PH Controller
PH Controller
PH Controller
D2C2-001-pH/pH-GB
pH/pH
7.57pH 7.63 pH
DULCOMETER
STOP
START
pH/pH
7.57 pH 7.63 pH
DULCOMETER
STOP
START
Type D
Type W
D2C A
___ ___ ___ ___ ___ ___ ___ ___ ___ ___ ___
Please enter the identity code of your device here!
Please completely read through the operating instructions! Do not discard! The warranty shall be invalidated by damage caused by operating errors!
Part No. 987661 ProMinent Dosiertechnik GmbH D-69123 Heidelberg Germany BA DM 142 01/07 GB
ProMinent
1
D2C A
D2C A
1 2 3 4 5 6 7 8 9 10
Device Identification / Identity Code .................................................................................................... 2 General User Information ...................................................................................................................... 3 Device Overview / Controls ................................................................................................................... 4 Functional Description .......................................................................................................................... 5 Display Symbols .................................................................................................................................... 6 Operation diagram ................................................................................................................................ 7 Operation Menu / Overview .................................................................................................................. 8 Operation Menu / Description ............................................................................................................. 12 Technical Terms .................................................................................................................................. 22 Faults / Notes / Troubleshooting ........................................................................................................ 24
IMPORTANT Please observe the parts of these operating instructions applicable to your particular version! This is indicated in the Section Device Identification / Identity Code. Correct measuring and metering is only possible in the case of impeccable operation of the sensor. The sensor has to be calibrated / checked regularly! In the event of a sensor failure, uncontrolled chemicals addition may result. We therefore recommend urgently to activate checkout time limits with automatic control shut-off! NOTE A form Documentation of controller settings type D2C, pH/pH is available under www.prominent.com/documentation_D2C_pH_pH for the purpose of documenting the controller settings.
"Up" Button
D2C2-002-pH/pH-GB
CHANGE button To change over within a menu level and to change from one variable to another within a menu point.
UP button To increase a displayed numerical value and to change variables (flashing display)
START/STOP button
STOP
START
BRANCH BACK button Back to permanent display or to start of relevant setting menu. DOWN button To decrease a displayed numerical value and to change variables (flashing display).
Start/Stop of control and metering function. ENTER button To accept, confirm or save a displayed value or status. For alarm acknowledgement.
Functional Description
NOTE Please refer to the description of the operating menu for a detailed description of the individual characteristics of the DULCOMETER D2C controller.
4.1
Operating Menu
The DULCOMETER D2C controller permits settings to be made in two different menus. All values are preset and can be changed in the complete operating menu. The controller is delivered with a restricted operating menu so that the DULCOMETER D2C controller can be used effectively in many applications from the very onset. If adaptations prove to be necessary, all relevant parameters can then be accessed by switching over to the complete operating menu. 4.2 Access Code
Access to the setting menu can be prevented by setting up an access code. The D2C controller is supplied with the access code 5000 which permits free access to the setting menu. The calibration menu remains freely accessible even if access to the settting menu is blocked by the code. 4.3 Control
The D2C can operate as a proportional controller or as a PID controller - depending on the device version (see identity code) and the setting. The controller variable is recalculated every second. Control procedures which require rapid correction of setpoint deviations (less than approx. 30 seconds) cannot be processed with this controller. Take cycle times into account in the case of solenoid valve control (pulse length). The control function (control variable output) can be deactivated via the pause function and the sample water control input. The calculation of the regulated variable starts again with the inactivation of the "pause" after expiry of the adjustable delay time "td". No fault treatment is performed with active "pause" function. 4.4 Fault messages
Faults to be acknowledged are shown in the permanent displays 1, 3 and 4 by the symbol " ". The corresponding fault messages and notes are shown in the permanent display 2. Faults/notes continuing after acknowledgement are shown alternatively. Faults automatically remedied through changing operating conditions are removed from the permanent display without necessitating acknowledgement. Chapter 10 includes an overview of fault messages and causes.
Display Symbols
The display of the DULCOMETER D2C controller uses the following symbols: Symbol Description Limit value transgression Relay 1 upper or zone Relay 1 lower Limit value transgression Relay 2 upper or zone Relay 2 lower Metering pump 1 (acid) Control off Control on Metering pump 2 (alkali) Control off Control on Solenoid valve 1 (acid) Control off Control on Solenoid valve 2 (alkali) Control off Control on Stop button pressed Manual metering Delay time td Fault Control starts after expiry of td Symbol left Symbol left Symbol right Symbol right Symbol left Symbol left Symbol right Symbol right left Symbol left Symbol right Symbol right Comment Symbol
pause
Operation diagram
Permanent display 1
Permanent display 2
The various menus are selectec with the CHANGE button The menu is started with the ENTER button
Permanent display 3
Permanent display 4
Calibration menu
Calibration notes
Various
Setting menus
Access code
D2C2-007-pH/pH-GB
NOTE Access to the setting menus can be barred with the access code! The number and scope of setting menus depends on the device version! If the access code is selected correctly in a setting menu, the following setting menus are also accessible! If within a period of 10 minutes no button is pushed, the unit automatically branches back from the setting menu to the permanent display 1.
