Omega Om CP Rhtemp2000 Manual
Omega Om CP Rhtemp2000 Manual
Omega Om CP Rhtemp2000 Manual
MADE IN
Shop online at
omega.com
e-mail: [email protected]
For latest product manuals:
omegamanual.info
OM-CP-PRHTEMP2000
OM-CP-RHTemp2000
Temperature,
Humidity andHumidity, and
Temperature
Pressure
Data Logger withData Logger
LCD Display
with LCD Display
OMEGAnet ® Online Service Internet e-mail
omega.com [email protected]
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI
regulations that apply. OMEGA is constantly pursuing certification of its products to the European New
Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any
errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, human applications.
Data Logger Manual
Table of Contents
Section 9: Maintenance...........................................................................25
Page 2
Data Logger Manual
CAUTION: DEVICE MUST BE USED ONLY IN A MANNER CONSISTENT WITH THIS MANUAL.
NOTICE: WHEN 230VAC SUPPLY USED, OMEGA SPECIFIES USE OF AC SUPPLY PART # T35-9-100R-3
MANUFACTURED BY ENG ELECTRIC. THIS SUPPLY IS AVAILABLE FROM OMEGA.
1.2 Specifications
Temperature
Range: -20 to +60ºC
Resolution: 0.1ºC
Calibrated Accuracy: ±0.5ºC (0 to +50ºC)
Humidity
Range: 0 to 95%RH
Resolution: 0.1%RH
Calibrated Accuracy: ±3%RH (±2%RH typical at 25ºC)
Specified Accuracy Range: +10 to +40°C, 10 to 80%RH
Dot-Matrix LCD
Dimensions: 2.5” x 1.375” (63mm x 35mm)
Text: Configurable channel text size
Indicators: Power, status, memory
Backlight: Configurable w/auto shut-off and contrast adjustment
Start/Stop Time:
Software programmable start time and date, up to six months in advance; programmable stop time
Memory:
131,071 readings per channel; 262,143 total readings; software configurable memory wrap
Reading Rate:
1 reading every 2 seconds to 1 every 24 hours
Calibration:
Digital calibration through software
Calibration Date:
Automatically recorded within device
Battery Type:
9V lithium battery included, user replaceable; optional AC adapter, 7-24 VDC, 100mA. For 230 VAC
operation use AC supply T35-9-100 R-3 made by ENG Electric (Available from MadgeTech).
Page 3
OMEGA OM-CP-RHTEMP2000
Battery Life:
30 days typical @ 1 min reading rate with continuous LCD and no backlight usage.
Data Format:
Date and time stamped ºC, ºF, K, ºR ; %RH, mg/ml water vapor concentration, dew point
Time Accuracy:
±1 minute/month (at 20º to 30°C)
Computer Interface:
PC serial or USB (interface cable required); 115,200 baud
Software:
Windows 95/98/ME/NT/2000/XP based software
Operating Environment:
-20 to +60ºC, 0 to 95%RH non-condensing
Dimensions:
4.8” x 3.3” x 1.25” (122mm x 84mm x 32mm)
Weight:
16 oz (440 g)
Enclosure:
Black anodized aluminum
Page 4
Data Logger Manual
Power Jack
Temperature
and Humidity
LED Sensor
Indicators Ventilation
Holes
Communication
LCD Screen Jack
Up Key
Page 5
OMEGA OM-CP-RHTEMP2000
Full Empty
Battery Battery
Battery status icons indicate the state of charge of the batteries. When the battery empty indicator is displayed, the battery
should be replaced.
Memory Memory
Empty Full
The memory status icons indicate the amount of memory left for data storage.
If the OM-CP-RHTemp2000 is in wraparound mode, the memory status icon will always indicate empty.
Indicates that the OM-CP-RHTemp2000 is taking readings. The icons cycle periodically.
The delay start icon indicates that a logging operation is scheduled for the future.
Page 6
Data Logger Manual
Page 7
OMEGA OM-CP-RHTEMP2000
Large: Up to two channels can be displayed which will fill the entire display area
Page 8
Data Logger Manual
Page 9
OMEGA OM-CP-RHTEMP2000
5.2.iii Time
The time screen displays current time data including the current time and date, time and date of last measured data, and
current time zone.
