Voyager Transport Incubator: Service Manual
Voyager Transport Incubator: Service Manual
Voyager Transport Incubator: Service Manual
Ma
ailing add
dress:
Intternational Biome
edical
82
206 Cross s Park Dr.
r.
Auustin, TX 78754
USSA
Au
uthorized
d represen
ntative in
n Europe for Regu
ulatory Afffairs:
Em
mergo Eu urope
Prrinsesseg
gracht 20
25
514 AP
Th
he Hague,, The Nettherlands
s
TABLE OF CONTENTS
LANGUAGE DISCLAIMER ................................................................................................................... 4
SECTION 1: GENERAL INFORMATION ........................................................................................... 11
INTRODUCTION ....................................................................................................................... 11
CLASSIFICATION .................................................................................................................... 12
SAFETY SUMMARY ................................................................................................................. 12
IMPORTANT SAFETY CONSIDERATIONS ............................................................................. 12
SYMBOLS ................................................................................................................................. 18
SECTION 2: SET-UP INSTRUCTIONS .............................................................................................. 19
UNPACKING AND ASSEMBLY OF INCUBATOR AND CART ............................................... 19
ASSEMBLY OF THE VENTILATION SYSTEM ........................................................................ 21
ASSEMBLY OF THE HANDLES .............................................................................................. 22
SECTION 3: OPERATING INSTRUCTIONS ...................................................................................... 23
FRONT PANEL DISPLAY FEATURES .................................................................................... 23
BACK PANEL FEATURES ....................................................................................................... 28
SIDE PANEL FEATURES ......................................................................................................... 29
POWER FEATURES ................................................................................................................. 30
POWERING UP THE INCUBATOR .......................................................................................... 32
POWERING OFF THE INCUBATOR ........................................................................................ 32
DISPLAY MODES ..................................................................................................................... 32
TEMPERATURE CONTROL..................................................................................................... 33
SETTING THE INFANT CHAMBER AIR TEMPERATURE ...................................................... 33
PULSE OXIMETER AND OXYGEN MONITOR CONTROL (OPTIONAL FEATURE) .............. 33
ALARMS / INDICATORS .......................................................................................................... 34
EXTERNAL LIGHTING ............................................................................................................. 37
SKIN TEMPERATURE PROBE ................................................................................................ 37
SECTION 4: PULSE OXIMETRY (OPTIONAL FEATURE) ................................................................ 38
PULSE OXIMETER PRINCIPLES OF OPERATION ................................................................ 38
PULSE OXIMETER SET-UP INSTRUCTIONS ......................................................................... 40
INITIAL SET-UP ............................................................................................................. 40
SETTING HIGH AND LOW %SpO2 ALARMS ............................................................... 40
SETTING HIGH AND LOW PULSE RATE ALARMS .................................................... 41
SETTING THE ALGORITHM MODE (MASIMO ONLY) ................................................. 41
SETTING THE AVERAGING MODE (MASIMO ONLY) ................................................. 42
SETTING THE PERFUSION INDEX (MASIMO ONLY) ................................................. 43
OPERATIONAL ALARMS AND WARNINGS........................................................................... 43
TEST OF OPERATIONAL ALARMS ........................................................................................ 45
PULSE OXIMETRY SENSORS ................................................................................................ 45
PULSE OXIMETRY TESTERS / SIMULATORS ....................................................................... 46
SECTION 5: OXYGEN MONITOR (OPTIONAL FEATURE) .............................................................. 47
OXYGEN MONITOR PRINCIPLES OF OPERATION ............................................................... 47
OXYGEN MONITOR SET-UP INSTRUCTIONS ....................................................................... 48
INITIAL SET-UP ............................................................................................................ 48
SETTING HIGH AND LOW OXYGEN ALARMS .......................................................... 48
OXYGEN MONITOR CALIBRATION ............................................................................ 49
The International Biomedical Voyager Infant Transport Incubator with optional PulseOx (referred to
herein as the incubator) provides a thermally regulated environment to support an infant’s temperature
requirements and has the capability to monitor vital information during transport. The pulse oximeter
and oxygen monitor measure pulse rate, oxygen saturation, and oxygen concentration and allow the
user to configure high and low alarm settings. The incubator circulates warmed air throughout the
infant chamber to maintain the temperature at a user-selected setpoint. Chamber doors and hand
ports provide quick and easy access to the infant. Positioning straps are provided to limit infant
movement within the infant chamber. Detachable air/oxygen carts or tank modules are available
which have space for up to four cylinders. There are no known contraindications associated with the
incubator.
NOTE: The Pulse oximetry and oxygen monitoring is only available if the PulseOx version of the
incubator was purchased.
This manual is designed to introduce the user to key features of the incubator, including safety issues,
instructions for use, equipment maintenance, and contact information. The manual should be read
and understood by all users before using the transport incubator.
Class I / Internally Powered, according to the type of protection against electric shock
Type B, according to the degree of protection against electric shock (i.e. the patient may not be
electrically isolated from earth)
Ordinary, according to the degree of protection against harmful ingress of water
Equipment is not suitable for use in the presence of a flammable anesthetic mixture with air or
with oxygen or nitric oxide.
Continuous operation for the mode of operation
The pulse oximeter cables and sensors are classified as type BF, per the EN60601-1 electrical
standard.
The skin temperature probe is classified as type BF, per EN60601-1 electrical standard (systems with
optional PulseOx).
The skin temperature probe is classified as type B, per EN60601-1 electrical standard (systems
without optional PulseOx).
SAFETY SUMMARY
The incubator is designed to be used by trained clinical users and/or biomedical engineers and
operated in a manner consistent with the instructions contained in this manual. Refer to any additional
training, procedures, requirements, or documentation beyond those identified here for operation and
policies required within the institution. All personnel operating the incubator must be familiar with the
warnings and operating procedures contained in this manual. International Biomedical is not to be
held responsible if the incubator is used in a manner inconsistent with the instructions herein.
Use of accessories, transducers, and cables other than those specified, with the exception of cables
and transducers sold by International Biomedical, may result in increased emissions or decreased
immunity of the incubator.
The transport incubator contains a sealed lead battery and must be recycled or disposed of properly.
Accessory equipment may also contain batteries or other items which present risks during disposal;
please consult the appropriate manual for proper disposal techniques.
Safety concerns or additional pertinent information will be displayed using warnings, cautions, and
notes, having the following significance:
The principal WARNING and CAUTION notices to be observed in use of this incubator are brought
together here for emphasis.
WARNINGS
OBSERVE BEST PRACTICE: The instructions in this manual in no way supersede established
medical procedures or staff preference concerning patient care.
ENSURE UNIT IS PROPERLY GROUNDED: To ensure grounding reliability, only connect the
power cord to a properly grounded, 3-wire hospital grade outlet of the proper voltage and
frequency. DO NOT USE EXTENSION CORDS. If the integrity of the connection is in doubt,
the incubator should be operated from its internal battery.
EXPLOSION HAZARD: Do not use the examination light, the pulse oximeter, or oxygen
monitor in the presence of flammable anesthetics or other flammable gases.
Avoid eye exposure. Direct light exposure may cause eye damage. Infants must wear eye
protection.
The use of oxygen may increase the noise level within the infant chamber.
Air and oxygen tanks are pressurized and must be properly secured.
The transport incubator is Type B equipment and the baby may not be electrically isolated from
earth. Care must be taken that additional equipment connected to the baby is electrically safe.
To ensure patient electrical isolation, connect only to other equipment with electronically
isolated circuits.
When the infant tray is removed, do not touch exposed circuitry and patient simultaneously.
This incubator was calibrated with the infant chamber originally supplied. If this chamber is
exchanged for an infant chamber of a different configuration or size, the temperature calibration
will be affected. Consult International Biomedical before returning the incubator to service.
The incubator should be turned off and AC or DC power disconnected when cleaning.
The use of devices which radiate high intensity electrical fields may affect the operation of the
transport incubator. Constant assessment of the patient and all life support equipment is
mandatory whenever interfering devices are operating on or near patient.
When using the transport incubator adjacent to or stacked with other equipment, observe the
operation of the transport incubator and the other equipment to ensure normal operation.
Do not use the pulse oximeter or oximetry sensors during magnetic resonance imaging (MRI)
scanning. Induced current could potentially cause burns. The pulse oximeter may affect the
MRI image and the MRI unit may affect the accuracy of the oximetry results.
If an alarm condition (other than exceptions listed herein) occurs while the MUTE button is
activated, only visual alarm indicators will be activated.
Pulse rate measurement is based on the optical detection of a peripheral flow pulse and
therefore may not detect arrhythmias. The pulse oximeter should not be used as a replacement
or substitute for ECG based arrhythmia analysis.
Connecting equipment to the outlets on this device creates a medical electrical system and the
user is responsible for continued compliance with the requirements of IEC 60601-1.
Do not modify this equipment without proper authorization from International Biomedical.
An Infant Transport Incubator should be used by appropriately trained personnel and under the
direction of qualified medical staff familiar with currently known risks and benefits of Infant
Transport Incubator use.
When this incubator is operated on battery in low ambient temperatures, the thermal cover must
be used.
Skin temperature probe is not a rectal probe. The skin temperature sensor is not to be used as
a rectal probe.
CAUTIONS
U. S. Federal and Canadian law restricts this device to sale by or on the order of a physician or
other licensed medical practitioner. Outside Canada and the U. S., check with local laws for
applicable restrictions.
Use of sharp objects on Front Display Panel will cause permanent damage and will void
warranty.
Do not touch, press, or rub the display panels with abrasive cleaning compounds, instruments,
brushes, rough-surface materials, or bring them into contact with anything that could scratch the
panel.
The incubator electronics contain static sensitive components that can be damaged by improper
handling. Use approved grounding techniques for work areas and service personnel.
The infant tray grounding tabs are sharp; use care when cleaning air flow system.
Do not move the incubator by pushing on the infant chamber. The infant chamber is not
designed to sustain the forces to push the incubator. Stress fractures in the infant chamber can
occur.
DO NOT leave the 12 volt cord attached to the DC connector of the incubator.
