Chapter 3 (Oscilloscope Function Generator)
Chapter 3 (Oscilloscope Function Generator)
The x-axis represents time, and the y-axis represents an instantaneous view of the
voltage of the input signal. To allow viewing signals across a wide frequency range,
the rate and speed at which the sweep of the x-axis occurs is configurable. The
sensitivity of the inputs can also be configured to accept signals from microvolt’s
peak-to-peak to many thousands of volts peak-to-peak.
Its functions are :
a) Shows and calculate the frequency and amplitude of an oscillating signal.
b) Shows the voltage and time of a particular signal.
c) Helps to troubleshoot any malfunction components of a project by looking at the
expected output after a particular component.
d) Shows the content of the AC voltage or DC voltage in a signal.
Some of the common waveforms that are measured using an oscilloscope are as
shown below.
Type of Oscilloscope
1. Analog
In an analog scope, the x-axis is controlled by an internal time base, and the y-axis is
directly controlled by the input signal. Analog uses continuously variable voltages.
Analog oscilloscope works by directly applying a voltage being measured to
an electron beam moving across the oscilloscope screen. The voltage deflects the
beam up and down proportionally, tracing the waveform on the screen.
2. Digital
In a digital model, the input voltage is sampled at a preset frequency. The x-axis
represents the samples along a timeline, and the y-axis shows the voltage levels of
each sample. Digital uses discrete binary numbers that represent voltage samples.
When probe touch the component , the signal travels directly to the vertical
deflection plates of the cathode ray tube (CRT). Voltage applied to these deflection
plates causes a glowing dot to move. (An electron beam hitting phosphor inside the
CRT creates the glowing dot.) A positive voltage causes the dot to move up while a
negative voltage causes the dot to move down.
The signal also travels to the trigger system to start or trigger a “horizontal sweep.”
Horizontal sweep is a term referring to the action of the horizontal system causing
the glowing dot to move across the screen. Triggering the horizontal system causes
the horizontal time base to move the glowing dot across the screen from left to right
within a specific time interval. Many sweeps in rapid sequence cause the movement
of the glowing dot to blend into a solid line. At higher speeds, the dot may sweep
across the screen up to 500,000 times each second.
Together, the horizontal sweeping action and the vertical deflection action traces a
graph of the signal on the screen. The trigger is necessary to stabilize a repeating
signal. It ensures that the sweep begins at the same point of a repeating signal,
resulting in a clear picture as shown in following figure. The attenuation or
amplification of the signal.
Use the volts/div control to adjust the amplitude of the signal before it is applied to
the vertical deflection plates.
• The time base. Use the sec/div control to set the amount of time per division
represented horizontally across the screen.
• The triggering of the oscilloscope. Use the trigger level to stabilize a repeating
signal, as well as triggering on a single event.
Also, adjusting the focus and intensity controls enables you to create a sharp, visible
display.
EE 101 MEASUREMENT Oscilloscope and Signal Generator / 4
Some of the systems that make up digital oscilloscopes are the same as those in
analog oscilloscopes; however, digital oscilloscopes contain additional data
processing systems. With the added systems, the digital oscilloscope collects data for
the entire waveform and then displays it.
When attach a digital oscilloscope probe to a circuit, the vertical system adjusts the
amplitude of the signal, just as in the analog oscilloscope.
Next, the analog-to-digital converter (ADC) in the acquisition system samples the
signal at discrete points in time and converts the signal’s voltage at these points to
digital values called sample points. The horizontal system’s sample clock determines
how often the ADC takes a sample. The rate at which the clock “ticks” is called the
sample rate and is measured in samples per second.
The sample points from the ADC are stored in memory as waveform points. More than
one sample point may make up one waveform point.
Together, the waveform points make up one waveform record. The number of
waveform points used to make a waveform record is called the record length. The
trigger system determines the start and stop points of the record. The display
receives these record points after being stored in memory.
Depending on the capabilities of your oscilloscope, additional processing of the
sample points may take place, enhancing the display. Pre-trigger may be available,
allowing you to see events before the trigger point. Fundamentally, with a digital
oscilloscope as with an analog oscilloscope, you need to adjust the vertical,
horizontal, and trigger settings to take a measurement.
