NX - Installation Planning 3.0
NX - Installation Planning 3.0
NX - Installation Planning 3.0
Order-No.: DD+DIS440.05E
*1Y4EJ61*
1 Piece Y4EJ6 MA1
NX
Type 4406/0303
Revision History
Revision Change Date Author Approver
1 Release of NX 1.0 Service Pack 1 2006/08 MG GK
“Hardware Prerequisites”: added UPS types
“ID Tablet”: Updated firmware level
“Digitizers”: Gridline suppression, added CR 35-X and CR 85-X
0 Initial Release 2006/04 MG GK
Chapter 3
List of Contents
1 Dimensions ...................................................................................1
5 Connectivity Aspects...................................................................7
5.1 General............................................................................................................. 7
5.2 IHE Profiles...................................................................................................... 7
5.3 ADB-tool .......................................................................................................... 8
5.4 Viewing Priors ................................................................................................. 8
5.5 Form to fill in the Network Parameters ......................................................... 9
7 Hospital Guidelines....................................................................12
7.1 General security guidelines ......................................................................... 12
7.1.1 Installing Antivirus Software on all Workstations ............................................ 12
7.1.2 Using Software Restriction Policies ................................................................ 12
7.1.3 Setting Account Lockout Policy....................................................................... 12
7.1.4 Disabling EFS (Encrypting File System) ......................................................... 13
7.1.5 Enabling O/S Auditing..................................................................................... 13
9 NX Licenses................................................................................ 16
9.1 NX Base License ...........................................................................................16
9.1.1 Worklist Overview............................................................................................16
9.1.2 Examination.....................................................................................................16
9.1.3 Viewing & Editing.............................................................................................17
9.1.4 Main Menu.......................................................................................................17
9.2 Product License: License 1 ..........................................................................17
9.3 DICOM store: License 1.1 .............................................................................18
9.4 DICOM Print: License 1.2..............................................................................18
9.5 Export: License 1.3........................................................................................19
9.6 Priors: License 1.5 ........................................................................................19
9.7 MPPS: License 1.6.........................................................................................19
9.8 Full Leg - Full Spine: License 1.7.................................................................20
9.9 MUSICA1: License 1.8...................................................................................20
9.10 MUSICA1 Advanced Processing: License 1.8.1 .........................................20
9.11 MUSICA1 Uro-Tomo: License 1.8.2..............................................................21
9.11.1 MUSICA1 Dental: License 1.8.3......................................................................21
9.11.2 MUSICA1 Paediatric: License 1.8.4 ................................................................21
9.12 MUSICA2: License 1.9...................................................................................22
9.12.1 MUSICA2 Advanced Processing: License 1.9.1 .............................................22
9.12.2 MUSICA2 Thorax: License 1.9.2 .....................................................................23
9.12.3 MUSICA2 Paediatric: License 1.9.3 ................................................................23
9.12.4 MUSICA2 Mammo: License 1.9.4 ...................................................................23
9.12.5 MUSICA2 Radiotherapy: License 1.9.5...........................................................23
9.13 Tools and Annotations: License 1.11 ..........................................................24
9.13.1 Advanced Tools and Annotations: License 1.11.1 ..........................................24
9.13.2 Advanced Measurements: License 1.11.2 ......................................................24
9.13.3 QA Tools: License 1.11.6 ................................................................................25
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
1 Dimensions
Platform dimensions:
Height: 41.14 cm, 16.2”
Width: 18.96 cm; 7.3”
Depth: 43.18 cm; 17”
Monitor dimensions:
BARCO MFCD 1219:
Height: 46.2 cm
Width: 45 cm
Depth: 21.4 cm
Net Weight 7.7 kg
2 Software Prerequisites
3 Hardware Prerequisites
NOTE:
• NX can be installed on hardware reused from previous CR QS installations.
• DX-S can only be connected to NX if a firewire connection is available on the pc.
The DELL 620 platform used for CR QS does not support firewire!
NOTE:
NX does not support grayscale monitors.
Coloured annotations are always displayed in colour on the NX workstation.
Supported monitors:
• Basic 17” LCD (Dell Standard Monitor)
• High Quality Barco MFCD 1219
• Touch Screen Barco MFCD 1219 TS
NOTE:
NX works with a fixed screen resolution (1280 x 1024).
Deviating from this resolution will cause the NX Viewer not to start up and an error
message will be displayed.
NOTE:
Possibility of connection problems, if there is a router or firewall between the segments
of the devices:
• NX workstation
• Digitizer
• Printers
• PACS
• RIS
When <ping> does not work, ask the hospital system administrator, for an
access-list or security rule which is blocking the traffic between the devices.
