Omega Om CP Rhtemp2000 Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 30

User’s Guide

MADE IN

Shop online at

omega.com
e-mail: [email protected]
For latest product manuals:
omegamanual.info

OM-CP-PRHTEMP2000
OM-CP-RHTemp2000
Temperature,
Humidity andHumidity, and
Temperature
Pressure
Data Logger withData Logger
LCD Display
with LCD Display
OMEGAnet ® Online Service Internet e-mail
omega.com [email protected]

Servicing North America:


U.S.A.: One Omega Drive, Box 4047
ISO 9001 Certified Stamford, CT 06907-0047
Tel: (203) 359-1660
FAX: (203) 359-7700
e-mail: [email protected]
Canada: 976 Bergar
Laval (Quebec) H7L 5A1, Canada
Tel: (514) 856-6928
FAX: (514) 856-6886
e-mail: [email protected]

For immediate technical or application assistance:


U.S.A. and Canada: Sales Service: 1-800-826-6342/1-800-TC-OMEGA®
Customer Service: 1-800-622-2378/1-800-622-BEST®
Engineering Service: 1-800-872-9436/1-800-USA-WHEN®

Mexico: En Español: (001) 203-359-7803


e-mail: [email protected]
FAX: (001) 203-359-7807
[email protected]
Servicing Europe:
Czech Republic: Frystatska 184, 733 01 Karvina´, Czech Republic
Tel: +420 (0)59 6311899
FAX: +420 (0)59 6311114
Toll Free: 0800-1-66342
e-mail: [email protected]
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: +49 (0)7056 9398-0
FAX: +49 (0)7056 9398-29
Toll Free in Germany: 0800 639 7678
e-mail: [email protected]
United Kingdom: One Omega Drive, River Bend Technology Centre
ISO 9002 Certified Northbank, Irlam, Manchester
M44 5BD United Kingdom
Tel: +44 (0)161 777 6611
FAX: +44 (0)161 777 6622
Toll Free in United Kingdom: 0800-488-488
e-mail: [email protected]

It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI
regulations that apply. OMEGA is constantly pursuing certification of its products to the European New
Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any
errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, human applications.
Data Logger Manual

Table of Contents

Section 1: Device Safety............................................................................3


1.1 Model information....................................................................3
1.2 Specifications............................................................................3

Section 2: Device Overview.......................................................................5

Section 3: Status Indicator Icons...............................................................6


3.1 Battery Status............................................................................6
3.2 Memory Status..........................................................................6
3.3 Running Indicator......................................................................6
3.4 Delay Start Indicator..................................................................6
3.5 Wait Icons.................................................................................6
3.6 Stop Icon...................................................................................6
3.7 External Power Icon...................................................................7
3.8 Reset Icon..................................................................................7

Section 4: Front Panel Overview...............................................................8


4.1 Changing the display units........................................................8
4.2 Changing the number, type, and size of channels viewed..........8
4.3 Checking the memory status.....................................................8
4.4 Checking power status..............................................................8
4.5 Changing the contrast...............................................................9

Section 5: OM-CP-RHTemp2000 Function Reference.............................10


5.1 Main Screen............................................................................10
5.2 Status Menu............................................................................10
5.3 Statistics Menu........................................................................10
5.4 Units.......................................................................................11
5.5 Setup Menu............................................................................11

Section 6: OM-CP-RHTemp2000 Screen Descriptions.............................14


6.1 Main Screen:...........................................................................14
6.2 Status Screens (Run Parameters):.............................................14
6.3 Status Screens (Memory Status):..............................................14
6.4 Status Screens (Time):.............................................................15
6.5 ID Parameters..........................................................................15
6.6 Statistics Menu Screen:............................................................16
6.7 Channel Statistics:...................................................................16
6.8 Type Statistics:.........................................................................16
6.9 Statistics Information Screen:...................................................17
6.10 Unit Selection Menu:.............................................................17
6.11 Device Configuration Menu:..................................................17
6.12 Adjust Visibility:.....................................................................18
6.13 Display Configuration:...........................................................18
6.14 Power Modes Screen:............................................................19
6.15 Display Update Mode Screen:................................................19

Revised 01/21/08 Page 1


OMEGA OM-CP-RHTEMP2000

6.16 Power Status Screen:.............................................................19


6.17 Device Information Screens (Minimum Device Range):..........20
6.18 Device Information Screens (Maximum Device Range):.........20
6.19 Device Information Screens (Device Version):........................20
6.20 Device Information Screens (Firmware Version):....................21
6.21 Calibration Information Screens (Calibration Date):...............21
6.22 Calibration Information Screens (Temperature Calibration):...21
6.23 Calibration Information Screens (Humidity Calibration):........22
6.24 Device Reset Screen (Hardware Reset):..................................22
6.25 Device Reset Screen (Power Interruption):.............................23

Section 7: Computer Interface................................................................24

Section 8: Wall Mounting........................................................................24

Section 9: Maintenance...........................................................................25

Section 10: Warranty/Disclaimer.............................................................26

Page 2
Data Logger Manual

Section 1: Device Safety

CAUTION: DEVICE MUST BE USED ONLY IN A MANNER CONSISTENT WITH THIS MANUAL.
NOTICE: WHEN 230VAC SUPPLY USED, OMEGA SPECIFIES USE OF AC SUPPLY PART # T35-9-100R-3
MANUFACTURED BY ENG ELECTRIC. THIS SUPPLY IS AVAILABLE FROM OMEGA.