Do not overtighten the infant chamber screws. Do not strip the aluminum into which these
screws are threaded.
Do not drip cleaning solution through the holes where the swell latches fit into the air flow
assembly.
The incubator MUST be plugged into AC power and the battery recharged after any battery
usage. The battery will sustain damage if drained of power and not placed on recharge soon.
CLEANING AND CARE: Do not autoclave, pressure sterilize, or gas sterilize the incubator,
cables, or sensors. Use cleaning solutions sparingly as excessive solution can flow into the
incubator and cause damage to internal components. Do not soak or immerse the incubator or
sensors in any liquid. Do not use petroleum-based, alcohol, acetone, or other harsh solvents.
See cleaning instructions of reusable sensors and cables in directions for their use.
Check alarm limits each time the system is used to ensure that they are appropriate for the
patient being monitored.
Use only parts, accessories, transducers, and cables designated by International Biomedical for
use with the transport incubator. Cables and accessories other than those supplied by
International Biomedical may result in unacceptable operation of the transport incubator and will
void the equipment warranty.
The Maxtec MAX-250E oxygen sensor is a sealed device containing a mild acid electrolyte,
lead (Pb), and lead acetate. Lead and lead acetate are hazardous waste constituents and
Pulse oximeter probes and cables are designed for use with specific monitors. Only use
Masimo sensors and patient cables for Masimo pulse oximetry model. Only use Nellcor
sensors and patient cables for Nellcor pulse oximetry model. Verify the compatibility of the
monitor, sensor, and cable before use, otherwise patient injury can result.
Only use International Biomedical cable and Maxtec sensor for oxygen monitoring.
Calibrate the oxygen sensor daily when in use or if environmental conditions change
significantly (i.e., Temperature, Humidity or Barometric Pressure).
Dropping or severely jarring the oxygen sensor after calibration may shift the calibration point
enough to require recalibration.
Oxygen sensor and pulse oximetry cables must be contained within cart or otherwise secured
when not in use.
Off (Standby)
Atte
ention, consult accompanyying documents
(Blue Background)
Alte
ernating Curre
ent
Dire
ect Current
Typ
pe B equipment (EN60601--1)
Typ
pe BF equipment (EN60601
1-1)
Rec
cycle or dispo
ose of properlyy, contains se
ealed lead ba
atteries
MANUAL
L EXPLA
ANATION
SYMBOL
REFERENC
R CE
ALARM
M Designates locatiion of alarms on front pane
el
HIGH
H TEMP High
h temperature
e alarm indica
ator
SYS FA
AIL Systtem failure ala
arm indicatorr
AIR FLO
O Air flow
f blockage
e alarm indica
ator
SENS
S FA
AIL Prim
mary temperatture sensor m
malfunction alarm indicatorr
Incu
ubator not connnected to AC
C nor DC, and
d battery pow
wer is
PWR
P FA
AIL belo
ow 10.1 volts
MUTE
E Mute
e button silen
nces audible a
alarms for app
proximately 1 minute
BAT CH
HG Battery charging indicator
AC OP
P Incu
ubator conneccted to AC po
ower
DC OP
P Incu
ubator conneccted to DC po
ower
Incu
ubator operati ng on interna
al battery due to no externa
al power
BAT OP
P conn
nected
LOW BA
AT Low
w battery indiccator
Figure
F 2 - 1 Voyagerr Incubatorr with Air/O
Oxy Cart and Access
sory Module
A. Chamber
B. Incubator /
Accessory
Module
M
C.
C Air/Oxy
Cart
C
ead these in
Re nstructions thoroughly
y before any
y assemblyy.
1. Open all boxes. Do not rem ems at this time. Identtify contentts (i.e. infan
move any ite nt chamber,,
main module,
m acc
cessory mo
odule, and air/oxygen
a ccart or strettcher).
2. Remov
ve the Air/O
Oxy Cart, ite
em C in Fig
gure 2 - 1, ffrom the bo
ox marked 2 38 ½ 22 ½
222-7311 (3
20 box
x). Set asid
de any loos
se items sto
ored with the
e air/oxyge
en cart in the shipping box.
3. Lock th
he two whe
eels with the
e black brakes. The lo
ocking whe
eels are to b
be on the side of the
incuba
ator facing the
t operato
or.
4. Remov
ve the incubator and accessory
a module,
m item
m B in Figu
ure 2 - 1 (sh
hipped toge
ether in box
x
310, or 42 27 20 box) and pla
222-73 ace on the a
air/oxy cartt. The stora
age opening
g on the
access
sory module is to be on
o the same
e side as th
he locking b
brakes.
6. Find th
he infant po
ositioning sttraps, usually shipped in the main
n module b
box. Loop the straps
through the slits in the side of
o the infant bed and V
Velcro into place. The
ere will be fo
our pieces of
wo complete
straps, making tw e belts. See Figure 2 - 2.
Figure 2 - 2 Po
ositioning Strap Mou
unting
7. Open the
t final bo mately 34 21 17, m
ox (approxim marked partt# 222-7312
2), remove the inner,
clear Plexiglas
P inffant chamb
ber out of th
he box, and place on th
he top of th
he main mo
odule, carefully
fitting iti in the gas
sket materia
al. The larg
ge opening on the side
e of the cha
amber shou
uld be place
ed
facing the front, with
w the cut out for the head door on the left..
9. e both cham
Secure mbers to the incubatorr by insertin
ng the four push/pull p
pins located
d at both ends
on the top of the main modu
ule (see Figure 2 - 3). Adjust the chambers until the pin
ns are easily
inserte
ed.
Figure
e 2 - 3 Cha
amber and
d Lamp Mo
ounting
10
0. Locate
e the lamp. The lamp base is loc
cated on the
e top right b
back cornerr of the maiin module, just
j
behind
d the infant chamber. See Figure
e 2 - 3. Inse
ert the conn
nector from
m the lamp tto the lamp
base.
Th
he airway su
upplies suc
ch as the ble
ender, flow
wmeter, and
d associated
d brackets are wrappe
ed and
shipped eithe
er in the Air//Oxy cart box or the se
eparate acccessories b
box.
Figu
ure 2 - 4 Blender
B Bra
acket Mou nting and Hose Loca
ation
3. Connect th
he green ho
ose with the
e straight fittting that is protruding from the slot in the
incubator to
t the side outlet
o of the
e blender.
6. Connect th
he short flow
wmeter hos
se to the bo
ottom cente
er outlet porrt on the ble
ender as
shown in Figure
F 2 - 5.
2. Unscrew the nuts on the four bolts located on both ends of the main module.
3. Slide the high rise handle (large inverted U-shape) on first and then the standard handle.
Replace the nuts, tighten.
4. Repeat the above steps to attach the two handles to the other end of the main module.
WARNING: Avoid direct sunlight or radiant heat which can cause a dangerous increase in
chamber air temperature.
CAUTION: Use of sharp objects on Front Display Panel will cause permanent damage and will
void warranty.
CAUTION: Do not touch, press, or rub the display panels with abrasive cleaning compounds,
instruments, brushes, rough-surface materials, or bring them into contact with
anything that could scratch the panel.
* NOT
TE: Ite
ems 18 - 25
5 are featurres exclusivve to the Pu
ulseOx verssion of the iincubator.
Yellow LED indicator that illuminates when the displayed temperature exceeds 39.0 C
Yellow LED indicator that illuminates when the temperature being sensed by the primary temperature
sensor is outside the normal temperature range of the incubator (10 - 45 C)
Yellow LED that illuminates when the chamber air temperature setpoint is set above 37 C
Yellow LED indicator that illuminates when the air flow is blocked
Yellow LED indicator that illuminates when the displayed temperature exceeds 38.5 C
6: AC Fail Indicator
Yellow LED indicator that illuminates when the incubator is unplugged from an AC power source
Yellow LED indicator that illuminates when the incubator is unplugged from AC and DC power and the
internal battery voltage falls below 10.1 volts
Seven segment digital display that displays chamber air temperature, baby temperature, battery life
status, and HI or LO temperature alarms
Indicator is illuminated when the Main Display Screen displays the current air temperature in the infant
chamber.
Indicator is illuminated when the user is adjusting the infant chamber air temperature set point.
Indicator is illuminated when the BABY TEMP button is pressed and held.
Indicator is illuminated when the Battery Life Display button is pressed and held.
When button is pressed, the Main Display Screen will display the temperature measured by the baby
temperature probe.
When button is pressed, the infant chamber air temperature can be adjusted.
Button is used to silence audible alarms for approximately one minute. The warning LED to the left of
the Mute Button will be illuminated for the duration of the mute cycle.
NOTE: The LED is only available with the PulseOx version of the incubator.
Bar graph will illuminate at the rate shown on the Pulse Rate Display Screen with a calculated
intensity.
Indicator is illuminated when the oxygen concentration lower alarm limit is set to a value below 21%.
Button used to enter modes to set pulse oximetry and oxygen monitor alarm limits and to calibrate the
oxygen sensor
Button used to display the status of the battery on the Main Display Screen
Yellow LED indicator that illuminates when the battery voltage is less than 11 volts
Green LED indicator that illuminates when the battery is actively being charged
Green LED indicator that illuminates when the incubator is on and connected to external AC power
Green LED indicator that illuminates when the incubator is on and connected to external DC power
Green LED indicator that illuminates when the incubator is on and no external power is applied,
indicating the unit is operating off of internal battery power
Button is used to turn the incubator on and off. When turned on, it enables the temperature display,
pulse oximetry and oxygen monitor displays, alarms, and infant chamber air temperature control.
1: Accessory
y Power Co
ord Conne
ectors
Au
uxiliary pow
wer cord con
nnectors alllow for additional equiipment to b
be connecte
ed to the AC
C input pow
wer
fro
om the incubator. Accessories plugged into these outle
ets will onlyy be powere
ed by AC w
when the
inccubator is plugged
p into
o AC. If the
e incubator is disconne
ected from AC, accesssories will rrevert to the
eir
ow
wn internal battery
b pow
wer, if availa
able.
