TAM BasicTheory Applications 2019
TAM BasicTheory Applications 2019
TAINSTRUMENTS.COM
Agenda – Basic TAM Course
•Calibration
TAINSTRUMENTS.COM
Thermal Measurements
TAINSTRUMENTS.COM
Thermal Analysis and Calorimetry
▪ Thermal analysis
▪ The change of a property as function of
temperature
▪ DSC (scanning mode), TGA, DMA, TMA (slow
scanning mode)
▪ Calorimetry
▪ The measurement of heat properties, i.e. heat
flow, heat and heat capacity as function of
time and temperature
TAINSTRUMENTS.COM
Thermal Analysis versus Calorimetry
ITC
TGA Solution
DSC
Calorimetry
TMA
TAM Reaction
DMA Calorimetry
TAINSTRUMENTS.COM
Calorimetric Range
1 nW 1 mW
-11 -9 -7 -5 -3 -1 1 3
-12 -10 -8 -6 -4 -2 0 2 4
1 pW 1 mW 1W
TAINSTRUMENTS.COM
Specific Sensitivity – a Comparison
Sample Specific
Sensitivity
amount sensitivity
Nanocalorimeter 10 nW 5g 2 nW/g
DSC 200 nW (0.2 mW) 10 mg 20 µW/g
TAINSTRUMENTS.COM
TAM versus DSC
•Complementary techniques
•High sensitivity in the range - nW vs. µW
•High specific sensitivity - g vs. mg
•Measurements in hours vs. minutes
•Absolute heat capacity determinations –
< 0.2 % vs. 1-5 %
•Most TAM experiments are done
isothermally
TAINSTRUMENTS.COM
INSTRUMENTATION
TAINSTRUMENTS.COM
TAM – Thermal Activity Monitor
TAINSTRUMENTS.COM
General Features of the TAM
▪ Thermostat
▪ Calorimeters (4mL, 20mL, 125mL)
▪ Sample handling (Ampoules)
▪ Accessories
▪ Software
TAINSTRUMENTS.COM
TAM – Isothermal and Scanning Thermostat
TAINSTRUMENTS.COM
TAM Temperature Modes
•Isothermal
•Step-wise scanning
•Slow scanning (max ±2 °C/hr)
Time
- Temperature profile
- Example heat flow data
TAINSTRUMENTS.COM
TAM Thermostat
Calorimeter
sink medium that prevents any
thermal event in a test sample or
from the room environment from inner
altering the constant temperature tank
bath
Peltier
cooler
fine heater
over flow
system
TAINSTRUMENTS.COM
TAM IV Thermostat Features
TAINSTRUMENTS.COM
The Flexibility of the TAM
Sample size
Nanocalorimeter
Multicalorimeter (4 mL) Microcalorimeter Multicalorimeter (20 mL)
Macrocalorimeter
Absolute Sensitivity
TAINSTRUMENTS.COM
Nanocalorimeter
TAINSTRUMENTS.COM
Detection System of the Nanocalorimeter
Thermopiles
(two on each side)
Heat Sink
Partition wall
Sample
TAINSTRUMENTS.COM
Nanocalorimeter – Heat Flow Stability over 24hr
30
20 Baseline stability/24h
Heat flow (nW)
0
Drift
-10
-20
-30
0 4 8 12 16 20 24
Time (h)
TAINSTRUMENTS.COM
Microcalorimeter
TAINSTRUMENTS.COM
Minicalorimeter (4 and 20 mL versions)
TAINSTRUMENTS.COM
Minicalorimeter (4 mL)
Sample
Ampoule
holder Outer
Steel
Cylinder
Thermopiles
Heat Sink
Reference
Ampoule
holder
TAINSTRUMENTS.COM
Minicalorimeter Measuring Assembly
Sample Side
Calibration Heater
Reference Side
TAINSTRUMENTS.COM
Minicalorimeter (unbalanced) - Baseline Stability
300
200
/ nW
100
P /nW
0
Heat flow
-100
-200
-300
0.00 4.00 8.00 12.00 16.00 20.00 24.00
Time /h
The absolute temperature in this measurement was 50.0 ˚C. Note: must acknowledge
25
that the calorimeter is not balanced due to the permanent reference plug.
TAINSTRUMENTS.COM
Multicalorimeter (4 mL)
26 TAINSTRUMENTS.COM
Minicalorimeters (4 mL) in TAM 48
27 TAINSTRUMENTS.COM
Macrocalorimeter
Lifter
Access tube
Thermal shunts
TAINSTRUMENTS.COM
Calorimeter Theory
TAINSTRUMENTS.COM
Definitions
30 TAINSTRUMENTS.COM
Rate Equation in Terms of Heat Flow
dC
= k f (c ) mol
m3 s
dt
dQ dC J mol J
= H m3 s = m3 s mol
dt dt
dQ
= H k f (c) = Heat flow signal from TAM
dt
TAINSTRUMENTS.COM
Heat Flow versus Time
18000
16000
flow (µW/g)
14000
Shows how the
Heatflow (µW/g)
12000
f(t, P, Q) reaction rate varies
10000
with time.
8000
Heat
6000
4000
2000
0
0 1 2 3 4
Time(d)
Time (d)
32 TAINSTRUMENTS.COM
Energy versus Time
18000
16000
14000
Shows how the
(J/g)
12000
extent of reaction
Energy (J/g)
10000
varies with time.
Energy
8000
6000
4000
2000
0
0 1 2 3 4
Time (d)
Time (d)
33 TAINSTRUMENTS.COM
Heat Flow versus Energy
18000
16000
(µW/g)
14000
(µW/g)
12000
flow
10000
Heatflow
8000
Heat
6000
4000
2000
0
0 5000 10000 15000 20000
Energy
Energy (J/g)
(J/g)
Shows how the reaction rate varies with the extent of reaction
(ex. 1 order → Q proportional to P with rate const., k, as proportionality constant)
34 TAINSTRUMENTS.COM
Calorimetric Unit
General Heat Balance Equation
dQ
To
dt
dQ dT
= + C
TS dt dt
dT Rate of Heat Rate of Heat Rate of Heat
C
dt Production = Exchange + Accumulation
TAINSTRUMENTS.COM
The Heat Balance Equation (in terms of temperature)
dQ dT
= +C Single calorimeter
dt dt
= k (T − To )
Note: dQ/dt depends on
(Newton's cooling law) the surrounding temperature.
dQ dT
= k (T − To ) + C
dt dt
TAINSTRUMENTS.COM
The Heat Balance Equation (in terms of voltage)
= k (T − To ) =
k U dT 1 dU
V T= + To =
g g dt g dt
dQ k C dV k C dV dV
= V+ = V + = V +
dt g g dt g k dt dt
Tian Equation
g = Seebeck coefficient (V/K)
dQ dV
= V + = calibration constant (W/V)
dt dt
= time constant (s)
TAINSTRUMENTS.COM
Twin Channel Calorimeter (as used in TAM)
S R
Sample Sample
holder holder
dQS dQR
TS CS TR CR
dt dt
To
TAINSTRUMENTS.COM
The Heat Balance Equation (Twin System)
Sample side
dQ S dT
= k S (TS − To ) + C S S
dt dt
Reference side
dQ R dTR
= 0 = k R (TR − To ) + C R
dt dt
Subtraction gives
dQS dTS dTR
= k S (TS − To ) + CS − k R (TR − To ) − CR
dt dt dt
39 TAINSTRUMENTS.COM
The Heat Balance Equation (Twin System)
dQ S d (TS − TR )
= k (TS − TR ) + C
dt dt
Assumptions:
k = kR = kS
C = CR = CS
Qseries DSC does not make this assumption in (T4 mode). Q20 and 29XX DSC and most
other commercial DSCs also make this assumption. High heat capacity and the thermal
stability within the calorimeters aid in the acceptance of the assumptions for the TAM.
