Wireless IR Handheld Operator
Wireless IR Handheld Operator
Manual
REV 05/2015
1
Dynaflox Wireless IR Handheld Operator manual
Content
2
Dynaflox Wireless IR Handheld Operator manual
Part1 Instruction
1.1 Spefication
Size 200mm*79mm*59mm
3
Dynaflox Wireless IR Handheld Operator manual
Part2 Operation
2.1 Keyboard
□ - 9□
0 Number key;
Enter
Dynaflox
2014.06.05 12:00:00
4
Dynaflox Wireless IR Handheld Operator manual
1. Long press“power”button to turn off;
2. Long time without operation, the operator will turn off as default setting(in main menu
could set the time interval);
3. Low battery voltage.
4.1V
Function menu
【1】 System communication
【2】 system information
【3】 system setup
【4】 system testing
【5】 operational procedure
【ESC】Quit
(1)system communication
Use USB cable to connect operator with PC, enter the menu, keep in communication
with the communication tools and operator.
(2)System information
Enter the menu and inspect the machine information、file info、system time and
battery status.
System time could not be modified in the menu, if requires modification,please use
communication tools;or enter IR meter reading system to change the time.
When the operator fully charged, the battery voltage is 4.2V. If battery voltage lower
than 3.8V, meter reading system could not get started, only could read the data in the
operator; If battery voltage lower than 3.3V, operator will turn off.
(3)System setting
In the menu, user could set LCD adjustment, FLASH Formatting,change password,
and shutdown time.Others are spare menu.
(4)System test
In the menu,user could make test for keypad, LCD and files. Others are spare menu.
5
Dynaflox Wireless IR Handheld Operator manual
(5)Operating Procedure
Operate the IR meter reading system.In start menu, ,press any key(except menu
F2)could enter IR meter reading system.
3.600V
In the reading process, please align the infrared and press key[←] as it shows when
there is a pause.If you want to cancel the reading process, press key[Esc] to stop and go
back to previous page.
Reading Result is stored in .CSV and .TXT format in the handset, you can export the
handset opened with EXCEL, file names are the current S / N number .CSV and S / N
number .TXT. For example, the current meter 8
S / N number is 70000012, the copy read
cumulative months after a successful flow generated file called 70000012.CSV; and copy
successful reading day cumulative flow generated after file name 70000012.TXT.
Dynaflox Wireless IR Handheld Operator manual
The function is querying meter’s latest Accumulative or Instantaneous Flow record which has
been saved..
Note: The current cumulative flow can store up to 499 records, which store records stored in a
file format in Flow.CSV. The current instantaneous flow can store up to 99 table records,
content is stored in the handset memory (after exiting power down the system or the content is
missing). If a table's current instantaneous flow is repeatedly stored, the previous records
stored once covered by the new query to the latest in a record store.
Adr digit number indicates the current meter settings (ie tag NO.), Can support up to eight
characters or 16 numbers, letters show; If this table is not set bit number is displayed S / N:
xxxx.
Parameters Setting
===============
1.User Setting->
2.Factory Setting->
3.Pulse Setting->
4.Current Setting->
5.DAC Debugging->
6.Factory Reset
7.ACC Sample Set->
9
Dynaflox Wireless IR Handheld Operator manual
MBUS Address Setting MBUS Baud Rate Setting Display Units Configuration
10
Dynaflox Wireless IR Handheld Operator manual
(2) Factory Setting
Factory Setting 1) Static Zero Setting: Water’s minor fluctuations will be seen as
================ zero to ensure measurement accuracy ;
1.Static Zero Setting 2) Password Setting: Water’s password (factory setting)
2.Password Setting changes, to prevent the operator operating ;
3.Correct Coefficient 3) Correct Coefficient: Test & Calibration(Not open to users);
4.Accumulation Set-> 4) Accumulation Set: Test & Calibration(Not open to users);
5.Synchronize Time 5) Synchronize Time: Set same time of hand-held operator and meter
6.Reverse Flow Enable 6) Reverse Flow Enable: Set in the direction of flow;
7.Low Flow Set 7) Low Flow Set: Test & Calibration(Not open to users);
8.Reverse Low Flow Set 8) Reverse Low Flow Set: Test & Calibration(Not open to users).
If has been Set Static Zero Success, then choose the option again it will shows the screen below:
Press[1] Set Static Zero again.
Press[2] Recover the parameters before Setting, and it shows the
Static Zero Setting returning screen below:
================
Static Zero Recover Static Zero Recover
1.Static Zero Setting ================ =================
2.Static Zero Recover
Enter Password:
xxxx Recover Success!
Press[Esc]Go Back
2. Password Modify
Dynaflox IrDA System Dynaflox IrDA System Dynaflox IrDA System Dynaflox IrDA System
================ ================ ================ ================
Positive Flow(m3): Positive Flow(m3): Positive Flow(m3): Positive Flow(m3):
+25.0000 +25.0000 +25.0000 +10.00
Press[←] Modify Enter Password: +EnterCumulative:
xxxx 10.00 Positive Flow success!
