1720D/L Low Range Process Turbidimeter Manual
1720D/L Low Range Process Turbidimeter Manual
1720D/L Low Range Process Turbidimeter Manual
© Hach Company, 1999. All rights reserved. Printed in the U.S.A. dd/dp 5/99 1ed
Rev.1, 11/99
TRADEMARKS OF HACH COMPANY
2
TABLE OF CONTENTS
CERTIFICATION ............................................................................................................................................................ 6
SAFETY PRECAUTIONS .............................................................................................................................................. 8
SPECIFICATIONS........................................................................................................................................................... 9
OPERATION ............................................................................................................................................................... 11
3
TABLE OF CONTENTS, continued
INSTALLATION/MAINTENANCE ......................................................................................................................67
4
TABLE OF CONTENTS, continued
GLOSSARY ................................................................................................................................................................ 91
5
CERTIFICATION
Hach Company certifies this instrument was tested thoroughly, inspected and found to
meet its published specifications when it was shipped from the factory.
The 1720D/L Sensor has been tested and is certified as indicated to the following
instrumentation standards using a Hach AquaTrend® and a Hach PS1201 Power Supply.
Product compliance with these standards cannot be guaranteed if another power source is
used.
Product Safety
The 1720D/L Sensor was tested with a Hach AquaTrend Interface and PS 1201
Power Supply.
Immunity
The 1720D/L Sensor was tested with a Hach AquaTrend Interface and PS 1201
Power Supply
EN 61326: 1998 (Electrical Equipment for measurement, control and laboratory use-EMC
requirements) per 89/336/EEC EMC: Supporting test records by Hach Company, certified
compliance by Hach Company.
Standards include:
IEC 1000-4-2:1995 (EN 61000-4-2:1995) Electro-Static Discharge Immunity (Criteria B)
IEC 1000-4-3:1995 (EN 61000-4-3:1996) Radiated RF Electro-Magnetic Field Immunity
(Criteria A)
IEC 1000-4-4:1995 (EN 61000-4-5:1995) Electrical Fast Transients/Burst (Criteria B)
IEC 1000-4-5:1995 (EN 61000-4-5:1995) Surge (Criteria B)
IEC 1000-4-6:1996 (EN 61000-4-6:1996) Conducted Disturbances Induced by RF Fields
(Criteria A)
IEC 1000-4-11:1994 (EN 61000-4-11:1994) Voltage Dip/Short Interruptions (Criteria B)
Emissions
The 1720D/L Sensor was tested with a Hach AquaTrend Interface and PS 1201
Power Supply
Per 89/336/EEC EMC: EN 61326:1998 (Electrical Equipment for measurement, control and
laboratory use-EMC requirements) Class “A” emission limits. Supporting test records Hewlett
Packard, Fort Collins, Colorado Hardware Test Center (A2LA # 0905-01), certified
compliance by Hach Company.
Standards include:
EN 61000-3-2 Harmonic Disturbances Caused by Electrical Equipment
EN 61000-3-3 Voltage Fluctuation (Flicker) Disturbances Caused by Electrical Equipment
6
CERTIFICATION, continued
This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions:
(1) this device may not cause harmful interference, and (2) this device must
accept any interference received, including interference that may cause
undesired operation.
This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with
the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference, in which case the user will be required to correct the interference at
his own expense. The following techniques of reducing the interference problems
are applied easily.
1. Disconnect the 1720D/L Sensor from its power source to verify that it is or is
not the source of the interference.
2. If the AC power supply for the 1720D/L Sensor is connected into the same
outlet as the device with which it is interfering, try another outlet.
3. Move the 1720D/L Sensor away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the interference.
7
SAFETY PRECAUTIONS
8
SPECIFICATIONS
Specifications subject to change without notice.
Response Time: For a full scale step change, initial response* in 1 minute,
15 seconds. Varies with flow rate, see the table below.
Sample Light Source: 870 nm 0.2 watt (Optical Power) Class 1 LED Product
Alarms**: Two turbidity set-point alarms, each equipped with an SPDT relay
with unpowered contacts rated for 5A resistive load at 230 VAC.
9
SPECIFICATIONS, continued
Sample Inlet Fitting: ¼" NPT female, ¼" compression fitting (provided)
Dimensions:
Mounting:
Shipping Weight:
10
OPERATION
DANGER
Handling chemical samples, standards, and reagents can be dangerous. Review the necessary Material Safety
Data Sheets and become familiar with all safety procedures before handling any chemicals.
DANGER
La manipulation des échantillons chimiques, étalons et réactifs peut être dangereuse. Lire les Fiches de
Données de Sécurité des Produits (FDSP) et se familiariser avec toutes les procédures de sécurité avant de
manipuler tous les produits chimiques.
PELIGRO
La manipulación de muestras químicas, estándares y reactivos puede ser peligrosa. Revise las fichas
de seguridad de materiales y familiarícese con los procedimientos de seguridad antes de manipular
productos químicos.
GEFAHR
Da das Arbeiten mit chemischen Proben, Standards und Reagenzien mit Gefahren verbunden ist,
empfiehlt die Hach Company dem Benutzer dieser Produkte dringend, sich vor der Arbeit mit sicheren
Verfahrensweisen und dem richtigen Gebrauch der Chemikalien vertraut zu machen und alle entsprechenden
Materialsicherheitsdatenblätter aufmerksam zu lesen.
PERIGO
A manipulação de amostras, padrões e reagentes químicos pode ser perigosa. Reveja a folha dos dados de
segurança do material e familiarize-se com todos os procedimentos de segurança antes de manipular
quaisquer produtos químicos.
11
12
SECTION 1 GENERAL DESCRIPTION
13
SECTION 1, continued
If you have ordered only the 1720D/L Turbidimeter Sensor (Cat. No. 54300-00)
ensure the following items are included in your shipping carton:
14
SECTION 1, continued
Lower overall system costs result when customized systems are created using the
AquaTrend Interface, up to eight 1720D/L sensors, and any or all of the optional
equipment listed in the previous section. Multiple 1720D/L turbidimeters can be
powered from one PS1201 power supply, however, all turbidimeters must be
within 100 feet (approximately 30 meters) of the PS1201.
15
SECTION 1, continued
Table 1 presents the maximum number of devices that may be added to any one
master AquaTrend Interface.
Additional Hach Sensors, Digital Display Modules, Serial I/O Modules, Signal
Output Modules, Signal Input Modules, and Remote AquaTrend Interfaces can all
be considered additional devices when using this manual.
Custom design your system using Hach NetSketcher™ Software. Access the
software at www.hach.com.
16
SECTION 1, continued
DANGER PELIGRO
The 1720D/L Turbidimeter is not Turbidímetro 1720D/L não é concebido para uso com amostras que sejam inflamáveis
designed for use with samples that ou explos-ivas. Se qualquer solução que não seja de água se usar neste produto, dever-
are flammable or explosive in se-á ensaiar a compati-bilidade da amostra/produto para garantir segurança ao usuário
nature. If any sample solution e desempenho correto do produto.
other than water is used in this
product, test the sample/product PERIGO
compatibility to assure user safety El Turbidímetro 1720D/L no está diseñado para usarse con muestras de naturaleza
and proper product performance. inflamable o explosiva. Si se empleara en este producto alguna solución de muestra que
no fuera a base de agua, ponga a prueba la compatibilidad de la muestra/producto, para
cerciorarse de la seguridad y del correcto funcionamiento del producto.
DANGER
Le turbidimètre 1720D/L n'est pas prévu pour utilisation avec des échantillons de
nature inflammable ou explosive. Pour toute solution d'échantillon autre que de l'eau
utilisée avec cet appareil, tester la compatibilité échantillon/appareil pour assurer la
sécurité de l'utilisateur et le fonctionnement correct de l'appareil.
GEFAHR
Das Trübungsmessgerät 1720D/L darf nicht in Verbindung mit Proben benutzt werden,
die entflammbar oder explosiv sind. Wenn irgendeine andere Lösung als Wasser in
diesem Gerät analysiert werden soll, muß die Proben/Gerät-Kompatibilität getestet
werden, um die Sicherheitdes Benutzers und korrektes Arbeiten des Gerätes zu
gewährleisten.
17
SECTION 1, continued
The sample inlet port is fitted with a ¼" NPT to ¼" compression fitting. The drain
fitting is a ½" NPT to ½" ID tubing hose barb.
18
SECTION 1, continued
19
SECTION 1, continued
Easily alternate between displaying data in numeric form and graphic format by
pressing the DISPLAY key. Because of the automatic-ranging feature with
automatic decimal point positioning, no range selection is required. Indicators for
turbidity level alarm conditions and certain critical system malfunctions also are
shown on the AquaTrend Interface display screen.
