Operation Manual
Operation Manual
Operation
Product
Overview
Basic
Operation
FDR-1000DRLZ/FDR-1000DRSZ
Operation Manual Troubleshooting
3rd Edition May 2009
Daily Inspection
and
Maintenance
Appendix
Maintenance
and Inspection
This Operation Manual provides necessary information for using the FUJIFILM Digital
Mammography FDR-1000DRLZ and FDR-1000DRSZ (hereinafter collectively referred to as
“exposure unit”), such as the equipment overview, operation procedures and precautions to
observe, as well as daily inspections and maintenance.
This Operation Manual contains descriptions of both types of the exposure units. Read this
manual after confirming the type of your exposure unit.
CAUTIONS
1 No part or all of this manual may be reproduced in any form without prior
permission.
2 The information contained in this manual may be subject to change without prior
notice.
3 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting
from installation, relocation, remodeling, maintenance, and repair performed by
dealers other than those specified by FUJIFILM Corporation.
4 FUJIFILM Corporation shall not be liable for malfunctions and damages of
FUJIFILM Corporation products due to products of other manufacturers not
supplied by FUJIFILM Corporation.
5 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting
from remodeling, maintenance, and repair using repair parts other than those
specified by FUJIFILM Corporation.
6 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
negligence of the precautions and operating methods contained in this manual.
7 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting
from use under environment conditions outside the range specified for this product,
such as the power supply, installation environment, etc., as described in this
manual.
8 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting
from natural disasters, such as fires, earthquakes, floods, lightning, etc.
Trademarks
FDR is a trademark or a registered trademark of FUJIFILM Corporation.
Chapter 4 Troubleshooting
This chapter describes actions should be taken if problems occur
in the the exposure unit.
Item title When using the FDR MS-1000 system, operations are required on the AWS. See “FUJIFILM Digital
Mammography System FDR MS-1000 Operation Manual” and “FUJIFILM Mammography Workstation
Index
FDR-1000AWS Operation Manual” for details.
Indicates the operation that is Allows you to open a
covered under the item. desired chapter
3.2.1 Starting Up the System
It takes approximately 30 minutes (including automatic calibration) until you can make an exposure after
quickly.
pressing the System On button.
NOTE
Before starting up the system, make sure to release compression by using the decompression button.
Otherwise, an error occurs.
Operation procedure
Basic Operation
A sequence of numbered
Breaker ON
operational steps to guide the
user through an operation.
Breaker OFF
Circuit breaker
2 Press the System On button of the control pad to turn the exposure stand on.
Calibration starts automatically after powering on the exposure stand.
System On button
NOTE
When it passes 1:00 a.m. without shutting down the system, calibration starts automatically.
Monitor
Power switch
Page number
Indicates the order of a particular
page in a chapter.
Page
We ask that you heed these usage precautions and use the equipment correctly.
4. If there is trouble with the equipment, do not attempt to fix it randomly. Instead, do what is
indicated and entrust repairs to a professional.
7. Other Items
7-1. When subjecting patients (particularly infants and pregnant women) to radiation, make
sure not to exceed the necessary time and dose. Also, ensure that radiation is contained
within the region for exposure.
7-2. For cleaning and disinfecting the parts that patients come in contact with in a normal
usage environment, slightly moisten a soft cloth (non-woven cloth: Kimwipe®) in neutral
detergent or ethanol for disinfection and wipe away any dirt. Do not use any other
cleansers or disinfectants.
Also, do not scrub the equipment’s surface vigorously.
7-3. Follow the Operation Manual and operate the equipment correctly.
Preventing explosion
WARNING
For Safe Operation
Because this equipment is not explosion-proof, do not use combustible and explosive
gases near the equipment.
WARNING
Flammable gasses may stay in the study room after disinfection. Ventilate the room well
before powering on the system following disinfection.
WARNING
WARNING
Do not use the equipment in a location where the equipment may be exposed to any
liquids. Never allow any liquid to spill on or into the equipment. Otherwise, electrical shock
may occur.
Should you suspect any liquid has penetrated into the equipment, immediately stop using
it and contact our official dealer.
WARNING
Do not use the equipment in a location where metal particles could come into the
equipment. This may cause an electrical shock.
