0% found this document useful (0 votes)
161 views74 pages

Corlyte Service Manual

Uploaded by

Ismael Morinigo
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
161 views74 pages

Corlyte Service Manual

Uploaded by

Ismael Morinigo
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 74

Service Manual

CorLYTE

ELECTROLYTE
ANALYZER
Na+/K+/Cl-/Ca++/Li+
Important Information!
This Service Manual contains important warnings and safety information to be observed by the user.

This instrument is only intended for one area of application which is described in the instructions. The
most important prerequisites for application, operation and safety, are explained to ensure smooth
operation. No warranty or liability claims will be covered if the instrument is applied in areas other
than those described or if the necessary prerequisites and safety measures are not observed.

The instrument is only to be operated by qualified personnel capable of serving these prerequisites.

Only accessories and supplies either delivered by or approved by your distributor are to be used with
the instrument.

Due to the operating principle of the instrument, analytical accuracy not only depends on correct
operation and function, but also upon a variety of external influences beyond the manufacturer’s
control. Therefore, the check results from this instrument must be carefully examined by an expert,
before further measures are taken based on the analytical results.

Instrument adjustment and maintenance with removed covers and connected power mains are only to
be performed by a qualified technician who is aware of the dangers involved.

Instrument repairs are only to be performed by the manufacturer or qualified service personnel.

Symbol Explanation

Attention symbol - Refer to the Operator's Manual or Service Manual for further
instructions. This symbol is located on the inside of the instrument.

Type B instrument symbol - An instrument of the B type falls under safety


categories I, II, or III, or has an internal power supply providing the required
insulation against discharge current and reliable ground connections.

Important Information!

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 2 of 74


Operating Safety Information
• This instrument falls under Safety Category I.
• This instrument is a Class B instrument.

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) this device may not cause harmful interference's, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
Warning: Changes or modifications to this unit not expressly approved by the party
responsible for compliance could void the user's authority to operate the equipment.

Note: This equipment has been checked and found to comply with the limits for a Class B
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio communication.
However, there is no guarantee that interference will not occur in a particular installation. If this
equipment does not cause harmful interference to radio or television reception, which can be
determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures :

• Reorient or relocate the receiving antenna


• Increase the separation between the equipment and receiver
• Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
• Consult the dealer or an experienced radio TV technician for help

Caution:
• The instrument is designed as a conventional device (closed, not waterproof type).
• Do not operate the instrument in an explosive environment or in the vicinity of explosive
anesthetic mixtures containing oxygen or nitrous oxide.
• This instrument is suitable for continuous operation.
• The power plug is to be plugged into a ground socket only. When using an extension cord, make
sure that it is of the proper size and is properly grounded.
• Any breakage of the ground lead inside or outside the instrument or a loose ground
connection can cause a hazardous condition when operating the instrument. Intentional
disconnection of the grounding is not permitted.
• When replacing the fuses, make sure that they are of the same type and rating as the
original fuses. Never use repaired fuses or short-circuit the fuse holders.

Operating Safety Information

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 3 of 74


Table of Contents
1. Introduction ..................................................................................................................................................... 5
2. General Description .......................................................................................................................................... 6
2.1. Reported Configurations ........................................................................................................................... 6
2.2. Operating Configurations .......................................................................................................................... 6
3. Software Operation ........................................................................................................................................... 9
3.1. Software Operation – CorLYTE................................................................................................................ 9
3.2. Service Menu .......................................................................................................................................... 26
4. Description of Modules ................................................................................................................................... 31
4.1. Mechanical Assemblies ........................................................................................................................... 31
4.1.1. Front Door Assembly...................................................................................................................... 31
4.1.2. Aspiration Assembly....................................................................................................................... 32
4.1.3. Electrode Tray Assembly................................................................................................................ 33
4.1.4. Peristaltic Pump Rotor E-728 ......................................................................................................... 33
4.1.5. Peristaltic Pump Motor E-727 ........................................................................................................ 34
4.1.6. Solenoid Valve Assembly ............................................................................................................... 35
4.1.7. Printer Assembly E-715 .................................................................................................................. 37
4.1.8. Power Supply Subassembly E-714 ................................................................................................. 38
4.1.9. CPU Board CE-730......................................................................................................................... 39
4.1.10. RFID Board CN-A4054 .............................................................................................................. 40
4.1.11. Valve Board E-732 ...................................................................................................................... 41
4.1.12. Display Board E-731 ................................................................................................................... 41
4.2. Fluid Pack ............................................................................................................................................... 42
4.2.1. Standard A ...................................................................................................................................... 42
4.2.2. Standard B and C ............................................................................................................................ 42
4.2.3. Reference Solution .......................................................................................................................... 42
4.2.4. Waste Liquids ................................................................................................................................. 42
4.2.5. Pinch Valve Tubing Set, E-711 ...................................................................................................... 44
4.3. Electronics .............................................................................................................................................. 46
4.3.1. CPU Board ...................................................................................................................................... 47
4.3.2. Valve Board E-732 ........................................................................................................................ 49
4.3.3. Display Board E-731....................................................................................................................... 49
4.3.4. Graphic Printer E-715 ..................................................................................................................... 50
5. Maintenance and Troubleshooting .................................................................................................................. 51
5.1. Maintenance ............................................................................................................................................ 51
5.2. Error Messages and Troubleshooting ...................................................................................................... 52
6. Adjustments .................................................................................................................................................... 60
6.1. Sample Sensor Adjustment ..................................................................................................................... 60
6.2. Temperature Adjustment ......................................................................................................................... 61
6.3. Display, Contrast and Backlight Adjustment ........................................................................................... 62
7. Illustrated Parts List ....................................................................................................................................... 63
8. Interface Specifications .................................................................................................................................. 68
8.1. Interface Information............................................................................................................................... 68
8.2. Example Data String Information ............................................................................................................ 69
8.3. Example Data String Information ............................................................................................................ 70
1 Introduction

1. Introduction

The Service Manual for the CorLYTE Electrolyte Analyzer contains the technical information
needed to ensure easy fault identification. This manual is intended to be complementary to the
Operator's Manual where detailed instructions for operation, maintenance and troubleshooting are
provided.

As for all clinical instrumentation, a thorough understanding of the principles of operation is


prerequisite to attempting service of this product. Training along with experience will enhance the
use of this manual.

Service and repair of the CorLYTE analyzer should be performed only by qualified repair
technicians. Care should be taken when removing the covers as hazardous voltages are exposed.
Use only accepted electronic check procedures and static protection when replacing and handling
all electronic parts.

This manual is divided into 9 chapters to facilitate location of technical information. Chapter 2
provides specifications and operating configuration information. Chapter 3 includes flow charts for
all system functions and detailed operation of system check procedures. In Chapter 4, all
mechanical, fluidic and electronic assemblies are described. Chapter 5 outlines routine maintenance
and troubleshooting procedures. Chapter 6 includes electronic adjustments and Chapter 7 provides
the system block diagram and all circuit schematic diagrams and wiring interconnection
information. In Chapter 7, part identification, location and description is provided by the illustrated
parts list. Finally, in Chapter 8 interface specifications are provided for the RS232 output.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 5 of 74


2 General Descriptions

2. General Description

Specifications

2.1.Reported Configurations
Measuring Range Resolution

Whole Blood
Serum
Plasma

Sodium Na+ 40 - 200 mmol/L 0.1 mmol/L

Potassium K+ 1.5 -15 mmol/L 0.01 mmol/L

Chloride Cl- 50 - 200 mmol/L 0.1 mmol/L

Calcium Ca++ 0.3 - 5.0 mmol/L 0.01 mmol/L

Lithium Li+ 0.2 – 5.5 mmol/L 0.001 mmol/L

Urine

Sodium Na+ 1 - 300 mmol/L 1.0 mmol/L

Potassium K+ 5 - 120 mmol/L 0.1 mmol/L


(60 - 120 mmol/L w. addit. dilution)

Chloride Cl- 1 - 300 mmol/L 1.0 mmol/L

Calcium and Lithium are not measured in urine samples.