EE 101 MEASUREMENT Oscilloscope and Signal Generator / 5
1. Display Controls
The display is usually a CRT or LCD panel which is laid out with both horizontal
and vertical reference lines referred to as the graticule. In addition to the screen,
most display sections are equipped with three basic controls, a focus knob, an
intensity knob and a beam finder button.
2. Vertical Controls
The vertical section controls the amplitude of the displayed signal. This section
carries a Volts-per-Division (Volts/Div) selector knob, an AC/DC/Ground selector
switch and the vertical (primary) input for the instrument. Additionally, this section
is typically equipped with the vertical beam position knob.
3. Horizontal Controls
The horizontal section controls the time base or “sweep” of the instrument. The
primary control is the Seconds-per-Division (Sec/Div) selector switch. Also
included is a horizontal input for plotting dual X-Y axis signals. The horizontal
beam position knob is generally located in this section.
4. Trigger Controls
The trigger section controls the start event of the sweep. The trigger can be set to
automatically restart after each sweep or it can be configured to respond to an
internal or external event. The principal controls of this section will be the source
and coupling selector switches. An external trigger input (EXT Input) and level
adjustment will also be included.
EE 101 MEASUREMENT Oscilloscope and Signal Generator / 6
A test probe (test lead, test prod, or scope probe) is a physical device used to connect
electronic test equipment to the device under test (DUT).
Peak-to-peak Voltage
10 Div
1 cycle
= 1 Peak Voltage
a. Refer figure 3.10, calculate the voltage peak-to-peak with the formula below.
If the setting VOLT/DIV = 1V.
Voltage peak–to-peak = Volt/Div X Div
Example :
Voltage peak–to-peak = Volt / Div X Div
= 1 Volt / Div X 10 Div
= 10 Vp-p
(T)
Ti
a. Refer figure 3.11, calculate the frequency with the formula below. If the setting
TIME/DIV = 0.2ms.
Where , T = Time/div X Div
So,
Example :
= 500Hz
Function generators are typically used in simple electronics repair and design;
where they are used to stimulate a circuit under test. A device such as an
oscilloscope is then used to measure the circuit's output. Function generators
vary in the number of outputs they feature, frequency range, frequency accuracy
and stability, and several other parameters.
Most function generators allow the user to choose the shape of the output from a
small number of options.
• Square wave - The signal goes directly from high to low voltage.
• Sine wave - The signal curves like a sinusoid from high to low voltage.
• Triangle wave - The signal goes from high to low voltage at a fixed rate.
The direct current (DC) offset control on a function generator varies the average
voltage of a signal relative to the ground.
EE 101 MEASUREMENT Oscilloscope and Signal Generator / 10
The frequency control of a function generator controls the rate at which output
signal oscillates. On some function generators, the frequency control is a
combination of different controls. One set of controls chooses the broad
frequency range (order of magnitude) and the other selects the precise frequency.
This allows the function generator to handle the enormous variation in frequency
scale needed for signals.
Function or Waveform
Group Offset Output Group
2. Output Group:
• The group consists of the attenuation buttons and the fused 50 ohm BNC
connector.
• The attenuation buttons are used to attenuate (decrease) the amplitude of the
signal by a factor measured in decibels.
• The fused 50 ohm BNC connector is used for connecting the function generator to
other equipment. The most common connection used to connect the function
generator to other equipment is a cable with a BNC connector on one end and
two alligator clips on the other end.
3. Amplitude Modulation
• Amplitude knob is used to increase the amplitude of waveform.
4. Sweep Group
• Sweep is used to set the frequency output.
Signal Generator
Osilloscope
volt/div volttime/div
/div
1. Connect the oscilloscope probe and signal generator, refer figure 3.14 and 3.15.
2. Switch on the signal generator and adjust the output level to produce a visible signal
on the oscilloscope screen.
3. Adjust TIME/DIV and VOLTS/DIV to obtain a clear display and investigate the
effects of pressing the waveform shape buttons.
4. The rotating FREQUENCY control and the RANGE switch are used together to
determine the frequency of the output signal.