5.4 Units
Pressing the F3 key while viewing the Main Screen will access the Units Selection screen. Here the measurement units can
be easily changed. In the Units Selection screen the F1 and F2 function keys correspond to the two channels temperature
and humidity. Selecting a channel by pressing its corresponding function key allows the user to change the units by either
pressing the function key repeatedly or using the UP and DOWN keys to scroll through the list of available units. Select-
ing OK (either by pressing the OK button or the F4 key) accepts and confirms the unit selection. Pressing the CANCEL
key cancels the unit changing action and reverts to the previously selected units. The OM-CP-RHTemp2000 offers most
commonly used units.
5.5.i Display
From the Display section, the number and identity of channels shown on the main screen can be changed and set, the
LCD screen’s contrast values can be changed and set, and the size the channel information appears on the main screen
can be changed and set. Pressing F1 while in the Setup Menu brings up the Display section.
Size
To change the size that each channel is displayed on the Main Screen view while viewing the Display Configuration
screen, first select the size configuration parameter by pressing the F2 key. This action will highlight the function
tab corresponding to the F2 key. Once this tab is highlighted the size can be increased or decreased using the UP
and DOWN keys. Sizes are as follows:
Small: Up to three channels can be displayed and appear much smaller than the available screen space.
Large: Up to two channels can be displayed which will fill the entire display area
Page 11
OMEGA OM-CP-RHTEMP2000
5.5.ii Power
The Power section displays information regarding battery and external power sources as well as giving options including
display and backlight status that could potentially save power and LED status options.
Display
The OM-CP-RHTemp2000’s LCD screen has options to be either on continuously or turn off automatically after five
minutes of inactivity. This option can be changed by first pressing the F1 key to highlight the display parameter
configuration tab. Once the tab is highlighted the option can be changed by pressing the F1 key repeatedly or
using the UP and DOWN keys to select either “On” or “Auto”.
Backlight
Like the display, the LCD screen’s backlight also has power saving options. The backlight can be either on continuously,
off, or set to automatically shut off after 30 seconds of inactivity. This option can be changed by first pressing the
F2 key to highlight the back light parameter configuration tab. Once the tab is highlighted the option can be
changed by pressing the F2 key repeatedly or using the UP and DOWN keys to select either “On”, “Off” or
“Auto”.
LED Modes
The function of the two LEDs on the upper left corner of the OM-CP-RHTemp2000 can be regulated by first pressing
the F3 key to highlight the LED parameter configuration tab. Once the tab is highlighted the option can be
changed by pressing the F3 key repeatedly or using the UP and DOWN keys to select either “Enabled” meaning the
LEDs will light to indicate device function or “Disabled” meaning the LEDs will never light.
Page 13
OMEGA OM-CP-RHTEMP2000
Displays information about run parameters including date of recording start (start date), time of recording start (start time),
stop date, stop time, recording rate, and current status.
For all status screens:
F1 = RUN: displays run parameters screen
F2 = MEM: displays memory status screen
F3 = TIME: displays time information screen
F4 = MORE: displays the second status screen
CANCEL = returns to main screen
OK = returns to main screen
UP = no function
DOWN = no function
Displays information about the device’s memory capabilities including percent of memory available (memory left), number of
readings taken so far (readings), max number of readings (max readings), and wrap.
Page 14
Data Logger Manual
Displays current time and date as well as registered time and date and time zone information.
For all status screens:
F1 = RUN: displays run parameters screen
F2 = MEM: displays memory status screen
F3 = TIME: displays time information screen
F4 = MORE: displays the second status screen
CANCEL = returns to main screen
OK = returns to main screen
UP = no function
DOWN = no function
6.5 ID Parameters
Page 15
OMEGA OM-CP-RHTEMP2000
Displays statistics (maximum recorded value, minimum recorded value, and average recorded value) based on channel
(Temperature or Humidity).
Displays statistics from both channels (Temperature and Humidity) sorted by type (maximum recorded value, minimum recorded
value, or average value).
