Auto Sampler Sil-Ht /HT: For Shimadzu High Performance Liquid Chromatograph Service Manual
Auto Sampler Sil-Ht /HT: For Shimadzu High Performance Liquid Chromatograph Service Manual
AUTO SAMPLER
SIL-HTA/HTC
FOR SHIMADZU HIGH PERFORMANCE
LIQUID CHROMATOGRAPH
SERVICE MANUAL
This manual is intended for the use of trained service engineers only .Unauthorized
personnel are advised not to attempt the procedures shown in this manual.
SHIMADZU CORPORATION
ANALITICAL &MEASURING INSTRUMENTS DIVISION
KYOTO , JAPAN
Blank Paper
Contents
Chapter1 Introduction
Introduction •••••••••••••••••••••••••••••••••••••• P1-2
Chapter2 Safety Instructions P2-1
2.1 Precautions for Safe Operation ••••••••••••••••••••• P2-2
2.1.1 Application Precaution •••••••••••••••••••••••• P2-2
2.1.2 Installation Site Precautions •••••••••••••••••••• P2-3
2.1.3 Installation Precautions •••••••••••••••••••••••• P2-4
2.1.4 Operation Precaution ••••••••••••••••••••••••• P2-5
2.1.5 Precautions for Inspection, Maintenance Adjustment , P2-6
and Care of the Instrument •••••••••••••••••
2.2 Static Electricity Precautions •••••••••••••••••••••• P2-7
2.1.2 Typical Causes of Static Electricity Accidents•••••••• P2-7
2.2.2 Preventing Static Electricity Accidents •••••••••••• P2-8
2.3 Precautions for Mobile Phase Selection and Use •••••••• P2-11
2.4 Warning Labels •••••••••••••••••••••••••••••••• P2-14
Chapter7 Options
7.1 List of options •••••••••••••••••••••••••••••••••• P7-2
7.2 2-ml Sample loop ••••••••••••••••••••••••••••••• P7-3
<Procedure for Tubing Connection> •••••••••••••••• P7-4
<Procedure for changing the setting of tubing volume> • P7-5
CONTENTS
Chapter1 Introduction
Introduction •••••••••••••••••••••••••••••••••••••• P1-2
1.Introduction
<Characteristics of this Service Manual>
This Service Manual describes the maintenance procedures for the SIL-HTA and
SIL-HTC.
This manual is to be used by Shimadzu-certified service personnel.
Untrained personnel should not perform the maintenance procedures described in
this manual.
<Note>
Parts of the text or diagrams in this manual may differ due to improvements in
the instrument.
Shimadzu does not handle all of the parts listed in this manual as service parts.
CONTENTS
2.Safety Instruction
2 . 1 P r ec a u t i o n f o r S a f e O p e r a t i o n s
• To ensure safe operation of the instrument, read these Safety Instructions carefully before use.
• Observe all of the WARNINGS and CAUTIONS described in this section. They are extremely important for
safety.
In this manual, warnings and cautions are indicated using the following conventions.
WARNING
Signifies danger of death, or serious injury, if the instructions given are not followed.
CAUTION
Signifies danger of minor injury, or of damage to objects, if the instructions given are not followed.
“Minor injury”: Injury not requiring hospitalization.
“Damage to objects”: Damage to other nearby objects.
2. 1 . 1 A p p l i c a t i o n P r e c a u t i o n s
WARNING
This instrument is an auto sampler for use with a high performance liquid chromatography system. Operate the
instrument for its intended purposes only. Using the instrument for any other purpose could cause accidents.
WARNING
• The solvents used in high performance liquid chromatography are flammable and toxic.
The room where the equipment is installed should be well-ventilated, other-wise, solvent
vapors could cause poisoning or ignite and cause a fire.
• Provide protective equipment near the unit. If solvent gets into the eyes, or on
the skin, it must be flushed away immediately. Provide equipment, such as eye
wash stations and safety showers, as close to the unit as possible.
WARNING
• The lab table or other surface on which this instrument is installed should be level,
stable, sufficiently strong to support the units weight, and have a depth of at least
60cm. Otherwise, the instrument could tip over or fall off the surface.
• Avoid installation sites that are exposed to corrosive gases or are excessively dusty.
These adverse conditions may be detrimental to maintaining the instrument perfor-
mance and may shorten its service life. 60cm
• Do not hold the front panel when you carry this unit, and front panel could be dam-
aged or removed. Also, be sure to carry the unit with two or more people.
SIL-HTA
Part Number Power Supply Voltage Power Consumption Frequency
228-41000-31 AC100V ± 10%
228-41000-32 AC110∼120V ± 10% 150VA 50/60Hz
228-41000-38 AC220∼240V ± 10%
SIL-HTC
Part Number Power Supply Voltage Power Consumption Frequency
228-41001-31 AC100V ± 10%
228-41001-32 AC110∼120V ± 10% 300VA 50/60Hz
228-41001-38 AC220∼240V ± 10%
• Ground the instrument. Grounding is necessary to prevent electrical shock in the event of an accidental electri-
cal discharge, and important for ensuring stable operation. See Section Instruction Manual
"5.3.1 Connecting to the power outlet”.
• Do not place heavy objects on the power cord, and keep any hot items away.
It could be damaged, resulting in fire, electric shock or malfunction, if the cord
becomes damaged, contact your Shimadzu representative immediately.
• Do not modify the cord in any way. Do not bend it excessively or pull on it. It could
become damaged, resulting in fire or electric shock. If the cord becomes damaged,
contact your Shimadzu representative.
2. 1 . 4 O p er a t i o n P r e c a u t i o n s
WARNING
• Take thorough measures to prevent buildup of static electricity. Section Instruction manual“1.2
Static Electricity Precautions”, p.6. Static buildup could result in fires or explo-
sions.
• Always wear goggles when handling solvents. If the solvent gets into the
eyes, blindness could result. Should solvent get into the eyes, flush immedi-
ately with large amount of water and get medical attention.
• Do not use flammable sprays (hair sprays, insecticide sprays, etc.) near this
instrument. They could ignite and cause a fire.
• Never remove the main cover. This may cause injury or malfunction of the instru-
ment. The main cover does not need to be removed for routine maintenance,
inspection and adjustment. Have your Shimadzu representative perform any
repairs requiring removal of the main cover.
• Replace fuses only with fuses of the proper type and capacity. Any other fuses
could cause a fire.
• If the power cord plug gets dusty, remove the plug from the power cord and wipe away the dust with a dry cloth.
If dust is allowed to accumulate, fire could result.
• Replacement parts must be of the specifications given Section 4.1 “Accessories”, 11.2 “Replacement Parts”. If
any other parts are used, they could become damaged and cause a malfunction.
• If any water gets into the instrument, wipe it away immediately to prevent rust. Never use alcohol or thinner sol-
vents for cleaning this instrument. They could cause discoloration.
2. 2 S t a ti c El e c t r i c i t y P r e c a u t i o n s
1 WARNING
Liquid chromatography (LC) uses flammable organic solvent(s) as the mobile phase. LC systems are also
often used where large amount of flammable substances are present. Thus, an accident can produce large
scale damage. Operators must be constantly on guard against accidents involving fire or explosion.
The major cause of these accidents is static electricity. Devising preventative measures for static can be dif-
ficult, because the symptoms before an accident vary and can be hard to detect, since such accidents occur
as a result of several simultaneous coincidences. Recommended methods for preventing static electricity
accidents are provided below. Take thorough safety measures based on this information.
2. 2 . 1 T y p i c a l C a s u s e o f S t a t i c E l e c t r i c i t y A c c i d e n t s
Solid
If electrostatically charged liquid is allowed
Potential Accident Situation
to accumulate in an electrically insulated Liquid flowing through thin
container, the charge will gradually tubing at high flow rates
*Air bubbles in liquid facilitate
increase, and can eventually reach several
generation of static electricity
thousand volts. If this happens and an elec-
trical conductor is brought within a certain Insulated containre of
polyethylene, etc.
distance of the container, electrical dis- Spark
Electrical
charge will occur, releasing thermal energy Discharge
Flammable gas present
which will ignite any flammable gas in the in container
Spark
Electrical
Discharge
Flammable organic solvent
with large electro-static
charge
TIP
• Take two or more of the following preventative measures.