Th
his replacea
able 15 A fu
use is a sec
condary tran
nsformer fu
use.
5: 3 A Acces
ssory Circu
uit Breakerrs
Th
hese 3 Amp
p circuit breakers are connected
c to
t the accesssory powe
er cord connector circu
uits.
6: 10 A Circu
uit Breakerrs
Th
hese 10 Am
mp circuit brreakers are connected
d to the batttery and the
e external D
DC input.
7: Potential Equalizatio
E on Stud
Th
his stud is used
u for pottential equa
alization.
8: Accessory
y Power Co
ord Conne
ectors
Au
uxiliary pow
wer cord con
nnectors alllow for additional equiipment to b
be connecte
ed to the AC
C input pow
wer
fro
om the incubator. Accessories plugged into these outle
ets will onlyy be powere
ed by AC w
when the
inccubator is plugged
p into
o AC. If the
e incubator is disconne
ected from AC, accesssories will rrevert to the
eir
ow
wn internal battery
b pow
wer, if availa
able.
SIDE PA
ANEL FE
EATURES
Th
he side panel of the inc
cubator is where
w the pulse
p oxime
eter sensor,, oxygen monitor senssor, and
tem
mperature probe
p conn
nect. Figure
e 3 - 3 show
ws a drawin
ng of the se
ensor panell and highlig
ghts importtant
fea
atures. A description
d of
o each of the
t features
s follows.
Figure
F 3 - 3 Incubato
or Side Pan
nel
Baby Te
emperature
e Oxygen
Cable Connector S
Sensor Connector
Pu
ulse Oxime
eter
Sen
nsor Conne
ector
The SpO2 and pulse rate sensor used with the Masimo or Nellcor pulse-oximeter connects to this
connection point. Refer to SECTION 4 for details on the pulse oximeter system.
The oxygen sensor used with the Oxygen Monitor connects to this connection point. Refer to
SECTION 5 for details on the oxygen monitor system.
POWER FEATURES
The incubator can be operated in one of three power modes: external AC power, external DC power,
or internal battery power.
AC Power Mode
AC power can be supplied using either 120 V or 230 V external AC power, depending on the incubator
configuration. When operating in AC power mode, the AC OP indicator (Figure 3 - 1, item 29) will be
illuminated. AC power has priority over all other power sources, external or internal. AC power should
be used for warming the incubator in preparation for transport and also to recharge the battery.
DC Power Mode
DC power can be supplied using a 12 V power source. When operating in DC power mode, the DC
OP indicator (Figure 3 - 1, item 30) will be illuminated. External DC power is supplied through a 3
conductor circular connector located on the back of the incubator (Figure 3 - 2, item 3). DC power has
priority over internal battery power. While operating in DC power mode, the battery will not be
charged.
Battery power will automatically be selected when neither AC power nor DC power are available.
When operating on internal battery, the BAT OP indicator (Figure 3 - 1, item 31) will be illuminated.
Connecting AC or DC power will supersede battery power operation. The battery charger operates
continually when the incubator is connected to an AC power source. When charging, the green BAT
CHG indicator (Figure 3 - 1, item 28) will be illuminated. The rate at which the battery charges
decreases when the incubator is powered on. A significantly low AC line voltage will also decrease the
rate of battery charge. It takes approximately 8 hours to fully charge a completely discharged standard
26 ampere-hour battery when the incubator is supplied AC power and is powered down. If the
incubator is on and warming up, the battery charge time is two to three times longer. The incubator
will maintain an infant chamber temperature of 37 C for three hours on a fully charged battery with an
ambient temperature of 20 C.
NOTE: When not in use, the incubator should be plugged into an AC source to recharge the
battery. See SECTION 6 on BATTERY CARE.
To determine how much battery life is left, the Battery Life Display Button (Figure 3 - 1, item 26) can be
pressed to display the status of the battery on the Main Display Screen (Figure 3 - 1, item 8). In this
mode, the Battery Life Display Indicator (Figure 3 - 1, item 12) will be illuminated. When operating on
AC power, the Main Display Screen will indicate the state of charge of the battery as an estimated
percentage (to the nearest 20%). A fully charged battery will display 100% and a battery which has
almost reached its safe discharge level will display 0%. When the incubator is operating from battery
power or external DC power, the Main Display Screen will indicate the battery voltage (to the nearest
0.3 volt). Regardless of the power source, error code E03 will be reported if the battery voltage is
outside of the 9.0 volt to 14.5 volt range. To continue normal operation, the battery voltage must be
above 10 0.3 volts.
For general operation, the use of external AC or DC power is HIGHLY RECOMMENDED. The internal
battery should be used only when the incubator has no other energy source available.
If external AC or DC power is lost while the incubator is powered on, the incubator will automatically
switch over to battery power with no interruption in incubator performance or degradation of pulse
oximeter and/or oxygen monitor accuracy. In the event of total power loss, all user-programmed
settings (e.g. temperature setpoint, alarm limits) will default to the most recently programmed values.
One of three power source indicators will be illuminated, indicating which source is currently powering
the incubator.
DISPLAY MODES
The incubator has the ability to display both infant chamber air temperature and infant temperature.
When the incubator is initially powered on, the Main Display Screen (Figure 3 - 1, item 8) displays the
setpoint temperature for a few seconds and then defaults to displaying the current air temperature in
the infant chamber. In this mode, the Air Temperature Display Indicator (Figure 3 - 1, item 9) is
illuminated. To display the infant temperature, the BABY TEMP button (Figure 3 - 1, item 13) can be
pressed. In this mode, the Baby Temperature Display Indicator (Figure 3 - 1, item 11) illuminates and
the infant temperature is displayed on the Main Display Screen. When the button is released, the
display reverts back to infant chamber air temperature.
1. Turn on the incubator using the Power Button (Figure 3 - 1, item 32).
2. After the Power-up Indicator Test is complete, press the Set Temperature Button (Figure 3 - 1,
item 15).
3. Use the Set Temperature Up and Down Arrow Buttons (Figure 3 - 1, items 14 and 16) to adjust
the temperature setting to the desired value. The temperature can be adjusted from 17.0 C to
38.9 C in 0.1 C increments.
4. When the desired value is set, either press the Set Temperature Button again or wait 5
seconds, after which time the Main Display Screen will return to displaying the measured infant
chamber air temperature.
WARNING: If an alarm condition (other than the exceptions listed herein) occurs while the
MUTE button is activated, only visual indicators will be activated.
Alarm Priority
System Fail Alarm LOW
Sensor Fail Alarm LOW
Air Temperature Setpoint >37C LOW
Air Flow Alarm Indicator LOW
High Temperature Indicator LOW
AC Fail Indicator LOW
Power Fail Indicator LOW
Alarms are indicated by a continuous beep to alert operator to inspect the device to determine the
cause of the alarm.
The SpO2 % LOW, O2 % HIGH, and O2 % LOW alarms are indicated by a series of 3 beeps followed
by a slight pause, and then repeated. This is to alert the operator to these specific alarms and their
Medium Priority.
When the incubator becomes disconnected from AC power, the AC Fail indicator (Figure 3 - 1, item 6)
illuminates and an audible alarm will sound. When the incubator becomes disconnected from both AC
and DC power and the internal battery voltage falls below 10.1 volts, the PWR FAIL indicator
(Figure 3 - 1, item 7) illuminates and a continuous audible alarm sounds. Power to maintain the
temperature in the infant chamber, and the accuracy of the O2 and the SpO2 monitors (if equipped), is
no longer available and all power to the incubator (other than to power the PWR FAIL indicator and
the audible alarm) is disabled. The indicator illuminates as long as the incubator remains on.
Pressing the MUTE button will not silence the audible alarm. The Power Button should be pressed to
turn the incubator off, silencing the audible alarm, and the incubator should be connected to AC power
to allow the battery to recharge.
In units with the Optional SpO2 feature, the Power Fail alarm is a Medium Priority Technical Alarm.
Temperature Alarms
When the measured air temperature is more than 1 C from the user-configured set point, an
intermittent audible alarm will sound every minute and the Main Display Screen will flash HI or LO
every 5 seconds until the incubator temperature is back to within 1 C of the set point.
When the measured air temperature exceeds 38.5 C, the HIGH TEMP alarm indicator (Figure 3 - 1,
item 5) illuminates and an intermittent audible alarm turns on. The incubator is still fully operational
and the operator, at his/her discretion, may ignore this alarm.
When the measured air temperature exceeds 39.0 C, the HIGH TEMP alarm indicator, the SYS FAIL
alarm indicator (Figure 3 - 1, item 1), and a continuous audible alarm turn on. The incubator’s heater
is disabled and the incubator must cool to below 39 C before the heater is re-enabled.
When the measured air temperature exceeds 39.2 C, the SYS FAIL alarm indicator and a continuous
audible alarm turn on. The incubator’s heater is disabled until the infant chamber temperature is
lowered, at which point the incubator must be turned off and then on again to reset the alarm.
The SENS FAIL indicator (Figure 3 - 1, item 2), coupled with a continuous audible alarm, indicates the
temperature being sensed by the primary temperature sensor is well outside the normal temperature
range of the incubator (10 - 45 C). The activation of this alarm may indicate a problem with the
temperature sensor or the control circuitry, in which case, the incubator needs to be serviced by
qualified personnel.
If the primary temperature sensor is open or indicates an extremely low temperature, the SENS FAIL
and SYS FAIL alarms will be activated, and the heater will be disabled.
If the user specifies a chamber set point temperature greater than 37 C, the Air Temperature Setpoint
> 37 C Indicator (Figure 3 - 1, item 3) will illuminate.
Below each of the pulse oximeter and oxygen monitor screens are two LEDs, labeled HI and LO.
When any of the three displayed values (oxygen saturation, pulse rate, or oxygen concentration)
exceed the user-programmed high alarm limit, the corresponding HI LED illuminates and an audible
alarm will sound. When any of the three displayed values goes below the user-programmed low alarm
limit, the corresponding LO LED illuminates and an audible alarm will sound.
When calibrating the oxygen monitor, if the low oxygen alarm limit is programmed below 21%, the
Oxygen Setpoint Low Indicator (Figure 3 - 1, item 23) is illuminated.
