GX 6000 Manual
GX 6000 Manual
GX 6000 Manual
Operator’s Manual
www.rkiinstruments.com
WARNING
Read and understand this instruction manual before
operating instrument. Improper use of the gas
monitor could result in bodily harm or death.
Chapter 2: Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Instrument Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Control Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Flashlight LED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Alarm LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Infrared Communications Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Buzzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Vibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Printed Circuit Boards (PCBs) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Flow Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Inlet Filter Holder. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Included Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Tapered Rubber Nozzle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Belt Clip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Rubber Boot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Wrist Strap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Sample Hose and Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Screen Protector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Other Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
VOC Zero Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
External Dilution Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
DIN Rail Mounting Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Chapter 3: Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Turning On the GX-6000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Performing a Demand Zero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Turning Off the GX-6000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
NOTE: RKI Instruments, Inc. recommends that you refer to ISA-RP12.13, Part II-1987
or an equivalent international recommended practice for guidance in the use of
combustible gas detection instruments.
Overview
This chapter briefly describes the GX-6000 gas monitor. This chapter also describes the
GX-6000 Operator’s Manual (this document). Table 1 at the end of this chapter lists the
specifications for the GX-6000.
C US
186718
Power • Three AA size alkaline batteries
Supply OR
• One lithium ion battery pack
Continuous • Alkaline Batteries: 8 Hours (Non Alarm Operation, Fully Charged)
Operating • Li-ion Battery Pack: 14 Hours (Non Alarm Operation, Fully Charged)
Hours
@ 25 °C
Case High-impact Plastic, RF Shielded, Dust and Weather Proof
Included • Belt clip
Accessories • Rubber boot
• Wrist strap
• Tapered nozzle
• Screen protector
• 3 foot hose and probe
• VOC zero filter (included for instruments with a low range PID sensor, a 10.0 eV/
benzene PID sensor, an 11.7 eV PID sensor or any PID/IR CO2 combination)
• Dehumidifier filter (included for instruments with an 11.7 eV PID sensor)
Overview
This chapter describes the GX-6000 instrument and accessories.
Instrument Description
The GX-6000 includes the case, LCD, control buttons, flashlight LED, alarm LEDs,
infrared communication port, buzzer, vibrator, printed circuit board, pump, flow chamber,
sensors, filters, inlet filter holder, and batteries.
Alarm LEDs
RKI GX-6000
Flow
Chamber
LCD DISP RESE T
A IR
LO CK
S H IF T
Control Buttons (P ANIC )
P OW ER / E N TER
Infrared Battery
Communications Pack
Port
Button Function(s)
Flashlight LED
A white LED is visible through a round, raised, frosted lens in the top of the case. This
LED can be used to provide extra light, if necessary. The RESET button turns this LED on
and off.
Alarm LEDs
Five sets of red alarm LEDs (light emitting diodes) border the LCD. The alarm LEDs alert
you to gas, low battery, and failure alarms.
Infrared Communications Port
An infrared (IR) communications port is located on the front of the case, below the
POWER/ENTER button. The data transmitted through the port is in standard IrDA
protocol. A computer’s infrared port or an IrDA/USB cable connected to a USB port can
be used to download data saved by the GX-6000 to a computer using the GX-6000 Data
Logger Management Program. See the GX-6000 Data Logger Management Program
operator’s manual for data logging and downloading instructions.
CAUTION: Sample hose lengths of more than 50 feet are not recommended for the
GX-6000 because of flow rate reduction.
Flow Chamber
The flow chamber is on the back of the GX-6000 and is held in place by four Phillips
screws. The flow chamber seals to the rubber sensor gasket which seals to the sensor faces
inside the GX-6000 and routes flow from the pump to the sensors and to the exhaust port
(on the top of the GX-6000 case).
Sensors
The GX-6000 uses five sensors to monitor combustible gas, oxygen (O2), carbon
monoxide (CO), hydrogen sulfide (H2S), and isobutylene (IBL) simultaneously. The
sensors are located inside the GX-6000 and are held in their sockets by the flow chamber.
The sensors use different detection principles, as described below.
Combustible Gas Sensor
The % LEL sensor detects combustible gas in the % LEL range. It uses a catalytic element
for detection. The reaction of gas with oxygen on the catalyst causes a change in the
resistance of the element which affects the current flowing through it. The current is
amplified by the GX-6000’s circuitry, converted to a measurement of combustible gas
concentration, and displayed on the LCD.
The standard calibration for the combustible gas sensor is to methane but the sensor will
still detect and respond to a variety of combustible gases.
Oxygen Sensor
The O2 sensor is a galvanic type of sensor. A membrane covers the cell and allows gas to
diffuse into the cell at a rate proportional to the partial pressure of oxygen. The oxygen
reacts in the cell and produces a voltage proportional to the concentration of oxygen. The
voltage is measured by the GX-6000’s circuitry, converted to a measurement of gas
concentration, and displayed on the LCD.
NOTE: Use of batteries or battery chargers not specified by RKI Instruments, Inc. will
compromise the CSA classification and may void the warranty. See pg.104.
CAUTION: Sample hose lengths of more than 50 feet are not recommended for the GX-
6000 because of flow rate reduction.
The probe includes a replaceable particle filter and hydrophobic filter disk that prevent
particulates and water from entering the GX-6000’s flow system. See “Replacing the
Probe’s Particle Filter and Hydrophobic Filter Disk” on page 112 for instructions to
replace the particle filter and hydrophobic filter disk.
Other Accessories
Several other accessories are available for the GX-6000. This section describes the VOC
zero filter, the dilution fitting, and the DIN rail. Detailed instructions regarding the use of
the dilution fitting are included in other parts of this manual. Data logging accessories are
briefly described in “Data Logging” on page 45.
VOC Zero Filter
A VOC zero filter is included as standard with GX-6000s that include:
• Low range PID sensor (PID-001L)
• 10.0 eV/benzene PID sensor (PID-003L)
• 11.7 eV PID sensor (PID-004)
• Any PID sensor and any CO2 sensor
The VOC zero filter scrubs out low levels of VOC gases using charcoal. Use the VOC
zero filter when:
• Performing an air adjust on a PID sensor in an area that may have a low-level VOC
background.
• Performing an air adjust on a PID/CO2 instrument (in combination with a CO2
scrubber).
Plug Plug
To GX-6000
1.38"
DIN Rail
DIN Rail
Clamp
Overview
This chapter explains how to use the GX-6000 to perform confined space entry
monitoring or general area monitoring in Normal Mode.
The GX-6000 can also operate in Leak Check Mode and Bar Hole Mode. See pg.170 for
Leak Check Mode instructions. See pg.180 for Bar Hole Mode instructions.
Start Up
This section explains how to start up the GX-6000, get it ready for operation, and turn it
off.
NOTE: The screens illustrated in this section are for a standard 4-gas + high range PID
unit. The screens displayed by your GX-6000 may be slightly different.
5 SEC
YES:ENTER
NO:DISP
CAL EXPRD set to CONFIRM CAL EXPRD set to CAL EXPRD set to
TO USE (factory setting) CANNOT USE NO EFFECT
LCD
4:07 4:07 4:07
CAL DATE PAST CAL DATE PAST CAL DATE PAST
Sound Buzzer sounds double pulsing tone Buzzer sounds double pulsing None
tone
Action • Option A, Perform calibration: The GX-6000 cannot be used until • Option A, Perform calibration:
Press and release POWER/ a successful calibration is per- To perform a calibration, press
ENTER to perform a calibra- formed. Press and release and release POWER/ENTER.
tion. The GX-6000 takes you POWER/ENTER to perform a cal- The instrument takes you
straight to the 4-gas Gas Select ibration. The instrument takes you straight to the 4-gas Gas Select
Screen in Calibration Mode’s straight to the 4-gas Gas Select Screen in Calibration Mode’s
AUTO CAL\GAS SELECT Screen in Calibration Mode’s AUTO CAL\GAS SELECT
menu. See pg.68 for calibration AUTO CAL\GAS SELECT menu. menu.
instructions. Even if the User password func- • Option B, Bypass message: To
If the calibration is successful, tion has been turned on, no pass- continue without performing a
the screen above will not appear word is required to perform a calibration, wait a few seconds
again until the unit is due for calibration. for the instrument to continue
calibration. If the calibration is with its startup sequence.
not successful, the screen above If you don’t press POWER/
will again appear in the startup ENTER, the instrument automati-
sequence. cally goes to the 4-gas Gas Select
• Option B, Bypass message: To Screen after 6 seconds. See pg.68
continue without performing a for calibration instructions.
calibration, press and release
RESET. If the calibration is successful, the
screen above will not appear again
until the unit is due for calibration.
If the calibration is not successful,
the screen above will again appear
in the startup sequence.
BUMP EXPRD set to CON- BUMP EXPRD set to BUMP EXPRD set to
FIRM TO USE (factory setting) CANNOT USE NO EFFECT
LCD
4:07 4:07 4:07
BUMP DATE PAST BUMP DATE PAST BUMP DATE PAST
Sound Buzzer sounds double pulsing Buzzer sounds double pulsing tone None
tone
Action • Option A, Perform bump test: The GX-6000 cannot be used until • Option A, Perform bump test:
Press and release POWER/ a successful bump test has been To perform a bump test, press
ENTER to perform a bump performed. Press and release and release POWER/ENTER.
test. The instrument takes you POWER/ENTER to perform a The instrument takes you
straight to the 4-gas Gas Select bump test. The instrument takes straight to the 4-gas Gas Select
Screen in Calibration Mode’s you straight to the 4-gas Gas Select Screen in Calibration
BUMP TEST menu. See pg.89 Screen in Calibration Mode’s Mode’s BUMP menu.
for bump test instructions. BUMP TEST menu. If you don’t • Option B, Bypass message: To
If the bump test is successful, press POWER/ENTER, the instru- continue without performing a
the screen above will not ment automatically goes to the 4- bump test, wait a few seconds
appear again until the unit is gas Gas Select Screen after 6 sec- for the instrument to continue
due for bump testing. If the onds. See pg.89 for bump test with its startup sequence.
bump test is not successful, the instructions.
screen above will again appear
in the startup sequence. If the bump test is successful, the
• Option B, Bypass message: To screen above will not appear again
continue without performing a until the unit is due for bump test-
bump test, press and release ing. If the bump test is not success-
RESET. ful, the screen above will again
appear in the startup sequence.
4/ 1/15
DD/MM/YYYY
4:07
VOLTAGE
4.5V
BATTERY TYPE
ALKALINE
LATCHING
8. The Active Gases Screen appears for a few seconds indicating which channels are
active and their target gas.
GAS GAS
SMART 1
CH4
%LEL
O2% Isobutylene
H2Sppm CO
ppm
P 10.6 eV
VOC
ppm
KIMBERLY TANK 5
SENSOR
CH4
FAIL
%LEL
If you wish to continue, press and release the RESET button to acknowledge the
failure. The gas reading for the failed sensor will be replaced by “- - - -”. Replace the
failed sensor as soon as possible.
12 . The GX-6000 is now monitoring for gas in Measuring Mode. The Measuring Mode
Screen appears displaying the current gas reading for each target gas.
If you have 2 PID sensors installed, they might both be displayed as “VOC”. The
channel that appears first is based on the following sensor priority.
4:07
CH4 O2
0
%LEL
20.9%
H2S CO
0.0
ppm ppm
0
VOC
0
ppm
NOTE: If you have both a PID and a CO2 sensor installed, you will need to follow the
directions on pg.209 instead of following the directions shown below.
a. Remove the plug from each end of the VOC zero filter.
b. Attach the VOC zero filter to the inlet fitting or probe. The filter does not have a
preferred flow direction.
c. Let the instrument draw through the VOC zero filter for 1 minute before
continuing.
4. Press and hold the ▲AIR button. The LCD prompts you to continue holding the
▲AIR button and the buzzer will pulse while you hold the button.
5. Continue to hold the ▲AIR button until the LCD prompts you to release it. The GX-
6000 will set the fresh air reading for all channels. Start up is complete and the unit is
now ready for monitoring.
6. If the VOC zero filter was used, remove the filter from the GX-6000’s inlet fitting or
probe. Reinstall the plugs on each end of the filter.
Turning Off the GX-6000
1. Press and hold the POWER/ENTER button.
2. TURN OFF will appear on the display and the buzzer will pulse for about five
seconds.
3. Release the button when TURN OFF disappears from the display.
NOTE: If the particle filter or hydrophobic filter become dirty or clogged, replace them.
If water enters the probe, dry out or replace the particle filter (if installed) and
shake any water out of the probe and off of the hydrophobic filter. If you notice
that water has entered the flow system through the probe, replace the probe’s
hydrophobic filter. See pg.112 for instructions to replace the particle filter and
the hydrophobic filter.
3. Wait 10 - 15 seconds and observe the display for gas readings. If a reading is
observed, allow the reading to stabilize to determine the gas concentrations present.
NOTE: Response time increases with the length of the sample hose. Long sample hoses
will require more time to show a response at the GX-6000. The maximum
sample hose length recommended for the GX-6000 is 50 feet. Consult RKI
Instruments, Inc. for longer sample hose lengths.
CAUTION: Sample hose lengths of more than 50 feet are not recommended for the GX-
6000 because of flow rate reduction and increased response time. Consult
RKI Instruments, Inc. for hose lengths longer than 50 feet.
The chart below illustrates how response time is affected by the sample hose length.
Table 4: GX-6000 Response Time vs. Sample Hose Length
• The GX-6000 provides the catalytic combustible sensor with some protection against
exposure to high levels of combustible gas which can damage the sensor. It does this
by turning off the combustible sensor power temporarily when it determines that an
over scale (more than 100 %LEL) concentration of combustible gas is present that
may damage the sensor. Nevertheless, concentrations of combustible gas of more than
100 %LEL can still affect the zero level or calibration of the combustible sensor if the
concentration is high enough.
CAUTION: Any rapid increase in the combustible gas reading on the catalytic
combustible channel followed by a declining or erratic reading may
indicate a gas concentration above the LEL which may be hazardous.
NOTE: If you have changed the factory-set channel assignments using the
CONFIGURE CH menu item in User Mode, the order of displayed sensors
may be different and you will need to remember how you set your instrument
up.
Oxygen Detection
Exposing the galvanic oxygen sensor to halogen gas or sulfides may shorten the sensor’s
life or cause malfunctions or inaccurate gas readings. Minimize the sensor’s exposure to
these gases as much as possible. If exposure occurs, allow the instrument to draw fresh air
and confirm that the readings return to fresh air values.
STATION ID
NUMBER 001
TANK 5
The first screen displays what snap log ID will be given to this particular set of data
and that you should press the POWER ENTER button to save a set of snap log data.
The snap log ID number increases sequentially with each set of snap log data taken.
The second screen displays what Station ID will be associated with this snap log and
that you should press the DISP/LOCK button to exit Snap Log Mode without saving
a set of snap log data. The third screen displays the year, month, day, and time of the
snap log. The fourth screen displays the current gas readings.
NOTE: If the GX-6000 detects an alarm condition while in Snap Log Mode, it will
automatically exit Snap Log Mode and return to Measuring Mode. You may
then reenter Snap Log Mode and take snap logs while the instrument is still in
alarm.
VALVE 3
3. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to scroll to the desired Station ID,
then press and release the POWER/ENTER button to return to the Snap Logging
Screen sequence.
To return to the Snap Logging Screen sequence without changing the Station ID,
press and release the DISP/LOCK button.
4. To take a snap log of the current gas readings, press and release the POWER/ENTER
button. The unit will display SAVED along the bottom of the screen before returning
to the Snap Logging Screen sequence.
To exit Snap Log Mode without taking a snap log or when you are finished recording
snap logs, press and release the DISP/LOCK button. The unit will immediately
return to the Measuring Mode Screen.
The data recorded in Snap Log Mode can be viewed in Display Mode. See pg.60 for
more information.
NOTE: False alarms may be caused by radio frequency (RF) or electromagnetic (EMI)
interference. Keep the GX-6000 away from RF and EMI sources such as radio
transmitters or large motors.
Alarm Indications
The GX-6000 buzzer will sound an alarm, the LEDs will flash, and the vibrator will pulse
when any sort of alarm condition or failure is encountered.
NOTE: If an alarm condition occurs while you are in Display Mode, the GX-6000 will
automatically bring up the alarm screen instead.
The table below summarizes the types of alarms produced by the GX-6000 and their
indications.
None of the Man Down alarm indications will happen if MAN DOWN is set to OFF in
Maintenance Mode (factory setting).
Table 6: Alarm Types and Indications
Man Down Warning 2 • Alarm LED arrays flash twice per Single pulsing tone twice
The WARNING 2 TIME second per second
defined in Maintenance Mode
has passed since the instrument
detected movement.
Man Down Alarm • MAN DOWN flashes at top of • High-low tone sounding
The ALARM TIME defined in screen in place of time twice per second
Maintenance Mode has passed • Alarm LED arrays flash in circle • Vibrator pulses twice
since the instrument detected sequence twice per second per second
movement.
Panic • Screen unaffected for 5 seconds • Single pulsing tone
User presses and holds the before alarm starts twice per second for 5
SHIFT ▼ (PANIC) button. • MAN DOWN flashes at top of seconds before alarm
screen in place of time starts
• Alarm LED arrays flash in circle • High-low tone sounding
sequence twice per second twice per second
• Backlight turns on • Vibrator pulses twice
per second
Responding to Alarms
This section describes response to gas, over range, battery, sensor failure, clock failure,
system failure, man down, and panic alarms.
Responding to Gas Alarms
1 . Determine which gas alarm has been activated.
2 . Follow your established procedure for an increasing gas condition or a decreasing
oxygen condition.
3 . Reset or silence the alarm as necessary or allowed. Table 7 below summarizes
resetting and silencing alarms for all ALARM LATCHING and ALARM
SILENCE combinations that are possible. See pg.151 and pg.151 for further
descriptions of these parameters.
PRUDENCE: Des lectures élevées hors échelle peuvent indiquer une concentration
explosive.
NOTE: If your instrument has both a low range PID sensor and a high range PID sensor
installed and if you are monitoring for gases in the higher range, the low range
PID channel may be in over range alarm while monitoring, depending on the
level of VOCs present.
3. Reset the alarm using the RESET button once the alarm condition has cleared.
The GX-6000 is fully functional during a low battery warning. However, only a limited
amount of operating time remains, approximately 1 - 2 hours. The amount of time depends
on how often the LCD backlight is used and how often the unit is responding to alarm
conditions. Recharge the Li-ion battery pack or replace the alkaline batteries as soon as
possible as described in “Replacing or Recharging the Batteries” on page 104.
NOTE: Alarms and the LCD back light consume battery power and reduce the amount
of operating time remaining.
1. Press and release the RESET button to continue into Measuring Mode.
CAUTION: There will be no datalogging function if you operate the instrument after a
clock failure.