1.1 Model information


• Model: OM-CP-RHTemp2000
• Description: Measure and record temperature and humidity, with LCD display
• Manufactured in the USA

1.2 Specifications

Temperature
Range: -20 to +60ºC
Resolution: 0.1ºC
Calibrated Accuracy: ±0.5ºC (0 to +50ºC)
Humidity
Range: 0 to 95%RH
Resolution: 0.1%RH
Calibrated Accuracy: ±3%RH (±2%RH typical at 25ºC)
Specified Accuracy Range: +10 to +40°C, 10 to 80%RH

Dot-Matrix LCD
Dimensions: 2.5” x 1.375” (63mm x 35mm)
Text: Configurable channel text size
Indicators: Power, status, memory
Backlight: Configurable w/auto shut-off and contrast adjustment

Start/Stop Time:
Software programmable start time and date, up to six months in advance; programmable stop time

Memory:
131,071 readings per channel; 262,143 total readings; software configurable memory wrap

Reading Rate:
1 reading every 2 seconds to 1 every 24 hours

Calibration:
Digital calibration through software

Calibration Date:
Automatically recorded within device

Battery Type:
9V lithium battery included, user replaceable; optional AC adapter, 7-24 VDC, 100mA. For 230 VAC
operation use AC supply T35-9-100 R-3 made by ENG Electric (Available from MadgeTech).

Page 3
OMEGA OM-CP-RHTEMP2000

Battery Life:
30 days typical @ 1 min reading rate with continuous LCD and no backlight usage.
Data Format:
Date and time stamped ºC, ºF, K, ºR ; %RH, mg/ml water vapor concentration, dew point
Time Accuracy:
±1 minute/month (at 20º to 30°C)
Computer Interface:
PC serial or USB (interface cable required); 115,200 baud
Software:
Windows 95/98/ME/NT/2000/XP based software
Operating Environment:
-20 to +60ºC, 0 to 95%RH non-condensing
Dimensions:
4.8” x 3.3” x 1.25” (122mm x 84mm x 32mm)
Weight:
16 oz (440 g)
Enclosure:
Black anodized aluminum

Page 4
Data Logger Manual

Section 2: Device Overview

Power Jack
Temperature
and Humidity
LED Sensor
Indicators Ventilation
Holes
Communication
LCD Screen Jack
Up Key

Black Okay Key


anodized
aluminum Down Key
enclosure
Function Keys Cancel Key

Page 5
OMEGA OM-CP-RHTEMP2000

Section 3: Important Status Icons


3.1 Battery Status

Full Empty
Battery Battery
Battery status icons indicate the state of charge of the batteries. When the battery empty indicator is displayed, the battery
should be replaced.

3.2 Memory Status

Memory Memory
Empty Full
The memory status icons indicate the amount of memory left for data storage.
If the OM-CP-RHTemp2000 is in wraparound mode, the memory status icon will always indicate empty.

3.3 Running Indicator

Indicates that the OM-CP-RHTemp2000 is taking readings. The icons cycle periodically.

3.4 Delay Start Indicator

The delay start icon indicates that a logging operation is scheduled for the future.

3.5 Wait Icons

The wait icon indicates that the device is busy.

3.6 Stop Icon

Indicates that the OM-CP-RHTemp2000 is not currently taking readings.

Page 6
Data Logger Manual

3.7 External Power Icon

Indicates that the OM-CP-RHTemp2000 is powered by an external source.

3.8 Reset Icon

Indicates that the device has been reset.

Page 7
OMEGA OM-CP-RHTEMP2000

Section 4: Front Panel Overview


4.1 Changing the display units
The OM-CP-RHTemp2000 comes with factory default display units of °C for temperature and %RH for humidity. These
units can be easily changed by pressing the F3 button in the main screen and then selecting either F1 for temperature
and F2 for humidity. After selecting a channel, the available units can be scrolled through by either pressing the channel’s
function key repeatedly or using the UP and DOWN keys.