Variables flash
Text 1 Selection 1 Text 2 Selection 2
The setting menus highlighted in grey and the adjustable parameters are only visible in the complete operating menu.
7.20 pH 1 7.50 pH 2
value 1 set point ctrlout temp. 1: 7.00 pH 7.20 pH 0% 25 C
value 2 temp. 2:
7.50 pH 25 C
calibration?
calibration MV1 zero p.: -3.1 mV slope at 25 C 53.70 mV/pH calibration MV2 zero p.: -3.1 mV slope at 25 C 58.32 mV/pH calibration MV2 buffer 1: 7.00 pH mea. val 2: 7.00 pH signal: -3.1 mV calibration MV2 probe in buffer 1 buffer: 7.00 pH temp. 2: 25.0 C calibration MV2 calib. 1 active please wait! -3.1 mV 25.0 C
calibration MV2 zero p.: -3.1 mV slope at 25 C 53.70 mV/pH calibration end
limits setting?
limit value 1
limit value 2
limit 2 upper MV 2 6.50 pH limit 1 lower MV 2 8.00 pH only with limit relay limit 2 upper MV 2 6.50 pH limit 1 lower MV 2 8.00 pH limit 2: upper MV : 0.50 pH LV relay 2: LV1 - active closed t on 0s t off 0s hysteresis limits: 0.05 pH checkout time limits: off hysteresis limit: 0.04 ppm checkout time limit: off
limit value
limit end
Automatic control
setpoint MV1: 7.20 pH dosing direction: acid/alkali
Manual control
manual dosing MV1: 0 % dosing direction: acid/alkali
control end
pumps setting?
relay setting?
relay 1
relay 2
relay end
mA output 1 measured value 1 4 mA = 2.00 pH 20 mA = 12.00 pH mA outputs setting? mA output 1 mA output 1 measured value 1 4 ... 20 mA mA output 1 correcting value 1 4 mA = 0.0 C 20 mA = 100.0 C mA output 1 regulated value 1 4 mA = 0.0 % 20 mA = 100.0 %
mA output 2
mA output end
10
Permanent display 1
11
Permanent displays
7.20 pH 1 7.50 pH 2
value 1 set point ctrlout temp. 1: 7.00 pH 7.20 pH 0% 25 C
value 2 temp. 2:
7.50 pH 25 C
The permanent displays 1 to 4 serve information on fault messages/causes (see also table on page 24) as well as operating values/settings. Calibration Calibration of pH sensors (measuring value 1, 2): The calibration of the pH sensors MV 1 and MV 2 uses a two-point calibration method (zero point, slope). As buffer pH 7 (zero point calibration) and pH 4 (slope calibration) are factory-set. If other buffers are to be used, the defaults in the complete operating menu (menu A, B) may be altered. If the temperature correction takes place manually or automatically, the actual buffer temperature must be entered or the Pt 100 immersed in the buffer solution. During calibration, control is stopped and metering is reduced to the set basic load. The output 0/4...20 mA (measuring value and correcting value) will be frozen. After successful calibration, all fault determinations relating to the measuring value are started again. The current sensor data (zero point/slope) will be displayed.
Menu A
calibration? calibration MV1 zero p.: -3.1 mV slope at 25 C 58.32 mV/pH calibration MV1 buffer 1: 7.00 pH mea. val: 7.00 pH signal: -3.1 mV calibration MV1 probe in buffer 1 buffer: 7.00 pH temp. 1: 25.0 C calibration MV1 calib. 1 active please wait! -3.1 mV 25.0 C
Menu B
calibration MV1 buffer 2: 4.00 pH mea. val: 4.00 pH signal: 157 mV calibration MV1 probe in buffer 2 buffer: 4.00 pH temp. 1: 25.0 C calibration MV1 calib. 2 active please wait! 157 mV 25.0 C
Menu A
calibration MV2 zero p.: -3.1 mV slope at 25 C 58.32 mV/pH calibration MV2 buffer 1: 7.00 pH mea. val 2: 7.00 pH signal: -3.1 mV calibration MV2 probe in buffer 1 buffer: 7.00 pH temp. 2: 25.0 C calibration MV2 calib. 1 active please wait! -3.1 mV 25.0 C
12
calibration MV2 zero p.: -3.1 mV slope at 25 C 53.70 mV/pH calibration end
Possible values Initial value Buffer values pH 7 pH 4 Increment pH 0.01 Lower value pH -2 Upper value pH 16 Remarks Error messages when both buffers too close (<2 pH-values)
Condition Buffer < pH 2 < -60 mV > +60 mV < 40 mV/pH > 65 mV/pH
Effect During calibration procedure: recalibrate buffer 2! Return to permanent display: Basic metering load Warning, old zero point and slope retained " " " " " " " " "
pH zero point low pH zero point high pH slope low pH slope high Measured value pH unsteady Measured value temp. unsteady
Limit values
When setting the checkout time, metering of the corresponding pump is stopped and an alarm is triggered through the alarm relay in the event of limit violations (LV measured value 1, LV measured value ) exceeding the set checkout time. The control time can be reset by pressing the enter key! For devices with limit relays, a limit value or a zone may be set for each measuring value, where the relay will switch.