022
3. If the error code is anything but 022 as shown above, the instrument cannot be used.
Contact RKI Instruments, Inc. as soon as possible.
If the error code is 022, you may press and release the RESET button to continue into
Measuring Mode if the instrument must be used temporarily.
CAUTION: There will be no datalogging function if you operate the instrument after a
022 system failure. Contact RKI Instruments, Inc. as soon as possible.
Data Logging
NOTE: The GX-6000 only logs data while in Normal Mode. If the GX-6000 is used in
Leak Check Mode or Bar Hole Mode, no downloadable data will be logged
while it is in either of these two modes.
The GX-6000 features the ability to log data to its internal memory and download it to a
computer via the infrared communications port on the front of the unit. It logs gas readings
in Measuring Mode, alarm data, and calibration data.
NOTE: The data logging function is separate from the snap logging function. For a
description of the snap logging function, see “Snap Log Mode” on page 37.
To utilize the GX-6000’s downloading capability, you will need the GX-6000 Data Logger
Management Program and a computer with an infrared port or a USB port that runs one of
the following operating systems: Windows 7, Windows 8, or Windows 10. If your
computer has an infrared port, then no additional accessories are needed to download data
from the GX-6000. If your computer does not have an infrared port but does have a USB
port, a USB/IrDA adapter cable can be used to download data from the GX-6000 using the
USB port. The GX-6000 Data Logger Management Program is available at
www.rkiinstruments.com/gx6000. The USB/IrDA adapter cable is available from RKI
Instruments, Inc.
The data logging capacity depends on how often the GX-6000 stores data, how many
channels are active, and how often the GX-6000 is turned on and off. The table below
illustrates how much data logging time is available for the various interval times. It
assumes that the unit is a standard four sensors plus PID unit, is only turned on once, and
there are no alarm occurrences. See “Updating the Data Log Interval Setting (LOG
INTERVAL)” on page 159 for instructions on setting the data logging interval time.
Interval
Data Logging Time
Time
10 seconds 10 hours
20 seconds 20 hours
30 seconds 30 hours
1 minute 60 hours
3 minutes 180 hours
5 minutes 300 hours
10 minutes 600 hours
For a complete description of the Data Logger Management Program and procedures for
downloading data to a computer, see the GX-6000 Data Logger Management Program
Operator’s Manual.
This section describes using the GX-6000 in Display Mode. See Table 10 below for a list
of Display Mode’s menu items, a short description of each item, and the page number for
further description.
Table 10: Display Mode Items
SELECT PID LIST (pg.48) Define the PID channel’s gas name
PEAK (pg.50) View and clear peak readings
STEL (pg.51) View STEL readings (H2S, CO, and high range PID only)
TWA (pg.51) View TWA readings (H2S, CO, and high range PID only)
NOTE: With the exception of the Pump Off Screen, each screen displays for 20
seconds. If you do not press a button within 20 seconds, the GX-6000
automatically returns to Measuring Mode. If the Pump Off Screen is used to
turn the instrument’s pump off, the Pump Off Screen will display until you turn
the pump back on.
10.6 eV
CHANGE :ENTER
KEEP :DISP
EXIT :RESET
1. With the PID Gas Name Screen displayed, press and release the POWER/ENTER
button. The top level of the PID Gas Name List will be displayed and the cursor will
be next to the letter corresponding to the first letter of the current setting. In the
example below, the PID channel is configured for isobutylene so the cursor is next to
“I”. Each letter has a list of gas names behind it.
At the top of the letter list, there is a User List item and a Recent List item. The User
List can hold 30 gas names. See pg.247 for instructions to edit the User List. The
Recent List can hold up to 8 gas names and is made up of the 8 most recently used
gas names. The Recent List is not user editable.
I s o b ut y l e n e
Peak Screen
The peak screen displays the highest (lowest for oxygen) concentrations detected since the
GX-6000 was turned on. Peak readings are stored in the GX-6000’s memory until a higher
level is detected (lower for oxygen), the peak reading is cleared, or the GX-6000 is turned
off.
The lunch break feature enables the GX-6000 to save peak readings when it is turned off
so it can continue them when it is turned on again. See “Turning On the GX-6000” on
page 25.
PEAK
CH4 O2
10
%LEL
15.0%
H2S CO
13.0
ppm
200
ppm
VOC
2000
ppm
PEAK PEAK
PEAK CLEAR PEAK CLEAR
HOLD RELEASE
STEL Screen
The STEL Screen displays the short term exposure limit (STEL) readings for H2S, CO,
and high range VOC only. The STEL reading is the average reading over the last 15
minutes.
STEL
H2S CO
2.0
ppm
10
ppm
VOC
10.0
ppm
TWA Screen
The TWA Screen displays the time weighted average (TWA) readings for H2S, CO, and
high range VOC only.
TWA
H2S CO
0.5
ppm
5
ppm
VOC
5.0
ppm
The TWA reading is the average reading over the last 8 hours. If 8 hours have not elapsed
since the last time the TWA reading was cleared, the average is still calculated over 8
hours. The missing time is assigned a 0 value for readings. If LUNCH BREAK is set to
OFF, the TWA is cleared when the GX-6000 is turned off.
If LUNCH BREAK is set to ON, the GX-6000 will remember TWA readings when it is
turned off so it can continue them when it is turned on again. See pg.25.
YES :ENTER
NO :DISP
EXIT :RESET
100.0
ppm
500
ppm
VOC
TWA 6000 AL1
ppm CH4 O2
10
%LEL
19.5%
H2S CO H2S CO
5.0
ppm
200
ppm
5.0
ppm
25
ppm
SHIFT ▼ (PANIC)
VOC VOC
40.0
ppm
400.0
ppm
STEL AL2
CH4 O2
▲ AIR 50
%LEL
23.5% ▲ AIR
H2S CO H2S CO
1.0
ppm
25
ppm ▲ AIR
30.0
ppm
50
ppm
VOC VOC
60.0
ppm
1000
ppm
2. Pressing and releasing the POWER/ENTER button while in a settings screen will
cause the instrument to simulate that condition. The buzzer, LED’s, and vibration
will activate. Press and release the POWER/ENTER button again to stop the
simulated alarm.
3. Press and release the DISP/LOCK button to return to the View Alarm Settings
Screen.
0:02
If LUNCH BREAK is set to ON, the time in operation will only be reset if you do not
choose to resume the peak and TWA measurements when the GX-6000 is turned on in the
Lunch Break Screen described in Step 3 on page 26. If you choose to resume the peak and
TWA measurements during startup, the GX-6000 will include the time in operation when
the unit was last turned off in the current time in operation.
5/12/2015
MM/DD/YYYY
1:57
4.1V
BATTERY TYPE
ALKALINE
129 HOUR
CAUTION: Once you clear the data logger, you cannot retrieve any data previously
stored in the data logger.
The LOG CLEAR screen asks if you want to clear the data logger memory.
DISP
LOG CLEAR
YES :ENTER
NO :DISP
EXIT :RESET
To continue through Display Mode without clearing the log data, press and release the
DISP/LOCK button.
YES : ENTER
NO : DISP
CONFIRM ?
2. If you do not want to clear the data logger memory, press and release the DISP/
LOCK button. The unit will return to the LOG CLEAR screen in Display Mode.
3. If you want to clear the data logger memory, press and release the POWER/ENTER
button. The unit will display the following screen as it clears the data.
DISP
LOG CLEAR
CLEAR
4. When the instrument has cleared the data, it will return to the LOG CLEAR screen
in Display Mode.
WARNING: The GX-6000 is not a gas monitoring device while the pump is off.
NOTE: The item appears in Display Mode only if the PUMP OFF DISP menu item in
Maintenance Mode is turned ON. The factory setting for PUMP OFF DISP is
OFF.
This feature can be used to conserve battery life if there is a significant period of time until
the next gas monitoring task, but you do not want to wait for the warm-up sequence to
complete if you turn the instrument off and on again.
DISP
PUMP OFF
YES :ENTER
NO :DISP
EXIT :RESET
1. In this screen, press and release the POWER/ENTER button to turn the pump off.
The fan symbol in the upper right corner will disappear. Once the pump is off, you
cannot exit this screen and return to Measuring Mode or continue in Display Mode
until you turn the pump back on.
1:44
PUMP OFF
OFF
2. To turn the pump back on and continue in Display Mode or return to Measuring
Mode, press and release the DISP/LOCK button.
KIMBERLY
The user ID provides a way to identify the GX-6000 user during a data logging session.
See the GX-6000 Data Logger Management Program Operator’s Manual for a detailed
description of data logging and the user ID.
The user ID list cannot be edited using the GX-6000 user interface. The GX-6000
Maintenance Data Loader Program is required to define or change user IDs in the user ID
list. For a detailed description of editing the list of user IDs stored in the GX-6000, see the
GX-6000 Maintenance Data Loader Program Operator’s Manual.
To select a different user ID:
1 . With the Select User ID Screen displayed, press and release the POWER/ENTER
button. The current user ID will be displayed and it will be flashing.
DISP
USER ID
KIMBERLY
2. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to scroll through the available user
IDs. The list will wrap around if you get to the top or bottom. Any of the user IDs in
the list that have not been changed from the factory setting will be shown as
USER_ID_XXX where the “XXX” is a number from 001 to 128.
NOTE: To exit the selection screen without saving a change, press and release the
DISP/LOCK button. You will return to the Select User ID screen without
saving the user ID change.
4. The unit will save the selected user ID as the current one and will return you to the
Select User ID Screen.
TANK 5
The station ID provides a way to identify a location where monitoring was done during a
data logging session. See the GX-6000 Data Logger Management Program Operator’s
Manual for a detailed description of data logging and the station ID.
The station ID list cannot be edited using the GX-6000 user interface. The GX-6000
Maintenance Data Loader Program is required to define or change station IDs in the
station ID list. For a detailed description of editing the list of station IDs stored in the
GX-6000, see the GX-6000 Maintenance Data Loader Program Operator’s Manual.
TANK 5
2. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to scroll through the available station
IDs. The list will wrap around if you get to the top or bottom. Any of the station IDs
in the list that have not been changed from the factory setting will be shown as
STATION_ID_XXX where the “XXX” is a number from 001 to 128.
3. When the desired station ID is displayed, press and release the POWER ENTER
button.
NOTE: To exit the selection screen without saving a change, press and release the
DISP/LOCK button. You will return to the Select Station ID screen without
saving the station ID change.
4. The unit will save the selected station ID as the current one and return you to the
Select Station ID Screen.
YES :ENTER
NO :DISP
EXIT :RESET
5/20/2015
NO DATA MM/DD/YYYY
3:40
NUMBER 001
TANK 5
2. If snap logs have been taken, the screen indicates the year, month, day, and time that
the most recent snap log was taken.
The number near the bottom of the screen indicates the snap log ID number. The first
snap log that is taken is given an ID of 001. The next snap log ID is 002. The ID
number increases sequentially with each set of snap log data.
The last line of the screen indicates the Station ID that was used for the snap log.
3. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to scroll through different snap
log IDs.
4. To view the data in a snap log ID, press and release the POWER/ENTER button.
REC DATA
CH4 O2
0
%LEL
20.9%
H2S CO
0.0
ppm
0
ppm
VOC
0
ppm
The gas readings that were taken during the snap log are displayed. You can use the
▲AIR and SHIFT ▼ (PANIC) buttons to scroll through the different snap logs. The
gas readings will change as you view different snap log data but the snap log ID is
not visible from this screen. To scroll through the data by viewing the snap log ID
number, press and release POWER/ENTER.
5. To return to the Snap Logging Screen, press and release the DISP/LOCK button.
YES :ENTER
NO :DISP
EXIT :RESET
1. With the Peak Bar Screen displayed, press and release the POWER/ENTER button.
The current setting will be displayed.
2. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to display the desired setting.
3. Press and release the POWER/ENTER button to save the setting and return to the
Peak Bar Screen.
To return to the Peak Bar Screen without saving any changes, press and release the
DISP/LOCK button.
YES :ENTER
NO :DISP
EXIT :RESET
Option Description
1. With the Gas Display Screen showing, press and release the POWER/ENTER
button. The available choices will appear.
DISP
GAS DISP
DISPLAY ALL
SCROLL AUTO
SCROLL MANUAL
NOTE: Even if INVERT SELECT is set to ON, the screens in Calibration Mode, User
Mode, and Maintenance Mode will not flip if the instrument is turned upside
down.
If INVERT SELECT is set to OFF, the instrument’s LCD will not flip and can only be
read when the instrument is right side up. A LOCK symbol will appear at the top of the
screen to indicate that the LCD’s position is fixed.
DISP
INVERT SELECT
YES :ENTER
NO :DISP
EXIT :RESET
1. With the Gas Display Screen showing, press and release the POWER/ENTER
button. The current setting will be displayed.
2. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to display the desired setting.
3. Press and release the POWER/ENTER button to return to the LCD Flip Screen.
To return to the LCD Flip Screen without saving any changes, press and release the
DISP/LOCK button.
PROTECTION
1. With the CAT (LEL) SENSOR PROTECTION screen displayed, press and release
the POWER/ENTER button. The current setting will be displayed along with a
warning about no %LEL alarms if the catalytic LEL sensor is turned off.
DISP
CAT (LEL) SENSOR
***WARNING***
NO LEL ALARMS IF
CAT (LEL) IS
DISABLED
ENABLED
2. Use the ▲AIR or SHIFT ▼ (PANIC) button to display the desired setting.
NOTE: If the catalytic LEL sensor is disabled and if the instrument is then turned off
and back on, the catalytic LEL sensor will be enabled again.
3. Press and release the POWER/ENTER button to return to the CAT (LEL) SENSOR
PROTECTION screen in Display Mode.
YES :ENTER
NO :DISP
EXIT :RESET
1. With the LCD Background Flip Screen showing, press and release the POWER/
ENTER button. The current setting will be displayed.
2. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to display the desired setting.
3. Press and release the POWER/ENTER button to return to the LCD Background Flip
Screen.
To return to the LCD Background Flip Screen without saving any changes, press and
release the DISP/LOCK button.
Language Screen
If the instrument’s language is changed to something other than English, the Language
Screen will appear in Display Mode. The Language Screen will not appear if the
instrument’s language is set to English. The Language Screen allows the user to change
the instrument’s language back to English.
DISP
LANGUAGE
CHANGE
TO ENGLISH
YES :ENTER
NO :DISP
EXIT :RESET
CHANGE
TO ENGLISH
CONFIRM ?
2. Press and release the POWER/ENTER button again to confirm the change.
To return to the Language Screen without changing the instrument’s language to
English, press and release the DISP/LOCK button.
Overview
This section describes the GX-6000 in Calibration Mode. In Calibration Mode, you can
move through a menu of screens to do the following:
• Perform a fresh air (zero) adjustment
• Perform a span adjustment on all channels simultaneously using auto calibration
• Perform a span adjustment on one channel at a time using single calibration
• Perform a bump test
NOTE: You can set up the GX-6000 to alert you during the startup sequence when
calibration is due. See pg.143.
If the combustible channel passes the above response test and does not require calibration,
the unit should still be calibrated periodically. The optimum frequency of calibration
depends heavily on how the GX-6000 is used. For example, instruments used daily may
need to be calibrated weekly or monthly, while instruments that are used only a few times
a year may need to be calibrated before each use. Typical calibration frequencies range
from monthly to quarterly. Make sure to perform the combustible channel response test as
described above and make sure to develop a calibration schedule tailored to your
application that takes this test and required calibration resulting from this test into account.
CAUTION: When using auto calibration, although the GX-6000 can be calibrated with
an oxygen concentration of up to 19.5%, RKI Instruments, Inc.
recommends that the multi-gas cylinder have an oxygen concentration in
the range of 10% - 16% oxygen.
NOTE: RKI Instruments, Inc. recommends that you dedicate a regulator for use with
chlorine (Cl2) gas and that you do not use that dedicated regulator for any other
gases, particularly hydrogen sulfide (H2S).
• Non-absorbent tubing
WARNING: If you are using a calibration kit that includes a gas bag and a fixed
flow regulator or dispensing valve, do not apply gas directly to the
GX-6000 with the regulator or dispensing valve or damage to the pump
will result.
To calibrate the %LEL, oxygen, CO, H2S, and PID sensors at the same time,
automatically, with no need for a zero-oxygen source, you can use the auto calibration
feature with either a 5-gas cylinder or a 4-gas cylinder and a PID cylinder. This chapter
includes instructions for both scenarios along with instructions for calibrating one channel
at a time using single calibration.
4. The Calibration Mode Screen displays with the cursor next to AIR CAL.
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
NORMAL MODE
NOTE: The following screens illustrate a five-gas GX-6000 for detection of CH4
(%LEL using catalytic sensor), oxygen, H2S, CO, and VOCs (high range PID
sensor). Your GX-6000 may display slightly different screens.
NOTE: You will only be prompted to enter a password if the PASSWORD menu item
in User Mode is set to ON. See pg.100 for instructions to update the
PASSWORD setting and to choose a password.
4. The Calibration Mode Screen displays with the cursor next to AIR CAL.
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
NORMAL MODE
5. Press and release the POWER/ENTER button. The Fresh Air Reading screen will
display.
AIR
CH4 O2
0
%LEL
20.9%
H2S CO
0.0
ppm
0
ppm
VOC
0
ppm
6. If you suspect any low-level VOC background in the area, you will need to use the
VOC zero filter.
NOTE: If you have both a PID and a CO2 sensor installed, you will need to follow the
directions on pg.209 instead of following the directions shown below.
a. Remove the plug from each end of the VOC zero filter.
b. Attach the VOC zero filter to the inlet fitting or probe. The filter does not have a
preferred flow direction.
c. Let the instrument draw through the VOC zero filter for 1 minute before
continuing.
9. Release the ▲AIR button. The fresh air adjustment will finish, the fresh air readings
will be displayed momentarily, and then the instrument will return to the Calibration
Mode Screen.
AIR AIR
AIR CAL AIR CAL
END
ZERO ADJUSTED
AIR CAL
CH4 O2 CAL MODE
0
%LEL
20.9% AIR CAL
AUTO CAL
H2S CO SINGLE CAL
0.0
ppm
0
ppm
NORMAL MODE
VOC
0
ppm
10 . If the VOC zero filter was used, remove the filter from the GX-6000’s inlet fitting or
probe. Reinstall the plugs on each end of the filter.
5. Press and release the POWER/ENTER button to display the Auto Cal Menu.
AUTO
AUTO CAL
CONCENTRATION
GAS SELECT
ESCAPE
The gas concentrations displayed in the Calibration Gas Values Screen must match
the gas concentrations listed on the 5-gas cylinder or the 4-gas calibration cylinder
and the PID cylinder.
If all concentrations match, go to Step 12.
