This wiki will describe both the settings available to the end user and the ones only accessible by inserting a
password, known only by our dealers, that will enable some features normally hidden from the end users.
The password is: CrE2u=eP
To login as a dealer, launch the software, click on the OEM button in the welcome page and insert the password.
This mode allows to change some advanced options and use the toolbox.
Index
User Manual
1. EC marking and Conformity declaration
2. Introduction
1. Presentation of the machine
2. User's Precautionary measures
3. Transportation and Storage
4. Machine Specification
5. Operation
6. Emergency Situations
7. Decommissioning
8. Residual Risk Relating to the machine
ScanWay 1.0
1. System Set Up
1. PC Minimum Requirements
2. Before Installing
3. Master License Dongle
4. Software Installation
5. Software Configuration
6. Manual connection to Exocad
2. Options Window
1. General Options
2. CAD Options
3. Service Mode
1. Scanner's Calibration
4. External Software Integrations
1. Dental Database and Orthodontic Cad Integration (VDDS)
2. Onyx Ceph Integration
Exocad
1. Updating the Exocad License and Adding Modules
2. Exoscan
3. Importing Intraoral Scan Data
1. iTero Importer
2. 3M Importer
3. Trios Importer
Neway 2
EC CE
Position of the label on the machine The product label attesting the Ec marking and the serial number of the product
is on the back of the scanner, near the back switch. Conformity Declaration All of our scanners are EC certified.
Intro 3
Intro
This manual refers to the 3D scanner and addresses the users, who, before performing any operation, use or
maintenance, must read and acquaint themselves with the material in the following pages in order to use the
machine correctly, avoiding the improper uses listed later on, before turning the machine on make sure the user is
suitable for the job.
The manufacturer refuses any responsibility for consequences deriving from:
• Use or installation not contemplated in this manual
• Disassembly, modifications or replacement of original parts or components with parts or components of different
origin without written agreements with the manufacturer
• Unsuitable power supply
• Serious inadequacy of the required maintenance
This manual reflects the technical standing at the time the machine was manufactured and it should not be
considered inadequate because based on new experiences it can be updated. The features and design of this machine
may be modified without any notice, for a constant improvement of the product.
Open Technologies S.r.l. reserve all their rights on this manual, no whole or partial reproduction is allowed
without their written authorization.
Norm References
In drafting this document, we have referred to what is reported in attachment I of Law n. 17 of 27/01/2010,
implementation of the 2006/42/EC Directive. In the planning, construction and realization of the machine a set of
technical norms have been used as reference to meet the requirements of the laws in force on the subject of safety on
the job.
Presentation 4
Presentation
NeWay is a sophisticated 3D optical scanner for the no contact acquisition of three-dimensional shapes.
The technology it uses is called active stereo vision: structured light patterns are projected on the surface of the
object and acquired by two video cameras. As the image belows shows, an object (on the right) is hit by the
projection of light patterns for 3D reconstruction.
Advanced image processing algorithms retrieve depth data of the acquired surface. This approach guarantees high
performances, flexibility and fast execution.
NeWay was specifically created to operate successfully in the dental field. The whole scanning phase is
computerized and is performed with the aid of a rotating table that allows the acquisition of different views of the
object.
Thanks to this system, you can digitize model, dies, impressions, articulators etc, very quickly.
Glossary
• 3D scanner – a machine created to acquire 3D images of opaque, solid objects;
• Bedplate – lower supporting structure, composed of screwed on steel plates;
• Rotating table – a support for the objects to be scanned located inside the equipment and having an intermittent,
rotating and tilting movement.
• Optical head – the component housing the optoelectronic components (2 cameras and a DMD projector) which
allow the acquisition of the object surface;
• Electric motor – actuator which rotates the rotating table;
• Electric panel – power device with a sectioning switch;
• Exposed person – any person within the whole or part of the danger zone;
• Operator – those people who have to install, operate, adjust, perform maintenance, clean, repair and transport
correctly the machine having received all the relative instructions and read this manual.
Technical Data
Presentation 5
Width mm 500
Depth mm 610
Height mm 540
Weight (empty) kg 18
Mains frequency Hz 50
Phases - 2P+T
Precautions
The user must: Keep all the fixed, moving protections, in their place, Avoid placing any liquids on the external
structure or nearby, Avoid moving, removing or modifying manually the rotating table; Check the electric switch
periodically, Use properly mounted sockets and do not use damaged or worn cables; Make sure the machine is
unplugged before moving it; Avoid inserting any metal objects, like clips, metal cables or flammable objects like
matches or pieces of paper inside the 3D scanner through the vents; Avoid touching the plug with wet hands; Avoid
connecting to the electric system more devices using just one socket; Avoid placing heavy objects on the electric
cable, bending it, pulling it or twisting it; Insert the plug all the way in; Connect to the scanner only grounded
devices; Put back in their place all the danger, warning and requirement signs that might come off. The manufacturer
declines any responsibility arising from accidents caused by the partial or total removal of these protection devices.
Trasport 6
Trasport
The buyer’s personnel can install the 3D scanner in the operating area after reading this manual.
