User Manual CA8335

Download as pdf or txt
Download as pdf or txt
You are on page 1of 116

C.

A 8335
THREE-PHASE electrical
net w o r k s A N A LY S E R +
QUALISTAR

ENGLISH User’s manual


Thank you for purchasing a C.A 8335 three-phase electrical networks analyser (Qualistar+). To obtain the best service from
your unit:
„„ read these operating instructions carefully,
„„ comply with the precautions for use.

WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger symbol appears.

Equipment protected by double insulation.

USB socket.

Earth.

The CE marking indicates conformity with European directives, in particular LVD and EMC.

Chauvin Arnoux has adopted an Eco-Design approach in order to design this appliance. Analysis of the complete
lifecycle has enabled us to control and optimize the effects of the product on the environment. In particular this ap-
pliance exceeds regulation requirements with respect to recycling and reuse.

The rubbish bin with a line through it indicates that, in the European Union, the product must undergo selective
disposal in compliance with Directive WEEE 2002/96/EC. This equipment must not be treated as household waste.

Definition of measurement categories:


„„ Measurement category IV corresponds to measurements taken at the source of low-voltage installations.
Example: power feeders, counters and protection devices.
„„ Measurement category III corresponds to measurements on building installations.
Example: distribution panel, circuit-breakers, machines or fixed industrial devices.
„„ Measurement category II corresponds to measurements taken on circuits directly connected to low-voltage installations.
Example: power supply to domestic electrical appliances and portable tools.

Precautions for use

This instrument and its accessories comply with safety standards IEC 61010-1, IEC 61010-031, and IEC 61010-2-032 for voltages
of 600V in category IV or 1000V in category III.
Failure to observe the safety instructions may result in electric shock, fire, explosion, and destruction of the instrument and of
the installations.
„„ The operator and/or the responsible authority must carefully read and clearly understand the various precautions to be taken
in use. Sound knowledge and a keen awareness of electrical hazards are essential when using this instrument.
„„ If you use this instrument other than as specified, the protection it provides may be compromised, thereby endangering you.
„„ Do not use the instrument on networks of which the voltage or category exceeds those mentioned.
„„ Do not use the instrument if it seems to be damaged, incomplete, or poorly closed.
„„ Before each use, check the condition of the insulation on the leads, housing, and accessories. Any item of which the insulation
is deteriorated (even partially) must be set aside for repair or scrapping.
„„ Use only the leads and accessories supplied. Using leads (or accessories) of a lower voltage or category reduces the voltage
or category of the combined instrument + leads (or accessories) to that of the leads (or accessories).
„„ Use personal protection equipment systematically.
„„ Keep your hands away from the terminals of the device.
„„ When handling the leads, test probes, and crocodile clips, keep your fingers behind the physical guard.
„„ Use only the mains power adaptor and battery pack supplied by the manufacturer. They include specific safety features.
„„ Some current sensors must not be placed on or removed from bare conductors at hazardous voltages: refer to the sensor
manual and comply with the handling instructions.

2
Contents

1. Getting started.........................................................4 10.4. Starting energy metering....................................74


1.1. Unpacking..............................................................4 10.5. Disconnection of energy metering......................75
1.2. Charging the battery...............................................5 10.6. Reset of energy metering...................................75
1.3. Choice of language................................................5 11. Screen snapshot mode.......................................76
2. Description of the device.....................................6 11.1. Screen snapshots...............................................76
2.1. Functions................................................................6 11.2. Handling of screen snapshots............................76
2.2. Overall view............................................................8 12. Help key ....................................................................77
2.3. On/Off switch.........................................................8 13. Data export software........................................78
2.4. Display ...................................................................9 14. General specifications .....................................79
2.5. Keypad keys.........................................................10 14.1. Environmental conditions...................................79
2.6. Connectors ..........................................................12 14.2. Mechanical conditions.......................................79
2.7. Power supply........................................................12 14.3. Compliance with international standards...........79
2.8. The stand..............................................................13 14.4. Electromagnetic compatibility............................80
2.9. Abbreviations........................................................13 14.5. Power supply......................................................80
3. Use ................................................................................15 15. Functional characteristics ...........................81
3.1. Start-up................................................................15 15.1. Reference Conditions.........................................81
3.2. Configuration........................................................15 15.2. Nominal current according to type of sensor.....81
3.3. Installation of leads...............................................16 15.3. Electrical characteristics ...................................81
3.4. Functions of the device........................................18 16. Appendices...............................................................91
4. Configuration .........................................................19 16.1. Mathematical formulae ......................................91
4.1. Configuration menu..............................................19 16.2. Distribution sources supported by the device.106
4.2. Display language..................................................19 16.3. Hysteresis.........................................................106
4.3. Date/Time.............................................................19 16.4. Minimum scale values for waveforms and
4.4. Display..................................................................20 minimum RMS values.......................................106
4.5. Calculation methods............................................21 16.5. Four-quadrant diagram ...................................107
4.6. Connection...........................................................24 16.6. Mechanism for triggering transient captures ..107
4.7. Sensors and ratios...............................................28 16.7. Capture conditions in Inrush Current mode.....107
4.8. Capture Mode .....................................................29 16.8. Glossary...........................................................108
4.9. Trend mode..........................................................31 17. MAINTENANCE .........................................................111
4.10. Mode Alarm mode..............................................33 17.1. Cleaning the casing..........................................111
4.11. Erase memory....................................................34 17.2. Maintenance of sensors...................................111
4.12. About..................................................................35 17.3. Replacing the battery.......................................111
5. Waveform Capture.................................................36 17.4. Replacing the screen film.................................112
5.1. Transient mode.....................................................36 17.5. Memory card....................................................113
5.2. Inrush current mode.............................................39 17.6. Metrological check...........................................113
6. Harmonic ...................................................................44 17.7. Repair...............................................................113
6.1. Phase-to-neutral voltage......................................44 17.8. Updating of the internal software.....................113
6.2. Current..................................................................45 18. Warranty ...............................................................114
6.3. Apparent power....................................................46 19. To order..................................................................115
6.4. Phase-to-phase voltage.......................................47 19.1. C.A 8335 three-phase electrical networks
6.5. Expert mode.........................................................48 analyser............................................................115
7. Waveform...................................................................50 19.2. Accessories......................................................115
7.1. Measurement of true RMS value..........................50 19.3. Spare parts.......................................................115
7.2. Measurement of total harmonic distortion...........52
7.3. Measurement of the peak factor..........................53
7.4. Measurement of extreme and mean voltage and
current..................................................................54
7.5. Simultaneous display...........................................56
7.6. Display of Fresnel diagram...................................58
8. Alarm mode................................................................60
8.1. Alarm mode configuration ...................................60
8.2. Programming an alarm campaign........................60
8.3. Display of the list of campaigns...........................61
8.4. Display of list of alarms........................................61
8.5. Deleting an alarm campaign.................................62
8.6. Erasing all alarm campaigns ...............................62
9. Trend mode ...............................................................63
9.1. Programming and starting recording....................63
9.2. Trend mode configuration....................................63
9.3. Viewing the recording list.....................................64
9.4. Deleting recordings..............................................64
9.5. Viewing the records..............................................64
10. Power and energy mode....................................71
10.1. 3L filter................................................................71
10.2. Filters L1, L2 and L3...........................................72
10.3. Filter Σ ................................................................73

3
1. Getting started

1.1. Unpacking
FICHE DE SÉCURITÉ DU C.A 6116 (FR)

Vous venez d’acquérir un contrôleur d’installation C.A 6116 et nous vous remercions de votre
confiance.
Pour obtenir le meilleur service de votre appareil :
 lisez attentivement cette notice de fonctionnement,
 respectez les précautions d’emploi.

ATTENTION, risque de DANGER !


L’opérateur s’engage à consulter la présente notice à chaque fois que ce symbole de
danger est rencontré.

Appareil protégé par une isolation double.


➂ ➀
Polarité du connecteur d’alimentation en tension continue.


Le marquage CE atteste la conformité aux directives européennes.

La poubelle barrée signifie que, dans l’Union Européenne, le produit doit faire l’objet

x5 x5
d’un tri sélectif des déchets pour le recyclage des matériels électriques et électroniques
conformément à la directive WEEE 2002/96/EC.

Définition des catégories de mesure :


 La catégorie de mesure IV correspond aux mesurages réalisés à la source de l’installation basse
tension.
 La catégorie de mesure III correspond aux mesurages réalisés dans l’installation du bâtiment.
 La catégorie de mesure II correspond aux mesurages réalisés sur les circuits directement branchés
à l’installation basse tension.
 La catégorie de mesure I correspond aux mesurages réalisés sur des circuits non reliés directement
au réseau.

PRÉCAUTIONS D’EMPLOI
Cet appareil est protégé contre des tensions accidentelles n’excédant pas 600 V par rapport à la terre en
catégorie de mesure III ou 300 V par rapport à la terre en catégorie de mesure IV. La protection assurée
par l’appareil peut-être compromise si celui-ci est utilisé de façon non spécifiée par le constructeur.

x5
 Respectez la tension et l’intensité maximales assignées ainsi que la catégorie de mesure.


 Ne dépassez jamais les valeurs limites de protection indiquées dans les spécifications.
 Respectez les conditions d’utilisation, à savoir la température, l’humidité, l’altitude, le degré de pollution
et le lieu d’utilisation.
 N’utilisez pas l’appareil ou ses accessoires s’ils paraissent endommagés.
 Pour le recharge de la batterie, utilisez uniquement le bloc adaptateur secteur fourni avec
l’appareil.


 Utilisez des accessoires de branchement dont la catégorie de surtension et la tension de service
sont supérieures ou égales à celles de l’appareil de mesure (600 V CAT III).
 Toute procédure de dépannage ou de vérification métrologique doit être effectuée par du personnel
compétent et agréé.
 Utilisez les moyens de protection adaptés.

04 - 2009
Code 691923A01 - Ed. 1


ATTESTATION DE VERIFICATION
CHECKING ATTESTATION
190, rue Championnet
75876 PARIS Cedex 18 Numéro de l'appareil :
FRANCE Equipment number

Type / Model :

Désignation de l'instrument :
Instrument designation

Vérifié par : Signature :


Tested by Signature

Établi en usine, ce document atteste que le produit ci-dessus a été vérifié et est conforme aux
conditions d'acceptation définies dans nos procédures de fabrication et de contrôle.

Tous les moyens de mesure et d'essai utilisés pour vérifier cet appareil sont raccordés aux


étalons nationaux et internationaux soit par l'intermédiaire d'un de nos laboratoires de métrologie
accrédités COFRAC soit par un autre laboratoire accrédité.


Après sa mise en service, cet instrument doit être vérifié à intervalle régulier
auprès d'un service de métrologie agréé.
Pour tout renseignement veuillez contacter notre service après vente et d'étalonnage.

At the time of manufacture, this document certifies that the above product have been verified and
complies with acceptance conditions defined in our manufacturing and testing procedures.

Every test or measuring equipment used to verify this instrument are related to national
and international standards through one of our laboratories of metrology certified by french COFRAC
equivalent to NAMAS in the UK or through another certified laboratory.

After being in use, this instrument must be recalibrated within regular intervals
by an approved metrology laboratory. Please contact our after sales and calibration department:

Service après vente et d'étalonnage TEL: +33 (2) 31 64 51 55 FAX: +33 (2) 31 64 51 72
After sales and calibration department e-mail: [email protected]
WEB : www.manumesure.com
www.chauvin-arnoux.com

ATTESTATION DE CONFORMITE
COMPLIANCE ATTESTATION

Nous certifions que ce produit a été fabriqué conformément aux spécifications


techniques de constuction applicables.

We certify that this product is manufactured in accordance with applicable


constructing specifications.

907 009 119 - 02/03

➃ ➄

No. Designation Quantity

1 Safety cables, black, banana-banana, straight-straight 5

2 Black crocodile clips. 5

3 User’s manual on CD-ROM. 1

4 Type A-B USB cord. 1

5 Specific mains power unit and mains cord. 1

6 No. 22 carrying bag 1

7 Sets of inserts and rings for marking the leads and current sensors according to phase. 12

8 Checking attestation. 1

9 Safety data sheets. 5

10 Power Analyser Transfer (PAT) software on CD-ROM. 1

4
1.2. Charging the battery
Before the first use, start by fully charging the battery.

120 V ± 10 %, 60 Hz
230 V ± 10 %, 50 Hz

Remove the cover from the receptacle and connect the plug
of the specific power supply unit to the device. Connect the
mains cord to the power supply unit and to mains.

The button lights; it will go out only when the plug is


disconnected.

When the battery is fully discharged, charging takes approximately 5 hours.

1.3. Choice of language


Before using the device, first choose the language in which you want the device to display its messages.

Press the green button to switch the device on.

Press the Configuration key.

Press the yellow key on the device corresponding to the desired language.

This key is used to go to the next page.

Figure 8: Configuration screen

5
2. Description of the device

2.1. Functions
The C.A. 8335 (Qualistar+) is a three-phase network analyzer with colour graphic display and built-in rechargeable battery.

It plays three roles, and can be used:


„„ to measure the RMS values, powers, and perturbations of electric distribution networks.
„„ to deliver a snapshot of the principal characteristics of a three-phase network
„„ to track the variations of various parameters over time.

The measurement uncertainty of the device is better than 1% (not counting the uncertainties due to the current sensors). The
device is also very flexible, with a choice of sensors allowing measurements ranging from a few milliamperes (MN93A) to several
kiloamperes (AmpFLEX™).

The device is compact and impact resistant.

The ergonomics and simplicity of its interface make using it pleasant.

The C.A 8335 is intended for the technicians and engineers of electrical installation and network inspection and maintenance teams.

2.1.1. Measurement functions


The principal measurements made are:
„„ The RMS values of AC voltages up to 1000 V between terminals. By using the ratios, the device can measure voltages up to
hundreds of gigavolts.
„„ The RMS values of AC currents up to 6500 amperes (neutral included). By using the ratios, the device can measure currents
up to hundreds of kiloamperes.
„„ The DC components of voltages and currents (neutral included).
„„ Minimum and maximum half-cycle RMS voltage and current values (excluding neutral).
„„ Peak voltage and current values (neutral included).
„„ The frequency of 50 Hz and 60 Hz networks.
„„ Current and voltage peak factors (neutral included).
„„ Calculation of the harmonic loss factor (FHL), application to transformers in the presence of harmonic currents.
„„ Calculation of the K factor (KF), application to transformers in the presence of harmonic currents.
„„ Measurement of total harmonic distortion with respect to the fundamental (THD in % f) of the current and of the voltages
(excluding neutral).
„„ Measurement of the total harmonic distortion with respect to the RMS AC value (THD in % r) for the current and the voltages
(neutral included)
„„ Active, reactive (capacitive and inductive), non-active, distortion, and apparent power, by phase and cumulative (excluding
neutral).
„„ Power factors (PF) and displacement factors (DPF or cos Φ) (excluding neutral).
„„ Measurement of the RMS distortion values (d) for the current and the voltages (excluding neutral).
„„ Short-term voltage flicker (PST) (excluding neutral).
„„ Measurement of the long-term flicker of the voltages (PLT) (excluding neutral).
„„ Active, reactive (capacitive and inductive), non-active, distortion, and apparent energy (excluding neutral).
„„ Current and voltage harmonics (excluding neutral) up to order 50: RMS value, percentage referred to the fundamental, (%f)
(excluding neutral), or the total RMS value (%r), minimum and maximum and sequence harmonics.
„„ Apparent power of each harmonic up to order 50 (excluding neutral): value, percentage referred to the fundamental apparent
power (%f) or the total apparent power (%r), minimum and maximum.
„„ Inrush currents, starting of motors.

6
2.1.2. Display functions
„„ Display of waveforms (voltages and currents).
„„ Inrush Current function: displays parameters useful for study of the starting of a motor.
„„ Instantaneous current and voltage at the instant designated by the cursor.
„„ Maximum instantaneous absolute value of the current and of the voltage (over the entire starting time).
„„ RMS value of the half-cycle (or lobe) of the current and voltage (excluding neutral) on which the cursor is positioned.
„„ Maximum half-cycle RMS current (over the entire starting time).
„„ Instantaneous network frequency at the instant designated by the cursor.
„„ Maximum, mean, and minimum network frequencies (over the entire starting time).
„„ Time at which starting of motor commenced.
„„ Screen captures (50 maximum).
„„ T
ransients function. Detection and recording of transients (up to 210) between user-defined start and stop dates and times.
Recording of 4 complete cycles (one before the triggering event and three after) in the 8 acquisition channels.
„„ Trend recording (data logging) function. 2GB memory with date-stamping and user-defined start and stop dates for record-
ing, with a maximum of 100 recordings. Display, in bar chart or curve form, of the means of many parameters vs. time, with
or without minima and maxima.
„„ Alarm function. List of recorded alarms (up to 16,000) exceeding thresholds defined in the configuration menu. User-defined
alarm monitoring start and stop times.

2.1.3. Configuration functions


„„ Date and time settings.
„„ Screen brightness and contrast settings.
„„ Choice of curve colours.
„„ Choice of management of switching off of the screen.
„„ Choice of calculation methods (non-active quantities broken down or not, choice of the unit of energy, choice of the coefficients
of calculation of the K factor, choice of reference for the level of harmonics, PLT calculation (sliding or not).
„„ Choice of distribution system (single-phase, two-phase, three-phase with or without neutral) and of the connection method
(standard, 2-element method or 2½-element method).
„„ Configuration of recording, alarms, inrush currents, and transients.
„„ Erasure of data (total or partial).
„„ Display of software and hardware version numbers.
„„ Choice of language.
„„ Display of current sensors detected or simulated (2-element connection method) and voltage and current ratio settings.

7
2.2. Overall view

Measurement connection
terminals (see §2.6.1)

Display
(see §2.4)
USB socket
Function keys (see §2.6.2)
(yellow keys)
(see §2.5.1)
Connector for the
mains power unit/bat-
Return /previous key tery charger
(see §2.5.2) (see §2.6.2)

Configuration key
(see §2.5.4) Confirm/Enter key
(see §2.5.2)

Screen snapshot key Navigation keys


(see §2.5.4) (see §2.5.2)

Help key Mode keys


(see §2.5.4) (violet keys)
(see §2.5.3)

On/Off switch
(see §2.3)

Figure 1: Overall view of Qualistar+

2.3. On/Off switch


Pressing the button powers up the device.

The device can operate either on its battery or on mains power.

Pressing the switch again turns the device off. If the device is recording, metering energy, or searching for transients, alarms,
and/or inrush current acquisition, it requests confirmation.

Select Yes or No on the corresponding yellow keys, then press the  key to validate.
„„ If No is selected, recording will continue.
„„ If Yes is selected, the data recorded until that point are saved and the device is turned off.

8
2.4. Display
2.4.1. Presentation
The backlit 320x240 (1/4 VGA) pixel graphic liquid crystal screen displays all measurements with their curves, the parameters of
the unit, the curves selected, the instantaneous values of the signals, and the type of measurement selected. When the device is
powered up, it automatically displays the Waveform screen. Information about this screen can be found in §7.

Battery charge level.


Reminder of the mode.
Current date and time.

Active mode screen. Frequency calculated over one


second.

Function keys.

Figure 2: example of a display screen

The management of switching off of the screen is chosen by the user in the Display Configuration Mode menu (see §4.4.3).

2.4.2. The function key icons


The display uses the following icons for the yellow function keys:

Icons Designation Icons Designation

V Phase-to-neutral voltage mode. Zoom in.


A Phase-to-neutral current mode. Zoom out.
VA Power mode.
Management of the contrast and brightness.
U Phase-to-phase voltage mode.
Choice of colours of the measurement channels.
VAR Management of the breakdown of the non-active
quantities. Management of the switching off of the screen
Wh Choice of unit of energy.
Recording programming mode.
FK Choice of coefficients of the K factor.
Recording look-up mode.
%f-%r Choice of reference for the level of harmonics
of the phases. Start of recording.
PLT Management of the long-term flicker calculation
Rapid programming and start of recording.
mode.
CF Display of the peak factors and of the curves. Disconnection of recording.

RMS Display of the RMS values and of the curves. Shut down function in progress prompt.
PEAK Display of the PEAK values and of the curves. Bin for deletions of elements.
THD Display of the level of harmonic distortion and Shortcut to the recording parameterizing mode
of the curves
PF… Display of PF, cos F (DPF), tan F, and F. Activate/deactivate selection of the transients
list display filter.
W… Display of powers and of the associated quanti-
ties (PF, cos F, DPF, tan F and FVA).
Wh… Display of the energy meters.
Activation and de-activation of the energy calcula-
tion.

9
Icons Designation Icons Designation

Display of mean values and extrema. Select all items.


Move the cursor to the first occurrence of the Unselect all items.
maximum phase-to-neutral voltage.
Move the cursor to the first occurrence of the Transient mode.
minimum phase-to-neutral voltage.
Inrush current mode.
Move the cursor to the first occurrence of the
maximum phase-to-phase voltage. Display of Fresnel diagram of the signals.

Move the cursor to the first occurrence of the >t=0< Move cursor to transient triggering time.
minimum phase-to-phase voltage. >t=-T< Move the cursor to one signal period before the
Move the cursor to the first occurrence of the triggering date of the transient.
maximum current. Energies consumed by the load.
Move the cursor to the first occurrence of the
minimum current. Energies generated by the load.

Move the cursor to the first occurrence of the Page screen 1 of the help function.
maximum instantaneous frequency.
Page screen 2 of the help function.
Move the cursor to the first occurrence of the
minimum instantaneous frequency. Page screen 3 of the help function.

Move the cursor to the first occurrence of the Page screen 4 of the help function.
maximum of the measurement displayed.
Previous configuration.
Move the cursor to the first occurrence of the
minimum of the measurement displayed. Next configuration.
Simultaneous display of all voltage and current Previous page screen.
measurements (RMS, DC, THD, CF, PST, PLT,
FHL, FK). Next page screen.

2.5. Keypad keys


2.5.1. Function keys (yellow keys)
These 6 keys activate the function or tool represented by the corresponding icon on the screen.

2.5.2. Navigation keys


A block of 4 arrow keys, a select key and a return key are used for navigation in the menus.

Item Function

 Up direction or navigation key.


 Down direction or navigation key.
 Right direction or navigation key.
 Left direction or navigation key.
 Confirms the selection.
Return key.

10
2.5.3. Mode keys (violet keys)
These give access to specific modes:

Item Function See

Waveform acquisition mode, with two sub-modes: transients mode (blackouts, interference, §5
etc.) and inrush current mode (starting of motor).
Harmonic curves display mode: representation of voltage, current, and power harmonics, order §6
by order; determination of harmonic currents produced by nonlinear loads, analysis of problems
caused by harmonics according to their order (overheating of neutrals, conductors, motors, etc.).
Display of voltage and current waveforms, display of minima and maxima of summary tables, §7
determination of phase rotation.
Alarm mode: list of recorded alarms exceeding the thresholds programmed in the configuration; §8
recording of network blackouts with half-cycle resolution (Vrms, Arms, Urms), determination of
energy consumption overshoots, monitoring of compliance with a power supply quality contract.
Trend mode: recording of the parameters selected in the Configuration menu. §9

Display of power and energy measurements § 10

Three keys are real-time mode keys: , and .

In each of these modes, the coloured circles on a white ground , in which the channel numbers or types are entered, are indi-
cators of saturation: the ground of the circle is coloured when the channel measured is potentially full .

When the identification disc corresponds to a simulated channel (for example in 4-wire three-phase with selection V1V2, 2½-ele-
ment method, or in 3-wire three-phase with selection A1A2, 2-element method; see connections in §4.6), this channel is potentially
full if at least one channel used in calculating it is potentially full.
Similarly, if the saturation disc corresponds to a phase-to-phase voltage channel, it is potentially full if at least one of the phase-
to-neutral voltage channels used in calculating it is potentially full.

2.5.4. Other keys


The other keys have the following functions:

Item Function See

Configuration key. §4
Snapshot of current screen and retrieval of screens already stored. § 11
Help key: provides information about the functions and the symbols used for the current dis- § 12
play mode.

11
2.6. Connectors
2.6.1. Connection terminals
Located on the top of the device, these connectors are distributed as follows:

4 current input terminals for current sensors (MN clamp, C 5 voltage input terminals.
clamp, AmpFLEX™, PAC clamp, E3N clamp, etc.).

Figure 3: the connection terminals

2.6.2. Side connectors


Located on the right side of the device, these connectors are used as follows:

USB connector. For connection to a PC.

Mains power connector: Recharges the battery and allows operation on mains
power.

Figure 4: the side connectors

2.7. Power supply


The battery icon in the top right corner of the screen shows the battery level. The number of bars is proportional to the charge level.

Battery charged.
Low battery.
Mobile bars: battery charging.
The condition of the battery is unknown because it has never been fully charged.
The device is connected to mains.

When the battery level is too low, the following message is displayed:

Press  to confirm the information. If you do not connect the device to mains, it switches itself off one minute after this message.

12
2.8. The stand
A retractable stand on the back of the Qualistar+ can be used to hold the device in a tilted position.

Metal ring. It is used to secure the device with padlock.

Retractable stand.

Battery.

Figure 5: stand and battery compartment cover

2.9. Abbreviations
Prefixes of International System (SI) units

Prefix Symbol Multiplies by

milli m 10-3
kilo k 103
Mega M 106
Giga G 109
Tera T 1012
Peta P 1015
Exa E 1018

13
Meanings of the symbols and abbreviations used:

Symbol Designation Symbol Designation

AC and DC components. RMS True RMS value (current or voltage).