WARNING
• Make sure to turn off the equipment before cleaning or disinfecting parts other than the
compression plate, exposure table, armrests, and face guard. Otherwise, you may
receive an electrical shock.
• During cleaning or disinfecting of the equipment, keep the temperature of the study
room 15°C to 35°C to ensure proper performance of the exposure unit.
Follow the specified procedure when turning off the equipment. Otherwise, the flat panel
sensor could be damaged by thermal shock.
CAUTION
Do not store magnetic media near the exposure stand or control cabinet of the exposure unit.
Otherwise, magnetism generated by the equipment may cause the data to be lost.
WARNING
Make sure that the devices to be connected to the equipment are authorized for
connection.
WARNING
Make sure that no patient is around the equipment before powering on it after an electrical
outage.
WARNING
Inspect the equipment before leading a patient to the equipment. Should you find any
failure, contact our official dealer.
WARNING
Before leading a patient to the equipment, make sure that the foot switch is away from the
path to the equipment and outside of the movable region of the swivel arm. Otherwise, the
patient or technologists may accidentally touch the foot switch in the way to the
equipment, resulting in an unintended operation.
WARNING
• Make sure to keep observing movements of both equipment and patient while
positioning the movable parts at the patient's body.
• When making an exposure of a seated patient, prepare a height-adjustable chair and
lower the seat to the end.
WARNING
Make sure that a patient’s hand or foot does not touch the swivel arm control panel,
control pad or foot switch.
WARNING
Exposure should be made in an environment where the technologist can visually and
aurally determine the patient’ s condition.
If the equipment malfunctions while adjusting the movable parts or performing a study
immediately press the emergency shut-down button and contact our official dealer.
WARNING
To ensure complete system isolation, never install any unauthorized accessories or other
For Safe Operation
such items.
When it is necessary to install authorized accessories or optional items, contact our
official dealer.
WARNING
Keep equipment other than the exposure stand out of reach of patients to ensure
appropriate system isolation.
WARNING
• Operate the equipment by following the instructions given in the Operation Manual.
Wrong operations could expose a technologist or patient to unnecessary X-ray exposure.
• Make sure that no one other than a technologist or patient is in the study room when
making X-ray exposures. If it is necessary for a care assistant or other person to remain
in the room, provide sufficient radiation protection.
WARNING
Use the longest possible focal spot to skin distance to minimize the absorbed dose.
WARNING
Reduce the radiation field size to the minimum necessary size for the exposure. If X-rays
are unavoidably radiated to areas outside of the subject region, take other protective
measures such as readjusting the collimator or wearing a Pb apron.
m
5
1.
Patient
WARNING
See “FUJIFILM Digital Mammography System FDR MS-1000 Operation Manual” for details on the
protective shield (optional).
WARNING
A patient’s or technologist’s hand, foot or other body parts may hit or be caught by the
potentially dangerous parts. Eliminate all dangers completely.
Swivel arm
This part rotates.
Face guard
Space between the
compression plate and
face guard or swivel
arm (when the face
guard is not installed)
Compression plate
Exposure table
This part moves up/down.
When operating the equipment by the foot switch, swivel arm control panel or control pad,
make sure that no hand, foot, or other part of the technologist's or patient's body is
caught between the movable parts.
WARNING
For Safe Operation
Press the emergency shut-down button to immediately turn off the equipment in an
emergency. This cuts off the power to the exposure stand, forcibly shutting down the
system.
The equipment has the three buttons: two on the exposure stand and one on the control pad.
Mammography-specific instructions
Only qualified technologists are allowed to operate the equipment. The technologists are required to fully
understand the principles of mammography in advance to perform operations described in this Operation
Manual.
Face guard
The face guard is designed to prevent the patient's head from entering the radiation field. It is not X-ray proof.
WARNING
When making an X-ray exposure without using the face guard, make sure that the
patient's head is physically positioned outside of the X-ray path.
Disinfection instructions
WARNING
Confirm that the respiratory density of disinfectant including solvent is under legal
regulation.
Certain disinfectants may damage health. When using a disinfectant, follow instructions
supplied by the manufacturers.
WARNING
Do not use any of the disinfectants described below as damage to the equipment may
result. Quality, performance and safety of the equipment cannot be assured.