• If large amounts of flammable solvents are collected in a large container, implement preventative measures 1, 2,
and 3 below.
Preventative Measure 1
Use a metal container for the waste liquid, and ground the container. This will ensure that the electrical
charges of the container and liquid pass to the ground.
TIP
• Be sure to ground the metal waste container properly. If the grounding wire is not properly attached or con-
nected to the ground, static electricity can build up in the container.
• Some metal containers have surfaces that are laminated or oxidized, and therefore do not conduct electricity.
After grounding the metal container, use a tester to verify that electricity is conducted to the ground
• If the liquid to be drained into the waste container is virtually non-conductive (10-10 S/m or less), it will be neces-
sary to add properly conductive, and therefore safe, liquid to the tank. This conductive liquid may be added
beforehand.
Cover the spaces between the tubing and the sides of the inlet and outlet openings of the waste container
with caps or other protective covering. This will prevent any sparks generated outside the container from
getting inside.
Preventative Measure 3
Keep electrostatically charged objects, including the human body, away from the waste container. To prevent
electrostatic charging of the human body, take the following precautions:
• Wear anti-static clothing and shoes.
• Ground the human body with anti-static wrist straps. For safety, the wrist strap should be connected to the
ground using an intervening resistor of about 1MΩ.
• Spread anti-static matting or the like on the floor, to make the floor conductive.
TIP
Persons who have not taken anti-static precautions should touch a grounded metal object before coming near the
waste container to drain static charges.
Preventative Measure 4
• Use tubing with an inner diameter of at least 2mm for drain lines with high flow rates.
TIP
Air bubbles in liquid can multiply the electrostatic charge by a factor of 20, 30 or more. Periodically check the tub-
ing connections for leaks.
If it is not possible to use a conductive waste container, take the following precautions:
1
• Ensure that the end of the inflow tubing is always submerged inside the container. Place some type of
grounded metal object, such as a ground wire connected to the instrument, into the liquid.
TIP
• The above precaution will be ineffective for low conductivity (less than 10-10 S/m) liquids.
• Use as small a container as possible to minimize damage in the event of fire.
• Ambient humidity exceeding 65% will prevent static. Keep the room at a proper humidity.
TIP
Anti-static equipment (anti-static clothing, shoes, and matting) and charge measurement equipment (potentiome-
ter) are sold by speciality manufacturers.
2. 3 P r e c a u t i o n s f o r M o b i l e P h a s e S e l e c t i o n a n d U s e
1
CAUTION
PEEK resin possesses high mechanical strength, and among synthetic resins, it is the most suitable for use
in high performance chromatographs. The following are safety-related considerations which require certain
limitations in the use of PEEK, as follows.
The following mobile phases should not be used when PEEK resin parts are used in the plumbing.
• concentrated sulfuric acid
• concentrated nitric acid
• dichloroacetic acid
• acetone
• tetrahydrofuran (THF)
• dichloromethane
• chloroform
• dimethyl sulfoxide (DMSO)
These mobile phases weaken the PEEK resin, which could result in cracked plumbing and mobile phase
leaks.
NOTE: Briefly using a weak solution of no more than 0.5% acetone in water (e.g. to check gradient performance)
will present no problems.
PEEK Tubing
1. The resistance to pressure of PEEK tubing varies depending on the internal diameter, manufacturing pro-
cess and solvents used. For this instrument, be sure to use the PEEK tubing having an internal diameter
of 0.13mm, provided standard with this instrument. Be advised that Shimadzu will assume no responsibil-
ity for any mishaps associated with the use of PEEK tubing provided by another manufacturer.
For example, the PEEK tubing might rupture if acetonitrile is delivered at a pressure continuously exceeding
27.5MPa (280kgf/cm2), posing considerable danger.
2. If PEEK tubing is bent to a curvature with an excessively short radius, that portion of tubing will be weak-
ened. Be sure that the curvature radius of the PEEK tubing is no shorter than 10mm. Avoid forcibly bend-
ing and fixing the tubing, and strive to obtain as natural a curvature as possible.
2. 4 W a r n i n g L a b e l s
1 To help users operate the unit safely, warning labels are affixed in places where special attention is required.
Should any of these labels peel off or become damaged, obtain replacements from Shimadzu.
Warning labelޓSIL-HTA㧦P/N 228-40955-01
ޓޓޓޓޓޓޓSIL-HT C㧦P/N 228-40955-02
Caution/Warning label
(P/N 228-38924-02)
CONTENTS
3 Fe a tu r e s a n d S y s t e m S p e c i f i c a t i o n s
3. 1 S ett i n g s a n d O p e r a t i o n
3. 1. 1 O p e r a ti o n s e t t i n g s
3. 1. 2 U n i t O p e r a t i o n
Sample injection
The sequence of operations involved when injecting samples is as shown in the flow and operation diagrams
below. Use these diagrams to become familiar with the operation of the unit.
2
<Flow line>
Sample loop
Rinsing port
Solvent delivery
Measuring pump pump
Low- High-
pressure pressure
valve valve
Sampling
needle
Mannual prime
Injection
port
Rinse solvent
Column
3
4 4 5
Rinse
Mannual
solvent Column
prime
2 Sampling
needle 2
1 2 1 Solvent
delivery
pump
Injection Low- High- 6
port pressure pressure
valve
5 valve
3
3
4 4 5
Rinse
Mannual
solvent Column
prime
1 2 1 Solvent
Sample delivery
2 pump
Low- High- 6
Injection pressure pressure
port valve
5 valve
3
3
4 4 5
Rinse
Mannual solvent Column
prime
3
4 4 5
Rinse
Mannual
solvent Column
prime
Rinse
Mannual
solvent Column
prime
Rinse
Mannual
solvent Column
prime
8. Standby (READY)
CONTENTS
4.Specifications
Injection system Full volume sample injection, variable injection volume (zero sample loss
during injection)
Injection volume setting range 0.1 to 100µL (standard), 1 to 2000µL (optional)
(0.1 to 0.9µL in 0.1µL increments, 1 to 2000µL in 1µL increments)
Number of samples processed SIL-HTC: Sample cooler model
350 (w/1mL sample vials), 140 (w/1.5mL sample vials), 100 (w/4mL sam-
ple vials),
384 (w/4 micro titer plates each with 96 wells), 1536 (w/4 micro titer plates
each with 384 wells)
384 (w/4 Deep well MTP each with 96 wells)
SIL-HTA: Model without sample cooler
350 (w/1mL sample vials), 210 (w/1.5mL sample vials), 100 (w/4mL sam-
ple vials),
384 (w/4 micro titer plates each with 96 wells), 1536 (w/4 micro titer plates
each with 384 wells)
384 (w/4 Deep well MTP each with 96 wells)
Sample vials 1.5mL glass, 1.1mL glass, 4mL glass,
0.3mL glass (w/plastic spacers), 0.3mL glass (4mL vial storage type),
1mL plastic, 0.2mL plastic, 4mL plastic, 96-well micro titer plate, 384-well
micro titer plate, 96-deep well
Injection volume repeatability RSD < 0.3% (at 10µL injection)
Carryover Less than 0.01% (using naphthalene analysis conditions)
Injection volume accuracy +/- 1% (at 50µL injection, n = 10)
Number of repeat injections 1 - 30 times/sample
Analysis time setting 0.01 minute steps (< 1000 minutes), 0.1 minute steps (> 1000 minutes)
Number of injection parameter steps 102 steps max.
Sample aspiration rate 0.1 to 15µL/sec (0.1µL/sec increments, 1 to 15µL/sec 1µL/sec incre-
ments)
Rinse aspiration rate Variable (1 to 35µL/sec, 1µL/increments)
Applicable pressure 35.0MPa max., 20.0MPa max. with PEEK tubing
Sample cooler System Direct cooling system (environment conditions: room temperature below
(SIL-HTC only) 30°C with humidity less than 70% when the cooler temperature is set to
4°C), dehumidification function built-in
Temperature setting 4 to 40°C (can be cooled down to 4°C when the room temperature is
range below 30°C and the humidity is below 70%.)
