Series 9000

Total Hydrocarbon Analyzer

Analyzer
The Baseline® Series 9000 is the candidate of choice
whenever accurate, reliable total hydrocarbon analysis is
required. The Series 9000 analyzer provides unparalleled
flexibility and offers continuous, fully automated gas
analysis over a broad range of concentrations.
Features
The 9000 has a dynamic range from less than 10 ppb on the
low range version and up to 100% as methane on the high • Flame ionization detector.
range version. The analyzer has a generous complement of • 3U case, 19” (48.3cm) rack mountable.
analog, digital and logic output capabilities. • Automatic calibration at user-defined
intervals.
The analyzer is based on a flame ionization detector (FID)
with FlowGuard electronic control that delivers a small • Internal, multipoint sampling option.
portion of the sample gas to the detector flame. During the • FlowGuard electronic control of fuel, air
combustion process, organic or hydrocarbon-based gases and sample.
in the sample are ionized then they are detected by the • Electronic back-pressure regulator with
instrument and reported as a concentration. sample bypass system ensures fast re-
sponse.
The Series 9000 can be configured with internal components
• Discrete, multilevel concentration & fault
for single or multipoint analysis of non-condensing gas
alarms.
samples. The automatic calibration feature enhances the
long-term analytical stability of the instrument. These • Analog output ranges programmable
features place the instrument well ahead of the competition from 1ppm – 100% full scale.
in performance, automation, and configurability. • Programmable relays for concentration
alarms, events and diagnostics.
Applications • Automatic FID ignition, with automatic
shut-off of sample, fuel and combustion
The Series 9000 is designed to continuously monitor the air.
total hydrocarbon content of non-condensing gases in a • Remote operation via RS-232 and Ether-
variety of applications, such as: net.
• Scrubber & oxidizer efficiency • Graphical display and easy to use menu
• Carbon bed breakthrough detection system.
• Contaminant analysis in pure/ultrapure inert gases
• Well logging
• Industrial hygiene & safety monitoring
• Fenceline (perimeter) monitoring around industrial sites
• LEL monitoring
www.Baseline-MOCON.com
Series 9000
Total Hydrocarbon Analyzer

Analytical Physical
Detector: Flame ionization detector (FID). Dimensions: 19.00” (48.3cm) W x 14.25” (36.2cm) D x
5.25” (13.3cm) H.
Ranges:
Weight: < 20 lb (9.07kg).
MDQ 0.01ppm, Full-scale 1 to 200ppm (Methane)
MDQ 0.1ppm, Full-scale 1 to 2,000 (Methane) Configuration: Bench-top or rack-mount 19” (48.3cm)
MDQ 0.3ppm, Full-scale 1 – 20,000ppm (Methane) panel.
MDQ 0.003%, Full-scale 0.01 – 100% (Methane) Connections: 1/4” (6.35mm) tube fitting connectors.
Repeatability: +/- 1% Full-scale response. Operating Temperature: 32 - 104 °F (0 - 40 °C).
Drift: +/- 1% of full-scale over 24 hours. Operating Humidity: 0 – 95% (non-condensing).
Response Time: < 5 seconds to 90% of final reading.
Electrical
Alarms: Multilevel concentration and fault alarms that
result in audible and visual alarms. Alarms may also be Power: 90 – 230 VAC, 50/60Hz, 3A.
mapped to relays to control external equipment. Display: 3.4” (8.6cm) x 4.5” (11.4cm) graphical display.
Sampling: Internal single or multipoint modules, for Relay Outputs: 5 programmable form A relays rated
pre-filtered (<0.1 microns) non-condensing samples, to 3A @ 230V AC (optional additional 9 relay output
with or without sample pump. board).

Calibration: Programmable automatic or manual Analog Outputs: 1 programmable 0-20mA or 4-20mA


calibration. isolated output (optional additional 3 analog output
Support Gas: Hydrogen 35 cc/min, Air 175 cc/min board).
(typical). Hydrocarbon content must be less than 1
Digital Outputs: RS-232, Ethernet.
ppm. Fuel blend options available, consult Baseline®.
Imputs: Optional digital input board for 3 contact
closure inputs. Supports start analysis, start calibration,
and analyze calibration gas as sample.
D011.1