13
limits setting?
limit value 1
limit value 2
limit 2 upper MV 2 6.50 pH limit 1 lower MV 2 8.00 pH only with limit relay limit 2 upper MV 2 6.50 pH limit 1 lower MV 2 8.00 pH limit 2: upper MV : 0.50 pH LV relay 2: LV1 - active closed t on 0s t off 0s hysteresis limits: 0.05 pH checkout time limits: off hysteresis limits: 0.20 pH checkout time limits: off
limit value
limit end
Initial value Type of limit transgression Limit value 1 Limit value 2 Limit value Limit value Measured value 1 Limit relay 1 lower upper upper pH 6.5 pH 8.0 zone*
Lower value
Upper value
pH -2.0 pH -2.0
pH 16 pH 16 *with regard to the setting zone, the difference between the limits and the set hysteresis should be >3 x.
pH -2.0 pH -2.0
pH 16 pH 16
14
Initial value Measured value Limit relay 2 Limit relay 1, 2 pH 0.5 LV 2 active closed
Remarks
Switch-on delay t on Switch-off delay t off Hysteresis limits Measured value 1 Measured value 2 Measured value Checkout time limits
0s 0s
pH 16 pH 16 pH 14 9999 s
Control
Automatic control
setpoint MV1: 7.20 pH dosing direction: acid/alkali
Manual control
manual dosing MV1: 0 % dosing direction: acid/alkali
control end
15
Lower value
Upper value
Remarks
Metering direction MV 1
acid/alkali
Setpoint 1/2 Control parameter xp Control parameter Ti Control parameter Td Additional load Manual metering Pumps
The maximum stroke value of the metering pumps should correspond to the stroke frequency of the metering pump used.
pumps setting?
Possible values Initial value Max. strokes/min. for pumps 1 and 2 Pump 2 180 Increment 1 Lower value 1 Upper value 500 Remarks
Metering direction
Measured value 1 Measured value 1 Measured value 2 Measured value off alkali acid alkali
16
relay setting?
relay 1
relay 2
relay end
Lower value
Upper value
Remarks
Measured value 1 Measured value 1 off as per identity code Limit Actuator Solenoid valve
Measured value 1 Measured value 1 off as per identity code Limit Actuator Solenoid valve 1s 1s 10 s 1s 9999 s period/2
10 s 1s
17
Possible values Increment automatic manual off automatic manual off 0.1 C 0.1 C
Lower value
Upper value
Remarks
Measured value 2
manual
0.0 C
100 C
18
mA output 1 measured value 1 4 mA = 2.00 pH 20 mA = 12.00 pH mA outputs setting? mA output 1 mA output 1 measured value 1 4 ... 20 mA mA output 1 correcting value 1 4 mA = 0.0 C 20 mA = 100.0 C mA output 1 regulated value 1 4 mA = 0.0 % 20 mA = 100.0 %
mA output 2
mA output end
Possible values Initial value Allocation mA-output 1 Increment Lower value Upper value Remarks
Measured value 1 Measured value 1 Measured value 2 Regulated value 1 Correcting value 1 Correcting value 2 off Measured value 2 Measured value 1 Measured value 2 Correcting value 1 Correcting value 2 Regulated value 1 off 4...20 mA 4...20 mA 0...20 mA 19
Allocation mA-output 2
Signal range
20
Permanent display 1
Possible values Initial value Alarm relay Pause active active closed Alarm off td: 10 min Control input sample water active closed Access code Language 5000 as per identity code Increment active not active active closed active open Alarm off Alarm on 1 min sample water off active closed active open 1 German English French Spanish Swedish Dutch restricted complete 1 9999 0 min 60 min Lower value Upper value Remarks
Operating menu
restricted
21
Technical Terms
Additive basic load: This results in the fact that the controller always generates a manipulated value corresponding to the additive basic load. This load may only be reset to 0 using the stop button. This function should not be activated when using PI or PID controllers. Calibration: By calibrating (adjusting), the measuring value readout will be adjusted to the actual sensor signal. Without calibration, a correct measurement is not possible. A calibration should be performed regularly (depending on application). If the DULCOMETER D2C operates with automatic or manual temperature correction, a Pt 100 must be immersed in the buffer solution or the buffer temperatures entered manually. The signal of the pH sensors (e.g. 59 mV/pH at 25 C) is influenced by temperature. High temperatures increase the signal while low temperatures reduce the signal. The resulting measurement errors can be compensated for by way of temperature correction. As a rule, no correction is necessary for measurements conducted within the range around pH 7. However, if the measuring range is further away from pH 7, automatic temperature correction via Pt 100 should be implemented in the case of fluctuating temperatures. Manual temperature correction can be used if temperatures are constant. Automatic temperature correction can only be used for both measuring sensors if the temperature of the measured water is identical. 