AC component only. t Relative date of time cursor.
DC component only. tan F Tangent of the phase shift of voltage with respect
Inductive phase shift. to current.
THD Total harmonic distortion (in %f or in %r).
Capacitive phase shift.
U Phase-to-Phase voltage.
° Degree.
U-h Phase-to-phase voltage harmonic.
-.+ Expert mode.
Ucf Phase-to-Phase voltage crest factor.
| | Absolute value.
Ud Phase-to-phase RMS distortion voltage.
S Sum of values.
Udc Phase-to-phase DC voltage.
% Percentage.
Upk+ Maximum peak value of the phase-to-phase voltage.
%f Fundamental value as reference
Upk- Minimum peak value of the phase-to-phase voltage.
%r Total value as reference
Urms True RMS phase-to-phase voltage.
FVA or FUA Phase shift of voltage with respect to current.
Uthdf Total phase-to-phase voltage harmonic distortion
A Current; also Ampere (unit). in %f.
A-h Current harmonic. Uthdr Total phase-to-phase voltage harmonic distortion
Acf Crest (peak) factor of current. in %r.
Ad RMS distortion current. Uunb Phase-to-phase voltage unbalance.
Adc DC current. V Phase-to-neutral voltage; also Volt (unit)
Apk+ Maximum peak value of the current. V-h Phase-to-neutral voltage harmonic.
Apk- Minimum peak value of the current. VA Apparent power.
Arms True RMS current. VA-h Apparent harmonic power.
Athdf Total harmonic distortion of current in % f. VAD Distortion power.
Athdr Total harmonic distortion of current in % r. VADh Distortion energy.
Aunb Current unbalance. VAh Apparent energy.
AVG Mean value (arithmetic mean). VAR Reactive or non-active power.
CF Peak factor (current or voltage). VARh Reactive or non-active energy.
cos F Cosine of the phase shift of voltage with respect Vcf Voltage crest (peak) factor.
to current (DPF – fundamental power factor or Vd Phase-to-neutral RMS distortion voltage.
displacement factor).
Vdc Phase-to-neutral DC voltage.
DC DC component (current or voltage).
Vpk+ Maximum peak value of the phase-to-neutral
DPF Displacement factor (cos F). voltage.
FHL Harmonic loss factor. Vpk- Minimum peak value of the phase-to-neutral
FK K factor. voltage.
Hz Frequency of network studied. Vrms True RMS phase-to-neutral voltage.
L Channel (Line). Vthdf Total harmonic distortion of phase-to-neutral
voltage in %f.
MAX Maximum value.
Vthdr Total harmonic distortion of phase-to-neutral
MIN Minimum value.
voltage in %r.
ms Millisecond.
Vunb Phase-to-neutral voltage unbalance.
PEAK Maximum (+) or minimum (-) peak instantaneous
W Active power.
or PK value of the signal.
Wdc DC power.
PF Power factor.
Wdch DC energy.
PLT Long-term flicker.
Wh Active energy.
PST Short-term flicker.

14
3. Use

3.1. Start-up
To switch the device on, press the button. It lights when pressed, then goes off if the mains power unit is not connected to
the device.

After the software check, the home page is displayed and indicates the software version of the device and its serial number.

Figure 6: Home page at start-up

Then the Waveform screen is displayed.

Figure 7: Waveform screen

3.2. Configuration
To configure the device, proceed as follows:
„„ Press . The configuration screen appears.
„„ Press  or  to select the parameter to be modified. Press  to enter the selected sub-menu.

Figure 8: Configuration screen

Then use the arrow keys ( or  and  or ) and the  key to validate. For more details, see §§4.3 to 4.10.

15
The following points must be checked or adapted for each measurement:
„„ Define the parameters of the calculation methods (see §4.5).
„„ Select the distribution system (single-phase to five-wire three-phase) and the connection method (2 wattmeters, 2 ½ elements,
standard) (see §4.6).
„„ Program the current ratios according to the type of current sensor connected (see §4.7).
„„ Program the voltage ratios (see §4.7).
„„ Define the transient triggering levels (transients mode and inrush current capture) (see §4.8).
„„ Define the values to be recorded (trend mode) (see §4.9).
„„ Define the alarm thresholds (see §4.10).

To return to the Configuration screen from a sub-menu, press the key.

3.3. Installation of leads


To identify the leads and input terminals, you may mark them in accordance with the usual phase/neutral colour code using the
coloured rings and inserts supplied with the device.
„„ Detach the insert and place it in the hole provided for it near the terminal (large hole for a current terminal; small hole for a
voltage terminal).

Insert for voltage Rings the same colour as


Insert for current terminal.
terminal. the terminal.

„„ Clip rings of the same colour to the ends of the lead you will be connecting to the terminal.
Twelve sets of rings and inserts of different colours are provided to enable you to harmonize the device with any of the phase/
neutral colour codes in force.

„„ Connect the measurement leads to the terminals of the device:

4 current inputs terminals. 5 voltage input terminals.

Figure 3: connection terminals

„„ Remember to define the transformation ratios of the current sensors and of the voltage inputs (see §4.7).

16
To make a measurement, you must program at least:
„„ the calculation method (see §4.5),
„„ the connection (see §4.6)
„„ and the ratios of the sensors (see §4.7).

The measuring leads must be connected to the circuit to be measured as shown by the following diagrams.

3.3.1. Single-phase network

Figure 9: 2-wire single-phase connection Figure 10: 2-wire single-phase connection

3.3.2. Split-phase network

Figure 11: 2-wire split-phase connection Figure 12: 3-wire split-phase connection Figure 13: 4-wire split-phase connection

3.3.3. Three-phase network

Figure 14: 3-wire three-phase Figure 15: 4-wire three-phase Figure 16: 5-wire three-phase
connection connection connection

In the case of a three-phase network, you are not obliged to connect all of the terminals in voltage or in current.

For 3-wire three-phase, indicate the 2 current sensors that will be connected: A1 and A2, or A2 and A3, or A3 and A1.
For 4- and 5-wire three-phase, indicate the voltages that will be connected: all 3 voltages (3V) or only 2 (V1 and V2, or V2 and
V3, or V3 and V1).

17
3.3.4. Connection procedure
„„ Switch the instrument on.
„„ Configure the device for the measurement to be made and the type of network concerned (see §4),
„„ Connect the leads and current sensors to the unit.
„„ Connect the earth and/or neutral lead to the network earth and/or neutral (when it is distributed) and connect the correspond-
ing current sensor,
„„ Connect the L1 phase lead to the network L1 phase and connect the corresponding current sensor.
„„ If applicable, repeat the procedure for phases L2 and L3 and for N.

Note: complying with this procedure reduces connection errors to a minimum and avoids wasting time.

Disconnection procedure:
„„ Proceed in the reverse of the order of connection, always finishing by disconnecting the neutral (when distributed).
„„ Disconnect the leads and switch the device off.

3.4. Functions of the device


Any screen can be saved (screen snapshot) by pressing the key (see §11).

You can press the help key at any time. The help screen will inform you about the functions and the symbols used for the
current display mode.

3.4.1. Waveform capture


With the device powered up and connected to the network, press .
You can display the Transients mode (see §5.1) or the Inrush current mode (see §5.2).

3.4.2. Display of harmonics


With the device powered up and connected to the network, press .
You can display the phase-to-neutral voltage (see §6.1), the current (see §6.2), the apparent power (see §6.3) or the phase-to-
phase voltage (see §6.4).

3.4.3. Waveform measurements


With the device powered up and connected to the network, press .

You can display the measurements of the true RMS value (see §7.1), the measurements of the total harmonic distortion (see §7.2),
the measurements of the peak factor (see §7.3), the extreme values in voltage and current (see §7.4), several values at once (see
§7.5), or the Fresnel diagram (see §7.6).

3.4.4. Alarm recording


With the device powered up and connected to the network, press .

You can configure the alarm mode (see §8.1), program an alarm campaign (see §8.2), look it up (see §8.4), or erase it (see §8.6).

3.4.5. Recording
With the device powered up and connected to the network, press .
You can configure recordings (see §9.2) and program them (see §9.1). You can also erase recordings (see §4.11).

3.4.6. Energy measurements


With the device powered up and connected to the network, press .
You can measure the energies consumed (see §10.1.3) or generated (see §10.1.4, §10.2.2, or §10.3.2).

18
4. Configuration

The Configuration key is used to configure the device. This must be done before each new type of measurement. The con-
figuration remains in memory, even after the device is switched off.

4.1. Configuration menu


The arrow keys (,, , ) are used to navigate in the Configuration menu and to parameterize the device.
A value that can be modified is flanked by arrows.

Most of the time, confirmation ( ) is necessary for the changes made by the user to be applied.

The return key ( ) is used to return to the main menu from a sub-menu.

Figure 8: the Configuration screen

4.2. Display language


To select the display language, press the yellow key under the corresponding icon on the screen (Figure 6).

The active language is identified by the icon on the yellow ground.

4.3. Date/Time
The menu defines the system date and time. The display is as follows:

Figure 17: Date/Time menu

With the Date/Time field highlighted in yellow, press  . To change a value, press  or . To move from one field to another,
press  or . To confirm, press  .

Proceed in the same way for the dating system (DD/MM/YY or MM/DD/YY) and the time system (12/24 or AM/PM). You see the
effect immediately in the display of the date.

To return to the Configuration menu, press .

Note: The date and time parameters cannot be configured while the device is recording, metering energy, or searching for tran-
sients, alarms, and/or inrush current acquisition.

19
4.4. Display
4.4.1. Contrast/Brightness
The menu is used to define the contrast and brightness of the display unit. The display is as follows:

Figure 18: the Contrast/Brightness menu

Use the arrow keys (,, , ) to change the contrast and brightness.
To return to the Configuration menu, press .

4.4.2. Colours
The menu is used to define the colours of the voltage and current curves. Press the yellow key corresponding to the icon.
There are 15 colours available: green, dark green, yellow, orange, pink, red, brown, blue, turquoise blue, dark blue, very light grey,
light grey, grey, dark grey, and black.

The display is as follows:

Figure 19: the Colours menu

Use the arrow keys (,, , ) to change the assignments of the colours.
To return to the Configuration menu, press .

4.4.3. Management of the switching off of the screen


The menu defines the management of the switching off of the screen. Press the yellow key corresponding to the icon.

Figure 124: the Management of Switching off of the Screen menu

Use the arrow keys (,) to choose the screen switching off mode: Automatic or Never.

20
The Automatic mode is used to save the battery. The display screen is switched off automatically after five minutes without action
on the keys if the device is powered only by its battery and if recording is in progress. The On/Off button blinks to indicate that
the device is still in operation. Pressing any key on the keypad relights the screen.

To return to the Configuration menu, press .

4.5. Calculation methods


The menu X= defines:
„„ The choice of breakdown or no breakdown of the non-active quantities,
„„ The choice of unit of energy,
„„ The choice of reference for the level of harmonics of the phases,
„„ The choice of coefficients of calculation for the factor,
„„ The choice of method of calculation of the long-term flicker.

4.5.1. Choice of calculation of non-active quantities


The VAR menu is used to choose whether or not to break down the non-active quantities (powers and energies).

Figure 20: the Methods of Calculation of Reactive Quantities menu

Use the arrow keys (,) to select broken down or not.


„„ Broken down: VAR corresponds to the fundamental reactive power and VAD to the distortion power.
„„ Not broken down: VAR corresponds to the non-active power; VAD does not exist.

Then validate with the  key. The device returns to the Configuration menu.

Note: The modification is impossible if the device is recording, metering energy, and/or searching for alarms.

21
4.5.2. Choice of unit of energy
The Wh menu defines the unit of display of energies.

Figure 21: the Choice of Unit of Energy menu

Use the arrow keys (,) to select the unit:


„„ Wh: Watt-hour.
„„ Joule.
„„ nuclear toe: nuclear tonne oil equivalent.
„„ non-nuclear toe: non-nuclear tonne oil equivalent.
„„ BTU: British Thermal Unit.

Then validate with the  key . The device returns to the Configuration menu.

4.5.3. Choice of coefficients of calculation of the K factor


The FK menu defines the coefficients used for the calculation of the K factor.

Figure 22: the Choice of Coefficients of Calculation of the K Factor menu

Use the arrow keys (,, , ) to fix the value of coefficients q and e:
„„ q: exponential constant that depends on the type of winding and the frequency.
The value of q can range from 1.5 to 1.7. The value of 1.7 is suitable for transformers having round or square conductors, in
all types of winding. The value of 1.5 is suitable for those in which the low-voltage windings are in tape form.
„„ e: ratio between the losses linked to eddy currents (at the fundamental frequency) and resistive losses (both evaluated at the
reference temperature). The value of e can range from 0.05 to 0.1.

The default values (q = 1.7 and e = 0.10) are suitable for most applications.

Then validate with the  key. The device returns to the Configuration menu.

Note: The modification is impossible if the device is recording and/or searching for alarms.

22
4.5.4. Choice of reference of the level of harmonics of the phases
The %f-%r menu defines the reference for the level of harmonics of the phases.

Figure 23: the Choice of Reference for the Level of Harmonics menu

Use the arrow keys (,) to fix the reference for the level of harmonics:
„„ %f: the reference is the value of the fundamental.
„„ %r: the reference is the total value.

Then validate with the  key. The device returns to the Configuration menu.

In the case of the level of harmonics of the V-h, A-h, and U-h phases, the fundamental and total values are RMS values. In the
case of the level of harmonics of the VA-h phases, the fundamental and total values are apparent power values.

Note: The modification is impossible if the device is recording and/or searching for alarms.

4.5.5. Choice of method of calculation of PLT


The PLT menu defines the method used to calculate the PLT (long-term flicker).

Figure 24: the Choice of Method of Calculation of PLT menu

Use the arrow keys (,) to choose sliding or non-sliding.


„„ Sliding: the PLT is calculated every 10 minutes. The first value is available 2 hours after the device is switched on, because it
takes 12 values of PST to calculate the PLT.
„„ Non-sliding: the PLT is calculated every 2 hours.

Then validate with the  key . The device returns to the Configuration menu.

Note: The modification is impossible if the device is recording and/or searching for alarms.

23
4.6. Connection
The menu is used to define how the device is connected, according to distribution system.

Figure 16: the Connection menu

Several electrical diagrams can be selected:


Use the arrow keys (,, , ) to choose a connection.

One or more types of network correspond to each distribution system.

Distribution system Source

Single-phase 2-wire (L1 and N) L1

Single-phase 2-wire non-earthed neutral

Single-phase 3-wire (L1, N and L1


earth)

Single-phase 3-wire earthed neutral


N
GND

L1

Split-phase 2-wire (L1 and L2) Split-phase 2-wire

L2

L1
3-phase open star 2-wire
L2

24
Distribution system Source

L1

Split-phase 3-wire non-earthed neutral N

L2

3-phase open star 3-wire non-earthed neutral L1

L2
Split-phase 3-wire (L1, L2 and
N)

3-phase high leg delta 3-wire non-earthed neutral L1


N

L2

3-phase open high leg delta 3-wire non-earthed neutral L1


N

L2

L1

Split-phase 4-wire earthed neutral N


GND
L2

3-phase open star 4-wire earthed neutral L1


GND

Split-phase 4-wire (L1, L2, N L2


and earth)

3-phase high leg delta 4-wire earthed neutral L1


N
GND
L2

3-phase open high leg delta 4-wire earthed neutral L1


N
GND
L2

25
Distribution system Source

L3

3-phase star 3-wire

L1

L2

L3

3-phase delta 3-wire


L1
L2

L3

3-phase open delta 3-wire


L1
3-phase 3-wire (L1, L2 and L3) L2

L3

3-phase open delta 3-wire earthed junction of phases


Indicate which 2 current sen- L1
sors will be connected: A1 and L2
A2, or A2 and A3, or A3 and A1.
L3

3-phase open delta 3-wire earthed corner of phase


L1

L2

L3

3-phase high leg delta 3-wire


L1

L2

L3

3-phase open high leg delta 3-wire


L1

L2

26
Distribution system Source

L3

N
3-phase star 4-wire non-earthed neutral

L1
3-phase 4-wire (L1, L2, L3
and N) L2

L3

3-phase open high leg delta 4-wire non-earthed neutral L1


Indicate which voltages will be N
connected: all 3 (3V) or only 2
(V1 and V2, or V2 and V3, or L2
V3 and V1).
L3

3-phase high leg delta 4-wire non-earthed neutral L1


N

L2

L3

N
3-phase star 5-wire earthed neutral

L1
3-phase 5-wire (L1, L2, L3, N GND

and earth) L2

L3

3-phase open high leg delta 5-wire earthed neutral L1


Indicate which voltages will be N
connected: all 3 (3V) or only 2 GND
(V1 and V2, or V2 and V3, or V3 L2
and V1).
L3

3-phase high leg delta 5-wire earthed neutral L1


N
GND
L2

Then validate with the  key . The device returns to the Configuration menu.

This makes it possible to connect the device to all existing networks.

Note: It is impossible to select a new connection if the device is recording, metering energy, or searching for transients, alarms,
and/or inrush current acquisitions.

27
4.7. Sensors and ratios
Note: The ratios cannot be changed if the device is recording, metering energy, or searching for transients, alarms, and/or inrush
current acquisitions.

4.7.1. Current sensors and ratios


A first screen A is used to define the current sensors and ratios. It automatically displays the current sensor models detected by
the device. It can also be used to define the transformation ratio (sensitivity) of certain current sensors (E3N clamp).

Figure 25: Current clamp and ratios screen in the Sensors and ratios menu

In the case of a 3-wire three-phase set-up where only two of the three current sensors required are connected, if these two sen-
sors are of the same type and have the same ratio, the device simulates the third sensor by assuming the same characteristics
as for the two others.

The various current sensors are:


MN93 clamp: 200 A.
MN93A clamp: 100 A or 5 A.

C193 clamp: 1000 A.

AmpFLEX™ A193: 6500 A.


MiniFLEX MA193: 6500 A.

PAC93 clamp: 1000 A.

E3N clamp: 100 A (sensitivity 10 mV/A).


E3N clamp: 10 A (sensitivity 100 mV/A).

Three phase adapter: 5 A.

If an MN93A clamp, 5A range, or an Adapter is used, the current ratio setting is proposed automatically.

Use the arrow keys (,, , ) to define the transformation ratios between the primary current (1A to 60,000A) and the second-
ary current (1A, 2A or 5A), then validate with the  key.

The primary current cannot be less than the secondary current.

4.7.2. Voltage ratios


A second screen, invoked by the V or U icon, defines the voltage ratios.

The programming of the ratio or ratios can be different or the same for all or for some channels.
The ratios to be programmed are phase-to-neutral voltage ratios when there is a neutral and phase-to-phase voltage ratios when
there is not.

4.7.3. To change the ratios, press the  key.

28
Figure 26: the Voltage Ratios screen in the Sensors and Figure 27: the Voltage Ratios screen in the Sensors and
ratios menu in the case of a set-up without neutral ratios menu in the case of a set-up with neutral

Use the arrow keys (,) to choose the configuration of the ratios.
„„ 3U 1/1 or 4V 1/1: all channels have the same 1/1 ratio.
„„ 3U or 4V: all channels have the same ratio, to be programmed.
„„ Press the  key, then use the , keys to highlight the ratio in yellow.

„„ Press the  key, then use the ,, and  keys to change the ratio. The primary voltage is in kV and the secondary
voltage is in V.

„„ 3V + VN: all channels have the same ratio and the neutral has a different ratio.
Proceed as when there is only one ratio, but perform the operation twice.
„„ U1+U2+U3 or V1+V2+V3+VN: each channel has a different ratio, to be programmed.
Proceed as when there is only one ratio, but perform the operation several times.

Validate with the  key. To return to the Configuration menu, press .

Note: the primary and secondary voltages can each be configured with a multiplier factor of 1/√3.

4.8. Capture Mode


The mode is used to configure the voltage thresholds, the current thresholds of the transient mode, and the current thresh-
olds of the inrush current mode.

4.8.1. Voltage thresholds of the transient mode

A first screen, displayed by pressing the V icon (or U, for set-ups without a neutral), is used to configure the voltage thresh-
olds.

The thresholds programmed can be the same for all channels or different for some or all of them.

Figure 28: the Current thresholds screen in the Transient Mode menu

To change the voltage thresholds, press  .

Use the arrow keys (,) to choose configuration of the thresholds.

29
„„ 4V or 3U: all channels have the same threshold.
„„ Press the  key, then use the , keys to highlight the value of the threshold in yellow.

„„ Press the  key, then use the ,, and  keys to change the threshold. The unit can be the V or the kV.

„„ 3V + VN: all channels have the same ratio and the neutral has a different ratio.
Proceed as when there is only one ratio, but perform the operation twice.
„„ V1+V2+V3+VN or U1+U2+U3: each channel has a different ratio, to be programmed.
Proceed as when there is only one ratio, but perform the operation several times.

Validate with the  key. To return to the Configuration menu, press .

Note: Changing the thresholds in the transient mode is impossible if the device is searching for transients.

4.8.2. Current thresholds of the transient mode

A second screen, displayed by pressing the A icon, is used to configure the current thresholds (independently of the current
sensors detected by the device).

The thresholds programmed can be the same for all channels or different for some or all of them.

Figure 29: the Voltage thresholds screen in the Transient Mode menu

To change the current thresholds, press  .

Use the arrow keys (,) to choose configuration of the thresholds.


„„ 4A: all current sensors have the same threshold.
„„ Press the  key, then use the , keys to highlight the value of the threshold in yellow.

„„ Press the  key, then use the ,, and  keys to change the threshold. The unit can be the A, the kA or the mA.

„„ 3A + AN: all current sensors have the same threshold and the one connected to the neutral has a different threshold.
Proceed as when there is only one ratio, but perform the operation twice.
„„ A1+A2+A3+AN: each current sensor has a different threshold, to be programmed.
Proceed as when there is only one ratio, but perform the operation several times.

Validate with the  key. To return to the Configuration menu, press .

Note: Changing the thresholds in the transient mode is impossible if the device is searching for transients.

30
4.8.3. Current thresholds of the inrush current mode

A third screen, displayed by pressing the icon, is used to define the inrush current thresholds. This involves programming
the triggering threshold and the inrush current capture stopping threshold (the stopping threshold being the triggering threshold
less the hysteresis).

Figure 30: the Inrush Current Thresholds screen in the Inrush Current Mode menu

To change the inrush current triggering threshold, press the  key.


Use the ,, and  keys to change the triggering threshold. The unit can be the A, the kA, or the mA.
Press the  key, then use the , keys to highlight the hysteresis in yellow.
Use the ,, and  keys to change the hysteresis and press the  key to validate.

Notes: for more information on the hysteresis, refer to §16.3. Configuring the hysteresis at 100% is equivalent to not having a
stop threshold (see §16.7).
Changing the thresholds in inrush current mode is impossible if the device is in inrush current capture.

To return to the Configuration menu, press .

4.9. Trend mode


The device has a recording function ( key, see §9), used to record measured and calculated values (Urms, Vrms, Arms, etc.).

Press the key of the Configuration mode and select the Trend Mode sub-menu.

Figure 31: The first screen of the Trend mode Figure 32: The second screen of the Trend mode

There are 4 possible programmable configurations , , and independent of one another. To go from one to the
other, use the or key.

To select the parameter to be recorded, move the yellow cursor to this parameter using the ,, and  keys then validate with
the  key. The selected parameter is identified by a red spot. The frequency (Hz) is always selected (black spot).

Note: If a quantity is displayed in red, it means that it is incompatible with the configuration chosen (connection selected, sensors
connected, ratios programmed, reference of the level of harmonics of the phases, breakdown of the non-active quantities).
For example, if no current sensor is connected, all current quantities will be in red.

To select all of the parameters of a page, press the key.


To unselect all of the parameters of a page, press the key.

31
To change configuration pages, press the or key.

The recordable values are:

Unit Designation

Urms RMS phase-to-phase voltage.


Udc RMS phase-to-neutral voltage.
Upk+ Maximum peak value of phase-to-phase voltage.
Upk- Minimum peak value of phase-to-phase voltage.
Ucf Crest (peak) factor of phase-to-phase voltage.
Uthdf Harmonic distortion of the phase-to-phase voltage with the RMS value of the fundamental as
reference.
Uthdr Harmonic distortion of the phase-to-phase voltage with the total RMS value without DC as reference.
Vrms RMS phase-to-neutral voltage.
Vdc DC phase-to-neutral voltage.
Vpk+ Maximum peak value of the phase-to-neutral voltage.
Vpk- Minimum peak value of the phase-to-neutral voltage.
Vcf Crest factor of phase-to-neutral voltage.
Vthdf Harmonic distortion of the phase-to-neutral voltage with the RMS value of the fundamental as
reference.
Vthdr Harmonic distortion of the phase-to-neutral voltage with the total RMS value without DC as refer-
ence.
Arms RMS current.
Adc DC current.
Apk+ Maximum peak value of the current.
Apk- Minimum peak value of the current.
Acf Crest factor of current.
Athdf Harmonic distortion of the current with the RMS value of the fundamental as reference.
Athdr Harmonic distortion of the current with the total RMS value without DC as reference.
W Active power.
Wdc DC power.
VAR Reactive or non-active power.
VAD Distortion power.
VA Apparent power.
PF Power factor.
cos F Cosine of the phase shift of the voltage with respect to the current (displacement factor or fun-
damental power factor – DPF).
tan F Tangent of the phase shift of the voltage with respect to the current.
PST Short-term flicker.
PLT Long-term flicker.
FHL Harmonic loss factor
FK K factor.
Vunb Phase-to-neutral voltage unbalance (set-up with neutral).Phase-to-phase voltage unbalance (set-
or Uunb up without neutral).
Aunb Current unbalance.
Hz Network frequency.
U-h Harmonics in phase-to-phase voltage.
V-h Harmonics in phase-to-neutral voltage
A-h Harmonics in current.
VA-h Harmonics in power.

