The ELISA Plate Reader is a user friendly micro plate Analyser. It is compact & lightweight. It is designed to measure and interpret
enzyme immunoassay results, both monochromatically and Bi-chromatically. It is intended for in vitro diagnostics.

CAT# : DIA2000

Special Features
The ELISA Reader can accommodate a flat bottom as well as a round
configuration.
The carriage is designed in a way that the plate automatically moves
smoothly and positions itself accurately in the optical measurement path.
Readings are taken continuously. The average value is calculated and results
are presented according to the option selected.
The ELISA Reader has a special provision, which allows it to be used even when
a printer is not available. Readings can be conveniently noted down manually.

The other special features of ELISA Reader are as follows:


• Option of Lamp saving mode
• Selection of both primary and secondary filters
• Latest technology with battery backup for 250 tests, and 100 plate data
• Robust built in 52-column thermal printer with 384 stationary heads
• Unique circuitry for long lamp life
• Alphanumeric Patients ID entry
• Editing of saved tests
• Human machine user interface: Touch panel, Keypad
• Multi-standard curve up to 12 standard calibrations with one blank optional
• Access to test by touch of key
• It takes 16 secs to read the plate
• With help of colour combination, we can distinguish the calibrator, control, high and low sample
• Blank is optional
• Setting of the Date and Time
• Capable of storing, deleting and recalling tests
• Multiple calibrator modes
• Selection of duplicates for both calibrators and samples
• Extensive software for cut off mode
• Selection of Positive, Equivocal, Negative cut-off
•S
 everal pre-programmed calculation modes will help to facilitate data processing of enzyme immunoassays.
These are menu driven modes for simple and error free operations:
- Absorbance mode
- Single Standard Mode
- Cut-off mode
- Multi-standard mode
- % Absorbance mode
- Uptake
- Ratio

12
Technical Specification

Human Machine Interface Touch panel / Keypad

Linear measurement range


0.000 to 3.000 Absorbance Units
Spectral Range
Filters Type of filter: narrow band Interference
Wave Length: 405nm, 450nm, 492nm, 630nm with two optional
filters namely – XXX and YYY (editable)
Half Bandwidth: 10nm ± 2nm
Selection: automatic by Stepper Motor

Drift <0.005A/hr

Photometric Linearity 2.5A

Photometric Accuracy +/- 2% or 0.007 (0 to 1.5 A), +/- 3% from 1.5 A to 3.0 A

Optical measurement 2 channels

Light Source Tungsten halogen lamp, 20 Watts

Display 5” TFT Colour Display

Curve Plotting STN 480*800 pixels, Graphical representation on printer

Printer Built in thermal printer 52 columns

Analysis Mode Absorbance, Single Standard, Cut-off, Multi-Standard, % Absorbance, Uptake, Ratio

Memory 2 MB RAM

Plate Carrier Movement Precisely through the stepper motor

Well Type Capable of reading U,V and Flat type wells

Connectivity
9600 baud, 8 data, 1 stop, no parity bits / USB
RS232 Serial Port / USB

Size (cm) 36 x 36 x 22 (lxbxh)

Weight (Approx.) 10 Kgs

Power/Wattage/ Voltage 75W, 115-230 Volt ± 10%, 50/60 Hz

Operating Conditions/
From + 18°C to 35°C, Up to 80%
Temperature /Relative Humidity

Storage Conditions From -10°c to 40°c, up to 80%