Astm D5454
Astm D5454
Astm D5454
for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
Copyright by ASTM Int'l (all rights reserved); Sun Dec 10 17:03:59 EST 2017
1
Downloaded/printed by
Universidad Industrial de Santander (Universidad Industrial de Santander) pursuant to License Agreement. No further reproductions authorized.
D5454 − 11´1
causing serious operating problems. Fuel heating value is 5.1.2.1 Pressure gages with bourdon tubes should be
reduced by water concentration. Water concentration levels are avoided as a result of water accumulation in the stagnant
therefore frequently measured in natural gas systems. A com- volume.
mon pipeline specification is 4 to 7 lb/MMSCF. This test 5.1.2.2 Sample purging is important to satisfactory response
method describes measurement of water vapor content with time. There must be a method to purge the sample line and
direct readout electronic instrumentation. sample cleanup system.
5.1.3 Electronics—Output from the sensor will be linearized
5. Apparatus for analog or digital display in desired units (usually lb/
5.1 The moisture analyzer and sampling system will have MMSCF). There must be an adjustment for calibration accu-
the following general specifications: racy available that can be used in the field if a suitable standard
5.1.1 Sampling System—Most errors involved with moisture is available. (This does not apply to instruments that assume
analysis can be eliminated with a proper sampling system. complete chemical reaction of water. Their accuracy still must
5.1.1.1 A pipeline sample should be obtained with a probe be verified as in Section 6.)
per Method D1145. The sample temperature must be main- 5.1.4 Power Supply—Analyzers for field use will have
tained 2°C (3°F) above the dewpoint of the gas to prevent rechargeable or easily replaceable batteries. (Warning—
condensation in the sample line or analyzer. Use of insulation Analyzers for use in hazardous locations because of combus-
or heat tracing is recommended at cold ambient temperatures. tible gas must be certified as meeting the appropriate require-
5.1.1.2 Analyzer sensors are very sensitive to contamina- ments.)
tion. Any contaminants injurious to the sensor must be re-
moved from the sample stream before reaching the sensor. This 6. Calibration
must be done with minimum impact on accuracy or time of
6.1 A calibration technique is described in Practice D4178
response. If the contaminant is an aerosol of oil, glycol, and so
that should be used to verify the accuracy of the analyzer. This
forth, a coalescing filter or semipermeable membrane separator
method uses the known vapor pressure of water at 0°C and
must be used.
mixes wet gas and dry gas to make up the total pressure so that
5.1.2 Construction—Sampling may be done at high or low
a standard gas of known water concentration is achieved.
pressure. All components subject to high pressure must be
rated accordingly. To minimize diffusion and absorption, all 6.1.1 Instruments very sensitive to sample flow must be
materials in contact with the sample before the sensor must be compensated for barometric pressure.
made of stainless steel. Tubing of 1⁄8-in. stainless steel is 6.2 A commercially made water vapor calibrator is shown in
recommended. (Warning —Use appropriate safety precau- Fig. 1, which uses essentially the same technique. This method
tions when sampling at high pressure.) is useful only between 5 to 50 lb/MMSCF.
Copyright by ASTM Int'l (all rights reserved); Sun Dec 10 17:03:59 EST 2017
2
Downloaded/printed by
Universidad Industrial de Santander (Universidad Industrial de Santander) pursuant to License Agreement. No further reproductions authorized.
D5454 − 11´1
6.3 Low-range water vapor standards may be obtained by dations prior to use. See Section 6. Verification of a dry
the use of water permeation tubes. Permeation rates must be instrument using dry compressed nitrogen to get a reading
established by tube weight loss. below 1 lb/MMSCF is recommended before field use.
6.4 Compressed gas water vapor standards may be used, 7.2 Sample Procedure—Sample as in 5.1.1.1. Use as short a
provided they are checked by an independent method once a sample line as practical. Purge the sample for 2 min before
month. valving to the sensor.
6.5 Calibrate the analyzer using one of the standards in 6.3 7.3 Reading—The time for a sensor to come to equilibrium
and 6.4 and respective procedures. Calibration must be at two is variable depending on its type and condition. The analyzer
points, one higher and one lower than average expected may require 20 min to stabilize. Some analyzers have an
readings. Some analyzers can have large nonlinear errors. Use external recorder output, and these can be used with a chart
the calibration adjustment if applicable. recorder to become familiar with the true equilibrium response
time.
7. Procedure 8. Precision and Bias
7.1 Preparation—The analyzer operation and calibration 8.1 Precision data is being prepared for this test method by
should be checked according to the manufacturer’s recommen- an interlaboratory study.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/
Copyright by ASTM Int'l (all rights reserved); Sun Dec 10 17:03:59 EST 2017
3
Downloaded/printed by
Universidad Industrial de Santander (Universidad Industrial de Santander) pursuant to License Agreement. No further reproductions authorized.