Button pressing chain:


Main Screen -> F3 -> F1(temp) or F2(humidity) -> function key repeatedly or UP and DOWN

4.2 Changing the number, type, and size of channels viewed


By default the OM-CP-RHTemp2000 displays recently measured values of both channels (temperature and humidity) on
its Main Screen with the two channels taking up the maximum amount of screen space available. Channels can, however,
be hidden or viewed on a smaller or larger scale.
To change the number and type of displayed channels:
From the Main Screen, press the F4 key to enter the Setup Menu and from this menu press the F1 key to enter the Dis-
play screen. On this screen, F1 corresponds to the Temperature channel and F2 to Humidity. Pressing these function keys
will cause the channels to scroll between “show” or “hide”. Channels displaying “show” will appear on the main screen
and channels displaying “hide” will not. Any number of channels between zero and two may be shown.
Button pressing chain:
Main Screen -> F4 -> F1 -> F1(temp) and F2(humidity)

To change the size of displayed channels:


From the Main Screen, press the F4 key to enter the Setup Menu and from this menu press the F1 key to enter the Dis-
play screen, then F4 to scroll to the next screen. Here the F2 key will change the size of the channels viewed. By pressing
F2 repeatedly the size parameter will scroll between three sizes:
Small: Up to two channels can be displayed and appear much smaller than the available screen space.
Medium: Up to two channels can be displayed

Large: Up to two channels can be displayed which will fill the entire display area

Button pressing chain:


Main Screen -> F4 -> F1 -> F4 -> F2 repeatedly to scroll

4.3 Checking the memory Status


A status icon appears on all screens representing memory, but further information including percent memory left and
number of readings taken can also be viewed. From the Main Screen press the F1 key to enter the Status screens then
press F2 to view memory status information.

Button pressing chain:


Main Screen -> F1 -> F2

4.4 Checking power status


A battery status and external power status (if available) icon appear on all screens, but percent battery power remaining
and external power presence as well as battery type, current battery voltage, and current external voltage can also be
viewed. From the Main Screen press F4 to view the Device Configuration Menu, F2 to access the power options, then F4
twice to view the Power Status screen, including battery power percent remaining and the presence of external power.
Battery type and battery voltage are also displayed, as well as external power voltage (if connected).

Page 8
Data Logger Manual

Button pressing chain:


Main Screen -> F4 -> F2 -> F4 -> F4

4.5 Changing the Contrast


The OM-CP-RHTemp2000’s LCD screen contrast values can be changed in two ways. One method is outlined in the
Function Reference Guide. A faster, simpler way involves simultaneously pressing the CANCEL and UP or DOWN button
in any screen.

Button pressing chain:


CANCEL + UP (to increase) or DOWN (to decrease)

Page 9
OMEGA OM-CP-RHTEMP2000

Section 5: OM-CP-RHTemp2000 Function Reference


5.1 Main Screen
The main screen of the OM-CP-RHTemp2000 features a real-time display of most recently measured temperature and
humidity. At the bottom of the main screen are tabs corresponding to each of the four function keys. These tabs are
used to access the four main function menus of the OM-CP-RHTemp2000: status, statistics, units, and setup. The left
side of the main screen and all subsequent screens of the device is where important status information icons can be found
(detailed in Section 3: Important Status Icons – page 7) including recording status, memory status, busy status, external
power status, and battery power status.

5.2 Status Menu


Pressing F1 on the main screen brings up the Status menu. The first screen that appears in the Status Menu is Run Param-
eters, but the Memory Status and Time screens can also be viewed by pressing the F2 and F3 keys respectively.

5.2.i Run Parameters


The Run Parameters screen displays important information regarding the device’s current recording session. These param-
eters include the time and date the recording session started (start time and start date), the time and date the recording
session will end (stop date and stop time) due to either lack of memory or preprogramming in the MadgeTech software.
The rate at which the OM-CP-RHTemp2000 is recording (rate) is also displayed. The device’s current status (either run-
ning or stopped) is the last parameter on the Run Parameters screen.

5.2.ii Memory Status


The Memory Status screen is where all information regarding the OM-CP-RHTemp2000’s memory. This screen displays the
percent of memory space currently available (memory left), the number of readings currently stored on the device (read-
ings), the maximum number of readings the device can record (max readings), as well as information about the wrap
feature displaying either “disabled” or the number of wrap readings currently stored in memory.

5.2.iii Time
The time screen displays current time data including the current time and date, time and date of last measured data, and
current time zone.

5.3 Statistics Menu


Pressing the F2 key while in the Main Screen brings up the Statistics Menu. From the Statistics Menu, statistics generated
from the conditions encountered by the device can be viewed in a variety of different styles including being sorted by
channel and by type. The Statistics Menu also displays important statistics information as well as the option to clear the
statistics at any time.

5.3.i Viewing statistics by channel


Pressing F1 while viewing the Statistics Menu brings up statistics sorted by channel. Here the F1and F2 keys provide the
ability to view statistics regarding temperature and humidity. Each channel screen displays the minimum, maximum, and
average values encountered by the device for a particular channel.