TAINSTRUMENTS.COM
Heat Capacity Balance versus Noise Level
4 mL
Mini
calorimeter
20 mL
Mini
calorimeter
Empty Balanced
(unbalanced)
TAINSTRUMENTS.COM
Choosing a Reference
Empty 20 mL calorimeter ( 60 s)
RH Perf./Titr. 206
1 g Water 41-44
1 g Glass beads 5-10
Heat capacity (C) of common materials:
(measured with TAM 2277 at 25 °C)
Water 4.18 J/K•g
*Use GPT to measure .
Sand (Quartz) 0.8 J/K•g
Glass 0.84 J/K•g
Stainless Steel 0.47 J/K•g Balance heat capacity
Aluminum 0.90 J/K•g (C) or time constant ()*
= C/k
See also EN 008
TAINSTRUMENTS.COM
Heat Capacity of Water Versus Temperature
4
.
2
2
4
.
2
1
HeatCapacity(J/Kg)
4
.
2
0
4
.
1
9
4
.
1
8
C p ,ri − 0.24 − C p ,a − C p , s m
= 100
C p ,ri + 7.18
Minicalorimeters:
Cp,ri , Cp,a and Cp,s are heat capacities of reference insert, ampoule and the substance
Heat capacity of the ampoule holder (reference side) = 7.42
Difference in heat capacity of the ampoule holders sample and reference side = 0.24
Mass (m) of substance used can be optimised so that approaches zero.
Maximum deviation for which the signal will lie within specifications is ±20%.
45 TAINSTRUMENTS.COM
Choosing a Reference – 4 mL Minicalorimeters
3320 10.8
46
4 ml St.Steel Ampoule threaded cap
TAINSTRUMENTS.COM
Choosing a Reference – 4 mL Minicalorimeters
Please refer to Table 1 that lists the total Cp of all 8 models of the 20 mL minicalorimeters.
For sample Cp calculations, utilize Table 2 for a list of some common sample materials.
Organic:
Lactose 1,22
Example 1: Sample of gunpowder loaded into a calorimeter with reference Glycine 1,32
number 6 (15.9 J/K). Urea 1,55
Glucose 1,24
Salicylic acid 1,16
Using a stainless steel threaded ampoule (2277-301) with Cp of 11.9 J/K the Gun powder 1,28
remaining heat capacity to balance the calorimeter is 4 J/K. Knowing the sample is
gun powder that has a Cp of 1.28 J/gK one can calculate that approximately 3.1 g Table 2: List of Sample Cp
2.5 g of gun powder must be loaded into the ampoule to best balance the
calorimeter to within 20% of the total Cp.
TAINSTRUMENTS.COM
Choosing a Reference – 20 mL Minicalorimeters
Please refer to Table 1 that lists the total Cp of all 8 models of the 20 mL minicalorimeters.
For sample Cp calculations, utilize Table 2 for a list of some common sample materials.
TAINSTRUMENTS.COM
When to Calibrate
49 TAINSTRUMENTS.COM
Calibration of TAM IV
50 TAINSTRUMENTS.COM
Purpose of Calibration
51 TAINSTRUMENTS.COM
Two Types of Calibrations
52 TAINSTRUMENTS.COM
Static or Dynamic Calibration
◆ For fast processes, i.e. when the response of the reaction is less
or in the range of the time constant of the system, a dynamic
calibration should be used. The thermal inertia of the calorimetric
system is taken into consideration. The displayed heat flow
always corresponds to the rate of heat production.
- ITC experiments: utilize Feedback mode or dynamic calibration
53 TAINSTRUMENTS.COM
Calibration Conditions
•Ampoule Experiments
▪ Empty ampoule lifters (or empty ampoules) should be in position in both sample and
reference side
54 TAINSTRUMENTS.COM
Static Calibration - Pulse
Exponential heat exchange
P (µW) t
Calibration Power ON P = Po exp(− )
= set µW pulse
Calibration
Power OFF
How to obtain
The time until P/Po = 1/e = 0.37
P/Po = 1/e
55 TAINSTRUMENTS.COM
Heat Flow Calibration Procedure
• Apply settings:
▪ Pulse calibration and integration(~0.5-1 h)
▪ Steady state calibration (2-3 h)
◆ Not for minicalorimeters
56 TAINSTRUMENTS.COM
Dynamic Mode
• Heat flow data will not reflect the true response of the sample for reactions with elapsed times
less than 10 min. For reactions where the slope of the heat flow time curve ( f) is changing
rapidly a dynamic correction can be applied to obtain the true response of the sample (P) using
the following formula (Tian’s equation)
df
P =f +τ
dt
• Dynamically corrected data represents the true data of the sample and has been calculated
from Heat flow data using the information about time constants obtained from a dynamic
calibration.
▪ The TAM Assistant software contains functions for considering the effects of the thermal inertia
(dynamic mode). TAM Assistant uses two time constants rather than one to get a better precision in the
correction (cf. Taylor expansion).
df df
P = f + ( 1 + 2 ) + 1 2 2
dt dt
1 and 2 are time constants obtained from dynamic calibration
57 TAINSTRUMENTS.COM
Dynamic Calibration
30
Calibration
power
Heat flow (µW)
20
Measured heat
10 flow
0
0 20 40
Time,min
TAINSTRUMENTS.COM
Calibration Results Example
Raw Corr
(Heat flow (Raw data
monitored recalculated
by the heat using info
flow sensor) about time
constants)
59 TAINSTRUMENTS.COM
Dynamic Calibration Procedure
60 TAINSTRUMENTS.COM
General Performance Test (GPT)
61 TAINSTRUMENTS.COM
Calorimeter Enthalpy Validation
TAINSTRUMENTS.COM
Ampoules and Accessories
TAINSTRUMENTS.COM
TAM IV – sample handling system
TAINSTRUMENTS.COM
Static measurements
•Stability
•Compatibility
•Reaction kinetics
•Amorphicity
•Polymorphism
•Curing
•Safety assessment
•Microorganism growth
•Etc...
TAINSTRUMENTS.COM
Closed, Sealed or Static Ampoules
66 TAINSTRUMENTS.COM
Titration setup
– Possibilities for adding and mixing
TAINSTRUMENTS.COM
Titration Ampoule
68 TAINSTRUMENTS.COM
Gas or liquid perfusion
• Stability
• Compatibility
• Absorption
• Reaction kinetics
•Oxidation
• Safety assessments
• Hydration
•Swelling
• Metabolism of living
small animals
• Etc...
70
TAINSTRUMENTS.COM
Glass Plate Holder
71
TAINSTRUMENTS.COM
Perfusion Insert
TAINSTRUMENTS.COM
Relative humidity perfusion
• Amorphicity assessments
• Polymorphism
• Stability
• Compatibility
•Absorption
• Reaction kinetics
• Safety assessments
• Hydration
•Swelling
• Etc...
TAINSTRUMENTS.COM
Verify the Humidifying System
▪ A salt solution with a known RH is loaded into the ampoule and heat flow is adjusted to
zero to calculate error in RH.