If one or more concentrations do not match, continue with Step 7.
If you do not want to continue with the concentration adjustment, press and release
the DISP/LOCK button to return to the Auto Cal Menu.
NOTE: The RKI 4-gas cylinder typically contains 12% O2 by volume. When using the
auto calibration method, be sure to set the “O2” auto calibration value to agree
with the concentration listed on the cylinder’s label, not zero.
7. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to place the cursor next to the channel
whose gas value you want to change.
8. Press and release the POWER/ENTER button. The auto cal value will begin to flash.
9. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to adjust the calibration gas setting to
the desired value.
NOTE: The calibration gas value cannot be set lower than the low alarm setting. If the
calibration gas value listed on the calibration cylinder is lower than the current
low alarm setting, enter Maintenance Mode and change the low alarm setting.
See pg.150 for instructions. If you need to change the alarm point setting only
to perform a calibration, make sure that you change the alarm point setting back
to its original value once the calibration has been performed.
10 . Press and release the POWER/ENTER button to save the change. The calibration gas
value stops flashing.
11 . Repeat Step 7 through Step 10 for any other channels that need to be changed.
13 . Press and release the POWER/ENTER button. The instrument returns to the Auto
Cal Menu.
14 . Use the ▲AIR and SHIFT ▼ (PANIC) buttons to move the cursor next to GAS
SELECT.
AUTO
AUTO CAL
CONCENTRATION
GAS SELECT
ESCAPE
AUTO AUTO
GAS SELECT GAS SELECT
CH4 50 %LEL
O2 12.0 % ▲ AIR or
H2S 25.0 ppm SHIFT ▼ (PANIC)
CO 50 ppm
VOC 100 ppm
AUTO
GAS SELECT
▲ AIR or ▲ AIRor
SHIFT ▼ (PANIC) SHIFT ▼ (PANIC)
ESCAPE
16 . With the standard four sensors screen displayed, press and release the POWER/
ENTER button to proceed to the Calibration In Process Screen with APPLYGAS
and the gas readings flashing.
APPLYGAS
CH4 O2
0
%LEL
20.9%
H2S CO
0.0
ppm
0
ppm
If you do not want to proceed with the calibration, press and release the DISP/LOCK
button to return to the Gas Select Screen.
If you do want to continue with the calibration, proceed to the next step.
20 . If all channels passed calibration and if no sensors are in a sensor life warning
condition or if SEN LIFE ALERT is set to OFF, the following screen sequence
occurs.
AUTO 4:07 AUTO
CH4 O2 GAS SELECT
50
%LEL
12.0%
PASS H2S CO
25.0
ppm
50
ppm VOC 100 ppm
The buzzer and alarm LEDs activate in a double pulsing pattern. Press and release
the RESET button to reset the alarm and continue to the Gas Select Screen. After
calibrating the PID channel by following the instructions below, attempt to calibrate
the standard channels again. If the failure continues, investigate the cause. See
“Troubleshooting” on page 102.
If SEN LIFE ALERT is set to ON (factory setting) and if a sensor is in a sensor life
warning condition, the following screen appears. The buzzer and LEDs activate in a
double pulsing pattern. Press and release POWER/ENTER to acknowledge the
sensor life warning.
SEN LIFE
WARNING
CO
AUTO
GAS SELECT
26 . Use the ▲AIR and SHIFT ▼ (PANIC) buttons to move the cursor next to the gas you
want to use for calibration.
27 . Press and release the POWER/ENTER button to proceed to the Calibration In
Process Screen for the PID channel with APPLYGAS and the gas reading flashing.
APPLYGAS
VOC
0
ppm
If you do not want to proceed with the calibration, press and release the DISP/LOCK
button to return to the Gas Select Screen.
If you do want to continue with the calibration, proceed to the next step.
28 . If you are using a 5-gas calibration cylinder, continue with Step 29.
If you are using a PID only calibration cylinder:
a. Connect the tubing from the regulator to the rigid tube on the probe. Allow the
GX-6000 to draw gas for one minute.
b. Press and release the POWER/ENTER button to set the span adjustment for the
PID channel to the programmed value. Continue with Step 30.
VOC
100
ppm
POWER/ENTER
to acknowledge
31 . If the PID channel passed calibration and if no sensors are in a sensor life warning
condition or if SEN LIFE ALERT is set to OFF, the following screen sequence
occurs.
AUTO 4:07 CAL
CAL MODE
AIR CAL
AUTO CAL
PASS SINGLE CAL
NORMAL MODE
VOC
100
ppm
VOC
FAIL
ppm
The buzzer and alarm LEDs activate in a double pulsing pattern. Press and release
the RESET button to reset the alarm and continue to the Gas Select Screen. Attempt
to calibrate the PID channel again. If the failure continues, investigate the cause. See
“Troubleshooting” on page 102.
If SEN LIFE ALERT is set to ON (factory setting) and if any sensor is in a sensor
life warning condition, the following screen appears. The buzzer and LEDs activate
in a double pulsing pattern. Press and release POWER/ENTER to acknowledge the
sensor life warning and continue to the Gas Select Screen.
SEN LIFE
WARNING
CO
AIR
CH4 O2
0
%LEL
20.9%
H2S CO
0.0
ppm
0
ppm
VOC
0
ppm
3. If you suspect any low-level VOC background in the area, you will need to use the
VOC zero filter.
NOTE: If you have both a PID and a CO2 sensor installed, you will need to follow the
directions on pg.209 instead of following the directions shown below.
a. Remove the plug from each end of the VOC zero filter.
b. Attach the VOC zero filter to the inlet fitting or probe. The filter does not have a
preferred flow direction.
c. Let the instrument draw through the VOC zero filter for 1 minute before
continuing.
4. To continue with the fresh air adjustment, press and hold the ▲AIR button.
If you do not want to continue, press and release the DISP/LOCK button and the unit
will return to the Calibration Mode Screen.
5. The GX-6000 will indicate that it is adjusting the zero reading for a few seconds and
then it will prompt you to release the ▲AIR button.
AIR AIR AIR
AIR CAL AIR CAL AIR CAL
ZERO ADJUSTED
0.0
ppm
0
ppm
NORMAL MODE
VOC
0
ppm
7. If the VOC zero filter was used, remove the filter from the GX-6000’s inlet fitting or
probe. Reinstall the plugs on each end of the filter.
Performing a Span Adjustment in Single Calibration
1. Install the demand flow regulator onto the calibration cylinder.
2. Connect the sample tubing to the demand flow regulator.
3. Install the sample hose and probe on the GX-6000 inlet fitting. Make sure the probe’s
two halves are tightened firmly together to avoid leaks that can affect the calibration.
See Figure 22 on page 112 for an illustration of the internal parts of the probe.
4. Move the cursor next to the SINGLE CAL menu item by using the ▲AIR or SHIFT
▼ (PANIC) button.
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
NORMAL MODE
6. Move the cursor next to the sensor you want to calibrate with the ▲AIR and SHIFT
▼ (PANIC) buttons. In the example above, the CH4 sensor is selected for span
adjustment.
If you do not want to proceed with the span adjustment, press and release the DISP/
LOCK button or place the cursor next to ESCAPE and press and release POWER/
ENTER to return to the Calibration Mode Screen.
To proceed with the calibration, continue with the next step.
7. For a CH4, O2, H2S, or CO channel continue to Step 10.
For a VOC channel, see Step 8 and Step 9 before continuing with Step 10.
8. Press and release the POWER/ENTER button. The target gas selection screen will
display. This screen allows you to select the gas you want to use for calibration.
Isobutylene will always be displayed on the top line. The second line will show the
gas that is currently selected in the PID Gas Name Screen in Display Mode. In the
example below, acetone is the currently selected gas. If isobutylene is the selected
gas, then isobutylene will appear twice in this list.
SINGLE
GAS SELECT
Isobutylene
Acetone
ESCAPE
9. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to move the cursor next to the gas you
want to use for calibration. Continue to Step 10.
11 . Connect the tubing from the demand flow regulator to the rigid tube on the probe.
Allow the GX-6000 to draw gas for one minute.
12 . If necessary, adjust the reading to match the cylinder concentration with the ▲AIR
and SHIFT ▼ (PANIC) buttons. If the reading cannot be adjusted to match the
cylinder concentration, see “Troubleshooting” on page 102.
13 . If SEN LIFE ALERT is set to ON (factory setting), and if a sensor is in a sensor life
warning condition, the following screen sequence occurs.
The buzzer and LEDs activate in a double pulsing pattern in the sensor life warning
screen. Press and release POWER/ENTER to acknowledge the warning.
POWER/ENTER to acknowledge
SINGLE SINGLE
SINGLE CAL
CH4
O2
PASS H2S
CO
VOC
ESCAPE
CAUTION: When calibrating the oxygen sensor, verify the concentration of oxygen
listed on the cylinder’s label. For oxygen-free samples (100% nitrogen for
example), set the oxygen calibration value to 0.0%.
17 . After the last channel is calibrated, disconnect the calibration tubing from the probe,
then unscrew the demand flow regulator from the calibration cylinder.
18 . With the Select Sensor Screen displayed, place the cursor next to ESCAPE using the
▲AIR or SHIFT ▼ (PANIC) button.
SINGLE
SINGLE CAL
CH4
O2
H2S
CO
VOC
ESCAPE
19 . Press and release the POWER/ENTER button to return to the Calibration Mode
Screen.
20 . Use the SHIFT ▼ (PANIC) button to place the cursor next to the NORMAL MODE
menu item, then press and release the POWER/ENTER button to return to
Measuring Mode.
NOTE: The BUMP DISP menu item in Maintenance Mode is factory set to OFF. The
BUMP TEST menu item will not appear unless BUMP DISP is set to ON. See
pg.145 for instructions.
NOTE: The GX-6000 uses parameters defined in the BUMP PARAMETERS menu
item of Maintenance Mode while performing a bump test. To view or update
these parameters, see pg.146.
If the combustible gas channel is calibrated to a gas other than methane, use an appropriate
multigas cylinder.
1 . Install the demand flow regulator onto the calibration cylinder.
2 . Connect the sample tubing to the demand flow regulator.
3 . Install the sample hose and probe on the GX-6000 inlet fitting. Make sure the probe’s
two halves are tightened firmly together to avoid leaks that can affect the calibration.
See Figure 22 on page 112 for an illustration of the internal parts of the probe.
4 . While in Measuring Mode, press and hold the SHIFT ▼ (PANIC) button, then press
the DISP/LOCK button and release both buttons.
5 . If the unit prompts you for the password, enter it by using the ▲AIR and SHIFT ▼
(PANIC) buttons to select each password number and then pressing and releasing
POWER/ENTER to enter the number and move on to the next one.
NOTE: You will only be prompted to enter a password if the PASSWORD menu item
in User Mode is set to ON. See pg.100 for instructions to update the
PASSWORD setting and to choose a password.
6. The Calibration Mode Screen displays with the cursor next to AIR CAL.
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
BUMP TEST
NORMAL MODE
8. Press and release the POWER/ENTER button. The first of the Gas Select Screens,
the standard four sensors screen, will be displayed. In addition to the standard four
sensors screen, there is a separate screen for each installed smart sensor, and an
ESCAPE screen that will take you back to the Calibration Mode menu. Use the
▲AIR and SHIFT ▼ (PANIC) buttons to display the gas(es) you want to test.
AUTO AUTO
GAS SELECT GAS SELECT
CH4 50 %LEL
O2 12.0 % ▲ AIR or
H2S 25.0 ppm SHIFT ▼ (PANIC)
CO 50 ppm
VOC 100 ppm
AUTO
GAS SELECT
▲ AIRor ▲ AIR
or
SHIFT ▼ (PANIC) SHIFT ▼ (PANIC)
ESCAPE
9. Before proceeding, confirm that the bump test gas value(s) are the same as the
concentration(s) in the calibration cylinder. If they are not, adjust the bump test gas
value(s) be entering the AUTO CAL menu item in Calibration Mode, changing the
value(s) there, and reentering the BUMP TEST menu item.
If you do not want to proceed with the bump test, press and release the DISP/LOCK
button to return to the Gas Select Screen.
If you do want to continue with the bump test, proceed to the next step.
11 . When AUTO CAL in the BUMP PARAMETERS menu item in Maintenance
Mode is set to OFF:
a. The instrument will indicate which channels passed or failed the bump test with
an OK (pass) or an NG (fail) to the right of the gas. You can scroll between the
bump test results and the bump test gas readings with the ▲AIR and SHIFT ▼
(PANIC) button.
BUMP BUMP
CH4 O2
CH4
O2
OK
NG
▲ AIR
or 50
%LEL
19.0%
SHIFT ▼ (PANIC)
H2S OK H2S CO
CO NG 25.0
ppm
13
ppm
b. If all channels passed: Press and release the POWER/ENTER button to return to
the Gas Select Screen. If START BUMP PASS is set to ON (factory setting) and
if your instrument is a 4-gas only, the instrument will automatically start the
warmup sequence.
c. If any channels failed: The buzzer will sound a double pulsing tone until you
press and release the POWER/ENTER button to return to the Gas Select Screen.
See “Troubleshooting” on page 102 to investigate the cause of the failure and
replace the failed sensor(s) if necessary.
d. Disconnect the tubing from the probe.
b. To return to the Gas Select Screen, press and release the POWER/ENTER button.
If START BUMP PASS is set to ON (factory setting) and if your instrument is a
4-gas only, the instrument will automatically start the warmup sequence.
c. Disconnect the tubing from the probe.
If any channel fails the bump test:
a. The instrument will beep and a calibration is immediately and automatically
started. Continue to apply the calibration gas. CAL will appear at the top of the
screen along with a countdown.
The calibration time counted down during a calibration initiated because of a
failed bump test is the difference between the gas time and the calibration time
defined in the BUMP SETTING menu item in Maintenance Mode.
60CAL
CH4 O2
49
%LEL
19.0%
H2S CO
23.0
ppm
32
ppm
CAL
▲ AIRor CH4 O2
SHIFT ▼ (PANIC) 50
%LEL
18.0% ▲ AIR
or
SHIFT ▼ (PANIC)
H2S CO
25.0
ppm
49
ppm
c. To return to the Gas Select Screen, press and release the POWER/ENTER button
at any time. If any channel failed the calibration, the buzzer will sound a double
pulsing tone until you press and release the POWER/ENTER button.
d. Disconnect the tubing from the probe.
13 . Unscrew the demand flow regulator from the calibration cylinder.
14 . If you want to bump test any more channels, repeat Step 10 - Step 13.
15 . Use the SHIFT ▼ (PANIC) button to display the ESCAPE screen, then press and
release the POWER/ENTER button to return to the Calibration Mode Menu.
16 . Use the SHIFT ▼ (PANIC) button to place the cursor next to the NORMAL MODE
menu option, then press and release the POWER/ENTER button to return to
Measuring Mode.
Overview
This section describes the GX-6000 in User Mode. See Table 12 below for a list of the
items found in User Mode, the page that the menu item’s instructions can be found on, and
a short description of the menu item.
Table 12: User Mode Menu Items
WARNING: The GX-6000 is not in operation as a gas detector while in User Mode.
PASSWORD PASSWORD
MODE SELECT
USER MODE
MAINTENANCE INPUT
PASSWORD
0 0 0 0
4. If PASSWORD has been set to OFF, the Mode Select screen will appear, prompting
you to choose the mode you’d like to enter.
5. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to put the cursor in front of USER
MODE and press and release the POWER/ENTER button. Continue to Step 8.
6. If PASSWORD has been set to ON in User Mode, a password screen will appear.
The first digit will be flashing.
7. Enter the password by using the ▲AIR and SHIFT ▼ (PANIC) buttons to select each
password number and then pressing and releasing the POWER/ENTER button to
enter it and move on to the next number until all of the numbers are entered.
Continue to Step 8.
8. The main menu displays. It has 7 menu items.
MAINTE
USER
DATE
DATE FORMAT
CONFIGURE CH
LANGUAGE
ROM/SUM
PASSWORD
START MEASURE
9. Use the ▲AIR or SHIFT ▼ (PANIC) button to move the cursor up and down the
menu items.
NOTE: The example screen shown below assumes a DATE FORMAT setting of MM/
DD/YYYY (factory setting). If your instrument’s DATE FORMAT is set to
something else, your screen will appear different.
5/ 1/2015
MM/DD/YYYY
12:55
3. Use ▲AIR or SHIFT ▼ (PANIC) to move the cursor next to the channel you want to
configure.
4. Press and release POWER/ENTER. The available configuration options are shown.
The current setting for the channel will be flashing.
USER
CHANNEL 1
CH4 (HC )
O2 (OX )
H2S (H2S )
CO (CO )
VOC (SM1)
ESCAPE
NOTE: Each channel must have a gas assignment. If you change one channel’s gas to
something else, the two affected channels will end up swapping places. In the
example above, when you select H2S as channel 1’s new gas, CH4, which used
to be channel 1, would automatically be assigned to channel 3, H2S’ old
channel.
7. Repeat Step 5 and Step 6 for any other channels you want to configure.
8. Use SHIFT ▼ (PANIC) to move the cursor next to the ESCAPE menu item.
NOTE: If you want to exit to the main menu without saving any channel configuration
changes, press and release DISP/LOCK.
9. Press and release POWER/ENTER to save the changes and return to the main menu.
If you do not wish to select a new language, press and release DISP/LOCK. The unit
will return to the main menu.
4. Press and release POWER/ENTER to save the new language setting and return to the
main menu. The GX-6000’s user interface will now be in the newly selected
language.
ROM 05017
SUM 62C7
ON
INPUT
PASSWORD
0 0 0 0
Overview
This chapter describes troubleshooting procedures for the GX-6000. It also includes
procedures for replacing and recharging the batteries and replacing various consumable
parts.
WARNING: RKI Instruments, Inc. recommends that service, calibration, and repair
of RKI Instruments be performed by personnel properly trained for this
work. Replacing sensors and other parts with original equipment does
not affect the intrinsic safety of the instrument.
Troubleshooting
The troubleshooting table describes error messages, symptoms, probable causes, and
recommended action for problems you may encounter with the GX-6000.
Table 13: Troubleshooting the GX-6000
• The LCD is blank. • The unit may have been 1. To turn on the unit, press and briefly
turned off. hold the POWER/ENTER button.
• The alkaline batteries may 2. If the unit does not turn on, replace
need to be replaced or Li-ion the alkaline batteries or recharge the
battery pack may need to be Li-ion battery pack.
recharged. 3. If the difficulties continue, contact
RKI Instruments, Inc. for further
instruction.
• The LCD shows • The H2S scrubber disk or 1. Replace the H2S scrubber disk or
abnormally high or charcoal filter may need to charcoal filter.
low readings but be replaced. 2. Recalibrate the unit.
other gas detection • The GX-6000 may need to 3. If the difficulties continue, replace
instruments do not. be recalibrated. the sensor for the affected channel(s)
and calibrate the affected channel(s).