20
SECTION 1, continued
21
SECTION 1, continued
The amount of light scattered is proportional to the turbidity of the sample. If the
turbidity of the sample is negligible, little light will be scattered and detected by
the photocell and the turbidity reading will be low. High turbidity, on the other
hand, will cause a high level of light scattering and result in a high reading.
Sample enters the turbidimeter body and flows through the baffle network of the
bubble trap. The flow allows bubbles to either cling to surfaces of the baffle
system or rise to the surface and vent to atmosphere. After traveling through the
bubble trap, sample enters the center column of the turbidimeter body,
rises into the measuring chamber and spills over the weir into the drain port.
Figure 5 presents a diagram of the sample flow path. A reading is taken every
three seconds.
22
SECTION 2 START UP
2.1 Introduction
This section provides instructions for initially placing the 1720D/L Turbidimeter
in operation or restarting the instrument after an extended shutdown. This
information is presented with the assumption that the user is controlling the
turbidimeter and peripheral devices with an AquaTrend® Interface. Some
familiarity with the AquaTrend Interface is also assumed; please read the
AquaTrend Interface Manual before beginning the installation.
1. Mount all devices to a pole, wall, or panel. (See Section 4.3 for turbidimeter
installation, specific instructions for the other devices are contained in
their manuals).
2. Make all wiring (power and network) connections (see the PS1201 Power
Supply Manual and the Startup Instruction Manual for 1720 Series
Turbidimeters and the AquaTrend Network System).
1. Review the AquaTrend Interface keypad and display (refer to the AquaTrend
Interface Manual).
3. Add devices to the network (refer to Section 2.4.1 and 2.4.2 and to the
AquaTrend Interface Manual).
23
SECTION 2, continued
ITEM DESCRIPTION
POWER LED Green LED lights when power is supplied to the unit.
Yellow LED blinks once every three seconds (corresponding to when a measurement is taken) in
STATUS LED
normal operation; the LED is constantly illuminated when a problem is detected by the microprocessor.
Yellow LED flashes once per second to indicate the device is not configured. The LED stops flashing
CONFIGURE LED
after the device is configured by the AquaTrend Interface.
Pressed to uniquely identify a sensor when more than one sensor is added from the AquaTrend
CONFIGURE BUTTON
Interface Add Device Menu. The configure LED is fully illuminated while this button is held down.
REMOTE A Digital Display Module (DDM) or Portable AquaTrend Interface may be connected here.
There are two procedures for adding device(s) to the network. They are Single
Sensor Systems (Section 2.4.1) and Multiple Sensor Systems (Section 2.4.2).
24
SECTION 2, continued
Each connected sensor must be added to the network. When power is initially
supplied to the new device, the configure LED should flash once per second. If
this is not the case, the device may already be configured. Refer to Section 2.11
and to Section 6.6 to unconfigure the sensor before re-installing it on the system.
During the following operation, an electronic signal to configure the sensor is sent
to the AquaTrend Interface. The configure light on the sensor will illuminate and
remain lit until the configuration is complete. Up to eight sensors may be added to
each AquaTrend Interface. If eight sensors have already been added, you will not
be able to add another sensor.
1. To add a device to the AquaTrend Interface network, first press the MENU
button to access the MAIN MENU.
2. Press the down ARROW key to move the pointer to the NETWORK MENU
option and press ENTER.
3. Select ADD DEVICE and press ENTER. If necessary, the display shows all
device categories currently on the system. Table 2 below shows device
categories. Refer to Table 1 on page 16 for information on the maximum
number of devices of each type allowed on the system.
4. Choose the device category and press ENTER.
25
SECTION 2, continued
Table 2 AquaTrend Interface Device Categories
Category Devices
SENSOR 1720D/L Turbidimeter, APA 6000, or other Hach Sensor
COMMUNICATIONS Serial I/O Module
OUTPUTS Signal Output Module
AQUATREND INTERFACE AquaTrend Interface or Remote AquaTrend Interface
DISPLAY Digital Display Module
5. If more than one unconfigured sensor is present on the network, a pop-up box
will appear asking the user to press the CONFIGURE button on the sensor
being added. Press the CONFIGURE button to continue with this procedure.
This button is located on the front face of the 1720D/L sensor head.
6. When the correct device is identified, select ADD DEVICE and press ENTER.
After a pause (approximately 2 seconds), the display will prompt for the entry
of a device name.
9. Press the MENU key to return to the MAIN MENU. You must now connect
the turbidity measurement to a channel before the measurement will be
shown on the AquaTrend display. See Section 10..
26
SECTION 2, continued
2. Move the pointer to AQUATREND MENU, then press the ENTER key.
6. Select ATTACH from the resulting pop-up screen and press the ENTER key.
7. Select the channel to which the measurement is to be attached and press the
ENTER key.
27
SECTION 2, continued
The 1720D/L Turbidimeter Sensor setup menu contains calibration, alarm setup,
recorder setup, and reading and diagnostic options. Move the pointer with the up
and down ARROW keys, then select the option by pressing the ENTER key.
Refer to Section 3.1 for detailed calibration information and to the sections below
for information on alarms, recorder outputs, measurement options,
and sensor diagnostics.
A selectable hysteresis setting also prevents alarms from cycling on and off when
the turbidity level is close to the setpoint.
• A history of the last 14 alarms or sensor warnings will be shown in the sensor
alarm log. For a description of the alarm log, see Section 2.8.
28
SECTION 2, continued
To access the alarm setup menu for alarm 1 or 2, perform the following steps:
1. Press the MENU key to access the MAIN MENU. Move the pointer to
SENSOR MENU and press the ENTER key.
2. Use the ARROW keys to select the appropriate sensor then press the
ENTER key.
3. Move the pointer to ALARM SETUP and press the ENTER key.
4. Select the appropriate alarm number (1 or 2) with the up and down ARROW
keys, and press the ENTER key.
5. The current settings of alarm, type, hysteresis, and set point will be displayed.
3. A popup window will appear with the pointer set to the current alarm type.
Use the up and down ARROW keys to select LOW, HIGH or SENSOR
WARNING. Press the ENTER key to accept the new type.
29
SECTION 2, continued
To change the alarm hysteresis setting for either alarm, perform the
following steps:
1. Access the ALARM SETUP menu and specific alarm as in Section 2.6.1.
3. A popup window will appear with the pointer set to current alarm hysteresis.
Use the up and down ARROW keys to select 2.5%, 5% or 10%.
Press the ENTER key to accept the new setting.
To change the alarm set point for either alarm, perform the following steps:
1. Access the ALARM SETUP menu as in Section 2.6.1, then perform the
following steps:
2. Move the pointer to SET POINT, and press the ENTER key.
3. A popup window will appear showing the current alarm set point.
Use the ARROW keys to change the value (up and down arrow to increase or
decrease the digit, right and left arrow to move to the next digit). Press the
ENTER key to accept the new set point.
30
SECTION 2, continued
3. Move the pointer to ALARM TO RELAY and press the ENTER key.
5. Select the sensor with the alarm of interest and press the ENTER key.
7. Select the name of the SOM containing the physical relay and press the
ENTER key.
For more information on attaching alarms to relays, refer to the Signal Output
Module (SOM) Instruction Manual (Cat. No. 51250-18).
31
SECTION 2, continued
2. Move the pointer to SENSOR MENU and press the ENTER key.
4. Move the pointer to HOLD OUTPUTS and press the ENTER key.
• To set the Hold Outputs timer to 30 minutes, press the ENTER key with the
pointer on SET TO 30 MINUTES.
• To reset the timer to zero (0) ending the hold output mode, move the
pointer to END HOLD OUTPUTS.
• To return to the Sensor Menu without changing the remaining displayed
time, press the EXIT key.
32
SECTION 2, continued
1. Press the MENU key to access the MAIN MENU. Move the pointer to SENSOR
MENU and press the ENTER key.
3. Move the pointer to SILENCE ALARMS and press the ENTER key.
4. The display shows SILENCE ALARMS and TIME REMAINING. Zero (0)
minutes indicates the alarms are not being silenced.
• To set the SILENCE ALARMS timer to 30 minutes, press the ENTER key
with the pointer on SET TO 30 MINUTES.
• To reset the timer to zero (0) ending the Silence Alarms mode, move the
pointer to END SILENCE ALARMS.
• To return to the SENSOR MENU without changing the remaining
displayed time, press the EXIT key.
