PSTC-Appendix C - Cleaning Test Surfaces
PSTC-Appendix C - Cleaning Test Surfaces
Appendix C
1. SCOPE
1.1 This method defines a cleaning procedure for use whenever the test method requires a clean
panel or surface.
2. MATERIALS
2.1 Absorbent cleaning material, either surgical gauze or tissue.
2.1.1 To be suitable, materials must be lint free during use, absorbent, contain no additives
that are soluble in the solvents listed in section 2.2, and be made exclusively from virgin
materials.
2.2 Solvents
acetone
2.2.3 Where toxicity and flammability requirements are paramount, a suitably blended mixture
of n-heptane and fluorinated hydrocarbon such as a refrigerant may meet requirements.
3. PROCEDURE
3.1 Dispense one of the solvents listed in 2.2 onto the panel, wiping it to dryness with fresh absor-
bent cleaning material. Repeat for a total of four washes with this solvent. See Figure 1.
3.4 During storage, panels should be protected from damage by covering with any non-stretching,
non-contaminating material such as tissue, soft felt or heavy fabric.
C-1
ADDENDUM
2 It is advisable to mark the “grip” end of each panel when using a tensile tester to do adhesion
testing. One should avoid gathering data from a scratched or roughened surface.
3 Handle test panels by the edges whenever possible. This will help keep skin oils out of the
testing area of the panels.
4 If you stack your panels when not using them, put a soft lint free tissue between the panels to
avoid scratching the testing surface. Or, consider using a test panel holder for storing these
panels – see Figure 2.
SUMMARY OF CHANGES
C-2
Figure 1. Panel cleaning procedure
C-3