▪ The difference between the expected RH and calculated RH defines the error
▪ The first salt solution is replaced with a second salt solution and the RH adjusted to
zero to calculate error in RH.
▪ The difference between the expected RH and adjusted RH defines the error
74 TAINSTRUMENTS.COM
Suitable Salt Solutions
(1983)
TAINSTRUMENTS.COM
RH Perfusion Calibration – Method I
P (µW)
1st Part 2nd Part 3rd Part
600 min 1+6xM
Waiting for Load
stable RH amp.
conditions Wait for
phys.
equilib.
RH Perfusion
empty.
Reference
empty.
Zero adj. if
necessary!!
Pause BL Pause M M M M M M M BL
Time
TAINSTRUMENTS.COM
LiCl at 25°C; RH from 5-15% in 1% steps
16 150
C: -79.1±0.1 µW
C: -74.7±0.05 µW
14 100 C: -57.0±0.04 µW
Relative humidity (%) C: -39.1±0.04 µW
C: -22.7±0.05 µW
C: -6.28±0.05 µW
10 0
8 -50
C: -63.3±0.04 µW
6 12 18 24 30 36 42
Time (ho ur)
TAINSTRUMENTS.COM
LiCl at 25°C; RH from 5-15% in 1% steps
RH Calibration Curve
78 TAINSTRUMENTS.COM
LiCl at 25°C; RH from 5-15% in 1% steps
RH Calibration Curve
Solve for y = 0
187.5 / 16.641
= 11.27%
LiCl at 25°C
Should be 11.3%
Correction factor
= 0.03%
79 TAINSTRUMENTS.COM
Dissolution
•Heat of dissolution
• Heat of wetting
• Amorphicity
• Polymorphisms
•Dissolution kinetics
• Etc...
Solution Calorimeter
(solid or liquid samples)
Microsolution ampoule
(solid sample)
TAINSTRUMENTS.COM
Micro Solution Ampoule
TAINSTRUMENTS.COM
Micro Solution Calorimeter Experiment
Zero adj. if
necessary!!
Sapphire tip
TAINSTRUMENTS.COM
Simultanoues pressure and heat flow
measurements
TAINSTRUMENTS.COM
Vacuum / Pressure Ampoule
85
TAINSTRUMENTS.COM
Large and heterogenous samples
•Batteries
Soil25 (3-26-10).rslt, 25 FYM Gluc Soil 23 (3-19-10).rslt, Sludge Glu
Soil 45 (3-19-10).rslt, (NH4)2SO4 Glu
•Environmental science
200
•Food applications
100
50
0
0 8 16 24 32 40 48
Time (ho ur)
TAINSTRUMENTS.COM
Battery testing
TAINSTRUMENTS.COM
Macrocalorimetry Accessories
TAINSTRUMENTS.COM
Accessories
TAINSTRUMENTS.COM
Voltage I/O module
Input Output
0 to 15 V (1 amp max) 0 to 14 V
TAINSTRUMENTS.COM
TAM Assistant
TAINSTRUMENTS.COM
TAM Assistant Allows You to:
•Control devices
•Run experiments
•View and edit results
•Perform analysis and calculations
•Create and edit reports
•21 CFR 11 compliant version available
93 TAINSTRUMENTS.COM
Temperature Watchdog
TAINSTRUMENTS.COM
Precision Solution Calorimeter
(SolCal)
TAINSTRUMENTS.COM
Precision Solution Calorimeter
•Heat of dissolution
▪amorphicity
TAINSTRUMENTS.COM
Precision Solution Calorimeter
97 TAINSTRUMENTS.COM
Heat Balance Equation - SolCal
+ k (T − Ts )
dQ dQF dT
Temperature of
surrounding
− − =C (1)
and in calorimeter
dt dt dt
Heat of Solution
Ts Heat exchanged
with surrounding
Stirring, heating Heat accumulated
T from termistor by the system
98 TAINSTRUMENTS.COM
Heat Balance at Baseline
+ k (T − Ts )
dQF dT
− =C (2)
dt dt
dT
At t C →0
dt
T T
= k (T − Ts )
dQF
Tsurr − (3)
dt
Substitution of eq. 3 into eq.1 gives
Time
− k (T − T )
dQ dT
− =C (4)
dt dt
dQF
99
dt is eliminated
TAINSTRUMENTS.COM
Heat of Solution
dT 1
− (T − T )
dQ
− = C (4b)
dt dt
T
C
where =
End k
T Integration of 4b
E
dQ
Heat of Solution = −Q = − dt
Start S
dt
E
− Q = C Tobs + (T − T ) dt
tS tE Time 1
S
TAINSTRUMENTS.COM
Heat of Solution
T − Q = C Tcorr
End
T
Tobs Tcorr = Tobs − Tadj
Start E
Tadj = (T − T ) dt
1
tS tE Time
S
TAINSTRUMENTS.COM
Baseline
+ k (T − T )
dT
T 0=C (5)
dt
by differentiation
T −t
T = T − Tnorm e (6)
−t
T0 T = T − Tnorm e
where
Tnorm = T0 − T
t0 (7)
Time
T,
102
TAINSTRUMENTS.COM
Break Experiment
Electrical Break
Calibration Experiment
Input QCal ?
Output Tobs
Cal
Tobs
Break
- -
Baseline Tadj
Cal
Tadj
Break
Analysis, T, t
Tcorr
Cal
Tcorr
Break
QCal
C= Q = C T Break
Tcorr
Cal reaction corr
103
TAINSTRUMENTS.COM
Isothermal versus Adiabatic Calorimetry
10
4
TAINSTRUMENTS.COM
Verify Performance of SolCal
• 100 mL water
• 3 - 50 J electronic calibrations
• Cp of system = 445 8 J/K
• Average of three within 0.2 J/K
• = C/k 8.200 ks
10
5
TAINSTRUMENTS.COM
Verify Performance of SolCal
TAINSTRUMENTS.COM
Chemical Calibration
10
7
TAINSTRUMENTS.COM
Example Report – SolCal experiment (page 1)
TAINSTRUMENTS.COM
Example Report – SolCal experiment (page 1)
TAINSTRUMENTS.COM
Example Calculation for KCl Experiment
Error Calculation:
Before break - Hsol(KCl)(T) = 234.77 J/g (from report)
Error = [(234.77 - 235.14)/235.14]100 = -0.16%
After break - Hsol(KCl)(T) = 234.78 J/g (from report)
Error = [(234.78 - 234.76)/234.76]100 = 0.001%
TAINSTRUMENTS.COM
Comments on KCl Experiment
▪ Usually calculation using the calibration before or after the break give
slightly different results of H. One reason is hydrodynamic stirring
changes after breaking the ampoule and may result in different
estimations of t and , which are essential for calculation of the
calibration constant. Secondly, the system may not have achieved
steady-state when the first calibration begins (check standard
deviation). For this reason the calculation that utilizes the calibration
after break is typically a better representation.
▪ Corrections for evaporation and condensation effects when breaking
the ampoule can be taken as negligible in this case.
▪ For additional sources of error, see reference by I. Wadso in the
Appendix of the Precision Solution Calorimeter Instruction Manual.
11
1
TAINSTRUMENTS.COM
Calculation – Heat of Solution for KCl
Concentration Dependence
Conditions:
Molecular weight: 74.5513 g/mol (KCl: NIST 1655)
Temperature = 25 °C = 298.15 K
KCl Molality = mKCl = 0.05551 - 0.15 mol/kg
Deviation from NIST certificate using the formula below should be less than 0.3 %
when concentrations are in the given range. Concentrations outside this range should
be regarded as an extrapolation and may result in an increase in error.