Temperature accu- ±3°C
racy
4.2 Others
Items Specifications
Liquid contact materials Stainless (316 SS), ceramic, Teflon (R), FEP, GFP, sapphire, PEEK
Ambient temperature 4~35°C
Ambient humidity ~80%
pH range 1~14
Dimensions and weight W540 × H415 × D500 mm, 40 kg (SIL-HTC), 35 kg (SIL-HTA)
Power supply SIL-HTA
Power
Part No. Power supply voltage Frequency
consumption
228-41000-31 AC100V ± 10%
228-41000-32 AC110 ~ 120V ± 10% 150VA 50/60Hz
228-41000-38 AC220 ~ 240V ± 10%
SIL-HTC
Power
Part No. Power supply voltage Frequency
consumption
228-41001-31 AC100V ± 10%
228-41001-32 AC110 ~ 120V ± 10% 300VA 50/60Hz
228-41001-38 AC220 ~ 240V ± 10%
Error display Exists (Error display and stop at the time of malfunction)
CONTENTS
Keyboard/Display
Specifies operation parameters.
Displays operation parameter
settings.
S IL -H T C
Outlet tubing
The column is
Outlet for connected here.
dehumidity
Power Switch
Turns the power ON/OFF, Pressing once turns
the power on. Pressing a second time turns the
power offf.
Leakage drain port
If leak occurs inside this unit, the
leaked solvent is discharged from Power Switch Solvent filter
this hole into waste bottle. Filters solids from the rinse
solvent.
(Pore size: 10 microns)
POWE POWE
R㨪 R㨪
Button in Button out
(power ON) (power OFF)
Low-pressure valve
Switch valve for determing the direction of
aspiration/delivery of sample, reagent and
rinse solvent.
Rinsing port
Needle is inserted here to clean
its outer surfaces.
Leak sensor
For detecting leaks inside
Injection port the unit.
Samples delivered by the sampling
needle pass from this port through the
high pressure valve, and into the column.
Power supply
connector
For connection of
power cord.
CONTENTS
6.Installation
SILHT SCL10Avp
Items Contents IM IM
External control Event cable - P5-29
connectors External control connectors - P3-6
(System controller) Connections - P5-30
Soft switching settings - P6-77,78
External control Event cable P31 -
connection Connecting the event cable P32 -
Rubber Spacer
Removing Procedure
(1) Temporarily remove the transparent front cover (only on models with a cooler).
(2) Use a hex wrench to remove the Z-mount shipping screw.
(3) Carefully lift and remove the rubber spacer used to secure the needle during s
hipment.
(4) Re-attach the front cover (only on models with a cooler).
Caution
To avoid damage to the sample loop, be sure that the customer fully understands t
he following caution.
The drive mechanism of the autosampler moves freely after the shipping screw is r
emoved. The actions listed below can deform the sample loop, which in the worst c
ase may cause tangling or damage to the loop during subsequent injections. These
actions must be avoided once the shipping screw is removed.
- Do not move the XY arm backward or forward while the Z-mount is
in the right most position.
- Do not move the Z-mount too far to the back and right.
- Do not leave the Z-mount at the back for more than 5 minutes.
- Do not apply sufficient force to distort the sample loop.
Tip
Channel 1(SIL) on the system controller is
reserved for SIL.
Inlet tubing
Outlet tubing
< Procedure of plumbing "Pipe, IN" and "Suction filter pipe of RINSE liquid"
Screw Screw
(4) Fix the plumbings onto the left clamp and draw out from instrument as photo.
CAUTION
The volume of waste liquid produced pre analysis is about 200uL ,under default
conditions(default values). If a large number of sampler is analyzed continuously,
provide a larger waste container . If the container is too small, it will overflow.
<Drain outlet>
Outlet for
dehumidity
Rinsing port drain
outlet
Assemble the drain tube by the use of the teflon tube in the standard accessory refering
to photo.
Assembling the
drain tube (unit 60 35
mm)
35
40
35
10
100
40
Cut the other end of the drain tubing so that it will penetrate a short distance into the
waste cntainer
Caution
The end of the drain tubing must not be submerged into the waste liquid .
Waste liquid will leak from the rinsing port .
If necessary , tape the tubing to the container’s rim to prevent submersion.
CONTENTS
Chapter7 Options
7.1 List of options •••••••••••••••••••••••••••••••••• P7-2
7.2 2-ml Sample loop ••••••••••••••••••••••••••••••• P7-3
<Procedure for Tubing Connection> •••••••••••••••• P7-4
<Procedure for changing the setting of tubing volume> • P7-5
7.Options
ASSY OP loop(P/N228-37552-91)
Loop volume
-91: 3927 +403ul
-383ul
7-3
7.Options
Ring part A
Ring part B
Caution
Insert the pipe until it hits the bottom of the connection hole. If the pipe dose not reach
the bottom, dead volume will occur and cause a peak to spread out.
(1)Turn on the power and operate the keys on the controller panel as below.
f5 key(menu)/5(maintenance)/func key/f3 key(SIL)/4(CALIBRATION),enter key
(2)Inputting five-digit password at this time displays “f2 OK”. Press the [f2]key on that
occasion. The password set at delivery is[00000].
(5) Setting is stored and the display returns to the CALIBRATION function menu pop-up
screen. Operate the keys on the controller panel as below
8(EXIT),[enter]/5(EXIT),[enter]/f5(MENU)
Caution
The inner volume of this loop is approx.4mL.
After the installation, replace the inside of sample loop with the mobile phase sufficiently
before injecting a sample.
7-6
7.Options
SCSI board
SCSI board
7-8
7.Options
2-channel board
Note
Both of DIP switch (SW1) should be set “ON” on the 1 st board and “OFF” on the
2 nd board.
2 channel board
Input connector
1 6
CONTENTS
8-1
8.FACTORY SCREEN
8-2
8.FACTORY SCREEN
8-3
8.FACTORY SCREEN
SILHTA/C
Ver
SN
(3) Enter the 5. MAINTENANCE screen in the MENU screen. Press the func key
once. Press the f3 key (SIL) to enter the SIL screen.
SIL screen
(4) Press the 5 key, and then press the enter key to enter the FACTORY screen.
FACTORY SCREEN
<Required Parts>
- Fixture (for 1.5 ml vial rack) P/N228-39802-02
- Fixture (INJ PORT teaching) P/N228-34416 one piece
- Right/left rack for 1.5 ml vials (standard accessories)
Jig
INJ PORT Teaching
Jig
For Rack Via
l.5ML
Type C Rack for 1.5 ml vials Type A Rack for 1.5 ml vials
<Adjustment Method>
(1) Set the racks for 1.5 ml vials (right/left) to the unit.
(2) Enter 1. RACK TEACHING screen in the FACTORY screen.
(Press the 1 key, enter key, and the f2 key in this order in the FACTORY screen to enter
the RACK TEACHING screen.)
8-5
8.FACTORY SCREEN
CORNER 2 CORNER 2
CORNER 1 CORNER 1
C Type Rack A Type Rack
(4) Mount the fixture (for 1.5 ml vial rack, INJ PORT teaching) on CORNER 1
(RACK-L) shown in the screen.
Adjustment standard
(Front/Rear/Right/Left) direction: Within 0.2 mm from the center of the Jig
(Up/Down): Within 0.3 mm from the end of the Jig
(6) Press the enter key after adjustment. The needle mount (Z mount) moves to
CORNER 2 (RACK-L).
(7) Mount the Jig (1.5 ml vial hook, INJ PORT teaching) on CORNER 2 (RACK-L) and
perform the adjustment in the same way.
Note 1.
Regarding the func key in the screen
f1 f3 f5
Note 2.
Teaching can be performed with 1.5 ml vial racks only.
