
The Filter Dial has four positions that index into place,
and are color coded to indicate the active filter. Refer
to Figure IN-7. The color coded index stops are as
follows: Figure IN-6 Illumination Angle Scale

Blue dot = Cobalt Blue
Red dot = Heat Absorbing
White dot = Open
Green dot = Red-free

Slit Rotation Slit Length Dial


Slit rotation is achieved by grasping the Slit Width/
Rotation Knob and twisting the slit body to the left or
right. The degree of rotation is indicated by the Slit Filter
Rotation Scale above the slit body. Refer to Figure IN-8. Dial

Figure IN-7 Filters and Slit Length

Slit Rotation
Scale

Figure IN-8 Slit Rotation Scale

15120-101 Rev. D 13
Cleaning & Maintenance
WARNING: Risk of electric shock. Always
disconnect the power cord from the wall
and the instrument before performing any
of the following care and maintenance
procedures.

External Cleaning
Clean the external surfaces of this instrument using a
clean, soft cloth moistened with a mild detergent solution
(1 cc of liquid dish soap to one liter of clean, filtered
water (filtered below 5 microns)). Refer to Figure MM-1.

Forehead / Chinrest Preparation Figure MM-1 Cleaning Main Unit


For hygienic reasons, wipe the forehead rest with an
alcohol wipe and change the chin rest papers after
Retaining
each patient.
Spring

Cleaning the Glide Plate


If the Glide Plate is dirty it may cause a rough feeling Bulb
when maneuvering the base of the slit lamp. Clean Retaining Holder
the Glide Plate with a soft cloth lightly dampened with Spring
a mild soap and water solution. Rotated
Figure MM-2 Securing Wire
Changing The Halogen Bulb
WARNING: Never remove a bulb that has
recently been in use as it will be very hot. Bulb
Wait until it has cooled and use gloves or
a thick cloth when handling any halogen
bulb. Bulb
Holder
WARNING: Never touch a Halogen Bulb with
bare hands as fingerprints will shorten
the bulb life.

1. Remove input power to the instrument.


2. Open the bulb door. Figure MM-3 Bulb
3. Swing the Retaining Spring away from the Bulb.
Refer to Figure MM-2.
4. Pull the Bulb Holder and Bulb from the unit.
Refer to Figure MM-3.
5. Replace the Bulb with the correct Bulb as indicated
in the Specifications section of this manual. Notch
6. Place the Bulb Holder back into the lamp housing.
Cut Out
Note: Position the Bulb Holder so the Cut Out in the
metal collar of the Bulb lines up with the Notch
in the lamp housing. Refer to Figure MM-4.

7. Move the Retaining Spring back into its original


position. Refer to Figure MM-2. Figure MM-4 Notches
8. Close the bulb door.

14 15120-101 Rev. D
Cleaning & Maintenance (continued)
Fuse Replacement
Replace the fuses in the Power Input Module with the
fuses indicated in the Specifications section of this
manual.

1. Remove input power to the instrument. Pull


2. Press down on the top tab in the middle of the Out
Power Input Module to release the Fuse Holder,
and gently pull out the Fuse Holder by gripping the Tabs
two small tabs. Refer to Figures MM-5 and MM-6.
3. Open the Door to the Fuse Holder by pulling it
down. Refer to Figure MM-6. Figure MM-5 Pull Out

Note: The Fuses will pop up when the door is open,


making removal easier. Fuses

4. Install new fuses into the Fuse Holder that is


indicated in the Specification section of this
manual.
5. Install the Fuse Holder by closing the door, and
pushing the Fuse Holder back until it snaps into
place.

Open
Door
Figure MM-6 Open Fuse Door

15120-101 Rev. D 15
Troubleshooting
The following chart outlines some common issues with the Xcel 255 Slit Lamp and some steps you can take
to correct the issue. If problems persist, please contact the Reichert as listed in the Introduction section of
this manual.