5.3.ii Viewing statistics by type


The OM-CP-RHTemp2000’s Statistics Menu also provides the option to view statistics by type. This can be done by press-
ing F2 while in the Statistics Menu. Here the function keys F1, F2, and F3 correspond to the three different types of statis-
tics; average, minimum, and maximum respectively. Each of the type statistics screens displays temperature and humidity
data of a particular type (either average, minimum, or maximum values).

5.3.iii Statistics Information


Pressing the F3 key while viewing the Statistics Menu screen brings up Statistics Information. This screen displays the
number of readings being considered within the statistics (readings), as well as the date and time the recording period
began. From this screen the statistics information can also be cleared. This is done by pressing the F1 key marked by a tab
labeled “CLEAR”. Upon pressing this key a confirmation message will appear with function tabs labeled “NO” and “YES”
corresponding to F1 and F2 respectively. Selecting “YES” by pressing the F2 key will confirm the statistics clear function.
Page 10
Data Logger Manual

5.4 Units
Pressing the F3 key while viewing the Main Screen will access the Units Selection screen. Here the measurement units can
be easily changed. In the Units Selection screen the F1 and F2 function keys correspond to the two channels temperature
and humidity. Selecting a channel by pressing its corresponding function key allows the user to change the units by either
pressing the function key repeatedly or using the UP and DOWN keys to scroll through the list of available units. Select-
ing OK (either by pressing the OK button or the F4 key) accepts and confirms the unit selection. Pressing the CANCEL
key cancels the unit changing action and reverts to the previously selected units. The OM-CP-RHTemp2000 offers most
commonly used units.

5.5 Setup Menu


Pressing the F4 key while in the Main Screen will display the Device Configuration screen. From this menu changes can be
made to most of the OM-CP-RHTemp2000’s display configuration including the screen contrast, size of the channel view
in the main screen, as well as display of the channels. Power status can also be viewed including battery power remaining
and the presence of external power, and options regarding the LCD display and the backlight can be modified. The setup
menu is also the place to find basic information regarding device identification, calibration parameters, and firmware
details.

5.5.i Display
From the Display section, the number and identity of channels shown on the main screen can be changed and set, the
LCD screen’s contrast values can be changed and set, and the size the channel information appears on the main screen
can be changed and set. Pressing F1 while in the Setup Menu brings up the Display section.

5.5.i.a Adjust Visibility


The first screen that appears in the Display section is the Adjust Visibility screen. Here the F1 and F2 keys correspond to
the temperature and humidity channels respectively. Pressing one of these function keys results in highlighting its co
responding channel function tab. While a channel is highlighted the corresponding display function can be toggled
between “Hide” and “Show”, with “Hide” indicating that the channel will not be displayed on the main screen and
“Show” indicating that it will be shown. Channels can be toggled by either repeatedly pressing the channel’s
corresponding function key or by using the UP and the DOWN keys.

5.5.i.b Display Configuration


By selecting “MORE” (either by pressing the F4 key on the Adjust Visibility Screen or by pressing the OK button) in the
Display section the Display Configuration screen can be viewed. Here LCD screen contrast values can be set as well as
the size the channels are displayed on the Main Screen.
Contrast
To change the LCD screen contrast view while viewing the Display Configuration screen, first select the contrast
configuration parameter by pressing the F1 key. This action will highlight the function tab corresponding to the F1
key. Once this tab is highlighted the contrast can be increased or decreased using the UP and DOWN keys.

Size
To change the size that each channel is displayed on the Main Screen view while viewing the Display Configuration
screen, first select the size configuration parameter by pressing the F2 key. This action will highlight the function
tab corresponding to the F2 key. Once this tab is highlighted the size can be increased or decreased using the UP
and DOWN keys. Sizes are as follows:

Small: Up to three channels can be displayed and appear much smaller than the available screen space.

Medium: Up to two channels can be displayed.

Large: Up to two channels can be displayed which will fill the entire display area

Page 11
OMEGA OM-CP-RHTEMP2000

5.5.ii Power
The Power section displays information regarding battery and external power sources as well as giving options including
display and backlight status that could potentially save power and LED status options.

5.5.ii.a Power Modes


The power modes screen displays information regarding the display, backlight, and LED mode as well as the options
to change their function.

Display
The OM-CP-RHTemp2000’s LCD screen has options to be either on continuously or turn off automatically after five
minutes of inactivity. This option can be changed by first pressing the F1 key to highlight the display parameter
configuration tab. Once the tab is highlighted the option can be changed by pressing the F1 key repeatedly or
using the UP and DOWN keys to select either “On” or “Auto”.

Backlight
Like the display, the LCD screen’s backlight also has power saving options. The backlight can be either on continuously,
off, or set to automatically shut off after 30 seconds of inactivity. This option can be changed by first pressing the
F2 key to highlight the back light parameter configuration tab. Once the tab is highlighted the option can be
changed by pressing the F2 key repeatedly or using the UP and DOWN keys to select either “On”, “Off” or
“Auto”.