<Required Parts>
- JIG (INJ PORT teaching) P/N228-36837
8-7
8.FACTORY SCREEN
(3) Perform the teaching on the injection port according to the teaching screen.
Adjustment standard
(Up/Down): Within 0.3 mm from the end of the JIG
(Front/Rear/Right/Left) direction: Should be in the center of the JIG
Within 0.3 mm
8-8
8.FACTORY SCREEN
Inner panel
8-9
8.FACTORY SCREEN
Press the f2 key (6V) for the high pressure valve adjustment.
Press the f5 key (LPV) for the low pressure valve adjustment.
(3) Perform the UP-SIDE and DOWN-SIDE adjustment of the valve slit according to
the screen instruction.
Bad
Rotor
High pressure valve <6V> UP-SIDE High pressure valve <6V> DOWN-SIDE
Low pressure valve <LPV> UP-SIDE Low pressure valve <LPV> DOWN-SIDE
8-11
8.FACTORY SCREEN
4. Deleting the teaching data of the injection port (ERASE INJ P-DATA)
Enter the 4. ERASE INJ. P-DATA screen in the FACTORY screen.
(Press the 4 key, and then press the enter key in the FACTORY screen.)
Press the f2 key and delete the teaching data of the injection port.
When the injection port is stored in abnormal position, execute this command and
delete the teaching data to perform the position adjustment again.
5. Deleting the teaching data of the rack vial position (ERASE RACK-DATA)
Enter 5. ERASE RACK-DATA screen in the FACTORY screen.
(Press the 5 key, and then press the enter key in the FACTORY screen.)
Press the f2 key and delete the rack vial teaching data.
(When the adjustment data of the vial position on the rack is stored in the abnormal
position, execute this command and delete the teaching data to perform position
adjustment again.)
8-12
8.FACTORY SCREEN
8-13
8.FACTORY SCREEN
8-14
8.FACTORY SCREEN
Note:
Firmware is loaded on the controller unit (SCL) and the sampler unit (SIL) of this
instrument. Upgrading must be conducted respectively.
<How to upgrade>
(1) Supply power with the f1 key held down and release the f1 key while the LED for
run and the LED for act are blinking alternately, so as to start up in the
maintenance screen.
Note:
Upgrade the version with the power connected to the main unit SILHTA/C being
OFF (disconnected).
(2) Enter the 5. MAINTENANCE screen in the MENU screen. Press the func key once.
(3) Insert the floppy disk for upgrading into the drive. Press the Ver up (f5 key) to
enter the version up screen.
f1 f2 f3 f4 f5 func
8-15
8.FACTORY SCREEN
(4) Select the program (SCLxxx.BNL) for upgrading, press SEL (f5 key) to start
upgrading.
(5) After the end of upgrading, the screen goes off. Turn off the power of the main unit
SILHTA/C, and then supply the power again.
Note:
- Do not turn OFF the power during upgrading. When the power is turned OFF
during upgrading. Turning OFF the power during upgrading not only fails
upgrading, but may damage ROM (SCL), requiring ROM replacement.
<Required parts>
- Event cable (P/N 228-39249)
- Upgrading VP kit
228-35530-91 (for 100V))-38 (for 200V), one set
Event cable
- PC for upgrading use, Upgrading software
Upgrading VP kit
<Upgrading procedure>
To execute upgrading, refer to the ROM Version Upgrading Kit Instruction Manual
P/N228-30305 of the LC-10Avp series.
(1) Connect the RS-232C cable of the upgrading fixture to the serial port of the PC. (If
there are two or more ports, either one will do.)
(2) Optical link REMOTE terminal of the unit SILHT and the fixture optical port for
upgrading are connected. Connect the power supply to the upgrading fixture using
the power supply adapter attached to the upgrading kit.
Connect the RS-232C
cable to the serial port of
the PC
Connect the optical link
REMOTE terminal of SILHT.
(3-1) Short-circuit the EVENT cable (black/gray) and connect it to the EXTcontrol
terminal at the lower rear surface of the main unit.
(3-2) Turn ON the power of the instrument and connect the event cable in (3-1) to
automatically start up the instrument in upgrading mode. When upgrading starts in
the screen, the auto-sampler does not perform initialization.
For the VP series, power is supplied while holding down the VP key.
As this instrument does not have the VP key, supply power with the above procedure.
Short-circuit the
EVENT cable
(black and gray).
Connect to the
EXTcontrol
terminal.
8-17
8.FACTORY SCREEN
program.
3. Set the floppy disk containing the upgrading data to the PC.
4. Specify the directory (normally A:¥) containing the upgrade file of the floppy in
"File/Change Directory", and press OK. (File name is not displayed until OK is
pressed.)
5. Upgrade file list appears in the "Files" box. Click on and select the target file.
6. Pressing the "Update" button starts upgrading. The total amount of byte and the
amount of the transferred byte are displayed below the dialog box.
7. Finally, a message dialog box that notifies the end of upgrading appears.
Note:
Do not turn OFF the power of the main unit SILHT right after the dialog box notifying
the completion of upgrading appears. When power supply is turned OFF right after
upgrading, ROM contents may be damaged. Turn OFF the power after 30 seconds or
more from the appearance of the dialog box showing upgrading completion.
The screen that appears when you press The screen that appears when
the f3 key to start up the system. the SLEEP key is pressed.
8-19
Blank Paper
Chapter9 Electrical system
CONTENTS
9.Electrical system
Storage parameter
board Items Page
PB-SCL1A Storage parameter in FLASH EEPROM P9-3
Storage memory in SRAM P9-3
PCB 10AVP CPU ASSY Storage parameter in FLASH EEPROM P9-3
Storage memory in SRAM P9-3
Board replacement
board Items Page
PB-SCL1A When the original FLASH P9-5
EEPROM(M9,M10) is mounted
To replace new FLASH P9-6
EEPROM(M9,M10)
PCB 10AVP CPU ASSY When the original FLASH P9-7
EEPROM(M3) is mounted
To replace new FLASH P9-8
EEPROM(M3)
9-2
9.Electrical system
right side
CAUTION:
When replacing PB-SCL1A , normally mount the original EEPROM (M9, M10). After
9-4
9.Electrical system
EEPROM is replaced, system information must be set again. (Refer to 9.2 Strage parameter)
PB-SCL1A
(2) Remove the connector on PB-SCL1A. Loosen the screw and remove PB-SCL1A
(3) Remove the FLASH EEPROM with special tool.
ROM1 P/N228-34650-92
exclusive-use tool
Removal and mounting of Using the exclusive use tool, remove two EEPROM
EEPROM (ROM1 M9) (ROM1 M9, ROM2 M10) of the board installed in the
(ROM2 M10) system and mount them to new board.
Mounting backup battery Before mounting the backup battery, check the voltage of
it. (New battery has voltage of 3.0 V or more.)
If the battery has been exhausted, replacement is
required.
Date and time setting Set the date using SET.D.T (f2 key) on the System screen.
Parameter setting Set the system, analysis parameter, SIL parameter (needle
stroke etc.) etc.
Note: When the power is turned OFF, SRAM memory is backed up by the power
supplied by the backup battery.
When the backup battery and the board are replaced, make setting again or load the file
saved in the floppy disk.
Mounting backup battery Before mounting the backup battery, check the voltage of
it. (New battery has voltage of 3.0 V or more.)
Mounting new board Mount a new board (new EEPROM (ROM1 M9, ROM2
(Setting DIP SW) M10) to the main unit.
Setting the date and time Set the date with SET.D.T (f2 key) in the system menu.
Initialization (Press f3 key With pressing f3 key, turn ON the power. While LED is
and turn ON the power, blinking, release f3 key and start with the check program
initialize with f2 key.) menu.
Press f2 key to check the memory and perform
initialization.
Input of serial number Start the factory menu. (With pressing f1 key, turn ON
the power, and release f1 key while LED is blinking.)
Input the serial number with S/N(f1 key) menu of SCL(f1
key) menu in MAINTENANCE screen (MENU 5).
Parameter setting Set the date and time, system, analysis parameter, and SIL
parameter (needle stroke etc.)