• Chloric disinfectant which is strongly corrosive to metals and rubber parts.
• Disinfectant whose uses on metals, plastics, and coating are forbidden according to the
instructions supplied with the disinfectant.
• Formalin gas and disinfectant sprays that may get inside the equipment.
Disinfectant ethanol is recommended for disinfection.
Carefully read the instructions and cautions supplied with the disinfectant before use.
WARNING
Conformance to standards
Laser notice
This equipment is the Class 1 laser product which conforms to IEC 60825-1:1993+A1:1997+A2:2001.
Classification
1) According to the type of protection against electrical shock
CLASS 1 EQUIPMENT
2) According to the degree of protection against electrical shock
Type B applied part
3) According to the degree of protection against harmful ingress of water
Main Unit: IPX0
Foot Switch: IPX1
4) According to the degree of safety of application in the presence of a flammable anesthetics mixture with
air or with oxygen or nitrous oxide.
Equipment not suitable for use in the presence of a flammable anesthetics mixture with air or with
oxygen or nitrous oxide.
5) According to the mode of operation
Continuous Operation
This equipment complies with the following standard regarding Electromagnetic Compatibility (EMC).
EMC standard : IEC 60601-1-2:2001+A1:2004 (EN 60601-1-2:2001+A1:2006)
These limits are designed to provide reasonable protection against harmful interference in a typical medical
installation.
For Safe Operation
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to other devices in the vicinity.
However, there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to other devices, which can be determined by tuning the
equipment off and on, the user is encouraged to try to correct the interference by one or more of the following
measures;
• Reorient or relocate the receiving device.
• Increase the separation between the equipment.
• Connect the equipment into an outlet on a circuit different from that to which the other device(s) are connected.
Consult a FUJIFILM dealer for help as necessary.
This Operation Manual contains the names of standards applied in IEC 60601-1-2.
The names of standards applied in EN 60601-1-2 are replaced as follows.
IEC 60601-1-2
Standards applied in EN 60601-1-2
(descriptions in this Operation Manual)
CISPR 11:1997+A1:1999+A2:2002 EN 55011
IEC 61000-3-2:2000+A1:2001+A2:2004 EN 61000-3-2
IEC 61000-3-3:1994+A1:2001+A2:2005 EN 61000-3-3
IEC 61000-4-2:1995+A1:1998+A2:2000 EN 61000-4-2
IEC 61000-4-3:2006 EN 61000-4-3
IEC 61000-4-4:2004 EN 61000-4-4
IEC 61000-4-5:2005 EN 61000-4-5
IEC 61000-4-6:2003+A1:2004+A2:2006 EN 61000-4-6
IEC 61000-4-8:1993+A1:2000 EN 61000-4-8
IEC 61000-4-11:2004 EN 61000-4-11
• T he use of accessories, transducers and cables other than those specified, with the exception of
transducers and cables sold by FUJIFILM Corporation as replacement parts for internal components, may
result in increased emissions or decreased immunity of the FDR-1000DRLZ/FDR-1000DRSZ.
• The FDR-1000DRLZ/FDR-1000DRSZ should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the FDR-1000DRLZ/FDR-1000DRSZ should be observed to verify
normal operation in the configuration in which it will be used.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the FDR-1000DRLZ/FDR-
1000DRSZ is used exceeds the applicable RF compliance, the FDR-1000DRLZ/FDR-1000DRSZ should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the FDR-1000DRLZ/FDR-1000DRSZ.
b Over the frequency range 150 kHz to 80 MHz, field strength should be less than 3 V/m.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
See “FUJIFILM Digital Mammography System FDR MS-1000 Operation Manual” for other related
precautions.
Minimizing exposure
For Safe Operation
• Use this equipment only for breasts. If used for other body parts, the patient may be subject to an
unnecessary X-ray dose or be injured.
• Before making an X-ray exposure, check if the patient has breast (mammary) implants. Inform the patient of
risks in an X-ray exposure of the breasts implants. Breast (mammary) implants affect the AEC function by
blocking X-rays or shifting breast tissues. For an exposure of the patient with breast (mammary) implants, use
manual mode. Take great care while compressing the patient's breasts because the bag(s) used for implants
may break when compressed.