32
The four last lines involve the recording of the harmonics of U, V, A and VA. You can select a range of orders of the harmonics to
be recorded (between 0 and 50) for each of these quantities, and within this range, if desired, only odd harmonics.

Note: The level of harmonics of order 01 will be displayed only if they concern values expressed in % r.

To change an order of harmonic, first select the parameter to be recorded (identified by a red spot), then move the yellow cur-
sor to this figure using the ,, and  keys, then validate with the  key. Change the value using the  and keys, then
validate with the  key.

Figure 33: The second screen of the Trend Mode during modification

Note: If a recording is in progress, the associated configuration cannot be modified and the selected values are identified by
black spots.

To return to the Configuration menu, press .

4.10. Mode Alarm mode


The screen defines the alarms used by the Alarm Mode function (see §7).

You can define a alarm on each of the following parameters:


Hz, Urms, Vrms, Arms, |Udc|, |Vdc|, |Adc|, |Upk+|, |Vpk+|, |Apk+|, |Upk-|, |Vpk-|, |Apk-|, Ucf, Vcf, Acf, Uthdf, Vthdf, Athdf, Uthdr,
Vthdr, Athdr, |W|, |Wdc|, |VAR|, VAD, VA, |PF|, |cos Φ|, |tan Φ|, PST, PLT, FHL, FK, Vunb (or Uunb for a three-phase source without
neutral), Aunb, U-h, V-h, A-h and |VA-h| (see the table of abbreviations in §2.9).

There are 40 programmable alarms.

To activate an alarm, move the yellow cursor to its number using the , keys, then validate with the  key. The active alarm is
identified by a red spot. An alarm that is not programmed (“?”) cannot be activated.

To program the alarm, move the yellow cursor using the ,, and  keys, then validate with the  key. Change the value,
then validate again.

Alarms active.

Alarm inactive.

Alarm not programmed.

Figure 34: The Alarm mode menu

33
„„ The type of alarm.
„„ The order of the harmonic (between 0 and 50), for |VA-h|, A-h, U-h and V-h only.
„„ The target of the alarm:
„„ 3L: 3 phases monitored individually,
„„ N: monitoring of neutral,
„„ 4L: 3 phases and neutral monitored individually,
„„ Σ: surveillance of the sum of the phases.
„„ The direction of the alarm (>or <) in the case of Hz, Urms, Vrms, Arms, |Udc|, |Vdc|, |Adc|, |Upk+|, |Vpk+|, |Apk+|, |Upk-|, |Vpk-|
and |Apk-|.
„„ The triggering threshold of the alarm (value and unit for Urms, Vrms, Arms, |Udc|, |Vdc|, |Adc|, |Upk+|, |Vpk+|, |Apk+|, |Upk-|,
|Vpk-|, |Apk-|, |W|, |Wdc|, |VAR|, VAD and VA).
„„ The triggering delay, or minimum duration above or below the alarm threshold: in minutes or seconds or, in the case of Vrms,
Urms and Arms (excluding the neutral), in hundredths of a second.
„„ The hysteresis: 1%, 2%, 5% or 10% (see §16.3).

To go from one page to the other, press the or key.

Each overshoot of an alarm will be recorded in a campaign of alarms.

Notes: The display in red of an alarm line means that the programmed quantity and/or target is incompatible with the configuration
chosen (connection selected, sensors connected, ratios programmed, calculation methods chosen).
The alarms on the level of harmonics of order 01 concern only the values expressed in % r.
If a search for alarms is in progress, the activated alarms cannot be modified and are identified by black spots. However,
new alarms (not yet programmed or not activated) can be activated.

To return to the Configuration menu, press .

4.11. Erase memory


The menu partially or totally deletes the data recorded in the device.

Figure 35: Erase memory menu

To select an item to be erased, move the yellow cursor to it using the ,, and  keys, then validate with the  key. The item
to be erased is identified by a red spot.

To select all items, press .


To unselect all items, press .

To proceed with the erasure, press the key , then confirm with the  key.

To return to the Configuration menu, press .

Note: Which erasures are possible depends on the recordings in progress (recording, metering of energy, search for transients,
alarms, and/or inrush current acquisition).

34
4.12. About
The About screen displays information concerning the device.

Figure 36: the About menu

To return to the Configuration menu, press .

35
5. Waveform Capture

The Waveform capture mode is used to display and to capture transients and inrush currents.

It contains two sub-modes:


„„ The transient mode (see §5.1)
„„ The inrush current mode (see §5.2)

Figure 37: the screen of the Waveform capture mode

To select a sub-mode, move the yellow cursor to it using the  and  keys, then validate with the  key.

To return to the Waveform capture screen, press .

5.1. Transient mode


The mode is used to record transients, to look up the list of recorded searches and the list of transients they contain, or erase
them. You can record up to 7 detections and 210 transients.

When the Transient mode is invoked:


„„ If no recording has been made, then the Detection schedule screen is displayed.
„„ If transients have been recorded, then the List of searches for transients is displayed.

Reminder of sub-mode used.


Display of the list of searches for
transients (see §5.1.2).
Memory indicator. The black bar
represents memory used; the white
bar represents memory available.
Programming a search (see §5.1.1).

Shortcut to the Configuration menu Starting a search.


to set the voltage and current trig-
gering thresholds (see §4.8).

Figure 38: the Detection schedule screen in Transient mode

36
5.1.1. Programming and starting a search
To program a search for a transient, enter the start date and time, the stop date and time, the number of transients to search for,
then the name of the search.

To change an item, move the yellow cursor to it using the  and  keys, then validate with the  key. Change the value using
the ,, and  keys, then validate again.

The name can be at most 8 characters long. Several searches can bear the same name. The available alphanumeric characters
are the uppercase letters from A to Z and the digits from 0 to 9. The last 5 names given (in the transient, trend, and alarm modes)
are kept in memory. When a name is entered, it may then be completed automatically.

Notes: The start date and time must be later than the current date and time.
The stop date and time must be later than the start date and time.

Once the programming is done, start the search by pressing the key. The icon of the status bar blinks to indicate that the
search has been started. The key replaces the key and can be used to stop the search before it is finished.

The message Detection on standby is displayed until the start time is reached. It is then replaced by the message Detection in
progress. When the stop time is reached, the Detection schedule screen returns with the ??? key . It is then possible to program
another search.

During a search for transients, only the stop date field can be modified. It is automatically highlighted in yellow.

To return to the Waveform capture screen, press .

5.1.2. Displaying a transient


To display the recorded transients, press . The List of Searches for Transients screen is displayed.

Display of sub-mode used.

Memory indicator. The black bar


represents memory used; the white
bar represents memory available.

Figure 39: the screen of the List of Searches for Transients

If the stop date is in red, it means that it does not match the stop date initially programmed:
„„ either because of a power supply problem (battery low or disconnection of the device supplied by mains only),
„„ or because the number of transients has been reached, thereby ending the search.

37
To select a search for transients, move the cursor to it using the  and  keys. The selected search is bolded. Then validate
with the  key. The device then displays a list of transients.

Transients display filter:


Triggering channel of the transient.
∀: all transients are displayed.
4 V: the transients triggered by an
Transient number. event in one of the 4 voltage chan-
nels are displayed.
4 A: the transients triggered by an
event in one of the 4 current chan-
Name of the search for transients.
nels are displayed.
L1, L2, or L3: the transients trig-
gered by an event on a particular
The icon is used to activate or phase are displayed (voltage or
deactivate the choice of a transient current).
list display filter. N: the transients triggered by an
event on the neutral current or neu-
tral voltage are displayed.

Figure 40: The Transient list screen in the case of a 5-wire three-phase set-up

To select a transient, move the cursor to it using the  and  keys. The selected field is bolded. Then validate with the  key.
The device displays the transients in the form of curves.

Location in the record of the zone Reminder of the number assigned to


displayed. the curve displayed; here, identifica-
tion disc 1 is filled in to indicate that
channel V1 triggered capture of the
Move the cursor to one period of transient.
the signal before the transient trig-
gering time. Selection of curves to be displayed.

Instantaneous value of the signals


Move the cursor to the transient according to the position of the cur-
triggering time. sor on the scale. To move the cursor
use the  or  key.

Figure 41: example of display of transients in the form of curves with a 5-wire three-phase connection

Note: The curves to be displayed selection filter is dynamic and depends on the connection chosen. For example, it proposes
(3U, 3A) for a 3-wire three-phase set-up

To return to the Transient list screen, press .

5.1.3. Deleting a search for transients


When the list of searches for transients is displayed (see figure 39), select the search to be erased. This is done by moving the
cursor to it using the  and  keys. The selected search is bolded.

Then press the key. Press  to validate or to cancel.

Note: A search for transients can be deleted only if it is not in progress.

To return to the Waveform capture screen, press the key.

5.1.4. Delete a transient


When the list of transients in a search is displayed (see figure 40), select the transient to be erased. This is done by moving the
cursor to it using the  and  keys. The selected transient is bolded.

Then press the key. Press  to validate or to cancel.

To return to the Waveform capture screen, press the key .

38
5.2. Inrush current mode
Still in the mode, the sub-mode is used to capture (record) inrush currents (voltage and current waveforms, network
frequency, half-cycle RMS voltages and currents except for the neutral) and to view and delete the recordings.

When the Inrush current mode is invoked:


„„ If no capture has been made, then the Capture schedule screen is displayed.
„„ If a capture has been made, then the Capture characteristics screen is displayed.

5.2.1. Programming the capture

Display of sub-mode used.

Display of the characteristics of the


capture (see §5.2.2).
Programming of the capture.
Rapid programming and starting of
a capture.
Shortcut to the Configuration menu Start of the capture.
to set the triggering thresholds (see
§4.8).

Figure 42: the Capture schedule screen in Inrush current mode

To start recording rapidly, press the key. Recording starts immediately with a current threshold of 0 A and a hysteresis of
100%.

Attention: The rapid starting of an inrush current modifies the configuration of the current threshold..

To program a capture, enter the triggering filter (3A, A1, A2, or A3), the start date and time, and the recording mode (RMS + PEAK
or RMS only).
„„ The RMS + PEAK recording mode is used to produce a trend recording of the half-period RMS values and a trend recording
of the samples (envelopes and waveforms). The maximum duration of such a recording depends on the frequency and is on
average about one minute.
„„ In the RMS only recording mode, the recording of the samples is eliminated in favour of a longer maximum capture duration.
This mode records only the half-period RMS values and its maximum duration is about ten minutes.

To modify an item, move the yellow cursor to it using the  and  keys, then validate with the  key. Change the value using
the ,, and  keys, then validate again.

Notes: The display of the triggering filter in red means that it is not available because of an incompatibility with the configuration
(connection, type of sensors, or current ratio).
For more information on the triggering filter, refer to §16.7.
It is not possible to program an alarm campaign if an inrush current capture is in progress.

Once the programming is done, start the capture by pressing the key. The icon of the status bar blinks to indicate that the
capture has been started. The key replaces the key and can be used to stop the capture before it is finished.

Attention: The voltage must be present before the inrush current strictly speaking for a stable and correct frequency lock.

The message capture pending is displayed until the start time is reached and the triggering conditions are satisfied. It is then
replaced by the message Capture in progress. The memory occupation indicator appears at the top of the screen. The indicator
disappears when the capture is completed.

If the capture is completed with a stop event (see conditions in §16.7) or if the recording memory of the device is full, the capture
stops automatically.

Note: The device can keep in memory only a single inrush current capture. If you wish to make another capture, first delete the
previous one.

To return to the Waveform capture screen, press .

39
5.2.2. Displaying the characteristics of the capture
To display the characteristics of the capture, press . The Capture parameters screen is displayed.

Display in PEAK mode (see §5.2.4).

Display in RMS mode (see §5.2.3).

Figure 43: The Capture parameters screen

If a capture duration is displayed in red, it means that it has been cut short:
„„ because of a power supply problem (battery low),
„„ or because the memory is full.
„„ or because of a measurement error.
„„ or because of an incompatibility between the quantity monitored and the configuration of the device (for example withdrawal
of a current sensor).

Choose the type of display, RMS or PEAK, by pressing the yellow key corresponding to the icon. The device then displays the
curves.

Note: the PEAK key is not displayed when the inrush current capture recording mode is RMS only.

5.2.3. True RMS current and voltage


The RMS mode displays the record of the trend of the true half-cycle RMS current and voltage and the frequency trend curve.

The display depends on the type of selection filter:


3V: displays the 3 voltages during the inrush current capture for set-ups with neutral.
3U: displays the 3 voltages during the inrush current capture for set-ups without neutral.
3A: displays the 3 currents during the inrush current capture.
L1, L2, L3: display the current and voltage in phases 1, 2 and 3 respectively (for set-ups with neutral only).
Hz: displays the evolution of the network frequency vs. time.

Below, three examples of display.

5.2.3.1. The 3A RMS display screen for a three-phase connection with neutral

Location of the zone displayed in MAX: maximum half-cycle RMS


the record. value of the inrush current capture.
Reminder of number assigned to
Scale of values in amperes. the curve displayed. Here, identifi-
cation disc 1 is filled in, indicating
Time cursor. Use the  or  key to that channel A1 triggered the inrush
move the cursor. current capture.

t: relative time position of the cursor A1, A2, A3: RMS values of currents
(t=0 corresponds to the start of the 1, 2, and 3 in the half-cycle at the
inrush current capture). position of the cursor.

Figure 44: The 3A RMS display screen for a three-phase connection with neutral

40
5.2.3.2. The 3A RMS display screen for a three-phase connection without neutral

Figure 45: The 3A RMS display screen for a three-phase connection without neutral

5.2.3.3. The L1 RMS display screen for a three-phase connection with neutral

MAX: maximum half-cycle RMS


Time cursor of the curve. Use the 
value of the inrush current capture.
or  key to move the cursor.

t: relative time position of the cursor


(t=0 corresponds to the start of the V1: RMS value of voltage 1 in the
inrush current capture). half-cycle at the position of the
cursor.
A1: RMS value of current 1 in the half-
cycle at the position of the cursor.

Figure 46: The L1 RMS display screen for a three-phase connection with neutral

Note: Filters L2 and L3 are used to display the recording of the true half-cycle RMS current and voltage of phases 2 and 3. The
screen is identical to the one displayed for filter L1.

The , , and keys are used to go to the first occurrence of a minimum or maximum voltage or current
value.

5.2.3.4. The RMS display screen in Hz for a three-phase connection without neutral

Figure 47: The Hz RMS display screen for a three-phase connection without neutral

The and keys are used to go to the first occurrence of a minimum or maximum frequency value.

41
5.2.4. Instantaneous inrush current
The PEAK mode is used to display the envelopes and waveforms of the inrush current capture.

The PEAK display of an inrush current capture provides two possible representations:
„„ envelope
„„ waveform.

The change from one of these representations to the other is automatic, as a function of the zoom level. If the zoom in is high
enough, the representation is of the “waveform” type.

The display depends on the type of display filter:


4V: displays the 4 voltages during the inrush current capture for set-ups with neutral (for a waveform type display only).
3U: displays the 3 voltages during the inrush current capture for set-ups without neutral (for a waveform type display only).
4A: displays the 4 currents during the inrush current capture (for a waveform type display only).
L1, L2 or L3: display the voltage and current of phases 1, 2, and 3, respectively (only for set-ups with neutral and for a waveform
type display).
N: displays the neutral current and neutral voltage during the inrush current capture (for a waveform type display only).
V1, V2, V3: displays the 3 voltages during the inrush current capture for set-ups with neutral (for an envelope type display only).
U1, U2, U3: displays the 3 voltages during the inrush current capture for set-ups without neutral (for an envelope type display only).
A1, A2, A3: displays the 3 currents during the inrush current capture (for an envelope type display only).

Below, three examples of display.

5.2.4.1. The 4A PEAK display screen for a 5-wire three-phase connection

Location of the zone displayed in


Reminder of number assigned to the
the record.
curve displayed. Here, identification
disc 3 is filled in to indicate that
MAX |PEAK|: absolute value. channel A3 triggered capture of the
inrush current.
Scale of values in amperes.

Time cursor. Use the  or  key to A1, A2, A3: instantaneous values of
move the cursor. currents 1, 2, and 3 at the position
of the cursor.

t: relative time position of the cursor


(t=0 corresponds to the start of the
inrush current capture).

Figure 48: The 4A PEAK display screen for a 5-wire three-phase connection

5.2.4.2. The 3A PEAK display screen for a 3-wire three-phase connection

Figure 49: The 3A PEAK display screen for a 3-wire three-phase connection

42
5.2.4.3. The A1 PEAK display screen for a three-phase connection without neutral
In the case shown, there is enough zoom out to force the envelope type of representation.

MAX |PEAK|: maximum instanta- Time cursor of the curve. Use the 
neous absolute value of the inrush or  key to move the cursor.
current capture.

t: relative time position of the cursor


(t=0 corresponds to the start of the
inrush current capture).
A1: maximum instantaneous cur-
rent of the half-cycle identified by
the cursor.

Figure 50: The A1 PEAK display screen for a three-phase connection without neutral

Note: Filters A2 and A3 display the record of the current envelope of phases 2 and 3. The screen is identical to the one displayed
for filter A1.

43
6. Harmonic

The Harmonic mode displays a representation of the harmonic levels of the voltage, current, and apparent power, order by
order. It can be used to determine the harmonic currents produced by nonlinear loads and analyze problems caused by harmon-
ics according to their order (overheating of neutrals, conductors, motors, etc.).

Analysis of the apparent power of Select the filters and the expert
the harmonics (see §6.3). mode (see §6.5). Use the  or 
key to select the display.
Analysis of harmonics of the current
(see §6.2).
Analysis of the harmonics of the
Analysis of harmonics of the phase- phase-to-phase voltage (see §6.4).
to-neutral voltage (see §6.1).

Figure 51: Harmonics mode screen

6.1. Phase-to-neutral voltage


The V sub-menu displays the harmonics of the phase-to-neutral voltage only for sources having a neutral.

The choice of curves to be displayed depends on the type of connection (see §4.6):
„„ Single-phase, 2-wire: no choice (L1)
„„ Single-phase, 3-wire: L1, N
„„ Split-phase, 3-wire: 2L, L1, L2
„„ Split-phase, 4-wire: 2L, L1, L2, N
„„ Three-phase, 4-wire: 3L, L1, L2, L3, -,+
„„ Three-phase, 5-wire: 3L, L1, L2, L3, N, -,+

The screen captures shown as examples are those obtained with a 5-wire three-phase connection.

6.1.1. The 3L phase-to-neutral voltage harmonics display screen

This information concerns the har- Display of the 3 phases 3L, of L1,
monic pointed to by the cursor. L2, L3, N, or of the expert mode
V-h03: harmonic number. (three-phase connection only - see
%: level of harmonics with the fun- §6.5). To select the display press the
damental RMS value as reference  or  key.
(%f) or the (total) RMS value as
reference (%r). The horizontal axis indicates the
V: RMS voltage of the harmonic in orders of the harmonics. The level of
question. the harmonic is given as a percent-
+000: phase shift with respect to the age with respect to the fundamental
fundamental (order 1). or to the total RMS value.
DC: DC component. 1 to 25: har-
Harmonic selection cursor. Use the  monics of order 1 to 25. When the
or  key to move the cursor. cursor exceeds order 25, order 26
to 50 appears.

Figure 52: example of 3L phase-to-neutral voltage harmonics display

44
6.1.2. The L1 phase voltage harmonics display screen

This information concerns the har- Display of the 3 phases 3L, of L1, L2,
monic pointed to by the cursor. L3, N, or of the expert mode (three-
V-h03: harmonic number. phase connection only - see §6.5).
%: level of harmonics with the fun- To select the display press  or .
damental RMS value as reference
(%f) or the (total) RMS value as
reference (%r).
V: RMS voltage of the harmonic in The horizontal axis indicates the
question. orders of the harmonics. The level of
+000: phase shift with respect to the the harmonic is given as a percent-
fundamental (order 1). age with respect to the fundamental
max – min: maximum and minimum or to the total RMS value.
levels of the harmonic in question. DC: DC component.
They are reset when the harmonic 1 to 25: harmonics of order 1 to 25.
Harmonic selection cursor. Use the  or  key When the cursor exceeds order 25,
number is changed or the  key is
to move the cursor. order 26 to 50 appears.
pressed.
THD: total harmonic distortion.
Vd: RMS distortion voltage. Indicator of the presence of non-zero
harmonics of order higher than 25.

Figure 53: example of display of harmonics of L1 phase-to-neutral voltage

Notes: Filters L2 and L3 display the harmonics of the phase-to-neutral voltage for phases 2 and 3, respectively. The screen is
identical to the one displayed for filter L1.
There is no phase shift or distortion value for the neutral channel.

6.2. Current
The A sub-menu displays the harmonics of the current.

6.2.1. The 3L current harmonics display screen

Display of the 3 phases 3L, of L1, L2,


This information concerns the har-
L3, N, or of the expert mode (three-
monic pointed to by the cursor.
phase connection only - see §6.5).
A-h05: harmonic number.
To select the display press  or .
%: level of harmonics with the fun-
damental RMS value as reference
(%f) or the (total) RMS value as
reference (%r). The horizontal axis indicates the
A: RMS current of the harmonic in orders of the harmonics. The level of
question. the harmonic is given as a percent-
+000: phase shift with respect to the age with respect to the fundamental
fundamental (order 1). or to the total RMS value.
Rang DC: DC component.
Harmonic selection cursor. Use the 1 to 25: harmonics of order 1 to 25.
 or  key to move the cursor. When the cursor exceeds order 25,
order 26 to 50 appears.

Figure 54: example of 3L display of current harmonics

45
6.2.2. The L1 current harmonics display screen

This information concerns the har- Display of the 3 phases 3L, of L1, L2,
monic pointed to by the cursor. L3, N, or of the expert mode (three-
A-h05: harmonic number. phase connection only - see §6.5).
%: level of harmonics with the fun- To select the display press  or .
damental RMS value as reference
(%f) or the (total) RMS value as
reference (%r). The horizontal axis indicates the
A: RMS current of the harmonic in orders of the harmonics. The level of
question. the harmonic is given as a percent-
+000°: phase shift with respect to age with respect to the fundamental
the fundamental (order 1). or to the total RMS value.
max – min: maximum and minimum Rang DC: DC component.
levels of the harmonic in question. 1 to 25: harmonics of order 1 to 25.
They are reset when the harmonic Harmonic selection cursor. Use the  or When the cursor exceeds order 25,
number is changed or the  key is  key to move the cursor. order 26 to 50 appears.
pressed
THD: total harmonic distortion.
Ad: RMS distortion current.

Figure 55: example of L1 display of harmonics of current

Notes: Filters L2 and L3 display the current harmonics of phases 2 and 3, respectively. The screen is identical to the one dis-
played for filter L1.
There is no phase shift or distortion value for the neutral channel.

6.3. Apparent power


The VA sub-menu displays the harmonics of the apparent power, for all connections except 3-wire three-phase.

The horizontal axis indicates the orders of the harmonics. The bars of the bar chart above the horizontal centreline signify harmonic
power consumed, those below it harmonic power generated.

6.3.1. The 3L apparent power harmonics display screen

This information concerns the har-


monic pointed to by the cursor.
VA-h03: harmonic number.
%: level of harmonics with the funda- Display of the 3 phases 3L, of L1,
mental apparent power as reference L2 or L3. To select the display press
(%f) or the (total) apparent power as the  or .
reference (%r).
+000: phase shift of the voltage The horizontal axis indicates the
harmonic with respect to the current orders of the harmonics. The level of
harmonic for the order in question. the harmonic is given as a percent-
: Indicator of energy generated age of the fundamental apparent
for this harmonic. power or the (total) apparent power.
: Indicator of energy consumed Rang DC: DC component.
for this harmonic. Harmonic selection cursor. Use the  1 to 25: harmonics of order 1 to 25.
or  key to move the cursor. When the cursor exceeds order 25,
order 26 to 50 appears.

Figure 56: example of 3L apparent power harmonics display

46
6.3.2. The L1 apparent power harmonics display screen

This information concerns the har-


monic pointed to by the cursor.
VA-h03: harmonic number. Display of the 3 phases 3L, of L1,
%: level of harmonics with the L2 or L3. To select the display press
fundamental apparent power as the  or .
reference (%f) or the (total) apparent
power as reference (%r). The horizontal axis indicates the
+000: phase shift of the voltage orders of the harmonics. The level of
harmonic with respect to the current the harmonic is given as a percent-
harmonic for the order in question. age of the fundamental apparent
min–max: maximum and minimum power or the (total) apparent power.
levels of the harmonic in question. Rang DC: DC component.
They are reset when the harmonic 1 to 25: harmonics of order 1 to 25.
number is changed or the  key is Harmonic selection cursor. Use the  or When the cursor exceeds order 25,
pressed.  key to move the cursor. order 26 to 50 appears.
Indicator of energy consumed
for this harmonic.

Figure 57: example of L1 apparent power of harmonics display

Note: Filters L2 and L3 display the apparent power of the harmonics for phases 2 and 3, respectively. The screen is identical to
the one displayed for filter L1.

6.4. Phase-to-phase voltage


The U sub-menu is available for all connections except 2- or 3-wire single-phase.. This sub-menu displays the harmonics of the
phase-to-phase voltage.