TAINSTRUMENTS.COM
Calculation – Heat of Solution for KCl
Temperature Dependence
Then…
1 − a cw
Wv = Wa
1 − a
( )
= Wa 1 + a 1 KCl − 1 cw Wa (1.000455)
KCl
Assume:
Concentration of KCl and
Ambient Temperature = 22 ± 1 C Temperature are also important
Ambient RH = 35 ± 15% for calculation of H!
Ambient Pressure = 750 ± 10 mm Hg
Density of Air = air = 0.0012 g/cm3
Density of KCl = KCl = 1.98 g/cm3
Density of Counter Weights = cw = 7.95 g/cm3 (e.g. brass 7.8 g/cm3, ss 8.0 – 8.4 g/cm3)
TAINSTRUMENTS.COM
Ideal experimental tips
TAINSTRUMENTS.COM
Ideal experimental tips
TAINSTRUMENTS.COM
Ideal experimental tips
TAINSTRUMENTS.COM
Experimental
Considerations
TAINSTRUMENTS.COM
TAM is a Non-Specific Technique
Process 2
▪ Individual contributions may be
Heat flow
distinguished by varying the Process 1
experimental conditions.
11
9
TAINSTRUMENTS.COM
Designing an Experiment
12
0
TAINSTRUMENTS.COM
Sample Handling Systems
12
1
TAINSTRUMENTS.COM
Tools Required - Disposable Glass Ampoules
TAINSTRUMENTS.COM
Sealing Disposable Glass Ampoules (3 & 4 mL)
TAINSTRUMENTS.COM
Disturbances with Disposable Glass Ampoules
•Can be minimized by
▪pre-storing the ampoules and lids at the operating temperature for 24 hours
▪Preparing the sample and the reference ampoule at the same time (use a
new reference for each measurement)
TAINSTRUMENTS.COM
Heat Seal Ampoules
•Completely sealed
•Heat seal ampoules of glass
•Special ampoule lifters required
Water circulation
inlet/outlet
•Stainless steel
▪resistance towards corrosion
▪should not be used for solvent with pH < 4
•Hastelloy
▪Improved resistance towards corrosion and acids
▪excellent for use with organic solvents
TAINSTRUMENTS.COM
Circlip Cap Ampoules
•Stainless steel
▪resistance towards corrosion
▪should not be used for solvent with pH < 4
•Hastelloy
▪Improved resistance towards corrosion and acids
▪excellent for use with organic solvents
TAINSTRUMENTS.COM
Equilibrium
•Thermal equilibrium
▪Within 60 min after loading
◆ depends on sample size – 20mL ampoules may take longer to
fully equilibrate
•Physical equilibrium
▪Depends on the sample and the pre-history
▪Might depend on the ampoule itself
•Chemical equilibrium
▪Slow/fast reactions
TAINSTRUMENTS.COM
Pre-history of the Sample
TAINSTRUMENTS.COM
Sample Geometry and Surface Area
TAINSTRUMENTS.COM
Sample Amount
TAINSTRUMENTS.COM
Kinetic Evaluation
TAINSTRUMENTS.COM
Choice of Reference Materials
TAINSTRUMENTS.COM
Baseline collection
TAINSTRUMENTS.COM
Baseline collection
TAINSTRUMENTS.COM
Baseline collection
TAINSTRUMENTS.COM
Baseline collection
TAINSTRUMENTS.COM
Baseline collection
TAINSTRUMENTS.COM
Baseline problems
TAINSTRUMENTS.COM
Baseline problems
TAINSTRUMENTS.COM
Baseline problems
TAINSTRUMENTS.COM
Baseline problems
TAINSTRUMENTS.COM
Baseline problems
TAINSTRUMENTS.COM
Baseline problems
TAINSTRUMENTS.COM
Baseline problems
TAINSTRUMENTS.COM
Baseline problems
TAINSTRUMENTS.COM
Baseline collection
TAINSTRUMENTS.COM
TAM Thermostat
Calorimeter
that prevents any thermal event in a test
sample or from the room environment
from altering the constant temperature
bath inner
tank
Peltier
cooler
fine heater
over flow
system
TAINSTRUMENTS.COM
Calorimeters
TAINSTRUMENTS.COM
Calorimeters
TAINSTRUMENTS.COM
Calorimeters
TAINSTRUMENTS.COM
TAM Applications
Pharmaceuticals
Life Science
Material Science
TAINSTRUMENTS.COM
TAM Applications
Stability
TAINSTRUMENTS.COM
TAM Stability Testing
Least stable X
exo
X
Heat Flow
X
Most stable
Time
TAINSTRUMENTS.COM
TAM Stability Testing
40°C 30°C
60
200
40
100
20
-100
-20
-200
-40
-300
-60
8 16 24 32 40 1 2 3
Time (ho ur) Time (da y )
TAINSTRUMENTS.COM
Active Pharmaceutical Ingredient Stability
Conditions:
60 TAM Temperature 50 °C
Side A: API and hygrostat filled w/ saturated NaCl
inside 4 mL SS screw cap ampoule
Lowering ampoule 40
Side B: Empty hygrostat inside 4 mL SS screw cap
into calorimeter ampoule
20
TAINSTRUMENTS.COM
Stability with TAM and Compared with HPLC
50°C
40°C
30°C
30°C
HPLC data
23°C
Ο TAM data
Otsuka T., Yoshioka S., Aso Y. and Terao T., Chem. Pharm. Bull., 42(1) 1994
TAINSTRUMENTS.COM
Hydrate Formation in Ethinyl Estradiol
600 Measuring
temperature: 45C
P/m /uW g-1
400
• Blue trace: 100 %RH
200
• Red trace: 95 % RH
0
0 20 40 60 80 100 120
• Green trace: 88 %RH
Time /h
600
400
Rate equation:
P/m /uW g-1
200
0 P = k (T ) f (q / H )
-200
-5 0 5 10 15 20 25 30
-1
q/m /J g
P = k (T ) f (q / H )
f (q / H ) = 2(1 − q / H ) [− ln(1 − q / H )]1/ 2
600
400
• Blue trace: experimental
P/m /uW g-1
200 data
0
• Black trace: fitting
-200
-5 0 5 10
-1
15 20 25 30 equation (k=0.0005 s-1)
-4 q/m /J g
x 10
Avrami rate constant
5
4
• Rate constant as a
3 function of relative humidity
2
1
0
82 84 86 88 90 92 94 96 98 100
Relative humidity (%)
RHcritical
TAINSTRUMENTS.COM
TAM Applications
Compatibility Testing
TAINSTRUMENTS.COM
Compatibility Measurements
A B Standard
mixture
Additional mixture
samples for special
tests
TAINSTRUMENTS.COM
Evalutation of Compatibility Measurements
A+B (measured)
A+B (theoretical)
Time (a.u.)
If the heat flow curve of A+B (measured) differs from A+B (expected), this is
an indication that the materials affect each other or are incompatible.