2 setpoints can be specified in the "control with dead zone" setting. If the measured value is outside the setpoint range defined in this way, the corresponding actuator is activated. If the measured values are within the setpoint range, the controller functions as if the setpoint were reached. In the case of a P-controller, a controlled variable is then no longer produced. In the case of PI-controllers, the last controlled variable is retained. In the limit value setting, the difference of both pH measurements is monitored. If monitoring time is activated, an alarm is automatically triggered via the alarm relay and the control variable is reset to 0 if values exceed the preset difference. This function serves the purpose of monitoring the pH measurement in critical applications. In this setting, the controller produces a controlled variable corresponding to the entry. It is retained up to the next change. It is independent of the measured variable and the set control parameters. This setting can be used for determining the time response (e.g. dead time...) of the controlled system.
Control parameter: The control parameters (xp, Ti, Td) determine the control characteristic (PID). Correction value:
Dead zone:
Limit value :
Manual control:
Metering direction: This value determines in which direction the controller is active. In case of the metering direction "acid", the controller generates a manipulated value when the specified value for pH is exceeded. If the metering direction is set to acid/alkali solution, the controller operates as a 2-side controller. Regulated value: Relay: The regulated value is the value (e.g. frequency, mA signal) the controller sends to the final controlling element, e.g. metering pump to reach again the set point. The relay (alarm, limit relay) switches when the corresponding prerequisites (e.g. alarm condition, limit violation) are given. The relay function can be set either as make contact (active = closed) or break contact (active = open). The relay may be reset pressing the stop button. Exception: limit value. Activation of solenoid valves (motor-driven pumps) is defined by the cycle time and the minimum on-state interval (minimum time) (pulse length control). The on-state interval always corresponds to at least the minimum time. However, it is increased up to the cycle time at a maximum depending on the control deviation and the control response. The cycle time itself defines the maximum possible on-state operations. For instance, an actuator is switched on a maximum of 60 times per hour when the
Solenoid valve:
22
Technical Terms
cycle time is at 60 seconds. The minimum time defines the minimum on-state interval duration. It should be selected as small as possible while, however, ensuring that metering is still possible within this time. Set point: Slope: Temperature correction: Ti (integralaction time): Td (derivative action time): xp-value: The set point is the value which is to be continuously maintained stable throughout the process via controlling. The slope of pH sensors should always be >50 (better >55) mV/pH. See correction value This term defines the integral (I) control response. The greater the Tn, the weaker the I-component. This term defines the differential (D) control response. The smaller the Td, the weaker the D-component. This value influences the proportional control behaviour. In case of a deviation of 1.4 pH (=10% of 14 pH) resp. 0.2 ppm (=10% of 2 ppm), a xp value of 10 %, for example, leads to a regulated value of 100%. If the xp value has to be increased to 20 %, the deviation must be double the value in order to reach a regulated value of 100 %. In case of control overshooting, the set xp value must be doubled. The zero point of pH sensors is theoretically 0 mV. In practice, for a good sensor function a zero point of 25 mV is acceptable. The zero point of the chlorine sensors is at 4 mA. Calibration is not necessary.
Zero point:
23
Fault
Fault text
Symbol
Effect on on control metering Basic load Basic load Stop no Stop yes
Remarks
Remedy
Calibration with error Measured value 2 Signal exceeded/ drops below value Calibration with error Limit transgression after checkout time
Check sensor, clean, replace if necessary, recalibrate if necessary Function detachable. Confirmation cancels the fault.
yes
Operation
Note text
Symbol
Remarks
Remedy
none none
Stop Stop
yes yes
Function switchable
Send in device
10
1999 ProMinent Dosiertechnik GmbH D-69123 Heidelberg Operating Instruction DULCOMETER D2C, Part 2, pH/pH Subject to technical modifications Printed in Germany Addresses and delivery information may be obtained from the manufacturer. ProMinent Dosiertechnik GmbH Im Schuhmachergewann 5-11 D-69123 Heidelberg Phone: +49 (6221) 842-0 Fax: +49 (6221) 842-419 [email protected] www.prominent.com
24