33
SECTION 2, continued
To review the entries in the 1720D/L alarm log, perform the following steps:
2. Move the pointer to ALARM LOG and press the ENTER key.
A list of the 1720D/L alarm log entries is displayed. For more information about a
particular alarm log entry, select the entry of interest and press the ENTER key.
Up to 14 entries can be stored in the alarm log. The latest entry is shown at the top
of the list. If more than 14 entries are stored, the newest entry will appear at the
top of the list and the oldest entry will be deleted. Any active sensor warnings are
automatically acknowledged (the number is removed from the lower portion of
the screen) when the sensor alarm log is first displayed.
The alarm log is maintained across power outages, so alarm events before a power
outage can be viewed after power is restored.
34
SECTION 2, continued
2.8.2 Power Up
A single POWER UP entry appears in the sensor alarm log. POWER UP is always
the first entry in the alarm log, regardless of the time/date power was restored to
the 1720D/L. No other occurrences of POWER UP appear in the alarm log.
Note: If the 1720D/L does not Each time power is applied to the1720D/L, or the 1720D/L is reset, the alarm log
receive the time/date from the records a POWER UP entry. The time and date of the power up event are also
Master AquaTrend, the power up displayed. The alarm log is maintained across power outages, so alarm events
time is displayed as 01/01/00 00:00.
before a power outage can be viewed after power is restored.
The sensor warnings that may be displayed are: A/D TIMEOUT, LIGHT LEAK,
BAD LAMP, and LOW SIGNAL. See Section 6.4 for more information on each
type of sensor warning. Sensor warnings are automatically acknowledged (the
number is removed from the lower portion of the screen) when the sensor alarm
log is first displayed.
35
SECTION 2, continued
1. Press the MENU key to access the MAIN MENU. Move the pointer to
SENSOR MENU and press the ENTER key.
5. Move the pointer to YES, and press ENTER. Reviewing the alarm log will
show NO ALARM ENTRIES until an alarm, power up, or sensor
warning occurs.
After recorders are set up, they must be attached to an analog output in an SOM as
described in Section 2.9.3. Other options, such as 0-20mA or 4-20mA selection,
loss of communication action, setting of analog output trim min/max, and analog
output tests are described in the Signal Output Module (SOM) Instruction Manual
(Cat. No. 51250-18).
36
SECTION 2, continued
If the minimum value is set greater than the maximum value, the output will invert
so that as turbidity levels increase, recorder mA values decrease. This is useful
when devices require control inversely proportional to turbidity.
2. Move the pointer to SENSOR MENU and press the ENTER key.
4. Move the pointer to RECORDER SETUP and press the ENTER key.
5. Select the appropriate recorder number (1 or 2) with the up and down ARROW
keys, and press the ENTER key.
7. A popup window will appear showing the current recorder minimum. Use the
up and down ARROW keys to change the value. Press the ENTER key to accept
the new minimum value.
37
SECTION 2, continued
If the minimum value is set greater than the maximum value, the output will invert
so that as turbidity levels increase, recorder mA values decrease. This is useful
when devices require control inversely proportional to turbidity.
1. Press the MENU key to access the MAIN MENU. Move the pointer to
SENSOR MENU and press the ENTER key.
3. Move the pointer to RECORDER SETUP and press the ENTER key.
4. Select the appropriate recorder number (1 or 2) with the up and down ARROW
keys, and press the ENTER key.
6. A popup window will appear showing the current recorder maximum. Use the
up and down ARROW keys to change the value. Press the ENTER key to accept
the new maximum value.
38
SECTION 2, continued
5. Select the name of the sensor to attach and press the ENTER key.
6. Select the appropriate sensor recorder number (1 or 2) with the up and down
ARROW keys, and press the ENTER key.
7. Select the name of the SOM containing the physical analog output and press
the ENTER key.
For more information on attaching recorders to analog outputs, see the Signal
Output Module (SOM) Manual, Cat. No. 51250-18.
39
SECTION 2, continued
2. Move the pointer to SENSOR MENU and press the ENTER key.
4. Move the pointer to MEASUREMENT OPTIONS and press the ENTER key.
5. Move the pointer to the option to be accessed and press the ENTER key.
Selecting the bubble reject mode will eliminate any momentary spikes in the
reading and yield a stable output. Bubble reject mode is recommended for normal
operation. The default setting is ON. To change the Bubble Rejection Mode,
proceed as follows:
2. Determine if the bubble reject mode is on or off by moving the pointer to the
BUBBLE REJECT line. If bubble reject is off, the display will show Off;
if it is on, the display will show On.
3. Press the ENTER key to move to the selection display. Change the displayed
mode by using the ARROW keys to toggle between the two options,
then press ENTER to accept.
40
SECTION 2, continued
3. Move the pointer to the appropriate signal averaging time and accept it with
the ENTER key. The selection will be retained until changed.
3. Select a print interval in the resulting pop-up box by moving the pointer to the
desired setting and then pressing the ENTER key to accept. The default setting
is one (1) minute. Selecting OFF disables the print interval feature.
41
SECTION 2, continued
A sample 4-hour printout from four 1720D/Ls (via an SIO) is shown below.
If two SIOs are being used, the information will be sent to both at the same
print interval.
2.10.4 Units
The units displayed can be changed using the steps below. All units are 1.0
multipliers of NTU, so a calibration can be performed in any of the reading units.
The mg/L units are intended for use with Kaolin Clay for reporting regulations
in Japan.
2. Select UNITS and press ENTER. Units of NTU, FNU, FTU, TE/F, and mg/L
are displayed.
42
SECTION 2, continued
When the time stamp feature is set to OFF, only the turbidity readings are printed
as shown below:
0.035
0.900
0.035
3. Move the pointer to the appropriate time stamp setting and accept it with the
ENTER key.
43
SECTION 2, continued
2. Move the pointer to the NETWORK MENU option and press the ENTER key.
3. Move the pointer to REMOVE DEVICE and press the ENTER key. After
choosing REMOVE DEVICE, the display may prompt for the selection of a
device type. All types of connected devices will be shown on the display.
4. If necessary, move the pointer to the device type that is to be removed and
press ENTER. All connected devices of that type will be displayed.
5. If necessary, move the pointer to the specific device to be removed and press
ENTER. A pop-up box will appear, so you can verify that this is the device you
want to remove.
6. Press ENTER to remove the device or move the pointer to CANCEL and press
ENTER if the displayed device is not the one to be removed.
44
SECTION 2, continued
2. Make all power and network wiring connections to the new device as if it
will be added to the system in the normal manner. See the PS1201 Manual
for instructions.
3. Supply power to the system and press the MENU key to access the
MAIN MENU.
4. Move the pointer to the NETWORK MENU option and press the ENTER key.
This means that although the AquaTrend Interface network connections of the old
device will be transferred to the new device, you will still need to set device
specific information.
45
SECTION 2, continued
6. Press ENTER to continue or, if you decide not to replace the device, move the
pointer to CANCEL and press the ENTER key.
If you chose CONTINUE, all settings accessed from menus controlled by that
device must be redone. This includes all options shown under the SENSOR
MENU. Settings accessed via the NETWORK MENU will remain.
All types of connected devices will be shown on the display. Move the pointer
to the device type that is to be replaced and press ENTER. All connected devices of
that type will be displayed. Move the pointer to the device to be replaced and
press ENTER.
A pop-up box will appear, confirming the device has been replaced.
2. Move the pointer to SENSOR MENU and press the ENTER key.
5. Use the ARROW keys (right or left arrows to move to another character space
in the name and up or down arrows to scroll through the available characters)
to create a new name (up to 12 characters). Press ENTER to accept the
displayed name.
46
SECTION 3 CALIBRATION
The 1720D/L Turbidimeter is factory-calibrated using StablCal® Stabilized
Formazin before shipment. The instrument must be recalibrated before use to
meet published accuracy specifications. In addition, recalibration is recommended
after any significant maintenance or repair and at least once every three months
during normal operation. The turbidimeter body and bubble trap must be
thoroughly cleaned before initial use and prior to each calibration.
See Section 5.1.3 on page 76. The sections that follow give instrument calibration
and verification information. In addition, details on preparing standards and using
StablCal Stabilized Formazin are presented.
Note: Performance Specifications Similarly, two calibration curve verification methods are presented in this manual,
are based on formazin or stabilized as described Table 4 (“A” denotes the best in each category).
formazin standards.
47
SECTION 3, continued
Note: Published specifications are For regulatory reporting, the user-prepared formazin or StablCal Stabilized
guaranteed only when calibration is Formazin calibration is recommended. The StablCal calibration is the simplest as
performed with user-prepared it involves the use of only one standard (1.00 or 20.0 NTU). The comparative
formazin or StablCal Stabilized
Formazin Standards.
calibration, although EPA-approved, is very susceptible to error and is not
recommended.