16
2
TAINSTRUMENTS.COM
Compatibility Between Wax and Mineral Wool
TAM Temperature: 50 °C
Aliquot of water added to ampoule
Ampoule
removed
164
Schmitt, E.A.; Peck, K.; Sun Y.; Geoffroy, J-M. Thermochim ACTA, 380, 175-183 (2001)
TAINSTRUMENTS.COM
Compatibility Experiment with TAM
TAM Temperature: 50 °C
Relative Humidity: 75%
Phipps, M.A.; Mackin, L.A. Pharm. Sci. Tech. Today, 3(1), 9-17 (2000)
TAINSTRUMENTS.COM
TAM Assistant Analysis (1 of 2)
TAINSTRUMENTS.COM
TAM Assistant Analysis (2 of 2)
Step 3
Step 4
TAINSTRUMENTS.COM
ASA with Mg-Stearate and Sucrose
TAM III at 40 °C
TAINSTRUMENTS.COM
Amine-Lactose Interactions
One approach to perform excipient compatibility screening is to add water to the powder
mixture. The graph shows the response of an amine-lactose interaction at different
temperatures with 20% water added.
Schmitt, Peck, Sun & Geoffroy, Thermochim. Acta, 380, 175-183, (2001).
TAINSTRUMENTS.COM
Drug - Excipient Stability
Time in days
Another approach to perform excipient compatibility screening is to compress (make
a tablet) with the ingredients. The figure shows the same compressed mixture at
several different temperatures and clearly the rate of reaction increases as the
temperature increases.
TAINSTRUMENTS.COM
Kinetic Analysis with TAM Assistant
Effect of Temperature
Isothermal Models
TAINSTRUMENTS.COM
TAM Applications
Amorphicity and Polymorphic Studies
TAINSTRUMENTS.COM
Definition of Crystallinity
17
4
TAINSTRUMENTS.COM
Relevance of Crystallinity
17
5
TAINSTRUMENTS.COM
Characterize Amorphicity with TAM
17
6
TAINSTRUMENTS.COM
The Microhygrostat Method
Microhygrostat:
Glass tube with pure
solvent or a solvent
saturated by a salt
Solid (e.g. sat. NaCl (aq))
sample
120
• Amorphous lactose
100
57 %RH • Exp temperature:
mass )/mW g-1
60
• Microhygrostat: Pure
40
water (100 %RH) and
sat. aq. NaBr (57
%RH)
20
0 2 4 6
Time (ho ur)
Time /h
TAINSTRUMENTS.COM
The Controlled RH Perfusion Ampoule
a
f
e
f
a) inlet of dry nitrogen
b) dry flow line
c) humidifiers and thermal
equilibrators c
d) sample ampoule
e) outlet
f) wet flow line
g) sample powder bed
d
g
TAINSTRUMENTS.COM
Moisture Induced Crystallization
RH ramp
80
0.3%
69.04mJ amorphous!
20
- 69.295mJ
-40
89.25mJ 69.04mJ
-100 40 → 30 %RH
0 2 4
Time/hour
Sapphire tip
TAINSTRUMENTS.COM
Dissolution of Lactose
18
4
TAINSTRUMENTS.COM
Heat of Solution for Different Lactose Forms
TAINSTRUMENTS.COM
Standard Curve <10%
18
7
TAINSTRUMENTS.COM
Benefits of TAM Over Other Techniques
Sensitivity
Detection limits below 1%, towards 0.1% is possible
Results are highly reproducible
Versatility
Adapt method to substance (different solvent vapors, sample
sizes and operating temperatures)
Sample throughput
The sample throughput attainable with TAM IV and TAM 48 can
not be achieved by any other method today.
18
8
TAINSTRUMENTS.COM
Polymorphism Introduction
18
9
TAINSTRUMENTS.COM
Main Issues
19
0
TAINSTRUMENTS.COM
→ Transformation of Tripalmitin
TAINSTRUMENTS.COM
→ Transformation of Tripalmitin
Note: the results were obtained below the DSC sensitivity level
The enthalpy does not change with temperature indicating that the same
transition has been observed at all the different temperature.
19
2
TAINSTRUMENTS.COM
Solution Calorimetry
Temp offset, mK
20
-20
Calibration 1
Calibration 2
Baseline
Baseline
Baseline
Baseline
Break
Pause
End
0 15 30 45 Time, min
T = 45 °C
19
3
TAINSTRUMENTS.COM
Heat of Solution
polymorph A solHm,A
tHm Solution
polymorph B solHm,B
19
4
TAINSTRUMENTS.COM
Which Direction? A → B or B → A?
195
TAINSTRUMENTS.COM
Transition Temperature of a Polymorphic Pair
(A and B)
Transition enthalpy from SolCal
+ solubility of form A and B
TAINSTRUMENTS.COM
Results with 4 drug compounds
Data from : Koji Urakami et al. (2002) A novel method for estimation of transition temperature for
polymorphic pairs in pharmaceuticals using heat of solution and solubility data. Chem. Pharm. Bull. 50 (2).
TAINSTRUMENTS.COM
Polymorphic Transformations
TAINSTRUMENTS.COM
Crystallization from a Super-Saturated Solution
ci
A
Typical solubility
solubility
curves for a
polymorphic pair cA B
c
cB
temperature
Typically the most soluble polymorph is not the most stable form (Form A)
TAINSTRUMENTS.COM
The Solvent Mediated Process
Initial dissolution of A
Dissolution of A
A→B
TAINSTRUMENTS.COM
SMPT with Microcalorimetry
dmA dmB
The Heat Flow: P= DHm + C H m
dt dt
TAINSTRUMENTS.COM
Water Slurries of a Meta-Stable Drug Compound
20
3
TAINSTRUMENTS.COM
Before and After
30
20
P/ W
10
-10
0 100 200 300
Time /h
FormA Form B
20
4
TAINSTRUMENTS.COM
TAM Applications
Temperature Scanning
TAINSTRUMENTS.COM
Applications of Slow Scanning
•Polymorphism
•Crystal Hydrates
•Heat Capacity
•Melting Behavior
•Glass Transitions
20
6
TAINSTRUMENTS.COM
Heat Capacity Determination
Temperature Empty ampoule Sapphire
25.6
Method:
1. Insert empty ampoule in sample side
2. Change the thermostat temperature by
25.4
typically 1°C (blue curve)
200
3. Restore the original temperature
4. Fill ampoule with sample
25.2 5. Repeat the temperature change with the
Temperature (°C)
Heat flow (µW)
-200
24.6
0 6 12 18 24 30
20
7 Time (hour) TAINSTRUMENTS.COM
Heat Capacity Determination with TAM III
10
27
26.83 J
9.579 J
The Geolgically heated water
Temperature (˚C)
-53.99 J
23 -30
Results 0 2 4 6 8 10
Time (hour) Time (h)
Lehto, Laine, Ylianttila, Hyysalo and, Jokela, J. Therm. Anal., Vol. 53, 685-695 (1998)
20
9
TAINSTRUMENTS.COM
→ Transformation of Tripalmitin
On heating
I. → transformation
at 40°C (exo)
II. Melting of form at
III
65°C (endo)
On cooling
III. Crystallisation of form
I (exo)
78 2.0
•TAM – 2 °C/hr (top)
•TAM – 1 °C/hr (bottom)
1.5
Temperature (°C)
76
Heat flow (mW)
1.0
74
0.5 •QDSC – 0.5, 2.0, and 10 °C/min
0.0
-0.5
72
-1.0
210 240 270 300 330 360
Time (min) 0
Hexatriacontane
Temperature, TAM III Thermostat #209
-5
78 1.0
76 0.5
––––––– Hexatriacontane
Temperature (°C)
––––––– Hexatriacontane
––––––– Hexatriacontane
-15
74 0.0
-20
72 -0.5 70 72 74 76 78 80 82
Exo Up Temperature (°C) Universal V4.5A TA Instruments
25 26 27 28 29 30
1
TAINSTRUMENTS.COM
Hexatriacontane - Function of Scanning Rate
212
TAINSTRUMENTS.COM
Other Methods used as Bioassays
TAINSTRUMENTS.COM
Rapid detection of Mycobacteria in culture
•In many developing countries, simple solid culture media are used, and
detection is achieved by visual inspection. Detection is very slow, since
one must wait until a visible colony appears.