• The sensor for the affected 4. If the difficulties continue, contact
channel(s) may need RKI Instruments, Inc. for further
replacement. instruction.
• The unit indicates • The probe tube is clogged. 1. Inspect the probe tube for any
flow failure and • The hydrophobic filter disk obstructions.
does not recover in the probe is dirty. 2. Inspect the hydrophobic filter disk in
when RESET is the probe and replace if necessary.
pressed and • The sample hose has a kink 3. Inspect the sample hose for kinks or
released. or obstruction. obstructions and replace if
• The pump is malfunctioning. necessary.
4. Attempt to set the low flow setpoint
as described on pg.161.
5. If difficulties continue, contact RKI
Instruments, Inc. for further
instruction.
• Auto calibration or • The auto calibration values 1. Check all calibration tubing for leaks
single calibration may not match the cylinder or for any bad connections.
fails. gas concentrations (auto 2. Make sure the GX-6000 has been
calibration only). properly set up for calibration.
• The charcoal filter is 3. Change the charcoal filter.
saturated causing an elevated 4. Verify that the calibration cylinder
CO reading. contains an adequate supply of fresh
test sample.
• The sample gas is not 5. Clean the PID sensor’s lamp.
reaching the sensors because 6. Replace the PID sensor’s electrode
of a bad connection. stack.
• The calibration cylinder may 7. If the fail condition continues,
be out of gas or is outdated. replace the sensor(s).
• For PID sensors, the lamp 8. If the difficulties continue, contact
may need to be cleaned. RKI Instruments, Inc. for further
• For PID sensors, the instruction.
electrode stack may need to
be replaced.
• The sensor for the affected
channel(s) may need
replacement.
• Heart symbol at the • A microprocessor error has • Contact RKI Instruments, Inc. for
top of the screen occurred. further instruction.
becomes steadily on
or disappears
• Instrument does not • Sensor gasket is not seated • Turn the instrument off and remove
go into low flow properly. the flow chamber. Ensure that the
alarm when inlet is sensor gasket’s flow fitting
plugged. connections are facing up and that the
gasket is secured under the gasket
aligning tabs.
Replace or recharge the batteries when the battery icon indicates that the unit is in low
battery warning. When in low battery warning, the lowest battery level indication bar
disappears and the battery icon will be flashing.
Replacing Alkaline Batteries
NOTE: Use Procell PC 1500 alkaline batteries or RKI Instruments, Inc. lithium-ion
battery pack 49-1619 to maintain the CSA classification of the GX-6000. Use
of other batteries will void the CSA classification and may void the warranty.
Do not mix old/new or different types of batteries.
NOTE: Utiliser Procell 1500 piles alcalines ou RKI Instruments, Inc. pack batterie
lithium-ion 49-1619 de maintenir la classification CSA de la GX-6000.
L’utilisation d’autres piles annule la classification CSA et peut annuler la
garantie. Ne mélangez pas les anciennes/nouvelles ou différents types de piles.
2. Turn the GX-6000 over so that the flow chamber and battery cover are facing up and
the LCD is facing down.
3. Push the battery cover latch toward the bottom of the instrument.
NOTE: Utiliser Procell 1500 piles alcalines ou RKI Instruments, Inc. pack batterie
lithium-ion 49-1619 de maintenir la classification CSA de la GX-6000.
L’utilisation d’autres piles annule la classification CSA et peut annuler la
garantie. Ne mélangez pas les anciennes/nouvelles ou différents types de piles.
WARNING: Do not remove the battery pack while the GX-6000 is on.
CAUTION: To be used only with lithium ion battery pack p/n 49-1619. Charge only
with RKI charger model BC-6000, p/n 49-2182, or RKI charger model BC-
6000DC, p/n 49-2183. Use of other rechargeable batteries or chargers or
charging of other rechargeable batteries in the GX-6000 will void the
warranty.
PRUDENCE: Pour être utilisé uniquement avec une batterie au lithium-ion p/n 49-1619.
Charge uniquement en fonction du modèle de chargeur RKI BC-6000, p/n
49-2182, ou modèle de chargeur BC-6000DC RKI, p/n 49-2183.
L’utilisation d’autres piles ou chargeurs rechargeables ou charger d’autres
batteries recharegables dans le GX-6000 annule la garantie.
The charger can be mounted to the wall using DIN rail, if desired. See pg.22 for
instructions.
1 . Make sure the GX-6000 is off.
2 . Confirm that the power adapter’s DC output cord is plugged into the charger.
AC Power
Supply
Power Supply
Jack (on
bottom of
charger)
DC12V
To AC
Outlet
DC Adapter
Power Supply
Jack (on
bottom of
charger)
DC12V
To Vehicle
12 VDC
Power Socket
DC Powered Charging Station
RKI GX-6000
D IS P R E SE T
A IR
LOC K
S H IF T
( P A N IC )
P O W E R /E N TE R
AC Power
Supply
Power Supply
Jack (on
bottom of
charger)
DC12V
To AC
Outlet
DC Adapter
Power Supply
Jack (on
bottom of
charger)
DC12V
To Vehicle
12 VDC
Power Socket
DC Powered Charging Station
Battery Pack
Figure 21: Putting the Battery Pack into the Charging Station
6. When a full charge has been reached, approximately 3 hours, the red LED on the
charging station will turn off. Remove the battery pack from the charging station and
unplug the charging station’s power cord from the AC outlet.
7. If necessary, reinstall the battery pack as described in Step 7 - Step 8 on pg.106.
Top Half
Gasket
Bottom Half
Sample Hose
Figure 22: Replacing the Particle Filter and Hydrophobic Filter Disk
3. Set the top half aside. Make sure the bottom half remains oriented with the sample
hose facing down.
4. The particle filter should be sitting on top of the hydrophobic filter disk. If it is not,
remove it from the top half of the probe.
CAUTION: If you are replacing filters in a probe intended for chlorine detection, do
not install the particle filter.
8. Carefully screw the top half onto the bottom half disk while keeping the probe
oriented with the probe tip facing up and the sample hose facing down. The particle
filter fits into a recess in the top half of the probe. Be sure the particle filter is not
compressed in any way.
9 . Tighten the halves together very firmly to ensure a seal.
10 . To test the seal, do the following.
a. Install the probe on the GX-6000.
b. Startup the GX-6000.
c. Confirm that a low flow alarm occurs when you cover the end of the probe tube
with your finger.
d. If a low flow alarm does not occur, hand tighten the probe further.
e. If a low flow alarm still does not occur when you cover the probe tube with your
finger, disassemble the probe, inspect the placement of the hydrophobic filter
disk, reassemble the probe, and re-test it.
Sensor G asket
Figure 23: Replacing the H2S Scrubber Disk and Charcoal Filter
6. Ensure that the sensor gasket is installed properly. The flow fitting connections need
to be facing up and cannot be skewed sideways. The gasket must be pushed down in
the center and secured under the gasket aligning tabs.
Failure to install the sensor gasket appropriately may result in inaccurate gas
readings.
Replacing a Sensor
1. Verify that the GX-6000 is off.
2. Use a small Phillips screwdriver to unscrew the four screws holding the flow
chamber to the rest of the GX-6000’s case.
3. Grasp the sides of the flow chamber and lift it away from the rest of the case. The
screws are captive screws and will not come out of the flow chamber.
4. Gently pull up the edges of the sensor gasket to loosen it from its connections. Be
sure not to lose any of the filters that are in it. The sensors will be exposed.
5. Locate the sensor you want to replace and remove it from its socket.
6. Carefully insert the replacement sensor in the correct socket.
a. Be sure that the new sensor is installed in the same position as the old sensor and
that the following sensor priority for the smart sensor 1 position is maintained.
Table 14: Priority for Smart Sensor 1 Position
For example, if your instrument is a 4-gas + low range PID + high range PID, the
low range PID sensor should be in the smart sensor 1 position and the high range
PID sensor should be in the smart sensor 2 position. If your instrument is a 4-gas
+ high range PID, the high range PID sensor should be in the smart sensor 1
position and a dummy sensor should be installed in the smart sensor 2 position.
See the figure below for the location of the smart sensor 1 and 2 positions.
CAUTION: When replacing a sensor, verify that the sensor is properly aligned with its
socket before inserting it into the socket. The CO and H2S sensors have
alignment slots which match up with alignment tabs in the sockets. The
combustible sensors each have two non-polarized contacts which must line
up with the contacts in their socket. Forcing a sensor into its socket may
damage the sensor or the socket.
7. If your instrument has a factory installed dummy sensor, ensure that it is still
installed correctly. Make sure that the flat side is up and the hollow side is down.
Filter Holder
CAUTION: The aluminum oxide cleaning kit is only for use with the 10.0 eV and
10.6 eV PID sensors’ lamps. An anhydrous methanol cleaning kit is used to
clean the 11.7 eV PID sensor’s lamp. DO NOT USE THE ALUMINUM
OXIDE CLEANING KIT TO CLEAN THE 11.7 eV PID SENSOR’S LAMP!
See “Appendix I: 11.7 eV PID Sensor” on page 239 for 11.7 eV lamp
cleaning kit information.
NOTE: Do not touch the lamp window with your fingers as this may contaminate the
window with finger oil. At this point it is recommended that the finger cots be
used on the fingers handling the lamp. Finger cots are included with the lamp
cleaning kit.
Lamp Window
Figure 29: Lamp Window Location
11 . If the spring in the lamp cavity comes out, place it back into the lamp cavity.
12 . Hold the lamp in one hand being careful not to touch the lamp window with your
fingers.
13 . With the other hand collect a small amount of aluminum oxide powder on a cotton
swab.
NOTE: Do not touch the new lamp window (the flat end) with your fingers as this may
contaminate the window with finger oil.
NOTE: At this point it is recommended that the finger cots be used on the fingers
handling the lamp. Finger cots are included with the lamp cleaning kit.
NOTE: Do not touch the new lamp window (the flat end) with your fingers as this may
contaminate the window with finger oil.
NOTE: At this point it is recommended that the finger cots be used on the fingers
handling the lamp. Finger cots are included with the lamp cleaning kit.
8. Hold the PID sensor steady on the working surface with one hand and using the other
hand, locate the tabs on the electrode stack removal tool and insert them into the slots
on the side of the PID sensor near the face.
Table 16 lists part numbers for the GX-6000’s replacement parts and accessories.
Table 16: General Parts List
47-5084RK USB/IrDA adapter module, Legasic, for use with all premier portables (without
USB cable)
47-5084RK-01 USB/IrDA adapter assembly, Legasic, for use with all premier portables (with
module and USB cable)
47-5085RK Cable, USB A to USB mini, 6 feet, for 47-5084RK USB/IrDA adapter module
47-5093 USB/IrDA adapter with cable and CD (not for use with Eagle 2)
49-0115RK AC adapter
49-1120RK AA size alkaline battery
49-1619 Li-ion battery pack
49-1620 Alkaline battery pack
49-2182-01 Battery charger, BC-6000, with 100 - 240 VAC adapter
49-2183 Battery charger, BC-6000, with 12 VDC adapter
51-1500RK PID sensor replacement lamp, 0 - 50,000 ppb, for PID-001 sensor
51-1502 PID sensor replacement lamp, 0 - 6,000 ppm, HPPM type, for PID-002 sensor
51-1502-01 PID sensor replacement lamp/electrode stack set, 0 - 6,000 ppm, HPPM type, for
PID-002 sensor
51-1503 PID sensor replacement lamp, for PID-001L and PID-002L sensor
65-7001RK Dummy sensor, LEL, O2, CO, or H2S sensor position
81-0143RK-04 Calibration cylinder, 5-gas (Proprietary blend of IBL, CH4, O2, H2S, CO),
34 liter aluminum
81-0154RK-02 Calibration cylinder, 4- gas (CH4/O2/ H2S/CO) 58 liter
81-1054RK Regulator, demand-flow type (for 58- and 103-liter aluminum or steel, and 34-liter
aluminum calibration cylinders)
81-1055RK Regulator, demand-flow type (for 17- and 34-liter steel cylinder)
81-6AAX-DLV Calibration kit: 34 liter four-gas (CH4/O2/H2S/CO) cylinder, 34 liter 10 ppm IBL in
air cylinder, demand flow regulator, calibration tubing
81-6ABI-DLV Calibration kit: 34 liter five-gas (IBL/CH4/O2/H2S/CO) cylinder, 34 liter 5 ppm
SO2 in N2 cylinder, demand flow regulator, calibration tubing
Overview
This appendix describes the GX-6000 in Maintenance Mode. The GX-6000 is factory-set
to suit most applications. Update settings in Maintenance Mode only if required for your
specific application. Maintenance Mode items and their factory settings are listed in
Table 17 below.
Table 17: Maintenance Mode Menu Items
0 0 0 0 - - - -
ENTER FOR
MAINTENANCE MODE
User PASSWORD OFF User PASSWORD ON User PASSWORD ON
Maint. PASSWORD ON Maint. PASSWORD ON Maint. PASSWORD OFF
or OFF
4. If User Mode’s PASSWORD has been set to OFF, the Mode Select screen will
appear, prompting you to choose the mode you’d like to enter.
5. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to put the cursor in front of
MAINTENANCE and press and release the POWER/ENTER button. A password
screen will appear. Continue to Step 7.
6. If User Mode’s PASSWORD has been set to ON, a password screen will appear.
7. The appearance of the password screen will be affected by Maintenance Mode’s
PASSWORD 4:07
INPUT INPUT
PASSWORD PASSWORD
0 0 0 0 - - - -
ENTER FOR
MAINTENANCE MODE
Maint. PASSWORD set to ON Maint. PASSWORD set to OFF
8. If Maintenance Mode’s PASSWORD is set to OFF, continue to Step 11.
9. If Maintenance Mode’s PASSWORD is set to ON (factory setting), enter the
password by using the ▲AIR and SHIFT ▼ (PANIC) buttons to select each password
number and then pressing and releasing the POWER/ENTER button to enter it and
move on to the next number until all of the numbers are entered. The password is
0006.
10 . If you enter an incorrect password, an error screen will appear.
PASSWORD
PASSWORD
ERROR
You must turn the unit off and reenter Maintenance Mode using the correct
password. Continue to Step 12.
11 . If Maintenance Mode’s PASSWORD is set to OFF, press and release the POWER/
ENTER button to enter Maintenance Mode. There is no need to enter a password.
The password screen appears even if PASSWORD is set to OFF so that a special
factory mode can be accessed, if necessary.
5/ 1/2015
MM/DD/YYYY
12:55
3. Use the ▲AIR or SHIFT ▼ (PANIC) buttons to move the cursor to the operation you
wish to perform.
See “Calibrating Using the Auto Calibration Method” on page 70 for instructions to
perform an automatic calibration.
NOTE: Even if START CAL PASS is set to ON (factory setting), the GX-6000 will
not automatically start its warmup sequence. That setting only applies to
Calibration Mode operation.
See “Calibrating Using the Single Calibration Method” on page 83 for instructions to
perform a single calibration.
NOTE: The BUMP DISP menu item in Maintenance Mode is factory set to OFF. The
BUMP TEST menu item will not appear unless BUMP DISP is set to ON. See
“Updating the Bump Display Setting (BUMP DISP)” on page 145 for
instructions.
Entering the BUMP TEST menu item brings you to the first of the Gas Select Screens.
See “Performing a Bump Test” on page 89 for instructions, starting at Step 8.
NOTE: Even if START BUMP PASS is set to ON (factory setting), the GX-6000 will
not automatically start its warmup sequence. That setting only applies to
Calibration Mode operation.
1. From the main menu, place the cursor next to CAL SETTING.
2. Press and release POWER/ENTER. The CAL SETTING menu appears.
Updating the Calibration Reminder Setting (CAL REMINDER)
ON (factory setting): The GX-6000 will give an indication at start up if it is due for
calibration. The type of indication will depend on the CAL EXPRD setting (see pg.143).
OFF: The GX-6000 will not give an indication at start up if it is due for calibration.
1 . From the CAL SETTING screen, place the cursor next to CAL REMINDER.
2 . Press and release POWER/ENTER.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the CAL
SETTING menu.
Updating the Calibration Interval (CAL INTERVAL)
This setting defines the amount of time between calibrations. The time can be set in 1 day
increments. The minimum setting is 1 day and the maximum setting is 365 days. The
factory setting is 90 days.
1 . From the CAL SETTING screen, place the cursor next to CAL INTERVAL.
2 . Press and release POWER/ENTER.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
Tip: Press and hold ▲AIR or SHIFT ▼ (PANIC) to rapidly scroll through the settings.
4. Press and release POWER/ENTER to save the setting and return to the CAL
SETTING menu.
Updating the Calibration Expired Setting (CAL EXPRD)
This item defines what indication is given during start up when calibration is due and
CAL REMINDER is set to ON.
CONFIRM TO USE (factory setting): The GX-6000 will give an indication at start up if
NO EFFECT: The GX-6000 will give an indication at startup that calibration is past due.
If desired, the POWER/ENTER button can be pressed to enter Calibration Mode and
perform a calibration but it is not necessary to acknowledge the calibration due indication.
The warm-up sequence will continue on its own.
1 . From the CAL SETTING screen, place the cursor next to CAL EXPRD.
2 . Press and release POWER/ENTER.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the CAL
SETTING menu.
Updating the Calibration Check Gas Setting (CAL CHECK GAS)
This item defines which sensors are used to update the calibration reminder screen.
4 GAS (factory setting): The calibration dates for only the standard 4 sensors (LEL, O2,
H2S, and CO) are used to determine if a calibration is due.
ALL GAS: The calibration dates for the standard 4 sensors along with the PID sensor and
any other installed smart sensors are used to determine if a calibration is due.
1 . From the CAL SETTING screen, place the cursor next to CAL CHECK GAS.
2 . Press and release POWER/ENTER.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the CAL
SETTING menu.
Updating the Auto Start After Successful Cal Setting
(START CAL PASS)
ON: If only the standard 4 sensors are installed, the GX-6000 automatically starts its
warmup sequence after a successful bump test in Calibration Mode. If any smart sensors
(PID, ESS-03, etc) are installed, this parameters has no effect on operation.
NOTE: The GX-6000 will not automatically start its warmup sequence from
Maintenance Mode, regardless of the START CAL PASS setting.
OFF: The GX-6000 does not automatically start its warmup sequence after a successful
bump test.
1 . From the CAL SETTING screen, place the cursor next to START CAL PASS.
2 . Press and release POWER/ENTER.
NOTE: BUMP DISP must be set to ON to have the GX-6000’s bump test date reset
after a successful bump test in an SDM-6000.
OFF (factory setting): The BUMP TEST menu item will not appear in Calibration Mode.