2.2. Operating Configurations

Sample type Whole blood, serum, plasma, urine

Sample device Syringe, sample cup, collection tube

Sample size 95 uL

Analysis time 50 seconds

Sample rate 60 per hour

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 6 of 74


2 General Descriptions
Measurement
Sodium (Na+) sensor Ion-selective, flow-through, glass capillary electrode

Potassium (K+) sensor Ion-selective, flow-through, liquid membrane electrode

Chloride (Cl-) sensor Ion-selective, flow-through, liquid membrane electrode

Calcium (Ca++) sensor Ion-selective, flow-through, liquid membrane electrode

Lithium (Li+) sensor Ion-selective, flow-through, liquid membrane electrode

Reference System Open liquid junction, flow-through electrode

Calibration Fully automatic


1 point with each sample
2 point every 4 hrs (Lithium, 3 point)
Standby Suspends Calibrations

Warm-up time 1 minute

Temperature Room temperature, 15 - 32° C, 60 - 90° F

Humidity Maximum 85% relative humidity, non-condensing

Data management Quality Control memory storage, 3 levels, 500 values each;
calculation of mean, standard deviation, and coefficient of
variation (CV)
Samples, 1000 measurements
Diagnostic Programs User-controlled diagnostics, YES/NO operation via the display or

Keyboard

Electronics Microprocessor-controlled

Display LCD dot-matrix, 2 line, and 16 characters per line

Printer Integral thermal printer, 16 character width

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 7 of 74


2 General Descriptions
Interface RS232C Serial Port,
USB Ports, 2
for Keyboard (provided) and Barcode Scanner (Optional)

Electrical requirements 100 - 240 V, 50/60 Hz,


1.4 A max.
Self-adjusting

Nominal power
Consumption 50 W

Dimensions
Height 12.2"; 335mm
Width 12.4"; 315m
Depth 12.0"; 295mm
Weight approx. 13 lbs; 6kg

Classifications

Safety category IB (according to EN 60601-1, IEC 601-1)


Device type

Mode of operation Continuous operation


Protection classification IP 20
Explosion protection The device is not designed for operation in explosive
environments.

Approvals CSA, IEC 1010 (TUV/GS), CE, FCC Class B

Data subject to change without notice. Technical information is supplied for general
informational purposes only.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 8 of 74


3 Software Operation

3. Software Operation

3.1. Software Operation – CorLYTE


Software operation of the CorLYTE analyzer can be accessed by YES/NO selection on the analyzer
front panel or through the function keys in the keyboard. The following flow chart diagrams are
provided to identify operating sequences of the CorLYTE Analyzer:

1. Overall Program Operating Flow


2. Main Menu
3. Measurement Sequence
4. Print Reports Menu
5. QC/Standard/Dialysate/Urine Sample Menu
6. Maintenance Menu
7. Operator Settings Menu
8. Instrument Settings Menu
9. Service Menu
10. Calibration Sequence

The SERVICE MENU is also provided with a more detailed description of each
check configurations for use in checking instrument subassemblies.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 9 of 74


3 Software Operation

1. OVERALL FLOW

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 10 of 74


3 Software Operation
2. MAIN MENU - HUMAN

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 11 of 74


3 Software Operation

3. MEASUREMENT SEQUENCE

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 12 of 74


3 Software Operation
4. A. PRINT REPORTS-SAMPLES

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 13 of 74


3 Software Operation

4.B PRINT REPORTS – QC ETC.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 14 of 74


3 Software Operation
5. QC-STD-DIALYSATE-URINE SAMPLE

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 15 of 74


3 Software Operation
6. MAINTENANCE

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 16 of 74


3 Software Operation
7.A. OPERATOR SETTINGS H TO V

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 17 of 74


3 Software Operation

7.B. OPERATOR SETTING H TO V

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 18 of 74


3 Software Operation

8.A. INSTRUMENT SETTINGS QC

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 19 of 74


3 Software Operation

8.B. INSTRUMENT SETTINGS OTHERS

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 20 of 74


3 Software Operation
9. SERVICE MENU

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 21 of 74


3 Software Operation
10.A. PERFORM CALIBRATION

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 22 of 74


3 Software Operation
10.B. PERFORM CALIBRATION

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 23 of 74


3 Software Operation
11. POWER-UP

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 24 of 74


3 Software Operation
12. KEYBOARD FUNCTIONAL KEYS

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 25 of 74


3 Software Operation

3.2. Service Menu


The SERVICE MENU (mode provides a menu of service check functions to assist in
failure diagnosis and identification. A logic flow chart is provided for reference of the
sequences included in the SERVICE MENU mode.

CHECK ELECTRODES

This mode allows for the measurement and display of electrode voltages. STD A, STD B
or an external sample can be measured to determine electrode performance and provide
measurement data useful in locating electrode calibration faults. Each electrode
configuration will be displayed and the electrode voltage associated with the sample type
selected can be observed.

Example Displays:

Std. A: Na=1582 Std. B: Na=237


K= 654 Cl=-904 K=-988 Cl=647

Ext: Na=1321
K=-598 Cl=-127

CHECK SAMPLE SENSOR

In this mode, the sample sensor output is displayed which is useful for electrical
adjustment or for troubleshooting sample sensor faults. A dry sample sensor should
display a value of approximately 100 with a range of 80 to 120. After verifying a dry
sensor, aspiration of a clear fluid (i.e., water) should increase the displayed value by at
least 40 units.

Example Display:

DRY 80 -120: 102 External: 153


YES=Pump/No=Exit Pump: NO=OFF

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 26 of 74


3 Software Operation

CHECK FLUID PACK SENSOR

In this check, the Fluid Pack Sensor detects whether the Fluid Pack is in place. “I”
indicates that the Fluid Pack is in place, “O” indicates that it is either not present or not
completely seated.

Example Display

Fluid Pack: I
I=In O=OUT

CHECK LANGUAGE

This mode displays the language selected for display and printout.

Example Display

Language:
English

To change the language, press the NO key until [INSTRUMENT SETTING?] is


displayed. Then select YES. Press NO to scroll through the menu until [SELECT
LANGUAGE?] is shown. Press YES. Upon entering this menu the language can be
set to English, German, French, Spanish, Italian, or Portuguese. Press YES when the
desired language is displayed on the screen. After confirming the selection, the
language will be reset automatically.

CHECK SAMPLE DOOR

This check verifies the operation of the sample door light gate and its associated
circuitry. The display indicates an “O” or “C” depending on the position of the door.

Example Display

Sample Door: C
C: Closed/O: Open

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 27 of 74


3 Software Operation
CHECK PUMP

This check turns on the peristaltic pump at each of the four pump speeds. The display
indicates for two seconds each pump speed while each pump speed is being checked.
Observe that the peristaltic pump changes speed as is indicated on the display.

* * * Check Pump * * * * * * Check Pump * * *


Very Slow Speed Slow Speed

* * * Check Pump * * * * * * Check Pump * * *


Medium Speed Fast Speed

VALVE CHECK

This turns on/off each solenoid valve as indicated and is useful to verify both electronic
control and mechanical operation of each solenoid valve.

CHECK INTERFACE

The Interface Check can be used to verify correct operation of the RS232C serial port.
To operate this check, jumper pin 2 and pin 3 of the rear interface connector prior to
invoking the check. During the interface check, the analyzer sends out a check string
and checks if it is received within a set period of time. If correct operation is observed
after completion of the check, INTERFACE_OK will be displayed. If not Interface
Error will be displayed.