• Stand at the rear of the protective shield (optional) while making X-ray exposures. When making an X-ray
exposure without using the protective shield (optional), install the protective window or take other measures to
protect the room adjacent to the study room.
• If not using the protective shield (optional), wear protective clothing (0.5 mm Pb equivalent) or stay in the low
level radiation area 1.5 m or more away at the rear of the patient back.
• Wear a film badge or pen dosimeter to monitor exposure doses you receive.
• Make sure that the X-ray radiation field is appropriate before making an X-ray exposure. Otherwise, a
misdiagnosis or an unnecessary increase in X-ray radiation dose may result.
• Make sure that no body part other than the breast to be exposed is in the radiation field. Adjust the equipment
if necessary so that the radiation field covers only the breast to be exposed, using the light field lamp as a
guide. Otherwise, the patient may subject to unnecessary X-ray exposure.
• Specify the correct compression force for an exposure. Otherwise, image quality degrades, resulting in a
misdiagnosis or an unnecessary increase in X-ray radiation dose.
• The damaged equipment should be ensured that the damaged equipment has no adverse effect on X-ray
exposures. Otherwise, a misdiagnosis or an unnecessary increase in X-ray radiation dose may result.
• Never apply a strong impact to the exposure table. Otherwise, a misdiagnosis or an unnecessary increase in
X-ray radiation dose may result.
• Only qualified technologists are allowed to make X-ray exposures.
Preventing accidents
• Keep the exposure table, compression plate or other parts touched by patients or technologists clean. If the
parts are soiled with patient's blood or other infectious material, disinfect the parts immediately with a
disinfectant, etc.
• Before making an X-ray exposure, make sure that there is no sign of abnormality on the exposure table. A
flaked or scratched surface could injure the patient.
• Before using the compression plate, make sure that it is not worn out. If the plate has any crack, replace it
with the new one. Otherwise, the plate may break while it is used and the sharp edges could injure the patient.
• Before adjusting the compression force on breasts, make sure that no part or all of the patient’s body is at the
risk of hitting the equipment or being caught by the equipment.
• Pay attention to the patient's behavior to prevent the patient from hanging onto or leaning against the swivel
arm. Otherwise, the patient may be injured or the arm may be deformed.
• Before adjusting the angle or height of the swivel arm, make sure that a part or all of the patient's body is not
at the risk of hitting the arm or being caught by the part.
Preventing malfunctions
• Do not use inappropriate detergent or disinfectant. Otherwise, it may cause cracks on the surface or inside of
the covers. Such detergent or disinfectant penetrates into the equipment and could cause fire.
• Use a disinfectant that will cause no damage to the equipment or harm to the human body when disinfecting
the equipment.
• Install a cover on the collimator when using the X-ray tube unit rotated ± 180°. If blood or other material
penetrates into the collimator, the equipment may malfunction.
• Pay attention to the patient's behavior and prevent the patient from hitting or leaning against the exposure
stand or swivel arm during image reading. Otherwise, image quality may be degraded, affecting image
interpretation.
• Do not calibrate the equipment with any objects left on the exposure table. The calibration may fail and
artifacts may appear on images. If artifacts appear, redo the calibration.
• Clean the exposure table, compression plate and other parts before making an exposure. Dirt or dust
adhering to the parts may produce abnormal images, resulting in insufficient image quality.
Other limitations
■ Exposure stand
3 1. System Label
4 4
1
2
2. Rating Label
3. Stop Label
When moving the swivel arm up/down time after time, allow a
60-second or longer interval between 16-second movements.
XXXXXXX
10139752
ExposureSW(フットタイプ) 126Y100016
XXXXXXX
See “FUJIFILM Digital Mammography System FDR MS-1000 Operation Manual” for details of optional
items.
14404867
14404866
14404859
14404858
14404862
14404863
14409863
14409864
14409803
See “FUJIFILM Digital Mammography System FDR MS-1000 Operation Manual” for details of optional
items.
2. Stop Label
9 7
Sign Meaning
Minimum necessary interval between successive
exposures
When making X-ray exposures one after another,
allow a longer interval between exposures than that
indicated in the graph “Minimum exposure interval
for energy input to X-ray tube”. Do not make
exposures more than 60 times in an hour. Otherwise,
X-ray exposure may be disabled for a long time due
to heat generated from the X-ray tube.