21
9
TAINSTRUMENTS.COM
Long-term Studies of GH on Lymphoid Cells
The proliferative response upon addition of human growth hormone, hGH
Heat Flow,µW
Heat flow, µW
50
0.4 nM GH
30
0.2 nM GH
20
PBS
10
5 10 15
Time,hour
Heat Flow,µW
20
PBS
0.2 nM GH
18
0.4 nM GH - 0.7 µW
16 5 nM - 2.2 µW
10 20 30
Time,min
22
2
TAINSTRUMENTS.COM
Drug Efficacy
Flow calorimetry: Leukemia (T-lymphoma) cells exposed to the anti-cancer drug methotrexate. The
final drug concentrations were (a) 0, (b) 0.2, (c) 0.5, (d) 1.0, (e) 2.0, (f) 4.0µM (ref 6).
a
b
c
d
e
f
Bermudez, Backman and Schon., Cell. Biophys. 20, 111-123, (1992). TAINSTRUMENTS.COM
Microorganism Detection
22
Trampuz, A. et al. JCM (2006), 44, 628
5
TAINSTRUMENTS.COM
S. epidermidis in 3 mL TSB
-4
1.75x10 3 104
3106 3 105 CFU
-4
CFU CFU 3000
1.50x10 300 CFU
CFU 3 CFU
30 CFU
-4
1.25x10
Heat flow (W)
-4
1.00x10
-5
7.50x10
-5
5.00x10
-5
2.50x10
0.3
0.00 CFU
0 CFU
0 4 8 12 16 20 24 28 32 36 40 44 48
Time (h) TAINSTRUMENTS.COM
S. aureus & S. epidermidis in TSB
-4
3.0x10
-4
2.5x10
Staphylococcus aureus
Heat flow (W)
-4
2.0x10
-4
1.5x10 Staphylococcus epidermidis
-4
1.0x10
-5
5.0x10
0.0
0 5 10 15 20 25 30
Time (h)
TAINSTRUMENTS.COM
Suspension of Microorganisms in TSB
S. aureus
P.
(ATCC 29213)
aeruginosa
E. coli
S. aureus
(ATCC 43300)
Prop.
acnes
S. epidermidis (clinical
C. albicans
isolate)
S. Epidermidis (ATCC
12228)
M. fortuitum (clinical isolate)
TAINSTRUMENTS.COM
Results
22
9
TAINSTRUMENTS.COM
Results
Tissue Sonicate
23
0
TAINSTRUMENTS.COM
Advantages of Calorimetry
23
1
TAINSTRUMENTS.COM
Enzymatic Reaction of Tyrosine
Tianhong Chen, Heather D. Embree, Li-Qun Wu, Gregory F. Payne “In Vitro Protein-Polysaccharide
Conjugation. Tyrosinase-Catalyzed Conjugation of Gelatin and Chitosan” Biopolymers, 64, 292-302
23
(2002)
2
TAINSTRUMENTS.COM
Oxidation of Tyrosine by Tyrosinase
Nitrogen
0 1 2 3 4
Time (hour)
23
Sigstad and Garcia, Thermochim. Acta 366, 149-155, (2001).
4
TAINSTRUMENTS.COM
Milk Fermentation
TAINSTRUMENTS.COM
Milk Fermentation
2.5
normal
2
1.5 +dextrose
1 +NaCl
0
0 50 100
time / h
TAINSTRUMENTS.COM
Carrot Juice Spoilage
TAINSTRUMENTS.COM
Carrot Juice Spoilage Measured at 25°C
0.5
55oC
heat production rate / mW/g
0.3 58oC
37-49oC
0.2
0.1
0
0 10 20 30 40 50 60 70
time / h
TAINSTRUMENTS.COM
Microbiological spoilage prevention
1
2
0
.
0
A
c
t
i
ve
Mi
l
l
ip
e
de
!
1
0
0
.
0
8
0
.
0
Heat flow (W)
6
0
.
0
N
o
t
i
c
ede
c
r
e
as
e
ds
i
g
na
ld
u
et
o
O2c
o
n
su
m
pt
i
on
4
0
.
0
C
o
n
d
i
ti
o
ns
:
T
A
MT
e
m
pe
r
a
t
ur
e2
5
°C
2
0
.
0
S
i
d
eA:
Mi
l
li
p
ed
es
e
a
le
d
in
s
i
d
e4m
L
S
Ss
c
r
ew
ca
p
am
p
ou
l
e
B
a
s
e
l
in
e
(
W
~
0) S
i
d
eB:
Em
pt
y4
mL
S
Ss
c
r
ew
ca
p
am
p
ou
l
e
0
.
0
1
0
0
00 2
0
0
00 3
0
0
00 4
0
0
00
T
i
me
(s
)
A
m
p
o
ul
e
lo
we
r
e
di
n
to
me
a
s
ur
i
n
gp
os
i
t
i
on
6
0
.
0
C
o
n
d
i
ti
o
ns
:
T
A
MA
i
r
Te
mp
e
r
a
tu
r
e2
5
°C
5
0
.
0 S
i
d
eA:
1
0mL
Y
e
as
t
i
nAd
d
/
Mi
xa
mp
o
u
l
e
S
i
d
eB:
1
0mL
W
a
te
r
i
ng
l
as
s
am
p
ou
l
e
4
0
.
0
Heat flow (mW)
I
n
j
ec
t
io
n
of
0
.1
mL
C
o
k
e
3
0
.
0
I
n
j
ec
t
io
n
of
0
.1
mL
D
i
et
Co
k
e
2
0
.
0
S
t
i
r
ri
n
g
1
0
.
0
0
.
0
2
0
0
0 7
0
0
0 1
2
0
00 1
7
0
00 2
2
0
00
T
i
me
(s
)
TAINSTRUMENTS.COM
Effect of Blanching on Turnip
TAINSTRUMENTS.COM
TAM Applications
Cement Hydration
TAINSTRUMENTS.COM
Sample Ampoules
▪ Static 20 mL disposable
glass ampoules.
▪ Mixing of solid/liquid
outside calorimeter
TAINSTRUMENTS.COM
Portland Cement Basics
TAINSTRUMENTS.COM
Portland Cement Basics
•Influence of concrete
admixtures
24
Dr. Sandberg, Grace Construction Products, US (2002)
8
TAINSTRUMENTS.COM
Cement Hydration
Cement Hydration
Conditions:
350.0 TAM Air Temperature 25 C
600 mW scale
300.0
Cement:Water ratio 1:1 (w/w)
250.0 Side A: ~5 g Cement samples in plastic ampoule
Baseline Side B: ~5 mL Water in plastic ampoule
200.0
150.0
100.0
50.0
0.0
TAINSTRUMENTS.COM
Effect of Contamination
25
Dr. L. Wadsö, University of Lund, Sweden (2002)
0
TAINSTRUMENTS.COM
TAM Air Reproducibility
25
Dr. Moro , Holcim Group Support, Switzerland (2002)
1
TAINSTRUMENTS.COM
Temperature Dependency of Cement Hydration
P (mW)
P reflects the rate of the process
Q (W)
Q reflects the extent of the process
Time (h)
25
Dr.
