Service Manual: Insightplus 9000 Phased Array Torso-Pelvis Coil
Service Manual: Insightplus 9000 Phased Array Torso-Pelvis Coil
Service Manual: Insightplus 9000 Phased Array Torso-Pelvis Coil
Document 780019
Revision B
GE Signa 1.5T
INSIGHTPLUS 9000
PHASED ARRAY TORSO-PELVIS COIL
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have
notation “damage in shipment” written on all copies of the freight or express bill before
delivery is accepted or “signed for” by a General Electric representative or a hospital receiving
agent. Whether noted or concealed, damage MUST be reported to the carrier immediately
upon discovery, or in any event, within 14 days after receipt, and the contents and containers
held for inspection by the carrier. A transportation company will not pay a claim for damage if
an inspection is not requested within this 14 day period.
Immediately complete a "Damage Loss Claim Form", available via MS Exchange Mail, after
the damage is found.
MS Exchange Path:
Outlook/Public Folder/All Public Folders/Medical Systems/!Global Initiatives/Information
Management/Forms/Common Forms/DAMAGE LOSS CLAIM FORM.
Send the completed form to the email address listed in the form.
For more information about the Transportation Claim Procedure, access the GE Medical
Systems Intranet and enter the following URL address (case sensitive):
ftp://3.87.40.2/globepro/qualsys/Docs/190016MF.PDF
Page 2 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
Page 3 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
Page 4 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
TABLE OF CONTENTS
Page 5 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
SECTION 1 – INTRODUCTION
Figure 2 shows a picture of the InsightPLUS 9000 Phased Array Torso-Pelvis Coil.
Page 6 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
1-2 Compatibility
This coil is compatible with the Signa HorizonTM and LX HorizonTM 1.5T.
1-3 Related Documentation
Operator’s Guide, GE Part Number 2298043-7.
Signa LX Service Methods CD, 2160623-1.
Page 7 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
Phase Correction
Phase Correction
To MRI System
Cable Assembly
Anterior Coil #3 Posterior Coil #3
Phase Correction
Phase Correction
Page 8 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
Anterior Section
Posterior Section
Preamp. 7
B13
Receiver 2
Preamp. 1
B1
Receiver 0
Preamp. 3
B5
Preamp. 5
B9
E2 E4 E6 E8
E1 E3 E5 E7 Preamp. 6
B11
Preamp. 4
B7
Preamp. 2
B3
Receiver 1
B15
Receiver 3
Active Element Inactive element
Page 9 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
Preamp. 7
B13
Preamp. 1
B1
Preamp. 3
B5 Receiver 2
Preamp. 5
B9
Receiver 0
E2 E4 E6 E8
E1 E3 E5 E7 Preamp. 6
B11
Receiver 1
Preamp. 4
B7
Receiver 3
Preamp. 2
B3
B15
Page 10 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
Page 11 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
Page 12 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
Step 1
Step 2
Step 3
Step 4
Step 5
4. At the magnet, press “Alignment Light” button to turn on the light. Move the cradle to
align the coil to the alignment marks as shown in Figure 8. Press “Landmark” button to
landmark the alignment.
Page 13 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
5. Move the coil to scan position by pushing the “Move to Scan” button, ensuring cable
does not get snagged.
6. At the console, set the protocols per the Signal section from the Signal and Noise
Protocols – Table 3-2-4.
7. Click [Save Series] to download the protocols, then click [Prepare to Scan].
8. Open [Display CVs] menu under [Research Operations] (click right mouse button). Set
the “saveinter” CV to “1” (saves the intermediate images so ROI measurements can be
performed).
9. Run [Auto Prescan]. Record the R1, R2 and TG values on the SNR Data Sheet (found
in the Appendix).
10. Run [Scan].
Page 15 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
Page 16 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
The SNR shall be calculated using the signal-to-noise ratios of the individual receiver
channels. Individual receiver SNR is defined as the mean of data within the signal ROI divided
by the standard deviation of data within the noise ROI:
Note: The SNR calculation uses the MEAN of the signal image and STANDARD DEVIATION
of the noise image.