As soon as the buyer receives the machine, he/she will have to check that the components have not suffered any
damages during transport, if any are noticed he will have to notify the carrier and the manufacturer immediately.
The scanner is in a protective cover made of a foam polystyrene shell placed inside a cardboard box.
When using the machine for the first time special attention must be paid to anything that might seem to have an
irregular operation and, if the case, contact Open Technologies S.r.l. immediately, reporting the anomaly.
Storage Conditions
If the 3D scanner is not installed inside the buyer’s premises but it is stored:
• outdoors - suitable covers and precautions must be arranged to protect the components of the machine from
weather conditions (rain, snow, fog, humidity etc.).Check periodically for any damages that might take place in
the course of time.
• indoors - no special precautions are required but just periodically check that the machine is stored correctly and
the environmental conditions below are respected:
TEMPERATURE
( -20 ÷ +45 )°C
HUMIDITY
( 5 ÷ 85 )%
Moving Instructions
Moving can be carried out with the help of trans pallet, trolleys or by hand.
Lifting
We advise the operator to lift the machine holding it by its opposite sides. The operator must pay attention when
lowering the machine onto a surface to avoid injures at the upper limbs.
Specs 7
Specs
Detailed Description of the Machine
The 3D scanner has a solid external structure. All the covers are integral part of the tool and must not be removed or
tampered with in any way so not to compromise the correct functioning of the apparatus.
The power connection and the power switch are on the back of the scanner. The frontal part of the scanner is the
access point to the scanning area framed by the optical head.
The optical head lodges the opto-electronic components for the acquisition of the object surface.
NeWay's optical head comprises two black-white video cameras and a DMD projector. The set up of these three
components allows one single scanning field.
Any other application not present here has not been contemplated by the manufacturer and, therefore, cannot
be performed.
Noise emission
The noise the machine produces, intended as the level of sound pressure in the workplace, is <80 dB (A).
Electrical System
See electrical diagram attached.
Operation 8
Operation
Loading
The loading operation of the objects to scan includes:
• Placing the objects inside the 3D scanner by laying them on the chosen model support, consequently attaching the
support to the magnetic joint at the base of the scanner's opening.
• Make sure the object is placed inside the scanning area of 120 x 100 mm. The scanning area is placed at a height
of 90 mm from the base of the scanning opening. The model holders and accessories are supplied to reach the
critical scanning height. Take care in using the correct support for each item.
• Scanning operations start using the management software installed on the computer connected to the 3D scanner.
Do not load the objects when the model holder is mounted on the joint, as to do so the reductor could suffer
damages from the pressure
Do not place objects on the model holders weighing more than 3kg.
The scanner does not work with transparent, light reflecting objects and with black objects.
Controls
Please read the Software Manual [1]] for the controls that the 3D scanner operational software can perform.
Stopping Measures
You stop the machine by:
• Switching off the power supply at the mains.
During maintenance operations, it is advisable, for safety purposes, to unplug the machine as well as switching it off.
• To operate the 3D scanner if the safety devices supplied with the installation are not present or working.
• To use the machine outdoors in the rain.
• To modify or tamper with the machine.
• To clean during scanning.
• To carry out ordinary maintenance without unplugging the machine.
ER 9
ER
Fire-fighting equipment
The parts of the machine, which can catch fire, are the electric components (electromotor, projector, cables, feeders).
In all these cases, the best extinguisher is carbon dioxide (CO2) or the ABC powder, keeping in mind that by using
the ABC powder you can permanently damage the electric and electronic components.
Using water is forbidden and it is only allowed when the electricity has been turned off in the whole building.
Using water to extinguish fires is forbidden and it is possible only with the electricity turned off.
Waste
For Ordinary deommisioning, the scanner does not produce any waste.
The disposal of the machine's material, for scrappage purpose, must be performed according to current legislation.
Ordinary
At the end of the work activity, the operator will:
• remove the objects from the scanning area,
• press the electric switch,
Scrapping disposal
During scrapping, separate the machine parts according to their material: steel, plastic, electric (cables, connectors)
electronic.
Scrapping operations must start with the disassembly of the plastic covers and then of the components (electromotor,
feeders, video cameras, electronic card drivers).
Risk 10
Risk
RISKS DANGEROUS DANGEROUS SITUATION DEALING WITH RISKS
AREA
Crushing Machine's base. Transport and moving. Placing the Warnings are in the instruction manual.
equipment on the work surface.
Loss of stability/ toppling over of Work Station Collaps of the surface supporting limitation instructions in the instruction manual
the machine the machine during its positioning
Electrical hazards
People coming in contact with Electric cables, Installation, adjustment, Warnings are in the instruction manual.
live elements (direct electric metallic base. calibration. Testing, maintenance,
contact). cleaning. Event of Cable breakage
People coming in contact with Metal structure of the Testing, maintenance, cleaning. Test periodically that the grounding system is
elements that become live in case apparatus. Event of Cable breakage efficient and in compliance with the regulatory
of fault (Indirect electric contact). guidelines. Test periodically the efficiency of the
safety switches.