6.4.1. The 3L phase-to-phase voltage harmonics display screen

This information concerns the har-


monic pointed to by the cursor.
U-h03: harmonic number. Display of the 3 phases 3L, of L1,
%: level of harmonics with the fun- L2, L3. To select the display press
damental RMS value as reference  or .
(%f) or the (total) RMS value as
reference (%r).
The horizontal axis indicates the
V: RMS voltage of the harmonic in
orders of the harmonics. The level of
question.
the harmonic is given as a percent-
+000: phase shift with respect to the
age with respect to the fundamental
fundamental (order 1).
or to the total RMS value.
Rang DC: DC component.
Harmonic selection cursor. Use the 1 to 25: harmonics of order 1 to 25.
 or  key to move the cursor. When the cursor exceeds order 25,
order 26 to 50 appears.

Figure 58: example of 3L phase-to-phase voltage harmonics display

47
6.4.2. The L1 phase-to-phase voltage harmonics display screen

This information concerns the har-


monic pointed to by the cursor.
Uh 03: harmonic number. Display of the 3 phases 3L, of L1,
%: level of harmonics with the fun- L2 or L3. To select the display press
damental RMS value as reference  or .
(%f) or the (total) RMS value as
reference (%r).
V: RMS voltage of the harmonic in The horizontal axis indicates the
question. orders of the harmonics. The level of
+000°: phase shift with respect to the harmonic is given as a percent-
the fundamental (order 1). age with respect to the fundamental
max – min: indicators of the maxi- or to the total RMS value.
mum and minimum of the level of Rang DC: DC component.
harmonics or by pressing the  key. Harmonic selection cursor. Use the  or  1 to 25: harmonics of order 1 to 25.
THD: total harmonic distortion. key to move the cursor. When the cursor exceeds order 25,
Ud: phase-to-phase RMS distortion order 26 to 50 appears.
voltage.

Figure 59: example of L1 phase-to-phase voltage harmonics display

Note: Filters L2 and L3 display the phase-to-phase voltage harmonics for phases 2 and 3, respectively. The screen is identical
to the one displayed for filter L1.

6.5. Expert mode


The Expert mode is available with a three-phase connection only. It is used to display the influence of the harmonics on the
heating of the neutral and on rotating machines. To display expert mode press the  or  key of the keypad. The selection is
highlighted in yellow and the screen simultaneously displays the expert mode.

From this screen, two sub-menus are available:


„„ V for three-phase set-ups with neutral or U for the three-phase set-up without neutral.
„„ A for the expert mode in current.

6.5.1. The phase-to-neutral voltage expert mode display screen


For three-phase set-ups with neutral, the V sub-menu displays the influence of the harmonics of the phase-to-neutral voltage on
the heating of the neutral and on rotating machines.

Harmonics inducing a negative Harmonics inducing a positive se-


sequence. quence.

Harmonics inducing a zero se-


quence.
%: level of harmonics with the fun-
damental RMS value as reference
(%f) or the (total) RMS value as
reference (%r).

Figure 60: The phase-to-neutral voltage expert mode screen (three-phase set-ups with neutral)

For three-phase set-ups without neutral, the U sub-menu displays the influence of the harmonics of the phase-to-phase voltage
on the heating of revolving machines.

48
6.5.2. The current expert mode display screen
The A sub-menu displays the influence of the harmonics of the current on the heating of the neutral and on rotating machines.

Harmonics inducing a negative Harmonics inducing a positive se-


sequence. quence.

Harmonics inducing a zero se-


quence.
%: level of harmonics with the fun-
damental RMS value as reference
(%f) or the (total) RMS value as
reference (%r).

Figure 61: the current expert mode screen

49
7. Waveform

The Waveform key is used to display the current and voltage curves, along with the values measured and those calculated
from the voltages and currents (except for power, energy, and harmonics).
This is the screen that appears when the device is powered up.

Display of the maximum and mini-


mum RMS values and of the peak Selection of the display filters. Use
values (see §7.4). the  or  key to select the display.

Measurement of the peak factor


(see §7.3). Simultaneous display of the follow-
ing measurements: RMS, DC, THD,
CF, PST, PLT, FHL and FK (see §7.5)
Measurement of total harmonic
distortion (see §7.2).
Display of the Fresnel diagram of the
signals (see §7.6).
Measurement of the true RMS value
(see §7.1).
Figure 62: Waveform mode screen

7.1. Measurement of true RMS value


The RMS sub-menu displays the waveforms over one period of the signals measured and the true RMS voltage and current.

The choice of curves to be displayed depends on the type of connection (see §4.6):
„„ Single-phase, 2-wire or Split-phase, 2-wire: no choice (L1)
„„ Single-phase, 3-wire:
„„ For RMS, THD, CF, and : L1 and N
„„ For : no choice (L1)
„„ Split-phase, 3-wire:
„„ For RMS, THD, CF, and : U, 2V, 2A, L1, L2
„„ For : 2V, 2A, L1, L2
„„ Split-phase, 4-wire:
„„ For RMS, THD, CF, and : U, 3V, 3A, L1, L2 N
„„ For : 2V, 2A, L1, L2
„„ Three-phase, 3wire: 3U, 3A
„„ Three-phase, 4-wire: 3U, 3V, 3A, L1, L2, L3
„„ Three-phase, 5-wire:
„„ For RMS, THD, CF, and : 3U, 4V, 4A, L1, L2, L3 and N
„„ For: 3U, 3V, 3A, L1, L2 and L3

The screen snapshots shown as examples are those obtained with a three-phase 5-wire connection.

50
7.1.1. The 3U RMS display screen
This screen displays the three phase-to-neutral voltages of a three-phase system.

RMS phase-to-phase voltages. Instantaneous values of the signals


at the position of the cursor.
t: time relative to the start of the
period.
Voltage axis with automatic scaling.
U1: instantaneous phase-to-phase
voltage between phases 1 and 2
(U12).
Instantaneous value cursor. Use the U2: instantaneous phase-to-phase
 or  key to move the cursor. voltage between phases 2 and 3
(U23).
U3: instantaneous phase-to-phase
voltage between phases 3 and 1
(U31).

Figure 63: The 3U RMS display screen

7.1.2. The 4V RMS display screen


This screen displays the three phase-to-neutral voltages and the neutral-to-earth voltage of a three-phase system.

Instantaneous values of the signals


RMS phase-to-neutral voltages. at the position of the cursor.
t: time relative to the start of the
period.
V1: instantaneous phase-to-neutral
Voltage axis with automatic scaling.
voltage of curve 1.
V2: instantaneous phase-to-neutral
voltage of curve 2.
Instantaneous value cursor. Use the
V3: instantaneous phase-to-neutral
 or  key to move the cursor.
voltage of curve 3.
VN: instantaneous value of the neu-
tral voltage.

Figure 64: The 4V RMS display screen

7.1.3. The 4A RMS display screen


This screen displays the three phase currents and the neutral current of a three-phase system.

Instantaneous values of the signals


RMS currents. at the intersection of the cursor and
of the curves.
t: time relative to the start of the
Current axis with automatic scaling. period.
A1: instantaneous current of phase
1.
Instantaneous value cursor. Use the A2: instantaneous current of phase
 or  key to move the cursor. 2.
A3: instantaneous current of phase
3.
AN: instantaneous value of the neu-
tral current.
Figure 65: The 4A RMS display screen

51
7.1.4. The RMS display screen for the neutral
This screen displays the neutral voltage with respect to earth and the neutral current.

RMS voltage and current.


Instantaneous values of the signals
at the position of the cursor.
Current and voltage axis with auto- t: time relative to the start of the
matic scaling. period.
VN: instantaneous neutral voltage.
AN: instantaneous neutral current.
Instantaneous value cursor. Use the
 or  key to move the cursor.

Figure 66: The RMS display screen for the neutral

Note: Filters L1, L2, and L3 display the current and voltage in phases 1, 2, and 3, respectively. The screen is identical to the one
displayed for the neutral.

7.2. Measurement of total harmonic distortion


The THD sub-menu displays the waveforms of the signals measured over one full cycle and the total voltage and current harmonic
distortion. The levels are displayed either with the fundamental RMS value as reference (%f) or with the RMS value without DC
as reference (%r), depending on which reference is chosen in the configuration menu.

7.2.1. The 3U THD display screen


This screen displays the phase-to-phase voltage waveforms for one period and the total harmonic distortion values.

Instantaneous values of the signals


Harmonic distortion for each curve. at the position of the cursor.
t: time relative to the start of the
period.
Voltage axis with automatic scaling. U1: instantaneous phase-to-phase
voltage between phases 1 and 2
(U12).
Instantaneous value cursor. Use the U2: instantaneous phase-to-phase
 or  key to move the cursor. voltage between phases 2 and 3
(U23).
U3: instantaneous phase-to-phase
voltage between phases 3 and 1
(U31).
Figure 67: The 3U THD display screen en 3U

7.2.2. The 4V THD display screen


This screen displays the phase-to-neutral voltage waveforms for one period and the total harmonic distortion values.

Harmonic distortion for each curve. Instantaneous values of the signals


at the position of the cursor.
t: time relative to the start of the
period.
Voltage axis with automatic scaling.
V1: instantaneous phase-to-neutral
voltage of curve 1.
V2: instantaneous phase-to-neutral
Instantaneous value cursor. Use the voltage of curve 2.
 or  key to move the cursor. V3: instantaneous phase-to-neutral
voltage of curve 3.
VN: instantaneous value of the neu-
tral voltage.

Figure 68: the 4V THD display screen

52
7.2.3. The 4A THD display screen
This screen displays the phase current waveforms for one period and the total harmonic distortion values.

Harmonic distortion for each curve. Instantaneous values of the signals


at the position of the cursor.
t: time relative to the start of the
Current axis with automatic scaling. period.
A1: instantaneous current of phase
1.
Instantaneous value cursor. Use the A2: instantaneous current of phase
 or  key to move the cursor. 2.
A3: instantaneous current of phase
3.
AN: instantaneous value of the neu-
tral current.
Figure 69: The 4A THD display screen

Note: Filters L1, L2, L3 and N display the total current and voltage harmonic distortion for phases 1, 2 and 3 and the neutral channel.

7.3. Measurement of the peak factor


The CF sub-menu displays the waveforms of the signals measured over one period and the voltage and current peak factors.

7.3.1. The 3U CF display screen


This screen displays the phase-to-phase voltage waveforms of one period and the peak factors.

Peak factor for each curve. Instantaneous values of the signals


at the position of the cursor.
t: time relative to the start of the
period.
Voltage axis with automatic scaling. U1: instantaneous phase-to-phase
voltage between phases 1 and2 (U12).
U2: instantaneous phase-to-phase
Instantaneous value cursor. Use the voltage between phases 2 and 3 (U23).
 or  key to move the cursor. U3: instantaneous phase-to-phase
voltage between phases 3 and 1 (U31).

Figure 70: The 3U CF display screen

7.3.2. The 4V CF display screen


This screen displays the phase-to-neutral voltage waveforms of one period and the peak factors.

Instantaneous values of the signals


Peak factor for each curve. at the position of the cursor.
t: time relative to the start of the
period.
Voltage axis with automatic scaling. V1: instantaneous phase-to-neutral
voltage of curve 1.
V2: instantaneous phase-to-neutral
voltage of curve 2.
Instantaneous value cursor. Use the
V3: instantaneous phase-to-neutral
 or  key to move the cursor.
voltage of curve 3.
VN: instantaneous value of the
phase-to-neutral voltage of the
neutral.
Figure 71: The 3V CF display screen

53
7.3.3. The 4A CF display screen
This screen displays the current waveforms of one period and the peak factors.

Peak factor for each curve. Instantaneous values of the signals


at the position of the cursor.
t: time relative to the start of the
Current axis with automatic scaling. period.
A1: instantaneous current of phase 1.
A2: instantaneous current of phase 2.
A3: instantaneous current of phase 3.
Instantaneous value cursor. Use the
AN: instantaneous value of the cur-
 or  key to move the cursor.
rent of the neutral.

Figure 72: The 4A CF display screen

Note: L1, L2, L3 and N display the current and voltage peak factors for phases 1, 2 and 3, respectively and the neutral channel.

7.4. Measurement of extreme and mean voltage and current


The sub-menu displays the one-second mean and half-cycle maximum and minimum RMS voltage and current and the in-
stantaneous positive and negative peak voltage and current.

Note: The MAX. and MIN. RMS measurements are calculated every half cycle (i.e. every 10 ms for a 50-Hz signal). The measure-
ments are refreshed every 250 ms.

7.4.1. The 3U Max.-Min. display screen


This screen displays the one-second mean and half-cycle maximum and minimum RMS values and the positive and negative
phase-to-phase voltage peaks.

Columns of values for each curve (1, 2, and 3).


MAX: maximum RMS phase-to-phase voltage since the switching on of the
device or since the last time the  key was pressed.
RMS: true RMS phase-to-phase voltage.
MIN: minimum RMS phase-to-phase voltage since the switching on of the device
or since the last time the  key was pressed.
PK+: maximum (positive) peak phase-to-phase voltage since the switching on
of the device or since the last time the  key was pressed.
PK-: minimum (negative) peak phase-to-phase voltage since the switching on of
the device or since the last time the  key was pressed.

Figure 73: The 3U Max.-Min. display screen

54
7.4.2. The 4V Max.-Min. display screen
This screen displays the one-second mean and half-cycle maximum and minimum RMS values and the positive and negative
peaks of the phase-to-neutral voltages and of the neutral.

Column of values for the neutral: RMS, PEAK+ and PEAK- parameters.

Columns of values for each voltage curve (1, 2 and 3).


MAX: maximum RMS phase-to-neutral voltage since the switching on of the
device or since the last time the  key was pressed.
RMS: true RMS phase-to-neutral voltage.
MIN: minimum RMS phase-to-neutral voltage since the switching on of the device
or since the last time the  key was pressed.
PK+: maximum peak phase-to-neutral voltage since the switching on of the device
or since the last time the  key was pressed.
PK-: minimum peak phase-to-neutral voltage since the switching on of the device
or since the last time the  key was pressed.

Figure 74: The 4V Max.-Min. display screen

7.4.3. The 4A Max.-Min. display screen


This screen displays the one-second mean and half-cycle maximum and minimum RMS values and the positive and negative
peak values of the phase and neutral currents.

Column of values for the neutral: RMS, PEAK+ and PEAK- parameters.

Columns of values for each current curve (1, 2 and 3).


MAX: maximum RMS current since the switching on of the device or since the
last time the  key was pressed.
RMS: true RMS current.
MIN: minimum RMS current since the switching on of the device or since the last
time the  key was pressed.
PK+: maximum peak current since the switching on of the device or since the
last time the  key was pressed.
PK-: minimum peak current since the switching on of the device or since the last
time the  key was pressed.
Figure 75: The 4A Max.-Min. display screen

7.4.4. The L1 Max.-Min. display screen


This screen displays the one-second mean and half-cycle maximum and minimum RMS values and the positive and negative
peaks of the phase-to-neutral voltage and of the current of phase 1.

The same information as for the phase-to-neutral voltage, but for the current.

Column of voltage values.


MAX: maximum RMS phase-to-neutral voltage since the switching on of the
device or since the last time the  key was pressed.
RMS: true RMS phase-to-neutral voltage.
MIN: minimum RMS phase-to-neutral voltage since the switching on of the device
or since the last time the  key was pressed.
PK+: phase-to-neutral maximum peak voltage since the switching on of the
device or since the last time the  key was pressed.
PK-: phase-to-neutral minimum peak voltage since the switching on of the device
or since the last time the  key was pressed.

Figure 76: The L1 Max.-Min. display screen

Note: L2 and L3 display the RMS, maximum, minimum, and mean values and the positive and negative peak values of the phase-
to-neutral voltage and of the current for phases 2 and 3, respectively.

55
7.4.5. The neutral Max.-Min. display screen
This screen displays the RMS values and the positive and negative peaks of the neutral relative to earth.

Column of voltage values. The same information as for the volt-


RMS: true RMS voltage. age, but for the current.
PK+: maximum peak voltage since
the switching on of the device or
since the last time the  key was
pressed.
PK-: minimum peak voltage since
the switching on of the device or
since the last time the  key was
pressed.

Figure 77: The neutral Max.-Min. display screen

7.5. Simultaneous display


The sub-menu displays all of the voltage and current measurements (RMS, DC, THD, CF, PST, PLT, FHL and FK).

7.5.1. 3U Simultaneous Display screen


This screen displays the RMS, DC, THD, and CF values of the phase-to-phase voltages.

Column of phase-to-phase voltages (phases 1, 2, and 3).


RMS: true RMS value calculated over 1 second.
DC: DC component.
THD: total harmonic distortion with the fundamental RMS value as reference (%f)
or with the total RMS value without DC as reference (%r).
CF: peak factor calculated over 1 second.

Figure 78: 3U simultaneous display screen

7.5.2. 4V Simultaneous Display screen


This screen displays the RMS, DC, THD, CF, PST and PLT values of the phase-to-neutral voltages and of the neutral.

RMS: true RMS value calculated over 1 second.


DC: DC component.
THD: total harmonic distortion with the fundamental RMS value as reference (%f)
or with the total RMS value without DC as reference (%r).
CF: peak factor calculated over 1 second.
PST: short-term flicker calculated over 10 minutes.
PLT: long-term flicker calculated over 2 hours.

Column of RMS and DC values along with the CF and the THD (%r) concerning
the neutral.

Figure 79: 4V simultaneous display screen

56
7.5.3. 4A Simultaneous display screen
This screen displays the RMS, DC (only if at least one of the current sensors can measure direct current), THD, CF, FHL and FK
values of the phase and neutral currents.

Columns of current values (phases 1, 2 and 3).


RMS: true RMS value calculated over 1 second.
DC: DC component.
THD: total harmonic distortion with the fundamental RMS value as reference (%f)
or with the total RMS value without DC as reference (%r).
CF: peak factor calculated over 1 second.
FHL: harmonic loss factor. For the oversizing of transformers to allow for harmonics.
FK: K factor.

Column of RMS values and (if the current sensor allows) DC values along with the
CF and the THD (%r) concerning the neutral.

Figure 80: 4A simultaneous display screen

Note: To make it possible to adjust the zero of current sensors measuring DC, the DC values are never cancelled.

7.5.4. L1 Simultaneous Display screen


This screen displays the RMS, DC, THD, and CF values of the phase-to-neutral voltage and of the current, PST and PLT of the
phase-to-neutral voltage, and FHL and FK of the current for phase 1.

Column of voltage values simple. Current column.


RMS: true RMS value calculated RMS, DC (if the current sensor al-
over 1 second. lows), THD and CF values.
DC: DC component. FHL: harmonic loss factor. For the
THD: total harmonic distortion with oversizing of transformers to allow
the fundamental RMS value as refer- for harmonics.
ence (%f) or with the total RMS value FK: K factor.
without DC as reference (%r).
CF: peak factor calculated over 1
second.
PST: short-term flicker calculated
over 10 minutes.
PLT: long-term flicker calculated
over 2 hours.
Figure 81: L1 simultaneous Display screen

Notes: The DC value of the current of phase 1 is displayed only if the associated current sensor can measure direct current.
L2 and L3 provide the simultaneous display of the current and voltage for phases 2 and 3, respectively.

7.5.5. Screen for Simultaneous Display of neutral


This screen displays the RMS, THD and CF voltage and current of the neutral, the DC component of the neutral voltage, and (if
the current sensor allows) the DC component of the neutral current.

57
7.6. Display of Fresnel diagram
The sub-menu displays a vector representation of the fundamentals of the voltages and currents. It indicates their associ-
ated quantities (modulus and phase of the vectors) and the unbalances of the voltages and currents.

Note: To allow the display of all vectors, those of which the modulus is too small to be pictured are shown even so, but their
name is followed by an asterisk (*).

7.6.1. The 3V Fresnel diagram display screen


This screen displays a vector representation of the fundamentals of the phase-to-neutral voltages and of the currents. It indicates
their associated quantities (modulus and phase of the phase-to-neutral voltage vectors) and the voltage unbalance. The reference
vector of the representation (at 3 o’clock) is V1.

Column of values for each vector


(1, 2, and 3).
|V1|, |V2| and |V3|: module of the
vectors of the fundamentals of the
phase-to-neutral voltages (phases Discs indicating potential saturation
1, 2 and 3). of the channel.
Φ12: phase angle of the fundamental
of phase 1 with respect to the fun-
damental of phase 2. Fresnel diagram.
Φ23: phase angle of the fundamental
of phase 2 with respect to the fun-
damental of phase 3.
Φ31: phase angle of the fundamental
of phase 3 with respect to the fun-
damental of phase 1. Vunb: voltage unbalance.

Figure 82: The screen Displaying the Fresnel diagram in 3V

7.6.2. The 3U Fresnel diagram display screen


This screen displays a vector representation of the fundamentals of the phase-to-phase voltages and of the currents. It indicates
their associated quantities (modulus and phase of the phase-to-phase voltage vectors) and the voltage unbalance. The reference
vector of the representation (at 3 o’clock) is U1.

The information displayed is identical to that described in §7.6.1 but relative to the phase-to-phase voltage.

7.6.3. The 3A Fresnel diagram display screen


For sources having a neutral, this screen displays the vector representation of the fundamental component of the phase-to-neutral
voltages and of the current. For 3-wire three-phase (source without neutral), this screen displays only the vector representation
of the fundamental components of the current. It indicates their associated quantities (modulus and phase of the current vectors)
and the current unbalance. The reference vector of the representation (at 3 o’clock) is A1.

The information displayed is identical to that described in §7.6.1 but relative to the current.

58
7.6.4. The L1 Fresnel diagram display screen
In the presence of the neutral, this screen displays a vector representation of the fundamentals of the phase-to-neutral voltages
and the currents of one phase. It indicates their associated quantities (modulus and phase of the current and phase-to-neutral
voltage vectors). The reference vector of the representation (at 3 o’clock) is the current vector.

|V1|: modulus of the vector of the


fundamental of the phase-to-neutral Discs indicating potential saturation
voltage of phase 1. of the channel.
|A1|: modulus of the vector of the fun-
damental of the current of phase 1.

ΦVA: phase angle of the fundamental


of the phase-to-neutral voltage of
phase 1 relative to the fundamental
of the current of phase 1.

Figure 83: The L1 Fresnel diagram display screen

Note: L2 and L3 display vector representations of the fundamentals of the phase-to-neutral voltages and the currents of phases
2 and 3, respectively. They indicate their associated quantities (modulus and phase of the current and phase-to-neutral
voltage vectors of phases 2 and 3, respectively). The reference vector of the representation (at 3 o’clock) is the current
vector (A2 and A3, respectively).

In the absence of the neutral (2-wire two-phase):

|U1|: modulus of the vector of the


fundamental component of the Discs indicating potential saturation
phase-to-phase voltage between of the channel.
phases 1 and 2 (U12 ).
|A1|: modulus of the vector of the fun-
damental of the current of phase 1.

ΦUA: phase shift of the fundamental


component of the phase-to-phase
voltage between phases 1 and 2
(U12 ) with respect to the fundamental
component of the current of phase 1.

Figure 84: the Fresnel diagram display screen in 2-wire two-phase

59
8. Alarm mode

The Alarm mode detects overshoots of thresholds on each of the following parameters:
Hz, Urms, Vrms, Arms, |Udc|, |Vdc|, |Adc|, |Upk+|, |Vpk+|, |Apk+|, |Upk-|, |Vpk-|, |Apk-|, Ucf, Vcf, Acf, Uthdf, Vthdf, Athdf, Uthdr,
Vthdr, Athdr, |W|, |Wdc|, |VAR|, VAD, VA, |PF|, |cos F|, |tan F|, PST, PLT, FHL, FK, Vunb, Uunb (for a three-phase source without
neutral) Aunb, U-h, V-h, A-h and |VA-h| (see the table of abbreviations in §2.9).

The alarm thresholds:


„„ must have been programmed in the Configuration / Alarm mode screen (see §4.10).
„„ must be active (marked with a red spot on that same screen).

Stored alarms can subsequently be transferred to a PC via the PAT application (see §13). You can capture over 16,000 alarms.

List of alarm campaigns (see §8.3).

Access to Alarm mode configuration


(see §8.1). Programming an alarm campaign
(see §8.2).

Figure 85: The Alarm Mode screen

The and icons have the following functions:


„„ : Validating the programming of a campaign and starting the alarm campaign.
„„ : Voluntary stoppage of alarm campaign.

8.1. Alarm mode configuration


The submenu displays the list of alarms configured (see §4.10). This shortcut key lets you define or change alarm configura-
tions.
The following information is displayed.

Press to return to the Programming a campaign screen.

8.2. Programming an alarm campaign


The submenu is used to specify the start and stop times for an alarm campaign (see figure 66).

To program an alarm campaign, enter the start date and time, the stop date and time, and the name of the campaign.

To modify an item, move the yellow cursor to it using the  and  keys, then validate with the  key. Change the value using
the ,, and  keys, then validate again.

The name can be at most 8 characters long. Several campaigns may have the same name. The available alphanumeric characters
are the uppercase letters from A to Z and the digits from 0 to 9. The last 5 names given (in the transient, trend, and alarm modes)
are kept in memory. When a name is entered, it may then be completed automatically.

Notes: The start date and time must be later than the current date and time.
The stop date and time must be later than the start date and time.
It is not possible to program an alarm campaign if an inrush current capture is in progress.

Once the programming is done, start the campaign by pressing the key. The icon of the status bar blinks to indicate that
the campaign has been started. The key replaces the key and can be used to stop the campaign before it is finished.
Alarms in progress (not yet ended) are recorded in the campaign if their duration is equal to or greater than their programmed
minimum duration.