Note: When the power is turned OFF, SRAM memory is backed up by the power
supplied by the backup battery.
When the backup battery is replaced and when the board is replaced, make setting again
CAUTION:
When replacing the PCB10AVPCPU (SIL-CPU board), normally mount the original
EEPROM (M3). After EEPROM is replaced, system information must be set again and
adjusted.
FLASH EEPROM
1.When the original EEPROM (M3) is mounted (PCB 10AVP CPU ASSY)
Removal and mounting of Using the exclusive-use tool, remove EEPROM M3 of the
EEPROM (ROM M3) board installed in the system and mount it to a new board.
Mounting backup battery Before mounting the backup battery, check the voltage of
it. (New battery has voltage of 3.0 V or more.)
9-7
9.Electrical system
(Initialization)
Mounting backup battery Before mounting the backup battery, check the voltage of
it. (New battery has voltage of 3.0 V or more.)
If the battery has been exhausted, replacement is
required.
Mounting new board Fix new EEPROM (M3) to a new board, and mount it to
the main unit
When ROM type is “MONITOR VP ROM”,
SIL(ROM)version upgrading is required
Initialization (ALLCLEAR) With pressing f1 key, turn ON the power. While LED is
(with pressing f1 key and blinking, release f1 key and start with the factory menu.
turn ON the power.
Execute FACTORY menu 7 ALL CLEAR of SIL menu
Initialize with FACTORY
(func f3 key)
menu.)
(Input of serial number) Start the factory menu. (With pressing f1 key, turn ON
the power, and release f1 key while LED is blinking.)
Input the serial number with S/N (f1 key) menu of SCL (f1
key) menu in MAINTENANCE screen (MENU 5).
9-8
9.Electrical system
W:WHITE
B:BLACK
CONTENTS
10.MAINTENANCE1
10.1 MAINTENANCE
<Consumable parts>
Service SILHT
Parts Name Maintenance Item Manual Instruction
Page (Page)
Needle Seal Replacement P11-6 P132
Measuring pump Seal Replacement - P135
Measuring pump Plunger Replacement - P139
LPV rotor and stator Replacement - P141
HPV rotor and stator Replacement - P145
Needle Replacement P11-4 P149
Fuse Replacement - P151
Panel Replacement - P152
Service SILHT
Maintenance Item Manual Instruction
Page (Page)
Piping(PEEK tube) P11-9 -
Sample loop P11-11 -
Removing the INJ port piping P11-10 -
<Maintenance List>
Parts Name Maintenance Item Service
Manual
Page
Z mount Removing Z mount P10-5
Removal of Z feed screw P10-10
Z feed screw grease-up P10-11
Z shaft oiling P10-11
Z sensor replacement P10-12
XY ARM Removing XY ARM P10-15
Teaching Injection Port P8-7
Rack Vial P8-5
Belt Adjustment X belt adjustment P10-17
Y belt adjustment P10-18
Z belt adjustment P10-18
Valve Mount Removal of valve mount P10-19
Removing and mounting the high-pressure valve P10-20
Adjustment of the high-pressure valve P8-9
Removing and mounting the low-pressure valve P10-21
Adjustment of the low-pressure valve P8-9
Disassembling the measuring pump (plunger, P10-23
spring, and bearing)
10.2 Z mount
Frame bar
Frame stay ASSY
2) Remove the frame stay ASSY (P/N228-41010-91) and frame bar (P/N228-37036).
Remove two flat head screws M3x4.
Frame bar
Front panel F
Side panel L
10-6
10.MAINTENANCE1
Panel L
Panel L2 Panel L3 Panel L1
Panel L
To remove Z cover, rotate Z motor with hand and lower the needle approx 25 to 28 mm
Remove three M3x6 screws and the spring.
Z cover
7) Remove the sample loop from Z mount (Remove the male nut bind fixing screw).
8) With pushing up the white connector, draw out the cable FPC-Z (P/N228-37541-91).
FPC-Z
White connector
the fence XY
11) Hold down X belt with your finger and remember the degree of tension.
12) Loosen X mount pulley screw located under the encoder cover of X arm.
13) Loosen four setscrews and pull out the shaft X (P/N 228-36363) to the right, and
remove Z mount (P/N228-37502-92). Setscrew M3x4 4 pcs
set screws
set screws
Z mount
6) Rotate Z feed screw pulley (counterclockwise viewed from above) and remove it.
10-10
10.MAINTENANCE1
the sensor
3) By pushing up the white connector from the board LC-2K INJ2, disconnect the cable
(FPC-Z).
4) Remove the cable (FPC-Z) from the fence XY. Remove the fence XY.
Remove the cable (FPC-Z)
from the fence XY
fixing screw
7) Remove the remained needle Z sensor and peel off FPC cable.
PCB INJ2
FPC-Z
upside of Z mount
FPC-Z
FPC-Z
side of Z mount
10) After having mounted the sensor, mount the feed screw, belt, Z motor, XY fence, and
cover.
1) Remove the front panel F (type C) and inner panel (type A).
(To perform the work with ease, remove the side plate L and
panels L1, L2, and L3
two screws
5) Loosen two screws sems M3x6, pull forward the shaft Y (P/N 228-36354) and remove
the ASSY base (SIL).
Screw sems M3x6 2 pcs
10-16
10.MAINTENANCE1
Z belt
Y belt
X belt
Adjust Y belt tension using Y motor fixing screw (sems M4x16 4 pcs).
1) To loosen Y motor fixing screw
Remove the valve mount.
Remove the board (SIL-MAIN LC2K-COOL).
Panel R (228-37511-91) (screw truss M4x10 5 pcs) should be removed.
Y belt
Z motor fixing
Z belt screws
Valve mount
valve mount
<Mounting>
1) Mount the valve from left side and fix it with screw. As the position is decided by the
pin, adjustment is not necessary.
2 )Mount the valve sensor holder (P/N 228-35981).
Take care so that the sensor may not get contact with the valve disc.
3) Apply the grease white (P/N017-30829-01 polirex #2 400G) thinly to whole
circumference of motor side gear. (At adjustment in the factory, 2 g of grease has been
applied.)
(Grease should not stick out from the gear so that the grease may not stick to the
sensor disc.)
Set the valve mount to the main unit and conduct teaching.(refer to 8.3)
gear(high-pressure valve)
low-pressure valve
valve mount
<Mounting>
1) Mount the valve from left side and fix it with screw. As the position is decided by the
pin, adjustment is not necessary.
2) Mount the valve sensor holder (P/N 228-35981). Take care so that the sensor may
not get contact with the valve disc.
3) Apply the grease white (P/N017-30829-01 polirex #2 400G) thinly to whole
circumference of motor side gear. (At adjustment in the factory, 2 g of grease has been
applied.)
(Grease should not stick out from the gear so that the grease may not stick to the
sensor disc.)
Set the valve mount to the main unit and conduct teaching. (refer to 8.3)
Whether the assembly plunger moves freely must be checked after head holder (14) is
installed. (Push the lower part of the assembly plunger SIL to check.)
seal holder
2) Draw out the seal holder (P/N228-37530-91), and then remove the plunger SIL
(sapphire) (P/N228-37531-91) with the plunger removing tool.
two cap
plunger SIL bolts
(sapphire)
3) To remove the head holder (P/N228-36335-01), loosen two cap bolts M4x30
alternately and remove the head holder with taking care so that the spring
(P/N228-36387-01) may not protrude.
Bearing is pushed and mounted with the head holder spring ring. When mounting
the bearing, pay attention to the direction of it.
When mounting the head holder, screw hole for mounting the plunger should come to
the center.
groove
slit disc
10-24
10.MAINTENANCE1
Groove side
Note:
After disassembling, adjust the position of slit disc.
Refer to the position adjustment of measuring pump slit disc
[Adjustment]
Turn ON the power. Then, turn OFF the power in the stop state of above 2 of
initialization.
[Criteria] 117degrees
Slit disc position should be in the range 72 to 117 degrees in 72 degrees
clockwise direction when viewed the measuring pump slit disc
from the back.