Thermal hazards
Burns caused by flames or Internal Projector or Testing, maintenance, cleaning. Follow the indications in the instruction manual.
explosions and by heat radiation. electronic components Event of components over-heating.
Radiation hazards
Infrared rays, visible light and Internal Scanning Normal Use. Event of beamer Follow the indications in the instruction manual.
ultraviolet. Area projecting unexpected amounts of
light intensity.
Restoring power after an Electric Circuit. Accidental Starting, short circuit, Switch the machine off and on again,
interruption power interruption. unauthorized use instructions in the manual
External influences on the electric Electric Circuit. Environmental conditions limitation instructions in the instruction manual
devices
Pc Min Requirements 11
Pc Min Requirements
Recommended Systems
The Computers connected to our scanners have to meet specific requirements.
In order to simplify the choice, we recommend this reference system:
1. Dell XPS (with Intel i7)
Minimum Requirements
Software
These are the minimum requirements our software needs:
Component Requirement
Scanner Connection
In order to work properly, the scanner needs to be correctly connected to the PC. Make sure to check here-below the
necessary connections for your scanning system.
NeWay scanning system can be connected to a Computer through only 2 cables:
• 1 USB 3.0 Port that has to be directly connected to the Computer's motherboard. Only Intel C220 or Better and
Rensas USB 3.0 are supported. Other USB chipsets could not properly work.
• 1 HDMI Port
Before SW1 12
Before SW1
Before Installing
The scanner's set up procedure is very straightforward. This paragraph describes the operations that need to be
carried out before installing the software.
They are enlisted in an ideal order and, even though it is not strictly necessary to follow such order, it is advisable to
do so:
1. Check the Graphic Card's (NVIDIA) drivers
2. Check the Windows Updates
3. Connect the Scanner
4. Configure the screen resolution.
Here it is possible to gather all the information one needs to know about the NVIDIA graphic card currently mounted
in the computer: the version of the drivers, here 347.52, and the product name, here GeForce GTX 980M.
WARNING! Not all NVIDIA Graphic Cards are supported. We have very specific Computer requirements that have
to be respected
for the software to function.
[1]
The list of supported Graphic Cards can be found in the Final User's Wiki at the page [PC Minimum Requirements ]”.
Before SW1 13
There are two ways one can update the NVIDIA drivers: the first is to download them directly from the NVIDIA
website and the other is to use the GeForce Experience App.
Once the new drivers have been downloaded to the computer, install them and reboot the system.
If the number of updates to install is not significant, it is possible to simply click on “Change Settings” in the left list
and set the automatic updates, as in this picture:
In this way, those few updates can be automatically installed when shutting down the system.
However, if there were many updates to install, after setting the automatic updates it still would be necessary to
Before SW1 15
install them and reboot the system before installing the software.
The Usb cable must be connected directly to the Computer's Motherboard to a USB 3.0 port. The Usb ports are
usually color-coded, and generally the Usb 3.0 ports are marked in blue.
Once everything is connected it is possible to check if the computer recognised the components correctly by opening
windows Device Manager. The Device manager can be easily found using the Windows Search tool.
As displayed here-below, in the section Universal Serial Bus Controllers there is a list of all the USB devices and
image acquisition devices recognised by the Computer:
The devices labelled as NET ICube_cam are the two cameras and the Usb Serial Converter is the control board.
Before SW1 18
ScanWay
From the Software Download Area [1] can be downloaded two versions of ScanWay:
ScanWay Full Setup
The full setup is the complete installer. This file .exe should be used for the first installation of the current
version on a computer. Therefore it will be used both when the current installation is the first ever on a new
computer or when, for example, the costumer already has the version 1.0 installed and wants to upgrade the
software to 1.5.
ScanWay Only Update Setup
The only update setup is the installer we release for bug fixes. It cannot be used if there is not already a
complete version of the same software installed in the computer. For example, it is not possible to use the 1.5
only update installer to update a 1.0 version or to install the software from scratch.
In both cases, in order to install is only necessary to follow a very simple installation wizard. The full installation
will take longer since it will install all the drivers necessary to the functioning of the scanner:
Software Instal SW1 21
Exocad
Open Technologies provides an Exocad installer that automatically configures the integration between the two
software and installs the Open Technologies implant library files. Should the automatic connection go wrong, the
page Manual connection to Exocad will provide all necessary information to set the link manually.
From the Exocad Download Area [2] can be downloaded various versions of the software:
Latest update (No Demo Data)
This installer can be used to update to the lastest old versions of the software. It does not contain the Exocad
example cases, or demo data, because they would overwrite the Dental Database of the customer.
Latest update (With Demo Data)
This installer can be used for new installations on computers in which there was never installed Exocad. The
Exocad example cases, or demo data, would infact overwrite the customer's DentalDB and delete all his files.
Previous Version
This is the previous version of the Exocad software.
Exocad Demo
This is the version that needs to be installed when a green dongle needs to be used.
Demo previous Version
previous version of the Demo software to be used with the green dongle.
During installation there are two steps that need to be set with particular care:
Software Instal SW1 22