Lab Report 3 Et
Lab Report 3 Et
SEMESTER II 2021/2022
PRACTICAL 3
SECTION 01
PERSON IN CHARGE:
Marks Distribution:
Ion exchange chromatography is known for its speciality to separates proteins based
on the charge of protein and the pH of the buffer used. In this case, the protein that was used
was Bovine Serum Albumin with pI value of ~4.7. The buffer that was used need to be either
above or below the isoelectric point of protein. This is to allow the separation or the binding
of the protein with the stationary phase to occur. This will also change the charge of the
protein. If the pH of the buffer is above the pI value it is called Anion exchange
chromatography. However, if it is below the pI value, then it is a cation exchange
chromatography. This is depending on the target proteins charge as if it is negatively charged,
it is called Anion exchange chromatography and if it is positively charged molecule is called
Cation exchanged chromatography.
2.0 OBJECTIVES
Firstly, 10ml column was prepared by using the glass column. The pump and the adaptor
were attached. Secondly, the flow rate was adjusted to 1-2 ml/ minute. Thirdly, 10 ml of Tris-
HCL buffer was used for washed the column. It should be same buffer to a height of 2cm to
avoid air bubbles and dry. Next, 1ml of protein was added into the column by using the
pipette. Other than that, 25ml of Tris-HCL buffer were added into the column and 5ml x 5
fractions were collect. Moreover, 25ml of 50mM NaCl were added into the column and 5ml x
5 fractions were collected to measure the absorbance. Furthermore, 25ml of 100mM NaCl
were added into the column and 5ml x 5 fraction was collect. Lastly, all the fraction were put
into the cuvette to analyse the absorbance at wavelength 600nm.
4.0 RESULT
OD600nm (optical density at 600nm wavelength) readings for the purification of BSA using
anion exchange chromatography. [blank absorbance: 0.000 A]
The flow rate was set to 1 to 2 ml/min, which is within the optimal range for an ion-
exchange column. The flow rate of a buffer through a resin is referred to as the flow rate. The
flow rate, also known as the residence time of a specific column at a certain flow rate,
determines the number of times proteins may interact with the column resin. The resolution
and capacity might be affected by the flow rate. This means that the longer the residence
time, the better the resin's capacity and resolution. The ion exchange adsorbent packed into
the glass column is pre-equilibrated in tris-HCL buffer before running the BSA protein
mixture by anion-exchange chromatography, as stated earlier, with the purpose of adjusting
the protein charge to become negatively charged. The pH of the BSA protein has been
determined to be 4.7 pH units. The isoelectric point's pH value, also known as the pH value,
is defined as the pH value at which all charged molecules in the protein solution have a net
charge of zero. Ion exchange is normally done at least one pH unit distant from the protein of
interest's pI to ensure that it is charged. When the pH reaches the pI, the molecule is
negatively charged. When the pH reaches the pI, the molecule becomes negatively charged,
requiring the use of an anion exchange resin, and vice versa. When flow rates increase,
however, pressure on the resin increases until the backpressure reaches lit, causing the
column to crush and the separation efficiency to decrease. Slower flow rates, on the other
hand, may give better resolution and capacity, but at the expense of degraded protein activity
in the chromatographic system.
The BSA protein mixture was then put into the column for ion exchange and ion
exclusion, where it was exposed to ion exchange and ion exclusion. The former occurred
when positively charged ions in the stationary phase were attached to their opposite charge
exchangeable ions (target protein), whereas the latter occurred when analytes of similar
charge repel the ions on the stationary phase's functional group, as in the case of the target
protein. Coulomb's law states that electrostatic forces cause interactions between ions in the
solute and oppositely charged ligands on the matrix in ion-exchange chromatography. If the
charges on both ions are the same, the force is repulsive; however, if the charges are
different, the force is attracting. After the buffer has passed through the column, it is washed
to remove any unbound contaminants as well as any uncharged or positively charged proteins
that may have collected on the column during the buffer's passage. Chromatography is done
in a positive mode, which means that the washing phase helps remove unwanted compounds
while the protein of interest is kept in a negative charge throughout the procedure. Washing
the chromatographic system with 2 column volumes of 100 ml of equilibration buffer and
collecting 5 ml fractions. In this step, utilise the same buffer, Tris-HCl, with the same pH to
help the target protein stay bound to the stationary phase. Because the total volume of Tris-
HCl equilibration buffer is 100 ml, and 5 ml fractions were obtained each time, there are 15
fractions. A spectrophotometer calibrated to wavelength 660nm was used to determine the
absorbance value of all fractions from the washing process. Samples 3 to 5 show absorbance
values of 0.004A, 0.007A, and 0.002A, respectively, according to the findings. This means
that the impurities have been thoroughly removed.