Chart of Common Errors

ISSUE PROBABLE CAUSE POSSIBLE SOLUTION


Incorrect input power supplied to Check the outlet to ensure proper power is
the Xcel 255 Slit Lamp. being supplied.
Lamp won’t turn on. Defective Power Cord. Replace the Power Cord.
Bulb may be blown out. Replace Bulb.
Defective Power Supply. Replace the Power Supply.
Rubber stopper may be attached
Remove the rubber stopper.
Slit Lamp won’t move. under the joystick.
Base Lock Screw may be tightened. Loosen the Base Lock Screw.
Rubber stopper may be attached
Remove the rubber stopper.
Rough base under the joystick.
movement. Bearings may be damaged. Replace the base.
Shaft may be damaged. Replace the base.
Fixation Light Harness not plugged Ensure the Fixation Light Harness is properly
Fixation light does not into the Power Supply Assembly. seated in the Power Supply Assembly.
light up.
Defective Power Supply. Replace the Power Supply.
Incorrect wattage for bulb being used. Replace with the proper Bulb.
Light too dim. Check bulb and ensure notch lines up with
Bulb not installed properly.
bulb housing.
Microscope not focused on focusing Install focusing rod and check to ensure
Double slit visible in rod before use. microscope is focused on it.
microscope. Check bulb and ensure notch lines up with
Bulb not installed properly.
bulb housing.

The following is a checklist of items that need to be assessed in order to determine if the Xcel 255 Slit Lamp
requires servicing.

• Check the outside of the slit lamp for any damage or missing components.
• Inspect the power cord for damage.
• Test the lamp by turning the lamp on and turning the light all the way to it’s brightest setting, and all
the way down to it’s lowest setting.
• Check to ensure all switches are functioning properly.
• Check the Filters by cycling through all the options.
• Check the Slit Wheel by cycling through all the options.
• Check the base movement.

16 15120-101 Rev. D
Specifications
Catalog Number 15120

Physical Dimensions
Size: Weight, unpacked: 23.0 lbs. (10.4 Kg)
Height: 19.8 in. (50.2 cm) Weight, packed: 52 lbs (23.64 Kg)
Width: 10.5 in. (26.7 cm)
Depth: 14.0 in. (35.6 cm)

Electrical
Voltage: 100-240 VAC
Power Input: Max 56-73VA
Frequency: 50/60 Hz
Fuses: T 1.6A L 250V
Halogen Bulb: P/N 15121 (6V, 20 W)

Operational Conditions
Environmental: 70°C
The environmental conditions are as follows:
Operating:
-20°C
Temperature 10° C (50° F) to 35° C (95° F)
Relative Humidity: 30% to 75% 80%
Atmospheric Pressure: 80 kPa (23.6 in. Hg) to
106 kPa (31.3 in. Hg)
10%
Transportation & Storage:
Temperature -20° C (-4° F) to +70° C (158° F). 106 kPa
Relative Humidity: 10% to 80% (non-condensing)
Atmospheric Pressure: 50 kPa (14.8 in. Hg) to
50 kPa
106 kPa (31.3 in. Hg)
Exposure to extreme temperature conditions indicated above must not exceed 15 weeks.
Microscope Galilean
Mag Change 3 Step Drum Rotation
Eyepiece 12.5X
Mag Ratio 10X 16X 25X
PD Range 54 - 75 mm
Diopter Adjustment +/- 5
Slit Illumination 6V 20W Halogen
Slit Width 0 - 13.5 mm
Slit Length 0 - 13.5 mm
Slit Apertures 0.6, 5.8, 9, 13.5 mm
Slit Rotation 0° - 180°
Filters Red Free, Heat Absorbing, Cobalt Blue

15120-101 Rev. D 17
Specifications (continued)
Movement Ranges
Longitudinal (In/Out) 90mm
Lateral (Left/Right) 107mm
Vertical (Up/Down) 30mm
Chinrest Range 80mm
Table Dimensions 18 1/2" x 12 10/16" (465 mm x 316 mm)

Disposal
This product does not generate any environmentally hazardous residues. At the end of its product life, follow
your local laws and ordinances regarding the proper disposal of this equipment.

Software Revision
There is no software installed in this unit.

Due to a policy of continuous development, we reserve the right to change specifications without notice.

18 15120-101 Rev. D
Guidance Tables

Table 201 – Guidance and Manufacturer’s Declaration


Electromagnetic Emissions
All Equipment and Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions

The Xcel 255 is intended for use in the electromagnetic environment specified below. The cus-
tomer or user of the Xcel 255 should ensure that it is used in such an environment.

