Compact Yet Robust: Chemistry Analyzer

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

BS-240E

Chemistry Analyzer

System function Reaction System


Automatic, Discrete, Random Access, Bench-top Reaction tray: 80 reusable cuvettes
STAT sample priority Reaction volume: 100~360μL
Throughput: Constant 240 photometric tests per Reaction temperature: 37℃ ± 0.1℃

hour, up to 400 T/H with ISE Cuvette Washing: Washing station with pre-warmed
Measuring principles: Absorbance photometry, detergent and de-ionized water

turbidimetry, ion selective electrode


technology ISE Module (optional)
Methodology: End-point, Fixed-time, Kinetic, Measuring K+, Na+, Cl-
optional ISE,
Single/Double reagent chemistries, Optical System
Mono-chromatic / bi-chromatic Light Source: Halogen-tungsten lamp
Original system pack reagent ready to use Wavelength: 12 wavelengths, 340nm, 380nm, 412nm,
Close system and open system is optional 450nm, 505nm, 546nm, 570nm, 605nm,
660nm, 700nm, 740nm, 800nm
Reagent/Sample Handling Absorption range: 0~3.5Abs, resolution 0.0001Abs
Reagent/Sample tray: 50 to 100 positions for reagents and 50 Stray Light: 4.9Abs

to 100 positions for samples in 24-hour


Control and Calibration
refrigerated compartment (2~12℃)
Reagent volume: R1: 100~200μL, step by 0.5μL Calibration modes: K factor, Linear (two points and multi-
R2: 10~200μL, step by 0.5μL points), Logit-Log 4P, Logit-Log 5P, Spline,
Sample volumne: 2~35μL, step by 0.1μL Exponential, Polynomial, Parabola,
Reagent/Sample probe: Liquid level detection, horizontal and Logit-Log3P, Broken line
vertical collision protection, inventory One key calibrator import function
checking, reagent pre-warming, Control Rules: Westgard multi-rule, Levey-Jennings,
optional clog detection Cumulative sum check, Twin plot
Probe cleaning: Automatic washing for interior and Operation Unit
exterior Operation system: Windows 10
Carry over < 0.05% Interface: RS-232
Automatic sample dilution: Pre-dilution and post-dilution
Mixing Unit: Independent mixing bar Working Conditions
BS-240E
Built-in Bar Code Reader (Optional)
Power Supply: 200~240V, 50/60Hz, ≤1300VA or Chemistry Analyzer
100~130V, 60Hz, ≤1300VA
Used for sample and reagent programming Dimension: 860 mm (length) × 660 mm (depth) ×
Be applicable to various bar code systems of Codabar, ITF 550 mm (height)
(Interleaved Two of Five), code128, code39, UPC/EAN, Code93
Capable to communicate with LIS in bi-directional mode
Weight:
Water Consumption:
115 kg
≤6.5 L/H Compact yet Robust
P/N:ENG-BS-240E-210285X8P-20170703
Large and flexible capacity Gratings photometer
Up to 100 sample positions
100μL minimum reaction volume
Up to 100 reagent positions
(50 fixed + 50 interchangeable)

A Whole with constant throughtput


New Generation of 240 photometric T/H

Considerate design HbA1c smart-sampling


More space to load samples function
Automatic preparation of hemolysate
BS-240E
Chemistry Analyzer

Independent mixing bar Optimized washing station

Waterfall probe cleaning Built-in barcode reader

Intelligent probe with optional Optional ISE module easy to access


clog detection

Constant throughput Intuitive software with more


functionalities
Complete traceability process Reagent menu
Complete calibration hierarchy and traceability chain are
based on ISO standard (EN/ISO17511) from reference system Hepatic Panel Cardiac panel
to routine measurement system. Alanine Aminotransferase (ALT) Creatine Kinase (CK)
Aspartate Aminotransferase (AST) Creatine Kinase-MB (CK-MB)
Traceability chain of Mindray measurement system (Glu) Alkaline Phosphatase (ALP) Lactate Dehydrogenase (LDH)
Traceability Material Calibration / Value Assignment Procedure Implementation Uncertainty Uc(y)
γ-GlutamylTransferase (γ-GT) α-Hydroxybutyrate Dehydrogenase(α-HBDH)

SI unit,mmol/L Direct Bilirubin (D-Bil) DSA Method High sensitive C-reaction protein( HS-CRP)
Direct Bilirubin (D-Bil)VOX Method
Reference material Primary Reference
measurement procedure Total Bilirubin (T-Bil) DSA Method Inorganic & Anemia
SRM 917C
Total Bilirubin (T-Bil)VOX Method Iron (Fe)
CDC Hexokinase
Total Protein (TP) Ferritin (FER)
Mindray Master
Calibrator reference method Albumin (ALB) Transferrin (TRF)
Total Bile Acids (TBA) Calcium (Ca)
Mindray Product Mindray standard Prealbumin (PA) Magnesium (Mg)
Calibrator Measurement system
Cholinesterase (CHE) Phosphate Inorganic (P)
α-L-fucosidase (AFU) Unsaturated iron binding capacity (UIBC)
End user’ Routine
Routine Sample 5'-nucleotidase (5'-NT)
Mindray Glucose-6-phosphate dehydrogenase (G6PD)

Renal Panel
Result Lipid Panel
Urea (UREA)
Total Cholesterol (TC)
Creatinine (CREA)Modified Jaffé Method
Triglycerides (TG)
External quality assurance for reference measurement Creatinine (CREA)Sarcosine Oxidase Method
HDL-Cholesterol (HDL-C)
Uric Acid (UA)
Mindray participates in RELA (External quality control for reference laboratory). LDL-Cholesterol (LDL-C)
Carbon dioxide (CO2)
Apolipoprotein A1 (ApoA1)
Microalbumin(MALB)
EQA for Mindray Reference laboratory——RELA Apolipoprotein B (ApoB)
β2-Microglobulin (β2-MG)
Mindray reference laboratory has passed RELA for 6 consecutive years. Lipoprotein(a) [Lp(a)]
Cystatin C (CysC)
Retinol binding protein( RBP)
Rheumatism Panel
C-reactive protein (CRP)
Immune Panel
Rheumatoid Factor (RF)
ALT GGT Immunoglobulin A (IgA)
B B Antibodies Against Streptolysin O (ASO)
3.4
Immunoglobulin G (IgG)
Immunoglobulin M (IgM)
3.2
2.8
Lung Panel
Complement C3 (C3)
Adenosine Deaminase (ADA)
3 2.6 Complement C4 (C4)
Angiotensin Converting Enzyme(ACE)
2.8
2.4 Diabetes Panel
2.6 Glucose (Glu) GOD-POD Method Pancreatitis Panel
2.2 Glucose (Glu) HK Method α-Amylase (α-AMY)
2.4

Hemoglobin A1c (HbA1c) Lipase (LIP)


1.5 1.6 1.7 1.8 1.9 2 1.68 1.76 1.64 1.92 2 2.08
A [umol/l] A [mmol/l]
limits of equivalence = +5,25% limits of equivalence = +5,25% Fructosamine (FUN)
β-Hydroxybutyrate(β-HB)
More RELA results please refer to: www.dgkl-rfb.de/81
RELA
All the items Mindray participate RELA

ALT AMY ALP CK GGT GLU LDH TB TP UA UREA