CHECK AMPLIFIER

The Amplifier Check displays ground, reference and output voltage of each input
amplifier. The values displayed for each electrode channel will be dependent on the
electrode inputs and may vary. Display of the ground should read between -10mV and
+10mV with the reference voltage reading between -2490mV and -2510mV. An
additional display indicating the temperature voltage and temperature in degrees C is
available to verify the temperature adjustment.

CHECK RFID SENSOR

A compatible Fluid Pack must be installed for this test, otherwise an error will be returned. This
option allows the RFID system to be verified. The RFID system includes the RFID board and a
compatible Fluid Pack. During the test, the instrument sends a signal to a tag in the Fluid Pack.
If the RFID transponder receives a signal from the Fluid Pack, TAG FOUND will be displayed.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 28 of 74


3 Software Operation
ENTER CONFIG CODE

CONFIG CODES are available and allow additional programming features. Following is a
listing of the available CONFIG CODES and functions. NEW CODES

Set Reset CONFIG CODES Description

AME EMA Dialysates in Acetate matrix can be analyzed.

BCE ECB Dialysates in Bicarbonate matrix can be analyzed.

CPC Connect PC to upload software.

CSC Clears all config codes, return to factory default

ECM MCE Allows the operator to run the instrument in the economy
mode. The instrument will enter the STANDBY MODE.
This occurs at the time of the next 2-point calibration
when no samples have been analyzed since the last 2-
point calibration.

HRS SRH Provides one additional digit of resolution for each


parameter for blood and serum samples. QC and Standard
samples are always displayed with high resolution. Urine
samples are always displayed in low resolution.

LCR Prints lease counter report.

MGC CGM This config code is available only in the model CorLYTE
analyzer and enables Ca++ values to be displayed in
mg/dL. The normal mode displays values in mmol/L. QC
and Standard results are always displayed in mmol/L.

PMI This code will print the model information of the


CorLYTE.

PSC Prints service codes.

QCF FCQ When this code is enabled, QC values will be


reported in direct ISE measurement instead of flame-
equivalent values. In this mode, correlation factors are
NOT applied to the measured values.

TBO OBT This config code disables the beep.

TDV VDT Allows printing of the electrode voltages for sample


measurements and calibrations.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 29 of 74


3 Software Operation

R14 V48 These codes allow the dealer to enter their own
password for those instruments will be released based
on number of tests.

RST Erases all data from memory including correlation


factors, QC Ranges, stored patient, QC data and lease
count.

ANALYZER RESET Press YES and NO keys simultaneously and


turn power off and back on to provide a total reset of the
instrument software. CAUTION! All QC Data and
Instrument Programming will be lost. Switching from
human to vet or vet to human resets all previously saved
data as well.

WARM-UP BYPASS Press YES and NO keys simultaneously during instrument


warm-up.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 30 of 74


4 Description of Modules

4. Description of Modules

4.1. Mechanical Assemblies

4.1.1. Front Door Assembly

The front door assembly is comprised of the following parts, Front Door, CN-4002, EMI
Window, CN-4193, Plastic Window, CN-4092 and Door Magnets.

The front door can be removed by moving the analyzer to the edge of a work surface so
the analyzer door, when opened, will extend past the edge of the work surface. First
remove small ground connector, then ground screw, external tooth washer, right angle
tab, and hex nut. Then with one hand, hold the analyzer door near the right side hinge pin
and, with the other hand, gently apply pressure to the middle rear area of the door. This
will allow the right hinge pin to clear the retaining hole in the main chassis. The door can
then be removed from the analyzer.

The window is comprised of two layers, the EMI window and the plastic window. Both
can be replaced in the field by gently pressing from the rear of the window and snapping
the EMI and plastic window out toward the front of the door. During replacement the
EMI window should be placed over the opening first, ensuring good contact of the EMI
window adhesive with the copper tape on the frame of the window. Next the Protective
window is applied over the EMI window.

Plastic Window

Front Door

Fig. 4-1

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 31 of 74


4 Description of Modules

4.1.2. Aspiration Assembly

This module is comprised of the Aspiration Subassembly, E-725 and the sample probe,
E-724.

The sample probe and fill port are designed for easy replacement. The sample probe can
be removed as a one-piece assembly including the intake tubing which connects the
sample probe to the electrode module through the sample sensor. To remove the sample
probe, first disconnect the intake tubing from the sample sensor on the right side of the
electrode tray. Then open the sample door and grasp the probe near the white holder
block and pull up to unsnap the sample probe from its position. To replace the assembly,
align a replacement sample probe in the white holder block and press the replacement
assembly into place. Connect the tubing to the sample sensor.

To remove the fill port assembly, pull the fill port holder towards the front for easy
access. Then, press the two plastic tabs on the fill port to allow removal of the fill port
from the fill port holder.

To remove the aspiration assembly, three screws must be removed. To remove the two
rear screws first ensure power is off and remove the rear panel. The rear panel can be
moved away from the unit for easier access by unplugging the power supply and ground
cables from the CPU board. The CPU board can be removed after disconnecting the
cables and the seven screws which secure the CPU board to the case. Remove the
grounding screw and star washer on the left side of the CPU board first. The cable on the
left is the valve board cable and the one at the top of the CPU board is the display cable.
Once the valve board is exposed, disconnect the aspiration assembly connector which is
the far left connection.

The two rear screws can now be removed using a hex key. Then, remove the front
securing screw, the sample intake tubing and the reagent supply tubing (red tag) from the
fill port. The aspiration assembly can now be removed from the front of the analyzer for
replacement. Reference the Illustrated Parts diagram in Section 8 for a detailed parts
diagram of the aspiration assembly.

Aspiration
Sub-assembly

Sample Probe

Fig. 4-2

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 32 of 74


4 Description of Modules

4.1.3. Electrode Tray Assembly

This assembly is comprised of the electrode tray, E-726, and the sample sensor, E-723.

To remove the electrode holder assembly from the analyzer, first unplug the sample
intake tubing on the right side of the assembly where the sample sensor is located. Then
disconnect the sample sensor cable, unplug the reference solution tubing connector and
last, disconnect the waste tube (green tag) from the left side of the electrode tray. The
electrode tray assembly can now be removed by pressing the two plastic tabs on the right
and left side of the module and sliding the module towards you. Continue to press the
plastic tabs to completely release the tray and remove the assembly from the analyzer.
The sample sensor (on the right) and the electrode locking mechanism can be removed
and replaced by removing the screws on the underside of the tray. The Illustrated Parts
List in Section 8 provides a detailed reference of the parts of this assembly.

Sample Sensor

Electrode Tray

Fig. 4-3

4.1.4. Peristaltic Pump Rotor E-728

Replacement of rotor assembly of the peristaltic pump can be accomplished by


removing the pump tubing and firmly pulling the rotor assembly away from the
instrument. To replace the rotor assembly, align the rotor with the flat on the motor
shaft and press the replacement roller into position.

Peristaltic Pump
Rotor

Fig. 4-4

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 33 of 74


4 Description of Modules

4.1.5. Peristaltic Pump Motor E-727

To remove the motor, first remove the rear panel and CPU Board as described in
Sections 4.1.8 and 4.1.9. Now, the electrical connector on the valve board can be
accessed to unplug the motor connector (See Figure 4-5). The motor can now be
removed by removing four screws located on the front of the housing near the
peristaltic pump.

Peristaltic Pump
Motor

Fig. 4-5

Fig. 4-5

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 34 of 74


4 Description of Modules

4.1.6. Solenoid Valve Assembly

The solenoid valve assembly is comprised of the following parts: the solenoid valve
subassembly, E-729, the crimp pin, CN-4155, and the indicator cap, CN-4018. There are
5 solenoid valves, one each for Standard A (A), Standard B (B), Standard C (C),
Reference Solution (R), and Air(V).