Other Alarms
When the airflow through the infant chamber is reduced due to an obstruction, such as a blanket, the
AIR FLO alarm indicator (Figure 3 - 1, item 4) illuminates and a continuous audible alarm turns on.
Due to reduced airflow, the incubator’s heater’s temperature will exceed a preset threshold and will be
disabled. After the airflow obstruction is removed and the heating element is allowed to cool, the
heater system will return to normal operation.
NOTE: A blanket should always be used inside the infant chamber between the infant and the
mattress.
When the battery voltage drops below 11 volts, the LOW BAT indicator (Figure 3 - 1, item 27)
illuminates and an intermittent audible alarm turns on. The battery will be able to supply the heater
requirements for only a few minutes after this alarm occurs. The LOW BAT alarm cannot be reset.
The incubator must be connected to AC power to allow the battery to recharge when this alarm occurs.
NOTE: When this alarm activates, the incubator has approximately fifteen (15) minutes of
operational power remaining. This estimate is for a good battery and an ambient
temperature of 20 C.
The light is intended to be used as an external light source. Do not place the light into the inner infant
chamber.
WARNING: Do not change the lamp and touch the patient simultaneously.
WARNING: EXPLOSION HAZARD: Do not use the examination light, the pulse oximeter,
or oxygen monitor in the presence of flammable anesthetics or other
flammable gases.
WARNING: Avoid eye exposure. Direct light exposure may cause eye damage. Infants
must wear eye protection.
WARNING: The skin temperature sensor is not a rectal probe. The skin temperature
sensor is not to be used as a rectal probe.
the change in volume of arterial blood in tissue (and hence, light absorption by blood) due to
changes in pulse (plethysmography)
the fluctuating absorbance of venous blood during arteriovenous shunting contributes to noise
during the pulse
Because oxyhemoglobin and deoxyhemoglobin differ in light absorption, the amount of red and
infrared light absorbed by the blood is related to hemoglobin oxygen saturation. The pulse oximeters
supported by the incubator, Masimo SET and Nellcor OxiMax, decompose the red and infrared
pulsatile absorbance signals into an arterial signal plus a noise component and calculate the ratio of
arterial signals. The ratio of the two arterial pulse-added absorbance signals is used to find the
oxygen saturation in an empirically derived equation in the pulse oximeter software.
Different pulse oximetry sensors can be used with the system. The user should consult the sensor’s
instructions for use to ensure the appropriate sensor is being used. In addition, only Masimo sensors
should be used with the Masimo pulse oximeter and only Nellcor sensors should be used with the
Nellcor pulse oximeter. Sensor accuracy data, found in SECTION 11 (PULSE OXIMETER
SPECIFICATIONS (OPTIONAL FEATURE)), are based on human blood studies in which the values
obtained for healthy adult volunteers in induced hypoxia states during motion and non-motion
conditions were compared to a laboratory co-oximeter.
Dyes, or any substance containing dyes, that change usual blood pigmentation
Severe anemia
Venous pulsations
WARNING: Do not use the pulse oximeter or oximetry sensors during magnetic
resonance imaging (MRI) scanning. Induced current could potentially cause
burns. The pulse oximeter may affect the MRI image and the MRI unit may
affect the accuracy of the oximetry results.
NOTE: The pulse oximeter should be considered an early warning device. As a trend towards
patient hypoxemia is indicated, blood samples should be analyzed by a laboratory co-
oximeter to completely understand the patient’s condition.
Pulse rate measurement is based on the optical detection of a peripheral flow pulse and therefore may
not detect arrhythmias. The pulse oximeter should not be used as a replacement or substitute for
ECG-based arrhythmia analysis.
NOTE: Loss of pulse signal can occur for many reasons including, but not limited to, when the
patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia; there
is arterial occlusion proximal to the sensor; or the patient is in cardiac arrest or is in
shock.
CAUTION: Check alarm limits each time the system is used to ensure that they are appropriate
for the patient being monitored.
INITIAL SET-UP
1. Turn the incubator power on using the Power Button (Figure 3 - 1, item 32).
2. Connect the pulse oximeter cable to the connector on the side of the incubator
(Figure 3 - 3). Connect pulse oximeter sensor to the cable and to a suitable site on the
patient being monitored.
3. To turn on the pulse oximeter displays, press the SET button (Figure 3 - 1, item 24) one
time and then either the up or down arrow button (Figure 3 - 1, item 25) until the %SPO2
screen displays “On”. If the screen shows “---”, there is no value available.
4. When the pulse oximeter is disabled, the %SPO2 and PULSE screens will display “OFF”.
All alarms related to the pulse oximeter will also be disabled.
1. Press the SET button (Figure 3 - 1, item 24) three times (or until the %SPO2 HI LED
illuminates) to set the high SpO2 alarm.
2. Use the up or down arrow buttons (Figure 3 - 1, item 25) to set the high alarm to the
desired setting.
3. Press the SET button four times (or until the %SPO2 LO LED illuminates) to set the low
SpO2 alarm.
4. Use the up or down arrow buttons to set the low alarm to the desired setting.
1. Press the SET button (Figure 3 - 1, item 24) five times (or until the PULSE HI LED
illuminates) to set the high Pulse alarm.
2. Use the up or down arrow buttons (Figure 3 - 1, item 25) to set the high alarm to the
desired setting.
a. Must be at least 1 bpm higher than the Pulse low alarm setting.
3. Press the SET button six times (or until the PULSE LO LED illuminates) to set the low
Pulse alarm.
4. Use the up or down arrow buttons to set the low alarm to the desired setting.
a. Must be at least 1 bpm lower than the Pulse high alarm setting.
Different algorithm modes can be used depending on the importance of SpO2 sensitivity vs.
probe-off detection. The system will power up in A1 Normal Mode as a default. When setting
the Algorithm Mode, the setting will be displayed on the %SPO2 screen.
1. Press the SET button (Figure 3 - 1, item 24) 10 times (or until HI and LO %SPO2 LEDs
are both illuminated) to set the Algorithm Mode.
2. Use the up or down arrow buttons (Figure 3 - 1, item 25) to choose one of the following
modes:
a. A1 = Normal Mode
ii. This mode is recommended for the sickest patients or during procedures
when clinician and patient contact is continuous.
c. A3 = APOD Mode
ii. This mode is recommended for use on patients that are at particular risk of
the sensor becoming detached (e.g. pediatric, combative, etc.)
The Averaging Mode is used to determine how many values will be used to calculate the SpO2
and pulse rate readings. A lower averaging rate uses fewer values to calculate the readings,
resulting in more unstable readings, but will more quickly alert the user to potentially alarming
patient conditions. The system will power up with C2 Middle Averaging Rate as a default.
When setting the Averaging Mode, the setting will be displayed on the %SPO2 screen.
1. Press the SET button (Figure 3 - 1, item 24) 11 times (or until the HI and LO PULSE
LEDs are both illuminated) to set the Averaging Mode.
2. Use the up or down arrow buttons (Figure 3 - 1, item 25) to choose one of the following
averaging rates:
1. Press and hold the down arrow button (Figure 3 - 1, item 25).
2. The perfusion index will be displayed on the %SPO2 screen. A display of “---” indicates
no value is available.
3. Release the down arrow button to switch back to Normal Operation Mode.
NOTE: Calibration and resetting the pulse oximeter algorithm or averaging mode
may cause a slight delay in readings.
NOTE: Setting alarm set points, calibration, and resetting pulse oximeter algorithm
or averaging methods will disable a portion of the pulse oximetry or oxygen
monitoring displays (Masimo only).
The PulseOx system updates the data being displayed on the display screens every second. After a
period of 5 seconds, if the system has not received an update, the screen will flash “PS”. If an update
still is unavailable after 15 seconds, “PS” will continue to flash and an audible alarm will sound until the
condition is corrected.
The system displays “LP” when there are very low amplitude arterial pulsations. It has been
suggested that at extremely low perfusion levels, pulse oximeters can measure peripheral saturation,
which may differ from central arterial saturation1. This “localized hypoxemia” may result from the
metabolic demands of other tissues extracting oxygen proximal to the monitoring site under conditions
of sustained peripheral hypoperfusion. (This may occur even with a pulse rate that correlates with the
ECG heart rate.)
NOTE: If the low perfusion message is frequently displayed, find a better-perfused monitoring
site. In the interim, assess the patient and, if indicated, verify oxygenation status through
other means.
When the signal quality is very low, the accuracy of the SpO2 measurement may be compromised, and
the system displays “LS” on the PULSE Display Screen. When the “LS” message appears, proceed
with caution and do the following:
Check the sensor and ensure proper sensor application. The sensor must be well secured to
the site to maintain accurate readings. Also, misalignment of the sensor’s emitter and detector
can result in smaller signals.
Determine if an extreme change in the patient’s physiology and blood flow at the monitoring site
occurred (e.g., an inflated blood pressure cuff, a squeezing motion, sampling of an arterial
blood specimen from the hand containing the pulse oximetry sensor, severe hypotension,
peripheral vasoconstriction in response to hypothermia, medications, or a spell of Raynaud’s
syndrome).
With neonates or infants, check that the peripheral blood flow to the sensor site is not
interrupted. Interruption, for example, may occur while lifting or crossing their legs during a
diaper change.
1
Severinghaus JW, Spellman MJ. Pulse Oximeter Failure Thresholds in Hypotension and Vasoconstriction.
Anesthesiology 1990; 73:532-537
1. After the sensor is attached to the patient, verify the patient alarms are functional by setting the
SpO2 and pulse rate high and low alarm limits beyond the patient readings.
b. A “HI” or “LO” visual indicator should flash on the appropriate display screen.
ii. Error code “E6” should flash on the %SpO2 display screen.
ii. Error code “E9” should flash on the %SpO2 display screen.
WARNING: If using pulse oximetry during full body irradiation, keep the sensor out of the
irradiation field. If the sensor is exposed to irradiation, the reading might be
inaccurate or the unit might read zero for the duration of the active irradiation
period.