1 . From the BUMP SETTING screen, place the cursor next to BUMP DISP.
2 . Press and release POWER/ENTER. The Bump Display Screen appears.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the BUMP
SETTING menu.
1. From the BUMP SETTING screen, place the cursor next to BUMP
PARAMETERS.
2 . Press and release POWER/ENTER. The Bump Parameters Menu Screen appears.
Updating the GAS TIME Setting
The GAS TIME is the amount of time that the instrument is exposed to gas during a bump
test. The available choices are 30 seconds (factory setting), 45 seconds, 60 seconds, and
90 seconds.
1 . From the BUMP PARAMETERS screen, place the cursor next to GAS TIME.
2 . Press and release POWER/ENTER. The current Gas Time value will begin to flash.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The value will stop flashing.
Updating the CHECK Setting
The CHECK is the bump test tolerance value and is represented as a percentage. It is the
percentage that the bump test reading can differ from the actual gas concentration. If the
bump test reading differs more, the bump test will fail. The available values are 10%,
20% 30% (factory setting), 40%, and 50%.
1 . From the BUMP PARAMETERS screen, place the cursor next to CHECK.
2 . Press and release POWER/ENTER. The current Check value will begin to flash.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The value will stop flashing.
Updating the CAL TIME Setting
The CAL TIME is the total time the instrument is exposed to calibration gas when a
bump test fails if AUTO CAL is set to ON. The bump test time is deducted from the
calibration time. For example, if the CAL TIME is set to 90 seconds and the GAS TIME
is set to 30 seconds, if the bump test fails, the GX-6000 will only be exposed to gas for an
additional 60 seconds. The available values are 90 seconds (factory setting), and 120
seconds.
1 . From the BUMP PARAMETERS screen, place the cursor next to CAL TIME.
NOTE: If BUMP REMINDER is set to ON, it is recommended that BUMP DISP also
be set to ON so that BUMP TEST appears in Calibration Mode.
OFF (factory setting): The GX-6000 will not give an indication at start up if it is due for
bump testing.
1 . From the BUMP SETTING screen, place the cursor next to BUMP REMINDER.
2 . Press and release POWER/ENTER. The current Bump Reminder setting will be
displayed.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The instrument will return to
the BUMP SETTING screen.
Updating the Bump Interval (BUMP INTERVAL)
This setting defines the amount of time between bump tests. The time can be set in 1 day
increments. The minimum setting is 0 days and the maximum setting is 30 days. The
factory setting is 30 days.
1 . From the BUMP SETTING screen, place the cursor next to BUMP INTERVAL.
2 . Press and release POWER/ENTER. The current Bump Interval setting will be
displayed.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The instrument will return to
the BUMP SETTING screen.
NOTE: If BUMP DISP is set to OFF, the bump test menu item will not appear in
Calibration Mode even though the instrument is prompting you to perform a
bump test. A bump test can always be performed in Maintenance Mode, if
necessary.
NO EFFECT: The GX-6000 will give an indication at startup that a bump test is past due.
If desired, the POWER/ENTER button can be pressed to enter Calibration Mode and
perform a bump test but it is not necessary to acknowledge the bump test due indication.
The warm-up sequence will continue on its own.
1 . From the BUMP SETTING screen, place the cursor next to BUMP EXPRD.
2 . Press and release POWER/ENTER. The current Bump Expired setting will be
displayed.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The instrument will return to
the BUMP SETTING screen.
Updating the Bump Check Gas Setting (BUMP CHECK GAS)
This item defines which sensors are used to update the bump test reminder screen.
4 GAS (factory setting): The bump test dates for only the standard 4 sensors (LEL, O2,
H2S, and CO) are used to determine if a bump test is due.
ALL GAS: The bump test dates for the standard 4 sensors along with the PID sensor and
any other installed smart sensors are used to determine if a bump test is due.
1 . From the BUMP SETTING screen, place the cursor next to BUMP CHECK GAS.
2 . Press and release POWER/ENTER. The current Bump Check Gas setting will be
displayed.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The instrument will return to
the BUMP SETTING screen.
NOTE: The GX-6000 will not automatically start its warmup sequence from
Maintenance Mode, regardless of the START BUMP PASS setting.
OFF: The GX-6000 does not automatically start its warmup sequence after a successful
bump test.
1 . From the BUMP SETTING screen, place the cursor next to START BUMP PASS.
2 . Press and release POWER/ENTER.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The instrument will return to
the BUMP SETTING screen.
Returning to the Maintenance Mode Menu (ESCAPE)
1. From the BUMP SETTING screen, place the cursor next to ESCAPE.
2. Press and release POWER/ENTER. The instrument will return to the main menu.
1. From the main menu, place the cursor next to ALARM SETTING.
2. Press and release POWER/ENTER. The ALARM SETTING menu appears.
ESCAPE
3. Move the cursor next to the channel of the alarm point you want to update. Press and
release POWER/ENTER. The channel’s alarm points and alarm operation (oxygen
only) are displayed.
ALARM - P ALARM - P ALARM - P
CH4 O2 H2S
L-H
AL 1 10 %LEL AL 1 19.5 % AL 1 5.0 ppm
AL 2 50 %LEL AL 2 23.5 % AL 2 30.0 ppm
STEL 15.0 ppm
TWA 10.0 ppm
ESCAPE ESCAPE ESCAPE
4. Move the cursor next to the alarm point or alarm operation (oxygen only) that you
want to update.
If you selected the oxygen channel, you can set the alarm rising/falling operation in
addition to the alarm points.
5 . Press and release POWER/ENTER. The alarm point or alarm operation (oxygen
only) will begin to flash.
6 . Use ▲AIR and SHIFT ▼ (PANIC) to adjust the alarm point or alarm operation
(oxygen only) to the desired setting. Keep the following in mind:
• Alarm 1 cannot be set higher than Alarm 2 and Alarm 2 cannot be set lower than
Alarm 1.
If you want to exit this screen without saving any change to the alarms, press and
release DISP/LOCK to return to the Channel Select Screen.
8 . Repeat Step 4 - Step 7 for any additional changes you want to make to the selected
channel’s alarm points.
9 . When you are done making changes to the selected channel, use SHIFT ▼ (PANIC)
to move the cursor next to ESCAPE and press and release the POWER/ENTER
button.
10 . Repeat Step 3 - Step 9 for any changes you want to make to additional channels.
11 . When you are done making changes to all channels, use SHIFT ▼ (PANIC) to move
the cursor next to ESCAPE.
12 . Press and release POWER/ENTER to return to the ALARM SETTING menu.
Updating the Alarm Latching Setting (ALARM LATCHING)
LATCHING (factory setting): The GX-6000 remains in alarm condition until the alarm
condition passes and the RESET button is pressed.
SELF RESET: The GX-6000 automatically resets an alarm when the alarm condition
passes.
1 . From the ALARM SETTING menu, place the cursor next to ALARM
LATCHING.
2 . Press and release POWER/ENTER. The Alarm Latching Screen appears.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the ALARM
SETTING menu.
Updating the Alarm Silence Setting (ALARM SILENCE)
ON (factory setting): Pressing and releasing the RESET button silences the buzzer when
the GX-6000 is in alarm. The LEDs continue to flash, the vibrator continues to pulse, and
the display continues to show the alarm. If you enter Display Mode during an alarm
condition, the buzzer will be silenced but the LEDs will continue to flash and the vibrator
will continue to pulse. If you return to Measuring Mode and there is still an alarm
condition, the LEDs will continue to flash, the vibrator will continue to pulse, and the
buzzer will remain off.
OFF: You cannot silence the buzzer. If you enter Display Mode during an alarm
condition, the buzzer will not be silenced, the LEDs will continue to flash, and the vibrator
will continue to pulse.
1 . From the ALARM SETTING menu, place the cursor next to ALARM SILENCE.
If you do not wish to select a new language, press and release DISP/LOCK. The unit
will return to the main menu.
4. Press and release POWER/ENTER to save the new language setting and return to the
main menu. The GX-6000’s user interface will now be in the newly selected
language.
5. If you select a language other than English, a prompt will appear during startup that
allows you to change the language back to English if desired.
WARNING: If the automatic fresh air feature is turned on, you must startup the GX-
6000 in a known fresh air environment, an environment free of toxic or
combustible gases and of normal oxygen content (20.9%). If this
feature is on and the GX-6000 is started up in the presence of a target
gas, the readings and alarms will not be accurate or reliable.
1. From the main menu, place the cursor next to AUTO ZERO.
AUTO RANGE (factory setting): The IR channel’s gas reading will be displayed in
%LEL until the gas level reaches 100 %LEL, or 5.0 %vol for methane. Once the gas
reading is above 100 %LEL, it is displayed in %vol. Alarm points exist for the %LEL
range but do not exist for the %vol range.
LOW ONLY: The IR channel’s gas reading will only be displayed in %LEL up to 100
%LEL.
1 . From the main menu, place the cursor next to IR AUTO RANGE.
2 . Press and release POWER/ENTER. The IR Auto Range Screen appears.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to move the cursor in front of the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the main menu.
5
MIN
1. From the main menu, place the cursor next to L./B. MODE.
2. Press and release POWER/ENTER. The Leak Check/Bar Hole Mode Screen
appears.
3. Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4. Press and release POWER/ENTER to save the setting and return to the main menu.
1. From the main menu, place the cursor next to MAN DOWN.
2. Press and release POWER/ENTER. The MAN DOWN MENU screen appears.
Updating the MAN DOWN Setting
ON: The Man Down alarm can be triggered if the instrument detects no motion for the
period of time defined in WARNING 1 TIME, WARNING 2 TIME, and ALARM
TIME below.
OFF (factory setting): The Man Down alarm cannot be triggered.
See “Measuring Mode, Alarms” on page 39 for a description of the Man Down alarm.
1 . From the MAN DOWN MENU screen, place the cursor next to MAN DOWN.
2 . Press and release POWER/ENTER. The Man Down Screen appears.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the MAN
DOWN MENU screen.
1. From the main menu, place the cursor next to LOG SETTING.
2. Press and release POWER/ENTER. The LOG SETTING menu appears.
Updating the Data Log Interval Setting (LOG INTERVAL)
This setting indicates how often the GX-6000 saves readings to the data logger. The
following interval times can be selected: 10 minutes, 5 minutes (factory setting), 3
minutes, 2 minutes, 1 minute, 30 seconds, 20 seconds, or 10 seconds.
1 . From the LOG SETTING screen, place the cursor next to LOG INTERVAL.
2 . Press and release POWER/ENTER. The Log Interval Screen appears.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the LOG
SETTING screen.
Updating the Log Clear Display Setting (LOG CLEAR DISP)
ON (factory setting): The LOG CLEAR Screen appears in Display Mode (see pg.55).
OFF: The LOG CLEAR Screen does not appear in Display Mode.
1 . From the LOG SETTING screen, place the cursor next to LOG CLEAR DISP.
2 . Press and release POWER/ENTER. The Log Clear Display Screen appears.
3 . Use ▲AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the LOG
SETTING screen.
OFF: The instrument does not indicate when a sensor is nearing the end of its useful life.
50
1500
3. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to get a reading of approximately
1500 on the bottom value. When you have adjusted the value as close as you can,
press and release the POWER/ENTER button.
4. The pump will turn on and two new values will be displayed. The top will be a
reference value and the bottom will reflect the pump’s current draw.
MAINTE
FLOW ADJUST
1500
2500
5. Connect a flow meter with a valve to the inlet of the instrument and adjust it to 0.2
LPM ± 0.1 LPM. This will be the low flow setpoint. You do not need to adjust
anything at the instrument.
ROM 05017
SUM 62C7
3. The first line displays the ROM number. The ROM number indicates the firmware
version number. In the above example, the ROM number is 05017. The bottom line
displays the firmware file’s checksum, 62C7 in the above example.
4. Press and release the POWER/ENTER button again to return to the main menu.
INPUT
PASSWORD
0 0 0 0
WARNING: When the GX-6000 is restored to its default configuration, the zero and
span values for each channel are reset. You must recalibrate all active
channels if you restore the GX-6000 to its default configuration.
1. From the main menu, place the cursor in front of RESTORE DEFAULT.
RESTORE ALL ?
YES : ENTER
NO : DISP
3. If you do not want to restore the default configuration, press and release DISP/LOCK
to return to the main menu.
YES : ENTER
NO : DISP
5. Press and release the POWER/ENTER button. The screen will indicate that the
default configuration has been restored and return to the main menu.
Overview
This appendix describes the GX-6000 in Maintenance Mode 2. The GX-6000 is factory-
set to suit most applications. Update settings in Maintenance Mode 2 only if required for
your specific application. A description of the Maintenance Mode 2 items is shown in
Table 18 below.
Table 18: Maintenance Mode 2 Menu Items
0 0 0 0 - - - -
ENTER FOR
MAINTENANCE MODE
User PASSWORD OFF User PASSWORD ON User PASSWORD ON
Maint. PASSWORD ON Maint. PASSWORD ON Maint. PASSWORD OFF
or OFF
4. If User Mode’s PASSWORD has been set to OFF, the Mode Select screen will
appear, prompting you to choose the mode you’d like to enter.
5. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to put the cursor in front of
MAINTENANCE and press and release the POWER/ENTER button. A password
screen will appear. Continue to Step 7.
6. If User Mode’s PASSWORD has been set to ON, a password screen will appear.
7. The appearance of the password screen will be affected by Maintenance Mode’s
PASSWORD setting but the procedure for entering Maintenance Mode 2 is
unaffected.
PASSWORD 4:07
INPUT INPUT
PASSWORD PASSWORD
0 0 0 0 - - - -
ENTER FOR
MAINTENANCE MODE
Maint. PASSWORD set to ON Maint. PASSWORD set to OFF
8. Enter the password for Maintenance Mode 2 by using the ▲AIR and SHIFT ▼
(PANIC) buttons to select each password number and then pressing and releasing the
POWER/ENTER button to enter it and move on to the next number until all of the
numbers are entered. The password for Maintenance Mode 2 is 2014.
PASSWORD
ERROR
You must turn the unit off and reenter Maintenance Mode 2 using the correct
password.
10 . The main menu displays.
MAINTE 2
MAINTENANCE 2
GAS COMB
HC SELECT
START MEASURE
11 . Use the ▲AIR or SHIFT ▼ (PANIC) button to move the cursor up and down through
the menu items.
NOTE: A smart sensor channel cannot be turned off if the smart sensor is still installed.
If you want to turn off a smart sensor’s channel, you must first remove the smart
sensor and replace it with a dummy sensor.
1. From the main menu, place the cursor next to GAS COMB.
2. Press and release POWER/ENTER. The Gas Combination Screen appears and all
channels are displayed.
FLASH
GAS COMB
HC ON
OX ON
H2S ON
CO ON
SM1 ON
SM2 OFF
ESCAPE
3. Move the cursor next to the channel you want to update. Press and release POWER/
ENTER. The channel’s current on/off setting will begin to flash.
4. Use ▲AIR and SHIFT ▼ (PANIC) to change the on/off setting.
5. If you want to continue with the change, press and release POWER/ENTER to
accept the setting.
If you want to exit this screen without saving the change, press and release DISP/
LOCK to return to the main menu.
6. Repeat Step 3 - Step 5 for any changes you want to make to additional channels.
7. When you are done making changes to all channels, use SHIFT ▼ (PANIC) to move
the cursor next to ESCAPE.
8. Press and release POWER/ENTER to return to the main menu.
Overview
This chapter explains how to operate the GX-6000 in Leak Check Mode. Leak Check
Mode is used to pinpoint small leaks of combustible gas from valves, flanges,
connections, and other potential leak points. When the GX-6000 is in Leak Check Mode,
only the catalytic combustible sensor and potentially the CO sensor, depending on the
setting of the CO DISPLAY parameter in Maintenance Mode, will be active.
If a GX-6000 is intended for tracking down leaks, it is shipped with Leak Check Mode or
both Leak Check Mode and Bar Hole Mode enabled so that the operator must choose
which operational mode to use when the unit is turned on (see pg.156).
CAUTION: If one or more sensors other than a catalytic combustible sensor is installed,
these sensors will not be active while the GX-6000 is in Leak Check Mode.
NOTE: In order for LEAK CHECK MODE to appear as a selection in the Mode
Select Screen, the L./B. MODE item in Maintenance Mode must be set to
LEAK CHECK MODE or LEAK CHECK MODE BAR HOLE MODE.
The factory setting is OFF.
1. Connect the tapered rubber nozzle or the sample hose and probe to the GX-6000’s
quick connect inlet fitting.
170 • Appendix C: Using the GX-6000 in Leak Check ModeGX-6000 Operator’s Man-
2. Press and briefly hold down the POWER/ENTER button. Release the button when
you hear a beep. The Normal Mode Select Screen displays.
4:07
MODE SELECT
NORMAL MODE
CH4 100 %LEL
O2 40.0 %
H2S 100.0 ppm
CO 500 ppm
VOC 6000 ppm
3. Use the ▲AIR or SHIFT ▼ (PANIC) button to scroll to the Leak Check Mode Select
Screen.
4:07
MODE SELECT
LEAK CHECK MODE
4. With the Leak Check Mode Select Screen displayed, press and release the POWER/
ENTER button to begin the Leak Check Mode startup sequence.
NOTE: If no button is pressed for 20 seconds, the unit will proceed into whichever
mode is displayed.
5. See “Turning On the GX-6000” on page 25 for a description of the remainder of the
warm-up sequence keeping the following in mind:
a. Even if LUNCH BREAK is set to ON, the Lunch Break Screen will not appear.
b. The Low Alarm, High Alarm, STEL, and TWA screens will not appear.
c. If the catalytic combustible sensor fails, it is not possible to enter Leak Check
Mode. Press and release the RESET button to acknowledge the failure and return
to the Normal Mode Select Screen. Replace the failed sensor.
If any other sensor that is installed fails, press and release the RESET button to
acknowledge the failure and continue to Leak Check Mode. Change the failed
sensor(s) as soon as possible for use in Normal Mode.
GX-6000 Operator’s Manual Appendix C: Using the GX-6000 in Leak Check Mode • 171
6. The GX-6000 is now monitoring for gas in Leak Check Mode. The pump is on and
one of the following screens appears depending on the setting of the CO DISPLAY
parameter in Maintenance Mode.
4:07 4:07
CH4 CH4
0 ppm CO
0 ppm
0 ppm
F .S .
5000
ppm
F .S .
5000
ppm
NOTE: For maximum sensor stability, allow 3-5 minutes for the sensor to warm up.
The small increment size in the lower range of a ppm measurement can cause
instability if the unit is not properly warmed up.
7. If you wish to change operating modes by returning to the Mode Select Screen, press
and hold the ▲AIR and DISP/LOCK buttons for 5 seconds.