The removal of each solenoid valve is identical. On the front of the analyzer, first
remove the indicator cap, which holds the crimp pin in place. Locate the opening of the
Valves indicator cap where the solenoid shaft is visible. Slide the indicator cap off the solenoid
valve shaft in the direction of the arrow. This exposes the crimp pin which can now be
removed from the solenoid shaft.
Valves control the movement of the liquid within the analyzer.
If the crimp pin and/or cap need to be replaced at this time, energize the solenoid as
described in Section 3.2 (Valve Check). This extends the solenoid shaft to the outer most
position for ease in replacing the crimp pin and indicator cap.

Indicator Cap
Valve R

Valve V

Valve C

Valve B

Valve A

Fig. 1-10

Crimp Pin

Fig.4-6

I SE PACK
ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 35 of 74

The self-contained ISE PACK is constructed to ensure that waste cannot s


4 Description of Modules

To remove the solenoid valve subassembly, first remove the crimp pin, indicator cap, and
the two screws on each side of the solenoid shaft. Remove the rear panel and the CPU
Board so that the valve board and solenoids can be accessed. Each solenoid has an
electrical connection to the valve board which must be unplugged prior to removal of the
solenoid. See figure below for connections.

Connectors

Fig.4-7

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 36 of 74


4 Description of Modules

4.1.7. Printer Assembly E-715

Ensure the power to the instrument is off before removing the printer. Partially slide the
paper tray out so that the paper roll can be torn from the printer. Completely remove the
paper roll and the paper tray. Slide fingers under the printer assembly and pull the printer
towards you along the printer board guides on the analyzer. This will disengage the
printer from the CPU board and enable removal of the assembly from the front of the
analyzer.

Removal of the printer should be performed for replacement and for removal of a paper
jam. To replace the printer, locate the printer board guiding tracks and slide a printer onto
the analyzer. Press firmly into place to ensure proper connection of the printer to the CPU
board.

Note: Never attempt to dislodge paper from the printer with a paper clip or similar
object to avoid damage to the print head or printer platen.

E-715 (Old Style) E-715 (New Style)

Fig.4-8

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 37 of 74


4 Description of Modules

4.1.8. Power Supply Subassembly E-714

The power supply subassembly is located on the rear panel and is comprised of the power
supply assembly and the main power receptable. To remove the rear panel assembly,
ensure that the power cord has been disconnected. Remove the four corner screws to
expose the rear panel. Disconnect the ground wires and connector from the CPU board
and front case. The power supply module located on the rear panel assembly can now be
removed by removing the four screws securing the power supply mounting plate to the
rear panel. The main power receptacle can be replaced by removing the two screws on
the rear panel holding this assembly.

Fig. 4-9

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 38 of 74


4 Description of Modules

4.1.9. CPU Board E-730

After the rear panel (Section 4.1.8) has been removed, the CPU Board is accessible for
removal. This circuit board can be removed by removing the seven screws with their
respective plastic washers securing the board to the housing and by disconnecting
electrical connectors from the power supply module, display board and valve board. The
electrode push pins can now be replaced by simply pulling the connecting pin from its
socket. When installing the board, make sure to place the plastic washers under the screw
heads.

Back Front

Pogo Pins

D1-D5

Fig. 4-10 Fig. 4-11

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 39 of 74


4 Description of Modules

4.1.10. RFID Board CN-A4054

RFID board is attached to the CPU board at JP10. This board must be connected all the
time to the CPU board for proper functioning of the instrument. DO NOT unplug the
RFID board while the CPU board is powered ON.

Fig. 4-12

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 40 of 74


4 Description of Modules

4.1.11. Valve Board E-732

The rear panel and CPU board, referenced in 4.1.9 and 4.1.10, must be removed to gain
access to the lamp board. Remove one screw with washer on the right side which holds
the board in place and remove the seven electrical connectors to slide this board out.

Fig. 4-13

4.1.12. Display Board E-731

The rear panel and CPU board, referenced in 4.1.9 and 4.1.10, must be removed to gain
access to the display board. To remove the display board, unplug the sample sensor cable
from the front, remove the ribbon cable from the CPU board. Then remove three screws
and 50 degree spacers to remove the display board.

Fig. 4-14

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 41 of 74


4 Description of Modules

4.2. Fluid Pack

4.2.1. Standard A

Standard A, STD A, is drawn to the electrode module by vacuum provided by the


peristaltic pump. When STD A is to be aspirated into the electrode module, solenoid
valve A is opened and solenoid valves B, C, V and R are closed. STD A is drawn from
the Fluid Pack, to the fill port, through the sample probe up to the sample sensor. When
the sample sensor detects fluid, solenoid valve V is opened and solenoid valve A is
closed as the peristaltic pump continues to pump STD A into the electrode module. At the
trailing edge of the STD A is air. When the sample sensor detects air, the peristaltic pump
stops. Standard A should fill the electrode module. During the time STD A is aspirated
into the electrode module, the reference housing is pressurized due to the peristaltic pump
providing pressure to the reference solution line and solenoid valve R being closed. A
small amount of reference solution is forced out through the reference junction to provide
contact to the sample.

4.2.2. Standard B and C

The process for STD B and STD C is identical to that of STD A with the exception of the
operation of solenoid valves B and C. These solenoid valves are operated in reverse order
of STD A aspiration.

4.2.3. Reference Solution

The reference solution is circulated back into the fluid pack. The reference housing has
an in and an out port. It is filled automatically using the second winding of the peristaltic
pump and solenoid valve R. The reference solution connector allows for the reference
housing tubing to be plugged into the reference solution circuit. As the peristaltic pump
aspirates reference solution, solenoid valve R is opened to allow excess solution to be
pumped into the reference return line of the Fluid Pack.

4.2.4. Waste Liquids

Calibration standards and sample waste are all pumped out of the left side of the
electrode module (green-banded tubing) through the peristaltic pump to the waste line
and waste bag of the Fluid Pack. An internal one-way valve is incorporated into the waste
bag to prevent waste products from leaking out of the Fluid Pack.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 42 of 74


4 Description of Modules

FLUIDIC DIAGRAM

Fig. 4-15

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 43 of 74


4 Description of Modules

4.2.5. Pinch Valve Tubing Set, E-711

The pinch valve tubing set is supplied as a pre-cut tubing harness with the fluidic
interconnections preassembled. Replacement of the tube set should be performed
annually. For removal of the tubing, first remove the Fluid Pack from the reagent
compartment. This will allow the removal of the white TPR block which is the
connection between the Fluid Pack and the tubing. Next, follow the procedure outlined in
Section 4.1.6 to remove the indicator cap and crimp pin from each solenoid valve to
enable fluidic tubing to be easily removed at each solenoid. Pull the two pump windings
of the peristaltic pump tube set off the pump rotor and pull out the white reference
solution tube connector. Remove the white TPR reference solution block by grasping the
front of the block and pulling it out of the analyzer. Disconnect the green tagged tube
from the electrode tray and the red tagged tube from the bottom of the fill port. All tubes
can now be removed and discarded.

Fit the replacement tubing harness and assemble in reverse order as disassembly.
Reference the Tube Diagram on the inside front cover of the analyzer for correct tube
locations.

Green Blue
Tube Tube

TPR

Fig. 4-16 Red


Tube

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 44 of 74


Valves control the movement of the liquid within the analyzer.
4 Description of Modules

Valve R

Valve V

Valve C

Valve B

Valve A

Fig. 1-10

R: reference V: vent A,B,C: Std A,B,C

Fig. 4-17

I SE PACK

The self-contained ISE PACK is constructed to ensure that waste cann


package.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 45 of 74


ig. 1-11
4 Description of Modules

4.3. Electronics

Block Diagram of CorLYTE

Sample Sensor Diver Sample Sensor

Sample Door Detector Door Sensor

Na Pre Amplifier Buzzer


K

Cl, Ca, Li
Fluid Pack Detection
Reference
Display PCB Board
Display Driver
ADC/1

Temperature Control Temperature sensor

Printer
Printer Driver

RTC

uP SPI Flash

RFID PCB

ADC/2
(Not Used) RS-232 Driver RS-232

USB Driver USB 1

USB 2
Power Supply
Peri Pump Driver
22 v DC/12.5 vDC
5v DC
-5 DC Solenoid Value Drivers Lamp and Valve
3.3 v DC PCB Module

CPU Board CN-4180

V V V V
al al al al
Power Supply Module ve ve ve ve
100-240 v Power Receptor Fuse F121SA/250 v A B C V
Main 1.6 AT +5v DC +24v DC
50-60 Hz 1.6 AT +12 v DC -12v DC

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 46 of 74


4 Description of Modules

4.3.1. CPU Board

Power Circuits

Power is supplied to this circuit board through connector JP3 from the power supply
assembly. Check points are provided to measure supply voltages. The following list
identifies the check points.