See the graph “Minimum exposure interval for energy
input to X-ray tube” in “A.1 Specifications”.
See “FUJIFILM Digital Mammography System FDR MS-1000 Operation Manual” for details of optional
items.
Sign Description
Stop
For Safe Operation
Power-OFF (partial)
Power-ON (partial)
Alternating current
Type B application
High voltage
Year of manufacture
Operation-on icon
Operation-off icon
Exposure icon
The installation requirements for the exposure unit are described below.
1580
427 1175
Maximum height of
the swivel arm 80
15
Vertical movement range
2350
70
0
600
920
1100
Unit: mm
565 265
For Safe Operation
570
1.7.3 Necessary Space for Installing the Exposure
Stand and Control Cabinet
The exposure stand and control cabinet should be installed in a location having the space shown in the
figures below at minimum.
Control cabinet
0(*1) Unit: mm
500
100
Cable length:
Approx. 4200
2690
Exposure stand
1500 1500
990
Top view
Unit: mm
2400 (*1)
2100 (*2)
(*1): When using the exposure unit without changing the factory settings
(*2): When the upper limit of the vertical movement was reduced 300 mm
at the time of installation
Right-side view
Page
Part names and functions of the exposure unit are described below.
■ Exposure stand
Emergency shut-down button
(Located at the left- and right-hand sides of the column)
Product Overview
Column
The swivel arm moves up or down this column.
Swivel arm
Mammography exposure unit equipped
with swivel function.
Progress bar
1 Compression foot Step on this switch to move the 4 Swivel arm down Step on this switch to lower the
switch* compression plate of the swivel arm foot switch swivel arm.
to strengthen breast compression.
5 Swivel arm up foot Step on this switch to raise the
2 Decompression Step on this switch to move the switch swivel arm.
foot switch compression plate of the swivel
arm to release breast compression. 6 Swivel arm control Use this to raise or lower the
foot switch swivel arm.
3 Compression/ Use this to start compression or
decompression decompression of the compression
switch plate of the swivel arm.
* Lightly stepping on the switch turns on the light field lamp, indicating the radiation field.
Product Overview
Press this to lower This button cannot the compression
the swivel arm. pressure.
be used.
Light field lamp button Compression plate
Press this to turn on or off This compresses breast
the light field lamp. during exposure.
Flat panel sensor
Exposure table Flat panel detector which
Set breast on this table during converts X-ray images
exposure. directly into electric signals.
■ Control cabinet
Principally no user operation is required.
Breaker ON
CAUTION
■ Control pad
Emergency shut-down button
Shut-down button in case of emergency.
Press this once to shut off the power to the
exposure stand and stop it forcibly.
Operation desk
Place the AWS on this.
Product Overview
See “FUJIFILM Digital Mammography System FDR MS-1000 Operation Manual” for optional items.
Page
2 Perform calibration.
Basic Operation
7 Perform exposures.
NOTE
Before starting up the system, make sure to release compression by using the decompression button.
Otherwise, an error occurs.
Basic Operation
Breaker ON
Breaker OFF
Circuit breaker
2 Press the System On button of the control pad to turn the exposure stand on.
Calibration starts automatically after powering on the exposure stand.
System On button
NOTE
When it passes 1:00 a.m. without shutting down the system, calibration starts automatically.
Monitor
Power switch
Power switch
Personal computer
If the exposure unit has not been used for a week or more, warm up the equipment after startup.
See “5.2.4 Warming Up the Exposure Stand” for details.
Swivel arm
Face guard
Basic Operation
2 Press the lock release in the left-hand arrow direction in the figure below to unlock the
compression plate.
NOTE
When the peripheral area of the compression plate socket is visible, the lock is released.
Peripheral area of the compression plate socket
Lock release
Compression plate
CAUTION
Be careful not to pinch your fingers when uninstalling the compression plates.
Lock release
Compression plate
NOTE
When the compression plate is correctly installed, the collimator automatically adjusts itself to produce the
radiation field of the plate and the light field lamp lights on the exposure table approximately 20 seconds.
Basic Operation
5 Confirm that the peripheral area of the compression plate socket is hidden under the lock
release.