Press[←] Set Again
Press[Esc] Go Back
5. Synchronize Time
Press [5]
Closed Opened
Enter Password:
xxxx Set Success!
Press[←] Set Again
Press[Esc] Go Back
13
Dynaflox Wireless IR Handheld Operator manual
8. Reverse Low Flow Set
Pulse setting
================= 1) Pulse Enable: If be opened, it means pulse output is enable.
1.Pulse Enable 2) Unit Configuration: A pulse flow through each occurrence,
2. Unit Configuration unit L.
Current setting
================== 1) 4-20mA Output Enable: If be opened, it means 4-20mA
1.4-20mA Output Enable output is enable.
2.4mA -Flow Rate 2) 4mA -Flow Rate: Output 4mA, the Instantaneous Flow;
3.20mA -Flow Rate 3) 20mA -Flow Rate: Output 4mA, the Instantaneous Flow;
Dynaflox IrDA System 1) Read the parameter of UL_LOOPC_Val, and also can have a
================= modification.
1.UL_LOOPC_Val 2) Read the parameter of AG_STAGE2_A2_Val, and also can
2.AG_STAGE2_A2_Val have a modification.
3.AG_STAGE2_R2_Val 3) Read the parameter of AG_STAGE2_R2_Val, and also can
have a modification.
*Contact us to get the detailed parameter.
15
Dynaflox Wireless IR Handheld Operator manual
Parameters Query
=================
*Remarks:
1.Software Version
Except software version, pipe structure and
2.MBUS Query
current temperature Query, other parameters
3.Unit Configuration
can be got in Parameter Setting. It’s convenient
4.Pulse Query
for user’s integrated Query.
5.Current Query
Among them, the software version number
6.Pipe Structure Query
can’t be changed.
7.Temperature Query
the above functions refer to the instruction of the hand-held operator screen
Format specification:
Pipe type name + “:” + “0,0;” + “data1,data2;” + “data3,data4;” + Enter[←] key, and
all punctuations in it are English punctuations, or it will not recognize this pipe.
Possible results of this operation:
1) If there is such pipe configuration data in pipe.txt file,it will display the corresponding
pipe configuration, as shown in the screen below:
16
Dynaflox Wireless IR Handheld Operator manual
2) If there is no such pipe configuration data in pipe.txt file, as shown in the screen below:
If press key [F1] now, you can add the corresponding Dynaflox IrDA system
pipe, enter the name of the pipe as format 1, then the ==================
data1, data2
pipe is automatically stored in the pipe.txt file and can be
data3, data4
queried in the future. Furthermore, the new pipe *Same Pipe Not Exist*
configuration data can also be directly modified in the Press [F1] Save Pipe
pipe.txt file. Wherein, data1, data2, data3 and data4 are Press[<-] Read Again
Press[Esc] Go Back
the parameters of the pipe configuration.
1. If there is no corresponding prompt in the interface, press [Esc] go back to the previous
menu.
2. If there is '->' symbol after the menu option indicates that there are menu selection after,
Without this flag should be aligned at this time meter infrared devices, select the
corresponding infrared copy after reading data.
3. In the key input string in the process, if a mistake you can press [Esc] to delete the
previous entry press the key [←] indicates that the input is completed; if no input directly
press the key [←]or[Esc] to no action direct return to the previous menu.
4. If you press the key [F1] to exit the system, the contents have been saved in History
Data Query -> View Saved Record -> Instantaneous Flow are emptied. After re-entering
the system, there is no contents in History Data Query-> View Saved Record -> Current
Instantaneous and need records stored in order to re-query.
17
Dynaflox Wireless IR Handheld Operator manual
communication tools。
correct COM port, and then click on the "Connect POS machine" connections. Once
connected, the software most lower-left corner displays the connection status, such
as“ ”。
18
Dynaflox Wireless IR Handheld Operator manual
4. Once connected, the software appears to the right handset file. As shown below:
3
(3)
(4)
5
20
Dynaflox Wireless IR Handheld Operator manual
[1] 上传手持机里文件到 PC 机
1.点击 连接到 POS 机(若链接成功左下角处显示手持机链接在 COM x)
2.点击 文件管理
3.选择 数据文件 存储位置所在
4.选择 要导出的文件
5.点击 上传到 PC 机 按钮(存储位置的文件架中不能有重名文件,否则上传不成功)
[2] 下载 PC 机里文件到手持机
(1) Click Connect to POS (if successfully connected, connect handheld to COMx will be
(1) Click Connect to POS (if successfully connected, connect handheld to COMx will be
Dynaflox follows the enterprise culture of "Respect, Harmony, Innovation and Integrity",
21
Dynaflox Wireless IR Handheld Operator manual
1. From the date of delivery, meters enjoy one year free warranty.
2. If user demands, the company will appoint professional technician to help users solve
3. Dynaflox will not be responsible for and will have no obligation under the Warranty to
3) The product is opened, modified or repaired by any person other than a technician from
4) Damages caused by failing to follow the instructions of the manual to install, use or
5) For the above reasons the maintenance expenses should be borne by users.
* The Company reserves the right to modify the information of the products. It is subject to change without notice
22