• Always clean the photocell window per the instructions in Section 5.1.2
Cleaning the LED Module and Photocell Window. Rinse the photocell with
deionized water and dry with a soft, lint-free cloth before calibrating.
• Always clean the turbidimeter body and calibration cylinder per the
instructions in Section 5.1.3 Cleaning the Turbidimeter Body and Bubble
Trap.
• Rinse the calibration cylinder and turbidimeter body with deionized water
immediately before use.
• Do not stir or mix dilution water. Noisy (fluctuating) readings will result.
• Gently invert StablCal standards for 1 minute before opening. Do not shake.
This ensures a consistent turbidity of the standard.
• If the 20.0 NTU standard is allowed to sit in the Calibration Cylinder for more
than 15 minutes, it must be remixed (gently swirled in the calibration
cylinder) before use to ensure a consistent turbidity.
• Discard all standards after use per the instructions on the container.
Never transfer the standard back into its original container.
Contamination will result.
48
SECTION 3, continued
1. Press the MENU key to access the MAIN MENU. Select SENSOR MENU and
press the ENTER key.
2. Select the sensor to be calibrated using the arrow keys and press the
ENTER key.
4. Move the pointer to ZERO ELECTRONICS and press the ENTER key.
5. Make sure the turbidimeter head is aligned on the calibration cylinder and
press the ENTER key to continue. The display will show PLEASE WAIT for
approximately 30 seconds while dark values are read. Do not remove the
turbidimeter head during this waiting period. If the head is removed while
PLEASE WAIT is shown on the display, repeat this step.
49
SECTION 3, continued
2. Press the MENU key to access the MAIN MENU. Select SENSOR MENU and
press the ENTER key.
3. Select the sensor to be calibrated using the arrow keys and press ENTER.
8. Remove the head assembly from the turbidimeter body. Clean the photocell
window per the instructions in Section 5.1.2 on page 75. Check the lamp and
lens for scaling or filming, and clean them if necessary (see Section 5.1
Scheduled Maintenance).
50
SECTION 3, continued
9. Place the head assembly on the calibration cylinder. Gently tap the cylinder to
free any bubbles adhering to the photocell window.
11. After waiting a minimum of one minute, press ENTER to accept the reading of
the StablCal standard. The displayed value is based on the previous
calibration, and will be updated after the calibration is accepted. The number
is displayed for stability purposes only.
12. If the calibration was performed correctly, the display will show
SUCCESSFUL CAL. Press ENTER to accept the calibration.
13. Input three alpha characters (such as initials) to complete the required
information and store the calibration in memory. If an improper calibration is
detected, the message QUESTIONABLE CAL will be displayed. See
Questionable Cal on page 80 in this manual for more information.
Note: The Hold Outputs feature is automatically set to 5 minutes after a calibration is
accepted or aborted, to allow time for the turbidimeter head to be placed on the body.
To cancel Hold Outputs immediately, see Section 2.6.6 Hold Outputs.
14. Replace the instrument head assembly on the turbidimeter body and press the
DISPLAY key. The instrument is now calibrated.
15. Discard the StablCal standard according to the directions on the standard
bottle. Rinse the calibration cylinder with deionized water and store upside
down to minimize dust buildup.
51
SECTION 3, continued
2. Press the MENU key to access the MAIN MENU. Select SENSOR MENU and
press the ENTER key.
3. Select the sensor to be calibrated using the arrow keys and press ENTER.
7. Remove the head assembly from the turbidimeter body. Clean the photocell
window per the instructions in Section 5.1.2 on page 75. Check the lamp and
lens for scaling or filming, and clean them if necessary (see Section 5.1
Scheduled Maintenance).
8. Place the head assembly on the calibration cylinder. Gently tap the cylinder to
free any bubbles adhering to the photocell window.
10. After waiting a minimum of one minute, press ENTER to accept the reading of
the dilution water. The value displayed is based on the previous calibration,
and will be updated after the calibration is accepted. The number is displayed
for stability purposes only.
52
SECTION 3, continued
11. The display will prompt DISPENSE 5 mL OF 4000 FORMAZIN INTO CAL
CYLINDER AND SWIRL. Attach a clean, dry pipet tip to the end of a
TenSette Pipet. Set the dispenser knob to the 5.0-mL position. Gently invert
the 4000-NTU formazin bottle 50 times to mix the solution. Depress the pipet
plunger and immerse the end of the tip one-half inch into the 4000-NTU
formazin. Slowly release the plunger to draw 5.0 mL of 4000-NTU formazin
into the tip.
12. Set the pipet knob to the 6-mL position to ensure all the 4000-NTU formazin
is dispensed. Lift the turbidimeter head assembly out of the calibration
cylinder and smoothly depress the pipet plunger to dispense the 4000-NTU
formazin into the calibration cylinder.
13. Gently swirl the formazin solution in the cylinder and replace the
head assembly.
15. After waiting a minimum of one minute, press ENTER to accept the value of
the 20-NTU standard. The value displayed is based on the previous
calibration, and will be updated after the calibration is accepted. The number
is displayed only for stability purposes.
16. If the calibration was performed correctly, the display will show
SUCCESSFUL CAL. Press ENTER to accept the calibration.
17. Input three alpha characters (such as initials) to complete the required
information and store the calibration in memory. If an improper calibration is
detected, the message QUESTIONABLE CAL will be displayed. See
Questionable Cal on page 80 in this manual for more information.
18. Replace the instrument head assembly on the turbidimeter body and press the
DISPLAY key. The instrument is now calibrated.
19. Discard the formazin standard according to the directions on the standard
bottle. Rinse the calibration cylinder with deionized water and store upside
down to minimize dust buildup.
Note: The Hold Outputs feature is automatically set to 5 minutes after a calibration is
accepted or aborted to allow time for the turbidimeter head to be placed on the body.
To cancel Hold Outputs immediately, see Section 2.6.6 Hold Outputs.
53
SECTION 3, continued
Three ICE-PIC/870 modules are available for use with the 1720D/L, however,
only the 20.0-NTU ICE-PIC/870 module is recommended for calibration where
specified accuracy is required across the full operating range of the instrument.
See REPLACEMENT PARTS AND REAGENTS on page 95.
2. Press the MENU key to access the MAIN MENU. Select SENSOR MENU and
press the ENTER key.
3. Select the sensor to be calibrated using the arrow keys and press ENTER.
54
SECTION 3, continued
8. Remove the protective cover from the ICE-PIC/870 module. Do not touch any
of the optical components of the ICE-PIC/870 module, handle the device by
the edges only. Place the ICE-PIC/870 module into the turbidimeter body with
the alignment tabs of the ICE-PIC/870 module resting on the alignment pins
of the turbidimeter body.
11. After waiting a minimum of one minute, press ENTER to accept the reading of
the ICE-PIC/870 module. The value displayed is based on the previous
calibration, and will be updated after the calibration is accepted. The number
is displayed for stability purposes only.
12. If the calibration was performed correctly, the display will show
SUCCESSFUL CAL. Press ENTER to accept the calibration.
13. Input three alpha characters (such as initials) to complete the required
information and store the calibration in memory. If an improper calibration is
detected, the message QUESTIONABLE CAL will be displayed. See
Questionable Cal on page 80 in this manual for more information.
Note: The Hold Outputs feature is automatically set to 5 minutes after a calibration is
accepted or aborted, to allow time for the turbidimeter head to be placed on the body.
To cancel Hold Outputs immediately, see Section 2.6.6 Hold Outputs.
14. Remove the turbidimeter head from the ICE-PIC/870 module and
immediately replace the protective cover on the ICE-PIC/870 module.
15. Replace the instrument head assembly on the turbidimeter body and press the
DISPLAY key. The instrument is now calibrated.
55
SECTION 3, continued
The accuracy of this method is valid only near the calibration point (the value read
by the laboratory turbidimeter). The specified accuracy of the 1720D/L may not
be valid over the full operating range when this calibration method is used.
6. Turn off the flow to the 1720D/L turbidimeter. Take care to avoid introducing
air bubbles into the sample.
8. Immediately measure the turbidity in the laboratory instrument and record the
stable reading. If the 1720D/L reading differs from the laboratory
turbidimeter reading by more than 5%, continue with the remaining steps of
this procedure.
9. Press the MENU key to access the MAIN MENU. Select SENSOR MENU and
press the ENTER key.