Performing a Demand Zero, Leak Check Mode
Before using the GX-6000, it is recommended to set the fresh air reading for the target gas
by performing a demand zero. This will set the CH4 channel to zero.
1. Find a fresh-air environment. This is an environment free of toxic or combustible
gases and of normal oxygen content (20.9%).
2. Turn on the unit as described above in “Turning On the GX-6000, Leak Check
Mode”.
3. Press and hold the ▲AIR button. The display prompts you to hold the ▲ AIR button.
4. Continue to hold the ▲AIR button until the display prompts you to release it. The
GX-6000 will set the fresh air reading for the CH4 channel. Start up is complete and
the unit is now ready for monitoring.
Leak Testing
In Leak Check Mode, the GX-6000 only displays combustible gas readings. The
increment of the reading is always 100 ppm but the display range can be adjusted by using
the DISP/LOCK button. The display range choices are 500 ppm, 1000 ppm, 2000 ppm,
and 5000 ppm. The readings are displayed in both numerical and bar graph form. As the
gas concentration increases from 0 ppm, the alarm LEDs begin to blink in unison with the
buzzer’s pulsing. The blinking/pulsing rate increases as the gas reading increases.
In a low-light environment, press and release any of the buttons to turn on the display
backlight. If CONFIRMATION is set to ON in the Maintenance Mode menu, the GX-
6000 beeps once every user-defined interval to confirm that it’s operating.
172 • Appendix C: Using the GX-6000 in Leak Check ModeGX-6000 Operator’s Man-
Setting the Display Range Value
The GX-6000’s Leak Check Mode has 4 display range ppm values to choose from: 500
ppm, 1000 ppm, 2000 ppm, and 5000 ppm.
NOTE: The GX-6000 is always detecting combustible gas up to 5000 ppm. The display
range value is only adjusting what part of the detection range is displayed on the
LCD.
GX-6000 Operator’s Manual Appendix C: Using the GX-6000 in Leak Check Mode • 173
Overscale Conditions
If the gas level goes over the display range, the gas reading will be replaced with “OVER”
and the LEDs and buzzer will pulse continuously until the overscale condition has cleared.
Once the gas level gets low enough, the instrument will start showing ppm readings again.
4:07
CH4
OVER ppm
F .S .
5000
ppm
PEAK
500 ppm
700ppm
F .S .
5000
ppm
3. If the highest reading recorded is higher than the display range value, “OVER” will
be displayed instead of a gas reading. If the display range value is changed to
something higher, the real value for the peak will be displayed.
4. To exit Peak Hold Mode and return to Leak Check Mode, press and release the
RESET button.
174 • Appendix C: Using the GX-6000 in Leak Check ModeGX-6000 Operator’s Man-
Snap Log Mode, Leak Check Mode Operation
The snap logging function in Snap Log Mode allows the user to record data at a specific
time and have it saved to the data logger. The data is assigned a snap log ID and is saved
with the station ID that was selected when the data was taken.
To enter Snap Log Mode and record snap log data:
1. Turn the GX-6000 on as described in “Turning On the GX-6000, Leak Check Mode”
on page 170.
2 . Select the desired display range. In the example below, 5000 ppm has been selected.
3 . Press and hold the SHIFT ▼ (PANIC) button, then press and hold the ▲AIR button
and hold both until you hear a beep. The Station ID Select Screen will display and
the current station ID will be in the middle of the screen.
REC DATA
REC. DATA
STATION ID
TANK 5
NO : DISP
4. To exit Snap Log Mode and return to Leak Check Mode, press and release the DISP/
LOCK button.
5. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to scroll to the desired station ID.
6. Press and release the POWER/ENTER button to continue to the Base Reading
Screen.
4:07
CH4
BASE
100 ppm
F .S .
5000
ppm
NOTE: If CO DISPLAY is set to ON, the CO reading is saved during a snap log but
will not be displayed in the Base Reading Screen, Peak Reading Screen, or
while viewing snap log data at the instrument. However, the saved CO
reading(s) can be viewed using the GX-6000 Datalogging Program.
GX-6000 Operator’s Manual Appendix C: Using the GX-6000 in Leak Check Mode • 175
7. To return to the Station ID Select Screen without taking a base reading, press and
release the RESET button. To exit Snap Log Mode completely and return to Leak
Check Mode, press and release the DISP/LOCK button.
8. In order to get a base, or background, reading, take the instrument a few feet away
from where you intend to test. To save the base reading, press and release the
POWER/ENTER button.
NOTE: If the gas concentration exceeds the display range value without exceeding
5000 ppm, “OVER” will be displayed but the actual gas reading at that time
will be recorded when you press and release POWER/ENTER. For example, if
the display range is 500 ppm but the gas level is 2300 ppm, the display will read
“OVER” but the instrument will save a reading of 2300 ppm.
If the gas concentration exceeds 5000 ppm in the Base Reading Screen, the
instrument will go into an overscale alarm. The full scale display range value
will be displayed but OVER will be recorded when you press POWER/ENTER.
The full scale display range value will continue to be displayed in the Peak
Reading Screen and OVER will again be recorded when you press POWER/
ENTER.
9. The unit will display SAVED before continuing to the Peak Reading Screen.
4:07
CH4
PEAK
1000 ppm
F .S .
5000
ppm
10. Take the instrument to the monitoring area. The instrument will retain and display
the highest gas concentration encountered.
To return to the Station ID Select Screen without taking a peak reading, press and
release the RESET button.
To save the peak reading, press and release the POWER/ENTER button. The reading
on the screen (the highest concentration encountered) will be saved.
NOTE: If the gas concentration exceeds the display range without exceeding 5000 ppm,
“OVER” will be displayed but the actual gas reading at that time will be
recorded when you press and release POWER/ENTER. For example, if the
display range is 500 ppm but the gas level is 2300 ppm, the display will read
“OVER” but the instrument will save a reading of 2300 ppm.
If the gas concentration exceeds 5000 ppm in the Peak Reading Screen, the
instrument will go into an overscale alarm. The full scale display range value
will be displayed and OVER will be recorded when you press POWER/
ENTER.
176 • Appendix C: Using the GX-6000 in Leak Check ModeGX-6000 Operator’s Man-
11 . The unit will display SAVED before returning to the Station ID Select Screen.
12. Repeat Step 5 through Step 11 to take additional snap log data.
13. To exit Snap Log Mode at any time without taking a snap log or when you are
finished recording snap logs, press and release the DISP/LOCK button. The unit will
immediately return to the Leak Check Mode Screen.
14. The data recorded in Snap Log Mode can be viewed in Display Mode. See the next
section for more information.
Viewing Snap Log Data in Leak Check Mode Operation
Snap log data can be viewed while in Leak Check Mode. If snap log data was taken while
in Measuring Mode, that data will also appear.
1. Turn the GX-6000 on as described in “Turning On the GX-6000, Leak Check Mode”
on page 170.
2. Press and hold the RESET button and then press and hold the DISP/LOCK button
and release both. The screen that appears will depend on whether or not any snap
logs have been taken.
REC DATA REC DATA
REC. DATA DATE
5/20/2015
NO DATA MM/DD/YYYY
3:40
NUMBER 001
TANK 5
GX-6000 Operator’s Manual Appendix C: Using the GX-6000 in Leak Check Mode • 177
ID Screen Data Screen
REC DATA 4:07
DATE CH4
5/20/2015
MM/DD/YYYY
3:47
POWER/
PEAK
1000 ppm
Leak Check
Mode Peak
ENTER Data
NUMBER
TANK 5
003 F .S .
5000
ppm
SHIFT ▼ SHIFT ▼
▲ AIR ▲ AIR
REC DATA 4:07
DATE CH4
5/20/2015
MM/DD/YYYY
3:45
POWER/ BASE 100 ppm
Leak Check
Mode Base
ENTER Data
NUMBER
TANK 5
002 F .S .
5000
ppm
SHIFT ▼ SHIFT ▼
▲ AIR ▲ AIR
REC DATA REC DATA
DATE CH4 O2
5/20/2015
0
%LEL
20.9%
MM/DD/YYYY POWER/ H2S CO Normal
3:40
ENTER 0.0
ppm
0
ppm
Mode
Data
NUMBER 001 VOC
TANK 5 0
ppm
178 • Appendix C: Using the GX-6000 in Leak Check ModeGX-6000 Operator’s Man-
Snap log data that was taken in Leak Check Mode operation will have 2 sequential
screens that are part of one data set. One screen is for the base reading and the other
screen is for the peak reading. “BASE” and “PEAK” will appear in the gas reading
screen for each type of reading, respectively. If a base data point appears without a
peak data point after it, it is because a peak data point was not taken.
NOTE: Even if CO DISPLAY was set to ON when the Leak Check snap logs were
taken, the CO reading will not be displayed. The CO readings can be viewed
using the GX-6000 Datalogging Program.
7. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to scroll through the different snap
log data screens. The gas readings will change but the snap log ID is not visible from
this screen.
8. You can also go back and forth between the ID and data screens by pressing and
releasing the POWER/ENTER button.
9. To return to Leak Check Mode, press and release the DISP/LOCK button.
Turning Off the GX-6000, Leak Check Mode
1. Press and hold the POWER/ENTER button.
2. TURN OFF will appear on the display and the buzzer will pulse for about five
seconds.
3. Release the button when TURN OFF disappears from the display.
GX-6000 Operator’s Manual Appendix C: Using the GX-6000 in Leak Check Mode • 179
Appendix D: Using the GX-6000 in Bar Hole Mode
Overview
This chapter explains how to operate the GX-6000 in Bar Hole Mode. Bar Hole Mode is
used to perform consistent checks of bar holes when tracking down underground gas
leaks. When the GX-6000 is in Bar Hole Mode, only the IR combustible and oxygen (if
installed and active) channels are displayed.
If a GX-6000 is intended for bar hole testing, it is shipped with Bar Hole Mode enabled so
that the operator must choose which operational mode to use when the unit is turned on.
In order to use a GX-6000 in Bar Hole Mode, an IR CH4 or an IR HC sensor must be
installed in the Smart Sensor 1 Position in the instrument. Bar Hole Mode will display an
oxygen reading if an oxygen sensor is installed and active but an oxygen sensor is not
required for Bar Hole Mode use.
LEL Sensor
Smart Sensor 1 CO Sensor
Position
H2S Sensor
Smart Sensor 2 CO
180 • Appendix D: Using the GX-6000 in Bar Hole Mode GX-6000 Operator’s Manual
Start Up, Bar Hole Mode
This section explains how to start up the GX-6000 in Bar Hole Mode and get it ready for
operation.
Turning On the GX-6000, Bar Hole Mode
WARNING: Gas alarms are not active when the GX-6000 is in Bar Hole Mode.
NOTE: In order for BAR HOLE MODE to appear as a selection in the Mode Select
Screen, the L./B. MODE item in Maintenance Mode must be set to BAR
HOLE MODE or LEAK CHECK MODE BAR HOLE MODE. The factory
setting is OFF.
1. Connect the tapered rubber nozzle or the sample hose and probe to the GX-6000’s
quick connect inlet fitting.
2. Press and briefly hold down the POWER/ENTER button. Release the button when
you hear a beep. The Normal Mode Select Screen displays.
4:07
MODE SELECT
NORMAL MODE
CH4 100 %LEL
O2 40.0 %
H2S 100.0 ppm
CO 500 ppm
VOC 6000 ppm
GX-6000 Operator’s Manual Appendix D: Using the GX-6000 in Bar Hole Mode • 181
3. Use the ▲AIR or SHIFT ▼ (PANIC) button to scroll to the Bar Hole Mode Select
Screen.
4:07
MODE SELECT
BAR HOLE MODE
4. With the Bar Hole Mode Select Screen displayed, press and release the POWER/
ENTER button to begin the Bar Hole Mode startup sequence.
NOTE: If no button is pressed for 20 seconds, the unit will proceed into whichever
mode is displayed.
5. See “Turning On the GX-6000” on page 25 for a description of the remainder of the
warm-up sequence keeping the following in mind:
a. Even if LUNCH BREAK is set to ON, the Lunch Break Screen will not appear.
b. The Low Alarm, High Alarm, STEL, and TWA screens will not appear.
c. If the IR CH4 or IR HC (whichever is installed) sensor fails, it is not possible to
enter Bar Hole Mode. Press and release the RESET button to acknowledge the
failure and return to the Normal Mode Select Screen. Replace the failed sensor.
If oxygen sensor fails, press RESET to acknowledge the failure and then
continue to Bar Hole Mode. No oxygen reading will be displayed. Replace the
failed sensor as soon as possible.
If any other installed sensor fails, press and release the RESET button to
acknowledge the failure and continue to Bar Hole Mode. Change the failed
sensor(s) as soon as possible for use in Normal Mode.
182 • Appendix D: Using the GX-6000 in Bar Hole Mode GX-6000 Operator’s Manual
6. The GX-6000 is now operating in Bar Hole Mode. The pump is off and the following
screen appears.
4:07
CH4
0
%LEL
O2
20.9%
VALVE 3
The combustible gas and oxygen channels are displayed along with the battery
charge level and station ID.
NOTE: If the oxygen sensor fails or if the oxygen channel is turned off, no oxygen
indication will display on the Bar Hole Mode screen.
GX-6000 Operator’s Manual Appendix D: Using the GX-6000 in Bar Hole Mode • 183
Performing a Bar Hole Test
1. Start up the GX-6000 as described in “Start Up, Bar Hole Mode” on page 181.
2. If you wish to change the station ID, press the SHIFT ▼ (PANIC) and DISP/LOCK
buttons and release when you hear a beep. The current station ID will be displayed
and it will be flashing. If you wish to exit the Station ID Select Screen without
making any changes, press and release the RESET button.
4:07
STATION ID
VALVE 3
3. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to scroll to the desired station ID and
press and release POWER/ENTER. The unit will return to the Bar Hole Measuring
Screen.
4:07
CH4
0
%LEL
O2
20.9%
VALVE 3
184 • Appendix D: Using the GX-6000 in Bar Hole Mode GX-6000 Operator’s Manual
5. Insert the probe into the bar hole and press and release the POWER/ENTER button.
The pump will turn on and the sample period will begin with the sample period
counting down in seconds in the lower middle portion of the display. “MEASURE”
will flash below the countdown. The combustible gas channel will automatically
switch between %LEL or %vol based on the level of gas present. The combustible
gas channel’s reading will also be displayed in graph format on the bottom of the
screen. If you wish to cancel the measurement, press and release the RESET button.
4:07
CH4
50 %LEL
O2
20.9%
30 SEC
MEASURE
6. At the end of the sample period, the pump will shut off and the buzzer will sound,
then the peak combustible gas reading and the minimum oxygen reading for the
sample period will be displayed. The peak combustible gas reading will also be
visually displayed on the graph along the bottom of the screen.
4:07
CH4
50 %LEL
O2
11.9%
PEAK
GX-6000 Operator’s Manual Appendix D: Using the GX-6000 in Bar Hole Mode • 185
• Press and release the DISP/LOCK button. The display will now indicate PURGE
along the bottom of the screen and the purge time will begin counting down from
30 seconds in the lower middle portion of the display. If you wish to cancel the
purge, press and release the RESET button.
4:07
CH4
0%LEL
O2
20.9%
30 SEC
PURGE
• When the purge is complete, the screen will return to the initial Bar Hole Mode
screen.
4:07
CH4
0%LEL
O2
20.9%
VALVE 3
8. If other bar holes will be tested, proceed to the next bar hole and repeat steps Step 5 -
Step 7.
9 . To cancel a bar hole measurement or fresh air purge that is in progress, press and
release the RESET button.
10 . To exit Leak Check Mode and enter another mode, press and hold the ▲AIR and
DISP/LOCK buttons for 5 seconds to access the Mode Select Screen.
Turning Off the GX-6000, Bar Hole Mode
1. Press and hold the POWER/ENTER button.
2. The unit will initiate a bar hole measurement. Keep holding the POWER/ENTER
button. The buzzer will pulse for about three seconds and TURN OFF will appear at
the bottom of the screen.
3. Release the button when TURN OFF disappears.
186 • Appendix D: Using the GX-6000 in Bar Hole Mode GX-6000 Operator’s Manual
Appendix E: Using the GX-6000 in Inert Mode
Inert Mode is used to measure the combustible gas and/or oxygen level in a purged
environment. In order for the instrument to operate in Inert Mode, the oxygen alarm must
be set to H-HH operation meaning that both alarms are increasing. The oxygen Low
Alarm High Alarm points are factory set at 5.0% and 10.0%.
It is recommended that an IR CH4 or IR HC sensor be installed if the instrument is going
to be used in Inert Mode since these sensors do not require oxygen to work properly. The
catalytic LEL sensor does not operate at oxygen concentrations below 10 %volume. If it is
necessary to use the catalytic LEL sensor while operating in Inert Mode, a dilution fitting
must be installed. Installing a dilution fitting will affect the oxygen reading since you’re
introducing oxygen into the sample.
NOTE: INERT MODE must be set to ON in Maintenance Mode for Inert Mode to be
an option in the Mode Select Screen at startup.
Alarms
The oxygen channel alarm points in Inert Mode are different from those in Normal Mode.
All other alarm points settings remain unchanged.
Since the application for Inert Mode is to detect a rising oxygen level in purged
environments, both oxygen alarms are rising in Inert Mode. The factory set alarm points
for Inert Mode are 5.0 %volume (Low Alarm) and 10.0 %volume (High Alarm). These
Inert Mode alarm points are user adjustable in Maintenance Mode.
1 . Follow the steps described in pg.150 to change the alarm points for Inert Mode. The
only difference between what’s described in that section and what the screen will
look like is that INERT will be an option in the Alarm Points Menu if INERT
MODE is set to ON in Maintenance Mode.
NOTE: If INERT MODE is set to OFF in Maintenance Mode, the INERT item will
not appear in the Alarm Points Menu.
ALARM - P
ALARM POINTS
CH4
O2
H2S
CO
VOC
INERT
ESCAPE
2. Select INERT (not O2) in order to view or change Inert Mode’s oxygen alarm
points.
GX-6000 Operator’s Manual Appendix E: Using the GX-6000 in Inert Mode • 187
Start Up and Operation
1. Connect the tapered rubber nozzle or the sample hose and probe to the GX-6000’s
quick connect inlet fitting.
2. Press and briefly hold down the POWER/ENTER button. Release the button when
you hear a beep.
3. The Normal Mode Select Screen displays. Use the ▲AIR or SHIFT ▼ (PANIC)
button to scroll to the Inert Mode Select Screen.
NOTE: INERT MODE must be set to ON in Maintenance Mode for Inert Mode to be
an option in the Mode Select Screen at startup.