If the slit disc position is not in this range, remove the disc fixing
screw, rotate the disc so that it comes within the range and fix it
again with the fixing screw.
(Fixing hole of disc is set in the interval of 45 degrees.)
If the slit position is near 72 degrees and it is difficult to judge the above mentioned
range, fix the disc on the 117 degrees side.
[Criteria] 24 degrees
disc slit
(Turn ON the power and on the READY state of completion of
initialization (on the notch of slit disc), turn OFF the power.
Rotate the slit disc by hand counterclockwise until it
mechanically stops. At this time, viewed from the rear of the
system, the slit of measuring pump should be in the range 24 to 360 degrees
360 degrees after passing the sensor position in the rotating
direction.
10-26
10.MAINTENANCE1
As soldering is usually difficult, the photosensor (high pressure valve, low pressure
valve, and measurement pump) attached to the valve mount is replaced by the set of
the sensor (FPC-Z) assembly.
Sensor replacement of the part having the following part number is also possible.
Photo interpreter: GP1S94 P/N061-7557-01
Note: After replacement of the sensor, it is required to perform the valve teaching for
high-pressure, low-pressure and adjust the measuring pump slit disc.
2) Remove the fixing screw of measuring pump slit sensor and remove the sensor.
3) Fix a new sensor with the screw. Next, remove the horizontal direction detecting
sensor of the measuring pump and mount the new sensor. Mount the measuring
pump motor.
With referring to the figure, remove the sensor in the order.
4)After having mounted the valve mount to the main unit, mount the following
connectors to SILMAIN board.
Connector CN21 Measuring pump
Connector CN19 Low-pressure valve motor
Connector CN18 High-pressure valve motor
Connector CN5 V sensor
5)Perform teaching for high-pressure valve and low-pressure valve and adjust the
position of slit disc of measuring pump
In this occasion, as the clearance between the cover and the slit is small, slide the cover
straight toward the rear of the unit and then remove it.
X encoder cover
When mounting and removing, take care not to touch the slit and bend it.
slit holder
0.3+/-0.2mm
photo interrupter
P/N061-7155-01
X PCB sensor
SIL-MAIN
LC2KCOOLASSY
PBSCL-1A
the rear panel
5) To remove the board, loosen five fixing screw (truss M4x10) of panel R
(P/N228-37511-91) and one screw (sems M3x6) of sensor cover XY (P/N228-36378-01).
panel R
10-30
10.MAINTENANCE1
6) Loosen the sems screw M3 and remove the sensor cover XY(P/N228-36378-01).
0.3+/-0.2mm
Adjust clearance between the sensor and the slit to the position 0.3 +/-0.2 mm from
the right of sensor viewed from the main unit front.
10-31
10.MAINTENANCE1
P/N228-37540-91 FPC-Y
photo interrupter
P/N061-7155-01
photo interrupter
P/N061-7155-01
LC2K INJ2
CAUTION:
On the SIL-HT, the SIL adjustment parameter is not written in the EEPROM of LC2K INJ2.
Even if the board is replaced, it is not necessary to change the adjustment parameter.
Remarks:
On the LC2010. the SIL parameter is written in the EEPROM of LC2K INJ2.
When replacing the board, the information "Load and Save" for the parameter is
necessary. Refer to LC2010 Service Manual.
Disconnect the connector and remove the board fixing screw from the front side then
replace the board.
10-34
10.MAINTENANCE1
rack sensor
rack sensor
3) Insert the Philips screwdriver into the portion from which INJ port ASSY was
removed, loosen the sensor fixing screw, then remove the sensor. (Philips screwdriver
of 150 mm length is necessary)
7) Remove the dehumidifying Peltier power connector and thermistor connector from
PCB. Work is made easier if the flat cable also is removed at this time.
flat cable
Remove duct.
10) Loosen main base fixing screws (two at the front and one at the back).
11) Release the harness via the two Peltier harness fixing clamps of the rack plate.
12) Remove the Peltier power supply connector of the rack plate from the PCB.
V mount
13) Lift up the base slightly and pull out the rack plate Peltier harness in a straight
line in the direction of the rear of the device.
14) Slide the Z mount to the left side of the device, now move it to the back and remove
the four rack plate fixing screws.
15) Place a hand under the Autosampler and push up the radiation fan of the rack
plate Peltier to be able to pull out the rack plate ASSY. The whole unit can be removed
if it is pulled upwards at an angle, but be careful not to hit the needle when doing so.
16) Degrease the area surrounding the main base Peltier mounting hole with acetone
and then coat the area with water repellent.
If white rust has already appeared on the base, wipe away the rust with a commercial
rust remover before commencing work. At this time, be careful not to scatter the
removed rust specks
17) Remove the 12 M4 truss screws on the rack plate to enable separation of the Peltier
unit. A slight amount of force is required in the separation task because the Peltier
unit is adhered to the rack plate with thermal compound.
thermistor transistor
2) Coat the thermistor with silicon compound before securing. Follow assembly procedure 1
to secure the transistor with the recovered insulation sheet sandwiched between the
transistor and rack plate. After securing is complete, coat the entire indentation with
silicon bond (do not leave any part uncoated).
3) Gather the thermistor and the harness of the transistor to the center rear and
bundle them there.
4) Fit the rack peltier heat insulator (P/N228-38662) to the square holes of the rack
plate. Have the thermistor and transistor gathered in 3) come out from the center.
5) Apply the thermal compound to the peltier and then mount it to a new rack plate.
Be sure to fully tighten the screw marked by an arrow mark first.
6) Semi-tighten all of the Peltier fixing screws and then fully tighten the screws marked by
arrows in following fig. (This is to prevent the thermal compound from rising up.)
Mount the cooling racks on both sides and set a vial in each #40. Push the vials down to
set them in place and make sure that the rise between rack and rack plate at the front end
of the rack plate is <…
0.8mm.
7) Stick heat insulation material to both ends of the rack plate in alignment with the rack
plate and front end.
Insulation
material
Insulation
material
8) Insert the long convex section of the pin recovered in assembly procedure 1 into the
rack plate hole. (Insert pin from the rear of the area shown)
9) Mount the rack plate ASSY to the main base. Be sure to pass the harness through the
convex hole at the rear of the rectangular hole of the main base when mounting. Place
Teflon washers on the three screws outside of the area marked by an arrow. Evenly tighten
the screws in the four corners.
10) In reverse order to item 13 of the dismantling procedure, pass the harness through the
notch of the main base. Also make sure that the sponge and Peltier fin marked by arrows
do not interfere with each other.
11) Secure the harness to the wiring fixing material removed in item 11 of the dismantling
procedure
12) Connect the rack plate power supply and thermistor connector in their original
positions.
13) Affix the packing seal around the rack plate so that the rack plate is covered.
14) Mount the rack sensor ASSY.
15) Mount the duct.
16) Insert the dehumidifying Peltier into the duct. And pass the harness between the base
and duct.
17) Connect the harness of the dehumidifying Peltier in its original position.
18) Secure the dehumidifying Peltier with the screws in the four corners and imbed sponge
caps in the screw holes.
19) Mount the fan on rear of sampler unit.
Take the connector section of the fan on the right side (when viewed from the rear of the
device) and trap it in the notch at the bottom left of the duct. Now pass the black covering
of the left-side fan’s connector through the notch.
20) Mount the sampler’s rear plate. Be sure to connect the fan.
21) Mount panel F (sampler’s transparent front plate), the two upper stays of the sampler
and the top plate of the Autosampler unit.
22) Mount the devices top plate and right side plate.
CONTENTS
11.Maintenace 2
Service SILHT
Maintenance Item Manual Instruction
Page (Page)
Piping(PEEK tube) P11-9 -
Sample loop P11-11 -
Removing the INJ port piping P11-10 -
11.2 Needle
<Repair part>
Contents of Needle assembly (P/N228-41024-92)
P/N Q’ty remarks
Part name
228-40817-04 1 length
needle 93mm
228-28430 1
ferule1.2F
228-36689 1
male nut needle
<Removing Needle>
For needle replacement, refer to Section 10.1.7 in the instruction manual.