Page 17 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
SNR Specification
The SNR measurements must be greater than or equal to the following specifications:
Page 18 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
15 1
B
A
15 1
Page 19 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
Page 20 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
Symptom #2: The coil does not pass SNR tests or exhibits poor image quality on patient
scans.
Page 21 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
SECTION 4 – MAINTENANCE
WARNING
Detach the coil connector from the scanner before attempting to clean the coil. Do not
touch the connectors with bare fingers. Never press a sharp object against the surface
of the connector. Do not reattach the connector after cleaning the coil until the coil has
dried completely. Electric shock may result if the coil is attached to the system during
cleaning or when it is wet.
CAUTION
Do not spray or pour cleaning solution directly on the coil. Do not submerge the coil in
the solution. The coil contains sensitive electronic components that could be damaged
by the solution. The coil cannot be sterilized and should be cleaned only according to
the procedure outlined in this section.
The following solutions are recommended for the coil and pad surfaces: (1) a ten percent
bleach solution (some discoloration may occur), (2) one ounce commercial dishwashing liquid
mixed with one gallon of water or (3) warm water. Apply cleaning solution to a soft cotton cloth
and proceed to clean. The user place a cotton sheet over the coil before positioning the
patient to prevent soiling of the coil. If the coil is soiled, clean the coil as described above.
Page 22 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
SECTION 5 – REPLACEMENT
Simple removals that are clearly obvious are not described here.
Unless otherwise noted, the steps for re-assembly are simply the reverse order of the steps
described for disassembly.
O-Ring Connector (1 of 2)
Containing DC Line SMB Plugs (1 through 4)
Page 23 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
O-Ring Connector (1 of 2)
Connecting DC Line
Figure 21: Anterior coil cable assembly.
Page 24 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
Page 25 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
SECTION 7 – APPENDIX
Date Comments
Date Comments
Date Comments
Page 26 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
7-2 Schematic
Passive Active
Blocking Blocking
Circuit Circuit
Anterior Coil Element 1
Output Cable
Passive Active
Blocking Blocking
Circuit Circuit
Posterior Coil Element 1
Page 27 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
Parameter Modes
Coil Name TPUPPER TPLOWER
Coil Type 3 3
Extremity Coil no no
Cable Loss 1.05 1.05
Coil Loss 1.72 1.72
Recon Scale Factor 1 1
Linear vrs Quadrature 1 1
Multiple Receiver Coil? yes yes
Number of Receivers 4 4
Starting Receiver ID 0 0
Ending Receiver ID 3 3
Multi-Coil Port Enable 9 6
Multi-Coil Port Error Enable 9 6
Additional Transmit Attenuation 0 0
Number of Fast Receivers 0 0
Starting Fast Receiver ID 4 4
Ending Fast Receiver ID 4 4
Start TA Value 90 90
Start RG Value 12 12
Multi-Coil Recon Enable 0 0
Head Default Freq. Direction 0 0
SCIC Axial 1.5 0.35 0.2 5.0 128 0.2 1.05 1.5 0.35 0.2 5.0 128 0.2 1.05
SCIC Saggital 1.5 0.35 0.2 5.0 128 0.2 1.05 1.5 0.35 0.2 5.0 128 0.2 1.05
SCIC Coronal 1.5 0.35 0.2 5.0 128 0.2 1.05 1.5 0.35 0.2 5.0 128 0.2 1.05
Page 28 of 29
USA INSTRUMENTS, INC. INSIGHTPLUS 9000 PHASED ARRAY TORSO-PELVIS COIL
REV B (06/21/01) SERVICE MANUAL
REVISION HISTORY
Rev Date Author Primary Reason for Change
A 06/11/01 Reva Zaretsky First Issue
B 06/21/01 Reva Zaretsky Revisions Requested by GE
Page 29 of 29