TEST POINT VOLTAGE TEST POINT VOLTAGE


TP1 +12 VDC TP2 -12 VDC
TP3 +24 VDC TP4 2.5 VREF
TP5 2.5 V REF TP6 PVCC
TP7 SAMPLE IN TP8 +5 VDC
TP9 TEMP OUT TP10 REF OUT
TP11 Cl-/Ca+/Li OUT TP12 K OUT
TP13 Na OUT TP14 NOT USED
TP15 NOT USED TP16 2.045 VDC
TP17 NOT USED TP18 -VREF
TP19 -1.2 V REF TP20 +2.5 VA
TP21 -2.5 VA TP22 3.3 VDC
TP23 10 V REF TP24 DGND
TP25 AGND

Moreover, light emitting diodes Dl through D5 located at the bottom of the CPU board
also provides indicators that each supply voltage is present on the board. Since each LED
is connected in series after the fuse, a blown fuse results in the respective LED off. The
following list identifies the fuses and LEDs for each respective DC supply voltage:

FUSE LED
F1 D1
F2 D2
F3 D3
F4 D4
F5 D5

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 47 of 74


4 Description of Modules

Microprocessor

The LM 3S5B91 Microprocessor represented as W6, can be located beneath the metal
shielding. This processor is responsible for controlling most of the peripherals, including:
ISE signal processing, printing, displaying, data storage, serial communication and USB
interfacing.

CPU Shielding

Four additional voltage regulator circuits are mounted inside the CPU board, -5VDC;
+5VDC; 3.3VDC and 24/12VDC (see block diagram). SMPS technology used to
generate these voltages requires a shield to protect the instrument from the 400 kHz and
the harmonics generated internally.

Fluid Pack Sensor

CorLYTE analyzer uses fluid packs for calibration, sample analysis, and waste storage.
To ensure that the fluid pack is present, an optical reflective sensor is used. The optical
sensor has an LED with a photo-detector. The photo-detector signal is used to determine
presence of fluid pack.

Valve Drivers

All the solenoids for the valves A; B; C; V and R are controlled from power transistors
Q1 to Q5. To open a solenoid, a pulse of voltage 22-24 VDC is applied for 0.5 seconds.
Then the voltage drops to 12.5 VDC to hold the solenoid. This feature is applied to
conserve energy and prevent overheating the valves.

Pump Motor Driver

The Pump driver circuit IC U7 provides the drive and control for the peristaltic pump.
The processor through commands: /PUMPEN and PUMPSTEP drive the IC U7 to
provide necessary control for the peristaltic pump stepper motor in the terminals PHASE
A; PHASE B; PHASE C and PHASE D.

Input Amplifiers

IC U27; U28; and U29 Provide high to low input impendence matching for each
electrode (Na; K; Cl; Ca; and Li) input. IC U25; U25; and U26 provide the offset values
and necessary gain for each electrode.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 48 of 74


4 Description of Modules

Temperature Sensor circuit

This circuit creates a temperature –depended voltage which is converted by software to


degree C. R190 can be adjusted to correct temperature setting.

Analog Channel and A/D Converter

The analog inputs from the amplifiers are connected to IC U21 and U22 ADC1 and
ADC2 converters. ADC 2 is in the schematic for future developments.

JP Connector explanation in the CPU board.

In order to fulfill all responsibilities of the CPU board and the program installed on it, is
quipped with a number of connectors to interface it with all peripherals. The description
of the JP connectors is as below:

JP1 - Graphic printer connector


JP2- NA
JP3 - Power connector receptacle.
JP4 - Display connector
JP5 – Valve board connector
JP6- NA
JP7 – Serial DB9 RS232 Connector
JP8 – USB1 Connector
JP9 – USB 2 Connector
JP10- RFID connector
JP11- JTAG connector

4.3.2. Valve Board E-732

The valve board contains the LED array used to illuminate the measuring chamber and
the interconnectors used for all solenoids, the peristaltic pump and the door sensor. A
ribbon cable from the CPU Board supplies the electrical connection for each signal
through connector JP5.

4.3.3. Display Board E-731

The graphic display module TH16032A has a 160x32 resolution. The backlight is
adjustable and allows for changing the contrast. In the same board are all connection for
YES and NO buttons as well sample sensor and temperature sensor. The connection of
the display board with CPU board is done through ribbon cable connected to JP4 in the
CPU board. The new display board is equipped with two potentiometers for controlling
the backlight and the contrast in the display.

Note: Graphics Display is not compatible as a direct replacement with previous


generation.
ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 49 of 74
4 Description of Modules

4.3.4. Graphic Printer E-715

A graphics printer with 192 dots of a TP series has improved the quality and the speed of
printing. The new schematic printer allows the selection of three different fonts and has a
speed 6 times faster that the previous generation.

Note: Graphics Display is not compatible as direct replacement with previous


generation.

Shielding and Grounding in CorLYTE

Beside the shield for eliminating the internal noise from the power supply, the entire
CorLYTE is equipped with a protective shield to prevent the influence of magnetic fields,
and other noise sources in the laboratory. The internal protective shield creates a
“Faraday Cage” that protects instrument measurements from outside noise. This shield
is also grounded with a grounding wire in the power receptacle in the power switch input.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 50 of 74


5 Maintenance and Troubleshooting

5. Maintenance and Troubleshooting


Maintenance procedures for the CorLYTE Electrolyte Analyzer require minimal time by
the operator to perform. The procedures outlined below should be performed by the
schedule indicated. Detailed instructions for these procedures can be found in the
Operator's Manual describing the correct method to perform routine analyzer
maintenance.

5.1. Maintenance

Daily Maintenance

- Perform cleaning cycle.


- Perform conditioning cycle.

Weekly Maintenance

- Clean sample probe and fill port.


- Clean analyzer surfaces.

Monthly Maintenance

-Clean reference electrode housing.

Six Month Maintenance

-Replace peristaltic pump tubing.

Annual Preventive Maintenance

- Replace complete tubing set.


- Replace fill port assembly
- Replace sample probe.
- Replace sample sensor Quad-ring.
- Check sample sensor voltage.
- Check sample temperature.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 51 of 74


5 Maintenance and Troubleshooting

5.2.Error Messages and Troubleshooting

STATUS: INVALID CALIBRATION

This message is only displayed when a calibration has been interrupted by pressing NO,
opening the sample door, or none of the electrodes are calibrated. Perform a calibration to
return to the H-READY Mode.

STD A NOT FOUND

This message is displayed when the sample sensor is unable to properly detect the STD A
solution in a programmed time period. As the analyzer draws a sample of STD A, the
leading edge of the solution is detected by the sample sensor. The sample sensor must
also detect a continuous flow of STD A Solution into the sample chamber, lasting several
seconds, without encountering air bubbles. After several seconds, the trailing edge of the
STD A sample is sensed, the peristaltic pump stops and measurement occurs.