NOTE
If the peripheral area is visible, the compression plate is not locked. Insert it further.
Lock release
Pole Peripheral area of the compression plate socket
CAUTION
• Make sure that the compression plate is securely installed. If it is not correctly locked, it
could drop off and cause injury to the patient or technologist.
• Be careful not to pinch your fingers when installing the compression plates.
Remark
Name Type
FDR-1000DRLZ FDR-1000DRSZ
Not used. — —
Basic Operation
Compression plate for magnification MAG Optional
NOTE
• If using the optional magnification table, attach it in this step.
See “FUJIFILM Digital Mammography System FDR MS-1000 Operation Manual” for details on
the magnification table.
• W
hen using the compression plate for rectangle spot magnification or for magnification, be sure to use
the magnification table (optional).
1 Press the System Off button of the control pad to turn the exposure stand off.
2 Confirm that the message “The line is being disconnected.” is displayed on the monitor,
and then shut down the system on the AWS.
Basic Operation
NOTE
When it passes 0:00 a.m. with the exposure stand turned on, calibration automatically starts at 1:00 a.m.
Since the system cannot be operated during calibration, wait until it finishes before starting operation.
3 Select in the “Patient Information Input window”, and then select from
the displayed menu.
Select . The application software of the AWS ends, and the power of the personal computer is
automatically turned off.
NOTE
As the AWS is not supposed to operate continuously all day long, turn the personal computer off once a
day by performing shutting down operation.
Do not press the power switch when turning off the personal computer.
Basic Operation
3.3.2 Emergency Shutdown Operation
1 Press one of the emergency shut-down buttons.
The exposure stand operation (swivel arm movement or X-ray radiation) stops.
The compression plate moves to the decompression position and stops.
If the compression plate is not released, rotate the compression adjustment knob to manually release it.
Before restoring the system, make sure that the patient is away from the exposure stand.
1 Turn the depressed emergency shut-down button clockwise to release its locked state.
2 Select [OK] in the displayed error message window to restore the system.
For example, the following message appears: “Emergency stop operation[00015] :
Pressing the button below will start restore processing on the exposure unit.”
NOTE
Information in an exposure or input in progress at the time of the shutdown cannot be restored.
Basic Operation
3 When the system is restored, the “Patient Information Input window” appears. Retry
operations from the patient information input.
5 Make sure that the system is shut down completely, wait 5 seconds and restart the system.
See “3.2.1 Starting Up the System” for details.
Page
When the exposure unit does not operate normally, see the relevant pages to take necessary actions
according to the symptom shown below.
Should you encounter any trouble you cannot handle, contact our official dealer.
Possible problems
This section describes necessary actions to clear error messages that may appear on the AWS monitor.
Troubleshooting
and the paddle. used with the exposure table. plate or exposure table.
The compression paddle is not set An incorrect compression plate is Install the correct compression
up correctly. installed or no compression plate plate.
is installed. This message is displayed while
no compression plate is installed.
Error
Message Cause Remedy
code
50950 Hardware error[50950] : Pressing the button Hardware error Restart the system.
below will start restore processing on the
exposure unit.
50951 Firmware error[50951] : Pressing the button Firmware error
below will start restore processing on the
exposure unit.
50952 Power Supply error[50952] : Pressing the Power Supply error
button below will start restore processing on
the exposure unit.
50953 Door switch error[50953] : Pressing the button Door switch error
below will start restore processing on the
exposure unit.
50954 Emergency stop operation[50954] : Pressing Emergency stop
the button below will start restore processing operation
on the exposure unit.
Troubleshooting
Troubleshooting
: Aborts calibration.
1 Remove the compression plate, and magnification table if installed, from the exposure stand
and confirm that there is no object on the exposure table.
See “3.2.2 Replacing the Compression Plate” and “FUJIFILM Digital Mammography System
FDR MS-1000 Operation Manual” for how to remove the compression plate and magnification table.
Troubleshooting
2 Select and then [Field Calibration] in the menu.
To abort it, select [Abort] within 20 seconds.
To switch the type to the Panel Check, select [Switch] within 20 seconds.
When 20 seconds have passed after selecting [Field Calibration], the calibration can no longer be
aborted or switched.