60
The Campaign on standby message is displayed until the start time is reached. It is then replaced by the message Campaign
running . When the stop time is reached, the Programming a Campaign screen returns with the key. You can then program
another campaign.

During an alarm campaign, only the stop date field can be modified. It is automatically highlighted in yellow.

8.3. Display of the list of campaigns


To display the list of campaigns performed, press the key. The List of Alarm Campaigns screen is displayed. The list can
contain up to 7 campaigns.

Name of the campaign.

Start date and time of the campaign.


Stop date and time of the campaign.

Figure 86: list of campaigns display screen

If the stop date of the campaign is in red, it means that it does not match the stop date initially programmed:
„„ either because of a power supply problem (battery low or disconnection of the device supplied by mains only),
„„ or because the memory is full.

8.4. Display of list of alarms


To select a campaign, move the cursor to it using the  and  keys. The selected field is bolded. Then validate with the  key.
The device then displays the alarms in list form.

Level of filling dedicated to the


alarm mode. The black part of the
bar corresponds to the fraction of Alarm duration.
memory used.
Extremum of the alarm detected
(minimum or maximum depending
Alarm date and time. on the programmed alarm direction).

Target of the alarm detected. The choice of filter is dynamic. It


depends on the connection chosen.
Type of alarm detected.

Figure 87: Alarm list screen

If an alarm duration is displayed in red, it means that it was cut short:


„„ because of a power supply problem (battery low),
„„ or because of a manual stoppage of the campaign (press on ) or deliberate switching off of the device (press on the key).
„„ or because the memory is full.
„„ or because of a measurement error.
„„ or because of an incompatibility between the quantity monitored and the configuration of the device (for example withdrawal
of a current sensor).

In the last two cases, the extremum is also displayed in red.

To return to the List of campaigns screen, press .

61
8.5. Deleting an alarm campaign
When the list of campaigns performed is displayed (see figure 86), select the campaign to be erased. This is done by moving the
cursor to it using the  and  keys. The selected campaign is bolded.

Then press the key. Press  to validate or to cancel.

Note: It is not possible to delete the alarm campaign in progress.

8.6. Erasing all alarm campaigns


Erasing all of the alarm campaigns is possible only from the Configuration menu, in the Erasure of Data sub-menu (see §4.11)

62
9. Trend mode

The Trend mode records changes to parameters previously specified in the Configuration / Trend mode screen (see §4.9).

Memory card usage. List of records (see §9.3).

Rapid programming and starting of


recording (see §9.1). Programming a recording (see §9.1).

Access to Trend mode configuration


(see §4.9). Starting a recording (see §9.1).

Figure 88: Trend mode screen

9.1. Programming and starting recording


The submenu specifies the characteristics of a recording (see figure 88).

To start a recording rapidly, press the key. Recording starts immediately; all measurements are recorded every second until
the memory or completely full. The configuration displayed is .

To program recording, before starting it, choose configuration to , enter the start date and time, the stop date and time,
the period, and the name of the recording.

To modify an item, move the yellow cursor to it using the  and  keys, then validate with the  key. Change the value using
the ,, and  keys, then validate again.

The integration period is the time over which the measurements of each recorded value are averaged (arithmetic mean). Possible
values for the period are: 1 s, 5 s, 20 s, 1 min, 2 min, 5 min, 10 min and 15 min.

The name can be at most 8 characters long. Several recordings may have the same name. The available alphanumeric characters
are the uppercase letters from A to Z and the digits from 0 to 9. The last 5 names given (in the transient, trend, and alarm modes)
are kept in memory. When a name is entered, it may then be completed automatically.

Notes: The start date and time must be later than the current date and time.
The stop date and time must be later than the start date and time.

Once the programming is done, start recording by pressing the key. If not enough memory is available, the device so indicates.
The icon of the status bar blinks to indicate that recording has begun. The key replaces the key and can be used to
stop the recording before it is finished.

The Recording on standby message is displayed until the start time is reached. It is then replaced by the message Recording
running . When the stop time is reached, the Program a Record screen returns with the key. You can then program a new re-
cording.

During the recording of a trend, only the stop date field can be modified. It is automatically highlighted in yellow.

9.2. Trend mode configuration


The submenu displays the list of trend recording configurations (see §4.9). This shortcut key lets you specify or modify the
trend recording configurations.

63
9.3. Viewing the recording list
The submenu displays the recording list already made.

Recording list memory usage. The


black part of the bar corresponds to
the fraction of memory used.

Recording name. Recording stop time.

Recording start time.

Figure 89: Recording list display screen

If the stop date is in red, it means that it does not match the stop date initially programmed, because of a power supply problem
(battery low or disconnection of the device supplied by mains only).

9.4. Deleting recordings


When the list of records is displayed (see figure 89), select the recording to be erased. This is done by moving the cursor to it
using the  and  keys. The selected recording is bolded.

Then press the key. Press  to validate or to cancel.

9.5. Viewing the records


9.5.1. Characteristics of the record
When the list of records is displayed (see figure 89), select the record to be displayed. This is done by moving the cursor to it
using the  and  keys. The selected recording is bolded. Then press  to validate.

The icon is used to navigate in


the following screen pages. It is also
possible to use the  or  key.
Types of measurement chosen in the
configuration used.

Figure 90: Recording list display screen

If a measurement does not appear in the tabs, it is because calculation of this measurement was incompatible with the configura-
tion chosen (connection, types of sensors, ratios programmed).

For example, if the calculation mode chosen during the programming is Non-active Quantities not broken down (see §4.5.1), the
VAD tab does not appear.

Press a yellow key to display the curve.

64
9.5.2. Trend curves

Date of the cursor. Position of the viewing window in


the record.

This screen is a partial view of the


trend curve. There are other screens
before and after the visible part. To select the display filter, press the
 or  key.
Use the  or  key to move the
cursor.

Figure 91: Vrms (4L) without MIN-AVG-MAX

The display period of this curve is one minute. Since the period of the record is one second, each point of this curve corresponds
to a value recorded in a one-second window once a minute. There is therefore a substantial loss of information (59 values out of
60), but the display is rapid.

Notes: Values of the cursor in red indicate saturated values.


Black dashes - - - - indicate erroneous values.
Red dashes - - - - indicate values missing in the memory card.

The MIN-AVG-MAX mode has been


activated.
To change the scale of the display
between 1 minute and 5 days.

Figure 92: Vrms (4L) with MIN-AVG-MAX

The display period of this curve is one minute. But with the MIN-AVG-MAX mode activated, each point of this curve represents
the arithmetic mean of 60 values recorded every second. This display is therefore more precise, because there is no loss of in-
formation, but slower (see the table of figure 108).

To stop the calculation of the MIN-AVG-MAX mode, press .

To return to the Characteristics of the Record screen, press .

To place the cursor on the first oc- To place the cursor on the first oc-
currence of the minimum value. currence of the maximum value.

Figure 93: Vrms (N) without MIN-AVG-MAX

Pressing the or key automatically sets the zoom in to the highest level (display period identical to the recording period)
and de-activates the MIN-AVG-MAX mode if it was activated.

65
Curve of the maxima.

Values of the cursor (minimum,


Curve of the mean. mean, and maximum).

Curve of the minima.

Figure 94: Vrms (N) with MIN-AVG-MAX

The display period of this curve is one minute. Each point of the mean curve represents the arithmetic mean of 60 values recorded
every second. Each point of the curve of the maxima represents the maximum of the 60 values recorded every second. Each
point of the curve of the minima corresponds to the minimum of the 60 values recorded every second.

This display is therefore more precise than the previous one.

Date of the cursor. Position of the viewing window in


the record.
This screen is a partial view of the
trend curve. There are other screens
before and after the visible part. To select the display filter, press the
 or  key.
Use the  or  key to move the
cursor.

Figure 95: Vrms (L1) without MIN-AVG-MAX

For each of the phases (L1, L2, and L3), at each recording of a value over one second (recording period), the device also records
the minimum half-cycle RMS value over one second and the maximum half-cycle RMS value over one second. These are the
three curves shown in the figure above.

The MIN-AVG-MAX mode has been


activated.

Figure 96: Vrms (L1) with MIN-AVG-MAX

This curve differs slightly from the previous one because, with the MIN-AVG-MAX mode, there is no loss of information.

66
Note: For the quantities (W, Wdc, VAR, VA, VAD, PF, cos Φ and tan Φ) and for a three-phase source without neutral, only the total
quantities are represented.

Figure 97: tan Φ (L1) without MIN-AVG-MAX for a three-phase connection with neutral

Figure 98: tan Φ (L1) with MIN-AVG-MAX

The sum of the powers of the three


phases (Σ) is presented in bargraph
form.

To change the scale of the display


between 1 minute and 5 days.

Figure 99: W (Σ) without MIN-AVG-MAX

For the energy curves, the quantities are expressed in Wh, J, toe, or BTU, depending on the unit chosen in the configuration of
the device (see §4.5.2).

Figure 100: W (Σ) with MIN-AVG-MAX

This curve differs slightly from the previous one because, with the MIN-AVG-MAX mode, there is no loss of information.

The MIN-AVG-MAX mode can be activated for the powers to display, above the curve, the mean power value at the cursor date
along with the maximum and minimum power values in the display period. Note that, in contrast with the other quantities, only
the curve of mean values is represented.

67
Starting date of the selection. Date of the cursor (ending date of
the selection). Use the  or  keys
to move the cursor.

Energy calculation mode. Pressing


this key lets you define the start of
the selection.

Figure 101: Wh (Σ) without MIN-AVG-MAX

The display period of this bar chart is one minute. Since the recording period is one second, each bar of this bar chart represents
a value recorded in a one-second window once a minute.
The energy calculation mode determines the sum of the powers on the selected bars.

Figure 102: Wh (Σ) with MIN-AVG-MAX

With the MIN-AVG-MAX mode activated, the display differs slightly from the previous one because there is no loss of information.

Date of the cursor. Position of the viewing window in


the record.

This screen is a partial view of the


trend curve. There are other screens
before and after the visible part. To select the display filter, press the
 or  key.
Use the  or  keys to move the
cursor.

Figure 103: cos Φ (L1) without MIN-AVG-MAX

The period of display of this curve is two hours. Since the recording period is one second, each point of this curve represents a
value recorded every second taken every two hours. There is therefore a substantial loss of information (7,199 out of 7,200), but
the display is rapid.

The MIN-AVG-MAX mode has been


activated.

Figure 104: cos Φ (L1) with MIN-AVG-MAX

68
This curve differs considerably from the previous one, because the MIN-AVG-MAX mode is activated. Each point of the mean
curve represents the arithmetic mean of 7,200 values recorded every second. Each point of the curve of the maxima represents
the maximum of the 7,200 values recorded every second. Each point of the curve of the minima corresponds to the minimum of
the 7,200 values recorded every second.
This display is therefore more precise, because there is no loss of information, but slower (see the table in figure 108).

The user can stop the loading of


the recorded values at any time by
pressing this key.

Figure 105: cos Φ (L1) loading/calculation of values.

The dashes indicate that the value is


not available at the cursor position.

Figure 106: cos Φ (L1) loading/calculation of values aborted.

Display of the record is not complete because reading was stopped before the end.

To change the scale of the display


between 1 minute and 5 days.

Figure 107: cos Φ (L1) loading/calculation of complete values without MIN-AVG-MAX


for a three-phase connection with neutral.

The display has not been stopped and is therefore complete.

69
The following table indicates the time needed to display the curve on screen as a function of the width of the display window for
a recording period of one second:

Typical waiting time for Typical waiting time for


Width of display window display with the display with the
Grid increment
(60 points or increments) MIN-AVG-MAX mode MIN-AVG-MAX mode
deactivated activated
5 days 2 hours 11 seconds 10 minutes
2,5 days 1 hour 6 seconds 5 minutes
15 hours 15 minutes 2 seconds 1 minute 15 seconds
10 hours 10 minutes 2 seconds 50 seconds
5 hours 5 minutes 1 second 25 seconds
1 hour 1 minute 1 second 8 seconds
20 minutes 10 seconds 1 second 2 seconds
5 minutes 5 seconds 1 second 1 second
1 minute 1 second 1 second 1 second
Figure 108: Display Times table

These times can be long, so it has been made possible to stop the display at any time by pressing the key.

It is also possible, at any time:


„„ to press the or key to change the scale of the display,
„„ to press the  or  key to move the cursor,
„„ to press the  or  key to change the display filter.

But note that this may restart the loading and/or calculation of the values from the beginning.

70
10. Power and energy mode

The key displays power- and energy-related measurements.

The sub-menus available depend on the filter.


„„ For 2- and 3-wire single-phase connections and for the 2-wire two-phase connection, only selection L1 is available. The filter
is therefore not displayed, but the display is the same as for L1.
„„ For the 3-wire three-phase connection, only the Σ selection is available. The filter is therefore not displayed, but the display
is the same as for Σ.

10.1. 3L filter
10.1.1. The Powers display screen
The W... sub-menu is used to display the powers.

Active power (P).

DC power (only if a DC current


sensor is connected).

Reactive power (Q).

Distortion power (D).

Apparent power (S).

Figure 109: the 3L Powers screen.

Note: This screen corresponds to the choice “non-active quantities broken down” in the AR tab of the Calculation Methods menu
of the Configuration mode. If the choice had been “non-active quantities not broken down”, then the VAD label (distortion
power) would have disappeared and the VAR label would have corresponded to the non-active power (N). This non-active
power (N) is unsigned and has no inductive or capacitive effect.

10.1.2. The Quantities Associated with the Powers display screen


The PF sub-menu... displays the quantities associated with the powers.

Power factor.

Fundamental power factor (also


called displacement factor - DPF).

Tangent of the phase shift.

Phase shift of the voltage with


respect to the current.

Figure 110: the Quantities Associated with the Powers screen in 3L

71
10.1.3. The Energies Consumed display screen
The sub-menu displays the meters of energy consumed by the load.

Active energy.

DC energy (only if a DC current sen- Inductive reactive effect .


sor is connected).

Reactive energy.

Distortion energy. Capacitive reactive effect .

Apparent energy.

Figure 111: the Energies Consumed display screen in 3L

Note: This screen corresponds to the choice “non-active quantities broken down” in the VAR tab of the Calculation Methods
menu of the Configuration mode. If the choice had been “non-active quantities not broken down”, then the VADh label
(distortion energy) would have disappeared and the VARh label would have corresponded to the non-active energy (N).
This non-active energy has no inductive or capacitive effect.

10.1.4. The Energies Generated display screen


The sub-menu displays the meters of the energy generated by the load.

Active energy.

DC energy (only if a DC current sen-


sor is connected). Inductive reactive effect .

Reactive energy.

Distortion energy. Capacitive reactive effect .

Apparent energy.

Figure 112: the Energies Generated display screen in 3L

Note: This screen corresponds to the choice “non-active quantities broken down” in the VAR tab of the Calculation Methods
menu of the Configuration mode. If the choice had been “non-active quantities not broken down”, then the VADh label
(distortion energy) would have disappeared and the VARh label would have corresponded to the non-active energy (N).
This non-active energy has no inductive or capacitive effect.

10.2. Filters L1, L2 and L3


10.2.1. The Powers and Associated Quantities display screen
The W... sub-menu displays the powers and associated quantities.

Active power (P).


Power factor (PF).
DC power.
Fundamental power factor (also
Reactive power (Q). called displacement factor - DPF).

Distortion power (D). Tangent of the phase shift.

Apparent power (S).


Phase shift of voltage with respect
to current.
Figure 113: the Powers and Associated Quantities display screen in L1

72
Notes: This screen corresponds to the choice “non-active quantities broken down” in the VAR tab of the Calculation Methods
menu of the Configuration mode. If the choice had been “non-active quantities not broken down”, then the VAD label
(distortion power) would have disappeared and the VAR label would have corresponded to the non-active power (N). This
non-active power has no inductive or capacitive effect.
Filters L2 and L3 display the same information for phases 2 and 3.
ΦUA is displayed for the 2-wire two-phase set-up.

10.2.2. The energy meters display screen


The Wh... sub-menu displays the energy meters.

Meters of the energy consumed by Meters of the energy generated by


the load. the load.

Active energy.

DC energy (only if a DC current sen-


Inductive reactive effect .
sor is connected).

Reactive energy.

Distortion energy. Capacitive reactive effect .

Apparent energy.

Figure 114: the Energies Consumed and Generated display screen in L1

Notes: This screen corresponds to the choice «non-active quantities broken down» in the VAR tab of the Calculation Methods
menu of the Configuration mode. If the choice had been «non-active quantities not broken down», then the VADh label
(distortion energy) would have disappeared and the VARh label would have corresponded to the non-active energy (N).
This non-active energy has no inductive or capacitive effect.
Filters L2 and L3 display the same information for phases 2 and 3.

10.3. Filter Σ
10.3.1. The Powers and Associated Quantities display screen
The W... sub-menu displays the powers and associated quantities.

Active power (P).

DC power. Power factor.

Reactive power (Q). Fundamental power factor (also


called displacement factor - DPF).

Distortion power (D).


Tangent.
Apparent power (S).

Figure 115: the Powers and Associated Quantities display screen in Σ

Note: This screen corresponds to the choice “non-active quantities broken down” in the VAR tab of the Calculation Methods
menu of the Configuration mode. If the choice had been “non-active quantities not broken down”, then the VAD label
(distortion power) would have disappeared and the VAR label would have corresponded to the non-active power (N). This
non-active power has no inductive or capacitive effect.

73
10.3.2. The Energy meters display screen
The Wh... sub-menu displays the energy meters.

Meters of energy consumed by the Meters of the energy generated by


load. the load.

Active energy.

DC energy (only if a DC current sen- Inductive reactive effect .


sor is connected).

Reactive energy.
Capacitive reactive effect .
Distortion energy.

Apparent energy.

Figure 116: the Sums of the Energies Consumed and Generated display screen in Σ

Notes: This screen corresponds to the choice “non-active quantities broken down” in the VAR tab of the Calculation Methods
menu of the Configuration mode. If the choice had been “non-active quantities not broken down”, then the VADh label
(distortion energy) would have disappeared and the VARh label would have corresponded to the non-active energy (N).
This non-active energy has no inductive or capacitive effect.
For the 3-wire three-phase set-up, only the display of total quantities is available, the method of calculation of the powers
used is the two-wattmeter method (see Appendix, §16.1.4.2).

10.4. Starting energy metering


To start an energy metering, press the key in an energies display window ( , , or Wh...).

The start date and time of the energy


metering.

The icon is used to suspend the


energy metering.

Figure 117: the Energy Metering Start-up screen in Wh

The blinking symbol indicates


that energy metering is in progress.

Figure 118: the Energy Metering screen in Joules

The diagram used is the 4-quadrant diagram (see §16.5).

Note: The non-nullity threshold is 11.6 kWh for non-nuclear toe and 3.84 kWh for nuclear toe.

74
10.5. Disconnection of energy metering
To suspend energy metering, press .

The stop date and time of the meter-


ing are displayed alongside the start
date and time.

Figure 119: the Energy Metering screen in toe

A disconnection of the metering is not definitive. It can be resumed by pressing the key again.

Note: If no recording is in progress, then disconnecting the energy metering leads to the appearance of the blinking symbol
in the status bar (in place of the symbol). Disconnecting the energy metering also leads to the replacement of the
key by the key.

10.6. Reset of energy metering


To suspend the metering, press the key. Then, to reset energy metering, press the key and validate with the  key. All
energy values (consumed and generated) are then reset.

75
11. Screen snapshot mode

The key can be used to take up to 50 screen snapshots and display the recorded snapshots.

Saved screens may then be transferred to a PC using the PAT application (Power Analyser Transfer).

11.1. Screen snapshots


Press the key for approx. 3 seconds to shoot any screen .

When a screen snapshot is taken, the icon of the active mode ( , , , , , , ) in the top strip of the display
unit is replaced by the icon. You can then release the key .

The device can record only 50 screen snapshots. If you want to record a 51st screen, the device informs you that snapshots must
be deleted by displaying the icon in place of .

11.2. Handling of screen snapshots


To enter screen snapshot mode, briefly press the key. The device then displays a list of recorded screen snapshots.

Indicator of available image memory. Icon for browsing in screen pages.


The black bar represents memory
used and the white bar represents
memory available. Display of the list of screen snap-
shots.
List of saved snapshots. Each icon
represents the type of screen re-
Deleting a screen snapshot.
corded. It is followed by the date and
time of the screen snapshot.

Figure 120: the List of Screen Snapshots display screen

11.2.1. Viewing a snapshot from the list


To view a snapshot, select it in the list of screen snapshots using the , ,  and  keys. The date and time of the selected
snapshot are bolded.

Press  to display the selected snapshot. The icon is displayed in alternation with the icon for the mode that was active
when the snapshot was taken ( , , , , , , ).

Press to return to the list of screen snapshots.

11.2.2. Deleting a snapshot from the list


To delete a snapshot, select it in the list of screen snapshots using the , ,  and  keys. The date and time of the selected
snapshot are bolded.

Press the key and validate by pressing  . The snapshot then disappears from the list.
To cancel the deletion, press rather than  .

76
12. Help key

The key provides information about the key functions and symbols used in the current display mode.

The following information is displayed:

Reminder of the mode used.


Reminder of the current sub-mode.

List of information concerning the


keys.
Help page 2.

Help page 1.

Figure 121: The help screen for the powers and energies mode, page 1

List of symbols used on the page.

Figure 122: the screen of the help page for the powers and energies mode, page 2

77
13. Data export software

There are two data export software programs:


„„ PAT (Power Analyser Transfer), supplied with the device, used to transfer the data recorded in the device to a PC.
„„ Dataview, optional, also used to transfer the data, which are then presented in the form of a report compliant with your coun-
try’s standards.

To install one of the two software programs, load the installation CD in the CD drive of your PC, then follow the instructions on
screen.

Then connect the device to the PC using the USB cord supplied, after removing the cover that protects the USB port on the device.

Switch the device on by pressing the key and wait for your PC to detect it.

The PAT transfer software automatically defines the communication rate between the PC and the device.

Note: All measurements recorded in the device can be transferred to the PC. The transfer does not erase the recorded data un-
less the user explicitly asks it to.

For directions for using the data export software, use its Help function or refer to its user manual.

78
14. General specifications

14.1. Environmental conditions


The following chart shows conditions relating to ambient temperature and humidity:

%RH

95

85

75 1 = Reference range.
2 = Range for use.
3 2 1 4 3 = Range for storage with batteries.
4 = Range for storage without batteries.
45

10
°C
-20 0 20 26 35 42,5 50 70

Caution: at temperatures above 40°C, the device must be powered by the battery alone OR by the mains power unit alone; use
of the device with both the battery AND the mains power unit is prohibited.

Altitude :
Use < 2 000 m
Storage < 10 000 m

Degree of pollution : 2.

Use indoors.

14.2. Mechanical conditions


Dimensions (L x W x H) 200 mm x 250 mm x 70 mm
Weight approximately 2 kg
Screen 118 mm x 90 mm, diagonal 148 mm

Tightness IP 50 as per EN 60529


Electrical IP 2X on the terminals

Fall 1 m as per IEC 61010-1

14.3. Compliance with international standards


The device is compliant with IEC 61010-1, 600 V category IV or 1000 V category III.
„„ the use of AmpFLEX™ or of MiniFLEX or C clamps keeps the “device + current sensor” system at 600 V category IV or
1000 V category III.
„„ the use of PAC, MN93, MN93A, or E3N clamps downgrades the “device + current sensor” system to 300 V category IV
or 600 V category III.
„„ the use of the 5 A adapter unit downgrades the “device + current sensor” system to 150 V category IV or 300 V category
III.

Double insulation between the inputs/outputs and earth.


Double insulation between the voltage inputs, the power supply, and the other inputs/outputs.

79
14.4. Electromagnetic compatibility
Emissions and immunity in an industrial setting compliant with IEC 61326-1.

14.5. Power supply


14.5.1. Mains power supply
This is a specific 600 Vrms, category IV – 1000 Vrms, category III external mains power supply unit.
Range of use: 230 V ± 10 % @ 50 Hz and 120 V ± 10 % @ 60 Hz.
Maximum input power: 65 VA.

14.5.2. Battery power supply


The device is supplied by a 9.6 V, 4000 mAh battery pack, comprising 8 rechargeable NiMH elements.

Battery 8 NiMH storage cells.


Capacity 4000 mAh nominal.
Nominal voltage 1,2 V per cell, or a total of 9.6 V.
Life at least 300 charge-discharge cycles.
Charging current 1 A.
Charging time approx. 5 hours
Service T° [0 °C ; 50 °C].
Charging T° [10 °C ; 40 °C].
Storage T° storage ≤ 30 days: [-20 °C ; 50 °C].
storage for 30 to 90 days: [-20 °C ; 40 °C].
storage for 90 days to 1 year: [-20 °C ; 30 °C].

14.5.3. Consumption

With the display unit brightness set to 50% 320 mA


In display-off stand-by mode 130 mA


14.5.4. Battery life
Battery life is 10 hours when the battery delivered with the device is fully charged, with the display screen on. If the display screen
is off (saving energy), battery life is more than 25 hours.

80
15. Functional characteristics

15.1. Reference Conditions


This table indicates the reference conditions of the quantities to be used by default in the characteristics indicated in § 15.3.4.