The second step involves elution of the required negatively charged protein with
gradually increasing salt concentration (linear salt gradient). The elution procedure employed
two NaCl salt buffer concentration levels: 50mM and 100mM. The initial eluent was a 25mL
50mM NaCl buffer, and protein with a low net charge was washed away. Following that, the
column was rinsed with additional 25mL of 100mM NaCl buffer. In addition, 5ml fractions
were taken from each of the ten fractions, for a total of 100 fractions collected. The
absorbance was measured with a spectrophotometer set at 660nm. According on the results of
the elution with 50mM salt banquet, sample 1 has -0.001A, sample 2 has -0.004A, sample 3
has -0.004A, sample 4 has -0.003A, and sample 5 has -0.000A. This section has all of the
negative value for the absorbance readings.
This is due to the reference's absorbance value is higher than the sample's, sample
measurements. The sample and reference have been mixed up, or the sample is very dilute
and close to the absorbance of the reference. In the meanwhile, the absorbance readings for
sample 1 was 0.010A, sample 2 was 0.002A, sample 3 was 0.016A, sample 4 was 0.005A,
and sample 5 was 0.005A for the elution with 100mM salt buffer. This indicated that all of
the strongly bound target negatively charged protein was clotted out and properly purified,
however there are some inconsistencies in the absorbance measurement from what we can
observed. The stationary phase resin had been regenerated and was ready for use in the
subsequent step.
In practise, protein purification is followed by 1D SDS-PAGE electrophoresis to
ensure that the protein purified contains the required protein. In this experiment, protein
purification is terminated at ion-exchange chromatography. 1D electrophoresis is a method
for separating proteins with molecular sizes ranging from 10 to 300 kilodaltons depending on
their molecular weight (kDa). Only one band will appear on the 1D SDS-PAGE result,
showing that the target protein was successfully purified. To ensure the precision and
consistency of the findings obtained during the experiment, we must take a few precautions
step to ensure that a consistent flow rate in the range of 1-2mV/min is maintained throughout
the experiment.
At that peak the LDH was pure. Ion exchange chromatography occurred due to
electrostatic attraction between buffer-dissolved charged proteins and oppositely charged
binding sites on a solid ion exchange adsorbent. Graph of the Absorbance plotted versus
fractions number at 600 nm. The fractions collected were assayed for total protein. The
values were plotted against the fraction number which is called elution profile. The peak on
the graph indicated where the fractions are that contain protein. Enzyme activity was
determined by performing an enzyme assay on each fraction that contain protein
In this experiment, a 10ml column was prepared using a glass column. The pump and
adaptor were connected. The flow rate was then reduced to 1-2 ml/min. Finally, the column
was washed with 10 ml of Tris-HCL buffer. To avoid air bubbles and dry, use the same
buffer up to a height of 2cm. The pipette was then used to add 1ml of protein to the column.
In addition, 25ml of Tris-HCL buffer was added to the column, and 5ml x 5 fractions were
collected. In addition, 25ml of 50mM NaCl was added to the column, and 5ml x 5 fractions
were collected to determine absorbance.
There are precaution steps to avoid the error that can affect the accuracy of result in
this experiment. First, handle the samples carefully to avoid contamination for high
sensitivity analysis. Next, as a stationary phase in the ion separation column, a packing
material modified with ion-exchange groups is used. As a result, it has greater
hydrophobicity, which can lead to the accumulation of hydrophobic components from the
sample, which can have negative consequences. Moreover, Handle the column with care, as
hitting it or exposing it to sudden pressure changes or high pressures can cause deterioration.
Precipitation should be avoided in the column. Because tris-buffers are temperature sensitive,
they must be adjusted to pH 7.8 at room temperature.
7.0 REFERENCES