Possible causes and remedies:


• Check for air leaks preventing the Standard from being drawn into the chamber.
Ensure that all O-rings are in place. There must be a continuous draw of standard
solution, free of air bubbles. Check solenoid V for proper sealing of the vent line.
Check fill port to ensure the absence of air leaks.
• Check for clots or crystals formed in the STD A tubing or electrode chamber.
• Ensure that the sample sensor is plugged in securely and perform Check Sample
Sensor to ensure sensor is operating correctly. Clean, adjust or replace sample sensor
to correct sample sensor operation.
• Replace peristaltic pump tube set to ensure proper sample aspiration.
• Check the fluid remaining in the Fluid Pack. If less than 5% remains, replace the
Fluid Pack.
• Prime system; look for fluid flow through system.

STD B NOT FOUND


STD C NOT FOUND

The system for drawing STD B or C is like that of STD A. If STD A is drawn properly
and STD B or C is not detected, check STD B tubing for crystallization. Perform checks
for STD A detection problems to correct. During the calibration sequence, STD B or C is
drawn into the measuring chamber prior to STD A, therefore check if STD A calibrant
can be properly detected to isolate possible fluidic faults. If neither STD A, B nor C are
detected check for air leaks or blockages within the fluidic path.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 52 of 74


5 Maintenance and Troubleshooting
INSPECT SAMPLE SENSOR

The sample sensor must provide a reading of 80 - 120 when air is detected.
Perform Check Sample Sensor to observe sample sensor response. When clear fluid (e.g.
water) is aspirated, the sample sensor display must indicate an increase of at least 40
units. For blood samples (not transparent), the reading should decrease by at least 40
units. The sample sensor is calibrated with air during each calibration.

Possible causes and remedies:


• Clean sample sensor using CLEANING procedure.
• Check for correct drying of measuring chamber during the wash cycle.
• Replace peristaltic pump tube set.
• Check and adjust sample sensor to correct adjustment. This adjustment can be
accessed through the rear cover vent slot and is located at the top center on the circuit
card.
• Replace sample sensor assembly.

INSPECT REFERENCE HOUSING

The reference electrode is pressurized by injecting STD A while R valve is closed. The
back pressure formed in the reference housing causes the STD A solution to flow back
toward the sample sensor. When the STD A solution is detected by the sample sensor, the
R valve is opened to release the pressure. If the sample sensor does not detect STD A as
it moves back, INSPECT REFERENCE HOUSING will be displayed.

Possible causes and remedies:


• Check for clogged reference housing and clean per monthly maintenance procedures.
• Check and ensure that reference tubing is securely connected at the tubing connector.
• Check for proper filling of the reference housing and ensure that reference solution
reagent level allows complete filling of the reference housing.
• Check for bubble-free aspiration of STD A.
• Check O-Rings to ensure they are present in all electrodes and sample sensor
• Prime CorLYTE and repeat calibration

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 53 of 74


5 Maintenance and Troubleshooting
SHUT DOOR

This message occurs when either of two conditions exists.

1. Sample door is not closed within 20 seconds after samples are in place.

2. Sample door has been opened and no sample is detected within 20 seconds.

Possible causes and remedies:


• Close sample door within the time allowed.
• Check needle mechanism light gate.
• Check whether sample sensor is plugged in.

SAMPLE NOT FOUND

This message is displayed when any of the following conditions occur:

1. During sampling, either a bubble in the sample is detected or the sample size is smaller than
the minimum sample required.

2. The sample door is opened and no sample is fed into the analyzer.

3. Sample door is not closed within 20 seconds after sample is detected.

Possible causes and remedies:


• Check whether the sample sensor is plugged in and perform CHECK SAMPLE
SENSOR to verify that it operates correctly.
• Check sample draw and look for clots or leaks in the sampling system.

CLEANING SOLUTION NOT FOUND


When the cleaning cycle has been initiated, the analyzer checks for proper aspiration of
the cleaning solution using the sample sensor. If solution is not detected in the proper
time, above error message will be displayed.

Possible causes and remedies:


• Ensure that a sufficient volume of cleaning solution is presented for aspiration and no air
bubbles are present as solution is aspirated.

• Check to ensure that the solution is being aspirated through the sampling mechanism
correctly. The sample path must be free of leaks or clots.

• Verify that sample sensor is properly plugged in and perform the CHECK SAMPLE
SENSOR to verify the correct operation of the sample sensor.

• Check pump tubing and replace as necessary to ensure a correct pump rate.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 54 of 74


5 Maintenance and Troubleshooting

CONDITIONER NOT FOUND


When the conditioning cycle has been initiated, the analyzer checks for proper aspiration
of the conditioning solution using the sample sensor. If the solution is not detected in the
proper time, this error message will be displayed.

Possible causes and remedies:


• Ensure that a sufficient volume of conditioning solution is presented for aspiration and no air
bubbles are present as solution is aspirated.

• Check to ensure that the solution is being properly aspirated through the sampling
mechanism. The sample path must be free of leaks or clots.

• Verify that sample sensor is correctly plugged in and perform the CHECK SAMPLE SENSOR
to verify its proper operation.

• Check pump tubing and replace as necessary to ensure correct pump rate.

INTERFACE CHECK ERROR

This message is displayed only during the ACTIVATE INTERFACE. Pins 2 and 3 must
be shorted together to perform the ACTIVATE INTERFACE. Make sure that the pins are
not shorted to chassis ground.

Possible cause and remedy:


• Call Product Support

PRINTER JAM OR DEFECT

This message is displayed when the printer head is jammed and tries to print. It is a
temporary message that is displayed for 2 seconds and then sample results are displayed.
To clear paper jam remove printer from the analyzer and remove jammed paper. Replace
printer and retry. The printer can be removed by grasping the rear edge of the printer
assembly and pulling it to the front. The printer can be removed with the Analyzer power
on. To free up jammed paper, turn the small gears on the left side of the printer module.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 55 of 74


5 Maintenance and Troubleshooting
INSPECT ELECTRODES

This message is displayed when either of the following conditions is present.

1. More than six aspirations of STD A does not result in three consistent STD A readings for all
electrodes.

2. The difference between voltage A and B is out of specification for all channels.

Possible causes and remedies:


• Electrodes are not plugged in. Check the slide mechanism to ensure it is in the correct
position and firmly pushed into electrode connectors.
• Perform MAINTENANCE.
• Check or replace reference electrode.
• Replace Fluid Pack.
• Check connection from shielding to ground to determine if electrical short is present.
• Call Product Support.

NA INVALID CAL
K INVALID CAL
CL INVALID CAL
CA INVALID CAL
LI INVALID CAL

These messages will be displayed when a specific electrode does not calibrate properly during a
calibration sequence. If more than six standard draws are required or the difference between STD
A and STD B voltages is out of range, the INVALID CALIBRATION message will be displayed
for each channel which exhibits either of these conditions. A copy of the calibration report
should be printed to provide the electrode output voltages measured during calibration. The
electrode voltage range table identifies correct voltage ranges for each electrode configuration. If
the difference between STD A and STD C is out of range for Li+ or Na+, LI INVALID CAL will
be displayed.

Possible causes and remedies:


• Check to ensure calibration reagents are transported correctly. Check for leaks, blockages,
bubbles, or improper pump flow rate.
• Check or replace Fluid Pack
• Check whether electrodes are plugged in properly. Check the slide mechanism to ensure
it is in the correct position and firmly pushed into electrode connectors. Clean and/or
replace electrode.
• If A-C is out of range for Na+, repeat the calibration or replace the Fluid Pack
• Check or replace CPU Card.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 56 of 74


5 Maintenance and Troubleshooting

ELECTRODE VOLTAGE RANGES


Allowed difference
Electrode Standard A Standard B Standard C A-B A-C
Na+ - 600 to +2400 -1600 to +2000 -600 to +2400 +250 to +680 -50 to +50
K+ - 700 to +1000 -2500 to +500 -700 to +1000 +470 to +1200 -40 to +40
Cl- - 3100 to +500 -1000 to +3000 -3100 to -100 -370 to -860 not used
Ca++ -3100 to +1000 -2300 to +2500 -3100 to +1000 -350 to -660 -150 to +150
Li+ - 3100 to +1900 -3600 to +1400 -2600 to +3400 +1 to +760 -1730 to -285

INSPECT FLUID PATH

After completion of sample measurement or calibration, the analyzer will monitor the
sample sensor to determine if the sample path has been cleared. This message will be
displayed if the sample path has not been cleared.