CAUTION: Oxygen sensor and pulse oximetry cables must be contained within cart or otherwise
secured when not in use.
Tissue damage can be caused by incorrect application or use of a pulse oximetry sensor, for example,
by wrapping the sensor too tightly. Inspect the sensor site as directed in the sensor instructions for
use to ensure skin integrity and correct positioning and adhesion of the sensor.
High ambient light sources such as surgical lights (especially those with a xenon light source), bilirubin
lamps, fluorescent lights, infrared heating lamps, and direct sunlight can interfere with the performance
of a SpO2 sensor. To prevent interference from ambient light, ensure that the sensor is properly
applied and that the sensor site is covered with opaque material if necessary. Failure to take this
precaution in high ambient light conditions may result in inaccurate measurements.
Temperature
In order for oxygen readings to be accurate, the oxygen monitor must be thermally stable when
calibrated and when measurements are taken. If temperature conditions change, adequate time must
be allowed for the sensor to stabilize and give accurate readings. For best results, calibrate the
oxygen monitor at the same temperature at which the monitor will be used.
Pressure
The readings from the oxygen monitor are proportional to the oxygen concentration only if pressure is
held constant. The oxygen monitor does not compensate for changes in barometric pressure.
Because the flow rate of the gas being sampled can affect the pressure at the sensor in that the back
pressure may change, the oxygen monitor should be calibrated at the same pressure as the sample
gas. Changes in elevation result in calibration error of approximately 1% of reading per 250 feet. In
general, calibration of the instrument should be performed when the geographic elevation at which the
product is being used changes by more than 500 feet. The oxygen monitor should not be calibrated at
a pressure above 2 atmospheres as this is beyond the sensor’s intended use.
Humidity
Condensation due to high humidity can adversely affect the oxygen monitor. The sensor should be
calibrated and used in environments < 95% relative humidity and placed upstream of any humidifier (if
used in a breathing circuit) to reduce the risk of water condensation.
CAUTION: Only use International Biomedical cable and Maxtec sensor for oxygen monitoring.
CAUTION: Use the oxygen monitor when oxygen is delivered to the infant.
CAUTION: Check alarm limits each time the system is used to ensure that they are appropriate
for the patient being monitored.
INITIAL SET-UP
1. Turn the incubator power on using the Power Button (Figure 3 - 1, item 32).
2. Connect the oxygen sensor cable to the connector on the side of the incubator
(Figure 3 - 3).
3. Connect the oxygen sensor to the cable and place in area to be monitored.
a. Place the oxygen sensor inside the infant chamber to monitor ambient oxygen
concentration within the chamber.
4. To turn on the Oxygen display, press the SET button (Figure 3 - 1, item 24) two times
and then either the up or down arrow buttons (Figure 3 - 1, item 25) until the OXYGEN
screen displays “On”. If the screen shows “---”, there is no value available.
5. When the Oxygen Monitor is disabled, the OXYGEN screen will display “OFF”. All
alarms related to the Oxygen Monitor will also be disabled.
1. Press the SET button (Figure 3 - 1, item 24) seven times (or until the OXYGEN HI LED
illuminates) to set the high oxygen alarm.
2. Use the up or down arrow buttons (Figure 3 - 1, item 25) to set the high alarm to the
desired setting.
4. Use the up or down arrow buttons to set the low alarm to the desired setting.
5. If the low oxygen alarm is set below 21.0%, the yellow “<21%” warning LED (Figure 3 - 1,
item 23) will be illuminated.
1. Press the SET button (Figure 3 - 1, item 24) nine times (or until the CAL LED illuminates,
Figure 3 - 1, item 21) to calibrate the oxygen monitor.
2. Use the up or down arrow buttons (Figure 3 - 1, item 25) to change the oxygen
percentage to the concentration of oxygen the sensor is exposed to during the calibration
(use 20.9% if exposed to room air). Expose the sensor to the calibration gas for several
minutes to ensure the sensor reading has stabilized.
NOTE: If using a calibration gas other than room air, procure and dispose of
gas according to hospital policy.
3. Once the calibration setting is set to the desired level, press both the up and down arrow
buttons simultaneously. “CAL” will be displayed on the OXYGEN screen to confirm that
calibration is complete.
CAUTION: Dropping or severely jarring the oxygen sensor after calibration may shift the
calibration point enough to require recalibration.
1. After the oxygen sensor is placed in the area to be monitored, verify the out-of-range alarms are
functional by setting the oxygen high and low alarm limits beyond the oxygen readings of the
area.
b. A “HI” or “LO” visual indicator should flash on the OXYGEN display screen.
2. Verify the sensor alarm is functional by disconnecting the oxygen sensor from the cable
attached to the incubator.
Use only Maxtec Max-250E oxygen sensors and International Biomedical supplied cables with the
incubator. Max-250E oxygen sensors offer quick response, stability, and life greater than 9000 hours.
CAUTION: The Maxtec MAX-250E oxygen sensor is a sealed device containing a mild acid
electrolyte, lead (Pb), and lead acetate. Lead and lead acetate are hazardous waste
constituents and should be disposed of properly, or returned to Maxtec or
International Biomedical for proper disposal or recovery.
CAUTION: Oxygen sensor and pulse oximetry cables must be contained within cart or otherwise
secured when not in use.
CAUTION: Only use International Biomedical cable and Maxtec sensor for oxygen monitoring.
DAILY MAINTENANCE
The following maintenance procedures should be performed on a daily basis by a knowledgeable
incubator transport person.
OPERATIONAL CHECK
1. Check the AC power cord and ensure that there are no cuts or severe bends in the cord,
that all three prongs on the plug are in good condition, and that the cord is securely
fastened in place. Replace the cord as needed.
2. Ensure the incubator is plugged in to an appropriate AC power source. Verify the BAT
CHG indicator (Figure 3 - 1, item 28) on the front panel is illuminated and that all other
indicators are off when the incubator power is off.
3. With the incubator connected to AC power, press the power button (Figure 3 - 1, item 32)
to turn the incubator on. The incubator will perform a test of all indicators, LEDs, and the
audible alarm. Ensure all the indicators and LEDs illuminate, and that the audible alarm
can be heard. Verify the AC OP indicator (Figure 3 - 1, item 29) remains illuminated
after completion of the test.
4. Place a hand inside the right side of the infant chamber and verify air flow. The fan
should be able to be heard at the left side of the incubator.
5. A few minutes after turning the incubator on, the air temperature inside the infant
chamber should begin to increase and will be displayed on the Main Display Screen.
Verify.
6. Turn on the external light and ensure it operates under AC power. Turn off the light.
8. Turn on the external light and ensure it operates under Battery power. Turn off the light.
10. If external 12 volt DC power is to be used, check the 12 volt DC power cord and ensure
that it has no cuts or severe bends and that the connectors have not been damaged.
11. Connect the incubator to a DC power source. Turn on the incubator and verify the DC
OP indicator (Figure 3 - 1, item 30) is illuminated and that the Main Display Screen still
displays the infant chamber air temperature.
12 Turn on the external light and ensure it operates under DC power. Turn off the light.
13. Disconnect the DC power connection and reconnect AC power to the incubator and turn
the incubator off.
CAUTION: DO NOT leave the 12 volt cord attached to the DC connector of the
incubator when not operating in DC power mode.
14. Inspect the mattress and positioning straps for damage. Repair or replace as needed.
WARNING: The transport incubator is Type B equipment and the baby may
not be electrically isolated from earth. Care must be taken that
additional equipment connected to the baby is electrically safe.
To ensure patient electrical isolation, connect only to other
equipment with electronically isolated circuits.
2. With the air and oxygen tank regulators connected to appropriate external devices, open
the valves on the air and oxygen tanks and ensure that each tank is fully pressurized and
ready for use.
3. Ensure that none of the hoses connecting the tanks to external devices are leaking or
kinked.
2. Ensure that the gasket between the chambers and airflow tray is in place. Check that
the grommets in the tubing through-holes on the infant chamber are also in place.
3. Ensure that the infant chamber is secured to the incubator with the two latches on both
ends of the infant chamber.
CAUTION: Do not overtighten the infant chamber screws. Do not strip the
aluminum into which these screws are threaded.
4. Ensure that the infant chamber is clean and ready for transport, as determined by the
transport team leader.
WARNING: This incubator was calibrated with the infant chamber originally
supplied. If this chamber is exchanged for an infant chamber of
a different configuration or size, the temperature calibration will
be affected. Consult International Biomedical before returning
the incubator to service.
CAUTION: When the incubator is attached to a rigid cart with an internal power
strip, the power cord must be contained within cart or secured on the
end of cart with cord holders when not in use.
2. Inspect the oxygen sensor cable and pulse oximeter cable for wear or damage.
Replace, if necessary.
CAUTION: CLEANING AND CARE: Do not autoclave, pressure sterilize, or gas sterilize
the incubator, cables, or sensors. Use cleaning solutions sparingly as
excessive solution can flow into the incubator and cause damage to internal
components. Do not soak or immerse the incubator or sensors in any liquid.
Do not use petroleum-based, alcohol, acetone, or other harsh solvents. See
cleaning instructions of reusable sensors and cables in directions for their
use.
Wash the chamber, infant mattress, and top surface of the airflow tray with mild soap or
detergent and water solution.
CAUTION: Do not drip cleaning solution through the holes where the swell latches fit into
the air flow assembly.
Household ammonia diluted in water in the concentrations recommended for hospital cleaning
may also be used. These solutions may be applied with a soft absorbent cloth, followed by
rinsing with clean water. Wipe away residue with soft chamois or cellulose sponge material.
CAUTION: The infant tray grounding tabs are sharp; use care when cleaning the air flow
system.
OPERATIONAL CHECK
1. Connect the incubator to an AC power source and turn the incubator power button on.
2. Adjust the setpoint to 38.9 C and observe the incubator as the temperature rises to
perform the following checks.
3. As the temperature reaches 38.5 0.1 C, the HIGH TEMP indicator should illuminate.
4. Adjust the setpoint to 37 C and allow the temperature to stabilize at that setting.
2. Press the Up Arrow button and see the display in the Pulse and %SpO2 change from
“OFF” to “On”.