LED Modes
The function of the two LEDs on the upper left corner of the OM-CP-RHTemp2000 can be regulated by first pressing
the F3 key to highlight the LED parameter configuration tab. Once the tab is highlighted the option can be
changed by pressing the F3 key repeatedly or using the UP and DOWN keys to select either “Enabled” meaning the
LEDs will light to indicate device function or “Disabled” meaning the LEDs will never light.

5.5.ii.b Display Update Mode


Selecting “MORE” by either pressing the F4 key or “OK” key while viewing the Power Modes screen brings up
the Display Update Mode screen. On this screen, options relating to the refresh rate of the screen are displayed.
The F1 key toggles between ‘auto’ and ‘reading’ modes. In ‘auto’ mode, the display is updated periodically,
whereas in ‘reading’ mode, the display only updates after a reading is taken. F2 selects the update period option.
Pressing the “UP” and “DOWN” keys increase and decrease the update period respectively.

5.5.ii.c Power Status


Selecting “MORE” by either pressing the F4 key or the OK key while viewing the Display Update Mode screen
brings up the Power Status screen. On this screen the battery type is displayed and should be changed by the
user based on whether a Lithium or Alkaline battery is used in the OM-CP-RHTemp2000. This can be changed by first
pressing the F1 button to highlight the type parameter configuration tab. Once the tab is highlighted, the battery
type can be changed by either pressing the F1 key repeatedly or by using the UP and DOWN arrows to select the
correct battery type. The percent battery power remaining, external power presence, battery voltage and external
power voltage are also displayed on the Power Status screen.

5.5.iii Device Information


The F3 key in the Setup Menu displays the Device Information screens. Here device identification information can be found
including serial number, product ID, revision, and subtype. Calibration parameters are also found under device informa-
tion including the date the device was last calibrated and when it will be due for calibration again. Firmware details are also
found here.
5.5.iii.a Device Range (Minimum)
The Minimum Device Range screen can be accessed by pressing the F1 key while viewing the Device Information
screen. This display indicates the minimum values which the OM-CP-RHTemp2000 is capable of detecting and logging
Values on this screen cannot be modified.
Page 12
Data Logger Manual

5.5.iii.b Device Range (Maximum)


The Maximum Device Range screen can be accessed by pressing the F2 key while viewing the Minimum Device
Range screen. This display indicates the maximum values which the OM-CP-RHTemp2000 is capable of detecting and
logging. Values on this screen cannot be modified.

5.5.iii.c Calibration Parameters


Pressing F3 while in the Maximum or Minimum Device Range screens will display the device calibration parameters.
This screen displays information indicating the most recent calibration date as well as the date the next calibration
will be due.

It is important to keep your device properly calibrated to ensure accurate readings.




5.5.iii.d Device Version
Pressing the F4 key while viewing the Maximum or Minimum Device Range screens will display the Device
Version screen, containing information such as firmware revision number and communications baud rate. Values
on this screen cannot be modified.

5.5.iii.e Firmware Version


Firmware details can be viewed by pressing the F2 key while viewing the Device Version screens. These details
include the firmware version number, date and time of firmware creation, and checksum.

Page 13
OMEGA OM-CP-RHTEMP2000

Section 6: OM-CP-RHTemp2000 Screen Descriptions

6.1 Main Screen:

Displays last measured values.

F1 = STATUS: goes to status screens


F2 = STATS: shows statistics menu
F3 = UNITS: goes to unit selection screen
F4 = SETUP: shows device configuration menu
CANCEL = no function
OK = no function
UP = no function
DOWN = no function

6.2 Status Screens (Run Parameters):

Displays information about run parameters including date of recording start (start date), time of recording start (start time),
stop date, stop time, recording rate, and current status.
For all status screens:
F1 = RUN: displays run parameters screen
F2 = MEM: displays memory status screen
F3 = TIME: displays time information screen
F4 = MORE: displays the second status screen
CANCEL = returns to main screen
OK = returns to main screen
UP = no function
DOWN = no function

6.3 Status Screens (Memory Status):

Displays information about the device’s memory capabilities including percent of memory available (memory left), number of
readings taken so far (readings), max number of readings (max readings), and wrap.

Page 14
Data Logger Manual

For all status screens:


F1 = RUN: displays run parameters screen
F2 = MEM: displays memory status screen
F3 = TIME: displays time information screen
F4 = MORE: displays the second status screen
CANCEL = returns to main screen
OK = returns to main screen
UP = no function
DOWN = no function

6.4 Status Screens (Time):

Displays current time and date as well as registered time and date and time zone information.
For all status screens:
F1 = RUN: displays run parameters screen
F2 = MEM: displays memory status screen
F3 = TIME: displays time information screen
F4 = MORE: displays the second status screen
CANCEL = returns to main screen
OK = returns to main screen
UP = no function
DOWN = no function

6.5 ID Parameters

Displays information relating to device identity.