This section describes supplementary contents.
“SIL” SCREEN
3) Pressing the [2] key and the [Enter] key in the ”MAINTENANCE” screen displays the
following “SYSTEM CHECK” screen.
Select “4.NEEDLE REPLACEMENT “,and the pop-up menu “REPLACE THE NEEDLE
f2 OK” appears.
4)Press the [f2] key. The needle move to the pre-detemined position and the following
pop-up menu appears.
Turn off the power.
Panel F ASSY
Loosen the panel screw with a Phillips head screwdriver , and pull out the panel,
slightly shifting it downward.
screw
screw
screw
7)Loosen and remove the male nut of the needle assembly with a spanner .
male nut
needle
<Consumable Part>
Needle seal (P/N228-33355-01)
P/N Q’ty Remarks
Part Name
228-33355-01 1 Material
Needle seal,PEEK 10XL :PEEK
Panel F ASSY
nut
the spanner14mm
Tip the spanner14mm is not provided.
Tip
When tightening the cap of new needle seal, tighten it manually up to hard to
move and then use the wrench with 90 degrees. Since the PEEK needle seal is
used, tightening angle of wrench is 90 degrees. If using Vespel needle seal
(LC2010), its angle is 45 degrees. So be careful that the tightening angle varies
depending on the material of needle seal.
“SIL” screen
Press the [4] key and the [enter] key in
the “SIL” screen.
Pressing th
The PEEK tubings are used for the piping (between INJ port and high-pressure valve;
and between high-pressure valve and column).
CAUTION
Since there are unusable solvent and pressure range for PEEK tube, refer to "Safety
Precautions" in the instruction manual.
Ferrule(SUS)
Note
Ferrule and male nut for PEEK tube are made of SUS.
2) Loosen the fixing screw of INJ port ASSY, and remove INJ port ASSY.
Drain pipe is attached to INJ port ASSY. Hold down the drain pipe fixed to INJ port
ASSY using the flat tip screwdriver and lift up INJ port ASSY and remove it.
Drain pipe
Note: When connecting the PEEK pipe using SUS ferrules and male nuts.
Tighten the male nut with a spanner referring to the amount described as follows:
Male nut with the outer diameter of 6 mm: Manually tighten the nut until it is hard
enough, then tighten for about 120 degrees further using a spanner.
Male nut with the outer diameter of 8 mm: Manually tighten the nut until it is hard
enough, then tighten for about 90 degrees further using a spanner.
11-10
11.MAINTENANCE2
F panel assy
Male nut
7)Loosen the screw of holder. Remove the sample loop from the groove on the holder
screw
When fixing the sample loop in the panel R, The needle must be located The upper end of the loop
the loop must pass above the sample at the inj port and the Z mount must satisfy the above
loop holder fixing screws as shown in Fig. 1. and the sample loop are The upper end of the loop
The position relation of the Teflon tube end satisfying the three-position relation.
surface and the holder end surface should -mentioned dimension
satisfy the above mentioned dimensions. Must be 15 mm or less against the Z mount top
surface shown in the figure.
The needle must not interfere with the bush when the needle is down.
3) Put the tube on the groove on the holder, then tighten the screw to fix the sample loop.
Mathc the end of the Teflon tube (outer diameter: 3mm) covering the sample loop with the
end of the holder ,then fix the sample loop so that its rightmost loop portion becomes
parallel to the panel.
the groove on the holder
rightmost loop portion
holder
Z mount
First turn of the loop
Black bush
5) Check that the distance from the pipe center of the loop’s first turn on the Z-mount side and the
Z-mount is properly maintained, as shown i.
3-7mm
6) Perform the Z-HOME operation and then turn off the power supply.
Pressing the [0] key with the “MENU” screen displayed shows “ANALYTICAL
SEQUENCE” screen. Pressing the [func] key twice and the [f1] key (Z-HOME) once
moves the Z-truck to home position.
38+/-3mm
First turn only
Sample loop shape checking
95-100mm
8) Measure the difference in height between the upper part of the loop’s first turn and the
Z-mount, to check if the requirement shown following fig is satisfied.
19+/-2mm
9) Move the loop on the side of panel R, so that the amount of the deviation between the
loop on panel R and that on the Z-mount falls within 5 mm.
11-17
Blank Paper
Chapter12 Trouble shooting
CONTENTS
12.Trouble shooting
TIP
Each message falls into one of the following three categories. The category is shown the “Type”
column.
Fatal : The unit stops operation.
Alarm : Press CE to clear the error message.
Warning : An error message is displayed, and alarm sounds . Unit operation continues .
ERR LEAK DETECT Alarm Cause: This message is displayed if detection value of
leak sensor is exceeded. Liquid leak may have
occurred, leak sensor may not be sufficiently
adjusted, or leak sensor may be defective.
Measures: Check liquid leakage.
Check cause of error message and take measure.
<Consumable parts>
Service SILHT
Parts Name Maintenance Item Manual Instruction
Page (Page)
Needle Seal Replacement P11-6 P132
Measuring pump Seal Replacement - P135
Measuring pump Plunger Replacement - P139
LPV rotor and stator Replacement - P141
HPV rotor and stator Replacement - P145
Needle Replacement P11-4 P149
Fuse Replacement - P151
Panel Replacement - P152
Service SILHT
Maintenance Item Manual Instruction
Page (Page)
Piping(PEEK tube) P11-9 -
Sample loop P11-10 -
Removing the INJ port piping P11-9 -
<Maintenance List>
Parts Name Maintenance Item Service
Manual
Page
Z mount Removing Z mount P10-5
Removal of Z feed screw P10-10
Z feed screw grease-up P10-11
Z shaft oiling P10-11
Z sensor replacement P10-12
XY ARM Removing XY ARM P10-15
Teaching Injection Port P8-7
Rack Vial P8-5
Belt Adjustment X belt adjustment P10-17
Y belt adjustment P10-18
Z belt adjustment P10-18
Valve Mount Removal of valve mount P10-19
Removing and mounting the high-pressure valve P10-20
Adjustment of the high-pressure valve P8-9
Removing and mounting the low-pressure valve P10-21
Adjustment of the low-pressure valve P8-9
Disassembling the measuring pump (plunger, P10-23
spring, and bearing)
Removing the measuring pump feed screw P10-24
12-5
Blank Paper
Chapter 13 Circuit Diagram
CONTENTS
13-2
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PB SCL-1A 1/7
13-3
13-4
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PB SCL-1A 2/7
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PB SCL-1A 3/7
13-5
13-6
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PB SCL-1A 4/7
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PB SCL-1A 5/7
13-7
13-8
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PB SCL-1A 6/7
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PB SCL-1A 7/7
13-9