3. Press the Down Arrow button and see the display in the Pulse and %SpO2 change from
“On” to “OFF”.
4. Check that the %SpO2 and Pulse HI and LO LEDs are on. Continue to press the SET
button and check that the LEDs correspond to the marked display.
2. Listen for the alarm. After hearing the alarm, press the MUTE button.
5. Turn off the Oxygen Monitor and plug the sensor back in.
TANK INSPECTION
1. Inspect the air and oxygen tanks for the date of the last hydrostatic test. Each tank must
be hydrostatically tested periodically according to the manufacturer’s recommendation.
The date of the last test is stamped on the top of the tank in month and year (mmyy).
2. Inspect the recharge record of each tank and, if the record indicates there have been a
maximum number of pressure cycles as per DOT (Department of Transportation)
specifications, the tank should be destroyed and another tank used.
3. Wash chamber with mild soap or detergent and water solution to clean the Plexiglas.
Household ammonia in water in the concentrations recommended for hospital cleaning is
also excellent. These solutions may be applied with a soft absorbent cloth, followed by
rinsing with clean water. Wipe away residue with soft chamois or cellulose sponge
material.
CAUTION: Any type of alcohol or solvents should never be used for cleaning the
infant chamber, as these solvents tend to soften the surface and often
cause crazing. Cleansing materials containing abrasives or strong
solvents as found in some window cleaning preparations should never
be used to clean the inner and outer infant chamber.
BATTERY TEST
The incubator is equipped with a 12 volt 26 ampere-hour battery. Under normal operating
conditions in an ambient temperature of 20 C, a new fully charged battery should operate the
incubator for over three hours. If the operating time is less than 1 ¼ hours, the battery needs to
be replaced.
The battery is specified for only a finite number of complete discharges. The test of the battery
will completely discharge it. The test can be performed as follows:
1. Ensure the battery is fully charged by connecting the incubator to AC power for 12 hours
with the power switch off.
3. After the incubator has reached operating temperature, allow an additional twenty
minutes for stabilization.
6. The difference between the noted times in steps 4 and 5 is the operational time. If this
time is less than 1 ¼ hour, the battery should be replaced.
2. Remove the infant mattress. Lift the swell latches on the infant tray and remove the
infant tray. Ensure the swell latches are securely attached to the infant tray. Clean the
infant mattress and infant tray with a hospital approved cleaning solution.
WARNING: Do not remove the infant tray and touch exposed circuitry and
patient simultaneously.
NOTE: The infant tray material is Kydex 100 and the support plate material is
conductive clear chromate aluminum.
3. Lift the two swell latches on impeller cover and remove it. Clean the cover with a
hospital approved cleaning solution.
CAUTION: Entry of fluids into fan motor, thermistors, or warming unit will
significantly reduce service life. Clean this area by wiping with a cloth
dampened with an appropriate cleaning solution and wipe dry.
WARNING: The infant tray grounding tabs are sharp, use care when cleaning
the air flow system.
CAUTION: Do not drip any cleaning solution through the holes where the swell
latches fit into the air flow assembly.
HARDWARE
1. Ensure that all screws on the incubator are properly tightened.
CAUTION: Do not strip the aluminum into which the screws are threaded.
CAUTION: The incubator MUST be plugged into AC power and the battery recharged after any
battery usage. The battery will sustain damage if drained of power and not
recharged.
Accessory equipment may also contain batteries which must be properly cared for. Consult the
accessory equipment user’s manuals for instructions.
The pulse oximetry sensors are single-use sensors and should be disposed of in a manner consistent
with local laws.
This is normal. The alarm indicates that the unit is no longer connected to AC power. Press the
MUTE button to clear it.
When a certain air flow rate is not met, the alarm sounds. Check for an obstruction, such as a blanket,
in the airflow pathway. Check the chamber for cracks, missing grommets, or missing gaskets. Verify
that the fan impeller is spinning. Verify that all swell latches are present and tightly secured.
Ensure the incubator is turned on. If problem still persists, the membrane panel has been damaged
and the overlay needs to be replaced. Contact International Biomedical for assistance.
Ensure the incubator is turned on; the light gets its power from the incubator and will not work unless
the incubator is on. If the light still doesn’t work with the incubator on, it needs to be replaced (e.g. an
LED may be bad, there may be a short in the circuitry). Contact International Biomedical for
assistance.
CAUTION: The incubator electronics contain static sensitive components that can be damaged
by improper handling. Use approved grounding techniques for work areas and
service personnel.
NOTE: Servicing the transport incubator requires special equipment and tools. Do not attempt
to service the transport incubator without proper equipment.
1. Turn the incubator power button off and unplug the AC power cord.
3. Remove the four latches that secure the infant chamber (two screws per latch).
4. Remove the two or four screws that secure the lamp assembly and gently pull the assembly to
expose the connector. Disconnect it and remove the lamp assembly.
5. Remove all remaining screws fastening the airflow tray assembly to the main frame (four along
the top front edge, three along the top rear edge).
6. Carefully lift the front edge of the airflow tray assembly just enough to gain sufficient access to
disconnect the two electrical cables and the ground cable. Disconnect both electrical cables
and the ground cable.
BATTERY REMOVAL
This section details the procedure to be followed to remove and replace the battery. The procedure
detailed to gain access to the electronics compartment must be followed before beginning this
procedure. Due to the weight of the battery and limited working area, exercise caution when inserting /
removing battery to avoid damage to circuit boards or cables.
CAUTION: The incubator electronics contain static sensitive components that can be damaged
by improper handling. Use approved grounding techniques for work areas and
service personnel.
1. Loosen the four quarter turn fasteners securing the battery and remove the battery clamping
bracket.
2. Remove the negative lead from the battery terminal. If necessary, the battery may be carefully
lifted or tilted to allow better access to the terminal.
3. Remove the positive lead from the battery terminal. If necessary, the battery may be carefully
lifted or tilted to allow better access to the terminal.
CAUTION: Replace battery only with International Biomedical part number 888-0071.
CAUTION: The incubator electronics contain static sensitive components that can be damaged
by improper handling. Use approved grounding techniques for work areas and
service personnel.
3. From underneath, remove the three screws securing the aluminum heatsink and spacer to the
floor of the electronics compartment. When reassembling, clean the bottom of the aluminum
extrusion and spacer and apply a thin coating of a suitable heatsink compound to all mating
surfaces.
NOTE: Accessory devices (e.g. ECG Monitor) may need to be removed in order to access the
control board hardware.
CAUTION: The incubator electronics contain static sensitive components that can be damaged
by improper handling. Use approved grounding techniques for work areas and
service personnel.
2. Disconnect the skin temp cable from the display board (J3).
3. Disconnect the oxygen sensor cable from the display board (J11).
NOTE: The SpO2 flex cable is secured to the PulseOx daughter board and needs to be
removed with the display PCB assembly.
5. Remove the two screws that secure the display board and front panel to the bracket above
them.
6. Pull the PC board towards the rear of the chassis and up. The display board, daughter board,
and the front panel will come out together as a unit.
7. Disconnect the membrane keypad connector from the display board (J4).
8. Remove the four visible screws and nylon spacers securing the PC board to the front panel
standoffs (one in each corner and one in the middle).
9. Remove the PulseOx daughter board in order to access the last Display PCB screw.
10. The display board may now be removed from the front panel.
CAUTION: The incubator electronics contain static sensitive components that can be damaged
by improper handling. Use approved grounding techniques for work areas and
service personnel.
1. Disconnect the PulseOx flex cable from the PulseOx daughter board.
3. Lift the PulseOx daughter board off the standoffs on the display board.
Oscilloscope
10 k resistor
Airborne temperature test circuit P/N 387-1190 (Available from International Biomedical)
1. Locate the J2 connector on the control board and connect the sensor simulator connector. If
not using a sensor simulator, leave the airflow tray connected to J2. Ensure JP1 on the control
board is set to route the Secondary Thermistor to the display board’s “measured temperature”
input. (Jumper JP1 should be on positions 1 and 3.) Connect the incubator to an appropriate
AC power source. Check the voltage at the positive battery terminal. It should be
approximately 13.7 0.2 volts with a fully charged battery.
2. Check the voltage at TP1 on control board. It should be 13.7 0.1 volts with a fully charged
battery. Adjust R1 if necessary. The BAT CHG indicator should be on.
4. Remove the sensor simulator. SENS FAIL and SYS FAIL alarms should activate. Reconnect
the sensor simulator to J2 on the control board.
5. Set the thermostat switch on the sensor simulator to OPEN (up position). The AIR FLO
indicator should be on and alarm should sound continuously. Set the thermostat switch to
CLOSED (down position).
6. Turn the incubator power off. Remove the fan impeller from underneath the mattress tray, turn
the unit on, adjust the temperature set point to 37.0 C, and verify that the AIR FLO alarm
activates under twelve minutes. The AIR FLO alarm may sound for a few minutes. Replace
the fan impeller.
7. With the incubator turned on, disconnect the power cord. An audible alarm indicating loss of
AC power should sound, which can be silenced by pressing the MUTE button. Turn the
incubator power off and disconnect the negative battery cable. Set the external power supply to
12 volts and connect the external power supply output to the battery cables, observing proper
polarity.
8. Turn the incubator power on. Connect the digital multimeter leads to the battery cables. Slowly
reduce the supply voltage. The LOW BAT warning LED should illuminate at 10.8 0.3 volts
and the audible alarm should sound at approximately 8 second intervals. The PWR FAIL
indicator should turn on at 10.1 0.1 volts and the incubator power should turn off. The audible
alarm should sound continuously. If these alarms turn on at incorrect voltages, see the
CONTROL BOARD CALIBRATION PROCEDURE.
9. Turn the incubator power off and disconnect the external supply. Reconnect the negative
battery cable to the battery. Connect one end of the red clip lead to TP7 of the control board
and run the other end outside the incubator. Similarly, connect one black clip lead to TP1 and
one black clip lead to TP2 of display board. Disconnect the sensor simulator from J2 on the
control board and connect the airflow tray cable to J2 on the control board. Connect heater and
ground wires also. Install the inner and outer infant chambers.