F1 = AVG: displays average recorded value for each channel
F2 = MIN: displays minimum recorded value for each channel
F3 = MAX: displays maximum recorded value for each channel
F4 = EXIT: Returns to Statistics Menu
Page 16
Data Logger Manual
Displays current statistics information including the number of recorded readings, start date, and start time.
F1 = CLEAR: Gives the option to clear all statistics
F2 = no function
F3 = no function
F4 = EXIT: returns to Statistics Menu
CANCEL = returns to Statistics Menu
OK = returns to Statistics Menu
UP = no function
DOWN = no function
Displays options for changing the viewing of the two channels on the main screen (either shows a particular channel or hides it).
F1 = TEMP: first highlights and then changes the viewing options of the temperature channel
F2 = HUMID: first highlights and then changes the viewing options of the humidity channel
F3 = no function
F4 = MORE: moves on to Display Configuration screen
CANCEL = return to Device Configuration Menu
OK = moves on to Display Configuration screen
UP = once channel parameter configuration tab is highlighted scrolls through available options
DOWN = once channel parameter configuration tab is highlighted scrolls through available options
Allows the user to change the contrast of the LCD display as well as the channel size.
Page 18
Data Logger Manual
Displays information regarding the devices different power modes including the display visibility, backlight options, and LED
modes.
F1 = DISPLAY: first highlights and then changes display visibility (On: full visibility or Auto: shuts off after 5 minutes of
inactivity)
F2 = BKLGHT: first highlights and then changes backlight options (On: backlight always on, Auto: backlight shuts off after
30 sec of inactivity, or Off: backlight always off)
F3 = LED: first highlights and then changes LED mode options
F4 = MORE: moves to Display Update Mode screen
CANCEL = Returns to Device Configuration Menu
OK = moves to Display Update Mode screen
UP = once parameter configuration tab is highlighted scrolls through available options
DOWN = once parameter configuration tab is highlighted scrolls through available options
Displays details about power available to the device including the battery type, battery voltage, and external voltage.
F1 = TYPE: highlights and then changes battery type (Lithium or Alkaline)
F3 = MORE: moves back to Power Modes screen
F4 = EXIT: Returns to Device Configuration Screen
CANCEL = Returns to Device Configuration Menu
Page 19
OMEGA OM-CP-RHTEMP2000
Displays values indicating minimum temperature and humidity detectable by the OM-CP-RHTemp2000.
For all Device Information Screens:
F1 = MIN: no function
F2 = MAX: shows Device Range (Maximum) screen
F3 = CAL: shows Calibration Information screens
F4 = MORE: shows further Device Information screens
CANCEL = Returns to Device Configuration Menu
OK = Returns to Device Configuration Menu
UP = No function
DOWN = No function
Displays values indicating maximum temperature and humidity detectable by the OM-CP-RHTemp2000.
F1 = VERS: no function
F2 = FIRM: shows Firmware Version screen
F3 = MORE: shows Device Range (Minimum) screen
F4 = EXIT: Returns to Device Configuration Menu
CANCEL = Returns to Device Configuration Menu
OK = Returns to Device Configuration Menu
UP = No function
DOWN = No function
Page 20
Data Logger Manual
Page 21
OMEGA OM-CP-RHTEMP2000
Page 22
Data Logger Manual
Page 23
OMEGA OM-CP-RHTEMP2000
1. Fully insert the male connector of the IFC110 interface cable into the female receptacle of the data logger. Insert
fully the RS232 connector into the Serial Port.
OR
2. Fully insert the male connector of the IFC200 interface cable into the female receptacle of the data logger. Fully
insert the female USB connector into the USB. (Please see the OMEGA Datalogger Software
manual for further information)
Note: Most OMEGA data loggers can use both OM-CP-IFC110 and OM-CP-IFC200 interface cables. For interface cable data
logger clarification contact technical support at 1-800-848-4286.
Page 24
Data Logger Manual
Section 9: Maintenance
BATTERY WARNING
Most OMEGA data loggers contain a lithium battery. Do not cut the battery open, incinerate, or recharge. Do
not heat lithium batteries above the specified operating temperature.* Dispose of the battery in accordance
with local regulations.
*See the individual specifications at www.omega.com.