For all status screens:
F1 = ID: no function
F3 = MORE: displays the first status screen
F4 = EXIT: returns to main screen
CANCEL = returns to main screen
OK = returns to main screen
UP = no function
DOWN = no function

Page 15
OMEGA OM-CP-RHTEMP2000

6.6 Statistics Menu Screen:

Displays options available within the statistics menu.


F1 = CHAN: goes to screen that sorts statistics by channel
F2 = TYPE: goes to screen that sorts statistics by type
F3 = INFO: goes to Statistics Information Screen
F4 = EXIT: returns to main screen
CANCEL = returns to main screen
OK = returns to main screen
UP = no function
DOWN = no function

6.7 Channel Statistics:

Displays statistics (maximum recorded value, minimum recorded value, and average recorded value) based on channel
(Temperature or Humidity).

F1 = TEMP…: Displays temperature statistics


F2 = HUMID…: Displays humidity statistics
F3 = no function
F4 = EXIT: returns to Statistics Menu Screen
CANCEL = returns to Statistics Menu
OK = returns to Statistics Menu
UP = no function
DOWN = no function

6.8 Type Statistics:

Displays statistics from both channels (Temperature and Humidity) sorted by type (maximum recorded value, minimum recorded
value, or average value).
F1 = AVG: displays average recorded value for each channel
F2 = MIN: displays minimum recorded value for each channel
F3 = MAX: displays maximum recorded value for each channel
F4 = EXIT: Returns to Statistics Menu
Page 16
Data Logger Manual

CANCEL = returns to Statistics Menu


OK = returns to Statistics Menu
UP = no function
DOWN = no function

6.9 Statistics Information Screen:

Displays current statistics information including the number of recorded readings, start date, and start time.
F1 = CLEAR: Gives the option to clear all statistics
F2 = no function
F3 = no function
F4 = EXIT: returns to Statistics Menu
CANCEL = returns to Statistics Menu
OK = returns to Statistics Menu
UP = no function
DOWN = no function

6.10 Unit Selection Menu:

Displays units currently being used for each channel

F1 = TEMP: selects temperature channel for unit change


F2 = HUMID: selects humidity channel for unit change
F3 = no function
F4 = OK: returns to Main Screen
Note: unit selection can be changed by pressing the corresponding function key repeatedly or using the UP and DOWN arrows
to select appropriate units.
CANCEL = returns to main menu without accepting changes
OK = accepts changes and returns to main menu

6.11 Device Configuration Menu:

Displays options available within the device configuration menu.


F1 = DISPLAY: enters Adjust Visibility screen
Page 17
OMEGA OM-CP-RHTEMP2000

F2 = POWER: enters Power Modes screen


F3 = INFO: goes to Device Information screens
F4 = EXIT: returns to main screen
CANCEL = returns to main screen
OK = returns to main screen
UP = no function
DOWN = no function

6.12 Adjust Visibility:

Displays options for changing the viewing of the two channels on the main screen (either shows a particular channel or hides it).
F1 = TEMP: first highlights and then changes the viewing options of the temperature channel
F2 = HUMID: first highlights and then changes the viewing options of the humidity channel
F3 = no function
F4 = MORE: moves on to Display Configuration screen
CANCEL = return to Device Configuration Menu
OK = moves on to Display Configuration screen
UP = once channel parameter configuration tab is highlighted scrolls through available options
DOWN = once channel parameter configuration tab is highlighted scrolls through available options

6.13 Display Configuration:

Allows the user to change the contrast of the LCD display as well as the channel size.

F1 = CNTRST: highlights the contrast parameter configuration tab


F2 = SIZE: highlights and then changes options of the channel size parameter
F3 = MORE: moves back to the Adjust Visibility screen
F4 = EXIT: returns to Display Configuration Menu
CANCEL = returns to Device Configuration Menu
OK = moves back to the Adjust Visibility screen
UP = when contrast parameter configuration tab is highlighted increases contrast value
DOWN = when contrast parameter configuration tab is highlighted increases decreases contrast value

Page 18
Data Logger Manual

6.14 Power Modes Screen:

Displays information regarding the devices different power modes including the display visibility, backlight options, and LED
modes.
F1 = DISPLAY: first highlights and then changes display visibility (On: full visibility or Auto: shuts off after 5 minutes of
inactivity)
F2 = BKLGHT: first highlights and then changes backlight options (On: backlight always on, Auto: backlight shuts off after
30 sec of inactivity, or Off: backlight always off)
F3 = LED: first highlights and then changes LED mode options
F4 = MORE: moves to Display Update Mode screen
CANCEL = Returns to Device Configuration Menu
OK = moves to Display Update Mode screen
UP = once parameter configuration tab is highlighted scrolls through available options
DOWN = once parameter configuration tab is highlighted scrolls through available options

6.15 Display Update Mode Screen:

Displays information regarding display refresh mode and refresh interval.