10. Using the calibrated temperature measuring device, monitor the temperature in the center of the
chamber 10 cm above the center of the mattress.
12. If the temperature indicated by the monitoring instrument is not 37.0 0.5 C, adjust R3 on the
control board until 37.0 C is achieved. This adjusts for the temperature offset between 10 cm
above the center of the mattress and the location of the control temperature sensor.
14. Adjust R9 until the incubator displayed temperature matches the measured temperature of the
measuring device, 0.5 C. Operate the incubator long enough to allow the displayed
temperature to stabilize and adjust R9 again as needed.
16. Use the two black clip leads to connect the 10 k resistor between TP1 and TP2 on the display
board, while the set point temperature is set at 37.0 C. (This effectively sets the temperature
to 39.9 C). Verify that the HIGH TEMP alarm activates at 38.5 0.1 C as indicated by the
front panel display. The alarm should sound at approximately 8 second intervals. Adjust R8 on
the control board as necessary.
17. As the temperature in the chamber continues to rise, verify that the SYS FAIL alarm activates at
39.0 0.2 C and that the heater is disabled (incubator cools gradually). Adjust R7 on the
control board as necessary. If this adjustment seems to have no effect, check that the
secondary SYS FAIL isn’t activating at too low a temperature (see step 18).
18. With the other black clip lead, ground TP7 on the control board by clipping on to the chassis.
The SYS FAIL alarm should turn off. Verify the SYS FAIL alarm reactivates at 39.2 0.1 C,
and that the heater is again disabled. Adjust R15 on the control board as necessary. This SYS
FAIL alarm should continue to be active even though the incubator cools below 39.2 C - the
incubator must be turned off, then back on, to cancel it.
19. Remove all clip leads. Allow time for the unit to re-establish equilibrium.
20. Disconnect the power cord. The BAT OP indicator should illuminate. The MUTE button will
cancel the AC FAIL alarm. Operate the unit from battery power for five minutes and verify that
the temperature maintained is the same as with AC operation.
CAUTION: The incubator electronics contain static sensitive components that can be damaged
by improper handling. Use approved grounding techniques for work areas and
service personnel.
1. Disconnect the AC power cord from any AC power source. Connect the temperature sensor
test circuit to J2 of the control board. Ensure JP1 on the control board is set to route the
Secondary Thermistor to the Display Board’s “measured temperature” input. (Jumper JP1
should be on positions 1 and 3.) Connect the decade box to the temperature sensor test circuit.
Set the temperature sensor test circuit switches as follows:
2. Reconnect the AC power cord to an appropriate AC power source. To set the battery charging
voltage, turn off the incubator power switch and check the voltage at TP1. With a fully charged
battery, TP1 should be 13.7 0.05 volts. Adjust R1 as necessary.
a. To check the maximum charging current, measure the voltage from TP14 to Z4 pin 7.
The voltage should be approximately 50 10 millivolts.
b. Turn the incubator power switch to on and allow 10 minutes for the voltage references to
stabilize.
d. To check the calibration of the setpoint temperature circuit, adjust the setpoint
temperature to 37.0 C. With a properly calibrated display board, the voltage at TP2
should be 3.399 0.005 volts. Adjust R2 on control board as necessary.
e. To check the calibration of the secondary temperature sensor, measure the voltage at
TP9. It should be 3.421 0.005 volts. Adjust R9 as necessary.
f. To check the SYS FAIL alarm, set the decade box to 1326 . The SYS FAIL alarm
should activate at 39.2 C (when switching from 1325 to 1326 100 ). Adjust R15
as necessary. Achieving the exact temperature and ohm values is not critical now as
this alarm will be readjusted more precisely during the system calibration.
g. Ensure JP1 on the control board is set to route the Primary Thermistor to the Display
Board’s “measured temperature” input. (Jumper JP1 should be on positions 1 and 2.)
Set the temperature sensor test circuit switches as follows:
h. To check the primary sensor calibration, measure the voltage at TP3. It should be 3.399
0.005 volts. Adjust R3 as necessary.
i. Set the decade box to 1356 100 . The HIGH TEMP alarm should activate at 38.5 C
when switching from 1356 to 1355 100 . Adjust R8 as necessary. Achieving the
exact temperature and ohm values is not critical now, as this alarm will be readjusted
more precisely during the system calibration.
3. Set the decade box to 1329 100 (the HIGH TEMP alarm will be on). The SYS FAIL alarm
should activate while switching at 39.0 C (when switching from 1329 to 1328 100 . Adjust
R7 as necessary. Achieving the exact temperature and ohm values is not critical now, as this
alarm will be readjusted more precisely during the system calibration.
4. Turn off the incubator and unplug the AC cord. Remove the negative lead from the battery
post. Connect the variable DC supply in place of the battery.
6. Decrease the voltage output of the DC supply to 10.5 volts. The low battery alarm should
activate.
7. Slowly decrease the applied voltage until the display turns off and the PWR FAIL indicator
illuminates. This should occur at 10.1 0.3 volts. If it does not, perform one of the following
procedures:
CAUTION: The incubator electronics contain static sensitive components that can be damaged
by improper handling. Use approved grounding techniques for work areas and
service personnel.
1. Turn the power switch on and observe for a couple of seconds; all LEDs should be illuminated,
all seven-segmented LED elements turned on, and the audible alarm is on.
3. Adjust R36 for a voltage of 1.809 0.005 volts from TP4 to TP13.
4. Attach decade resistor box across the ground and input lugs of the skin temperature connector.
Set decade resistor box to 1.355 . Adjust R30 for a reading of 3.399 0.005 volts from TP5 to
TP11. The display should show a reading of 37.0 0.1 C when the BABY TEMP button is
pushed.
CONTROL BOARD
This section describes the theory of operation of the incubator’s control board.
TEMPERATURE REGULATION
The air inside the incubator is heated and its temperature is sensed by the components of the air flow
tray. The air flow tray contains the circulating fan, the temperature sensors, the heater, and the heater
switching transistor. The heater power is controlled with pulse width modulation. There are two
separate thermistors in the airflow tray. The first or primary thermistor is used to measure and control
the incubator’s temperature. The second or secondary thermistor is used for display control and an
independent indicator of high temperature.
On the incubator control board are three identical temperature sensor circuits (Z3B, Z9B, Z10B, and
their associated components). Two of these circuits (Z3B and Z10B) convert the two thermistors’
nonlinear resistance values into a voltage proportional to the temperature of the thermistors. The third
bridge circuit (Z9B) converts the voltage output of the setpoint circuit on the display board into a
voltage proportional to the temperature setting.
The difference between the measured temperature and the setpoint temperature is amplified by Z10A
producing an error voltage. Z7B compares this error voltage to a triangular wave produced by a 555
timer oscillator circuit, Z5, R60, and C9. This comparison produces a square wave, the duty cycle of
which is proportional to the error voltage. This square wave is then used to drive the base of the
heater switching transistor. Diode CR12 limits the maximum value of the error voltage seen by
comparator Z7B. The transistor amplifier formed by Q4, R54, R55, R56, R57, and C16, causes this
maximum value limit to decrease approximately as the square of the applied heater voltage. This
results in a relatively constant maximum heater power even though the heater voltage may vary. The
applied heater voltage will vary due to changing AC line conditions, varying states of charge of the
internal battery, or variations in the external DC voltage.
ALARMS
Four comparators monitor the voltage derived from the primary thermistor. Comparator Z11A
activates if the voltage is greater than 5 volts, indicating a shorted thermistor. Comparator Z11D
activates if the voltage is less than 0.9 volts indicating an open thermistor. Activation of either of these
two comparators will turn on the SENS FAIL indicator, sound the audible alarm, and disable the heater
by turning Q3 off.
Comparator Z11B activates when a voltage corresponding to a temperature of 39 C is reached. This
turns on the SYS FAIL indicator, sounds the audible alarm, and disables the heater.
Comparator Z7A monitors the voltage level of the power source. If this voltage drops below 10.8 volts,
battery operation is assumed and the LOW BAT alarm is turned on. Z13 is also enabled causing the
intermittent audible alarm to sound.
Without the application of external power, turning the power switch off will charge a capacitor (C3) on
the display board through R72. Then when the power switch is turned on, this capacitor is switched to
pin 8 of Z1. This provides sufficient power to Z1 to allow it to turn on Q1, if the battery voltage is
sufficient. When Q1 turns on, power is then supplied to Z1 through CR9.
VOLTAGE REFERENCE
Operational amplifier Z9A amplifies the 0.2 volts reference generated by Z8. Potentiometer R4 adjusts
the gain to provide 5 volts at pin 1 of Z9, TP4. This establishes the voltage reference for the control
loop amplifier circuits and all alarm circuits except the fail-safe circuit. The voltage reference for the
fail-safe circuit is generated by Z2 and adjusted by means of R10 to be 5 volts at TP10.
The battery charging is interrupted when the heater is on to prevent excessive current draw on the
transformer. This is accomplished by means of R35 and CR15. When the heater on signal at TP12
goes high, the heater is turned on and causes a current through R35. The current increases the
voltage at the inverting input of Z6 to exceed the 200 millivolt reference level, causing Z6 to turn off
Q5. During the portion of the duty cycle when the heater signal at TP12 output is low, the heater is
turned off and battery charging resumes. Battery charging will be totally interrupted if the duty cycle of
the heater reaches 100%. With average line voltages the maximum duty cycle is limited to less than
65%. When the battery is depleted, maintaining a voltage of 13.7 volts at TP1 would force a charging
current high enough to damage the battery. Therefore, a current limiting circuit is provided. This
circuit is formed by Q6, Z4, and associated components. Zener diode CR14 establishes a reference
voltage and the potentiometer R14 is adjusted to provide a reference of negative 50 millivolts at TP14
with respect to the AC rectified voltage. The resistor R28 is connected to the AC rectified voltage and
supplies the battery charging current. As long as the voltage across R28 is less than 50 millivolts or
the charging current is less than 5 amps, Z4 keeps Q6 turned off. When the voltage across R28
exceeds 50 millivolts, Z4 causes the drain current of Q6 to be increased until the voltage drop across
R27 matches the voltage drop across R28. The drain current of Q6 flows through R34 producing an
additional voltage to the inverting input of Z6. The output of Z6 responds by lowering the voltage to Q5
and the battery charging current. The control loop responds to current rather than voltage and
regulates to maintain a constant current of 5 amps. The current regulation is maintained until the
battery terminal voltage rises sufficiently to cause TP1 to rise above 13.7 volts and voltage regulation
resumes.