56
SECTION 3, continued
10. Select the sensor to be calibrated using the arrow keys and press ENTER.
11. Select CALIBRATION from the menu choices and press ENTER.
12. Move the pointer to STABLCAL/COMPARATIVE CAL and press ENTER to
begin the calibration procedure. During calibration, the instrument will
automatically switch to a 6-second signal average. After calibration, the signal
average will revert to the value set prior to calibration.
14. The display will prompt PLACE TURB HEAD ON CAL CYLINDER OR
BODY. Ensure the 1720D/L head is placed on the turbidimeter body.
16. After waiting a minimum of one minute, press ENTER to accept the reading.
The value displayed is based on the previous calibration, and will be updated
after the calibration is accepted. The number is displayed for stability
purposes only.
17. If the calibration was performed correctly, the display will show
SUCCESSFUL CAL. Press ENTER to accept the calibration.
18. Input three alpha characters (such as initials) to complete the required
information and store the calibration in memory. If an improper calibration is
detected, the message QUESTIONABLE CAL will be displayed. See
Questionable Cal on page 80 in this manual for more information.Press the
DISPLAY key. The instrument is now calibrated.
Note: The Hold Outputs feature is automatically set to 5 minutes after a calibration is
accepted or aborted, to allow time for the turbidimeter head to be placed on the body.
To cancel Hold Outputs immediately, see Section 2.6.6 Hold Outputs.
57
SECTION 3, continued
3. Repeat step 2 for each additional turbidimeter (up to 8 total). If more than
8 turbidimeters are being calibrated, discard the StablCal standard after
8 turbidimeter calibrations according to the directions on the standard bottle.
Rinse the calibration cylinder with deionized water and dry with a lint free
cloth. Gently mix and pour a new 1-liter bottle of 20.0-NTU StablCal standard
solution for the next eight turbidimeters.
4. After all calibrations are complete, discard the StablCal standard according to
the directions on the standard bottle. Rinse the calibration cylinder with
deionized water and store upside down to minimize dust buildup.
58
SECTION 3, continued
Use StablCal Standards for this procedure. Do not use other turbidity standards as
they may not yield accurate results due to the nature of the materials used in the
manufacture of these products.
2. Place a container underneath the turbidimeter body. Remove the drain plug
from the bottom of the body (release the lock on the bottom of the plug and
move the plug back and forth to ease it out) and drain the sample.
3. Perform the cleaning procedure specified in Section 5.1.3 on page 76. If the
turbidimeter body is not clean this verification will not work.
5. Gently invert a 1-liter bottle of 1.0 NTU StablCal Standard 50 times. Invert
carefully to avoid creating bubbles in the standard. Do not shake!
6. Carefully pour just enough of the 1.0 NTU StablCal Standard into the
turbidimeter body to bring the level up to the overflow weir (inside the
turbidimeter body). This will equal approximately one liter.
7. Check the photocell for film or debris. Clean if necessary (see Section 5.1.3
on page 76).
8. Place the turbidimeter head on the body. Make sure the head is seated
properly by ensuring it is lined up with the bubble trap cover.
9. Wait approximately 3-10 minutes for any bubbles to dissipate and the reading
to stabilize.
10. The displayed value should be 1.00 ±0.05 NTU (0.95-1.05 NTU).
Note: The accuracy of the StablCal Standard is ±0.05 NTU (0.95-1.05 NTU).
11. If the reading is not 1.00 ±0.05 NTU (0.95-1.05 NTU), recalibrate
the instrument.
59
SECTION 3, continued
Three ICE-PIC/870 modules are available, however only the 20.0-NTU ICE-PIC/
870 module is recommended for verification of the 1720D/L turbidimeter.
See REPLACEMENT PARTS AND REAGENTS on page 95.
1. Activate the Hold Outputs feature to hold attached SOM outputs at their
pre-verification settings. See Section 2.6.6 Hold Outputs on page 32.
3. Use a soft, lint-free cloth to remove ALL moisture from the photodetector and
the turbidimeter head. Do not allow any moisture to remain on the
photodetector or turbidimeter head. Moisture introduced into the components
of the ICE-PIC/870 module will negatively affect performance. The module
will not operate properly if the optical components have scratches, smudges,
water drops, or other debris on them.
4. Place the ICE-PIC/870 module into the turbidimeter body with the alignment
tabs of the ICE-PIC/870 module resting on the alignment pins of the
turbidimeter body.
5. Remove the protective cover from the ICE-PIC/870 module. Do not touch any
of the optical components of the ICE-PIC/870 module. Handle the device by
the edges only.
12. Deactivate the Hold Outputs feature. See Section 2.6.6 Hold Outputs on
page 32.
60
SECTION 3, continued
To view the details of the last eight calibrations, perform the following steps:
1. Press the MENU key to access the MAIN MENU. Select SENSOR MENU and
press the ENTER key.
2. Select the desired sensor using the arrow keys and press the ENTER key.
3. Select CALIBRATION from the menu choices and press ENTER.
4. Move the pointer to CAL HISTORY and press ENTER. The details of the latest
calibration are displayed.
5. If the details of another calibration are available, the prompt shows VIEW
PREVIOUS CAL. Press ENTER to view the previous calibration. Repeat this
step until all calibration details have been viewed, or press EXIT to return to
the calibration menu.
61
SECTION 3, continued
To print the details of the last 8 calibrations, perform the following steps:
1. Press the MENU key to access the MAIN MENU. Select SENSOR MENU and
press the ENTER key.
2. Select the desired sensor using the arrow keys and press the ENTER key.
4. Move the pointer to PRINT CAL HISTORY and press ENTER. The display
shows PLEASE WAIT while the calibration history log is printed.
62
SECTION 3, continued
The following screen is displayed in the calibration history log after a restore
default gain operation has been performed.
ATTENTION
Pour se familiariser avec les précautions à prendre lors de la manipulation, les dangers
et les procédures d'urgence, toujours lire les Fiches de Données de Sécurite de
manipuler les récipients, les réservoirs et les systèmes de distribution contenant les
réactifs chimiques et les solutions étalons. Il est toujours recommandé de porter des
lunettes de protection lorsqu'un contact avec les produits chimiques est possible.
WARNHINWEIS
Es wird dringend empfohlen, die Sicherheitsdatenblätter vor der Handhabung von
Behältern, Tanks und Zufuhrsystemen, die chemische Reagenzien und
Standardsubstanzen enthalten, aufmerksam durchzulesen, damit Sie sich mit den beim
Umgang mit diesen Chemikalien notwendigen Vorsichtsmaßnahmen, Risiken und
Notfallschutzmaßnahmen vertraut machen, Es wird empfohlen, in allen Situationen, in
denen mit einem Kontakt mit Chemikalien zu rechnen ist, eine Schutzbrille zu tragen.
63
SECTION 3, continued
Formazin is the recognized primary standard for turbidity. Formazin standards can
be prepared at turbidity values up to 4000 NTU. A 4000 NTU formazin stock
solution is available and can be diluted to the appropriate NTU values, see Section
3.6.2 Preparing Dilute Formazin Standards for dilution instructions.
Reagents and apparatus are available from Hach. Obtain descriptions and catalog
numbers from REPLACEMENT PARTS AND REAGENTS on page 95.
5. Allow the solution to stand for 48 hours at 25 ±1 °C (68 to 72 °F). During this
time, the white polymer suspension will develop. The resulting standard is
4000 NTU.
6. Immediately before dilution, invert the flask containing the stock suspension
to mix.
7. Dilute the stock suspension before use with ultra-filtered deionized water to
achieve a standard of the desired NTU value.
64
SECTION 3, continued
Example:
To prepare a 20.0 NTU standard, use a Class A pipet to measure 5.00 mL of
4000 NTU stock solution into a 1-liter flask. Fill to the mark with ultra-filtered
dilution water. Stopper the flask and invert several times to mix.
Notes
• Formazin standards at 2 NTU are stable for only a short time. Prepare
standards immediately before use and discard after use. These standards
cannot be stored for long-term use.
65
66
INSTALLATION/MAINTENANCE
Some of the following manual sections contain information in the form of warnings, cautions
and notes that require special attention. Read and follow these instructions carefully to avoid
personal injury and damage to the instrument. Only personnel qualified to do so, should
conduct the installation/maintenance tasks described in this portion of the manual.
Certains des chapitres suivants de ce mode d’emploi contiennent des informations sous
la forme d’avertissements, messages de prudence et notes qui demandent une attention
particulière. Lire et suivre ces instructions attentivement pour éviter les risques de blessures
des personnes et de détoriation de l’appareil. Les tâches d’installation et d’entretien décrites
dans cette partie du mode d’emploi doivent être seulement effectuées par le personnel qualifié
pour le faire.