If the calibration is aborted, redo it from the start.
3 When the message requesting an X-ray radiation appears, press the release radiation switch.
The exposure conditions are automatically set to the following values.
CAUTION
If plates with a prickly feel or cracks are used, they may break during
compression and cause serious injury to patients. Make sure to
replace them.
3 Motor condition • Make sure that the motor works smoothly by operating the swivel
arm control panel and foot switches.
• If the motor does not work smoothly, turn off the exposure stand and
contact our official dealer.
4 Height and angle adjustment • Make sure that you cannot raise, lower, or rotate the swivel arm when
of the swivel arm the compression force is set to 30N or more.
• If the swivel arm operates in the above mentioned condition, turn off
the exposure stand and contact our official dealer.
TIP
During starting up of the exposure stand, the release radiation switch
briefly turns yellow to inspect LED.
1 Remove the compression plate and magnification table from the exposure stand to clear
the space between the radiation source and exposure table.
2 Press (exposure mode setting button) and select [Semi] (semi auto).
3 Specify the tube voltage with (tube voltage setting button) and the target/filter with
(target/filter setting button).
The recommended settings are as follows:
•23 kV and Mo/Mo
4 Make an exposure.
The release radiation switch lights for a while (approximately one second).
The displayed mAs is under 20 mAs.
CAUTION
If the error message does not appear, shut down the exposure stand and contact our official
dealer.
8 Follow the message and select [OK] to start the restore processing.
When the restore processing finishes, the AEC normally operates.
■ Cleaning/Disinfection
Turn off the exposure stand before cleaning or disinfection.
Note that there is no need to turn off the exposure stand before cleaning the following items that touch
patients during studies.
• Compression plate
• Exposure table/Magnification table
• Armrests
• Face guard
CAUTION
Make sure to use a recommended detergent to clean the equipment. Improper detergent may
damage the equipment.
CAUTION
Make sure not to let detergent get inside the equipment during cleaning.
Disinfection
Ethanol for disinfection is recommended.
Parts to be disinfected:
• Compression plate
• Exposure table/Magnification table
• Face guard
WARNING
Certain disinfectants may damage health. When using a disinfectant, follow instructions
supplied by the manufacturers.
CAUTION
Do not use any of the disinfectants described below as damage to the equipment may result.
Quality, performance and safety of the equipment cannot be assured.
• Chloric disinfectant which is strongly corrosive to metals and rubber parts.
• Disinfectant whose uses on metals, plastics, and coating are forbidden according to the
instructions supplied with the disinfectant.
• Formalin gas and disinfectant sprays that may get inside the equipment.
CAUTION
The compression plates and face guard gradually wear out due to disinfectant. Make sure to
perform daily inspection. Do not use these parts if any defects are found. Otherwise, the
patient may be injured.
WARNING
If the swivel arm still operates, turn off the breaker that supplies power to the exposure
stand and contact our official dealer.
2 After a minute, confirm that changes of the value are within the range of ±20N.
WARNING
If the compression force changed more than ± 20N, turn off the exposure stand and contact
our official dealer.
“FDR Mammography QC Program” (optional) is available for you to easily perform quality control
of multiple items, especially those related to image quality.
Contact our official dealer for details.
2 Set the exposure menu items. Enter any information for making a test exposure.
See “FUJIFILM Digital Mammography System FDR MS-1000 Operation Manual” for details.
Daily Inspection and Maintenance
Page
Maximum
Operation time
Minimum
temperature
ambient
Reference
Attenuation Equivalent
• Exposure table : 0.1 mmAl
• Magnification table: 0.15 mmAl
• Inner cover: 0.01 mmAl
Operating Mode
Continuous operation with intermittent loading
Continuous Operation
60 exposures/hour (28 kV/100 mAs)
Main Specifications
Room temperature 20 °C
Average rotor power 124 W
400
350
Min. exposure interval [s]
300
250
200
150
100
50
0
0 2000 4000 6000 8000 10000 12000 14000 16000 18000 20000
Input for exposure [kV * mA * s]
Main Specifications
Direct conversion
X-ray Conversion Material
a-Se
Effective Image Area
(1) FDR-1000DRSZ
177 mm (vertical) x 237 mm (horizontal)
(2) FDR-1000DRLZ
236.4 mm (vertical) x 296.4 mm (horizontal)
Pixel Resolution
50 x 50 μm/pix
Number of Pixels
(1) FDR-1000DRSZ
3540 (vertical) x 4740 (horizontal) = 16,779,600 pixels
(1) FDR-1000DRLZ
4728 (vertical) x 5928 (horizontal) = 28,027,584 pixels
Density Resolution
16 bit
Spatial Frequency Limit
10.0 lp/mm
Calibration Frequency
At least 1 time/day
The external view and weight of the exposure unit are described below.