13.Circuit Diagram
PB SCL-4 P/N228-34369-91
Fig. PB SCL-4
13-10
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PB SCL-4 1/1
13-11
13.Circuit Diagram
13-12
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PCB 10AVP CPU 1/2
13-13
13-14
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PCB 10AVP CPU 2/2
13.Circuit Diagram
13-17
13-18
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PCB SIL MAIN 3/10
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PCB SIL MAIN 4/10
13-19
13-20
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PCB SIL MAIN 5/10
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PCB SIL MAIN 6/10
13-21
13-22
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PCB SIL MAIN 7/10
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PCB SIL MAIN 8/10
13-23
13-24
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PCB SIL MAIN 9/10
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PCB SIL MAIN 10/10
13-25
13.Circuit Diagram
13-26
13.Circuit Diagram
13-31
13-32
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PCB LC2K-COOL ASSY 3/5
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PCB LC2K-COOL ASSY 4/5
13-33
13-34
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PCB LC2K-COOL ASSY 5/5
13.Circuit Diagram
13-35
13-36
SIL-HTA/HTC Service Manual
Fig. CIRCUIT DIAGRAM PCB PB-SCL2 1/1
13.Circuit Diagram
13-39
13-40
Fig. CIRCUIT DIAGRAM PCB PB-SCL5 1/1
CONTENTS
14-2
14.Parts Configuration and Parts Number
14-3
14.Parts Configuration and Parts Number
14-4
14.Parts Configuration and Parts Number
228-41001-31 SIL-HTC100V
129 228-16004-03 1
72 228-40915-91 1 1.6C
CABLE,DOOR-INJ1 130 228-18565 1
74 228-23013-01 1 MALE NUT PEEK
KEY 132 03738881-51 2
75 228-34553 1 SPACER,CL-305N
ROD SW 133 228-37280-91 1
76 037-50493-01 2 HARNESS INJ1-COOL
BUSHING,BS-1PK 134 228-38936-91 1
77 228-36809 2 CABLE PEL-POWER
TUBE JOINT 135 228-41212 2
78 228-39621 5 FILTER,K5BRC26-30-13MG
PYPE HOLDER 136 07473307-01 1
79 228-37285-91 1 BATTERY,CR2032
CABLE,SR-COOL 137 228-40923-91 1
80 075-08032-03 1 ROM,SIL-HT
CORE,EMI FILTER SFC-6 138 072-60334-01 1
107 037-38860-33 2 CLAMP,LFC-16N-0
SPACER C305 139 072-60324-02 2
110 016-43201 20MM CLIP MFC-3000
PE TUBE 7x1 140 018-17101-03 0.19
111 016-43201 330 SHIELD,FINGER 97-540-02
PE TUBE 7x1 MM 141 072-60306 1
112 035-61561-11 1 CABLE TIE,TY-33M
FITTING L 1252 145 072-60093-01 2
113 228-16000 10 GROMMET,BLANK 11
FERRULE 1.6F 146 228-39789-01 1
114 228-16001-03 6 KNURLED KNOB,M4X10
MALE NUT 1.6MN,W6 148 023-92054-01 1
115 228-16001 4 WASHER,SST WAVE #4
MALE NUT 1.6MN 149 228-41033 1
116 228-41030 1 DOOR SENSOR HOLDER
PYPE ,DGU-LV#4
117 228-41380-92 1
PYPE,P-HV#6
118 22841380-92 1
PEEK TUBE0.25ID
119 228-40526-01 3
SPECER,SAMPLER DOOR(1.0)
119 228-40526-02 2
SPACER SAMPLER DOOR(1.5)
119 228-40526-03 2
SPACER SAMPLER DOOR(2.0)
121 072-60333-02 3
CLIP,WIRE SD-10
123 016-31385-05 15MM
VINYL-T 1.5x0.4 BLUE
125 228-41076 1
INNNER PANEL SIDE
128 228-28413 1
CONNECTOR 1.6/3.0
14-7
14.Parts Configuration and Parts Number
14-11
14.Parts Configuration and Parts Number
14-13
14.Parts Configuration and Parts Number
14-14
14.Parts Configuration and Parts Number
14-16
14.Parts Configuration and Parts Number
14-17
14.Parts Configuration and Parts Number
14-18
14.Parts Configuration and Parts Number
14-19
14.Parts Configuration and Parts Number
14-20
14.Parts Configuration and Parts Number
14-21
14.Parts Configuration and Parts Number
14-22
14.Parts Configuration and Parts Number
14-23
14.Parts Configuration and Parts Number
14-24
14.Parts Configuration and Parts Number
14-25
14.Parts Configuration and Parts Number
14-26
14.Parts Configuration and Parts Number
14-27
14.Parts Configuration and Parts Number
14-28
14.Parts Configuration and Parts Number
14-29
14.Parts Configuration and Parts Number
40 228-38419 1
BEARING
41 228-38421 1
NUT
43 228-40166-91 1
MACHING PART
70 228-41075-05 1
LABEL
14-30
14.Parts Configuration and Parts Number
14-31
14.Parts Configuration and Parts Number
59 228-38419 1
BEARIN
60 228-38421 1
NUT
62 228-40165-92 1
MACHING PARTS
14-32
14.Parts Configuration and Parts Number
14-34
14.Parts Configuration and Parts Number
14-35
14.Parts Configuration and Parts Number
14-37
14.Parts Configuration and Parts Number
14-38
14.Parts Configuration and Parts Number
14-39
14.Parts Configuration and Parts Number
14-40
Consumable parts
P/N228-33355-01
NEEDLE SEAL PEEK
P/N228-35145
GFP,SEAL,42429
Fig.1 Fig.2
COATED NEEDLE
FERRULE 1.2F
P/N228-21984-01
SUCTION FILTER NEW Fig.6
Fig.5
MALE NUT NEEDLE
P/N228-41024-92 COATED NEEDEL ASSY
14-41
14.Parts Configuration and Parts Number
Repair Parts
<Electrical Parts>
P/N228-34369-91 Fig.10
PB SCL-4
P/N228-40900-91
PCB SIL MAIN Fig.9
P/N228-34505-91
PB SCL-5 Fig.11
P/N228-34366-91 Fig.12
PB SCL-3
14-42
14.Parts Configuration and Parts Number
P/N228-34363-92
PB SCL-2 Fig.13 Fig.14
P/N208-94869-03
Cable, LCD
P/N228-40905-91
PCB CONTRAST
P/N228-34565 Fig.16
Fig.15 Key Top 1, SCL
P/N228-34566
KEY TOP 2, SCL Fig.17
P/N078-12114-11 Fig.18
LCD,DMF50174NB-FW6
14-43
14.Parts Configuration and Parts Number
<Valve assy>
P/N228-35010-91
PLUNGER,SIL ASSY
Fig.19 Fig.20
P/N228-36445-01
MOTOR,HPV
P/N228-36917-01 Fig.22
P/N228-36446-01 STATOR 5PV
MOTOR LPV
Fig.21
Fig.23
Fig.24
P/N228-41020-91
P/N228-37506-91 HPV VALVE ASSY
LPV VALVE ASSY
14-44
14.Parts Configuration and Parts Number
P/N228-37542-91
FPC-V ASSY
P/N228-38654 Fig.26
Fig.25
STATOR PACKING,LPV
P/N228-37530-91 Fig.28
ASSY SEAL HOLDER
P/N228-39247-92
LEAK SENSOR ASSY , 700MM
Fig.27
P/N228-36444-01 Fig.29
MOTOR MEASURING PUMP
14-45
14.Parts Configuration and Parts Number
<XY truck>
P/N228-36376
P/N228-36362-02 SLIT
Fig.30 Fig.31
MOTOR X
Fig.33
P/N228-37355-91
P/N228-37320-91 PCB LC2K-XENC ASSY
PCB LC2K-INJ2 ASSY
Fig.32
P/N228-37365-91
PCB LC2K-XHP ASSY
(X PCB sensor)
Fig.34
Fig.35
P/N228-37509-91
SLIT HOLDER ASSY
14-46
14.Parts Configuration and Parts Number
<Z truck>
P/N228-36607-01
MOTOR Z Fig.37
Fig.36
P/N228-37541-91
FPC-Z ASSY
14-47
14.Parts Configuration and Parts Number
P/N228-36355-01 P/N228-37360-91
MOTOR Y PCB,RACK ASSY
Fig.39 Fig.40
P/N228-37508-92
RACK PLATE ASSY(C-TYPE)
P/N228-37508-91 Fig.42
RACK PLATE ASSY(A-TYPE)
Fig.41
P/N228-40528-91
Fig.43 HARNESS INJ1-INJ2 Fig.44
P/N228-37540-91
FPC-Y ASSY
14-48
14.Parts Configuration and Parts Number
<Parts for SIL-HTc>
Fig.45 Fig.46
P/N228-37389-91 P/N228-38718
AUX HEATER-COOL ASSY PELTIER UNIT
P/N228-38677-91 Fig.48
THERMISTOR SIL ASSY
P/N228-37537-91 Fig.47 (For Rack plate assy of SIL-HTc)
Dehumidifier , Peltier ASSY
P/N228-38677-92
P/N228-37330-91 THERMISTOR ,SIL ASSY
PCB LC2K-COOL ASSY Fig.49 (For dehumidifier of SIL-HTc) Fig.50
14-49
Blank Paper