100% found this document useful (1 vote)
79 views24 pages

Alibre 1310 29 Q Scs Cald Corr Corh Cors Sts 8 Jewels Folder 1 - Replacement of Power Cell

The document provides instructions for servicing a Calibre 1310 watch movement. It contains 3 folders that describe: 1) Replacing the power cell, including cleaning contacts and replacing damaged parts. 2) Adjusting the watch rate using a trimmer and DELTATEST to speed up or slow down the movement. 3) Standard procedures for exchanging the entire movement or modules, including checks for power consumption and hand positioning. Precautions are given to avoid damaging delicate electronic components.

Uploaded by

paulisac77
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
79 views24 pages

Alibre 1310 29 Q Scs Cald Corr Corh Cors Sts 8 Jewels Folder 1 - Replacement of Power Cell

The document provides instructions for servicing a Calibre 1310 watch movement. It contains 3 folders that describe: 1) Replacing the power cell, including cleaning contacts and replacing damaged parts. 2) Adjusting the watch rate using a trimmer and DELTATEST to speed up or slow down the movement. 3) Standard procedures for exchanging the entire movement or modules, including checks for power consumption and hand positioning. Precautions are given to avoid damaging delicate electronic components.

Uploaded by

paulisac77
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 24

Technical guide

CALIBRE
1310
29 Q SCS CALD CORR CORH CORS STS 8 jewels
Folder 1 - Replacement of power cell

o/ 29.00 mm

Movement height 6.35 mm

Jewel number 8
Frequency 32'768 A/h

IMPORTANT
To be observed strictly

The watch must in no case be demagnetized.

In order not to disturb the proper functioning of


the watch the electronic module (fig. 1.1) must not
be touched with the fingers.

The new power cell should be manipulated by


means of tweezers with insulated tips (fig. 2.1).

EQUIPMENT Electronic module fig 1.1

Insulated tweezers.

fig 2.1

English
Cal. 1310 1973 1
REPLACEMENT OF POWER CELL

1.1. Open the case.

1.2. Remove and clean the back.

1.3. Remove magnetic protection.

1.4. Remove power cell bridle No. 1310.9179 by


unscrewing its screw.

1.5. Remove by suction any possible metal waste


caused by the screw of the power cell bridle.

1.6. Remove faulty power cell by turning the


watch over.

1.7. Check cleanliness of the two power cell


contacts. If necessary, clean them. If the contacts
are oxidized, it is necessary in the case of :

the power cell bridle (positive pole): to replace the


bridle No 1310.9179. Fig. 3.1

the lower power cell contact (negative pole): to


clean carefully the bridle by means of a buff. COMPONENTS

1.8. Check new power cell. Even if minute No 1310.9179. Power cell bridle, with screw
leakages are apparent, it should not be used. Place fitted.
power cell No 9903 in position, positive pole (+)
uppermost (fig. 3.1).
No 9903. Power cell, tension 1.35 V.
1.9. Refit power cell bridle and screw thoroughly.

1.10. Fit magnetic protection.

1.11. Grease water-resistance gasket. Close the


case.

Cal. 1310 1973 2


Technical guide

CALIBRE
1310
29 Q SCS CALD CORR CORH CORS STS 8 jewels
Folder 2 - Adjustment of the rate

IMPORTANT
To be observed strictly

The watch must in no case be demagnetized.

In order not to disturb the proper functioning of


the watch, the electronic module (fig. 1.2) must
not be touched either with a tool - except key for
trimmer - or with the fingers.

EQUIPMENT

Key for trimmer


DELTATEST

Electronic module fig. 1.2

English
Cal. 1310 1973 3
ADJUSTMENT OF THE RATE Note

2.1. Open the case. Adjustment can only be effected if the daily rate
is lower than the possibilities offered by the
2.2. Remove and clean the back. trimmer.

2.3. Place watch on captor, movement side The guide-mark printed on the trimmer should
uppermost, and press button "l Hz" of not be moved outside the limits indicated by fig.
DELTATEST. 2.2. The total correction area allows a rate
adjustment of 10 seconds.
Note
If the daily rate is higher than the possible
Generally, a measuring time of 10 seconds and correction limits, the electronic module should
a measuring accuracy of 1/100th second will be be replaced (see folder 3-1310).
used.