Parameter Reference conditions


Ambient temperature 23 ± 3 °C
Humidity (relative humidity) [45 %; 75 %]
Atmospheric pressure [860 hPa ; 1060 hPa]
Phase-to-neutral voltage [50 Vrms ; 1000 Vrms] without DC (< 0.5 %)
[30 mVrms ; 1 Vrms] without DC (< 0.5 %) The values of A nom
Standard current circuit input voltage
„„ Anom ⇔ 1 Vrms are given in the table
(Except FLEX type current sensors)
„„ 3 × Anom ÷ 100 ó 30 mVrms below.
[11,73 mVrms ; 254.15 mVrms] without DC (< 0.5 %)
Rogowski current circuit input voltage
„„ 6500 Arms ⇔ 254.15 mVrms to 50 Hz
(FLEX type current sensors)
„„ 300 Arms ⇔ 11.73 mVrms to 50 Hz
Frequency of electrical network 50 Hz ± 0,1 Hz or 60 Hz ± 0,1 Hz
0° (active power and energy)
Phase shift
90° (reactive power and energy)
Harmonics < 0,1 %
Voltage unbalance < 10 %
Voltage ratio 1 (unity)
Current ratio 1 (unity)
Power supply Battery only
Electric field < 1 V.m-1
Magnetic field < 40 A.m-1

15.2. Nominal current according to type of sensor


Current sensor Nominal RMS current
(3 × Anom ÷ 100) [A]
(except FLEX) (Anom) [A]
C clamp 1000 30
PAC clamp 1000 30
MN93 clamp 200 6
MN93A clamp (100 A) 100 3
E3N clamp (10 mV/A) 100 3
E3N clamp (100 mV/A) 10 0.3
MN93A clamp (5 A) 5 0.15
5 A adapter 5 0.15

15.3. Electrical characteristics


15.3.1. Voltage input characteristics
Range for use: 0 Vrms to 1000 Vrms AC+DC phase-to-neutral and neutral-to-earth.
0 Vrms to 2000 Vrms AC+DC phase-to-phase.
(on condition of compliance with 1000 Vrms with respect to earth in Cat III).
Input impedance: 969 kW (between phase and neutral and between neutral and earth).
Admissible overload: 1200 Vrms constant.
2000 Vrms for one second.

81
15.3.2. Current input characteristics
Operating range: [0 V ; 1 V]
Input impedance: 1 MW.
Admissible overload: 1.7 Vrms constant.

FLEX type current sensors (AmpFLEX™ and MiniFLEX) switch the current input to an integrator assembly (‘Rogowski’ channel)
capable of interpreting the signals from Rogowski sensors. The input impedance is reduced to 12.4 kW in this case.

15.3.3. Bandwidth
Measurement channels: 256 points per period, i.e.:
„„ At 50 Hz: 6.4 kHz (256 × 50 ÷ 2).
„„ At 60 Hz: 7.68 kHz (256 × 60 ÷ 2).

The analogue 3-dB bandwidth exceeds 10 kHz.

15.3.4. Characteristics of the device alone (excluding the current sensor)

82
Quantities concerning currents and voltages
Measurement range without ratio
(with unity ratio) Display resolution Maximum intrinsic
Measurement
(with unity ratio) error
Minimum Maximum
Frequency 40 Hz 70 Hz 0.01 Hz ±(0.01 Hz)
0.1 V
±(0.5 % + 0.2 V)
V < 1000 V
simple 5 V 1,200 V (1)
1V
±(0.5 % + 1 V)
RMS V ≥ 1000 V
voltage(5) 0.1 V
±(0.5 % + 0.2 V)
U < 1000 V
compound 5 V 2,400 V (2)
1V
±(0.5 % + 1 V)
U ≥ 1000 V
0.1 V
±(1 % + 0.5 V)
V < 1000 V
simple 5 V 1,697  V (3)
1V
Direct ±(1 % + 1 V)
V ≥ 1000 V
voltage
(DC)(6) 0.1 V
±(1 % + 0.5 V)
U < 1000 V
compound 5 V 3,394 V (4)
1V
±(1 % + 1 V)
U ≥ 1000 V
0.1 A
±(0.5 % + 0.2 A)
C clamp A < 1000 A
1 A 1,200 A
PAC clamp 1A
±(0.5 % + 1 A)
A ≥ 1000 A
MN93 clamp 0.2 A 240 A 0.1 A ±(0.5 % + 0.2 A)
0.01 A
±(0.5 % + 0.02 A)
E3N clamp (10 mV/A) A < 100 A
0.1A 120 A
MN93A clamp (100 A) 0.1 A
±(0.5 % + 0.1 A)
A ≥ 100 A
RMS
current(5) 0.001 A
±(0.5 % + 0.002 A)
A < 10 A
E3N clamp (100 mV/A) 0.01 A 12 A
0.01 A
±(0.5 % + 0.01 A)
A ≥ 10 A
MN93A clamp (5 A)
0.005 A 6A 0.001 A ±(0.5 % + 0.002 A)
5 A adapter
0.1 A
AmpFLEX™ A < 1000 A
10 A 6500 A ±(0.5 % + 1 A)
MiniFLEX 1A
A ≥ 1000 A
0.1 A
A < 1000 A
PAC clamp 1A 1200 A (4) ±(1 % + 1 A)
1A
A ≥ 1000 A
0.01 A
Direct A < 100 A
current E3N clamp (10 mV/A) 0.1 A 169.7 A (3)
±(1 % + 0.1 A)
(DC)(6) 0.1 A
A ≥ 100 A
0.001 A
A < 10 A
E3N clamp (100 mV/A) 0.01 A 16.97 A (3)
±(1 % + 0.01 A)
0.01 A
A ≥ 10 A
±(1 % + 5 ct)
Peak factor (PF) CF < 4
1 9.99 0.01
(voltage and current) ±(5 % + 2 ct)
CF ≥ 4
(1) In 1,000 Vrms, category III, provided that the voltage between each of the terminals and earth does not exceed 1,000 Vrms.
(2) Two-phase (opposite phases) – same note as (1).
((3) 1200 x √2 ≈ 1697; 2400 x √2 ≈ 3394; 120 x √2 ≈ 169,7; 12 x √2 ≈ 16,97;
(4) Limitation of the PAC clamp.
(5) Total RMS value and RMS value of the fundamental
(6) DC harmonic component (n= 0)

83
Measurement range without ratio
(with unity ratio) Display resolution
Measurement Maximum intrinsic error
(with unity ratio)
Minimum Maximum
0.1 V
V < 1,000 V
simple 5 V 1,200 V (1) ±(0.8 % + 1 V)
1V
RMS ½ V ≥ 1,000 V
voltage 0.1 V
U < 1,000 V
compound 5 V 2,400 V (2)
±(0.8 % + 1 V)
1V
U ≥ 1,000 V
0.1 V
V < 1,000 V
simple 5 V 1,697  V (3) ±(3 % + 2 V)
1V
Peak V ≥ 1,000 V
voltage 0.1 V
U < 1,000 V
compound 5 V 3,394 V (3)
±(3 % + 2 V)
1V
U ≥ 1,000 V
0.1 A
C clamp A < 1,000 A
1 A 1,200 A ±(1 % + 1 A)
PAC clamp 1A
A ≥ 1,000 A
MN93 clamp 0.2 A 240 A 0.1 A ±(1 % + 1 A)
0.01 A
E3N clamp (10 mV/A) A < 100 A
0.1A 120 A ±(1 % + 0.1 A)
MN93A clamp (100 A) 0.1 A
A ≥ 100 A
RMS ½
current 0.001 A
A < 10 A
E3N clamp (100 mV/A) 0.01 A 12 A ±(1 % + 0.01 A)
0.01 A
A ≥ 10 A
MN93A clamp (5 A)
0.005 A 6A 0.001 A ±(1 % + 0.01 A)
5 A adapter
0.1 A
AmpFLEX™ A < 1,000 A
10 A 6,500 A ±(2.5 % + 5 A)
MiniFLEX 1A
A ≥ 1,000 A
1A
C clamp A < 1,000 A
1A 1,697 A (3)
±(1 % + 2 A)
PAC clamp 1A
A ≥ 1,000 A
MN93 clamp 0.2 A 339.4 A (3) 0.1 A ±(1 % + 2 A)
0.01 A
E3N clamp (10 mV/A) A < 100 A
0.1 A 169.7 A (3)
±(1 % + 0.2 A)
MN93A clamp (100 A) 0.1 A
A ≥ 100 A
Peak
current) 0.001 A
A < 10 A
E3N clamp (100 mV/A) 0.01 A 16.97 A (3) ±(1 % + 0.02 A)
0.01 A
A ≥ 10 A
MN93A clamp (5 A)
0.005 A 8.485 A (3) 0.001 A ±(1 % + 0.02 A)
5 A adapter
0.1 A
AmpFLEX™ A < 1,000 A
10 A 9,192 A (3) ±(3 % + 5 A)
MiniFLEX 1A
A ≥ 1,000 A
Severity of short-term flicker (PST) 0 12 0,01 See the corresponding table
Severity of long-term flicker (PLT) 0 12 0,01 Uncertainty of PST
(1) In 1,000 Vrms, category III, provided that the voltage between each of the terminals and earth does not exceed 1,000 Vrms.
(2) Two-phase (opposite phases) – same note as (1).
(3) 1200 x √2 ≈ 1697; 2400 x √2 ≈ 3394; 240 x √2 ≈ 339,4; 120 x √2 ≈ 169,7; 12 x √2 ≈ 16,97; 6 x √2 ≈ 8,485; 6500 x √2 ≈ 9192;

84
Quantities concerning powers and energies
Measurement range without ratio
(with unity ratio) Display resolution
Measurement Maximum intrinsic error
(with unity ratio)
Minimum Maximum
±(1 %)
cos F ≥ 0.8
Excluding FLEX
±(1.5 % + 10 ct)
Active 0.2 ≤ cos F < 0.8
5 mW (3)
7,800 kW (4)
4 digits at most (5)
power (1) ±(1 %)
AmpFLEX™ cos F ≥ 0.8
MiniFLEX ±(1.5 % + 10 ct)
0.5 ≤ cos F < 0.8
±(1 %)
sin F ≥ 0.5
Excluding FLEX
±(1.5 % + 10 ct)
Reactive 0.2 ≤ sin F < 0.5
5 mvar (3)
7,800 kvar (4)
4 digits at most (5)
power (2) ±(1.5 %)
AmpFLEX™ sin F ≥ 0.5
MiniFLEX ±(2.5 % + 20 ct)
0.2 ≤ sin F < 0.5
±(3 % + 20 ct)
Excluding FLEX if ∀ n ≥ 1, τn ≤(100 ÷ n) [%]
or
Distortion
5 mvar (3) 7,800 kvar (4) 4 digits at most (5) ±(2 % +(nmax × 0,5 %) + 50 ct)
power (7)
AmpFLEX™ THDA ≤ 20 %f
MiniFLEX ±(2 % +(nmax × 0,7 %) + 10 ct)
THDA > 20 %f
Apparent power 5 mVA (3) 7,800 kVA (4) 4 digits at most (5) ±(1 %)
±(1.5 %)
cos F ≥ 0.5
Peak factor (PF) -1 1 0.001
±(1.5 % + 10 ct)
0.2 ≤ cos F < 0.5
±(1 %)
Excluding FLEX
cos F ≥ 0.8
±(1.5 %)
Active 0.2 ≤ cos F < 0.8
1 mWh 9,999,999 MWh (6) 7 digits at most (5)
energy (1) AmpFLEX™ ±(1 %)
MiniFLEX cos F ≥ 0.8
±(1.5 %)
0.5 ≤ cos F < 0.8
±(1 %)
sin F ≥ 0.5
Excluding FLEX
±(1.5 %)
Reactive 0.2 ≤ sin F < 0.5
1 mvarh 9,999,999 Mvarh (6) 7 digits at most (5)
energy (2) ±(1.5 %)
AmpFLEX™ sin F ≥ 0.5
MiniFLEX ±(2 %)
0.2 ≤ sin F < 0.5
±(5.5 %)
Excluding FLEX
Distortion THDA ≤ 20 %f
1 mvarh 9,999,999 Mvarh (6) 7 digits at most (5)
energy AmpFLEX™ ±(1.5 %)
MiniFLEX THDA > 20 %f
Apparent energy 1 mVAh 9,999,999 MVAh (6) 7 digits at most (5) ±(1 %)
(1) The stated uncertainties on the active power and energy measurements are maxima for |cos F| = 1 and typical for the other phase differences.
(2) The stated uncertainties on the reactive power and energy measurements are maxima for |sin F| = 1 and typical for the other phase shifts.
(3) With MN93A clamp (5 A) or 5 A adapter.
(4) With AmpFLEX™ or MiniFLEX.
(5) The resolution depends on the current sensor used and on the value to be displayed.
(6) The energy is equivalent to more than 146 years of the associated maximum power (unity ratios).
(7) nmax is the highest order for which the harmonic ratio is non-zero.

85
Quantities associated with power
Measurement range
Measurement Display resolution Maximum intrinsic error
Minimum Maximum
Phase differences of funda-
-179° 180° 1° ±(2°)
mentals
±(1°) for F
cos F (DPF) -1 1 0.001
±(5 ct) for DPF
0.001
tan F < 10
tan F -32.77 (1) 32.77 (1) ±(1°) for F
0.01
tan F ≥ 10
Unbalance
(UNB)
0% 100 % 0.1 % ±(1 %)

(1) |tan F| = 32,767 corresponds to F = ±88.25° + k × 180° (k being a natural number)

86
Quantities concerning the spectral composition of the signals
Measurement range
Measurement Display resolution Maximum intrinsic error
Minimum Maximum
0.1 %
1500 %f tn < 1000 %
Voltage harmonic ratio (tn) 0% ±(2.5 % + 5 ct)
100 %r 1%
tn ≥ 1000 %
0.1 % ±(2 % + (n × 0.2 %) + 10 ct)
Current harmonic ratio (tn) 1500 %f tn < 1000 % n ≤ 25
0%
(non-FLEX) 100 %r 1% ±(2 % + (n × 0.6 %) + 5 ct)
tn ≥ 1000 % n > 25
0.1 % ±(2 % + (n × 0.3 %) + 5 ct)
Current harmonic ratio (tn) 1500 %f tn < 1000 % n ≤ 25
0%
(AmpFLEX™ & MiniFLEX) 100 %r 1% ±(2 % + (n × 0.6 %) + 5 ct)
tn ≥ 1000 % n > 25
Total voltage harmonic distortion (THD)
0% 999.9 % 0.1 % ±(2.5 % + 5 ct)
(referred to the fundamental) of voltage
±(2.5 % + 5 ct)
si ∀ n ≥ 1. tn ≤ (100 ÷ n) [%]

Total current harmonic distortion (THD) or


(referred to the fundamental) of current 0% 999.9 % 0.1 % ±(2 % + (nmax × 0.2 %) + 5 ct)
(non-FLEX) nmax ≤ 25
±(2 % + (nmax × 0.5 %) + 5 ct)
nmax > 25
±(2.5 % + 5 ct)
si ∀ n ≥ 1. tn ≤ (100 ÷ n2) [%]

Total current harmonic distortion (THD) or


(referred to the fundamental) of current 0% 999.9 % 0.1 % ±(2 % + (nmax × 0.3 %) + 5 ct)
(AmpFLEX™ & MiniFLEX) nmax ≤ 25
±(2 % + (nmax × 0.6 %) + 5 ct)
nmax > 25
Total voltage harmonic distortion (THD)
(referred to the signal without DC) of 0% 100 % 0.1 % ±(2.5 % + 5 ct)
voltage
±(2.5 % + 5 ct)
si ∀ n ≥ 1. tn ≤ (100 ÷ n) [%]

Total current harmonic distortion (THD) or


(referred to the signal without DC) of 0% 100 % 0.1 % ±(2 % + (nmax × 0.2 %) + 5 ct)
current (non-FLEX) nmax ≤ 25
±(2 % + (nmax × 0.5 %) + 5 ct)
nmax > 25
±(2.5 % + 5 ct)
si ∀ n ≥ 1. tn ≤ (100 ÷ n2) [%]

Total current harmonic distortion (THD) or


(referred to the signal without DC) of 0% 100 % 0.1 % ±(2 % + (nmax × 0.3 %) + 5 ct)
current (AmpFLEX™ & MiniFLEX) nmax ≤ 25
±(2 % + (nmax × 0.6 %) + 5 ct)
nmax > 25
±(5 % + (nmax × 0.3 %) + 5 ct)
nmax ≤ 25
Harmonic loss factor (FHL) 1 99.99 0.01
±(10 % + (nmax × 0.6 %) + 5 ct)
nmax > 25
±(5 % + (nmax × 0.3 %) + 5 ct)
nmax ≤ 25
K factor (FK) 1 99.99 0.01
±(10 % + (nmax × 0.6 %) + 5 ct)
nmax > 25
Phase shifts of harmonics (order ≥ 2) -179° 180° 1° ±(1.5° + 1° x (n ÷ 12.5)

Note : nmax is the highest order for which the harmonic ratio is non-zero.

87
Measurement range
(with unity ratio) Display resolution
Measurement Maximum intrinsic error
(with unity ratio)
Minimum Maximum
0.1 V
V < 1000 V
simple 5 V 1200 V (1) ±(2.5 % + 1 V)
RMS 1V
harmonic V ≥ 1000 V
voltage 0.1 V
(order n ≥ 2) U < 1000 V
compound 5 V 2400 V (2) ±(2.5 % + 1 V)
1V
U ≥ 1000 V
0.1 A ±(2 % + (n x 0.2%) + 1 A)
C clamp A < 1000 A n ≤ 25
1 A 1200 A
PAC clamp 1A ±(2 % + (n x 0.5%) + 1 A)
A ≥ 1000 A n > 25
±(2 % + (n x 0.2%) + 1 A)
n ≤ 25
MN93 clamp 0.2 A 240 A 0.1 A
±(2 % + (n x 0.5%) + 1 A)
n > 25
0.01 A ±(2 % + (n x 0.2%) + 0.1 A)
E3N clamp (10 mV/A) A < 100 A n ≤ 25
0.1A 120 A
RMS MN93A clamp (100 A) 0.1 A ±(2 % + (n x 0.5%) + 0.1 A)
harmonic A ≥ 100 A n > 25
current 0.001 A ±(2 % + (n x 0.2%) + 0.01 A)
(order n ≥ 2) A < 10 A n ≤ 25
E3N clamp (100 mV/A) 0.01 A 12 A
0.01 A ±(2 % + (n x 0.5%) + 0.01 A)
A ≥ 10 A n > 25
±(2 % + (n x 0.2%) + 0.01 A)
MN93A clamp (5 A) n ≤ 25
0.005 A 6A 0.001 A
5 A adapter ±(2 % + (n x 0.5%) + 0.01 A)
n > 25
0.1 A ±(2 % + (n x 0.3%) + 1 A + (Afrms(3) x 0.1%))
AmpFLEX™ A < 1000 A n ≤ 25
10 A 6500 A
MiniFLEX 1A ±(2 % + (n x 0.6%) + 1 A + (Afrms(3) x 0.1%))
A ≥ 1000 A n > 25
0.1 V
V < 1000 V
simple (Vd) 5 V 1200 V (1) ±(2.5 % + 1 V)
1V
RMS V ≥ 1000 V
distortion
voltage 0.1 V
U < 1000 V
compound (Ud) 5 V 2400 V (2)
±(2.5 % + 1 V)
1V
U ≥ 1000 V
0.1 A
C clamp A < 1000 A
1 A 1200 A ±((nmax x 0,4%) + 1 A)
PAC clamp 1A
A ≥ 1000 A
MN93 clamp 0.2 A 240 A 0.1 A ±((nmax x 0,4%) + 1 A)
0.01 A
E3N clamp (10 mV/A) A < 100 A
0.1A 120 A ±((nmax x 0,4%) + 0,1 A)
MN93A clamp (100 A) 0.1 A
RMS A ≥ 100 A
distortion
current (Ad) 0.001 A
(4) A < 10 A
E3N clamp (100 mV/A) 0.01 A 12 A ±((nmax x 0,4%) + 0,01 A)
0.01 A
A ≥ 10 A
MN93A clamp (5 A)
0.005 A 6A 0.001 A ±((nmax x 0,4%) + 0,01 A)
5 A adapter
0.1 A
AmpFLEX™ A < 1000 A
10 A 6500 A ±(nmax x 0,4%) + 1 A)
MiniFLEX 1A
A ≥ 1000 A
(1) In 1,000 Vrms, category III, provided that the voltage between each of the terminals and earth does not exceed 1,000 Vrms.
(2) Two-phase (opposite phases) – same note as (1).
(3) RMS value of the fundamental.
(4) nmax is the highest order for which the harmonic ratio is non-zero.

88
Severity of short-term flicker

Maximum intrinsic error of the short-term flicker severity measurement (PST)


Rectangular variations
120 V lamp 230 V lamp
per minute
60 Hz network 50 Hz network
(50% duty cycle)
1 PST ∈ [0.5 ; 4] ± 5% PST ∈ [0.5 ; 4] ± 5%
2 PST ∈ [0.5 ; 5] ± 5% PST ∈ [0.5 ; 5] ± 5%
7 PST ∈ [0.5 ; 7] ± 5% PST ∈ [0.5 ; 8] ± 5%
39 PST ∈ [0.5 ; 12] ± 5% PST ∈ [0.5 ; 10] ± 5%
110 PST ∈ [0.5 ; 12] ± 5% PST ∈ [0.5 ; 10] ± 5%
1620 PST ∈ [0.25 ; 12] ± 15% PST ∈ [0.25 ; 10] ± 15%

Range of current and voltage ratios

Ratio Minimum Maximum

100 9,999,900 x √3
Voltage
1,000 x √3 0,1

Current (1)
1 60,000 / 1

(1) Only for the MN93A clamp (5 A) and the 5 A adapter.

Measurement ranges after application of the ratios

Measurement range
Measurement Minimum Maximum
with minimum ratio(s) with maximum ratio(s)

RMS simple 58 mV 207.8 GV


& RMS ½ voltage compound 58 mV 415.7 GV

Direct voltage (DC) simple 58 mV 293.9 GV


& peak voltage compound 58 mV 587.9 GV
RMS & RMS ½ current 5 mA 360.0 kA
Peak current 5 mA 509.1 kA
Active power 0.289 mW 74.82 PW
Reactive power 0.289 mvar 74.82 Pvar
Distortion power 0.289 mvar 74.82 Pvar
Apparent power 0.289 mVA 74.82 PVA
Active energy 1 mWh 9,999,999 EWh (1)
Reactive energy 1 mvarh 9,999,999 Evarh (1)
Distortion energy 1 mvarh 9,999,999 Evarh (1)
Apparent energy 1 mVAh 9,999,999 EVAh (1)
(1) The energy corresponds to more than 15,000 years of the associated maximum power (maximum ratios).

89
15.3.5. Current sensor characteristics (after linearization)
Sensor errors are offset by a typical correction inside the device. This typical correction, applied to the phase and amplitude,
depends on the type of sensor connected (detected automatically) and the gain in the current acquisition channel used.

The RMS current measurement error and phase error are additional errors (which must therefore be added to the device errors),
indicated as influences on the calculations carried out by the analyser (powers, energies, power factors, tangents, etc.).

Type of sensor RMS current (Arms) Maximum error for Arms Maximum error for F

AmpFLEX™ A193 [10 A ; 100 A[ ±(3 %) ±(1°)


6,500 A [100 A ; 6,500 A] ±(2 %) ±(0.5°)
MiniFLEX MA193 [10 A ; 100 A[ ±(3 %) ±(1°)
6,500 A [100 A ; 6,500 A] ±(2 %) ±(0.5°)
[1 A ; 3 A[ -
±(0.8 %)
C193 clamp [3 A ; 10 A[ ±(1°)
1,000 A [10 A ; 100 A[ ±(0.3 %) ±(0.5°)
[100 A ; 1,200 A] ±(0.2 %) ±(0.3°)
[1 A ; 10 A[ -
±(1.5 % + 1 A)
PAC93 clamp [10 A ; 100 A[ ±(2°)
1,000 A [100 A ; 800 A[ ±(3 %)
±(1.5°)
[800 A ; 1,200 A] ±(5 %)
[0.5 A ; 2 A[ -
±(3 % + 1 A)
MN93 clamp [2 A ; 10 A[ ±(6°)
200 A [10 A ; 100 A[ ±(2.5 % + 1 A) ±(3°)
[100 A ; 240 A] ±(1 % + 1 A) ±(2°)
[100 mA ; 300 mA[ -
MN93A clamp ±(0.7 % + 2 mA)
[300 mA ; 1 A[ ±(1.5°)
100 A
[1 A ; 120 A] ±(0.7 %) ±(0.7°)
E3N clamp 100A [0 A ; 40 A[ ±(2 % + 50 mA)
±(0.5°)
Sensitivity 10 mV/A [40 A ; 100 A] ±(5 %)
E3N clamp 10A
[0 A ; 10 A] ±(1.5 % + 50 mA) ±(1°)
Sensitivity 100 mV/A
[5 mA ; 50 mA[ ±(1 % + 0.1 mA) ±(1.7°)
MN93A clamp
[50 mA ; 500 mA[ ±(1 %)
5A ±(1°)
[500 mA ; 6 A] ±(0.7 %)
[5 mA ; 50 mA[ ±(1 %) ±(1°)
5 A adapter
[50 mA ; 6 A] ±(0.5 %) ±(0°)



90
16. Appendices

This section presents the mathematical formulae used in calculating various parameters.

16.1. Mathematical formulae


16.1.1. Network frequency and sampling
Sampling is controlled by (locked to) the network frequency so as to deliver 256 samples per cycle from 40 Hz to 70 Hz. This
locking is essential for many calculations, including reactive power, distorting power, fundamental power factor, unbalance, and
harmonic factors and angles.

