This document describes the procedure for determining the Hall coefficient using a Hall effect experiment. The apparatus includes a gaussmeter, electromagnet, constant current source, semiconductor specimen, and voltmeter. Current is passed through the specimen and a magnetic field is applied perpendicular to it. The Hall voltage developed is measured and used to calculate the Hall coefficient based on the specimen's thickness, current passed, and applied magnetic field. Repeating the experiment for different currents allows plotting a graph to determine the Hall coefficient. Then the applied magnetic field can be measured using the known Hall coefficient.

Uploaded by

Hypo Kadambari
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
245 views5 pages

Hall Effect PDF

This document describes the procedure for determining the Hall coefficient using a Hall effect experiment. The apparatus includes a gaussmeter, electromagnet, constant current source, semiconductor specimen, and voltmeter. Current is passed through the specimen and a magnetic field is applied perpendicular to it. The Hall voltage developed is measured and used to calculate the Hall coefficient based on the specimen's thickness, current passed, and applied magnetic field. Repeating the experiment for different currents allows plotting a graph to determine the Hall coefficient. Then the applied magnetic field can be measured using the known Hall coefficient.

Uploaded by

Hypo Kadambari
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 5

Page 1 of 6

Hall Effect

Aim: -
1) To determine the Hall coefficient (RH)

Apparatus:-

1) Gauss meter with probe


2) Electromagnet
3) Constant current source to Pass current through electromagnet
4) Specimen of semi-conducting material with connecting terminals to Pass current (IX)through
it [Green wire terminals] and to measure the Hall voltage (VH) [Red wire terminals]
5) Constant current source to Pass current (IX) through semiconductor specimen
6) Milli-ammeter to measure IX
7) Voltmeter to measure Hall voltage (VH) [5,6 &7 are fixed to Hall effect board]
8) Connecting terminals.
VH t cm3 / Coulumb
Formulae :- Hall coefficient RH 
I X BY
Where VH = Hall voltage developed between the upper and lower faces of the specimen (V)
IX = Current sent through the specimen (A)
t = Thickness of the specimen (Distance between the side faces) (cm)
Theory :- As per the statement of Hall effect “when a magnetic field is applied perpendicular to
a current carrying current conductor, a potential difference is developed between the opposite
faces of the conductor which is perpendicular to both current and magnetic field directions.”
Consider a uniform thick semiconductor placed with its length parallel to X-axis. A
current IX is passed through the conductor along X-axis. A magnetic field BY is applied along Y –
axis, then the charge carriers experience a force (F) perpendicular to X-Y plane i.e. along Z- axis
as per the Fleming’s Left Hand rule as shown in the figure. If electrons are the charge carriers,
they accumulate at the upper surface. This surface acquires negative charge while the lower
surface gets positive charge and some potential difference is developed between these two
surfaces. This is electrostatic field. This voltage is called Hall Voltage (VH) and the Electric field
is called Hall electric field (EH). (In this case the Hall voltage VH is negative, this –ve sign
indicates the charge carriers are negatively charged i.e. electrons. If it is +ve the charge carriers
are holes.
Page 2 of 6

These F and EH act in opposite directions. i.e. The forces due to magnetic field and
electrostatic field on charge carrier act in opposite direction. Ultimately, the net force on the
charge carrier becomes zero.
∴ q (Vd X BY) + q EH = 0 (or) (Vd X BY) + EH = 0
Here Vd is the drift velocity of the charge carrier = J/nq.
Where J = Current density q = Charge of the carrier n = No. of charge carriers/cm3
Here is another physical quantity called Hall coefficient which is equal to the reciprocal
of the amount of charge per unit volume.
1 VH 1
Hallcoefficient RH   EH &  ( )
nq JB d (I X A) BY
Y

VH 1
R ( ) 
VH t
H
d (I X td ) BY IXB Y
Where t = Thickness of the specimen d = Height A = Area of cross-section

This experiment can be used to find 1) the nature of the charge carrier 2) the Hall
voltage 3) current through the specimen 4) applied magnetic field 5) Conductivity of the
conductor etc. But for the present it is confined to measure the unknown magnetic field.
Description :- This experimental set-up consists of 3 main instrumental parts.

1) Digital Gaussmeter with Hall probe :- Hall probe cable is to be plugged-in to the socket
of the digital gaussmeter and the power should be given to the gaussmeter. This probe
also operates basing on the principle of Hall effect. A small current sent through the Hall
probe develops a small Hall voltage when it is placed in a magnetic field and the Hall
voltage is amplified by an amplifier whose output is calibrated in Gauss which directly
gives the magnetic induction (B) value in the gaussmeter.
Page 3 of 6

2) Electromagnet with constant current supply :- Two insulated Copper wires are wound on
two soft iron bars whose faces are facing each other. When a D.C. current (in amperes)
from a constant current source is sent through the coils, the faces of the iron bars acts as
the two poles of a magnet (electromagnet) creating a magnetic field in between them.
The gap between poles can be varied , in general, the gap should be 1 cm.