2.4. Read instant rate shown on DELTATEST.

Fig. 2.2

Cal. 1310 1973 4


2.5. Where correction of the daily rate is possible,
adjust said rate by means of trimmer, using the key
specially devised for this purpose (fig. 3.2). This key
only must be used in order not to damage the
trimmer. Excess pressure should be avoided during fig. 3.2
this operation.

By moving the guide-mark of the trimmer in a


clockwise direction, a gain is obtained; by moving it
in the opposite direction, a loss.

Adjust the watch in relation to the ambient


temperature, according to fig. 4.2.

In order that the temperature of the watch may be


equal to the ambient temperature, it is necessary to
wait about one hour before the measuring is effected.

The values indicated in fig. 4.2. are valid for


a watch with closed back.

Turn the trimmer, bearing in mind a gain of 0.10 to


0.20 sec. as compared with the desired adjustment
according to fig. 4.2, in order to compensate the loss
caused by closing of the case.

2.6. Grease the water-resistance gasket. Close the


case.

2.7. Set the watch and check frequency on


DELTATEST after 24 hours.

fig. 4.2

Cal. 1310 1973 5


Technical guide

CALIBRE
1310
29 Q SCS CALD CORR CORH CORS STS 8 jewels
Folder 3 - Standard exchange of movement or modules

IMPORTANT
To be observed strictly

The watch must in no case be demagnetized.

Do not touch the electronic module with the


fingers, nor unsolder the elements.

The motor module must IN NO CASE be


disassembled; the characteristics of the
magnetic circuit would be disturbed.

Do not clean the electronic module and motor


module in bath:.

EQUIPMENT

Movement holder 1310.

Key for trimmer 1301.

Tool for adjusting motor eccentric.

Deltatest measuring apparatus.

Alitest checking apparatus.

Feed by substitute power cell 1310.

Microscope.

English
Cal. 1310 1973 6
EXCHANGE OF MOVEMENT

3.1. Open the case. 3.10. Checking of the second hand jump; this
latter hand being positioned on an even
3.2. Remove upper magnetic protection No. number, push the hand-setting stem; 1
1310.9230 by turning the watch over. second after this function the hand should
make its first jump. If this is not so, take off
3.3. Remove hand-setting stem No. the second hand and move it by 1 second
1310.9030 by pressing the head of the before refitting.
setting lever staff (red plastic) No.
1310.9044. 3.11. Proceed with casing-up of movement.
3.4. Remove the two fixing bridle screws 3.12. Set up feed by substitute power cell and
No. 2533 and the bridles. check by means of Alitest that the watch
functions between 1.15 V and 1.40 V.
3.5. Extract movement from case.
3.13. Measure current. Consumption should
3.6. Take off the three hands. not exceed 15 µA under a tension of 1.35 V.
Important : Avoid moving the second hand Should one of the conditions mentioned not
backwards so as not to damage the retaining apply, then the movement must be returned
click. to the Omega agent.
3.7. Remove simultaneously the dial, lower 3.14. Place power cell in position (see folder
magnetic protection No. 1310.9225 and dial 1-1310).
rest (green plastic) by lifting the whole with
the aid of a screw-driver inserted near the 3.15. Fit magnetic protection.
dial feet in the two main plate countersinks.
3.16. Proceed with adjustment of the rate
3.8. Fit a power cell on the new movement (see folder 2-1310).
(see folder 1-1310)

3.9. Pull the hand-setting stem and place


dial and hands in position; turn hour and
minute hands backwards in order to
compensate the backlash before fitting the
second hand. This latter hand will be fitted
on an even number index.

Cal. 1310 1973 7


EXCHANGE OF ELECTRONIC MODULE

3.17. Open the case.

3.18. Remove upper magnetic protection No.


1310.9230 by turning the watch over.

3.19. Remove power cell bridle and power cell


(point 1, fig. 1.3.)

3.20. Unscrew the 2 motor connecting screws No.


2377 (2), earth screw No. 2589 (3), electronic
module screw No. 2589 (4) and electronic
module screw No. 2377 (5).

3.21. Extract faulty electronic module No.


1310.9100.

3.22. Place new electronic module in position.


Important : For this operation, the hand-setting
stem must be pushed in.
Fig. 1.3
3.23. Tighten the 5 screws according to op. 3.20
and fig. 1.3.