Algunos de los capítulos del manual que presentamos contienen información muy importante
en forma de alertas, notas y precauciones a tomar. Lea y siga cuidadosamente estas
instrucciones a fin de evitar accidentes personales y daños al instrumento. Las tareas
de instalación y mantenimiento descritas en la presente sección deberán ser efectuadas
únicamente por personas debidamente cualificadas.
Einige der folgenden Abschnitte dieses Handbuchs enthalten Informationen in Form von
Warnungen, Vorsichtsmaßnahmen oder Anmerkungen, die besonders beachtet werden
müssen. Lesen und befolgen Sie diese Instruktionen aufmerksam, um Verletzungen von
Personen oder Schäden am Gerät zu vermeiden. In diesem Abschnitt beschriebene
Installations- und Wartungsaufgaben dürfen nur von qualifiziertem Personal
durchgeführt werden.
67
68
SECTION 4 INSTALLATION
DANGER PELIGRO
This instrument should be Este instrumento debe ser instalado por personal técnico capacitado para asegurar elo
installed by qualified technical cumplimiento con todos los códigos eléctricos y de plomería aplicables.
personnel to ensure adherence to
all applicable electrical and PERIGO
plumbing codes. Este instrumento deve ser instalado por pessoal técnico qualificado para asergurar o
cumprimento de todas as normas elétricas e de canalização aplicáveis.
DANGER
Cet appareil doit être installé par du personnel technique qualifié, afin d'assurer le
repect de toutes les normes applicables d'électricité dt de plomberie.
GEFAHR
Um zu gewährleisten, daß alle elektrischen und sanitärinstallationstechnischen
VDE-Vorschriften und gegebenefalls die Zusatzvorschriften der zuständigen
Elektrizitäts- und Wasserwerke erfüllt werden, darf dieses Gerät nur von geschultem
Fachpersonal installiert werden.
The AquaTrend® Interface Module and PS1201 Power Supply enclosures are
designed for indoor installation. Ambient temperatures may range from 0 to 40 °C
(32 to l04 °F) but best performance will result if temperatures do not
change rapidly.
69
SECTION 4, continued
70
SECTION 4, continued
Clean the turbidimeter body and bubble trap before mounting using the
instructions supplied in Section 5.1.3 Cleaning the Turbidimeter Body and Bubble
Trap. Slotted mounting brackets are integral parts of the turbidimeter body.
Hardware to mount the body is not supplied with the instrument. Allow at least
22 cm (approximately 10 inches) clearance for removal of the head assembly and
bubble trap cover from the top of the turbidimeter body.
Leave enough room below the turbidimeter body to remove the bottom plug and
to place a container under the drain when calibrating or cleaning the turbidimeter.
Note: Make sure the top of the Install two ¼-20 bolts 10-¾ inches apart (on center), see Figure 7. Leave at least
turbidimeter body is level. ¼-inch of the bolt head exposed. Make sure these bolts are installed level. Hang
the slotted mounting brackets of the turbidimeter body onto the bolts.
The rear portion of the head assembly has a molded “lip” which may be used to
hang the head assembly on the turbidimeter body edge for routine maintenance.
See Figure 3.
71
SECTION 4, continued
DANGER
Le turbidimètre 1720D/L n'est pas prévu pour utilisation avec des échantillons de
nature inflammable ou explosive. Pour toute solution d'échantillon autre que de l'eau
utilisée avec cet appareil, tester la compatibilité échantillon/appareil pour assurer la
sécurité de l'utilisateur et le fonctionnement correct de l'appareil.
GEFAHR
Das Trübungsmeßgerät 1720D/L darf nicht in Verbindung mit Proben benutzt werden,
die entflammbar oder explosiv sind. Wenn irgendeine andere Lösung als Wasser in dem
Gerät analysiert werden soll, muß die Proben/Gerät-Kompabilität getestet werden, um
die Sicherheites Benutzers und korrektes Arbeiten des Gerätes zu gewährleisten.
Note: When setting the flow rate, One-fourth inch O.D. rigid or semi-rigid tubing is recommended for sample lines.
take care to avoid sweeping air Run them as directly as possible between the turbidimeter body and the sampling
“micro-bubbles” through the internal point to minimize sample flow lag time.
bubble trap. Observe the sample
flow inside the turbidimeter body. If
Install sample line taps into larger process pipes to minimize interference from
small air bubbles can be seen
flowing up through the center, air bubbles or pipeline bottom sediment. A tap projecting into the center of the
reduce the flow rate. pipe is ideal. Figure 8 shows examples of good and poor methods of installing a
sample tap.
72
SECTION 4, continued
The required flow rate is 250-750 mL/minute (4.0 to 11.9 gal/hour). Flow rate into
the turbidimeter may be controlled with a flow restriction device on the inlet line.
Flow rates below 250 mL/min will reduce response time and cause inaccurate
readings. Flow rates above 750 mL/min will cause the turbidimeter to overflow,
indicating the flow rate is too high. See Section 6.7 on page 84.
73
74
SECTION 5 MAINTENANCE
CAUTION PRECAUCION
Maintenance should be performed El mantenimiento debe ser realizado por personal técnico capacitado a fin de cumplir
by qualified technical personnel to con todos los códigos eléctricos y de plomería aplicables.
ensure adherence to all applicable
electrical and plumbing codes. PRECAUÇÃO
A manutenção deve ser efetuada por pessoal técnico qualificado a fim de se ajustar a
todas as normas elétricas e de canalização aplicáveis.
PRUDENCE
Toute maintenance doit être effectuée par du personnel technique qualifié, afin
d'assurer le respect de toutes les normes applicables d'électricité et de plomberie.
VORSICHT
Um zu gewährleisten, daß alle elektrischen und sanitärinstallationstechnischen
Anschlüsse den VDE-Vorschriften und gegebenenfalls den Zusatzvorschriften der
zuständigen Elektrizitäts- und Wasserwerke entsprechen, dürfen Wartungsarbeiten nur
von geschultem Fachpersonal vorgenommen werden.
Note: Although not required when It is very important to maintain the cleanliness of the interior and exterior of the
using the ICE-PIC™/870, it is good turbidimeter body, head assembly, the integral bubble trap, and the surrounding
practice to clean the turbidimeter area. Doing so will ensure accurate, low-level turbidity measurements.
body each time a calibration or
calibration verification is performed.
Clean the body before calibration verification and before calibration (especially at
levels of 1.0 NTU or lower).
5.1.1 Calibration
Check and/or perform a calibration periodically (as experience dictates)
using one of the methods described in Section 3.1. Determine when the instrument
was last calibrated using the instructions in Section 3.4 on page 61.
75
SECTION 5, continued
2. Remove the head assembly and bubble trap cover from the body. Remove the
bubble trap by lifting it vertically. Set it aside to be cleaned separately.
3. Drain the body by removing the plug from the bottom of the body.
4. Replace the drain plug and fill the body to the weir with cleaning solution.
This cleaning solution can consist of dilute chlorine solution (25 mL of
household bleach in 3.78 liters of water) or a laboratory detergent such as
Liqui-nox (1 mL detergent in 1 liter of water).
6. Remove the drain plug again and thoroughly flush the turbidimeter body with
ultra-filtered deionized water. Clean and replace the plug.
2. Using a test tube brush such as Cat. No. 690-00, clean each surface.
3. Rinse the bubble trap thoroughly with ultra-filtered deionized water and
reinstall it in the turbidimeter body.
4. Replace the bubble trap cover and head assembly on the top of the body.
6. Calibrate the instrument using one of the methods in Section 3.2 on page 49.
If the above cleaning procedures have been performed and the turbidimeter
readings are still noisy, the bottom plate and gasket may need to be removed and
cleaned. Carefully perform the following procedure to ensure the integrity of the
turbidimeter body is maintained.
2. Remove the head assembly, bubble trap cover, and bubble trap (by lifting it
vertically) from the body.
3. Drain the body by removing the plug from the bottom of the body.
76
SECTION 5, continued
5. With the body turned upside-down, remove the two Phillips-head screws
holding the bottom plate.
6. Lift the bottom plate off the body; set the gasket aside for use in reassembly.
7. Use a soft brush and a dilute cleaning solution (as prepared above) to clean the
bottom plate and inside surfaces of the turbidimeter body. Rinse the entire
body and bottom plate with ultra-filtered deionized water.
12. Replace the bubble trap, bubble trap cover, and head assembly on the top of
the body.