The OM-CP-RHTemp2000 does not have any user-serviceable parts except the battery which should be replaced periodi-
cally. The battery life is affected by battery type, ambient temperature, sample rate, sensor selection, offloads and LCD
display usage. The OM-CP-RHTemp2000 has a battery status indicator on the LCD display. If the battery indication is low,
or if the device seems to be inoperable, it is recommended that the battery be changed.
To change the battery, locate and remove the four (4) 3/32 hex screws on the back of the unit. Separate the halves and
the battery compartment is now visible. Remove the old 9V battery from the battery clips and replace with a new 9V bat-
tery. OMEGA recommends using 9V lithium battery. An alkaline battery is acceptable, but will yield a shorter battery life.
Page 25
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and
workmanship for a period of 61 months from date of purchase. OMEGA’s WARRANTY adds
an additional one (1) month grace period to the normal five (5) year product warranty to
cover handling and shipping time. This ensures that OMEGA’s customers receive maximum
coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer
Service Department will issue an Authorized Return (AR) number immediately upon phone or
written request. Upon examination by OMEGA, if the unit is found to be defective, it will be
repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting
from any action of the purchaser, including but not limited to mishandling, improper
interfacing, operation outside of design limits, improper repair, or unauthorized modification.
This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows
evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture
or vibration; improper specification; misapplication; misuse or other operating conditions
outside of OMEGA’s control. Components in which wear is not warranted, include but are not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However,
OMEGA neither assumes responsibility for any omissions or errors nor assumes
liability for any damages that result from the use of its products in accordance with
information provided by OMEGA, either verbal or written. OMEGA warrants only
that the parts manufactured by the company will be as specified and free of
defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY
KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL
IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total
liability of OMEGA with respect to this order, whether based on contract, warranty,
negligence, indemnification, strict liability or otherwise, shall not exceed the
purchase price of the component upon which liability is based. In no event shall
OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1)
as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or
activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or
with any nuclear installation or activity, medical application, used on humans, or misused in
any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER
language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from
any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department.
BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN
AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT
(IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be
marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to
prevent breakage in transit.
FOR WARRANTY RETURNS, please have FOR NON-WARRANTY REPAIRS, consult
the following information available BEFORE OMEGA for current repair charges. Have the
contacting OMEGA: following information available BEFORE
1. Purchase Order number under which contacting OMEGA:
the product was PURCHASED, 1. Purchase Order number to cover the
2. Model and serial number of the product COST of the repair,
under warranty, and 2. Model and serial number of the
3. Repair instructions and/or specific product, and
problems relative to the product. 3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible.
This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2007 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without
the prior written consent of OMEGA ENGINEERING, INC.
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at omega.comsm
TEMPERATURE
�
� Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
�
� Wire: Thermocouple, RTD & Thermistor
�
� Calibrators & Ice Point References
�
� Recorders, Controllers & Process Monitors
�
� Infrared Pyrometers
FLOW/LEVEL
�
� Rotameters, Gas Mass Flowmeters & Flow Computers
�
� Air Velocity Indicators
�
� Turbine/Paddlewheel Systems
�
� Totalizers & Batch Controllers
pH/CONDUCTIVITY
�
� pH Electrodes, Testers & Accessories
�
� Benchtop/Laboratory Meters
�
� Controllers, Calibrators, Simulators & Pumps
�
� Industrial pH & Conductivity Equipment
DATA ACQUISITION
�
� Data Acquisition & Engineering Software
�
� Communications-Based Acquisition Systems
�
� Plug-in Cards for Apple, IBM & Compatibles
�
� Datalogging Systems
�
� Recorders, Printers & Plotters
HEATERS
�
� Heating Cable
�
� Cartridge & Strip Heaters
�
� Immersion & Band Heaters
�
� Flexible Heaters
�
� Laboratory Heaters
ENVIRONMENTAL
MONITORING AND CONTROL
�
� Metering & Control Instrumentation
�
� Refractometers
�
� Pumps & Tubing
�
� Air, Soil & Water Monitors
�
� Industrial Water & Wastewater Treatment
�
� pH, Conductivity & Dissolved Oxygen Instruments M4612-0108
M4597/1207
DOC-MAN040A-02