T170R Applanation Tonometer Structure and Specification: User Manual
T170R Applanation Tonometer Structure and Specification: User Manual
T170R Applanation Tonometer Structure and Specification: User Manual
Range of Measurement: 0 mmHg ~ 80mmHg (0 kPa ~ 10.64kPa) Test on position “ 4 ” with 1 Counterweight on
Moving Range of the Rings: 1.53×2=3.06mm。 1, Put one Counterweight in the Seat of the Counterweight.
Prism Diameter: 7mm。 2, Set the Dial Drum at the number “ 4 ”.
Moving Range of Prism: 0 ~ 3mm。 3, Turn the Knob till the Prism Arm stays freely between the two extreme positions.
Dimension: 190mm×80mm×80mm。 4, If the scale line "4" is out of contact with the Basic Line, the tonometer is out of accuracy, shown
Net Weight: 480g。 as below:
Working Temperature: 0℃~40℃ / 32℉~104℉。 Counterweight
Working Relative Humidity: 10%~85%。
Storage Temperature: 0℃~55℃ / 32℉~131℉。
Storage Relative Humidity: 10%~95%。
version: 1.0 Within Accuracy Within Accuracy Out of Accuracy
20140210 Installation Test on position “ 6 ” with 2 Counterweights on
serial number:2013-08-007 1, Keep 2 Counterweights in the Seat of the Counterweight.
Mounting the Tonometer onto the Slit lamp 2, Set the Dial Drum at the number “ 6 ”.
Shown as below: 3, Turn the Knob till the Prism Arm stays freely between the two extreme positions.
Tonometer
Acknowledgment 4, If the scale line "6" is out of contact with the Basic Line, the tonometer is out of accuracy, shown
as below:
Many thanks you for making the decision of purchasing our products known as the best Counterweight
price-performance and almost trouble-free products in the global market of Ophthalmic Instrument Mounting Screw
and Vision Machinery!
For having the smooth and pleasant experience in using our products, it is recommended to read the
Within Accuracy Within Accuracy Out of Accuracy
User Manual first after unpacking.
Mounting Adaptor Test on position “ 0” with 3 Counterweights on
Packing list Round Body type Slip Lamp
1, Keep 3 Counterweights in the Seat of the Counterweight.
2, Set the Dial Drum at the number “ 0 ”.
1, Applanation Tonometer (including Mounting Rod ) 3, Turn the Knob till the Prism Arm stays freely between the two extreme positions.
2, User Manual The Mounting Adapter provided in the packing is for the Slit Lamp with the round body only.
4, If the scale line "0" is out of contact with the Basic Line, the tonometer is out of accuracy, shown
3, Accessory Box (including Prism Arm x1, Prism x 3, Weight Seat x 1, Counterweight x 3 ) ( shown as above).
as below:
4, Mounting Adaptor (including Mounting Screw x 1 ) When coming across any problem with the Mounting Adapter, please contact your service or
Counterweight
supplier.
The Tonometer can also stay mounted on the Slit Lamp when not in use.