Electromagnetic Environment
Emissions Test Compliance
- Guidance -
The Xcel 255 uses RF energy only for its internal func-
RF Emissions Group 1 tion. Therefore, its RF emissions are very low and are
CISPR 11 Class B not likely to cause any interference in nearby electronic
equipment.
Harmonics
Class A The Xcel 255 is suitable for use in all establishments,
IEC 61000-3-2
including domestic, and those directly connected to the
Flicker public low-voltage power network that supplies build-
Complies ings used for domestic purposes.
IEC 61000-3-3

15120-101 Rev. D 19
Guidance Tables (continued)

Table 202 – Guidance and Manufacturer’s Declaration


Electromagnetic Immunity
All Equipment and Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity

The Xcel 255 is intended for use in electromagnetic environment specified below. The customer
or user of the Xcel 255 should ensure that it is used in such an environment.

Immunity IEC 60601 Compliance Electromagnetic


Test Test Level Level Environment - Guidance
Floors should be wood, concrete or ceramic
ESD ±6kV Contact ±6kV Contact
tile. If floors are synthetic, the R/H should be at
EN/IEC 61000-4-2 ±8kV Air ±8kV Air
least 30%.

EFT ±2kV Mains ±2kV Mains Mains power quality should be that of a typical
EN/IEC 61000-4-4 ±1kV I/Os ±1kV I/Os commercial or hospital environment.

Surge ±1kV Differential ±1kV Differential Mains power quality should be that of a typical
EN/IEC 61000-4-5 ±2kV Common ±2kV Common commercial or hospital environment.

>95% Dip
>95% Dip
for 0.5 Cycle Mains power quality should be that of a typical
for 0.5 Cycle
60% Dip commercial or hospital environment. If the user
Voltage 60% Dip
for 5 Cycles of the Xcel 255 requires continued operation
Dips/Dropout for 5 Cycles
30% Dip during power mains interruptions, it is recom-
EN/IEC 61000-4-11 30% Dip
for 25 Cycles mended that the Xcel 255 be powered from an
for 25 Cycles
>95% Dip uninterruptible power supply or battery.
Note 1
for 5 Seconds
Power Frequency
50/60Hz Power frequency magnetic fields should be that
3A/m 3A/m
Magnetic Field of a typical commercial or hospital environment.
EN/IEC 61000-4-8

Note 1: During a 5 second loss of power, the Xcel 255 powers off, but recovers once power is restored; meeting the
manufacturers performance requirements.

20 15120-101 Rev. D
Guidance Tables (continued)

Table 204 – Guidance and Manufacturer’s Declaration


Electromagnetic Immunity
Equipment and Systems that are NOT Life-supporting
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Xcel 255 is intended for use in the electromagnetic environment specified below. The cus-
tomer or user of the Xcel 255 should ensure that it is used in such an environment.

Immunity IEC 60601 Compliance Electromagnetic


Test Test Level Level Environment - Guidance
Conducted RF 3 Vrms (V1) = 3 Vrms Portable and mobile RF communications
IEC 61000-4-6 150 kHz to 80 equipment should be used no closer to any
MHz part of the Xcel 255, including cables, than
the recommended separation distance cal-
culated from the equation applicable to the
frequency of the transmitter.

Radiated RF 80 MHz to 2.5 (E1) = 3 V/m Recommended Separation Distance:


IEC 61000-4-3 GHz @ 3V/m
d=(3.5/V1)(Sqrt P)

d=(3.5/E1)(Sqrt P)
80 to 800 MHz

d=(7/E1)(Sqrt P)
800 MHz to 2.5 GHz

Where P is the max output power rating of


the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in meters (m).

Field strengths from fixed RF transmitters,


as determined by an electromagnetic site
survey, should be less than the compliance
levels in each frequence range. (V1 and E1)

Interference may occur in the vicinity


of equipment marked with the following
symbol.

Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.

* Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. The measured field strength in the location in which the ME Equipment or
ME System should be observed to verify normal operation. If abnormal performance is observed, additional
measures many be necessary, such as re-orienting or relocating the ME Equipment or ME System.

* Over the frequency range 150 kHz to 80 MHz, field strengths should be less then [V1] V/m.

15120-101 Rev. D 21
Guidance Tables (continued)

Table 206 – Recommended Separation Distances between


Portable and Mobile RF Communications Equipment and the Xcel 255 for ME
Equipment and ME Systems that are NOT Life-supporting.

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

Recommended Separation Distances for between


Portable and Mobile RF Communications Equipment and the Xcel 255
The Xcel 255 is intended for use in the electromagnetic environment in which radiated RF distur-
bances are controlled. The customer or user of the Xcel 255 can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF Communi-
cations Equipment (transmitters) and the Xcel 255 as recommended below, according to the
maximum output power of the communications equipment.