77
78
SECTION 6 TROUBLESHOOTING
6.1 Introduction
The 1720D/L Turbidimeter incorporates a number of self-diagnostic functions to
monitor key performance parameters. Continuous diagnostic functions are used to
detect system failures and to actuate sensor warnings. The Alarm Log stores
pertinent information about alarms and sensor warnings, and in many cases,
indicates the nature of the failure.
If the power LED is off, there is no power being supplied to the 1720D/L or the
wiring is not properly connected. Refer to the PS1201 Power Supply Manual
(Cat. No. 52010-18) for detailed wiring information.
The configure LED also lights when the CONFIGURE button is pressed and held.
6.2 Underrange/Overrange
If the sample being measured is outside the instrument range (0-100 NTU),
an underrange or overrange condition may occur.
79
SECTION 6, continued
If the calibration is accepted at this point, the error entry in the CAL HISTORY
screen will display one of the following errors. These errors can only be cleared
by recalibrating the instrument.
The only error that will occur in the StablCal/Comparative Cal procedure is
Bad Gain.
80
SECTION 6, continued
6.4 Troubleshooting
The following table presents sensor warnings displayed in the Alarm Log, their
possible causes, and corrective actions. See Section 2.8 on page 33 for information
on accessing the Alarm Log.
81
SECTION 6, continued
The following table presents additional malfunctions which may not be recorded
in the Alarm Log.
Deionized water turbidity is Access Calibration History for turbidity value of ultra-filtered water.
High Readings
greater than 0.5 NTU Verify the flow is between 250 - 750 mL/min.
Recalibrate the instrument.
Readings shown as Reconnect network wires.
Network wires disconnected or
horizontal dashes on the
shorted Check for pinched or shorted network wires.
display
Sensor Menu not Network wires disconnected or Reconnect network wires.
responding shorted Check for pinched or shorted network wires.
82
SECTION 6, continued
3. Select the sensor to be cold started and press the ENTER key.
4. Select SENSOR DIAGNOSTICS from the Sensor Menu and press the
ENTER key.
5. Move the pointer to SENSOR COLD START and press the ENTER key.
83
SECTION 6, continued
Proceed with caution! 1720D/L connections to the AquaTrend, SOM, and SIO will be
removed by this procedure.
The manual unconfigure procedure does not change the following settings:
• Calibrations and calibration standard values
• Settings for Measurement Options
• Alarm levels, recorder min/max
• Sensor name
1. Remove power from the 1720D/L; ensure the green Power LED is off.
2. Press and hold the configure button on the 1720D/L head (see Figure 3).
3. While holding the configure button down, restore power to the 1720D/L.
The configure LED will begin to flash, indicating the 1720D/L is ready to be
added to the AquaTrend network. See Multiple Sensor System on page 25 for
information on adding the sensor to the network. Before adding the 1720D/L to
the AquaTrend, it must be removed from the AquaTrend database. See Removing
a Device on page 44.
Reduce the flow rate into the turbidimeter inlet line with a restriction device or
remove the clog in the outlet.
84
APPENDIX A RS232 COMMANDS FOR THE 1720D/L
The following RS232 remote command set is available when an SIO is attached to
the AquaTrend® network. Window’s Terminal or a SCADA program may be used
to send/receive these commands. Refer to the SIO Manual (Cat. No. 52074-18) for
more information on setting the baud rate, stop bits, etc.
Each command must be prefixed with the sensor number (1-8) or the SIO returns
BAD CHANNEL. Sensor replies are prefixed with [x] where ‘x’ is the sensor
number (1-8).
1RCH?
[1]1720DL V2.3 #1_1720DL
[1]DATE: 01/01/00 00:00 ***
[1] STABLCAL/COMPARATIVE CAL
[1] CAL STD: 20.00
[1] GAIN: 1.000
[1] ERROR: NONE
1RIS?
[1]1720DL V2.3 #1_1720DL
[1]ALARM 1
[1] TYPE: High
[1] HYSTERESIS: 5%
[1] SET POINT: 100.0
[1]ALARM 2
[1] TYPE: High
[1] HYSTERESIS: 5%
[1] SET POINT: 100.0
[1]RECORDER 1
[1] MIN: 0.000
[1] MAX: 100.0
[1]RECORDER 2
[1] MIN: 0.000
[1] MAX: 100.0
[1]BUBBLE REJECT: On
[1]SIGNAL AVG: 30 Sec
[1]PRINT INTERVAL: Off
[1]TIME STAMP: On
[1]UNITS: NTU
85
APPENDIX A, continued
RS232 Instrument ID
This command recalls the instrument ID string.
86
APPENDIX B 1720D/L SCADA SYSTEM CONNECTIONS
As an example, consider the Hach Aqua View+™ data acquisition system, which
accepts 4-20mA inputs via IM420 modules. These devices convert two 4-20mA
channels to RS485 signals, which are in turn transmitted to Aqua View+.
After an IM420 module is installed in Aqua View+, the span (min/max) value
must be set to match the span of the 1720D/L. Units of NTU may also be entered
in the IM420 configure screen. For more information on Aqua View+ and the
IM420 input module, see the Aqua View+ Manual.
87
APPENDIX B, continued
Once connected to the SCADA computer, the data may be transmitted in one of
two ways.
1. To have the 1720D/L initiate the data transmission, set the print interval of
each 1720D/L to the desired print interval (1, 15, 30, or 60 minutes).
See Section 2.10.3 on page 41 for more information.Once set, the output
format is:
[1]0.035
Where 1 corresponds to the sensor number (1-8) and 0.035 is the sensor
reading. The brackets [ ] may be used as steering characters to ensure the
command is correct.
2. To have the SCADA system initiate the data transmission, the SCADA system
must send the following command:
1RMR?
Where 1 is the sensor number and RMR? is the RS232 Measurement Reading
command. The response is the same as in part A above. Note that each sensor
must be instructed to return its current value (up to eight sensors).
Many SCADA packages provide LonWorks drivers for direct connection to each
1720D/L. Refer to the respective SCADA package’s user manual for specific
information on interfacing to LonWorks nodes.
88
APPENDIX C SIO EXAMPLES
When multiple 1720D/L instruments are connected on the network, their outputs
are combined into a single data stream on the SIO. Each line of the data stream
begins with [X], where “X” is the sensor number sending the data. Each line ends
with a carriage return.
• Time Stamp = ON
Note that each sensor generates the sensor reading (as opposed to being requested
by the SIO), hence the data is not necessarily output in sensor order (i.e., sensor 2
sends before sensor 1 in this example).
89
APPENDIX C, continued
Next consider the same data, but with Time Stamp set to OFF:
[2]0.080
[1]0.244
[2]0.080
[1]0.244
[2]0.091
[1]0.244
[2]0.091
[1]0.444
[2]0.085
[1]0.469
[2]0.085
[1]0.497
[2]0.122
[1]0.572
[2]0.222
[1]0.782
[2]0.081
[1]0.521
[2]0.080
[1]0.245
[2]0.080
[1]0.244
Using Windows Terminal to record the data, a Microsoft Excel graph (such as the
one shown in Figure 9) can be created.
ÿ ÿþýüûúù
üþ
ý ÿþýüûúù
óöïõøýø ò ø
ü
ü
óöïõøÿø ò ø óöïõøýø ø
ü ü þ
NTU
óöïõøÿø ø
ü
ü
üý
üÿ
ü
üÿ üÿ üÿ üÿ ü ÿþ üÿ üÿ ü ýü ü ýÿ ü ýý üý
ÿúÿú þ ÿúÿú þ ÿúÿú þ ÿúÿú þ ÿúÿú þ ÿúÿú þ ÿúÿú þ ÿúÿú þ ÿúÿú þ ÿúÿú þ ÿúÿú þ
TIME
õò
90
GLOSSARY
91
GLOSSARY, continued
92
GENERAL INFORMATION
93
94
REPLACEMENT PARTS AND REAGENTS
Description Unit Cat. No.
