DI-60AE Tiger
Digital Panel Meter
• Modular construction with interchangeable input and ✔ TIGER FAMILY: More than 75 different Plug-in
output modules. I-Series Input Signal Conditioners are approved for
• Able to receive single and dual input signals from one of more Texmate’s Tiger Family of meters.
than 75 input signal conditioners. See www.texmate.com for an up to date listing.
• Optional output module that can support a relay module with up to
six relays; an analog output module with user selectable isolated Description
voltage (0-10 VDC) or isolated current (0-20 mA); and a serial out-
put module with either isolated RS-232, isolated RS-485, or non-
The Texmate Tiger is a super-intelligent digital panel meter with pro-
isolated TTL logic level serial output.
grammable logic and advanced digital signal processing capabilities.
• Auto-sensing high or optional low voltage AC / DC power supply.
Its innovative modular design creates a highly flexible meter able to
• Built-in 5 / 10 / 24 VDC sensor excitation voltage.
process almost any input signal type from one of more than 75 sin-
• Optional red or green, 7- or 15-segment, 0.56” high LEDs gle and dual input signal conditioner types. With relay, analog, and
• Easy to program using front panel buttons or PC software. serial communications output options available on an output module,
• Internal program safety lockout switch to prevent tampering. the meter becomes a powerful instrument for monitoring, measure-
• Provision for external LOCK, HOLD, and TEST switch. ment, control, and communication applications.
• Six programmable setpoints with front panel annunciator LEDs. An auto-sensing high or optional low voltage AC / DC power supply
• Smart digital filtering and input averaging. circuit means the meter can be powered from almost any power
• Optional NEMA-4 front cover. supply source. User selectable 5, 10, or 24 VDC excitation voltages
• Seven-level display and annunciator brightness control. are available to power the input module.
• Can perform dual channel tasks. The meter uses smart digital filtering and input averaging. The
• Can be customized for O.E.M. applications. meter’s resident software is designed to allow maximum user flexi-
bility through logical menus. All programming takes place in the
• Make your O.E.M. meter perform complex functions but appear
simple to use by hiding set-up codes, etc. main or setpoint programming modes. The meter has a wide range
of built-in software features. For more details on software features
• Texmate developed PC software for meter programming (serial
output module required). see Specifications on Page 2.
Serial Output Module: ..Communication: Isolated ASCII, or ModBus, Conversion Rate: ........Typically 10 samples per second for 50 Hz;
or Profibus with direct printer interface 5 samples per second for 60 Hz
Isolated RS-232, or isolated RS-485, or Temp. Coefficient: ......Typically 30 ppm/ °C
non-isolated TTL serial output:
Mains Input Polarity: ..Auto
Data: 7/8 bits
Baud: 300, 600, 1200, 2400, 4800, 9600, Decimal Point:..............Front panel, user programmable to five
19200, or 57.6 k. positions
Parity: Off, odd, or even Warm Up Time: ............10 minutes
Bus Address: 1 to 255
Operating Temp.: ........0-50 °C
Display Range:
DI-50E ............................–19999 to 32766, or –19990 to 99990 with Case Dimensions: ......3.78" W x 1.89" H x 4.84" D (96 x 48 x 117
fixed LSD zero selected mm)
DI-60E, DI-60AE............–32766 to 32766, or –199990 to 327660 Case Material: ..............Polycarbonate
with fixed LSD zero selected. Totalizer and
Lens Cover: ..................NEMA-4, (optional)
counter functions to 999999
Input Impedance:........Depends on input signal conditioner Application Functions Include:
selected (1 MΩ for DCV ) Linearization ..................32-point linearization is used to convert
non-linear curves to straight sections over
Display: ........................7 segment, 0.56” (14.2 mm) LEDs for the length of the curve.
DI-50T and DI-60T. 14 segment, 0.56”
(14.2mm) for DI-60AT. Red standard, or Totalizator ......................The totalizator calculates the total of a
optional green LEDs. process signal being metered by accumu-
lating an input process variable over time.
Annunciators: ..............Six red LEDs on front panel; one annunci-
ator per setpoint Counter Function ..........A multitude of counter applications are
supported.
Power Requirements
Standard High Voltage: Between 85-265 VAC or 95-370 VDC Timer ..............................Timer functions are supported in either time-
up, time-down, or real-time clock modes.
Optional Low Voltage:....Between 9-32 VAC or 10-60 VDC
Math Function ................The math function can calculate the sum,
Excitation Voltage: ........User selectable 5 / 10 / 20 VDC
difference, ratio, and product of two inputs.
Setpoints:......................Setpoint 1 activates Relay 1 This, for example, can be used to produce
Setpoint 2 activates Relay 2 a tare reading during weighing operations.
Setpoint 3 activates Relay 3 Peak and Valley ............The meter can retain peak and valley
Setpoint 4 activates Relay 4 (min/max) information and recall this on
the front panel.
Setpoint 5 resets selected register or acti-
vates Relay 5 Remote Setpoint Input ..Remote setpoint input via channel 2.
Setpoint 6 resets selected register or acti- Primary Input
vates Relay 6 Compensation................Provides compensation to the primary input
channel (CH1) via channels 1, 2, 3, 4
Front Panel Functions When the SETPOINT LOCKOUT switch is set to the OFF position,
the setpoints can be programmed.
Program Button Once the setpoint values have been entered and the SETPOINT
The PROGRAM button is used to move from one program step to LOCKOUT switch set to position 1, the setpoints can be viewed but not
the next.When pressed at the same time as the UP button, it initiates changed. All other programming functions, except setpoint control func-
the main programming mode. When pressed at the same time as tions, can be carried out provided the PROGRAM LOCKOUT switch is
the DOWN button, it initiates the setpoint programming mode. set to OFF. See the Display PCB without Faceplate and Bezel diagram.