Max Output Power Separation (m) Separation (m) Separation (m)


of Transmitter 150kHz to 80 MHz 80 to 800 MHz 800MHz to 2.5GHz
(W) d=(3.5/V1)(Sqrt P) d=(3.5/E1)(Sqrt P) d=(7/E1)(Sqrt P)
0.01 0.1166 0.1166 0.2333
0.1 0.3689 0.3689 0.7378
1 1.1666 1.1666 2.3333
10 3.6893 3.6893 7.3786
100 11.6666 11.6666 23.3333
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d)
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (w) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.

22 15120-101 Rev. D
Warranty
This product is warranted by Reichert Technologies against defective material and workmanship under
normal use for a period of one year from the date of invoice to the original purchaser. (An authorized dealer
shall not be considered an original purchaser.) Under this warranty, Reichert’s sole obligation is to repair or
replace the defective part or product at Reichert’s discretion.
This warranty applies to new products and does not apply to a product that has been tampered with, altered
in any way, misused, damaged by accident or negligence, or which has had the serial number removed,
altered or effaced. Nor shall this warranty be extended to a product installed or operated in a manner not in
accordance with the applicable Reichert instruction manual, nor to a product which has been sold, serviced,
installed or repaired other than by a Reichert factory, Technical Service Center, or authorized Reichert
Dealer.
Lamps, bulbs, charts, cards and other expendable items are not covered by this warranty.
All claims under this warranty must be in writing and directed to the Reichert factory, Technical Service
Center, or authorized instrument dealer making the original sale and must be accompanied by a copy of the
purchaser’s invoice.
This warranty is in lieu of all other warranties implied or expressed. All implied warranties of merchantability
or fitness for a particular use are hereby disclaimed. No representative or other person is authorized to make
any other obligations for Reichert. Reichert shall not be liable for any special, incidental, or consequent
damages for any negligence, breach of warranty, strict liability or any other damages resulting from or
relating to design, manufacture, sale, use or handling of the product.
PATENT WARRANTY
If notified promptly in writing of any action brought against the purchaser based on a claim that the
instrument infringes a U.S. Patent, Reichert will defend such action at its expense and will pay costs and
damages awarded in any such action, provided that Reichert shall have sole control of the defense of any
such action with information and assistance (at Reichert’s expense) for such defense, and of all negotiation
for the settlement and compromise thereof.
PRODUCT CHANGES
Reichert reserves the right to make changes in design or to make additions to or improvements in its
products without obligation to add such to products previously manufactured.
CLAIMS FOR SHORTAGES
We use extreme care in selection, checking, rechecking and packing to eliminate the possibility of error. If
any shipping errors are discovered:
1. Carefully go through the packing materials to be sure nothing was
inadvertently overlooked when the unit was unpacked.
2. Call the dealer you purchased the product from and report the
shortage. The materials are packed at the factory and none
should be missing if the box has never been opened.
3. Claims must be filed within 30 days of purchase.
CLAIMS FOR DAMAGES IN TRANSIT
Our shipping responsibility ceases with the safe delivery in good condition to the transportation company.
Claims for loss or damage in transit should be made promptly and directly to the transportation company.
If, upon delivery, the outside of the packing case shows evidence of rough handling or damage, the
transportation company’s agent should be requested to make a “Received in Bad Order” notation on the
delivery receipt. If within 48 hours of delivery, concealed damage is noted upon unpacking the shipment
and no exterior evidence of rough handling is apparent, the transportation company should be requested to
make out a “Bad Order” report. This procedure is necessary in order for the dealer to maintain the right of
recovery from the carrier.

15120-101 Rev. D 23
Reichert Technologies
3362 Walden Ave
Depew, NY 14043
USA

Toll Free: 888-849-8955


Phone: 716-686-4500
Fax: 716-686-4555
Email: [email protected]
www.reichert.com

AMETEK GmbH
Business Unit Reichert
Carl-von-Linde-Strasse 42
85716 Unterschleissheim/Munich
Germany
Email: [email protected]

Tel: +49 (89) 315 8911 0


Fax: +49 (89) 315 891 99

ISO-9001/13485 Registered

15120-101 Rev. D

March 12, 2013

You might also like