3.24. Follow operations in the order 3.12 to 3.16.

EXCHANGE OF MOTOR MODULE

3.25. Follow operations in the order 3.1 to 3.5.

3.26. Release retaining click No. 1310.9028


(point 6, fig. 2.3.) by slightly turning the click
stud (6a).

3.27. Remove the second hand.

3.28. Unscrew the wheel train bridge screw No.


2377 (7) and withdraw the wheel train bridge No.
1310.9001.

3.29. Remove second wheel No. 1310.9010 and


third wheel No. 1310.9014.
Fig. 2.3

Cal. 1310 1973 8


3.30.
Unscrew the 2 motor connecting screws (2),
motor module screw No. 2377 (8)
and motor eccentric locking screw No. 2582
(9).

3.31.
Extract faulty motor module No. 1310.9200.

3.32.
Place new motor module in position.

3.33.
Important : Screw very slightly the 2 motor
connecting screws (2) and motor module
screw (8).

3.34. Fig. 3.3


Fit simultaneously the third and second
wheels.

3.35.
Fit wheel train bridge and tighten its screw.

3.36.
Adjustment of motor-second wheel backlash
by motor eccentric No. 1310.9227. Turn wheel-train bridge direction
motor eccentric clockwise until wheel train is Fig. 3.3. (B)
stopped, using the tool specially devised for
this purpose (fig. 3.3.). If locking of the motor
is impossible, the eccentric should be left in
its position of maximum eccentricity (fig. 3.3
(B). The following operation 3.37 will not be
necessary.

3.37.
Turn eccentric slowly in the opposite
direction until the wheel train
turns normally, i.e. always in the same
direction (the wheel train is free, without Important : If a faulty operation has occurred
backlash). in adjusting the backlash, it will be necessary
to recommence the entire adjustment as from
Important : Check rotation of wheel train operation 3.36, and to follow the indications
curing one minute at least. scrupulously.

3.38. 3.39.
The backlash between motor and second Tighten the motor eccentric locking-screw.
wheel is then obtained by turning the motor
eccentric through an angle of 40°. To secure 3.40.
this angle accurately, place tool on eccentric, Lock motor module screw and the 2 motor
hold with one hand the upper part of the connecting screws.
instrument, and with the other hand turn
handle arrow-wise (fig. 3.3.) as far as tool 3.41.
banking. Adjustment of click by microscope : This
For the movements which have not been operation is carried out in "stop second"
stopped by operation 3.36, check that the position of the hand setting stem. Before
wheeltrain is always turning in the same adjustment, turn the hand setting stem arrow-
direction, then effect the 40° recoil. wise in order to compensate the backlash.

Cal. 1310 1973 9


3.42. Slightly unscrew the screw No. 2564
(point 11, fig. 4.3.) for retaining click plate.

3.43. Engage click (6) on retaining wheel


(12). The tip of the blade must be at the top
of a tooth, as close as possible, without
touching it (position a, fig. 5.3.). If necessary,
move retaining click plate No. 1310.9026
(13) by turning the eccentric of click bearing
plate No. 1310.9029 (14).

3.44. Lock screw of retaining click plate.


Check again the click position after having
turned the stem according to fig. 5.3.

3.45. Start the watch by pushing hand setting


stem. Engage click in retaining wheel by
turning the click stud until the blade touches
the tooth (c). Turn slightly further the click
stud in order to tighten the blade (position b), Fig. 4.3
and so that a space (d) appears between the
end of the click and the tooth (e).

Important : The distance (d) between the


end of the blade and the tooth (e) must be as
small as possible. It is necessary to check
during one minute at least the passage of the
60 teeth of the retaining wheel. The click
must never touch the tooth at point (e).

3.46. Fit second hand according


to op. 3.9. and 3.10.

3.47. Follow operation in the order 3.11 to


3.16.

Fig. 5.3

Cal. 1310 1973 10


Technical guide

CALIBRE
1310
29 Q SCS CALD CORR CORH CORS STS 8 jewels
Folder 4 - Diagnostic Disassembling Cleaning
Reassembling and lubrication of movement

IMPORTANT - To be observed strictly

The watch must in no case be demagnetized.

Do not touch the electronic module with the


fingers, nor unsolder the elements.

The motor module must IN NO CASE be


disassembled; the characteristics of the
magnetic circuit would be disturbed.