The instantaneous frequency is measured by analysing eight consecutive positive-going zero crossings on the signal in question
after digital low-pass filtering and digital suppression of the DC component (i.e. 7 periods filtered). The time of the zero crossing
is determined precisely by linear interpolation between two samples to achieve resolution better than 0.002%.

The device is capable of calculating an instantaneous frequency simultaneously on each of the 3 phases in voltage (phase-to-
neutral for distribution systems with neutral and phase-to-phase for distribution systems without neutral) or in current. It then
chooses one from among two or three of them as the official instantaneous frequency.

The network frequency over one second is determined (approximately) as the reciprocal of the arithmetic mean of the instantane-
ous periods.

The signals are acquired using a 16-bit converter and (for current acquisition) dynamic gain switches.

16.1.2. Waveform mode


16.1.2.1. RMS values of half-cycle voltage and current (excluding neutral)
Half-cycle RMS phase-to-neutral voltage of phase (i+1) with i ∈ [0; 2].
( Zéro suivant ) −1
1
Vdem[i ] =
NechDemPer
⋅ ∑ V [i][n]
n = Zéro
2

Half-cycle RMS phase-to-phase voltage of phase (i+1) with i ∈ [0 ; 2].


( Zéro suivant ) −1
1
Udem[i ] =
NechDemPer
⋅ ∑ U [i][n]
n = Zéro
2

Half-cycle RMS current of phase (i+1) with i ∈ [0 ; 2].


( Zéro suivant ) −1
1
Adem[i ] =
NechDemPer
⋅ ∑ A[i][n]
n = Zéro
2

Notes: these values are calculated for each half-cycle so as not to miss any fault.
NechDemPer is the number of samples in the half cycle.

16.1.2.2. Minimum and maximum half-cycle RMS values (excluding neutral)


Minimum and maximum RMS phase-to-neutral voltages of phase (i+1) with i ∈ [0 ; 2].
Vmax [i] = max(Vdem[i]), Vmin[i] = min(Vdem[i])

Minimum and maximum RMS phase-to-phase voltages of phase (i+1) with i ∈ [0 ; 2].
Umax [i] = max(Udem[i]), Umin[i] = min(Udem[i])

Minimum and maximum RMS currents of phase (i+1) with i ∈ [0 ; 2].


Amax [i] = max(Adem[i]), Amin[i] = min(Adem[i])

Note: The duration of the evaluation is left to the user’s discretion (reset by pressing the  key in the MAX-MIN mode).

91
16.1.2.3. Severity of Short-term flicker – 10 minutes (excluding neutral)
Method based on the IEC61000-4-15 standard.

The input values are half-period RMS voltages (phase-to-neutral for distribution systems with neutral, phase-to-phase for distribu-
tion systems without neutral). Blocks 3 and 4 are generated digitally. The classifier of block 5 has 128 levels.

The value of PST[i] is updated every 10 minutes (phase (i+1) with i ∈ [0; 2]).

Note: The user can reset the PST calculation by pressing the  key in the Summary mode.

16.1.2.4. Severity of LONG-term flicker – 2 hours (excluding neutral)


Method based on the IEC61000-4-15 standard.
1

3 ∑ PST [i][n]3

PLT [i ] = n =0
12
12

The values of PST[i][n] are consecutive and 10 minutes apart. The value of PLT[i] (phase (i+1) with i ∈ [0; 2]) calculated in a two-
hour window is updated either:
„„ every 10 minutes (Sliding long-term flicker – Configuration >Calculation methods >PLT)
„„ or every 2 hours (Non-sliding long-term flicker – Configuration >Calculation methods >PLT)

Note: The user can reset the PLT calculation by pressing the  key in the Summary mode.

16.1.2.5. Peak values (neutral included except Upp and Upm – reassessment every second
Positive and negative phase-to-neutral peak voltages of phase (i+1) with i ∈ [0; 3] (i = 3 ⇔ neutral).
Vpp[i] = max(V[i][n]), Vpm[i] = min(V[i][n]) n ∈ [0; N]

Positive and negative phase-to-phase peak voltages of phase (i+1) with i ∈ [0 ; 2].
Upp[i] = max(U[i][n]), Upm[i] = min(U[i][n]) n ∈ [0 ; N]

Positive and negative peak currents of phase (i+1) with i ∈ [0; 3] (i = 3 ⇔ neutral).
App[i] = max(A[i][n]), Apm[i] = min(A[i][n]) n ∈ [0; N]

Note: The duration of the evaluation is left to the user’s discretion (reset by pressing the  key in the MAX-MIN mode).

16.1.2.6. Peak factors (neutral included except Ucf – over one second)
Peak factor of phase-to-neutral voltage of phase (i+1) with i ∈ [0; 3] (i = 3 ⇔ neutral).
max( Vpp[i], Vpm[i] )
Vcf [i ] =
1 NechSec −1
⋅ ∑ V [i ][n]
2

NechSec n =0

Peak factor of phase-to-phase voltage of phase (i+1) with i ∈ [0 ; 2].


max( Upp[i], Upm[i] )
Ucf [i ] =
1 NechSec −1
⋅ ∑ U [i ][n]
2

NechSec n =0

Peak factor of current of phase (i+1) with i ∈ [0; 3] (i = 3 ⇔ neutral).


max( App[i], Apm[i] )
Acf [i ] =
1 NechSec −1
⋅ ∑ A[i ][n]
2

NechSec n =0

Note: NechSec is the number of samples in the second. The duration of evaluation of the peak values here is one second.

92
16.1.2.7. Rms values (neutral included except Urms – over one second)
RMS phase-to-neutral voltage of phase (i+1) with i ∈ [0; 3] (i = 3 ⇔ neutral).
NechSec −1
1
Vrms[i ] =
NechSec
⋅ ∑ V [i][n]
n =0
2

RMS phase-to-phase voltage of phase (i+1) with i ∈ [0 ; 2].


NechSec −1
1
Urms[i ] =
NechSec
⋅ ∑ U [i][n]
n =0
2

RMS current of phase (i+1) with i ∈ [0; 3] (i = 3 ⇔ neutral).


NechSec −1
1
Arms[i ] =
NechSec
⋅ ∑ A[i][n]
n =0
2

Note: NechSec is the number of samples in the second.

16.1.2.8. Reverse unbalances (three-phase connection – over one second)


These are calculated from the filtered RMS vector values (over one second) VFrms[i] and AFrms[i] for distribution systems with
neutral and UFrms[i] and AFrms[i] for distribution systems without neutral. (Ideally the fundamental vectors of the signals).

j
Note: The operations are vector operations in complex notation with a = e 3

Forward phase-to-neutral voltage (vector) in a distribution system with neutral


1
Vrms + = (VFrms[0] + a ⋅ VFrms[1] + a 2 ⋅ VFrms[2])
3

Reverse phase-to-neutral voltage (vector) in a distribution system with neutral


1
Vrms − = (VFrms[0] + a 2 ⋅ VFrms[1] + a ⋅ VFrms[2])
3

Phase-to-neutral voltage unbalance in a distribution system with neutral


Vrms −
Vunb =
Vrms +

Forward phase-to-phase voltage (vector) in a distribution system without neutral


1
Urms + = ( UFrms[0] + a ⋅ UFrms[1] + a 2 ⋅ UFrms[2])
3

Reverse phase-to-phase voltage (vector) in a distribution system with neutral


1
Urms − = ( UFrms[0] + a 2 ⋅ UFrms[1] + a ⋅ UFrms[2])
3

Phase-to-phase voltage unbalance in a distribution system with neutral


Urms −
Uunb =
Urms +

Forward current (vector)


1
Arms + = (AFrms[0] + a ⋅ AFrms[1] + a 2 ⋅ AFrms[2])
3

Reverse current (vector)


1
Arms − = (AFrms[0] + a 2 ⋅ AFrms[1] + a ⋅ AFrms[2])
3

93
Current unbalance (vector)
Arms −
Aunb =
Arms +

16.1.2.9. Fundamental rms values (excluding neutral – over one second)


These are calculated from the filtered vector (instantaneous) values. A digital filter made up of 6 low-pass order 2 infinite pulse
response Butterworth filters and one high-pass order 2 infinite pulse response Butterworth filter is used to extract the fundamental
components.

16.1.2.10. Fundamental angular values (excluding neutral – over one second)


These are calculated from the filtered vector (instantaneous) values. A digital filter made up of 6 low-pass order 2 infinite pulse
response Butterworth filters and one high-pass order 2 infinite pulse response Butterworth filter is used to extract the fundamental
components. The angular values calculated are those between:
„„ 2 phase-to-neutral voltages
„„ 2 line currents
„„ 2 phase-to-phase voltages
„„ One phase-to-neutral voltage and one line current (distribution systems with neutral)
„„ One phase-to-phase voltage and one line current (2-wire two-phase distribution systems)

16.1.3. Harmonic mode


16.1.3.1. FFT (neutral included except for Uharm and VAharm – over 4 consecutive periods every second)
These calculations are carried out by FFT (16 bits), 1024 points over four cycles, with a rectangular window (see IEC61000-4-7).
From the real parts bk and the imaginary parts ak, the harmonic factor is calculated for each order (j) and for each phase (i) Vharm[i]
[j], Uharm[i][j] and Aharm[i][j] with respect to the fundamental and the angles Vph[i][j], Uph[i][j] and Aph[i][j] with respect to the
fundamental. For the neutral-earth voltage and the neutral current, the level of harmonics is calculated for each order (j) Vharm[3]
[j] and Aharm[3][j] with respect to the total RMS value (AC+DC) of the complete signal (the harmonic angles are not calculated).

Note: The calculations are performed sequentially: {V1; A1} then {V2; A2} then {V3; A3} then {UN ; AN} then {U1; U2} and finally
{U3}. In the case of a 2-wire two-phase distribution source, the couple {V1; A1} is replaced by the couple {U1; A1}.

ck
The level in % with respect to the fundamental [% f] ⇔ τ k = 100
c4

ck
The level in % with respect to the total RMS value [% r] ⇔ τ k = 5050
100


m =0
C 42m

 ak 
The angle with respect to the fundamental in degrees [°] ⇔ ϕ k = arctan  −ϕ4

 bk 

c = bk + ja = a k2 + bk2
 k k
 1024
1  kπ 
bk

= ∑
512 s =0
Fs ⋅ sin 
 512
s +ϕk 


with  1024
1  kπ 
ak

= ∑
512 s =0
Fs ⋅ cos
 512
s +ϕk 

 1024
c 1
 0 = ∑
1024 s =0
Fs

k k
ck is the amplitude of the component of order m = with a frequency f k = f 4 .
4 4
Fs is the sampled signal at the fundamental frequency f 4 .

94
co is the DC component.
k
k is the index of the spectral spike (the order of the harmonic component is m = ).
4

Note: The power harmonic factors are calculated by multiplying the phase-to-neutral voltage harmonic factors by the current
harmonic factors. The power harmonic angles (VAharm[i][j] and VAph[i][j]) are calculated by differentiating the phase-to-
neutral voltage harmonic angles with the current harmonic angles. In the case of a 2-wire two-phase distribution source,
the phase-to-neutral voltage V1 is replaced by the phase-to-phase voltage U1 and one obtains the harmonic power levels
UAharm[0][j] and the harmonic power angles UAph[0][j].

16.1.3.2. Harmonic distortions


Two global values giving the relative quantity of the harmonic are calculated:
„„ the THD as a proportion of the fundamental (also noted THD-F),
„„ the THD as a proportion of the total RMS-AC value (also noted THD-R).

Total harmonic distortion of phase (i+1) with i ∈ [0; 2] (THD-F)


50
50 50
50 50
50

∑Vharm[i][n] ∑Uharm[i][n] ∑ Aharm[i][n]


2 2 2

Vthdf [i] = n=2


, Uthdf [i] = n=2
, Athdf [i] = n=2

Vharm[i ][1] Uharm[i ][1] Aharm[i ][1]

Total harmonic distortion of channel (i+1) with i ∈ [0; 3] (THD-R).


50
50 50
50 50
50

∑Vharm[i][n] ∑Uharm[i][n] ∑ Aharm[i][n]


2 2 2

Vthdr[i] = n=2
50
50
, Uthdr[i] = n=2
50
50
, Athdr[i] = n=2
50
50

∑Vharm[i][n] ∑Uharm[i][n] ∑ Aharm[i][n]


2 2 2

n =1 n =1 n =1

The THD as a proportion of the RMS-AC value (THD-R) is also called the distortion factor (DF).

16.1.3.3. Harmonic loss factor (without neutral – over 4 consecutive periods every second)
Harmonic loss factor of the phase (i+1) with i ∈ [0; 2]
n =50
50

∑n ⋅ Aharm[i ][n]
2 2

FHL[i] = n =1
n =50
50

∑ Aharm[i ][n]
2

n =1

16.1.3.4. K factor (excluding neutral – over 4 consecutive cycles every second)


K factor for phase (i+1) with i ∈ [0; 2], e ∈ [0.05; 0.1] and q ∈ [1.5; 1.7]
n =5050

∑n ⋅ Aharm[i ][n]
q 2

e
FK [i] = 1 +
FK . n=2
n =5050
1+ e
∑ Aharm[i][n]
2

n =1

16.1.3.5. Sequence harmonics (over 3 × (4 consecutive cycles) every second)


Negative-sequence harmonics
7

1
2 ∑ Aharm[i][3 j + 2]

j =0
Aharm − =
3 i =0
Aharm[i ][1]

95
Three-phase systems with neutral
7

1
2 ∑ Vharm[i][3 j + 2]

j =0
Vharm − =
3 i =0
Vharm[i ][1]

Three-phase systems without neutral


7

1
2 ∑ Uharm[i][3 j + 2]

j =0
Uharm − =
3 i =0
Uharm[i ][1]

Zero-sequence harmonics
7

1
2 ∑ Aharm[i][3 j + 3]

j =0
Aharm 0 =
3 i =0
Aharm[i ][1]

Three-phase systems with neutral


7

1
2 ∑ Vharm[i][3 j + 3]

j =0
Vharm 0 =
3 i =0
Vharm[i ][1]

Three-phase systems without neutral


7

1
2 ∑ Uharm[i][3 j + 3]

j =0
Uharm 0 =
3 i =0
Uharm[i ][1]

Positive -sequence harmonics


7

1
2 ∑ Aharm[i][3 j + 4]

j =0
Aharm + =
3 i =0
Aharm[i ][1]

Three-phase systems with neutral


7

1
2 ∑ Vharm[i][3 j + 4]

j =0
Vharm + =
3 i =0
Vharm[i ][1]

Three-phase systems without neutral


7

1
2 ∑ Uharm[i][3 j + 4]

j =0
Uharm + =
3 i =0
Uharm[i ][1]

16.1.4. Power
Powers without neutral – over one second

16.1.4.1. Distribution system with neutral


Active power of phase (i+1) with i ∈ [0; 2].
NechSec −1
1
W[i ] =
NechSec
⋅ ∑ V [i][n]⋅ A[i][n]
n =0

96
DC power of phase (i+1) with i ∈ [0 ; 2].
Wdc[i ] = Vdc[i ]⋅ Adc[i ]

Apparent power of phase (i+1) with i ∈ [0; 2].

Reactive power of phase (i+1) with i ∈ [0; 2] (Non-active quantities broken down).
NechSec −1
1 NechPer
VAR[i ] = VARF[i ] =
NechSec
⋅ ∑n =0
VF
V
F [i ][ n −
4
F [i ][n]
] ⋅ AF
A

Distortion power of the phase (i+1) with i ∈ [0; 2] (Non-active quantities broken down).
D [i] = VA[i] 2 − W[i ]2 − VARF[i ]2
VAD
A

Non-active power of phase (i+1) with i ∈ [0;2] (Non-active quantities broken down).
R [i] = VA
VAR
A A [i ]2 − W[i ]2
V

Total active power


W[3] = W[0] + W[1] + W[2]

Total DC power
Wdc[3] = Wdc[0] + Wdc[1] + Wdc[2]

Total apparent power


VA[3] = VA[0] + VA[1] + VA[2]

Total reactive power (Non-active quantities broken down)


VAR[3] = VARF[3] =VARF[0] + VARF[1] + VARF[2]

Total distortion power (Non-active quantities broken down)


D [3] = VA
VAD
A A [3]2 − W[3]2 − VARF[3]2
V

Total non-active power (Non-active quantities non broken down)


R [3] = VA
VAR
A A [3]2 − W[3]2
V

16.1.4.2. Three-phase system without neutral


Three-phase distribution systems without neutral are considered as a whole (no phase-by-phase power calculation). The device
therefore displays only the total quantities.

The two-wattmeter method (Aron method or two-element method) is applied for the calculation of the total active power, of the
total reactive power and of the total DC power.

a) Reference in L1
Active power, Wattmeter 1
NechSec −1
1
W[0] =
NechSec
⋅ ∑ U [2][n]⋅ A[2][n]
n =0

Active power, Wattmeter 2


NechSec −1
1
W[1] =
NechSec
⋅ ∑ − U [0][n]⋅ A[1][n]
n =0

Reactive power, Wattmeter 1


NechSec −1
1 NechPer
VARF[0] =
NechSec
⋅ ∑
n =0
UF
U
F [2][ n −
4
F [2][n]
] ⋅ AF
A

97
Reactive power, Wattmeter 2
NechSec −1
1 NechPer
VARF[1] =
NechSec
⋅ ∑
n =0
− UF
U
F [0][ n −
4
F [1][n]
AF
]⋅ A

DC power, Wattmeter 1
Wdc[0] = Udc[2]⋅ Adc[2]

DC power, Wattmeter 2
Wdc[1] = −Udc[0]⋅ Adc[1]

b) Reference in L2
Active power, Wattmeter 1
NechSec −1
1
W[0] =
NechSec
⋅ ∑ U [0][n]⋅ A[0][n]
n =0

Active power, Wattmeter 2


NechSec −1
1
W[1] =
NechSec
⋅ ∑ − U [1][n]⋅ A[2][n]
n =0

Reactive power, Wattmeter 1


NechSec −1
1 NechPer AF
VARF[0] =
NechSec
⋅ ∑
n =0
UF
U
F [0][ n −
4
F [0][n]
]⋅ A

Reactive power, Wattmeter 2


NechSec −1
1 NechPer AF
VARF[1] =
NechSec
⋅ ∑
n =0
−UF
U
F [1][ n −
4
F [2][n]
]⋅ A

DC power, Wattmeter 1
Wdc[0] = Udc[0]⋅ Adc[0]

DC power, Wattmeter 2
Wdc[1] = −Udc[1]⋅ Adc[2]

c) Reference in L3
Active power, Wattmeter 1
NechSec −1
1
W[0] =
NechSec
⋅ ∑ − U [2][n]⋅ A[0][n]
n =0

Active power, Wattmeter 2


NechSec −1
1
W[1] =
NechSec
⋅ ∑ U [1][n]⋅ A[1][n]
n =0

Reactive power, Wattmeter 1


NechSec −1
1 NechPer AF
VARF[0] =
NechSec
⋅ ∑
n =0
UF
−U
F [2][ n −
4
F [0][n]
]⋅ A

Reactive power, Wattmeter 2


NechSec −1
1 NechPer AF
VARF[1] =
NechSec
⋅ ∑
n =0
UF
U
F [1][ n −
4
F [1][n]
]⋅ A

98
DC power, Wattmeter 1
Wdc[0] = −Udc[2]⋅ Adc[0]

DC power, Wattmeter 2
Wdc[1] = Udc[1]⋅ Adc[1]

d) Calculation of the total quantities


Total active power
W[3] = W[0] + W[1]

Total DC power
Wdc[3] = Wdc[0] + Wdc[1]

Total apparent power


1
A [3] =
VA
V U rms 2 [0] + U rms 2 [1] + U rms 2 [2] Arms 2 [0] + Arms 2 [1] + Arms 2 [2]
3

Note: This is the total apparent RMS power as defined in IEEE 1459-2010 for distribution systems without neutral.

Total reactive power (Non-active quantities broken down – Configuration >Methods of calculation >VAR)
R [3] = V ARF[3] = V ARF[0] + V ARF[1]
VAR
A

Total distortion power (Non-active quantities broken down – Configuration >Calculation methods >VAR)
D [3] = VA[3]2 − W[3]2 − VARF[3]2
VAD
A

Total non-active power (Non-active quantities not broken down – Configuration >Calculation methods >VAR)
R [3] = VA
VAR
A A [3]2 − W[3]2
V

16.1.4.3. Two-phase systems without neutral


Two-phase distribution systems without neutral (2-wire two-phase) are treated as single-phase distribution systems having their
voltage reference on L2 rather than N (neutral).

Active power
NechSec −1
1
W[0] =
NechSec
⋅ ∑ U [0][n]⋅ A[0][n]
n =0

DC power
Wdc[0] = Udc[0]⋅ Adc[0]

Apparent power
A [0] = Urms[0]⋅ Arms[0]
VA
V

Reactive power (Non-active quantities broken down – Configuration >Calculation methods >VAR)
NechSec −1
1 NechPer
VAR[0] = VARF[0] =
NechSec
⋅ ∑n =0
U
F [0][ n −
UF
4
F [0][n]
A
] ⋅ AF

Distortion power (Non-active quantities broken down – Configuration >Calculation methods >VAR)
D [0] = VA[0] 2 − W[0]2 − VARF[0]2
VAD
A

Non-active power (Non-active quantities not broken down – Configuration >Calculation methods >VAR)
R [0] = V
VAR
A A [0]2 − W[0]2
VA

99
16.1.5. Power ratios (excluding neutral – over one second)
a) Distribution system with neutral

Power Factor of phase (i+1) with i ∈ [0; 2].

Fundamental power factor of phase (i+1) or cosine of the angle of the phase-to-neutral voltage fundamental of phase (i+1) with
respect to the current fundamental of phase (i+1) with i ∈ [0; 2]

Note: The fundamental power factor is also called the displacement factor.

Tangent of phase (i+1) or tangent of the angle of the phase-to-neutral voltage fundamental of phase (i+1) with respect to the cur-
rent fundamental of phase (i+1) with i ∈ [0; 2]

Total power factor


W [3]
PF [3] =
PF
VA
V
A [3]

Total fundamental power factor


F [3]
WF
W
DPF[3] =
F [3]2 + VARF [3]2
WF
W

With :
NechSec −1 NechSec −1 NechSec −1
F [3] =
WF
W ∑VF
n =0
F [0][n] + ∑VF
F [0][n] ⋅ AF
V A V F [1][n] + ∑VF
F [1][n] ⋅ AF
A
n =0
F [2][n] ⋅ AF
V F [2][n]
A
n =0

NechSec −1 NechSec −1 NechSec −1


NechPer NechPer NechPer
VARF [3] = ∑
n=0
VF
V
F [0][ n −
4
F [0][n] +
] ⋅ AF
A ∑
n=0
V
F [1][ n −
VF
4
F [1][n] +
] ⋅AF
A ∑
n=0
VF [2][ n −
VF
4
F [2][n]
] ⋅ AF
A

Note: The fundamental power factor is also called the displacement factor.

Total tangent
VARF[3]
Tan [3] =
F [3]
WF
W

b) Three-phase system without neutral

Total power factor


W [3]
PF [3] =
PF
V
A [3]
VA

Total fundamental power factor


F [3]
WF
W
DPF[3] =
F [3]2 + VARF[3]2
WF
W

100
With :
If reference on L1
NechSec −1 NechSec −1
1 1
F [3] =
WF
W
NechSec
⋅ ∑
n =0
U [2][n]⋅ A[2][n] +
NechSec
⋅ ∑ − U [0][n]⋅ A[1][n]
n =0

If reference on L2
NechSec −1 NechSec −1
1 1
F [3] =
W
WF
NechSec
⋅ ∑ U [0][n]⋅ A[0][n] + NechSec ⋅ ∑ − U [1][n]⋅ A[2][n]
n =0 n =0

If reference on L3
NechSec −1 NechSec −1
1 1
F [3] =
WF
W
NechSec
⋅ ∑
n =0
− U [2][n]⋅ A[0][n] +
NechSec
⋅ ∑ U [1][n]⋅ A[1][n]
n =0

Note: The fundamental power factor is also called the displacement factor.

Total tangent
VARF[3]
Tan [3] =
F [3]
WF
W

c) Two-phase system without neutral

Two-phase distribution systems without neutral (2-wire two-phase) are treated as single-phase distribution systems having their
voltage reference on L2 rather than N (neutral).

Power factor
W [0]
PF [0] =
PF
VA
V
A [0]

Fundamental power factor


F [0]
WF
W
DPF[0] =
F [0]2 + VARF[0]2
WF
W

With :
NechSec −1
1
F [0] =
WF
W
NechSec
⋅ ∑UF
F [0][n]⋅ AF
U
n =0
F [0][n]
A

Note: The fundamental power factor is also called the displacement factor.

Tangent
VARF[0]
Tan [0] =
F [0]
WF
W

101
16.1.6. Energies
Energies excluding neutral – over Tint with refresh every second

16.1.6.1. Distribution system with neutral


Note: Tint is the power integration time in energy calculations; the beginning and end of this period are user-controlled.

Consumed DC energy of phase (i+1) with i ∈ [0; 2]


Wdc[i ][n]
Tint
Wdch[0][i ] = ∑ n
3600
with Wdc[i][n] ≥ 0

Total consumed DC energy


Wdch[0][3] = Wdch[0][0] + Wdch[0][1] + Wdch[0][2]

a) Energies consumed other than DC (W[i][n] ≥ 0)


Consumed active energy of phase (i+1) with i ∈ [0; 2].
W [i ][n]
Tint
h [0][i ] =
W
Wh
∑n
3600

Consumed apparent energy of phase (i+1) with i ∈ [0; 2].