DISPLAY BOARD
This section details the functions provided by the circuitry on the display board. Microcontroller U4 is
the chief component of the incubator display portion of the board. When powered on, dedicated
firmware processes key inputs received from the membrane switch panel; continuously monitors the
incubator air temperature, skin temperature, and battery voltage inputs; selects and updates the
seven-segment digits, display mode LEDs, and temperature setpoint; and provides nonvolatile
memory to restore system settings when power is cycled.
Meanwhile, microcontroller U19 is the chief component of the pulse oximeter and oxygen monitor.
When powered on, dedicated firmware processes key inputs received from the membrane switch
panel and continuously monitors the oxygen content and serial communications from the pulse
oximetry board. It continuously selects and updates the seven-segment digits for %SpO2, Pulse, and
Oxygen concentration, as well as the pulse bar graph and the pulse-oximeter / oxygen display mode
LEDs. It provides early warning for too low of a low oxygen alarm setpoint and drives the alarm
output. It also provides nonvolatile memory to restore system settings when power is cycled.
TEMPERATURE SETPOINT
The temperature setpoint is entered from the membrane keypad as described in the Operator’s
Manual. The microcontroller firmware then converts this setpoint to a voltage output, which is written
to the serial digital-to-analog converter U6. This voltage in turn controls the current across fixed
resistor R16 (inserted between TP1 and TP2) and thereby emulates the passive resistor model of an
ordinary thumbwheel switch. The equivalent resistance is input to a temperature sensor circuit on the
control board. The microcontroller U4 also monitors the voltage out of U6 directly to guard against any
failure in the temperature setpoint circuit.
SETPOINT ALARM
This alarm activates when the incubator temperature is more than 1 from the setpoint. The display
will flash HI or LO every 5 seconds. An audio alarm sounds every minute until the incubator
temperature is within 1 of the setpoint.
NOTE: Dimensions and weights are approximate. Height dimensions are given such that when
added the approximate total height is given.
Fo
our draw lattches at the
e corners off the incuba
ator are inte
ended to be
e utilized to mount the incubator to
t
an
n interface that will sec
curely fix the
e incubatorr system in a hospital ssetting or e
emergency vehicle as
ap
ppropriate. It is the cus
stomer’s re
esponsibility
y to ensure that any sa
afety requirrements havve been
ad
ddressed on
n the installation. Unle
ess otherwise specified maximum
m additional equipmen
nt weight
lim
mited to 10 lbs.
l
Figure
F 11 - 1
ELECTR
RICAL SPECIF
S ICATIO
ONS
AC
C Power ------------------------------------------------------------ 1 20 VAC 50
0 - 400 Hz 6 Amps
230 VAC 50
0 - 400 Hz 6 Amps
Acccessories ---------------
- ----------------------------------------- 3 Amps
Exxternal DC Power
P ----------------------------------------------- 1 2 volts 10 A
Amps
Intternal Batte
ery ---------------------------------------------------- O
One 12 volt 26 AH Sea
aled Lead/A
Acid type
Ba
attery Life ---------------------------------------------------------- A
Approx. 200
0 cycles
No
ominal Batte
ery Recharrge (90%) Time
T -------------------- 8 Hr. on AC,, Unit Off
No
ominal Batte
ery Operatiion Time ----------------------------- 3 Hr. Chamb
ber at 37 C Ambient 2
20 C
Exxamination Light ------------------------------------------------- 5 Watt
EMC SPECIFICATIONS
EMC Compliance
The Transport Incubator has been tested and found to comply with limits for electromagnetic
interference and suseptibility as defined by IEC 60601-1-2. However, this equipment may radiate
radio frequency (RF) energy and may cause harmful interference to other devices. The Transport
Incubator is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Transport Incubator can help prevent electromagnetic
interference by operating the device in the environments and with the minimum separation distances
specified in the Service Manual. Additionally, periodic maintenance as specified by International
Biomedical will allow the device to continue to provide basic safety and essential performance.
2
As determined by EN60601-2-20, clause 201.12.1.107. Time to rise 11 C, when control temperature is set 12 C
above ambient.
3
As determined by EN60601-2-20, clause 201.12.4.2.101. Measured 15 cm from 4% CO2 mixture administered at rate
of 750 ml/min, 10 cm above center of mattress.
60% Dip for 5 60% Dip for 5 Mains power quality should be
Cycles Cycles that of a typical commercial or
hospital environment. If the user
30% Dip for 25 30% Dip for 25 of the Transport Incubator
Voltage
Cycles Cycles requires continued operation
Dips/Dropout
during power mains interruption, it
IEC 61000-4-11
> 95% Dip for 5 During the 5 is recommended that the
Seconds Second event, the Transport Incubator be powered
Transport from an uninterruptible power
Incubator supply or internal battery.
switches to
internal battery
power.
Power Frequency Power Frequency magnetic fields
50/60 Hz Magnetic should be that of a typical
30 A/m 30 A/m
Field commercial or hospital
IEC 61000-4-8 environment.
ESSENTIAL PERFORMANCE
The accuracy of the set temperature to the real temperature will be maintained within 2 C at
ambient temperatures between 10 and 20 C and within 1.5 C at an ambient temperature of
25 C 1 C for normal operation.
In the event that the temperature is outside of the prescribed range, an audible alarm will be
supplied.
The warm up time for the incubator with large chamber, as defined by 60601-2-20 section
201.12.1.107 shall be 20 minutes* - as defined in our IFU.
The indication of the temperature shall be accurate within 1 C when measured with a standard
thermometer at a point 10 cm above the middle of the mattress.
The O2 Monitor shall be accurate within 2.5% + 2.5% of gas level (Volume Fraction of Gas
Level).
When the power supply falls outside of the normal values for operation, the incubator shall
switch over to its internal battery, and shall display an indicator that the device is operating on
battery power.
For incubators equipped with SpO2 capability, the SpO2 accuracy shall be less than or equal to
4% over the range of 70-100% SaO2.
For incubators equipped with SpO2 capability, the Pulse Rate accuracy shall be less than or
equal to 5 bpm over the range of 25-240 bpm.
For incubators equipped with SpO2 capability, SpO2 level low shall create an audible and visual
alarm.
For incubators equipped with SpO2 capability, alarms shall be generated in the event of probe
or patient cable failures, or if the SpO2 system is incapable of updating the measured data for a
period of 30 seconds.
4
Masimo sensors have been validated for pulse rate accuracy for the range of 25 - 240 bpm in bench top testing against
a Biotek Index 2 simulator. This variation equals plus or minus one standard deviation, which encompasses 68% of
the population.
5
Masimo SET technology with LNCS sensors have been validated for no motion accuracy in human blood studies on
healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies in the range of
70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard
deviation, which encompasses 68% of the population. The saturation accuracy of the neonatal sensors was validated
on adult male and female volunteers with light to dark skin pigmentation and 1% was added to account for the
properties of fetal hemoglobin.
6
Masimo SET technology with LNCS sensors has been validated for motion accuracy in human blood studies on
healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies while
performing rubbing and tapping motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between
1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory
co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation, which encompasses 68%
of the population.
7
The Masimo SET Technology with LNCS Neo sensors has been validated for neonatal motion accuracy in human
blood studies on healthy adult volunteers in induced hypoxia studies while performing rubbing and tapping motions, at
2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in
induced hypoxia studies in the range of 70-100% SpO2 against a laboratory CO-oximeter and ECG monitor. 1% has
been added to the saturation accuracy to account for the effects of fetal hemoglobin. This variation equals plus or
minus one standard deviation, which encompasses 68% of the population.
8
Masimo SET technology has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2
simulator and Masimo’s simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5%
for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation, which
encompasses 68% of the population.
INTERFERING SUBSTANCES
Interferent Volume % Dry Interference in Oxygen Reading
Nitrous Oxide 75% < 2%
Carbon Dioxide 10% < 2%
Halothane 5% < 2%
Enflurane 5% < 2%
Isoflurane 5% < 2%
Helium 70% < 2%
Sevoflurane 6% < 2%
Desflurane 15% < 2%
9
The NELL-1 PulseOximeter board has been validated, in part, by studies conducted on healthy adult male and female
volunteers, spanning a range of skin pigmentations and ranging in age from 18-50 years old, under controlled
laboratory conditions over the saturation range of 70% to 100%. SpO2 was compared to SaO2 measured to blood CO-
oximetry, and pulse rate was compared to EKG heart rate. Clinical functionality has been demonstrated on a
population of hospitalized neonate and infant patients (MAX sensors: 1-23 days old, weight from 750-4100 grams,
Softcare sensors: 24-40 weeks old, weight from 710-5000 grams).
10
The NELL-1 PulseOximeter board has been validated for low perfusion accuracy using signals supplied by a patient
simulator. SpO2 and pulse rate values were varied across the monitoring range over a range of weak signal conditions
and compared to the known true saturation and pulse rate of the input signals.
Emergo Europe
Prinsessegracht 20
2514 AP
The Hague, The Netherlands
SYSTE
EM WIR
RING DIAGRAM
M, NELLCOR PULSE
EOX (00
02-0016)
CONTR
ROL BO
OARD SCHEM
S MATIC (0
001-518
81)
PULS
SEOX DISPLA
D AY BOA
ARD SCHEMAT
TIC (001
1-5193))
SCHE
EMATIC, NON-P
PULSE
EOX DIS
SPLAY BOARD
D (001-5
5192)
AIRFLO
OW DIA
AGRAM
M (002-4
4000)