Do not clean the electronic module and motor


module in baths.

EQUIPMENT

Movement holder 1310.

Key for trimmer 1301.

Tool for adjusting motor eccentric.

Deltatest measuring apparatus.

Alitest checking apparatus.

Feed by substitute power cell 1310.

Microscope.

English
Cal. 1310 1973 11
Cal. 1310 1973 12
Cal. 1310 1973 13
Cal. 1310 1973 14
CLEANING INSTRUCTIONS

4.4.1.0.
Motor module 1310.9200.

Do not disassemble or clean motor module in


baths.
Standard exchange if necessary.

4.5.1.0.
Electronic module 1310.9100.

Do not disassemble or clean electronic module in


baths.
Standard exchange if necessary.

4.6.10.0.
Date indicator driving wheel 1310.9062.
Date Maltese cress 1310.9064.

The cleaning machine is not recommended for


these magnetic components. They should be
cleaned in the appropriate benzine bath which
must be absolutely clean and free from any
metallic particles. Check, however, the cleanliness
of the magnets.

4.6.13.0.
Minute wheel 1310.9020.

Important : NEVER clean minute wheel in the


cleaning machine. Proceed as indicated under
4.6.10.0.

Furthermore, the following should not be cleaned


in the baths
Day disc 1310.9967.
Date indicator 1310.9066.
Power cell 9903.
Upper magnetic protection 1310.9230.

The remainder of the movement, including the


motor eccentric 1310.9227 left on the plate

(only remove it in case of replacement) may be


cleaned in baths as per the usual procedure.

Cal. 1310 1973 15


Cal. 1310 1973 16
Cal. 1310 1973 17
Cal. 1310 1973 18
4.9.5.0.

Place one of the three fingers of the day disc


driving gear No. 1310.9057 opposite the post of
the date indicator driving wheel (magnetic mobiles
Nos. 1310.9062 and 1310.9064). Fit the two
magnetic mobiles Nos. 1310.9062 and 1310.9064.

Check on positioning : make the day disc driving


gear No. 1310. 9057 turn slightly until the three
fingers are visible; one of the three magnetic teeth
must then be located under the upper pinion of the
day disc driving gear No. 1310.9057.

4.9.6.0.

Fit the date driving gear No. 1310.9053.

Check on positioning : make the date driving


gear No. 1310.9053 turn until the finger drives the
two mobiles previously fitted. When the driving of
these two mobiles is accomplished, one of the
three magnetic teeth must again be located under
the upper pinion of the mobile No. 1310.9057.

Cal. 1310 1973 19


Technical guide

CALIBRE
1310
29 Q SCS CALD CORR CORH CORS STS 8 jewels
Folder 5 - 6 Application of Deltatest and Alitest

Note :

The numbers in parentheses ( ) refer to the


numbers appearing in the instructions for use
of the Deltatest ODT 3 and Alitest OAT 2
apparatus.

Indications in brackets [ ] refer to the


Deltatest ODT 1 and Alitest OAT 1.

All apparatus which do not bear the model


number on the back plate correspond with the
ODT 1 and OAT 1.

English
Cal. 1310 1973 20
5.1 Measuring of the rate (Deltatest) 6.4 Analysis of motor - measuring of the coil
resistance
Press key (2) of keyboard " < 15' Hz" ["1 Hz ".]
Note:
Move the watch in order to secure maximum
signal. If the movement is under tension, it is necessary to
isolate the electronic module before measuring the
Regulate amplification (6) by observing the level resistance.
control provided by the instrument hand (7).
Optimal adjustment is secured when the hand is at Unscrew the two motor connecting-screws.
the beginning of the black zone [green zone]
Slide an insulating leaf under the terminals of the
Select by means of (3) the measuring time of 12 module (A and B).
[10] seconds.
Measure R by applying the probes (16) to the
Select by means of (4) the measuring accuracy of terminals A and B.
.99 [1/100.]
The resistance of the motor
6.1 Measuring of the power cell tension (Alitest)
should be between 2 and 2.2 K. If this is not the
With power cell placed on table, connect black case, the motor module must be changed.
probe to top of power cell (-) and red probe to its
case (+). 6.5 Measuring of the rate on Deltatest using
feed from Alitest
The tension value must be in the region of 1.35V;
if it is below 1.25V, the power cell must be Connect Alitest to "Input"
replaced.
of Deltatest by means of the special cable.
6.2 Measuring of the current (Alitest)
Proceed in identical manner to 6.2 : Measuring of
the current.
This measuring is effected by using substitute
power cell feed (15) connected with "Output" by
means of the feeding clip (13). Press button Atten. Deltatest (12) [Switch on back
of Alitest in position 1].
Press buttons (2) [1.35 V] relating to feed tension
of the watch and (8) [µA] enabling the current to
be measured.