F1 = MODE: first highlights and then changes update mode (Auto: screen refreshes periodically or Reading: refreshes only
after a reading is taken)
F2 = TIME: first highlights and then changes refresh interval.
F4 = MORE: moves to Power Status screen
CANCEL = Returns to Device Configuration Menu
OK = moves to Power Status screen
UP = once parameter configuration tab is highlighted scrolls through available options
DOWN = once parameter configuration tab is highlighted scrolls through available options

6.16 Power Status Screen:

Displays details about power available to the device including the battery type, battery voltage, and external voltage.
F1 = TYPE: highlights and then changes battery type (Lithium or Alkaline)
F3 = MORE: moves back to Power Modes screen
F4 = EXIT: Returns to Device Configuration Screen
CANCEL = Returns to Device Configuration Menu
Page 19
OMEGA OM-CP-RHTEMP2000

OK = moves back to Power Modes screen


UP = if type parameter configuration tab is highlighted, scrolls through available options
DOWN = if type parameter configuration tab is highlighted, scrolls through available options

6.17 Device Information Screens (Minimum Device Range):

Displays values indicating minimum temperature and humidity detectable by the OM-CP-RHTemp2000.
For all Device Information Screens:
F1 = MIN: no function
F2 = MAX: shows Device Range (Maximum) screen
F3 = CAL: shows Calibration Information screens
F4 = MORE: shows further Device Information screens
CANCEL = Returns to Device Configuration Menu
OK = Returns to Device Configuration Menu
UP = No function
DOWN = No function

6.18 Device Information Screens (Maximum Device Range):

Displays values indicating maximum temperature and humidity detectable by the OM-CP-RHTemp2000.

6.19 Device Information Screens (Device Version):

Displays device version information.

F1 = VERS: no function
F2 = FIRM: shows Firmware Version screen
F3 = MORE: shows Device Range (Minimum) screen
F4 = EXIT: Returns to Device Configuration Menu
CANCEL = Returns to Device Configuration Menu
OK = Returns to Device Configuration Menu
UP = No function
DOWN = No function

Page 20
Data Logger Manual

6.20 Device Information Screens (Firmware Version):

Displays device firmware version information.


F1 = VERS: shows Device Version screen
F2 = FIRM: no function
F3 = MORE: shows Device Range (Minimum) screen
F4 = EXIT: Returns to Device Configuration Menu
CANCEL = Returns to Device Configuration Menu
OK = Returns to Device Configuration Menu
UP = No function
DOWN = No function

6.21 Calibration Information Screens (Calibration Date):

Displays date of last calibration and due date of next calibration.


For all Device Information Screens:
F1 = DATE: no function
F2 = TEMP: shows temperature calibration information
F3 = HUMID: shows humidity calibration information
F4 = EXIT: Returns to Device Configuration Menu
CANCEL = Returns to Device Configuration Menu
OK = Returns to Device Configuration Menu
UP = No function
DOWN = No function

6.22 Calibration Information Screens (Temperature Calibration):

Displays calibration information for temperature channel.


F1 = DATE: shows calibration date information
F2 = TEMP: no function
F3 = HUMID: shows humidity calibration information
F4 = EXIT: Returns to Device Configuration Menu

Page 21
OMEGA OM-CP-RHTEMP2000

CANCEL = Returns to Device Configuration Menu


OK = Returns to Device Configuration Menu
UP = No function
DOWN = No function

6.23 Calibration Information Screens (Humidity Calibration):

Displays calibration information for humidity channel.


F1 = DATE: shows calibration date information
F2 = TEMP: shows temperature calibration information
F3 = HUMID: no function
F4 = EXIT: Returns to Device Configuration Menu
CANCEL = Returns to Device Configuration Menu
OK = Returns to Device Configuration Menu
UP = No function
DOWN = No function

6.24 Device Reset Screen (Hardware Reset):

Displayed as notification when a hardware reset has occurred.


F1 = OK: accepts notification and displays main screen
F2 = no function
F3 = no function
F4 = no function
CANCEL = no function
OK = accepts notification and displays main screen
UP = no function
DOWN = no function

Page 22
Data Logger Manual

6.25 Device Reset Screen (Power Interruption):

Displayed as notification when power is interrupted during device operation.


F1 = OK: accepts notification and displays main screen
F2 = no function
F3 = no function
F4 = no function
CANCEL = no function
OK = accepts notification and displays main screen
UP = no function
DOWN = no function

Page 23
OMEGA OM-CP-RHTEMP2000

Section 7: Computer Interface:

1. Fully insert the male connector of the IFC110 interface cable into the female receptacle of the data logger. Insert
fully the RS232 connector into the Serial Port.
OR
2. Fully insert the male connector of the IFC200 interface cable into the female receptacle of the data logger. Fully
insert the female USB connector into the USB. (Please see the OMEGA Datalogger Software
manual for further information)
Note: Most OMEGA data loggers can use both OM-CP-IFC110 and OM-CP-IFC200 interface cables. For interface cable data
logger clarification contact technical support at 1-800-848-4286.

