Mini-Doppler Flow System: This Product Is Available Through Associates 800-221-0111
Mini-Doppler Flow System: This Product Is Available Through Associates 800-221-0111
Mini-Doppler
Flow System
GAMMEX1430 LE
The 1430 LE is designed to provide a ground texture and composite film scanning surface
sonographer with advanced design that has improved transmission properties.
features that allow for both Doppler Unique to the 404GS LE phantom are the closely
and B-Mode ultrasound system tests in spaced pin targets. By spacing them closer together,
one unit. This product consists of the flow system, our targets are ideal for testing a range of high
phantom, and electric flow controller meter. frequency transducers. Resolution patterns and all
The B-Mode Phantom component of the 1430 LE is vertical and horizontal pins are made of 0.1 mm nylon
based on our 404GS LE model that provides advanced fibers for better detail definition and spatial
technology for measuring image quality on small parts resolution. Three grey scale targets of -6 dB, +6 dB
and intra-cavity ultrasound scanning systems. With and +12 dB relative to the background material are
added grey scale targets this small parts phantom also included. A series of anechoic cysts with diameters of
measures contrast, temporal resolution and system 1, 2, 4 and 7 mm permit easy evaluation of system
linearity. The phantom incorporates the Gammex noise and distortion parameters.
Tissue Mimicking gel which provides a smoother back- continued .....
SPECIFICATIONS
Two 4 mm vessels are incorporated into the system to
meet with FDA Doppler sensitivity recommendations. Tissue Mimicking Material
Speed of Sound . . 1540 ±10 m/s
One vessel is parallel to the scan plan at 2 cm, and
Attenuation . . . . 0.5 or 0.7 ±0.05 dB/cm/MHz
the other vessel descends 35 degrees from the surface
Grey Scale Targets
and is used for measuring Doppler sensitivity and for
Contrast . . . . . . . -6, +6, and +12 dB relative
developing scanner techniques. to background
Diameter . . . . . . 7 mm
This self-contained system provides a broad range of
Speed of Sound . . 1540 ±10 m/s
available flow rates and test objects. The enhanced
Attenuation . . . . 0.5 or 0.7 ±0.05 dB/cm/MHz
microprocessor-based flow controller meter produces
Anechoic Cysts
accurate flow rates from 1 to 17 ml/sec corresponding Diameters . . . . . 1, 2, 4, and 7 mm
to 0 to 174 cm/sec measurable velocities . Speed of Sound . . 1540 ±10 m/s
Attenuation . . . . 0.05 ±0.01 dB/cm/MHz
The 1430 LE provides a realistic test medium
Blood Mimicking Fluid
for assessing:
Density . . . . . . . 1.03 g/cc
• Maximum signal penetration Speed of Sound . . 1550 ±10 m/sec
Scatter Size . . . . 4.7 micron average
• Channel isolation of directional discrimination
Vessels
• Registration accuracy of duplex sample gates and Size. . . . . . . . . . 4 mm inside diameter
similarities between B-Mode and color flow images. Location . . . . . . horizontal 2 cm below surface;
diagonal at 35 degrees from 1.5 cm
• Flow rate visibility below surface
The Gammex 1430 LE is an all-in-one portable system Electronic Flow Control System
Measurable
powered by a built-in LI-ion rechargeable battery.
Velocities . . . . . . 0-174 cm/sec
The universal charger is included with the system. Pulsatile
Flow Mode . . . . . preprogrammed for
0.25-2 seconds/pulse
Pin Targets
Diameter of
nylon lines . . . . . 0.1 mm (0.004 in)
Vertical Spacing. . 5 mm at 1 cm to 9 cm deep
Horizontal
Spacing . . . . . . . 10 mm at 1 cm and 5 cm deep
Container Construction
Size. . . . . . . . . . 22.9x15.9x21.9 cm (9x6.25x8.625 in)
Weight. . . . . . . . 4.4 kg (9.5 lbs)
AC input voltage . 110 VAC – 240 VAC, 50 – 60 Hz
Battery . . . . . . . LI-ion, rechargeable,
universal charger and plugs included
All acoustic measurements at 4.5 MHz, 22° C
JRT Associates
5 Nepperhan Avenue, Suite 2B
Elmsford, NY 10523
800-221-0111