1720D/L Sensor Head Assembly ................................................................................................ each................. 54303-00
Formazin Primary Standard ................................................................................................... 500 mL................... 2461-49
Filter, 0.45 µm, to produce ultra-filtered water for cleaning and calibration .............................. each................. 26705-00
Filter, 0.2 µm, to produce ultra-filtered water for calibration standard preparation.................... each................. 23238-10
PS1201 Power Supply ................................................................................................................. each................. 52010-00
Lid, Bubble Trap, 1720D/L ......................................................................................................... each................ 52012-00
OPTIONAL ACCESSORIES
Calibration Cylinder.................................................................................................................... each................. 44153-00
Cap, Connector Receptacle ......................................................................................................... each................. 52100-00
Card, IR Detection, 700 nm to 1300nm band width, 0.750 × .75 inch active area ..................... each................. 54335-00
Communication Cable, 22 gauge, 2-conductor, shielded....................................................... per foot................. 52157-00
Communication Cable, 22 gauge, 2-conductor, shielded......................................................... 100 ft. .................52157-10
Communication Cable, 22 gauge, 2-conductor, shielded......................................................... 250 ft..................52157-25
Communication Cable, 22 gauge, 2-conductor, shielded......................................................... 500 ft..................52157-50
Communication Cable, 22 gauge, 2-conductor, shielded....................................................... 1000 ft.................. 52157-51
Communication and Power Cable, 20 gauge, 4-conductor, shielded..................................... per foot................. 52158-00
Communication and Power Cable, 20 gauge, 4-conductor, shielded....................................... 100 ft.................. 52158-10
Communication and Power Cable, 20 gauge, 4-conductor, shielded....................................... 250 ft.................. 52158-25
Communication and Power Cable, 20 gauge, 4-conductor, shielded....................................... 500 ft.................. 52158-50
Communication and Power Cable, 20 gauge, 4-conductor, shielded..................................... 1000 ft.................. 52158-51
Deionized (demineralized) water .................................................................................................. 1 L..................... 272-56
Digital Display Module (DDM).................................................................................................. each................. 52400-00
Drain plug for the 1720D/L body................................................................................................ each................. 44116-00
Formazin Calibration Kit includes:
Calibration Cylinder, TenSette® Pipet, 4000 NTU Formazin Primary Standard (500 mL) . each................. 44156-00
Floor Stand for 1720D/L/PS1201 ............................................................................................... each................. 52160-00
Flow meter, 500–700 mL/min..................................................................................................... each................. 40282-00
ICE-PIC™/870 Module for Calibration and Calibration Verification
20.0 NTU .............................................................................................................................. each................. 52260-00
1.0 NTU ................................................................................................................................ each................. 52270-00
0.50 NTU .............................................................................................................................. each................. 52280-00
Mounting Kit, pole/wall for Portable AquaTrend®..................................................................... each................. 51413-00
Portable AquaTrend® Interface................................................................................................... each................. 51400-00
Power Cord Kit, 10A-125V, 1.8 m (6 feet), UL/CSA listed ....................................................... each................. 46306-00
Power Cord Kit, 10A-230V, 1.8 m (6 feet), European-style plug, VDE approved ..................... each................. 46308-00
Repeater Kit, AquaTrend® .......................................................................................................... each.................52200-00
95
REPLACEMENT PARTS AND REAGENTS, continued
96
HOW TO ORDER
97
REPAIR SERVICE
98
WARRANTY
Hach warrants most products against defective materials or workmanship for at least one
year from the date of shipment; longer warranties may apply to some items.
99
HACH COMPANY
WORLD HEADQUARTERS
P.O. Box 389
Loveland, Colorado 80539-0389
Telephone: (970) 669-3050
FAX: (970) 669-2932
All devices must be electrically connected to the network before setting up communications between devices on the networ
NETWORK ADD DEVICE If more than one type of device is available, select Sensor as the Press ENTER to add the selected sensor, Accept t
MENU device type. or press the EXIT key to cancel. name or
If more than one sensor is available, press the configure button name of
on the device you want to add.
ght and left arrow keys to move the cursor right and left along the row of characters; press the up and down arrow keys to change the character. Press ENTE
ress EXIT to cancel.
NETWORK REMOVE If more than one type of device is Select the specific device to Press ENTER to remove the Remove all phys
MENU DEVICE available, select sensor as the device remove. device, EXIT to cancel. connections.
type.
NETWORK REPLACE If more than one type of device is Select the existing Select the Press ENTER to Redo de
MENU DEVICE available, select sensor as the device device to replace. unconfigured device replace the device, informati
type. that will replace the EXIT to cancel
existing device.*
one unconfigured device exists on the network, press the configure button on the selected device.
Sensors have various numbers and types of alarm or recorder signals available. Define the conditions under which an alarm
be triggered through the setup menu of the sensor containing that alarm or recorder.
ALARM TO ATTACH SELECT SENSOR, then SELECT ALARM 1 or 2, Attach an alarm to a relay
RELAY then SELECT SOM, then SELECT RELAY 1 or 2
DETACH SELECT SENSOR, then SELECT ALARM 1 or 2 Detach an alarm from a relay
REC TO ATTACH SELECT SENSOR, then SELECT RECORDER 1 or 2, Attach a recorder to an analog output
ANALOG then SELECT SOM, then SELECT ANALOG OUTPUT 1 or 2
OUTPUT DETACH SELECT SENSOR, then SELECT RECORDER Disconnect recorder/analog output
Select an alarm group: List of last 14 alarms will be displayed. Active alarms cause the corresponding a
1 through 8 or N for network or AquaTrend alarms Use the up or down arrow keys to highlight the alarm indicator to be illuminated at the bottom
of interest and press ENTER to review details of the display.
alarm or press EXIT to cancel. Active as well as no-longer-active alarms
the log.
ement Options
MEASUREMENT BUBBLE REJECT Select this option to turn bubble reject ON or OFF. When bubble reject is ON, momentary spikes
he user-prepared calibration method as follows:
Note: The Hold Outputs feature is automatically enabled during the zero electronics and calibration procedure, holding atta
outputs at their precalibration settings.
1. Zero the electronics as described in the 1720D/L Turbidimeter Instruction Manual.
2. Press the MENU key to access the MAIN MENU. Select SENSOR MENU and press the ENTER key.
3. Select the sensor to be calibrated using the arrow keys and press ENTER.
4. Select CALIBRATION from the menu choices and press ENTER.
5. Move the pointer to USER PREPARED CAL and press ENTER to begin the calibration procedure. During calibration,
the instrument will automatically switch to a 6-second signal average. After calibration, the signal average will revert to
set prior to calibration.
6. The display will prompt FILL CAL CYLINDER WITH 1 LITER OF DI WATER. Accurately measure 1.0 liter of low turb
water and pour it into a clean, dry calibration cylinder. Pour carefully to minimize bubbles.
Note: Refer to the 1720D/L Manual for information on preparing dilution water.
7. Remove the head assembly from the turbidimeter body. Clean the photocell window per the instructions in the 1720D/L
Turbidimeter Instruction Manual. Check the lamp and lens for scaling or filming, and clean them if necessary.
8. Place the head assembly on the calibration cylinder. Gently tap the cylinder to free any bubbles adhering to the photoce
9. Press ENTER to begin reading the dilution water value or, to exit without updating the calibration, move the pointer to A
CALIBRATION and press the ENTER key.
10. After waiting a minimum of one minute, press ENTER to accept the reading of the dilution water. The value displayed is b
previous calibration, and will be updated after the calibration is accepted. The number is displayed for stability purposes
11. The display will prompt DISPENSE 5 mL OF 4000 FORMAZIN INTO CAL CYLINDER AND SWIRL. Attach a clean, dry
the end of a TenSette® Pipet. Set the dispenser knob to the 5-mL position. Gently invert the 4000-NTU formazin bottle se
to mix the solution. Depress the pipet plunger and immerse the end of the tip one-half inch into the 4000-NTU formazin
release the plunger to draw 5 mL of 4000-NTU formazin into the tip.
12. Set the pipet knob to the 6-mL position to ensure all the 4000-NTU formazin is dispensed. Lift the turbidimeter head as
of the calibration cylinder and smoothly depress the pipet plunger to dispense the 4000-NTU formazin into the calibratio
13. Gently stir the formazin solution in the cylinder and replace the head assembly.
14. Press ENTER to begin reading the 20-NTU standard, or to exit without updating the calibration, move the pointer to ABO
CALIBRATION and press the ENTER key.
15. After waiting a minimum of one minute, press ENTER to accept the value of the 20-NTU standard. The value displayed
the previous calibration, and will be updated after the calibration is accepted. The number is displayed only for stability
16. If the calibration was performed correctly, the display will show SUCCESSFUL CAL. Press ENTER to accept the calibra
three alpha characters (such as initials) to complete the required information and store the calibration in memory.
17. Replace the instrument head assembly on the turbidimeter body and press the DISPLAY key. The instrument is now cal
18. Discard the formazin standard according to the directions on the standard bottle. Rinse the calibration cylinder with deio
and store upside down to minimize dust buildup.
Note: The Hold Outputs feature is automatically set to 5 minutes after a calibration is accepted or aborted, to allow time for
turbidimeter head to be placed on the body. To cancel Hold Outputs immediately, refer to the 1720D/L Turbidimeter
Instruction Manual.