A [i ][n]
Tint
V
VAh[0][i ] = ∑ VA3600
n

Consumed inductive reactive energy of phase (i+1) with i ∈ [0; 2].


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
VARF [i ][n]
Tint
VARhL[0][i ] = ∑ n
3600
with VARF[i][n] ≥ 0

Consumed capacitive reactive energy of phase (i+1) with i ∈ [0; 2].


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
− VARF [i ][n]
Tint
VARhC[0][i ] = ∑ n
3600
with VARF[i][n] < 0

Consumed distortion energy of phase (i+1) with i ∈ [0; 2]


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
VAD[i ][n]
Tint
VADh[0][i ] = ∑ n
3600

Consumed non-active energy of phase (i+1) with i ∈ [0; 2]


(Non-active quantities not broken down – Configuration >Calculation methods >VAR)
VAR[i ][n]
Tint
VARh[0][i ] = ∑ n
3600

Total consumed active energy


Wh[0][3] = Wh[0][0] + Wh[0][1] + Wh[0][2]

Total consumed apparent energy


VAh[0][3] = VAh[0][0] + VAh[0][1] + VAh[0][2]

Total consumed reactive inductive energy


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
VARhL[0][3] = VARhL[0][0] + VARhL[0][1] + VARhL[0][2]

102
Total consumed capacitive reactive energy
(Non-active quantities broken down – Configuration >Calculation methods >VAR)
VARhC[0][3] = VARhC[0][0] + VARhC[0][1] + VARhC[0][2]
Total consumed distortion energy
(Non-active quantities broken down – Configuration >Calculation methods >VAR)
VADh[0][3] = VADh[0][0] + VADh[0][1] + VADh[0][2]

Total consumed non-active energy


(Non-active quantities not broken down – Configuration >Calculation methods >VAR)
VARh[0][3] = VARh[0][0] + VARh[0][1] + VARh[0][2]

b) Generated DC energy of phase (i+1) with i ∈ [0; 2]


− Wdc[i ][n]
Tint
Wdch[1][i ] = ∑ n
3600
with Wdc[i][n] < 0

c) Total generated DC energy


Wdch[1][3] = Wdch[1][0] + Wdch[1][1] + Wdch[1][2]

d) Energies generated other than DC (W[i][n] < 0)


Generated active energy of phase (i+1) with i ∈ [0; 2].
− W [i ][n]
Tint
h [1][i ] =
Wh
W ∑n
3600

Generated apparent energy of phase (i+1) with i ∈ [0; 2].


A [i ][n]
Tint
VA
V
VAh[1][i ] = ∑ n
3600

Generated inductive reactive energy of phase (i+1) with i ∈ [0; 2].


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
− VARF [i ][n]
Tint
VARhL[1][i ] = ∑ n
3600
with VARF[i][n] < 0

Generated capacitive reactive energy of phase (i+1) with i ∈ [0; 2].


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
VARF [i ][n]
Tint
VARhC[1][i ] = ∑ n
3600
with VARF[i][n] ≥ 0

Generated distortion energy of phase (i+1) with i ∈ [0; 2]


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
VAD[i ][n]
Tint
VADh[1][i ] = ∑ n
3600

Generated non-active energy of phase (i+1) with i ∈ [0; 2]


(Non-active quantities not broken down – Configuration >Calculation methods >VAR)
VAR[i ][n]
Tint
VARh[1][i ] = ∑ n
3600

Total generated active energy


Wh[1][3] = Wh[1][0] + Wh[1][1] + Wh[1][2]

Total generated apparent energy


VAh[1][3] = VAh[1][0] + VAh[1][1] + VAh[1][2]

103
Total generated inductive reactive energy
(Non-active quantities broken down – Configuration >Calculation methods >VAR)
VARhL[1][3] = VARhL[1][0] + VARhL[1][1] + VARhL[1][2]

Total generated capacitive reactive energy


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
VARhC[1][3] = VARhC[1][0] + VARhC[1][1] + VARhC[1][2]

Total generated distortion energy


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
VADh[1][3] = VADh[1][0] + VADh[1][1] + VADh[1][2]

Total generated non-active energy


(Non-active quantities not broken down – Configuration >Calculation methods >VAR)
VADh[1][3] = VADh[1][0] + VADh[1][1] + VADh[1][2]

16.1.6.2. Distribution system without neutral


We speak here only of total energies, with:
„„ Three-phase systems without neutral: i = 3
„„ Two-phase systems without neutral: i = 3 or i = 0 (this is the same thing - see remark below)

Note: Two-phase distribution systems without neutral (2-wire two-phase) are treated as single-phase distribution systems having
their voltage reference on L2 rather than N (neutral).

Total consumed DC energy


Wdc[i ][n]
Tint
Wdch[0][i ] = ∑ n
3600
with Wdc[i][n] ≥ 0

a) Total energies consumed other than DC (W[i][n] ≥ 0)


Total consumed active energy
W [i ][n]
Tint
h [0][i ] =
Wh
W ∑n
3600

Total consumed apparent energy


A [i ][n]
Tint
VA
V
VAh[0][i ] = ∑ n
3600

Total consumed inductive reactive energy


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
VARF [i ][n]
Tint
VARhL[0][i ] = ∑ n
3600
with VARF[i][n] ≥ 0

Total consumed capacitive reactive energy


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
− VARF [i ][n]
Tint
VARhC[0][i ] = ∑ n
3600
with VARF[i][n] < 0

Total consumed distortion energy


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
VAD[i ][n]
Tint
VADh[0][i ] = ∑ n
3600

104
Total consumed non-active energy
(Non-active quantities not broken down – Configuration >Calculation methods >VAR)
VAR[i ][n]
Tint
VARh[0][i ] = ∑ n
3600

b) Total generated DC energy


− Wdc[i ][n]
Tint
Wdch[1][i ] = ∑ n
3600
with Wdc[i][n] < 0

c) Total energies generated other than DC (W[i][n] < 0)


Total generated active energy
− W [i ][n]
Tint
h [1][i ] =
Wh
W ∑n
3600

Total generated apparent energy


A [i ][n]
Tint
VA
V
VAh[1][i ] = ∑ n
3600

Total generated inductive reactive energy


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
− VARF [i ][n]
Tint
VARhL[1][i ] = ∑ n
3600
with VARF[i][n] < 0

Total generated capacitive reactive energy


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
VARF [i ][n]
Tint
VARhC[1][i ] = ∑ n
3600
with VARF[i][n] ≥ 0

Total generated distortion energy


(Non-active quantities broken down – Configuration >Calculation methods >VAR)
VAD[i ][n]
Tint
VADh[1][i ] = ∑ n
3600

Total generated non-active energy


(Non-active quantities not broken down – Configuration >Calculation methods >VAR)
VAR[i ][n]
Tint
VARh[1][i ] = ∑ n
3600

105
16.2. Distribution sources supported by the device
See connections in §4.6.

16.3. Hysteresis
Hysteresis is a screening principle that is often used after detection of a threshold stage in Alarm mode (see §4.10) and in
Inrush current mode (see §5.2). A correct hysteresis setting avoids repeated changes of state when the measurement oscillates
about the threshold.

16.3.1. Surge detection


With a hysteresis of 2%, for example, the return level for surge detection is equal to (100% - 2%) or 98% of the reference volt-
age threshold.

Maximum

Threshold
Hysteresis

Return level

Duration

16.3.2. Undervoltage or blackout detection


With a hysteresis of 2%, for example, the return level for undervoltage detection is equal to (100% + 2%) or 102% of the Uref
voltage threshold.

Duration
Return level
Hysteresis

Threshold

Minimum

16.4. Minimum scale values for waveforms and minimum RMS values
Minimum scale value
Minimum RMS values
(waveform mode)
Phase-to-neutral and phase-to-phase voltages 8 V (1) 5 V (1)
AmpFLEX™ 90 A 10 A
MiniFLEX 90 A 10 A
C clamp 8A 1A
PAC clamp 8A 1A
MN93 clamp 2A 0.2 A
MN93A clamp (100 A) 0.8 A 0.1 A
E3N clamp (10 mV/A) 0.8 A 0.1 A
E3N clamp (100 mV/A) 0.08 A 0.01 A
MN93A clamp (5 A) 0.04 A (1)
0.005 A (1)
5 A adapter 0.04 A (1) 0.005 A (1)
(1) Value to be multiplied by the ratio in effect (if not unity).

106
16.5. Four-quadrant diagram
This diagram is used for power and energy measurements (see §9).

Figure 123: Four-quadrant diagram

Here VAR is the fundamental reactive power (and not the non-active power).

16.6. Mechanism for triggering transient captures


The sampling rate is a constant 256 samples per cycle. When a transient capture is started, each sample is compared to the
sample from the preceding cycle. The preceding cycle defines the mid-point of the trigger envelope and is used as reference. As
soon as a sample is outside the envelope, the triggering event occurs; the representation of the transient is then captured by the
device. The cycle preceding the event and the three following cycles are saved to memory.

Here is a graphic representation of the transient capture triggering mechanism:

Reference period

Top of envelope

Cycle monitored

Bottom of envelope

Triggering event

The half-width of the envelope for the voltage and current is equal to the threshold programmed in the Transient mode of the
configuration (see §4.8).

16.7. Capture conditions in Inrush Current mode


The capture depends on a triggering (start) event and a stop event. If a capture ends with a stop event or if the recording memory
of the device is full, the capture stops automatically.

The capture stop threshold is calculated as follows:


[Stop threshold [A]] = [Start threshold [A]] × (100 – [stop hysteresis [%]]) ÷ 100

107
Here are the conditions for triggering and stopping captures:

Triggering filter Triggering and stop conditions

Triggering condition ⇔ [A1 half-cycle RMS value] > [Triggering threshold]


A1
Stop condition ⇔ [A1 half-cycle RMS value] < [Stop threshold]
Triggering condition ⇔ [A2 half-cycle RMS value] > [Triggering threshold]
A2
Stop condition ⇔ [A2 half-cycle RMS value] < [Stop threshold]
Triggering condition ⇔ [A3 half-cycle RMS value] > [Triggering threshold]
A3
Stop condition ⇔[A3 half-cycle RMS value] < [Stop threshold]
Triggering condition ⇔ [[the half-cycle RMS value of one current channel]> [Triggering threshold]
3A
Stop condition ⇔ [the half-cycle RMS values of all current channels] < [Stop threshold]

16.8. Glossary
AC and DC components.
AC component only.
DC component only.
Φ Phase shift of the phase-to-neutral voltage with respect to the phase-to-neutral current.
Inductive phase shift.
Capacitive phase shift.
° Degree.
-.+ Expert mode.
| | Absolute value.
Σ Values of the system.
% Percentage.
%f Fundamental value as reference
%r Total value as reference
ΦVA Phase shift of the voltage with respect to the current.
A Current or the unit “ampere”.
A-h Harmonics in current.
Acf Peak factor of the current.
Ad RMS distortion current.
Adc DC current.
Ah Harmonic of the current.
Apk+ Maximum peak value of the current.
Apk- Minimum peak value of the current.
Arms RMS current.
Athd Total harmonic distortion of the current.
Athdf Harmonic distortion of the current with the RMS value of the fundamental as reference.
Athdr Harmonic distortion of the current with the total RMS value without DC as reference.
Aunb Unbalance in current.
AVG Mean value (arithmetic mean).
Blackout: fall of voltage, at some point in the power network, to below a specified threshold.
BTU British Thermal Unit.
CF Peak factor (Crest Factor) in current or voltage: ratio of the peak value of a signal to the RMS value.
Channel and phase: a measurement channel corresponds to a difference in potential between two conductors. A phase is a
single conductor. In polyphased systems, a measurement channel may be between two phases, or between a
phase and neutral, or between a phase and earth, or between neutral and earth.
cos Φ Cosine of the phase shift of the voltage with respect to the current (displacement factor – DPF).
Dip threshold: specific voltage used to define the beginning and end of a voltage dip.
DC DC component (current or voltage).
DPF Displacement factor (cos Φ).
E Exa (1018 )

108
FK K factor. Used to quantify the effect of a load on a transformer.
FHL Harmonic loss factor.
Flicker a visual effect of voltage variations.
Frequency number of full voltage or current cycles in one second.
Fundamental component: component at the fundamental frequency.
G Giga (109 )
Harmonics in electrical systems, voltages and currents at frequencies that are multiples of the fundamental frequency.
Hysteresis difference between thresholds for reciprocal changes of state.
Hz Frequency of the network.
k kilo (103 )
L Channel (Line).
m milli (10-3 )
ms millisecond.
M Mega (106 )
MAX Maximum value.
MIN Minimum value.
Nominal voltage: Reference voltage of a network.
non-nuclear toe: non-nuclear tonne oil equivalent.
nuclear toe: nuclear tonne oil equivalent.
Order of a harmonic: ratio of the frequency of the harmonic to the fundamental frequency; a whole number.
P Peta (1015 )
Passband range of frequencies in which the response of a device exceeds some specified minimum.
PF Power Factor: ratio of active power to apparent power.
Phase temporal relationship between current and voltage in alternating current circuits.
PK or PEAK. Maximum (+) or minimum (-) peak value of the signal.
PLT Long-term severity. The device calculates its PLT over 2 hours.
PST Short-term severity. The device calculates its PST over 10 minutes.
RMS RMS (Root Mean Square) value of current or voltage. Square root of the mean of the squares of the instantaneous
values of a quantity during a specified interval.
Temporary surge at industrial frequency: temporary increase in voltage at some point in the power network, to above a speci-
fied threshold.
t Relative date of the Time cursor.
T Tera (1012 )
tan Φ Tangent of the phase shift of the voltage with respect to the current.
THD Total Harmonic Distortion. The total harmonic distortion describes the proportion of the harmonics of a signal with
respect to the fundamental RMS value (%f) or with respect to the total RMS value without DC (%r).
U Phase-to-phase voltage.
U-h Harmonics in phase-to-phase voltage.
Ucf Peak factor of the phase-to-phase voltage.
Ud Phase-to-phase RMS distortion voltage.
Udc Phase-to-phase DC voltage.
Uh Harmonic of the phase-to-phase voltage.
Upk+ Maximum peak value of phase-to-phase voltage.
Upk- Minimum peak value of phase-to-phase voltage.
Urms Phase-to-phase RMS voltage.
Uthd Total harmonic distortion of the phase-to-phase voltage.
Uthdf Harmonic distortion of the phase-to-phase voltage with the RMS value of the fundamental as reference.
Uthdr Harmonic distortion of the phase-to-phase voltage with the total RMS value without DC as reference.
Uunb Unbalance in phase-to-phase voltage.
V Phase-to-neutral voltage or the unit “volt”.
V-h Harmonics in phase-to-neutral voltage
VA Apparent power.
VA-h Harmonics in power.
VAD Distortion power.

109
VADh Distortion energy.
VAh Apparent energy.
VAR Reactive or non-active power.
VARh Reactive or non-active energy.
Vcf Peak factor of the voltage.
Vd Phase-to-neutral RMS distortion voltage.
Vdc Phase-to-neutral DC voltage.
Vpk+ Maximum peak value of the phase-to-neutral voltage.
Vpk- Minimum peak value of the phase-to-neutral voltage.
Vh Harmonic of the phase-to-neutral voltage.
Voltage dip: temporary fall of voltage, at some point in the power network, to below a specified threshold.
Voltage unbalance in a polyphased electric power network (UNB - unbalance): state in which the RMS voltages between
conductors (fundamental component) and/or the phase differences between successive conductors are not all equal.
Vrms RMS phase-to-neutral voltage.
Vthd Total harmonic distortion of phase-to-neutral voltage.
Vthdf Total harmonic distortion of phase-to-neutral voltage with the fundamental RMS value as reference.
Vthdr Total harmonic distortion of phase-to-neutral voltage with the total RMS value without DC as reference.
Vunb Phase-to-neutral voltage unbalance.
W Active power.
Wdc DC power.
Wdch Dc energy.
Wh Watt-hour.
Wh Active energy.

110
17. MAINTENANCE

Except for the battery and the memory card, the instrument contains no parts that can be replaced by personnel
who have not been specially trained and accredited. Any unauthorized repair or replacement of a part by an “equivalent”
may gravely impair safety.

17.1. Cleaning the casing


Disconnect the unit completely and switch it OFF.

Use a soft cloth, dampened with soapy water. Rinse with a damp cloth and dry rapidly with a dry cloth or forced air. Do not use
alcohol, solvents, or hydrocarbons.

17.2. Maintenance of sensors


Current sensors must be maintained and calibrated as follows:
„„ Clean with a sponge and soapy water, rinse with a sponge and clean water, and dry rapidly.
„„ Keep the air gaps of the clamps (MN93, MN93A, C193, PAC93 and E3N) perfectly clean using a cloth. Lightly oil visible metal
parts to avoid rust.

17.3. Replacing the battery


For safety reasons, replace the battery only with the original model (see §19.3).

„„ Do not throw the battery into a fire.


„„ Do not expose the battery to a temperature in excess of 100°C.
„„ Do not short-circuit the terminals of the battery pack.

Remove the old battery.


„„ To eliminate all risk of electric shock, disconnect the power supply cord and measurement leads of the device.
„„ Turn the device over, raise the stand, and prop it up against the small yellow stops.
„„ Use a coin to unscrew the two quarter-turn screws on the back of the housing.

Small yellow stops.

„„ Using a flat screwdriver, remove the cover from the compartment.

111
„„ Turn the device over and hold the battery as it slides out of its compartment.
„„ Disconnect the battery connector without pulling on the wires.

Note: The Qualistar+ preserves the date-time function for approximately 24 hours without the battery.

Spent batteries must not be treated as ordinary household waste. Take them to the appropriate recycling collection point.

Fitting the new battery.


„„ Connect the new battery. The connector is error-proofed to prevent reversals of polarity.
„„ Place the battery in its compartment and arrange the wires so that they do not protrude.
„„ Put the battery compartment cover back in place and screw the 2 quarter-turn screws back in.

Warning: If the battery is disconnected, it must then be fully recharged, even if it is not replaced, so that the device will know the
battery charge condition (this information is lost when the battery is disconnected).

17.4. Replacing the screen film


Proceed as follows to replace the screen film of the device:
„„ Remove the old screen film.
„„ Remove the protection plastic film from the new screen film using the white tab.
„„ Place the adhesive side of the film against the screen of the device. Smooth the film with a clean cloth to remove any air bubbles.

112
17.5. Memory card
Before withdrawing or inserting the memory card, make sure that the device is disconnected and off.

17.6. Metrological check


Like all measuring or testing devices, the instrument must be checked regularly.

This instrument should be checked at least once a year. For checking and calibration, contact one of our accredited metrology
laboratories (information and contact details available on request), at our Chauvin Arnoux subsidiary or the branch in your country.

Note: The Information sub-menu of the Configuration menu displays the date of adjustment and the date of the next adjustment,
as in the example below:

Figure 36: the About menu

17.7. Repair
For all repairs before or after expiry of warranty, please return the device to your distributor.

17.8. Updating of the internal software


With a view to providing, at all times, the best possible service in terms of performance and technical upgrades, Chauvin Arnoux
invites you to update the embedded software of the device by downloading the new version, available free of charge on our web site.

Our site:
http://www.chauvin-arnoux.com
Sign in and open your account.
Then go to “Software support space”, then “Freely available software”, then “C.A 8335”.

Connect the device to your PC using the type A-B USB cord provided.

The embedded software update requires compatibility with the hardware version of the device, indicated in the About sub-menu
of the Configuration menu (see figure 36 above).

Warning: the update of the embedded software erases all data (configuration, alarm campaigns, snapshots, inrush current cap-
ture, transient detection, trend recordings. Save any data you want to keep to a PC using the PAT software (see §13)
before updating the embedded software.

113
18. Warranty

Except as otherwise stated, our warranty is valid for three years starting from the date on which the equipment was sold. Extract
from our General Conditions of Sale provided on request.

The warranty does not apply in the following cases:


„„ Inappropriate use of the equipment or use with incompatible equipment;
„„ Modifications made to the equipment without the explicit permission of the manufacturer’s technical staff;
„„ Work done on the device by a person not approved by the manufacturer;
„„ Adaptation to a particular application not anticipated in the definition of the equipment or not indicated in the user’s manual;
„„ Damage caused by shocks, falls, or floods.

114
19. To order

19.1. C.A 8335 three-phase electrical networks analyser


C.A 8335 without clamp ........................................................................................................................................... P01160577
C.A 8335 MN .............................................................................................................................................................. P01160571
C.A 8335 MN93A ........................................................................................................................................................ P01160572
C.A 8335 AMP450 ...................................................................................................................................................... P01160573
C.A 8335 AMP800 ...................................................................................................................................................... P01160574
C.A 8335 PAC ............................................................................................................................................................. P01160575
C.A 8335 C193 ............................................................................................................................................................ P01160576
C.A 8335 MiniFLEX .................................................................................................................................................... P01160581

The device is delivered with:


„„ 1 no. 22 shoulder bag
„„ 5 black straight-straight safety cables 3 m long.
„„ 5 black crocodile clips.
„„ 1 specific PA 30 W mains power unit with mains cord.
„„ 1 set of 12 inserts and rings to identify voltage phases and leads and current phases and leads.
„„ 1 USB A/B cord, 1.80m long, with ferrite.
„„ 1 Power Analyser Transfer (PAT) software.
„„ 1 checking attestation.
„„ user’s manuals on CD (one per language)
„„ safety sheets (one per language)

and, when clamps are included:


„„ 4 current sensors (one type from among the 7 possible).

19.2. Accessories
5 A adapter unit (three-phase) ..................................................................................................................................... P01101959
MN93 clamp ................................................................................................................................................................ P01120425B
MN93A clamp .............................................................................................................................................................. P01120434B
PAC93 clamp ............................................................................................................................................................... P01120079B
C193 clamp ................................................................................................................................................................. P01120323B
AmpFLEX™ A193 450 mm ......................................................................................................................................... P01120526B
AmpFLEX™ A193 800 mm ......................................................................................................................................... P01120531B
MiniFLEX MA193 250 mm ........................................................................................................................................... P01120580
E3N clamp ................................................................................................................................................................... P01120043A
E3N clamp adapter ..................................................................................................................................................... P01102081
Mains power unit + E3N clamp ................................................................................................................................... P01120047
Dataview software ....................................................................................................................................................... P01102095

19.3. Spare parts


9.6 V 4 Ah NiMH battery pack ..................................................................................................................................... P01296024
USB-A USB-B cord ..................................................................................................................................................... P01295293
PA30W mains power unit ............................................................................................................................................ P01102057
Screen protection film ................................................................................................................................................. P01102059
No. 22 carrying bag...................................................................................................................................................... P01298056
No. 21 carrying bag...................................................................................................................................................... P01298055
Set of 5 black banana-banana straight-straight safety cables ......................................................................... Please consult us
Set of 5 black crocodile clips ............................................................................................................................ Please consult us
Set of 12 inserts and rings to identify voltage phases and leads and current phases and leads ............................... P01102080

115
11 - 2012

Code 692272D02 - Ed. 1

DEUTSCHLAND - Chauvin Arnoux GmbH SCHWEIZ - Chauvin Arnoux AG


Straßburger Str. 34 - 77694 Kehl / Rhein Moosacherstrasse 15 - 8804 AU / ZH
Tel: (07851) 99 26-0 - Fax: (07851) 99 26-60 Tel: 044 727 75 55 - Fax: 044 727 75 56
ESPAÑA - Chauvin Arnoux Ibérica S.A. UNITED KINGDOM - Chauvin Arnoux Ltd
C/ Roger de Flor, 293 - 1a Planta - 08025 Barcelona Unit 1 Nelson Ct - Flagship Sq - Shaw Cross Business Pk
Tel: 902 20 22 26 - Fax: 934 59 14 43 Dewsbury, West Yorkshire - WF12 7TH
Tel: 01924 460 494 - Fax: 01924 455 328
ITALIA - Amra SpA MIDDLE EAST - Chauvin Arnoux Middle East
Via Sant’Ambrogio, 23/25 - 20050 Macherio (MI) P.O. BOX 60-154 - 1241 2020 JAL EL DIB (Beirut) - LEBANON
Tel: 039 245 75 45 - Fax: 039 481 561 Tel: (01) 890 425 - Fax: (01) 890 424
ÖSTERREICH - Chauvin Arnoux Ges.m.b.H CHINA - Shanghai Pu-Jiang - Enerdis Instruments Co. Ltd
Slamastrasse 29/2/4 - 1230 Wien 3 F, 3 rd Building - N° 381 Xiang De Road - 200081 SHANGHAI
Tel: 01 61 61 9 61-0 - Fax: 01 61 61 9 61-61 Tel: +86 21 65 21 51 96 - Fax: +86 21 65 21 61 07
SCANDINAVIA - CA Mätsystem AB USA - Chauvin Arnoux Inc - d.b.a AEMC Instruments
Box 4501 - SE 18304 TÄBY 200 Foxborough Blvd. - Foxborough - MA 02035
Tel: +46 8 50 52 68 00 - Fax: +46 8 50 52 68 10 Tel: (508) 698-2115 - Fax: (508) 698-2118

http://www.chauvin-arnoux.com

190, rue Championnet - 75876 PARIS Cedex 18 - FRANCE


Tél. : +33 1 44 85 44 85 - Fax : +33 1 46 27 73 89 - [email protected]
Export : Tél. : +33 1 44 85 44 38 - Fax : +33 1 46 27 95 59 - [email protected]

You might also like