2
Johansson , Thermometric AB, Sweden (2002) TAINSTRUMENTS.COM
Admix Ampoule -Two Identical Ampoules
Excellent reproducibility !
25
Dr.
3 Moro , Holcim Group Support, Switzerland (2002)
TAINSTRUMENTS.COM
Admix Ampoule Experiment
0.25
––––––– Stucco_w/c 0.50_1
–––– Stucco_w/c 0.50_2
––––––– Stucco_w/c 0.50_5% Salt
0.14hr
0.20
0.15
0.04hr
0.39hr
0.10
0.46hr
0.04hr
0.05
0.00
0.0 0.2 0.4 0.6 0.8 1.0
Time (hr) Universal V4.4A TA Instruments
25
4
TAINSTRUMENTS.COM
Admix Ampoule Experiment
0.12
––––––– Stucco w/c 0.35
–––– Stucco w/c 0.50
0.10
0.08
0.06
0.04
0.02
0.00
0.0 0.5 1.0 1.5 2.0
Time (hr) Universal V4.4A TA Instruments
TAINSTRUMENTS.COM
TAM Applications
TAINSTRUMENTS.COM
Stability with TAM
TAM Temperature: 40 °C
Sodium Percarbonate QC test (AN 22034)
TAINSTRUMENTS.COM
Stability of Organic Peroxides
200
Specific heat flow
150
Heat flow (µW/g)
0
0 0.5 1 1.5 2 2.5 3 3.5 4
Time (weeks)
200
Specific heat flow
150
Heat flow (µW/g)
100
Prediction of the 4 week behaviour from
data obtained during 0.1-2 week using the multi-
50
exponentional function obtained by the fitting procedure.
0
0 0.5 1 1.5 2 2.5 3 3.5 4
Time (weeks)
25
9
TAINSTRUMENTS.COM
Repeatability
1.4
Relative specific heat flow
1.2
1.0
0.8
0.6
0.4
0.2
0.0
0 0.5 1 1.5 2 2.5 3 3.5 4
Time (weks)
26
0
TAINSTRUMENTS.COM
Stabilization of an Energetic Plasticizer (NPN)
26
Wingborg and Eldsäter, Propellants, Explosives and Pyrotechnics 27, 314 -319, (2002).
1
TAINSTRUMENTS.COM
NATO Standard
Conditions:
40 TAM at 25 °C
Side A: Activated carbon inside 4 mL SS
RH perfusion ampoule
30 Side B: Empty 4 mL SS cir-clip ampoule
Heat Flow (We-6)
Carbon B
20
Carbon C
Carbon F
10
25 ppm Ethanol Adsorption onto Activated Carbon
0
0 0.5 1 1.5 2 2.5 3
-10
Time (Hours)
Note: Each sample weighed approximately 15 mg. Heat flow signal was not normalized.
TAINSTRUMENTS.COM
Percarbonate – Vacuum/Pressure Ampoule
220
1.8
200
Heat flow (µW)
1.6
Pressure (bar)
180
1.4
160
1.2
2 4 6 8
Time (day)
TAINSTRUMENTS.COM
Moisture Sorption - Vacuum/Pressure Ampoule
TAINSTRUMENTS.COM
Oxidation of Polymers
•Important intermediates
▪Hydroperoxides and radicals
26
6
TAINSTRUMENTS.COM
Characterization of Oxidation
26
7
TAINSTRUMENTS.COM
Polyamide 6
O H
HO C CH2 N H HO
5 n
Polyamide 6
• Polyamide 6 film from Nyltech, Italy
• Thickness: 40 µm
• Crystallinity: 40 %
• Melting point: 210oC
• Glass transition temp: 70oC (dry), -10oC (100 % r.h.)
• Stabilizer: Irganox 1098 (hindered phenol)
• Sensitive to oxidation!
268
TAINSTRUMENTS.COM
Response to Nitrogen / Oxygen
Heat flow curves for unstabilized polyamide 6 film exposed to nitrogen and
oxygen at 110C. After four days, the gases were switched .
700
Gas switch
600
Heatflow ( W/g)
500
400
oxygen
300
200
100 nitrogen
0
0 2 4 6 8
Ageing time (d)
Influence of Irganox 1098 on the Heat Flow time curve for PA6. The presence of
the stabiliser suppresses the heat flow signal indicating the preventing action.
o
Unstabilised 120 C air
400
300
100 ppm
Heat flow (W/g)
200 ppm
200
400 ppm
100
0
0 1 2 3
Time (d)
C C C C C C
H H H H H H
TAINSTRUMENTS.COM
Rate-Sensitive Techniques
27
2
TAINSTRUMENTS.COM
ICL - Oxidation of Unstabilized PP Powder in Air
1.2 o
120 C
▪ The ”oxidation
o o o o
130 C 110 C 100 C 90 C
1
rate” is not constant but
CL intensity (a.u)
vary in a characteristic
0.8 way.
TAINSTRUMENTS.COM
TAM ICL 100oC
20000 1.2
o
tmax=0.92 d 100 C
18000
1
16000
CL intensity (a.u)
14000
Heatflow (µW/g)
0.8
12000
10000 0.6
8000
0.4
6000
4000
0.2
2000
0 0
0 0.5 1 1.5 2
TAINSTRUMENTS.COM
TAM: Heat Flow versus Energy
Energy (J/g)
TAINSTRUMENTS.COM
Curing of Epoxy in TAM Air
3
(mW/g)
25 °C ▪ Jeffamine D-230
hardener from Huntsman
0
6 12 18 24 30 36
Time (ho ur)
TAINSTRUMENTS.COM
Curing of Epoxy in TAM Air
0.04
––––––– TAMAir_25
––––––– TAMAir_30
––––––– TAMAir_35
0.78hr
––––––– TAMAir_40
––––––– TAMAir_50
0.03
Normalized Heat Flow (W/g)
0.02
1.68hr
7.58hr
2.48hr
0.01
3.14hr
0.00
0 5 10 15 20 25 30
Time (hr) Universal V4.4A TA Instruments
3000
• ESP 110 – usually cured
at 150oC
50oC
• Closed 3 ml glass
ampoules
1 2 3
Time (d)
0
1 0 10 20 30 40 50
1
log10(Heat prod. rate / mW/g )
Q(t)/Q(max)
Time / h
0.5
0.8
0
0.6
Extent of reaction
-0.5
0.4
-1
0.2
-1.5
P rate Q extent
0 0.5 1 1.5 2 0
0 20 40 60
log10(Time / h)
Time / h
TAINSTRUMENTS.COM
Battery Discharge
TAINSTRUMENTS.COM
Self-Discharge (No Load) of Alkaline Battery
10
8
Heat flow (µW)
▪ TAM III at 25 °C
4
0
0 2 4 6 8 10 12
Time (ho ur)
TAINSTRUMENTS.COM
Larger Sample Applications
•Environmental science
•Food applications Soil 17 (3-15-10).rslt, Ca(NO3)2 Glu
Soil25 (3-26-10).rslt, 25 FYM Gluc
Soil 16 (3-19-10).rslt, Straw+N Glu
Soil 23 (3-19-10).rslt, Sludge Glu
Soil 45 (3-19-10).rslt, (NH4)2SO4 Glu
100
50
0
0 8 16 24 32 40 48
Time (ho ur)
TAINSTRUMENTS.COM
RH Perfusion – Step RH
Lowering ampoule
into calorimeter
Conditions:
A TAM at 40 C
50 PBO mass = 0.00490 g
Forceps were used to tease small fiber
bundles from a ~11 cm piece.
