Analytical Method Validation - Pharmaceutical Guidelines
Analytical Method Validation - Pharmaceutical Guidelines
1. Principle
1.1 This appendix presents some information on the characteristics that should be considered during validation of analytical methods. Approaches
other than those specified in this appendix may be followed and may be acceptable. Manufacturers should choose the validation protocol and
procedures most suitable for testing of their product.
1.2 The manufacturer should demonstrate (through validation) that the analytical procedure is suitable for its intended purpose.
1.3 Analytical methods, whether or not they indicate stability, should be validated.
1.4 The analytical method should be validated by research and development before being transferred to the quality control unit when appropriate.
2. General
2.1 There should be specifications for both, materials and products. The tests to be performed should be described in the documentation on standard
test methods.
2.2 Specifications and standard test methods in pharmacopoeias (“pharmacopoeial methods”), or suitably developed specifications or test methods
(“non-pharmacopoeial methods”) as approved by the national drug regulatory authority may be used.
2.3 Well-characterized reference materials, with documented purity, should be used in the validation study.
2.4 The most common analytical procedures include identification tests, assay of drug substances and pharmaceutical products, quantitative tests for
content of impurities and limit tests for impurities. Other analytical procedures include dissolution testing and determination of particle size.
2.5 The results of analytical procedures should be reliable, accurate and reproducible. The characteristics that should be considered during validation
of analytical methods are discussed in paragraph 6.
2.6 Verification or revalidation should be performed when relevant, for example, when there are changes in the process for synthesis of the drug
substance; changes in the composition of the finished product; changes in the analytical procedure; when analytical methods are transferred from
one laboratory to another; or when major pieces of equipment instruments change.
2.7 The verification or degree of revalidation depend on the nature of the change(s).
2.8 There should be evidence that the analysts, who are responsible for certain tests, are appropriately qualified to perform those analyses (“analyst
proficiency”).
3. Pharmacopoeial methods
3.1 When pharmacopoeial methods are used, evidence should be available to prove that such methods are suitable for routine use in the laboratory
(verification).
3.2 Pharmacopoeial methods used for determination of content or impurities in pharmaceutical products should also have been demonstrated to be
specific with respect to the substance under consideration (no placebo interference).
4. Non-pharmacopoeial methods
4.1 Non-pharmacopoeial methods should be appropriately validated.
5. Method validation
5.1 Validation should be performed in accordance with the validation protocol. The protocol should include procedures and acceptance criteria for all
characteristics. The results should be documented in the validation report.
5.2 Justification should be provided when non-pharmacopoeial methods are used if pharmacopoeial methods are available. Justification should
include data such as comparisons with the pharmacopoeial or other methods.
5.3 Standard test methods should be described in detail and should provide sufficient information to allow properly trained analysts to perform the
analysis in a reliable manner. As a minimum, the description should include the chromatographic conditions (in the case of chromatographic tests),
reagents needed, reference standards, the formulae for the calculation of results and system suitability tests.
https://www.pharmaguideline.com/2010/12/analytical-method-validation.html#gsc.tab=0 1/3
6/15/2020 Analytical Method Validation : Pharmaceutical Guidelines
— linearity
— range
— accuracy
— precision
— detection limit
— quantitation limit
— robustness.
6.1.1 Accuracy is the degree of agreement of test results with the true value, or the closeness of the results obtained by the procedure to the true
value. It is normally established on samples of the material to be examined that have been prepared to quantitative accuracy. Accuracy should be
established across the specified range of the analytical procedure.
Note: it is acceptable to use a “spiked” placebo where a known quantity or concentration of a reference material is used.
6.1.2 Precision is the degree of agreement among individual results. The complete procedure should be applied repeatedly to separate, identical
samples drawn from the same homogeneous batch of material. It should be measured by the scatter of individual results from the mean (good
grouping) and expressed as the relative standard deviation (RSD).
6.1.2.1 Repeatability should be assessed using a minimum of nine determinations covering the specified range for the procedure e.g. three
concentrations/ three replicates each, or a minimum of six determinations at 100% of the test concentration.
6.1.2.2 Intermediate precision expresses within-laboratory variations (usually on different days, different analysts and different equipment).
If reproducibility is assessed, a measure of intermediate precision is not required.
6.1.2.3 Reproducibility expresses precision between laboratories.
6.1.3 Robustness (or ruggedness) is the ability of the procedure to provide analytical results of acceptable accuracy and precision under a variety of
conditions. The results from separate samples are influenced by changes in the operational or environmental conditions. Robustness should be
considered during the development phase, and should show the reliability of an analysis when deliberate variations are made in method parameters.
6.1.3.1 Factors that can have an effect on robustness when performing chromatographic analysis include:
— stability of test and standard samples and solutions;
— reagents (e.g. different suppliers);
— different columns (e.g. different lots and/or suppliers);
— extraction time;
— variations of pH of a mobile phase;
— variations in mobile phase composition;
— temperature; and
— flow rate.
6.1.4 Linearity indicates the ability to produce results that are directly proportional to the concentration of the analyte in samples. A series of samples
should be prepared in which the analyte concentrations span the claimed range of the procedure. If there is a linear relationship, test results should
be evaluated by appropriate statistical methods. A minimum of five concentrations should be used.
6.1.5 Range is an expression of the lowest and highest levels of analyte that have been demonstrated to be determinable for the product. The
specified range is normally derived from linearity studies.
6.1.6 Specificity (selectivity) is the ability to measure unequivocally the desired analyte in the presence of components such as excipients and
impurities that may also be expected to be present. An investigation of specificity should be conducted during the validation of identification tests, the
determination of impurities and assay.
6.1.7 Detection limit (limit of detection) is the smallest quantity of an analyte that can be detected, and not necessarily determined, in a quantitative
fashion. Approaches may include instrumental or non-instrumental procedures and could include those based on:
— visual evaluation;
— signal to noise ratio;
— standard deviation of the response and the slope;
— standard deviation of the blank; and
— calibration curve.
6.1.8 Quantitation limit (limit of quantitation) is the lowest concentration of an analyte in a sample that may be determined with acceptable accuracy
and precision. Approaches may include instrumental or non-instrumental procedures and could include those based on:
— visual evaluation;
— signal to noise ratio;
— standard deviation of the response and the slope;
— standard deviation of the blank; and
— calibration curve.
https://www.pharmaguideline.com/2010/12/analytical-method-validation.html#gsc.tab=0 2/3
6/15/2020 Analytical Method Validation : Pharmaceutical Guidelines
6.2 Characteristics (including tests) that should be considered when using different types of analytical procedures are summarized in Table 1.