Up Button Error Message “Err”
When in the operational display, pressing the UP button initiates the Error messages usually occur during calibration procedures. The
viewing setpoints 2, 4, 6, and peak readings routine. Once into the three most likely causes of an error message are:
routine, pressing the UP button moves through the steps of the routine. 1) The full scale and zero signals were too similar. The high input (full
When setting a displayed parameter during programming, the UP scale) signal must be at least 1000 counts greater than the low
button is used to increase the value of the displayed parameter. input (zero) signal (positive and negative values are allowed).
2) The scaling requirement exceeded the capability of the meter
Down Button (–19999 to +32766).
When in the operational display, pressing the DOWN button initiates
3) No input signal present, or incorrect connections.
the viewing setpoints 1, 3, 5, and valley readings routine. Once
into the routine, pressing the DOWN button moves through the Rear Panel Functions
steps of the routine.
When setting a displayed parameter during programming, the DOWN
Program Lockout Pin
button is used to decrease the value of the displayed parameter. By connecting the LOCK pin (pin 8 on the main PCB) to the COM-
MON pin (pin 11 on the main PCB), the set-up program is locked out.
Annunciator LEDs The programmed meter parameters and setpoints can be viewed
The annunciator LEDs may be programmed to indicate the alarm sta- but not changed when the meter is in the "locked out" mode.
tus.They are labeled from left to right: SP1, SP2, SP3, SP4, SP5, SP6.
Hold Reading Pin
UP
TEXMATE Button When the HOLD pin (pin 9) is connected to the COMMON pin (11)
the displayed reading is frozen. However, A/D conversions continue
and as soon as pin 9 is disconnected from pin 11 by the switch, the
Prog.
updated reading is instantly displayed when released by a switch
SP1 SP2 SP3 SP4 SP5 SP6
DOWN contact. The HOLD pin can also be used to reset the following:
Button
PROGRAM LED Annunciators • Reset of total while in the total display.
Button P for Setpoints 1-6
• Reset of peak while in the peak display.
Display with Faceplate and Bezel
• Reset of valley while in the valley display.
• Reset of tare, peak, and valley while in the tare function.
Seven Segment LED Displays • If the relays are latched, they are also reset with the above total,
The five, seven segment LED displays are used to display the meter peak, valley, or tare setting.
input signal readings. They also display the programming codes
and settings during programming. The display is available in red, Serial Output
Function Pins
18 17 16
AC/DC Power
Pin 18 for Dual
Analog output ONLY
When the PROGRAM LOCKOUT switch is set to the OFF position, Input Module
(See specific input module data
sheet for connection details)
2
1
Common Pin
To activate the HOLD, TEST and RESET, or LOCK pins from the
SETPOINT rear of the meter, the respective pins have to be connected to the
LOCKOUT
Switch COMMON pin (pin 11).
Display PCB without Faceplate and Bezel
Installation
DI-60E OA____
Add to the basic model number the order code suffix for each standard option required. The last suffix is to
indicate how many different special options and or accessories that you may require to be included with this product.
Ordering Example: DI-60T-DR-PS1-IA01-AIC-R1-OA2 plus CR-CHANGE and a OP-N4X/96X48, $225 + N/C + N/C + 35 + 50 + 30 + 5+ 20 = $365
IT02 . . .RTD, 100Ω Pt. 2/3/4-wire . . . . . . . . ..... .... . . . . . . . . . . . . . . . . .$35
BASIC MODEL NUMBER IT12 . . .RTD, 120Ω Nickel 2/3/4-wire . . . . . . ..... .... . . . . . . . . . . . . . . . . .$45
DI-50E . . . 96x48mm, Tiger, 5 Digit, E Version . . . . . . . . . .$180 IT13 . . .RTD, 10Ω Copper 2/3/4-wire . . . . . . ..... .... . . . . . . . . . . . . . . . . .$40
DI-60AE . . 96x48mm, Tiger, 6 Digit, Alphanumeric, E Version .$275 IW01 . . .AC-Watts, 300V/1A AC, Single Phase 2-wire ... . . . . . . . . . . . . . . . . .$65
DI-60E . . . 96x48mm, Tiger, 6 Digit, E Version . . . . . . . . . .$225 IW02 . . .AC-Watts, 300V/5A AC, Single Phase 2-wire ... . . . . . . . . . . . . . . . . .$65
IW03 . . .DC-Watts, 10V/50mV DC (0-100.00) ..... .... . . . . . . . . . . . . . . . . .$65
Standard Options for this Model Number
ANALOG OUTPUT
Order Code Suffix Description List
AIC . . .Isolated 16 Bit Current Output, 4-20mA . . . . . . . . . . . . . . . .$50
DISPLAY AIV . . .Isolated 16 Bit Voltage Output, 0-10VDC . . . . . . . . . . . . . . .$50
DR . .Red LED, 0.56 inch high . . . . . . . . . . . . . . . . . . . .N/C S2 . . .Isolated RS-232 (Requires Cable, See Options) . . . . . . . . . .$65
DB . . . .Super–bright Red LED, 0.56 inch high . . . . . . . . . . . . . . . . . . .$25 S4 . . .Isolated RS-485 (Requires Cables, See Options) . . . . . . . . .$85
DG . . .Green LED, 0.56 inch high . . . . . . . . . . . . . . . . . . . . . . . . . . .$10 S5 . . .Isolated ModBus, ASCII Code Compatible . . . . . . . . . . . . . .$95