4:07 4:07
MODE SELECT MODE SELECT
NORMAL MODE ▲ AIR
or INERT MODE
CH4 100 %LEL SHIFT ▼ (PANIC)
O2 40.0 %
H2S 100.0 ppm
CO 500 ppm
VOC 6000 ppm
4. With the Inert Mode Select Screen displayed, press and release the POWER/ENTER
button to begin the Inert Mode startup sequence.
NOTE: If no button is pressed for 20 seconds, the unit will proceed into whichever
mode is displayed.
5. The warm-up will proceed as described in “Turning On the GX-6000” on page 25.
6. Once the warm-up sequence has finished, the instrument will be operating in Inert
Mode. Since the oxygen concentration in fresh air is above both Inert Mode oxygen
alarm points, the GX-6000 will go into alarm if turned on in a fresh air environment.
4:07
CH4 O2
0
%LEL
0.5%
H2S CO
0.0
ppm
0
ppm
VOC
0
ppm
7. If you wish to change operating modes by returning to the Mode Select Screen, press
and hold the ▲AIR and DISP/LOCK buttons for 5 seconds.
188 • Appendix E: Using the GX-6000 in Inert Mode GX-6000 Operator’s Manual
Appendix F: ESS-03 Toxic Sensors
Overview
The ESS-03 sensors are used to monitor levels of a variety of toxic gases. This appendix
describes the GX-6000’s ESS-03 sensors. It also includes instructions to replace an ESS-
03 sensor. Table 19 below lists the available ESS-03 sensors.
Table 19: ESS-03 Sensor Specifications
Chlorine 0 - 10.0 ppm 0.05 ppm 0.50 ppm 1.00 ppm 1.00 ppm 0.50 ppm
(Cl2)
Hydrogen 0 - 15.0 ppm 0.1 ppm 5.00 ppm 10.0 ppm 4.7 ppm N/A
Cyanide
(HCN)
Nitrogen 0 - 20.00 ppm 0.05 ppm 3.00 ppm 6.00 ppm N/A 3.00 ppm
Dioxide
(NO2)
Sulphur 0 - 99.90 ppm 0.05 ppm 2.00 ppm 5.00 ppm 5.00 ppm 2.00 ppm
Dioxide
(SO2)
NOTE: If you are attempting to detect chlorine, be sure the hose/probe does not have a
particle filter installed. Also be sure that the hose is not longer than 30 feet.
Description
The ESS-03 is a smart sensor that stores sensor parameters including the target gas,
detection range, alarm points, and calibration settings in its memory. So a sensor can be
calibrated at the factory and shipped as a replacement sensor without the need to calibrate
the sensor when it is installed as long as it is installed during the sensor’s valid calibration
period which is typically 3 months. In addition, you can change an existing ESS-03
channel from one type of ESS-03 sensor to another and the GX-6000 will automatically
load all the sensor parameters and configure the ESS-03 channel for the new sensor. See
pg.198 for instructions to replace or change an ESS-03 sensor.
The ESS-03 sensor is a cylindrical sensor with a diffusion opening on the front and a
connector on the back. The ESS-03 sensor is installed in one of the two smart sensor
sockets and is held in place by the sensor gasket and the flow chamber.
Cl2 • HCl
• SO2
HCN • H2
• SO2
SO2 • H2
The table below indicates some of the gases that will cause a negative response and a
decreased reading for the affected sensor. For example, if you are attempting to detect SO2
but NO2 is also present, the instrument’s SO2 reading will be lower than the environment’s
actual SO2 level.
HCN • NO2
NH3 • HCl
• NO2
NO2 • CO
• SO2
SO2 • NO2
ESS-03 Calibration
An ESS-03 channel can be calibrated using the auto calibration method or the single
calibration method.
NOTE: ESS-03 calibration must be done with the hose and probe attached. If you are
calibrating a chlorine sensor, be sure the hose/probe does not have a particle
filter installed.
1. To calibrate the standard 4 sensors and/or the high range PID sensor, follow the
instructions outlined on pg.70.
NOTE: Be sure to calibrate the ESS-03 sensor with the hose and probe attached.
2. When you review the auto calibration values, confirm that the ESS-03’s auto
calibration gas value matches the calibration cylinder.
a. If you need to change an HCN channel’s auto calibration value, the procedure is
slightly different since PH3 can be used to calibrate the HCN channel. When you
move the cursor to HCN in the Concentration Screen and then press and release
the POWER/ENTER button, the Alternative Gas Screen will appear. The HCN
concentration will be flashing.
AUTO
CONCENTRATION
ALTERNATIVE GAS
PH3 / 9 PH3 concentration
1.1 ppm
HCN concentration
HCN 10.0 ppm
ESCAPE
b. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to change the HCN concentration.
As the HCN concentration changes, the PH3 concentration will also change. If
you are using HCN to calibrate the HCN sensor, adjust the HCN concentration
until it matches the concentration listed on your cylinder. If you are using PH3 to
calibrate the HCN sensor, adjust the HCN concentration until the PH3
concentration matches the concentration listed on your cylinder.
4. If you were using a 5-gas (standard 4-gas + ESS-03) calibration cylinder to calibrate
the standard 4 sensors, continue to Step 5.
If you were using a 5-gas (standard 4-gas + high range PID) calibration cylinder or a
4-gas calibration cylinder/high range PID-only calibration cylinder combination to
calibrate the standard 4 sensors and/or the high range PID sensor:
a. Remove the tubing from the rigid tube on the probe.
b. Unscrew the 5-gas (standard 4-gas + high range PID), 4-gas cylinder, or high
range PID cylinder from the regulator.
c. Screw the ESS-03 calibration cylinder onto the demand flow regulator. Continue
with Step 6.
5. If you were using a 5-gas (standard 4-gas + ESS-03) calibration cylinder to calibrate
the standard 4 sensors:
a. Leave the tubing connected to the rigid tube on the probe. Continue with Step 6.
6. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to move the cursor next to GAS
SELECT.
AUTO
AUTO CAL
CONCENTRATION
GAS SELECT
ESCAPE
▲ AIRor ▲ AIR
or
SHIFT ▼ (PANIC) SHIFT ▼ (PANIC)
AUTO AUTO
GAS SELECT GAS SELECT
▲ AIR
or
SHIFT ▼ (PANIC)
8. With the ESS-03 channel’s Gas Select Screen displayed, press and release the
POWER/ENTER button.
9. If you are calibrating anything other than an HCN sensor, continue to Step 12.
11 . Use the ▲AIR and SHIFT ▼ (PANIC) buttons to move the cursor in front of the gas
you will use for calibration then press and release the POWER/ENTER button.
12 . The Calibration In Process Screen for the ESS-03 channel will display. APPLYGAS
and the gas reading will be flashing.
APPLYGAS
HCN
0.0
ppm
If you do not want to proceed with the calibration, press and release the DISP/LOCK
button to return to the Gas Select Screen.
If you do want to continue with the calibration, proceed to the next step.
13 . If you are using a 5-gas (standard 4-gas + ESS-03) calibration cylinder, continue with
Step 14.
If you are using an ESS-03 only calibration cylinder:
a. Connect the tubing from the regulator to the rigid tube on the probe. Allow the
GX-6000 to draw gas for one minute.
b. Press and release the POWER/ENTER button to set the span adjustment for the
ESS-03 channel to the programmed value. Continue with Step 15.
PASS
CO
HCN
9.0
ppm
ESCAPE
POWER/ENTER
to acknowledge
16 . If the ESS-03 channel passed calibration and if no sensors are in a sensor life
warning condition or if SEN LIFE ALERT is set to OFF, the following screen
sequence occurs.
AUTO 4:07 AUTO
GAS SELECT
PASS
HCN
9.0
ppm
ESCAPE
APPLYGAS
HCN
FAIL
ppm
The buzzer and alarm LEDs activate in a double pulsing pattern. Press and release
the RESET button to reset the alarm and continue to the Gas Select Screen. Attempt
to calibrate the ESS-03 sensor again. If the failure continues, investigate the cause.
See “Troubleshooting” on page 102.
If SEN LIFE ALERT is set to ON (factory setting) and if any sensor is in a sensor
life warning condition, the following screen appears. The buzzer and LEDs activate
in a double pulsing pattern. Press and release POWER/ENTER to acknowledge the
sensor life warning and continue to the Gas Select Screen.
SEN LIFE
WARNING
CO
NOTE: RKI Instruments, Inc. recommends that you dedicate a regulator for use with
chlorine (Cl2) gas and that you do not use that dedicated regulator for any other
gases, particularly hydrogen sulfide (H2S).
NOTE: RKI Instruments, Inc. recommends that you dedicate a regulator for use with
chlorine (Cl2) gas and that you do not use that dedicated regulator for any other
gases, particularly hydrogen sulfide (H2S).
For example, if your instrument is a 4-gas + ESS-03 Cl2 + ESS-03 NH3, the Cl2 sensor
should be in the smart sensor 1 position and the NH3 sensor should be in the smart sensor
2 position. If your instrument is a 4-gas + ESS-03 NH3, the NH3 sensor should be in the
smart sensor 1 position and a dummy sensor should be installed in the smart sensor 2
position. See the figure below for the location of the smart sensor 1 and 2 positions.
Smart Sensor 1
Position
CO
Smart Sensor 2
Position H2 S
H2S Scrubber
SO2 Sensor
33-0176 Teflon filter for SO2 sensor (need to replace 33-0176 and 33-7135 at the same time)
33-7135 H2S scrubber disk for SO2 sensor (need to replace 33-0176 and 33-7135 at the same
time)
33-7135-05 H2S scrubber disk for SO2 sensor, pack of 5 (need to replace 33-0176 and 33-7135
at the same time)
80-0006-XXP Sample hose with integral probe, with hydrophobic filter and particle filter, no
scrubber section. For use with NH3, HCN, NO2, and SO2 sensors. Replace “XX”
with length in feet. 3 foot hose is standard. Available lengths for the GX-6000 are 3,
5, 10, 15, 20, 25, 30, 35, 40, 45, and 50 feet.
80-0006-XXP-01 Sample hose with integral probe, no particle filter, no scrubber section. For use
with Cl2 sensor. Replace “XX” with length in feet. 3 foot hose is standard.
Available lengths for the GX-6000 are 3, 5, 10, 15, 20, 25, and 30 feet.
80-0172 Probe, with hydrophobic filter and particle filter, no scrubber section. For use with
NH3, HCN, NO2, and SO2 sensors.
80-0172-01 Probe, with hydrophobic filter, no particle filter, no scrubber section. For use with
Cl2 sensor.
81-0142RK-02 Calibration cylinder, 5-gas (SO2, CH4, O2, H2S, CO), 58-liter aluminum
81-0142RK-04 Calibration cylinder, 5-gas (SO2, CH4, O2, H2S, CO), 34-liter aluminum
Overview
The infrared sensors are used to monitor levels of methane, combustible gas, and carbon
dioxide. This appendix describes the GX-6000’s infrared sensors and includes instructions
to use a GX-6000 that has one or more infrared sensors installed. It also includes
instructions to replace an infrared sensor. The table below lists the available infrared
sensors along with their specifications.
Table 24: Infrared Sensor Specifications
IR HC Target Gases
The infrared HC sensor is a general hydrocarbon sensor. It is setup for and factory-
calibrated to isobutane.
The infrared HC sensor is known to not or to poorly respond to the following combustible
gases:
• acetylene
• benzene
• hydrogen
IR CH4 Target Gases
The infrared methane sensor is setup for and factory-calibrated to methane. There are
other gases that the sensor will still detect and respond to. There are also gases that the
methane sensor will not detect or respond to. Lists of the gases falling in each of these
respective categories can be found below. Consult RKI Instruments, Inc. for combustible
gases not listed below.
The infrared methane sensor is known to respond to the following combustible gases:
• acetone
• ethane
• ethanol
Description
IR Sensor
The IR sensor is a smart sensor that stores sensor parameters including the target gas,
detection range, alarm points, and calibration settings in its memory. So a sensor can be
calibrated at the factory and shipped as a replacement sensor without the need to calibrate
the sensor when it is installed as long as it is installed during the sensor’s valid calibration
period which is typically 6 months. In addition, you can change an existing IR channel
from one type of IR sensor to another and the GX-6000 will automatically load all the
sensor parameters and configure the IR channel for the new sensor. See “Replacing an IR
Sensor” on page 210 for instructions to replace or change an IR sensor.
The IR sensor is a cylindrical sensor with a diffusion opening on the front and a connector
on the back. The IR sensor is installed in one of the two smart sensor sockets and is held in
place by the sensor gasket and the flow chamber.
Plug Plug
To GX-6000
• When you perform a demand zero during start up, operation, or calibration, you must
use the CO2 scrubber provided with the instrument to remove background CO2 from
the air being sampled. See “Performing a Demand Zero for Carbon Dioxide Sensors”
below.
If you have a PID sensor in your instrument, see “Performing a Demand Zero for
Instruments with a Carbon Dioxide Sensor and a PID Sensor” on page 209.
• Since there is a background of CO2 in air, do not use the AUTO ZERO feature that
can be turned on and off in Maintenance Mode. The factory setting for the feature is
OFF.
Performing a Demand Zero for Carbon Dioxide Sensors
When setting the zero reading, the carbon dioxide scrubber shipped with the GX-6000
allows you to eliminate carbon dioxide normally found in fresh air. To perform a demand
zero, do the following:
1 . Remove the tubing from the side of the scrubber that does not have an arrow pointing
to it.
2 . Connect the tubing to the GX-6000’s inlet fitting. Be sure the arrow on the filter is
pointing toward the inlet fitting.
CO2 Scrubber
To GX-6000
To GX-6000
NOTE: Do not store the CO2 scrubber and VOC zero filter connected together. They
must be separated and appropriately plugged for storage.
NOTE: If there is a PID sensor installed in your instrument and you are calibrating an
IR CH4 sensor with %vol concentrations, either calibrate the PID sensor first or
allow enough time for the %vol CH4 gas to flush out before calibrating the PID
sensor.
IR Bump Testing
1. See “Performing a Bump Test” on page 89 for instructions to perform a bump test.
Replacing an IR Sensor
When replacing a sensor, you may either replace it with another of the same sensor or you
may install a different IR sensor. If a different one is installed, the GX-6000 will load the
sensor parameters and configure the IR channel for the new sensor.
1 . Follow the instructions listed in “Replacing a Sensor” on page 115.
Be sure that the new sensor is installed in the same position as the old sensor and that the
following sensor priority for the smart sensor 1 position is maintained.
Table 26: Priority for Smart Sensor 1 Position
For example, if your instrument is a 4-gas + high range PID + IR HC, the high range PID
sensor should be in the smart sensor 1 position and the IR HC sensor should be in the
smart sensor 2 position. If your instrument is a 4-gas + IR HC, the IR HC sensor should be
in the smart sensor 1 position and a dummy sensor should be installed in the smart sensor
2 position. See the figure below for the location of the smart sensor 1 and 2 positions.
CO
Smart Sensor 2
Position H2S
Parts List
Table 27: Infrared Sensors Parts List
33-6092 VOC zero filter, charcoal, with tubing stubs and plugs (for instruments that
include an IR CO2 sensor and a PID sensor)
Overview
This appendix describes the GX-6000’s 10.0 eV/benzene sensor (RKI part number PID-
003L).
The 10.0 eV/benzene PID sensor detects VOCs in the 0-100.00 ppm range when used in
Measuring Mode. It is optimized for benzene detection. The standard calibration for the
10.0 eV/benzene PID sensor is to benzene but it can be factory setup for and calibrated to
other gases. Regardless of the calibration gas, the 10.0 eV/benzene sensor will still detect
and respond to a variety of volatile organic compounds (VOCs) while used in Measuring
Mode. Even though the 10.0 eV/benzene sensor can be used for general VOC detection, if
your application is for general VOC detection and is not related to monitoring for benzene
specifically, RKI Instruments, Inc. recommends that the low range or high range 10.6 eV
PID sensor be used instead.
The 10.0 eV/benzene PID sensor detects VOCs in the 0-50.00 ppm range when used in
Benzene Select Mode. The filter tube that needs to be used in Benzene Select Mode allows
benzene to pass through but scrubs out all other VOCs except for nitrobenzene,
cyclohexane, n-octane, and n-pentane. Common applications for the 10.0 eV/benzene PID
sensor include process monitoring where only benzene is present or petroleum industry
monitoring where other VOCs may be present.
See the table below for 10.0 eV/benzene sensor specifications.
Table 28: 10.0 eV/Benzene PID Sensor Specifications
Sensor Description
The PID sensor is a cylindrical sensor with a diffusion opening on the front and 3 pins on
the back. It is installed in a white housing that has three sockets on the bottom that mate
with the GX-6000 instrument. The PID sensor must always be installed in the first smart
sensor position which is located in the top left corner of the sensor block.
Cap
Tip-Breaking Feature
Tip Reservoir
Breakaway Feature
Tube Holder with Tube
GAS
Exploded View
H2Sppm CO
ppm
P 10.0 eV
BNZppm
Display Mode
Display Mode for a 10.0 eV/benzene PID instrument appears and functions as described in
“Chapter 4: Display Mode” on page 47 with the following exceptions:
• A Benzene Select Mode screen is now the first item in Display Mode. Benzene Select
Mode is described in “Benzene Select Mode” on page 217.
DISP
BENZENE SELECT
YES :ENTER
NO :DISP
EXIT :RESET
10.0 eV
CHANGE :ENTER
KEEP :DISP
EXIT :RESET
Measuring Mode
The 10.0 eV/benzene sensor detects VOCs in the 0-100.00 ppm range in Measuring Mode.
See “Measuring Mode, Normal Operation” on page 33 for a complete description of
Measuring Mode.
If the 10.0 eV/benzene PID sensor is one of two PID sensors installed in your instrument,
as shipped from the factory, the 10.0 eV/benzene PID channel will say “BNZ” and the low
or high range PID channel will say “VOC”.
If you change the target gas of the 10.0 ev/benzene PID sensor to something other than
benzene in Display Mode, both channels will appear as “VOC”. In that case, the channel
that appears first will be based on the following sensor priority.
High Priority 10.6 eV low range PID
10.0 eV/benzene PID
11.7 eV PID
NOTE: If you have changed the factory-set channel assignments using the
CONFIGURE CH menu item in User Mode, the order of displayed sensors
may be different and you will need to remember how you set your instrument
up.
NOTE: A PID-003L 10.0 eV/benzene sensor must be installed in order for the Benzene
Select Mode screen to appear.
A tube holder and filter tube are necessary for monitoring an area in Benzene Select
Mode.
Be sure to perform an appropriate calibration for Benzene Select Mode use as described
on pg.225 before continuing.
1. Enter Display Mode by pressing the DISP/LOCK button while in Measuring Mode.
The Benzene Select Mode Screen appears.
