DIGITAL MULTIMETER
WARNING
READ AND UNDERSTAND THIS MANUAL BEFORE
USING THE INSTRUMENT
Failure to understand and comply with the
WARNINGS and operating instructions can result in
serious or fatal injuries and/or property damage.
WARNING
1. To avoid electrical shock hazard and/or damage of the
instrument. Do not measure voltages that might exceed
500V above earth ground.
2. Before the use of instrument, inspect test leads,
connectors and probes for cracks, breaks, or crazes in the
insulation.
DC VOLTAGE MEASUREMENT
1. Connect red lead to "V Ω mA" jack. Black lead to "COM"
jack.
2. Set RANGE swich to desired DCV position. If the voltage
to be measured is unknown, set switch to the highest range
and reduce it until satisfactory reading is obtained.
3. Connect test leads to device or circuit being measured.
4. Turn on power of the device or circuit being measured,
voltage value will appear on Digital Display along with the
voltage polarity.
AC VOLTAGE MEASUREMENT
1. Connect red lead to "V Ω mA" jack. Black lead to "COM"
jack.
2. Set RANGE switch to desired ACV position.
3. Connect test leads to device or circuit being tested.
4. Read voltage value on Digital Display.
DC CURRENT MEASUREMENT
1. Connect red lead to "V Ω mA" jack. Black lead to "COM"
jack. (For measurements between 200mA and 10A, connect
red lead to "10A" jack.)
2. Set RANGE switch to desired DCA position.
3. Open the circuit to be measured and connect test leads
IN SERIES with the load in which current is to be measured.
4. Read current value on Digital Display.
RESISTANCE MEASUREMENT
1. Connect red lead to "V Ω mA" jack. Black lead to "COM"
jack.
2. Set RANGE switch to desired Ω position.
3. If the resistance being measured is connected to a circuit,
turn off power and discharge all capacitors before
measurement.
4. Connect test leads to circuit being measured.
5. Read resistance value on Digital Display.
DIODE MEASUREMENT
1. Connect red lead to "V Ω mA" jack. Black lead to "COM"
jack.
2. Set RANGE switch to position.
3. Connect the red test lead to the anode of the diode to be
measured and the black test lead to the cathode.
4. The forward voltage drop in mV will be displayed. If the
diode is reversed, figure "1" will be shown.
TRANSISTOR hFE MEASUREMENT
1. Set RANGE switch to hFE position.
2. Determine whether the transistor is NPN or PNP type and
locate the Emitter, Base and Collector leads. Insert the leads
into the proper holes of the hFE Socket on the front panel.
3. The meter will display the approximate hFE value at the
condition of base current 10 µ A and Vce 2.8V.
DC VOLTAGE
RANGE RESOLUTION ACCURACY
200mV 100µV ± 1.0% of rdg ± 2D
2000mV 1mV ± 1.2% of rdg ± 2D
20V 10mV ± 1.2% of rdg ± 2D
200V 100mV ± 1.2% of rdg ± 2D
1000V 1V ± 1.5% of rdg ± 2D
OVERLOAD PROTECTION: 250V rms AC for 200mV range
and 1000V DC or 750Vrms AC for other ranges
AC VOLTAGE
RANGE RESOLUTION ACCURACY
200V 100mV ± 2.0% of rdg ± 10D
750V 1V ± 2.0% of rdg ± 10D
OVERLOAD PROTECTION: 1000V DC or 750Vrms AC for all
ranges
RESPONSE: average responding, calibrated in rms of a sine
wave
FREQUENCY RANGE: 40Hz - 400Hz
DC CURRENT
RANGE RESOLUTION ACCURACY
200µA 100nA ± 1.5% of rdg ± 2D
2000µA 1µA ± 1.5% of rdg ± 2D
20mA 10µA ± 1.5% of rdg ± 2D
200mA 100µA ± 2.0% of rdg ± 2D
10A 10mA ± 3.0% of rdg ± 2D
OVERLOAD PROTECTION: 500mA 250V fuse (10A range
unfused)
MEASURING VOLTAGE DROP: 200mV
RESISTANCE
RANGE RESOLUTION ACCURACY
200Ω 100mΩ ± 1.2% of rdg ± 2D
2000Ω 1Ω ± 1.2% of rdg ± 2D
20kΩ 10Ω ± 1.2% of rdg ± 2D
200kΩ 100Ω ± 1.2% of rdg ± 2D
2000kΩ 1kΩ ± 1.5% of rdg ± 2D
MAXIMUM OPEN CIRCUIT VOLTAGE: 2.8V
OVERLOAD PROTECTION: 15 seconds maximum 250Vrms
AC or DC 250V on all ranges