Possible causes and remedies:


• Check for blockages in the sample path, especially in the sample probe, the tubing to the
sample sensor and in the sample sensor.
• Check sample sensor is securely plugged in and check sample sensor under SERVICE MENU
to ensure correct operation.
• Check peristaltic pump and tubing are correct and all tubing is properly in place.
Run Prime Function, Run Calibration



In case the analyzer displays arrows up or arrows down instead of sample results, the
concentration of the sample is outside of the measurement range.

Possible causes and remedies:


• If the sample is a urine sample, arrows up instead of the K+ result indicate that further dilution
of the specimen is required.
• Check for proper sample preparation.
• Ensure that the sample is correctly aspirated into the measuring chamber and ensure small air
bubbles are not present.
• Check for proper aspiration of STD A.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 57 of 74


5 Maintenance and Troubleshooting

On the CorLYTE, the calibration report will print an arrow up or down instead of the
actual temperature, if the temperature measured is out of range (10.0°C -40.0°C) or when
calibration values are higher or lower than specifications. The temperature sensor is
located in the right side electrode holder.

Possible causes and remedies:


• Ensure that sample sensor cable is securely plugged in.

• Check whether room temperature is within specified limits (15°C to 32°C / 60°F to 90°F).

• Perform temperature adjustment procedure.

• Replace CPU board or right side electrode holder.

ERR

The analyzer will display ERR in place of sample results when the analyzer is unable to
obtain valid voltage readings from the electrode (A/D over- or underflow).

Possible causes and remedies:


• Ensure that the electrodes are securely in place and plugged into the analyzer.

• Check to ensure proper filling of the reference electrode has occurred.

• Ensure proper sampling and proper sample preparation. Check for air bubbles in the sample.

• Check Amplifier to verify electrode voltages

EXECUTE MAINTENANCE

This message will be printed at the end of a sample report when cleaning or conditioning
has not been performed within the last 24 hours. This message is printed only. Perform
maintenance to clear this message.

CHANGE

When the monitored fluid level in the Fluid Pack is less then 0% remaining, the analyzer
will display change FLUID PACK. Replace the Fluid Pack following the instructions
outlined in the Operator Manual.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 58 of 74


5 Maintenance and Troubleshooting
VALVE OVERTEMP! INSPECT VALVES

In case one of the valve solenoids overheats or is not connected properly, the message
VALVE OVERTEMP! INSPECT VALVES will be displayed. Each solenoid
incorporates a thermo-fuse, which will reset automatically.

Possible causes and remedies:


• Turn power off for at least one hour.

• Check fuse F3 (24V).

• Check whether room temperature is within specified limits (15°C to 32°C / 60°F to 90°F).

• Ensure all the pinch valves are connected properly.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 59 of 74


6 Adjustments

6. Adjustments
The CorLYTE Electrolyte Analyzer has been designed to require minimal electronic
adjustment. It is recommended to adjust the sample sensor circuit after changing the
sample sensor during routine troubleshooting or during annual maintenance procedures.
The temperature adjustment is active and requires the temperature adjustment check plug
for correct calibration of the temperature circuit.

6.1. Sample Sensor Adjustment

Press F7 in the keyboard to access SERVICE MENU and here you need to access
SAMPLE SENSOR CHECK. By confirming, instruments will remove solutions from
sampling path. On the top right corner of the screen, a number between the ranges of 0 to
240 is shown. Sample sensor potentiometer (R189) is located on the top of the CPU
board (See Fig. 6-1).

Adjust R189 until the value 100 is reached in the display. After this adjustment, open
sample door, press YES or ENTER, and introduce water in the sample probe. When
liquid arrives in the sample sensor, the number on the screen will change. Displayed
values must be 50 points or higher that the adjusted value (80-120 range).

Note: R189 can be accessed through the rear cover vent slots.

R189 S.S. ADJ.

Fig. 6-1

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 60 of 74


6 Adjustments

6.2. Temperature Adjustment

Press F7 in the keyboard to access SERVICE MENU. Scroll down SERIVCE MENU
and select CHECK AMPLIFIER. If electrodes are installed, the instrument will display
the amplifier mV status. If electrodes are not installed, the instrument will display 4095
as its status.

By pressing NO in the Touch Pad or NEXT in the keyboard twice, the instrument will
show temperature screen in mV and in C degree. (Upper row displays mV values and
bottom row displays temperature in degrees Celsius.

Dye to the closed door, the instrument in the measuring chamber has a temperature
higher than room temperature.

Please insert a probe of a calibrated thermometer, in the top of the sample sensor. Adjust
R190 (located in the top right corner of the CPU board) to match values of the
thermometer reading with displayed values in the second row in the screen (See Fig. 6-2).

R190 TEMP. ADJ.

Fig. 6-2

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 61 of 74


6 Adjustments

6.3.Display, Contrast and Backlight Adjustment

The new display board E-731 is equipped with contrast and backlight adjustment possibilities.
This adjustment is done in the factory and is not in normal procedure to adjust or readjust in the
field.

Fig. 6-3 shows the location of these potentiometers in the display board. R1 is for the contrast
and R2 is for backlight.

Note: R1 and R2 potentiometers are NOT accessible through the rear cover. Any adjustment needs
to be done with the back cover open.

Fig. 6-3

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 62 of 74


7 Illustrated Parts

7. Illustrated Parts List


CorLYTE Assembly Drawing

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 63 of 74


7 Illustrated Parts

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 64 of 74


7 Illustrated Parts

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 65 of 74


7 Illustrated Parts

CorLYTE Fluidic Diagram

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 66 of 74


7 Illustrated Parts

Item Replacement Parts Description


E-724 PROBE, SAMPLE
AV-BP5014D KIT, SHUTDOWN
E-723 SENSOR, SAMPLE
E-711 KIT, PINCH TUBING
AV-BP5194D KIT, STARTUP
CN-4018 INDICATOR, FLUID FLOW, CLIP-ON
CN-4021 TRAY, ISE ELECTRODE
CN-4070 TRAY, PRINTER PAPER
CN-A1076 PROBE, BODY SUB-ASSEMBLY
E-725 ASSEMBLY, ASPIRATION
E-726 SUB-ASSEMBLY, TRAY, ELECTRODE
E-727 PUMP, PERISTALTIC PUMP
CN-A4013 PIN, CRIMP, FLUID PATH
E-728 SUB-ASSEMBLY, ROTOR, PERISTALTIC PUMP
E-729 VALVE, SOLENOID, SUBASSEMBLY
E-731 BOARD, DPB
E-732 BOARD, VPB
E-714 SUBASSEMBLY, POWER SUPPLY
E-715 SUBASSEMBLY, PRINTER
E-730 BOARD, CPU
CN-A4054 BOARD, RFID
DD-0201 KEYBOARD
DD-0202 BARCODE SCANNER

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 67 of 74


8 Interface Specifications

8. Interface Specifications

8.1. Interface Information

The CorLYTE Electrolyte analyzer is equipped with a standard serial interface output.
This interface output is intended to be used with standard commercially available
computer systems.

The data transmitted through the serial interface port employs the ASCII code.

The serial interface is terminated on the rear cover with a 9-pin male DB-9 connector.

The signal levels are as follows:


- Binary 1 = -12V to -3V
- Binary 0 = +3V to +12V

Two stop bits follow the eight data bits to complete the 10 bit word.

The baud rate is set at 115200 Baud fixed.

The maximal recommended cable length is 40 feet.