3) Hall effect board with Hall probe semi-conductor specimen mounted on sun mica PCB :-
A specimen of rectangular semi-conductor slab in which Hall effect is to be studied is
fixed to a printed circuit board (PCB) with the help of 4 supporting terminals. Out of 4
terminals, 2 terminals are along the length and these (Middle & green) terminals are
connected to the current source of the Hall effect board. The other 2 terminals are along
the width and these (Red) terminals are connected to the Voltmeter of the Hall effect
board. The Hall effect board has 2 uses. A) To pass current (IX) through the specimen &
to measure that current. B) To measure the Hall voltage (VH) developed across the
specimen. To meet these two purposes a two mode switch is arranged to the digital meter
of the board. First mode is to measure the current (IX), sent through the specimen and the
second mode is to measure the Hall voltage (VH) developed across the specimen.
Procedure :- This experiment comprises of two parts
1) Measurement of Hall coefficient (RH)
2) Measurement of applied magnetic field (By) by using Hall effect (Application of Hall effect)
Measurement of Hall coefficient (RH)
In the 1st part give the power supply to the gaussmeter. Keep the range switch of the
gaussmeter in minimum range and adjust the zero adjustment such that the reading in the
gaussmeter shows zero. If the gaussmeter does not come to zero, then keep its value at minimum
and take it as zero error. This zero error is to be corrected while taking the final reading of
magnetic induction (BY).
Now adjust the distance between the poles of the electromagnet equal to 1cm &
pass 1A current through the electromagnet from its constant current source, then some magnetic
field (BY) is created between the poles of electromagnet. This magnetic field or magnetic
induction (BY) is measured by the gaussmeter by keeping its probe between the poles of the
electromagnet. The position of the probe is adjusted such that the flat faces of the probe are
perfectly vertical and perpendicular to the magnetic field. Then the gaussmeter shows maximum
value.
Page 4 of 6

Keep the mode switch of the current source of the semiconductor specimen in the current
mode (Here also the zero error and zero correction are to be made) and pass 0.5 mA (or) 1 mA
current (IX) through the specimen. Place this specimen between the poles of the electromagnet
such that the faces of the specimen are perfectly vertical and note the voltage (V 1) developed
between the upper and lower faces of the specimen after turning the mode switch of the current
source of the specimen in to voltage mode. Note the values of IX, V1 in the table – 1 and calculate
VH. Repeat the experiment for different values of IX by increasing its value in equal intervals of
0.5 mA.

Graph :- Draw the graph by taking the current through the specimen IX on X- axis and Hall
voltage VH on Y- axis. This gives a straight line passing through the origin. Take a particular
value of IX and note its corresponding value of VH
Note the thickness (t) of the specimen which was noted on the board of the specimen.
Substitute the values of IX, VH, t and BY in the formula – 1 and calculate Hall coefficient (RH)
Measurement of applied magnetic field (By)
In the 2nd part of the experiment keep the current through the semi-conductor specimen
(IX) at constant value of 1 mA and pass 1A current through the electromagnet and measure the
voltages V1 and V2 by placing the specimen between the poles of the electromagnet & from that
calculate the Hall voltage VH (as measured in the 1st part). Also measure the magnetic induction
(BY1) with gaussmeter by placing its probe between the poles of electromagnet.
Substitute the values of IX, VH, t and Hall coefficient RH (as calculated in 1st part ) in the
formula – 2 and calculate value of applied magnetic field (BY2). The experiment is repeated by
increasing the current through the electromagnet (it means by changing the magnetic field B Y ) in
equal intervals of 0.5 A. Note the values in the table – 2.
In the 2nd part the experiment the magnetic field BY is measured with gaussmeter as BY1
and also measured by using the Hall effect as BY2. These two are compared in the table – 2.

Precautions :-
1) Electromagnet power supply should be connected to a 3 pin main sockethaving good earth
connection.
2) Switch “ON” or “OFF” the power supply at zero current position.
3) Adjust the distance between the poles of the magnet nearly 1 cm, then only the gaussmeter
shows correct reading.
Page 5 of 6
Table - 1
Thickness of the semi-conductor specimen t = cm.
Current through the Electromagnet = 1 A
Applied magnetic field measured with gauss meter BY =

Current through Measurement of Hall Voltage


S. No. The specimen VH (m V) Hall Coefficient
RH (cm3/c)
IX (m A)

0.5

1.0

1.5

Measurement of Slope (VH/I) by plotting I Vs VH graph:

Result:

Hall Coefficient RH = - - - - - - - - - - cm3/coulumb