Section 8: Wall Mounting

Power Supply Cable Slot

Page 24
Data Logger Manual

Section 9: Maintenance

BATTERY WARNING
Most OMEGA data loggers contain a lithium battery. Do not cut the battery open, incinerate, or recharge. Do
not heat lithium batteries above the specified operating temperature.* Dispose of the battery in accordance
with local regulations.
*See the individual specifications at www.omega.com.

The OM-CP-RHTemp2000 does not have any user-serviceable parts except the battery which should be replaced periodi-
cally. The battery life is affected by battery type, ambient temperature, sample rate, sensor selection, offloads and LCD
display usage. The OM-CP-RHTemp2000 has a battery status indicator on the LCD display. If the battery indication is low,
or if the device seems to be inoperable, it is recommended that the battery be changed.

To change the battery, locate and remove the four (4) 3/32 hex screws on the back of the unit. Separate the halves and
the battery compartment is now visible. Remove the old 9V battery from the battery clips and replace with a new 9V bat-
tery. OMEGA recommends using 9V lithium battery. An alkaline battery is acceptable, but will yield a shorter battery life.

Page 25
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and
workmanship for a period of 61 months from date of purchase. OMEGA’s WARRANTY adds
an additional one (1) month grace period to the normal five (5) year product warranty to
cover handling and shipping time. This ensures that OMEGA’s customers receive maximum
coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer
Service Department will issue an Authorized Return (AR) number immediately upon phone or
written request. Upon examination by OMEGA, if the unit is found to be defective, it will be
repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting
from any action of the purchaser, including but not limited to mishandling, improper
interfacing, operation outside of design limits, improper repair, or unauthorized modification.
This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows
evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture
or vibration; improper specification; misapplication; misuse or other operating conditions
outside of OMEGA’s control. Components in which wear is not warranted, include but are not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However,
OMEGA neither assumes responsibility for any omissions or errors nor assumes
liability for any damages that result from the use of its products in accordance with
information provided by OMEGA, either verbal or written. OMEGA warrants only
that the parts manufactured by the company will be as specified and free of
defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY
KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL
IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total
liability of OMEGA with respect to this order, whether based on contract, warranty,
negligence, indemnification, strict liability or otherwise, shall not exceed the
purchase price of the component upon which liability is based. In no event shall
OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1)
as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or
activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or
with any nuclear installation or activity, medical application, used on humans, or misused in
any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER
language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from
any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.

RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department.
BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN
AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT
(IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be
marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to
prevent breakage in transit.
FOR WARRANTY RETURNS, please have FOR NON-WARRANTY REPAIRS, consult
the following information available BEFORE OMEGA for current repair charges. Have the
contacting OMEGA: following information available BEFORE
1. Purchase Order number under which contacting OMEGA:
the product was PURCHASED, 1. Purchase Order number to cover the
2. Model and serial number of the product COST of the repair,
under warranty, and 2. Model and serial number of the
3. Repair instructions and/or specific product, and
problems relative to the product. 3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible.
This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2007 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without
the prior written consent of OMEGA ENGINEERING, INC.
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at omega.comsm
TEMPERATURE

� Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies

� Wire: Thermocouple, RTD & Thermistor

� Calibrators & Ice Point References

� Recorders, Controllers & Process Monitors

� Infrared Pyrometers

PRESSURE, STRAIN AND FORCE



� Transducers & Strain Gages

� Load Cells & Pressure Gages

� Displacement Transducers

� Instrumentation & Accessories

FLOW/LEVEL

� Rotameters, Gas Mass Flowmeters & Flow Computers

� Air Velocity Indicators

� Turbine/Paddlewheel Systems

� Totalizers & Batch Controllers

pH/CONDUCTIVITY

� pH Electrodes, Testers & Accessories

� Benchtop/Laboratory Meters

� Controllers, Calibrators, Simulators & Pumps

� Industrial pH & Conductivity Equipment

DATA ACQUISITION

� Data Acquisition & Engineering Software

� Communications-Based Acquisition Systems

� Plug-in Cards for Apple, IBM & Compatibles

� Datalogging Systems

� Recorders, Printers & Plotters

HEATERS

� Heating Cable

� Cartridge & Strip Heaters

� Immersion & Band Heaters

� Flexible Heaters

� Laboratory Heaters

ENVIRONMENTAL
MONITORING AND CONTROL

� Metering & Control Instrumentation

� Refractometers

� Pumps & Tubing

� Air, Soil & Water Monitors

� Industrial Water & Wastewater Treatment

� pH, Conductivity & Dissolved Oxygen Instruments M4612-0108
M4597/1207
DOC-MAN040A-02

You might also like