The pin assignment is as follows in RS232 port are:


pin 1... sample ground... GND
pin 2...RS232IN ... . RxD
pin 3 ... RS232OUT... . TxD
pin 4...NC
pin 5... Digital ground... DGND
pin 6...NC
pin 7.. RTS
pin 8...CTS
pin 9...Print VCC
(NC=Not Connected)

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 68 of 74


8 Interface Specifications

Software

The patient sample data is sent at the end of each measurement, the calibration report is
sent at the end of each calibration.

The interface and the printer setting are activated on the menu “instrument setting”

Note: The arrow up (e.g. out of normal range) is sent as HEX 18 (↑), the arrow down as HEX19
(↓) and the ° (degree) is sent as HEX1A (→).

8.2. Example Data String Information

Automatic Calibration Report

sx* CorLYTE*crcrlfELECTROLYTE ISEcrcrlfn3JAN92


10:51crcrlf*CALIBR REPORT *crcrlfcrcrlfDailyMaintenancecrcrlf
Performed Last : crcrlfl)2JAN92 10:3 5 crcrlfcrcrlf
STD AcrcrlfNa = -112mV (3)crcrlfK = -1392mV (3)
crcrlfCl = -106mV (3)crcrlfcrcrlfDifference A-Bcrcrlf
Na = 1402mV ()cr crlfK = 1032mV ()crcrlfCl = -1006mV ( )
crcrlfcrcrlfFluid Pack:crcrlf6 8% Remainingcrcrlfcrcrlfex

Serum Sample Report

sx* CorLYTE*crcrlfELECTROLYTE ISEcrcrlf1D3JAN92


10:5 9crcrlfcrcrlf Name:..... crcrlf .... crcrlfSample:
SERUMcrcrlfcrcrlf Sample No.13 crcrlfcrcrlfNa= 159soh mmol/LcrcrlfK
= 5.4mmol/LcrcrlfCl= 122soh mmol/Lcrcrlfcrcrlf*PERFORM DAILY
*crcrlf*MAINTENANCE ! *crcrlfcrcrlfex

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 69 of 74


8 Interface Specifications

8.3. Example Data String Information

A serial cable with straight through configuration is required for this process.
To turn on communication, enter LIS service code in [SERVICE LOGIN?].

Serial communication protocol

Result Format

PARAMETER NO OF EXAMPLE
CHARACTERS
START 1 (02 H)
CODE1 1 ‘S’ SAMPLE
CODE2 3 ANA ANALYSIS
CODE3 3 000 RESERVED FOR FUTHER USE
DATA 5 00037
LENGTH
DATE 6 ddmmyy
TIME 4 hhmm
RACK NOS 4 0000
TUBE NOS 4 0000
TEST NOS 5 00010
ID 16 PATIENT ID
NAME 16 PATIENT NAME
PATIENT 2 Human, Control, Animal
TYPE
SAMPLE 1 B Blood
TYPE
RESULT 1 33 Electrode Result (see break up)
RESULT 2 33 Electrode Result (see break up)
RESULT 3 33 Electrode Result (see break up)
RESULT N 33 Electrode Result (see break up)
Temperature 8 Temperature
STOP 1 (03 H)

Start Code
Hexadecimal value [02] marks the start of data packet

Analysis Code
Description Code1 Code3 Code3
Sample Analysis S ANA 000
Calibration C ION 000
Data Length
Determines the data length that follows after this block

Date
Represents date in DD-MM-YY format
ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 70 of 74
8 Interface Specifications
Time
Represents time in HHMM format

Rack Number
Represent the rack number. Not used here. Reserved for future use!

Tube Number
Represent the sample tube number. Not used here. Reserved for future use!

Test Number
Default internal sample test number give to each sample.

Patient Id
Operator assigned 16 characters long Patient Id. Default Patient Id is 16 white spaces.

Patient Name
Operator assigned 16 characters long Patient Name. Default Patient name is 16 white spaces.

Patient Type
Patient Type Patient Type Code
Control Sample 01
Standard Sample 02
Human Sample 03
Dog Sample 04
Cat Sample 05
Cow Sample 06
Horse Sample 07
Pig Sample 08
Sheep Sample 09
Other Animal 10
Sample Type
Sample Type Sample Type Code
Serum 1
Blood 2
Urine 3
Standard 4
QC1 5
QC2 6
QC3 7
Bicarbonate 8
Acetate 9

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 71 of 74


8 Interface Specifications
Sample Result [X]

Ion result is sent out in 33 characters. Table below shows the result format:
PARAMETER NO OF EXAMPLE
CHARACTERS
ION CODE 2 01
Unit Code 1 M
Value Sign 1 + Positive number
Value 7 0100000 1000 times Conc result
Correlation Flag 1 D Default
Error Flag 1 L Low
Sample milli-volt sign 1 + Positive number
Sample milli-volt 7 0343400 1000 times Sample mV
Sample try Count 2 01 Number of tries for successful result
Standard milli-volt 1 + Positive number
sign
Standard milli-volt 7 0343400 1000 times Sample mV
Standard try Count 2 01 Number of tries for successful result

Ion Code List


Ion Code Code
Parameter
Na 01
K 02
Cl 03
Ca 04
Li 05
Na (Std C) 06
K (Std C) 07
Cl (Std C) 08
Ca (Std C) 09
Li (Std C) 10

Unit Code List


Unit Code Unit Code
mmol/dL M
mg/dL G

Value Sign
This character represents if value is positive or negative. “+” represents positive and “-” negative
numbers

Value
Ion concentration value is sent out in 7 Digits. This number is 1000 times the concentration
determined in the analysis. For example, 01234560 represent value of 1234.56

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 72 of 74


8 Interface Specifications
Correlation Flag
Correlation Type Flag Code
Default D
User Correlation U
MGL Correlation (for M
Ca)
Flame Correlation F

Error Flag
Correlation Type Flag Code
MV Range Error V
Noise Error N
Drift Error D
Range Low L
Range High H
Too Low <
Too High >
Not Calibrated X
Not Consistent !
Cannot Measure @
Li Cannot Measure M

Reading 1 mV Sign
This character represents if value is positive or negative. “+” represents positive and “-” negative
numbers. During calibration, this represents STD-A. During sample analysis, it represents
Sample mV.

Reading 1 mV Value
Ion concentration value is sent out in 7 Digits. This number is 1000 times the concentration
determined in the analysis. For example, 01234560 represent value of 1234.56. During
calibration, this represents STD-A. During sample analysis, it represents Sample mV.

Reading 1 mV Try Count


The number represents number of tries instrument ran the standard fluid before producing stable
results. During calibration, this represents STD-A.

Reading 2 mV Sign
This character represents if value is positive or negative. “+” represents positive and “-” negative
numbers. During calibration, this represents STD-B. During sample analysis, it represents
Standard Fluid mV.

Reading 2 mV Value
Ion concentration value is sent out in 7 Digits. This number is 1000 times the concentration
determined in the analysis. For example, 01234560 represent value of 1234.56. During
calibration, this represents STD-B. During sample analysis, it represents Standard Fluid mV.
ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 73 of 74
8 Interface Specifications

Reading 2 mV Try Count


The number represents number of tries instrument ran the standard fluid before producing stable
results. During calibration, this represents STD-B.

Temperature
PARAMETER NO OF EXAMPLE
CHARACTERS
Constant 1 T
Unit Code 1 C Celsius
Value 7 0100000 100 times Actual temperature
Error Flag 1 L Low

Temperature Unit Code


Units Unit Code
Celsius C
Kelvin K

Temperature Value
100 times recorded temperature value is give these 5 characters.

Temperature Flag
Flag Name Code
Low L
High H

Stop Code
Hexadecimal value [03] marks end of data pack.

ECO 7116 SOP05-5057F Rev. 01 Effective Date: 06/16/14 Page 74 of 74

You might also like