DISP
BENZENE SELECT
YES :ENTER
NO :DISP
EXIT :RESET
2. Press and release the POWER/ENTER button to enter Benzene Select Mode.
INSTALL TUBE
AND HOLDER
YES :ENTER
NO :DISP
4. Be sure a filter tube is installed in the tube holder and then attach the tube holder to
the instrument’s inlet fitting. See pg.235 for instructions.
5. Press and release the POWER/ENTER button to confirm the tube and holder have
been installed.
6. You will be prompted to start a benzene measurement.
TUBE
START
MEASUREMENT
YES :ENTER
NO :DISP
7. Press and release the POWER/ENTER button to start the benzene measurement.
8. The benzene reading is displayed at the top of the screen. An automatically-
determined, temperature-based countdown is displayed at the bottom of the screen.
The number to the left of the countdown (“5” in the example below) is the
temperature code.
TUBE
BNZ
0.00 ppm
MEASURING
EXIT :RESET
5 45SEC
YES :ENTER
NO :DISP
43SEC
• Press the POWER/ENTER button to confirm the measurement cancel.
• Press the DISP/LOCK button to return to the measuring screen without canceling
the measurement.
10 . Once the measurement is over, the final reading is displayed and you are asked if you
want to repeat the measurement.
TUBE
BNZ
0.00 ppm
REPEAT ?
YES :ENTER
NO :DISP
STEL :RESET
• To repeat a measurement, continue to Step 11.
• To not repeat a measurement and return to Measuring Mode, continue to Step 12.
• To start a STEL measurement, continue to Step 13.
a. You will be prompted to change the tube before continuing. See pg.235 for
instructions.
TUBE
CHANGE TUBE
YES :ENTER
NO :DISP
EXIT :RESET
b. Once the tube is changed, press and release the POWER/ENTER button to start
the measurement. Return to Step 8.
12 . To not start another measurement and to return to Measuring Mode, press and release
the DISP/LOCK button.
a. You will be prompted to remove the tube holder from the instrument’s inlet
fitting.
TUBE
REMOVE HOLDER
YES :ENTER
NO :DISP
START STEL
MEASUREMENT
YES :ENTER
NO :DISP
a. Leave the tube holder installed but there is no need to change the tube.
b. Press and release the POWER/ENTER button to start a STEL measurement.
c. The STEL measurement takes 15 minutes and a countdown displays at the
bottom of the screen. The STEL reading is displayed at the top of the screen and
the instantaneous reading is displayed below it.
TUBE
BNZ STEL
0.00 ppm
0.00 ppm
STEL MEASURING
EXIT :RESET
900SEC
d. To cancel the STEL measurement, press and release the RESET button. Then
press and release the POWER/ENTER button.
TUBE
BNZ STEL
0.00 ppm
0.00 ppm
CANCEL ?
YES :ENTER
NO :DISP
YES :ENTER
NO :DISP
REMOVE HOLDER
YES :ENTER
NO :DISP
YES :ENTER
NO :DISP
EXIT :RESET
3. Press and release the POWER/ENTER button. The screen that appears will depend
on whether or not any data has been saved to the snap logger.
5/20/2015
NO DATA MM/DD/YYYY
3:40
NUMBER 001
TANK 5
SHIFT ▼ SHIFT ▼
▲ AIR ▲ AIR
REC DATA REC DATA
DATE BNZ
5/20/2015
MM/DD/YYYY
3.20 ppm Benzene
POWER/ Select Mode
3:45
ENTER Standard Data
BENZENE SELECT
NUMBER 002
TANK 5
SHIFT ▼ SHIFT ▼
▲ AIR ▲ AIR
REC DATA REC DATA
DATE CH4 O2
5/20/2015
0
%LEL
20.9%
MM/DD/YYYY POWER/ H2S CO Measuring
3:40
ENTER 0.0
ppm
0
ppm
Mode
Data
NUMBER 001 VOC
TANK 5 0
ppm
Option 1 1. AUTO CAL or SINGLE CAL with benzene, without tube and
Benzene Select Mode holder
benzene detection 2. TUBE CAL with benzene, with tube and holder
Option 2 1. AUTO CAL or SINGLE CAL with isobutylene (or other
target gas)**
2. enter CAL CODE
Measuring Mode, AUTO CAL with benzene (no TUBE CAL or CAL CODE required)
benzene detection
Measuring Mode, Option 1 AUTO CAL or SINGLE CAL with benzene
general VOC
detection* Option 2 AUTO CAL or SINGLE CAL with isobutylene (or other target
gas)**
* Even though it is possible to use the 10.0 eV/benzene PID sensor for general VOC detection, RKI
Instruments, Inc. recommends using the low range or high range 10.6 eV sensor instead.
** In order to calibrate to a gas other than benzene or isobutylene, you must select that other gas for the 10.0
eV/benzene sensor in the PID Gas Name Screen in Display Mode as described in
“PID Gas Name Screen” on page 48. You must also select the appropriate calibration gas during the
calibration procedure.
If you change the target gas of the 10.0 ev/benzene PID sensor to something other than
benzene in Display Mode, both channels will appear as “VOC”. In that case, the
channel that appears first will be based on the following sensor priority. Be sure to
select the correct PID sensor for calibration.
High Priority 10.6 eV low range PID
10.0 eV/benzene PID
11.7 eV PID
NOTE: If you have changed the factory-set channel assignments using the
CONFIGURE CH menu item in User Mode, the order of displayed sensors
may be different and you will need to remember how you set your instrument
up.
If you change the target gas of the 10.0 ev/benzene PID sensor to something other than
benzene in Display Mode, both channels will appear as “VOC”. In that case, the
channel that appears first will be based on the following sensor priority. Be sure to
select the correct PID sensor for calibration.
NOTE: If you have changed the factory-set channel assignments using the
CONFIGURE CH menu item in User Mode, the order of displayed sensors
may be different and you will need to remember how you set your instrument
up.
NOTE: If you begin to use filter tubes from a new box and if the calibration code shown
on the new box is different than the calibration code shown on the old box, you
must perform a TUBE CAL operation using one of the tubes from the new box.
To calibrate the 10.0 eV/benzene sensor in TUBE CAL, you will need:
• A benzene calibration cylinder. RKI Instruments, Inc. recommends a concentration of
5 ppm.
• A demand-flow regulator to provide adequate sample gas flow.
• Non-absorbent tubing.
• Tube holder and filter tube.
Be sure that you are doing a complete calibration as appropriate for your application and
for the calibration gas you have available (see pg.225).
2. Use the SHIFT ▼ (PANIC) button to scroll to TUBE CAL and then press and release
the POWER/ENTER button.
3. The pump will stop and the instrument will prompt you to install the filter tube and
tube holder.
CAL
INSTALL TUBE
AND HOLDER
YES :ENTER
NO :DISP
4. Be sure a filter tube is installed in the tube holder and then attach the tube holder to
the instrument’s inlet fitting. See pg.235 for instructions.
5. Press and release the POWER/ENTER button to enter the TUBE CAL Menu.
TUBE CAL
GAS CALIBRATION
AUTO CAL
SINGLE CAL
ESCAPE
6. Use the SHIFT ▼ (PANIC) button to place the cursor next to AUTO CAL.
8. To view or change the benzene concentration to be used for the TUBE CAL:
a. Use the SHIFT ▼ (PANIC) button to place the cursor next to
CONCENTRATION.
b. Press and release the POWER/ENTER button. The current concentration will be
shown and it will be flashing.
TUBE CAL
CONCENTRATION
5 45SEC
14 . If the calibration passes, the screen will indicate PASS and will return you to the
TUBE CAL Auto Calibration Menu.
15 . If the calibration fails, a failure screen will appear and the instrument will go into
alarm.
TUBE CAL
BNZ
FAIL ppm
Press and release the RESET button to clear the alarm and return to the TUBE CAL
Auto Calibration Menu. See “Troubleshooting” on page 102.
16 . Use the SHIFT ▼ (PANIC) button to place the cursor next to ESCAPE, then press
and release the POWER/ENTER button to return to the TUBE CAL Menu.
REMOVE HOLDER
YES :ENTER
NO :DISP
2. Use the SHIFT ▼ (PANIC) button to place the cursor next to TUBE CAL and then
press and release the POWER/ENTER button.
INSTALL TUBE
AND HOLDER
YES :ENTER
NO :DISP
4. Be sure a filter tube is installed in the tube holder and then attach the tube holder to
the instrument’s inlet fitting. See pg.235 for instructions.
5. Press and release the POWER/ENTER button to enter the TUBE CAL Menu.
TUBE CAL
GAS CALIBRATION
AUTO CAL
SINGLE CAL
ESCAPE
6. Screw the demand flow regulator onto the benzene calibration cylinder.
7. Attach the calibration tubing to the demand flow regulator.
8. Attach the other end of the calibration tubing to the tube holder’s inlet fitting.
9. Use the SHIFT ▼ (PANIC) button to place the cursor next to SINGLE CAL.
10 . Press and release the POWER/ENTER button.
5 45SEC
12 . At the end of the countdown, use the ▲AIR or SHIFT ▼ (PANIC) button to adjust
the reading to match the concentration listed on the calibration cylinder.
13 . Press and release the POWER/ENTER button.
14 . If the calibration passes, the screen will indicate PASS and will return you to the
TUBE CAL Menu.
If the reading cannot be adjusted to match the cylinder concentration, see
“Troubleshooting” on page 102.
15 . Use the SHIFT ▼ (PANIC) button to place the cursor next to ESCAPE, then press
and release the POWER/ENTER button. You will be prompted to remove the filter
tube and tube holder.
TUBE CAL
REMOVE HOLDER
YES :ENTER
NO :DISP
NOTE: If you begin to use filter tubes from a new box and if the calibration code shown
on the new box is different than the calibration code shown on the old box, you
must enter the new calibration code before using the new filter tubes.
2. Use the SHIFT ▼ (PANIC) button to place the cursor next to CAL CODE.
3. Press and release the POWER/ENTER button.
4. The current calibration code will be displayed and it will be flashing.
CAL CODE
INPUT
CAL CODE : C
Maintenance
Replacing a Sensor
See “Replacing a Sensor” on page 115 for sensor replacement instructions.
PID Sensor Maintenance
See pg.119 for PID sensor maintenance instructions keeping in mind the following
differences in spare parts.
CAUTION: The aluminum oxide cleaning kit described on pg.119 is only for use with
the 10.0 eV and 10.6 eV PID sensors’ lamps. DO NOT USE THE
ALUMINUM OXIDE CLEANING KIT TO CLEAN THE 11.7 eV PID
SENSOR’S LAMP!
Replacing a Tube
When to Replace Tube
RKI Instruments, Inc. recommends that the tube be replaced before starting a new
measurement in Benzene Select Mode. If discoloration of the filter material extends
beyond the “3/4” line (into the side without the text and arrow), the tube needs to be
replaced. Even though discoloration can be a measure of when to replace the tube, not all
gases cause the filter material to discolor which is why RKI Instruments, Inc. recommends
that the tube be replaced before starting a new measurement in Benzene Select Mode.
If the tubes are being stored in a refrigerator, take them out of the refrigerator and let them
sit overnight before using them.
Pull
GAS
GAS
Locking Feature
13-0550 Lanyard for tube holder (for new style tube holder with knurled tip-
breaker)
13-0700 Knurled tip-breaker on tube holder (for new style tube holder with knurled
tip-breaker)
13-0701 Black cap for knurled tip-breaker on tube holder (for new style tube holder
with knurled tip-breaker)
33-0563 PID sensor electrode stack, 10.0 eV/benzene sensor
33-7128 Filter tubes, 1 box, 10 tubes per box
33-7128-05 Filter tubes, 5 boxes, 10 tubes per box
51-1503 PID sensor replacement lamp, 10.0 eV/benzene sensor
80-0174 Tube holder
81-0100RK-03 Calibration cylinder, benzene, 5 ppm in air, 103 liter
81-0100RK-04 Calibration cylinder, benzene, 5 ppm in air, 34 liter aluminum
81-0104RK-03 Calibration cylinder, isobutylene, 10 ppm in air, 103 liter
81-0104RK-04 Calibration cylinder, isobutylene, 10 ppm in air, 34 liter aluminum
81-6XAX-DLV Calibration kit: 34 liter 10 ppm IBL in air cylinder, demand flow regulator,
calibration tubing
81-6XZX Calibration kit: 103 liter 5 ppm benzene in air cylinder, demand flow
regulator, calibration tubing
81-6XZX-DLV Calibration kit: 34 liter 5 ppm benzene in air cylinder, demand flow
regulator, calibration tubing
82-0300RK Aluminum oxide powder PID lamp cleaning kit, with electrode stack
removal tool, for 10.0 eV and 10.6 eV PID sensors’ lamps ONLY
Overview
This appendix describes the GX-6000’s 11.7 eV PID sensor (RKI part number PID-004).
The 11.7 eV PID sensor detects VOCs in the 0-1000 ppm range. The standard calibration
for the 11.7 eV PID sensor is to isobutylene but it can be factory setup for and calibrated to
other gases. Regardless of the calibration gas, the 11.7 eV PID sensor will still detect and
respond to a variety of volatile organic compounds (VOCs) while used in Measuring
Mode.
The 11.7 eV PID sensor is capable of detecting VOCs that the 10.6 eV low range and high
range PID sensors are not, such as:
• acrylonitrile
• carbon tetrachloride
• chloroform
• methanol
• refrigerants (R-32A, R-143)
See the table below for 11.7 eV PID sensor specifications.
Table 30: 11.7 eV PID Sensor Specifications
H2Sppm CO
ppm
P 11.7 eV
VOC
ppm
Display Mode
Display Mode for a 11.7 eV PID instrument appears and functions as described in
“Chapter 4: Display Mode” on page 47 with the following exception:
• The PID Gas Name Screen operates as described on pg.48 but appears as shown
below. If there are 2 PID sensors installed in your GX-6000, the PID Gas Name
Screens will appear in Display Mode in the following order regardless of their position
in the flow system: 10.0 eV/benzene, 10.6 eV, 11.7 eV.
DISP
SELECT PID LIST
PID SET TO
Isobutylene
11.7 eV
CHANGE :ENTER
KEEP :DISP
EXIT :RESET
NOTE: If you have changed the factory-set channel assignments using the
CONFIGURE CH menu item in User Mode, the order of displayed sensors
may be different and you will need to remember how you set your instrument
up.
• When turning off the GX-6000 after performing a calibration, follow the instructions
on pg.242.
NOTE: If you have changed the factory-set channel assignments using the
CONFIGURE CH menu item in User Mode, the order of displayed sensors
may be different and you will need to remember how you set your instrument
up.
• When turning off the GX-6000 after performing a calibration, follow the instructions
on pg.242.
Maintenance
Replacing the Dehumidifier Filter
RKI Instruments, Inc. recommends replacing the dehumidifier filter when it looks wet.
Replacing a Sensor
See pg.115 for sensor replacement instructions.
Replacing the PID Sensor’s Lamp and Electrode Stack
RKI Instruments, Inc. recommends that you replace the electrode stack every time you
replace the lamp. However, if you replace only the lamp and notice that the instrument is
operating properly, it is not necessary to change the electrode stack.
See pg.123 and pg.127 for PID sensor maintenance instructions keeping in mind the
following differences in spare parts.
CAUTION: The aluminum oxide cleaning kit described on pg.119 is only for use with
the 10.0 eV and 10.6 eV PID sensors’ lamps. DO NOT USE THE
ALUMINUM OXIDE CLEANING KIT TO CLEAN THE 11.7 eV PID
SENSOR’S LAMP!
Cleaning Kit
The 11.7 eV PID sensor cannot be cleaned with the same cleaning kit as the 10.0 eV and
10.6 eV PID sensors!
A special cleaning kit for the 11.7 eV PID sensor is not included with the instrument. See
Table 31 on page 246 for the 11.7 eV lamp cleaning kit ordering information.
The 11.7 eV PID sensor lamp cleaning kit includes the following items:
• an electrode stack removal tool
• a small vial of anhydrous methanol (you may purchase your own anhydrous
methanol but do NOT use anything with water in it)
• 10 cotton swabs
• 10 finger cots
When using the anhydrous methanol, try to minimize the vial’s exposure to air by leaving
the cap on as much as possible. Discard the anhydrous methanol 3 months after opening or
after leaving the cap off for an extended period of time.
CAUTION: Using the anhydrous methanol more than 3 months after opening it or after
the cap has been off for an extended period of time may damage the 11.7 eV
lamp.
07-0220 Rubber cap, black, for dehumidifier filter and GX-6000 inlet fitting
33-0564 PID sensor electrode stack, 11.7 eV sensor
33-2125-01 Dehumidifier filter with tubing
51-1504 PID sensor replacement lamp, 11.7 eV sensor
81-0103RK-03 Calibration cylinder, isobutylene, 100 ppm in air, 103 liter
81-0103RK-04 Calibration cylinder, isobutylene, 100 ppm in air, 34 liter aluminum
81-0143RK-02 Calibration cylinder, 5-gas (Proprietary blend of IBL, CH4, O2, H2S, CO),
58-liter aluminum
81-0143RK-04 Calibration cylinder, 5-gas (Proprietary blend of IBL, CH4, O2, H2S, CO),
34-liter aluminum
81-6ABX-DLV Calibration kit: 34 liter five-gas (IBL/CH4/O2/H2S/CO) cylinder, demand
flow regulator, calibration tubing
82-0301 Anhydrous methanol PID lamp cleaning kit, with electrode stack removal
tool, for 11.7 eV sensor’s lamp ONLY
PID-004 PID sensor, 11.7 eV
Overview
This appendix explains how to set up a User List for the GX-6000’s PID sensor(s). This
User List appears in the PID Gas Name Screen in the GX-6000’s Display Mode. The PID
User List is not related to the GX-6000’s User ID list.
Materials
In order to set up the GX-6000’s PID User List, you will need:
• IrDA/USB cable
• GX-6000
• PID Gas List Editor program
• computer with Windows 7, Windows 8, or Windows 10
Procedure
Installing the PID Gas List Editor Program
1. Double click the setup.exe file.
2. Follow the Setup Wizard prompts to install the program.
Connecting a GX-6000
1. Plug your IrDA/USB cable into your computer.
2. Launch the PID Gas List Editor Program (appears as “GX-6000PID” in the program
list).
3. Align the IrDA/USB cable with the GX-6000’s IrDA window (located directly below
the POWER/ENTER button).
4. Press and release the POWER/ENTER button to turn on the GX-6000.
5. When a connection is established, the GX-6000’s LCD will say “PC TRANSMIT”.
The PID Gas List Editor Program does not show any connection indication.
10.6 eV .pgl
10.0 eV .pxl
11.7 eV .pul
Absorbent cartridges
This warranty covers instruments and parts sold to users only by authorized distributors,
dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of
this gas monitor and our warranty is limited to replacement of parts or our complete
goods.