2. Click <Properties>.
figure 1
3. Click <Configure>.
figure 2
figure 3
5 Connectivity Aspects
5.1 General
5.2 IHE Profiles
5.3 ADB-tool
Information about the Agfa DICOM Bridge tool can be found on
MedNet / Healthcare Library:
General Info / Connectivity & Application / To Agfa / User Manual / Agfa DICOM Bridge-
User Guide
NOTE:
This functionality is only available if WEB1000 is ready for it.
The preconditions and commercial and technical procedures are described in:
"HEALTHCARE SERVICES EUROPE - NX - WEB1000 Integration:
Viewing Prior Studies - Commercial Procedure and Technical Documentation"
and is available on MedNet / Healthcare Library:
-> Professional Services – EMEA
-> Integration Services
-> Clinical Viewing Integration & Consulting
NOTE:
It is not allowed to give hostnames with an underscore.
NX Workstation
System Type
Computer Name
Workgroup Name
IP Address
Subnet Mask
Default Gateway
Windows License Key
6.3 Certificates
In case the hospital wants to work with SSL-communciation, some actions are imminent:
• You should ask whether the hospital has its own CA (Certification authority), or
whether Agfa needs to supply certificates for its devices.
• Try to collect the certificates of the devices with which the NX system will need to
communicate prior to the installation. If not, it is possible that you’ll loose a lot of
time during the installation because you cannot test the SSL connection unless you
have all certificates.
• Also, provide the NX system cerficate up-front to the hospital (or in case they have
their own CA, ask them to generate the NX system certificate).
• Arrange the hospital personnel or Agfa service in such a way that either the NX
certificate is installed on all systems with which NX needs to communicate or make
sure that somebody is present at the time of installation that is able to do this (and
has the permissions to do this).
7 Hospital Guidelines
7.2.4 Replacing the "Everyone" Group with "Authenticated Users" on File Shares
"Everyone" in the context of Windows XP security, means anyone who gains access to
your network can access the data. Never assign the "Everyone" Group to have access to
a file share on your network, use "Authenticated Users" instead. This is especially
important for printers, who have the "Everyone" Group assigned by default.
8 Release Environment
8.1 ID Tablet
The ID Tablet firmware that needs to be installed on the ID Tablet is "1.12.07" or higher.
NOTE:
In case the installed ID Tablet firmware is lower than version “1.12.07”, stability
problems with the ID Tablet like
• a lost connection between workstation and ID Tablet or
• not recognized cassettes
are to be expected.
8.2 Digitizers
Gridline Suppression
For the ADC Compact Plus, CR25.0, CR75.0, CR 35-X and CR 85-X digitizers the use of
Standard Resolution cassettes in combination with Gridline Suppression is not supported.
9 NX Licenses
9.1.2 Examination
The examination pane allows you to compose an exam for one patient with one or more
exposures and link that data with the patient data. It is possible to edit the patient or
exposure data, add or copy an exposure or reject an image. The preview shows the
images as they are digitised. When finished they can be rotated or flipped. You can add a
‘left’- or ‘right’-marker. It is possible to have more than one patient active and to toggle
between them.
NOTE:
This base configuration does not include:
• printing
• exporting
• sending
of images to any destination without at least one of the concerned licenses.
NOTE:
Automatic archiving of rejected images is not part of this license.
See 9.15.
Enable: Default window level must be disabled when this license is active
Disable: Default window level is enabled when this licence is inactive
NOTE:
This license is not available for NX 1.0.
The MUSICA2 Radiotherapy license gives the ability to use the MUSICA2 Radiotherapy
processing parameters.
When activated it will be possible to:
• Process exposure types with the MUSICA2 Radiotherapy processing parameters.
Disable: When this license is disabled, the dose statistics will be saved. The
rationale behind this is that the dose statistics should be available
immediately upon enabling the license. Therefore the task will always
be executed in the background, independent of the activation of the
license.
NOTE:
This license is not available for NX 1.0.
When choosing a site for the CR QS System, observe the following principles for
positioning monitors:
NOTE:
If the customer wants to work with P-values, the ambient light conditions in the room
have to be considered when setting up the monitor.
For the functional description of P-values refer to chapter 2 in this service manual.
For the configuration of P-values on the QS system refer to chapter 3.7 in the service
manual.
figure 4
figure 5
figure 6
figure 7
NOTE:
For detailed information on ergonomic working conditions, please consult the
respective local regulations.
Published by