Diaphragm Pumps For In-Vitro Diagnostics
Diaphragm Pumps For In-Vitro Diagnostics
Diaphragm Pumps For In-Vitro Diagnostics
In-Vitro Diagnostics
6311
6313
www.gd-thomas.com
Diaphragm Pumps for
In-Vitro Diagnostics
P3 6311 Series
P6 6313 Series
RETURN TO CONTENTS 2
6311 Series
6311 Series
TYPICAL APPLICATIONS
Diaphragm
Diaphragm
RETURN TO CONTENTS 3
Diaphragm Liquid
Diaphragm Liquid Pump 6311BLDC
Pump6311 BLDC
Flow 190 ml/min
Pressure height 40 m H2O
17
Suction height 3 m H2O
0)
(0.1 7)
2.5 .1
3x .3 (0
4 5 (0.2)
t=
14.5 (0.57)
22.5 (0.89)
9.5 (0.37)
Housing: Molex 22-01-2045
M5 (2x) Terminals: Molex 08-50-0031 Wire
length: 400 ± 10 (15.75 +/- 0.39)
5)
(1.2
Dimensions in mm (inch)
.7
31
41.9 (1.65)
37.2 (1.46)
16 (0.63)
400 +/- 10 (15.75 +/- 0.39)
65.4 (2.57)
ELECTRICAL DATA
GENERAL DATA
Ambient temperature 15 to 40 °C
Media temperature 5 to 55 °C
Weight 130 g
Pneumatic connection thread size M5
The technical data is based on the use of PTFE measuring tubes Shore D 55, 6 x 1 mm
(O.D. x wall thickness), 65 cm length / 8 x 1 mm, 15 cm length / 6 x 1 mm, 10 cm length.
150 150
100 100
50 50
0 0
0 5 10 15 20 25 30 35 40 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0
Pressure height [m H2O] Control voltage [V]
RETURN TO CONTENTS 4
Wiringdiagram
Wiring diagram(Brushless
(BrushlessDC
DCmotor)
motor)
Function Colour Control
12 V DC (10 - 18 V, reverse polarity protected)
Pin 1 V Supply Red
24 V DC (10 - 28 V, reverse polarity protected)
Pin 2 Ground Black Ground for V supply (PIN 1) and Speed Control (PIN 3)
Pin 3 Speed Control White Speed control range 0 - 5 V | PWM: 6 - 20 kHz
18 pulses per revolution
Pulse time “LOW”: 195 µs
Pin 4 Tacho out Green
Output level “LOW”: 0 / Max. 0.5 V
Output level “HIGH”: Min. 4 V / Max. 5 V
74HC...
FG 3K9
Vin 10K 68K IC
470p
10K
56K
100n
0
t
195µs
Options / Accessories
RETURN TO CONTENTS 5
3
6313 Series
6313 Series
TYPICAL APPLICATIONS
Diaphragm
Diaphragm
RETURN TO CONTENTS 6
Diaphragm
Diaphragm Liquid Pump6313
Liquid Pump 6313BLDC
BLDCStandard
Standard
Max. liquid flow 0.35 l/min FLOW CURVES
Max. gas flow 2.3 l/min
Pneumatic performance
Pressure height (liquid) 10.0 m H2O Flow rate [l/min]
2.5
Final vacuum (gas) 70 %
63130002 (EPDM) Gas
2.0
Intended use Aspiration of liquids 63130004 (PTFE) Gas
1.5
1.0
0.5
0
-1000 -800 -600 -400 -200 0 200 400 600 800 1000 1200 1400
Pressure [mbar rel.]
Hydraulic performance
Flow rate [l/min]
0.4
63130002 (EPDM) Liquid
0.3 63130004 (PTFE) Liquid
0.2
0.1
0
0 1 2 3 4 5 6 7 8 9 10
Pressure [m H2O]
Max. flow [l/min] 0.35 Pin 1 V Supply Red 12 V DC (max. range 8.0 - 14.0 V)
Max. cont. pressure [m H2O] 10.0 24 V DC (max. range 8.0 - 26.4 V)
Max. suction height [m H2O] 7.0 Pin 2 Tacho out Green 3 pulses / cycle, signal current 2mA
- Output level ‚Low‘: min. 0V / max. 0.5V
PNEUMATIC DATA* - Output level ‚High‘: min. 2V / max. 6V
Max. flow [l/min] 2.3 (EPDM) / 1.6 (PTFE) l/min Pin 3 Rot. direction Yellow Input voltage range 0 - 5.5V
Max. cont. pressure [mbar rel.] 1.100 - Input level ‚Low‘: min. 0V / max. 0.5V (CW)
0,297
7,55
2,880
73,15
RETURN TO CONTENTS 7
GENERAL CONTACT
Please check out all our brands for your mission-critical flow control technologies:
OEM Pumps Vacuum Pumps & Systems Syringe Pumps Lab Automation Syringes
www.gd-thomas.com www.welchvacuum.com www.tricontinent.com www.zinsser-analytic.com www.microsyringes.com