Exposure stand Approx. 920-1,580 Approx. 1,175 Approx. 1,930-2,350 Approx. 345
Control cabinet Approx. 265 Approx. 565 Approx. 570 Approx. 21.5
NOTE
Dimensions and weight are subject to change due to the specification change, etc.
Main Specifications
16.5 cm
plate (Low)
0.5 cm
0.5 cm
Chest wall edge Chest wall edge
18 x 24 compression 24 cm
24 cm
plate (High)
18 x 24 compression
18 cm
16.5 cm
plate (Low)
0.5 cm
0.5 cm
Chest wall edge Chest wall edge
Main Specifications
2D biopsy compression 24 cm
24 cm
plate
18 cm
16.5 cm
0.5 cm
0.5 cm
Compression plate,
rectangle spot
9 cm
9 cm
0.5 cm
Compression plate,
rectangle spot
12.4cm
magnification
12.4cm
0.5cm
16.5 cm
16.5cm
0.5 cm
0.5cm
Chest wall edge Chest wall edge
8 cm
0.5 cm
Chest wall edge
1 When the compression plate is correctly installed, the collimator automatically adjusts itself to
produce the radiation field of the plate and the light field lamp lights on the exposure table
approximately 20 seconds. Check that the light field size corresponds to the radiation field size
of the compression plate.
Main Specifications
NOTE
If the compression plate is not correctly installed, the light filed lamp does not light. In this case, check and
correct the compression plate installation.
See “3.2.2 Replacing the Compression Plate” for how to replace the compression plate.
NOTE
Check that the installed compression plate is displayed.
Remark
Name Type
FDR-1000DRLZ FDR-1000DRSZ
Not used. — —
A.4.1 Coverage
■ Subjects
[Auto] : Automatically controls the target/filter, X-ray tube voltage and mAs.
[Semi] : Automatically controls the mAs.
Manually select the target/filter and X-ray tube voltage.
[Manu] : The automatic control is not executed.
Manually select the above parameters.
See “FUJIFILM Mammography Workstation FDR-1000AWS Operation Manual”for details on the
exposure modes.
23 kV - 35 kV
• AEC area
Main Specifications
The AEC detects X-rays in the shadowed area shown in the figure below.
11cm
9.8cm
2.5cm
0.5cm
NOTE
To detect X-rays properly, object must cover AEC area at least 2.5cm from chest wall side.
Dose level
Compressed
breast H-mode L-mode W-mode
thickness
(mm) Tube Tube Tube
Target/Filter Target/Filter Target/Filter
voltage voltage voltage
20 Mo/Mo 26 Mo/Mo 26 W/Rh 26
In [Manu] (manual mode), specify the exposure conditions using the following table as a guide since the
exposure is not controlled automatically.
Compressed breast
Tube voltage (kV) Target/Filter mAs
thickness (mm)
20 26 Mo/Mo 22
30 27 Mo/Mo 28
40 28 Mo/Rh 45
45 28 Mo/Rh 56
50 28 Mo/Rh 63
Main Specifications
60 30 Mo/Rh 80
70 29 W/Rh 220
* The above values are the recommended parameters for exposures in L-mode (low dose).
* First select an appropriate target/filter for the thickness and density of the breast to be examined, and
then select the suitable tube voltage (kV) and mAs.
(*) First the incident dose is estimated from the exposure conditions (target/filter) and actual dose (mAs/kV) of an
exposure. Then the absorbed dose is calculated from the incident dose and compressed breast thickness by
following the EUREF guidelines.
The absorbed dose can be automatically calculated only when compressed breast thickness is between 20 to
100 mm.
Main Specifications