The value measured should not exceed 15 µA.

6.3 Checking of the functioning limit tensions


(Alitest)

This checking is effected by using substitute


power cell feed (15) connected with "Output" by
means of the feeding cable (13).

Press buttons (4) [Uadj] and (9) [Vadj].

Check proper functioning of movement within the


tension limits of 1.15 V and 1.40 V (tensions are
regulated by means of knob (5)).

Note

The motor is said to function normally when the


second wheel makes one complete revolution in
one minute. It is important to check the passage of
each of the 60 teeth.

Cal. 1310 1973 21


Technical guide

CALIBRE
1310
29 Q SCS CALD CORR CORH CORS STS 8 jewels
Technical information No. 8

CALIBER 1310

SUPPLEMENT TO DIAGNOSIS AND


ANALYSIS ACCOMPANYING THE
TECHNICAL GUIDE

Following up requests by our General Agents, we


are presenting a refined check-method to make
servicing of
caliber 1310 easier. The contents of the technical
guide, folder 0 to 6, remain valid.

1. CHECK DELTATEST

Accepted rate : + 0.10 at + 0.30 s./d.


Regulation beyond the range of the Trimmer :
replace electronic module.

2. BATTERY : 1,35 V

Out of tolerance . replace battery SSIH 9903


(or Vartachron 508)
(Check cleanliness of battery contact points)

English
Cal. 1310 1973 22
3. MOTOR COIL RESISTANCE : approximately 7. WHEEL TRAIN AND CALENDAR
2,2 k CHECKING
Motor module removed. Make wheels spin by means
Out of tolerance : exchange motor module (1310-9200) of a dust blower - they must spin freely and make the
date change with only small force from dust blower.
4. CONSUMPTION OF WHOLE WATCH
8. CHECK OF PARTS : CONDITION,
CLEANLINESS
Limits : maximum 15 µA with bi-polar
circuit (straight connections)
Change defective parts, clean movement if necessary
(see technical guide 4.4.1.0)
maximum 7.5 µA C-Mos circuit (T-shaped
connections)
9. ADJUSTEMENT
Above accepted limit : slide insulator blade between
motorcontacts, detect and replace definitive module. To insure a perfect backlash between the second
wheel and the motor worm gear, proceed with
(Limit for motor only 3 µA) operations 3.36 to 3.38 of the technical guide 3-1310
and check thereafter the following procedure .
5. FUNCTIONING LIMITS (WHOLE WATCH)

Lower limit : 1.20 V upper limit : 1.40 V

Beyond accepted limits :

Check

a) position and tension of retaining click


b) backlash of motor screw and second wheel (see pt. 9)
c) endshake and freedom of rotor
d) cleanliness of motor worm-gear

6. FUNCTIONING LIMITS MOTOR MODULE


WITHOUT WHEEL TRAIN

Lower limit 1.15 V upper limit : 1.45 V


Beyond accepted limits : exchange motor module
(1310-9200)

Cal. 1310 1973 23


9.1. Adjust the retaining click into position shown
by figure 1, the end of the click to be used as
guiding-mark to observe the movement of the
second wheel.

9.2. Put hand setting stem into setting position and


turn it slightly in both directions to observe the
movement of the second wheel in comparison
with the click; to be suitable, this movement must
be 1/4 to 1/2 of the distance between two teeth
(figure 2). Repeat this control twice after 20
seconds
(= 120°).

9.3. If the backlash has to be modified, turn the


eccentric for motor in the direction suitable to the
required correction.

IMPORTANT

Whilst moving the crown, a displacement of the


rotor might be observed by microscope; this
deplacement increases the movement of the
second wheel and therefore, falsifies the
measuring. The first movement of the wheel must
be considered and not the movement caused by
addition of the endshake of the rotor.

Cal. 1310 1973 24

You might also like