25 N2 flow rate 120 mL/hr
Normalized Heat Flow (mW/g)
RH set to 0% - Desorption
B
-25
-50
-75
-100
Poly(p-phenylene-2,6-benzobisoxazole) fiber
25 30 35 40 45 50 55 60 65 70 75 80 85 90
Time (h)
TAINSTRUMENTS.COM
RH Perfusion – Step RH
Conditions:
70
A TAM at 40 C
PBO mass = 0.00490 g
Forceps were used to tease small fiber
60 bundles from a ~11 cm piece.
N2 flow rate 120 mL/hr
Normalized Heat Flow (mW/g)
50
40
30
10 RH set to 90%
30 31 32 33 34
Time (h)
28
4
TAINSTRUMENTS.COM
RH Perfusion – Step RH
0
B RH set to 0%
Singnal returns to baseline
(0 W) after 19 hr.
Normalized Heat Flow (mW/g)
-25
-75
Conditions:
TAM at 40 C
PBO mass = 0.00490 g
Forceps were used to tease small fiber
-100
bundles from a ~11 cm piece.
N2 flow rate 120 mL/hr
70 71 72 73 74
Time (h)
28
5
TAINSTRUMENTS.COM
Pop Quiz!!!
PV Work
Conditions:
5.0 TAM Air Temperature 25 C ▪ Lowering ampoules will
600 mW scale
4.0 Side A: Empty glass ampoule w/ 1mL syringe and canula
create frictional heat and
Side B: Empty glass ampoule
1.0 mL injection
work (or pressure) on the
3.0
calorimeter due to small air
2.0
0.5 mL injection
Baseline (~0 W) gap between ampoule and
Heat flow (mW)
-5.0
TAINSTRUMENTS.COM
TAM Applications
TAINSTRUMENTS.COM
ITC in Biochemistry/Biophysics
20
15
Heat flow (µW)
10
0 1 2 3
Time (ho ur)
TAINSTRUMENTS.COM
ITC in Biochemistry/Biophysics
28
9
TAINSTRUMENTS.COM
Strength of TAM-ITC Relative to Other Methods
29
0
TAINSTRUMENTS.COM
Binding Characterization
•Binding Affinity, K
▪Tightest binding measurable approaches nanomolar
Weakest binding for biological macromolecules millimolar
•Binding Stoichiometry, n
•Binding Thermodynamics, H, G, S and Cp
H measure of the heat released or absorbed
Cp measure of the temperature dependence of H (dH/dT)
G measure of the equilibrium constant (G = -RTlnK)
S measure of ”order” in the system (G = H-TS)
29
1
TAINSTRUMENTS.COM
Determination of Binding Constants
G = -RTlnKa = H-TS
M+L ML
K = [ML] / [L][M]
Q = ([ML] / [M]total) · H
Q = (K · [L] · H) / (1 + K · [L])
kJ/mol
Calculated
dQ/dAmount
30 <Current>
20
10
0
0.0 0.6 1.2 1.8 [L]/[M],
TAINSTRUMENTS.COM
Choosing a Binding Model
• Independent? (one or several identical sites bind the same ligand with the same
enthalpy and Ka, independent of each other)
• Multiple? (two or more binding sites, each capable of binding ligands, but with different
enthalpies and Ka).
◆ Possibly fit using several Independent models
• Cooperative? (two or more binding sites. The binding of the first ligand affects the
binding of succeeding ligands)
K=
H = 4x
6
K = 10
H = 3x
5
K = 10
H = 2x
4
K = 10
H = 1x
▪ Generally want to obtain stoichiometry (n), enthalpy (H) and binding constant (Ka) from one
experiment
▪ Enthalpy is directly measured. Receptor should be saturated with ligand at end of titration
▪ To obtain Ka: 10 < Ka[M]T < 1000
▪ [M]T is typically 10 – 100 mM, Ka is typically 103 to 109 M-1
▪ Reminder: KD = 1/Ka
TAINSTRUMENTS.COM
ITC Experimental Design
296
TAINSTRUMENTS.COM
ITC Experimental Design
• According to the ligand binding theory, the heat evolved for each injection depends on two variables, i.e. r and
X where r = Ka·[Mt] and X = [Lt]/[Mt].
• This left side of this equation (Independent model) can be plotted versus X for different r values, see the graph.
It can be observed that the shape of the binding curve is ‘nice’ for an r value of 50-100. For an unknown sample
it is important that the binding curves shows a good shape to obtain highest accuracy in the fitting procedure in
determining the reaction enthalpy and the r value. From the r value the equilibrium constant can be found since
r = Ka·[Mt]. In practice binding curves with a good shape is obtained for r values in the range: 10 < r < 1000.
1000 and
1
50
5
0.8
1 dq 1 1 + r ( X − 1)
= 1−
(1/VDH)dq/dL
V H dLt 2 (r − X r + 1)2 + 4 X r
0.6
1
0.4
0.2
0.1
TAINSTRUMENTS.COM
ITC Experimental Design
29
8
TAINSTRUMENTS.COM
Hints for ITC Experiments
29
9
TAINSTRUMENTS.COM
Hints for ITC Experiments
30
0
TAINSTRUMENTS.COM
Hints for ITC Experiments
TAINSTRUMENTS.COM
Hints for ITC Analysis
• Solvent protonation
▪Study different buffers at the same pH
▪Plot observed enthalpy versus ionization enthalpy
◆ Slope gives the number of protons released (if negative)
30
2
TAINSTRUMENTS.COM
Proton Linkage
•M + L + B ML + BH+; Hreaction
•M + L ML + H+; Hbinding
•B + H+ BH+; Hproton
200
-200
Hreaction
P/ nW
-400
-600
Hproton
-800
IGF-I to the soluble extracellular
-1.000 IGF-I receptor
0 50 100 150 200
Time/min
TAINSTRUMENTS.COM
DSC may be Helpful for ITC Results Correction
-1
0.75 mM
ligand there is, or the tighter it 0.15 mM
Excess Cp / kJ mol K
-1
50
binds, the more Tm increases. 1 mM
40
1.25 mM
▪ Can determine binding 30
0.075 mM
1.5 mM
constant at Tm. 20
0.05 mM
0 mM
▪ But, useful if very slow or very 10
24
18
12
0 20 40 60 Time,min
P/µW
P/µW
2 2
1 1
0 0
-1 -1
0 10 20 30 40 0 10 20 30 40
t/min t/min
30
7
TAINSTRUMENTS.COM
CMC of Sodium Dodecyl Sulfate (37 °C)
Accu m u la ted h ea t (m J )
-5
-10
Heat flow (µW)
-20
-10
-30
-15
0 50 100
Vo lu m e (µl)
-20
0 1 2 3 4 5 6
Time (hour)
TAINSTRUMENTS.COM
Need Assistance?
•Call TA Instruments
▪Phone: 302-427-4000 M-F 8-4:30 EST
•Check out our Website: www.tainstruments.com
TAINSTRUMENTS.COM
Thank You
TAINSTRUMENTS.COM